Class 2 Device Recall PowerLOAD Cot Fastening System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall PowerLOAD Cot Fastening Systemsee related information
Date Initiated by FirmMarch 12, 2018
Date PostedMay 01, 2018
Recall Status1 Terminated 3 on May 12, 2020
Recall NumberZ-1389-2018
Recall Event ID 79630
510(K)NumberK113598 
Product Classification Stretcher, wheeled - Product Code FPO
ProductPower-LOAD Cot Fastening System, Model #6390, Catalog #6390-000-000, UDI: 07613327261523 Power-LOAD is intended to assist with loading and unloading of a compatible wheeled stretcher (ambulance cot) to and from a transport vehicle and to secure the ambulance cot during transport. The device has a maximum safe working load of 870 lb (395 kg), which includes the weight of the ambulance cot, patient, and equipment attached to the cot (such as oxygen bottles, monitors, and pumps). The intended users of the device are trained professionals, including emergency medical service and medical care center personnel, as well as medical first responders, service technicians and installers.
Code Information Serial numbers: 160139700  160139701  160139702  160139704  160139705  160139706  160139707  160139708  160139709  160139710  160139711  160139712  160139713  160139714  160139743  160139744  160139745  160139746  160139747  160139748  160139749  160139750  160139751  160139752  160139753  160139754  160139755  160139756  160139804  160139805  160139806  160139807  160139808  160139809  160139810  160139811  160139812  160139813  160139814  160139815  160139816  160139817  160139818  160139819  160139820  160139821  160139822  160139823  160139824  160139825  160139826  160139827  160139828  160139829  160139830  160139831  160139832  160139833  160139834  160139835  160139836  160139837  160139838  160139839  160139894  160139895  160139896  160139897  160139898  160139899  160139900  160139901  160139902  160139903  160139904  160139905  160139906  160139907  160139908  160139909  160139910  160139911  160139912  160139913  160139914  160139915  160139916  160139917  160139918  160139919  160139920  160139921  160139922  160139923  160139924  160139925  160139926  160139927  160139928  160139929  160139930  160139931  160139982  160139983  160139984  160139985  160139986  160139987  160139988  160139989  160139990  160139991  160139992  160139993  160139994  160139995  160139996  160139997  160139998  160139999  160140027  160140028  160140029  160140030  160140031  160140032  160140033  160140034  160140035  160140036  160140037  160140038  160140039  160140040  160140041  160140042  160140043  160140044  160140045  160140046  160140047  160140048  160140049  160140075  160140076  160140077  160140078  160140079  160140080  160140081  160140082  160140083  160140084  160140085  160140086  160140087  160140088  160140089  160140090  160140091  160140127  160140128  160140129  160140130  160140131  160140132  160140133  160140134  160140135  160140136  160140137  160140138  160140139  160140140  160140141  160140142  160140143  160140144  160140145  160140146  160140340  160140341  160140342  160140343  160140344  160140345  160140346  160140347  160140348  160140349  160140350  160140351  160140352  160140353  160140354  160140395  160140396  160140397  160140398  160140399  160140749  160140751  160140752  160140753  160140754  160140755  160140756  160140757  160140758  160140759  160140760  160140793  160140794  160140795  160140796  160140797  160140798  160140799  160140800  160140801  160140802  160140803  160140804  160140805  160140806  160140807  160140808  160140809  160140810  160140811  160140812  160140813  160140814  160140815  160140816  160140817  160140818  160140819  160140820  160140821  160140822  160140823  160140824  160140825  160140826  160140827  160140828  160140829  160140830  160140831  160140832  160140833  160140834  160140835  160140836  160140837  160140838  160140839  160140840  160140878  160140879  160140880  160140881  160140882  160140883  160140884  160140885  160140886  160140887  160140888  160140889  160140890  160140891  160140892  160140893  160140894  160140922  160140923  160140924  160140925  160140926  160140927  160140928  160140929  160140930  160140931  160140932  160140933  160140934  160140935  160140936  160140937  160140938  160140939  160140940  160140941  160140962  160140963  160140964  160140965  160140966  160140967  160140968  160140969  160140970  160140971  160140972  160140973  160140974  160140975  160140976  160140977  160140978  160141031  160141032  160141033  160141035  160141036  160141037  160141039  160141040  160141041  160141042  160141043  160141044  160141045  160141046  160141047  160141048  160141550  160141551  160141557  160141558  160141559  160141560  160141561  160141562  160141563  160141564  160141565  160141566  160240050  160240051  160240052  160240053  160240054  160240055  160240056  160240057  160240058  160240059  160240060  160240061  160240062  160240063  160240064  160240065  160240066  160240400  160240401  160240402  160240403  160240404  160240405  160240406  160240407  160240408  160240409  160240410  160240411  160240412  160240450  160240462  160240463  160240464  160240465  160240466  160240467  160240468  160240469  160240470  160240471  160240472  160240473  160240474  160240475  160240476  160240496  160240497  160240498  160240499  160240500  160240501  160240502  160240503  160240504  160240505  160240506  160240507  160240508  160240509  160240510  160240511  160240512  160240513  160240514  160240515  160240516  160240517  160240518  160240519  160240520  160240521  160240522  160240523  160240524  160240525  160240526  160240527  160240564  160240565  160240566  160240567  160240568  160240569  160240570  160240571  160240572  160240573  160240574  160240575  160240576  160240577  160240578  160240579  160240580  160240581  160240582  160240583  160240584  160240585  160240586  160240587  160240588  160240589  160240590  160240591  160240592  160240593  160240594  160240595  160240596  160240614  160240707  160240708  160240709  160240710  160240711  160240712  160240713  160240714  160240715  160240716  160240717  160240718  160240719  160240720  160240721  160240722  160240723  160240724  160240725  160240726  160241140  160241141  160241142  160241143  160241144  160241145  160241146  160241147  160241148  160241149  160241150  160241151  160241152  160241153  160241154  160241155  160241156  160241157  160241158  160241159  160241190  160241191  160241192  160241193  160241194  160241195  160241196  160241197  160241198  160241199  160241200  160241201  160241202  160241203  160241204  160241205  160241206  160241207  160241208  160241209  160241210  160241246  160241247  160241248  160241249  160241250  160241251  160241252  160241253  160241254  160241255  160241256  160241257  160241258  160241259  160241260  160241261  160241262  160241263  160241264  160241265  160241314  160241315  160241316  160241317  160241318  160241319  160241320  160241321  160241322  160241323  160241324  160241325  160241326  160241327  160241328  160241329  160241330  160241331  160241332  160241333  160241356  160241357  160241358  160241359  160241360  160241361  160241362  160241363  160241364  160241365  160241366  160241367  160241368  160241369  160241370  160241371  160241372  160241373  160241374  160241375  160241376  160241377  160241378  160241379  160241380  160241381  160241382  160241383  160241384  160241385  160241386  160241387  160241388  160241389  160241390  160241391  160241392  160241393 160241394  160241395  160241396  160241397  160241398  160241462  160241463  160241464  160241465  160241466  160241467  160241468  160241469  160241470  160241471  160241472  160241473  160241474  160241475  160241476  160241477  160241478  160241479  160241480  160241481  160241482  160241483  160241484  160241485  160241486  160241541  160241542  160241543  160241544  160241545  160241546  160241547  160241548  160241549  160241550  160241551  160241552  160241553  160241554  160241555  160241556  160241557  160241558  160241559  160241560  160241561  160241562  160241563  160241564  160241565  160241606  160241607  160241608  160241609  160241610  160241611  160241612  160241613  160241614  160241615  160241616  160241617  160241618  160241619  160241620  160241621  160241622  160241623  160241624  160241625  160241626  160241627  160241628  160241629  160241630  160241631  160241632  160241633  160241634  160241635  160241636  160241637  160241638  160241639  160241640  160241641  160241642  160241643  160241644  160241645  160241646  160241647  160241648  160241649  160241650  160241651  160241652  160241653  160241654  160241655  160339106  160339107  160339108  160339109  160339110  160339111  160339112  160339113  160339114  160339115  160339116  160339117  160339118  160339119  160339120  160339121  160339122  160340158  160340159  160340160  160340161  160340162  160340163  160340164  160340165  160340166  160340167  160340168  160340169  160340170  160340171  160340172  160340173  160340174  160340175  160340176  160340177  160340178  160340179  160340180  160340181  160340182  160340183  160340184  160340185  160340186  160340187  160340188  160340189  160340191  160340192  160340193  160340194  160340195  160340196  160340197  160340198  160340199  160340200  160340201  160340202  160340203  160340204  160340205  160340206  160340207  160340208  160340209  160340210  160340211  160340212  160340213  160340214  160340215  160340216  160340217  160340218  160340219  160340220  160340221  160340222  160340223  160340224  160340225  160340227  160340228  160340229  160340230  160340231  160340232  160340233  160340234  160340236  160340237  160340238  160340239  160340274  160340278  160340279  160340280  160340281  160340282  160340283  160340284  160340285  160340286  160340287  160340289  160340290  160340291  160340292  160340293  160340294  160340295  160340296  160340297  160340300  160340301  160340302  160340303  160340304  160340305  160340306  160340307  160340308  160340309  160340310  160340311  160340312  160340313  160340314  160340315  160340316  160340317  160340318  160340319  160340320  160340321  160340322  160340323  160340324  160340325  160340326  160340327  160340328  160340329  160340330  160340331  160340332  160340333  160340334  160340335  160340336  160340337  160340338  160340339  160340340  160340341  160340342  160340343  160340344  160340345  160340346  160340347  160340348  160340349  160340350  160340351  160340352  160340353  160340354  160340355  160340356  160340357  160340358  160340359  160340360  160340361  160340362  160340363  160340364  160340365  160340366  160340367  160340368  160340369  160340370  160340371  160340374  160340375  160340376  160340377  160340378  160340379  160340380  160340381  160340382  160340383  160340384  160340385  160340386  160340387  160340388  160340389  160340390  160340391  160340392  160340393  160340417  160340418  160340419  160340420  160340421  160340422  160340423  160340424  160340425  160340426  160340427  160340428  160340429  160340430  160340485  160340486  160340487  160340488  160340489  160340490  160340491  160340492  160340493  160340494  160340495  160340496  160340497  160340498  160340499  160340500  160340501  160340502  160340503  160340504  160340505  160340506  160340507  160340508  160340509  160340790  160340791  160340792  160340794  160340795  160340796  160340797  160340798  160340799  160340800  160340801  160340802  160340803  160340804  160340805  160340806  160340807  160340808  160340846  160340847  160340848  160340849  160340850  160340851  160340852  160340853  160340854  160340855  160340856  160340857  160340858  160340859  160340860  160340861  160340862  160341017  160341018  160341019  160341020  160341021  160341022  160341023  160341024  160341025  160341026  160341027  160341028  160341029  160341030  160341031  160341032  160341033  160341034  160341035  160341036  160341037  160341038  160341039  160341040  160341041  160341042  160341043  160341044  160341045  160341046  160341047  160341048  160341049  160341050  160341051  160341052  160341053  160341054  160341055  160341056  160341057  160341058  160341059  160341060  160341061  160341062  160341063  160341064  160341065  160341066  160341067  160341068  160341069  160341070  160341071  160341072  160341150  160341151  160341152  160341153  160341154  160341155  160341156  160341157  160341158  160341159  160341160  160341161  160341162  160341163  160341164  160341165  160341166  160341167  160341168  160341169  160341170  160341171  160341172  160341173  160341174  160341250  160341251  160341252  160341253  160341254  160341255  160341256  160341257  160341258  160341259  160341260  160341261  160341262  160341263  160341264  160341265  160341266  160341267  160341268  160341269  160439110  160439111  160439112  160439113  160439114  160439115  160439116  160439117  160439118  160439119  160439120  160439121  160439122  160439123  160439124  160439125  160439126  160439127  160439128  160439129  160439130  160439131  160439132  160439133  160439134  160439186  160439187  160439188  160439189  160439190  160439191  160439193  160439194  160439195  160439196  160439197  160439198  160439199  160439200  160439201  160439202  160439203  160439204  160439205  160439206  160439207  160439208  160439209  160439210  160439211  160439212  160439213  160439214  160439215  160439216  160439217  160439218  160439219  160439220  160439221  160439222  160439223  160439224  160439225  160439285  160439286  160439287  160439288  160439289  160439290  160439291  160439292  160439293  160439294  160439295  160439296  160439297  160439298  160439299  160439300  160439301  160439302  160439303  160439304  160439305  160439306  160439307  160439308  160439309  160439310  160439311  160439312  160439313  160439314  160439315  160439316  160439317  160439318  160439319  160439320  160439321  160439322  160439323  160439324  160439325  160439326  160439329  160439330  160439332  160439333  160439334  160439340  160439341  160439342  160439343  160439344  160439345  160439346  160439347  160439348  160439349  160440103  160440104  160440105  160440106  160440107  160440108  160440109  160440110  160440111  160440112  160440113  160440114  160440115  160440116  160440117  160440118  160440119  160440120  160440121  160440122  160440123  160440124  160440125  160440126  160440127  160440128  160440129  160440130  160440131  160440132  160440133  160440134  160440135  160440136  160440137  160440138  160440139  160440140  160440141  160440143  160440144  160440145  160440146  160440147  160440148  160440149  160440150  160440151  160440152  160440153  160440154  160440155  160440156  160440157  160440158  160440159  160440160  160440161  160440162  160440163  160440164  160440165  160440166  160440196  160440197  160440198  160440199  160440200  160440201  160440202  160440203  160440204  160440205  160440206  160440207  160440208  160440209  160440277  160440278  160440279  160440280  160440281  160440282  160440283  160440284  160440285  160440287  160440288  160440289  160440290  160440291  160440292  160440293  160440294  160440295  160440296  160440297  160440298  160440299  160440300  160440301  160440790  160440791  160440792  160440793  160440794  160440795  160440796  160440797  160440798  160440799  160440800  160440801  160440802  160440803  160440804  160440805  160440806  160440807  160440808  160440809  160440810  160440811  160440812  160440813  160440814  160440815  160440816  160440817  160440818  160440825  160440826  160440855  160440856  160440857  160440858  160440859  160440860  160440861  160440862  160440863  160440865  160440866  160440867  160440868  160440930  160440936  160440937  160440938  160440939  160440940  160440941  160440942  160440943  160440944  160440945  160440946  160440947  160440948  160440949  160440970  160440971  160440972  160440973  160440974  160440975  160440976  160440977  160440978  160440979  160440980  160440981  160440982  160440983  160440984  160440985  160440986  160440987  160440988  160440989  160441011  160441012  160441013  160441014  160441015  160441016  160441017  160441018  160441019  160441020  160441021  160441022  160441023  160441024  160441025  160441026  160441027  160441028  160441040  160441041  160441042  160441043  160441044  160441045  160441046  160441047  160441048  160441049  160441050  160441051  160441052  160441053  160441054  160441055  160441056  160441057  160441058  160441059  160441060  160441061  160441062  160441063  160441064  160441742  160441743  160441744  160441745  160441781  160441782  160441793  160441794  160539098  160539099  160539100  160539101  160539102  160539350  160539351  160539352  160539353  160539354  160539355  160539356  160539357  160539358  160539359  160539360  160539361  160539362  160539363  160539364  160539365  160539366  160539367  160539368  160539369  160539370  160539371  160539372  160539373  160539374  160539375  160539427  160539428  160539429  160539430  160539431  160539432  160539433  160539434  160539435  160539436  160539437  160539438  160539439  160539440  160539441  160539442  160539443  160539444  160539445  160539446  160539484  160539485  160539486  160539522  160539534  160539535  160539536  160539537  160539538  160539539  160539540  160539541  160539542  160539543  160539544  160539545  160539546  160539547  160539548  160539549  160539550  160539551  160539552  160539553  160539554  160539555  160539556  160539557  160539558  160539559  160539560  160539561  160539562  160539622  160539623  160539624  160539625  160539626  160539627  160539628  160539629  160539630  160539631  160539632  160539633  160539634  160539635  160539636  160539637  160539638  160539639  160539640  160539641  160539688  160539689  160539690  160539691  160539692  160539693  160539694  160539695  160539696  160539697  160539698  160539699  160540502  160540503  160540504  160540505  160540506  160540507  160540508  160540509  160540510  160540511  160540512  160540513  160540514  160540515  160540516  160540517  160540518  160540519  160540520  160540521  160540522  160540523  160540529  160540530  160540531  160540532  160540533  160540534  160540535  160540536  160540537  160540538  160540539  160540540  160540541  160540542  160540543  160540544  160540545  160540546  160540547  160540548  160540549  160540589  160540590  160540591  160540592  160540593  160540594  160540595  160540596  160540597  160540598  160540599  160540600  160540601  160540602  160540603  160540604  160540605  160540606  160540607  160540608  160540609  160540610  160540611  160540612  160540613  160540614  160540615  160540616  160540617  160540618  160540619  160540620  160540621  160540622  160540623  160540624  160540625  160540626  160540627  160540628  160540629  160540630  160540631  160540632  160540633  160540634  160540635  160540636  160540637  160540638  160540639  160540640  160540641  160540642  160540643  160540644  160540645  160540646  160540647  160540648  160540649  160540680  160540681  160540682  160540683  160540684  160540685  160540686  160540687  160540688  160540689  160540690  160540691  160540692  160540693  160540694  160541100  160541101  160541102  160541103  160541104  160541105  160541106  160541107  160541108  160541109  160541110  160541111  160541112  160541113  160541114  160541115  160541116  160541117  160541118  160541119  160541120  160541121  160541122  160541123  160541124  160541125  160541126  160541127  160541128  160541129  160541130  160541131  160541132  160541133  160541134  160541135  160541136  160541450  160541451  160541452  160541453  160541454  160541455  160541456  160541457  160541458  160541459  160541460  160541461  160541462  160541463  160541464  160541465  160541466  160541467  160541468  160541469  160541470  160541471  160541472  160541473  160541474  160541475  160541476  160541477  160541478  160541479  160541480  160541481  160541482  160541483  160541484  160541485  160541486  160541487  160541488  160541489  160541490  160541491  160541492  160541493  160541494  160541495  160541583  160541584  160541585  160541586  160541587  160541588  160541589  160541590  160541591  160541592  160541593  160541594  160541595  160541596  160541597  160541598  160541599  160541600  160541601  160541602  160541603  160541604  160541605  160541606  160541607  160541608  160541646  160541647  160541648  160541649  160541650  160541651  160541652  160541653  160541654  160541655  160541691  160541692  160541693  160541694  160541695  160541696  160541697  160541698  160541699  160541700  160541721  160541722  160541723  160541724  160541725  160541726  160541727  160541729  160541730  160541731  160541732  160541733  160541734  160541735  160541736  160541737  160541738  160541739  160541740  160541741  160541742  160541743  160541744  160541745  160541746  160541747  160541748  160541749  160541750  160541751  160541752  160541753  160541754  160541755  160541756  160541757  160541758  160541759  160541760  160541761  160541762  160541763  160541764  160541765  160541766  160541767  160541768  160541769  160541770  160541771  160541772  160541797  160541798  160541799  160639078  160639079  160639080  160639081  160639082  160639083  160639084  160639085  160639086  160639087  160639088  160639100  160639101  160639102  160639103  160639104  160639105  160639106  160639107  160639108  160639109  160639110  160639111  160639112  160639113  160639114  160639115  160639116  160639117  160639118  160639119  160639120  160639121  160639122  160639123  160639124  160639160  160639161  160639162  160639163  160639164  160639165  160639166  160639167  160639168  160639169  160639170  160639171  160639172  160639173  160639174  160639175  160639176  160639177  160639178  160639179  160639180  160639181  160639182  160639183  160639184  160639198  160639250  160639251  160639252  160639253  160639254  160639255  160639256  160639257  160639258  160639259  160639260  160639261  160639262  160639263  160639264  160639265  160639266  160639267  160639268  160639269  160639318  160639319  160639320  160639322  160639323  160639324  160639325  160639326  160639327  160639328  160639329  160639330  160639331  160639332  160639333  160639334  160639335  160639336  160639337  160639338  160639339  160639340  160639341  160639342  160639343  160639344  160639352  160639353  160639354  160639355  160639356  160639357  160639358  160639359  160639360  160639361  160639362  160639363  160639364  160639365  160639366  160639367  160639368  160639369  160639370  160639371  160639372  160639373  160639374  160639375  160639376  160639377  160639378  160639379  160639380  160639381  160639382  160639383  160639384  160639385  160639386  160639387  160639388  160639389  160639390  160639391  160639781  160639782  160639783  160639848  160639849  160639850  160639851  160639852  160639853  160639854  160639855  160639856  160639857  160639858  160639859  160639860  160639861  160639862  160639863  160639864  160639865  160639866  160639867  160639887  160639888  160639889  160639890  160639891  160639892  160639893  160639894  160639895  160639896  160639897  160639898  160639899  160639900  160639901  160639902  160639903  160639904  160639905  160639906  160639907  160639908  160639909  160639910  160639911  160639912  160639913  160639914  160639915  160639916  160639917  160639918  160639919  160639920  160639921  160639922  160639923  160639924  160639925  160639926  160639927  160639928  160639987  160639988  160639989  160639990  160639991  160639992  160639993  160639994  160639995  160640000  160640001  160640002  160640003  160640004  160640005  160640006  160640007  160640008  160640009  160640010  160640011  160640012  160640013  160640014  160640015  160640016  160640017  160640018  160640019  160640020  160640021  160640022  160640023  160640024  160640025  160640026  160640027  160640028  160640029  160640030  160640031  160640075  160640076  160640077  160640078  160640079  160640080  160640081  160640082  160640083  160640084  160640085  160640086  160640087  160640088  160640089  160640090  160640091  160640092  160640093  160640094  160640095  160640499  160640513  160640514  160640515  160640516  160640517  160640518  160640519  160640520  160640521  160640522  160640523  160640524  160640525  160640526  160640527  160640528  160640529  160640530  160640531  160640532  160640533  160640534  160640535  160640536  160640537  160640538  160640539  160640540  160640541  160640542  160640543  160640544  160640545  160640546  160640547  160640548  160640549  160640600  160640601  160640602  160640603  160640604  160640605  160640606  160640607  160640608  160640700  160640701  160640702  160640703  160640704  160640705  160640706  160640707  160640708  160640709  160640710  160640711  160640712  160640713  160640714  160640715  160640716  160640717  160640718  160640719  160640720  160640752  160640753  160640754  160640755  160640756  160640757  160640758  160640759  160640760  160640761  160640762  160640763  160640764  160640765  160640766  160640767  160640768  160640769  160640770  160640771  160640772  160640773  160640774  160640775  160640776  160640777  160640778  160640779  160640780  160640781  160640782  160640783  160640784  160640785  160640786  160640787  160640788  160640789  160640790  160640791  160640792  160640793  160640794  160641265  160641266  160641267  160641268  160641269  160641270  160641271  160641272  160641273  160641274  160641275  160641276  160641277  160641278  160641279  160641280  160641281  160641282  160641283  160641284  160641285  160641286  160641287  160641288  160641289  160641290  160641291  160641292  160641293  160641294  160641295  160641330  160641331  160641332  160641333  160641334  160641335  160641336  160641337  160641338  160641339  160641348  160641349  160641350  160641351  160641352  160641353  160641354  160641355  160641356  160641357  160641358  160641359  160641360  160641361  160641362  160641363  160641364  160641365  160641366  160641367  160641368  160641369  160641370  160641371  160641372  160641373  160641374  160641375  160641376  160641377  160641378  160641379  160641380  160641381  160641382  160641383  160641384  160641385  160641386  160641387  160641388  160641389  160641390  160641391  160641392  160641393  160641394  160641395  160641396  160641397  160641398  160641399  160739015  160739016  160739017  160739018  160739019  160739020  160739021  160739022  160739023  160739050  160739051  160739052  160739053  160739054  160739055  160739056  160739057  160739058  160739059  160739060  160739061  160739062  160739063  160739064  160739065  160739066  160739067  160739068  160739069  160739070  160739071  160739072  160739073  160739074  160739075  160739076  160739077  160739078  160739079  160739080  160739081  160739082  160739083  160739084  160739085  160739086  160739139  160739140  160739141  160739142  160739143  160739144  160739145  160739146  160739147  160739148  160739150  160739151  160739152  160739153  160739154  160739155  160739156  160739157  160739158  160739159  160739160  160739161  160739162  160739163  160739164  160739165  160739166  160739167  160739168  160739169  160739170  160739171  160739172  160739173  160739174  160739175  160739176  160739177  160739178  160739194  160739195  160739196  160739197  160739198  160739199  160739225  160739226  160739227  160739228  160739229  160739230  160739231  160739232  160739233  160739234  160739235  160739236  160739237  160739238  160739239  160739240  160739241  160739242  160739243  160739244  160739245  160739246  160739247  160739248  160739249  160739275  160739276  160739277  160739278  160739279  160739280  160739281  160739282  160739283  160739284  160739285  160739286  160739287  160739288  160739289  160739290  160739291  160739292  160739293  160739294  160739295  160739296  160739297  160739298  160739299  160739300  160739301  160739302  160739303  160739304  160739305  160739306  160739307  160739308  160739309  160739310  160739311  160739312  160739313  160739314  160739315  160739316  160739317  160739319  160739320  160739321  160739322  160739323  160739324  160739325  160739326  160739327  160739328  160739776  160739777  160739778  160739779  160739780  160739781  160739782  160739783  160739784  160739785  160739786  160739787  160739788  160739789  160739790  160739791  160739792  160739793  160739794  160739795  160739796  160739838  160739839  160739840  160739841  160739843  160739844  160739845  160739846  160739847  160739848  160739849  160739850  160739851  160739852  160739853  160739854  160739855  160739856  160739857  160739858  160739859  160739900  160739901  160739902  160739903  160739904  160739905  160739906  160739907  160739908  160739909  160739910  160739911  160739912  160739913  160739914  160739915  160739916  160739917  160739918  160739919  160739920  160739921  160739922  160739923  160739924  160739925  160739926  160739927  160739928  160739929  160739930  160739931  160739932  160739933  160739934  160739989  160739990  160739991  160739992  160739994  160739995  160739996  160739997  160739998  160739999  160740256  160740257  160740258  160740259  160740260  160740261  160740262  160740263  160740264  160740265  160740343  160740344  160740345  160740346  160740347  160740615  160740616  160740617  160740618  160740619  160740620  160740621  160740622  160740623  160740624  160740625  160740626  160740627  160740628  160740629  160741245  160741246  160741508  160741509  160741510  160741511  160741512  160741513  160741514  160741515  160741516  160741544  160741547  160741550  160741551  160741552  160741553  160741554  160741555  160741556  160741557  160741558  160741559  160741560  160741561  160741562  160741563  160741564  160741565  160741566  160741567  160741568  160741569  160741650  160741651  160741652  160741653  160741654  160741655  160741656  160741657  160741658  160741659  160741660  160741661  160741662  160741663  160741664  160741665  160741666  160741692  160741693  160741694  160741695  160741696  160741697  160741698  160741699  160741745  160741746  160741747  160741750  160741751  160741752  160741753  160741754  160741755  160741756  160741757  160741758  160741759  160741760  160741761  160741762  160741763  160741764  160741765  160741766  160741767  160741768  160741769  160741823  160741824  160741825  160741826  160741827  160741828  160741829  160741830  160741831  160741832  160741833  160741834  160742180  160742181  160742182  160742183  160742184  160742185  160742186  160742187  160742188  160742198  160742199  160839175  160839176  160839177  160839178  160839179  160839180  160839181  160839182  160839183  160839184  160839185  160839186  160839187  160839188  160839189  160839190  160839191  160839192  160839193  160839194  160839195  160839230  160839231  160839232  160839233  160839234  160839249  160839300  160839301  160839302  160839303  160839304  160839305  160839306  160839307  160839308  160839309  160839310  160839311  160839312  160839313  160839314  160839315  160839316  160839317  160839318  160839319  160839320  160839321  160839322  160839323  160839324  160839350  160839351  160839352  160839353  160839354  160839355  160839356  160839357  160839358  160839359  160839360  160839361  160839362  160839363  160839364  160839365  160839366  160839367  160839368  160839369  160839443  160839444  160839445  160839450  160839451  160839452  160839453  160839454  160839455  160839456  160839457  160839458  160839459  160839474  160839475  160839476  160839477  160839478  160839479  160839487  160839488  160839489  160839490  160839491  160839496  160839497  160839498  160839499  160839510  160839511  160839512  160839513  160839514  160839515  160839516  160839517  160839518  160839519  160839520  160839521  160839522  160839523  160839524  160839525  160839526  160839534  160839535  160839536  160839537  160839538  160839539  160839540  160839541  160839542  160839543  160839544  160839546  160839547  160839549  160839638  160839639  160839640  160839641  160839642  160839643  160839644  160839645  160839646  160839930  160839931  160839932  160839933  160839934  160839935  160839936  160839937  160839938  160839939  160839940  160839941  160839942  160839943  160840000  160840001  160840002  160840003  160840004  160840005  160840006  160840007  160840008  160840009  160840010  160840041  160840042  160840043  160840044  160840045  160840047  160840048  160840049  160840050  160840051  160840052  160840053  160840054  160840055  160840056  160840057  160840058  160840059  160840060  160840061  160840062  160840063  160840064  160840065  160840066  160840067  160840068  160840069  160840070  160840071  160840072  160840073  160840074  160840097  160840098  160840100  160840101  160840102  160840103  160840104  160840105  160840106  160840107  160840108  160840109  160840110  160840111  160840112  160840113  160840114  160840115  160840116  160840122  160840123  160840124  160840125  160840126  160840127  160840128  160840129  160840130  160840131  160840132  160840133  160840134  160840135  160840136  160840137  160840180  160840181  160840182  160840183  160840184  160840185  160840186  160840187  160840188  160840189  160840190  160840191  160840192  160840193  160840194  160840195  160840196  160840467  160840468  160840469  160840470  160840471  160840472  160840473  160840474  160840475  160840476  160840477  160840478  160840479  160840544  160840545  160840546  160840547  160840548  160840586  160840587  160840588  160840589  160840590  160840591  160840592  160840593  160840594  160840595  160840596  160840599  160840850  160840851  160840852  160840853  160840854  160840855  160840856  160840857  160840858  160840859  160840860  160840861  160840862  160840863  160840864  160840865  160840866  160840867  160840868  160840869  160840870  160840871  160840872  160840873  160840874  160840936  160840937  160840938  160840939  160840940  160840941  160840942  160840943  160840944  160840945  160840946  160840949  160840980  160840981  160840982  160840983  160840984  160840985  160840986  160840987  160840988  160840989  160840990  160840991  160840992  160840993  160840994  160840995  160840996  160840997  160840998  160840999  160841030  160841031  160841032  160841033  160841034  160841035  160841036  160841037  160841038  160841039  160841040  160841041  160841042  160841097  160841098  160841100  160841101  160841102  160841103  160841104  160841105  160841106  160841107  160841108  160841109  160841110  160841111  160841112  160841113  160841114  160841115  160841116  160841117  160841118  160841119  160841120  160841121  160841122  160841123  160841124  160841125  160841126  160841127  160841128  160841129  160841138  160841139  160841140  160841141  160841149  160841210  160841211  160841212  160841213  160841214  160841215  160841216  160841217  160841218  160841219  160841220  160841221  160841222  160841223  160841224  160841225  160841226  160841227  160841228  160841229  160841230  160841231  160841232  160841233  160841234  160841235  160841244  160841245  160841246  160841247  160841248  160841249  160841600  160841601  160841602  160841603  160841604  160841605  160841606  160841607  160841608  160841609  160841610  160841650  160841651  160841652  160841653  160841654  160841655  160841656  160841657  160841658  160841659  160841660  160841661  160841662  160841663  160841664  160841665  160841666  160841667  160841668  160841669  160841710  160841711  160841712  160841713  160841714  160841715  160841716  160841717  160841718  160841719  160841720  160841721  160841722  160841723  160841724  160841725  160841726  160841727  160841728  160841729  160841730  160841731  160841732  160841733  160841734  160939183  160939184  160939185  160939186  160939187  160939188  160939189  160939190  160939191  160939192  160939193  160939194  160939195  160939196  160939197  160939198  160939199  160939200  160939201  160939202  160939203  160939204  160939205  160939206  160939207  160939208  160939209  160939210  160939211  160939212  160939213  160939214  160939215  160939216  160939250  160939251  160939252  160939253  160939254  160939255  160939256  160939257  160939258  160939259  160939260  160939261  160939262  160939263  160939264  160939265  160939266  160939267  160939268  160939269  160939270  160939271  160939272  160939273  160939274  160939275  160939276  160939277  160939278  160939279  160939289  160939290  160939291  160939292  160939293  160939294  160939295  160939296  160939297  160939298  160939299  160939300  160939301  160939302  160939303  160939304  160939305  160939306  160939335  160939336  160939337  160939338  160939339  160939340  160939341  160939342  160939343  160939344  160939345  160939346  160939347  160939348  160939349  160939490  160939491  160939492  160939493  160939494  160939495  160939496  160939497  160939498  160939650  160939695  160940006  160940007  160940008  160940009  160940010  160940011  160940012  160940013  160940014  160940015  160940016  160940017  160940018  160940019  160940041  160940042  160940043  160940044  160940045  160940046  160940047  160940048  160940049  160940299  160940448  160940449  160940500  160940501  160940502  160940503  160940504  160940505  160940506  160940508  160940509  160940510  160940511  160940512  160940513  160940514  160940515  160940516  160940517  160940535  160940536  160940537  160940538  160940539  160940540  160940563  160940564  160940565  160940566  160940567  160940568  160940569  160940570  160940571  160940572  160940573  160940574  160940575  160940577  160940578  160940579  160940580  160940581  160940582  160940583  160940584  160940585  160940586  160940587  160940588  160940589  160940590  160940591  160940592  160940593  160940594  160940595  160940596  160940597  160940598  160940599  160940600  160940601  160940602  160940603  160940604  160940800  160940801  160940802  160940803  160940804  160940809  160940810  160940811  160940812  160940813  160940814  160940815  160940816  160940817  160940818  160940819  160940820  160940821  160940822  160940823  160940824  160940825  160940826  160940827  160940828  160941950  160941951  160941952  160941953  160941954  160941955  160941956  160941957  160941958  160941959  160941960  160941961  160941962  160941963  160941964  160941965  160941966  161039025  161039026  161039027  161039028  161039029  161039030  161039031  161039032  161039033  161039034  161039035  161039036  161039037  161039038  161039039  161039040  161039041  161039042  161039043  161039044  161039045  161039046  161039047  161039048  161039049  161039050  161039051  161039052  161039053  161039054  161039084  161039085  161039086  161039087  161039088  161039089  161039090  161039091  161039092  161039093  161039094  161039095  161039096  161039097  161039098  161039200  161039201  161039202  161039203  161039204  161039205  161039206  161039207  161039208  161039209  161039210  161039211  161039212  161039213  161039214  161039215  161039216  161039217  161039218  161039219  161039232  161039233  161039234  161039235  161039236  161039237  161039238  161039239  161039240  161039241  161039242  161039243  161039244  161039280  161039281  161039282  161039283  161039284  161039285  161039286  161039287  161039288  161039289  161039290  161039363  161039364  161039365  161039366  161039367  161039368  161039369  161039370  161039371  161039372  161039373  161039374  161039375  161039376  161039377  161039378  161039379  161039380  161039381  161039382  161039383  161039384  161039395  161039396  161039397  161039398  161039399  161039500  161039501  161039502  161039503  161039504  161039505  161039506  161039507  161039508  161039509  161039510  161039511  161039512  161039513  161039514  161039515  161039516  161039517  161039518  161039519  161039520  161039626  161039627  161039628  161039629  161039630  161039631  161039632  161039633  161039634  161039635  161039636  161039637  161039638  161039639  161039640  161039641  161039642  161039643  161039644  161039645  161039646  161039647  161039648  161039649  161039825  161039826  161039827  161039828  161039829  161039830  161039831  161039832  161039833  161039834  161039835  161039836  161039837  161039838  161039839  161039840  161039841  161039842  161039843  161039844  161040100  161040101  161040102  161040103  161040104  161040105  161040106  161040107  161040108  161040109  161040110  161040111  161040112  161040113  161040114  161040115  161040116  161040117  161040118  161040119  161040120  161040121  161040122  161040123  161040124  161040125  161040146  161040147  161040148  161040149  161040150  161040151  161040152  161040153  161040154  161040155  161040178  161040179  161040180  161040181  161040182  161040183  161040184  161040185  161040186  161040187  161040188  161040189  161040190  161040191  161040192  161040193  161040194  161040195  161040196  161040197  161040198  161040199  161040249  161040448  161040463  161040464  161040465  161040466  161040467  161040468  161040469  161040470  161040471  161040472  161040473  161040474  161040475  161040476  161040477  161040478  161040479  161040480  161040486  161040487  161040488  161040489  161040490  161040491  161040492  161040493  161040540  161040541  161040542  161040543  161040544  161040545  161040546  161040547  161040548  161040549  161040555  161040556  161040557  161040558  161040559  161040560  161040561  161040562  161040563  161040564  161040565  161040566  161040567  161040568  161040569  161040570  161040571  161040572  161040573  161040574  161040575  161040576  161040577  161040578  161040579  161040580  161040581  161040595  161040596  161040597  161040598  161040599  161040600  161040601  161040602  161040603  161040604  161040605  161040606  161040607  161040608  161040682  161040683  161040684  161040685  161040686  161040687  161040688  161040689  161040690  161040691  161040692  161040693  161040694  161040695  161040696  161040697  161040698  161040699  161040831  161040832  161040833  161040834  161040835  161040836  161040837  161040838  161040839  161040840  161040841  161040842  161040843  161040844  161041300  161041301  161041302  161041303  161041304  161041305  161041306  161041307  161041308  161041309  161041310  161041311  161041312  161041313  161041314  161041315  161041316  161041317  161041318  161041319  161041320  161041321  161041322  161041323  161041324  161041325  161041326  161041327  161041328  161041329  161041330  161041331  161041332  161041333  161041334  161041335  161041336  161041337  161041338  161041339  161041340  161041341  161041342  161041343  161041344  161041345  161041346  161041347  161041348  161041349  161041360  161041361  161041362  161041363  161041364  161041365  161041366  161041367  161041368  161041369  161041370  161041371  161041372  161041373  161041374  161041375  161041376  161041377  161041378  161041379  161041380  161041381  161041382  161041383  161041384  161041385  161041386  161041395  161041396  161041397  161041398  161041875  161041876  161041877  161041878  161041879  161041880  161041881  161041882  161041883  161041884  161041885  161041886  161041942  161041943  161041944  161041945  161041946  161041947  161041948  161041949  161139357  161139358  161139391  161139392  161139393  161139394  161139395  161139396  161139397  161139819  161139920  161139921  161139922  161139923  161139924  161139925  161139959  161139960  161139961  161139962  161139963  161139964  161139965  161139966  161139967  161139968  161139997  161140000  161140001  161140002  161140003  161140004  161140005  161140006  161140007  161140008  161140009  161140010  161140034  161140035  161140036  161140037  161140038  161140039  161140100  161140101  161140102  161140103  161140104  161140105  161140106  161140107  161140108  161140109  161140110  161140111  161140112  161140113  161140114  161140115  161140116  161140117  161140118  161140119  161140120  161140121  161140122  161140123  161140124  161140125  161140126  161140127  161140128  161140129  161140130  161140131  161140132  161140133  161140134  161140135  161140136  161140157  161140158  161140159  161140160  161140161  161140162  161140163  161140164  161140165  161140166  161140167  161140168  161140169  161140190  161140191  161140192  161140193  161140194  161140195  161140196  161140197  161140198  161140199  161140300  161140301  161140302  161140303  161140304  161140305  161140306  161140307  161140308  161140309  161140310  161140311  161140546  161140547  161140548  161140570  161140571  161140572  161140573  161140574  161140575  161140576  161140577  161140578  161140579  161140580  161140581  161140582  161140583  161140584  161140585  161140586  161140587  161140588  161140589  161140590  161140591  161140592  161140593  161140594  161140595  161140596  161140597  161140598  161140599  161140640  161140641  161140642  161140643  161140644  161140645  161140646  161140647  161140648  161140649  161140700  161140701  161140702  161140703  161140704  161140705  161140706  161140707  161140708  161140709  161140710  161140711  161140712  161140713  161140714  161140715  161140716  161140717  161140718  161140719  161140720  161140721  161140722  161140723  161140724  161140725  161140726  161140727  161140728  161140729  161140730  161140731  161140732  161140733  161140734  161140735  161140736  161140737  161140738  161140739  161140740  161140741  161140746  161140747  161140748  161140749  161140797  161140798  161140799  161140850  161140851  161140852  161140853  161140854  161140855  161140856  161140857  161140858  161140859  161140860  161140861  161140862  161140863  161140864  161140865  161140990  161140991  161140992  161141029  161141030  161141031  161141032  161141033  161141034  161141035  161141036  161141037  161141038  161141039  161141040  161141041  161141042  161141043  161141044  161141045  161141046  161141047  161141048  161141070  161141071  161141072  161141073  161141074  161141075  161141076  161141077  161141078  161141079  161141080  161141081  161141082  161141083  161141084  161141085  161141086  161141087  161141088  161141090  161141091  161141092  161141093  161141094  161141095  161141096  161141097  161141098  161141099  161141100  161141126  161141127  161141128  161141129  161141130  161141131  161141132  161141133  161141134  161141135  161141136  161141137  161141138  161141139  161141140  161141141  161141142  161141143  161141144  161141145  161141146  161141147  161141148  161141149  161141330  161141331  161141332  161141333  161141334  161141335  161141336  161141337  161141338  161141339  161141340  161141341  161141342  161141343  161141344  161141345  161141346  161141347  161141348  161141349  161141500  161141501  161141502  161141503  161141504  161141505  161141506  161141507  161141508  161141509  161141510  161141511  161141512  161141513  161141514  161141515  161141516  161141517  161141518  161141519  161141520  161141521  161141522  161141523  161141524  161141590  161141591  161141592  161141593  161141594  161141595  161141596  161141597  161141598  161141599  161141635  161141636  161141637  161141638  161141639  161141640  161141641  161141642  161141643  161141644  161141645  161141646  161141647  161141648  161141649  161141663  161141664  161141665  161141666  161141667  161141668  161141669  161141670  161141750  161141751  161141752  161141753  161141754  161141755  161141756  161141774  161141775  161141776  161141777  161141778  161141779  161141780  161141781  161141782  161141783  161141784  161141785  161141786  161141787  161141788  161141789  161141790  161141791  161141792  161141793  161141798  161141799  161141805  161141806  161141807  161141808  161141809  161141810  161141811  161141812  161141813  161141814  161141815  161141816  161141817  161141818  161141819  161141820  161141821  161141822  161141823  161141824  161141825  161141826  161141827  161141828  161141829  161141830  161141831  161141832  161141833  161141834  161141835  161141836  161141837  161141838  161141839  161141840  161141841  161141842  161141843  161141844  161141845  161141846  161141847  161141848  161141849  161141850  161141851  161141852  161141853  161141854  161141855  161141886  161141887  161141888  161141889  161141890  161141891  161141894  161141895  161141896  161141897  161141898  161141899  161142165  161142166  161142167  161142168  161142169  161142170  161142171  161142172  161142173  161142174  161142175  161142176  161142177  161142178  161142179  161142180  161142181  161142182  161142183  161142184  161142185  161142186  161142187  161142188  161142189  161142190  161142191  161142193  161239050  161239051  161239052  161239053  161239054  161239055  161239056  161239057  161239058  161239059  161239060  161239061  161239062  161239063  161239064  161239065  161239066  161239067  161239068  161239069  161239070  161239071  161239072  161239073  161239074  161239075  161239076  161239077  161239078  161239079  161239080  161239081  161239082  161239083  161239084  161239085  161239086  161239087  161239088  161239089  161239090  161239136  161239137  161239138  161239139  161239140  161239141  161239142  161239143  161239144  161239145  161239146  161239147  161239148  161239149  161239390  161239391  161239392  161239393  161239394  161239395  161239396  161239397  161239398  161239516  161239517  161239518  161239519  161239520  161239521  161239522  161239523  161239524  161239525  161239526  161239527  161239528  161239529  161239530  161239531  161239532  161239533  161239534  161239535  161239536  161239537  161239538  161239539  161239540  161239541  161239542  161239543  161239544  161239545  161239546  161239547  161239548  161239549  161239550  161239551  161239552  161239553  161239599  161239700  161239701  161239702  161239703  161239704  161239705  161239706  161239707  161239708  161239709  161239710  161239711  161239712  161239713  161239714  161239715  161239716  161239717  161239718  161239719  161239720  161239721  161239722  161239723  161239724  161239725  161239726  161239727  161239728  161239845  161239846  161239847  161239848  161239849  161239850  161239851  161239852  161239853  161239854  161239855  161239856  161239857  161239858  161239859  161239860  161239861  161239862  161239863  161239864  161239865  161239866  161239867  161239868  161239869  161239870  161239871  161239872  161239873  161239874  161239875  161239876  161239877  161239878  161239879  161239880  161239881  161239882  161239883  161239884  161239885  161239886  161239887  161239888  161239889  161239890  161239891  161239892  161239893  161239894  161239895  161239896  161239897  161239898  161239899  161239920  161239921  161239922  161239923  161239924  161239925  161239926  161239927  161239928  161239929  161239945  161239946  161239947  161239948  161239949  161240240  161240241  161240242  161240243  161240244  161240245 161240246  161240247  161240248  161240250  161240251  161240252  161240253  161240254  161240255  161240256  161240257  161240258  161240259  161240260  161240261  161240262  161240263  161240264  161240265  161240266  161240267  161240268  161240269  161240270  161240271  161240272  161240295  161240296  161240297  161240298  161240299  161240350  161240351  161240352  161240353  161240354  161240355  161240356  161240357  161240358  161240359  161240360  161240361  161240362  161240363  161240364  161240365  161240366  161240367  161240368  161240369  161240370  161240399  161240448  161240449  161240536  161240537  161240538  161240539  161240540  161240541  161240542  161240543  161240544  161240545  161240750  161240751  161240752  161240753  161240754  161240755  161240756  161240757  161240758  161240759  161240760  161240761  161240762  161240763  161240764  161240765  161240766  161240767  161240768  161240769  161240770  161240771  161240772  161240773  161240774  161240775  161240776  161240777  161240778  161240779  161241200  161241201  161241202  161241203  161241204  161241205  161241326  161241327  161241328  161241329  161241330  161241331  161241332  161241333  161241334 160739842
FEI Number 1831750
Recalling Firm/
Manufacturer
Stryker Medical Division of Stryker Corporation
3800 E Centre Ave
Portage MI 49002-5826
For Additional Information ContactVictoria Haney
269-389-8306
Manufacturer Reason
for Recall
The power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn February 27, 2018, affected customers received an Urgent Medical Device Correction Notification via FedEx notifying them of the issue and potential hazard. For distributed units, the four Floor Plate Attachment Brackets that hold the anchor in place will be replaced in the field by Strykers field service personnel. Customers will be asked to return a business reply form to acknowledge that they have received and understand the notification. For further questions, please call (269) 389-8306.
Quantity in Commerce4724
DistributionWorldwide Distribution - USA (nationwide) and to the countries of : Australia, Canada, France, Germany, Hong Kong, Italy, India, Netherlands, Portugal, Saint Martin, Switzerland, United Kingdom
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FPO
-
-