| | Class 2 Device Recall LUSTER HIP |  |
| Date Initiated by Firm | May 16, 2012 |
| Date Posted | June 29, 2012 |
| Recall Status1 |
Terminated 3 on March 05, 2013 |
| Recall Number | Z-1917-2012 |
| Recall Event ID |
61749 |
| 510(K)Number | K983136 |
| Product Classification |
Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
|
| Product | LUSTER HIP SZ 3
Packaging: The product was packaged within inner and outer blister trays, sealed with coated Tyvek lidding. The stems were packaged within vinyl protectors, with foam taper protectors.
Intended Use: The LUSTER Cemented Polished Femoral Hip Stems product are forged CoCr polished cemented stems which are used as part of a total hip arthroplasty. These stems were indicated for use with all DePuy bearings and acetabular cups. |
| Code Information |
Product code:152180003 and Lot numbers: A46D71000, A52C11000, A88D11000, AY2CE1000, B1XGV1000, BA3BF1000, BC6BC1000, BX9B71000, Y97FM1000, and Z5NBJ1000. |
| FEI Number |
1818910
|
Recalling Firm/ Manufacturer |
DePuy Orthopaedics, Inc. 700 Orthopaedic Dr Warsaw IN 46582-3994
|
| For Additional Information Contact | 574-267-8143 |
Manufacturer Reason for Recall | DePuy Orthopaedics, Inc. is issuing a Field Safety Notice for the LUSTER Cemented Polished Femoral Hip Stems and the ULTIMA LX Cemented Polished Femoral Hip Stems that were implanted with any metal-on-metal (MoM) articulation (including a head of CoCr and an acetabular cup where the cup is of an all metal construction (monoblock or modular)).The LUSTER Cemented Polished Femoral Hip Stems and the U |
FDA Determined Cause 2 | Device Design |
| Action | DePuy sent an "URGENT RECALL INFORMATION-FIELD SAFETY NOTICE" letter dated May 16, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact the firm at 574-372-7333 for questions regarding this notice. |
| Quantity in Commerce | (85) LUSTER Cemented Polished Femoral Hip Stems in 69 lots |
| Distribution | Nationwide Distribution-including the states of CA, CO, IA, NC, OH, TX, WA, WI, and WV. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JDI
|
|
|
|