| | Class 2 Device Recall Philips HeartStart MRx |  |
| Date Initiated by Firm | June 20, 2012 |
| Date Posted | June 27, 2012 |
| Recall Status1 |
Terminated 3 on March 16, 2017 |
| Recall Number | Z-1896-2012 |
| Recall Event ID |
62333 |
| 510(K)Number | K051134 |
| Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
| Product | Philips HeartStart MRx monitor/defibrillators
Models: M3535A, M3536A, M3536M4, M3536M5, and M3536M6
Product Usage:
The Heartstart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician |
| Code Information |
Manufactured between the dates of July 12, 2011 through September 2, 2011 and January 19, 2012 through May 22, 2012. Serial numbers within the ranges of: US00550047 to US00551654 and US00556357 to US00559493 respectively. There are also some additional MRx units that were repaired and may contain compromised components. These MRx unit serial numbers are identified below: US00210180, US00210183, US00210738, US00211000, US00213281 US00214706, US00316827, US00318163, US00319241, US00328491 US00329811, US00333101, US00536832, US00538140, US00543161 US00543825, US00544081, US00544214, US00546107, US00548046 |
Recalling Firm/ Manufacturer |
Philips Healthcare Inc. 3000 Minuteman Road Andover MA 01810
|
| For Additional Information Contact | 978-687-1501 |
Manufacturer Reason for Recall | HeartStart MRx Unexpected Pads/Paddles ECG Failure
|
FDA Determined Cause 2 | Process control |
| Action | Philips issued an Urgent Voluntary Medical Device Correction/Field Safety Notice letter on June 20, 2012. The letter informs customers of the issue, identifies details (including models and software versions applicable) of units affected, gives instructions on actions to be taken until the correction is available, and identifies what action Philips plans to eliminate or remedy the issue. The letter states that a Philips Healthcare representative will contact customers to arrange for the replacement of the Power PCA in all the affected devices.
Philips will complete the installation free of charge. For questions or support concerning this issue contact your local Philips representative at 1-800-722-9377. |
| Quantity in Commerce | 2914 units |
| Distribution | Worldwide Distribution - US (nationwide) and the countries of: AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, COLOMBIA, EL SALVADOR, ETHIOPIA, FRANCE, GERMANY, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOREA, REPUBLIC OF, LEBANON, LIBYAN ARAB JAMAHIRIYA, MEXICO, MOROCCO, NETHERLANDS, NETHERLANDS ANTILLES, NICARAGUA, NORWAY, PERU, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, and VIET NAM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MKJ
|
|
|
|