• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall EPIQ 7 Ultrasound System, WITHOUT the Pediatric Cardiology option

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall EPIQ 7 Ultrasound System, WITHOUT the Pediatric Cardiology optionsee related information
Date Initiated by FirmApril 14, 2015
Date PostedMay 19, 2015
Recall Status1 Terminated 3 on August 08, 2016
Recall NumberZ-1632-2015
Recall Event ID 71105
510(K)NumberK132304 
Product Classification System, imaging, pulsed doppler, ultrasonic - Product Code IYN
ProductEPIQ 7 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 7G, EPIC 7C, EPIQ 7W. Catalog Number: 795200 / 795201 Part Number: 989605386721 With: 453561726491 (1.0) 453561728121 (1.0.1) 453561736781 (1.1) 453561750021 (1.1.1) 453561753631 (1.1.2) 453561772251 (1.2) 453561772631 (1.2.1) 453561786591 (1.2.2) 453561785101 (1.3) 453561800601 (1.3.1) 453561805211 (1.3.2) Diagnostic ultrasound imaging and fluid flow analysis
Code Information ********************SYSTEM BASE ID IN THE US*****************************  US114B0056, US114B0057, US114B0160, US114B0187, US114B0188, US114B0189, US114B0224, US114B0282, US114B0283, US114B0303, US114B0318, US114B0355, US114B0393, US114B0394, US114B0400, US114B0401, US114B0403, US114B0435, US114B0468, US114B0495, US114B0527, US114B0528, US114B0529, US114B0537, US114B0538, US114B0540, US114B0541, US114B0542, US114B0544, US114B0545, US114B0547, US114B0548, US114B0552, US114B0569, US114B0570, US114B0583, US114B0612, US114B0702, US114B0705, US114B0706, US114B0707, US114B0709, US114B0710, US114B0711, US114B0712, US115B0010, US115B0011, US115B0013, US115B0014, US115B0043, US115B0044, US115B0045, US115B0046, US115B0047, US115B0048, US115B0049, US115B0090, US115B0125, US115B0166, US115B0167, US115B0264, US115B0267, US115B0269, US115B0270, US115B0377, US115B0378, US115B0480, US115B0481, US115B0482, US115B0483, US115B0519, US115B0520, US115B0586, US115B0592, US115B0593, US115B0594, US115B0595, US115B0656, US115B0657, US115B0658, US214B0005, US214B0015, US214B0016, US214B0030, US214B0247, US214B0248, US214B0249, US214B0433, US214B0518, US214B0521, US214B0522, US214B0527, US214B0571, US214B0572, US214B0573, US215B0115, US215B0159, US215B0160, US215B0161, US215B0170, US215B0195, US215B0196, US215B0197, US215B0297, US215B0298, US215B0300, US215B0451, US215B0452, US215B0470, US215B0476, US215B0477, US215B0478, US215B0479, US215B0480, US215B0481, US215B0482, US215B0522, US215B0525, US215B0526, US215B0527, US215B0528, US215B0554, US215B0610, US215B0611, US215B0612, US215B0632, US215B0633, US215B0634, US215B0708, US215B0709, US215B0719, US215B0729, US215B0852, US215B0853, US215B0854, US215B0856, US215B0860, US215B0893, US215B0896, US215B0934, US215B0935, US215B0943, US215B0944, US215B0945, US215B0946, US215B0947, US215B1013, US215B1014, US215B1015, US215B1016, US215B1017, US215B1018, US215B1019, US215B1032, US215B1033, US215B1034, US215B1035, US215B1036, US215B1037, US215B1038, US215B1039, US215B1040, US215B1041, US313B0383, US313B0385, US314B0012, US314B0013, US314B0015, US314B0017, US314B0018, US314B0019, US314B0094, US314B0106, US314B0301, US314B0320, US314B0396, US314B0423, US314B0448, US314B0483, US314B0484, US314B0486, US314B0772, US314B0773, US314B0774, US314B0797, US315B0038, US315B0045, US315B0057, US315B0059, US315B0187, US315B0191, US315B0233, US315B0234, US315B0236, US413B0001, US413B0002, US413B0038, US413B0039, US413B0052, US413B0053, US413B0088, US413B0089, US413B0090, US413B0105, US413B0129, US413B0130, US413B0131, US413B0148, US413B0150, US413B0152, US413B0165, US413B0186, US413B0187, US413B0194, US413B0195, US413B0196, US413B0197, US413B0198, US413B0200, US413B0201, US413B0258, US413B0280, US413B0282, US413B0340, US413B0383, US413B0384, US413B0385, US413B0399, US413B0400, US413B0401, US413B0402, US413B0434, US413B0435, US413B0436, US413B0438, US413B0455, US413B0457, US413B0491, US413B0533, US413B0534, US413B0536, US413B0537, US413B0553, US413B0554, US413B0557, US413B0558, US414B0003, US414B0063, US414B0156, US414B0265, US414B0266, US414B0301, US414B0302, US414B0333, US414B0334, US414B0335, US414B0336, US414B0337, US414B0338, US414B0476, US414B0522, US414B0525, US414B0530, US414B0553, US513B0002, US513B0003, US513B0004, US513B0065, US513B0174, US513B0176, US513B0177, US513B0226, US513B0228, US513B0257, US513B0259, US513B0260, US513B0261, US513B0262, US513B0263, US513B0264, US513B0266, US513B0295, US513B0324, US513B0350, US513B0351, US513B0415, US513B0440, US513B0442, US513B0443, US513B0444, US513B0445, US513B0446, US513B0534, US513B0535, US513B0536, US513B0537, US513B0540, US513B0562, US513B0563, US513B0620, US513B0621, US513B0623, US514B0173, US514B0216, US514B0242, US514B0270, US514B0313, US514B0314, US514B0417, US514B0418, US514B0419, US514B0423, US514B0492, US514B0493, US514B0509, US514B0607, US514B0624, US514B0625, US514B0626, US514B0631, US514B0676, US514B0682, US514B0722, US613B0082, US613B0085, US613B0123, US613B0124, US613B0125, US613B0126, US613B0127, US613B0141, US613B0142, US613B0143, US613B0144, US613B0156, US613B0157, US613B0159, US613B0176, US613B0177, US613B0178, US613B0179, US613B0278, US613B0286, US613B0287, US613B0288, US613B0289, US613B0290, US613B0291, US613B0300, US613B0301, US613B0404, US613B0405, US613B0454, US613B0486, US613B0488, US614B0011, US614B0147, US614B0213, US614B0284, US614B0312, US614B0322, US614B0323, US614B0324, US614B0325, US614B0413, US614B0414, US614B0415, US614B0433, US614B0434, US614B0476, US614B0477, US614B0478, US614B0507, US614B0508, US614B0509, US614B0513, US614B0615, US614B0616, US614B0618, US614B0619, US614B0620, US614B0629, US614B0777, US614B0804, US614B0805, US614B0806, US614B0807, US614B0808, US614B0813, US614B0839, US614B0840, US614B0841, US614B0862, US614B0863, US614B0864, US614B0865, US614B0866, US614B0867, US614B0868, US614B0911, US713B0199, US713B0207, US713B0224, US713B0225, US713B0226, US713B0227, US713B0228, US713B0264, US713B0273, US713B0274, US713B0275, US713B0276, US713B0277, US713B0278, US713B0288, US713B0289, US713B0290, US713B0291, US713B0306, US713B0307, US713B0308, US713B0339, US713B0340, US713B0341, US713B0342, US713B0343, US713B0344, US713B0345, US713B0364, US713B0365, US713B0398, US713B0399, US713B0400, US713B0401, US713B0402, US713B0403, US713B0404, US713B0405, US713B0438, US713B0442, US713B0443, US713B0444, US713B0445, US713B0492, US713B0493, US713B0494, US713B0495, US713B0496, US713B0498, US713B0524, US713B0525, US713B0526, US713B0527, US713B0528, US714B0052, US714B0053, US714B0054, US714B0060, US714B0065, US714B0136, US714B0157, US714B0186, US714B0189, US714B0190, US714B0243, US714B0248, US714B0284, US714B0325, US714B0326, US714B0357, US714B0394, US714B0395, US714B0396, US714B0397, US714B0399, US714B0400, US714B0402, US714B0408, US714B0409, US714B0410, US714B0411, US714B0412, US714B0413, US714B0414, US714B0415, US714B0416, US714B0435, US714B0436, US714B0437, US714B0457, US714B0467, US714B0515, US714B0535, US714B0547, US714B0577, US714B0626, US714B0627, US714B0628, US714B0629, US714B0630, US714B0631, US714B0632, US714B0633, US714B0634, US714B0648, US714B0665, US714B0666, US714B0667, US813B0017, US813B0018, US813B0019, US813B0020, US813B0022, US813B0023, US813B0024, US813B0025, US813B0122, US813B0123, US813B0125, US813B0127, US813B0128, US813B0129, US813B0131, US813B0132, US813B0134, US813B0135, US813B0230, US813B0231, US813B0233, US813B0234, US813B0235, US813B0236, US813B0237, US813B0238, US813B0239, US813B0240, US813B0241, US813B0423, US813B0424, US813B0425, US813B0426, US813B0427, US813B0428, US813B0429, US813B0430, US813B0431, US813B0432, US813B0433, US813B0434, US814B0018, US814B0064, US814B0065, US814B0066, US814B0108, US814B0109, US814B0110, US814B0131, US814B0160, US814B0177, US814B0182, US814B0184, US814B0185, US814B0206, US814B0207, US814B0258, US814B0276, US814B0278, US814B0280, US814B0295, US814B0296, US814B0364, US814B0399, US814B0401, US814B0402, US814B0494, US814B0525, US814B0526, US814B0556, US814B0560, US814B0561, US814B0562, US814B0563, US814B0580, US814B0597, US814B0600, US913B0033, US913B0034, US913B0035, US913B0036, US913B0037, US913B0038, US913B0039, US913B0040, US913B0052, US913B0053, US913B0054, US913B0057, US913B0060, US913B0061, US913B0062, US913B0063, US913B0064, US913B0065, US913B0066, US913B0106, US913B0107, US913B0108, US913B0109, US913B0110, US913B0111, US913B0112, US913B0113, US913B0182, US913B0186, US913B0232, US913B0247, US913B0248, US913B0249, US913B0250, US913B0258, US913B0350, US913B0351, US913B0352, US913B0395, US913B0396, US913B0397, US913B0398, US913B0399, US913B0400, US913B0401, US913B0522, US914B0009, US914B0028, US914B0029, US914B0080, US914B0081, US914B0082, US914B0083, US914B0084, US914B0173, US914B0178, US914B0185, US914B0195, US914B0196, US914B0198, US914B0232, US914B0236, US914B0414, US914B0415, US914B0459, US914B0460, US914B0461, US914B0462, US914B0463, US914B0464, US914B0465, US914B0466, US914B0467, US914B0468, US914B0469, US914B0487, US914B0488, US914B0489, US914B0490, US914B0491, US914B0494, US914B0495, US914B0496, US914B0497, US914B0525, US914B0530, US914B0531, US914B0532, US914B0533, US914B0534, US914B0545, US914B0546, US914B0626, US914B0690, US914B0695, US914B0736, US914B0776, US914B0790, US914B0791, US914B0792, US914B0830, US914B0831, US914B0832, US914B0833, US914B0834, US914B0847, US914B0903, US914B0904, US914B0944, US914B0945, US914B0946, US914B0947, US914B0948, US914B0951, US914B0952, USD13B0017, USD13B0213, USD13B0311, USD13B0413, USD13B0424, USD13B0425, USD13B0451, USD13B0465, USD13B0469, USD13B0470, USD13B0533, USD13B0534, USD13B0551, USD13B0561, USD13B0567, USD13B0581, USD13B0582, USD13B0626, USD13B0627, USD13B0628, USD13B0629, USD13B0630, USD13B0631, USD13B0632, USD13B0633, USD13B0658, USD13B0665, USD13B0666, USD13B0704, USD13B0705, USD13B0721, USD13B0778, USD13B0780, USD13B0783, USD14B0238, USD14B0239, USD14B0337, USD14B0338, USD14B0345, USD14B0351, USD14B0460, USD14B0461, USD14B0505, USD14B0506, USD14B0507, USD14B0516, USD14B0583, USD14B0585, USD14B0590, USD14B0593, USD14B0766, USD14B0767, USD14B0866, USD14B0873, USD14B0944, USD14B0945, USD14B0946, USD14B0947, USD14B0948, USD14B0949, USD14B0950, USD14B0958, USD14B0959, USD14B0960, USD14B0961, USD14B0995, USD14B0996, USD14B1029, USD14B1030, USD14B1033, USD14B1035, USN13B0037, USN13B0201, USN13B0204, USN13B0334, USN13B0397, USN13B0398, USN13B0413, USN13B0422, USN13B0466, USN13B0467, USN13B0468, USN13B0515, USN13B0563, USN13B0566, USN13B0568, USN13B0651, USN14B0024, USN14B0025, USN14B0026, USN14B0028, USN14B0095, USN14B0293, USN14B0305, USN14B0306, USN14B0307, USN14B0393, USN14B0406, USN14B0407, USN14B0412, USN14B0413, USN14B0414, USN14B0415, USN14B0416, USN14B0417, USN14B0447, USN14B0525, USN14B0526, USN14B0527, USN14B0528, USN14B0529, USN14B0530, USN14B0531, USN14B0532, USN14B0533, USN14B0534, USN14B0594, USN14B0595, USN14B0643, USN14B0644, USN14B0676, USN14B0677, USN14B0678, USN14B0679, USN14B0680, USN14B0681, USN14B0682, USN14B0690, USN14B0691, USN14B0692, USN14B0775, USN14B0776, USN14B0892, USO13B0053, USO13B0207, USO13B0210, USO13B0241, USO13B0286, USO13B0287, USO13B0288, USO13B0289, USO13B0290, USO13B0394, USO13B0449, USO13B0450, USO13B0451, USO13B0452, USO13B0598, USO13B0602, USO13B0631, USO14B0011, USO14B0027, USO14B0028, USO14B0045, USO14B0046, USO14B0047, USO14B0048, USO14B0087, USO14B0154, USO14B0155, USO14B0192, USO14B0203, USO14B0204, USO14B0267, USO14B0269, USO14B0270, USO14B0342, USO14B0369, USO14B0371, USO14B0372, USO14B0373, USO14B0374, USO14B0375, USO14B0402, USO14B0409, USO14B0440, USO14B0441, USO14B0442, USO14B0443, USO14B0444, USO14B0445, USO14B0463, USO14B0464, USO14B0524, USO14B0526, USO14B0554, USO14B0555, USO14B0564, USO14B0668, USO14B0669, USO14B0713, USO14B0723, USO14B0751, USO14B0752, USO14B0757, USO14B0758, USO14B0759, USO14B0772, USO14B0773, USO14B0915, USO14B0917, and USO14B1085.   ***********************SYSTEM BASE ID IN CANADA********************************  US115B0271, US115B0272, US115B0438, US115B0439, US115B0444, US115B0445, US115B0472, US115B0473, US115B0474, US115B0517, US214B0443, US214B0444, US215B0517, US215B0557, US215B0558, US215B0629, US215B0723, US215B0728, US215B0730, US215B0930, US215B0931, US215B0932, US215B0933, US314B0174, US413B0437, US413B0556, US414B0332, US414B0393, US414B0496, US414B0520, US514B0229, US514B0230, US914B0242, US914B0628, US914B0691, USD14B0093, USD14B0382, USN14B0360, USN14B0893, USN14B0923, USO14B0152, USO14B0153, and USO14B1083.   *********************SYSTEM BASE IN OTHER COUNTRIES**********************  US114B0090, US114B0091, US114B0092, US114B0113, US114B0126, US114B0167, US114B0201, US114B0202, US114B0203, US114B0205, US114B0234, US114B0235, US114B0236, US114B0237, US114B0264, US114B0265, US114B0280, US114B0281, US114B0324, US114B0325, US114B0326, US114B0347, US114B0354, US114B0396, US114B0397, US114B0433, US114B0438, US114B0439, US114B0470, US114B0476, US114B0488, US114B0506, US114B0507, US114B0568, US114B0693, US114B0708, US114B0740, US114B0741, US115B0007, US115B0008, US115B0009, US115B0039, US115B0040, US115B0041, US115B0085, US115B0086, US115B0087, US115B0096, US115B0126, US115B0127, US115B0128, US115B0129, US115B0130, US115B0131, US115B0168, US115B0169, US115B0170, US115B0171, US115B0172, US115B0200, US115B0201, US115B0205, US115B0207, US115B0375, US115B0440, US115B0441, US115B0478, US115B0486, US115B0492, US115B0493, US115B0587, US115B0588, US115B0589, US115B0590, US115B0591, US115B0604, US115B0659, US115B0660, US214B0001, US214B0095, US214B0163, US214B0199, US214B0229, US214B0230, US214B0231, US214B0244, US214B0395, US214B0398, US214B0400, US214B0401, US214B0405, US214B0429, US214B0430, US214B0431, US214B0442, US214B0448, US214B0449, US214B0473, US214B0515, US214B0528, US214B0531, US215B0021, US215B0076, US215B0092, US215B0096, US215B0097, US215B0112, US215B0113, US215B0114, US215B0126, US215B0127, US215B0185, US215B0201, US215B0294, US215B0334, US215B0338, US215B0339, US215B0340, US215B0363, US215B0435, US215B0436, US215B0437, US215B0438, US215B0440, US215B0448, US215B0473, US215B0474, US215B0484, US215B0492, US215B0518, US215B0519, US215B0524, US215B0581, US215B0631, US215B0681, US215B0710, US215B0711, US215B0712, US215B0722, US215B0726, US215B0727, US215B0849, US215B0851, US215B0862, US215B0898, US215B0899, US215B0937, US215B1028, US314B0061, US314B0095, US314B0136, US314B0154, US314B0172, US314B0173, US314B0204, US314B0239, US314B0302, US314B0303, US314B0307, US314B0614, US314B0795, US314B0809, US314B0839, US314B0847, US315B0047, US315B0080, US315B0081, US315B0237, US315B0304, US413B0199, US414B0002, US414B0203, US414B0208, US414B0255, US414B0256, US414B0257, US414B0303, US414B0309, US414B0378, US414B0470, US414B0473, US414B0551, US414B0579, US414B0606, US414B0609, US414B0613, US414B0654, US513B0001, US513B0193, US513B0224, US513B0258, US513B0320, US513B0387, US513B0388, US513B0558, US513B0559, US513B0560, US514B0109, US514B0111, US514B0147, US514B0231, US514B0232, US514B0234, US514B0237, US514B0266, US514B0267, US514B0268, US514B0278, US514B0279, US514B0311, US514B0312, US514B0322, US514B0323, US514B0324, US514B0336, US514B0362, US514B0369, US514B0422, US514B0489, US514B0510, US514B0584, US514B0586, US514B0590, US514B0622, US514B0623, US514B0664, US514B0674, US514B0723, US514B0724, US613B0081, US613B0083, US613B0158, US613B0245, US613B0299, US613B0317, US613B0359, US613B0366, US614B0009, US614B0010, US614B0150, US614B0180, US614B0181, US614B0182, US614B0183, US614B0184, US614B0185, US614B0186, US614B0187, US614B0188, US614B0229, US614B0230, US614B0238, US614B0239, US614B0245, US614B0319, US614B0320, US614B0321, US614B0328, US614B0381, US614B0417, US614B0426, US614B0427, US614B0428, US614B0435, US614B0454, US614B0456, US614B0621, US614B0623, US614B0625, US614B0666, US614B0667, US614B0733, US614B0737, US614B0778, US614B0779, US614B0797, US614B0798, US614B0815, US713B0085, US713B0086, US713B0118, US713B0119, US713B0120, US713B0121, US713B0122, US713B0124, US713B0125, US713B0126, US713B0127, US713B0148, US713B0149, US713B0161, US713B0163, US713B0164, US713B0165, US713B0166, US713B0171, US713B0172, US713B0192, US713B0193, US713B0194, US713B0195, US713B0206, US713B0209, US713B0324, US713B0325, US713B0326, US713B0327, US713B0328, US713B0358, US713B0360, US713B0362, US713B0363, US713B0406, US713B0436, US713B0437, US713B0439, US713B0440, US713B0441, US713B0497, US713B0581, US713B0582, US713B0583, US713B0584, US713B0586, US713B0587, US713B0588, US713B0589, US713B0593, US713B0594, US713B0595, US714B0001, US714B0002, US714B0003, US714B0006, US714B0007, US714B0037, US714B0039, US714B0047, US714B0049, US714B0058, US714B0059, US714B0102, US714B0137, US714B0156, US714B0168, US714B0181, US714B0183, US714B0210, US714B0211, US714B0212, US714B0213, US714B0231, US714B0232, US714B0233, US714B0234, US714B0235, US714B0236, US714B0239, US714B0281, US714B0283, US714B0316, US714B0317, US714B0343, US714B0344, US714B0345, US714B0355, US714B0438, US714B0455, US714B0458, US714B0459, US714B0460, US714B0466, US714B0533, US714B0534, US714B0546, US714B0691, US714B0692, US714B0728, US813B0016, US813B0021, US813B0032, US813B0037, US813B0038, US813B0121, US813B0124, US813B0151, US813B0152, US813B0153, US813B0154, US813B0155, US813B0156, US813B0157, US813B0158, US813B0159, US813B0160, US813B0162, US813B0163, US813B0195, US813B0196, US813B0232, US813B0401, US813B0402, US813B0405, US813B0457, US814B0019, US814B0054, US814B0060, US814B0061, US814B0097, US814B0098, US814B0099, US814B0100, US814B0104, US814B0135, US814B0136, US814B0159, US814B0186, US814B0205, US814B0242, US814B0261, US814B0273, US814B0275, US814B0304, US814B0306, US814B0307, US814B0416, US814B0417, US814B0419, US814B0421, US814B0422, US814B0521, US814B0554, US814B0555, US814B0559, US814B0589, US814B0591, US814B0599, US814B0622, US913B0011, US913B0013, US913B0026, US913B0027, US913B0059, US913B0099, US913B0100, US913B0101, US913B0102, US913B0103, US913B0104, US913B0105, US913B0130, US913B0136, US913B0137, US913B0138, US913B0139, US913B0177, US913B0178, US913B0179, US913B0181, US913B0202, US913B0218, US913B0251, US913B0252, US913B0253, US913B0254, US913B0255, US913B0301, US913B0302, US913B0303, US913B0332, US913B0333, US913B0336, US913B0337, US913B0339, US913B0340, US913B0349, US913B0385, US913B0408, US913B0458, US913B0487, US913B0488, US913B0489, US913B0490, US913B0504, US913B0505, US914B0013, US914B0031, US914B0044, US914B0045, US914B0049, US914B0077, US914B0078, US914B0079, US914B0127, US914B0129, US914B0130, US914B0131, US914B0186, US914B0227, US914B0228, US914B0229, US914B0230, US914B0231, US914B0276, US914B0277, US914B0384, US914B0388, US914B0389, US914B0390, US914B0391, US914B0392, US914B0393, US914B0412, US914B0492, US914B0493, US914B0625, US914B0627, US914B0631, US914B0725, US914B0788, US914B0789, US914B0846, US914B0899, US914B0900, US914B0901, US914B0902, US914B0906, US914B0907, USD13B0050, USD13B0061, USD13B0084, USD13B0085, USD13B0144, USD13B0150, USD13B0151, USD13B0152, USD13B0215, USD13B0243, USD13B0244, USD13B0252, USD13B0253, USD13B0349, USD13B0350, USD13B0351, USD13B0352, USD13B0356, USD13B0382, USD13B0417, USD13B0637, USD14B0025, USD14B0102, USD14B0103, USD14B0107, USD14B0143, USD14B0144, USD14B0148, USD14B0149, USD14B0241, USD14B0243, USD14B0336, USD14B0383, USD14B0397, USD14B0398, USD14B0458, USD14B0459, USD14B0463, USD14B0464, USD14B0579, USD14B0581, USD14B0582, USD14B0584, USD14B0587, USD14B0589, USD14B0648, USD14B0653, USD14B0658, USD14B0745, USN13B0013, USN13B0034, USN13B0035, USN13B0036, USN13B0153, USN13B0420, USN13B0428, USN13B0429, USN13B0430, USN13B0436, USN13B0437, USN13B0507, USN13B0514, USN13B0554, USN13B0555, USN13B0556, USN13B0557, USN13B0581, USN13B0590, USN13B0591, USN13B0642, USN13B0643, USN13B0646, USN14B0020, USN14B0134, USN14B0139, USN14B0167, USN14B0168, USN14B0181, USN14B0182, USN14B0183, USN14B0184, USN14B0232, USN14B0233, USN14B0299, USN14B0358, USN14B0359, USN14B0385, USN14B0386, USN14B0441, USN14B0443, USN14B0444, USN14B0446, USN14B0517, USN14B0519, USN14B0520, USN14B0522, USN14B0583, USN14B0584, USN14B0585, USN14B0590, USN14B0592, USN14B0621, USN14B0650, USN14B0652, USN14B0653, USN14B0670, USN14B0671, USN14B0683, USN14B0684, USN14B0685, USN14B0772, USN14B0777, USN14B0797, USN14B0798, USN14B0799, USN14B0800, USN14B0801, USN14B0802, USN14B0895, USN14B0896, USN14B0897, USN14B0899, USN14B0900, USN14B0906, USN14B0924, USN14B0934, USN14B1022, USN14B1027, USO13B0018, USO13B0019, USO13B0200, USO13B0201, USO13B0202, USO13B0203, USO13B0214, USO13B0265, USO13B0268, USO13B0294, USO13B0295, USO13B0296, USO13B0304, USO13B0313, USO13B0322, USO13B0323, USO13B0439, USO13B0459, USO13B0567, USO13B0633, USO13B0634, USO13B0635, USO13B0636, USO13B0637, USO13B0638, USO13B0639, USO13B0640, USO13B0641, USO13B0645, USO13B0672, USO14B0010, USO14B0044, USO14B0049, USO14B0050, USO14B0083, USO14B0084, USO14B0085, USO14B0088, USO14B0089, USO14B0111, USO14B0112, USO14B0114, USO14B0119, USO14B0120, USO14B0121, USO14B0147, USO14B0148, USO14B0149, USO14B0150, USO14B0151, USO14B0189, USO14B0190, USO14B0191, USO14B0264, USO14B0324, USO14B0367, USO14B0404, USO14B0405, USO14B0420, USO14B0421, USO14B0422, USO14B0423, USO14B0424, USO14B0518, USO14B0519, USO14B0523, USO14B0525, USO14B0527, USO14B0556, USO14B0670, USO14B0671, USO14B0672, USO14B0711, USO14B0717, USO14B0770, USO14B0818, USO14B0841, USO14B0843, USO14B0926, USO14B1084, USO14B1089, USO14B1119, USO14B1121, USO14B1128, USO14B1129, USO14B1131, USO14B1167, USO14B1168, USO14B1169, and USO14B1172.
Recalling Firm/
Manufacturer
Philips Ultrasound, Inc.
22100 Bothell Everett Hwy
Bothell WA 98021-8431
For Additional Information ContactPhilips Customer Service
800-722-9377
Manufacturer Reason
for Recall
When EPIQ 7 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).
FDA Determined
Cause 2
Software design
ActionPhilips sent the Urgent-Medical Device Correction-Philips Model EPIQ5 Ultrasound system and the Urgent-Medical Device Correction-Philips Model EPIQ7 Ultrasound system letters, both dated 2015 APR 14, to their US consignees. Philips sent the Urgent-Field Safety Notice-Philips Model EPIQ5 Ultrasound system letter and the Urgent-Field Safety Notice-Philips Model EPIQ7 Ultrasound system letter (both letters dated 2015 APRIL 14) to their foreign consignees. The EPIQ5 or EPIQ7 Ultrasound systems WITHOUT the Pediatric Cardiology option are affected. Philips issued two Field Change Orders (FCO) to their Field Service Engineer to install upgraded software: -FCO 79500340 for EPIQ5 -FCO 79500341 for EPIQ7 Customers with questions can call Philips Customer Service at 1-800-722-9377. Field Service Engineers (FSE) to install software version 1.3.3 per FCO 79500340 instructions for EPIQ5 5 and per FCO 79500341 instructions for EPIQ7.
Quantity in Commerce1692 units total (902 units in the US and 790 units outside the US)
DistributionWorldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = IYN
-
-