 Class 2 Device Recall SOUNDSTAR eco Cathetersee related information
Date Initiated by FirmSeptember 10, 2015
Date PostedOctober 01, 2015
Recall Status1 Terminated 3 on May 13, 2016
Recall NumberZ-0003-2016
Recall Event ID 72167
510(K)NumberK140318 
Product Classification Catheter, ultrasound, intravascular - Product Code OBJ
ProductSOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. For intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.
Code Information Catalog #: 10439236, 10439072, 10439011, 10438577
FEI Number 3003113629
Recalling Firm/
Manufacturer
Biosense Webster, Inc.
15715 Arrow Hwy
Irwindale CA 91706-2006
For Additional Information ContactVadim Kastin
909-839-8500
Manufacturer Reason
for Recall
Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
FDA Determined
Cause 2
Device Design
ActionA Customer Notification Letter and Acknowledgement Form was distributed to customers on 09/10/15. One letter and form was sent via express mail, addressed to the Risk Management Office and one additional letter and form was sent by the firm's account representative to the Electrophysiology (EP/Cardiology Lab directly at the customer site. The firm's notification letter stated that they want to emphasize the following statement "The intra-cardiac ultrasound image will disappear if the CARTO 3 EP Navigation System power is disrupted and this may present a safety issue if the EP is using the ultrasound to monitor the patien during EP procedure. The ultrasound image will not reappear until the CARTO 3 System is restored." The firm will be updating the product labeling to further reinforce the precautionary statement. The firm requests that customers take the following actions: 1. Read the Field Safety Notification carefully. 2. Pass on this notification to anyone in the facility that needs to be informed of this issue, including appropriate clinical personnel involved in the use of SOUNDSTAR eco Catheters. 3. Review, complete, sign and return the attached Acknowledgement Form in accordance with the instructions on the form. 4. Maintain a copy of this letter with the product. 5. Maintain awareness of the Field Safety Notification For any questions the firm states to contact the BWI sales representative.
Quantity in Commerce1035 (U.S.)
DistributionWorldwide Distribution. US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = OBJ
-
-