| | Class 2 Device Recall MicroScan Pos Breakpoint Combo Panel Type 23 |  |
| Date Initiated by Firm | December 01, 2015 |
| Create Date | December 30, 2015 |
| Recall Status1 |
Terminated 3 on July 29, 2016 |
| Recall Number | Z-0562-2016 |
| Recall Event ID |
72779 |
| Product Classification |
Panels, test, susceptibility, antimicrobial - Product Code LTT
|
| Product | MicroScan Pos Breakpoint Combo Panel Type 23, Part No. B1017-206
MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes. |
| Code Information |
2016-05-06 Part No. B1017-206 |
| FEI Number |
2050012
|
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
| For Additional Information Contact | Rula Melconian 714-961-4295 |
Manufacturer Reason for Recall | Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive. |
FDA Determined Cause 2 | Device Design |
| Action | The firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 1, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to discontinue use and discard any remaining inventory of lots listed; return the attached form for product replacement; retain your inventory of the other lot numbers of the panel types, as they are not impacted by this issue; discuss content of the letter with your laboratory director/staff, retain the notification as part of your laboratory QS documentation; if you have forwarded any of the affected products to another laboratory provide them a copy of the letter, and complete and return the enclosed Response Form within 10 days via mail to: Beckman Coulter, Inc., 1584 Enterprise Blvd., West Sacramento, CA 95691, Attn: Quality Systems & Compliance or Fax: 916-374-2119 or email: MicrobiologyCustomersHCUS@beckman.com.
Customers with questions are instructed to contact their Customer Support Representative at http://www.beckmancoulter.com or by phone at (800) 677-7226 in US and Canada. |
| Quantity in Commerce | 715 units total (616 units in US) |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Austria, Belgium, Brazil, Canada, Germany, Italy, Japan, Kazakhstan, Kuwait, Mexico, Myanmar, Netherlands, Norway, Panama, Poland, Portugal, Romania, Russian Federation, Spain, Taiwan, Thailand and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|