Class 2 Device Recall NAMIC (TM) Convenience Kit |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
January 09, 2015 |
Date Posted |
February 26, 2015 |
Recall Status1 |
Terminated 3 on March 10, 2017 |
Recall Number |
Z-1189-2015 |
Recall Event ID |
70322 |
Product Classification |
Cardiovascular procedure kit - Product Code OEZ
|
Product |
NAMIC Convenience Kit, UPN H749602301481, REF/Catalog No. 60230148, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588.
NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems. |
Code Information |
Lots 4245786, 4248982, 4252309, 4260234, 4268141, 4279218, 4288402, 4298225, 4314481, 4320621, 4329211, 4337129, 4344176, 4353785, 4361710, 4372511, 4384175, 4393701, 4398980, 4403990, 4423304, 4457003, 4466333, 4469607, 4476888, 4485248, 4492442, 4498101, 4501790, 4507503, 4514700, 4528475, 4538380, 4548574, 4557974, 4565199, 4575422, 4579968, 4597587, 4613549, 4616353, 4636559, 4639117, 4642346, 4643978, 4644874, 4645905, 4649110, 4651282, 4651492, 4663423, 4667161, 4668890, 4671604, 4675060, 4676056, 4677276, 4678679, 4681618, 4685667, 4687664, 4690797, 4692101, 4693222, 4694581, 4697711, 4699670, 4701307, 4704014, 4706057, 4706979, 4707971, 4709279, 4721006, 4722335, 4725913, 4731836, 4731978, 4735580, 4739507, 4746790, 4749728, 4754079, 4757753, 4760072, 4763823, 4767249, 4769237, 4773706, 4774613, 4778708, 4783821, 4790213, 4792502, 4794061, 4799220, 4802066, 4804507, 4814927, 4817339, 4819622, 4820297, 4822238, 4828380, 4831134, 4833530, 4835324. |
Recalling Firm/ Manufacturer |
Navilyst Medical, Inc 10 Glens Falls Technical Park Glens Falls NY 12801
|
For Additional Information Contact |
518-792-4112
|
Manufacturer Reason for Recall |
The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
|
FDA Determined Cause 2 |
Packaging |
Action |
Navilyst Medical sent an Urgent Medical Device Recall Notification letter dated January 9, 2015 to US customers (end users) via Federal Express. On January 20, 2015, Recall Notification Packages were sent via Federal Express to 2 International consignees (distributors). the letter identified the affected product, problem and actions to be taken. Customers are instructed to:
" Immediately remove the recall product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location).
" Segregate this product in a secure location for return to Navilyst Medical.
" Immediately forward a copy of this recall notification to all sites to which you have distributed affected product.
" Complete and return the Reply Verification Tracking Form. |
Quantity in Commerce |
2502 units |
Distribution |
Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam. |