• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Brightview XCT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Brightview XCTsee related information
Date Initiated by FirmJanuary 31, 2017
Date PostedMarch 16, 2017
Recall Status1 Terminated 3 on July 02, 2018
Recall NumberZ-1479-2017
Recall Event ID 76483
510(K)NumberK062298 K080927 
Product Classification diagnostic scanner - Product Code KPS
Product882482: BrightView XCT 882454 BrightView X upgrade to XCT, designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors).
Code Information 6000013, 6000025, 6000029, 6000036, 6000038, 6000045, 6000051, 6000065, 6000006, 6000070, 6000074, 6000072, 6000078, 6000081, 6000083, 6000084, 6000086, 6000090, 6000113, 6000127, 6000173, 6000174, 6000185, 6000182, 6000188, 6000183, 6000187, 6000186, 6000201, 6000215, 6000236, 6000234, 6000246, 6000296, 6000332, 6000334, 6000351, 6000368, 6000370, 6000380, 6000012, 6000152, 6340009, 6000011, 6000024, 6000175, 6000031, 6000253, 6000042, 6000052, 6000030, 6000002, 6000004, 6000069, 6000076, 6000142, 6000004A, 6000122, 6000016, 6000158, 6000189, 6000093, 6000005, 6000095, 6000092, 6000022, 6000035, 6000019, 6000023, 6000010, 6000111, 6000040, 6000056, 6000060, 6000057, 6000026, 6000054, 6000033, 6000028, 6000114, 6000003, 6000049, 6000059, 6000110, 6000020, 6000062, 6000058, 6000162, 6000018, 6000043, 6000064, 6000071, 6000027, 6000037, 6000085, 6000047, 6000017, 6000015, 6000063, 6000061, 6000048, 6000055, 6000091, 6000044, 6000106, 6000100, 6000067, 6000102, 6000096, 6000286, 6000103, 6000088, 6000107, 6000068, 6000041, 6000050, 6000104, 6000077, 6000066, 6000089, 6000094, 6000099, 6000130, 6000117, 6000080, 6000097, 6000087, 6000134, 6000073, 6000079, 6000053, 6000105, 6000244, 6000101, 6000116, 6000218, 6000121, 6000148, 6000108, 6000138, 6000143, 6000098, 6000150, 6000082, 6000118, 6000123, 6000120, 6000277, 6000147, 6000144, 6000128, 6000119, 6000109, 6000112, 6000208, 6000115, 6000125, 6000209, 6000288, 6000139, 6000160, 6000159, 6000140, 6000129, 6000184, 6000145, 6000146, 6000161, 6340008, 6000141, 6000131, 6000135, 6000132, 6000361, 6000151, 6000133, 6000238, 6000191, 6000155, 6000149, 6000153, 6000156, 6000163, 6000136, 6000254, 6000257, 6000177, 6000154, 6000270, 6000225, 6000170, 6000196, 6000126, 6000157, 11000021, 6000165, 6000164, 6000167, 6000190, 6000168, 6000166, 6000169, 6000172, 6000178, 6000405, 6000353, 6000214, 6000198, 6000197, 6000231, 6000180, 6000193, 6000192, 6000213, 6000176, 6340010, 6000179, 6000194, 6000171, 6000203, 6000195, 6000199, 6000204, 6000233, 6000212, 6000200, 6000181, 6000360, 6000202, 6000217, 6000412, 6000205, 6000210, 6000227, 6000272, 6000211, 6000230, 6000219, 6000220, 6000229, 6000228, 6000237, 6000235, 6000207, 6000206, 6000250, 6000222, 6000226, 6000046, 6000331, 6000223, 6000221, 6000287, 6000240, 6000224, 6000216, 6000014, 6000232, 6000239, 6000283, 6000249, 6000258, 6000245, 6340011, 6000256, 6000241, 6000278, 6000242, 6000243, 6000266, 6000285, 6000247, 6000259, 6000251, 6000263, 6000284, 6000345, 6000260, 6000265, 6000279, 6000255, 6000262, 6000269, 6000276, 6000274, 6000252, 6000271, 6000268, 6000267, 6000312, 6000261, 6000273, 6000006A, 6000264, 6000280, 6000281, 6000290, 6000311, 6000289, 6000275, 6000325, 6000310, 6000309, 6000416, 6000335, 6000356, 6000307, 6000302, 6000292, 6000291, 6000303, 6000294, 6000295, 6000282, 6000321, 6000297, 6000300, 6000320, 6000305, 6000301, 6000314, 6000298, 6000344, 6000364, 6000306, 6000308, 6000315, 11000111, 6000338, 6000318, 6000317, 6000299, 6000316, 6000293, 6000341, 6000313, 6000319, 6000330, 6000337, 6000350, 6000326, 6000357, 6000328, 6000327, 6000385, 6000324, 6000322, 6000336, 11000082, 6000329, 6000304, 6000323, 6000339, 6000343, 6000377, 6000403, 6000348, 6000346, 6000349, 6000378, 6000347, 6000367, 6000371, 6000383, 6000352, 6000366, 6000354, 6000355, 6000358, 6000359, 6000362, 6000365, 6000404, 6000363, 6000376, 6000411, 6000393, 6000381, 6000407, 6000369, 6000408, 6000420, 6000401, 6000387, 6000340, 6000422, 6000373, 6000372, 6000333, 6000379, 6000417, 6000375, 6000388, 6000384, 6000390, 6000400, 6000382, 6000399, 6000394, 6000395, 6000389, 6000386, 6000409, 6000396, 6000342, 6000424, 6000425, 6000423, 6000391, 6000392, 6000402, 6000398, 6000428, 6000406, 6000410, 6000414, 6000413, 6000427, 6000421, 6000418, 6000374, 6000426, 6000415, 6000419, 6000075, 6000032 & 6000039. 
FEI Number 1525965
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information ContactMs. Holly Wright Lee
440-483-2015
Manufacturer Reason
for Recall
Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
FDA Determined
Cause 2
Software design
ActionConsignees received a customer information letter via certified mail, 1/31/2017. International consignees were notified by Philips Healthcare Global Market representatives. Field Service Engineers are contacting consignees to conduct appropriate corrections. Non-responders will be contacted multiple times.
Quantity in Commerce1218 total
DistributionWorldwide distribution. US nationwide, ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CAYMEN ISLANDS, CHILE, CHINA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, FINLAND, FRANCE, GABON, GERMANY, GREAT BRITAIN, GREECE, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, & VIETNAM.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = KPS
510(K)s with Product Code = KPS
-
-