| | Class 2 Device Recall Masimo SET uSpO2 Cable |  |
| Date Initiated by Firm | September 30, 2014 |
| Date Posted | November 13, 2014 |
| Recall Status1 |
Terminated 3 on March 28, 2016 |
| Recall Number | Z-0213-2015 |
| Recall Event ID |
69504 |
| 510(K)Number | K121914 |
| Product Classification |
Oximeter - Product Code DQA
|
| Product | Masimo SET uSpO2 Cable
The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that provides continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate when installed in a compatible OEM host system.
Shipping/Package: All Pulse Oximetry cables are shipped non-sterile in a cardboard carton. They are intended to be used non-sterile. |
| Code Information |
PART_ID LOT_ID 3311 11CE2 3311 E11EA7 3311 11EMM 3311 E11EDJ 3311 11FAU 3311 11FAV 3311 11CE2 3311 11FAW 3311 11GBY 3311 11HGF 3311 11HE6 3311 11J80 3311 11JCJ 3311 11HNO 3311 11JCK 3311 11K37 3412 11MA7 3311 R11KFY 3311 11K37 3311 11KQM 3311 11NDZ 3311 12B95 3412 12CP1 3412 11AM7 3449 E12HR3 3450 E12HR4 3412 12CP1 3635 12KP4 3636 12KP5 3635 12KQ7 3636 12KQ8 3450 E12KJY 3449 E12KS3 3449 E12KT4 3450 E12KS4 3635 12MAG 3450 E12KT5 3450 E12MA2 3578 E12MA6 3577 E12ME6 3449 E12KS3 3450 E12KS4 3578 R12N2U 3578 E12NB2 3577 E12NA9 3578 E12NB2 3450 E12HR4 3577 R12N2W 3449 13AYG 3450 R13BRX 3450 R13BE2 3577 R12N2W 3577 E12NA9 3315 E12JS1 3635 12N17 3450 13C57 3635 12N1S 3449 E12HR3 3578 13C59 3450 13DPZ 3578 E12NB2 3578 13C59 3636 13CNZ 3635 12N1S 3449 13EPF 3635 13C4K 3449 13C56 3449 13AYG 3449 13EPF 3636 13FPN PART_ID LOT_ID SERIAL_NUMBER 3449 2062190-001 3315 E12JS1 3449 V25060 3449 V25061 3449 V25062 3449 V25063 3449 V25064 3449 V25065 3449 V25066 3449 V25067 3449 V25068 3449 V25069 3449 V25070 3449 V25071 3449 V25072 3449 V25073 3449 V25074 3449 V25075 3449 V25076 3449 V25077 3449 V25078 3449 V25079 3449 V25080 3449 V25081 3449 V25082 3449 V25083 3449 V25084 3449 V25085 3449 V25086 3449 V25087 3449 V25088 3449 V25089 3449 V25136 3449 V25152 3449 V25154 3449 V25155 3449 V25156 3449 V25158 3449 V25164 3449 V25167 3449 V25170 3449 V25172 3449 V25174 3449 V25175 3449 V25179 3449 V25181 3449 V25255 3449 V25257 3449 V25258 3449 V25259 3449 V25262 3449 V25267 3577 V14404 3577 V14422 3577 V14464 3577 V14425 3577 V14434 3577 V14457 3577 V14465 3577 V14467 3577 V16984 3577 V17047 3577 V17088 3577 V17089 3577 V17090 3577 V17091 3577 V17092 3577 V17093 3577 V17094 3577 V17095 3577 V17097 3577 V17098 3577 V17099 3577 V17102 3577 V17103 3577 V17104 3577 V17105 3577 V17106 3577 V17107 3577 V17111 3577 V19603 3577 V19612 3577 V19615 3577 V19616 3577 V19617 3577 V19618 3577 V17116 3577 V14448 3577 V14466 3577 V14402 3577 V14405 3577 V14408 3577 V14409 3577 V14410 3577 V14411 3577 V14413 3577 V14414 3577 V14415 3577 V14419 3577 V14423 3577 V14427 3577 V14428 3577 V14430 3577 V14433 3577 V14437 3577 V14440 3577 V14444 3577 V14447 3577 V14449 3577 V14454 3577 V14455 3577 V14456 3577 V14462 3577 V18343 3577 V19440 3577 V19997 3577 V19425 3577 V19443 3578 V18683 3578 V18998 3636 V14341 3636 V14358 3636 V14363 3636 V14364 3636 V14368 3636 V14543 3636 V25406 3636 V25420 3636 V25429 3636 V25447 3636 V25451 3636 V25456 3636 V25458 3636 V25485 3636 V25486 3636 V25504 3636 V25505 3636 V25511 3636 V25512 3636 V25515 3636 V25521 3636 V25522 3636 V25524 3636 V25527 3636 V25528 3636 V25529 3636 V25538 3636 V25542 3636 V25545 3636 V25546 3636 V25550 3636 V25564 3636 V25573 3636 V25576 3636 V25577 3636 V25578 3577 V19435 3577 V17082 3577 V17113 3577 V19610 3578 V18637 3578 V18658 3578 V18684 3578 V18739 3578 V18956 3578 V19467 3577 V17100 3577 V17121 3577 V17123 3577 V19601 3577 V19445 3577 V19613 3577 V19614 3577 V17101 3577 V19447 3577 V19448 3577 V19606 3577 V19608 3577 V17087 3577 V17109 3577 V19605 3577 V19607 3577 V19609 3577 V13301 3315 E12JS1 3577 V19299 3577 V19300 3577 V19301 3577 V19302 3577 V19303 3577 V19304 3577 V19305 3577 V19306 3577 V19307 3577 V19308 3577 V19309 3577 V19310 3577 V19311 3577 V19312 3577 V19313 3577 V19314 3577 V19316 3577 V19317 3577 V19318 3577 V19319 3577 V19320 3577 V19321 3577 V19322 3577 V19323 3577 V19324 3577 V19325 3577 V19326 3577 V19327 3577 V19328 3577 V19329 3577 V19330 3577 V19331 3577 V19332 3577 V19333 3577 V19335 3577 V19336 3577 V19337 3577 V19338 3577 V19339 3577 V19340 3577 V19346 3577 V19350 3577 V19360 3577 V19361 3577 V19368 3577 V19371 3577 V19381 3577 V19391 3577 V19392 3577 V19394 3578 V24322 3578 V24327 3578 V24579 3578 V24600 3578 V24621 3578 V24687 3578 V24688 3578 V24689 3578 V24690 3578 V24691 3578 V24694 3578 V24695 3578 V24696 3578 V24697 3578 V24698 3578 V24699 3578 V24700 3578 V24701 3578 V24702 3578 V24703 3578 V24704 3578 V24705 3578 V24706 3578 V24845 3578 V24818 3578 V24839 3577 V19444 3578 V24849 3577 V18378 3577 V19359 3577 V19373 3577 V19375 3577 V19384 3577 V19387 3577 V18379 3636 R13FZJ 3577 V19364 3578 V24313 3578 V24315 3578 V24316 3578 V24318 3578 V24321 3578 V24598 3578 V24615 3578 V24617 3578 V24796 3578 V24799 3578 V24800 3578 V24804 3578 V24807 3578 V24808 3578 V24809 3578 V24810 3578 V24811 3578 V24812 3578 V24813 3578 V24814 3578 V24815 3578 V24816 3578 V24817 3578 V24819 3578 V24820 3578 V24821 3578 V24822 3578 V24823 3578 V24824 3578 V24825 3578 V24826 3578 V24827 3578 V24828 3578 V24829 3578 V24830 3578 V24831 3578 V24832 3578 V24833 3578 V24834 3578 V24835 3578 V24836 3578 V24837 3578 V24841 3578 V24842 3578 V24844 3578 V24846 3578 V24847 3578 V24848 3578 V24850 3578 V24854 3578 V24605 3577 19388V 3577 V19388 3577 V26963 3577 V26965 3577 V26966 3577 V26972 3577 V26975 3577 V26976 3577 V26977 3577 V26978 3577 V26979 3577 V26956 3577 V26962 3577 V26967 3577 V26970 3577 V26973 3577 V26901 3577 V26916 3577 V26931 3577 V26959 3577 V26958 3577 V26964 3577 V26968 3577 V26969 3577 V26971 3577 V26974 3577 V26980 3577 V26981 3577 V26982 3577 V26983 3577 V26985 3577 V26986 3577 V26987 3577 V26988 3577 V26990 3577 V27135 3577 V27136 3577 V27139 3577 V27140 3577 V27141 3577 V26957 3578 V24320 3449 V27000 3449 V27001 3449 V27002 3449 V27003 3449 V27004 3449 V27005 3449 V27006 3449 V27007 3449 V27008 3449 V27009 3449 V27010 3449 V27011 3449 V27012 3449 V27013 3449 V27015 3449 V27016 3449 V27017 3449 V27018 3449 V27019 3449 V27020 3449 V27021 3449 V27022 3449 V27023 3449 V27024 3449 V27025 3449 V27026 3449 V27027 3449 V27028 3449 V27029 3449 V27030 3449 V27031 3449 V27032 3449 V27033 3449 V27034 3449 V27035 3449 V27036 3449 V27037 3449 V27038 3449 V27039 3449 V27040 3449 V27041 3449 V27042 3449 V27043 3449 V27044 3449 V27045 3449 V27046 3449 V27047 3449 V27048 3449 V27049 3449 V27050 3449 V27051 3449 V27052 3449 V27053 3449 V27054 3449 V27055 3449 V27056 3449 V27057 3449 V27058 3449 V27060 3449 V27061 3449 V27062 3449 V27063 3449 V27066 3449 V27068 3449 V27069 3449 V27070 3449 V27071 3449 V27072 3449 V27073 3449 V27076 3449 V27077 3449 V27078 3449 V27081 3449 V27082 3449 V27084 3449 V27085 3449 V27086 3449 V27089 3449 V27090 3449 V27093 3449 V27095 3449 V25147 3449 V25148 3449 V25160 3449 V25161 3449 V25162 3449 V25163 3449 V25165 3449 V25166 3449 V25168 3449 V25169 3449 V25171 3449 V25173 3449 V25186 3449 V25187 3449 V25188 3449 V25190 3449 V25157 3577 V13359 3577 V14967 3577 V14178 3577 V14197 3577 V18293 3577 V18331 3577 V18333 3577 V18334 3577 V18336 3577 V18338 3577 V18339 3577 V18340 3577 V18344 3577 V18345 3577 V19427 3577 V19433 3577 V19437 3577 V19438 3577 V19439 3577 V19496 3577 V19497 3577 V19546 3577 V19547 3577 V19548 3577 V19598 3577 V19988 3577 V19989 3577 V19992 3577 V19993 3577 V19994 3577 V19995 3577 V19996 3577 V13300 3577 V14169 3577 V14232 3577 V14436 3577 V14964 3577 V15030 3577 V15037 3577 V15054 3577 V15066 3577 V15068 3577 V14407 3578 V24314 3578 V24324 3578 V24459 3578 V24463 3578 V24840 3578 V24843 3578 V24851 3578 V24853 3577 V13292 3577 V14452 3577 V14954 3577 V15050 3577 V15061 3578 V259924 3577 V15025 3577 V17080 3577 V19602 3577 V19604 3577 V19611 3577 V19619 3578 V25913 3578 V25916 3578 V25918 3578 V25919 3578 V25921 3578 V25923 3578 V25928 3578 V25930 3578 V25931 3578 V24127 3578 V24153 3578 V24155 3578 V24725 3578 V24732 3578 V24735 3578 V24745 3578 V24752 3578 V24764 3578 V24772 3578 V24774 3578 V24777 3578 V24778 3578 V24786 3578 V24789 3578 V24791 3578 V24792 3578 V24795 3578 V24797 3578 V24798 3578 V24801 3578 V24802 3578 V24803 3578 V24805 3578 V24806 3577 V18288 3577 V19138 3577 V19426 3577 V19441 3577 V19600 3577 V19998 3577 V15060 3577 V26828 3577 V26830 3577 V26831 3577 V26832 3577 V26833 3577 V26834 3577 V26837 3577 V26838 3577 V26839 3577 V26840 3577 V26842 3577 V26844 3577 V26850 3577 V26851 3577 V26852 3577 V26855 3578 V24770 3578 V24771 3578 V24793 3578 V24794 3577 V15033 3577 V26853 3577 V14443 3577 V14446 3577 V14955 3577 V15021 3577 V15040 3578 V24780 3578 V24782 3578 V24784 3578 V24787 3315 E12JS1 3578 V26567 3578 V24769 3578 V24788 3578 V24781 3577 V18391 3577 V18467 3577 V18471 3577 V19342 3577 V19344 3577 V19382 3577 V26760 3577 V26761 3577 V26762 3577 V26763 3577 V26764 3577 V26765 3577 V26766 3577 V26767 3577 V26768 3577 V26769 3577 V26770 3577 V26771 3577 V26772 3577 V26773 3577 V26774 3577 V26775 3577 V26776 3577 V26777 3577 V26778 3577 V26779 3577 V26780 3577 V26781 3577 V26782 3577 V26784 3577 V26785 3577 V26786 3577 V26787 3577 V26788 3577 V26789 3577 V26790 3577 V26791 3577 V26792 3577 V26793 3577 V26794 3577 V26795 3577 V26796 3577 V26797 3577 V26798 3577 V26799 3577 V26800 3577 V26801 3577 V26802 3577 V26803 3577 V26804 3577 V26805 3577 V26806 3577 V26807 3577 V26808 3577 V26809 3577 V26810 3577 V26811 3577 V26812 3577 V26814 3577 V26815 3577 V26816 3577 V26817 3577 V26819 3577 V26820 3577 V26821 3577 V26822 3577 V26823 3577 V26824 3577 V26825 3577 V26826 3577 V26829 3577 V26841 3577 V26854 3577 V26856 3577 V26845 3577 V26846 3577 V26847 3577 V26848 3577 V26849 3577 V26860 3577 V26862 3577 V26863 3577 V26864 3577 V26865 3577 V26866 3577 V26867 3577 V26868 3577 V26869 3577 V26870 3577 V26871 3577 V26872 3577 V26873 3577 V26874 3577 V26875 3577 V26876 3577 V26877 3577 V26878 3577 V26879 3577 V26880 3577 V26881 3577 V26882 3577 V26883 3577 V26884 3577 V26885 3577 V26886 3577 V26887 3577 V26888 3577 V26889 3577 V26890 3577 V26892 3577 V26893 3577 V26894 3577 V26895 3577 V26896 3577 V26897 3577 V26898 3577 V26899 3577 V26900 3577 V26902 3577 V26903 3577 V26904 3577 V26905 3577 V26906 3577 V26907 3577 V26908 3577 V26909 3577 V26910 3577 V26911 3577 V26912 3577 V26913 3577 V26914 3577 V26915 3577 V26917 3577 V26918 3577 V26919 3577 V26920 3577 V26921 3577 V26922 3577 V26923 3577 V26924 3577 V26925 3577 V26926 3577 V26927 3577 V26928 3577 V26929 3577 V26930 3577 V26932 3577 V26933 3577 V26934 3577 V26935 3577 V26950 3577 V26951 3577 V26952 3577 V26953 3577 V26954 3577 V26955 3577 V27245 3577 V27246 3577 V27247 3577 V27248 3577 V27249 3577 V27250 3577 V27251 3577 V27252 3577 V27253 3577 V27254 3577 V27255 3577 V27256 3577 V27257 3577 V27258 3577 V27259 3577 V27260 3577 V27261 3577 V27262 3577 V27264 3577 V27265 3577 V27266 3577 V27267 3577 V27268 3577 V27269 3577 V27270 3577 V27271 3577 V27272 3577 V27273 3577 V27274 3577 V27275 3577 V27276 3577 V27277 3577 V27278 3577 V27279 3577 V27280 3577 V27281 3577 V27282 3577 V27283 3577 V27284 3577 V27285 3577 V27286 3577 V27287 3577 V27288 3577 V27289 3577 V27290 3577 V27291 3577 V27292 3577 V27293 3577 V27294 3577 V27295 3577 V27296 3577 V27297 3577 V27298 3577 V27299 3577 V27300 3577 V27301 3577 V27302 3577 V27303 3577 V27304 3577 V27305 3577 V27306 3577 V27307 3577 V27308 3577 V27309 3577 V27310 3577 V27312 3577 V27313 3577 V27314 3577 V27315 3577 V27316 3577 V27317 3577 V27318 3577 V27319 3577 V27320 3577 V27322 3577 V27323 3577 V27324 3577 V27325 3577 V27326 3577 V27327 3577 V27328 3577 V27329 3577 V27330 3577 V27331 3577 V27332 3577 V27333 3577 V27334 3577 V27335 3577 V27336 3577 V27337 3577 V27338 3577 V27339 3577 V27340 3577 V27341 3577 V27342 3577 V27343 3577 V27344 3577 V27345 3577 V27346 3577 V27347 3577 V27348 3577 V27349 3577 V27350 3577 V27351 3577 V27352 3577 V27353 3577 V27354 3577 V27355 3577 V27356 3577 V27357 3577 V27358 3577 V27359 3577 V27360 3577 V27361 3577 V27362 3577 V27363 3577 V27364 3577 V27365 3577 V27366 3577 V27367 3577 V27368 3577 V27369 3577 V27370 3577 V27371 3577 V27372 3577 V27373 3577 V27374 3577 V27375 3577 V27376 3577 V27377 3577 V27378 3577 V27379 3577 V27380 3577 V27381 3577 V27382 3577 V27383 3577 V27384 3577 V27385 3577 V27386 3577 V27387 3577 V27388 3577 V27389 3577 V27390 3577 V27391 3577 V27392 3577 V27393 3577 V27394 3577 V27395 3577 V27396 3577 V27397 3577 V27398 3577 V27399 3577 V27400 3577 V27401 3577 V27402 3577 V27403 3577 V27404 3577 V27405 3577 V27406 3577 V27407 3577 V27408 3577 V27409 3577 V27410 3577 V27411 3577 V27412 3577 V27413 3577 V27414 3577 V27415 3577 V27416 3577 V27417 3577 V27418 3577 V27419 3577 V27420 3577 V27421 3577 V27422 3577 V27423 3577 V27424 3577 V27425 3577 V27426 3577 V27427 3577 V27428 3577 V27429 3577 V27430 3577 V27431 3577 V27432 3577 V27433 3577 V27434 3577 V27435 3577 V27436 3577 V27437 3577 V27438 3577 V27439 3577 V27440 3577 V27441 3577 V27442 3577 V27443 3577 V27444 3577 V27445 3577 V27447 3577 V27448 3577 V27449 3577 V27450 3577 V27451 3577 V27452 3577 V27453 3577 V27454 3577 V27455 3577 V27456 3577 V27457 3577 V27458 3577 V27459 3577 V27460 3577 V27461 3577 V27464 3577 V27465 3577 V27466 3577 V27467 3577 V27468 3577 V27469 3577 V27470 3577 V27471 3577 V27472 3577 V27473 3577 V27474 3577 V27475 3577 V27476 3577 V27478 3577 V27479 3577 V27480 3577 V27481 3577 V27482 3577 V27483 3577 V27484 3577 V27485 3577 V27486 3577 V27487 3577 V27488 3577 V27489 3577 V27490 3577 V27491 3577 V27492 3577 V27493 3577 V27494 3577 V27495 3577 V27496 3577 V27497 3577 V27498 3577 V27499 3577 V27500 3577 V27501 3577 .V18378 3577 .V18391 3577 V18382 3577 V18383 3577 V18384 3577 V18385 3577 V18386 3577 V18387 3577 V18388 3577 V18390 3577 V18392 3577 V18393 3577 V18394 3577 V18395 3577 V18396 3577 V18397 3577 V18398 3577 V18399 3577 V18401 3577 V18402 3577 V18403 3577 V18404 3577 V18405 3577 V18406 3577 V18407 3577 V18408 3577 V18409 3577 V18410 3577 V18411 3577 V18412 3577 V18413 3577 V18414 3577 V18415 3577 V18416 3577 V18417 3577 V18421 3577 V18423 3577 V18426 3577 V18427 3577 V18428 3577 V18429 3577 V18430 3577 V18431 3577 V18433 3577 V19334 3577 V19341 3577 V19345 3577 V19348 3577 V19396 3577 V19397 3577 V15046 3578 V18667 3578 V26243 3578 V27148 3578 V27149 3577 V28503 3577 V28595 3578 V26300 3578 V26550 3578 V26561 3578 V26568 3578 V26569 3578 V26571 3578 V26245 3578 V27147 3578 V27150 3578 V27151 3577 V18341 3577 V18342 3577 V19544 3577 V19999 3315 E12JS1 3578 V26556 3578 V26557 3578 V26558 3578 V26565 3578 V26566 3578 V27169 3635 13FTX 3315 E12JS1 3577 V28618 3578 U26559 3577 V28501 3577 V28510 3577 V28523 3449 V27200 3449 V27201 3449 V27202 3449 V27203 3449 V27204 3449 V27205 3449 V27206 3449 V27208 3449 V27209 3449 V27210 3449 V27211 3449 V27212 3449 V27213 3449 V27214 3449 V27215 3449 V27216 3449 V27217 3449 V27219 3449 V27221 3449 V27222 3449 V27223 3449 V27224 3449 V27225 3449 V27226 3449 V27227 3449 V27228 3449 V27229 3449 V27230 3449 V27231 3449 V27232 3449 V27233 3449 V27234 3449 V27235 3449 V27236 3449 V27237 3449 V27238 3449 V27242 3449 V27244 3449 V27943 3449 V27944 3449 V27946 3449 V14252 3449 V25159 3449 V25737 3449 V25741 3449 V25752 3449 V25762 3449 V25766 3449 V25767 3449 V25768 3449 V25769 3449 V25770 3449 V25771 3449 V25772 3449 V25774 3449 V25775 3449 V25776 3449 V25778 3449 V25780 3449 V25781 3449 V25782 3449 V25783 3449 V25784 3449 V25785 3449 V25786 3449 V25787 3449 V25788 3449 V25789 3449 V25790 3449 V25792 3449 V25794 3577 V26783 3577 V26843 3577 V26984 3412 12CP1 3577 V19990 3578 V26563 3449 V25856 3449 V25861 3449 V27207 3449 V27218 3449 V27220 3449 V27240 3449 V27241 3449 V27243 3449 V27942 3449 V27945 3449 V28183 3449 V28184 3449 V28185 3449 V28186 3449 V28187 3449 V28188 3449 V28189 3449 V28190 3449 V28191 3449 V28192 3449 V28193 3449 V28194 3449 V28195 3449 V28196 3449 V28197 3449 V28198 3449 V28199 3449 V28200 3449 V28201 3449 V28202 3449 V28203 3449 V28204 3449 V28205 3449 V28206 3449 V28207 3449 V28208 3449 V28210 3449 V28211 3449 V28212 3449 V28213 3449 V28214 3449 V28215 3449 V28216 3449 V28217 3449 V28218 3449 V28219 3449 V28220 3449 V28221 3449 V28222 3449 V28223 3449 V28224 3449 V28225 3449 V28226 3449 V28227 3449 V28228 3449 V28229 3449 V28230 3449 V28232 3449 V25137 3577 V28586 3315 E12JS1 3577 V19596 3578 V25929 3578 V25933 3578 V25934 3578 V25935 3578 V26091 3578 V26108 3578 V24319 3578 V24601 3578 V24607 3578 V24693 3578 V24692 3449 V25857 3449 V25864 3578 V25927 3578 V26170 3315 E12JS1 3577 V28353 3577 V28354 3577 V28387 3577 V28394 3577 V28395 3577 V28397 3577 V28401 3577 V28402 3577 V28403 3577 V28404 3577 V28405 3577 V28406 3577 V28407 3577 V28408 3577 V28409 3577 V28522 3577 V28527 3577 V28530 3577 V28533 3577 V28535 3577 V28536 3577 V28537 3577 V28538 3577 V28587 3577 V28592 3577 V28616 3577 V28617 3577 V28619 3577 V28620 3577 V28621 3578 V25925 3578 V26142 3578 V26155 3578 V24325 3578 V24619 3577 V19545 3577 V12960 3577 V14970 3577 V19987 3578 V26560 3578 V26562 3578 V26564 3449 V25855 3449 V25858 3449 V25859 3449 V25860 3449 V25862 3449 V25863 3449 V28844 3449 V28845 3449 V28862 3449 V28866 3578 V26111 3577 V18380 3577 V18470 3577 V19343 3577 V19347 3577 V19349 3577 V19376 3577 V19378 3577 V19385 3577 V19386 3577 V28660 3577 V28661 3577 V28662 3577 V28663 3577 V28664 3577 V28665 3577 V28666 3577 V28667 3577 V28668 3577 V28669 3577 V28670 3577 V28671 3577 V28672 3577 V28673 3577 V28674 3577 V28675 3577 V28676 3577 V28678 3577 V28688 3577 V28697 3577 V28607 3315 E12NC1 3449 V28817 3449 V28820 3449 V28821 3449 V28832 3449 V28835 3449 V28838 3449 V28847 3449 V28856 3449 V28865 3449 V28868 3577 V28649 3577 V28650 3577 V28651 3577 V28652 3577 V28653 3577 V28654 3577 V28657 3577 V28658 3577 V28659 3577 V28702 3315 E12JS1 3315 E12JS1 3315 E12NC1 3315 E12NC1 3315 13J06 3449 V28816 3449 V28818 3449 V28843 3449 V28854 3578 V26536 3578 V26525 3578 V26524 3578 V26531 3578 V26570 3578 V26551 3578 V26574 3577 V28008 3577 V28009 3577 V28010 3577 V28011 3577 V28012 3577 V28013 3577 V28014 3577 V28015 3577 V28016 3577 V28017 3577 V28018 3577 V28019 3577 V28020 3577 V28022 3577 V28023 3577 V28024 3577 V28025 3577 V28026 3577 V28027 3577 V28029 3577 V28031 3577 V28032 3577 V28033 3577 V28034 3577 V28035 3577 V28036 3577 V28037 3577 V28038 3577 V28040 3577 V28041 3577 V28042 3577 V28043 3577 V28304 3577 V28305 3577 V28306 3577 V28307 3577 V28308 3577 V28309 3577 V28310 3577 V28311 3577 V28313 3577 V28314 3577 V28315 3577 V28316 3577 V28317 3577 V28318 3577 V28319 3577 V28320 3577 V28321 3577 V28322 3577 V28323 3577 V28324 3577 V28325 3577 V28326 3577 V28327 3577 V28328 3577 V28329 3577 V28330 3577 V28331 3577 V28333 3577 V28334 3577 V28335 3577 V28336 3577 V28337 3577 V28338 3577 V28339 3577 V28340 3577 V28642 3577 V28643 3577 V28644 3577 V28645 3577 V28646 3577 V28355 3577 V28356 3577 V28357 3577 V28358 3577 V28359 3577 V28360 3577 V28361 3577 V28362 3577 V28363 3577 V28364 3577 V28365 3577 V28366 3577 V28367 3577 V28368 3577 V28369 3577 V28370 3577 V28371 3577 V28372 3577 V28373 3577 V28374 3577 V28375 3577 V28376 3577 V28377 3577 V28378 3577 V28379 3577 V28380 3577 V28381 3577 V28382 3577 V28383 3577 V28384 3577 V28385 3577 V28386 3577 V28388 3577 V28389 3577 V28390 3577 V28391 3577 V28392 3577 V28393 3577 V28396 3577 V28398 3577 V28399 3577 V28400 3577 V28411 3577 V28412 3577 V28413 3577 V28414 3577 V28415 3577 V28416 3577 V28417 3577 V28418 3577 V28419 3577 V28420 3577 V28421 3577 V28422 3577 V28423 3577 V28424 3577 V28425 3577 V28426 3577 V28427 3577 V28428 3577 V28429 3577 V28430 3577 V28431 3577 V28432 3577 V28433 3577 V28434 3577 V28435 3577 V28436 3577 V28437 3577 V28438 3577 V28439 3577 V28440 3577 V28441 3577 V28442 3577 V28443 3577 V28444 3577 V28445 3577 V28446 3577 V28447 3577 V28448 3577 V28449 3577 V28450 3577 V28451 3577 V28452 3577 V28453 3577 V28454 3577 V28455 3577 V28456 3577 V28457 3577 V28458 3577 V28459 3577 V28460 3577 V28461 3577 V28462 3577 V28463 3577 V28464 3577 V28466 3577 V28467 3577 V28468 3577 V28469 3577 V28471 3577 V28472 3577 V28473 3577 V28474 3577 V28475 3577 V28476 3577 V28477 3577 V28478 3577 V28479 3577 V28480 3577 V28481 3577 V28482 3577 V28483 3577 V28484 3577 V28465 3577 V28470 3577 V28485 3577 V28486 3577 V28487 3577 V28488 3577 V28489 3577 V28490 3577 V28491 3577 V28494 3577 V28495 3577 V28496 3577 V28497 3577 V28498 3577 V28499 3577 V28500 3577 V28502 3577 V28504 3577 V28505 3577 V28506 3577 V28507 3577 V28508 3577 V28509 3577 V28511 3577 V28512 3577 V28513 3577 V28514 3577 V28515 3577 V28516 3577 V28517 3577 V28518 3577 V28519 3577 V28520 3577 V28521 3577 V28524 3577 V28525 3577 V28526 3577 V28528 3577 V28529 3577 V28531 3577 V28532 3577 V28534 3577 V28539 3577 V28540 3577 V28541 3577 V28542 3577 V28543 3577 V28544 3577 V28545 3577 V28546 3577 V28547 3577 V28548 3577 V28549 3577 V28550 3577 V28551 3577 V28552 3577 V28553 3577 V28554 3577 V28555 3577 V28556 3577 V28557 3577 V28558 3577 V28559 3577 V28560 3577 V28561 3577 V28562 3577 V28563 3577 V28564 3577 V28565 3577 V28566 3577 V28567 3577 V28568 3577 V28569 3577 V28570 3577 V28571 3577 V28572 3577 V28573 3577 V28574 3577 V28575 3577 V28576 3577 V28577 3577 V28578 3577 V28579 3577 V28580 3577 V28581 3577 V28582 3577 V28583 3577 V28584 3577 V28585 3577 V28588 3577 V28589 3577 V28590 3577 V28591 3577 V28593 3577 V28594 3577 V28596 3577 V28597 3577 V28599 3577 V28600 3577 V28603 3577 V28608 3577 V28609 3577 V28610 3577 V28611 3577 V28612 3577 V28613 3577 V28614 3577 V28615 3577 V13186 3577 V14200 3577 V14450 3577 V15055 3577 V28309. 3577 V28322. 3315 E12JS1 3578 V26509 3578 V26515 3578 V26526 3578 V26530 3578 V26533 3578 V26539 3578 V26541 3578 V26542 3578 V26545 3578 V26547 3449 V28258 3449 V28262 3449 V28263 3449 V28265 3449 V28266 3449 V28268 3449 V28269 3449 V28270 3449 V28271 3449 V28272 3449 V28273 3449 V28274 3449 V28275 3449 V28280 3449 V28281 3449 V28282 3449 V28283 3449 V28284 3449 V28285 3449 V28286 3449 V28287 3449 V28288 3449 V28289 3449 V28290 3449 V28291 3449 V28292 3449 V28293 3449 V28294 3449 V28295 3449 V28296 3449 V28297 3449 V28298 3449 V28299 3449 V28300 3449 V28301 3449 V28302 3449 V28744 3449 V28745 3449 V28746 3449 V28747 3449 V28748 3449 V28749 3449 V28750 3449 V28751 3449 V28752 3449 V28753 3449 V28754 3449 V28755 3449 V28756 3449 V28757 3449 V28758 3449 V28759 3449 V28760 3449 V28761 3449 V28762 3449 V28763 3449 V28764 3449 V28765 3449 V28766 3449 V28767 3449 V28768 3449 V28769 3449 V28770 3449 V28771 3449 V28776 3449 V28777 3449 V28778 3449 V28779 3449 V28780 3449 V28781 3449 V28782 3449 V28783 3449 V28784 3449 V28785 3449 V28786 3449 V28787 3449 V28788 3449 V28789 3449 V28790 3449 V28791 3449 V28792 3449 V28793 3449 V28794 3449 V28796 3449 V28797 3449 V28801 3449 V28802 3449 V28803 3449 V28804 3449 V28805 3449 V28806 3449 V28807 3449 V28808 3449 V28809 3449 V28811 3449 V28812 3449 V28810 3449 V28704 3449 V28705 3449 V28706 3449 V28707 3577 V28647 3577 V28656 3577 V28655 3635 12N1S 3635 12N1S 3635 13GKC 3635 13GTC 3635 13GTC 3635 V15194 3635 V25730 3635 V25738 3635 V25759 3635 V25795 3635 V25801 3635 V25817 3635 V25822 3635 V25838 3635 V25842 3635 V25852 3635 V25871 3635 V27067 3635 V27074 3635 V27087 3635 V27094 3449 V28231 3449 V28233 3449 V28234 3449 V28235 3449 V28236 3449 V28237 3449 V28238 3449 V28239 3449 V28240 3449 V28241 3449 V28242 3449 V28243 3449 V28244 3449 V28245 3449 V28246 3449 V28247 3449 V28248 3449 V28249 3449 V28250 3449 V28251 3449 V28252 3449 V28253 3449 V28254 3449 V28255 3449 V28256 3449 V28257 3449 V28259 3449 V28260 3449 V28264 3449 V28726 3449 V28727 3449 V28728 3449 V28729 3449 V28730 3449 V28731 3449 V28732 3449 V28733 3449 V28734 3449 V28737 3449 V28738 3449 V28739 3449 V28740 3449 V28741 3449 V28742 3449 V28743 3449 V28772 3449 V28773 3449 V28774 3449 V28775 3449 V28926 3449 V28927 3449 V28928 3449 V28930 3449 V28931 3449 V28932 3449 V28933 3449 V28934 3449 V28935 3449 V28936 3449 V28937 3449 V28938 3449 V28939 3449 V28940 3449 V28941 3449 V28942 3449 V28870 3449 V28871 3449 V28872 3449 V28873 3449 V28874 3449 V28875 3449 V28876 3449 V28877 3449 V28878 3449 V28879 3449 V28880 3449 V28882 3449 V28883 3449 V28884 3449 V28885 3449 V28886 3449 V28887 3449 V28888 3449 V28889 3449 V28893 3449 V28895 3449 V28896 3449 V28897 3449 V28898 3449 V28899 3449 V28900 3449 V28901 3449 V28902 3449 V28903 3449 V28904 3449 V28905 3449 V28906 3449 V28908 3449 V28909 3449 V28910 3577 V18448 3577 V18449 3577 V18450 3577 V18451 3577 V18454 3577 V18455 3577 V18461 3577 V18463 3577 V18464 3577 V18466 3577 V18468 3577 V18472 3577 V18465 3577 V18469 3315 E12JS1 3635 12N1S 3635 13FTX 3635 13GTC 3636 R13FZJ 3636 13GBJ 3636 13HCG 3577 V18424 3577 V18436 3577 V18440 3577 V18446 3577 V18447 3577 V18452 3577 V18453 3577 V18458 3577 V18459 3577 V18460 3578 V15481 3578 V15482 3578 V18632 3578 V18636 3578 V18638 3578 V18647 3578 V18654 3578 V18657 3578 V18659 3578 V18661 3578 V18664 3578 V18669 3578 V18677 3578 V18679 3578 V18681 3578 V18737 3578 V18950 3578 V19465 3578 V19480 3578 V19481 3577 V18400 3577 V18420 3636 R13FZJ 3636 13GBJ 3577 V18432 3577 V18434 3577 V18438 3577 V18444 3577 V18445 3577 V18389 3577 V18418 3577 V18419 3577 V18437 3577 V18441 3577 V18443 3577 V18435 3577 V18439 3578 V25915 3578 V26554 3577 V18422 3577 V18425 3578 V19475 3636 R13FZJ 3636 13GBJ 3578 V26506 3578 V26535 3578 V26520 3578 V26532 3578 V27852 3578 V27887 3578 V27888 3578 V27891 3578 V18655 3578 V18674 3578 V18680 3449 V28881 3449 V28890 3449 V28891 3449 V28892 3449 V28894 3577 V12676 3577 V13179 3577 V13264 3577 V14424 3577 V14435 3577 V14459 3577 V14958 3577 V14960 3577 V14990 3577 V15018 3577 V15034 3577 V15049 3577 V15051 3577 V15058 3577 V19351 3577 V19352 3577 V19353 3577 V19355 3577 V19356 3577 V19357 3577 V19366 3577 V19374 3577 V19380 3577 V19389 3577 V19395 3577 V19354 3577 V19358 3578 V27851 3578 V27878 3449 V28800 3449 V28907 3449 N28795 3577 V19362 3577 V19365 3577 V19369 3577 V19390 3577 V19393 3578 V27871 3577 V19370 3577 V19377 3577 V19379 3577 V19372 3577 V30931 3577 V30932 3577 V30933 3577 V30920 3577 V30924 3577 V30926 3577 V30929 3577 V30934 3577 V19363 3577 V19367 3578 V27865 3577 V28057 3577 V28052 3577 V28039 3577 V28045 3577 V28046 3577 V28051 3577 V28054 3577 V28044 3578 V27859 3578 V27867 3578 V27839 3578 V27853 3578 V27857 3578 V27841 3578 V27844 3449 V32501 3449 V32502 3449 V32503 3449 V32504 3449 V32505 3449 V32844 3449 V32846 3449 V32847 3449 V32848 3449 V32850 3449 V32851 3449 V32852 3449 V32853 3449 V32855 3449 V32856 3449 V32857 3449 V32858 3449 V32859 3449 V32860 3449 V32861 3449 V32863 3449 V32864 3449 V32865 3449 V32866 3449 V32867 3449 V32868 3449 V32869 3449 V32871 3449 V32872 3449 V32873 3449 V32874 3449 V32875 3449 V32878 3449 V32879 3449 V32881 3449 V32882 3449 V32884 3449 V32885 3449 V32886 3449 V32887 3449 V32888 3449 V32889 3449 V32890 3449 V32891 3449 V32892 3449 V32893 3449 V32894 3449 V32895 3449 V32896 3449 V32898 3449 V32899 3449 V32900 3449 V32902 3449 V32903 3449 V32904 3449 V32905 3449 V32906 3449 V32907 3449 V32908 3449 V32909 3449 V32910 3449 V32911 3449 V32912 3449 V32913 3449 V32914 3449 V32915 3449 V32917 3449 V32919 3449 V32920 3449 V32921 3449 V32923 3449 V32924 3449 V32925 3449 V32926 3449 V32927 3449 V32928 3449 V32930 3449 V32931 3449 V32932 3449 V32933 3449 V32934 3449 V32935 3449 V32937 3449 V32938 3449 V32939 3449 V32940 3449 V32941 3449 V32942 3449 V32943 3449 V32944 3449 V32945 3449 V32946 3449 V32947 3449 V32948 3449 V32949 3449 V32951 3449 V32952 3449 V32953 3449 V32954 3449 V32956 3449 V31500 3449 V31501 3449 V31502 3449 V31503 3449 V31504 3449 V31505 3449 V31506 3449 V31507 3449 V31508 3449 V31509 3449 V31510 3449 V31511 3449 V31512 3449 V31513 3449 V31514 3449 V31515 3449 V31516 3449 V31517 3449 V31518 3449 V31519 3449 V31520 3449 V31521 3449 V31522 3449 V31523 3449 V31524 3449 V31525 3449 V31526 3449 V31527 3449 V31528 3449 V31529 3449 V31530 3449 V31531 3449 V31532 3449 V31533 3449 V31534 3449 V31535 3449 V31536 3449 V31537 3449 V31538 3449 V31540 3449 V31541 3449 V31542 3449 V31543 3449 V31544 3449 V31545 3449 V31546 3449 V31547 3449 V31548 3449 V31557 3449 V31558 3449 V31560 3449 V31562 3449 V31563 3449 V31564 3449 V31565 3449 V31566 3449 V31567 3449 V31568 3449 V31569 3449 V31571 3449 V31573 3449 V31574 3449 V31575 3449 V31576 3449 V31577 3449 V31579 3449 V31580 3449 V31581 3449 V31582 3449 V31583 3449 V31584 3449 V31585 3449 V31586 3449 V31587 3449 V31588 3449 V31589 3449 V31590 3449 V31591 3449 V31592 3449 V31593 3449 V31594 3449 V31595 3449 V31596 3449 V31597 3449 V31598 3449 V31599 3449 V31600 3449 V31602 3449 V31603 3449 V31604 3449 V31605 3449 V31606 3449 V31607 3449 V31608 3449 V31609 3449 V32520 3449 V32521 3449 V32522 3449 V32523 3449 V32524 3577 V28047 3577 V28049 3577 V28050 3577 V28053 3577 V28030 3577 V28048 3577 V28055 3577 V28056 3577 V28058 3577 V31020 3577 V31025 3577 V31031 3577 V31044 3577 V31046 3577 V31047 3577 V31048 3577 V31055 3577 V31105 3577 V31060 3577 V31099 3577 V31106 3577 V31114 3577 V30900 3577 V30903 3577 V30904 3577 V30905 3577 V30906 3577 V30908 3577 V30911 3577 V30912 3577 V30913 3577 V30914 3577 V30915 3577 V30916 3577 V30917 3577 V30919 3577 V30922 3577 V30923 3577 V30925 3577 V30927 3577 V30930 3577 V31086 3577 V31095 3577 V31096 3577 V31098 3577 V31101 3577 V31102 3577 V31109 3577 V31121 3577 V31123 3577 V31124 3577 V31125 3577 V31127 3577 V31128 3577 V31129 3577 V31130 3577 V31131 3577 V31132 3577 V31133 3577 V31134 3577 V31135 3577 V31136 3577 V31137 3577 V31138 3449 U32809 3578 V24323 3578 V24838 3577 V31071 3577 V31088 3577 V31103 3577 V31104 3577 V31110 3577 V31111 3577 V31116 3577 V31118 3577 V31119 3577 V31122 3577 V19436 3636 V25493 3636 V25506 3636 V26601 3636 V26602 3636 V26603 3636 V26604 3636 V26607 3636 V26610 3636 V26612 3636 V26615 3636 V26616 3636 V26618 3636 V26619 3636 V26620 3636 V26628 3636 V26630 3636 V26633 3636 V26637 3636 V26638 3636 V26644 3636 V26645 3636 V26648 3636 V26649 3636 V26651 3636 V26653 3636 V26655 3636 V26657 3636 V26658 3636 V26660 3636 V26664 3577 V18337 3577 V19430 3577 V19432 3577 V19991 3577 V31094 3577 V31120 3578 V27849 3635 V31693 3635 V31698 3635 V31699 3635 V31701 3635 V31702 3635 V31706 3635 V31715 3635 V31718 3635 V31719 3635 V31726 3635 V31727 3635 V31730 3635 V31734 3635 V31736 3635 V31746 3635 V31747 3635 V31751 3635 V31759 3635 V31760 3635 V32964 3635 V32966 3635 V32967 3635 V32969 3635 V32972 3635 V32973 3635 V32974 3635 V32975 3635 V32977 3635 V32980 3635 V32985 3635 V31682 3635 V31683 3635 V31684 3635 V31685 3635 V31687 3635 V31688 3635 V31689 3635 V31690 3635 V31692 3635 V31696 3635 V31700 3635 V31703 3635 V31704 3635 V31707 3635 V31708 3635 V31710 3635 V31713 3635 V31720 3635 V31721 3635 V31722 3635 V31723 3635 V31724 3635 V31725 3635 V31735 3635 V31749 3635 V31750 3635 V31753 3635 V31754 3635 V31756 3635 V31757 3635 V31758 3635 V31761 3635 V31762 3635 V32808 3635 V32810 3635 V32813 3635 V32814 3635 V32817 3635 V32820 3635 V32823 3635 V32828 3635 V32831 3635 V32832 3635 V32837 3635 V32842 3635 V32961 3635 V32962 3635 V32963 3635 V32965 3635 V31755 3577 V31117 3577 V31045 3577 V31065 3577 V31093 3577 V31107 3577 V31113 3578 V27854 3578 V27886 3577 V31006 3577 V31032 3577 V31040 3577 V31052 3577 V31053 3577 V31057 3577 V31061 3577 V31064 3577 V31066 3577 V31068 3577 V31097 3577 V31100 3577 V31108 3577 V31112 3577 V31115 3578 V27843 3578 V27846 3578 V27850 3578 V27860 3578 V27864 3577 V31000 3577 V31019 3577 V31030 3577 V31036 3577 V31051 3635 V31681 3635 V31686 3635 V31691 3635 V31694 3635 V31695 3635 V31705 3635 V31709 3635 V31711 3635 V31714 3635 V31728 3635 V31729 3635 V32796 3635 V32819 3635 V32833 3635 V32971 3635 V28798 3635 V28799 3635 V28813 3635 V28814 3635 V28823 3635 V28826 3635 V28830 3635 V28833 3635 V28837 3635 V32298 3635 V32299 3635 V32300 3635 V32301 3635 V32302 3635 V32303 3635 V32304 3635 V32305 3635 V32306 3635 V32307 3635 V32308 3635 V32309 3635 V32310 3635 V32311 3635 V32312 3635 V32313 3635 V32314 3635 V32315 3635 V32316 3635 V32816 3635 V27977 3635 V27978 3635 V27979 3635 V27980 3635 V27981 3635 V27982 3635 V27983 3635 V27984 3635 V27985 3635 V27987 3635 V27988 3635 V27989 3635 V27990 3635 V27991 3635 V27992 3635 V27993 3635 V27994 3635 V27995 3635 V27996 3635 V27997 3635 V27998 3635 V27999 3635 V28000 3635 V28001 3635 V28002 3635 V28006 3635 V28815 3635 V31731 3635 V31732 3635 V31737 3635 V31738 3635 V31740 3635 V31741 3635 V31742 3635 V31743 3635 V31744 3635 V31745 3635 V32279 3635 V32280 3635 V32281 3635 V32282 3635 V32283 3635 V32284 3635 V32286 3635 V32287 3635 V32290 3635 V32291 3635 V32293 3412 E14CXC 3449 V32733 3449 V32734 3449 V32735 3449 V32736 3449 V32737 3449 V32738 3449 V32739 3449 V32740 3449 V32741 3449 V32742 3449 V32743 3449 V32744 3449 V32745 3449 V32746 3449 V32747 3449 V32748 3449 V32749 3449 V32750 3449 V32751 3449 V32752 3449 V32753 3449 V32754 3449 V32755 3449 V32756 3449 V32757 3449 V32758 3449 V32759 3449 V32760 3449 V32761 3449 V32762 3449 V32763 3449 V32764 3449 V32765 3449 V32766 3449 V32767 3449 V32768 3449 V32769 3449 V32770 3449 V32771 3449 V32772 3449 V32773 3449 V32774 3449 V32776 3449 V32777 3449 V32778 3449 V32779 3449 V32780 3449 V32781 3449 V32782 3449 V32783 3449 V32784 3449 V32785 3449 V32786 3449 V32787 3449 V32788 3449 V32789 3449 V32790 3449 V32791 3449 V32792 3449 V32793 3449 V32794 3449 V32795 3449 V32797 3449 V32798 3449 V32801 3449 V32802 3449 V32803 3449 V32804 3449 V32805 3449 V32806 3449 V32807 3449 V32811 3449 V32812 3449 V32815 3449 V32821 3449 V32822 3449 V32824 3449 V32825 3449 V32826 3449 V32829 3449 V32830 3449 V32834 3449 V32835 3449 V32836 3449 V32838 3449 V32839 3449 V32840 3449 V32841 3449 V32843 3449 V32506 3449 V32507 3449 V32508 3449 V32509 3449 V32510 3449 V32511 3449 V32512 3449 V32513 3449 V32514 3449 V32515 3449 V32516 3449 V32517 3449 V32518 3449 V32519 3449 V32845 3449 V32849 3449 V32862 3449 V32877 3449 V32880 3449 V32897 3449 V32918 3449 V32950 3449 V32957 3449 V31539 3449 V31550 3449 V31572 3449 V31578 3449 V31610 3449 V31611 3449 V31612 3449 V31613 3449 V31614 3449 V31615 3449 V31616 3449 V31617 3449 V31618 3449 V31619 3449 V31620 3449 V31621 3449 V31622 3449 V31623 3449 V31624 3449 V31625 3449 V31626 3449 V31627 3449 V31628 3449 V31629 3449 V31630 3449 V31631 3449 V31632 3449 V31633 3449 V31635 3449 V31636 3449 V31637 3449 V31638 3449 V31639 3449 V31640 3449 V31641 3449 V31642 3449 V31643 3449 V31644 3449 V31645 3449 V31646 3449 V31647 3449 V31648 3449 V31649 3449 V31650 3449 V31651 3449 V31652 3449 V31653 3449 V31654 3449 V31655 3449 V31656 3449 V31657 3449 V31658 3449 V31659 3449 V31660 3449 V31661 3449 V31662 3449 V31663 3449 V31664 3449 V31665 3449 V31666 3449 V31667 3449 V31668 3449 V31669 3449 V31670 3449 V31671 3449 V31672 3449 V31673 3449 V31674 3449 V31675 3449 V31676 3449 V31677 3449 V31678 3449 V31679 3449 V31680 3449 V31631523 3449 V32799 3449 V32818 3449 V32827 3449 V32278 3449 V32285 3449 V32292 3449 V32296 3449 V32999 3449 V33002 3449 V33003 3449 V33004 3449 V33008 3449 V33009 3578 V27869 3578 V27874 3578 V27875 3578 V27882 3578 V27863 3577 V31023 3577 V31049 3578 V27861 3523 V32321 3523 V32323 3523 V32324 3523 V32330 3523 V32338 3523 V32339 3523 V32340 3523 V32345 3523 V32347 3523 V32348 3523 V32349 3523 V32350 3523 V32351 3523 V32352 3523 V32353 3523 V32354 3523 V32356 3523 V32357 3523 V32358 3523 V32359 3523 V32360 3523 V32363 3523 V32364 3523 V32365 3523 V32366 3523 V32367 3523 V32368 3523 V32369 3523 V32370 3523 V32371 3523 V32372 3523 V32373 3523 V32374 3523 V32375 3523 V32376 3523 V32377 3523 V32378 3523 V32379 3523 V32380 3523 V32381 3523 V32382 3523 V32383 3523 V32384 3523 V32385 3523 V32386 3523 V32387 3523 V32388 3523 V32389 3523 V32390 3523 V32391 3523 V32392 3523 V32393 3523 V32394 3523 V32395 3523 V32396 3523 V32397 3523 V32398 3523 V32399 3523 V32400 3523 V32403 3523 V32404 3523 V32406 3523 V32409 3523 V32413 3523 V32415 3577 V31013 3577 V31035 3577 V310550 3578 V26510 3578 V26513 3315 E12JS1 3577 V30999 3577 V31003 3577 V31005 3577 V31021 3577 V31022 3577 V31028 3577 V31033 3577 V31037 3577 V31043 3577 V31001 3577 V31002 3577 V31014 3577 V31024 3577 V31027 3577 V31007 3577 V31011 3577 V31026 3577 V31041 3577 V31042 3578 V27840 3578 V27876 3577 V28648 3577 V28680 3577 V31004 3577 V31008 3577 V31015 3577 V31016 3577 V31038 3578 V26508 3578 V26511 3578 V26514 3578 V26516 3578 V26517 3578 V26523 3578 V26527 3578 V26529 3578 V26540 3578 V26546 3577 V31012 3577 V31018 3577 V31039 3577 V28341 3577 V28303 3577 V28343 3577 V28348 3577 V31009 3577 V31010 3578 V25926 3578 V27842 3578 V27856 3578 V27858 3578 V27866 3577 V28350 3577 V28342 3577 V28344 3577 V28345 3577 V28346 3577 V28349 3577 V28352 3577 V28351 3577 V31835 3522 V33087 3522 V33114 3522 V33120 3577 V31830 3577 V31831 3577 V31839 3577 V31841 3577 V31847 3577 V31877 3577 V31878 3577 V31880 3577 V32164 3577 V32167 3577 V32168 3577 V32174 3577 V32177 3577 V32179 3577 V32181 3577 V32186 3577 V32190 3577 V32191 3577 U31824 3578 V25914 3578 V25920 3578 V25922 3578 V25917 3578 V25924 3578 V25932 3578 V26501 3578 V26518 3577 V28694 3577 V28699 3577 V28700 3577 V28701 3577 V28347 3577 V28679 3578 V26519 3578 V26502 3578 V26503 3578 V26504 3578 V26505 3578 V26507 3578 V26521 3578 V26522 3578 V26534 3578 V26537 3578 V27868 3577 V28689 3578 V27881 3522 V33083 3522 V33084 3522 V33085 3522 V33086 3522 V33088 3522 V33089 3522 V33090 3522 V33091 3522 V33092 3522 V33093 3522 V33094 3522 V33095 3522 V33096 3522 V33097 3522 V33098 3522 V33099 3522 V33100 3522 V33101 3522 V33102 3522 V33103 3522 V33104 3522 V33105 3522 V33106 3522 V33107 3522 V33108 3522 V33109 3522 V33110 3522 V33111 3522 V33112 3522 V33113 3522 V33115 3522 V33116 3522 V33117 3522 V33118 3522 V33119 3522 V33121 3522 V33122 3522 V33123 3522 V33124 3522 V33125 3522 V33126 3522 V33127 3522 V33128 3522 V33129 3522 V33130 3522 V33131 3522 V33132 3523 V33203 3523 V33205 3523 V33207 3523 V33208 3523 V33210 3523 V33211 3523 V33214 3523 V33215 3523 V33216 3523 V33217 3523 V33218 3523 V33219 3523 V33220 3523 V33221 3523 V33222 3523 V33223 3523 V33224 3523 V33225 3523 V33226 3523 V33227 3523 V33228 3523 V33229 3523 V33230 3523 V33231 3523 V33232 3523 V33233 3523 V33234 3523 V33235 3523 V33236 3523 V33237 3523 V33238 3523 V33239 3523 V33240 3523 V33241 3523 V33242 3523 V33243 3523 V32318 3523 V32319 3523 V32320 3523 V32322 3523 V32325 3523 V32326 3523 V32327 3523 V32329 3523 V32331 3523 V32332 3523 V32334 3523 V32335 3523 V32336 3523 V32337 3523 V32341 3523 V32342 3523 V32343 3523 V32344 3523 V32346 3523 V32401 3523 V32402 3523 V32405 3523 V32407 3523 V32410 3523 V32411 3523 V32414 3523 V32416 3523 V33012 3523 V33013 3523 V33014 3523 V33015 3523 V33016 3523 V33017 3523 V33018 3523 V33019 3523 V33020 3523 V33021 3523 V33029 3523 V33030 3523 V33031 3523 V33032 3523 V33038 3523 V33039 3523 V33040 3523 V33041 3523 V33042 3523 V33043 3523 V33044 3523 V33045 3523 V33046 3523 V33048 3523 V33049 3523 V33050 3523 V33051 3523 V33052 3523 V33053 3523 V33054 3523 V33055 3523 V33056 3523 V33057 3523 V33058 3523 V33059 3523 V33061 3523 V33062 3523 V33064 3523 V33066 3523 V33068 3523 V33069 3523 V33071 3523 V33244 3523 V33245 3523 V33341 3523 V33342 3523 V33343 3523 V33344 3523 V33345 3523 V33346 3523 V33348 3523 V33349 3523 V33350 3523 V33351 3523 V33352 3523 V33246 3523 V33248 3523 V33249 3523 V33250 3523 V33251 3523 V33252 3523 V33253 3523 V33254 3523 V33255 3523 V33256 3523 V33257 3523 V33258 3523 V33259 3523 V33260 3523 V33261 3523 V33262 3523 V33263 3523 V33264 3523 V33265 3523 V33266 3523 V33267 3523 V33268 3523 V33269 3523 V33270 3523 V33271 3523 V33272 3523 V33273 3523 V33274 3523 V33275 3523 V33276 3523 V33278 3523 V33279 3523 V33280 3523 V33281 3523 V33282 3523 V33284 3523 V33285 3523 V33286 3523 V33288 3523 V33347 3523 V33289 3523 V33290 3523 V33291 3523 V33292 3523 V33293 3523 V33294 3523 V33295 3523 V33296 3523 V33297 3523 V33298 3523 V33299 3523 V33300 3523 V33301 3523 V33302 3523 V33303 3523 V33304 3523 V33305 3523 V33306 3523 V33307 3523 V33308 3523 V33309 3523 V33310 3523 V33311 3523 V33312 3523 V33313 3523 V33314 3523 V33315 3523 V33316 3523 V33317 3523 V33318 3523 V33319 3523 V33320 3523 V33321 3523 V33322 3523 V33323 3523 V33324 3523 V33325 3523 V33326 3523 V33327 3523 V33328 3523 V33329 3523 V33331 3577 V28682 3578 V27883 3522 V34587 3522 V34588 3522 V34589 3522 V34590 3522 V34591 3522 V34592 3522 V34593 3522 V34594 3522 V34595
3522 V34596 3522 V34597 3522 V34598 3522 V34599 3522 V34600 3522 V34601 3522 V34602 3522 V34603 3522 V34604 3522 V34605 3522 V34606 3522 V34607 3522 V34608 3522 V34609 3522 V34610 3522 V34611 3522 V34612 3522 V34613 3522 V34614 3522 V34615 3522 V34616 3522 V34617 3522 V34618 3522 V34619 3522 V34620 3522 V34621 3522 V34622 3522 V34623 3522 V34624 3522 V34625 3522 V34626 3522 V34627 3522 V34628 3522 V34629 3522 V34630 3522 V34631 3522 V34632 3522 V34633 3522 V34634 3522 V34635 3522 V34636 3522 V34637 3522 V34638 3522 V34639 3522 V34640 3522 V34641 3522 V34642 3522 V34643 3522 V34644 3522 V34646 3522 V34647 3522 V34648 3522 V34649 3522 V34650 3522 V34651 3522 V34652 3522 V34653 3522 V34654 3522 V34655 3522 V34656 3522 V34657 3522 V34658 3522 V34659 3522 V34660 3522 V34661 3522 V34662 3522 V34663 3522 V34664 3522 V34665 3522 V34666 3522 V34667 3522 V34668 3522 V34669 3522 V34670 3522 V34671 3522 V34672 3522 V34673 3522 V34674 3522 V34675 3522 V34676 3522 V34677 3522 V34678 3522 V34679 3522 V34680 3522 V34681 3522 V34682 3522 V34683 3522 V34684 3522 V34685 3522 V34686 3522 V34687 3522 V34688 3522 V34689 3522 V34690 3522 V34691 3522 V34692 3522 V34693 3522 V34694 3522 V34695 3522 V34696 3522 V34697 3522 V34698 3522 V34536 3522 V34537 3522 V34538 3522 V34539 3522 V34540 3522 V34544 3522 V34545 3522 V34546 3522 V34547 3522 V34548 3522 V34549 3522 V34550 3522 V34551 3522 V34552 3522 V34553 3522 V34554 3522 V34555 3522 V34556 3522 V34557 3522 V34558 3522 V34559 3522 V34560 3522 V34561 3522 V34562 3522 V34563 3522 V34564 3522 V34565 3522 V34566 3522 V34567 3522 V34568 3522 V34569 3522 V34570 3522 V34571 3522 V34572 3522 V34573 3522 V34574 3522 V34575 3522 V34576 3522 V34577 3522 V34578 3522 V34579 3522 V34580 3522 V34532 3522 V34533 3522 V34534 3522 V34535 3522 V34422 3522 V34423 3522 V34424 3522 V34425 3522 V34426 3522 V34427 3522 V34428 3522 V34429 3522 V34430 3522 V34431 3522 V34432 3522 V34433 3522 V34434 3522 V34435 3522 V34436 3522 V34437 3522 V34438 3522 V34439 3522 V34440 3522 V34441 3522 V34442 3522 V34443 3522 V34444 3522 V34445 3522 V34446 3522 V34447 3522 V34448 3522 V34449 3522 V34450 3522 V34451 3522 V34452 3522 V34453 3522 V34454 3522 V34455 3522 V34456 3522 V34457 3522 V34458 3522 V34459 3522 V34460 3522 V34461 3522 V34462 3522 V34463 3522 V34464 3522 V34465 3522 V34466 3522 V34467 3522 V34468 3522 V34469 3522 V34470 3522 V34471 3522 V34472 3522 V34473 3522 V34474 3522 V34475 3522 V34476 3522 V34477 3522 V34478 3522 V34479 3522 V34480 3522 V34481 3522 V34482 3522 V34483 3522 V34484 3522 V34485 3522 V34486 3522 V34487 3522 V34488 3522 V34489 3522 V34490 3522 V34491 3522 V34492 3522 V34493 3522 V34494 3522 V34495 3522 V34496 3522 V34497 3522 V34498 3522 V34499 3522 V34500 3522 V34501 3522 V34502 3522 V34503 3522 V34504 3522 V34505 3522 V34506 3522 V34507 3522 V34508 3522 V34509 3522 V34510 3522 V34511 3522 V34512 3522 V34513 3522 V34514 3522 V34515 3522 V34516 3522 V34517 3522 V34518 3522 V34519 3522 V34520 3522 V34521 3522 V34522 3522 V34523 3522 V34524 3522 V34525 3522 V34526 3522 V34527 3522 V34528 3522 V34529 3522 V34530 3522 V34531 3522 V34407 3522 V34408 3522 V34409 3522 V34410 3522 V34411 3522 V34412 3522 V34413 3522 V34414 3522 V34415 3522 V34416 3522 V34417 3522 V34418 3522 V34419 3522 V34420 3522 V34421 3522 V34342 3522 V34343 3522 V34344 3522 V34345 3522 V34346 3522 V34347 3522 V34348 3522 V34349 3522 V34350 3522 V34351 3522 V34352 3522 V34353 3522 V34354 3522 V34355 3522 V34356 3522 V34357 3522 V34358 3522 V34359 3522 V34360 3522 V34361 3522 V34362 3522 V34363 3522 V34364 3522 V34365 3522 V34366 3522 V34367 3522 V34368 3522 V34369 3522 V34370 3522 V34371 3522 V34372 3522 V34373 3522 V34374 3522 V34375 3522 V34376 3522 V34377 3522 V34378 3522 V34379 3522 V34380 3522 V34381 3522 V34382 3522 V34383 3522 V34384 3522 V34385 3522 V34386 3522 V34387 3522 V34388 3522 V34389 3522 V34390 3522 V34391 3522 V34392 3522 V34393 3522 V34394 3522 V34395 3522 V34396 3522 V34397 3522 V34398 3522 V34399 3522 V34400 3522 V34401 3522 V34402 3522 V34403 3522 V34404 3522 V34405 3522 V34406 3522 V34215 3522 V34216 3522 V34217 3522 V34218 3522 V34219 3522 V34220 3522 V34221 3522 V34222 3522 V34223 3522 V34224 3522 V34225 3522 V34226 3522 V34227 3522 V34228 3522 V34229 3522 V34230 3522 V34231 3522 V34232 3522 V34233 3522 V34234 3522 V34236 3522 V34237 3522 V34238 3522 V34239 3522 V34241 3522 V34243 3522 V34245 3522 V34246 3522 V34247 3522 V34248 3522 V34249 3522 V34250 3522 V34251 3522 V34252 3522 V34253 3522 V34254 3522 V34255 3522 V34256 3522 V34257 3522 V34258 3522 V34259 3522 V34260 3522 V34261 3522 V34262 3522 V34263 3522 V34264 3522 V34265 3522 V34266 3522 V34267 3522 V34268 3522 V34269 3522 V34270 3522 V34271 3522 V34272 3522 V34273 3522 V34274 3522 V34275 3522 V34276 3522 V34277 3522 V34278 3522 V34279 3522 V34280 3522 V34281 3522 V34282 3522 V34283 3522 V34284 3522 V34285 3522 V34286 3522 V34287 3522 V34288 3522 V34289 3522 V34290 3522 V34291 3522 V34292 3522 V34293 3522 V34294 3522 V34295 3522 V34296 3522 V34297 3522 V34298 3522 V34299 3522 V34300 3522 V34301 3522 V34302 3522 V34303 3522 V34304 3522 V34305 3522 V34306 3522 V34307 3522 V34308 3522 V34309 3522 V34310 3522 V34311 3522 V34312 3522 V34313 3522 V34314 3522 V34315 3522 V34316 3522 V34317 3522 V34318 3522 V34319 3522 V34320 3522 V34321 3522 V34322 3522 V34323 3522 V34324 3522 V34325 3522 V34326 3522 V34327 3522 V34328 3522 V34329 3522 V34330 3522 V34331 3522 V34332 3522 V34333 3522 V34334 3522 V34335 3522 V34336 3522 V34337 3522 V34338 3522 V34339 3522 V34340 3522 V34341 3578 V27838 3578 V27845 3578 V27855 3578 V27872 3578 V27873 3578 V27877 3578 V27879 3578 V27884 3578 V27885 3578 V27889 3578 V27890 3578 V27893 3578 V27894 3636 V33353 3636 V33354 3636 V33355 3636 V33356 3636 V33357 3636 V33358 3636 V33359 3636 V33360 3636 V33361 3636 V33362 3636 V33363 3636 V33364 3636 V33365 3636 V33366 3636 V33367 3636 V33368 3636 V33369 3636 V33370 3636 V33371 3636 V33372 3636 V33373 3636 V33374 3636 V33375 3636 V33376 3636 V33377 3636 V33378 3636 V33379 3636 V33380 3636 V33381 3636 V33382 3636 V33383 3636 V33384 3636 V33385 3636 V33386 3636 V33387 3636 V33388 3636 V33390 3636 V33393 3636 V33394 3636 V33396 3636 V33399 3636 V33400 3636 V33401 3636 V33403 3636 V33406 3636 V33408 3636 V33409 3636 V33410 3636 V33413 3636 V33415 3522 V34135 3522 V34136 3522 V34137 3522 V34138 3522 V34139 3522 V34140 3522 V34141 3522 V34142 3522 V34143 3522 V34144 3522 V34145 3522 V34146 3522 V34147 3522 V34148 3522 V34149 3522 V34150 3522 V34151 3522 V34152 3522 V34153 3522 V34154 3522 V34155 3522 V34156 3522 V34157 3522 V34158 3522 V34159 3522 V34160 3522 V34161 3522 V34162 3522 V34163 3522 V34164 3522 V34165 3522 V34166 3522 V34167 3522 V34168 3522 V34169 3522 V34170 3522 V34171 3522 V34172 3522 V34173 3522 V34174 3522 V34175 3522 V34176 3522 V34177 3522 V34178 3522 V34179 3522 V34180 3522 V34181 3522 V34182 3522 V34183 3522 V34184 3522 V34185 3522 V34186 3522 V34187 3522 V34188 3522 V34189 3522 V34190 3522 V34191 3522 V34192 3522 V34193 3522 V34194 3522 V34195 3522 V34196 3522 V34197 3522 V34198 3522 V34199 3522 V34200 3522 V34201 3522 V34202 3522 V34203 3522 V34204 3522 V34205 3522 V34206 3522 V34207 3522 V34208 3522 V34209 3522 V34210 3522 V34211 3522 V34212 3522 V34003 3522 V34004 3522 V34005 3522 V34006 3522 V34007 3522 V34008 3522 V34009 3522 V34010 3522 V34011 3522 V34012 3522 V34013 3522 V34014 3522 V34015 3522 V34016 3522 V34017 3522 V34018 3522 V34019 3522 V34020 3522 V34021 3522 V34022 3522 V34023 3522 V34024 3522 V34025 3522 V34026 3522 V34027 3522 V34028 3522 V34029 3522 V34030 3522 V34031 3522 V34032 3522 V34033 3522 V34034 3522 V34035 3522 V34036 3522 V34037 3522 V34038 3522 V34039 3522 V34040 3522 V34041 3522 V34042 3522 V34043 3522 V34044 3522 V34045 3522 V34046 3522 V34047 3522 V34048 3522 V34049 3522 V34050 3522 V34051 3522 V34052 3522 V34053 3522 V34054 3522 V34055 3522 V34056 3522 V34057 3522 V34058 3522 V34059 3522 V34060 3522 V34061 3522 V34062 3522 V34063 3522 V34064 3522 V34065 3522 V34066 3522 V34067 3522 V34068 3522 V34069 3522 V34071 3522 V34072 3522 V34073 3522 V34074 3522 V34075 3522 V34076 3522 V34077 3522 V34078 3522 V34079 3522 V34080 3522 V34082 3522 V34083 3522 V34084 3522 V34085 3522 V34087 3522 V34088 3522 V34089 3522 V34090 3522 V34091 3522 V34092 3522 V34093 3522 V34094 3522 V34095 3522 V34096 3522 V34097 3522 V34098 3522 V34099 3522 V34100 3522 V34101 3522 V34102 3522 V34103 3522 V34104 3522 V34105 3522 V34106 3522 V34107 3522 V34108 3522 V34109 3522 V34110 3522 V34111 3522 V34112 3522 V34113 3522 V34114 3522 V34115 3522 V34116 3522 V34117 3522 V34118 3522 V34119 3522 V34120 3522 V34121 3522 V34122 3522 V34123 3522 V34124 3522 V34125 3522 V34126 3522 V34128 3522 V34129 3522 V34130 3522 V34131 3522 V34132 3522 V34133 3522 V34134 3522 V33951 3522 V33952 3522 V33953 3522 V33954 3522 V33955 3522 V33956 3522 V33957 3522 V33958 3522 V33959 3522 V33960 3522 V33961 3522 V33962 3522 V33963 3522 V33964 3522 V33965 3522 V33966 3522 V33967 3522 V33968 3522 V33969 3522 V33970 3522 V33971 3522 V33972 3522 V33973 3522 V33974 3522 V33975 3522 V33976 3522 V33977 3522 V33978 3522 V33979 3522 V33980 3522 V33981 3522 V33982 3522 V33983 3522 V33984 3522 V33985 3522 V33986 3522 V33987 3522 V33988 3522 V33989 3522 V33990 3522 V33991 3522 V33992 3522 V33993 3522 V33994 3522 V33995 3522 V33996 3522 V33997 3522 V33998 3522 V33999 3522 V34000 3522 V34001 3522 V34002 3578 V27870 3578 V27723 3578 V27724 3578 V27741 3577 V28695 3577 V28698 3578 V27725 3578 V27728 3578 V27729 3578 V27847 3578 V27848 3636 V33391 3636 V33392 3636 V33395 3636 V33397 3636 V33398 3636 V33402 3636 V33404 3636 V33405 3636 V33407 3636 V33411 3636 V33412 3636 V33414 3636 V33416 3636 V33417 3636 V33418 3636 V33419 3636 V33420 3636 V33421 3636 V33423 3636 V33424 3636 V33427 3636 V33430 3636 V33433 3636 V33434 3636 V33435 3636 V33436 3636 V33437 3636 V33438 3636 V33439 3636 V33440 3636 V33441 3636 V33442 3577 V28677 3577 V28683 3577 V28685 3577 V28686 3577 V28687 3577 V28690 3577 V28691 3577 V28692 3577 V28693 3577 V28696 3577 V28681 3577 V28684 3577 V31062 3577 V31067 3578 V27726 3578 V27727 3577 V31969 3577 V31970 3577 V31973 3577 V31974 3578 V27730 3578 V27731 3578 V27732 3578 V27733 3578 V27734 3578 V27735 3578 V27743 3315 V29017 3315 V29018 3315 V29020 3315 V29021 3315 V29022 3315 V29023 3315 V29024 3315 V29025 3315 V29027 3315 V29028 3315 V29029 3315 V29033 3315 V29035 3315 V29036 3315 V29038 3315 V29039 3315 V29040 3315 V29041 3315 V29042 3315 V29043 3315 V29046 3315 V29048 3315 V29050 3315 V29059 3315 V29065 3577 V31975 3578 V27738 3522 V33894 3522 V33895 3522 V33896 3522 V33897 3522 V33899 3522 V33902 3522 V33903 3522 V33904 3522 V33905 3522 V33906 3522 V33907 3522 V33908 3522 V33909 3522 V33910 3522 V33911 3522 V33912 3522 V33916 3522 V33917 3522 V33918 3522 V33919 3522 V33921 3522 V33922 3522 V33923 3522 V33924 3522 V33943 3522 V33946 3522 V33813 3522 V33814 3522 V33815 3522 V33816 3522 V33817 3522 V33818 3522 V33819 3522 V33820 3522 V33821 3522 V33822 3522 V33823 3522 V33824 3522 V33825 3522 V33826 3522 V33827 3522 V33828 3522 V33829 3522 V33830 3522 V33831 3522 V33832 3522 V33833 3522 V33834 3522 V33835 3522 V33836 3522 V33837 3522 V33838 3522 V33839 3522 V33840 3522 V33841 3522 V33842 3522 V33843 3522 V33844 3522 V33845 3522 V33846 3522 V33847 3522 V33848 3522 V33849 3522 V33850 3522 V33851 3522 V33852 3522 V33853 3522 V33854 3522 V33855 3522 V33856 3522 V33857 3522 V33858 3522 V33859 3522 V33860 3522 V33861 3522 V33862 3522 V33863 3522 V33864 3522 V33865 3522 V33866 3522 V33867 3522 V33868 3522 V33869 3522 V33770 3522 V33771 3522 V33772 3522 V33773 3522 V33774 3522 V33775 3522 V33776 3522 V33777 3522 V33778 3522 V33779 3522 V33780 3522 V33781 3522 V33782 3522 V33783 3522 V33784 3522 V33785 3522 V33786 3522 V33787 3522 V33788 3522 V33789 3522 V33790 3522 V33792 3522 V33793 3522 V33794 3522 V33795 3522 V33796 3522 V33797 3522 V33798 3522 V33799 3522 V33800 3522 V33801 3522 V33802 3522 V33805 3522 V33806 3522 V33808 3522 V33809 3522 V33810 3522 V33811 3522 V33812 3522 V33871 3522 V33872 3522 V33873 3522 V33874 3522 V33875 3522 V33876 3522 V33877 3522 V33878 3522 V33879 3522 V33880 3522 V33881 3522 V33882 3522 V33883 3522 V33884 3522 V33885 3522 V33886 3522 V33887 3522 V33888 3522 V33889 3522 V33890 3522 V33891 3522 V33892 3522 V33893 3522 V33700 3522 V33701 3522 V33702 3522 V33703 3522 V33704 3522 V33705 3522 V33706 3522 V33707 3522 V33708 3522 V33709 3522 V33710 3522 V33711 3522 V33712 3522 V33713 3522 V33714 3522 V33715 3522 V33716 3522 V33717 3522 V33718 3522 V33719 3522 V33720 3522 V33721 3522 V33722 3522 V33723 3522 V33724 3522 V33725 3522 V33726 3522 V33727 3522 V33728 3522 V33729 3522 V33913 3522 V33914 3522 V33915 3522 V33920 3522 V33925 3522 V33926 3522 V33927 3522 V33928 3522 V33929 3522 V33930 3522 V33931 3522 V33932 3522 V33933 3522 V33934 3522 V33935 3522 V33936 3522 V33937 3522 V33938 3522 V33939 3522 V33941 3522 V33942 3522 V33944 3522 V33945 3522 V33947 3522 V33948 3522 V33949 3522 V33950 3522 V33730 3522 V33731 3522 V33732 3522 V33733 3522 V33734 3522 V33735 3522 V33736 3522 V33737 3522 V33738 3522 V33739 3522 V33740 3522 V33741 3522 V33742 3522 V33743 3522 V33744 3522 V33745 3522 V33746 3522 V33747 3522 V33748 3522 V33749 3522 V33750 3522 V33751 3522 V33752 3522 V33753 3522 V33754 3522 V33755 3522 V33756 3522 V33757 3522 V33758 3522 V33759 3522 V33760 3522 V33761 3522 V33762 3522 V33763 3522 V33764 3522 V33765 3522 V33766 3522 V33767 3522 V33768 3522 V33769 3522 V33791 3522 V33803 3522 V33804 3522 V33541 3522 V33543 3522 V33544 3522 V33545 3522 V33546 3522 V33547 3522 V33548 3522 V33549 3522 V33550 3522 V33551 3522 V33553 3522 V33555 3522 V33557 3522 V33561 3522 V33564 3522 V33568 3522 V33569 3522 V33570 3522 V33571 3522 V33572 3522 V33573 3522 V33574 3522 V33575 3522 V33576 3522 V33577 3522 V33578 3522 V33579 3522 V33580 3522 V33581 3522 V33582 3522 V33583 3522 V33584 3522 V33585 3522 V33586 3522 V33587 3522 V33588 3522 V33589 3522 V33591 3522 V33592 3522 V33593 3522 V33594 3522 V33595 3522 V33596 3522 V33597 3522 V33598 3522 V33599 3522 V33600 3522 V33601 3522 V33602 3522 V33603 3522 V33604 3522 V33605 3522 V33606 3522 V33607 3522 V33608 3522 V33609 3522 V33610 3522 V33611 3522 V33612 3522 V33613 3522 V33614 3522 V33615 3522 V33616 3522 V33617 3522 V33618 3522 V33619 3522 V33620 3522 V33621 3522 V33622 3522 V33623 3522 V33624 3522 V33625 3522 V33626 3522 V33627 3522 V33628 3522 V33629 3522 V33630 3522 V33631 3522 V33632 3522 V33634 3522 V33635 3522 V33636 3522 V33637 3522 V33638 3522 V33640 3522 V33641 3522 V33642 3522 V33643 3522 V33644 3522 V33645 3522 V33646 3522 V33648 3522 V33649 3522 V33652 3522 V33653 3522 V33655 3522 V33656 3522 V33657 3522 V33658 3522 V33659 3522 V33660 3522 V33661 3522 V33662 3522 V33663 3522 V33664 3522 V33665 3522 V33666 3522 V33667 3522 V33668 3522 V33669 3522 V33670 3522 V33671 3522 V33672 3522 V33673 3522 V33674 3522 V33675 3522 V33676 3522 V33677 3522 V33678 3522 V33679 3522 V33680 3522 V33681 3522 V33682 3522 V33683 3522 V33684 3522 V33685 3522 V33686 3522 V33688 3522 V33689 3522 V33690 3522 V33691 3522 V33692 3522 V33693 3522 V33694 3522 V33695 3522 V33696 3522 V33697 3522 V33698 3522 V33699 3315 V28160 3315 V28161 3315 V28163 3315 V28167 3315 V28174 3315 V28175 3315 V29019 3315 V29026 3315 V29030 3315 V29031 3315 V29032 3315 V29034 3315 V29037 3315 V29044 3315 V29045 3315 V29047 3315 V29049 3315 V29051 3315 V29052 3315 V29053 3315 V29054 3315 V29055 3315 V29056 3315 V29057 3315 V29058 3315 V29060 3315 V29061 3315 V29063 3315 V29064 3315 V29066 3449 V34902 3449 V34903 3449 V34904 3449 V34906 3449 V34907 3449 V34909 3449 V34911 3449 V34912 3449 V34913 3449 V34914 3449 V34920 3449 V34921 3449 V34922 3449 V34925 3449 V34929 3449 V34931 3449 V34933 3449 V34934 3449 V34935 3449 V34936 3449 V34937 3449 V34939 3449 V34940 3449 V34941 3449 V34942 3449 V34943 3449 V34944 3449 V34945 3449 V34946 3449 V34947 3449 V34948 3449 V34949 3449 V34951 3449 V34952 3449 V34953 3449 V34954 3449 V34955 3449 V34956 3449 V34957 3449 V34958 3449 V34959 3449 V34960 3449 V34961 3449 V34962 3449 V34963 3449 V34964 3449 V34965 3449 V34966 3449 V34967 3449 V34968 3449 V34969 3449 V34970 3449 V34971 3449 V34972 3449 V34973 3449 V34974 3449 V34975 3449 V34976 3449 V34977 3449 V34978 3449 V34979 3449 V34980 3449 V34981 3449 V34982 3449 V34983 3449 V34984 3449 V34985 3449 V34986 3449 V34987 3449 V34988 3449 V34989 3449 V34990 3449 V34991 3449 V34992 3449 V34993 3449 V34994 3449 V34995 3449 V34996 3449 V34997 3449 V34998 3449 V34999 3449 V35000 3449 V35001 3449 V35002 3449 V35003 3449 V35004 3449 V35007 3449 V35008 3449 V35009 3449 V35010 3449 V35011 3449 V35012 3449 V35013 3449 V35014 3449 V35015 3449 V35016 3449 V35017 3449 V35018 3449 V35019 3449 V35020 3449 V35021 3449 V35022 3449 V35023 3449 V35024 3449 V35025 3449 V35026 3449 V35027 3449 V35028 3449 V35029 3449 V35030 3449 V35031 3449 V35032 3449 V35033 3449 V35034 3449 V35035 3449 V35036 3449 V35037 3449 V35038 3449 V35039 3449 V35040 3449 V35041 3449 V35042 3449 V35043 3449 V35044 3449 V35046 3449 V35047 3449 V35048 3449 V35049 3449 V35050 3449 V35051 3449 V35052 3449 V35053 3449 V35054 3449 V35055 3449 V35056 3449 V35057 3449 V35060 3449 V35066 3449 V35079 3449 V35113 3449 V35114 3449 V35115 3449 V35116 3449 V35117 3449 V35118 3449 V35224 3449 V35239 3449 V35243 3449 V35244 3449 V35251 3449 V35058 3449 V35059 3449 V35061 3449 V35062 3449 V35063 3449 V35064 3449 V35065 3449 V35067 3449 V35068 3449 V35069 3449 V35070 3449 V35071 3449 V35072 3449 V35073 3449 V35074 3449 V35075 3449 V35077 3449 V35078 3449 V35080 3449 V35081 3449 V35082 3449 V35083 3449 V35084 3449 V35085 3449 V35086 3449 V35087 3449 V35088 3449 V35089 3449 V35090 3449 V35091 3449 V35092 3449 V35093 3449 V35094 3449 V35095 3449 V35096 3449 V35097 3449 V35098 3449 V35099 3449 V35101 3449 V35102 3449 V35103 3449 V35104 3449 V35105 3449 V35106 3449 V35107 3449 V35108 3449 V35109 3449 V35111 3449 V35112 3449 V35120 3449 V35121 3449 V35122 3449 V35123 3449 V35124 3449 V35125 3449 V35126 3449 V35127 3449 V35128 3449 V35129 3449 V35130 3449 V35131 3449 V35132 3449 V35133 3449 V35134 3449 V35135 3449 V35136 3449 V35137 3449 V35138 3449 V35139 3449 V35140 3449 V35141 3449 V35142 3449 V35143 3449 V35144 3449 V35146 3449 V35147 3449 V35148 3449 V35149 3449 V35150 3449 V35151 3449 V35152 3449 V35153 3449 V35154 3449 V35155 3449 V35156 3449 V35157 3449 V35158 3449 V35159 3449 V35160 3449 V35161 3449 V35162 3449 V35163 3449 V35164 3449 V35165 3449 V35166 3449 V35167 3449 V35168 3449 V35169 3449 V35170 3449 V35171 3449 V35172 3449 V35173 3449 V35174 3449 V35175 3449 V35176 3449 V35177 3449 V35178 3449 V35179 3449 V35180 3449 V35181 3449 V35182 3449 V35183 3449 V35184 3449 V35185 3449 V35186 3449 V35187 3449 V35188 3449 V35189 3449 V35190 3449 V35191 3449 V35192 3449 V35193 3449 V35194 3449 V35195 3449 V35196 3449 V35197 3449 V35198 3449 V35199 3449 V35200 3449 V35201 3449 V35202 3449 V35203 3449 V35204 3449 V35205 3449 V35206 3449 V35207 3449 V35208 3449 V35209 3449 V35210 3449 V35211 3449 V35212 3449 V35213 3449 V35214 3449 V35215 3449 V35216 3449 V35217 3449 V35218 3449 V35219 3449 V35220 3449 V35221 3449 V35222 3449 V35223 3449 V35225 3449 V35226 3449 V35227 3449 V35228 3449 V35229 3449 V35230 3449 V35231 3449 V35232 3449 V35233 3449 V35234 3449 V35235 3449 V35236 3449 V35237 3449 V35238 3449 V35240 3449 V35241 3449 V35242 3449 V35245 3449 V35247 3449 V35248 3449 V35249 3449 V35250 3449 V35252 3449 V35253 3449 V35254 3449 V35255 3449 V35256 3449 V35257 3449 V35258 3449 V35259 3449 V35260 3449 V35262 3449 V35263 3449 V35264 3449 V35265 3449 V35266 3449 V35267 3449 V35268 3449 V35269 3449 V35270 3449 V35271 3449 V35272 3449 V35273 3449 V35274 3449 V35275 3449 V35276 3449 V35277 3449 V35278 3449 V35279 3449 V35280 3449 V35281 3449 V35282 3449 V35284 3449 V35285 3449 V35286 3449 V35287 3449 V35288 3449 V35289 3449 V35290 3449 V35291 3449 V35292 3449 V35293 3449 V35294 3449 V35295 3449 V35296 3449 V35297 3449 V35298 3449 V35299 3449 V35300 3449 V35302 3449 V35304 3449 V35305 3449 V35261 3449 V35283 3449 V35301 3449 V35303 3449 V35306 3449 V35307 3449 V35308 3449 V35309 3449 V35310 3449 V35311 3449 V35312 3449 V35313 3449 V35314 3449 V35315 3449 V35316 3449 V35317 3449 V35318 3449 V35319 3449 V35320 3449 V35321 3449 V35322 3449 V35323 3449 V35324 3449 V35325 3449 V35326 3449 V33000 3577 V31814 3577 V31827 3577 V31844 3577 V31846 3577 V31851 3577 V31853 3577 V31861 3577 V31972 3449 V34900 3449 V34905 3449 V34908 3449 V34910 3449 V34915 3449 V34916 3449 V34917 3449 V34918 3449 V34923 3449 V34924 3449 V34926 3449 V34928 3449 V34930 3449 V34932 3449 V34938 3449 V33007 3449 V33010 3449 V32485 3449 V32489 3449 V33011 3577 V31813 3577 V31818 3577 V31819 3577 V31849 3577 V31856 3578 V26235 3578 V26238 3578 V26240 3578 V26241 3578 V26242 3578 V26248 3578 V26292 3578 V26295 3578 V26297 3578 V26298 3635 V32483 3635 V32484 3635 V32487 3635 V32491 3635 V32492 3635 V32496 3635 V33000 3635 V33005 3635 V33006 3635 V33466 3635 V33468 3635 V33471 3635 V33473 3635 V33475 3635 V33477 3635 V33478 3635 V33481 3635 V33483 3635 V33484 3635 V33488 3578 V26227 3578 V26230 3578 V26231 3578 V26232 3578 V26234 3578 V26237 3578 V26239 3578 V26246 3578 V26228 3578 V26233 3578 V26236 3578 V27772 3578 V27773 3636 V35615 3636 V35616 3636 V35617 3636 V35618 3636 V35620 3636 V35622 3636 V35623 3636 V35624 3636 V35625 3636 V35626 3636 V35628 3636 V35629 3636 V35630 3636 V35631 3636 V35632 3636 V35634 3636 V35635 3636 V35638 3636 V35641 3636 V35642 3636 V35643 3636 V35645 3636 V35646 3636 V35647 3636 V35648 3636 V35649 3636 V35650 3636 V35651 3636 V35652 3636 V35653 3636 V35654 3636 V35655 3636 V35656 3636 V35657 3636 V35658 3636 V35660 3636 V35661 3636 V35662 3636 V35663 3636 V35666 3636 V35667 3636 V35672 3636 V35674 3636 V35675 3636 V35676 3636 V35677 3636 V35679 3636 V35680 3636 V35684 3636 V35685 3636 V35686 3636 V35691 3636 V35693 3636 V35694 3636 V35696 3636 V35697 3636 V35698 3636 V35701 3636 V35702 3636 V35708 3636 V35716 3636 V35717 3636 V35726 3636 V35742 3636 V35743 3636 V35748 3636 V35750 3636 V35752 3636 V35756 3636 V35762 3636 V35765 3636 V35768 3636 V35769 3636 V35771 3636 V35773 3636 V36049 3636 V36054 3636 V36057 3636 V36058 3636 V36060 3636 V36061 3636 V36062 3636 V36063 3636 V36064 3636 V36066 3636 V36068 3636 V36070 3636 V36071 3578 V27744 3578 V27745 3578 V27746 3578 V27747 3578 V27748 3578 V27749 3578 V27750 3578 V27751 3578 V27752 3578 V27753 3578 V27754 3578 V27755 3578 V27756 3578 V27757 3578 V27758 3578 V27759 3578 V27760 3578 V27761 3578 V27762 3578 V27763 3578 V27764 3578 V27765 3578 V27766 3578 V27767 3578 V27768 3578 V27769 3578 V27770 3578 V27771 3578 V27774 3578 V27775 3578 V27776 3578 V27777 3412 V32427 3412 V32429 3412 V32443 3412 V32445 3412 V32449 3412 V32450 3412 V32457 3412 V32459 3412 V32467 3412 V32468 3449 V35327 3449 V35328 3449 V35329 3449 V35330 3449 V35331 3449 V35332 3449 V35333 3449 V35334 3449 V35335 3449 V35336 3449 V35337 3449 V35338 3449 V35339 3449 V35340 3449 V35341 3449 V35342 3449 V35343 3449 V35344 3449 V35345 3449 V35346 3449 V35347 3449 V35348 3449 V35349 3449 V35350 3449 V35351 3449 V35352 3449 V35353 3449 V35354 3449 V35355 3449 V35356 3449 V35357 3449 V35358 3449 V35362 3449 V35363 3449 V35364 3449 V35365 3449 V35366 3449 V35367 3449 V35368 3449 V35370 3449 V35371 3449 V35372 3449 V35374 3449 V35375 3449 V35376 3449 V35377 3449 V35378 3449 V35379 3449 V35380 3449 V35381 3449 V35382 3449 V35383 3449 V35384 3449 V35385 3449 V35386 3449 V35387 3449 V35389 3449 V35396 3449 V35398 3449 V35399 3449 V35400 3449 V35401 3449 V35402 3449 V35403 3449 V35404 3449 V35405 3449 V35406 3449 V35407 3449 V35408 3449 V35409 3449 V35410 3449 V35411 3449 V35412 3449 V35413 3449 V35414 3449 V35415 3449 V35416 3449 V35417 3449 V35418 3449 V35419 3449 V35420 3449 V35421 3449 V35422 3449 V35423 3449 V35424 3449 V35425 3449 V35426 3449 V35427 3449 V35428 3449 V35429 3449 V35430 3449 V35431 3449 V35432 3449 V35433 3449 V35434 3449 V35435 3449 V35436 3449 V35437 3449 V35438 3449 V35440 3449 V35441 3449 V35442 3449 V35443 3449 V35444 3449 V35445 3449 V35446 3449 V35447 3449 V35448 3449 V35449 3449 V35450 3449 V35451 3449 V35452 3449 V35453 3449 V35454 3449 V35455 3449 V35456 3449 V35457 3449 V35458 3449 V35459 3449 V35460 3449 V35461 3449 V35462 3449 V35463 3449 V35464 3449 V35465 3449 V35466 3449 V35467 3449 V35470 3449 V35471 3449 V35476 3449 V35477 3449 V35478 3449 V35479 3449 V35480 3449 V35481 3449 V35482 3449 V35483 3449 V35484 3449 V35485 3449 V35486 3449 V35487 3449 V35488 3449 V35489 3449 V35490 3449 V35491 3449 V35492 3449 V35493 3449 V35494 3449 V35495 3449 V35496 3449 V35497 3449 V35498 3449 V35556 3449 V35557 3449 V35558 3449 V35559 3449 V35562 3449 V35566 3449 V35571 3449 V35572 3449 V35573 3449 V35576 3449 V35577 3449 V35579 3449 V35580 3449 V35581 3449 V35582 3449 V35583 3449 V35584 3449 V35585 3449 V35586 3449 V35587 3449 V35588 3449 V35589 3449 V35590 3449 V35591 3449 V35592 3449 V35593 3449 V35594 3449 V35595 3449 V35596 3449 V35597 3449 V35598 3449 V35600 3449 V35601 3449 V35602 3449 V35603 3449 V35604 3449 V35605 3449 V35606 3449 V35607 3449 V35608 3449 V35609 3449 V35610 3449 V35611 3449 V35612 3449 V35613 3449 V35614 3523 V33063 3523 V33072 3523 V33074 3523 V33077 3523 V33065 3523 V33067 3523 V33070 3523 V33073 3523 V33075 3523 V33204 3523 V33206 3523 V33209 3523 V33212 3523 V33213 3523 V33340 3523 V33330 3523 V33332 3523 V33334 3523 V33335 3523 V33336 3523 V33337 3523 V33333 3523 V36177 3523 V36178 3523 V36179 3523 V36180 3523 V36182 3523 V36183 3523 V36184 3523 V36185 3523 V36186 3523 V36187 3523 V36188 3523 V36189 3523 V36190 3523 V36191 3523 V36192 3523 V36193 3523 V36194 3523 V36195 3523 V36196 3523 V36198 3523 V36199 3523 V36200 3523 V36202 3523 V36203 3523 V36205 3523 V36206 3523 V36207 3523 V36208 3523 V36209 3523 V36211 3523 V36212 3523 V36213 3523 V36215 3523 V36216 3523 V36217 3523 V36218 3523 V36219 3523 V36220 3523 V36221 3523 V36223 3523 V36226 3523 V36228 3523 V36229 3523 V36230 3523 V36231 3523 V36232 3523 V36279 3523 V36280 3523 V36281 3523 V36282 3523 V36283 3523 V36284 3523 V36285 3523 V36286 3523 V36287 3523 V36288 3523 V36289 3523 V36292 3523 V36293 3523 V36294 3523 V36295 3523 V36296 3523 V36297 3523 V36298 3523 V36299 3523 V36300 3523 V36301 3523 V36302 3523 V36303 3523 V36305 3523 V36307 3523 V36309 3523 V36310 3523 V36311 3523 V36312 3523 V36313 3523 V36315 3523 V36317 3523 V36318 3523 V36319 3523 V36320 3523 V36321 3523 V36322 3523 V36323 3523 V36324 3523 V36325 3523 V36328 3523 V36329 3523 V36348 3523 V36349 3523 V36350 3523 V36351 3523 V36352 3523 V36353 3523 V36354 3523 V36355 3523 V36356 3523 V36357 3523 V36358 3523 V36359 3523 V36360 3523 V36361 3523 V36362 3523 V36363 3523 V36364 3523 V36365 3523 V36366 3523 V36367 3523 V36369 3523 V36370 3523 V36371 3523 V36372 3523 V36373 3523 V36374 3523 V36375 3523 V36376 3523 V36077 3523 V36078 3523 V36278 3523 V36291 3523 V36306 3523 V36308 3523 V36314 3523 V36316 3523 V36326 3523 V36327 3523 V36330 3523 V36331 3523 V36332 3523 V36333 3523 V36334 3523 V36335 3523 V36336 3523 V36337 3523 V36338 3523 V36339 3523 V36340 3523 V36341 3523 V36342 3523 V36343 3523 V36344 3523 V36345 3523 V36346 3523 V36347 3523 V36233 3523 V36234 3523 V36235 3523 V36236 3523 V36237 3523 V36238 3523 V36239 3523 V36240 3523 V36241 3523 V36242 3523 V36243 3523 V36244 3523 V36245 3523 V36246 3523 V36247 3523 V36249 3523 V36251 3523 V36252 3523 V36253 3523 V36254 3523 V36255 3523 V36256 3523 V36259 3523 V36260 3523 V36261 3523 V36263 3523 V36264 3523 V36265 3523 V36267 3523 V36268 3523 V36270 3523 V36272 3523 V36275 3523 V36276 3523 V36377 3523 V36383 3523 V36384 3523 V36385 3523 V36386 3523 V36387 3523 V36389 3523 V36390 3523 V36391 3523 V36392 3523 V36441 3523 V36443 3523 V36444 3523 V36445 3523 V36446 3523 V36447 3523 V36451 3523 V36452 3523 V36453 3523 V36455 3523 V36456 3523 V36457 3523 V36458 3523 V36459 3523 V36461 3523 V36463 3523 V36464 3523 V36465 3523 V36466 3523 V36467 3523 V36469 3523 V36470 3523 V36577 3523 V36578 3523 V36579 3523 V36580 3523 V36583 3523 V36585 3523 V36586 3523 V36587 3523 V36248 3523 V36250 3523 V36258 3523 V36262 3523 V36269 3523 V36271 3523 V36273 3523 V36274 3523 V36378 3523 V36379 3523 V36380 3523 V36381 3523 V36382 3523 V36393 3523 V36394 3523 V36398 3523 V36399 3523 V36400 3523 V36401 3523 V36402 3523 V36403 3523 V36405 3523 V36407 3523 V36408 3523 V36409 3523 V36410 3523 V36411 3523 V36412 3523 V36413 3523 V36414 3523 V36415 3523 V36416 3523 V36417 3523 V36418 3523 V36419 3523 V36420 3523 V36422 3523 V36423 3523 V36424 3523 V36425 3523 V36427 3523 V36428 3523 V36429 3523 V36430 3523 V36432 3523 V36433 3523 V36434 3523 V36435 3523 V36436 3523 V36437 3523 V36438 3523 V36439 3523 V36440 3523 V36448 3523 V36449 3523 V36462 3523 V36468 3523 V36582 3523 V36584 3523 V36115 3523 V36117 3523 V36118 3523 V36119 3523 V36120 3523 V36122 3523 V36123 3523 V36124 3523 V36125 3523 V36126 3523 V36127 3523 V36128 3523 V36129 3523 V36131 3523 V36132 3523 V36134 3523 V36135 3523 V36136 3523 V36137 3523 V36138 3523 V36139 3523 V36140 3523 V36141 3523 V36142 3523 V36143 3523 V36144 3523 V36145 3523 V36146 3523 V36147 3523 V36148 3523 V36149 3523 V36150 3523 V36151 3523 V36152 3523 V36153 3523 V36154 3523 V36156 3523 V36157 3523 V36158 3523 V36159 3523 V36160 3523 V36161 3523 V36162 3523 V36163 3523 V36164 3523 V36165 3523 V36166 3523 V36167 3523 V36169 3523 V36170 3523 V36173 3523 V36174 3523 V36175 3523 V36176 3523 V36471 3523 V36472 3523 V36473 3523 V36474 3523 V36475 3523 V36476 3577 V31876 3577 V31821 3577 V31822 3577 V31825 3577 V31828 3577 V31829 3577 V31845 3577 V31848 3577 V31854 3577 V31855 3577 V31857 3577 V31859 3577 V31862 3577 V31863 3577 V31864 3577 V31865 3577 V31866 3577 V31867 3577 V31868 3577 V31869 3577 V31870 3577 V31872 3577 V31873 3577 V31874 3577 V31875 3577 V31826 3577 V31852 3577 V31836 3577 V31838 3635 V32486 3635 V32488 3635 V32490 3635 V32493 3635 V32494 3635 V32495 3635 V32497 3635 V33467 3635 V33469 3635 V33470 3635 V33472 3635 V33474 3635 V33476 3635 V33480 3635 V33482 3635 V33485 3635 V33486 3635 V33487 3635 V33489 3635 V33490 3635 V33491 3635 V33492 3635 V33493 3635 V33494 3635 V33496 3635 V33497 3635 V33498 3635 V35514 3635 V35515 3635 V35517 3635 V35518 3635 V35519 3635 V35520 3635 V35522 3635 V35523 3635 V35524 3635 V35525 3635 V35526 3635 V35528 3635 V35529 3635 V35530 3635 V35531 3635 V35532 3635 V35533 3635 V35535 3635 V35536 3635 V35537 3635 V35538 3635 V35539 3635 V35540 3635 V35541 3635 V35542 3635 V35543 3635 V35544 3635 V35545 3635 V35546 3635 V35547 3635 V35548 3635 V35549 3635 V35550 3635 V35551 3635 V35552 3635 V35553 3635 V35554 3635 V35555 3636 V35619 3636 V35621 3636 V35627 3636 V35633 3636 V35636 3636 V35637 3636 V35639 3636 V35640 3636 V35659 3636 V35665 3636 V35668 3636 V35669 3636 V35670 3636 V35673 3636 V35678 3636 V35681 3636 V35682 3636 V35683 3636 V35687 3636 V35688 3636 V35689 3636 V35690 3636 V35695 3636 V35699 3636 V35700 3636 V35703 3636 V35705 3636 V35706 3636 V35709 3636 V35710 3636 V35711 3636 V35712 3636 V35713 3636 V35714 3636 V35715 3636 V35718 3636 V35719 3636 V35720 3636 V35721 3636 V35722 3636 V35723 3636 V35724 3636 V35725 3636 V35727 3636 V35728 3636 V35729 3636 V35730 3636 V35731 3636 V35732 3636 V35733 3636 V35734 3636 V35735 3636 V35736 3636 V35737 3636 V35738 3636 V35739 3636 V35740 3636 V35741 3636 V35744 3636 V35745 3636 V35746 3636 V35747 3636 V35749 3636 V35751 3636 V35753 3636 V35754 3636 V35755 3636 V35757 3636 V35758 3636 V35759 3636 V35760 3636 V35761 3636 V35763 3636 V35764 3636 V35767 3636 V36050 3636 V36051 3636 V36052 3636 V36053 3636 V36055 3636 V36056 3636 V36069 3636 V35766 3636 V35775 3636 V35777 3636 V35779 3636 V35951 3636 V35973 3636 V35974 3636 V35987 3636 V35990 3636 V36000 3636 V36010 3636 V36013 3636 V36014 3636 V36024 3636 V36027 3636 V36039 3636 V36043 3636 V36048 3523 V36204 3523 V36498 3523 V36500 3523 V36503 3523 V36504 3523 V36507 3523 V36509 3523 V36510 3523 V36511 3523 V36513 3523 V36515 3523 V36079 3523 V36080 3523 V36081 3523 V36083 3523 V36085 3523 V36086 3523 V36087 3523 V36088 3523 V36089 3523 V36090 3523 V36091 3523 V36092 3523 V36093 3523 V36094 3523 V36095 3523 V36096 3523 V36097 3523 V36098 3523 V36099 3523 V36100 3523 V36101 3523 V36102 3523 V36103 3523 V36105 3523 V36106 3523 V36107 3523 V36108 3523 V36109 3523 V36110 3523 V36111 3523 V36112 3523 V36113 3523 V36114 3523 V36116 3523 V36197 3523 V36210 3523 V36227 3523 V36257 3523 V36477 3523 V36478 3523 V36479 3523 V36480 3523 V36481 3523 V36484 3523 V36485 3523 V36486 3523 V36487 3523 V36488 3523 V36490 3523 V36491 3523 V36493 3523 V36494 3523 V36495 3523 V36496 3523 V36497 3523 V36501 3523 V36502 3523 V36505 3523 V36512 3523 V36514 3523 V36516 3523 V36517 3523 V36518 3523 V36519 3523 V35785 3523 V35786 3523 V35787 3523 V35788 3523 V35789 3523 V35790 3523 V36201 3523 V36431 3523 V36442 3523 V36460 3523 V36482 3523 V36499 3523 V36520 3523 V36521 3523 V36522 3523 V36523 3523 V36526 3523 V36527 3523 V36528 3523 V36529 3523 V36530 3523 V36531 3523 V36532 3523 V36534 3523 V36537 3523 V36538 3523 V36539 3523 V36540 3523 V36541 3523 V36543 3523 V36544 3523 V36545 3523 V36546 3523 V36548 3523 V36549 3523 V36550 3523 V36551 3523 V36552 3523 V36553 3523 V36554 3523 V36556 3523 V36557 3523 V36559 3523 V36560 3523 V36561 3523 V36562 3523 V36563 3523 V36564 3523 V36565 3523 V36566 3523 V36567 3523 V36568 3523 V36569 3523 V36570 3523 V36571 3523 V36572 3523 V36573 3523 V36688 3523 V36689 3523 V36690 3523 V36691 3523 V36692 3523 V36693 3523 V36694 3523 V36695 3523 V36696 3523 V36697 3523 V36698 3523 V36699 3523 V36700 3523 V36702 3523 V36703 3523 V36704 3523 V36705 3523 V36706 3523 V36707 3523 V36708 3523 V36709 3523 V36710 3523 V36711 3523 V36712 3523 V36713 3523 V36715 3523 V36716 3523 V36717 3523 V36718 3523 V36719 3523 V36720 3523 V36721 3523 V36722 3523 V36723 3523 V36724 3523 V36725 3523 V36726 3523 V36727 3523 V36728 3578 V26262 3578 V27736 3578 V27739 3578 V27740 3578 V27742 3635 V36664 3635 V36665 3635 V36666 3635 V36667 3635 V36668 3635 V36669 3635 V36670 3635 V36671 3635 V36672 3635 V36673 3635 V36675 3635 V36676 3635 V36677 3635 V36678 3635 V36679 3635 V36680 3635 V36681 3635 V36682 3635 V36683 3635 V36684 3635 V36685 3635 V36686 3635 V36687 3635 V36749 3635 V36750 3635 V36751 3635 V36752 3635 V36753 3635 V36754 3635 V36755 3635 V36756 3635 V36757 3635 V36758 3635 V36760 3635 V37093 3315 V28155 3315 V28162 3449 V35578 3577 V31842 3577 V32201 3577 V32169 3577 V32172 3577 V32185 3577 V32195 3577 V32202 3577 V32170 3577 V32182 3577 V32184 3577 V32189 3577 V32199 3577 V32200 3522 V33552 3522 V33554 3522 V33556 3522 V33558 3522 V33559 3522 V33560 3522 V33562 3522 V33563 3522 V33565 3522 V33566 3522 V33567 3522 V33633 3522 V33639 3522 V33647 3522 V33650 3522 V33651 3522 V33654 3522 V33687 3577 V32188 3577 V31770 3577 V31771 3577 V31772 3577 V31773 3577 V31774 3577 V31775 3577 V31777 3577 V31778 3577 V31779 3577 V31780 3577 V31781 3577 V31782 3577 V31783 3577 V31784 3577 V31785 3577 V31787 3577 V31788 3577 V31789 3577 V31790 3577 V31791 3577 V31792 3577 V31793 3577 V31794 3577 V31795 3577 V31796 3577 V31797 3577 V31798 3577 V31799 3577 V31800 3577 V31801 3577 V31802 3577 V31803 3577 V31804 3577 V31805 3577 V31806 3577 V31807 3577 V31808 3577 V31809 3577 V31810 3577 V31811 3577 V31812 3577 V31786 3577 V32197 3577 V31884 3577 V31885 3577 V31887 3577 V31888 3577 V31889 3577 V31890 3577 V31891 3577 V31892 3577 V31893 3577 V31894 3577 V31896 3577 V31897 3577 V31898 3577 V31900 3577 V31901 3577 V31902 3577 V31903 3577 V31904 3577 V31905 3577 V31906 3577 V31907 3577 V31908 3577 V31909 3577 V31911 3577 V31912 3577 V31913 3577 V31914 3577 V31915 3577 V31916 3577 V31917 3577 V31918 3577 V31919 3577 V31920 3577 V31921 3577 V31922 3577 V31923 3577 V31964 3577 V31966 3577 V31968 3577 V31895 3577 V31910 3577 V31965 3577 V32180 3578 V27713 3578 V27779 3578 V27785 3578 V27787 3578 V27789 3578 V27790 3578 V27791 3578 V27797 3578 V27798 3578 V27802 3578 V27803 3578 V27792 3578 V27799 3578 V27800 3578 V27804 3578 V27778 3578 V27783 3578 V27788 3578 V27793 3578 V27794 3578 V27780 3578 V27784 3578 V27786 3578 V27795 3578 V27796 3578 V27801 3578 V27900 3578 V27910 3577 V32173 3577 V31882 3577 V31883 3577 V32166 3577 V32194 3577 V32203 3577 V32198 3578 V27705 3578 V27706 3578 V27708 3578 V27710 3578 V27711 3578 V27712 3578 V27714 3578 V27715 3578 V27716 3578 V27717 3578 V27719 3578 V27781 3577 V32196 3577 V31375 3577 V31384 3577 V31386 3577 V31397 3577 V31398 3577 V31399 3577 V31400 3577 V31401 3577 V31402 3577 V31403 3577 V31404 3577 V31405 3577 V31406 3577 V31407 3577 V31410 3577 V31413 3577 V31414 3577 V31416 3577 V31417 3577 V31420 3577 V31388 3635 V40900 3635 V40901 3635 V40903 3635 V40904 3635 V40905 3635 V40906 3635 V40907 3635 V40909 3635 V40910 3635 V40912 3635 V40914 3635 V40915 3635 V40916 3635 V40917 3635 V40919 3635 V40921 3635 V40922 3635 V40923 3635 V40924 3635 V40925 3635 V40926 3635 V40927 3635 V40928 3635 V40929 3635 V40930 3635 V40931 3635 V40932 3635 V40933 3635 V40934 3635 V40935 3635 V40936 3635 V40937 3635 V40938 3635 V40939 3635 V40940 3635 V40941 3635 V40942 3635 V40943 3635 V40944 3635 V40945 3635 V40946 3635 V40947 3635 V40948 3635 V40949 3635 V40950 3635 V40951 3635 V40952 3635 V40954 3635 V40955 3635 V40956 3635 V40957 3635 V40958 3635 V40959 3635 V40960 3635 V40961 3635 V40964 3635 V40965 3635 V40966 3635 V40967 3635 V40968 3635 V40969 3635 V40970 3635 V40971 3635 V40972 3635 V40973 3635 V40974 3635 V40975 3635 V40976 3635 V40977 3635 V40978 3635 V40979 3635 V40980 3635 V40981 3635 V40982 3635 V40983 3635 V40984 3635 V40985 3635 V40986 3635 V40987 3635 V40988 3635 V40989 3635 V40990 3635 V40991 3635 V40993 3635 V40994 3635 V40995 3635 V40996 3635 V40997 3635 V40998 3635 V40999 3635 V41000 3635 V41001 3635 V41002 3635 V41003 3635 V41004 3635 V41005 3635 V41006 3635 V41007 3635 V41008 3635 V41010 3577 V31383 3577 V31385 3577 V31387 3577 V32585 3577 V32175 3577 V32183 3577 V32187 3449 V35472 3449 V35560 3449 V35561 3449 V36651 3449 V36652 3449 V36654 3449 V36656 3449 V36657 3449 V36658 3449 V36659 3449 V36660 3449 V36661 3449 V36662 3449 V36663 3577 V31392 3577 V31393 3577 V31394 3577 V31396 3577 V31395 3577 V31374 3577 V33168 3577 V33194 3577 V33198 3577 V33141 3577 V33143 3577 V33154 3577 V33161 3577 V33167 3577 V33170 3577 V33172 3577 V33175 3577 V33178 3577 V33179 3577 V33171 3577 V33174 3577 V33177 3577 V33181 3577 V33182 3577 V33183 3577 V33184 3577 V33185 3577 V33186 3577 V33187 3577 V33188 3577 V33189 3577 V33190 3577 V33191 3577 V33192 3577 V33193 3577 V33195 3577 V33196 3577 V33197 3577 V33199 3578 V27683 3578 V27816 3578 V27819 3578 V27822 3578 V27830 3578 V27820 3578 V27821 3578 V27826 3578 V27828 3578 V27829 3577 V31376 3577 V31382 3577 V31389 3577 V31391 3577 V33166 3577 V33169 3577 V33176 |
| FEI Number |
3001452206
|
Recalling Firm/ Manufacturer |
Masimo Corporation Forty Parker 40, 50 & 60 Parker Irvine CA 92618
|
| For Additional Information Contact | 949-297-7000 |
Manufacturer Reason for Recall | Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device. Additionally when a high SpO2 value is measured, a low SpO2 value would be displayed on the monitoring device. |
FDA Determined Cause 2 | Employee error |
| Action | An "Urgent: Advisory Notice," dated 09/30/2014, was sent to customers. The letter informed customers of the problem identified and the actions to be taken.
Customers were instructed to evaluate their inventory of Oximetry Cable(s) using the following method: Use the Oximetry Cable to take a SpO2 measurement on a healthy individual with an expected SpO2 value above 90% or review SpO2 measurements take with the Oximetry Cable and verify that measurement results were consistent with expected SpO2 values. If the SpO2 measurement(s) is consistent with expected SpO2 value, the Oximetry Cable is not affected and can continue to be used. Customers are instructed to complete and return Attachment 2, Tracking/Verification Form, to Masimo in order to document that the Oximetry Cable(s) is functioning as intented. If the assessment results in SpO2 measurement that is not consistent with expected SpO2 value, then the Oximetry Cable should be removed from service, segregated and returned to Masimo. Customers are instructed to call 1-800-326-4890 and select option 2 for Technical Services. Customers are to determine with Technical Services whether they will return the affected Oximetry Cable(s) to Masimo for replacement or refund. If returning for replacement customers are suppose to obtain a Return Material Authorization Number (RMA number). |
| Quantity in Commerce | 21,754 units (21,222) in the U.S. and Puerto Rico) |
| Distribution | Worldwide distribution including; nationwide (U.S. and Puerto Rico) and the countries of Argentina, Austria, Brazil, Canada, China, Denmark, France, Gernany, Hong Kong, India, Italy, Japan, Jordan, Kuwait, Mexico, Netherlands, Poland, Portugal, South Korea, Spain, Switzerland, Turkey, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DQA
|
|
|
|