• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Promote Quadra

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Promote Quadrasee related information
Date Initiated by FirmApril 16, 2018
Create DateJune 29, 2018
Recall Status1 Terminated 3 on March 08, 2024
Recall NumberZ-2323-2018
Recall Event ID 80212
PMA NumberP030054 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductPromote Quadra, Sterile EO, Model # / Part #: CD3221-36/100025054, 60019195; CD3223-36P/100013763; CD3239-40/100029115, 100029254, 100037181; CD3239-40Q/100029116, 100029244, 100037115
Code Information Serial #: 1062415 584948 584973 844882 1059669 585788 844881 585793 585797 844884 585785 585777 1059667 585796 584135 584051 584008 584133 583946 583999 583991 584136 583996 583982 583943 584078 583980 584005 584007 584106 583989 584128 583950 584028 584060 584115 583968 584064 584067 584069 584029 584077 584024 584116 584058 584044 584111 583930 583964 584065 583994 584032 583966 583990 583979 584062 568379 583959 584056 568359 584019 584015 584126 584014 583978 583975 584038 584050 584013 584049 583954 584127 584031 584080 584119 568348 584071 584070 584021 584000 583957 584122 584083 583987 584081 584035 583976 584072 583997 583985 584040 583963 584017 584054 584026 584002 568380 584068 584034 583972 584043 584023 583981 584125 584001 585635 585143 585420 585127 585405 585289 585400 585422 585379 585370 585442 585319 585357 585129 585423 585414 585445 585435 585428 585099 585415 585438 585446 585444 585140 585436 585409 585365 585408 585705 585608 585696 585724 585715 585760 585732 585708 585629 585698 585693 585699 585697 585734 585725 585686 585681 585719 585710 585721 585740 585443 585694 585677 585746 585722 585680 585763 585712 585703 585727 585690 585700 585632 585729 585702 585747 585748 585678 585354 585348 585388 585744 585753 585146 585689 585711 585687 585695 585728 585749 585723 585758 585755 585726 585720 585688 585434 585752 585278 585733 585757 585761 585714 585736 585684 585739 585704 585387 585685 585691 585759 585682 585619 585706 585756 585741 585738 585750 585754 585590 584463 584461 584431 584489 584330 584521 584353 584368 584446 584508 584090 584495 584356 584170 583948 584483 584412 584476 584418 568405 584454 584473 584464 584513 584487 584444 584434 584465 584422 568370 568483 584171 584459 584485 585631 585742 585717 585764 585766 585701 585582 585120 585299 585567 585605 585612 585574 585595 585560 585094 585576 585628 585585 585583 585617 585251 585075 585573 585623 585633 585571 585566 585360 585578 585614 585613 585286 585133 585592 585615 585607 585557 585427 585577 585630 585581 585579 585610 585584 585626 585559 585565 585620 585572 585274 585621 585611 585616 585153 585627 585624 585563 585598 585568 585589 585636 585601 585606 585593 585618 585596 585564 585599 585358 585586 585558 585600 585561 585597 585591 585602 585316 585625 585609 585604 585603 585575 585244 585569 585562 585570 585730 568470 568511 571905 571964 571869 571923 571903 568443 568510 568452 571970 571955 571947 571946 571872 571912 568502 567733 567728 568465 571958 571922 568442 568444 571921 571950 571916 584360 584492 584514 584518 584506 584447 584500 584482 568413 584468 584430 584511 584493 584458 584445 584504 584451 584088 584457 584419 584442 584415 584471 584424 584507 584148 584156 583988 567697 584157 583931 568400 584149 584087 584076 567683 584134 567686 584161 583977 583942 568389 583928 584046 584138 584012 583983 584164 584025 584123 567672 568361 584074 584041 583955 584073 568374 584030 583984 584066 584129 584121 584107 584098 568420 583956 584063 583925 584096 583992 583986 567688 584085 584095 584042 584061 584142 583971 584347 583937 583998 584027 584047 583935 584143 584160 584117 584105 584155 568350 567667 584154 584057 584159 584094 567679 584141 568375 568365 583936 568418 584018 583944 584342 584100 583951 583953 584153 567678 568363 567698 584147 584371 568351 584112 584367 584048 584151 584139 584393 583940 568415 585160 585271 585130 585156 585142 585083 585062 585097 585113 585115 585154 585117 585086 585077 585126 585092 585121 585087 585125 585137 585074 585124 585110 585078 585155 585158 585161 585136 585085 585159 585108 585068 585273 585091 585254 585063 585300 585082 585123 585135 585105 585093 585107 585081 585281 585301 585303 585269 585252 585112 585233 585243 585245 585157 585315 585268 585386 585371 585309 585267 585280 585152 585304 585412 585238 585429 585227 585312 585298 585410 585306 585393 585383 585433 585283 585282 585287 585285 585247 585275 585308 585134 585266 585317 585255 585426 585138 585261 585418 585441 585411 585241 585314 585222 585265 585307 585263 585373 585284 585368 585066 585421 585232 585258 585249 585223 585382 585128 585235 585272 585305 585401 585353 585228 585296 585090 585239 585150 585432 585102 585104 585065 585311 585290 585277 585288 585257 585297 585116 585259 585229 585149 585310 585237 585246 585291 585234 585260 585224 585253 585262 585248 585250 585240 585279 585225 585230 585313 585302 585270 585226 585141 585397 585242 585294 585231 584515 584484 584082 584510 584417 584350 584383 583993 584011 584479 584466 584339 584130 583995 584414 584150 584124 584091 584499 584335 584516 584324 584363 584327 584319 584477 584509 584437 583962 584490 584425 584435 584439 584373 584517 584496 584351 584388 584158 584089 584416 584433 584341 584354 584427 584086 584358 584520 584449 584470 584467 584423 584016 584420 584140 584352 584382 584355 584426 584440 583941 584450 584480 584502 584093 584503 584421 584436 584456 584488 584162 584498 584346 584113 584369 584053 584453 584481 584163 584389 583960 584337 583970 584394 584413 584370 584331 584102 584120 584079 584384 584380 584092 584391 584055 584345 584146 584322 584452 584494 584340 584366 584333 584386 584039 584432 584486 584329 584462 584104 584474 584387 584475 584478 584375 584381 583926 584338 584376 583927 584318 584378 584144 584392 584009 584361 584460 584344 584390 584357 584372 584108 584325 584323 584349 584321 584326 584320 584455 584020 584377 584152 584114 584491 584365 584145 584379 584317 584101 584131 584118 584348 584103 584097 584059 584343 584362 584037 584385 584099 584519 584501 584010 584441 584497 584438 584505 585132 585098 585106 585071 585089 585131 585114 584395 585111 585103 585145 585122 585073 585095 585069 585080 585148 585070 585088 585064 585151 585084 585079 584168 584184 584173 584182 584188 584179 584181 584178 571868 571949 584176 584165 584169 584183 584175 584167 1051871 1098436 1039528 1039519 1034491 1006834 1014497 1101651 1006856 1025230 1018840 1031658 1034487 836810 836842 1030478 1078260 1006818 785240 1004259 1006843 1014481 806099 806080 806087 806103 806119 806113 806102 806093 806109 806086 806108 804000 806100 804008 806106 816990 806105 806095 806097 806094 806096 806118 806090 806083 1025233 806092 806114 806112 806110 806081 806116 1093882 841046 1093892 1056487 1044414 1014495 1093893 1093890 1093878 1062563 1093889 1104359 1059428 1098449 1098441 1070691 1071577 1086605 1098448 1059462 1070704 1078657 1098451 1044387 1051833 1051862 1044412 1044394 1047828 1051853 1082251 1071953 1006860 1106332 1039517 836812 836796 836802 836814 842542 1048918 836798 836808 836837 836838 836811 836787 836774 836789 836801 836835 836800 836825 836821 836797 1014491 836792 836828 836779 1034492 1025209 1051839 1044400 1047822 1048934 1031662 1047826 1044409 1044396 1051852 1059449 1034499 1039532 1025238 1062575 1059431 1031656 1056472 1031654 1056475 1059441 1030475 1030480 1034484 1056481 1034489 1025235 1051865 1030488 1034494 1044395 1047841 1034483 1044406 1044407 1056476 1059440 1044415 1014494 1025234 1051863 836781 1082250 1014468 1025206 813298 1018860 1057975 813310 1071951 1047834 1039535 1034913 1025208 1051850 1071562 1014500 1059438 1051846 1075600 1057977 1025221 1051855 1039518 1071576 1039522 1023708 1051858 1048930 1051840 1070695 4034493 1078665 836775 1071582 1034922 790927 1047838 1057885 1024061 1057980 1057886 1034486 1075605 1070693 808928 1006849 1113762 1057897 1014474 1057976 1057883 836813 1113748 1075620 1057887 1057981 790928 1025205 1024066 1057983 1024075 1069108 1057896 1090757 1104357 1070686 1104358 1104365 1104340 1075623 1071583 1071949 1082244 1075610 1082248 1078668 1071579 1070687 1075633 1075632 1075611 1082253 1070683 1070697 1075637 1070690 1075622 1078258 1075626 1078667 1075615 1070699 1078671 1078259 1070696 1082256 1091493 1091492 1091498 1091497 1091488 1070702 1091486 1091496 1091490 1093883 1091495 1091485 1086602 1093885 1090761 1091491 1091484 1093880 1090763 1039520 1006836 1069104 1004257 1018853 1078663 1004272 1004281 808921 813312 808941 810801 1097774 810817 1101654 810807 808939 810800 808913 810812 810822 808944 810815 1006826 810802 810805 813294 810803 808931 808940 813308 813292 810828 810804 1097767 810816 810826 813293 1018847 813299 813311 810811 810825 813296 813291 813297 813290 813295 810813 1014466 813305 813306 813314 813300 810808 813309 813303 813301 813304 813302 1086608 1006846 1025212 1018842 1014486 836771 1030484 780113 1024057 1025215 1051843 1006812 1024064 1030483 1101650 842548 816207 1030481 1004282 1025226 1044403 1075628 842553 1059429 1024072 1075629 1101653 1004255 1006816 836772 1078660 1004261 1006857 1082247 810806 1024063 1030474 1107350 1113771 814988 1091489 1025232 1059444 1034921 1071956 1047845 1034917 1075640 1018863 790586 1047831 1048931 1025227 1034912 842536 1034911 1034501 1034482 1062560 1024070 1090760 1086601 1090762 1034914 1070705 1069129 1098440 816206 816201 816988 816995 1006815 808911 1018862 1025239 1097766 1097772 1044393 1098437 816209 1071958 816197 808916 1024060 1075603 841058 1075641 1004276 1082257 1030485 1004274 1075607 841052 842541 1034919 1034923 1004280 1004273 842538 1104343 842539 1004269 1004262 1004279 1004249 842537 1025213 1004247 1004270 1047830 1004267 1018852 1078661 1069127 1107360 785255 1047837 1024068 1051873 1004246 1024073 1051866 1051872 1051849 1051874 816195 841059 1051836 1104360 1024062 1059459 1059458 1024074 1024065 1034496 1051842 1034918 1051868 1107366 1047839 1047821 836834 1107365 1051658 1051835 1069120 1069097 1051857 842549 1051837 1034909 836799 1006827 1082246 1059443 1044397 841065 1018846 1113774 1113796 1113787 1113779 1113767 1113765 1113747 1113761 1113758 1113790 1113754 1113769 1113782 1107363 1113798 1113778 1113751 1113766 1113799 1113791 1107362 1104350 1062567 1078662 1078666 1062572 1104341 1104339 1104354 1104355 1104361 1104346 1106327 1104368 1104342 1104352 1104348 1104356 1104345 836816 1051848 1057973 1057971 1057884 1113750 1056471 836807 1025223 1057974 1057891 1069111 1014476 790942 1047842 1051860 1014492 841043 1047820 1059439 1024056 1047848 1044408 1044398 1059437 1057895 1113788 1057978 1057984 1034488 1071944 1057889 1113755 1057985 808675 1025211 1048923 1048920 1086600 1031660 810810 836803 842543 1048921 1014489 1071573 1106328 1078673 1048922 1069105 1069109 1069103 1014498 1014472 1069130 1014503 1047825 1107368 1047844 1051867 1014490 841066 1093876 1059456 1014479 1006821 1004264 842546 816991 814992 814983 814987 814982 814995 814979 1004265 813307 814985 1004258 816998 816981 808920 814978 814991 816982 816983 814994 814981 816992 814984 814980 816985 814986 816997 814993 1048928 808934 816994 816996 814997 814989 814996 813313 1006839 1082254 1078670 1014499 1004268 841041 1031661 836790 1093894 1059457 1056491 1004252 1106326 1047836 1048929 1106324 1062577 1048932 1025225 842534 1018849 842556 1069110 1051654 1078664 1078674 1014477 841042 1059432 1078659 1070689 1051844 1006855 1034920 1047840 1062576 1059446 1098446 785257 790573 790940 790564 790566 790929 785250 790591 790589 790563 790930 785248 785228 790932 790923 790944 790921 790920 779879 790575 785249 1006828 1018864 790926 790580 790935 790593 790917 790574 1097769 785231 790937 790569 785235 790918 785258 790910 790934 790924 790579 785241 790912 790911 1113783 790919 790592 790584 790562 790922 790567 790570 790560 790582 790916 790588 790915 790925 790939 1097768 785229 790571 790561 790941 790583 785244 1014487 1062580 1025216 1014493 1014482 1056480 1057892 1025220 1014501 1106330 1107364 1106323 1107369 1107356 1107355 1109930 1107370 1106329 1107357 1106331 836840 1093886 1093881 1098438 1098435 1098452 1098439 1097771 1098445 1097777 1090756 1098444 1097759 1097773 1093887 841055 1018857 1024058 1051847 1006829 1051864 1051856 1034497 1051830 1024071 1034485 790578 1044405 1034915 1044399 810821 1006837 1059435 1006859 1048925 1025224 1014469 1044404 790577 816993 1086607 816989 1075621 1014502 1062561 1075636 1014480 1004260 1062571 1075608 1062570 1069107 1071568 1051838 1031653 1057894 836782 1062573 1018861 1034490 1071581 1075601 1062574 1071948 1071563 1057893 1062579 1075606 1069134 1069102 1075617 1069116 1078257 1069117 1062566 1101648 1062565 1101646 1006819 1047829 1051834 1025218 1039531 1093884 1039533 810829 1071960 1039538 1039521 836783 1044402 1030482 1025231 1047832 1039541 1047824 810820 810827 836830 790568 1056469 1056473 1034493 841047 1048933 1048935 1059460 1056470 1056489 1004277 1025219 1039534 1048926 1059454 1059452 1004254 1056486 1056474 1056477 1014496 1006814 1056478 1039529 841068 1051829 1051869 836795 1101649 1101652 1018848 1047833 1018854 1006848 1018851 1044417 1018841 806091 1018856 1018843 1006830 836844 1051870 1004256 803999 1018859 808936 1031650 1006822 841062 780124 780105 781687 781702 780137 779891 780125 779889 780119 779864 779890 779875 779860 779861 780115 779873 781682 779874 781703 779867 785226 781681 780138 781686 779884 779898 781693 781690 780121 779870 785238 780133 780114 785243 779866 785253 780129 781684 780143 780128 781701 781696 781683 779863 781689 785247 1113781 785227 779880 785263 785254 779893 780134 785259 780104 785256 785237 779877 780126 781695 785234 780117 780131 785239 1113786 781699 779897 781688 785224 781697 785246 785245 785252 785262 785261 781698 779871 785260 785233 785230 785236 779872 779859 779881 781691 785251 785232 785225 816194 816203 816200 816198 816205 816202 1006853 1047849 816208 1071959 816987 816986 1044413 1018844 1056485 794076 1034916 1014484 1006850 1004275 1004271 1006862 1093877 1093875 1004263 1006823 1006842 1006851 1006833 1004250 1006838 1006832 1006845 1006840 1006817 1006852 1006835 1014485 1051861 1104370 1071580 1078655 1059442 1051841 1024067 1006844 836805 836804 1006847 1006824 1031652 841049 1024069 781692 1004248 1047835 1069126 1069125 1069113 1069112 1075634 1070682 1034500 1044389 1069101 1069121 1069095 1069114 1075604 1069115 1069124 1014504 1044411 1071957 1059434 1006863 1044410 1034910 1044416 1069122 1069119 1069096 1069133 1044392 1034498 1014471 1018850 1071575 1071567 1071570 1071941 1059436 841063 1069131 1104353 1014473 1018855 1025228 1014467 1059448 1093874 1104363 1097765 1107346 1090759 1082245 1098433 1097770 1107349 1107359 1104366 1104351 1104344 1107367 1101647 1107347 1107354 1106325 1104367 1104347 1097763 1097775 1098434 1093891 1104364 1093888 1091487 1006861 808929 1048924 1069100 1048927 1071565 1069099 1071952 1056483 1070685 1006820 1075619 1069128 836843 1044391 1025214 1006841 1070694 1062569 1030473 1069106 1069123 1075613 1025229 1062582 1057972 841072 1004266 1062568 1039515 1059450 1062562 1051859 1086606 836786 804007 1039530 1048919 836784 1039525 1018858 1006813 808923 1062578 808924 808917 804013 842532 842551 836820 836780 836793 836809 836819 836788 1039536 836823 836841 1030487 836770 836791 836777 1039527 841040 836776 836827 836839 836822 1070701 836778 836818 836817 836824 1030476 1039540 1031657 836806 836826 836785 836773 836815 836833 836794 1006831 1006858 1098443 1075616 1071945 1018845 1098450 1071571 1075609 1044388 1014470 1044390 1039514 1047843 1014475 1004253 1025222 1098447 1047823 814990 1069098 836831 1070688 1070703 1031651 1025210 1059453 1006825 1044401 1034495 1014483 1014488 1090754 1004283 810823 806101 808915 808930 808910 808925 810809 808671 806104 810824 806111 1024059 808922 808919 806085 806084 806098 808927 806089 806082 806117 808933 808938 806088 808932 808912 808677 806115 1006854 808943 808674 808935 1098442 808926 808672 810819 804005 808914 808676 806107 808937 1113752 1109928 1113776 1113770 1113793 1113800 1113792 1113756 779894 1113789 1097762 780142 1097761 1113773 779868 1086609 1071942 1070692 1059451 1082252 1059461 1075618 1078656 1075612 1071947 1070681 1059433 1071566 1075638 1062581 1075635 1070684 1071578 1071950 1059447 1059430 1082255 1082249 1071955 1086603 1078261 1075631 1075627 1057979 1078669 1051831 1075602 1057982 1075639 1078658 1075630 1075625 1062564 1071943 1071946 1071564 1057888 1075624 1075614 1070698 1071954 1078672 1056482 1071572 836832 1091494 1071574 1059455 1031655 810814 1059445 1047846 1039524 1056484 790581 1025237 1004251 841064 842540 841071 841076 842545 842552 842544 842533 841057 841074 1030486 841050 842547 841051 841067 841060 841054 1097776 841075 842554 841056 842555 842535 841069 1056488 1030479 1030477 842550 841045 841070 841048 1005286 1005287 818746 818748 818751 818753 818749 818747 830446 830416 1118339 1005291 838029 1007863 818316 790712 800343 838018 827551 827553 804733 804724 823285 1007850 804731 1005306 838022 1005294 818333 823283 1007849 830439 1007862 818736 818734 818732 818331 818325 818733 827548 818326 818324 818327 818329 823286 827542 823296 830420 823293 823274 800316 800334 800351 800336 800371 800367 800325 800368 800332 800365 800323 800329 800313 800322 800331 800320 1005289 1005273 800312 800335 800373 818738 838014 800578 830414 830443 830429 796017 800359 800358 827556 827558 827547 827561 823295 800338 838024 800310 800339 800363 830433 830444 830432 800370 800324 800577 800362 838028 838013 843621 1046205 830441 843617 830424 843616 843619 843615 843635 830437 843629 818745 823269 823266 827554 827543 823272 830419 823300 830430 1117624 1005304 1005293 800305 830425 800350 800352 800369 800308 800347 800574 800342 800307 800306 830417 827562 800571 800340 1005274 823299 823282 823297 823265 823267 1019260 796006 1007851 1007846 1007847 1005302 1005280 800337 800344 830447 818743 830422 830421 823281 823280 800326 804729 800580 800333 800372 800317 804735 804732 804734 1007867 804718 804730 1007857 800345 804717 804722 818318 1007865 838011 818322 843624 818752 818323 804726 843627 800361 800309 800349 800356 818740 838020 838025 1005271 1005269 1005288 1046202 830436 830427 1117621 1118343 1109827 800327 800354 818737 800573 830428 823279 818317 827550 830442 818742 843618 843620 843630 1005292 800576 800318 1007855 823277 823270 823292 823288 823289 823264 818315 790716 790707 795991 1005281 827564 1005307 795616 795603 795592 796002 827555 796020 795618 827567 795990 796007 827552 794872 827559 794871 818321 794157 827557 827560 823278 827544 827563 818320 830435 1048237 830746 796012 1086437 1045666 1008047 1019436 1045651 1039403 1068801 1053319 826534 1028657 1028652 1026221 1054670 1030587 1039392 1054610 1028655 1039387 1007870 1039404 1039372 1039390 1039389 836532 1054669 1002118 1074154 1058230 1074142 1086447 1086630 1096228 1086478 1095939 1086637 1086460 1085217 1086459 1063722 1002148 1002101 1045657 1095949 1086465 1096208 1078639 1056611 1016016 830739 1035532 1039373 1095009 1063725 1048286 826574 818936 821527 821498 1008063 831031 1062770 1086479 1086614 836552 1003705 1056620 1063719 1001926 1056607 842638 1048259 1080932 1013678 1081376 1019262 1081373 804377 801597 802802 804382 802906 801591 801572 801596 802808 801582 799749 801570 802806 802905 802789 802918 803175 803285 801592 803164 802921 802915 799068 804375 799732 801581 803301 803299 802917 803161 802793 801587 803173 802810 803176 803295 802807 799705 795998 802922 802902 802900 802908 802812 801595 802923 803165 804374 802909 803178 803155 802910 799034 802805 799158 801583 804388 1030580 803163 803166 802904 801574 804386 796027 802795 799720 802804 803159 801577 795979 802794 802920 802798 803281 803167 803283 804393 802913 802792 803156 803296 802813 804384 802911 803172 803157 803170 802901 803171 801593 801573 803280 802800 801571 803303 803300 803282 803154 803288 804391 1056596 803286 802799 802912 802796 801578 801594 803294 799041 803290 803291 803289 803284 803279 801589 803160 803302 803162 801585 803297 801584 801590 801588 799717 801576 799747 804385 798371 802899 803174 803287 802791 803177 1009051 1007033 1008033 830754 1009091 1008060 1009057 1008032 1009074 1009068 1009083 1001944 1009077 1100233 1009075 1063716 1008062 1009056 1016065 1034369 1006991 1008065 1007000 1009088 1008061 1060400 1009063 1009085 1008046 1008049 1075858 1053331 1008057 1009064 1009066 1008051 1008039 1009069 1008045 1009071 1009070 1009090 1045674 1008038 1006997 1008029 1009061 1009078 1009062 1009094 1009052 1009073 1008056 1008043 1078645 1009084 1009076 1016073 1007866 1048229 1101664 1009055 1009080 1008031 1007034 1008034 1008054 1008050 1008028 1008044 1057194 1009059 814082 1009058 1009081 1019397 1085219 1056589 1080941 1063718 1101665 1025447 818896 822641 822594 822637 822610 826551 818899 821537 822645 817768 822612 822606 822611 822614 822634 822584 822586 822639 826601 822625 822616 821538 1044572 822596 1034367 822636 822595 1030583 822590 822635 822609 1034371 822587 822585 822605 822599 822589 822592 822601 822598 822591 822627 822618 821516 1044585 822633 821551 826607 1048292 822640 821508 1057157 822597 826528 822600 821545 826611 818944 1045658 821500 822615 1019426 1087481 822624 822617 822626 822632 822593 1035531 822582 822646 822603 822608 822604 816577 821499 822644 822631 1019389 826561 822643 822620 822630 1045670 1117627 1117619 1117628 1117633 1117622 1109838 1117629 1117620 1117636 1117617 1109822 1100255 1054616 1019240 1016071 1086632 1051736 1008059 1014743 1093049 1034414 1063720 1095960 1019717 1077064 1079933 1063742 1016006 817781 831021 835051 830719 830727 830734 1009089 830717 830729 830987 835067 1009092 835026 835050 835027 831018 830726 835061 1008035 830713 835037 830991 835032 835066 835054 830738 836519 1007028 835030 830722 831032 1068804 830748 831036 830743 835048 835041 830720 835036 835031 830985 835046 830733 835071 835074 836539 1044550 835034 830712 830728 831009 830983 830990 835059 831028 830753 835056 835033 831022 831037 830736 835028 835057 831001 831035 836524 835058 835072 835064 831033 831010 830742 830730 830745 835040 830995 835039 830744 830716 831019 830731 1075712 831005 830740 836540 830735 830737 835043 831023 831003 836518 830747 831029 830725 830711 835053 1062756 1048260 1053318 1060383 1060393 1063730 1057160 1051742 1070084 1057846 1054674 1056622 1049536 1054622 1044552 1048257 1044578 1056588 1062779 1062784 1045650 1057847 1062771 1048225 1057195 1044594 1030604 1057180 1063661 1045700 1035541 1063715 1060376 1062772 1056624 1059755 1044596 1048242 1048304 1048234 1056609 1063669 1062757 1054620 1053006 1035542 1045655 1044579 1062763 1053320 1059768 1056616 1056590 1063658 1057184 1063655 1060379 1044581 1060399 1060409 1062740 1063707 1057189 1059740 1051739 1054681 1052999 1024400 1045683 1095948 1035553 1093059 1016010 1074145 1096227 1095933 1069757 1062767 1093062 1085227 1096229 1093058 1093043 1081369 1063749 1019435 1019716 1019733 1030613 1028631 1028644 1080959 1070090 1019245 1024395 1024404 830758 1024412 1056587 1075720 1024416 1062746 1048283 1048280 1058228 1048281 1056604 1044600 808202 1048223 1058233 1062768 1080922 1058387 1060386 1062744 1057848 1059726 1078603 1028609 1048262 1057845 1057843 1049548 1074140 1019239 1048311 1060402 1028650 1024411 1062788 1058217 1056595 1026226 1075874 1074161 1051725 1056580 1019685 822622 1002137 840913 1048244 1070101 1060391 1024388 1045697 1045663 1057159 1060382 1070095 1019263 1060387 1044554 1048273 1056608 1060392 1078613 1078641 1054231 1078621 1019394 1039388 1078614 1034421 1045681 1078634 1039376 1078605 1078610 1078640 1054245 1028632 1098511 1019703 1098646 1078631 1093052 1060377 1063651 1059724 1093087 1078628 1098519 1078629 1019679 1098503 1026219 1008048 1028618 1078648 840885 1078644 1078627 1075714 1086631 1075706 1075882 840898 1078642 1078612 1019437 1078618 1019735 1019440 1056592 1075843 1063710 1034419 1074146 1054221 1062769 1056633 1044545 1078630 1078624 1030603 1019424 1054242 795620 795626 1002099 795999 796028 795579 795595 794141 794857 795974 795617 789241 798364 794135 795599 795625 794854 794847 794853 794841 795604 794858 795982 795605 795573 796015 795984 795628 795606 794846 794147 795988 798376 793210 795981 794863 794867 794137 794842 795583 794852 795588 794850 794851 795585 795622 1001939 795587 794836 794843 795571 794139 795973 795619 796000 1056630 794129 794869 794838 795627 794150 794855 795572 796016 793352 796014 795623 1087487 795576 794874 795975 790717 794865 794155 793365 795615 795970 795586 795589 793212 795578 1077053 788936 795978 795987 795989 794859 791455 793214 791464 790728 791438 795570 1035544 795986 795575 795593 794145 794128 794132 795597 795594 794125 795612 794122 795601 795596 795621 794864 796001 795613 794856 795608 795591 795569 836527 780307 794146 794875 795590 795997 794156 794159 794861 796003 794844 794870 794849 795994 794143 794123 795584 795598 794131 1034375 794149 791165 795611 795581 795577 795980 798366 794126 795995 794153 795574 794845 1059770 1045680 840829 1070096 1081372 1025429 1039374 1086462 1060394 1045690 1054682 1019241 1054677 1034415 836522 1025420 1026233 1025442 1007013 1063653 1054675 1039378 818895 1016874 1024387 1075724 1051721 1051729 1075707 1051727 1060375 1045678 816575 1081371 1019447 1019705 1045687 1054611 1069758 1054612 1054621 1059744 1075881 1051720 1045696 1048254 1009086 1028607 1048289 1048285 1003700 835029 1039383 1039409 1026220 1019694 1013688 840847 1035552 1058219 1085225 1060390 1111492 1026206 1026225 1054608 1086441 1051747 1079929 1074130 1045692 1056629 1059750 1019732 1086461 1093120 1062752 1019396 1030582 1030589 1059759 1028621 1075868 1081380 1079916 1068774 1080934 1080927 1078638 1074143 1079934 1080957 1078653 1085228 1069763 1069771 1082679 1069759 1069761 1068773 1082671 1075846 1074158 1079931 1074138 1078654 1081370 1074148 1079936 1081364 1075722 1074131 1074157 1080925 1075872 1075845 1080952 1078625 1075850 1068781 1075839 1081379 1070087 1080945 1058221 1080937 1081381 1068789 1085220 1081375 1063732 1080935 1070088 1074132 1081374 1079913 1075713 1075880 1068795 1086636 1086634 1077061 1086621 1093094 1093082 1093061 1086470 1086629 1093119 1086623 1093042 1093067 1078616 1087478 1080944 1068772 1093044 1093063 1093054 1093093 1093079 1093077 1086626 1086635 1086639 1086473 1086471 1086433 1085224 1093046 1093045 1093100 1086615 1093104 1093117 1086627 1093056 1087477 1093105 1093084 1085226 1087482 1086442 1086467 1093074 1086446 1086469 1093114 1093081 1093069 1093112 1093089 1028648 1069774 1068799 1028599 1093103 1086476 1034396 1093102 1086475 793369 1057186 1034422 795983 1006999 1014735 1086620 1057171 1100242 1051740 1093113 1080954 1034377 814136 813088 814140 808874 808873 808889 813252 1101663 814077 814141 814142 813238 814151 814147 814137 813087 813085 814156 814058 813237 813251 808871 814057 813232 813239 808876 1045689 813107 813076 813225 814056 813102 813069 813245 813235 813068 813109 813230 813070 813257 808381 785044 814068 1063659 813067 808357 813250 813234 814699 813253 808880 814146 807693 813064 813219 808886 813254 813103 808371 808872 808360 813220 813244 813098 813228 813074 1045659 813084 814163 813223 808870 813231 814085 813091 813075 808888 814073 813243 813096 813229 1063728 813083 813078 813092 813063 813259 813211 813242 813241 814064 808882 807706 813226 807696 813089 807893 813072 807692 813086 808864 808388 813100 813256 1062751 1077062 1068767 813258 813224 813218 813090 813216 1087483 813080 813101 813213 808883 813066 813240 808879 813110 813255 808860 813081 808365 814084 808196 814159 813104 1063733 1087479 1007029 830724 1063657 1014738 1019430 830993 1014723 1019687 1019393 1024408 835069 822621 1062780 842620 818871 835049 1019707 1019390 1015995 1026236 1019730 822628 1044563 830751 1017065 795992 1002107 1098505 1017063 795580 796004 794120 830989 1056599 795969 1025419 1062741 1015990 1099212 1019683 1019701 1019409 1015455 1100251 802801 1025424 1007010 1099198 1098647 1019723 1016041 1048277 1013727 1099200 1002133 1100514 1019709 1048258 1028605 1093086 1039397 1008053 1019419 1028637 1024402 1075841 1113847 826555 826569 1002163 813247 826585 1016040 836537 1034373 1078601 1090767 1074139 1085218 1074147 1078619 1074159 1078600 1028598 1028633 1024414 1025423 1035550 1056584 1078620 813097 1044541 1025439 1035535 1049549 1078646 1062748 1059728 1081365 1054673 1044589 1044588 1025417 1049539 817770 1078608 1048298 1035549 1013715 1039399 1063665 1048250 1048268 1035524 1078622 1045646 1068777 1025418 1060398 794140 1026223 1098521 1078598 1059716 1045652 1056631 1078647 1025430 1078626 1086618 1034413 808868 1044576 1063747 1049544 1048267 1056601 1056637 1034417 1025437 1026212 1049540 1024391 1051730 1078623 1056626 1085221 1056581 1019243 1090765 1025421 1053333 1063738 1019678 1034397 1026211 1059754 1028606 1069766 1051714 1028619 1079917 826578 826532 826597 826553 1063744 826541 826529 826562 826612 826537 826549 826527 826535 826530 826587 1075709 826552 826557 826598 826576 1054215 826542 826595 826539 826603 826580 826548 826554 826558 826602 826531 826526 821540 826579 1057849 826564 1054614 826568 826599 826547 826593 826581 826545 822629 1096202 826544 826550 1024394 1075851 826591 826559 826609 826588 826604 826536 826592 1096205 826575 1060381 826572 826583 826590 826582 826556 826605 826560 826594 826573 826589 826610 826540 826565 826546 826586 1057197 826571 826570 826538 1028626 826577 826608 1044598 1048232 1062749 1086444 1016072 1080956 1059720 1024419 1024415 1078615 1094989 1048221 1014715 1024410 1030597 1028651 1026229 1080936 1068808 1039371 835044 836531 842616 822623 822638 1002121 1001937 1003693 1103272 1048314 1002097 1001948 1002117 1002142 1016019 793223 1002161 1060389 1002128 1093076 1003711 1002103 1068779 1002139 1002123 1026217 1001927 1003690 1019414 1003701 1001930 1002116 1002098 1002096 1002122 1002110 1001925 1002115 1095011 1002109 1001945 808384 802907 1002138 1094998 1002135 1001929 1001932 1001924 1001935 1002141 807877 1002162 1094992 1001941 1002134 1019688 1019698 1003702 842644 1048246 1001942 1070098 1002124 1003715 822602 1001949 803292 1002164 1003697 1015440 1001923 1100254 1080948 1001947 1002154 1002149 1002132 1002140 1002145 1002151 1002125 1001922 1002126 1002152 1002144 1002102 1002153 1002131 1002108 794121 1002106 1002157 1002158 1002119 842639 1002143 1001940 1002156 1002136 1002113 1002146 1056583 1019445 1002159 1003706 1002127 1003694 785062 1001950 1003704 1093098 1025451 1001934 1016056 840921 1001928 1003698 1001943 1003710 1003714 1003692 1074137 1003713 1002160 1030572 840866 1014706 1059753 840832 1001938 1014710 1025428 1026218 1013709 1077056 1057187 1059733 1016877 1057169 1026208 1075835 840930 840905 840839 1013711 1093080 840927 836517 1098531 1045661 840842 1008064 1059764 840882 1062777 1048239 840828 826533 1074136 835073 1074134 1056625 1063724 1056594 840937 1063668 1087476 1048265 1048247 1002165 1075704 1008041 1049541 1049535 1087484 1056610 1059765 1077051 840938 1045662 1087486 1069773 1003695 840908 1013691 1019724 840854 1009072 1019427 1059766 1009065 831025 1030574 1048256 1030595 1024405 1030592 1030569 1015991 1019706 1016000 1019719 1030593 1030607 1030614 1048236 1048287 1016055 1028608 1035526 1044546 1028643 1044543 1057165 1063652 813222 1063734 1052995 1048291 1016067 1056614 1062766 1057166 1060411 1070093 1075864 1057842 1063706 816546 1014740 1057844 1070100 1049545 1093083 1063663 1063660 1113854 1063723 1002130 1075838 1025441 1051715 1063717 1024393 1035543 1002155 1019404 1062754 1051723 1062785 1111481 1095976 1113825 1113862 1113860 1113824 1113842 1113846 1113823 1113856 1113853 1113834 1113840 1113852 1109829 1111495 1113831 1100519 1113844 1113841 1113829 1113832 1113822 1113838 1113836 1113839 1113857 1113861 1113851 1113859 1113845 1113833 1113850 1113826 1113863 1113828 1113837 1060396 1093050 1080950 1049142 1057175 1053323 1093065 1059741 1057163 1086477 1059743 1093060 1093057 1074144 1045691 1045656 1093091 1093048 1075853 1079922 1068809 1080955 1079935 1100520 1100235 1098662 1103247 1098536 1103252 1103253 1103259 1103248 1103265 1103257 1100245 1101667 1104709 1100525 1100536 1100518 1103258 1100252 1100535 1103255 1100538 1100527 1100234 1103262 1103254 1103267 1103268 1103261 1100256 1099202 1103274 1103250 1101673 1104710 1101666 1100246 1103263 1100253 1101672 1100236 1103260 1096214 1103266 1100238 1103256 1100541 1103245 1098500 1100257 1103269 1100230 1100247 1101662 1101671 1100241 1100513 1103264 1104708 1054248 1060397 1019410 1058227 1075708 1079930 1007856 1024398 1035525 1019434 1028624 1005303 1086452 1060384 818941 1070085 1034383 1019736 831002 1045682 1094994 1016018 1058388 1058222 1086430 1019729 1086610 1062791 1075837 1058226 1045667 1058391 1058218 1058390 1058231 821525 1019702 1003699 1079924 1060404 1028603 1095012 1034380 1095013 1045649 1045660 1024396 1058386 1002111 1087485 1028610 1034376 1034388 1026209 1034372 1058225 1075710 1068793 1048288 1045673 1045653 1060395 1054619 1093090 831027 1060372 1028617 1035548 1044583 1060378 1053322 1075721 807879 807705 818905 1070094 1075702 1086443 1056602 1014722 1095934 1048305 1053001 1054227 798358 1056600 1068765 1096209 1049547 1045644 1045672 1075865 1096225 1068794 1048308 1045671 1048296 1013725 1028629 1075867 1059722 1039381 1068769 1054606 1085222 1075836 1014708 1016066 1054607 1063746 1075717 1007869 1048312 1074160 1078602 1078611 1077063 1081378 1086622 1086455 1054239 1093073 1086431 1054671 1078599 1095962 1002112 1095974 1093108 1079914 1044575 1054613 1086625 1063752 1095965 1086457 1086617 1019731 1086468 1095935 817748 814701 816509 814075 816580 816552 1030591 814071 808869 817762 814702 813246 814148 814723 814704 814720 816518 816533 814079 816516 816539 816586 808875 814150 813105 814711 814707 816578 816582 817753 817744 814066 817759 813215 816563 816508 814722 814698 814161 814714 814065 816519 814164 816561 816564 814078 813062 816569 814725 813217 816562 813099 814726 1030581 814063 814160 814718 817752 814152 813227 816548 814165 816547 814713 814080 816565 1030588 816530 817772 817755 816534 814157 814153 816517 817776 816535 813248 814162 817743 814139 816541 816523 816566 816560 814061 816520 816521 814727 814710 817754 816540 808884 814721 816559 808881 814709 817750 816513 808866 816568 816545 808887 813061 816584 816558 816514 1017067 814144 817747 816570 816532 817774 814062 817742 814081 814724 814067 814069 813233 814060 814719 817780 813214 813079 816536 808865 814712 816542 816557 814149 816524 808885 816544 1054618 813082 814158 816510 816522 814072 813212 816550 816579 817769 816507 813093 1039400 813249 816554 816585 814083 814143 813108 1030584 816543 814715 816515 814070 808863 808877 816529 814076 816528 816511 1039377 814706 814716 816526 1039406 816531 814703 814708 816512 817745 816527 813073 1093096 816571 814074 816576 816590 813077 814059 816537 816581 1053004 1001931 1095971 1095957 1062783 1025443 1075844 1098529 1051728 1024413 1056627 1063741 1044544 1049146 1048226 1025438 1060403 1080931 1075866 840862 840835 1059769 1045699 1056603 1016001 830980 1075859 1049145 1075862 1030601 1051735 1019713 1059748 1086436 1059767 1059729 1028602 1039407 1060388 1059736 1015999 1062773 830988 835055 1025432 1062765 1049144 1016062 1028613 1003703 1069769 1014741 1048300 1063731 1075711 840851 802797 840902 1059760 1075726 1075723 1062747 1025431 1034391 1025436 842615 1075703 822588 1078597 1075705 1049143 1070099 1079926 1017068 840893 1035545 1034404 1080929 1056605 1014729 1030611 1044560 1051734 1034382 1034407 1019727 1019421 1016030 1014728 1034393 813221 1074153 1060406 1034378 1051718 1016074 1017074 1068802 1017070 1034424 1028656 1014721 1045668 1051737 1045679 1035537 1034395 1014742 1014726 1019422 1069760 1035530 1019696 1034384 1048266 1051731 1009054 1013718 1070089 1014717 1014725 1048303 1045701 840837 1057156 1019734 1087475 1051712 1034425 1007005 1056619 1093066 1093101 1093095 1095946 1093068 1025446 1095968 786805 791185 790704 786803 791171 790721 791177 786651 786822 790711 786310 788927 787963 788926 786652 789260 788910 787960 788602 780287 785469 788913 791449 790725 780040 789267 786816 787949 788589 787976 789249 791172 787967 789243 788935 788912 791169 789259 787955 789251 789261 788917 787975 791170 789244 787959 788922 788606 790708 788603 787961 787952 790726 787978 785467 786799 789268 787977 788582 788933 787950 787957 788604 786645 788931 787972 787966 788598 788588 786292 788925 790722 787970 788930 789247 786299 791164 788915 791186 786309 788919 791160 787969 786818 787962 787953 789263 789262 790727 786306 1060407 786647 791176 785480 786801 787971 786800 787964 790734 789240 1014712 788914 786808 788592 789264 791188 786307 786636 791441 791435 1026234 786637 788594 788937 791190 791189 788916 786625 790731 788607 791442 788924 789253 788608 789258 789252 785478 787954 788932 789256 788918 790733 788934 789269 786806 790724 789255 787951 787965 788923 791439 788920 791181 791447 786795 788581 791178 788586 786819 788590 789248 791191 786632 787956 791174 788580 786810 791161 786817 791443 788929 786821 786807 791194 791193 788939 790732 790701 786820 788599 788605 788938 788921 788596 785073 788601 787973 790710 790709 786648 786627 788583 790706 791167 781566 788609 786291 788591 786811 786809 789246 787968 788911 788584 786796 788597 790723 780008 788595 790713 791168 786815 788593 787958 791175 789257 790705 791436 788587 786813 786628 788600 1045693 1013702 1016063 1013733 1016035 1074151 1016058 1101669 1013693 1013720 1059723 1016052 816567 1079918 1016015 1016036 1016033 1100524 1013722 1013694 1014736 1013706 1014709 1016007 1013713 1016868 1059745 1013703 1059749 1016023 1014744 1013732 1016057 1014737 1082674 1013704 1016038 1013685 818903 1014714 1013708 1044573 1009053 1016003 1101668 1013689 1013696 1013676 1013684 1015997 1013681 1014734 1016026 1013731 826543 1014718 1016047 1016009 1013700 1014733 1017072 1014731 1013701 1016037 1013695 1016043 1013719 1013716 1013728 1014732 1014724 1016051 1016050 1044592 1016054 1016049 1013714 1016064 1015992 1068807 1016042 1016002 1013683 1016061 1013712 1013682 1016044 1016024 1016032 1016014 1016059 1016008 1014720 1093070 1013692 1111496 1111486 1111479 1111480 1111474 1111482 1109824 1096222 1111485 1111478 1111475 1111487 1104714 1109828 1109834 1096217 1104706 1111477 1109837 1109831 1111491 1109820 1109833 1111494 1104712 1111489 1111484 1104715 1104707 1111490 1109826 1096220 1109825 1052996 1096215 1051741 1034381 1063671 1075852 1095961 1028658 1086613 1026227 1098523 1096207 1098532 1098644 1098502 1094991 1098524 1098648 1096216 1099195 1095008 1090764 1100526 1095004 1099207 1094993 1098530 1098645 1095964 1098656 1096194 1095005 1098498 1094982 1095954 1095945 1098522 1095970 1098520 1098506 1098499 1095973 1098516 1098650 1098654 1094996 1098508 1098537 1094980 1098526 1095007 1093071 1099196 1099206 1099209 1098657 1098660 1099205 1099214 1096199 1099201 1094979 1100515 1094984 1096226 1094997 1098504 1100539 1099197 1098652 1099208 1099199 1099203 1098514 1100237 1095937 1095955 1095967 1095977 1098659 1099204 1096196 1099213 1095972 1095001 1099211 1098517 1096195 1094999 1095969 1100232 1100258 1093111 1098513 1098651 1098515 1095006 1095002 1098649 1095956 1100540 1098663 1098501 1095950 1095942 1095963 1098655 1098525 1096224 1096231 1096206 1095938 1096218 1098661 1096210 1098528 1098509 1095952 1096213 1100516 1095003 1096193 1094990 1095947 1096211 1095940 1100250 1098518 1100248 1096221 1096200 1095010 1093115 1063737 1094995 1048313 1057164 1086456 1086464 1051717 1045648 840939 1019442 1069776 1051732 1079927 1060401 1053002 1019700 1030571 1053005 1014745 1016070 1014716 818907 1024386 1019691 822583 1045647 1014719 1093085 1087480 1016870 1016017 1007848 794127 817777 816574 816555 821509 818931 818893 821566 818919 816556 818915 818872 818918 818898 816553 816588 816589 818950 1062760 818869 818939 821557 818932 818937 821526 821546 818906 818925 818874 821544 817782 821534 818940 1007861 818945 816593 818876 818885 818920 821518 818938 817773 818933 1053335 818888 818929 818926 818881 818873 818890 821555 821511 818954 816583 821524 821561 821510 1007845 817771 821567 816551 821506 821542 821562 818870 818916 1096230 817765 818908 817787 818864 818930 818913 818879 818886 818923 817756 816525 821521 826584 813106 818882 818928 817779 818883 821517 816587 818901 818914 818863 818946 817788 818947 818877 821528 818891 818942 816549 816592 814154 818902 821512 816572 817751 818904 814705 818897 818875 818868 817786 818934 817749 818956 817767 821559 818922 817763 816538 1039382 821539 821554 821548 818935 817790 818862 817746 816591 821565 821507 821547 817764 818878 818900 818887 817789 821549 818880 817785 817784 821533 821530 817760 821513 821504 818894 818865 821523 821501 821502 818866 817757 821541 818924 818955 818892 818951 1025425 818948 818910 821520 821552 818952 821531 821543 818867 1019392 821536 821556 821558 821515 821505 821550 821535 822619 818949 817775 822642 821529 822613 821563 817761 822607 817766 817778 818921 1034374 1062753 1093075 1063721 1063754 1059732 1093097 1044595 1035547 1045676 1019704 1028600 1048249 1048295 1048224 1056638 1068771 1056618 1019725 1069770 1019443 1060410 1044586 1019418 1063726 1019388 1068766 1034426 1093106 799731 1080924 1058220 1080921 1080933 1056628 1056593 1080939 1059756 1068768 1080926 1094986 1080943 1095000 1068800 1068797 1052998 818889 1054225 1044577 1057161 1057192 1057183 1028601 1045675 1044547 830992 1044557 1057181 1053332 1062764 1044549 1044570 1044566 1057193 1054249 1054226 1054222 1019417 1054235 1019712 1078617 1048299 1034401 1045686 1034409 1075869 1034420 1068796 821564 1048240 1057190 1075719 1057182 1028659 1028612 1057188 1057176 1068791 1049534 1060380 1070086 840871 1069767 1113855 1053321 1054243 1049542 1057196 1044574 1034412 1019439 1068783 1075855 1057191 1057179 1075871 1057185 1048272 1044556 1024420 1026210 787974 1026231 830741 830721 831020 808216 1063714 803293 1079920 835045 821522 803168 818953 1054605 1094987 1063753 1026214 1057162 1026232 1030602 1048302 1086445 1086624 1048231 1054223 1048235 1048297 1016029 1039395 1039408 840880 1054233 1048243 1056636 1026224 1039393 1048241 1056613 1019697 1048248 1039396 1039370 1048307 1003708 1001946 1068784 1048293 1054238 1086638 1013729 1014711 1034423 1054246 1039379 1063729 1016045 1056579 1034398 1028622 1080946 1039375 1048306 1039386 1019433 1045645 813260 1019444 1019408 1019416 1028646 1016871 840890 1013717 1028636 1028641 1017066 1019718 1001936 1026228 1016039 1016022 1016034 1019406 1079915 795610 1009067 1049141 1016060 840876 1019680 1008055 1016031 1015998 1019407 1019728 808363 1014739 1028640 1019722 814138 1019423 1019431 1019715 813236 1100240 1016068 1016053 1100521 1017071 1100534 1019413 814155 1006994 1048274 814145 1019432 1063670 1019395 1019686 1019412 1015441 1034402 1028623 1002120 1013710 1048309 1019399 1019405 1057158 1034385 1028634 1053329 1013686 1019695 1034400 1054230 1048294 1019402 1034418 1019714 1028635 1019682 1001933 1009087 1075715 813071 1034405 1026215 1053334 1053326 799751 1024401 1052994 799761 799752 818884 799733 1019710 1051745 1077069 1054676 799706 799762 818917 799757 1113849 1028596 1034408 1028620 1028649 1008067 1019391 1054680 835035 1008036 1002129 785061 786295 785052 786640 786644 786631 785043 785441 786804 785479 785454 785074 785055 786629 781568 835038 785472 785054 785461 785049 785048 780005 785483 785047 785452 1045695 785442 786288 785463 1054617 781558 785468 785038 781414 785464 780020 785457 785042 785462 785453 785031 786284 785030 785434 786313 786301 786290 785481 780273 786642 786302 786797 786802 785051 785066 785470 786293 786296 786311 786285 780282 786305 785064 786304 786812 786303 786635 785476 785466 785033 785072 785050 786638 786308 778191 785032 785482 781409 785450 785437 785060 785041 786300 786294 785053 786794 781561 785063 785459 785477 785067 785076 785075 785448 786643 786633 785471 785449 786287 780031 781419 785058 785039 785034 786297 785440 786653 785458 785456 781570 786286 785065 785036 780033 781562 785045 780310 781557 785438 785445 835042 780276 781556 786634 785439 785029 785056 786798 785057 785070 786626 786312 785455 785473 786641 785071 785035 786814 786650 785443 785068 785451 785040 785447 781418 785059 780283 778301 780039 786624 785037 785446 785069 786630 785077 786649 785460 750344 785474 785435 831008 830752 821519 830749 826563 826600 831016 831000 830715 1035551 831015 821532 831007 830723 830760 830999 831012 831024 831004 1026216 830759 1057167 831014 830997 826596 830757 830978 1016048 830994 1034389 831006 830981 830755 830732 1034392 830979 830998 830982 830750 831017 830718 831030 798349 798352 802914 798360 798350 799145 799043 794839 801575 799708 799750 799146 796026 801569 799753 798361 798367 799164 801598 799050 799740 799756 795996 798369 799741 799079 795976 798357 798355 799063 799061 799036 798354 799053 799162 799149 1030578 1075879 799090 796019 798374 796024 799059 796005 799044 799040 795977 798372 799052 798356 799159 796021 799156 799084 798359 801580 799758 796023 799056 799144 799150 799707 799727 799724 799080 799072 799086 796013 798353 798377 799046 799093 799725 799081 1070102 799083 799160 799739 799065 799042 798370 799087 796011 799710 799728 799077 799148 798351 799152 799078 799037 799171 795993 798365 799047 799082 799147 796022 799745 799074 799737 799089 799163 799039 799161 799172 796008 799057 799092 799049 796009 799051 794866 799153 799154 799730 796010 799721 826566 799070 799173 799045 799055 799054 795985 798363 798378 799064 799743 799062 1094981 802811 799170 798373 799168 799067 799091 799066 799151 799073 799709 799167 799038 798362 799715 799726 799071 796025 799069 799718 799714 799763 799712 799088 799085 799734 799075 799736 799166 799165 799742 799713 799735 799754 799729 799711 799058 799755 799048 799157 799169 799746 799748 799060 799155 801579 794142 801586 798368 798375 795971 1007024 1007035 1078635 1006984 1048228 1007015 1028597 1100243 1007017 1051733 1078633 1007031 842634 1093088 1078650 1007021 1007012 1007026 840924 1007019 1006990 1006995 1086454 1053330 840884 1006985 1007025 1007018 1002105 1081367 1006993 1003707 1007020 1007006 1013697 1019693 1006998 1001921 1078643 1007022 1006988 1007011 1039401 1007027 1101661 1003712 1007002 1054609 1006992 1007014 1007003 1006987 1045677 1007036 1003691 1002147 1006986 1007030 1007009 1007001 1007037 1024417 1007038 1007023 1019428 1007008 1006996 1007032 1007004 1007007 1007016 1039384 1051746 1035529 1054232 1039405 1051724 1035538 1096203 1096192 1082677 1056598 1094985 1086458 1051713 1086628 1095959 1077057 1054224 1048251 1094983 1086448 1075834 1054672 1014707 1028625 1045664 1028645 831026 831013 1002104 1002100 830714 802790 840919 1002114 1013698 840886 1019235 1100517 840901 1016869 1028638 1019446 1006989 1013680 1078652 1016005 1015994 1078637 1028653 840877 1019681 1044597 1117640 1051722 1062755 1079919 1063727 1051726 1063751 1117631 1063667 1044561 1019690 1035539 1015993 1039385 1059734 1045685 1117626 1051716 1035540 1030608 1016021 1087489 1048269 1068770 1068778 1051719 1053324 826606 1044555 1048282 1013723 1013699 1054229 1013707 1048252 1063666 1054247 1013705 1044567 1048276 1044542 1044553 1075847 1044558 830996 1044587 1025426 1054217 1013674 1057177 1028628 1044580 830984 1063709 1054236 1048227 1016004 1039380 1028630 1056597 1019400 1013730 1063735 1028642 1013677 1103270 1059758 1026222 1035534 1068782 1059757 1068785 1068775 1019689 1062789 1082673 813065 1063656 1062790 813095 1077055 1082676 1030596 785475 1063673 1077052 1030594 1086472 1086435 1078607 1045654 1103271 1075849 1063748 1077059 1016046 1044569 1059751 1086449 1059742 1053316 1075856 1062774 1053327 1059747 1059752 1059761 1063739 1069775 1026213 1054241 1082678 836520 1059763 1063713 1062781 1078632 1068790 1015444 1078604 1078596 1074155 1048270 1059762 1014727 1015996 1024397 1062759 1086451 1052997 1100522 1054220 1077060 1019425 1054244 816573 1059725 1013675 1077065 1062743 1053328 1019726 1059731 1068806 1059730 1045698 1048278 1060373 1062739 1063755 1063675 1117639 1062787 1063711 1109835 1099210 1111476 1100528 1093092 1090766 1093064 1093110 1095966 1098653 1100537 1086463 1096204 1100239 1096198 1093055 1100231 1096223 1109830 1086466 1103246 1098527 1104713 1093053 1109821 1101674 1109832 1101670 1095936 1100244 1100523 1098534 1098512 1096201 1103251 1103249 1095951 1098533 1048233 1048284 1026235 1048271 842609 1048261 1080942 1078651 1051738 1078636 1095944 1030598 1054615 840881 1056585 1096197 1044548 1063712 1078609 1075848 1063740 1079923 1028604 1059727 1095953 1019257 1045703 1086453 1034370 1059737 1082680 1048275 1019708 1048290 1039402 1034387 1002150 1030612 1075840 1070091 1079928 1086434 1045684 1063743 1070097 1008037 1098535 1016873 1063664 1062762 818911 1062792 1098510 1053317 836528 1062758 1044568 1063662 1057178 1053000 1057173 1019737 1054679 1056582 1054219 1034410 1034368 1030570 1062742 1035536 1044564 1062750 1048230 1058224 1048301 1075718 1034394 1070092 814700 1016028 1030586 1044562 1030600 1060408 1026230 1051744 1049537 1044565 1034403 1034399 1049538 1044591 1035546 1063654 818943 1034411 1098658 1074141 821560 1019420 1019711 842645 821553 1048264 1019438 842643 836510 836550 836508 836535 836548 836549 836511 836515 1013679 1017075 831011 840834 836551 836512 840907 836536 1016875 836545 836525 1080923 836521 836534 836533 836544 836509 1019429 836523 836542 1048279 1093107 836541 1111473 1069762 1056621 1016013 1008066 1045694 1048222 1048238 1056612 1049543 1049546 1026207 1069768 1059739 1019398 1028611 1048255 1028654 1054237 1028614 1048310 1034386 1100249 1056591 1035528 1034390 1051743 1044584 1070083 1056586 1093078 1044593 1059718 830986 1048263 1048245 1044559 1054240 804389 1082672 1058223 1039398 1093072 1057168 1028616 1093041 1062745 1075842 1053325 1096219 1044571 1056634 1056606 1069764 1034406 1049550 1044551 1045688 1025448 1093051 1093047 1056632 1019684 1053003 1045702 1069765 1019401 1054218 1057170 1093109 1054228 1080953 1016069 1057172 842641 1016872 1063745 1054234 1019258 1056623 1080938 1062782 1094988 1095941 1075863 1015439 807872 807694 808219 807687 808206 807707 808197 808212 808358 807697 808195 808214 807709 808380 807873 807885 807704 807886 808205 802809 807887 807897 808192 808385 808215 808209 807888 808369 807702 807884 808359 808199 807685 807700 808211 807896 807688 807876 1074150 808213 808377 807898 804387 807689 807875 803298 808210 808366 808368 808207 807684 807900 808387 807708 808217 808379 808203 804378 808194 1068803 808208 807895 807690 808190 807686 807894 807892 808191 807703 808362 808193 802903 804390 807710 1081368 807880 808201 808204 807874 808378 802919 808376 807691 1044599 807891 807695 808200 807878 808374 808382 807683 807881 1058229 808361 1030605 808370 808364 807698 807682 808218 807681 807889 808372 804392 1024407 807871 807890 808375 808355 803158 808386 808383 808373 804383 808356 807899 807883 808198 799035 808389 1081362 1016876 1111472 1117638 1109839 1103273 1109836 1113830 1111488 1113827 1111483 1117623 1111493 1113848 1015454 1030585 1075725 1058232 1062775 1059717 1063708 1080947 1080920 1068786 1085223 1079932 1068787 1081377 1079937 1075876 1080949 1074135 1080930 1075860 1077054 1077070 1075877 1078595 1086612 1063674 1054216 1049551 1068805 1078649 1075833 1082675 1086450 1069772 1060374 1080928 1063672 1075878 1086438 1062776 1058389 1059719 1074149 1087488 1080951 1063750 1074156 1075875 1059738 1075716 1075873 1074152 1077058 1075854 1062778 1068792 1075861 1086616 1068798 1081363 1074133 1086439 1060385 1086474 1075870 1095943 1079925 1025450 1077068 1095958 1048253 1096212 1079921 1025427 1059721 1013721 1077066 1014730 1086432 1019720 1093116 1078606 1034379 817783 1062761 1016011 1025422 1025444 1059746 1086619 1009082 1081366 1068780 821503 1056615 1025445 1080958 1062793 1025449 1025435 1025440 1025433 1025434 1056635 793205 793231 765678 793355 793346 793372 789254 793211 793234 791440 793364 791448 791456 791444 790714 790719 791459 793224 793350 790729 793225 791182 793213 788928 793362 793220 794154 794133 794158 789242 790720 791187 789250 830756 791457 793217 793229 791451 791162 793345 794144 793215 793356 793232 793222 791453 791192 793374 791446 790703 793216 793233 794136 793368 791180 794130 793227 793207 793367 793206 793221 791454 790730 793228 791183 791461 794124 793219 793370 793359 793353 790718 793208 791450 793230 791460 791462 791166 790700 793358 793218 793360 793351 791173 789245 793361 793347 789265 791445 793366 793348 791452 790702 793209 793373 794148 793371 793354 794138 791463 791458 1019254 791163 794134 1024392 1030610 1030606 1030599 807699 1014713 1008052 842632 1086440 1030575 1008030 791184 1019699 1086611 1028615 1086633 821514 1093118 842622 840895 840894 840830 840889 840910 1019403 842635 842626 836513 840831 842608 1030576 842614 840922 840923 840916 840849 840929 1013726 786823 840903 840858 1030573 840850 840833 840865 840843 840934 840869 840867 840878 840933 840936 842637 842627 840838 1039391 840845 840857 842630 840861 840917 840900 840853 840873 842631 1013724 840911 842633 842625 840879 840906 840875 840912 1016012 840852 840844 842612 842642 842636 840914 840883 1016020 840931 840915 842646 842629 840840 831034 1059735 842621 840925 842628 1039394 840856 840860 842624 840836 840892 840899 842607 840870 840920 840868 840909 840859 1013690 840874 1013687 842618 840887 840904 840841 842613 840935 840848 842619 840897 842617 840863 1026205 840855 840926 840872 1034416 840932 840846 1030609 842611 840928 1005308 816193 790943 794074 790572 790933 808918 841044 785242 794073 790938 1023710 1005309 794079 804003 804006 1023705 1023709 1023706 816199 816204 1017432 804011 804001 804004 804009 804012 794075 790576 804010 790590 790594 1005310 841073 790914 794072 790585 717812 749597 749578 749613 749654 741151 749596 749563 743168 741204 750384 750395 750191 750160 741142 750166 750171 749598 749559 749643 750179 749610 749558 750194 750320 749552 750218 750275 750153 749555 749603 750173 750201 741197 750159 743211 750132 750340 750386 749580 750155 750307 750352 750237 749614 743205 750198 749632 749617 750419 741153 741190 749639 749567 749653 750315 750119 749658 749607 749630 749547 750156 749659 749615 749544 750298 750301 750184 749557 749577 749619 750316 749624 749571 750351 743230 749554 749635 750181 750207 750375 743206 749581 750214 750129 750196 750161 749548 749595 750120 749633 750123 749588 749600 750163 749656 750139 749660 750151 750118 750288 749545 743229 750180 750309 749599 749605 749622 743203 750148 750195 749627 750304 750236 750157 743172 743183 750141 749594 750169 749585 750164 749592 749648 741115 750222 749609 750140 749625 749647 750317 749576 749586 743227 750206 749574 749626 750186 749560 750134 749631 749637 749612 749642 750414 750277 750341 749651 750291 743226 750219 750182 750365 750131 750204 749566 750121 750209 750296 749553 750200 750225 743214 749562 749572 749556 743228 749575 750212 749573 750224 749657 749618 749645 741170 749593 750145 743232 749655 750323 749551 750208 749608 749549 749550 750226 750305 749602 741180 750221 750170 750274 749616 750382 749634 750154 750227 749628 750122 749620 749661 750190 750165 750310 749565 716118 716114 716113 716110 716111 719109 727582 721336 721351 721366 721396 721332 721343 721379 721309 719101 724468 721403 724397 717182 721375 719117 721330 717205 724442 719119 719143 724386 724498 719098 724418 718553 724389 721406 721393 718615 716615 719083 721361 721348 721388 721335 721386 721391 721367 719160 721394 721347 716434 724438 718609 724432 719131 719120 719092 719155 724401 718585 719102 724474 717763 718554 717781 721340 721352 717815 721389 724382 718595 718569 721395 719078 721381 719150 719079 719159 721315 719104 721342 719152 716582 721390 721328 721365 719100 721320 719161 717135 719162 721398 718614 724430 718532 721329 721325 724454 721382 724463 721314 719090 717773 718621 717784 721380 719088 724447 724443 724471 717745 719145 719114 719171 721383 721370 724485 724475 719097 717757 718530 719089 721354 724458 724426 717744 718586 721350 721356 724416 724415 716443 719167 724452 724404 724448 717736 724451 724500 718603 719096 717817 719156 719148 717786 719111 721392 724390 719125 721405 717728 721371 721339 717734 719137 721326 721378 719123 718581 721345 719141 719077 721316 721397 719116 719073 719076 721344 721402 717787 719133 724428 721333 718584 721377 719122 717729 717780 717148 721349 719166 717779 717730 717162 721387 719142 721400 719126 719115 717175 719108 724461 719118 719134 717122 721331 721313 721357 719165 717134 718617 719149 719124 721384 721362 721404 721358 721372 721355 724425 719163 719169 721327 721337 724410 721363 717766 718583 721310 719151 719084 719130 721319 716432 718547 718545 717788 765665 765716 765759 763616 757449 763692 763672 765706 763652 763657 763662 757392 765703 763675 763639 763680 765696 763636 765722 763599 765736 765754 763649 765702 763684 765776 763683 765719 763582 763623 763654 763619 763585 763664 763626 765767 765714 763663 763690 763666 765738 757457 763612 757412 763605 763628 763655 763669 757445 763588 750398 750415 763641 763650 765713 765708 765662 757389 757433 757391 763689 757420 765727 763681 763635 765707 765763 763589 763592 763624 763586 750412 763598 763602 763617 765698 763584 763597 763697 763590 765785 765758 765710 765773 763608 763633 763618 765718 765681 763638 763587 765780 763695 765743 765762 765704 763702 757393 757471 763613 763611 765679 763699 765691 763659 763682 763686 763693 765664 763622 763696 763595 763661 763632 763706 763621 763625 763676 765734 763704 763671 763694 765712 763583 765670 765683 763606 763609 763698 765729 765764 763603 765781 763668 763674 763620 765751 749561 763615 763600 763596 763691 750185 765786 763667 765700 765782 763642 765748 765735 750397 765709 763644 763629 765732 763594 765757 763673 763703 765740 765752 765692 763601 763614 763660 763593 765750 765667 763631 765761 765784 765721 763700 765755 765746 765742 763685 763640 765725 763647 765771 765673 765715 763678 763643 763604 765769 763679 763677 763670 765745 763591 763648 750187 765724 724453 727670 727610 721317 721353 721324 730697 721307 724413 727634 717813 730663 730643 721308 719153 727593 724470 718576 730655 719146 730690 727628 727614 724450 716442 718571 717185 727631 727580 730624 727666 724402 716648 717164 724490 721323 727625 727609 727606 730647 724473 721364 724419 719129 724497 724392 727576 724478 724494 717186 727645 724437 730683 727640 730679 727605 718526 727662 718527 730693 727638 727590 727646 724424 724469 717756 727648 727573 727624 727649 719139 727575 727654 724445 716444 717777 730639 717176 730701 724444 724499 724395 727585 724483 727615 717764 727635 727621 727597 724467 717805 727663 727611 724477 724493 724489 717179 727623 727653 730661 727627 719144 727579 724459 724388 727613 724385 717809 724417 727594 724431 727644 724393 724403 727616 727642 724446 727672 727630 727641 727636 727660 724407 727596 724486 724496 730712 730696 724439 721373 724429 730641 716608 718568 730646 721376 716628 727584 730650 730651 727587 724399 721312 727602 719135 730638 717149 727643 716479 730626 727639 724408 724406 727650 730657 730644 719128 727592 727626 721385 730613 727581 727657 730678 730669 717177 730688 724435 724409 730640 727619 727604 718533 730616 727669 717775 730617 730694 721360 724457 730648 721369 730653 724460 730698 730627 730687 724400 724462 730666 718616 730705 721338 727608 730618 719087 716646 716468 730631 724383 727633 727658 727655 727661 727656 730695 719157 727578 717156 724420 724501 716441 724465 730702 719081 730691 727647 730659 730667 730676 727622 724476 719105 724411 724434 724482 724412 724427 724466 716433 724479 719074 718608 727617 730632 724423 718620 724491 717155 724436 724414 724480 718604 727659 727651 717184 724396 727586 727612 724464 724488 721399 716631 730668 730664 730681 730625 730633 727664 727671 727632 721334 721346 727618 730689 730680 727601 717807 730652 730636 730628 730635 724472 717789 721322 724455 724398 724495 718590 724384 717204 724441 724433 724456 727574 727620 727629 727603 724387 716466 721341 741167 741171 743213 741195 741121 743167 743215 741203 741192 741189 743191 741122 741200 741188 741139 727577 741196 741175 735327 735322 743169 741172 741135 743207 741118 741132 741145 743189 743181 743197 743164 743195 743184 741147 741148 743225 743204 741131 743198 741134 743201 741208 741183 741141 730658 735279 741149 735242 741178 741169 741157 741144 741155 741184 741154 741160 743174 741123 735280 741194 743192 735302 743212 741193 741127 743178 741165 741138 741199 741205 741133 741206 741119 741182 741120 741176 743190 741156 741191 741116 743217 741201 743209 741158 741174 741185 741150 741143 741117 741159 741181 735234 743163 743165 741136 743180 743220 741161 743188 743194 741162 741198 743199 741128 743182 741129 741163 733938 735295 733958 733962 724391 724484 735324 735249 733899 743187 735260 735273 735320 735257 730660 743179 735284 735252 735299 733916 724405 743193 741210 741179 735235 735289 735301 735287 735306 735248 735303 735300 733927 733922 733897 730710 735329 741137 743196 741187 735297 735274 741164 735269 735332 743208 735310 743170 743202 735286 730700 735298 735304 743224 735276 733970 735308 743171 735272 735266 716455 743210 735246 735317 735321 735275 727588 735278 733928 735315 741152 716438 719072 717814 718537 716661 717147 719086 717792 717803 716449 719107 717169 716112 716618 716636 717760 717210 717171 718557 717211 717753 717749 718573 717180 717738 718580 718572 717725 716606 717743 718540 717761 716481 717770 718597 717794 716610 718552 716613 719170 718601 716671 718562 716639 717197 716658 718522 718605 718563 717150 716605 716454 716659 716591 717208 716656 717799 717140 716457 718600 718598 716474 717741 718538 718531 717740 717778 717161 718534 717207 718613 716598 717722 717818 719075 717819 717194 717157 717765 716626 716609 718582 717178 717132 716452 717165 718594 716451 717146 718599 717747 718550 718596 716592 716657 716589 716604 717769 717723 717732 716584 717759 716115 716117 717153 717168 717133 716477 717203 716601 716603 718539 716651 716473 717126 716436 719164 716467 716586 717783 716445 717200 718544 718560 717762 718541 718612 717172 717771 716463 717220 717748 717159 718610 717139 716641 716612 716437 717181 716465 717199 716665 716594 717797 717142 716642 717735 718559 717726 717138 716472 717796 718543 716664 716655 716459 716652 717219 716597 716450 716471 717221 716476 716616 716590 717202 716670 716634 718548 718551 717733 717215 716458 717727 717209 717127 717201 717193 716625 716600 717772 717187 716470 716448 717143 716435 718523 716650 716649 719091 717821 718611 716643 717742 717768 717810 718555 717774 716644 719094 716645 717785 716617 717758 716621 716462 716654 716480 716447 718528 717123 717806 716611 716461 717754 718570 717795 718589 717746 718591 718578 717189 717183 717160 717152 717791 717173 718618 717125 717145 716440 716667 716614 716627 717213 716640 718558 718567 717798 717776 718535 717124 717800 717144 717214 716666 716653 718574 719127 717195 716593 719085 717739 718602 717192 717816 716662 717804 717196 716638 717151 716632 716588 716668 718587 716637 716619 719093 716453 716602 717206 716660 717166 716583 716635 718565 718524 716464 717131 717198 717137 717782 718575 716460 717191 716478 717793 717154 716596 716663 717129 718579 716669 716439 716622 717724 718525 717750 716599 718593 716456 716623 716475 717216 718619 717174 716629 717163 718564 719112 716119 717802 716624 718556 717217 717811 718536 717128 717158 716607 717737 716630 717751 718542 717167 718607 718561 719136 716595 719132 719082 717767 717218 717136 781420 778174 733887 772200 774256 774261 774372 774369 774325 772139 774332 774367 774329 772172 774254 775144 775141 774265 774296 772131 774384 774356 775117 774359 775135 775151 774301 774386 774402 774341 774351 774346 772151 774282 775149 774262 774280 775109 774390 774306 774264 775145 774278 774269 774340 774307 774396 774258 774279 775143 765756 774370 774304 774276 774324 775162 774326 774286 774364 774317 774392 775108 775102 775128 774293 774363 775096 775158 775110 775111 774350 775113 774270 775159 774374 772205 774388 774285 774313 775106 774347 775140 774355 775138 774272 774383 775125 775123 774393 775157 774268 774343 775122 774310 774380 772188 774263 774297 772146 765778 774338 775114 774260 775118 775124 775093 775095 775131 774377 775160 775121 775132 774357 774358 774284 775112 774302 774400 775155 775100 774275 765747 775116 774311 775153 774316 775136 772206 774274 775097 774379 774365 774342 774321 774333 774344 774322 774366 774259 774371 774381 774294 774399 775154 775103 772199 774328 774271 774288 774403 774257 772179 774375 774320 774395 774373 774354 774330 774335 774352 774385 774336 774334 774362 775161 774290 774308 774255 774266 774323 774398 775098 775115 775120 774339 774345 774368 774353 774281 774376 774291 774391 775107 775127 774303 775148 774277 774289 774349 774283 772173 775142 775134 774394 774401 774389 772195 774314 774360 774267 775099 775137 733878 716446 735312 735311 730623 735328 735267 730711 735271 733961 733904 730673 735293 730704 733944 735323 717130 730614 735250 735233 733908 733907 735247 735256 730672 730621 730707 733915 733888 733930 733909 735270 735262 733959 730671 733934 733895 735292 735290 730634 733879 735236 730622 735255 735313 735319 733971 735263 735264 733975 733952 733891 719147 733977 733894 733947 733939 735265 735285 735239 735243 735288 733914 733925 733932 735291 733890 733883 733900 733976 735331 733933 716620 735283 733941 727591 733931 733919 735307 735316 719138 718577 733910 717808 735258 735309 730654 733885 724481 735296 727607 730699 730642 724422 733893 733906 733884 733905 733913 733918 730670 733912 733974 735305 727665 733950 733946 735241 733937 735277 730685 733957 716633 717755 730692 733945 733923 733901 733949 733960 733954 733963 721359 730630 733892 733968 735259 730674 735261 733898 730620 724487 735325 730677 733967 730645 733889 733951 733903 733942 735268 730684 735254 730686 733943 730709 730706 733969 733880 724421 733886 733929 730656 717170 735330 733972 733936 730703 733940 733973 733911 733955 733917 730665 733965 733902 735253 733964 730675 733948 730662 730708 730637 716587 735318 735282 733956 727652 735281 735240 733920 733924 735244 733921 733881 750130 750125 750150 750168 750329 750387 750142 749636 750366 750372 750289 750235 750371 750146 750278 750286 749568 749584 750373 750314 750354 750213 750126 750325 750193 750167 750331 750390 750284 750338 750322 749587 750178 750285 749611 749621 750177 750345 750368 750360 750396 750197 750355 750333 750335 750205 750216 750279 750290 750302 750202 750233 750401 750321 750343 750359 750287 750273 750370 750402 750336 749590 750229 750406 750389 750319 750292 750174 750175 750149 750346 750294 750378 750411 749641 750405 750379 750223 749623 750172 750383 750392 750217 750136 750215 750364 750326 750337 750388 749583 750231 750385 750376 750152 750276 750367 750281 750143 750418 750409 750228 750311 750230 749638 750342 749570 765737 765728 772193 772202 772211 765699 772196 765688 765777 772223 772208 772191 765677 772228 765731 765730 765685 772159 772220 772186 772164 772170 765686 765749 772158 765753 772183 772214 765726 765684 765690 765666 772181 765765 765669 772134 765672 772198 765689 765693 772184 765701 772204 763637 765697 765680 765766 765671 765733 765705 765739 772147 763610 765695 772145 772130 772148 765674 772209 763658 772161 772129 772140 772201 772174 763687 772157 765720 765760 772167 763607 772137 772165 772192 772221 765783 772207 772166 772212 772225 772132 772144 765687 765694 765768 765779 772142 772169 772163 772175 763653 772203 772187 772224 765723 765675 772162 765741 772197 772213 772136 772190 772215 772171 772168 765663 772143 772185 772160 772227 772149 772210 772226 772217 772177 772222 772153 772180 772194 772156 772133 772218 772219 772182 772154 772150 772141 724394 775104 778519 719103 781578 781426 750313 730629 781422 730682 781564 781416 730649 781565 778320 778169 750332 733882 727637 717188 750327 757461 750318 757469 757436 757387 757418 757416 750361 750138 750220 750328 749601 750295 749569 757402 750297 750211 750283 750404 750393 750380 749564 749606 750282 750176 749646 757450 750137 749649 750188 750306 750293 749591 750362 750303 757401 749652 750334 750357 757481 757415 757454 757472 757479 757437 757422 757465 749662 750374 750124 757425 757447 750299 757435 757414 757456 757442 750369 750158 757394 750330 750399 757426 757409 757413 757395 750381 757406 757434 757478 757466 757432 757419 757474 750417 757468 757388 757444 757463 750403 757427 757429 757407 757404 749644 749589 757473 757405 757403 757458 757475 750353 757438 750324 749629 750339 750348 757428 757467 750192 757400 757385 757448 757417 749604 757451 757398 757408 757477 757446 757399 757480 750128 749582 750312 757431 757462 757440 757421 757443 757430 757455 750363 750135 750407 749579 750300 750358 750147 750189 750127 750133 778310 778725 778312 778313 781581 778528 778737 781412 778188 781417 778193 778518 774348 774312 781582 774287 775150 778733 778304 765774 780297 780011 778529 778533 778171 775094 774309 778727 774292 778747 778306 781423 781583 774382 780315 779994 780306 780318 781567 778305 774319 778728 778317 775101 778741 778523 778183 781408 778512 778302 778724 778731 778735 778176 778300 778323 778531 778532 778736 778181 778186 781579 778748 778530 775133 774295 780305 778521 781576 778309 778179 775156 778743 775126 778192 774327 778178 775139 778177 778525 772155 774378 774315 781554 781429 778514 778308 778516 775152 778734 780004 780022 778509 778187 781430 781413 778190 778513 778314 778175 778303 778180 778740 778307 778182 778527 774305 780037 778185 780024 778315 774300 778316 778515 780272 780292 780300 781410 778517 780042 780277 778738 780010 781577 781405 781404 781411 781431 780293 780028 781433 780017 780290 780280 780034 779998 780271 772178 780014 781432 780279 781407 781563 780308 780309 780035 780296 778726 779997 780286 780013 781559 780278 780274 717731 779999 780016 780036 780026 781424 780281 781574 780009 780041 780002 780270 778170 781572 780304 778172 780030 781571 717790 780021 780312 778311 780298 780023 778739 780289 772138 778322 780029 778173 778730 780012 780025 781560 781421 780294 780317 781580 781573 781575 780285 780275 778321 780302 780301 778522 779996 781427 780295 781406 780311 781569 781415 780313 778189 780000 780027 763665 780269 780015 775119 774299 780288 780019 780001 781555 778742 778524 780299 780018 780284 780303 778732 778520 779995 780038 778526 778319 780316 774273 780291 778744 778511 780032 780003 778745 780006 778746 778184 780007 778729 780043 780314 778318 741130 750111 750104 741672 750115 741671 741710 741677 741705 750113 741675 741681 750112 741686 741682 741709 750099 741668 741714 750089 750096 741719 741708 741718 741693 741673 741685 716701 721843 721796 718744 721759 721780 721778 721806 721757 721835 718731 718749 721771 716963 718795 721827 721831 718734 718813 721812 721788 721855 721822 721839 721840 721795 718776 721825 721826 718771 718722 716687 718785 721849 721760 721808 721773 721844 721769 721828 716930 721851 721790 721804 721772 721793 721852 721762 716680 718765 721841 718799 718812 716928 721837 721766 721802 716966 718815 716956 721776 718735 721820 721765 721803 721775 721799 721821 718778 721794 721829 721785 721810 721815 721781 721830 721809 718782 716720 721798 721853 718747 721856 718780 763543 756868 757731 753108 756919 763535 756923 756906 756878 763522 756926 763528 756890 756912 757728 763540 757738 756853 756876 756870 763575 756918 756864 756852 756939 757729 756938 756893 757746 756867 763538 763541 763527 763568 757750 756917 756883 763581 756888 763561 763553 757743 757742 763552 763573 763556 756930 763549 763517 763521 763567 756933 756881 756932 757724 757751 756899 750088 756874 756937 757722 763550 756879 756925 757739 756891 757725 757723 756928 763539 756927 756902 756905 757744 757733 753103 757730 756916 763570 721801 721850 718788 727979 727967 716960 721845 721807 716688 721854 721777 721787 716940 721774 727977 721847 727966 721800 727981 716942 718748 727971 721842 721805 727975 721832 721758 721838 716694 727978 716968 721764 727982 727976 727964 716949 721814 721817 721818 718811 718750 721819 718753 718772 721834 727972 727974 721767 716713 716948 716955 727980 727973 721782 727968 716952 727969 718743 721763 718819 721848 727963 734142 741713 741669 734144 741715 741680 741678 734138 741702 734157 734155 741723 741712 741691 741703 741700 741716 734130 741688 734141 741676 741724 741722 734101 741699 734133 741692 741694 741695 734153 734111 741697 734152 734105 721791 741674 734126 716690 716967 718757 716685 716934 716950 718807 716938 716927 716702 716709 716922 716708 718793 718730 718818 716681 716924 718732 718791 716710 718745 716712 718773 718786 718764 718761 716944 716678 718809 716683 718820 718766 718770 716965 718740 716715 718751 718806 716943 716932 716719 718741 716700 716962 718779 718789 716707 716929 716682 716672 716706 718802 716964 718763 716939 718781 718726 716697 718816 718768 716691 718742 716954 718754 716933 716959 716674 716947 718798 718746 718727 718783 716971 716946 716705 718769 716961 716935 716970 718737 718821 718817 716925 718729 716936 716718 716693 718775 716969 716699 718797 716937 716721 718794 718804 718805 718760 716926 716675 716958 716931 718736 718777 718755 718803 718752 716676 718723 716673 718733 718796 716703 718808 716957 718762 716711 718784 716945 716698 716695 718759 716941 718756 716689 716717 718728 716677 716953 716696 718800 718725 718724 770705 770644 770663 770740 770708 770669 770652 770710 770702 770716 770673 770714 770715 770682 770668 770744 770712 770745 779865 716951 734139 734150 734099 734143 734136 721824 718801 734148 734145 718790 727970 716704 716716 716923 734108 734103 734107 734140 734151 734120 716714 734098 716686 734146 734131 734124 734100 734149 721770 734109 734113 734125 734116 734132 734102 734118 734104 734117 721797 734156 734127 734110 734134 734106 734123 727965 718810 734135 734119 734129 734121 718758 734137 718787 734114 756901 753107 756871 756920 750098 750100 753113 750105 753110 753105 753111 750101 750094 750097 756857 756894 750103 750107 750102 750095 750093 753112 756941 750091 756882 753106 756877 756855 756866 753116 753115 756860 750117 756880 750110 753109 750106 750108 750109 756872 770703 763531 770723 763532 763547 763569 763516 770683 770642 770677 770693 770687 770732 763555 770688 763571 770724 770722 763524 770707 763537 763563 763580 763576 763520 763557 763551 770694 770680 763515 763562 770700 770701 770665 770646 763514 756904 763529 763526 763565 770639 763578 763574 770658 770649 770737 770648 763530 770709 763512 770695 770653 770647 770746 770671 770736 763577 770730 770674 770676 770748 763533 763546 770667 770733 770672 770729 770706 770643 763519 763518 770728 763572 763536 770742 763566 763534 770656 763545 763525 770731 770738 770661 763523 770685 770743 763554 763564 763513 763548 770713 763560 763559 770681 770699 770726 763558 763544 779888 741689 741701 770741 753114 756873 756856 756858 756875 756913 756903 750114 756934 753104 756921 757726 757747 756940 757741 756859 756929 756908 756931 756896 756910 756861 756924 756884 756915 757735 756886 756922 756907 756887 756869 757732 756854 756885 756900 757749 756897 757737 756895 757727 757734 756911 757745 756914 756936 756862 756898 756892 757748 757736 770657 780109 779876 780139 779883 779887 770696 770678 779885 779882 780130 770704 770641 770670 770698 779869 770691 770645 770662 780123 770679 770735 770717 779892 780108 770721 770650 770697 779862 780118 770675 780116 770659 770727 770660 779895 780120 780127 770654 780141 779886 780140 770666 780136 770719 780135 770734 770692 780107 770711 770651 770690 770725 770739 780122 780132 780110 770718 770689 770720 770747 779896 780111 568512 571874 568424 571960 571930 571939 571913 571969 571973 571900 571895 568518 568453 571972 568426 571866 567720 567721 568514 567718 571967 568441 571892 567730 568513 571971 571962 585377
Recalling Firm/
Manufacturer
St Jude Medical Inc.
15900 Valley View Ct
Sylmar CA 91342-3577
For Additional Information ContactKelly Morrison
818-493-3591
Manufacturer Reason
for Recall
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
FDA Determined
Cause 2
Software design
ActionOn 04/16/2018, Important Medical Device Advisory notifications were mailed via overnight service to U.S. physicians and hospitals notifying customers of the availability of the programmer software update and associated ICD/CRT-D firmware update. Advisory notifications were mailed to customers outside the U.S. on different dates depending on geography and local regulatory requirements. Prophylactic replacement of affected devices is not recommended. The firm recommends firmware upgrades for all eligible patients at the next regularly scheduled visit or when appropriate depending on the preferences of the patient and physician. For devices not eligible for the cybersecurity firmware update the firm recommends a discussion of the risks of cybersecurity vulnerabilities and proven benefits of remote monitoring with patients at their next regularly scheduled visit. RF communication may be permanently disabled in devices not eligible for firmware updates during an in-clinic device interrogation with a programmer that has received the software update. Customers with questions about the firmware update can call the customer technical support hotline at 1-800-436-5056.
Quantity in Commerce10425
DistributionU.S. Nationwide distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = LWS
-
-