|
Class 2 Device Recall Merge Eye Station |
|
Date Initiated by Firm |
December 20, 2018 |
Date Posted |
February 25, 2019 |
Recall Status1 |
Terminated 3 on February 28, 2023 |
Recall Number |
Z-0939-2019 |
Recall Event ID |
81869 |
510(K)Number |
K982689
|
Product Classification |
Camera, ophthalmic, ac-powered - Product Code HKI
|
Product |
The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System
Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.
|
Code Information |
UDI: 00842000100140. Serial Numbers: a0L3200000BVmMo, a0L3200000DyTSA, a0L3200000AiuLX, a0L3200000BUou2, a0L0e00000EPuiz, a0L3200000BVZ7i, a0L3200000Ajd0n, a0L3200000C04Ol, a0L3200000Aj6EX, a0L32000009gkeu, a0L3200000BUOXv, a0L3200000Aj6TP, a0L3200000BTnqr, a0L3200000AjTBg, a0L32000008q1mk, a0L0e00000DzQ3a, a0L3200000B2Csy, a0L32000009hJZk, a0L32000009n0mi, a0L0e00000E0aFN, a0L3200000DyFdA, a0L0e00000EPw0a, a0L3200000C03vm, a0L3200000DyF4P, a0L3200000AjY5P, a0L32000009n1Mk, a0L3200000B1MV5, a0L0e00000EPw6s, a0L32000008qtIF, a0L3200000BVzZ6, a0L3200000BTLiZ, a0L3200000Aj6bv, a0L0e00000E0C04, a0L3200000AihSR, a0L3200000AihUh, a0L3200000BUc9x, a0L3200000BUcAH, a0L32000009hLmm, a0L3200000AivBM, a0L3200000AjYPs, a0L32000009h8o4, a0L0e00000E07hv, a0L32000008quuk, a0L3200000DyNRS, a0L3200000B19Jt, a0L3200000B0uso, a0L0e00000E0CSc, a0L3200000B09PT, a0L3200000B09PJ, a0L3200000B09Pd, a0L3200000BTlfb, a0L3200000DiqQx, a0L32000009hJrd, a0L32000009mlt9, a0L32000009md53, a0L32000009nTWm, a0L0e00000EPxQj, a0L3200000BWKuB, a0L3200000B1N0q, a0L3200000Dyg4I, a0L0e00000Dzwru, a0L3200000BzVK2, a0L3200000AjnnW, a0L3200000DySwx, a0L3200000BgdOF, a0L32000009h6WM, a0L32000009h6WR, a0L3200000B0CXr, a0L3200000Aitxy, a0L32000009hJ6P, a0L3200000DyGJu, a0L60000007LRa6, a0L3200000DyFXC, a0L60000007OcXf, a0L3200000AjkWQ, a0L3200000Di2CQ, a0L3200000BWZGK, a0L0e00000Dkz6R, a0L3200000AjbF3, a0L3200000DyGL3, a0L3200000DyGKd, a0L3200000DyGKs, a0L3200000BTYBz, a0L3200000AjSo7, a0L3200000AjmbA, a0L0e00000DxTju, a0L32000009eaXj, a0L3200000AivsT, a0L32000009hA2T, a0L32000009hZDo, a0L3200000B0vwx, a0L3200000DiGem, a0L32000008qVOb, a0L3200000AjlpT, a0L0e00000DzgWg, a0L32000009mzAO, a0L3200000DiF29, a0L3200000DiF1z, a0L3200000AjYmf, a0L3200000DiSYx, a0L0e00000E0TgS, a0L3200000BzJCA, a0L3200000AjbIq, a0L3200000B0A6r, a0L3200000Aiiis, a0L0e00000DxSED, a0L0e00000EPw0k, a0L3200000DyGKE, a0L3200000DyGKT, a0L3200000DyGKY, a0L0e00000DzNNe, a0L32000009h8mS, a0L3200000B0Un4, a0L3200000B0Un9, a0L3200000B0Umz, a0L3200000B0UnE, a0L0e00000EPvHa, a0L32000009mmxx, a0L3200000AjVHx, a0L3200000DyFYj, a0L0e00000E0CWU, a0L3200000Diqej, a0L3200000AjVby, a0L32000009mpFG, a0L32000009mqPj, a0L3200000BWZRT, a0L0e00000EPtf1, a0L32000009h5NQ, a0L0e00000DkuPo, a0L3200000AifYF, a0L0e00000EPwfs, a0L3200000AjTri, a0L0e00000EPvs7, a0L32000009hJJp, a0L0e00000DzNNy, a0L3200000DhRjV, a0L0e00000DzXBa, a0L3200000BzIUD, a0L3200000AihtL, a0L32000009hLHD, a0L3200000DycpT, a0L3200000Dyd75, a0L60000007OvfM, a0L32000009hXaG, a0L3200000B1z3N, a0L3200000Ajq8e, a0L3200000Bzsim, a0L3200000Bzsjf, a0L32000009g59e, a0L3200000BToIj, a0L32000009nGPH, a0L3200000DiGFZ, a0L0e00000EPqr6, a0L3200000Bg6tH, a0L0e00000DzNPz, a0L3200000AjMEo, a0L3200000BV8Dk, a0L3200000AjMbE, a0L0e00000DzQdh, a0L3200000AijMO, a0L3200000B1ZXY, a0L0e00000EPwCg, a0L3200000AjPKO, a0L32000009h6Sx, a0L3200000B0CEa, a0L3200000B0CF9, a0L3200000BgJfJ, a0L3200000Ajp2G, a0L32000009nGVt, a0L3200000B0S7X, a0L3200000B0Cwo, a0L0e00000EPtfV, a0L0e00000DzV9w, a0L0e00000DzgiM, a0L32000009hKk6, a0L0e00000E0lM5, a0L3200000DhOvR, a0L32000009hVqm, a0L3200000Ajfc7, a0L0e00000EPwYN, a0L0e00000DzNSU, a0L3200000BVTjP, a0L3200000B1O4R, a0L32000009nWFW, a0L3200000B2Rmk, a0L0e00000DzVCW, a0L3200000AiiWw, a0L0e00000DqRSC, a0L32000009eHT9, a0L32000009hKk1, a0L32000009nVXX, a0L0e00000E0CMe, a0L3200000Aiwq1, a0L32000009h9dO, a0L32000009nGjy, a0L3200000Di1aa, a0L0e00000Dl0rE, a0L0e00000Dl1Jr, a0L3200000B0hDN, a0L3200000DiqN0, a0L32000009haQu, a0L0e00000DkuPK, a0L3200000B0CZd, a0L32000009hYJt, a0L3200000BWlgS, a0L0e00000E07gx, a0L0e00000EPuG2, a0L0e00000EPuFx, a0L32000009hMxY, a0L3200000B0u8J, a0L3200000BVUTo, a0L32000009hJuN, a0L3200000B0UNR, a0L3200000AjAkv, a0L3200000Ajnb9, a0L3200000BVmPE, a0L3200000DiTgA, a0L3200000BWkbl, a0L0e00000EfMBR, a0L32000009hKc9, a0L0e00000DzO9k, a0L3200000DhguX, a0L3200000Ajp39, a0L32000009h59x, a0L0e00000E0Sd7, a0L3200000AisHf, a0L32000009eHg3, a0L0e00000DzPyQ, a0L0e00000DzPzJ, a0L3200000DigHt, a0L0e00000DzQ2I, a0L3200000AjkZC, a0L0e00000DqRTF, a0L0e00000DqRTe, a0L3200000BVmIC, a0L3200000AjPzb, a0L0e00000DqRTU, a0L32000008qpH3, a0L32000008qpHD, a0L32000008qpGy, a0L0e00000Dzwoq, a0L0e00000E0kP3, a0L3200000BVcwJ, a0L3200000DiRkQ, a0L3200000DiRkG, a0L0e00000EPv0F, a0L0e00000EPv0A, a0L3200000BWNOD, a0L3200000AjPGP, a0L0e00000EfMDc, a0L3200000AjV66, a0L3200000AjPD0, a0L3200000B1KJW, a0L3200000B1KJM, a0L32000009nUI7, a0L0e00000EPrer, a0L3200000BVynO, a0L3200000DyNVK, a0L32000008qV53, a0L0e00000E0ltx, a0L3200000BVmij, a0L3200000BUmKq, a0L32000009hXyD, a0L32000009hXxo, a0L3200000BWkCU, a0L3200000BVVac, a0L3200000B1KJl, a0L3200000BV8DV, a0L3200000B1KJR, a0L3200000BUo9h, a0L0e00000EPt0y, a0L3200000Ajkr2, a0L3200000AjTRX, a0L3200000B0jAs, a0L0e00000DxY42, a0L3200000DhDXm, a0L3200000DhsCj, a0L3200000BWl3a, a0L32000009moWy, a0L3200000BVKpV, a0L3200000BUx34, a0L3200000BUx2z, a0L3200000DyF7d, a0L3200000BUx2a, a0L3200000Bgbdg, a0L3200000BUYqX, a0L3200000BWNQG, a0L3200000Bzsb1, a0L32000009ngLZ, a0L32000009ngPx, a0L32000009ngQl, a0L3200000BUx3E, a0L3200000DhqKP, a0L32000009hY1q, a0L3200000AjMVk, a0L3200000AisQc, a0L32000009hW71, a0L32000009hKtX, a0L3200000BWky0, a0L3200000BWkxg, a0L3200000BWkwa, a0L3200000BWkxq, a0L3200000BWkxv, a0L32000008rXRq, a0L32000009hXIW, a0L0e00000DxSLJ, a0L32000009mdAh, a0L3200000B21Uq, a0L3200000B1nIO, a0L3200000BVXjc, a0L3200000BVXjm, a0L0e00000EPwYr, a0L3200000Di3H1, a0L32000009hMo2, a0L3200000B0UB0, a0L3200000BWBJP, a0L0e00000EPsXc, a0L32000009nFoF, a0L0e00000E0Bpu, a0L3200000DisEg, a0L3200000AjZUl, a0L32000009hZn7, a0L3200000AjegQ, a0L3200000Dhryh, a0L3200000B0vDR, a0L32000009hJPR, a0L32000009h6dj, a0L3200000DhggD, a0L0e00000E0C3r, a0L0e00000DzxLG, a0L3200000Bz6LS, a0L32000009hZh6, a0L0e00000Dkz05, a0L0e00000DzQHF, a0L3200000B0D1f, a0L0e00000DzVCC, a0L3200000B19uW, a0L32000009n0w5, a0L0e00000EPwmF, a0L32000009nDM1, a0L3200000BTOG3, a0L0e00000Dkyte, a0L3200000BTOGD, a0L3200000BgdIO, a0L3200000B1apU, a0L3200000BTOFy, a0L32000009gtEq, a0L3200000BUwZ9, a0L3200000AjehT, a0L0e00000E0mzD, a0L3200000DhDh3, a0L32000009h9mG, a0L3200000AjN8S, a0L32000009hLSL, a0L32000009hLSa, a0L3200000BUlrM, a0L32000009hYik, a0L32000009hZnH, a0L0e00000DzQeV, a0L3200000AixBW, a0L3200000DiGzi, a0L3200000DiGzx, a0L3200000DiH1q, a0L3200000Dighg, a0L3200000Dighb, a0L3200000BVxRI, a0L3200000BVxQt, a0L3200000BVxRS, a0L3200000BVxRN, a0L3200000DiGyQ, a0L3200000DiIEq, a0L3200000DiIBF, a0L3200000DiIAW, a0L3200000DiIEK, a0L3200000DiIFR, a0L3200000DiIFF, a0L3200000DiI8u, a0L3200000DiI8a, a0L3200000DiIE5, a0L3200000DiIB0, a0L3200000DiIEg, a0L3200000BVn2X, a0L3200000Di2do, a0L3200000Di2dZ, a0L3200000Di2eh, a0L3200000Di2dF, a0L3200000Di2ie, a0L3200000Di2j8, a0L0e00000DqRRO, a0L0e00000DqRRJ, a0L3200000DyaP8, a0L3200000DyaPr, a0L3200000DipDP, a0L3200000DipDU, a0L32000009hNAh, a0L3200000AjBEA, a0L3200000B2RoH, a0L3200000BzLg5, a0L3200000BzKLe, a0L3200000DhgNk, a0L3200000DhEbU, a0L0e00000EfLoI, a0L3200000B0tAC, a0L3200000Dpw2q, a0L3200000Dykp5, a0L0e00000DzxL1, a0L0e00000DxV8c, a0L0e00000DzNRl, a0L32000009nGRc, a0L32000009hIx7, a0L3200000BUcNx, a0L32000009hZnM, a0L3200000DyNth, a0L3200000BWBPm, a0L3200000B0SF2, a0L3200000DyPqR, a0L0e00000EfKJE, a0L3200000B2Eu9, a0L0e00000DzO7Z, a0L32000009hIxW, a0L32000009hYBL, a0L0e00000E0Bij, a0L32000009hK2S, a0L32000008pHlh, a0L60000008pCXi, a0L32000008pO3Z, a0L3200000DyIQ0, a0L3200000Bg4iC, a0L0e00000EPu4L, a0L3200000BVTqg, a0L0e00000EPs5i, a0L0e00000DzO8I, a0L3200000Aj9Af, a0L60000008p87v, a0L0e00000DzXFI, a0L3200000BVTsX, a0L0e00000DzXFX, a0L3200000B0vFN, a0L3200000B1kSl, a0L3200000Ajpyn, a0L3200000BUlH5, a0L3200000BV8dg, a0L3200000DyNrB, a0L0e00000Dzboz, a0L3200000Ajgrh, a0L3200000AiuS8, a0L3200000AiuyH, a0L3200000AiwYK, a0L3200000Aj5z5, a0L3200000Aj5zA, a0L3200000Aj61f, a0L3200000Aj61a, a0L3200000Aji1r, a0L0e00000DzQRd, a0L3200000B1KIY, a0L3200000BVTsh, a0L0e00000E0CCo, a0L0e00000DzQGH, a0L3200000Aj8aE, a0L3200000AjA3K, a0L3200000AjBG6, a0L3200000DzDpv, a0L0e00000DzNCl, a0L0e00000DzQO0, a0L3200000AjozO, a0L0e00000EPtrW, a0L3200000DykPy, a0L0e00000EPpTN, a0L0e00000EPpi8, a0L0e00000EPtrM, a0L0e00000E0Bpp, a0L3200000Aj8Qg, a0L3200000B1KIn, a0L0e00000E0mlz, a0L0e00000E0CXD, a0L3200000AjoCZ, a0L3200000BVmV0, a0L3200000BWlDd, a0L0e00000DzOB2, a0L3200000Dy3n9, a0L0e00000DxPD2, a0L0e00000DzXEF, a0L3200000DyVe9, a0L32000009hJNF, a0L0e00000DqRbP, a0L0e00000E0hr9, a0L3200000B186W, a0L3200000BWlCT, a0L3200000BVmTO, a0L32000009hJ2E, a0L32000009hYSp, a0L0e00000EPuII, a0L3200000Bgb9h, a0L3200000DpwMM, a0L3200000BziPr, a0L0e00000DzQ5z, a0L0e00000EPtgs, a0L0e00000EPtgn, a0L3200000BTZVI, a0L0e00000DzNNA, a0L0e00000DkuTX, a0L3200000AjegV, a0L0e00000Dzgk3, a0L3200000B1KIE, a0L3200000B1KI9, a0L3200000B1KIJ, a0L0e00000DqRZs, a0L32000009gkit, a0L3200000Dj3TN, a0L3200000DpkOH, a0L3200000C03yW, a0L32000009mmzZ, a0L32000009nUtq, a0L0e00000Dl5zm, a0L3200000AjSK9, a0L3200000AjSM7, a0L0e00000DzQZ7, a0L0e00000EPvdq, a0L32000009mq3U, a0L0e00000E0CAx, a0L0e00000EPrKD, a0L0e00000EPrK8, a0L32000009hIvw, a0L3200000BWYT3, a0L0e00000DzQfd, a0L32000008peLC, a0L3200000Aj624, a0L0e00000EPpSt, a0L0e00000DxPtS, a0L0e00000Dl5zr, a0L3200000B0wf2, a0L0e00000DzQ3f, a0L3200000AjhJF, a0L0e00000DzcfG, a0L0e00000EPsXh, a0L3200000Bz7pS, a0L3200000Ajkmq, a0L3200000AjXH2, a0L0e00000E0BcB, a0L0e00000EPrx5, a0L3200000B0woQ, a0L0e00000EPqIL, a0L0e00000EPqMN, a0L32000009n2qT, a0L3200000B0f1C, a0L3200000B0f1b, a0L3200000B0S8a, a0L3200000B0T7a, a0L3200000BWl28, a0L0e00000EPu3r, a0L0e00000DzbyW, a0L3200000B1Zcv, a0L32000009hIf1, a0L3200000Dyke1, a0L3200000DhDCk, a0L3200000AjeiM, a0L3200000AjTtA, a0L0e00000DqRbK, a0L0e00000Dl5zh, a0L0e00000DzQ3x, a0L0e00000E0C0d, a0L32000009hIfi, a0L3200000B0tqM, a0L3200000B18Ce, a0L3200000AifGK, a0L3200000AifGF, a0L32000009h8kM, a0L3200000DhFZ4, a0L3200000BTOG8, a0L3200000B2Fgn, a0L0e00000DxS0B, a0L32000009ekM1, a0L0e00000DxPfB, a0L3200000B19N2, a0L3200000DyP9F, a0L0e00000E0btG, a0L3200000Aihc2, a0L32000009hY8b, a0L0e00000E0h97, a0L3200000AjYzA, a0L3200000BVSms, a0L32000009nTPM, a0L32000009n1Qw, a0L32000009nExk, a0L3200000BVYoq, a0L32000009nTzg, a0L3200000BVc7t, a0L3200000BVc8I, a0L3200000BVc8S, a0L3200000BVc5M, a0L3200000BVc7y, a0L3200000BVc83, a0L3200000DhCjz, a0L3200000BUNbm, a0L3200000BUNe7, a0L3200000BUNd4, a0L3200000BUNbI, a0L3200000BVYzj, a0L32000009hMk1, a0L32000009hZSD, a0L3200000AifcR, a0L3200000BTnsU, a0L3200000BTnso, a0L3200000DyFko, a0L3200000DyFmG, a0L0e00000DxPlY, a0L32000008pd76, a0L3200000B16Cm, a0L3200000BUxUl, a0L3200000BzUMv, a0L32000009hJ6o, a0L3200000AjASC, a0L0e00000DzNFu, a0L3200000AiggP, a0L3200000Aiipi, a0L32000009hNOH, a0L3200000Aige9, a0L3200000DhrGg, a0L3200000Dhbt5, a0L3200000AjAJn, a0L3200000AjAKR, a0L3200000AjAK2, a0L3200000Aigda, a0L3200000Aifnw, a0L3200000Aiip4, a0L3200000DpljY, a0L3200000B1zaL, a0L3200000B16MP, a0L3200000DyGJz, a0L3200000DisvM, a0L3200000Aiwl6, a0L32000009ha9q, a0L32000009nE1z, a0L3200000BUl3E, a0L32000009hZ7R, a0L3200000AjVLG, a0L3200000BUm77, a0L32000009nEr6, a0L0e00000EPxLF, a0L3200000BVahZ, a0L32000009hX70, a0L32000009hJGH, a0L3200000Bgb9r, a0L32000008pHtq, a0L3200000B0veu, a0L0e00000E0CWK, a0L3200000BUm7M, a0L3200000DidmF, a0L3200000Didll, a0L3200000Didnc, a0L3200000BVW8d, a0L3200000BVW97, a0L0e00000EPrRY, a0L3200000B0OJX, a0L0e00000EPrFh, a0L0e00000DzgfI, a0L0e00000EfL3R, a0L0e00000EfL3W, a0L0e00000EPwre, a0L3200000BWaHH, a0L3200000Ajkq4, a0L32000009hMSr, a0L3200000AiijM, a0L3200000B0tTZ, a0L32000008qpGo, a0L32000008qpGt, a0L32000008qaaX, a0L32000008pZJi, a0L3200000AjkaK, a0L0e00000EPx2S, a0L3200000BVmNb, a0L3200000BVKme, a0L0e00000EPx2c, a0L0e00000DkuT3, a0L3200000Aigf2, a0L32000009haOZ, a0L32000009mp2z, a0L0e00000DxQEa, a0L0e00000E0lKY, a0L0e00000E0lKE, a0L3200000B1a5Z, a0L3200000BVbN4, a0L3200000AjhGF, a0L3200000AiiLy, a0L0e00000DzQ2h, a0L3200000Ajh6K, a0L3200000AisNY, a0L3200000Bgbi2, a0L32000009hVxe, a0L32000009hVy3, a0L32000009hVyS, a0L3200000AiimE, a0L32000009gwY2, a0L0e00000EPqDG, a0L3200000BVXWN, a0L3200000BVXWI, a0L3200000B1nQd, a0L3200000BzW13, a0L3200000B2CsP, a0L3200000BTyHe, a0L0e00000EPwIj, a0L3200000AjhCX, a0L3200000BTbQq, a0L3200000Aig8T, a0L3200000Aiw33, a0L3200000DynjL, a0L3200000AjTxC, a0L3200000Aivsn, a0L32000009hJOY, a0L3200000BWYIG, a0L3200000B22dy, a0L3200000BTybd, a0L3200000DhDen, a0L3200000AiiA3, a0L0e00000E0kSM, a0L0e00000DqRT0, a0L3200000AjnIa, a0L0e00000EfMTG, a0L3200000BTmsS, a0L3200000B0h3V, a0L32000009hLiy, a0L0e00000Dkz4u, a0L0e00000Dkyyn, a0L3200000BgLqL, a0L3200000AjjTm, a0L3200000AjS6i, a0L3200000AjOLD, a0L3200000BgJkx, a0L0e00000EfKmz, a0L3200000DyIxV, a0L3200000AjPaG, a0L3200000AjmXw, a0L3200000B1kbk, a0L3200000B0B6s, a0L3200000BUOzK, a0L3200000BUm48, a0L32000008qpGP, a0L32000009h60O, a0L3200000B0i5q, a0L3200000BWN3n, a0L3200000BWN5B, a0L3200000AjLAG, a0L3200000DyU4c, a0L0e00000Dl1Cl, a0L0e00000E0BxK, a0L32000009hXeU, a0L3200000Aj629, a0L3200000DzIIq, a0L3200000DzIIl, a0L32000009hXeK, a0L32000009n1pe, a0L0e00000DzQep, a0L3200000BUcIY, a0L3200000DiseL, a0L0e00000DzwyH, a0L32000009hX7A, a0L3200000BVWmP, a0L0e00000DzVCv, a0L3200000Dynno, a0L3200000DiqbW, a0L32000009eZco, a0L3200000DyItp, a0L3200000Aj6oe, a0L3200000AjM8N, a0L0e00000Dl1RC, a0L3200000B0f6G, a0L0e00000EPqwa, a0L32000009hVu6, a0L32000008qpGj, a0L3200000B09OB, a0L3200000AjnNY, a0L3200000B20ld, a0L3200000DyHC3, a0L3200000AiwN7, a0L3200000BWmpV, a0L3200000B0w8C, a0L3200000B0Sr6, a0L3200000B1KiP, a0L32000009hapG, a0L0e00000EfKhB, a0L3200000DyOJH, a0L3200000AjqBE, a0L0e00000DzgkI, a0L3200000BV6eb, a0L3200000BV8cJ, a0L3200000AiiYx, a0L3200000AjhWA, a0L3200000Aj87N, a0L3200000BgdHL, a0L3200000BWa9s, a0L3200000BgM7L, a0L3200000BgbBT, a0L3200000Bg6XB, a0L0e00000DqRU8, a0L0e00000DzQEz, a0L3200000AjpGS, a0L0e00000EPu4z, a0L3200000B1Y8Q, a0L3200000Ajk9l, a0L3200000Di3cq, a0L3200000AjeEV, a0L3200000DyJIr, a0L3200000BVJwX, a0L32000009hIst, a0L0e00000Dl0cx, a0L3200000BVHcC, a0L32000009hYJj, a0L3200000Ajm62, a0L32000008qtHv, a0L3200000Di4RX, a0L32000009hKvB, a0L3200000B0Ayn, a0L3200000Aj914, a0L3200000AjZVc, a0L0e00000DxZBY, a0L3200000Di3Fo, a0L3200000B218m, a0L3200000BWBlr, a0L3200000DySF8, a0L32000009hImk, a0L3200000BVmUW, a0L3200000Dyi3I, a0L32000009hA37, a0L3200000BzLRT, a0L3200000C02sQ, a0L32000009n2tX, a0L3200000DyPps, a0L3200000AiwEu, a0L0e00000EPsWo, a0L32000009mchG, a0L0e00000EPphA, a0L0e00000Dl1I5, a0L3200000AigoM, a0L3200000BVU1o, a0L0e00000Dkz6M, a0L3200000BVZuY, a0L32000009hNLb, a0L32000009n0iW, a0L3200000BUxSF, a0L3200000BV7Wv, a0L32000008qvEm, a0L3200000B2GbM, a0L3200000AjZJo, a0L0e00000EfJmj, a0L3200000B16Bo, a0L3200000DzHbs, a0L0e00000DzPwy, a0L0e00000DzPwj, a0L0e00000DzPwt, a0L3200000Ajjba, a0L0e00000DxR4b, a0L3200000AjdH9, a0L0e00000DzgiW, a0L3200000AjMQ6, a0L3200000AjMR4, a0L3200000BTluh, a0L3200000BTNp3, a0L3200000B20Wy, a0L3200000Dykyy, a0L3200000BVa5h, a0L3200000BWlT8, a0L3200000AjYa4, a0L3200000Ajaxs, a0L3200000AigET, a0L0e00000Dl0Bc, a0L3200000AjkPm, a0L0e00000Dl0Dn, a0L3200000DirIM, a0L0e00000EPu3X, a0L32000009hYBQ, a0L3200000B0ihC, a0L3200000AjNFq, a0L3200000DyWZW, a0L3200000B0eZl, a0L0e00000EPraL, a0L0e00000DxRZA, a0L32000009hM6S, a0L3200000BWkUy, a0L3200000Aiufr, a0L32000009mz67, a0L32000009hZGi, a0L0e00000DzgWq, a0L0e00000DzxMx, a0L0e00000EPr0D, a0L0e00000DxPpp, a0L3200000DhPHT, a0L3200000BTmZD, a0L0e00000DzlGS, a0L0e00000Dzbou, a0L3200000Dj4dg, a0L3200000BWkVI, a0L3200000BUNqA, a0L3200000DhpuB, a0L32000009ngPJ, a0L0e00000Dzwsn, a0L3200000DyqkE, a0L0e00000E0khR, a0L3200000Aitp5, a0L3200000BVYs4, a0L32000009ha19, a0L0e00000Dl5tt, a0L0e00000EPtpa, a0L3200000Ajbnf, a0L3200000Ajbs6, a0L3200000Ajbs1, a0L3200000B0fFK, a0L3200000Aje6m, a0L32000009hM0V, a0L3200000B0SmJ, a0L3200000AjVgP, a0L0e00000E0By8, a0L0e00000E0Bvi, a0L32000008qv9I, a0L32000008qv4h, a0L32000009hX7j, a0L32000009h9aE, a0L3200000BVn5C, a0L32000009e66V, a0L3200000Aj9ri, a0L3200000BVo5n, a0L3200000BVn4x, a0L3200000Aj9t5, a0L3200000Ajbaa, a0L3200000Aj9g1, a0L3200000Aj9gB, a0L3200000BVo5J, a0L0e00000EPsYQ, a0L32000009n1hH, a0L3200000BVo6C, a0L0e00000DxS0z, a0L3200000B0PYk, a0L0e00000EPtgE, a0L3200000BWbMz, a0L32000009mnxA, a0L0e00000Dl0VN, a0L0e00000Dl19c, a0L32000009nVre, a0L0e00000Dl1Az, a0L3200000Ajkg3, a0L32000009n30A, a0L3200000B0QR9, a0L3200000BTyCS, a0L3200000BgdVb, a0L0e00000E0kJP, a0L32000008rNGA, a0L3200000AihQV, a0L3200000Bz8Ab, a0L3200000Ajm9f, a0L32000009ng6Y, a0L3200000AiiNp, a0L0e00000EfL8R, a0L32000009h528, a0L3200000AjZ6p, a0L32000009nGFd, a0L32000009n0DG, a0L3200000AjkVw, a0L3200000BUl2k, a0L3200000B0Np6, a0L0e00000DkymE, a0L3200000DhS47, a0L32000009hIxq, a0L32000009md4Z, a0L0e00000Dl0Hz, a0L0e00000DkyrT, a0L0e00000Dl0Hp, a0L0e00000Dl0Hu, a0L0e00000Dl0I9, a0L3200000AifR8, a0L3200000Dhqyz, a0L3200000B21lP, a0L0e00000DzOBC, a0L3200000BTmHc, a0L0e00000E0BLA, a0L3200000B0U1R, a0L3200000B0U0n, a0L3200000B0U17, a0L3200000DyFXq, a0L3200000DyKil, a0L0e00000Dzbun, a0L3200000BWLpj, a0L32000009mdB6, a0L32000009mdAr, a0L32000009mdBB, a0L3200000DyFX2, a0L3200000DyFXl, a0L0e00000EPtxe, a0L0e00000EPwI5, a0L3200000BWO6L, a0L0e00000E0CYL, a0L3200000B1pSc, a0L0e00000EPrIg, a0L32000008pw81, a0L3200000BVTgG, a0L3200000B0ifE, a0L3200000B18AJ, a0L3200000AjogU, a0L3200000Bg7Nm, a0L3200000BzuJl, a0L60000007OXJQ, a0L0e00000EPutT, a0L32000009hK2H, a0L3200000Ajk9q, a0L3200000AjqD6, a0L3200000B0Ql9, a0L3200000DyHLS, a0L3200000Aj8Rt, a0L32000009hYBp, a0L3200000DpkqA, a0L3200000DyFYo, a0L3200000B0Ulh, a0L32000009h5Rw, a0L3200000B1Y5g, a0L0e00000DxQ38, a0L3200000BVIfK, a0L3200000DhCAh, a0L0e00000Dkz70, a0L32000009h5VN, a0L0e00000EPwlM, a0L3200000BVTfX, a0L0e00000EPr9K, a0L3200000B1KGw, a0L0e00000EPrAI, a0L0e00000E0BkL, a0L3200000BWBnO, a0L0e00000Dzbta, a0L32000009hJnR, a0L0e00000DqRZE, a0L32000009hX3w, a0L3200000AjfOU, a0L3200000AjjLs, a0L32000009nUHn, a0L3200000AjpNF, a0L3200000BUZR4, a0L3200000AjaSq, a0L0e00000EPrfp, a0L3200000BgR0I, a0L3200000BgR0S, a0L3200000BgR0D, a0L3200000BgR0X, a0L32000009nGQU, a0L3200000B19GF, a0L3200000BzVBr, a0L3200000BVY9F, a0L32000008qvLb, a0L32000008qpGZ, a0L0e00000EPq9Y, a0L0e00000EPq9O, a0L0e00000EPq9T, a0L0e00000EPq9J, a0L0e00000EPq9E, a0L0e00000EPto8, a0L0e00000EPscc, a0L0e00000Dkz1H, a0L0e00000Dkz2F, a0L0e00000Dl10f, a0L3200000Ajhtn, a0L3200000B2RzI, a0L32000009hNXQ, a0L0e00000DxY55, a0L3200000B0OEO, a0L3200000DyNJa, a0L32000008pH9X, a0L3200000Dhr5z, a0L3200000BVJFh, a0L32000009h9e2, a0L3200000BVKnN, a0L32000009hX7t, a0L3200000BTNBM, a0L3200000AjPyi, a0L32000009hA1d, a0L3200000AiuMk, a0L32000009hKKE, a0L0e00000EPsYp, a0L3200000Aitep, a0L3200000B1Klq, a0L0e00000DxP45, a0L3200000B2CZF, a0L3200000DpkhX, a0L32000008pbCK, a0L3200000DiFTa, a0L3200000BVUm7, a0L3200000BVXWm, a0L3200000BVXWh, a0L3200000AijvM, a0L3200000AijI0, a0L3200000AixJx, a0L32000009gYXr, a0L0e00000EPuoJ, a0L3200000DyPo6, a0L3200000B2QVC, a0L60000005LLQf, a0L3200000DyGTf, a0L3200000DyGS3, a0L32000009hYsJ, a0L3200000DyGTL, a0L3200000DyFYF, a0L3200000DyFYA, a0L0e00000DxT8L, a0L32000009gkfs, a0L3200000BUG6D, a0L3200000B0CZ9, a0L3200000Ajdak, a0L3200000Aj6i9, a0L3200000BVyAO, a0L32000008pZIe, a0L3200000AiwEa, a0L3200000AiwHp, a0L3200000DhPIC, a0L32000008qrs1, a0L0e00000Dl0gu, a0L3200000BWLFQ, a0L3200000BWLFB, a0L3200000BWAI4, a0L32000008pw9T, a0L32000009md4o, a0L3200000DyNPR, a0L3200000DhbAm, a0L3200000DhbAN, a0L3200000DyPro, a0L0e00000DzNIP, a0L3200000BWA1i, a0L0e00000Dzgf3, a0L3200000DyFiW, a0L3200000DyFfw, a0L0e00000E0C41, a0L3200000B0Umk, a0L32000009n3K7, a0L3200000B0U6m, a0L3200000B0U7B, a0L3200000DyLQp, a0L3200000B1my2, a0L32000009mcLl, a0L3200000Dpw1O, a0L32000009nUrR, a0L32000009hM9C, a0L32000009hKZs, a0L0e00000Dl0Bh, a0L3200000Aiv0I, a0L3200000BV9Vc, a0L3200000B0iGa, a0L0e00000E0gyT, a0L0e00000EfLJ0, a0L3200000Aj6ep, a0L3200000B2GLq, a0L32000009hX9V, a0L3200000BTbl9, a0L3200000BTMhy, a0L32000009n0Qa, a0L3200000B17w6, a0L3200000DyOtf, a0L3200000DyOu9, a0L32000008pKtJ, a0L0e00000DzQFx, a0L0e00000DzQFs, a0L3200000DyOux, a0L3200000DyOv7, a0L3200000DyOvq, a0L3200000Ajm8w, a0L0e00000E0SdW, a0L32000009hMXd, a0L3200000BVoob, a0L3200000BTad3, a0L3200000AjYAK, a0L0e00000DzNIj, a0L32000009hKwx, a0L3200000BWZQz, a0L3200000BUak3, a0L3200000DykWQ, a0L3200000DykVw, a0L3200000AjViM, a0L32000009nTTt, a0L3200000DhFbZ, a0L3200000DhF2F, a0L3200000DhNew, a0L3200000DhNer, a0L3200000DhNf1, a0L32000009mmXo, a0L32000009nGQy, a0L3200000BTaTO, a0L3200000B1Xxt, a0L3200000DhdZN, a0L3200000DhdZr, a0L3200000DhdYU, a0L3200000DhdZh, a0L3200000DhdZw, a0L3200000DhdZD, a0L3200000BgNch, a0L3200000AjnHD, a0L3200000BWZQu, a0L3200000BW02U, a0L3200000BzLLd, a0L32000009hX2y, a0L0e00000DzQEk, a0L3200000BzJOt, a0L0e00000DzVBi, a0L32000009nU2R, a0L3200000B1Znb, a0L3200000AjVxg, a0L32000009hJDr, a0L3200000BzhZ6, a0L3200000B09U9, a0L0e00000Dl0mx, a0L0e00000Dkyl1, a0L0e00000EfKgh, a0L3200000Bg6f0, a0L32000009hZ7v, a0L3200000Bzu7S, a0L32000008pUze, a0L32000008pUzy, a0L0e00000DxYph, a0L32000009mpoW, a0L3200000DyT5R, a0L0e00000E0M5O, a0L32000009n0Yo, a0L3200000DyDHm, a0L32000009mplR, a0L0e00000DxTU2, a0L0e00000E0CSX, a0L3200000BTZcU, a0L0e00000EPpDo, a0L3200000DhSCV, a0L3200000Ajd0d, a0L3200000BTlbK, a0L3200000B0jiZ, a0L0e00000EPqsd, a0L3200000BTyEe, a0L32000009hWVI, a0L3200000Dy3oR, a0L0e00000E0mcT, a0L0e00000E0mlk, a0L32000009nUQB, a0L3200000Ajhe7, a0L3200000Ajj12, a0L0e00000DxU41, a0L3200000AiimO, a0L3200000BVW8E a0L3200000Aihb4, a0L32000008rXSF, a0L32000008rXRx, a0L32000008rXSP, a0L3200000AihZr, a0L3200000B1zXx, a0L3200000BVZpa, a0L3200000AifYA, a0L60000008p4aW, a0L32000009hJo0, a0L3200000BUoJ3, a0L32000009hKnF |
Recalling Firm/ Manufacturer |
Merge Healthcare, Inc. 900 Walnut Ridge Dr Hartland WI 53029-8347
|
For Additional Information Contact |
262-367-0700
|
Manufacturer Reason for Recall |
Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
|
FDA Determined Cause 2 |
Device Design |
Action |
On about 12/20/2018 Merge Healthcare notified customers via "Urgent Field Safety Notice." The notice informed customers that certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. A leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
Instructions included to immediately cease use of all Eye Station hardware and that a Merge Healthcare representative will contact each customer to schedule a correction once the modification is available.
Questions regarding the recall notification can be directed to customer support at Eyecaresupport@us.ibm.com or 877-741-5369. |
Quantity in Commerce |
1,252 Units |
Distribution |
Distribution nationwide to Alabama, Alaska, Arizona, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, , Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West , Virginia, Wisconsin.
International distribution to Algeria, Austria, Belgium, Canada, Czech Republic, Denmark, Germany, Israel, Netherlands, Puerto Rico, United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HKI and Original Applicant = OPHTHALMIC IMAGING SYSTEMS
|
|
|
|