| | Class 2 Device Recall ROTICULATOR Single Use Stapler |  |
| Date Initiated by Firm | May 09, 2012 |
| Date Posted | June 21, 2012 |
| Recall Status1 |
Terminated 3 on February 12, 2014 |
| Recall Number | Z-1847-2012 |
| Recall Event ID |
61942 |
| 510(K)Number | K855047 |
| Product Classification |
Staple, implantable - Product Code GDW
|
| Product | Covidien ROTICULATOR 30-4.8 Single Use Stapler
REF# 017617
The RETICULA TOR TM Single Use Staplers have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomosis |
| Code Information |
LOT #'s BEGINNING WITH THE FOLLOWING LETTER/NUMBER COMBINATIONS P7E P8B P8L P9H POE PIB PIL P7F P8C P8M P9J POF PIC PlM P7G P8D P9A P9K POG PlD P2A P7H P8E P9B P9L POH PIE P2B0153X P7J P8F P9C P9M POJ PlF P7K P8G P9D POA POK PlG P7L P8H P9E POB POL PlH P7M P8J P9F POC POM PlJ P8A P8K P9G POD PIA PlK Also Best Practice Kits Ref#'s: OOZ0800, OOZ0976, OOZ1401, OOZ1407, OOB0117, OOBOI22 NOTE: THE SPECIFIC LOT NUMBERS LISTED BELOW AND LOT NUMBERS ENDING IN ""RMX" and "RRMX" ARE NOT AFFECTED BY THIS RECALL AND ARE ACCEPT ABLE FOR USE. P2A0194XR, P2B0488XR, P2B0153XR |
| FEI Number |
1219930
|
Recalling Firm/ Manufacturer |
Covidien LP
60 Middletown Ave
North Haven
CT
06473-3908
|
| For Additional Information Contact | 203-492-8165 |
Manufacturer Reason for Recall | Potential for the sterility barrier to be compromised. |
FDA Determined Cause 2 | Package design/selection |
| Action | The firm, Covidien, sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 16, 2012 via federal express to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately identify and quarantine all affected inventory (note: units from affected lots may have been incorporated into Covidien BEST PRACTICE KITS; evaluate kits for affected lots). All affected ROTICULATOR units must be returned. The customers were also instructed to complete and return the attached ROTICULATOR Single Use Stapler Recalled Product Return form via fax to: (800) 895-6140 or email to: SDFeedback@Covidien.com, even if you do not have any of the affected units, and include a copy of completed form with products being returned. Customers who received product directly from Covidien should contact Covidien Customer Service at SDFeedback@Covidien.com or (800) 722-8772, option 1, prior to returning affected products to Covidien, Field Returns Department, at 195 McDermott Road, North Haven, CT 06473; if received from distributor, return form and contact distributor directly.
If you have any questions or concerns, please do not hesitate to contact Covidien Representative or Covidien Quality Assurance at (203) 492-5232. |
| Quantity in Commerce | 28248 units |
| Distribution | Worldwide distribution: USA (Nationwide) and countries including; Africa, Argentina, Australia, Belgium, Brazil, Canada, China Costa Rica, CROATIA, Czech Republic,Dominion Republic, ESPA¿A France, Germany, Greece Guadeloupe, Guatemala, Honduras Hong Kong, Hungary, India, Indonesia, Ireland, Italy, Japan, Korea ,Malaysia, Mexico, Peru, Portugal, Puerto Rico, Spain, Sri Lanka, Switzerland, Taiwan , Thailand, and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = GDW
|
|
|
|