 Class 2 Device Recall Transpac IV Monitoring Kitsee related information
Date Initiated by FirmSeptember 27, 2016
Date PostedNovember 15, 2016
Recall Status1 Terminated 3 on July 11, 2017
Recall NumberZ-0570-2017
Recall Event ID 75491
510(K)NumberK061573 
Product Classification Transducer, blood-pressure, extravascular - Product Code DRS
ProductCath Lab Kit w/3 Port "ON" Manifold (600 psi), Transpac IV, 2 Admin Sets and 8 cc Control Syringe, Item No. 46300-01 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet Reservoir system is a blood sampling system assembled directly to the patient monitoring line (Transpac units). The SafeSet reservoir is used on patients where periodic blood sampling is required.
Code Information 3235660 3264173            
FEI Number 2025816
Recalling Firm/
Manufacturer
ICU Medical, Inc.
951 Calle Amanecer
San Clemente CA 92673-6212
For Additional Information Contact
949-366-2183
Manufacturer Reason
for Recall
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
FDA Determined
Cause 2
Device Design
ActionICU Medical sent an Urgent Medical Device Recall Notification letter dated September 27, 2016, to all affected customers to inform them of the expanded recall of the monitoring kits with Transpac IV transducers. Customers were informed of updated communications from August 10, 2016, when ICU Medical sent a communication indicating a potential leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers. Based on continued evaluation, ICU Medical is expanding the notification to now include certain other monitoring kits that contain the Transpac IV transducers. A small number of these devices may have the potential to leak or potentially introduce small amounts of air into the line. Customers with questions were instructed to contact ICU Customer Service Monday through Friday between the hours of 8:30 AM and 4:00 PM Pacific time: 866-829-9025 and select option 2, or e-mail the following address: productreturns@icumed.com. Customers are instructed to complete the attached Medical Device Recall Response Form and fax to (801) 264-1755 or e-mail to recall@icumed.com. Customers with questions were instructed to call 866-829-9025.
Quantity in Commerce270 units
DistributionNationwide Distribution
Total Product Life CycleTPLC Device Report