|
Class 2 Device Recall Fresenius Crit Line |
|
Date Initiated by Firm |
December 19, 2014 |
Date Posted |
February 03, 2015 |
Recall Status1 |
Terminated 3 on February 18, 2021 |
Recall Number |
Z-1047-2015 |
Recall Event ID |
70288 |
510(K)Number |
K022536
|
Product Classification |
Accessories, blood circuit, hemodialysis - Product Code KOC
|
Product |
Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen saturation and percent change in blood volume calculation during hemodialysis treatment. |
Code Information |
Serial Numbers: 1C31M140038 1C31M140040 1C31M140041 1C31M140042 1C31M140033 1C31M140034 1C31M140035 1C31M140030 1C31M140075 1C31M140031 1C31M140027 1C31M140054 1C31M140057 1C31M140060 1C31M140039 1C31M140056 1C32M140005, 1C32M140059, 1C32M140058, 1C32M140014, 1C32M140013, 1C32M140004, 1C32M140012, 1C32M140011, 1C32M140009, 1C32M140008, 1C32M140003, 1C32M140007, 1C32M140006, 1C32M140023, 1C32M140021, 1C32M140022, 1C32M140020, 1C31M140073, 1C31M140050 |
Recalling Firm/ Manufacturer |
Fresenius Medical Care Holdings, Inc. 920 Winter St Waltham MA 02451-1521
|
For Additional Information Contact |
800-662-1237
|
Manufacturer Reason for Recall |
Potential for misinterpretation of the graphic display of the Blood Volume ( BV) slope
|
FDA Determined Cause 2 |
Device Design |
Action |
FMCRTG, LLC representatives contacted Clinics on 12/19/14 by visit/telephone phone) and removed all CLiC units. |
Quantity in Commerce |
35 units |
Distribution |
CT, NY |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KOC and Original Applicant = FRESENIUS MEDICAL CARE
|
|
|
|