| | Class 2 Device Recall Philips HeartStart MRx monitor/defibrillators |  |
| Date Initiated by Firm | June 20, 2012 |
| Date Posted | September 10, 2012 |
| Recall Status1 |
Terminated 3 on February 15, 2017 |
| Recall Number | Z-2347-2012 |
| Recall Event ID |
62328 |
| 510(K)Number | K051134 |
| Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
| Product | Philips HeartStart MRx monitor/defibrillators
Models: M3535A and M3536A, M3536J, M3536M, M3536MC, M3536M1, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6
The Heartstart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician |
| Code Information |
Philips HeartStart MRx monitor/defibrillators manufactured prior to June 2012 and used in external transport and Emergency Medical Service (EMS) environments, distributed worldwide |
| FEI Number |
1218950
|
Recalling Firm/ Manufacturer |
Philips Healthcare Inc. 3000 Minuteman Road Andover MA 01810
|
| For Additional Information Contact | 978-687-1501 |
Manufacturer Reason for Recall | Therapy Cable Connection wear may lead to a malfunction in detection of defibrillation pads/paddles therapy Cables
|
FDA Determined Cause 2 | Equipment maintenance |
| Action | Philips notified customers on June 20, 2012, by an Urgent Medical Device Correction notification/Field Safety Notice and a revised Instructions for Use addendum for each HeartStart MRx unit.
The Urgent Medical Device Correction notification/Field Safety Notice informs customers of the issue, identifies details of units affected and and identifies the action. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice:
Carefully read the attached HeartStart MRx Instructions for Use Addendum. Immediately implement ongoing therapy connection inspection on all of your MRx devices to detect wear from higher than expected levels of stress. The HeartStart MRx Instructions for Use Addendum describes how a user can identify wear. If wear is detected, remove affected devices immediately from use and contact Philips to arrange for service.
WARNING:
The service life of your therapy cables/external paddles is up to three years. To maintain reliable performance and reduce the possibility of failure during patient use, replace them every three years from the time they were initially placed into service or if they fail the inspection criteria in the attached Instructions for Use Addendum
The addendum provides the customer with complete user information regarding ongoing inspections of port.
In addition, service support will be provided for Therapy Port and Cables requirement replacements.
Further information or support concerning this issue,contact your local Philips representative at 1-800-722-9377.
For questions regarding this recall calll 978-687-1501. |
| Quantity in Commerce | 58,701 units |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of :
AFGHANISTAN
ALBANIA
ARGENTINA
ARUBA
AUSTRALIA
AUSTRIA
AZERBAIJAN
BAHRAIN
BANGLADESH
BELGIUM
BOLIVIA
BOSNIA AND HERZEGOWINA
BRAZIL
BRUNEI DARUSSALAM
BULGARIA
CAMBODIA
CANADA
CHILE
CHINA
COLOMBIA
COSTA RICA
CROATIA (local name: Hrvatska)
CYPRUS
CZECH REPUBLIC
DENMARK
ECUADOR
EGYPT
EL SALVADOR
ESTONIA
ETHIOPIA
FINLAND
FRANCE
GERMANY
GREECE
GUADELOUPE
HONG KONG
ICELAND
INDIA
INDONESIA
IRAQ
IRELAND
ISRAEL
ITALY
JAPAN
JORDAN
KENYA
KOREA, REPUBLIC OF
KUWAIT
LAOS
LATVIA
LEBANON
LIBYAN ARAB JAMAHIRIYA
LITHUANIA
LUXEMBOURG
MACAU
MACEDONIA
MALAYSIA
MALTA
MEXICO
MOROCCO
NAMIBIA
NETHERLANDS
NETHERLANDS ANTILLES
NEW CALEDONIA
NEW ZEALAND
NICARAGUA
NIGERIA
NORWAY
OMAN
PAKISTAN
PANAMA
PAPUA NEW GUINEA
PARAGUAY
PERU
PHILIPPINES
POLAND
PORTUGAL
PUERTO RICO
QATAR
REUNION
ROMANIA
RUSSIAN FEDERATION
SAUDI ARABIA
SINGAPORE
SLOVENIA
SOUTH AFRICA
SPAIN
SRI LANKA
SWEDEN
SWITZERLAND
TAIWAN, PROVINCE OF CHINA
TANZANIA, UNITED REPUBLIC OF
THAILAND
TRINIDAD AND TOBAGO
TUNISIA
TURKEY
UGANDA
UKRAINE
UNITED ARAB EMIRATES
UNITED KINGDOM
UZBEKISTAN
VENEZUELA
VIET NAM
YEMEN |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MKJ
|
|
|
|