| | Class 2 Device Recall Nuclear, Magnetic Resonance Imaging |  |
| Date Initiated by Firm | March 16, 2018 |
| Date Posted | April 18, 2018 |
| Recall Status1 |
Terminated 3 on April 10, 2020 |
| Recall Number | Z-1439-2018 |
| Recall Event ID |
79451 |
| 510(K)Number | K945945 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | ACS-NT, Model 78108
Product Usage:
Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities. |
| Code Information |
(Added 6/25/18): 5097 5426 5519 5507 5255 5059 5379 5257 5316 5144 5002 5300 5440 5482 5014 5492 5019 5423 5256 5293 5382 5319 5420 5030 5477 5488 5193 5085 5347 5287 5195 5481 5370 5474 5360 5352 5106 5135 5495 5313 5413 5245 5142 5378 5411 5214 5151 5053 5222 5201 5381 5246 5306 5429 5441 5324 5273 5048 5124 5211 5453 5277 5533 5239 5248 5349 5414 5311 5513 5204 5215 5434 5577 5296 5259 5271 5512 5373 5542 5290 5091 5446 5266 5406 5551 5090 5455 5123 5261 5323 5390 5254 5353 5502 5241 5514 5238 5035 5105 5333 5454 5226 5334 5328 5417 5298 5451 5288 5200 5077 5250 5342 5309 5172 5432 5081 5199 5458 5573 5568 5268 5110 5472 5052 5009 5461 5022 5297 5263 5158 5078 5230 5558 5428 5303 5376 5515 5520 5269 5394 5253 5282 5042 5155 5084 5047 5530 5501 5383 5497 5016 5329 5162 5143 5491 5470 5566 5071 5079 5258 5125 5283 5129 5475 5574 5149 5018 5421 5473 5227 5330 5264 5132 5141 5327 5265 5351 5121 5216 5236 5180 5487 5478 5116 5160 5345 5301 5522 5083 5391 5021 5552 5314 5363 5024 |
| FEI Number |
1218950
|
Recalling Firm/ Manufacturer |
Philips Electronics North America Corporation 3000 Minuteman Rd Andover MA 01810-1032
|
| For Additional Information Contact | Austin O'Connell 800-722-9377 |
Manufacturer Reason for Recall | Potential risk for helium gas inside the MR examination room during a magnet quench |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On March 16, 2018, Philips Healthcare distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via certified mailing. On May 24, 2018, the recall letter was distributed to additional affected customers identified by the firm.
**Action to be Taken by Customer**:
During a quench, a large amount of helium evaporates and is vented outside the building through a venting system. A quench causes immediate removal of the static magnetic field. A quench can occur spontaneously or can be induced if there is an emergency. In case helium gas escapes into the examination room during a magnet quench ensure strict adherence to the Instructions for Use:
Release of helium gas in the examination room"
If helium gas is not vented properly after the Magnet Emergency Off button is used or during a quench (for example if the helium vent pipe is blocked) a high concentration of helium gas may build up in the examination room. This gas forms clouds of cold mist. Helium gas dilutes the oxygen in the air. High concentrations of helium gas can lead to suffocation.
If helium enters the examination room:
" Immediately remove all persons from the examination room.
" Do not switch off air circulation and ventilation in the examination room.
" Do not reenter the examination room until it is confirmed that the air oxygen content is at a safe level.
"Emergency procedures"
The User is required to establish emergency procedures for the following situations:
" A medical emergency
" A fire
" An emergency that requires immediate removal of the magnetic field
" The release of helium gas into the examination room
Philips MRI systems have an Emergency Table Stop button in case there is an emergency during tabletop movement.
(Instructions for Use R5.3)
Philips will schedule an inspection of all MRI systems that may be affected by this notice. If applicable the spare metal burst disk on site will also be replaced during this inspection. Metal Burst disks on all aff |
| Quantity in Commerce | 8,205 units in total |
| Distribution | Worldwide Distribution - US Nationwide; International - Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia,
Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Chile,
China, Colombia, Congo (Democratic Republic of the), Costa Rica,
C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic,
Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary,
Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan,
Kazakhstan, Kenya, Republic of Korea - South Korea, Kuwait, Kyrgyzstan,
Latvia, Lebanon, Lesotho, Libya, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Mauritius, Mexico, Monaco, Mongolia,
Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Caledonia,
New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of
Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Qatar,
Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore,
Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland,
Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Venezuela,
Viet Nam, Yemen & Zimbabwe. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = LNH
|
|
|
|