| | Class 2 Device Recall Aisys Anesthesia System |  |
| Date Initiated by Firm | September 27, 2018 |
| Date Posted | November 09, 2018 |
| Recall Status1 |
Terminated 3 on April 12, 2021 |
| Recall Number | Z-0411-2019 |
| Recall Event ID |
81182 |
| 510(K)Number | K110213 K170872 |
| Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
| Product | Aisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE. |
| Code Information |
Model Number, UDI Number, (Serial Number): 2097674-001-US, Not Applicable, (5470345-30); 2097674-001-US, Not Applicable, (5470345-31); 2097674-001-US, Not Applicable, (5470345-32); 1011-9000-000, Not Applicable, (ANAK00774); 1011-9000-000, Not Applicable, (ANAK00775); 1011-9000-000, Not Applicable, (ANAL01201); 1011-9000-000, Not Applicable, (ANAL01234); 1011-9000-000, Not Applicable, (ANAL01235); 1011-9000-000, Not Applicable, (ANAL01236); 1011-9000-000, Not Applicable, (ANAL01237); 1011-9000-000, Not Applicable, (ANAL01238); 1011-9000-000, Not Applicable, (ANAL01240); 1011-9000-000, Not Applicable, (ANAL01494); 1011-9000-000, Not Applicable, (ANAL01495); 1011-9000-000, Not Applicable, (ANAL01551); 1011-9000-000, Not Applicable, (ANAM01023); 1011-9000-000, Not Applicable, (ANAM01024); 1011-9000-000, Not Applicable, (ANAM01025); 1011-9000-000, Not Applicable, (ANAM01026); 1011-9000-000, Not Applicable, (ANAM01027); 1011-9000-000, Not Applicable, (ANAM01104); 1011-9000-000, Not Applicable, (ANAM01105); 1011-9000-000, Not Applicable, (ANAP00438); 1011-9000-000, Not Applicable, (ANAP00439); 1011-9000-000, Not Applicable, (ANAQ01188); 1011-9000-000, Not Applicable, (ANAR05264); 1011-9000-000, Not Applicable, (ANAR05265); 1011-9000-000, Not Applicable, (ANAR05610); 1011-9000-000, Not Applicable, (ANAR05611); 1011-9000-000, Not Applicable, (ANAS00202); 1011-9000-000, Not Applicable, (ANAS00203); 1011-9000-000, Not Applicable, (ANAS00204); 2097674-001-US, Not Applicable, (5479630-1); ANAK00798, Not Applicable (1011-9000-000); ANAK00799, Not Applicable (1011-9000-000); ANAK00803, Not Applicable (1011-9000-000); ANAK00807, Not Applicable (1011-9000-000); ANAK00814, Not Applicable (1011-9000-000); ANAK00839, Not Applicable (1011-9000-000); ANAK00864, Not Applicable (1011-9000-000); 5481223-1, Not Applicable (2097674-001-US); 5481223-10, Not Applicable (2097674-001-US); 5481223-11, Not Applicable (2097674-001-US); 5481223-12, Not Applicable (2097674-001-US); 5481223-2, Not Applicable (2097674-001-US); 5481223-3, Not Applicable (2097674-001-US); 5481223-4, Not Applicable (2097674-001-US); 5481223-5, Not Applicable (2097674-001-US); 5481223-6, Not Applicable (2097674-001-US); 5481223-7, Not Applicable (2097674-001-US); 5481223-8, Not Applicable (2097674-001-US); 5481223-9, Not Applicable (2097674-001-US); ANAP00712, Not Applicable (1011-9000-000); ANAP00713, Not Applicable (1011-9000-000); ANAP00714, Not Applicable (1011-9000-000); ANAN01125, Not Applicable (1011-9000-000); ANAN01138, Not Applicable (1011-9000-000); 4652664-1, Not Applicable (2097674-001-US); 4652664-10, Not Applicable (2097674-001-US); 4652664-11, Not Applicable (2097674-001-US); 4652664-12, Not Applicable (2097674-001-US); 4652664-13, Not Applicable (2097674-001-US); 4652664-2, Not Applicable (2097674-001-US); 4652664-3, Not Applicable (2097674-001-US); 4652664-4, Not Applicable (2097674-001-US); 4652664-5, Not Applicable (2097674-001-US); 4652664-6, Not Applicable (2097674-001-US); 4652664-7, Not Applicable (2097674-001-US); 4652664-8, Not Applicable (2097674-001-US); 4652664-9, Not Applicable (2097674-001-US); ANAK00156, Not Applicable (1011-9000-000); ANAK00157, Not Applicable (1011-9000-000); ANAK00158, Not Applicable (1011-9000-000); ANAK00159, Not Applicable (1011-9000-000); ANAK00511, Not Applicable (1011-9000-000); ANAK00512, Not Applicable (1011-9000-000); ANAK00513, Not Applicable (1011-9000-000); ANAK00514, Not Applicable (1011-9000-000); ANAN01051, Not Applicable (1011-9000-000); ANAN01052, Not Applicable (1011-9000-000); ANAP00545, Not Applicable (1011-9000-000); ANAP00546, Not Applicable (1011-9000-000); ANAP00708, Not Applicable (1011-9000-000); ANAL01400, Not Applicable (1011-9000-000); ANAL01401, Not Applicable (1011-9000-000); ANAM00449, Not Applicable (1011-9000-000); ANAM01089, Not Applicable (1011-9000-000); 5453450-1, Not Applicable (2097674-001); 5453450-2, Not Applicable (2097674-001); 5453450-3, Not Applicable (2097674-001); 5453450-4, Not Applicable (2097674-001); 5453450-5, Not Applicable (2097674-001); 5453450-6, Not Applicable (2097674-001); 5453450-7, Not Applicable (2097674-001); 5453450-8, Not Applicable (2097674-001); 5480532-1, Not Applicable (2097674-001); 5447751-1, Not Applicable (2097674-001); 5447751-10, Not Applicable (2097674-001); 5447751-11, Not Applicable (2097674-001); 5447751-12, Not Applicable (2097674-001); 5447751-13, Not Applicable (2097674-001); 5447751-14, Not Applicable (2097674-001); 5447751-15, Not Applicable (2097674-001); 5447751-16, Not Applicable (2097674-001); 5447751-17, Not Applicable (2097674-001); 5447751-18, Not Applicable (2097674-001); 5447751-19, Not Applicable (2097674-001); 5447751-2, Not Applicable (2097674-001); 5447751-20, Not Applicable (2097674-001); 5447751-21, Not Applicable (2097674-001); 5447751-22, Not Applicable (2097674-001); 5447751-23, Not Applicable (2097674-001); 5447751-24, Not Applicable (2097674-001); 5447751-3, Not Applicable (2097674-001); 5447751-4, Not Applicable (2097674-001); 5447751-5, Not Applicable (2097674-001); 5447751-6, Not Applicable (2097674-001); 5447751-7, Not Applicable (2097674-001); 5447751-8, Not Applicable (2097674-001); 5447751-9, Not Applicable (2097674-001); 5452649-1, Not Applicable (2097674-001); 5452649-2, Not Applicable (2097674-001); 5474323-1, Not Applicable (2097674-DE); ANAR05021, Not Applicable (1011-9000-000); 5473625-5, Not Applicable (2097674-NL); 5473625-6, Not Applicable (2097674-NL); 5473625-7, Not Applicable (2097674-NL); ANAK00330, Not Applicable (1011-9000-000); ANAK00332, Not Applicable (1011-9000-000); ANAK00334, Not Applicable (1011-9000-000); ANAL00806, Not Applicable (1011-9000-000); ANAL00807, Not Applicable (1011-9000-000); ANAL00808, Not Applicable (1011-9000-000); ANAL00809, Not Applicable (1011-9000-000); ANAN01010, Not Applicable (1011-9000-000); ANAN01011, Not Applicable (1011-9000-000); ANAN01012, Not Applicable (1011-9000-000); ANAN01013, Not Applicable (1011-9000-000); ANAN01014, Not Applicable (1011-9000-000); ANAP00399, Not Applicable (1011-9000-000); ANAP01133, Not Applicable (1011-9000-000); ANAR00344, Not Applicable (1011-9000-000); ANAR00896, Not Applicable (1011-9000-000); ANAR05978, Not Applicable (1011-9000-000); ANAR05979, Not Applicable (1011-9000-000); ANAR05980, Not Applicable (1011-9000-000); ANAR05981, Not Applicable (1011-9000-000); ANAR05982, Not Applicable (1011-9000-000); ANAR05983, Not Applicable (1011-9000-000); ANAR05984, Not Applicable (1011-9000-000); ANAS00709, Not Applicable (1011-9000-000); ANAS00710, Not Applicable (1011-9000-000); ANAS00711, Not Applicable (1011-9000-000); 5472401-2, Not Applicable (2097674-001); ANAP00132, Not Applicable (1011-9000-000); 5472557-4, Not Applicable (2097674-001); 5472557-5, Not Applicable (2097674-001); 5472557-6, Not Applicable (2097674-001); ANAN01045, Not Applicable (1011-9000-000); ANAN01046, Not Applicable (1011-9000-000); ANAR00327, Not Applicable (1011-9000-000); ANAS00345, Not Applicable (1011-9000-000); 5454764-3, Not Applicable (2097674-001); ANAP00584, Not Applicable (1011-9000-000); ANAP00585, Not Applicable (1011-9000-000); ANAP00586, Not Applicable (1011-9000-000); ANAP00587, Not Applicable (1011-9000-000); ANAP00588, Not Applicable (1011-9000-000); ANAP00589, Not Applicable (1011-9000-000); ANAP00825, Not Applicable (1011-9000-000); ANAP01215, Not Applicable (1011-9000-000); ANAS00738, Not Applicable (1011-9000-000); ANAL01390, Not Applicable (1011-9000-000); ANAL01391, Not Applicable (1011-9000-000); ANAS00391, Not Applicable (1011-9000-000); ANAR00492, Not Applicable (1011-9000-000); ANAR00493, Not Applicable (1011-9000-000); ANAR00494, Not Applicable (1011-9000-000); ANAR00495, Not Applicable (1011-9000-000); ANAR00496, Not Applicable (1011-9000-000); ANAR00497, Not Applicable (1011-9000-000); ANAR00498, Not Applicable (1011-9000-000); ANAR00499, Not Applicable (1011-9000-000); ANAR00500, Not Applicable (1011-9000-000); ANAR00501, Not Applicable (1011-9000-000); ANAR00756, Not Applicable (1011-9000-000); ANAR00757, Not Applicable (1011-9000-000); ANAL00705, Not Applicable (1011-9000-000); ANAL01040, Not Applicable (1011-9000-000); ANAL01041, Not Applicable (1011-9000-000); ANAL01042, Not Applicable (1011-9000-000); ANAL01043, Not Applicable (1011-9000-000); ANAL01044, Not Applicable (1011-9000-000); ANAL01045, Not Applicable (1011-9000-000); ANAL01046, Not Applicable (1011-9000-000); ANAL01047, Not Applicable (1011-9000-000); ANAL01048, Not Applicable (1011-9000-000); ANAL01050, Not Applicable (1011-9000-000); ANAM00335, Not Applicable (1011-9000-000); ANAL00192, Not Applicable (1011-9000-000); ANAL00754, Not Applicable (1011-9000-000); ANAL00755, Not Applicable (1011-9000-000); ANAL00756, Not Applicable (1011-9000-000); ANAL00757, Not Applicable (1011-9000-000); ANAL00764, Not Applicable (1011-9000-000); ANAL00769, Not Applicable (1011-9000-000); ANAL00776, Not Applicable (1011-9000-000); ANAL00777, Not Applicable (1011-9000-000); ANAL00778, Not Applicable (1011-9000-000); ANAL00779, Not Applicable (1011-9000-000); ANAL00780, Not Applicable (1011-9000-000); ANAL00781, Not Applicable (1011-9000-000); ANAL00782, Not Applicable (1011-9000-000); ANAL00783, Not Applicable (1011-9000-000); ANAL00784, Not Applicable (1011-9000-000); ANAL00785, Not Applicable (1011-9000-000); ANAL00786, Not Applicable (1011-9000-000); ANAL00787, Not Applicable (1011-9000-000); ANAL00788, Not Applicable (1011-9000-000); ANAL00789, Not Applicable (1011-9000-000); ANAL00790, Not Applicable (1011-9000-000); ANAL00791, Not Applicable (1011-9000-000); ANAL00792, Not Applicable (1011-9000-000); ANAL01541, Not Applicable (1011-9000-000); ANAL01542, Not Applicable (1011-9000-000); ANAP00810, Not Applicable (1011-9000-000); ANAP00811, Not Applicable (1011-9000-000); ANAS00938, Not Applicable (1011-9000-000); ANAP01070, Not Applicable (1011-9000-000); 5473409-1, Not Applicable (2097674-FR); 5473409-2, Not Applicable (2097674-FR); 5473409-3, Not Applicable (2097674-FR); ANAP00517, Not Applicable (1011-9000-000); ANAP00518, Not Applicable (1011-9000-000); ANAP00519, Not Applicable (1011-9000-000); ANAM01291, Not Applicable (1011-9000-000); ANAM01292, Not Applicable (1011-9000-000); ANAM01293, Not Applicable (1011-9000-000); ANAN00430, Not Applicable (1011-9000-000); ANAN00431, Not Applicable (1011-9000-000); ANAN00432, Not Applicable (1011-9000-000); ANAN00436, Not Applicable (1011-9000-000); ANAN00437, Not Applicable (1011-9000-000); ANAP00300, Not Applicable (1011-9000-000); ANAQ00766, Not Applicable (1011-9000-000); ANAS00753, Not Applicable (1011-9000-000); ANAP00492, Not Applicable (1011-9000-000); ANAP00493, Not Applicable (1011-9000-000); ANAM00628, Not Applicable (1011-9000-000); ANAL01029, Not Applicable (1011-9000-000); 5479515-1, Not Applicable (2097674-JA); 5479515-2, Not Applicable (2097674-JA); 5479515-3, Not Applicable (2097674-JA); 5479515-4, Not Applicable (2097674-JA); 5479515-5, Not Applicable (2097674-JA); 5479515-6, Not Applicable (2097674-JA); 5479515-7, Not Applicable (2097674-JA); 5479515-8, Not Applicable (2097674-JA); 5479515-9, Not Applicable (2097674-JA); 5480673-1, Not Applicable (2097674-JA); 5480673-2, Not Applicable (2097674-JA); 5480673-3, Not Applicable (2097674-JA); 5480673-4, Not Applicable (2097674-JA); 5480673-5, Not Applicable (2097674-JA); 5480673-6, Not Applicable (2097674-JA); 5480673-7, Not Applicable (2097674-JA); 5480673-8, Not Applicable (2097674-JA); 5483069-1, Not Applicable (2097674-JA); ANAQ01435, Not Applicable (1011-9000-000); ANAM01312, Not Applicable (1011-9000-000); ANAM01313, Not Applicable (1011-9000-000); ANAM01315, Not Applicable (1011-9000-000); ANAM01317, Not Applicable (1011-9000-000); ANAM01318, Not Applicable (1011-9000-000); ANAM01320, Not Applicable (1011-9000-000); ANAM01324, Not Applicable (1011-9000-000); ANAM01325, Not Applicable (1011-9000-000); ANAL01313, Not Applicable (1011-9000-000); ANAL01314, Not Applicable (1011-9000-000); ANAL01315, Not Applicable (1011-9000-000); ANAL01316, Not Applicable (1011-9000-000); ANAL01317, Not Applicable (1011-9000-000); ANAL01318, Not Applicable (1011-9000-000); ANAL01319, Not Applicable (1011-9000-000); ANAL01320, Not Applicable (1011-9000-000); ANAM01302, Not Applicable (1011-9000-000); ANAM01304, Not Applicable (1011-9000-000); ANAN00487, Not Applicable (1011-9000-000); ANAN00488, Not Applicable (1011-9000-000); ANAN00489, Not Applicable (1011-9000-000); ANAN00490, Not Applicable (1011-9000-000); ANAN00491, Not Applicable (1011-9000-000); ANAN00492, Not Applicable (1011-9000-000); ANAR05143, Not Applicable (1011-9000-000); ANAR05144, Not Applicable (1011-9000-000); 5450447-2, Not Applicable (2097674-NL); 5450447-4, Not Applicable (2097674-NL); 5454351-1, Not Applicable (2097674-NL); 5454351-2, Not Applicable (2097674-NL); ANAR00198, Not Applicable (1011-9000-000); ANAR00199, Not Applicable (1011-9000-000); ANAR00200, Not Applicable (1011-9000-000); ANAR00279, Not Applicable (1011-9000-000); ANAM01311, Not Applicable (1011-9000-000); ANAM01314, Not Applicable (1011-9000-000); ANAM01316, Not Applicable (1011-9000-000); ANAM01319, Not Applicable (1011-9000-000); ANAM01321, Not Applicable (1011-9000-000); ANAM01322, Not Applicable (1011-9000-000); ANAM01323, Not Applicable (1011-9000-000); 5456012-1, Not Applicable (2097674-001); 5456012-2, Not Applicable (2097674-001); 5456012-3, Not Applicable (2097674-001); 5456719-1, Not Applicable (2097674-001); ANAN01065, Not Applicable (1011-9000-000); ANAN01066, Not Applicable (1011-9000-000); 5476468-1, Not Applicable (2097674-001); 5476468-10, Not Applicable (2097674-001); 5476468-11, Not Applicable (2097674-001); 5476468-12, Not Applicable (2097674-001); 5476468-13, Not Applicable (2097674-001); 5476468-14, Not Applicable (2097674-001); 5476468-15, Not Applicable (2097674-001); 5476468-16, Not Applicable (2097674-001); 5476468-17, Not Applicable (2097674-001); 5476468-2, Not Applicable (2097674-001); 5476468-3, Not Applicable (2097674-001); 5476468-4, Not Applicable (2097674-001); 5476468-5, Not Applicable (2097674-001); 5476468-6, Not Applicable (2097674-001); 5476468-7, Not Applicable (2097674-001); 5476468-8, Not Applicable (2097674-001); 5476468-9, Not Applicable (2097674-001); Additional Serial Numbers - Model Number, UDI Number, (Serial Number): 2097674-001-US, Not Applicable, (APWV01271); 2097674-001-US, Not Applicable, (APWU01812); 2097674-001-US, Not Applicable, (ANAR00714); 2097674-001-US, Not Applicable, (ANAR00741); 2097674-001-US, Not Applicable, (ANAS00330); 2097674-001-US, Not Applicable, (ANAS01319); 2097674-001-US, Not Applicable, (ANAS01320); 2097674-001-US, Not Applicable, (APWU00486); 2097674-001-US, Not Applicable, (APWU00487); 2097674-001-US, Not Applicable, (APWU00575); 2097674-001-US, Not Applicable, (APWU01267); 2097674-001-US, Not Applicable, (APWX00154); 2097674-001-US, Not Applicable, (APWX00155); 2097674-001-US, Not Applicable, (APWX00156); 2097674-001-US, Not Applicable, (APWX00157); 1011-9000-000, Not Applicable, (ANAS00456); 2097674-001-US, Not Applicable, (ANAM00764); 2097674-001-US, Not Applicable, (ANAM00765); 2097674-001-US, Not Applicable, (ANAM00766); 2097674-001-US, Not Applicable, (ANAM00767); 2097674-001-US, Not Applicable, (ANAM00768); 2097674-001-US, Not Applicable, (ANAM00769); 2097674-001-US, Not Applicable, (ANAM00770); 2097674-001-US, Not Applicable, (ANAM00771); 2097674-001-US, Not Applicable, (ANAM00772); 2097674-001-US, Not Applicable, (ANAM00773); 2097674-001-US, Not Applicable, (ANAM00774); 2097674-001-US, Not Applicable, (ANAM00775); 2097674-001-US, Not Applicable, (APWV00946); 2097674-001, Not Applicable, (ANAL00264); 2097674-001, Not Applicable, (ANAL00265); 2097674-001, Not Applicable, (ANAL00266); 2097674-001, Not Applicable, (ANAK00415); 2097674-001, Not Applicable, (ANAK00703); 2097674-001, Not Applicable, (ANAK00705); 2097674-001, Not Applicable, (ANAK00707); 2097674-001, Not Applicable, (ANAK00709); 2097674-001, Not Applicable, (ANAK00410); 2097674-001, Not Applicable, (ANAK00411); 2097674-001, Not Applicable, (ANAK00413); 2097674-001, Not Applicable, (ANAK00414); 2097674-001, Not Applicable, (ANAK00416); 2097674-001, Not Applicable, (ANAK00417); 2097674-001, Not Applicable, (ANAK00421); 2097674-001, Not Applicable, (ANAL00893); 2097674-001, Not Applicable, (APWU02246); 2097674-001, Not Applicable, (APWV00404); 2097674-001, Not Applicable, (ANAK00412); 2097674-001, Not Applicable, (ANAK00418); 2097674-001, Not Applicable, (ANAK00419); 2097674-001, Not Applicable, (ANAK00420); 2097674-001, Not Applicable, (ANAK00704); 2097674-001, Not Applicable, (ANAK00706); 2097674-001, Not Applicable, (ANAK00708); 2097674-001, Not Applicable, (APWW00334); 1011-9000-000, Not Applicable, (ANAR05371); 2097674-DE, Not Applicable, (ANAS00870); 1011-9000-000, Not Applicable, (ANAR00333); 1011-9000-000, Not Applicable, (ANAJ00122); 1011-9000-000, Not Applicable, (ANAQ01270); 1011-9000-000, Not Applicable, (ANAQ01271); 1011-9000-000, Not Applicable, (ANAQ01272); 1011-9000-000, Not Applicable, (ANAQ01273); 1011-9000-000, Not Applicable, (ANAQ01274); 1011-9000-000, Not Applicable, (ANAQ01275); 1011-9000-000, Not Applicable, (ANAQ01276); 1011-9000-000, Not Applicable, (ANAQ01277); 1011-9000-000, Not Applicable, (ANAR00187); 1011-9000-000, Not Applicable, (ANAR00188); 1011-9000-000, Not Applicable, (ANAR00189); 1011-9000-000, Not Applicable, (ANAR00190); 1011-9000-000, Not Applicable, (ANAR00191); 1011-9000-000, Not Applicable, (ANAR00192); 1011-9000-000, Not Applicable, (ANAR00334); 1011-9000-000, Not Applicable, (ANAR00701); 1011-9000-000, Not Applicable, (APWU02039); 1011-9000-000, Not Applicable, (APWW01068); 2097674-001, Not Applicable, (ANAQ00825); 1011-9000-000, Not Applicable, (ANAR05370); 2097674-FR, Not Applicable, (ANAN01100); 1011-9000-000, Not Applicable, (ANAT00157); 1011-9000-000, Not Applicable, (ANAT00158); 1011-9000-000, Not Applicable, (ANAT00156); 1011-9000-000, Not Applicable, (ANAR05372); |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha
WI
53188-1615
|
| For Additional Information Contact | 800-437-1171 |
Manufacturer Reason for Recall | Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure. |
FDA Determined Cause 2 | Software design |
| Action | GE Healthcare notified customers on about 09/27/2018 via "URGENT MEDICAL DEVICE CORRECTION" letter. The letter informed customers that they could continue to use the anesthesia device. The issue can only occur if the patient stops spontaneous breathing in PSVPro ventilation mode after completion of lung recruitment breaths. Stopping the Cycling Procedure resolves the issue. Continue to monitor patient as is normal for Apnea.
A GE Healthcare representative will contact each customer to arrange for the correction.
Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 4,207 units total |
| Distribution | Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Macedonia, Mexico, Netherlands, New Zealand, Nigeria, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|