 Class 2 Device Recall MX 16 Slicesee related information
Date Initiated by FirmMarch 06, 2015
Date PostedJune 11, 2015
Recall Status1 Terminated 3 on March 13, 2017
Recall NumberZ-1759-2015
Recall Event ID 71134
510(K)NumberK091195 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductMX 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
Code Information Model #: 728131; Serial #s: EP16E130006, EP16E130007, EP16E130009, EP16E130010, EP16E130012, EP16E130013, EP16E130014, EP16E130015, EP16E130016, EP16E130017, EP16E130020, EP16E130021, EP16E130022, EP16E130023, EP16E130025, EP16E130026, EP16E130027, EP16E130029, EP16E130030, EP16E130031, EP16E130033, EP16E130034, EP16E130035, EP16E130036, EP16E130037, EP16E130040, EP16E130041, EP16E130042, EP16E130043, EP16E130044, EP16E130045, EP16E130046, EP16E130047, EP16E130048, EP16E130049, EP16E130050, EP16E130051, EP16E130052, EP16E130053, EP16E130055, EP16E130056, EP16E130057, EP16E130058, EP16E130062, EP16E130063, EP16E130064, EP16E130065, EP16E130066, EP16E130067, EP16E130068, EP16E130069, EP16E130070, EP16E130071, EP16E130072, EP16E130073, EP16E130074, EP16E130075, EP16E130077, EP16E130078, EP16E130079, EP16E130080, EP16E130082, EP16E130083, EP16E130086, EP16E130087, EP16E130088, EP16E130089, EP16E130092, EP16E130093, EP16E130094, EP16E130095, EP16E130097, EP16E130098, EP16E130099, EP16E130100, EP16E130102, EP16E130103, EP16E130105, EP16E130106, EP16E130107, EP16E130109, EP16E130110, EP16E140001, EP16E140002, EP16E140003, EP16E140007, EP16E140008, EP16E140009, EP16E140010, EP16E140011, EP16E140012, EP16E140014, EP16E140015, EP16E140016, EP16E140017, EP16E140018, EP16E140019, EP16E140021, EP16E140022, EP16E140024, EP16E140025, EP16E140027, EP16E140029, EP16E140035, EP16E140039, EP16E140041, EP16E140042, EP16E140050, EP16E140057, EP16E140059, EP16E140075, EP16EB110006, EP16EB120006, EP16EB120007, EP16EB120008, EP16EB120009, EP16EB130001, EP16EB130002, EP16EB130003, P16E090027 & P16EM09002
FEI Number 1525965
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information ContactMr. Yuchol Kim
440-483-2015
Manufacturer Reason
for Recall
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn 3/9/2015 the firm sent Urgent Field Safety Notices Medical Device Correction letters to their customers.
Quantity in Commerce125 units
DistributionWorldwide Distribution-US (nationwide) including DC & Puerto Rico and the following countries: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cambodia, Cameroon, Cayman Islands, China, Colombia, Democratic Congo, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, LAO People's Dem. Rep., Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malawi, Malaysia, Malta, Mauritius, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Philippines, Poland, Portugal, R¿union, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia & Montenegro, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, and Zambia.
Total Product Life CycleTPLC Device Report