Class 2 Device Recall CODMAN CERTAS Programmable Valve

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CODMAN CERTAS Programmable Valvesee related information
Date Initiated by FirmMay 16, 2013
Date PostedJanuary 29, 2014
Recall Status1 Terminated 3 on June 02, 2015
Recall NumberZ-0864-2014
Recall Event ID 65179
510(K)NumberK112156 K113526 
Product Classification Shunt, central nervous system and components - Product Code JXG
ProductCodman Certas Programmable Valve In Line Valve with SiphonGuard Device, Unitized Bactiseal Catheter and Accessories Product Code: 82-8807 Product Usage: The CODMAN CERTAS Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular perssure drainage of CSF. The CODMAN CERTAS Therapy Management System (TMS) allows the non-invasive reading or adjustment of the valve setting.
Code Information Lot Codes:  CLPCNB CMBCPT CMCC2W CMDB6L CMDDJV CMKB24 CMLCKN CMMC3H CMNBZH CNCB6N CNCCLJ CNCCZ3 CNDB8C CNDBRJ CNGBK4 CNHCZR CNJB1G CNJC4T CNJCMR CNKBZD CNLBVM CNLBZC CNMBFL CNNDH1 CNMBFM CNMDNK CNPBRL CNPCLV CPBBGT CNCCLH CNCCZ2 CNDB69 CNDCV0 CPBCR3 CPCBHC CPCBHD CPDCC3 CMMCGV CNBB11 CPBCR4 CPCCMN CPDBYL 
FEI Number 3002947440
Recalling Firm/
Manufacturer
Codman & Shurtleff, Inc.
325 Paramount Drive
Raynham MA 02767
For Additional Information ContactSame
508-880-8100
Manufacturer Reason
for Recall
Codman Certas Programmable Valves used for hydrocephalus may not operate properly
FDA Determined
Cause 2
Device Design
ActionCodman issued an Urgent Notice dated May 16, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Mitigation instructions are provided and if not followed or are not successful, either issue could lead to symptoms associated with under- or over-drainage. Units not implanted are to be returned. A U.S Acknowledgement Form and fax are requested to be completed and faxed to: 1-888-943-4193, or e-mail the form to codman8894@stericycle.com. If you have any questions on concerns in regards to this recall, please contact your Codman Representative or Stericycle at 1-888-943-2395.
Quantity in Commerce746 units
DistributionWorldwide Distribution - US Nationwide, Australia, Austria, Belgium, Brazil, Switzerland, Czech Republic, Germany, Spain, Finland, France, Great Britain, Israel, Italy, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa, and UK.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.