 Class 2 Device Recall PREMIUM O.B. PACKsee related information
Date Initiated by FirmJuly 10, 2015
Date PostedSeptember 03, 2015
Recall Status1 Terminated 3 on April 12, 2017
Recall NumberZ-2577-2015
Recall Event ID 71721
Product Classification Obstetrical kit - Product Code OKV
ProductPREMIUM O.B. PACK CUSTOMED 5/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .
Code Information Code:9001297  Lots: 141015609 exp. 10/31/15 141115976 exp. 11/30/15 150116757 exp. 2/28/16 150116758 exp. 2/28/16 150116756 exp. 2/28/16 150619380 exp. 6/30/16 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
Manufacturer Reason
for Recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
FDA Determined
Cause 2
Packaging
ActionCustomed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.
Quantity in Commerce390 packs
DistributionDistributed Only in Puerto Rico.