| | Class 2 Device Recall STERRAD NX Sterilization System |  |
| Date Initiated by Firm | January 09, 2017 |
| Date Posted | February 14, 2017 |
| Recall Status1 |
Terminated 3 on November 02, 2017 |
| Recall Number | Z-1197-2017 |
| Recall Event ID |
76194 |
| 510(K)Number | K042116 K142454 K151725 |
| Product Classification |
Sterilizer, chemical - Product Code MLR
|
| Product | STERRAD NX Sterilization System, Product Code: 10033, 10033-002 |
| Code Information |
Serial No 33100270 33110267 10033070599 33110166 33102348 10033070293 10033070296 10033070115 10033060166 33111764 33120476 10033050564 33130497 10033070381 10033070382 10033060659 33101036 33110266 33110308 33101084 33101087 33120355 33110445 33120379 1033160123 33101078 10033060377 33090564 33081028 33120585 33090532 10033060637 33130516 33110446 33110894 33111832 33111336 33111363 33111573 33111580 33110148 33101330 33120120 10033060898 33100023 33090562 33101545 33101573 33090528 10033070039 10033060706 33100367 10033050651 10033050652 10033050653 10033060178 10033070271 33110891 33120063 33111632 33081444 33081449 33111445 33101574 33120131 33130325 10033080299 33101333 33101352 1033142129 33130407 33111725 10033060230 33111710 33120162 33120165 33110448 33110454 10033080184 10033080185 10033080035 33101110 33101113 33101121 10033070022 10033070023 10033070026 10033070143 10033070144 10033070168 33110864 33110870 33101124 33101130 33090338 10033060582 10033070514 10033060285 33120537 10033070064 1033150476 33110057 33090500 33090510 33090515 33090524 33102301 33102359 10033070618 10033060837 10033060836 10033050216 33111413 33100063 33120518 33130559 1033142050 1033150314 33090190 33090192 33100316 33111679 10033070772 33110592 33100205 33100233 10033060811 33090428 33090397 33090404 10033050496 10033060620 10033060933 10033070075 33120294 10033080224 33110745 10033060510 33111674 10033050452 33102368 33120617 10033070889 10033070894 1033142096 33111551 33111579 10033080232 33120395 33111461 33111463 33111473 33111478 33111479 33111483 10033060502 33120721 33100483 33111084 1033150315 10033071045 33111277 33111280 33100208 33100209 33120481 33120482 33140044 33101457 33101458 10033060150 10033060275 10033080324 10033080325 33100422 33100460 33100471 10033070493 33100418 33081236 33101474 33101478 33101406 33110829 33110852 10033070673 10033070674 33090113 10033050252 33111028 33111040 33100448 10033080157 33110052 33100227 33110047 10033060702 10033060703 33111521 33120221 33102328 33100124 33101105 1033150350 33120121 33111499 33110836 33100400 33120217 33102213 33102216 33111974 10033080079 10033080084 33111604 33111517 10033060983 10033070664 10033070665 10033060665 33111258 33111264 33100330 33100332 33100534 10033070651 33111566 33120347 10033070748 10033050640 33110936 33111398 33111402 33110034 1033142064 33110327 10033070448 10033080313 33111393 33111405 33090360 33100078 33100084 33111382 10033060121 33081426 10033070368 33090001 33111874 33090438 33090460 33090463 33100336 33100372 33110196 33110771 33110743 33111523 33111854 10033060315 33081023 33081024 33101079 33101080 10033050647 10033080068 33120717 10033080023 33111200 33111457 33101000 33120073 33100518 33100522 33102330 33140025 33100053 33111454 33110045 33110050 33110054 33111446 33100037 33111895 10033070436 33120089 1033142060 33102134 33102135 33111356 10033080076 10033070751 33100378 10033070571 10033070108 33111671 33111676 33101562 33111105 33110197 10033070428 33102292 33102308 33102314 33102320 33102326 33101526 33110587 33120318 33102321 33110132 10033050184 10033050580 33100604 33100607 33111178 1033150345 33120102 33120104 33102285 10033050271 33100544 33110100 33110117 33110123 33110124 33110141 33111713 33120672 10033080348 10033070341 10033060913 33090334 33100167 10033070380 33111920 33110101 10033070182 33090277 33110300 33110301 33102204 10033080247 33110483 33110799 33110409 33110410 33130529 10033050238 33110150 33120729 10033060940 33090495 10033060558 10033070121 10033070122 33101650 33101658 33102115 33110820 33090177 33090212 33101138 10033050672 33120154 10033070394 33100272 10033060614 33081042 33111176 33111177 33111186 10033080066 33100090 33100094 33120169 10033060531 10033070348 33090261 10033050661 33101032 33101034 10033050630 10033050631 33110347 33120283 33120330 33101582 33111726 33102011 10033050185 33101639 33110343 33110381 33111111 33111113 10033060924 33090571 33100004 33100164 33110477 33110478 33110482 33110484 33111555 10033060184 33102139 33081201 33111389 33111400 33101307 10033060339 33100360 33100362 33101076 33110439 33120048 33120049 33101358 33101362 33100447 1033150061 10033070944 10033060731 10033060732 10033060734 33130049 10033070695 10033050528 10033050538 33110947 10033050434 33110393 10033060461 10033060463 33101054 33111915 33111932 1033150615 10033050287 33111424 33111916 1033141125 33110142 10033050542 33111662 33110498 33111734 33100356 33111727 33111559 33111694 33111698 33090549 33111203 33111227 33110081 33110654 33100166 33130015 33101604 33110536 33110550 1033150480 33100015 1033160326 1033160331 1033160333 10033070209 33090356 33100079 10033080091 10033050618 33101640 33130402 1033150303 10033070073 33101500 33130272 33130088 33081045 33130554 10033070987 33111701 33130400 10033071036 33101166 33110687 33110690 33110704 33130019 10033070667 1033150031 1033150043 33090083 10033060240 10033080252 33110572 33110596 33111414 33110900 33110901 33111051 10033060922 1033141023 33111441 33110928 10033070345 10033070776 33101123 33111898 33081354 33100238 10033050481 10033050426 33090454 10033070515 10033070518 10033060553 1033141004 1033142068 10033070437 10033070438 10033080033 33110268 33110378 10033080319 10033060872 33110639 33110676 33100424 33110416 33110406 33090489 10033050539 33111729 33120124 1033150018 33100317 10033060379 33111651 33100530 33111469 10033070171 33090517 10033060566 33100245 33110250 33110254 33102030 33081415 33081417 10033070435 33081386 10033060547 10033060779 10033060707 10033070457 33111796 10033070081 10033070085 33120423 33120648 33120386 33120391 33111347 10033050598 10033080065 33111668 33101446 33101055 10033060488 33102103 33100610 33102298 33120627 10033070883 10033080219 33111520 33130282 10033060519 33100482 33111884 33101613 33102121 10033060517 33120923 10033050414 33111689 33130143 33111919 33120463 1033150063 33090078 33100252 33110586 33090459 33110972 33110974 33101396 33101398 1033142112 10033070602 33090171 33110565 33110579 33100137 33101527 33111954 10033060283 33110231 33111819 33120191 33101575 33101579 33101584 33101588 10033070492 33130237 10033060358 33101131 1033160045 33120333 10033080295 33111041 33100619 10033060571 10033070446 10033060380 10033060677 10033070272 10033050160 10033060977 10033060978 10033060671 33102268 33102272 33120649 1033160252 10033060451 10033060452 10033050205 33120111 33120125 10033080249 33120082 33120396 33120526 33100606 10033060293 33090211 10033060342 33100171 33101179 33110118 33101019 33101040 33101471 10033070409 10033070421 10033060886 10033080292 33101310 10033060812 10033060819 33110456 10033070423 33100585 33111000 33100097 33100100 33100101 33100107 10033050574 33110027 1033160050 10033080115 33130106 33110553 33110279 10033080142 10033060613 33102288 10033050284 10033070410 33111860 33120043 10033060841 33120084 33101077 33110385 10033070134 33111233 33111255 33111261 33111266 33111268 33111290 33101069 10033070658 33110570 33110320 33110321 33111257 33111411 33111533 33111540 33111541 33111542 33111006 33111012 33111017 33110465 33110475 33110480 33110489 33100157 10033070373 33111492 33120197 33120589 33101511 33081307 33111353 10033070226 10033070227 10033060211 33090116 10033060353 10033060354 33120224 33130485 10033070600 10033060371 10033060372 33100218 33100265 10033060261 33111508 33090306 33102177 33102192 33081452 10033071029 10033080064 10033060980 10033080113 33110356 10033060202 33120293 10033060429 10033070795 33111760 33111779 33111939 33120503 33120505 33130437 33111217 33090179 10033060962 33111565 33111852 33111901 33090471 10033050654 33101482 33101484 33081035 33111090 10033070964 10033070965 33110352 33110547 33101134 10033070981 10033070583 10033070586 10033060952 10033060506 33090509 33090557 1033150071 33120777 33100123 33100152 10033060892 10033070767 10033070769 33081276 10033070582 33100304 10033060813 33111685 33110740 10033080105 10033080121 33120625 33120633 33120645 1033150040 33090080 33110772 10033060458 33111392 10033050555 10033060337 33130359 33101355 33111918 33120335 10033080040 10033050559 33111417 10033070089 10033060460 33102283 33120385 33110950 1033160037 1033160039 10033070623 33120530 33101139 33101140 10033070387 33101665 10033060758 33110915 10033050249 10033070118 33110326 10033071017 33111293 33111307 33111401 10033050576 33081188 33081189 33081191 33081192 33100085 10033070076 33110742 33120292 10033070400 10033070189 33110875 33110885 10033080318 33101663 33110739 33110768 33102362 10033060249 33102387 33110001 33110003 33110754 33110764 33110495 33090483 33120524 33111763 10033080134 10033060663 33111993 33110211 33101428 33101430 33110844 33110845 33110348 33130224 10033070749 33110632 33110642 10033060492 10033060209 33090429 10033070969 33081267 33090566 10033070973 10033060743 10033060497 10033060472 33110826 33111092 10033060956 33111886 1033160101 33100453 33101151 33100470 33120272 33120135 33120136 33110629 33101155 33120156 33110568 33100088 33100131 33100286 33100289 33100290 33120839 10033050472 33111391 10033060603 33090146 33130499 10033070396 33101045 33100206 33111956 33081303 10033050482 10033060500 10033060723 10033060729 33110190 10033070837 33130176 10033060144 10033060815 10033050613 33110225 10033071006 33111707 10033050405 33110715 33101435 33110424 33111681 33101120 10033060405 33102390 33110002 10033050644 10033070455 33120677 33100558 33100595 33100617 33111428 33111451 33111468 33111539 33111549 10033060970 33100089 33111718 33102303 33102324 10033050553 33102122 33102257 1033142082 33100242 10033050655 10033050656 10033080221 10033050552 33090543 10033070575 33090106 1033141072 33081124 33130266 33100629 33101005 33100413 33102214 33111439 33130466 33130495 33120147 1033150634 10033050648 10033070287 33102156 1033150100 10033080122 33110622 10033060252 10033060967 33100109 33100121 10033070722 10033050565 33111842 33100216 33100421 10033060514 33120885 33090120 33101342 33120047 33120055 33120056 33120065 33120146 33120161 33110006 33120845 10033070653 10033060345 10033060409 10033070062 10033070801 33111164 33111172 33111462 10033080090 1033141026 33110973 33110982 10033060288 10033060951 33102277 33110943 33081272 33100112 33100116 33100213 33100243 33110729 10033050411 10033070024 10033070032 33120354 33120398 33120452 33110303 10033060827 10033070429 10033070188 33090148 33090153 33102344 10033070785 10033070874 10033070878 10033060248 1033141102 10033060109 10033070179 10033070181 33081253 33101012 33130385 10033080327 33111083 10033050673 10033070454 33111484 33111491 10033070430 33101157 10033060616 10033050581 33081095 33110237 10033060818 33120242 10033070104 33100344 33110382 1033141007 33110353 33110367 33090073 33090565 33100001 10033070879 10033070899 10033070968 10033070862 10033070865 10033050692 33090010 33120487 33120790 33100132 10033080321 33110411 33101466 33101467 10033060512 33120251 10033060652 33100102 33090480 33101403 33110812 10033060281 10033060548 33110260 33120758 33090516 10033060137 33100025 33102173 33110033 33110455 33110458 33110883 33110888 10033070836 10033080203 33102392 1033160117 33111386 33100162 1033150078 33100237 33100284 33100302 33100307 33100308 33100314 10033060260 33102241 33120671 33081088 33100068 10033060201 10033060204 33111416 1033142072 33110636 33110451 33110453 33110457 33110459 33110463 1033160081 33120356 10033050637 33102194 33102206 10033060164 10033060181 10033060253 33110363 33110964 1033142165 33090393 33100269 33100273 1033150034 33101518 33102193 1033160139 10033070685 33120458 33081086 1033150112 10033070082 33120415 33120417 33120418 33120428 33110958 33110819 10033060984 10033070838 10033070839 33110961 33110975 33110978 33110981 33110984 33100567 33130506 33100276 10033060581 33081389 33100613 33101399 33111219 10033080107 33102116 33102119 33120502 33090143 10033050416 33111827 1033142167 33102003 10033071002 33081019 33100013 33100014 33100020 33111369 33111373 33130315 33102357 33102365 33102375 33111564 33110162 33110173 33111959 33100511 33100533 10033060247 33111429 33111148 33101175 10033080138 33101198 10033080204 33090355 33111282 33101492 33111253 33111498 10033070091 10033050483 33081428 10033060294 10033070513 33102049 33102050 33102067 33102070 33102080 1033142110 1033150095 1033150096 10033080130 10033080164 33101150 33102367 33111720 33111908 33090027 10033060389 33120725 33120732 33111283 33100577 10033070668 33120948 33101111 33090494 33090183 33111650 33102378 33101445 33102006 10033070544 10033060719 10033060720 33110857 33101637 33090132 33111859 33120635 33081145 33081265 33102304 33111548 33111669 33081133 33100226 33130162 33101100 1033150617 33120406 33081231 33100563 33100565 33100566 33100569 33100584 33101315 10033080214 10033080215 33140031 10033070815 10033070826 33111079 33130284 33111558 10033070576 33081367 1033142005 10033080043 33101366 33110505 33100295 33130484 33101135 33101136 1033160107 33110103 33110104 33111243 10033060945 10033060946 33110650 33102167 33101495 10033070634 33111567 10033070426 10033080223 33102383 33110843 10033070231 1033150026 1033150057 33120170 10033060462 10033060496 10033060503 10033060268 10033050659 33090488 33110837 33111138 33090467 1033150521 33081211 33081275 33081283 33101498 1033141090 10033050447 10033050450 33111465 33101408 33120166 10033070860 10033070260 10033050530 10033070388 33101423 33101424 33110134 10033070041 33100012 10033060770 33102053 33101326 33100409 10033070353 33102110 33102124 33120871 33140075 1033150080 33130465 10033050230 33081165 33120768 33081051 33111865 10033070402 10033070403 33102163 33111450 33090269 10033050279 33120112 33100229 33100253 33110851 10033050491 10033050495 33110510 33110652 10033050253 33110730 33102201 33102211 10033060959 10033070425 33101572 33111214 33130014 33100547 33090082 33081277 33120822 33101353 33110220 33110230 33120020 33111224 33111247 1033150111 33102278 33090089 10033070963 33101590 33110084 33102069 33102123 33120660 33120662 1033160119 1033160121 33130105 10033070393 33081230 33101507 10033070378 33110683 10033080220 33090367 10033070422 33130133 1033141012 1033142127 10033070867 10033060369 10033060574 1033150456 10033070342 10033070451 10033070463 10033050225 10033080298 33120491 33101002 10033080290 10033060947 10033060948 33090540 33090541 33090546 33090552 33090556 33120527 33120531 33120532 10033070222 33090493 33120182 33111904 33111800 33120590 33111784 1033150021 10033050480 10033080191 33101636 33100549 33111716 33120640 33111495 33111519 33101631 33110708 33101021 33101037 1033142047 33130232 33120681 33111708 33100455 33100463 10033060617 33120913 10033050611 10033050612 10033070033 10033080230 33111326 33111856 33111864 33100425 33102202 33111099 33111372 33110834 33110835 33111184 33111221 33120179 10033050493 33100153 33130174 10033060412 33102087 1033142076 10033070207 33110911 33111490 10033050440 33102195 33102207 33110523 33110530 10033070482 10033070483 33111406 10033060344 33100429 33102270 33101354 33081328 1033150457 10033060244 33100141 33111789 10033070612 33111774 33081043 33081085 33120150 33120152 10033060239 33101013 10033070848 10033070849 10033071034 33110223 33110242 33111743 33130470 33131000 33130167 10033070027 33111407 10033060359 10033060390 33100125 33100126 10033060404 10033060424 33110013 33110039 10033070367 33101532 10033070453 10033070456 33102120 33111683 10033080181 33110256 10033080060 33110571 33110573 10033050567 33102276 33102294 33101450 33102092 33120450 33120472 33120497 33120499 1033141159 33110500 10033060340 33100087 33100118 10033070487 33090502 33102208 33110282 33110283 33100046 33110516 33090008 10033060130 33090479 33111370 1033141139 33111949 10033050584 33140037 10033050575 10033080058 10033080059 33120905 10033060560 33110634 33110643 10033050534 10033070792 10033070835 33110628 33111096 10033050629 33090384 10033070978 10033060507 33110801 10033070782 33110335 33120144 33110630 33110697 33101525 33111069 33111097 33110396 33111620 33110716 33110233 33111488 33111493 33110417 10033060440 33090538 33120072 33120079 1033150474 33110823 33120042 33120060 33110585 33090022 33110905 10033060760 33111368 33110907 33111666 33100512 33120521 33111791 33110869 33110818 33111123 33101014 33101018 33100589 33100608 33101046 33101047 33120542 33120557 33120540 33120545 33081001 33120289 33111387 33111526 33111527 33100042 33102101 33101010 33120543 33120546 33110330 33111545 33101558 10033060214 33110470 10033060607 33101023 33110192 33100605 33111635 33111660 33111680 33102299 1033141083 33110567 33120130 33081405 33111633 33111649 33111100 33111672 33111183 33110621 33110638 33090322 33090325 33140017 33101493 10033080099 33101171 33111106 33111817 33101581 10033050486 33130227 10033060900 10033050536 33130507 10033060258 33110376 33110377 33110228 33120895 33111166 1033141117 10033070111 33102382 33102009 10033050671 33100066 33100067 33111630 33111637 33111641 33111788 10033050595 10033080141 10033060710 33111297 1033150004 10033070221 33100007 33102279 33102295 10033060931 10033060775 33111114 33111127 33130491 33111879 33090118 10033070038 1033150202 33120673 10033060876 33110043 33090443 33120613 33120651 33101653 33120365 33101489 1033150059 33102338 33100259 33120378 10033060941 10033050418 33110415 33110420 33101475 33100297 33101086 10033050561 10033080255 10033060388 10033050687 33120451 33111922 33101477 33130324 33111173 10033060776 10033060894 33111821 1033141063 33110388 33110474 33110383 10033070736 33111964 33120371 33090040 33110397 33111170 33101509 33100419 33120298 33110855 33110854 10033050624 33101412 33101417 33101638 10033070054 10033070106 10033070390 33110657 33110667 33111110 33111112 1033141014 33110532 33120233 33111043 33111048 33111550 33081027 33130334 10033050571 33120600 33120601 33102339 33102306 33081321 33110691 33100019 10033060232 33111470 33100072 33100077 33100081 33100082 33100083 33100093 10033060407 10033060949 10033060950 33110061 33110096 10033060976 10033070528 10033071043 33111654 10033060577 33100515 33100525 33102305 1033141008 33101331 33100504 33120652 33102309 33100573 33100578 33110522 33101668 10033060264 10033050639 33111677 33110882 33111538 33101554 33102054 33120285 33102379 33120186 33100134 10033060508 33120563 33110163 33120548 33111721 33140038 10033060117 33110827 1033141101 33110425 33102369 33090161 33110537 33110538 10033060907 33100542 33100342 33100393 33100474 33111237 33101536 33102133 33102251 33102253 33102329 33120460 33110384 33120611 10033080281 33110944 10033070047 33110839 33110841 33110351 33081081 10033080236 10033080240 10033060357 33110419 33090044 33130164 33111507 10033080336 33111228 33111675 33111695 33110925 33140052 33110932 33120569 33130329 33090346 10033060505 33090292 1033142065 33120929 33111055 33110295 33120171 33120349 10033060966 10033080245 1033160053 10033050569 33090164 33090165 33110442 1033160304 33120887 33111868 33110520 1033141097 33100230 33100462 33100521 33111806 10033050432 33101059 10033070013 10033070078 10033070132 10033070135 10033070139 33090130 33111693 1033141107 10033050577 33102358 33120238 33120370 33120373 33120375 33120376 33111306 33111360 33111361 10033070385 10033070590 33110130 10033070123 33110780 1033142138 33111013 33120078 33120083 10033070051 10033060865 33102010 33102012 33101065 1033160273 33110789 10033070735 1033160083 10033060599 10033080186 10033070141 33111862 33110701 10033050273 10033060686 33110673 10033070692 33120348 33120363 1033141103 33120512 10033070745 10033060787 33081025 33100055 33110311 33111656 10033070946 10033060755 10033060431 10033060432 33090451 10033060621 33100231 33100315 33120616 10033060509 1033141104 33090529 1033160002 33120483 10033060816 33111736 33111299 10033060218 33081418 10033050669 33100536 33120189 33101432 33101440 33090210 1033141067 10033070376 1033160189 10033050585 1033142025 33081323 33100398 33100407 33110817 33110832 33102236 33102343 33102350 33101405 33101422 33111776 33111016 33111034 10033070983 33110674 33102312 10033060433 33100532 33100548 33090242 33090244 33090248 33090296 33081134 33111031 10033060436 10033060348 33110733 33110791 10033050603 10033050604 1033150068 33110217 10033070605 10033070301 33110816 33110847 33130353 33100516 1033142069 33120952 1033150072 33081089 33120511 33120190 10033050514 10033070990 10033070017 33130321 33101058 33101066 33120434 1033141076 33110560 10033060835 10033060862 33111133 33110008 10033070029 1033142017 1033142018 33120455 33120909 33102170 33120533 33120549 33120938 10033050632 33120466 10033080280 10033060175 33101481 33081429 33110641 33111529 33111536 33130020 33100557 10033070494 33101407 10033060822 10033080053 10033070452 33100175 33111790 33120087 10033080051 33110702 33102286 33102342 10033050628 33081402 33101007 33101039 33090084 33111589 33111459 10033080284 1033160010 10033080317 10033070490 33111024 33111039 33120115 33090357 33100163 33102376 33120937 33102176 33120114 33130120 33100140 33111807 33110447 33120506 10033080050 33102282 33120046 33120067 10033071035 1033150467 10033050638 33110184 33110219 33111497 1033160082 33102275 33120951 10033050683 33100402 10033070433 10033060368 33130279 10033050544 10033070057 33102223 10033060777 33120826 33100169 33101567 33101371 33101521 33101571 33101370 33101379 33110794 10033071011 33101420 10033070273 33111688 33111697 33090504 33102189 33081423 33081431 33101448 33120544 10033071005 33120891 33110619 33110633 10033070270 33102063 10033070556 33110703 1033142093 1033142094 33101375 33101389 10033070628 10033070636 33130026 33120015 33111755 33111768 33101627 33101632 33101669 33101671 33120377 33120401 33120404 33102057 33101101 33101109 33111829 33111848 33111855 33110802 33110804 33111383 33100091 33102073 33120928 10033070524 10033070534 10033080106 33110924 33110986 33110994 10033080314 33110264 33111248 33111251 33090108 33111225 33111150 33111204 33111305 33111325 33111333 33120018 33120057 33120068 33120075 33111162 33111160 33111252 33120077 33100129 33100138 33130172 33111909 33101043 33101074 33081266 33111869 33111260 10033060227 33100048 33100086 33120838 10033050220 33120478 1033142162 33101380 33110618 33110625 33120295 10033070210 10033070840 33111902 10033070804 33120560 33120570 33102008 33120488 33120535 33111063 33111095 33102354 33111621 33090405 33101190 33102243 33130435 33101336 1033150448 1033150471 10033050442 10033050443 33101063 33101104 33090274 10033060712 1033141128 1033141130 33081006 33081007 10033060670 33130406 33102230 33102233 33130387 33130392 10033050431 10033070694 33111730 33111738 33111050 33111065 33130273 10033060638 33101595 33101596 10033070228 33102060 33130468 33081048 33130428 33101082 10033060467 10033060468 33101195 10033080225 10033060884 10033060889 33130113 33130116 33130349 33130013 10033070298 10033070299 33110942 33110965 10033060583 33090295 33090109 33100311 10033070690 33102370 33081148 33101030 33130182 33130219 10033070434 33120906 10033070578 10033070579 1033160231 33110960 33120529 1033160145 1033160126 1033160133 33130257 33130262 33120901 33130366 33120794 33120808 33101649 33101112 33140047 33101351 33120835 33120789 33130293 33130154 10033060642 33120496 33120797 33110993 33110995 33110939 33110940 33120829 10033070758 33100553 33110099 33102093 33120700 33102079 33111899 33110783 33111851 33081247 33101662 33102098 33130388 33111858 33130557 33102142 33102157 33102162 33110562 33110566 33110574 10033071026 33130458 10033070761 33090353 1033142044 33110989 33110990 33102384 33101431 33120631 33101626 33130122 33120889 33120910 33081047 33120414 1033142119 1033142121 33111778 33111822 10033050502 10033070391 10033070392 10033071031 33120615 10033050500 10033070828 33101303 33101304 33130235 10033050448 10033050449 10033060736 10033060737 10033060738 10033060739 10033060742 33111422 1033150444 10033060651 10033060602 10033060606 33111867 33090358 33100572 33102113 33102114 1033150407 1033150412 10033060746 33101038 33111765 33110286 10033070236 10033070237 33102219 33102240 33081241 33081259 10033060669 33081466 33090009 33090323 33090328 10033060668 33101578 10033070759 33110709 33110751 33110696 33110731 33110750 33110760 10033050503 33111581 33111612 33111437 33111449 33110921 33110949 33111066 33111958 33081187 33081208 33081216 33081178 33081207 10033070147 10033070148 33102186 33102203 33110241 33110249 33110265 33101332 33101335 10033050217 10033050218 1033142013 33110261 33090513 33090551 33081167 10033060955 33130080 33102289 33102300 10033060212 10033070621 33110903 33110910 1033150402 33101564 33101577 33110860 33111664 10033060635 33111485 33130417 33111438 33110637 33081166 33102291 33102346 1033141105 33101625 33101630 33101634 33101652 33102022 33102071 1033141084 33110685 33110755 10033060443 33110177 10033060714 10033070450 10033050617 33120767 33090560 1033142085 10033060469 33090293 33110861 33110878 33101115 33101128 10033070649 10033060580 33081150 33081183 10033050248 33110501 33100029 33130138 33101340 33101344 33100612 33100615 33110476 33111758 33120160 33102055 33090523 33100382 33100456 33100481 33100493 33110744 33120528 33100348 33130043 33102014 10033070661 10033070662 10033080103 33111936 33090342 33100172 10033071016 10033071018 33101576 33101556 33102255 10033070035 33120098 33120388 33100329 10033060295 33110748 33101156 33130270 10033070886 33081310 33130053 33110886 33111980 1033150427 10033060272 33100054 33101486 33090458 1033142134 33111780 33102302 33130096 1033150308 33120422 10033050219 33101397 33090087 33110624 33110644 33130206 33110270 33101529 33120907 33120234 33110243 10033080274 33111955 10033070888 33120232 33110133 33120749 33111088 10033071024 33110316 33111244 33111728 10033070469 33120880 33110588 33110059 33111872 33120592 10033060532 33111544 33111546 33111547 33111610 33111646 33110452 33102171 10033070223 33120469 33111395 33090490 33110931 33100510 10033070624 10033070431 33111715 33111394 10033070825 33101356 10033050579 10033050660 33111830 33090015 33090026 33120133 33111737 33111769 10033081260 33090147 33120690 33102313 33130386 33111808 33100293 10033080086 10033080092 10033080108 33110172 33110174 33100192 33100379 33110340 33111238 33111239 10033070401 33111570 1033150201 33110859 33110966 33110983 33111652 33101173 33101006 33120268 33081406 33111474 33140010 33130087 33130511 10033071004 10033060100 10033060934 33111608 33120149 33101552 33110185 33110391 33110164 33130146 33111124 33101603 10033070055 33102317 33111818 33090206 33111943 33111957 10033070553 10033060817 10033050439 1033141113 33100391 10033050183 1033160279 33111025 33102181 10033050188 10033080093 1033141055 33111431 33130097 33110193 33110214 10033050469 10033060267 33100183 33140069 10033060996 33100256 1033150418 33110036 33130339 10033060578 33101008 33110491 33100154 1033142103 33100056 33100062 33111838 33101176 33101177 10033070449 33102239 33101655 33101387 33110795 33102315 1033142111 33130225 33110608 33110627 33130371 33110514 10033050463 10033060262 33111642 33140068 33120352 33110293 33110306 33100032 33100033 33130338 33130390 33111276 10033050550 10033070072 33081040 10033060251 10033070783 33111590 33111464 10033070619 33110635 33110468 33102196 33110485 33110488 33101553 1033150509 33101479 1033142058 33111249 10033070355 33110106 10033070346 10033070659 33110540 10033050560 10033060101 33110395 10033060997 33110433 33111794 33111801 33102293 33100537 33111659 33110761 33090352 33120576 33130191 33100592 10033080256 10033080257 33111618 33110757 33110765 33110421 33102310 10033070292 33090060 33081123 33110332 33100556 33102261 33111824 10033070060 33120646 10033060480 10033060481 10033070760 33130076 1033150334 33140019 33090264 10033070214 10033070216 33120525 33101053 33111032 1033142071 33081286 33081287 33081222 33081223 10033050446 33110756 33130509 33101009 33111584 1033150035 10033070522 10033070414 33110735 33110737 33110738 33110741 33110781 33110784 33120837 10033070399 33102385 33102391 33090464 10033070595 10033080096 33100008 33120119 33100408 33120421 33111963 10033080250 10033060493 33081443 33102263 1033142008 10033050291 1033150200 33100491 33110479 33110713 33100377 33100396 33101328 33111271 33111296 10033060797 33100300 33120168 10033060170 33081082 33110449 10033070398 1033150320 33110831 33130103 1033141074 10033060489 33100005 1033160047 33120520 33111640 1033142140 33100325 33081383 1033142164 10033071015 10033071019 33100185 33101162 33102371 1033150090 1033160169 33110354 33130476 33130512 10033070580 33111731 33110645 1033142049 1033150332 10033080126 33100538 33120691 33100350 10033060158 10033060208 33130021 10033050460 10033070256 10033070309 1033150400 1033150409 33100145 33100150 33110276 33110357 33102015 33110821 33110272 33110275 33130528 33110494 33090107 33100319 33102025 10033060762 33120222 33120277 33110392 33111408 33110782 33111458 1033150076 1033150446 33120612 33100387 33090548 10033080154 10033080155 10033050668 33081038 33110178 33101347 33120201 10033080243 33111471 33120710 33110373 33100574 33110705 33081217 33081218 10033070288 33110649 33110808 33111723 10033070340 1033150359 1033150452 33081373 33081374 33102191 33101441 33081421 33100207 33120301 33120827 33120831 33102323 33101364 33111970 33110919 10033070729 10033050467 10033050479 33111350 33111355 33101566 10033060242 10033050456 33111648 33101510 33090275 10033070725 33130451 33090332 1033142041 33111600 33111602 10033070213 33101003 33101197 33101560 33101569 33111883 33110450 1033141075 33111930 33110102 33130201 33110389 10033050180 33111781 33101142 33111364 33111143 10033060479 33120240 33111015 33090339 33101551 33081302 33110235 10033050594 33111607 33111609 33090536 33111117 10033070046 1033150414 33102086 33102088 33110695 33081457 33110858 33110481 33110215 33081227 33111950 10033070873 33111799 10033080145 10033080172 33100181 33111103 33110058 33111960 33111379 10033070688 33111748 10033060349 10033060925 33102325 33111873 33101041 1033150493 1033150494 33130028 33111686 1033150048 10033060188 33110236 33110240 1033141029 10033050304 33120678 33120862 10033050451 33110545 10033070771 33110497 33110503 10033071014 33090380 10033080003 33100554 10033060590 33081233 10033060717 33120336 33110576 33110575 33110584 33100579 33111134 33111154 33111655 33102143 33110205 10033060485 33102377 1033160087 1033142062 10033060385 33102126 33110953 33120832 33111531 33110331 33100065 33102271 33102131 10033070962 33110758 33100369 1033142023 33110048 33110991 33110992 33101341 33101350 33120924 33100070 1033160019 33110578 10033080320 33110908 33120420 33090348 33110557 33111543 10033080162 33111494 33110871 33111927 33111762 33101400 10033060259 10033070967 33120638 33090019 33100220 10033070014 1033160028 10033050578 33120471 33120062 33101565 33110009 33090330 33100099 33110595 33130494 10033070137 10033070138 33101346 33120437 33111093 1033160091 33110822 10033050302 33111286 33110917 1033150413 33111983 10033070008 10033070020 33100635 33101057 33090512 33111320 33102158 1033150013 1033142010 33111568 33111571 33101160 33101165 10033070208 10033060286 33081422 10033070065 33130238 33111583 33130566 1033150628 10033060562 33111085 1033150432 33120307 33140073 33100570 33120878 33101555 33102220 10033080229 1033150305 33120783 33120430 33110591 33120024 33111678 33110807 33111853 33120245 33110299 33120045 10033050621 33101531 33101547 33101548 10033060716 33102179 33111207 33111213 33110496 33110509 33120515 33110933 33110954 33101539 33101546 33101561 33090189 1033142021 33120128 33120818 1033142113 33110531 1033160023 33081220 33081274 33101384 10033080054 10033080056 10033080170 10033080173 10033080174 33110798 33110803 33090101 10033060656 33111574 33110151 33120108 33100224 33081197 10033050240 33090048 33130051 10033060958 33081034 33081409 10033060757 10033060634 33100035 33100038 33081176 33120663 33101385 10033050646 1033150065 33100219 33110846 33110874 10033060765 33120894 33120915 10033060106 10033060149 1033141077 10033070183 10033070887 33100119 33111109 33130031 10033080160 33111724 33111746 33081037 33111074 33111075 33111087 33120315 10033060417 33101508 1033150443 33130009 33130036 33120425 33081346 33111180 33111182 33110129 33111340 33120776 33120006 10033060328 10033060329 33100340 33110753 33110773 33101020 1033160263 1033160140 33101476 33111968 33120138 10033060226 10033060363 10033080089 33130355 33081232 10033070440 10033070441 10033070442 10033070443 10033070764 33102017 33120095 33130510 33120795 33081420 33090347 33090525 33102061 33111086 10033070606 10033070608 10033060476 33111304 33111582 33111115 33120059 33120508 33110011 33110014 33110521
10033060455 10033060456 10033060457 33110427 10033070716 33101334 10033080246 33111880 33111900 33111512 1033160288 33110018 10033080041 33110675 33100451 33111430 33111448 33111466 1033141109 33110372 33110380 33102160 33110274 33120900 33111575 10033060490 10033060495 33100098 33110229 33110209 33111528 1033142014 33120176 33120183 33110212 33110234 10033070411 10033070412 33081130 33081156 33120493 1033150060 33090005 33120926 10033080032 10033070379 33102345 10033060612 10033070743 1033150025 1033150029 10033080244 33102353 10033070996 10033070997 33081224 33081228 33090531 1033150205 10033050424 33120916 1033142001 33090121 33111700 1033160017 33110564 33110614 33120207 33120213 33110095 33102254 33090287 33120090 33110434 33111223 33120303 33090024 33100143 33100146 33100147 33110198 33110221 33102327 33120329 33110239 33120076 1033160147 33130255 33101447 10033080042 10033080045 33102212 1033150077 10033070354 33120116 33101094 33101129 33090227 33101543 33101549 33120106 33100135 33110776 33081292 33081298 33110065 33110066 1033142024 33100003 33100024 10033070128 10033080067 33081388 10033060630 33120139 33081126 10033070224 10033070225 33111222 10033070966 33111475 33101557 33110806 33110815 1033150301 10033060320 33110153 10033050582 1033142148 33111500 10033070912 10033060391 33110569 10033070413 10033080303 10033060633 10033060713 10033070432 33120945 33130331 33100497 10033070527 1033141017 33101068 1033142038 33102316 10033050269 10033070488 33111562 33111585 33102205 33081404 1033141045 33111717 33081395 10033050237 33120367 10033050688 33100201 33102013 33111053 33090412 33090340 10033050415 10033080104 33120436 33120440 33120467 33110407 33110408 33081193 10033070486 33110194 33090554 33100080 33110853 33120035 1033141153 1033150618 10033070903 33090337 1033160298 10033050173 33110526 33111881 33102164 33100039 33102097 1033150062 10033070100 33110736 33110717 33140001 33110374 33100411 33111631 33110170 33110176 10033050428 10033080293 10033070864 33110161 33110175 33100071 10033050182 33102165 33100052 10033070067 33111722 33102141 33110607 10033050215 10033070120 10033070489 33130498 33081044 10033070360 33120013 33120014 10033070185 10033071010 33102033 33110077 33140066 33111825 33111837 10033050589 10033050614 10033050615 33110631 10033050529 33102166 10033070142 10033060314 33111636 33110517 33102389 33110258 33081467 10033080241 10033070871 1033150104 33111167 1033141021 33120151 33120153 33102048 33102052 33102068 33111187 33111196 33110796 33110814 33101305 33120039 33100309 33110049 33111335 33100262 33130431 33130445 33100075 33090506 10033050517 33110309 33081091 33090272 33090425 33120586 10033050407 10033060343 10033070119 10033060780 10033050260 10033050264 33090217 10033060645 10033070869 33110778 10033070205 33101419 33090256 33130110 33120622 33140002 10033070289 33120707 33120712 1033150639 10033060217 33100246 33100261 10033060243 33081174 33081175 33100423 33101193 10033070988 10033071025 33081390 33100263 33100280 33111840 33111530 33111534 33111537 33120500 10033060796 10033060808 10033060810 33100590 33111703 33110307 10033060465 33111516 33110677 33110684 33110698 33111316 33120462 33100151 10033060918 10033060919 33090007 33111419 33102083 33101394 33120080 33111313 33110790 33110556 33081229 33081235 33090263 10033060783 10033060784 1033150206 33090074 33110976 33111882 10033050607 10033060722 10033080182 33111071 10033060331 33120264 33101580 10033080101 10033080128 33090514 33090537 10033060998 33090508 1033150044 33111349 33110749 10033050406 10033070254 33090058 33090081 33111976 33081438 33081446 1033142160 10033050433 10033060302 33110767 10033070680 10033070686 33101585 33081022 33102222 33111850 33110191 10033070594 33081132 1033141020 33111091 33111615 33120094 33101623 33101667 10033070389 33120575 10033070090 10033070092 10033070093 10033070249 10033070617 33102264 10033080052 33120577 33120269 10033070757 33111080 10033060475 10033070265 10033070775 33120297 33090045 10033070095 33130082 1033141062 10033050470 33102153 33102154 10033050665 33120339 33110430 33120647 10033070881 10033050461 33120921 33100139 33110017 33110022 33110032 33111925 33110512 10033060215 33110490 10033060400 10033070408 33111136 10033050508 10033050566 33090570 33110759 33110824 33102020 33102095 33102100 33120644 33101343 33101349 1033150022 33120142 33120791 33090199 33100160 33100165 33120054 1033150405 33101088 10033060271 33111749 33101372 10033070275 33060910 33102386 10033060938 33111705 33101022 33120723 33120628 10033060573 33120882 33130135 33130171 10033050280 10033070152 10033070154 10033070255 33120942 33140035 1033150017 33110111 33101418 33111456 33110091 33110113 33110527 33110726 33130125 10033050693 33081297 33101308 33090567 33100015 33100106 33100120 33100122 33110710 10033060831 10033080098 33120479 33110626 33110387 33090455 33140028 10033080347 33120044 33081039 10033050554 33101016 33101309 33110653 1033160188 33102249 33102265 33111910 33111946 33110714 10033050658 33100250 33110418 33110423 33110429 33111861 33081087 10033070525 10033050468 10033050404 33110662 33110046 33090499 33090521 33100492 10033070655 33090452 10033060709 33110800 33100618 33100627 33102199 33100517 1033150641 10033050400 33130151 33130111 33110359 33110364 33110323 10033070063 10033050583 33100626 33100279 33100306 1033142152 33111037 33100432 33100434 33100173 33110315 33120023 10033060477 33111965 33102347 33120863 33120290 1033141042 33130474 10033071021 10033070116 33110507 33111108 33111661 33110920 10033060564 33102175 10033070037 10033070195 33111739 33110524 1033150421 33110317 33130187 33110262 33110294 33110203 10033071030 10033071042 33110539 33140020 33081029 33111380 1033142161 33110548 1033150322 33090290 33100182 33111049 33101537 33101586 33110259 33111875 33111878 33111314 33110541 33111375 10033050573 33081324 33101118 33110360 33111014 33111876 33130023 1033142046 33090349 33100496 33110412 33110064 1033150472 33111947 33111951 33102081 33111321 33120299 33111797 33100161 10033070418 33120053 33111525 10033060161 33081225 33131007 33140041 33101106 33101108 33102281 10033060609 33110838 10033080202 33101083 33110840 33111696 10033080258 33100459 33120123 33130332 10033050429 33101395 33101402 33110350 33110082 33111232 33102210 33101499 10033070383 10033070646 33101464 33101523 33111578 33111616 33111913 33140067 33100277 33110206 10033060256 33120374 33111302 33111315 33100047 10033070114 33110946 33111002 33090470 33120004 10033061020 33120489 33102351 33110128 33111557 33100397 33081000 33110672 33111687 33110213 33110227 10033060769 33090426 33110181 33100616 33100620 33100632 10033060789 33110938 10033050587 33120117 1033141156 10033060592 33101589 33101661 33120103 33120140 33102077 33102109 33081380 33120641 10033050245 1033141106 10033050623 10033070626 33110828 33110525 33110529 1033141114 33120041 10033080095 33111563 33120605 33111785 33120051 33081075 33101633 33102107 33111052 33111062 33101456 33110848 33110922 10033070034 33090294 33101382 33101485 33110361 10033060971 10033060972 10033060973 10033050562 10033050563 33111554 1033160235 33110877 33110896 33120237 33120243 1033142166 33101516 10033060501 33110656 33110862 1033150003 33111288 33110339 33111759 33100176 10033080216 33120334 33120875 33100500 10033071023 33120840 33120884 33111135 33111159 10033060942 33130291 33120081 33111269 33111560 1033160269 10033060159 33111064 10033060675 10033060676 10033060567 10033060664 33130163 10033080120 10033080218 33100417 33100502 33130354 1033150307 33102341 33130454 1033160063 10033080272 33110333 33130367 33111740 33102274 33120132 33120157 33110310 33110314 33110313 33110329 33120311 10033070297 33100339 10033070459 1033141001 1033141047 10033060944 33110904 33101017 33111535 10033060579 33081365 1033150330 10033060618 33100485 10033050597 10033060304 33101359 33101666 33101672 33102136 33102137 33110159 33110473 33111094 33110207 10033050228 1033160282 33120773 33120925 33101159 10033070187 33111510 33110912 33111810 33111834 33100180 10033060682 33100395 33110083 33090424 33110956 33110583 33110647 10033080306 10033080309 33101095 33101114 10033060119 33102065 33111415 33111142 33111144 33111146 33111149 10033050641 33100634 33120209 10033060608 10033060766 33110766 33131001 10033080080 33110724 10033060839 33081312 33100144 1033150348 33110298 33110616 33110884 33102032 33090547 1033142077 33111524 33130248 33130269 33130330 10033080062 10033060711 33090376 1033141079 33101325 10033060376 33101164 33120536 10033050568 33090354 33111967 33120196 1033160255 1033160271 33110328 33110345 33111741 33120595 33111426 33111432 33120220 33120034 33110131 33110137 10033060228 10033060330 33120096 33090194 1033141030 1033160021 33110543 10033060849 33100111 10033060632 33110544 33110546 33111682 10033060105 33110926 33090226 33120353 10033060112 33110365 33090028 33100443 33102258 10033060257 33120614 33120184 33100249 33110934 33130001 10033050626 10033080278 10033080279 33102355 10033080158 33110769 33130239 33102159 33110438 33110394 33101540 33120629 10033060185 33111991 33120185 33110202 10033060419 33101490 10033060347 33120218 33130374 33111396 33120387 1033141129 33120716 1033141118 1033150318 33102161 33120464 33130186 33101598 33102064 33102074 33120670 10033070670 33100359 33110182 1033142059 33101622 33110168 33101587 10033070162 33111966 1033150204 33110370 33110461 33090173 10033060449 33120431 1033141137 33120109 33100215 33110375 33120893 33120908 33102172 33102188 33110200 33110216 33110609 33110155 33120306 33120877 33110089 33110015 33110999 33111019 33101125 10033070816 10033070472 10033071000 10033071001 33090266 33120143 33120145 33120174 33120753 33130072 33130073 33130077 10033060943 33110020 33110025 33110029 33081387 33111706 10033080110 33090053 10033080282 10033050591 10033060118 10033060622 10033070744 1033142035 33101388 33130249 33111890 33120852 33130044 33102296 33110145 33111911 33100479 33110842 33110850 10033060350 10033070343 10033070344 33130320 33110996 10033070444 10033080024 1033150041 33100475 33081403 33100374 33100376 10033050645 33110941 33110273 33110188 33110519 10033050605 33120017 10033050208 10033071027 10033070980 33120215 10033070098 33110079 33120227 33111653 10033050455 33120380 33120383 1033160111 1033160178 33101488 10033060441 10033060447 33090498 10033050609 33111102 33110440 33110504 10033080135 33120085 33111897 33081308 33110528 33120040 33120432 33120746 10033060923 33090111 33101345 33081015 33081016 33081017 33081190 33081194 33081138 33081327 10033060897 33110085 33110107 33130074 33130198 33120223 10033050261 33102388 33120922 10033060266 33111036 33111503 33090351 33110278 33111888 33090315 33110833 33110927 10033070110 33101338 33110414 33110342 33110349 33111030 33120344 33120514 10033060658 33120634 10033050201 33111941 10033060809 33101191 10033060176 33090535 33090539 33090544 33090545 33110090 33111337 33111804 33100232 33100234 10033060470 1033150326 33090071 10033070468 33110918 33100203 33120587 33110499 33110502 33102259 33090427 33110722 33100414 33120313 33111378 33110660 33110669 1033142043 33120208 1033160024 1033150343 33111752 33081369 33130490 33100105 33100110 33120309 10033060975 33100524 33100546 33090235 33110688 33110693 33111847 33110105 10033070478 1033160034 33110549 33110552 33120510 33110098 10033070473 33100050 33110063 1033150101 33090416 33130149 33130560 1033160007 10033060801 33111893 33102221 33102238 33120784 33100095 33100096 10033071013 33111767 33110792 10033060317 1033150082 33111384 1033141022 33101044 33102169 33110623 33130147 10033070476 33120517 1033150637 33081382 33120230 10033050419 33090175 10033070025 1033160141 10033060561 10033080335 33120212 1033150512 33110238 33110895 33110897 33110062 10033050596 33110810 33101544 33090518 33120624 33111914 10033060383 10033050494 33100318 33100335 1033141133 33090520 10033070763 10033080156 33100495 33100539 33111761 33102246 1033142135 33111587 33130119 10033070779 33110493 33120760 33090533 33111151 33111152 33120853 33101337 33120210 33100187 33110344 1033141085 10033070283 1033150020 33081366 33101425 33101438 33101439 33110977 33120618 33120655 33120661 33110218 33100239 10033070741 10033070726 33102322 33102333 10033080300 33090167 33081144 33120620 33111072 33081399 33090492 10033060586 33101444 1033160281 33120164 33081177 33101376 33102247 10033050570 33120604 10033050436 10033050437 10033060937 33111622 33110288 33090466 10033060162 10033070304 33110297 33081296 33100520 10033070357 33120093 33050422 33050532 33120052 33110004 33110005 33111504 33102335 33102374 33140008 33120578 33102174 33111969 33111948 33111365 33101427 10033080063 10033060234 33111952 33120771 1033142104 33120172 10033060154 33111891 33100435 33110777 1033142056 33100267 33110097 10033050263 33100247 33120603 33090374 33090377 1033150300 1033160312 33101119 1033142172 33090561 33140071 33101559 10033080028 33111409 33111691 33111692 10033070551 1033160118 33120199 1033141134 10033070244 10033070352 33081118 33090017 33110195 33111514 33130515 10033050441 33110399 33110880 1033142086 10033060265 33110167 33120321 33111511 33120927 33090558 33100368 33111617 10033070516 33140034 1033141027 10033060173 33111828 33120602 33120358 33120173 33081285 33120757 33090344 33130169 10033070549 33120276 1033150046 1033150051 33100278 33110706 33110707 33110312 33090163 33100603 33120092 33111663 33111665 33100545 33110204 33081391 33120288 33101487 10033060386 1033150094 10033060908 33110322 33101673 33111433 1033141053 33100158 1033150008 33090527 33090534 33100288 33100353 33090260 33101329 1033160089 33120726 10033070477 10033080333 10033080334 33111501 33120562 33101188 33110617 33110640 33120742 33111836 10033050489 10033070404 33110969 33100346 33100092 33101628 33102198 33090283 10033060192 33110135 33110171 10033070972 33130226 10033060681 1033150070 10033070083 33100602 33100225 33100294 33102231 33120591 33111496 33110680 33110682 33120141 33110797 10033070555 33110208 33090465 33140004 33100380 33110422 33100594 10033060526 10033060963 10033060964 10033060965 33101473 33110937 10033060826 33130117 33090043 1033141011 33111935 33100320 33120897 10033070465 33102197 33102200 33081050 1033160125 33100305 33100555 33110338 33110511 33110515 33120881 10033070898 33090006 10033070537 33111826 33140103 33110108 33110109 10033050213 33110443 10033050236 10033060995 1033142171 33130537 10033070984 33120091 33100235 33110558 33110561 33100128 33100494 10033060522 33090023 10033060520 33090511 33120552 10033070484 33101170 33100168 33100174 10033080189 33140043 1033150019 33110721 33101324 33101360 10033050649 33110464 33110469 33110472 10033050619 10033050620 33090102 33111753 33111803 10033050627 10033060207 33120319 1033150091 33102267 33100006 33100331 33130070 33110224 33111329 33130268 1033160085 10033050556 33081412 33081080 33081234 10033060689 33111561 33130514 1033150351 33130327 33100550 33120074 1033150356 33081009 33100312 33100069 33111553 1033150103 10033080070 10033050610 33100275 33101072 33120320 33090385 33090563 1033141078 1033141147 33120899 10033050202 33110762 33111814 33102232 33100559 33130508 33120192 33081020 33102180 10033060932 33111777 33111611 33111614 33102307 33130426 33130535 1033142124 33081008 10033060575 33120070 33100430 10033070074 1033160036 10033080222 33110988 33100301 10033050459 10033050430 33090406 10033050164 33101472 33120574 1033142169 10033050492 10033070151 10033070993 10033070995 10033070308 33100343 33100364 33120286 33130503 10033060466 10033070384 10033050535 33111823 33111846 33081282 33111843 10033050423 10033070660 10033071009 10033060237 33111073 33110139 33110149 33111502 33120551 10033050179 10033060225 10033080088 33110923 10033060313 1033142002 10033050181 33110651 10033050590 33111613 33120816 33120163 10033050408 33120366 10033060471 10033050662 33111711 33110222 33130504 33100334 33120302 33081439 10033070861 33081149 10033060128 10033050438 33111833 10033050622 33111667 33111733 10033050409 10033070951 33130318 10033060103 33090225 33101075 33100583 33140030 33110603 33110086 33110088 33100337 33110431 33110436 33120785 33111629 10033050681 33100581 33101361 33101414 33120538 1033160044 10033060190 10033050642 1033150623 10033050664 10033060205 33111658 10033070030 33110435 33110881 33102280 33102287 10033070215 33090400 10033060800 10033060565 33111434 10033060450 33101194 33071033 10033060148 33111940 33100399 33120275 33102360 33110080 10033060195 10033060591 33120439 33081180 10033060569 33110092 1033150408 10033060274 33111747 10033070204 10033070261 33100505 33130024 33130030 10033060241 33111742 1033141018 33111601 33111699 10033060610 10033060751 10033060920 33100142 33100159 33120155 10033060312 33120720 10033070206 33120896 33120940 33111619 33111141 33120203 33140074 10033060220 33120328 33100040 33111586 10033080131 10033070109 33090061 33110699 33110513 33090233 10033080326 10033060559 10033060426 33081427 33110720 10033060986 10033060899 10033060473 10033060478 10033060600 10033060438 1033141016 10033060985 33111397 10033060688 10033060434 10033060957 33120382 33130314 10033060735 10033060136 10033060604 10033060605 33081414 33110060 33090140 10033060572 33111772 33090526 10033080171 10033060300 33110658 10033070547 10033060654 33100523 10033060631 33102332 10033060636 10033050633 10033070002 33110051 1033150619 10033060292 10033060756 33100328 33130132 33101163 33111588 10033060867 33111487 33102334 33120865 1033142118 33111961 33081239 10033060961 33111744 33111757 10033060991 33110355 1033141048 33090530 33111657 33120555 10033070885 33111751 33111820 10033060648 33111917 33111921 33111923 33111931 33111944 33100506 33100526 33111509 1033142170 10033070080 33111844 10033070133 10033070086 10033070088 33081440 33111945 10033070259 10033070601 10033070050 33111235 33100486 10033070191 33100076 33110154 33120134 33081325 10033070070 10033070071 10033070285 33110805 10033070170 1033160234 33120101 33100281 10033070977 33120066 33120474 10033070286 33120284 10033070235 10033070211 10033070806 10033070397 33120010 10033070155 10033070178 10033070683 10033070691 1033141086 33111905 33111912 10033070370 10033070371 33110346 33100130 10033070068 33111877 33110963 33101189 10033060767 33100527 10033070842 33100155 1033141046 10033070587 33090029 1033150050 33130391 10033071022 10033070467 10033070581 10033080087 10033070471 10033070548 10033070554 10033070474 10033070979 33100027 10033070552 33111513 1033160156 33100254 10033070625 33090012 33050413 10033070872 33102155 33111098 10033070656 33100338 10033070607 10033070369 10033070374 10033070645 10033060960 10033070844 10033080081 10033070689 10033070882 10033070884 33110016 33110019 33090059 10033080139 33120381 10033070765 10033070766 33111815 33111046 33090282 33102096 33111505 33111467 33110898 33110390 33120750 33120763 10033060905 10033070015 10033070058 33090396 33110559 10033070875 10033080217 10033070740 10033070880 33111089 33130022 10033080242 33100375 10033071044 10033070604 10033071020 33130526 10033071032 33111137 33120828 33110140 33110462 33110471 33110487 1033160033 33120912 33102284 33100291 33100292 10033070084 10033070087 10033080114 10033080100 10033080111 10033080102 33081002 10033070059 33120105 33090219 33090569 33111343 1033141031 10033080009 10033080226 33120226 33100351 33110362 10033080213 33120278 33111231 10033060888 33120346 33081030 10033070991 10033070992 33120071 33100352 10033080308 10033080287 33100061 33101506 33101520 33120485 10033080304 10033080069 33081090 33090030 33101187 10033080332 33111603 33081004 33100326 10033060794 33120334 33081059 10033070947 33100406 33081041 33090365 33081153 10033060909 33081084 33120920 33101374 33081076 33111745 1033150056 33120359 33110492 33110580 33081181 10033070526 33111812 33111953 33111962 33110466 33110467 33100236 33100244 33111029 33111719 33100394 33081005 33090278 33081273 33111906 33101323 33101327 33081396 33081264 33081186 1033142168 1033150341 33090301 33110460 33100323 33100373 33081408 10033050606 33081196 33081281 33110793 33081127 10033080316 33110711 10033070824 10033070834 33110506 33110589 33090020 33101496 33090127 33090129 33120919 33110879 33110902 33110210 33081456 33090021 33130361 33081351 33120473 33120486 33100370 33081416 33081243 33130323 33081413 33101429 33081393 10033080082 10033070846 33090117 33090205 33110615 33140029 33101449 1033150451 33081419 33090103 1033160015 10033070892 1033141050 33101015 33090049 33090368 10033050676 10033070042 10033070113 33110809 33090016 33100296 1033150333 33090553 33090159 33111145 33130444 33081353 33102118 33100366 10033060629 33100043 33090119 33090142 10033050258 33090232 10033050172 33090257 33090258 33090259 33090262 33110169 33140061 33111569 33111576 10033050161 1033160248 1033160249 33090187 33120801 33120158 1033160208 1033150632 1033160237 33130183 33130175 33111246 33100170 33090418 33090141 33070361 33090198 10033060124 33120803 33090208 33090213 33090178 33102245 33090218 33100551 33111849 33110040 33090409 33100287 33100324 33100509 33090184 33111522 33090284 33100431 33090297 33100009 1033160060 33110534 33081131 33111984 33090559 33111208 33130104 33130221 1033160220 33090414 1033150069 1033142070 33111841 33100104 33090419 33130005 33130016 33110951 33110970 33110971 33120148 33110232 1033150516 1033160222 33101533 33100433 33090568 10033060454 10033060715 33100358 33100333 33120626 33100454 33120886 33101568 33100178 10033080254 33100136 33100064 33102235 10033050667 33101618 33100576 33111515 33120796 33100214 33101383 33100355 33100630 33100313 33100535 33100401 33101067 1033150049 33100282 1033142120 33100484 33130129 33100508 33100426 33100450 33100599 33100420 33100439 33100488 33120484 33101098 33101421
33101073 33100633 33101312 33102027 33102034 33102036 33102058 33102059 33102075 33120389 33130531 33102262 33120516 33111323 33111577 33101144 33110138 1033150014 33120033 33111420 33102062 33100437 33101302 1033141112 33130434 1033150045 33120558 33111896 33110292 10033070447 33100575 33101411 33101659 33111712 33101610 33101538 33102129 33101621 33102140 33110763 10033080180 33110226 33130351 1033142006 10033080206 33101306 33101141 33102021 33102297 33110336 33110341 33102178 10033070950 33102250 33110952 33102132 33102127 33102128 33102130 33102102 33102104 33102084 33110305 10033050289 33111506 33110024 33110296 33111410 33140022 33090350 33130137 1033150439 33120943 33110849 33140036 33111831 33111835 33101161 33101168 33111008 33111421 33111122 33110980 33130313 33111938 33111942 33110856 33110867 33110863 33110700 33111081 33130107 10033060446 33110611 33110646 33102215 33110811 33111067 33130008 33111385 33111161 33110987 33111147 33111007 33111026 33111291 33111452 33120005 33111399 33120429 10033060384 33111220 33120118 33111444 33111647 33120107 33120110 33111442 33111371 33111047 33111068 33120345 33100113 33110979 1033142036 1033150093 33120099 33140003 33101386 33111274 33120167 33090207 33111709 33111903 33111889 33111894 33130027 33120593 10033070243 33120310 33111907 33111644 33110369 10033050301 33120252 1033150358 1033150436 33081032 33120465 33120061 33120214 33120492 33120316 33120324 33110026 33110056 33110087 10033060814 33120064 33120393 33120097 33130047 33120219 33120225 33111924 33130427 33120317 33120216 33111139 33120402 33120621 33111809 33130100 33120580 10033060930 33120337 33130145 33120129 33090522 33120573 1033160067 33120122 10033060439 33120642 1033150520 33120175 33120519 33120550 33130075 1033142083 33120326 33120400 33120426 33120069 33130346 33110187 33120714 33120793 33102152 33120686 33120609 33140059 33120507 33120509 33120561 33120588 33120610 33120323 33120534 33120728 33120745 33111702 33120666 33100519 33120792 1033160090 33120736 33120722 33111934 33111270 10033060721 33120930 33120868 33120718 33130039 33130054 33120944 33120800 33130178 10033050525 33130004 33130086 33120844 33130199 33130567 33130012 33130102 33130071 33120954 33130290 33120946 33130115 33111327 33110551 33130209 33111518 33130156 33120583 33130112 1033142154 1033142155 33130368 33130307 33130430 33090222 10033080118 33130173 33130095 33130181 33130286 1033160065 33130317 1033150102 33130433 10033060672 10033060673 33140039 33111684 33140076 10033060464 33130308 33130128 33090234 1033150612 33130360 33130401 33120449 33130397 33130393 33130336 10033060619 33130396 33130456 10033060864 33081237 1033142003 1033142011 33110955 33130467 1033150466 33130011 33101025 33130500 33130501 1033150064 33120738 33130403 33130561 33130563 1033160093 33130565 1033142139 33130502 1033141144 33130505 33140027 1033160253 33140053 33140055 1033141015 33130513 33130517 1033141032 33130010 33130037 1033141095 33111045 1033142009 10033070682 33130029 1033142067 1033141028 33100427 33120468 33101042 33111792 33111798 33111816 33120058 1033141141 33130352 1033142105 1033142078 1033142079 1033141115 1033141126 1033141064 1033150625 1033142057 1033142115 1033142144 1033142153 1033142045 33140009 33111673 1033142090 1033150447 1033150461 1033142032 33100365 10033050572 1033150489 1033142022 33110577 10033050547 1033142128 1033150622 1033150016 1033150109 1033150302 1033150207 1033150006 33110554 1033150083 33110432 1033150353 33101035 33101103 10033060987 1033150329 1033150037 1033150086 1033142123 33100465 33120338 1033150074 10033060593 1033150406 1033150430 1033160092 1033150511 1033150338 1033150079 1033150475 1033150470 1033150425 1033150464 1033150510 1033150478 1033160203 1033141034 1033160004 1033150606 1033150429 1033150524 1033150503 1033150505 1033160109 1033150636 1033150638 1033150633 1033150613 1033150449 1033150501 1033160086 1033160277 1033160306 1033160049 1033160300 1033160210 1033160149 1033150106 1033160084 1033160069 1033160026 1033150630 1033160094 1033160095 1033160194 1033160192 1033160276 |
| FEI Number |
3003662624
|
Recalling Firm/ Manufacturer |
Advanced Sterilization Products 33 Technology Dr Irvine CA 92618-2346
|
| For Additional Information Contact | 949-453-6400 |
Manufacturer Reason for Recall | Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration. |
FDA Determined Cause 2 | Software design |
| Action | Advanced Sterilization will send an Field Safety Notice dated January 9, 2017,to inform customers that Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration.
ASP is sending the letter to remind customers to follow the instructions in the STERRAD NX System Users Guide. Specifically, after cycle initiation, the user should only open the chamber door of the STERRAD NX System when
prompted by the Graphical User Interface (GUI) per the Users Guide. Customers are instructed to complete and return the attached Business Reply Card (double-sided) for their acknowledgement within 3 business days of receipt of this notification letter via mail by using the pre-paid postage, by email
at ASP5816@stericycle.com, or by fax to Stericycle: 8662297775.
For questions regarding the business reply card, customers are instructed to contact Stericycle at (877) 497- 2425 and reference event # 5816.
For further questions regarding this recall please call (949) 453-6400, |
| Quantity in Commerce | 7,694 units total (5,935 units in US) |
| Distribution | Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Australia, China, Hong Kong, India, Indonesia, Korea (South), Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Taiwan, Thailand, Vietnam, Canada, Belgium, Czech Republic, Egypt, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Netherlands, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, Japan, Argentina, Bolivia, Brazil, Chile, Colombia, Ecuador, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MLR
|
|
|
|