| | Class 2 Device Recall ApexPro Telemetry Server System |  |
| Date Initiated by Firm | August 30, 2019 |
| Date Posted | October 23, 2019 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0175-2020 |
| Recall Event ID |
83813 |
| 510(K)Number | K000779 K021325 K032369 K033365 K080251 |
| Product Classification |
Detector and alarm, arrhythmia - Product Code DSI
|
| Product | ApexPro Telemetry Server System. Also identified as Modification To: ApexPro Telemetry System and ApexPro FH Telemetry System - Product Usage: These systems are typically deployed in sub acute care areas in hospitals or clinical sites where patient mobility can enhance the effectiveness of the medical procedures administered. |
| Code Information |
1. Product Code: 3F, No UDI Number, Serial Numbers: M13F1043G, B23F1531G, B23F1532G, B23F2216G, A23F1290G, B23F1324G, M13F1152G, M13F1153G, M13F1154G, M13F1155G, M13F1164G, M13F1165G, M13F1166G, M13F1168G, M13F1169G, L13F0687G, D23F1925G, D23F1968G, E23F2074G, E23F2075G, B23F1402G, M13F1059G, M13F1173G, M13F1109G, C23F1815G, A23F1260G, J13F0272G, C23F1607G, D23F1981G, M13F0968G, L13F0683G, L13F0576G, J13F0274G, B23F1302G, K13F0437G, C23F1793G, D23F1943G, M13F1094G, B23F1495G, K13F0481G, L13F0571G, B23F1304G, C23F1808G, D23F1899G, M13F1151G, J13F0098G, J13F0099G, J13F0100G, C23F1594G, M13F1099G, B23F1439G, J13F0229G, L13F0600G, K13F0499G, K13F0500G, K13F0501G, L13F0623G, L13F0624G, L13F0625G, B23F13886, B23F1388G, B23F1390G, D23F1851G, D23F1861G, J13F0369G, M13F1075G, M13F1084G, M13F0891G, M13F0892G, A23F1229G, E23F2036G, L13F0776G, A23F1258G, B23F1315G, B23F1496G, B23F1501G, B23F1502G, B23F1506G, B23F1508G, C23F1784G, F13F0006G, J13F0006G, L13F0864G, L13F0865G, L13F0866G, L13F0668G, J13F0291G, J13F0292G, J13F0300G, M13F0950G, M13F1150G, B23F1313G, M13F0995G, M13F1044G, C23F1699G, J13F0293G, J13F0294G, J13F0295G, J13F0374G, J13F0375G, J13F0378G, J13F0379G, L13F0721G, L13F0729G, L13F0767G, L13F0768G, L13F0769G, M13F1062G, M13F1063G, E23F2068G, M13F1069G, D23F1852G, K13F0415G, K13F0426G, B23F1454G, D23F1971G, D23F1987G, K13F0411G, K13F0387G, L13F0755G, M13F1141G, M13F1142G, J13F0333G, J13F0203G, D23F1936G, M13F0991G, B23F1452G, B23F1509G, D23F1980G, D23F1982G, E23F2064G, J13F0308G, M13F1032G, M13F1034G, M13F1046G, M13F1047G, L13F0669G, A23F1266G, A23F1267G, A23F1297G, C23F1589G, B23F1357G, B23F1358G, M13F1074G, D23F1859G, D23F1933G, F23F2081G, F23F2090G, F23F2091G, F23F2092G, F23F2111G, A23F1263G, B23F1362G, C23F1690G, D23F1877G, D23F1888G, D23F1997G, D23F2014G, J13F0250G, J13F0252G, J13F0255G, K13F0462G, K13F0464G, K13F0518G, K13F0533G, L13F0691G, L13F0692G, L13F0741G, M13F0917G, M13F0927G, M13F0939G, M13F0972G, M13F0973G, M13F0993G, M13F0994G, M13F1033G, M13F1045G, M13F1048G, M13F1051G, M13F1070G, M13F1106G, M13F1107G, M13F1108G, M13F1114G, M13F1146G, D23F1867G, L13F0631G, C23F1809G, D23F1835G, L13F0607G, C23F1777G, J13F0222G, J13F0219G, C23F1800G, C23F1801G, C23F1802G, J13F0072G, B23F1346G, B23F1347G, B23F1348G, D23F1853G, M13F1073G, D23F1854G, D23F1855G, D23F1856G, B23F1516G, B23F1526G, B23F1527G, C23F1649G, C23F1779G, K13F0566G, J13F0245G, M13F0940G, D23F1958G, E23F2079G, E23F2072G, K13F0431G, K13F0432G, K13F0435G, K13F0563G, K13F0564G, K13F0565G, A23F1250G, A23F1257G, B23F1316G, J13F0313G, J13F0314G, J13F0315G, A23F1246G, A23F1249G, C23F1728G, C23F1757G, D23F1994G, M13F1040G, L13F0616G, L13F0861G, J13F0273G, A23F1282G, K13F0475G, C23F1767G, C23F1769G, C23F1770G, C23F1772G, D23F1846G, J13F0367G, J13F0368G, D23F1946G, B23F1510G, B23F1512G, B23F1513G, B23F1517G, B23F1521G, B23F1529G, B23F1530G, B23F1534G, B23F1539G, B23F1544G, C23F1725G, C23F1726G, K13F0524G, L13F0724G, L13F0730G, A23F1196G, A23F1198G, B23F1473G, L13F0809G, L13F0810G, L13F0811G, L13F0813G, L13F0814G, C23F1817G, C23F1818G, D23F1839G, M13F0949G, J13F0048G, J13F0189G, J13F0280G, J13F0281G, J13F0348G, M13F1049G, M13F1050G, C23F1723G, D23F1967G, K13F0402G, K13F0406G, K13F0407G, M13F0969G, M13F0975G, M13F0941G, M13F0948G, J13F0324G, B23F1518G, B23F1538G, M13F1052G, M13F1053G, M13F1066G, M13F1067G, K13F0506G, E23F2030G, B23F1423G, B23F1498G, C23F1750G, C23F1756G, D23F1864G, D23F1879G, D23F1960G, D23F2614G, M13F1130G, M13F1131G, D23F2001G, E23F2069G, M13F0976G, M13F1054G, M13F1136G, M13F1137G, M13F1148G, M13F1149G, A23F1286G, M13F1030G, M13F1031G, B23F1309G, J13F0323G, J13F0271G, L13F0614G, L13F0615G, L13F0663G, M13F0945G, M13F0946G, A23F1294G, B23F1528G, M13F1098G, C23F1786G, A23F1224G, A23F1240G, A23F1269G, B23F1329G, B23F1546G, B23F1547G, D23F1841G, D23F1845G, D23F1848G, D23F1947G, D23F1953G, D23F1954G, D23F1956G, D23F2013G, D23F2016G, D23F2017G, J13F0242G, J43F0481G, L13F0612G, M13F1122G, C23F1585G, L13F0637G, A23F1291G, A23F1293G, C23F1791G, L13F0713G, L13F0714G, L13F0685G, B23F1363G, C23F1789G, C23F1790G, C23F1813G, D23F1926G, D23F1966G, D23F1969G, D23F2000G, D23F2018G, D23F2019G, J13F0026G, J13F0247G, J13F0316G, J13F0317G, J13F0382G, J13F0383G, J13F0384G, K13F0390G, K13F0463G, L13F0753G, L13F0840G, L13F0841G, L13F0842G, M13F1056G, M13F1160G, B23F1372G, D23F1908G, D23F2012G, E23F2031G, L13F0594G, L13F0606G, M13F1110G, M13F1111G, M13F1112G, M13F1113G, M13F1115G, M13F1116G, M13F1117G, M13F1118G, M13F1119G, K13F0465G, K13F0482G, K13F0483G, L13F0596G, L13F0602G, D23F1998G, L13F0772G, M13F1089G, M13F1143G, M13F0928G, F23F2080G, C23F1785G, C23F1795G, D23F1913G, A23F1247G, A23F1248G, A23F1251G, A23F1252G, A23F1253G, B23F1305G, B23F1306G, B23F1314G, B23F1317G, B23F1318G, B23F1320G, B23F1376G, B23F1407G, B23F1408G, B23F1409G, B23F1410G, B23F1457G, B23F1458G, B23F1459G, B23F1460G, B23F1461G, B23F1462G, B23F1463G, B23F1464G, B23F1472G, B23F1474G, B23F1481G, B23F1482G, B23F1491G, B23F1492G, B23F1493G, B23F1494G, C23F1695G, C23F1696G, C23F1697G, C23F1698G, C23F1700G, C23F1701G, C23F1735G, C23F1749G, L13F0599G, D23F1871G, F23F2101G, J13F0343G, J13F0344G, L13F0695G, L13F0696G, L13F0760G, L13F0761G, M13F1082G, C23F1740G, C23F1747G, B23F1505G, D23F1860G, B23F1497G, C23F1774G, K13F0396G, K13F0398G, K13F0399G, K13F0400G, K13F0401G, L13F0646G, M13F0894G, B23F1387G, J13F0084G, J13F0102G, J13F0342F, J13F0342G, K13F0395G, K13F0494G, L13F0588G, K13F0517G, K13F0425G, J13F0004G, J13F0005G, J13F0077G, J13F0078G, J13F0206G, J13F0207G, D23F1865G, E23F2025G, D23F1988G, D23F1989G, D23F1993G, M13F0996G, A23F1296G, C23F1587G, C23F1604G, D23F1945G, L13F0699G, L13F0700G, L13F0701G, L13F0702G, L13F0703G, L13F0704G, L13F0705G, L13F0706G, J13F0223G, J13F0232G, J13F0233G, J13F0234G, L13F0790G, L13F0791G, M13F1128G, M13F1129G, M13F1057G, M13F1058G, D23F2009G, E23F2073G, E23F2076G, E23F2077G, E23F2078G, K13F0561G, J13F0126G, B23F1351G, L13F0860G, F23F2084G, F23F2085G, F23F2086G, F23F2089G, D23F1847G, D23F1880G, D23F1881G, D23F1928G, D23F1944G, C23F1755G, B23F1361G, B23F1364G, L13F0773G, M13F0965G, M13F0966G, M13F0967G, L13F0650G, L13F0651G, L13F0652G, J13F0211G, B23F1338G, A23F1287G, A23F1288G, B23F1322G, B23F1426G, B23F1427G, B23F1430G, B23F1431G, B23F1432G, B23F1433G, B23F1438G, B23F1441G, K13F0491G, L13F0632G, D23F2007G, L13F0694G, J13F0340G, M13F1140G, L13F0778G, L13F0779G, L13F0781G, C23F1729G, J13F0246G, L13F0666G, A23F1285G, B23F1379G, B23F1380G, B23F1381G, B23F1382G, B23F1392G, B23F1401G, B23F1429G, B23F1434G, B23F1436G, B23F1422G, C23F1753G, J13F0096G, B23F1524G, D23F2004G, D23F2021G, J13F0090G, J13F0205G, A23F1272G, L13F0722G, A23F1292G, B23F1319G, B23F1418G, B23F1420G, B23F1475G, K13F0474G, L13F0821G, L13F0822G, L13F0823G, M13F1095G, M13F1096G, M13F1100G, M13F1101G, M13F1102G, M13F1127G, M13F1097G, A23F1234G, J13F0231G, M13F1175G, J13F0239G, J13F0329G, J13F0365G, J13F0377G, B23F1368G, B23F1523G, M13F1158G, C23F1751G, C23F1692G, D23F1894G, D23F1895G, D23F1897G, A23F1182G, A23F1289G, B23F1389G, B23F1503G, D23F1842G, M13F1083G, J13F0243G, J13F0341G, B23F1469G, B23F1476G, C23F1724G, C23F1731G, J13F0248G, J13F0249G, B23F1310G, B23F1344G, K13F0490G, K13F0495G, K13F0496G, K13F0502G, M13F1156G, K13F0514G, L13F0728G, J13F0269G, A23F1237G, J13F0251G, J13F0279G, J13F0289G, J13F0290G, L13F0848G, M13F0947G, J13F0235G, B23F1399G, B23F1400G, K13F0478G, K13F0479G, K13F0480G, A23F1298G, B23F1349G, M13F1055G, M13F1065G, M13F1086G, B23F1425G, C23F1797G, C23F1598G, J13F0093G, J13F0094C, J13F0094G, K13F0428G, K13F0436G, L13F0580G, M13F0992G, M13F1076G, B23F1411G, C23F1657G, B23F1340G, B23F1359G, B23F1417G, B23F1468G, B23F1480G, B23F1543G, C23F1703G, C23F1754G, C23F1792G, C23F1794G, C23F1798G, C23F1820G, D23F1872G, B23F1542G, F23F2097G, F23F2099G, F23F2100G, J13F0350G, J13F0351G, C23F1804G, L13F0592G, L13F0638G, A23F1223G, A23F1277G, A23F1278G, A23F1279G, A23F1280G, B23F1378G, J13F0349G, L13F0708G, L13F0709G, L13F0710G, L13F0711G, L13F0723G, L13F0836G, B23F1448G, M13F0926G, J13F0228G, L13F0593G, B23F1391G, M13F0944G, K13F0545G, L13F0754G, L13F0757G, L13F0758G, L13F0759G, M13F0986G, J13F0240G, K13F0552G, K13F0553G, K13F0554G, L13F0610G, C23F1733G, C23F1775G, C23F1776G, C23F1778G, B23F1500G, L13F0667G, L13F0677G, L13F0689G, L13F0690G, J13F0204G, J13F0210G, J13F0270G, J13F0339G, B23F1446G, D23F1907G, D23F1909G, D23F1972G, D23F1973G, D23F1984G, D23F1986G, D23F1990G, E23F2023G, L13F0777G, L13F0820G, L13F0827G, L13F0850G, L13F0851G, L13F0852G, L13F0853G, L13F0854G, L13F0855G, M13F1139G, K13F0413G, K13F0520G, L13F0587G, L13F0589G, L13F0603G, L13F0604G, K13F0537G, B23F1323G, L13F0796G, L13F0797G, L13F0800G, L13F0801G, L13F0803G, J13F0241G, A23F1228G, M13F0974G, M13F0979G, F23F2114G, F23F2115G, F23F2116G, F23F2117G, F23F2118G, F23F2119G, J13F0253G, J13F0254G, J13F0262G, J13F0263G, J13F0264G, J13F0275G, J13F0276G, J13F0277G, J13F0302G, J13F0303G, J13F0304G, J13F0305G, C23F1592G, D23F1911G, D23F1912G, D23F1914G, D23F1927G, M13F1132G, M13F1133G, J13F0196G, D23F2006G, B23F1339G, B23F1341G, B23F1442G, B23F1449G, B23F1451G, B23F1535G, C23F1732G, C23F1734G, C23F1745G, C23F1746G, D23F1991G, L13F0693G, M13F1068G, A23F1230G, A23F1233G, A23F1235G, D23F1974G, J13F0038G, J13F0286G, J13F0370G, J13F0372G, J13F0373G, K13F0388G, K13F0393G, K13F0394G, K13F0416G, K13F0417G, K13F0418G, K13F0420G, K13F0421G, K13F0422G, K13F0423G, K13F0424G, L13F0654G, A23F1190G, L13F0597G, A23F1284G, K13F0515G, M13F0962G, A23F1226G, B23F1331G, B23F1334G, B23F1335G, B23F1336G, B23F1337G, B23F1350G, B23F1352G, B23F1353G, B23F1354G, B23F1367G, B23F1369G, B23F1370G, B23F1444G, B23F1445G, B23F1450G, B23F1453G, D23F1970G, E23F2026G, J13F0097G, J13F0318G, J13F0327G, J13F0357G, A23F1281G, A23F1283G, B23F1332G, A23F1225G, A23F1227G, B23F1326G, B23F1343G, B23F1345G, B23F1356G, B23F1360G, B23F1375G, B23F1383G, B23F1384G, B23F1385F, B23F1385G, B23F1455G, B23F1456G, B23F1467G, B23F1470G, B23F1471G, B23F1875G, C23F1691G, C23F1806G, C23F1810G, C23F1816G, D23F1843G, D23F1866G, D23F1885G, D23F1904G, D23F1905G, D23F1937G, D23F1948G, D23F1949G, D23F1950G, D23F1951G, D23F1952G, D23F1957G, D23F1959G, D23F2015G, J13F0168G, J13F0366G, L13F0817G, L13F0818G, L13F0819G, L13F0825G, L13F0826G, L13F0857G, M13F1157G, A23F1206G, J13F0216G, J13F0217G, K13F0438G, K13F0458G, K13F0459G, M13F1170G, D23F1964G, J13F0236G, J13F0237G, A23F1212G, B23F1540G, D23F1849G, J13F0042G, J13F0069G, J13F0085G, J13F0091G, J13F0092G, J13F0106G, J13F0107G, J13F0118G, J13F0119G, J13F0175G, J13F0177G, J13F0182G, J13F0198G, J13F0199G, J13F0208G, J13F0209G, J13F0212G, J13F0213G, J13F0214G, J13F0215G, J13F0296G, J13F0299G, J13F0301G, J13F0307G, J13F0309G, J13F0310G, J13F0311G, J13F0312G, J13F0360G, K13F0486G, K13F0487G, K13F0488G, L13F0815G, L13F0824G, A23F1202G, B23F1393G, J13F0170G, J13F0171G, J13F0265G, J13F0285G, K13F0403G, K13F0538G, L13F0774G, L13F0775G, M13F1159G, M13F1171G, F23F2082G, F23F2083G, D23F1999G, B23F1465G, D23F2011G, L13F0611G, K13F0454G, B23F1507G, D23F1836G, J13F0079G, J13F0080G, J13F0202G, J13F0244G, J13F0345G, J13F0346G, J13F0347G, J13F0354G, J13F0355G, J13F0356G, K13F0460G, K13F0508G, M13F1134G, M13F1135G, B23F1435G, B23F1478G, B23F1479G, L13F0756G, B23F1419G, D23F2008G, A23F1195G, B23F1321G, B23F1504G, J13F0001G, J13F0002G, J13F0331G, J13F0332G, J13F0361G, J13F0362G, J13F0363G, L13F0843G, L13F0844G, M13F0884G, M13F0885G, M13F0886G, M13F0887G, M13F0987G, B23F1327G, B23F1325G, J13F0278G, L13F0640G, D23F1870G, A23F1188G, L13F0712G, M13F0997G, M13F1120G, M13F1121G, M13F1123G, M13F1124G, M13F1125G, M13F1126G, J13F0328G, J13F0376G, L13F0635G, L13F0636G, J13F0201G, J13F0221G, J13F0230G, M13F1178G, L13F0572G, J13F0218G, K13F0441G, K13F0511G, L13F0780G, M13F0935G, M13F1072G, J13F0322G, C23F1780G, B23F1533G, B23F1386G, B23F1377G, J13F0191G, J13F0192G, D23F1890G, D23F1891G, C23F1742G, C23F1743G, C23F1744G, C23F1748G, C23F1758G, D23F1955G, D23F1992G, L13F0737G, L13F0750G, L13F0751G, L13F0752G, D23F2020G, J13F0238G, A23F1265G, B23F1447G, B23F1519G, A23F1264G, B23F1514G, B23F1548G, B23F1549G, B23F1553G, B23F1554G, B23F1555G, C23F1586G, C23F1593G, C23F1595G, C23F1596G, C23F1597G, C23F1605G, C23F1606G, C23F1819G, D23F1863G, D23F1869G, D23F1910G, D23F2002G, D23F2005G, E23F2024G, E23F2061G, J13F00081G, J13F0065G, J13F0066G, J13F0081G, J13F0082G, J13F0083G, J13F0101G, J13F0103G, J13F0256G, J13F0257G, J13F0259G, J13F0261G, K13F0392G, K13F0405G, K13F0419G, L13F0003G, L13F0573G, L13F0574G, L13F0575G, L13F0688G, L13F0733G, L13F0738G, L13F0740G, M13F0882G, M13F1041G, M13F1060G, M13F1061G, C23F1603G, D23F1965G, E23F2065G, B23F1311G, B23F1330G, A23F1295G, M13F1138G, M13F1144G, M13F1145G, L13F0725G, M13F0888G, J13F0095G, J13F0184G, J13F0224G, J13F0225G, J13F0226G, A23F1191G, A23F1199G, A23F1208G, A23F1209G, A23F1210G, A23F1211G, A23F1218G, A23F1219G, A23F1220G, D23F1850G, D23F1858G, D23F1862G, K13F0489G, M13F1161G, M13F0961G, B23F1301G, J13F0297G, D23F1934G, F23F2109G, B23F1333G, D23F1939G, D23F1940G, K13F0449G, D23F1962G, E23F2027G, C23F1713G, C23F1787G, D23F1827G, D23F1828G, J13F0044G, J13F0185G, J13F0193G, J13F0200G, L13F0581G, L13F0681G, M13F0898G, M13F0899G, D23F1832G, J13F0064G, J13F0186G, J13F0195G, L13F0598G, L13F0605G, L13F0648G, L13F0673G, L13F0674G, L13F0676G, M13F0936G, M13F0959G, K13F0569G, J13F0128G, J13F0141G, J13F0282G, J13F0283G, J13F0330G, K13F0410G, K13F0461G, M13F0953G, M13F0956G, M13F0963G, M13F0964G, M13F0970G, M13F0971G, J13F0029G, L13F0707G, L13F0655G, C23F1727G, C23F1805G, C23F1807G, C23F1811G, K13F0386G, J13F0023G, J13F0122G, B23F1525G, M13F0960G, M13F0977G, J13F0114G, J13F0123G, J13F0129G, J13F0194G, J13F0227G, K13F0468G, K13F0469G, K13F0476G, L13F0577G, L13F0578G, M13F0985G, M13F0895G, M13F1104G, M13F1105G, J13F0160G, A23F1241G, A23F1242G, A23F1245G, B23F1421G, D23F1941G, L13F0579G, L13F0670G, M13F0931G, B23F1397G, C23F1668G, J13F0138G, M13F0981G, C23F1717G, J13F0121G, C23F1722G, J13F0137G, L13F0837G, C23F1721G, J13F0139G, J13F0140G, J13F0054G, J13F0154G, L13F0839G, J13F0144G, A23F1201G, C23F1645G, J13F0125G, J13F0055G, J13F0150G, J13F0143G, J13F0149G, C23F1718G, J13F0052G, M13F0952G, J13F0104G, J13F0130G, J13F0131G, J13F0146G, J13F0147G, L13F0834G, L13F0835G, M13F0877G, C23F1591G, J13F0153G, J13F0157G, C23F1648G, B23F1303G, A23F1238G, A23F1244G, C23F1644G, M13F0930G, J13F0142G, K13F0412G, D23F1838G, C23F1658G, J13F0053G, J13F0117G, J13F0151G, J13F0156G, J13F0158G, J13F0159G, L13F0867G, L13F0868G, L13F0869G, L13F0871G, M13F0873G, M13F0874G, M13F0875G, M13F0879G, M13F0880G, M13F0881G, J13F0046G, J13F0113G, J13F0148G, J13F0162G, J13F0163G, L13F0831G, L13F0833G, L13F0830G, A23F1243G, J13F0152G, M13F0929G, J13F0105G, L13F0832G, C23F1643G, C23F1623G, C23F1642G, C23F1647G, L13F0872G, J13F0112G, L13F0870G, J13F0155G, J13F0145G, J13F0161G, M13F0876G, J13F0180G, J13F0181G, J13F0187G, J13F0188G, C23F1781G, J13F0178G, C23F1655G, C23F1656G, C23F1652G, C23F1803G, D23F1868G, D23F1900G, D23F2003G, M13F1176G, F23F2102G, F23F2103G, B23F1415G, J13F0041G, J13F0050G, L13F0672G, B23F1373G, E23F2034G, M13F0942G, J13F0043G, J13F0040G, J13F0057G, J13F0059G, L13F0675G, B23F1487G, M13F0914G, M13F0915G, A23F1262G, D23F1831F, D23F1831G, D23F1935G, B23F1374G, M13F1036G, L13F0828G, L13F0829G, L13F0802G, L13F0838F, L13F0838G, C23F1680G, C23F16826, C23F1682G, B23F1978G, D23F1978G, B23F1371G, C23F1584G, C23F1662G, C23F1663G, C23F1665G, C23F1681G, C23F1683G, C23F1684G, C23F1689G, C23F1708G, C23F1709G, C23F1738G, J13F0050GA, A23F1216G, J13F0051G, J13F0056G, J13F0058G, M13F0058G, M13F1093G, J13F0039G, M13F0955G, M13F1147G, D23F1932G, J13F0024G, J13F0025G, J13F0047G, L13F0629G, L13F0630G, A23F1271G, H13F0934G, B23F1488G, E23F2037G, B23F1489G, C23F1660G, J13F0027G, J13F0031G, J13F0035G, L13F0671G, B23F1365G, L13F0763G, D23F1929G, K13F0455G, M13F0902G, D23F1930G, C23F1615G, B23F1395G, B23F1398G, D23F1975G, L13F0583G, C23F1737G, C23F1788G, E23F2060G, L13F0858G, L13F0859G, M13F0896G, M13F0909G, J13F0019G, J13F0020G, K13F0457G, L13F0582G, M13F0920G, M13F1037G, B23F1485G, B23F1486G, C23F1617G, L13F0585G, B23F1405G, D23F1875G, B23F1403G, B23F1483G, B23F1490G, E23F2029G, M13F0901G, L13F8192G, J13F0033G, J13F0037G, J13F0061G, J13F0062G, M13F0900G, E23F2062G, D23F1976G, D23F1977G, C23F1616G, D23F1996G, J13F0032G, F23F2106G, L13F0584G, K13F0568G, J13F0063G, L13F0764G, H13F0567G, L13F06716, B23F1404G, M13F0897G, C23F1714G, M13F0918G, C23F1711G, C23F1712G, E23F2038G, E23F2039G, G23F1711G, F23F2095G, E23F2032G, E23F2033G, E23F2035G, F23F2107G, B13F3997G, A23F1217G, B23F1406G, J13F0021G, L13F0653G, J13F0022G, F23F2105G, C23F1685G, F23F2094G, F23F2093G, J13F0028G, M13F0921G, B23F1414G, K13F0492G, L13F0697G, L13F0726G, L13F0727G, C23F1716G, M13F1088G, M13F1087G, D23F1837G, L13F0731G, L13F0659G, L13F0660G, D23F1844G, M13F0910G, M13F0911G, M13F0912G, D23F1902G, D23F1903G, C23F1608G, D23F1979G, F23F2104G, J13F0076G, J13F0011G, L13F0845G, A23F1222G, L13F0783G, K13F0472G, K13F0473G, L13F0863G, J13F0045G, J13F0012G, M13F0958G, K13F0470G, K13F0471G, K13F0485G, K13F0484G, C23F1599G, C23F1600G, C23F1601G, C23F1602G, C23F1670G, C23F1671G, L13F0732G, J13F0013G, D23F1857G, A23F1183G, A23F1184G, L13F0856G, M13F1028G, C23F1760G, C23F1761G, C23F1762G, C23F1609G, C23F1613G, M13F1017G, D23F1915G, M13F1023G, M13F1004G, C23F1636G, D23F1919G, C23F1759G, D23F1840G, M13F1020G, M13F1021G, M13F1022G, M13F1024G, M13F1027G, C23F1625G, C23F1632G, C23F1633G, C23F1635G, E23F2058G, E23F2059G, C23F1763G, C23F1766G, C23F1627G, D23F2022G, E23F2063G, M13F0988G, M13F1025G, M13F1026G, D23F1916G, D23F1917G, D23F1918G, D23F1920G, D23F1921G, D23F1923G, D23F1924G, M13F0999G, M13F1000G, M13F1001G, M13F1002G, M13F1003G, E23F2053G, E23F2054G, C23F1628G, M13F1019G, E23F2051G, M13F1007G, M13F1011G, M13F1012G, C23F1629G, C23F1672G, C23F1673G, C23F1674G, C23F1676G, C23F1677G, C23F1679G, C23F1688G, C23F1705G, E23F2043G, C23F1764G, C23F1765G, E23F2070G, C23F1614G, E23F2052G, C23F1687G, C23F1704G, E23F2050G, C23F1612G, C23F1634G, F23F2044G, C23F1610G, C23F1626G, C23F1630G, B23F1511G, L13F0684G, L13F0686G, M13F0937G, D23F1834G, A23F1254G, A23F1255G, A23F1259G, A23F1261G, J13F0088G, J13F0115G, D23F1833G, C23F1621G, J13F0016G, M13F1064G, J13F0086G, D23F1983G, K13F0439G, E23F2040G, D23F1822G, D23F1823G, D23F1824G, D23F1825G, J13F0018G, L13F0619G, L13F0620G, L13F0622G, L13F0736G, L13F0784G, J13F0164G, J03F0164G, J13F0338G, L13F0847G, L13F0849G, M13F0890G, L13F0627G, L13F0643G, L13F0644G, L13F0645G, L13F0656G, L13F0657G, L13F0658G, J13F0306G, C23F1637G, C23F1638G, M13F1103G, C23F1782G, M13F0904G, M13F0905G, M13F0906G, M13F1077G, M13F1078G, M13F1079G, M13F1080G, M13F1081G, M13F1090G, M13F1091G, M13F1092G, D23F1874G, E23F2028G, K13F0467G, L13F0789G, E23F2071G, K13F0555G, J13F0172G, J13F0173G, J13F0176G, J13F0353G, J13F0068G, J13F0120G, J13F0127G, J13F0165G, J13F0371G, K13F0445G, K13F0446G, K13F0447G, C13F1620G, C23F1619G, C23F1620G, K13F0452G, K13F0453G, M13F1177G, E23F2041G, C23F1646G, C23F1730G, D23F1821G, D23F1961G, L13F0716G, B23F1416G, B23F1522G, B23F1556G, C23F1567G, D23F1906G, C23F1571G, K13F0442G, K13F0443G, K13F0444G, B23F1557G, B23F1552G, C23F1578G, C23F1579G, C23F1580G, C23F1562G, C23F1560G, C23F1569G, C23F1572G, C23F1573G, L13F0590G, L13F0591G, L13F0595G, B23F1558G, C23F1564G, C23F1566G, C23F1561G, C23F1568G, C23F1574G, B23F1550G, B23F1559G, C23F1563G, C23F1565G, C23F1570G, C23F1577G, C23F1581G, C23F1582G, C23F1583G, C23F1618G, F23F2110G, K13F0546G, K13F0547G, L13F0626G, C23F1575G, B23F1394G, C23F1702G, D23F1901G, J13F0015G, L13F0662G, C23F1669G, C23F1736G, A23F1194G, K13F0466G, M13F0913G, M13F0932G, M13F0916G, B23F1299G, A23F1193G, F23F2108G, L13F0762G, B23F1515G, B23F1520G, J13F0135G, J13F0136G, B23F1484G, C23F1661G, C23F1710G, C23F1715G, D23F1829G, D23F1830G, D23F1931G, L13F0586G, A23F1179G, L13F0601G, M13F0938G, J13F0110G, C23F1667G, A23F1204G, M13F0883, M13F0883G, M13F0883S, C23F1659G, A23F1203G, L13F0771G, A23F1268G, E23F2042G, M13F0923G, C23F1640G, C23F1641G, D23F1826G, K13F0427G, K13F0429G, K13F0434G, K13F0440G, B23F1443G, K13F0448G, B23F1413G, C23F1654G, K13F0430G, C23F1653G, L13F0765G, L13F0766G, A23F1192G, A23F1200G, J13F0111G, L13F0661G, M13F0957G, B23F1412G, L13F0816G, C23F16549, D23F1963G, A23F1205G, J13F0174G, C23F1639G;
2. Product Code: 4T, UDI Number (Serial Numbers): Not Applicable (F24T0254G), Not Applicable (F24T0259G), Not Applicable (F24T0264G), Not Applicable (F24T0279G), Not Applicable (A44T4691G), Not Applicable (A44T4713G), Not Applicable (C44T4931G), Not Applicable (D34T2574G), Not Applicable (H34T3263G), Not Applicable (H34T3264G), Not Applicable (H34T3266G), Not Applicable (H34T3267G), Not Applicable (H34T3268G), Not Applicable (H34T3270G), Not Applicable (H34T3272G), Not Applicable (H34T3274G), Not Applicable (M24T1660G), Not Applicable (M24T1661G), Not Applicable (M24T1662G), Not Applicable (M24T1672G), Not Applicable (M24T1710G), Not Applicable (M24T1716G), Not Applicable (M24T1718G), Not Applicable (M24T1721G), Not Applicable (M34T4314G), Not Applicable (M34T4315G), Not Applicable (M34T4316G), Not Applicable (M34T4321G), Not Applicable (M34T4360G), Not Applicable (M34T4368G), Not Applicable (M34T4482G), Not Applicable (M34T4215G), Not Applicable (M34T4216G), Not Applicable (J34T3725G), Not Applicable (J34T3752G), Not Applicable (J34T3755G), Not Applicable (M34T4340G), Not Applicable (G24T0384G), Not Applicable (K24T1151G), Not Applicable (L24T1274G), Not Applicable (L24T1284G), Not Applicable (L24T1377G), Not Applicable (L24T1401G), Not Applicable (L34T4001G), Not Applicable (A44T4739G), Not Applicable (G24T0475G), Not Applicable (A44T4573G), Not Applicable (C34T2261G), Not Applicable (B44T4778G), Not Applicable (F24T0081G), Not Applicable (F24T0292G), Not Applicable (F24T0294G), Not Applicable (F24T0296G), Not Applicable (F24T0297G), Not Applicable (F24T0318G), Not Applicable (F34T2928G), Not Applicable (K24T1003G), Not Applicable (L34T4174G), Not Applicable (F34T3021G), Not Applicable (C34T2336G), Not Applicable (F24T0242G), Not Applicable (L34T4041G), Not Applicable (M34T4479G), Not Applicable (F24T0084G), Not Applicable (G24T0564G), Not Applicable (M24T1736G), Not Applicable (M34T4449G), Not Applicable (M34T4450G), Not Applicable (B34T1952G), Not Applicable (A34T1762G), Not Applicable (F24T0124G), Not Applicable (F24T0241G), Not Applicable (F34T2959G), Not Applicable (J24T0609G), Not Applicable (J24T0826G), Not Applicable (K24T1143G), Not Applicable (K34T3974G), Not Applicable (M34T4381G), Not Applicable (A34T1853G), Not Applicable (B44T4800G), Not Applicable (D44T5014G), Not Applicable (K34T3833G), Not Applicable (K34T3834G), Not Applicable (K34T3837G), Not Applicable (K34T3843G), Not Applicable (K34T3858G), Not Applicable (K34T3859G), Not Applicable (K34T3966G), Not Applicable (M24T1560G), Not Applicable (M24T1665G), Not Applicable (M24T1673G), Not Applicable (M24T1692G), Not Applicable (M34T4413G), Not Applicable (M34T4415G), Not Applicable (M34T4442G), Not Applicable (M34T4443G), Not Applicable (M34T4444G), Not Applicable (M34T4445G), Not Applicable (L34T4036G), Not Applicable (C34T2256G), Not Applicable (K24T1165G), Not Applicable (L24T1309G), Not Applicable (M24T1528G), Not Applicable (C34T2324G), Not Applicable (C34T2326G), Not Applicable (C34T2328G), Not Applicable (C34T2341G), Not Applicable (A34T1825G), Not Applicable (A34T1826G), Not Applicable (A34T1827G), Not Applicable (A34T1829G), Not Applicable (A34T1830G), Not Applicable (A34T1831G), Not Applicable (A34T1832G), Not Applicable (A34T1848G), Not Applicable (A34T1896G), Not Applicable (A34T1899G), Not Applicable (A34T1900G), Not Applicable (A34T1903G), Not Applicable (B34T2033G), Not Applicable (C34T2298G), Not Applicable (C44T4984G), Not Applicable (G24T0512G), Not Applicable (J34T3789G), Not Applicable (J34T3791G), Not Applicable (C34T2275G), Not Applicable (C34T2361G), Not Applicable (C34T2365G), Not Applicable (C34T2378G), Not Applicable (F24T0247G), Not Applicable (J24T0867G), Not Applicable (J24T0873G), Not Applicable (J34T3597G), Not Applicable (J34T3598G), Not Applicable (J34T3604G), Not Applicable (J34T3605G), Not Applicable (J34T3606G), Not Applicable (J34T3628G), Not Applicable (J34T3631G), Not Applicable (J34T3669G), Not Applicable (K34T3871G), Not Applicable (F34T2927G), Not Applicable (M24T1608G), Not Applicable (M24T1647G), Not Applicable (D34T2530G), Not Applicable (D34T2534G), Not Applicable (D34T2538G), Not Applicable (J34T3517G), Not Applicable (J24T0831G), Not Applicable (C34T2265G), Not Applicable (C34T2267G), Not Applicable (F24T0269G), Not Applicable (J34T3816G), Not Applicable (B34T1947G), Not Applicable (B34T1951G), Not Applicable (J24T0846G), Not Applicable (J24T0862G), Not Applicable (J24T0876G), Not Applicable (H34T3437G), Not Applicable (A44T4581G), Not Applicable (B34T1984G), Not Applicable (B34T1989G), Not Applicable (B44T4905G), Not Applicable (B44T4913G), Not Applicable (C44T4930G), Not Applicable (F24T0054G), Not Applicable (F24T0239G), Not Applicable (G24T0414G), Not Applicable (G24T0506G), Not Applicable (G24T0520G), Not Applicable (G24T0521G), Not Applicable (G24T0524G), Not Applicable (H34T3284G), Not Applicable (J24T0738G), Not Applicable (J24T0833G), Not Applicable (J24T0836G), Not Applicable (M24T1668G), Not Applicable (M24T1671G), Not Applicable (D34T2454G), Not Applicable (D34T2535G), Not Applicable (D34T2545G), Not Applicable (F24T0053G), Not Applicable (D44T5027G), Not Applicable (K24T1117G), Not Applicable (K24T1118G), Not Applicable (K24T1190G), Not Applicable (K24T1191G), Not Applicable (K24T1196G), Not Applicable (K24T1197G), Not Applicable (K24T1200G), Not Applicable (B44T4880G), Not Applicable (B44T4882G), Not Applicable (J24T0642G), Not Applicable (J24T0813G), Not Applicable (M24T1569G), Not Applicable (B34T2021G), Not Applicable (M24T1540G), Not Applicable (J34T3601G), Not Applicable (M24T1603G), Not Applicable (M24T1749G), Not Applicable (M34T4339G), Not Applicable (M34T4493G), Not Applicable (M34T4494G), Not Applicable (K34T3968G), Not Applicable (D34T2629G), Not Applicable (B34T2006G), Not Applicable (M24T1609G), Not Applicable (M24T1663G), Not Applicable (A44T4676G), Not Applicable (B34T1991G), Not Applicable (B34T1998G), Not Applicable (B34T1999G), Not Applicable (B34T2105G), Not Applicable (B34T2107G), Not Applicable (B34T2108G), Not Applicable (B34T2109G), Not Applicable (D34T2440G), Not Applicable (H34T3419G), Not Applicable (J24T0704G), Not Applicable (J24T0790G), Not Applicable (J24T0792G), Not Applicable (J24T0799G), Not Applicable (J24T0854G), Not Applicable (J34T3807G), Not Applicable (K34T3839G), Not Applicable (K34T3840G), Not Applicable (K34T3841G), Not Applicable (K34T3844G), Not Applicable (L24T1304G), Not Applicable (F34T3027G), Not Applicable (K34T3920G), Not Applicable (L34T4102G), Not Applicable (L34T4103G), Not Applicable (L34T4104G), Not Applicable (L34T4105G), Not Applicable (L34T4106G), Not Applicable (K24T1214G), Not Applicable (K24T1215G), Not Applicable (K24T1225G), Not Applicable (L24T1362G), Not Applicable (H34T3477G), Not Applicable (K24T1266G), Not Applicable (L34T4005G), Not Applicable (M24T1455G), Not Applicable (F34T2841G), Not Applicable (M34T4432G), Not Applicable (F24T0308G), Not Applicable (F24T0309G), Not Applicable (G24T0417G), Not Applicable (G24T0420G), Not Applicable (J34T3521G), Not Applicable (G24T0462G), Not Applicable (G24T0470G), Not Applicable (B34T1979G), Not Applicable (B34T1986G), Not Applicable (B34T1987G), Not Applicable (L24T1294G), Not Applicable (L24T1303G), Not Applicable (L24T1315G), Not Applicable (H34T3361G), Not Applicable (K24T1193G), Not Applicable (K24T1198G), Not Applicable (B34T2043G), Not Applicable (B34T2047G), Not Applicable (B34T2055G), Not Applicable (B34T2202G), Not Applicable (L24T1391G), Not Applicable (F24T0278G), Not Applicable (F24T0149G), Not Applicable (F24T0211G), Not Applicable (G24T0324G), Not Applicable (H34T3435G), Not Applicable (M24T1512G), Not Applicable (J24T0900G), Not Applicable (L34T4109G), Not Applicable (M34T4446G), Not Applicable (M34T4481G), Not Applicable (M34T4539G), Not Applicable (H34T3441G), Not Applicable (K34T3978G), Not Applicable (B44T4781G), Not Applicable (F34T2978G), Not Applicable (B44T4745G), Not Applicable (B44T4751G), Not Applicable (D34T2470G), Not Applicable (K24T1184G), Not Applicable (A34T1781G), Not Applicable (A34T1784G), Not Applicable (A34T1785G), Not Applicable (A34T1789G), Not Applicable (A34T1791G), Not Applicable (C34T2407G), Not Applicable (J24T0823G), Not Applicable (D34T2509G), Not Applicable (K34T3886G), Not Applicable (K34T3928G), Not Applicable (K34T3960G), Not Applicable (F24T0276G), Not Applicable (B34T2163G), Not Applicable (B44T4783G), Not Applicable (C34T2235G), Not Applicable (F24T0256G), Not Applicable (J24T0671G), Not Applicable (J24T0700G), Not Applicable (J24T0879G), Not Applicable (J24T0884G), Not Applicable (J34T3686G), Not Applicable (J34T3688G), Not Applicable (J34T3689G), Not Applicable (J34T3693G), Not Applicable (K34T3900G), Not Applicable (K34T3902G), Not Applicable (A44T4704G), Not Applicable (A44T4707G), Not Applicable (K24T1033G), Not Applicable (K24T1039G), Not Applicable (A44T4637G), Not Applicable (A44T4639G), Not Applicable (B44T4747G), Not Applicable (F24T0203G), Not Applicable (G34T3207G), Not Applicable (B34T2019G), Not Applicable (D34T2473G), Not Applicable (D34T2475G), Not Applicable (D34T2491G), Not Applicable (D34T2493G), Not Applicable (D34T2495G), Not Applicable (D34T2518G), Not Applicable (D34T2527G), Not Applicable (D34T2541G), Not Applicable (D34T2546G), Not Applicable (D34T2570G), Not Applicable (D34T2602G), Not Applicable (D34T2604G), Not Applicable (D34T2605G), Not Applicable (D34T2606G), Not Applicable (M34T4318G), Not Applicable (M24T1600G), Not Applicable (M24T1700G), Not Applicable (D34T2482G), Not Applicable (D34T2516G), Not Applicable (D34T2573G), Not Applicable (F34T2854G), Not Applicable (F34T2858G), Not Applicable (F34T2863G), Not Applicable (F34T2864G), Not Applicable (F34T2865G), Not Applicable (F34T2867G), Not Applicable (G24T0448G), Not Applicable (K24T1231G), Not Applicable (K24T1239G), Not Applicable (M24T1677G), Not Applicable (M24T1687G), Not Applicable (M24T1706G), Not Applicable (M24T1712G), Not Applicable (C34T2351G), Not Applicable (C34T2372G), Not Applicable (J34T3653G), Not Applicable (H34T3486G), Not Applicable (J34T3681G), Not Applicable (E34T2726G), Not Applicable (E34T2727G), Not Applicable (E34T2729G), Not Applicable (E34T2732G), Not Applicable (E34T2737G), Not Applicable (F24T0288G), Not Applicable (F24T0289G), Not Applicable (F24T0290G), Not Applicable (F24T0291G), Not Applicable (M24T1504G), Not Applicable (M24T1511G), Not Applicable (M24T1555G), Not Applicable (M24T1561G), Not Applicable (M34T4390G), Not Applicable (B44T4909G), Not Applicable (G24T0543G), Not Applicable (G24T0545G), Not Applicable (H24T0594G), Not Applicable (H34T3265G), Not Applicable (A34T1803G), Not Applicable (F24T0271G), Not Applicable (F24T0284G), Not Applicable (F24T0285G), Not Applicable (F24T0298G), Not Applicable (F24T0299G), Not Applicable (F24T0300G), Not Applicable (F24T0301G), Not Applicable (F24T0303G), Not Applicable (F24T0316G), Not Applicable (G24T0484G), Not Applicable (G24T0488G), Not Applicable (G24T0490G), Not Applicable (G24T0499G), Not Applicable (G24T0542G), Not Applicable (J24T0692G), Not Applicable (J24T0709G), Not Applicable (J24T0710G), Not Applicable (J24T0711G), Not Applicable (J24T0712G), Not Applicable (J24T0716G), Not Applicable (J24T0717G), Not Applicable (J24T0718G), Not Applicable (J24T0722G), Not Applicable (J24T0830G), Not Applicable (K24T1004G), Not Applicable (K24T1014G), Not Applicable (K24T1080G), Not Applicable (F24T0236G), Not Applicable (G24T0411G), Not Applicable (B44T4904G), Not Applicable (F24T0130G), Not Applicable (F24T0272G), Not Applicable (F24T0273G), Not Applicable (G24T0529G), Not Applicable (G24T0538G), Not Applicable (J24T0630G), Not Applicable (J24T0651G), Not Applicable (J24T0691G), Not Applicable (J24T0721G), Not Applicable (J24T0723G), Not Applicable (K24T1007G), Not Applicable (K24T1116G), Not Applicable (K24T1119G), Not Applicable (K24T1120G), Not Applicable (K24T1122G), Not Applicable (K24T1123G), Not Applicable (L24TE1JG), Not Applicable (H34T3277G), Not Applicable (B34T2029G), Not Applicable (B34T2203G), Not Applicable (C34T2268G), Not Applicable (C34T2269G), Not Applicable (C34T2281G), Not Applicable (C34T2397G), Not Applicable (F24T0074G), Not Applicable (F24T0077G), Not Applicable (F24T0078G), Not Applicable (F24T0079G), Not Applicable (G24T0468G), Not Applicable (M24T1442G), Not Applicable (M24T1513G), Not Applicable (M34T4452G), Not Applicable (M34T4453G), Not Applicable (A34T1870G), Not Applicable (B34T1942G), Not Applicable (B34T2111G), Not Applicable (D34T2542G), Not Applicable (E34T2722G), Not Applicable (E34T2746G), Not Applicable (E34T2765G), Not Applicable (F24T0002G), Not Applicable (F24T0003G), Not Applicable (F24T0004G), Not Applicable (F24T0006G), Not Applicable (F24T0007G), Not Applicable (F24T0008G), Not Applicable (F24T0009G), Not Applicable (F24T0010G), Not Applicable (F24T0011G), Not Applicable (F24T0012G), Not Applicable (F24T0014G), Not Applicable (F24T0015G), Not Applicable (F24T0017G), Not Applicable (F24T0026G), Not Applicable (F24T0028G), Not Applicable (F24T0033G), Not Applicable (F24T0035G), Not Applicable (F24T0036G), Not Applicable (F24T0042G), Not Applicable (F24T0043G), Not Applicable (F24T0068G), Not Applicable (F24T0069G), Not Applicable (F24T0133G), Not Applicable (F24T0134G), Not Applicable (F34T2818G), Not Applicable (F34T2862G), Not Applicable (F34T3003G), Not Applicable (G24T0343G), Not Applicable (G24T0367G), Not Applicable (G24T0502G), Not Applicable (G24T0504G), Not Applicable (H34T3373G), Not Applicable (J24T0678G), Not Applicable (K24T1160G), Not Applicable (K24T1188G), Not Applicable (K34T3838G), Not Applicable (K34T3865G), Not Applicable (M24T1481G), Not Applicable (M24T1482G), Not Applicable (M24T1573G), Not Applicable (M24T1682G), Not Applicable (M24T1686G), Not Applicable (M34T4533G), Not Applicable (A44T4598G), Not Applicable (A44T4627G), Not Applicable (A44T4714G), Not Applicable (A44T4716G), Not Applicable (B44T4758G), Not Applicable (B44T4769G), Not Applicable (B44T4809G), Not Applicable (H34T3329G), Not Applicable (F34T2951G), Not Applicable (F24T0020G), Not Applicable (F24T0233G), Not Applicable (J24T0617G), Not Applicable (C34T2334G), Not Applicable (G24T0395G), Not Applicable (J24T0637G), Not Applicable (J24T0638G), Not Applicable (J24T0639G), Not Applicable (J34T3662G), Not Applicable (J34T3763G), Not Applicable (M34T4406G), Not Applicable (M34T4407G), Not Applicable (M34T4408G), Not Applicable (M34T4428G), Not Applicable (M34T4429G), Not Applicable (A44T4732G), Not Applicable (C34T2363G), Not Applicable (E34T2731G), Not Applicable (L34T2731G), Not Applicable (L34T4086G), Not Applicable (L34T4125G), Not Applicable (B34T2191G), Not Applicable (L34T4151G), Not Applicable (M34T4531G), Not Applicable (A34T1874G), Not Applicable (A34T1883G), Not Applicable (A34T1884G), Not Applicable (B34T1944G), Not Applicable (F34T2904G), Not Applicable (G24T0349G), Not Applicable (H34T3305G), Not Applicable (J24T0753G), Not Applicable (K24T1015G), Not Applicable (K24T1016G), Not Applicable (K24T1008G), Not Applicable (K24T1034G), Not Applicable (K24T1035G), Not Applicable (E34T2739G), Not Applicable (E34T2740G), Not Applicable (E34T2742G), Not Applicable (E34T2744G), Not Applicable (C34T2290G), Not Applicable (J24T0941G), Not Applicable (L34T4026G), Not Applicable (L34T4079G), Not Applicable (M24T1503G), Not Applicable (G24T0463G), Not Applicable (C34T2299G), Not Applicable (C34T2301G), Not Applicable (J24T0765G), Not Applicable (J24T0767G), Not Applicable (J24T0832G), Not Applicable (J24T0834G), Not Applicable (J24T0863G), Not Applicable (J24T0887G), Not Applicable (K34T3927G), Not Applicable (A44T4547G), Not Applicable (A44T4564G), Not Applicable (A44T4565G), Not Applicable (B34T2048G), Not Applicable (C34T2228G), Not Applicable (C34T2283G), Not Applicable (C34T2325G), Not Applicable (E34T2724G), Not Applicable (F24T0249G), Not Applicable (G34T3077G), Not Applicable (M34T4334G), Not Applicable (M34T4356G), Not Applicable (M34T4434G), Not Applicable (G34T3099G), Not Applicable (G34T3102G), Not Applicable (G34T3109G), Not Applicable (E34T2653G), Not Applicable (C44T4938G), Not Applicable (G24T0509G), Not Applicable (F24T0126G), Not Applicable (H34T3369G), Not Applicable (H34T3370G), Not Applicable (H34T3372G), Not Applicable (H34T3374G), Not Applicable (H34T3376G), Not Applicable (H34T3377G), Not Applicable (H34T3380G), Not Applicable (J34T3800G), Not Applicable (L34T4054G), Not Applicable (H34T3301G), Not Applicable (H34T3302G), Not Applicable (H34T3346G), Not Applicable (J24T0707G), Not Applicable (J24T0759G), Not Applicable (J24T0840G), Not Applicable (J24T0880G), Not Applicable (J24T0905G), Not Applicable (J24T0929G), Not Applicable (J24T0931G), Not Applicable (F34T3012G), Not Applicable (F34T3019G), Not Applicable (M34T4221G), Not Applicable (A34T1765G), Not Applicable (A34T1808G), Not Applicable (A34T1809G), Not Applicable (A34T1815G), Not Applicable (A34T1816G), Not Applicable (A34T1822G), Not Applicable (F34T3004G), Not Applicable (F34T3007G), Not Applicable (M24T1739G), Not Applicable (G24T0368G), Not Applicable (M24T1745G), Not Applicable (M24T1750G), Not Applicable (J24T0610G), Not Applicable (J34T3754G), Not Applicable (J34T3806G), Not Applicable (K34T3823G), Not Applicable (K34T3870G), Not Applicable (K34T3899G), Not Applicable (B44T4801G), Not Applicable (C34T2354G), Not Applicable (C34T2355G), Not Applicable (F24T0235G), Not Applicable (G24T0338G), Not Applicable (J24T0679G), Not Applicable (K24T1220G), Not Applicable (M24T1483G), Not Applicable (M34T4504G), Not Applicable (M34T4505G), Not Applicable (F24T0051G), Not Applicable (H34T3395G), Not Applicable (H34T3396G), Not Applicable (H34T3420G), Not Applicable (J34T3600G), Not Applicable (F24T0244G), Not Applicable (F24T0245G), Not Applicable (F24T0253G), Not Applicable (F34T2835G), Not Applicable (F24T0125G), Not Applicable (F24T0128G), Not Applicable (K24T1148G), Not Applicable (D34T2522G), Not Applicable (D34T2559G), Not Applicable (D34T2576G), Not Applicable (B34T1960G), Not Applicable (C34T2241G), Not Applicable (G34T3062G), Not Applicable (G34T3202G), Not Applicable (G34T3144G), Not Applicable (F24T0270G), Not Applicable (A44T4687G), Not Applicable (K24T0999G), Not Applicable (K24T1042G), Not Applicable (K24T1043G), Not Applicable (M34T4378G), Not Applicable (M34T4486G), Not Applicable (B44T4899G), Not Applicable (D34T2438G), Not Applicable (A34T1879G), Not Applicable (G34T3211G), Not Applicable (L34T4021G), Not Applicable (B44T4744G), Not Applicable (B44T4750G), Not Applicable (B44T4752G), Not Applicable (B44T4754G), Not Applicable (B44T4799G), Not Applicable (B44T4804G), Not Applicable (F24T0047G), Not Applicable (F24T0048G), Not Applicable (L34T4082G), Not Applicable (L34T4084G), Not Applicable (L34T4085G), Not Applicable (L34T4087G), Not Applicable (L34T4088G), Not Applicable (L34T4090G), Not Applicable (B44T4782G), Not Applicable (E34T2668G), Not Applicable (E34T2669G), Not Applicable (L34T4081G), Not Applicable (L34T4089G), Not Applicable (B34T2005G), Not Applicable (K24T1183G), Not Applicable (B34T2065G), Not Applicable (B34T2066G), Not Applicable (B34T2067G), Not Applicable (B34T2069G), Not Applicable (B34T2070G), Not Applicable (B34T2071G), Not Applicable (B34T2072G), Not Applicable (B34T2154G), Not Applicable (B34T2155G), Not Applicable (D34T2524G), Not Applicable (E34T2747G), Not Applicable (E34T2752G), Not Applicable (F34T2888G), Not Applicable (F34T2990G), Not Applicable (M24T1584G), Not Applicable (M24T1585G), Not Applicable (J34T3674G), Not Applicable (B34T1930G), Not Applicable (B34T1938G), Not Applicable (B34T2135G), Not Applicable (C34T2231G), Not Applicable (C34T2237G), Not Applicable (F34T2824G), Not Applicable (K24T1101G), Not Applicable (K24T1126G), Not Applicable (K24T1127G), Not Applicable (K24T1186G), Not Applicable (K24T1224G), Not Applicable (K24T1236G), Not Applicable (K24T1238G), Not Applicable (L24T1327G), Not Applicable (L34T4025G), Not Applicable (M34T4435G), Not Applicable (B34T1977G), Not Applicable (D34T2480G), Not Applicable (D34T2555G), Not Applicable (D34T2560G), Not Applicable (F24T0001G), Not Applicable (F24T0002), Not Applicable (F24T0005G), Not Applicable (F24T0013G), Not Applicable (F24T0016G), Not Applicable (F24T0025G), Not Applicable (F24T0027G), Not Applicable (F24T0028), Not Applicable (F24T0029G), Not Applicable (F24T0030G), Not Applicable (F24T0031G), Not Applicable (F24T0032), Not Applicable (F24T0032G), Not Applicable (F24T0034G), Not Applicable (F24T0037G), Not Applicable (F24T0038G), Not Applicable (F24T0039G), Not Applicable (F24T0040G), Not Applicable (F24T0059G), Not Applicable (F24T0067G), Not Applicable (F24T0072G), Not Applicable (F34T2834G), Not Applicable (F34T2870G), Not Applicable (F34T2872G), Not Applicable (F34T3035G), Not Applicable (H34T3285G), Not Applicable (H34T3292G), Not Applicable (H34T3334G), Not Applicable (L24T1322G), Not Applicable (M34T4394G), Not Applicable (C44T4925G), Not Applicable (C44T4947G), Not Applicable (F24T0221G), Not Applicable (F24T0222G), Not Applicable (J34T3775G), Not Applicable (L34T4057G), Not Applicable (L34T4058G), Not Applicable (L34T4059G), Not Applicable (L34T4060G), Not Applicable (L34T4061G), Not Applicable (L34T4063G), Not Applicable (L34T4065G), Not Applicable (L34T4066G), Not Applicable (F24T0260G), Not Applicable (F24T0261G), Not Applicable (F24T0262G), Not Applicable (F24T0263G), Not Applicable (F24T0265G), Not Applicable (F24T0266G), Not Applicable (F24T0267G), Not Applicable (G34T3119G), Not Applicable (M34T4471G), Not Applicable (C34T2396G), Not Applicable (L24T1310G), Not Applicable (J34T3770G), Not Applicable (J24T0859G), Not Applicable (M34T4470G), Not Applicable (E34T2748G), Not Applicable (E34T2750G), Not Applicable (E34T2757G), Not Applicable (E34T2763G), Not Applicable (E34T2764G), Not Applicable (E34T2773G), Not Applicable (E34T2781G), Not Applicable (E34T2783G), Not Applicable (F34T2840G), Not Applicable (F34T2934G), Not Applicable (F34T2936G), Not Applicable (F34T2937G), Not Applicable (F34T2939G), Not Applicable (F34T2952G), Not Applicable (F34T2953G), Not Applicable (F34T2954G), Not Applicable (F34T2955G), Not Applicable (F34T2956G), Not Applicable (G34T3138G), Not Applicable (G34T3145G), Not Applicable (M34T4469G), Not Applicable (A34T1806G), Not Applicable (A34T1807G), Not Applicable (A34T1810G), Not Applicable (C44T4924G), Not Applicable (D34T2436G), Not Applicable (L24T1351G), Not Applicable (L24T1358G), Not Applicable (L24T1365G), Not Applicable (M34T4483G), Not Applicable (J34T3785G), Not Applicable (J24T0892G), Not Applicable (J34T3772G), Not Applicable (J34T3776G), Not Applicable (J34T3777G), Not Applicable (F34T3040G), Not Applicable (G34T3127G), Not Applicable (G34T3128G), Not Applicable (G34T3132G), Not Applicable (K24T1044G), Not Applicable (M24T1606G), Not Applicable (M24T1611G), Not Applicable (M34T4235G), Not Applicable (M34T4409G), Not Applicable (M34T4410G), Not Applicable (M34T4411G), Not Applicable (M34T4456G), Not Applicable (B44T4841G), Not Applicable (E34T2778G), Not Applicable (E34T2779G), Not Applicable (F24T0189G), Not Applicable (F24T0196G), Not Applicable (F34T2873G), Not Applicable (F34T2895G), Not Applicable (J24T0695G), Not Applicable (J24T0881G), Not Applicable (G24T0536G), Not Applicable (K34T3877G), Not Applicable (K34T3882G), Not Applicable (L34T4111G), Not Applicable (M24T1533G), Not Applicable (M34T4322G), Not Applicable (M34T4524G), Not Applicable (F24T0071G), Not Applicable (G34T3139G), Not Applicable (E34T2685G), Not Applicable (E34T2688G), Not Applicable (E34T2689G), Not Applicable (E34T2691G), Not Applicable (E34T2693G), Not Applicable (E34T2715G), Not Applicable (F34T2086G), Not Applicable (F34T2976G), Not Applicable (F34T2986G), Not Applicable (F34T2987G), Not Applicable (F34T2992G), Not Applicable (F34T2993G), Not Applicable (F34T2994G), Not Applicable (F34T2995G), Not Applicable (F34T3022G), Not Applicable (M24T1693G), Not Applicable (A34T1768G), Not Applicable (C34T2352G), Not Applicable (C44T4983G), Not Applicable (D34T2547G), Not Applicable (F24T0287G), Not Applicable (J24T0860G), Not Applicable (K24T1246G), Not Applicable (L24T1281G), Not Applicable (M24T1456G), Not Applicable (M24T1471G), Not Applicable (M24T1506G), Not Applicable (M24T1525G), Not Applicable (M24T1526G), Not Applicable (M24T1618G), Not Applicable (M24T1636G), Not Applicable (M24T1674G), Not Applicable (M24T1675G), Not Applicable (F24T0131G), Not Applicable (F24T0135G), Not Applicable (F24T0218G), Not Applicable (F24T0219G), Not Applicable (F24T0220G), Not Applicable (J34T3796G), Not Applicable (B34T2097G), Not Applicable (B44T4851G), Not Applicable (G24T0391G), Not Applicable (G24T0392G), Not Applicable (D34T2628G), Not Applicable (B34T2201G), Not Applicable (B34T2056G), Not Applicable (B34T2204G), Not Applicable (C34T2216G), Not Applicable (F34T3032G), Not Applicable (J34T3520G), Not Applicable (J34T3546G), Not Applicable (J34T3558G), Not Applicable (J34T3559G), Not Applicable (J34T3560G), Not Applicable (J34T3561G), Not Applicable (M34T4420G), Not Applicable (K24T1210G), Not Applicable (K24T1211G), Not Applicable (K24T1222G), Not Applicable (C34T2409G), Not Applicable (J34T3585G), Not Applicable (J34T3595G), Not Applicable (J34T3783G), Not Applicable (M34T4260G), Not Applicable (M34T4269G), Not Applicable (J34T3764G), Not Applicable (J34T3765G), Not Applicable (F34T2857G), Not Applicable (K24T1029G), Not Applicable (K24T1273G), Not Applicable (B34T2004G), Not Applicable (J34T3575G), Not Applicable (J34T3787G), Not Applicable (J34T3811G), Not Applicable (J34T3813G), Not Applicable (H24T0598G), Not Applicable (H24T0599G), Not Applicable (L34T4100G), Not Applicable (M24T1423G), Not Applicable (M24T1451G), Not Applicable (M24T1458G), Not Applicable (B44T4807G), Not Applicable (C34T2358G), Not Applicable (F24T0304G), Not Applicable (F24T0305G), Not Applicable (F24T0306G), Not Applicable (F24T0307G), Not Applicable (D34T2434G), Not Applicable (D34T2439G), Not Applicable (D34T2565G), Not Applicable (K24T1099G), Not Applicable (K34T3937G), Not Applicable (M34T4461G), Not Applicable (M34T4535G), Not Applicable (M34T4537G), Not Applicable (B34T2058G), Not Applicable (B34T2060G), Not Applicable (B34T2062G), Not Applicable (J24T0801G), Not Applicable (L34T4013G), Not Applicable (L34T4046G), Not Applicable (A44T4685G), Not Applicable (C34T2305G), Not Applicable (D34T2455G), Not Applicable (F24T0019G), Not Applicable (M24T1598G), Not Applicable (M24T1644G), Not Applicable (M24T1645G), Not Applicable (G24T0528G), Not Applicable (G24T0531G), Not Applicable (J24T0714G), Not Applicable (J24T0715G), Not Applicable (J24T0741G), Not Applicable (J24T0742G), Not Applicable (J34T3803G), Not Applicable (H34T3423G), Not Applicable (M24T1713G), Not Applicable (F34T2980G), Not Applicable (F34T3026G), Not Applicable (F34T3042G), Not Applicable (F24T0055G), Not Applicable (F34T3025G), Not Applicable (F34T3047G), Not Applicable (H34T3442G), Not Applicable (B34T1980G), Not Applicable (J24T0732G), Not Applicable (A34T1925G), Not Applicable (K24T1229G), Not Applicable (L24T1399G), Not Applicable (B34T2089G), Not Applicable (M24T1439G), Not Applicable (M24T1446G), Not Applicable (M24T1448G), Not Applicable (M24T1449G), Not Applicable (M24T1450G), Not Applicable (M24T1452G), Not Applicable (M24T1453G), Not Applicable (M24T1454G), Not Applicable (M24T1464G), Not Applicable (F34T3009G), Not Applicable (F34T3011G), Not Applicable (F34T3013G), Not Applicable (F34T3015G), Not Applicable (F34T3016G), Not Applicable (J34T3690G), Not Applicable (J34T3691G), Not Applicable (J34T3692G), Not Applicable (L24T1366G), Not Applicable (L24T1367G), Not Applicable (L24T1368G), Not Applicable (L24T1369G), Not Applicable (C34T2383G), Not Applicable (C34T2386G), Not Applicable (C34T2387G), Not Applicable (C34T2388G), Not Applicable (C34T2393G).Not Applicable (C34T2394G), Not Applicable (F24T0199G), Not Applicable (F24T0200G), Not Applicable (F24T0202G), Not Applicable (K24T1149G), Not Applicable (K24T1253G), Not Applicable (B34T1963G), Not Applicable (F24T0238G), Not Applicable (J24T0800G), Not Applicable (C34T2315G), Not Applicable (C34T2318G), Not Applicable (J24T0693G), Not Applicable (J24T0694G), Not Applicable (J24T0699G), Not Applicable (J24T0802G), Not Applicable (J24T0803G), Not Applicable (J24T0845G), Not Applicable (B34T1982G), Not Applicable (L24T1297G), Not Applicable (A44T4595G), Not Applicable (B34T2007G), Not Applicable (B34T2010G), Not Applicable (B34T2011G), Not Applicable (B34T2012G), Not Applicable (B34T2013G), Not Applicable (B34T2015G), Not Applicable (B34T2017G), Not Applicable (M24T1411G), Not Applicable (M24T1414G), Not Applicable (M24T1415G), Not Applicable (M24T1416G), Not Applicable (M24T1420G), Not Applicable (M24T1431G), Not Applicable (M34T4491G), Not Applicable (B34T1966G), Not Applicable (B34T1971G), Not Applicable (B34T2150G), Not Applicable (B44T4840G), Not Applicable (B44T4849G), Not Applicable (B44T5023G), Not Applicable (C34T2302G), Not Applicable (D44T5023G), Not Applicable (F24T0215G), Not Applicable (B34T2199G), Not Applicable (B34T2200G), Not Applicable (A44T4710G), Not Applicable (G34T3180G), Not Applicable (G34T3181G), Not Applicable (G34T3182G), Not Applicable (A44T4638G), Not Applicable (A44T4640G), Not Applicable (B44T4773G), Not Applicable (D34T2579G), Not Applicable (D34T2582G), Not Applicable (D34T2635G), Not Applicable (D44T5037G), Not Applicable (F34T3024G), Not Applicable (F34T3029G), Not Applicable (F34T3033G), Not Applicable (F34T3044G), Not Applicable (G24T0340G), Not Applicable (G24T0341G), Not Applicable (G24T0342G), Not Applicable (G24T0344G), Not Applicable (G24T0345G), Not Applicable (G24T0346G), Not Applicable (G24T0350G), Not Applicable (G24T0351G), Not Applicable (G24T0352G), Not Applicable (G24T0353G), Not Applicable (G24T0354G), Not Applicable (G24T0355G), Not Applicable (G24T0357G), Not Applicable (G24T0358G), Not Applicable (G24T0359G), Not Applicable (G24T0381G), Not Applicable (G24T0382G), Not Applicable (G24T0454G), Not Applicable (G24T0461G), Not Applicable (G24T0466G), Not Applicable (G24T0467G), Not Applicable (G24T0469G), Not Applicable (G24T0471G), Not Applicable (G24T0472G), Not Applicable (G24T0473G), Not Applicable (G24T0496G), Not Applicable (G24T0497G), Not Applicable (G24T0498G), Not Applicable (G34T3068G), Not Applicable (G34T3071G), Not Applicable (G34T3072G), Not Applicable (G34T3073G), Not Applicable (G34T3074G), Not Applicable (G34T3075G), Not Applicable (G34T3076G), Not Applicable (G34T3078G), Not Applicable (G34T3079G), Not Applicable (G34T3080G), Not Applicable (G34T3081G), Not Applicable (G34T3082G), Not Applicable (G34T3083G), Not Applicable (G34T3085G), Not Applicable (G34T3087G), Not Applicable (G34T3088G), Not Applicable (G34T3089G), Not Applicable (G34T3091G), Not Applicable (G34T3117G), Not Applicable (H34T3454G), Not Applicable (J24T0733G), Not Applicable (J34T3562G), Not Applicable (K24T1208G), Not Applicable (K24T1219G), Not Applicable (B34T1992G), Not Applicable (B34T1994G), Not Applicable (B34T2053G), Not Applicable (B34T2054G), Not Applicable (B34T2059G), Not Applicable (G34T3092G), Not Applicable (G34T3174G), Not Applicable (G24T0515G), Not Applicable (K24T1223G), Not Applicable (L34T4012G), Not Applicable (K34T3919G), Not Applicable (D44T5029G), Not Applicable (J24T0697G), Not Applicable (J24T0897G), Not Applicable (J24T0901G), Not Applicable (K24T1227G), Not Applicable (M34T4512G), Not Applicable (H34T3318G), Not Applicable (C44T4957G), Not Applicable (D34T2497G), Not Applicable (G34T3172G), Not Applicable (G34T3177G), Not Applicable (J24T0804G), Not Applicable (J24T0930G), Not Applicable (J24T0773G), Not Applicable (J24T0794G), Not Applicable (J24T0653G), Not Applicable (G24T0373G), Not Applicable (J34T3702G), Not Applicable (J34T3706G), Not Applicable (J24T0727G), Not Applicable (A44T4567G), Not Applicable (B44T4788G), Not Applicable (B44T4808G), Not Applicable (B44T4813G), Not Applicable (C34T2243G), Not Applicable (C44T4943G), Not Applicable (D34T2504G), Not Applicable (D34T2508G), Not Applicable (D34T2512G), Not Applicable (D44T5028G), Not Applicable (E04TMLJ1B3), Not Applicable (F34T2896G), Not Applicable (G24T0552G), Not Applicable (G24T0562G), Not Applicable (G34T3155G), Not Applicable (H34T3347G), Not Applicable (H34T3368G), Not Applicable (J24T0757G), Not Applicable (J24T0853G), Not Applicable (J24T0856G), Not Applicable (J24T0891G), Not Applicable (J24T0894G), Not Applicable (J34T3805G), Not Applicable (J34T3808G), Not Applicable (K24T1009G), Not Applicable (K24T1154G), Not Applicable (K34T3884G), Not Applicable (L34T4139G), Not Applicable (L34T4140G), Not Applicable (L34T4141G), Not Applicable (L34T4143G), Not Applicable (M24T1607G), Not Applicable (M24T1610G), Not Applicable (M34T4229G), Not Applicable (M34T4511G), Not Applicable (K34T3932G), Not Applicable (D44T5019G), Not Applicable (D44T5020G), Not Applicable (J34T3515G), Not Applicable (J34T3518G), Not Applicable (J34T3526G), Not Applicable (J34T3528G), Not Applicable (J34T3529G), Not Applicable (J34T3534G), Not Applicable (A34T1906G), Not Applicable (A34T1912G), Not Applicable (B34T2132G), Not Applicable (J34T3661G), Not Applicable (J34T3663G), Not Applicable (B34T2044G), Not Applicable (B34T2112G), Not Applicable (B34T2113G), Not Applicable (C34T2255G), Not Applicable (C34T2257G), Not Applicable (C34T2260G), Not Applicable (C34T2264G), Not Applicable (C34T2270G), Not Applicable (C34T2278G), Not Applicable (D34T2462G), Not Applicable (D34T2584G), Not Applicable (D34T2586G), Not Applicable (D34T2612G), Not Applicable (D34T2617G), Not Applicable (K34T3879G), Not Applicable (C44T4959G), Not Applicable (F34T2922G), Not Applicable (M34T4495G), Not Applicable (M34T4496G), Not Applicable (M34T4375G), Not Applicable (M34T4457G), Not Applicable (K34T3922G), Not Applicable (D34T2469G), Not Applicable (D34T2478G), Not Applicable (F34T2912G), Not Applicable (F34T2916G), Not Applicable (F34T2917G), Not Applicable (F34T2918G), Not Applicable (M34T4440G), Not Applicable (M34T4441G), Not Applicable (M34T4447G), Not Applicable (M34T4464G), Not Applicable (M34T4465G), Not Applicable (J34T3536G), Not Applicable (J34T3537G), Not Applicable (J34T3538G), Not Applicable (J34T3539G), Not Applicable (J34T3541G), Not Applicable (J34T3543G), Not Applicable (J34T3545G), Not Applicable (J34T3694G), Not Applicable (J34T3695G), Not Applicable (J34T3696G), Not Applicable (J34T3768G), Not Applicable (L24T9KWG), Not Applicable (L24TAJYG), Not Applicable (L24TDZJG), Not Applicable (J34T3564G), Not Applicable (J34T3566G), Not Applicable (K24T1153G), Not Applicable (K24T1083G), Not Applicable (H34T3326G), Not Applicable (G24T0464G), Not Applicable (G24T0575G), Not Applicable (G34T3095G), Not Applicable (A34T1779G), Not Applicable (A34T1864G), Not Applicable (B34T2080G), Not Applicable (D34T2499G), Not Applicable (F24T0121G), Not Applicable (L34T4038G), Not Applicable (L34T4145G), Not Applicable (A44T4566G), Not Applicable (J34T3678G), Not Applicable (J34T3709G), Not Applicable (J34T3710G), Not Applicable (J34T3711G), Not Applicable (J34T3712G), Not Applicable (J34T3713G), Not Applicable (J34T3714G), Not Applicable (J34T3715G), Not Applicable (J34T3716G), Not Applicable (J34T3719G), Not Applicable (L34T4147G), Not Applicable (M34T4202G), Not Applicable (M34T4205G), Not Applicable (M34T4207G), Not Applicable (M34T4209G), Not Applicable (M34T4210G), Not Applicable (M34T4211G), Not Applicable (M34T4212G), Not Applicable (M34T4213G), Not Applicable (M34T4214G), Not Applicable (F34T2962G), Not Applicable (J34T3649G), Not Applicable (J34T3650G), Not Applicable (J34T3652G), Not Applicable (J34T3657G), Not Applicable (J34T3659G), Not Applicable (J34T3682G), Not Applicable (J34T3683G), Not Applicable (J34T3700G), Not Applicable (J34T3721G), Not Applicable (J34T3722G), Not Applicable (J34T3732G), Not Applicable (J34T3749G), Not Applicable (J34T3750G), Not Applicable (J34T3756G), Not Applicable (M34T4193G), Not Applicable (M34T4195G), Not Applicable (M34T4196G), Not Applicable (M34T4197G), Not Applicable (M34T4198G), Not Applicable (M34T4199G), Not Applicable (M34T4200G), Not Applicable (M34T4201G), Not Applicable (M34T4203G), Not Applicable (M34T4204G), Not Applicable (M34T4208G), Not Applicable (M34T4251G), Not Applicable (M34T4313G), Not Applicable (M34T4317G), Not Applicable (M34T4431G), Not Applicable (M34T4433G), Not Applicable (E34T2705G), Not Applicable (J24T0665G), Not Applicable (J24T0677G), Not Applicable (J24T0740G), Not Applicable (F24T0317G), Not Applicable (G24T0474G), Not Applicable (G24T0480G), Not Applicable (B44T4850G), Not Applicable (B44T4852G), Not Applicable (B44T4857G), Not Applicable (B44T4859G), Not Applicable (B44T4861G), Not Applicable (B44T4864G), Not Applicable (B44T4867G), Not Applicable (B44T4870G), Not Applicable (B44T4871G), Not Applicable (B44T4873G), Not Applicable (G24T0567G), Not Applicable (G24T0573G), Not Applicable (J24T0963G), Not Applicable (K24T1056G), Not Applicable (K24T1057G), Not Applicable (K24T1144G), Not Applicable (B34T2037G), Not Applicable (B34T2039G), Not Applicable (A44T4587G), Not Applicable (B44T4814G), Not Applicable (F34T2946G), Not Applicable (F34T2975G), Not Applicable (J34T3658G), Not Applicable (J34T3701G), Not Applicable (J34T3703G), Not Applicable (J34T3704G), Not Applicable (J34T3708G), Not Applicable (J34T3744G), Not Applicable (J34T3746G), Not Applicable (M34T4248G), Not Applicable (M34T4278G), Not Applicable (M34T4279G), Not Applicable (M34T4280G), Not Applicable (M34T4283G), Not Applicable (M34T4284G), Not Applicable (M34T4285G), Not Applicable (M34T4286G), Not Applicable (M34T4287G), Not Applicable (J34T3675G), Not Applicable (J34T3677G), Not Applicable (L34T4146G), Not Applicable (L34T4167G), Not Applicable (L34T4173G), Not Applicable (E34T2741G), Not Applicable (J34T3625G), Not Applicable (J34T3626G), Not Applicable (J34T3627G), Not Applicable (J34T3629G), Not Applicable (M34T4254G), Not Applicable (B44T4756G), Not Applicable (B34T2115G), Not Applicable (C34T2284G), Not Applicable (L24T1280G), Not Applicable (D34T2468G), Not Applicable (M34T4239G), Not Applicable (K24T1097G), Not Applicable (H34T3480G), Not Applicable (H34T3300G), Not Applicable (H34T3306G), Not Applicable (H34T3310G), Not Applicable (B44T4823G), Not Applicable (G24T0386G), Not Applicable (G24T0459G), Not Applicable (G24T0460G), Not Applicable (H34T3350G), Not Applicable (H34T3351G), Not Applicable (H34T3399G), Not Applicable (H34T3438G), Not Applicable (H34T3439G), Not Applicable (H34T3440G), Not Applicable (H34T3443G), Not Applicable (H34T3444G), Not Applicable (H34T3445G), Not Applicable (H34T3446G), Not Applicable (H34T3447G), Not Applicable (H34T3448G), Not Applicable (H34T3449G), Not Applicable (H34T3476G), Not Applicable (H34T3478G), Not Applicable (H34T3484G), Not Applicable (M24T1680G), Not Applicable (M24T1717G), Not Applicable (M24T1727G), Not Applicable (B34T2035G), Not Applicable (G34T3229G), Not Applicable (G34T3230G), Not Applicable (H34T3299G), Not Applicable (H34T3307G), Not Applicable (H34T3366G), Not Applicable (G24T0519G), Not Applicable (G24T0577G), Not Applicable (G24T0580G), Not Applicable (B34T1962G), Not Applicable (A44T4690G), Not Applicable (A44T4692G), Not Applicable (A44T4693G), Not Applicable (A44T4702G), Not Applicable (A44T4703G), Not Applicable (A44T4706G), Not Applicable (B44T4853G), Not Applicable (B44T4854G), Not Applicable (B44T4855G), Not Applicable (B44T4856G), Not Applicable (B44T4858G), Not Applicable (B44T4860G), Not Applicable (D44T5015G), Not Applicable (H34T3463G), Not Applicable (M24T1746G), Not Applicable (D34T2614G), Not Applicable (J34T3563G), Not Applicable (J34T3565G), Not Applicable (M24T1580G), Not Applicable (J34T3753G), Not Applicable (A44T4561G), Not Applicable (C34T2330G), Not Applicable (C34T2364G), Not Applicable (C34T2367G), Not Applicable (H34T3261G), Not Applicable (M34T4301G), Not Applicable (M34T4302G), Not Applicable (M34T4303G), Not Applicable (M34T4304G), Not Applicable (M34T4305G), Not Applicable (M34T4306G), Not Applicable (M34T4307G), Not Applicable (M34T4309G), Not Applicable (M34T4310G), Not Applicable (M34T4311G), Not Applicable (M34T4349G), Not Applicable (M34T4351G), Not Applicable (M34T4352G), Not Applicable (B34T2206G), Not Applicable (C34T2382G), Not Applicable (C34T2401G), Not Applicable (D34T2597G), Not Applicable (B34T2031G), Not Applicable (B44T4907G), Not Applicable (L34T4009G), Not Applicable (L34T4014G), Not Applicable (L34T4015G), Not Applicable (L34T4018G), Not Applicable (L34T4019G), Not Applicable (K24T1261G), Not Applicable (L24T1296G), Not Applicable (L24T1317G), Not Applicable (L24T1318G), Not Applicable (L24T1320G), Not Applicable (L24T1321G), Not Applicable (M24T1520G), Not Applicable (M24T1524G), Not Applicable (B34T1941G), Not Applicable (B34T1949G), Not Applicable (B34T2104G), Not Applicable (E34T2684G), Not Applicable (J24T0749G), Not Applicable (J24T0754G), Not Applicable (J24T0755G), Not Applicable (J24T0756G), Not Applicable (J24T0785G), Not Applicable (K34T3855G), Not Applicable (M24T1696G), Not Applicable (M24T1697G), Not Applicable (M24T1698G), Not Applicable (M24T1699G), Not Applicable (M34T4522G), Not Applicable (A44T4544G), Not Applicable (M34T4536G), Not Applicable (A34T1804G), Not Applicable (A34T1846G), Not Applicable (B34T2024G), Not Applicable (B34T2026G), Not Applicable (F24T0283G), Not Applicable (F24T0312G), Not Applicable (F24T0313G), Not Applicable (F34T2947G), Not Applicable (F34T2981G), Not Applicable (F34T3051G), Not Applicable (H34T3483G), Not Applicable (J34T3707G), Not Applicable (D34T2441G), Not Applicable (G24T0481G), Not Applicable (J34T3581G), Not Applicable (C34T2366G), Not Applicable (J24T0779G), Not Applicable (L24T1359G), Not Applicable (G24T0576G), Not Applicable (G24T0402G), Not Applicable (B44T4760G), Not Applicable (B44T4763G), Not Applicable (B44T4770G), Not Applicable (G24T0559G), Not Applicable (G24T0508G), Not Applicable (C44T4945G), Not Applicable (M24T1536G), Not Applicable (M24T1664G), Not Applicable (C34T2344G), Not Applicable (C34T2379G), Not Applicable (C34T2392G), Not Applicable (G24T0574G), Not Applicable (H34T3280G), Not Applicable (L34T4101G), Not Applicable (M34T4194G), Not Applicable (M34T4466G), Not Applicable (M34T4517G), Not Applicable (M34T4518G), Not Applicable (M34T4530G), Not Applicable (M34T4532G), Not Applicable (B34T1953G), Not Applicable (B34T1978G), Not Applicable (B34T1990G), Not Applicable (B34T2001G), Not Applicable (D34T2556G), Not Applicable (J34T3603G), Not Applicable (A34T1843G), Not Applicable (F24T0280G), Not Applicable (F24T0281G), Not Applicable (F24T0282G), Not Applicable (F24T0319G), Not Applicable (F24T0320G), Not Applicable (F24T0321G), Not Applicable (K24T1017G), Not Applicable (M34T4538G), Not Applicable (J24T0701G), Not Applicable (B44T4862G), Not Applicable (B44T4872G), Not Applicable (E34T2758G), Not Applicable (E34T2780G), Not Applicable (E34T2793G), Not Applicable (G34T3186G), Not Applicable (M34T4534G), Not Applicable (J24T0837G), Not Applicable (K24T1010G), Not Applicable (D34T2476G), Not Applicable (D34T2477G), Not Applicable (A44T4635G), Not Applicable (H34T3357G), Not Applicable (H34T3332G), Not Applicable (K24T1257G), Not Applicable (J24T0906G), Not Applicable (J24T0925G), Not Applicable (M24T1436G), Not Applicable (C44T4932G), Not Applicable (C44T4934G), Not Applicable (A34T1770G), Not Applicable (C34T2362G), Not Applicable (A44T4542G), Not Applicable (D34T2484G), Not Applicable (F24T0246G), Not Applicable (G34T3090G), Not Applicable (J24T0899G), Not Applicable (K34T3832G), Not Applicable (L34T4010G), Not Applicable (L34T4024G), Not Applicable (L34T4030G), Not Applicable (L34T4042G), Not Applicable (L34T4043G), Not Applicable (L34T4045G), Not Applicable (L34T4052G), Not Applicable (M34T4336G), Not Applicable (M34T4341G), Not Applicable (D34T2496G), Not Applicable (M34T4423G), Not Applicable (K34T3921G), Not Applicable (K34T3924G), Not Applicable (K34T3950G), Not Applicable (M24T1582G), Not Applicable (M24T1588G), Not Applicable (M24T1593G), Not Applicable (M24T1597G), Not Applicable (M24T1601G), Not Applicable (M24T1641G), Not Applicable (M24T1642G), Not Applicable (A34T1872G), Not Applicable (B44T4797G), Not Applicable (D34T2562G), Not Applicable (F24T0277G), Not Applicable (F34T2855G), Not Applicable (G24T0449G), Not Applicable (G34T3141G), Not Applicable (G34T3232G), Not Applicable (G34T3233G), Not Applicable (G34T3234G), Not Applicable (G34T3237G), Not Applicable (G34T3238G), Not Applicable (G34T3245G), Not Applicable (G34T3246G), Not Applicable (G34T3248G), Not Applicable (G34T3250G), Not Applicable (H34T3283G), Not Applicable (H34T3312G), Not Applicable (H34T3313G), Not Applicable (H34T3314G), Not Applicable (H34T3321G), Not Applicable (H34T3322G), Not Applicable (H34T3331G), Not Applicable (H34T3342G), Not Applicable (H34T3344G), Not Applicable (H34T3360G), Not Applicable (H34T3364G), Not Applicable (H34T3367G), Not Applicable (H34T3411G), Not Applicable (H34T3413G), Not Applicable (H34T3416G), Not Applicable (H34T3421G), Not Applicable (H34T3427G), Not Applicable (H34T3428G), Not Applicable (H34T3429G), Not Applicable (H34T3430G), Not Applicable (H34T3431G), Not Applicable (H34T3432G), Not Applicable (H34T3433G), Not Applicable (H34T3436G), Not Applicable (H34T3450G), Not Applicable (H34T3451G), Not Applicable (H34T3452G), Not Applicable (H34T3453G), Not Applicable (H34T3455G), Not Applicable (H34T3456G), Not Applicable (H34T3457G), Not Applicable (H34T3458G), Not Applicable (H34T3461G), Not Applicable (H34T3489G), Not Applicable (J24T0613G), Not Applicable (J24T0614G), Not Applicable (J24T0615G), Not Applicable (J24T0616G), Not Applicable (J24T0848G), Not Applicable (J24T0850G), Not Applicable (J24T0868G), Not Applicable (J24T0869G), Not Applicable (J34T3586G), Not Applicable (M34T4256G), Not Applicable (M34T4484G), Not Applicable (M34T4485G), Not Applicable (K24T1212G), Not Applicable (L24TDWSG), Not Applicable (G34T3135G), Not Applicable (K24T1040G), Not Applicable (L34T4003G), Not Applicable (M24T1460G), Not Applicable (M24T1491G), Not Applicable (M24T1499G), Not Applicable (M24T1501G), Not Applicable (M24T1502G), Not Applicable (M24T1605G), Not Applicable (M24T1634G), Not Applicable (M24T1637G), Not Applicable (M24T1638G), Not Applicable (F24T0120G), Not Applicable (K24T1121G), Not Applicable (J24T0626G), Not Applicable (J24T0628G), Not Applicable (J24T0629G), Not Applicable (K24T1262G), Not Applicable (K34T3835G), Not Applicable (K34T3836G), Not Applicable (K34T3842G), Not Applicable (L24TE71G), Not Applicable (M24T1554G), Not Applicable (M34T4395G), Not Applicable (K24T1053G), Not Applicable (M24T1535G), Not Applicable (M24T1576G), Not Applicable (B34T1943G), Not Applicable (F24T0223G), Not Applicable (G24T0516G), Not Applicable (A44T4596G), Not Applicable (B34T2022G), Not Applicable (B34T2028G), Not Applicable (F34T2832G), Not Applicable (F34T2909G), Not Applicable (F34T2982G), Not Applicable (K34T3987G), Not Applicable (K34T3988G), Not Applicable (K34T3992G), Not Applicable (K34T3994G), Not Applicable (M24T1438G), Not Applicable (E34T2772G), Not Applicable (M24T1655G), Not Applicable (K34T3822G), Not Applicable (K34T3830G), Not Applicable (B44T4811G), Not Applicable (B44T4812G), Not Applicable (B44T4832G), Not Applicable (B44T4833G), Not Applicable (L34T4132G), Not Applicable (L34T4133G), Not Applicable (K24T1159G), Not Applicable (A34T1873G), Not Applicable (A34T1880G), Not Applicable (E34T2655G), Not Applicable (F24T0216G), Not Applicable (J24T0664G), Not Applicable (J24T0825G), Not Applicable (J34T3655G), Not Applicable (J34T3665G), Not Applicable (K34T3916G), Not Applicable (L24T1355G), Not Applicable (L24T1364G), Not Applicable (L34T4073G), Not Applicable (L34T4074G), Not Applicable (A44T4715G), Not Applicable (K34T3964G), Not Applicable (K34T3969G), Not Applicable (L34T4053G), Not Applicable (M34T4520G), Not Applicable (M34T4527G), Not Applicable (M34T4528G), Not Applicable (B44T4779G), Not Applicable (B44T4839G), Not Applicable (B44T4843G), Not Applicable (B44T4844G), Not Applicable (B44T4845G), Not Applicable (B44T4846G), Not Applicable (K24T1038G), Not Applicable (E34T2784G), Not Applicable (G24T0339G), Not Applicable (G24T0348G), Not Applicable (M24T1679G), Not Applicable (F24T41255G), Not Applicable (G24T0332G), Not Applicable (G24T0393G), Not Applicable (K24T1155G), Not Applicable (M24T1737G), Not Applicable (M24T1738G), Not Applicable (L34T4129G), Not Applicable (L34T4134G), Not Applicable (L34T4136G), Not Applicable (D44T5013G), Not Applicable (C34T2398G), Not Applicable (D34T2430G), Not Applicable (D34T2431G), Not Applicable (D34T2536G), Not Applicable (D34T2539G), Not Applicable (E34T2660G), Not Applicable (F34T2970G), Not Applicable (F34T2971G), Not Applicable (F34T2972G), Not Applicable (F34T2973G), Not Applicable (F34T2977G), Not Applicable (F34T2979G), Not Applicable (G24T0397G), Not Applicable (G24T0438G), Not Applicable (G34T3094G), Not Applicable (G34T3111G), Not Applicable (G34T3112G), Not Applicable (G34T3113G), Not Applicable (G34T3114G), Not Applicable (G34T3115G), Not Applicable (G34T3116G), Not Applicable (G34T3121G), Not Applicable (G34T3122G), Not Applicable (G34T3149G), Not Applicable (J24T0645G), Not Applicable (J24T0646G), Not Applicable (J24T0656G), Not Applicable (L24TALKG), Not Applicable (L34T4047G), Not Applicable (B34T2051G), Not Applicable (B34T2052G), Not Applicable (B34T2079G), Not Applicable (J24T0864G), Not Applicable (J34T3508G), Not Applicable (A34T1904G), Not Applicable (A34T1907G), Not Applicable (B34T1928G), Not Applicable (B34T1932G), Not Applicable (B34T1934G), Not Applicable (B34T1935G), Not Applicable (B34T1939G), Not Applicable (B44T4863G), Not Applicable (B44T4869G), Not Applicable (C34T2239G), Not Applicable (C34T2244G), Not Applicable (F34T2893G), Not Applicable (G24T0478G), Not Applicable (H34T3341G), Not Applicable (J34T3687G), Not Applicable (J34T3739G), Not Applicable (J34T3790G), Not Applicable (L34T4099G), Not Applicable (L34T4177G), Not Applicable (M24T1577G), Not Applicable (D34T2472G), Not Applicable (H34T3496G), Not Applicable (B44T4753G), Not Applicable (B44T4755G), Not Applicable (D34T2591G), Not Applicable (D34T2600G), Not Applicable (M24T1468G), Not Applicable (M24T1565G), Not Applicable (M24T1681G), Not Applicable (M24T1690G), Not Applicable (M34T4270G), Not Applicable (J24T0815G), Not Applicable (J24T0824G), Not Applicable (J34T3795G), Not Applicable (B34T2036G), Not Applicable (B44T4865G), Not Applicable (B44T4866G), Not Applicable (B44T4868G), Not Applicable (B44T4906G), Not Applicable (F34T3036G), Not Applicable (F34T3048G), Not Applicable (F34T3053G), Not Applicable (G34T3105G), Not Applicable (M24T1751G), Not Applicable (M24T1753G), Not Applicable (F34T3020G), Not Applicable (A34T1927G), Not Applicable (M24T1741G), Not Applicable (M24T1742G), Not Applicable (M24T1748G), Not Applicable (M24T1756G), Not Applicable (M24T1757G), Not Applicable (M24T1758G), Not Applicable (F24T0274G), Not Applicable (F24T0275G), Not Applicable (J24T0620G), Not Applicable (J34T3736G), Not Applicable (K24T1141G), Not Applicable (K24T1145G), Not Applicable (K24T1146G), Not Applicable (L34T4169G), Not Applicable (K34T3901G), Not Applicable (K34T3905G), Not Applicable (A34T1839G), Not Applicable (A34T1842G), Not Applicable (J24T0621G), Not Applicable (J24T0736G), Not Applicable (E34T2776G), Not Applicable (F34T2871G), Not Applicable (F34T3054G), Not Applicable (F34T3055G), Not Applicable (G34T3101G), Not Applicable (G34T3106G), Not Applicable (G34T3120G), Not Applicable (G34T3151G), Not Applicable (G34T3153G), Not Applicable (H34T3464G), Not Applicable (H34T3466G), Not Applicable (H34T3479G), Not Applicable (H34T3482G), Not Applicable (H34T3488G), Not Applicable (J34T3651G), Not Applicable (K24T1073G), Not Applicable (K24T1074G), Not Applicable (K24T1075G), Not Applicable (K24T1076G), Not Applicable (K24T1077G), Not Applicable (K24T1078G), Not Applicable (K24T1079G), Not Applicable (K24T1081G), Not Applicable (K24T1082G), Not Applicable (K24T1084G), Not Applicable (K24T1086G), Not Applicable (K34T3917G), Not Applicable (M34T4261G), Not Applicable (B34T2131G), Not Applicable (A44T4686G), Not Applicable (B34T1931G), Not Applicable (L24T1384G), Not Applicable (L24T1387G), Not Applicable (M34T4510G), Not Applicable (M34T4513G), Not Applicable (M34T4514G), Not Applicable (M34T4515G), Not Applicable (L24T1290G), Not Applicable (A34T1861G), Not Applicable (D34T2506G), Not Applicable (D34T2507G), Not Applicable (D34T2554G), Not Applicable (D34T2558G), Not Applicable (D34T2583G), Not Applicable (D34T2585G), Not Applicable (D34T2588G), Not Applicable (D34T2636G), Not Applicable (J24T0661G), Not Applicable (J24T0687G), Not Applicable (B34T2138G), Not Applicable (A44T4662G), Not Applicable (B44T4749G), Not Applicable (H34T3492G), Not Applicable (H34T3497G), Not Applicable (J34T3525G), Not Applicable (L24T1308G), Not Applicable (M24T1466G), Not Applicable (A44T4552G), Not Applicable (A44T4553G), Not Applicable (A34T1787G), Not Applicable (A34T1788G), Not Applicable (A34T1790G), Not Applicable (F24T0255G), Not Applicable (F24T0258G), Not Applicable (G24T0369G), Not Applicable (G24T0401G), Not Applicable (G24T0404G), Not Applicable (G24T0410G), Not Applicable (G24T0421G), Not Applicable (G24T0500G), Not Applicable (G24T0501G), Not Applicable (G24T0588G), Not Applicable (G24T0589G), Not Applicable (G24T0590G), Not Applicable (H34T3358G), Not Applicable (H34T3359G), Not Applicable (H34T3362G), Not Applicable (J24T0622G), Not Applicable (J24T0623G), Not Applicable (J24T0632G), Not Applicable (J24T0763G), Not Applicable (J24T0764G), Not Applicable (J24T0841G), Not Applicable (J24T0844G), Not Applicable (J24T0849G), Not Applicable (J24T0851G), Not Applicable (J24T0861G), Not Applicable (J24T0866G), Not Applicable (J24T0871G), Not Applicable (J24T0882G), Not Applicable (J24T0885G), Not Applicable (J24T0886G), Not Applicable (J34T3698G), Not Applicable (J34T3784G), Not Applicable (J34T3786G), Not Applicable (J34T3797G), Not Applicable (J34T3798G), Not Applicable (J34T3799G), Not Applicable (J34T3801G), Not Applicable (K24T1063G), Not Applicable (K24T1090G), Not Applicable (K24T1091G), Not Applicable (K24T1103G), Not Applicable (K24T1173G), Not Applicable (L24T1300G), Not Applicable (L24T1311G), Not Applicable (J24T0828G), Not Applicable (C34T2376G), Not Applicable (D34T2567G), Not Applicable (D34T2572G), Not Applicable (H34T3417G), Not Applicable (J24T0680G), Not Applicable (J24T0683G), Not Applicable (J24T0684G), Not Applicable (J24T0685G), Not Applicable (J24T0686G), Not Applicable (J24T0690G), Not Applicable (M24T1676G), Not Applicable (A44T4545G), Not Applicable (A44T4550G), Not Applicable (A44T4551G), Not Applicable (B34T2025G), Not Applicable (B44T4757G), Not Applicable (B44T4889G), Not Applicable (B44T4891G), Not Applicable (D34T2427G), Not Applicable (D34T2531G), Not Applicable (D34T2590G), Not Applicable (D34T2592G), Not Applicable (F24T0080G), Not Applicable (F24T0085G), Not Applicable (H34T3408G), Not Applicable (K24T1234G), Not Applicable (K34T3898G), Not Applicable (L24T1328G), Not Applicable (M34T4265G), Not Applicable (M34T4324G), Not Applicable (M34T4330G), Not Applicable (M34T4345G), Not Applicable (M34T4373G), Not Applicable (M34T4380G), Not Applicable (M34T4389G), Not Applicable (M34T4393G), Not Applicable (M34T4418G), Not Applicable (M34T4419G), Not Applicable (A34T1913G), Not Applicable (A44T4634G), Not Applicable (B44T4824G), Not Applicable (F24T0248G), Not Applicable (J24T0865G), Not Applicable (J24T0870G), Not Applicable (J24T0893G), Not Applicable (A34T1840G), Not Applicable (A34T1841G), Not Applicable (J34T3757G), Not Applicable (J34T3758G), Not Applicable (J34T3760G), Not Applicable (J34T3767G), Not Applicable (C44T4979G), Not Applicable (J24T0675G), Not Applicable (J24T0708G), Not Applicable (G34T3108G), Not Applicable (A44T4548G), Not Applicable (L34T4135G), Not Applicable (L34T4137G), Not Applicable (L34T4144G), Not Applicable (L34T4148G), Not Applicable (L34T4150G), Not Applicable (L34T4154G), Not Applicable (L34T4166G), Not Applicable (M34T4252G), Not Applicable (E34T2794G), Not Applicable (E34T2795G), Not Applicable (F34T2894G), Not Applicable (J24t0922G), Not Applicable (J24T0923G), Not Applicable (J24T0926G), Not Applicable (J24T0928G), Not Applicable (J34T3506G), Not Applicable (J34T3532G), Not Applicable (A44T4624G), Not Applicable (A44T4625G), Not Applicable (J34T3815G), Not Applicable (M24T1429G), Not Applicable (M24T1476G), Not Applicable (G34T3179G), Not Applicable (G34T3188G), Not Applicable (G34T3189G), Not Applicable (C34T2331G), Not Applicable (C34T2342G), Not Applicable (C34T2369G), Not Applicable (C34T2377G), Not Applicable (C34T2404G), Not Applicable (J34T3584G), Not Applicable (G24T0412G), Not Applicable (G24T0554G), Not Applicable (G24T0568G), Not Applicable (M24T1510G), Not Applicable (M24T1599G), Not Applicable (M24T1602G), Not Applicable (J34T3608G), Not Applicable (J34T3726G), Not Applicable (J34T3728G), Not Applicable (G24T0544G), Not Applicable (G24T0551G), Not Applicable (G24T0553G), Not Applicable (H34T3426G), Not Applicable (A44T4557G), Not Applicable (B34T2129G), Not Applicable (B34T2133G), Not Applicable (C34T2306G), Not Applicable (D34T2521G), Not Applicable (D34T2630G), Not Applicable (D34T2532G), Not Applicable (F24T0024G), Not Applicable (F24T0041G), Not Applicable (G24T0388G), Not Applicable (G24T0565G), Not Applicable (G24T0566G), Not Applicable (H34T3393G), Not Applicable (J34T3613G), Not Applicable (M34T4401G), Not Applicable (M34T4404G), Not Applicable (H34T3390G), Not Applicable (H34T3391G), Not Applicable (K34T3965G), Not Applicable (K34T3971G), Not Applicable (B34T2152G), Not Applicable (B34T2185G), Not Applicable (F24T0217G), Not Applicable (G24T0525G), Not Applicable (M24T1484G), Not Applicable (M24T1486G), Not Applicable (M24T1487G), Not Applicable (M24T1550G), Not Applicable (G24T0507G), Not Applicable (F24T0250G), Not Applicable (F24T0251G), Not Applicable (F24T0252G), Not Applicable (F34T2890G), Not Applicable (F34T2892G), Not Applicable (K34T3915G), Not Applicable (L34T4039G), Not Applicable (L34T4048G), Not Applicable (L34T4050G), Not Applicable (L34T4062G), Not Applicable (L34T4070G), Not Applicable (L34T4071G), Not Applicable (L34T4072G), Not Applicable (L34T4076G), Not Applicable (L34T4078G), Not Applicable (F24T0142G), Not Applicable (F24T0143G), Not Applicable (F24T0144G), Not Applicable (C34T2329G), Not Applicable (C34T2337G), Not Applicable (C34T2339G), Not Applicable (C34T2340G), Not Applicable (J24T0625G), Not Applicable (J24T0644G), Not Applicable (M34T4348G), Not Applicable (M34T4353G), Not Applicable (C34T2321G), Not Applicable (H34T3348G), Not Applicable (H34T3392G), Not Applicable (H34T3398G), Not Applicable (H34T3400G), Not Applicable (H34T3422G), Not Applicable (J34T3592G), Not Applicable (L24T1286G), Not Applicable (L24T1301G), Not Applicable (G24T0563G), Not Applicable (G24T0569G), Not Applicable (G24T0570G), Not Applicable (G24T0572G), Not Applicable (K34T3977G), Not Applicable (L34T3997G), Not Applicable (L34T3999G), Not Applicable (L34T4016G), Not Applicable (F24T0293G), Not Applicable (A44T4629G), Not Applicable (A44T4631G), Not Applicable (B44T4836G), Not Applicable (G24T0491G), Not Applicable (G24T0547G), Not Applicable (K24T1128G), Not Applicable (M24T1479G), Not Applicable (M24T1485G), Not Applicable (M34T4506G), Not Applicable (M34T4507G), Not Applicable (L34T4075G), Not Applicable (K24T1245G), Not Applicable (A44T4563G), Not Applicable (A44T4720G), Not Applicable (M24T1649G), Not Applicable (M24T1653G), Not Applicable (M24T1654G), Not Applicable (M24T1656G), Not Applicable (M24T1657G), Not Applicable (M24T1658G).Not Applicable (C34T2384G), Not Applicable (K24T1037G), Not Applicable (K24T1104G), Not Applicable (K24T1107G), Not Applicable (K24T1109G), Not Applicable (K24T1110G), Not Applicable (K24T1111G), Not Applicable (D34T2457G), Not Applicable (D34T2464G), Not Applicable (D34T2529G), Not Applicable (D34T2540G), Not Applicable (H24T0602G), Not Applicable (M24T1413G), Not Applicable (M24T1459G), Not Applicable (J34T3574G), Not Applicable (A44T4588G), Not Applicable (A44T4592G), Not Applicable (A44T4626G), Not Applicable (A44T4666G), Not Applicable (A44T4668G), Not Applicable (A44T4678G), Not Applicable (A44T4679G), Not Applicable (A44T4681G), Not Applicable (A44T4682G), Not Applicable (A44T4683G), Not Applicable (A44T4688G), Not Applicable (A44T4689G), Not Applicable (A44T4695G), Not Applicable (B34T2093G), Not Applicable (B44T4903G), Not Applicable (C34T2312G), Not Applicable (C34T2345G), Not Applicable (E34T2782G), Not Applicable (F24T0123G), Not Applicable (F24T0257G), Not Applicable (F34T2838G), Not Applicable (F34T2886G), Not Applicable (F34T2887G), Not Applicable (F34T2943G), Not Applicable (F34T2957G), Not Applicable (F34T2958G), Not Applicable (F34T2960G), Not Applicable (F34T2963G), Not Applicable (G34T3064G), Not Applicable (G34T3067G), Not Applicable (G34T3069G), Not Applicable (G34T3070G), Not Applicable (G34T3093G), Not Applicable (G34T3118G), Not Applicable (G34T3142G), Not Applicable (G34T3146G), Not Applicable (G34T3147G), Not Applicable (G34T3150G), Not Applicable (G34T3152G), Not Applicable (G34T3156G), Not Applicable (G34T3183G), Not Applicable (G34T3187G), Not Applicable (G34T3201G), Not Applicable (G34T3203G), Not Applicable (G34T3210G), Not Applicable (G34T3239G), Not Applicable (G34T3240G), Not Applicable (G34T3241G), Not Applicable (G34T3242G), Not Applicable (G34T3243G), Not Applicable (G34T3244G), Not Applicable (G34T3247G), Not Applicable (G34T3249G), Not Applicable (J34T3654G), Not Applicable (J34T3656G), Not Applicable (J34T3660G), Not Applicable (J34T3817G), Not Applicable (K34T3821G), Not Applicable (K34T3824G), Not Applicable (K34T3827G), Not Applicable (K34T3861G), Not Applicable (K34T3869G), Not Applicable (K34T3887G), Not Applicable (K34T3891G), Not Applicable (K34T3918G), Not Applicable (K34T3961G), Not Applicable (L34T4156G), Not Applicable (L34T4107G), Not Applicable (F34T2843G), Not Applicable (E34T2666G), Not Applicable (E34T2667G), Not Applicable (E34T2775G), Not Applicable (F34T2889G), Not Applicable (F24T0150G), Not Applicable (F24T0151G), Not Applicable (F24T0152G), Not Applicable (F24T0204G), Not Applicable (G24T0330G), Not Applicable (G24T0371G), Not Applicable (G24T0511G), Not Applicable (G24T0513G), Not Applicable (E34T2665G), Not Applicable (G24T0534G), Not Applicable (G24T0537G), Not Applicable (A34T1817G), Not Applicable (F24T0075G), Not Applicable (F24T0076G), Not Applicable (F24T0224G), Not Applicable (F24T0225G), Not Applicable (F24T0237G), Not Applicable (G24T0457G), Not Applicable (K24T1267G), Not Applicable (K24T1269G), Not Applicable (K24T1270G), Not Applicable (M24T1477G), Not Applicable (M24T1539G), Not Applicable (M24T1557G), Not Applicable (F24T0073G), Not Applicable (F24T0086G), Not Applicable (F34T3034G), Not Applicable (F34T3037G), Not Applicable (F34T3039G), Not Applicable (G34T3173G), Not Applicable (G34T3175G), Not Applicable (H34T3319G), Not Applicable (L24T1361G), Not Applicable (M24T1478G), Not Applicable (M24T1684G), Not Applicable (M24T1714G), Not Applicable (M34T4476G), Not Applicable (C44T4939G), Not Applicable (C44T4946G), Not Applicable (C44T4951G), Not Applicable (C44T4954G), Not Applicable (C44T4978G), Not Applicable (G24T0485G), Not Applicable (G24T0487G), Not Applicable (H34T3275G), Not Applicable (H34T3325G), Not Applicable (G34T3143G), Not Applicable (H34T3269G), Not Applicable (H34T3271G), Not Applicable (H34T3273G), Not Applicable (H34T3308G), Not Applicable (K24T1135G), Not Applicable (K24T1216G), Not Applicable (K24T1218G), Not Applicable (H34T3425G), Not Applicable (L34T4168G), Not Applicable (L34T4171G), Not Applicable (L34T4172G), Not Applicable (L34T4176G), Not Applicable (F24T0023G), Not Applicable (G34T3148G), Not Applicable (H34T3371G), Not Applicable (D34T2608G), Not Applicable (K24T1181G), Not Applicable (M34T4455G), Not Applicable (M34T4477G), Not Applicable (D34T2458G), Not Applicable (G34T3134G), Not Applicable (A44T4623G), Not Applicable (B44T4831G), Not Applicable (D44T5038G), Not Applicable (D44T5039G), Not Applicable (F24T0315G), Not Applicable (F34T2967G), Not Applicable (F34T3008G), Not Applicable (F34T3014G), Not Applicable (G24T0479G), Not Applicable (H34T3296G), Not Applicable (M24T1571G), Not Applicable (M34T4238G), Not Applicable (M34T4282G), Not Applicable (M34T4403G), Not Applicable (M34T4414G), Not Applicable (M34T4416G), Not Applicable (M34T4427G), Not Applicable (M34T4480G), Not Applicable (M34T4488G), Not Applicable (M34T4490G), Not Applicable (M34T4492G), Not Applicable (M34T4500G), Not Applicable (M34T4502G), Not Applicable (M34T4508G), Not Applicable (M34T4519G), Not Applicable (J24T0939G), Not Applicable (J24T0951G), Not Applicable (D44T5007G), Not Applicable (D44T5008G), Not Applicable (D44T5009G), Not Applicable (D44T5010G), Not Applicable (L24T1374G), Not Applicable (L24T1375G), Not Applicable (L24T1378G), Not Applicable (L24T1381G), Not Applicable (L24T1382G), Not Applicable (L24T1383G), Not Applicable (L24T1392G), Not Applicable (L24T1396G), Not Applicable (L24T1397G), Not Applicable (L24T1402G), Not Applicable (L24T1403G), Not Applicable (F24T0311G), Not Applicable (J24T0702G), Not Applicable (J24T0703G), Not Applicable (K24T1134G), Not Applicable (M34T4281G), Not Applicable (K34T3897G), Not Applicable (A44T4651G), Not Applicable (A44T4657G), Not Applicable (H34T3471G), Not Applicable (H34T3473G), Not Applicable (H34T3495G), Not Applicable (J34T3519G), Not Applicable (J34T3590G), Not Applicable (M34T3495G), Not Applicable (F24T0295G), Not Applicable (G24T0465G), Not Applicable (G34T3154G), Not Applicable (F34T2964G), Not Applicable (C34T2233G), Not Applicable (C34T2411G), Not Applicable (D34T2443G), Not Applicable (D34T2445G), Not Applicable (D34T2544G), Not Applicable (D34T2599G), Not Applicable (E34T2649G), Not Applicable (E34T2650G), Not Applicable (E34T2652G), Not Applicable (K24T1254G), Not Applicable (K24T1268G), Not Applicable (M24T1517G), Not Applicable (M24T1521G), Not Applicable (E34T2738G), Not Applicable (J34T3510G), Not Applicable (A44T4554G), Not Applicable (M24T1515G), Not Applicable (G24T0447G), Not Applicable (B34T1945G), Not Applicable (B34T2100G), Not Applicable (B34T2101G), Not Applicable (C44T4937G), Not Applicable (D34T2596G), Not Applicable (E34T2791G), Not Applicable (F24T0243G), Not Applicable (J34T3727G), Not Applicable (K34T3881G), Not Applicable (K34T3954G), Not Applicable (M24T1562G), Not Applicable (M24T1563G), Not Applicable (M24T1564G), Not Applicable (M24T1581G), Not Applicable (M24T1614G), Not Applicable (M24T1615G), Not Applicable (M24T1651G), Not Applicable (M24T1755G), Not Applicable (C44T4941G), Not Applicable (K24T1232G), Not Applicable (K34T3955G), Not Applicable (L24T1388G), Not Applicable (A44T4628G), Not Applicable (K24T1115G), Not Applicable (L34T4069G), Not Applicable (L34T4077G), Not Applicable (M34T4240G), Not Applicable (M34T4241G), Not Applicable (M34T4246G), Not Applicable (M34T4247G), Not Applicable (M34T4250G), Not Applicable (C34T2346G), Not Applicable (C34T2347G), Not Applicable (C34T2348G), Not Applicable (C34T2349G), Not Applicable (C34T2373G), Not Applicable (C34T2374G), Not Applicable (C34T2375G), Not Applicable (J24T0752G), Not Applicable (J24T0761G), Not Applicable (J24T0847G), Not Applicable (J24T0875G), Not Applicable (J34T3792G), Not Applicable (J34T3793G), Not Applicable (J34T3794G), Not Applicable (D34T2561G), Not Applicable (J34T3567G), Not Applicable (J34T3568G), Not Applicable (K24T1069G), Not Applicable (K24T1147G), Not Applicable (K24T1242G), Not Applicable (K24T1247G), Not Applicable (M24T1505G), Not Applicable (M24T1715G), Not Applicable (M24T1719G), Not Applicable (M24T1720G), Not Applicable (M24T1723G), Not Applicable (A44T4664G), Not Applicable (L34T4004G), Not Applicable (L34T4115G), Not Applicable (M34T4245G), Not Applicable (B34T2106G), Not Applicable (B34T2110G), Not Applicable (C34T2242G), Not Applicable (C34T2247G), Not Applicable (C34T2282G), Not Applicable (G24T0523G), Not Applicable (K24T1032G), Not Applicable (K24T1209G), Not Applicable (A44T4696G), Not Applicable (A44T4701G), Not Applicable (B34T1956G), Not Applicable (B34T2194G), Not Applicable (C34T2240G), Not Applicable (C44T4956G), Not Applicable (C44T4962G), Not Applicable (F24T0231G), Not Applicable (F24T0232G), Not Applicable (G24T0390G), Not Applicable (G24T0476G), Not Applicable (G24T0494G), Not Applicable (H34T3251G), Not Applicable (H34T3253G), Not Applicable (H34T3256G), Not Applicable (H34T3257G), Not Applicable (H34T3343G), Not Applicable (H34T3345G), Not Applicable (J24T0647G), Not Applicable (J24T0648G), Not Applicable (J24T0669G), Not Applicable (J24T0696G), Not Applicable (J24T0698G), Not Applicable (J24T0706G), Not Applicable (J34T3633G), Not Applicable (J34T3636G), Not Applicable (K24T0998G), Not Applicable (K24T1031G), Not Applicable (K24T1052G), Not Applicable (K24T1054G), Not Applicable (K24T1055G), Not Applicable (K24T1062G), Not Applicable (K24T1064G), Not Applicable (K24T1067G), Not Applicable (K34T3885G), Not Applicable (K34T3888G), Not Applicable (K34T3889G), Not Applicable (K34T3956G), Not Applicable (M24T1531G), Not Applicable (M24T1532G), Not Applicable (A34T1897G), Not Applicable (A34T1915G), Not Applicable (G24T0581G), Not Applicable (G24T0548G), Not Applicable (B34T2057G), Not Applicable (B34T2196G), Not Applicable (B34T2197G), Not Applicable (B44T4806G), Not Applicable (B44T4816G), Not Applicable (B44T4817G), Not Applicable (B44T4818G), Not Applicable (C34T2251G), Not Applicable (E34T2759G), Not Applicable (K34T3933G), Not Applicable (E34T2751G), Not Applicable (E34T2787G), Not Applicable (G34T3123G), Not Applicable (G34T3124G), Not Applicable (G34T3126G), Not Applicable (K34T3958G), Not Applicable (F34T2825G), Not Applicable (G34T3129G), Not Applicable (G34T3130G), Not Applicable (G34T3131G), Not Applicable (J34T3673G), Not Applicable (K24T1203G), Not Applicable (L34T4119G), Not Applicable (L34T4153G), Not Applicable (B44T4911G), Not Applicable (D44T5012G), Not Applicable (K34T3979G), Not Applicable (M34T4233G), Not Applicable (M34T4236G), Not Applicable (M34T4237G), Not Applicable (M34T4242G), Not Applicable (M34T4243G), Not Applicable (J24T0822G), Not Applicable (F24T0147G), Not Applicable (F34T2880G), Not Applicable (G34T3140G), Not Applicable (K34T3951G), Not Applicable (J24T0670G), Not Applicable (J24T0935G), Not Applicable (C44T4960G), Not Applicable (D34T2517G), Not Applicable (H34T3276G), Not Applicable (K24T1248G), Not Applicable (K24T1250G), Not Applicable (K24T1271G), Not Applicable (K34T3894G), Not Applicable (K34T3936G), Not Applicable (K34T3972G), Not Applicable (K34T3973G), Not Applicable (M24T1470G), Not Applicable (M24T1578G), Not Applicable (M24T1578GG), Not Applicable (M24T1579G), Not Applicable (F24T0175G), Not Applicable (F24T0182G), Not Applicable (F24T0183G), Not Applicable (K34T3826G), Not Applicable (K34T3831G), Not Applicable (M24T1461G), Not Applicable (F34T30506), Not Applicable (F34T3050G), Not Applicable (A44T4589G), Not Applicable (A44T4684G), Not Applicable (B34T2000G), Not Applicable (B44T4885G), Not Applicable (B44T4886G), Not Applicable (B44T4896G), Not Applicable (C34T2357G), Not Applicable (C34T2370G), Not Applicable (C34T2416G), Not Applicable (D34T2500G), Not Applicable (E34T2755G), Not Applicable (E34T2766G), Not Applicable (E34T2777G), Not Applicable (F24T0018G), Not Applicable (F24T0087G), Not Applicable (F24T0088G), Not Applicable (F24T0122G), Not Applicable (F34T2755G), Not Applicable (F34T2830G), Not Applicable (F34T2831G), Not Applicable (F34T2837G), Not Applicable (F34T2842G), Not Applicable (F34T2844G), Not Applicable (F34T2845G), Not Applicable (F34T2846G), Not Applicable (F34T2848G), Not Applicable (F34T2849G), Not Applicable (F34T2851G), Not Applicable (F34T2853G), Not Applicable (F34T2948G), Not Applicable (F34T2949G), Not Applicable (F34T2950G), Not Applicable (F34T2996G), Not Applicable (G24T0458G), Not Applicable (G24T0489G), Not Applicable (G24T0526G), Not Applicable (G24T0527G), Not Applicable (G34T3065G), Not Applicable (G34T3066G), Not Applicable (H34T3472G), Not Applicable (J24T0705G), Not Applicable (J24T0735G), Not Applicable (J24T0751G), Not Applicable (J24T0760G), Not Applicable (J24T0762G), Not Applicable (J24T0771G), Not Applicable (J24T0772G), Not Applicable (J24T0778G), Not Applicable (J24T0780G), Not Applicable (J24T0786G), Not Applicable (J24T0789G), Not Applicable (J24T0796G), Not Applicable (J24T0798G), Not Applicable (J24T0808G), Not Applicable (J24T0809G), Not Applicable (J24T0811G), Not Applicable (J24T0816G), Not Applicable (J24T0829G), Not Applicable (J24T0857G), Not Applicable (J24T0858G), Not Applicable (J24T0883G), Not Applicable (J24T0895G), Not Applicable (J24T0903G), Not Applicable (J24T0904G), Not Applicable (J24T0909G), Not Applicable (J24T0914G), Not Applicable (J24T0915G), Not Applicable (J24T0918G), Not Applicable (J24T0920G), Not Applicable (J24T0921G), Not Applicable (J24T0933G), Not Applicable (J34T3579G), Not Applicable (J34T3666G), Not Applicable (J34T3679G), Not Applicable (J34T3685G), Not Applicable (K24T1093G), Not Applicable (K24T1098G), Not Applicable (K24T1138G), Not Applicable (K24T1139G), Not Applicable (K24T1252G), Not Applicable (L24T1279G), Not Applicable (L24T1298G), Not Applicable (M24T1457G), Not Applicable (M24T1472G), Not Applicable (M24T1507G), Not Applicable (M24T1508G), Not Applicable (M24T1509G), Not Applicable (M24T1542G), Not Applicable (M24T1551G), Not Applicable (M24T1566G), Not Applicable (M24T1583G), Not Applicable (M24T1587G), Not Applicable (M24T1590G), Not Applicable (M24T1592G), Not Applicable (M24T1683G), Not Applicable (M24T1685G), Not Applicable (M24T1694G), Not Applicable (M24T1695G), Not Applicable (M24T1701G), Not Applicable (M24T1703G), Not Applicable (M24T1707G), Not Applicable (M24T1708G), Not Applicable (M24T1709G), Not Applicable (M24T1728G), Not Applicable (M34T4228G), Not Applicable (M34T4525G), Not Applicable (C34T2414G), Not Applicable (J24T0818G), Not Applicable (J34T3766G), Not Applicable (K24T1132G), Not Applicable (K24T1133G), Not Applicable (K34T3864G), Not Applicable (K34T3866G), Not Applicable (L24T1295G), Not Applicable (M24T1430G), Not Applicable (M34T4255G), Not Applicable (M34T4472G), Not Applicable (M34T4473G), Not Applicable (B44T4768G), Not Applicable (A34T1792G), Not Applicable (M24T1519G), Not Applicable (C44T4955G), Not Applicable (B44T4765G), Not Applicable (B44T4774G), Not Applicable (D44T5011G), Not Applicable (E34T2728G), Not Applicable (E34T2745G), Not Applicable (E34T2760G), Not Applicable (E34T2789G), Not Applicable (E34T2797G), Not Applicable (F34T2868G), Not Applicable (F34T2940G), Not Applicable (H34T3282G), Not Applicable (A34T1819G), Not Applicable (F34T3010G), Not Applicable (F34T3017G), Not Applicable (K24T1161G), Not Applicable (L24T1287G), Not Applicable (L24T1289G), Not Applicable (L24T1293G), Not Applicable (L24T1312G), Not Applicable (L24T1313G), Not Applicable (L34T4008G), Not Applicable (M24T1434G), Not Applicable (C44T4921G), Not Applicable (C44T4922G), Not Applicable (C44T4933G), Not Applicable (C44T4952G), Not Applicable (C44T4953G), Not Applicable (C44T4958G), Not Applicable (C34T2234G), Not Applicable (F24T0022G), Not Applicable (F24T0145G), Not Applicable (G24T0518G), Not Applicable (H34T3465G), Not Applicable (H34T3467G), Not Applicable (J34T3502G), Not Applicable (J34T3504G), Not Applicable (J34T3505G), Not Applicable (J34T3512G), Not Applicable (J34T3569G), Not Applicable (G24T0333G), Not Applicable (C44T4923G), Not Applicable (C44T4949G), Not Applicable (D44T5041G), Not Applicable (J34T3578G), Not Applicable (J34T3580G), Not Applicable (M24T1538G), Not Applicable (K34T3981G), Not Applicable (L34T4049G), Not Applicable (L34T4083G), Not Applicable (L34T4023G), Not Applicable (D34T2624G), Not Applicable (F24T0240G), Not Applicable (B34T1959G), Not Applicable (B34T1993G), Not Applicable (K34T3849G), Not Applicable (M24T1514G), Not Applicable (M24T1541G), Not Applicable (M24T1553G), Not Applicable (M24T1556G), Not Applicable (M24T1558G), Not Applicable (M24T1589G), Not Applicable (M24T1643G), Not Applicable (M24T1648G), Not Applicable (M24T1650G), Not Applicable (M24T1704G), Not Applicable (M24T1705G), Not Applicable (M24T1740G), Not Applicable (M24T1747G), Not Applicable (F24T0146G), Not Applicable (L24T1282G), Not Applicable (L24T1283G), Not Applicable (L24T1285G), Not Applicable (L24T1288G), Not Applicable (L24T1379G), Not Applicable (L24T1380G), Not Applicable (L24T1394G), Not Applicable (L24T1395G), Not Applicable (M34T4222G), Not Applicable (M34T4232G), Not Applicable (K24T1174G), Not Applicable (A34T1911G), Not Applicable (F24T0129G), Not Applicable (L34T4027G), Not Applicable (L34T4037G), Not Applicable (M24T1426G), Not Applicable (M24T1702G), Not Applicable (D44T5040G), Not Applicable (H34T3493G), Not Applicable (J34T3514G), Not Applicable (J34T3516G), Not Applicable (M34T4223G), Not Applicable (D34T2603G), Not Applicable (D34T2625G), Not Applicable (D34T2626G), Not Applicable (E34T2690G), Not Applicable (E34T2692G), Not Applicable (K34T3895G), Not Applicable (K34T3923G), Not Applicable (L34T4011G), Not Applicable (L34T4029G), Not Applicable (F24T0268G), Not Applicable (E34T2743G), Not Applicable (J34T3596G), Not Applicable (J34T3788G), Not Applicable (L24T1350G), Not Applicable (L24T1353G), Not Applicable (C34T2250G), Not Applicable (C34T2253G), Not Applicable (E34T2682G), Not Applicable (G24T0331G), Not Applicable (G34T3209G), Not Applicable (M34T4357G), Not Applicable (J24T0624G), Not Applicable (B34T1981G), Not Applicable (B44T4842G), Not Applicable (B44T4901G), Not Applicable (C34T2292G), Not Applicable (C34T2295G), Not Applicable (F34T2965G), Not Applicable (F34T2966G), Not Applicable (F34T2968G), Not Applicable (F34T2969G), Not Applicable (F34T2974G), Not Applicable (F34T2988G), Not Applicable (F34T2989G), Not Applicable (L34T4031G), Not Applicable (M24T1567G), Not Applicable (M24T1568G), Not Applicable (M24T1570G), Not Applicable (M24T1572G), Not Applicable (M24T1574G), Not Applicable (M24T1678G), Not Applicable (M24T1635G), Not Applicable (M24T1639G), Not Applicable (B34T2032G), Not Applicable (B34T2040G), Not Applicable (C34T2273G), Not Applicable (C34T2280G), Not Applicable (C34T2332G), Not Applicable (C34T2333G), Not Applicable (C34T2371G), Not Applicable (E34T2723G), Not Applicable (A44T4705G), Not Applicable (D44T5026G), Not Applicable (A44T4717G), Not Applicable (A44T4719G), Not Applicable (A44T4727G), Not Applicable (K24T1192G), Not Applicable (A44T4697G), Not Applicable (D44T5021G), Not Applicable (D44T5022G), Not Applicable (M34T4268G), Not Applicable (M34T4377G), Not Applicable (F34T2925G), Not Applicable (G34T3227G), Not Applicable (G34T3235G), Not Applicable (H34T3259G), Not Applicable (H34T3262G), Not Applicable (M34T4267G), Not Applicable (M34T4320G), Not Applicable (M34T4384G), Not Applicable (A44T4594G), Not Applicable (B44T4761G), Not Applicable (B44T4847G), Not Applicable (K34T3867G), Not Applicable (K34T3984G), Not Applicable (K34T3985G), Not Applicable (K34T3986G), Not Applicable (K34T3989G), Not Applicable (A44T4582G), Not Applicable (A44T4656G), Not Applicable (F34T3018G), Not Applicable (F34T3043G), Not Applicable (F34T3045G), Not Applicable (F34T3046G), Not Applicable (D34T2610G), Not Applicable (G24T0555G), Not Applicable (G24T0579G), Not Applicable (J24T0634G), Not Applicable (L24T1319G), Not Applicable (M24T1752G), Not Applicable (M24T1754G), Not Applicable (F34T3030G), Not Applicable (F34T3049G), Not Applicable (G34T3061G), Not Applicable (K24T1249G), Not Applicable (M24T1530G), Not Applicable (M24T1534G), Not Applicable (M34T4487G), Not Applicable (M34T4489G), Not Applicable (D34T2520G), Not Applicable (F34T2819G), Not Applicable (F34T2881G), Not Applicable (F34T2882G), Not Applicable (F34T2885G), Not Applicable (F34T2891G), Not Applicable (G24T0370G), Not Applicable (G24T0415G), Not Applicable (G24T0416G), Not Applicable (G24T0418G), Not Applicable (G24T0493G), Not Applicable (J24T0776G), Not Applicable (J24T0788G), Not Applicable (K24T1006G), Not Applicable (M24T1613G), Not Applicable (G34T3205G), Not Applicable (A34T1812G), Not Applicable (A34T1920G), Not Applicable (A34T1923G), Not Applicable (A34T1924G), Not Applicable (A34T1926G), Not Applicable (A44T4574G), Not Applicable (A44T4575G), Not Applicable (A44T4576G), Not Applicable (A44T4584G), Not Applicable (A44T4585G), Not Applicable (A44T4586G), Not Applicable (A44T4590G), Not Applicable (A44T4591G), Not Applicable (A44T4593G), Not Applicable (A44T4597G), Not Applicable (A44T4599G), Not Applicable (A44T4604G), Not Applicable (A44T4605G), Not Applicable (A44T4606G), Not Applicable (A44T4607G), Not Applicable (A44T4609G), Not Applicable (A44T4610G), Not Applicable (A44T4611G), Not Applicable (A44T4612G), Not Applicable (A44T4615G), Not Applicable (A44T4616G), Not Applicable (A44T4617G), Not Applicable (A44T4618G), Not Applicable (A44T4619G), Not Applicable (A44T4621G), Not Applicable (A44T4622G), Not Applicable (A44T4630G), Not Applicable (A44T4632G), Not Applicable (A44T4636G), Not Applicable (A44T4641G), Not Applicable (A44T4646G), Not Applicable (A44T4647G), Not Applicable (A44T4648G), Not Applicable (A44T4649G), Not Applicable (A44T4650G), Not Applicable (A44T4652G), Not Applicable (A44T4653G), Not Applicable (A44T4654G), Not Applicable (A44T4655G), Not Applicable (A44T4658G), Not Applicable (A44T4659G), Not Applicable (A44T4660G), Not Applicable (A44T4661G), Not Applicable (A44T4669G), Not Applicable (A44T4672G), Not Applicable (A44T4675G), Not Applicable (A44T4677G), Not Applicable (A44T4718G), Not Applicable (A44T4726G), Not Applicable (A44T4728G), Not Applicable (B34T1933G), Not Applicable (B44T4780G), Not Applicable (D34T2501G), Not Applicable (D34T2505G), Not Applicable (D34T2514G), Not Applicable (D34T2528G), Not Applicable (E34T2749G), Not Applicable (E34T2753G), Not Applicable (F34T2866G), Not Applicable (F34T2991G), Not Applicable (G24T0510G), Not Applicable (G34T3063G), Not Applicable (G34T3137G), Not Applicable (G34T3204G), Not Applicable (G34T3236G), Not Applicable (H34T3402G), Not Applicable (H34T3403G), Not Applicable (H34T3404G), Not Applicable (H34T3405G), Not Applicable (H34T3406G), Not Applicable (H34T3407G), Not Applicable (H34T3409G), Not Applicable (H34T3410G), Not Applicable (H34T3412G), Not Applicable (H34T3424G), Not Applicable (H34T3475G), Not Applicable (H34T3481G), Not Applicable (H34T3485G), Not Applicable (H34T3490G), Not Applicable (H34T3491G), Not Applicable (J24T0919G), Not Applicable (J24T0924G), Not Applicable (J34T3667G), Not Applicable (J34T3668G), Not Applicable (J34T3705G), Not Applicable (J34T3781G), Not Applicable (J34T3802G), Not Applicable (K24T1255G), Not Applicable (K34T3934G), Not Applicable (K34T3935G), Not Applicable (L34T4035G), Not Applicable (L34T4158G), Not Applicable (L34T4163G), Not Applicable (M34T4262G), Not Applicable (M34T4263G), Not Applicable (M34T4264G), Not Applicable (M34T4288G), Not Applicable (M34T4325G), Not Applicable (M34T4326G), Not Applicable (M34T4327G), Not Applicable (M34T4332G), Not Applicable (M34T4333G), Not Applicable (M34T4335G), Not Applicable (M34T4338G), Not Applicable (M34T4342G), Not Applicable (M34T4344G), Not Applicable (M34T4374G), Not Applicable (M34T4376G), Not Applicable (M34T4379G), Not Applicable (M34T4382G), Not Applicable (M34T4385G), Not Applicable (M34T4386G), Not Applicable (M34T4387G), Not Applicable (M34T4388G), Not Applicable (M34T4391G), Not Applicable (M34T4392G), Not Applicable (M34T4448G), Not Applicable (M34T4451G), Not Applicable (B44T4784G), Not Applicable (G34T3084G), Not Applicable (G34T3086G), Not Applicable (G34T3107G), Not Applicable (G34T3171G), Not Applicable (G34T3185G), Not Applicable (J34T3542G), Not Applicable (E34T2762G), Not Applicable (F34T2826G), Not Applicable (F34T2827G), Not Applicable (F34T2911G), Not Applicable (L24T1306G), Not Applicable (L24T1307G), Not Applicable (M24T1518G), Not Applicable (M34T4227G), Not Applicable (M34T4400G), Not Applicable (M34T4405G), Not Applicable (M34T4467G), Not Applicable (M34T4468G), Not Applicable (M34T4523G), Not Applicable (E34T2656G), Not Applicable (G24T0443G), Not Applicable (G24T0514G), Not Applicable (J34T3738G), Not Applicable (J34T3769G), Not Applicable (K24T1156G), Not Applicable (M34T4206G), Not Applicable (M34T4422G), Not Applicable (M34T4462G), Not Applicable (M34T4606G), Not Applicable (M34T4474G), Not Applicable (E34T2767G), Not Applicable (F34T3031G), Not Applicable (L24T1389G), Not Applicable (M24T1421G), Not Applicable (M24T1424G), Not Applicable (M24T1425G), Not Applicable (M24T1433G), Not Applicable (B34T2170G), Not Applicable (C34T2263G), Not Applicable (C34T2266G), Not Applicable (C34T2296G), Not Applicable (J24T0775G), Not Applicable (K24T1066G), Not Applicable (K24T1217G), Not Applicable (K34T3962G), Not Applicable (K34T3963G), Not Applicable (G24T0546G), Not Applicable (J24T0643G), Not Applicable (J24T0649G), Not Applicable (J24T0650G), Not Applicable (J24T0652G), Not Applicable (K24T1114G), Not Applicable (H34T3287G), Not Applicable (H34T3288G), Not Applicable (H34T3289G), Not Applicable (H34T3294G), Not Applicable (H34T3295G), Not Applicable (H34T3297G), Not Applicable (H34T3317G), Not Applicable (H34T3328G), Not Applicable (K34T3931G), Not Applicable (K34T3976G), Not Applicable (M34T4253G), Not Applicable (M34T4396G), Not Applicable (J34T3522G), Not Applicable (J34T3599G), Not Applicable (J34T3602G), Not Applicable (K24T1235G), Not Applicable (K24T1264G), Not Applicable (K24T1265G), Not Applicable (B44T4916G), Not Applicable (B44T4917G), Not Applicable (B44T4918G), Not Applicable (F34T2905G), Not Applicable (F34T2906G), Not Applicable (K34T3845G), Not Applicable (K34T3846G), Not Applicable (A44T4729G), Not Applicable (L34T4028G), Not Applicable (K34T3982G), Not Applicable (D34T2618G), Not Applicable (K24T1182G), Not Applicable (K34T3957G), Not Applicable (M24T1480G), Not Applicable (D34T2633G), Not Applicable (K34T3825G), Not Applicable (K34T3828G), Not Applicable (L34T4175G), Not Applicable (H34T3255G), Not Applicable (D24T2492G), Not Applicable (D34T2492G), Not Applicable (D34T2510G), Not Applicable (D34T2542J), Not Applicable (D34T2569G), Not Applicable (E34T2659G), Not Applicable (L24T1305G), Not Applicable (M24T1735G), Not Applicable (B44T4894G), Not Applicable (H24T1228G), Not Applicable (K24T1228G), Not Applicable (C44T4965G), Not Applicable (D44T5030G), Not Applicable (A44T4671G), Not Applicable (C34T2417G), Not Applicable (A44T4699G), Not Applicable (B44T4785G), Not Applicable (E34T2754G), Not Applicable (F34T2929G), Not Applicable (D34T2543G), Not Applicable (C34T2225G), Not Applicable (H34T3460G), Not Applicable (B34T2098G), Not Applicable (A34T1889G), Not Applicable (M34T4350G), Not Applicable (D34T2486G), Not Applicable (D34T2598G), Not Applicable (C44T4940G), Not Applicable (C44T4982G), Not Applicable (D44T5034G), Not Applicable (F34T2900G), Not Applicable (J24T0743G), Not Applicable (F24T0093G), Not Applicable (F24T0094G), Not Applicable (F24T0095G), Not Applicable (F24T0109G), Not Applicable (F24T0114G), Not Applicable (F24T0115G), Not Applicable (G24T0403G), Not Applicable (G24T0409G), Not Applicable (G24T0419G), Not Applicable (G24T0539G), Not Applicable (H24T0591G), Not Applicable (H24T0592G), Not Applicable (H24T0593G), Not Applicable (J24T0604G), Not Applicable (L24T1338G), Not Applicable (L24T1344G), Not Applicable (L34T4092G), Not Applicable (M34T4343G), Not Applicable (F34T3023G), Not Applicable (J34T3531G), Not Applicable (K24T1092G), Not Applicable (K34T3853G), Not Applicable (K34T3856G), Not Applicable (K34T3857G), Not Applicable (M34T4259G), Not Applicable (L24T1357G), Not Applicable (A34T1786G), Not Applicable (A34T1855G), Not Applicable (A34T1908G), Not Applicable (A34T1909G), Not Applicable (A44T4560G), Not Applicable (B34T1948G), Not Applicable (B34T1954G), Not Applicable (B34T1976G), Not Applicable (C34T2232G), Not Applicable (J24T0943G), Not Applicable (A44T4620G), Not Applicable (A44T4644G), Not Applicable (A34T1772G), Not Applicable (A34T1773G), Not Applicable (J34T3733G), Not Applicable (J34T3734G), Not Applicable (J34T3735G), Not Applicable (A44T4731G), Not Applicable (A44T4733G), Not Applicable (A44T4734G), Not Applicable (A44T4736G), Not Applicable (A44T4738G), Not Applicable (B44T4767G), Not Applicable (B44T4830G), Not Applicable (H34T3415G), Not Applicable (H34T3418G), Not Applicable (M34T4509G), Not Applicable (B44T4775G), Not Applicable (A44T4721G), Not Applicable (K24T1230G), Not Applicable (G24T0561G), Not Applicable (B34T1995G), Not Applicable (D34T2467G), Not Applicable (D34T2533G), Not Applicable (D34T2550G), Not Applicable (D34T2551G), Not Applicable (K34T3854G), Not Applicable (B44T4794G), Not Applicable (C34T2230G), Not Applicable (B44T4893G), Not Applicable (B44T4766G), Not Applicable (B44T4835G), Not Applicable (B44T4819G), Not Applicable (D34T2446G), Not Applicable (K34T3850G), Not Applicable (K34T3893G), Not Applicable (L34T4067G), Not Applicable (B44T4802G), Not Applicable (A34T1856G), Not Applicable (B34T1975G), Not Applicable (B44T4837G), Not Applicable (L34T4002G), Not Applicable (L34T4006G), Not Applicable (L34T4007G), Not Applicable (L34T4017G), Not Applicable (L34T4022G), Not Applicable (L34T4149G), Not Applicable (L34T4152G), Not Applicable (J24T0793G), Not Applicable (B34T2063G), Not Applicable (B34T2193G), Not Applicable (C44T4963G), Not Applicable (C44T4964G), Not Applicable (G24T0363G), Not Applicable (G24T0364G), Not Applicable (G24T0365G), Not Applicable (G24T0366G), Not Applicable (G24T0396G), Not Applicable (G24T0482G), Not Applicable (G24T0483G), Not Applicable (G24T0486G), Not Applicable (G24T0492G), Not Applicable (G24T0517G), Not Applicable (G24T0522G), Not Applicable (H24T0595G), Not Applicable (J24T0726G), Not Applicable (J24T0783G), Not Applicable (J24T0784G), Not Applicable (K24T1001G), Not Applicable (K24T1041G), Not Applicable (K24T1068G), Not Applicable (K24T1088G), Not Applicable (K24T1185G), Not Applicable (K34T3875G), Not Applicable (K34T3878G), Not Applicable (L34T4155G), Not Applicable (D34T2593G), Not Applicable (L34T3996G), Not Applicable (L34T3998G), Not Applicable (E44T5052G), Not Applicable (E44T5053G), Not Applicable (E44T5054G), Not Applicable (D44T5042G), Not Applicable (G34T3164G), Not Applicable (K24T1013G), Not Applicable (K24T1022G), Not Applicable (K24T1023G), Not Applicable (A34T1798G), Not Applicable (L24T1331G), Not Applicable (L24T1334G), Not Applicable (L24T1335G), Not Applicable (L24T1349G), Not Applicable (A34T1875G), Not Applicable (C34T2309G), Not Applicable (F34T2984G), Not Applicable (G24T0389G), Not Applicable (G24T0444G), Not Applicable (G24T0477G), Not Applicable (H34T3500G), Not Applicable (J34T3623G), Not Applicable (M24T1545G), Not Applicable (M24T1544G), Not Applicable (M24T1407G), Not Applicable (B34T2124G), Not Applicable (C44T4993G), Not Applicable (H34T3330G), Not Applicable (B44T4821G), Not Applicable (C34T2224G), Not Applicable (C34T2356G), Not Applicable (D34T2568G), Not Applicable (E34T2661G), Not Applicable (E44T5051G), Not Applicable (F24T0160G), Not Applicable (F24T0166G), Not Applicable (F24T0167G), Not Applicable (F24T0170G), Not Applicable (F24T0186G), Not Applicable (F24T0191G), Not Applicable (F24T0193G), Not Applicable (F24T0194G), Not Applicable (F24T0205G), Not Applicable (F24T0206G), Not Applicable (F44T5065G), Not Applicable (F44T5066G), Not Applicable (G24T0584G), Not Applicable (G34T3192G), Not Applicable (G34T3193G), Not Applicable (G34T3194G), Not Applicable (G34T3195G), Not Applicable (G34T3196G), Not Applicable (G34T3197G), Not Applicable (G34T3198G), Not Applicable (G34T3199G), Not Applicable (G34T3200G), Not Applicable (J24T0667G), Not Applicable (J34T3737G), Not Applicable (J34T3740G), Not Applicable (J34T3748G), Not Applicable (K24T1176G), Not Applicable (L24T1299G), Not Applicable (M24T1612G), Not Applicable (M24T1616G), Not Applicable (M24T1617G), Not Applicable (M24T1621G), Not Applicable (M24T1623G), Not Applicable (M24T1624G), Not Applicable (M24T1625G), Not Applicable (M24T1626G), Not Applicable (M24T1628G), Not Applicable (B34T1961G), Not Applicable (C34T2229G), Not Applicable (C44T5003G), Not Applicable (D44T5018G), Not Applicable (J34T3577G), Not Applicable (J34T3587G), Not Applicable (M34T4460G), Not Applicable (M34T4417G), Not Applicable (M34T4499G), Not Applicable (C34T2353G), Not Applicable (C34T2368G), Not Applicable (E34T2687G), Not Applicable (E34T2735G), Not Applicable (E34T2736G), Not Applicable (F44T5072G), Not Applicable (B34T2134G), Not Applicable (E34T2714G), Not Applicable (F24T0154G), Not Applicable (F24T0169G), Not Applicable (B34T2116G), Not Applicable (F24T0174G), Not Applicable (F24T0207G), Not Applicable (F24T0208G), Not Applicable (F24T0209G), Not Applicable (F24T0286G), Not Applicable (J34T3699G), Not Applicable (K24T1244G), Not Applicable (F24T0159G), Not Applicable (F24T0172G), Not Applicable (B34T2130G), Not Applicable (K24T1150G), Not Applicable (M24T1516G), Not Applicable (F24T0195G), Not Applicable (F24T0192G), Not Applicable (C34T2227G), Not Applicable (C44T4995G), Not Applicable (E34T2769G), Not Applicable (F24T0021G), Not Applicable (F44T5074G), Not Applicable (J24T0728G), Not Applicable (J24T0730G), Not Applicable (K24T1169G), Not Applicable (F24T0162G), Not Applicable (M34T4182G), Not Applicable (M24T1473G), Not Applicable (F24T0190G), Not Applicable (J24T0731G), Not Applicable (F24T0148G), Not Applicable (L24T1329G), Not Applicable (M24T1689G), Not Applicable (J24T0729G), Not Applicable (K24T1162G), Not Applicable (M24T1469G), Not Applicable (L24T1291G), Not Applicable (L24T1292G), Not Applicable (G24T0323G), Not Applicable (M24T1622G), Not Applicable (F44T5064G), Not Applicable (F44T5067G), Not Applicable (F44T5068G), Not Applicable (B34T2144G), Not Applicable (F44T5063G), Not Applicable (B44T4820G), Not Applicable (F34T2829G), Not Applicable (F34T3038G), Not Applicable (F44T5073G), Not Applicable (G24T0378G), Not Applicable (E34T2730G), Not Applicable (F34T2822G), Not Applicable (F34T2828G), Not Applicable (J34T3644G), Not Applicable (B34T2147G), Not Applicable (J34T3570G), Not Applicable (F44T5075G), Not Applicable (J34T3523G), Not Applicable (J34T3524G), Not Applicable (J34T3624G), Not Applicable (J34T3646G), Not Applicable (C34T2319G), Not Applicable (F44T5061G), Not Applicable (F44T5062G), Not Applicable (J34T3751G), Not Applicable (M34T4399G), Not Applicable (M34T4436G), Not Applicable (M34T4437G), Not Applicable (M34T4438G), Not Applicable (M34T4439G), Not Applicable (M34T4397G), Not Applicable (E34T2770G), Not Applicable (F24T0064G), Not Applicable (M24T1604G), Not Applicable (K24T1259G), Not Applicable (M34T4257G), Not Applicable (J24T0746G), Not Applicable (F24T0187G), Not Applicable (M24T1620G), Not Applicable (C34T2286G), Not Applicable (G24T0372G), Not Applicable (G24T0440G), Not Applicable (C34T2287G), Not Applicable (C34T2293G), Not Applicable (J34T3676G), Not Applicable (J34T3745G), Not Applicable (F24T0184G), Not Applicable (F24T0185G), Not Applicable (J34T3680G), Not Applicable (J24T0724G), Not Applicable (C34T2294G), Not Applicable (B34T2160G), Not Applicable (C44T4996G), Not Applicable (F24T0163G), Not Applicable (F24T0168G), Not Applicable (J34T3742G), Not Applicable (M24T1619G), Not Applicable (M24T1627G), Not Applicable (C34T2317G), Not Applicable (F24T0158G), Not Applicable (F24T0226G), Not Applicable (H34T3290G), Not Applicable (H34T3291G), Not Applicable (J34T3609G), Not Applicable (J34T3610G), Not Applicable (J34T3731G), Not Applicable (C44T4985G), Not Applicable (F24T0070G), Not Applicable (F24T0165G), Not Applicable (H34T3298G), Not Applicable (F24T0157G), Not Applicable (B34T2117G), Not Applicable (B34T2123G), Not Applicable (B34T2146G), Not Applicable (B34T2148G), Not Applicable (B34T2157G), Not Applicable (B44T4887G), Not Applicable (B44T4890G), Not Applicable (C34T2316G), Not Applicable (C44T4999G), Not Applicable (C44T5000G), Not Applicable (C44T5002G), Not Applicable (F24T0171G), Not Applicable (F24T0173G), Not Applicable (J34T3640G), Not Applicable (J34T3641G), Not Applicable (J34T3645G), Not Applicable (F44T5060G), Not Applicable (F44T5070G), Not Applicable (A34T1794G), Not Applicable (F44T5071G), Not Applicable (F24T0164G), Not Applicable (G44T5077G), Not Applicable (J24T0758G), Not Applicable (M24T1527G), Not Applicable (F24T0188G), Not Applicable (J24T0608G), Not Applicable (J24T0662G), Not Applicable (B44T4787G), Not Applicable (J34T3607G), Not Applicable (J34T3743G), Not Applicable (M24T1591G), Not Applicable (M34T4298G), Not Applicable (C34T2238G), Not Applicable (C44T4994G), Not Applicable (M34T4299G), Not Applicable (C44T4948G), Not Applicable (E44T5058G), Not Applicable (K24T1168G), Not Applicable (B34T2168G), Not Applicable (C34T2343G), Not Applicable (F24T0161G), Not Applicable (J34T3617G), Not Applicable (F44T5069G), Not Applicable (B34T2162G), Not Applicable (C34T2310G), Not Applicable (B34T2142G), Not Applicable (B34T2119G), Not Applicable (B34T2156G), Not Applicable (C34T2335G), Not Applicable (C44T5001G), Not Applicable (J34T3642G), Not Applicable (C34T2311G), Not Applicable (C34T2313G), Not Applicable (J34T3639G), Not Applicable (J34T3643G), Not Applicable (B44T4884G), Not Applicable (C44T5006G), Not Applicable (J34T3637G), Not Applicable (B34T2143G), Not Applicable (C34T2338G), Not Applicable (B34T2149G), Not Applicable (C34T2320G), Not Applicable (C44T4986G), Not Applicable (B34T2145G), Not Applicable (C34T2314G), Not Applicable (J34T3638G), Not Applicable (B44T4829G), Not Applicable (J24T0911G), Not Applicable (J24T0913G), Not Applicable (K24T1237G), Not Applicable (G24T0455G), Not Applicable (E34T2734G), Not Applicable (G24T0530G), Not Applicable (A34T1761G), Not Applicable (A34T1777G), Not Applicable (A34T1793G), Not Applicable (A34T1918G), Not Applicable (F34T2919G), Not Applicable (F34T2921G), Not Applicable (A34T1796G), Not Applicable (A34T1878G), Not Applicable (A44T4642G), Not Applicable (F34T2884G), Not Applicable (F34T3028G), Not Applicable (H34T3469G), Not Applicable (J34T3723G), Not Applicable (J34T3724G), Not Applicable (J34T3729G), Not Applicable (J34T3730G), Not Applicable (A34T1862G), Not Applicable (A34T1782G), Not Applicable (A34T1795G), Not Applicable (B34T2077G), Not Applicable (E34T2683G), Not Applicable (F34T2913G), Not Applicable (F34T2920G), Not Applicable (F34T2923G), Not Applicable (G24T0433G), Not Applicable (K24T1163G), Not Applicable (G34T3125G), Not Applicable (G24T0430G), Not Applicable (K34T3389G), Not Applicable (J24T0912G), Not Applicable (F24T0212G), Not Applicable (J24T0719G), Not Applicable (B44T4915G), Not Applicable (J24T0878G), Not Applicable (D44T5016G), Not Applicable (H34T3494G), Not Applicable (J34T3615G), Not Applicable (J34T3616G), Not Applicable (J34T3618G), Not Applicable (M34T4184G), Not Applicable (J24T0666G), Not Applicable (G24T0585G), Not Applicable (H34T3381G), Not Applicable (L24T1342G), Not Applicable (M24T1631G), Not Applicable (A34T1887G), Not Applicable (E34T2676G), Not Applicable (L24T1332G), Not Applicable (A24T1026G), Not Applicable (A34T1834G), Not Applicable (A34T1835G), Not Applicable (A34T1836G), Not Applicable (A34T1837G), Not Applicable (K24T1026G), Not Applicable (K24T1027G), Not Applicable (G24T0533G), Not Applicable (K24T0553G), Not Applicable (K24T0984G), Not Applicable (A34T1797G), Not Applicable (M34T4179G), Not Applicable (A34T1849G), Not Applicable (A34T1852G), Not Applicable (G34T3059G), Not Applicable (D34T2426G), Not Applicable (A34T1838G), Not Applicable (B34T2999G), Not Applicable (J34T4405G), Not Applicable (F34T2859G), Not Applicable (H34T3353G), Not Applicable (K34T3945G), Not Applicable (G24T0541G), Not Applicable (M34T4258G), Not Applicable (H34T3385G), Not Applicable (D44T0531G), Not Applicable (D44T5031G), Not Applicable (C44T4936G), Not Applicable (G44T4936G), Not Applicable (A34T1876G), Not Applicable (D44T5044G), Not Applicable (A34T1869G), Not Applicable (B34T1869G), Not Applicable (B34T1940G), Not Applicable (B34T194OG), Not Applicable (G34T3060G), Not Applicable (A34T850G), Not Applicable (F24T0105G), Not Applicable (D34T2425G), Not Applicable (G34T3213G), Not Applicable (K34T3944G), Not Applicable (C44T4988G), Not Applicable (A44T4602G), Not Applicable (A44T4603G), Not Applicable (C44T4928G), Not Applicable (A34T1885G), Not Applicable (G34T3220G), Not Applicable (A34T1771G), Not Applicable (A34T1851G), Not Applicable (A44T4578G), Not Applicable (C34T2420G), Not Applicable (E34T2681G), Not Applicable (F34T2800G), Not Applicable (F34T2801G), Not Applicable (H34T3336G), Not Applicable (K34T3947G), Not Applicable (L24T1336G), Not Applicable (L24T1339G), Not Applicable (M24T1465G), Not Applicable (B44T4878G), Not Applicable (C44T4975G), Not Applicable (G24T03756), Not Applicable (G24T0375G), Not Applicable (K24T1025), Not Applicable (K24T1025G), Not Applicable (H34T3352G), Not Applicable (K24T1018G), Not Applicable (A34T1799), Not Applicable (A34T1799G), Not Applicable (A34T1858G), Not Applicable (G34T3219G), Not Applicable (K34T3847G), Not Applicable (K34T8347G), Not Applicable (L24T1333G), Not Applicable (L24T1347G), Not Applicable (B44T4777G), Not Applicable (C34T2419G), Not Applicable (C34T2449G), Not Applicable (C44T4927G), Not Applicable (L24T1341), Not Applicable (L24T1341G), Not Applicable (G34T3214G), Not Applicable (G34T3212G), Not Applicable (A34T1847G), Not Applicable (B44T4776G), Not Applicable (G24T0337G), Not Applicable (L24T1337G), Not Applicable (L24T1346G), Not Applicable (M34T4398G), Not Applicable (E34T2686G), Not Applicable (G34T3221G), Not Applicable (G34T3222G), Not Applicable (H34T3386G), Not Applicable (L24T1343G), Not Applicable (M34T4503G), Not Applicable (K24T0995G), Not Applicable (A34T3354G), Not Applicable (A34T3355G), Not Applicable (H34T3354G), Not Applicable (H34T3355G), Not Applicable (F24T0098G), Not Applicable (A34T1850G), Not Applicable (A44T4600G), Not Applicable (L34T4094G), Not Applicable (A44T4577G), Not Applicable (C44T4991G), Not Applicable (C44T4992G), Not Applicable (C44T4976G), Not Applicable (G24T0452G), Not Applicable (H24T0597G), Not Applicable (F34T2907G), Not Applicable (K24T1172G), Not Applicable (J24T0672G), Not Applicable (J24T0673G), Not Applicable (L24T1372G), Not Applicable (J34T3557G), Not Applicable (C34T2285G), Not Applicable (G24T0432G), Not Applicable (F34T2803G), Not Applicable (C44T4950G), Not Applicable (G34T3218G), Not Applicable (E34T2679G), Not Applicable (H34T3337G), Not Applicable (D34T2453G), Not Applicable (G34T3216G), Not Applicable (E34T2670G), Not Applicable (J24T0720G), Not Applicable (C34T2276G), Not Applicable (C34T2277G), Not Applicable (B44T4827G), Not Applicable (H34T3338G), Not Applicable (G24T0450G), Not Applicable (G24T0451G), Not Applicable (J24T0745G), Not Applicable (D34T2526G), Not Applicable (J34T3622G), Not Applicable (J34T3647G), Not Applicable (D34T2423G), Not Applicable (K24T1205G), Not Applicable (M34T4192G), Not Applicable (J24T0744G), Not Applicable (K24T1002G), Not Applicable (C44T4926G), Not Applicable (C44T4935G), Not Applicable (C44T4967G), Not Applicable (C44T4968G), Not Applicable (C44T4969G), Not Applicable (C44T4970G), Not Applicable (C44T4971G), Not Applicable (C44T4972G), Not Applicable (E34T2671G), Not Applicable (G44T4968G), Not Applicable (F24T0089G), Not Applicable (F24T0090G), Not Applicable (F24T0091G), Not Applicable (H24T0601G), Not Applicable (A34T1801G), Not Applicable (M24T1463G), Not Applicable (M34T4294G), Not Applicable (G34T3057G), Not Applicable (A34T1868G), Not Applicable (D34T2452G), Not Applicable (F34T2804G), Not Applicable (K24T1070G), Not Applicable (A44T4570G), Not Applicable (D34T2627G), Not Applicable (A44T4572G), Not Applicable (K24T1020G), Not Applicable (G24T0436G), Not Applicable (G24T0437G), Not Applicable (A44T4569G), Not Applicable (B44T4798G), Not Applicable (G24T0329G), Not Applicable (D44T5035G), Not Applicable (D44T5043G), Not Applicable (G34T3217G), Not Applicable (C34T2222G), Not Applicable (F34T2901G), Not Applicable (G34T3096G), Not Applicable (A34T1890G), Not Applicable (H24T0596G), Not Applicable (C34T2323G), Not Applicable (D34T2502G), Not Applicable (A34T1775G), Not Applicable (A34T1871G), Not Applicable (D34T2646G), Not Applicable (M34T4290G), Not Applicable (A34T1802G), Not Applicable (C44T4990G), Not Applicable (B44T4828G), Not Applicable (B44T4874G), Not Applicable (B44T4875G), Not Applicable (B44T4877G), Not Applicable (G34T3165G), Not Applicable (L34T4093G), Not Applicable (J34T3761G), Not Applicable (F34T2897G), Not Applicable (F34T2898G), Not Applicable (F34T2899G), Not Applicable (F34T2908G), Not Applicable (K24T1071G), Not Applicable (A44T4722G), Not Applicable (C34T3388G), Not Applicable (C34T3609G), Not Applicable (C34T3622G), Not Applicable (D44T5047G), Not Applicable (D44T5048G), Not Applicable (D44T5049G), Not Applicable (D44T5050G), Not Applicable (M34T4180G), Not Applicable (M34T4292G), Not Applicable (M34T4293G), Not Applicable (J34T3501G), Not Applicable (A34T1821G), Not Applicable (B34T2986G), Not Applicable (F34T2802G), Not Applicable (K24T1170G), Not Applicable (K24T1171G), Not Applicable (K24T1206G), Not Applicable (L24T1371G), Not Applicable (M34T4291G), Not Applicable (M24T1546G), Not Applicable (E34T2678G), Not Applicable (M34T4277G), Not Applicable (G34T3166G), Not Applicable (G34T3191G), Not Applicable (G24T0327G), Not Applicable (G24T0532G), Not Applicable (K34T3943G), Not Applicable (E44T5055G), Not Applicable (E44T5056G), Not Applicable (B44T4876G), Not Applicable (M34T4181G), Not Applicable (B34T2182G), Not Applicable (A44T4571G), Not Applicable (H34T3387G), Not Applicable (D34T2432G), Not Applicable (B34T2181G), Not Applicable (A44T4549G), Not Applicable (A34T1800G), Not Applicable (D34T2548G), Not Applicable (H34T3474G), Not Applicable (F34T2805G), Not Applicable (F24T0092G), Not Applicable (M34T4188G), Not Applicable (M34T4189G), Not Applicable (M34T4190G), Not Applicable (M34T4217G), Not Applicable (G24T0431G), Not Applicable (M24T1408G), Not Applicable (M34T4218G), Not Applicable (G34T3097G), Not Applicable (C34T2410G), Not Applicable (M34T4187G), Not Applicable (M34T4191G), Not Applicable (B34T1958G), Not Applicable (B34T2102G), Not Applicable (B34T2103G), Not Applicable (B34T2140G), Not Applicable (F24T0101G), Not Applicable (F24T0102G), Not Applicable (G34T3168G), Not Applicable (G34T3170G), Not Applicable (H34T3260G), Not Applicable (H34T3278G), Not Applicable (H34T3281G), Not Applicable (H34T3311G), Not Applicable (J24T0795G), Not Applicable (J34T3591G), Not Applicable (K24T1058G), Not Applicable (M24T1652G), Not Applicable (M34T4421G), Not Applicable (D34T2448G), Not Applicable (D34T2456G), Not Applicable (G34T3215G), Not Applicable (L34T4113G), Not Applicable (L34T4118G), Not Applicable (K34T3925G), Not Applicable (A44T4700G), Not Applicable (A44T4708G), Not Applicable (A44T4709G), Not Applicable (F24T0213G), Not Applicable (K24T1050G), Not Applicable (G24T0540G), Not Applicable (H34T3383G), Not Applicable (H34T3384G), Not Applicable (H34T3462G), Not Applicable (J34T3773G), Not Applicable (M34T4271G), Not Applicable (M34T4272G), Not Applicable (C44T4980G), Not Applicable (F24T0136G), Not Applicable (F24T0137G), Not Applicable (F24T0138G), Not Applicable (F24T0139G), Not Applicable (F24T0140G), Not Applicable (F24T0176G), Not Applicable (F24T0177G), Not Applicable (F24T0179G), Not Applicable (F24T0180G), Not Applicable (F24T0181G), Not Applicable (J24T0688G), Not Applicable (F24T0153G), Not Applicable (A44T4712G), Not Applicable (G24T0427G), Not Applicable (G24T0439G), Not Applicable (J34T3573G), Not Applicable (E34T2712G), Not Applicable (F34T2914G), Not Applicable (B34T1996G), Not Applicable (F24T0052G), Not Applicable (M24T1522G), Not Applicable (F24T0127G), Not Applicable (F24T0201G), Not Applicable (M24T1529G), Not Applicable (F24T0107G), Not Applicable (K34T3930G), Not Applicable (B34T2164G), Not Applicable (B34T2165G), Not Applicable (B34T2166G), Not Applicable (B34T2167G), Not Applicable (B34T2169G), Not Applicable (C34T2236G), Not Applicable (F24T0197G), Not Applicable (C34T2308G), Not Applicable (M34T4185G), Not Applicable (M34T4273G), Not Applicable (J34T3619G), Not Applicable (M34T4183G), Not Applicable (L34T4091G), Not Applicable (M24T1659G), Not Applicable (M34T4274G), Not Applicable (M24T1629G), Not Applicable (C34T2322G), Not Applicable (G24T0383G), Not Applicable (G24T0550G), Not Applicable (C34T2418G), Not Applicable (D34T2422G), Not Applicable (A34T1886G), Not Applicable (M24T1467G), Not Applicable (C44T4961G), Not Applicable (A34T1863G), Not Applicable (M24T1406G), Not Applicable (E34T2672G), Not Applicable (E34T2673G), Not Applicable (E34T2674G), Not Applicable (E34T2675G), Not Applicable (E34T2677G), Not Applicable (G34T3056G), Not Applicable (C34T2258G), Not Applicable (M24T1547G), Not Applicable (G34T3223G), Not Applicable (M24T1409G), Not Applicable (M34T4219G), Not Applicable (G34T3225G), Not Applicable (G34T3098G), Not Applicable (M34T4220G), Not Applicable (M24T1548G), Not Applicable (D34T2421G), Not Applicable (G34T3224G), Not Applicable (K24T1005G), Not Applicable (B44T4789G), Not Applicable (B44T4790G), Not Applicable (B44T4791G), Not Applicable (B44T4792G), Not Applicable (B44T4793G), Not Applicable (J24T0674G), Not Applicable (D34T2424G), Not Applicable (A34T1844G), Not Applicable (C44T4989G), Not Applicable (M34T4275G), Not Applicable (D34T2607G), Not Applicable (M24T1462G), Not Applicable (M34T4186G), Not Applicable (L34T4095G), Not Applicable (C34T2212G), Not Applicable (M34T4178G), Not Applicable (A34T1881G), Not Applicable (J44T5093G), Not Applicable (F34T2850G), Not Applicable (C34T2248G), Not Applicable (C34T2249G), Not Applicable (E34T2657G), Not Applicable (L24T1385G), Not Applicable (L24T1405G), Not Applicable (C34T2307G), Not Applicable (H44T5080G), Not Applicable (H44T5081G), Not Applicable (G34T3163G), Not Applicable (H44T5078G), Not Applicable (L44T5100G), Not Applicable (L44T5101G), Not Applicable (M24T1492G), Not Applicable (B34T2159G), Not Applicable (K34T3995G), Not Applicable (L44T5096G), Not Applicable (J44T5085G), Not Applicable (L24T1398G), Not Applicable (D34T2645G), Not Applicable (G34T3110G), Not Applicable (B34T2183G), Not Applicable (J44T5088G), Not Applicable (J44T5090G), Not Applicable (J44T5092G), Not Applicable (L44T5095G), Not Applicable (M24T1493G), Not Applicable (M24T1494G), Not Applicable (F34T2903G), Not Applicable (H44T5079G), Not Applicable (M34T4354G), Not Applicable (H34T3293G), Not Applicable (L34T4034G), Not Applicable (F34T2878G), Not Applicable (B34T1988G), Not Applicable (B44T4815G), Not Applicable (E34T2662G), Not Applicable (D34T2450G), Not Applicable (G24T0535G), Not Applicable (L24T1323G), Not Applicable (L24T1324G), Not Applicable (M34T4244G), Not Applicable (J34T3533G), Not Applicable (J34T3509G), Not Applicable (M34T4230G), Not Applicable (B34T1997G), Not Applicable (C34T2415G), Not Applicable (C44T4966G), Not Applicable (D34T2525G), Not Applicable (E34T2768G), Not Applicable (E34T2771G), Not Applicable (F24T0049G), Not Applicable (F24T0050G), Not Applicable (F24T0118G), Not Applicable (F24T0119G), Not Applicable (F24T0155G), Not Applicable (F34T2869G), Not Applicable (F34T2875G), Not Applicable (F34T2876G), Not Applicable (H34T3320G), Not Applicable (K24T0979G), Not Applicable (K34T3929G), Not Applicable (L24T1276G), Not Applicable (M34T4224G), Not Applicable (J24T0689G), Not Applicable (K34T3952G), Not Applicable (E34T2785G), Not Applicable (C34T2359G), Not Applicable (C34T2390G), Not Applicable (C34T2400G), Not Applicable (C34T2402G), Not Applicable (C34T2403G), Not Applicable (C34T2405G), Not Applicable (F34T2821G), Not Applicable (F34T2823G), Not Applicable (F34T3005G), Not Applicable (J24T0839G), Not Applicable (J24T0877G), Not Applicable (J24T0888G), Not Applicable (J24T0889G), Not Applicable (J24T0890G), Not Applicable (J34T3571G), Not Applicable (J34T3593G), Not Applicable (M34T4359G), Not Applicable (F34T2997G), Not Applicable (F34T3001G), Not Applicable (H34T3258G), 00840682109260M24T1559G (M24T1559G), Not Applicable (B34T2128G), Not Applicable (B44T4910G), Not Applicable (J34T3588G), Not Applicable (J34T3612G), Not Applicable (B34T2120G), Not Applicable (B44T4879G), Not Applicable (B44T4881G), Not Applicable (F24T0065G), Not Applicable (F34T2820G), Not Applicable (J34T3697G), Not Applicable (K34T3863G), Not Applicable (K34T3983G), Not Applicable (G24T0398G), Not Applicable (G24T0429G), Not Applicable (F24T0044G), Not Applicable (J24T0725G), Not Applicable (L34T4032G), Not Applicable (G24T0376G), Not Applicable (L34T4164G), Not Applicable (M34T4463G), Not Applicable (J34T3507G), Not Applicable (K24T1195G), Not Applicable (M24T1422G), Not Applicable (L34T4117G), Not Applicable (K34T3906G), Not Applicable (K34T3907G), Not Applicable (K34T3908G), Not Applicable (K34T3909G), Not Applicable (K34T3910G), Not Applicable (K34T3911G), Not Applicable (K34T3912G), Not Applicable (K34T3938G), Not Applicable (K34T3939G), Not Applicable (M34T4459G), Not Applicable (M34T4516G), Not Applicable (J24T0842G), Not Applicable (J24T0843G), Not Applicable (J24T0852G), Not Applicable (B34T2177G), Not Applicable (B34T2178G), Not Applicable (B34T2179G), Not Applicable (C44T4929G), Not Applicable (C44T4942G), Not Applicable (M34T4297G), Not Applicable (A34T1859G), Not Applicable (K24T1059G), Not Applicable (K24T1089G), Not Applicable (K24T1094G), Not Applicable (K24T1095G), Not Applicable (K24T1096G), Not Applicable (K24T1102G), Not Applicable (K24T1178G), Not Applicable (K24T1187G), Not Applicable (K34T3818G), Not Applicable (K34T3819G), Not Applicable (L24T1326G), Not Applicable (D44T5046G), Not Applicable (D34T2640G), Not Applicable (D34T2641G), Not Applicable (D34T2642G), Not Applicable (M24T1729G), Not Applicable (M24T1730G), Not Applicable (M24T1731G), Not Applicable (M24T1732G), Not Applicable (M24T1733G), Not Applicable (M24T1734G), Not Applicable (K24T1240G), Not Applicable (K24T1241G), Not Applicable (F24T0096G), Not Applicable (B34T2172G), Not Applicable (B34T2173G), Not Applicable (J34T3621G), Not Applicable (J34T3648G), Not Applicable (A34T1780G), Not Applicable (A34T1860G), Not Applicable (A34T1888G), Not Applicable (B34T2127G), Not Applicable (C34T2289G), Not Applicable (C34T2303G), Not Applicable (H34T3254G), Not Applicable (H34T3315G), Not Applicable (H34T3316G), Not Applicable (H34T3414G), Not Applicable (J24T0806G), Not Applicable (J24T0807G), Not Applicable (J34T3511G), Not Applicable (M34T4249G), Not Applicable (M34T4266G), Not Applicable (K24T1213G), Not Applicable (A44T4614G), Not Applicable (G34T3167G), Not Applicable (C34T2350G), Not Applicable (C44T4987G), Not Applicable (A34T1916G), Not Applicable (H34T3375G), Not Applicable (H34T3378G), Not Applicable (J34T3547G), Not Applicable (J34T3548G), Not Applicable (J34T3549G), Not Applicable (J34T3550G), Not Applicable (J34T3551G), Not Applicable (J34T3552G), Not Applicable (J34T3553G), Not Applicable (J34T3554G), Not Applicable (J34T3555G), Not Applicable (J34T3556G), Not Applicable (M34T4312G), Not Applicable (M34T4497G), Not Applicable (M34T4498G), Not Applicable (M34T4501G), Not Applicable (A34T1882G), Not Applicable (L34T4122G), Not Applicable (F34T2932G), Not Applicable (M34T4226G), Not Applicable (F34T2930G), Not Applicable (F34T2933G), Not Applicable (F34T2935G), Not Applicable (F34T2941G), Not Applicable (J34T3804G), Not Applicable (K34T3874G), Not Applicable (K34T3883G), Not Applicable (L34T4110G), Not Applicable (L34T4112G), Not Applicable (M24T1445G), Not Applicable (D34T2490G), Not Applicable (D34T2553G), Not Applicable (D34T2639G), Not Applicable (K34T3862G), Not Applicable (K34T3868G), Not Applicable (C34T2245G), Not Applicable (C34T2246G), Not Applicable (H34T3286G), Not Applicable (C44T4997G), Not Applicable (F24T0046G), Not Applicable (F24T0056G), Not Applicable (F24T0063G), Not Applicable (F24T0112G), Not Applicable (F34T2879G), Not Applicable (G24T0503G), Not Applicable (G24T0571G), Not Applicable (H34T3279G), Not Applicable (J24T0633G), Not Applicable (J24T0654G), Not Applicable (J24T0655G), Not Applicable (J24T0659G), Not Applicable (J24T0947G), Not Applicable (K24T1199G), Not Applicable (K34T3941G), Not Applicable (L24T1302G), Not Applicable (K24T1051G), Not Applicable (K24T1061G), Not Applicable (D34T2471G), Not Applicable (J24T0791G), Not Applicable (B44T4908G), Not Applicable (J24T0658G), Not Applicable (J24T0660G), Not Applicable (J34T3594G), Not Applicable (J34T3611G), Not Applicable (J34T3614G), Not Applicable (M24T1537G), Not Applicable (M24T1575G), Not Applicable (F34T3002G), Not Applicable (C34T2327G), Not Applicable (F34T2961G), Not Applicable (G24T0441G), Not Applicable (L24T1325G), Not Applicable (B44T4822G), Not Applicable (D34T2447G), Not Applicable (C34T2291G), Not Applicable (A34T1866G), Not Applicable (F34T3041G), Not Applicable (A44T4694G), Not Applicable (F34T2915G), Not Applicable (F34T2998G), Not Applicable (H34T3459G), Not Applicable (J34T3771G), Not Applicable (J34T3778G), Not Applicable (J34T3779G), Not Applicable (J34T3780G), Not Applicable (J34T3782G), Not Applicable (J24T0668G), Not Applicable (B44T4795G), Not Applicable (F24T0103G), Not Applicable (J24T0657G), Not Applicable (J24T0748G), Not Applicable (B34T2046G), Not Applicable (F24T0198G), Not Applicable (D34T2564G), Not Applicable (G34T3136G), Not Applicable (G24T0422G), Not Applicable (G24T0424G), Not Applicable (G24T0442G), Not Applicable (A34T1769G), Not Applicable (F34T2999G), Not Applicable (F34T3000G), Not Applicable (F34T3006G), Not Applicable (M24T1725G), Not Applicable (M24T1726G), Not Applicable (C44T5004G), Not Applicable (B34T2176G), Not Applicable (B34T2180G), Not Applicable (F24T0099G), Not Applicable (B34T1965G), Not Applicable (C34T2223G), Not Applicable (K24T1060G), Not Applicable (M34T4372G), Not Applicable (H34T3499G), Not Applicable (E34T2680G), Not Applicable (B34T2174G), Not Applicable (L24T1373G), Not Applicable (A44T4743G), Not Applicable (G24T0445G), Not Applicable (G24T0446G), Not Applicable (M24T1543G), Not Applicable (M24T1549G), Not Applicable (C44T5005G), Not Applicable (H34T3335G), Not Applicable (L24T1345G), Not Applicable (A34T1891G), Not Applicable (G24T0405G), Not Applicable (G24T0406G), Not Applicable (G24T0407G), Not Applicable (G24T0408G), Not Applicable (D44T5033G), Not Applicable (E34T2798G), Not Applicable (E34T2799G), Not Applicable (K34T3940G), Not Applicable (K34T3946G), Not Applicable (K34T3949G), Not Applicable (K24T1152G), Not Applicable (A34T1805G), Not Applicable (A34T1820G), Not Applicable (A34T1824G), Not Applicable (B34T2045G), Not Applicable (B44T4920G), Not Applicable (C44T4998G), Not Applicable (D34T2503G), Not Applicable (F34T2860G), Not Applicable (F34T2861G), Not Applicable (F34T2902G), Not Applicable (G24T0328G), Not Applicable (G24T0423G), Not Applicable (G24T0426G), Not Applicable (G24T0435G), Not Applicable (G24T0456G), Not Applicable (G34T3058G), Not Applicable (G34T3190G), Not Applicable (H34T3388G), Not Applicable (J34T3762G), Not Applicable (K34T3948G), Not Applicable (M34T4295G), Not Applicable (F24T0108G), Not Applicable (L34T4097G), Not Applicable (L34T4108G), Not Applicable (L34T4114G), Not Applicable (L34T4116G), Not Applicable (E34T2713G), Not Applicable (F34T2833G), Not Applicable (K24T1272G), Not Applicable (D34T2575G), Not Applicable (K24T1243G), Not Applicable (F34T2836G), Not Applicable (J24T0838G), Not Applicable (L24T1277G), Not Applicable (L24T1278G), Not Applicable (L24T1523G), Not Applicable (J24T0611G), Not Applicable (C34T2408G), Not Applicable (C34T2412G), Not Applicable (D34T2474G), Not Applicable (D34T2557G), Not Applicable (D34T2631G), Not Applicable (E34T2664G), Not Applicable (K34T3872G), Not Applicable (K34T3876G), Not Applicable (K34T3926G), Not Applicable (M34T4231G), Not Applicable (K34T3880G), Not Applicable (A34T1857G), Not Applicable (J24T0810G), Not Applicable (D34T2563G), Not Applicable (E34T2733G), Not Applicable (D34T2601G), Not Applicable (D34T2621G), Not Applicable (D34T2621G1), Not Applicable (E34T2658G), Not Applicable (D34T2498G), Not Applicable (D34T2566G), Not Applicable (D34T2523G), Not Applicable (G24T0326G), Not Applicable (L34T4055G), Not Applicable (G24T0413G), Not Applicable (B34T2068G), Not Applicable (C34T2406G), Not Applicable (D34T2449G), Not Applicable (D34T2571G), Not Applicable (D34T2622G), Not Applicable (J34T3576G), Not Applicable (J34T3583G), Not Applicable (L34T4040G), Not Applicable (D34T2619G), Not Applicable (D34T2620G), Not Applicable (J34T3759G), Not Applicable (B44T4892G), Not Applicable (A34T1774G), Not Applicable (B34T1936G), Not Applicable (G34T3103G), Not Applicable (A34T1823G), Not Applicable (B34T1955G), Not Applicable (K24T1030G), Not Applicable (H34T3468G), Not Applicable (B44T4838G), Not Applicable (B44T4883G), Not Applicable (A34T1818G), Not Applicable (A34T1865G), Not Applicable (A44T4673G), Not Applicable (A34T1767G), Not Applicable (L34T4000G), Not Applicable (M24T1428G), Not Applicable (C24T1046G), Not Applicable (A44T4698G), Not Applicable (B34T2122G), Not Applicable (H34T3333G), Not Applicable (K34T3975G), Not Applicable (M34T4454G), Not Applicable (B34T2083G), Not Applicable (J24T0797G), Not Applicable (F24T0062G), Not Applicable (B44T4897G), Not Applicable (K24T1177G), Not Applicable (A44T4663G), Not Applicable (M24T1594G), Not Applicable (M24T1595G), Not Applicable (A34T1783G), Not Applicable (D44T5024G), Not Applicable (A44T4740G), Not Applicable (B34T2118G), Not Applicable (K34T3953G), Not Applicable (G34T3184G), Not Applicable (K24T1046G), Not Applicable (K24T1157G), Not Applicable (K24T1194G), Not Applicable (K24T1202G), Not Applicable (K24T1204G), Not Applicable (M24T1444G), Not Applicable (F24T0106G), Not Applicable (G34T3133G), Not Applicable (K24T1137G), Not Applicable (C34T2226G), Not Applicable (K24T1164G), Not Applicable (B44T4803G), Not Applicable (A34T1845G), Not Applicable (J24T0940G), Not Applicable (J24T0944G), Not Applicable (B44T4759G), Not Applicable (B44T4772G), Not Applicable (K24T1047G), Not Applicable (K24T1065G), Not Applicable (A34T1867G), Not Applicable (A44T4633G), Not Applicable (A44T4674G), Not Applicable (G34T3169G), Not Applicable (K24T1049G), Not Applicable (K24T1045G), Not Applicable (M24T1474G), Not Applicable (G24T0425G), Not Applicable (K24T1140G), Not Applicable (B34T2125G), Not Applicable (B34T2002G), Not Applicable (B34T2003G), Not Applicable (B34T2942G), Not Applicable (F34T2942G), Not Applicable (K24T1179G), Not Applicable (K24T1201G), Not Applicable (K24T1263G), Not Applicable (L24T1400G), Not Applicable (A44T4680G), Not Applicable (F24T0061G), Not Applicable (B34T1957G), Not Applicable (B34T1969G), Not Applicable (B44T4825G), Not Applicable (A34T1776G), Not Applicable (B44T4786G), Not Applicable (C34T2413G), Not Applicable (B44T4796G), Not Applicable (J34T3630G), Not Applicable (J34T3632G), Not Applicable (B44T4762G), Not Applicable (B44T4771G), Not Applicable (J24T0603G), Not Applicable (M34T4234G), Not Applicable (C34T2297G), Not Applicable (G24T0399G), Not Applicable (B44T4805G), Not Applicable (G34T3100G), Not Applicable (J24T0605G), Not Applicable (J24T0606G);
3. Product Code: GU, No UDI Number, Serial Numbers: K0GU2966G, K0GU2976G, L0GU3003G, L0GU3004G, L0GU3007G, L0GU3008G, L0GU3009G, L0GU3010G, M0GU3113G, M0GU3148G, C1GU3291G, C1GU3292G, D9GU0690G, E9GU0800G, G9GU1036G, G9GU1043G, G9GU1049G, G9GU1050G, G9GU1051G, G9GU1052G, M9GU1713G, M9GU1723G, M9GU1730G, C9GU0649G, H9GU1240G, H9GU1241G, H9GU1242G, J0GU2875G, C0GU2061G, H9GU1096G, H9GU1098G, M0GU3094G, C1GU3331G, G0GU2522G, M0GU3118G, F0GU2452G, F0GU2453G, F0GU2454G, F0GU2456G, F0GU2475G, D9GU0715G, D9GU0716G, D9GU0750G, D9GU0751G, D9GU0752G, K8GU0028G, L0GU3052G, H9GU1102G, H9GU1137G, F9GU0969G, F9GU0970G, H9GU1127G, J9GU1352G, J9GU1353G, C9GU0583G, C9GU0584G, J0GU2839G, J0GU2840G, K8GU0019G, K8GU0020G, K9GU1390G, L8GU0137G, L8GU0180G, L8GU0182G, L8GU0183G, L8GU0184G, L8GU0185G, L8GU0186G, L8GU0196G, L8GU0197G, M8GU0328G, M8GU0329G, M8GU0338G, M8GU0345G, M8GU0346G, M8GU0362G, M9GU1679G, M9GU1680G, E9GU0869G, E9GU0870G, E9GU0875G, E9GU0876G, E9GU0877G, E9GU0878G, H9GU1196G, J9GU1339G, L9GU1539G, K8GU0045G, A9GU0413G, A9GU0414G, E0GU2227G, E0GU2239G, E0GU2242G, E0GU2264G, E0GU2271G, E0GU2272G, E0GU2273G, L0GU3036G, M0GU0095G, M0GU0096G, K0GU2908G, B0GU1918G, B0GU1919G, B1GU3224G, D9GU0719G, D9GU0720G, G9GU1024G, G9GU1025G, G9GU1026G, H9GU1205G, B9GU0529G, C9GU0621G, C9GU0622G, F9GU0963G, F9GU0964G, J0GU2880G, L0GU3026G, E0GU2246G, J0GU2720G, G0GU2485G, B9GU0499G, B9GU0500G, F0GU2362G, F0GU2363G, H9GU1101G, H9GU1151G, M9GU1151G, A1GU0230G, C0GU2003G, E0GU2318G, L9GU1547G, L9GU1548G, L9GU1549G, L9GU1550G, L9GU1551G, L9GU1569G, F9GU0929G, F9GU0930G, A0GU1873G, A0GU1874G, B0GU1981G, E0GU2206G, B1GU3220G, E0GU2285G, A1GU3189G, A9GU0365G, A9GU0366G, A9GU0367G, B1GU3270G, D0GU2128G, E0GU2207G, E0GU2208G, E0GU2209G, E0GU2210G, E0GU2211G, E0GU2212G, E0GU2213G, E0GU2217G, E0GU2218G, E0GU2322G, E0GU2323G, F0GU2401G, H9GU1174G, H9GU1175G, K0GU2925G, L9GU1545G, M9GU1728G, J0GU2742G, J0GU2743G, J0GU2744G, J0GU2745G, H9GU1227G, H9GU1228G, J9GU1328G, J9GU1329G, J9GU1363G, J9GU1369G, K9GU1429G, K9GU1435G, B9GU0451G, E0GU2255G, E0GU2345G, E0GU2346G, J0GU2850G, M8GU0351G, L8GU0133G, L8GU0134G, L8GU0135G, L8GU0136G, C9GU0581G, H9GU1139G, H9GU1157G, H9GU1158G, H9GU1159G, H9GU1160G, H9GU1161G, H9GU1162G, H9GU1163G, H9GU1164G, H9GU1165G, H9GU1166G, M8GU0228G, B0GU1976G, M8GU0359G, E9GU0859G, F9GU0911G, G9GU1047G, H0GU2717G, H0GU2719G, J0GU2721G, J0GU2722G, J0GU2723G, L8GU0170G, E0GU2288G, B0GU1927G, B0GU1936G, B0GU1937G, J0GU2836G, J0GU2837G, M8GU0243G, J0GU2789G, J9GU1350G, J9GU1351G, L9GU1600G, L9GU1601G, L9GU1602G, L9GU1603G, M0GU3155G, M0GU3156G, B9GU0501G, B9GU0502G, B9GU0503G, J9GU1356G, J9GU1362G, A0GU1813G, A0GU1823G, A0GU1824G, A0GU1825G, A0GU1826G, A0GU1827G, J0GU2854G, K0GU2927G, K0GU2928G, K0GU2929G, C9GU0620G, J9GU1315G, K8GU0116G, L8GU0160G, L8GU0161G, B9GU0455G, C9GU0633G, C9GU0634G, C9GU0638G, C9GU0639G, C9GU0646G, D9GU0709G, D9GU0710G, D9GU0711G, D9GU0717G, D9GU0746G, D9GU0747G, D9GU0748G, D9GU0749G, D9GU0758G, E9GU0855G, E9GU0856G, F9GU0899G, F9GU0900G, F9GU0901G, F9GU0902G, F9GU0903G, F9GU0904G, F9GU0905G, F9GU0906G, F9GU0965G, G9GU0646G, H0GU2619G, H0GU2665G, H9GU1090G, H9GU1100G, H9GU1203G, H9GU1204G, J9GU1259G, J9GU1334G, L8GU0227G, A0GU1894G, B9GU0485G, J9GU1371G, L8GU0201G, B9GU0443G, B9GU0527G, C9GU0655G, B9GU0431G, H0GU2695G, H0GU2696G, M8GU0342G, C9GU0663G, K9GU1420G, K9GU1421G, C9GU0566G, C9GU0568G, D9GU0760G, D9GU0761G, J9GU1316G, J9GU1335G, J9GU1336G, D9GU0725G, K9GU1494G, K9GU1495G, K9GU1496G, K9GU1497G, K9GU1498G, K9GU1499G, M8GU0298G, B1GU3242G, B9GU0531G, B9GU0534G, B9GU0535G, C1GU3321G, D9GU0703G, F0GU2402G, F0GU2403G, F0GU2479G, F9GU0944G, F9GU0988G, J0GU2892G, L9GU1511G, M0GU3092G, M0GU3093G, M0GU3103G, M9GU1677G, H9GU1193G, C1GU3313G, C1GU3314G, C9GU0585GR, D0GU2176G, F0GU2355G, F0GU2411G, F0GU2450G, H9GU1169G, H9GU1171G, C0GU2041G, D9GU0790G, K8GU0063G, A0GU1875G, A0GU1876G, A0GU1877G, G9GU1039G, A1GU3177G, A1GU3178G, B1GU3244G, B1GU3245G, H9GU1167G, H9GU1168G, H9GU1176G, H9GU1177G, H9GU1178G, J9GU1380G, L0GU3062G, L0GU3063G, M0GU3096G, M0GU3151G, M0GU3152G, K8GU0052G, M8GU0336G, C1GU3333G, E0GU2338G, L0GU3015G, L0GU3016G, L0GU3017G, L0GU3018G, M9GU1690G, D0GU2135G, G9GU1079G, D0GU2138G, E0GU2247G, E0GU2251G, H0GU2620G, A1GU3200G, K8GU0088G, K9GU1440G, K9GU1441G, L8GU0126G, L8GU0131G, A0GU1769G, A0GU1788G, A0GU1789G, A9GU0411G, B9GU0444G, H0GU2684G, K9GU1455G, M9GU1663G, M9GU1664G, M9GU1665G, K9GU1474G, K9GU1475G, M9GU1735G, M9GU1736G, M9GU1737G, M9GU1738G, M9GU1739G, M9GU1740G, M9GU1741G, J0GU2877G, M0GU3137G, B0GU1942G, L0GU3033G, L0GU3034G, A9GU0008G, A9GU0415G, A9GU0416G, A9GU0417G, B0GU1988G, B9GU0436G, B9GU0437G, B9GU0438G, C9GU0665G, D9GU0738G, H0GU2635G, H0GU2636G, C0GU2084G, C0GU2085G, G9GU1077G, K9GU1436G, D9GU0712G, D9GU0739G, E9GU0739G, A9GU0387G, A9GU0388G, A9GU0405G, A9GU0406G, E0GU2320G, C9GU0625G, C9GU0626G, C9GU0627G, C9GU0628G, H9GU1103G, H9GU1206G, K9GU1476G, K9GU1477G, L9GU1543G, M9GU1636G, J0GU2807G, J0GU2808G, J0GU2841G, J0GU2842G, M8GU0290G, M8GU0291G, M8GU0293G, M8GU0294G, A9GU0420G, B9GU0440G, F9GU0913G, E0GU2316G, B9GU0495G, E9GU0798G, F0GU2391G, F0GU2392G, F9GU0931G, G9GU1003G, G9GU1004G, G9GU1005G, G9GU1082G, H9GU1212G, E0GU2308G, J9GU1330G, K9GU1433G, M9GU1659G, M9GU1695G, A0GU1872G, H0GU2694G, H9GU1099G, K9GU1444G, K9GU1450G, H9GU1105G, F0GU2438G, F0GU2439G, F0GU2442G, F0GU2443G, F0GU2455G, F0GU2457G, A9GU0426G, B9GU0457G, B9GU0458G, B9GU0487G, C9GU0657G, L8GU0172G, L8GU0173G, M8GU0355G, G0GU2569G, G0GU2570G, J0GU2823G, J0GU2824G, A0GU1752G, A0GU1758G, A0GU1759G, A9GU0392G, A9GU0393G, A9GU0394G, A9GU0395G, A9GU0398G, A9GU0401G, B9GU0468G, B9GU0469G, B9GU0496G, B9GU0497G, B9GU0498G, B9GU0506G, B9GU0507G, B9GU0510G, B9GU0511G, B9GU0512G, B9GU0519G, B9GU0536G, B9GU0537G, C9GU0589G, D9GU0787G, D9GU788G, E0GU2286G, E9GU0821G, E9GU0848G, E9GU0849G, E9GU0850G, E9GU0851G, E9GU0852G, E9GU0853G, E9GU0854G, E9GU0857G, E9GU0865G, E9GU0881G, E9GU0882G, F9GU0908G, F9GU0920G, F9GU0921G, F9GU0922G, F9GU0923G, F9GU0925G, F9GU0926G, F9GU0996G, F9GU0997G, G9GU1013G, G9GU1014G, G9GU1015G, G9GU1016G, G9GU1017G, G9GU1019G, G9GU1020G, G9GU1037G, G9GU1061G, G9GU1919G, H9GU1121G, H9GU1122G, H9GU1123G, H9GU1124G, K8GU0048G, L8GU0147G, L8GU0165G, L8GU0211G, L8GU0212G, L8GU0217G, M8GU0244G, M8GU0245G, M8GU0246G, M8GU0247G, M8GU0248G, M8GU0249G, M8GU0250G, M8GU0264G, M8GU0265G, M8GU0266G, M8GU0267G, M8GU0268G, M8GU0269G, M8GU0270G, M8GU0271G, M8GU0272G, M8GU0273G, M8GU0274G, M8GU0276G, M8GU0277G, M8GU0278G, M8GU0279G, M8GU0280G, M8GU0281G, M8GU0285G, M8GU0286G, M8GU0292G, M8GU0295G, M8GU0296G, M8GU0297G, M8GU0299G, M8GU0300G, M8GU0332G, M8GU0347G, B1GU3274G, F0GU2373G, J9GU1368G, K0GU2965G, K9GU1432G, A0GU1878G, B9GU0523G, C1GU3318G, F0GU2357G, F0GU2358G, F0GU2395G, C9GU0015G, C9GU0661G, C9GU0662G, G9GU1028G, J0GU2814G, J0GU2815G, J0GU2817G, L9GU1552G, L9GU1553G, M8GU0321G, A0GU1848G, A0GU1849G, B0GU1940G, M8GU0275G, M0GU3121G, M0GU3122G, M0GU3124G, M0GU3125G, M0GU3126G, D0GU2107G, J0GU2857G, K9GU1502G, E0GU2205G, E9GU0840G, M9GU1714G, M9GU1715G, M9GU1716G, M9GU1717G, M9GU1718G, B9GU0517G, B0GU1950G, B0GU1951G, B0GU1952G, B0GU1953G, B0GU1954G, C0GU2089G, C0GU2090G, C0GU2091G, C0GU2092G, C0GU2093G, J0GU2811G, J0GU2812G, J9GU1377G, J9GU1378G, F0GU2432G, F0GU2433G, F0GU2434G, C0GU1999G, F9GU0095G, F9GU0955G, K8GU0024G, K8GU0025G, K8GU0057G, E9GU0842G, E9GU0847G, E9GU0956G, F9GU0956G, D9GU0707G, F9GU0973G, F9GU0974G, F9GU0975G, J9GU1264G, J9GU1265G, G9GU1027G, G9GU1029G, G9GU1030G, G9GU1031G, C1GU3329G, F9GU0907G, M8GU0323G, M8GU0324G, E0GU2276G, A0GU1865G, G0GU2524G, G0GU2542G, F9GU0966G, F9GU0967G, B0GU1959G, E0GU2233G, H9GU1189G, H9GU1190G, H9GU1191G, H9GU1192G, H9GU1992G, L9GU1624G, L9GU1625G, M9GU1652G, M9GU1691G, M9GU1694G, D9GU0701G, F0GU2426G, F0GU2427G, J0GU2835G, J0GU2838G, F0GU2353G, F0GU2354G, B9GU0538G, B9GU0539G, A0GU1835G, F0GU2474G, F9GU1000G, J0GU2816G, L8GU0190G, L8GU0191G, E1GU3373G, L8GU0194G, L8GU0195G, L8GU0220G, L8GU0221G, B9GU0447G, B0GU1973G, B0GU1974G, B0GU1975G, L8GU0188G, L8GU0189G, L8GU0198G, C9GU0669G, C9GU0670G, C9GU0671G, C9GU0672G, G9GU1075G, G9GU1076G, J0GU2891G, J9GU1252G, C9GU0551G, B0GU1967G, C9GU0610G, C9GU0611G, C9GU0612G, C9GU0615G, C9GU0616G, D0GU2174G, E0GU2244G, E0GU2248G, E0GU2282G, F0GU2430G, L8GU0193G, F0GU2451G, F0GU2367G, E9GU0803G, E9GU0808G, E9GU0814G, E9GU0871G, E9GU0872G, E9GU0873G, E9GU0874G, L9GU1577G, M0GU3135G, G9GU1044G, E1GU3372G, A0GU1866G, L8GU0204G, L8GU0205G, J9GU1305G, A9GU0377G, A9GU0378G, A9GU0379G, A9GU0380G, A9GU0381G, A9GU0382G, A9GU0385G, A9GU0386G, A9GU0396G, A9GU0397G, B9GU0528G, C9GU0673G, C9GU0674G, C9GU0675G, C9GU0676G, C9GU0680G, D9GU0741G, D9GU0754G, D9GU0759G, G0GU2486G, G0GU2487G, G0GU2513G, G9GU1089G, H9GU1093G, H9GU1201G, H9GU1202G, J9GU1359G, K9GU1427G, K9GU1428G, L8GU0200G, L9GU1517G, M8GU0333G, J9GU1256G, J9GU1257G, L9GU1542G, E9GU0816G, F9GU0982G, G9GU1058G, G9GU1059G, H9GU1210G, H9GU1211G, L9GU1525G, L9GU1558G, M9GU1710G, C0GU2050G, C0GU2051G, L0GU3041G, C0GU2029G, A0GU1841G, F0GU2470G, F0GU2473G, K8GU0023G, B0GU1906G, B0GU1907G, B0GU1908G, B0GU1909G, B0GU1910G, B0GU1911G, B0GU1912G, B0GU1913G, C0GU2031G, C0GU2086G, B0GU1930G, B0GU1932G, B0GU1933G, B0GU1966G, B0GU1968G, B1GU3273G, D0GU2121G, E9GU0797G, F0GU2429G, F0GU2477G, F9GU0950G, F9GU0951G, F9GU0958G, G0GU2604G, H0GU2650G, H0GU2651G, H0GU2667G, H0GU2668G, H0GU2669G, H9GU1138G, H9GU1215G, H9GU1216G, H9GU1217G, H9GU1218G, J9GU1307G, J9GU1308G, J9GU1309G, J9GU1310G, J9GU1327G, K0GU2955G, K9GU1423G, K9GU1425G, K9GU1430G, K9GU1431G, K9GU1467G, K9GU1468G, K9GU1469G, K9GU1470G, L0GU3066G, L0GU3067G, L9GU1554G, L9GU1606G, L9GU1607G, M8GU0357G, M8GU0358G, H0GU2677G, J0GU2728G, J0GU2787G, J0GU2788G, J0GU2821G, J0GU2822G, J0GU2827G, J0GU2828G, J0GU2829G, J0GU2830G, J0GU2832G, J0GU2833G, M0GU3147G, G0GU2595G, G0GU2611G, H0GU2685G, E0GU2200G, A0GU1750G, F0GU2393G, F0GU2394G, F0GU2404G, J0GU2810G, L0GU3012G, G0GU2582G, J9GU1343G, J9GU1344G, J9GU1345G, J9GU1347G, J9GU1348G, J9GU1349G, L9GU1618G, D9GU0691GR, H9GU1197G, J9GU1251G, L9GU1570G, F0GU2407G, F0GU2428G, J9GU1375G, K9GU1403G, K0GU2946G, A0GU1748G, A0GU1756G, A0GU1757G, A0GU1796G, A0GU1799G, A0GU1805G, A0GU1816G, A0GU1850G, A0GU1854G, A0GU1870G, H9GU1148G, H9GU1229G, H9GU1247G, H9GU1248G, H9GU1249G, H9GU1250G, J9GU1253G, J9GU1254G, J9GU1255G, J9GU1272G, J9GU1273G, J9GU1278G, J9GU1279G, J9GU1280G, K9GU1381G, K9GU1382G, K9GU1500G, L9GU1538G, L9GU1541G, L9GU1578G, L9GU1579G, L9GU1580G, L9GU1581G, M9GU1644G, M9GU1645G, M9GU1646G, M9GU1707G, M9GU1722G, M9GU1729G, C9GU0613G, L0GU2999G, K8GU0026G, B1GU3216G, B9GU0449G, D0GU2179G, D9GU0702G, E0GU2265G, F1GU3384G, L8GU0222G, L8GU0224G, L8GU0225G, M8GU0340G, M8GU0341G, M8GU0343G, H9GU1133G, L9GU1546G, C1GU3328G, G0GU2504G, G0GU2505G, C9GU0577G, C9GU0588G, E0GU2219G, E0GU2220G, J0GU2796G, J0GU2819G, E9GU0819G, E9GU0860G, E9GU0861G, E9GU0862G, E9GU0863G, E9GU0864G, C0GU2072G, E1GU3371G, G0GU2492G, G0GU2493G, G0GU2581G, H0GU2637G, J0GU2748G, J0GU2774G, J0GU2775G, J0GU2794G, J0GU2795G, J0GU2797G, J0GU2820G, C9GU0591G, C9GU0597G, E9GU0868G, G0GU2613G, H0GU2666G, G9GU1021G, G9GU1022G, D1GU3348G, D1GU3349G, E0GU2224G, E0GU2225G, E0GU2226G, B0GU1956G, H9GU1172G, H9GU1173G, H9GU1181G, H9GU1182G, H9GU1184G, H9GU1194G, H9GU1195G, K9GU1395G, K9GU1396G, K9GU1398G, K9GU1399G, K9GU1400G, K9GU1401G, K9GU1424G, C0GU2068G, J9GU1262G, J9GU1263G, K8GU0064G, K8GU0065G, K8GU0066G, K8GU0070G, K8GU0071G, K8GU0086G, K8GU0087G, K8GU0091G, K8GU0092G, K8GU0093G, K8GU0095G, L9GU1518G, L9GU1508G, L9GU1509G, E0GU2223G, E0GU2230G, E0GU2234G, E0GU2237G, E0GU2238G, E0GU2250G, F0GU2467G, M0GU3101G, M0GU3109G, M0GU3111G, M0GU3112G, L9GU1559G, L9GU1560G, L9GU1561G, L9GU1562G, L9GU1563G, L9GU1564G, A9GU0371G, A9GU0372G, A9GU0373G, A9GU0400G, A9GU0422G, A9GU0423G, A9GU0424G, B9GU0505G, C9GU0573G, D9GU0742G, D9GU0743G, M8GU0239G, M8GU0240G, M8GU0241G, M8GU0353G, M8GU0356G, C0GU2073G, G0GU2573G, B0GU1914G, B0GU1915G, C9GU0642G, C9GU0643G, D0GU2162G, E0GU2216G, H9GU1245G, K8GU0112GR, M9GU1661G, M9GU1662G, M9GU1671G, G0GU2500G, M8GU0301G, M8GU0302G, M8GU0304G, M8GU0322G, D0GU2178G, A0GU1775G, L9GU1520G, M0GU3134G, M0GU3143G, M8GU0242G, E0GU2328G, E0GU2330G, G0GU2503G, J9GU1370G, J9GU1374G, K9GU1422G, M8GU0229G, M8GU0232G, M8GU0233G, M8GU0234G, M8GU0235G, M8GU0236G, M8GU0237G, E9GU0815G, E9GU0992G, F9GU0992G, K8GU0049G, K8GU0050G, K8GU0082G, K8GU0084G, K8GU0085G, F9GU0971G, L9GU1594G, L9GU1595G, M0GU3145G, A0GU1885G, K8GU0079G, L9GU1582G, M9GU1649G, M9GU1650G, M9GU1672G, B0GU1969G, B0GU1970G, B0GU1971G, L0GU3005G, D9GU0791G, D9GU0792G, L0GU2987G, L0GU3035G, B0GU1916G, L9GU1587G, F0GU2374G, C9GU0555G, C9GU0556G, C9GU0560G, C9GU0561G, C9GU0562G, C9GU0563G, C9GU0565G, C9GU0598G, C9GU0677G, C9GU0678G, C9GU0679G, D9GU0696G, D9GU0697G, E9GU0884G, E9GU0885G, F0GU2435G, G9GU1034G, H9GU1112G, J0GU2890G, K0GU2938G, K0GU2983G, K9GU1397G, L9GU1557G, E0GU2342G, F0GU2416G, K9GU1451G, K9GU1452G, K9GU1453G, K9GU1456G, A9GU0390G, A9GU0391G, B0GU1920G, B0GU1921G, C1GU3319G, D1GU3347G, M8GU0363G, C9GU0607G, L0GU3072G, F0GU2408G, D9GU0722G, E0GU2198G, A0GU1751G, G0GU2601G, L9GU1585G, F9GU1001G, G0GU2589G, H0GU2699G, H0GU2700G, H0GU2701G, H0GU2702G, G0GU2574G, G0GU2575G, G0GU2576G, C0GU2083G, M9GU1675G, M9GU1676G, D0GU2108G, A0GU1801G, B0GU1986G, B0GU1987G, A9GU0418G, A9GU0425G, E9GU0846G, H9GU1136G, M8GU0348G, M9GU1670G, M8GU0330G, C9GU0599G, K0GU2918G, K9GU1411G, K9GU1412G, K9GU1454G, M0GU3153G, H9GU1183G, J9GU1318G, J0GU2740G, J0GU2741G, F9GU0961G, F9GU0962G, M0GU3157G, M0GU3158G, B0GU1960G, K8GU0099G, K8GU0106G, K8GU0107G, K9GU0106G, E0GU2253G, E0GU2254G, B1GU3250G, J9GU1372G, J9GU1373G, J9GU1376G, K9GU1413G, B9GU0522G, H0GU2692G, H0GU2693G, J9GU1266G, K8GU0047G, L8GU0192G, J0GU2780G, J0GU2806G, A0GU1809G, A0GU1810G, A0GU1830G, A0GU1831G, A0GU1840G, C0GU2040G, E0GU2278G, E0GU2317G, J9GU1298G, J9GU1299G, K9GU1442G, K9GU1443G, L9GU1528G, L9GU1529G, L9GU1530G, L9GU1531G, E9GUO890G, F0GU2469G, H9GU1239G, J9GU1286G, J9GU1287G, J9GU1288G, K8GU5049G, M9GU1638G, B9GU0453G, C9GU0592G, C9GU0593G, D9GU0744G, D9GU0745G, G9GU1041G, B1GU3248G, L9GU1626G, H9GU1222G, A0GU1814G, B9GU0486G, J9GU1303G, J9GU1304G, B9GU0518G, H9GU1115G, H9GU1116G, H9GU1117G, L8GU0226G, H9GU1223G, H9GU1243G, H9GU1244G, G0GU2520G, J0GU2738G, J0GU2739G, J0GU2879G, E0GU2275G, E0GU2295G, A0GU1845G, B0GU1902G, B0GU1903G, J9GU1364G, J9GU1365G, B9GU0442G, L8GU0202G, A9GU0383G, A9GU0384G, D0GU2131G, D0GU2132G, L9GU1535G, L9GU1536G, L9GU1537G, F9GU0984G, F9GU0985G, F9GU0916G, F9GU0960G, G0GU2515G, G0GU2516G, G0GU2517G, G9GU1007G, G9GU1008G, H0GU2676G, H0GU2683G, H9GU1220G, J0GU2777G, J0GU2798G, J0GU2799G, J0GU2800G, M0GU3106G, M0GU3107G, M0GU3108G, M0GU3127G, M9GU1704G, M9GU1705G, A9GU0404G, C0GU2046G, C0GU2047G, C0GU2048G, C0GU2049G, C0GU2059G, C0GU2060G, C1GU3330G, E0GU2231G, E0GU2232G, E9GU0828G, E9GU0829G, G9GU1060G, H9GU1125G, H9GU1126G, J0GU2858G, J0GU2876G, K0GU2954G, L9GU1604G, L9GU1605G, M9GU1706G, G9GU1054G, G9GU1055G, G9GU1056G, G9GU1057G, H0GU2655G, H0GU2656G, H0GU2657G, H0GU2658G, H0GU2715G, H0GU2716G, J0GU2844G, J0GU2845G, J9GU1274G, J9GU1300G, J9GU1317G, J9GU1366G, K8GU0072G, K8GU0073G, K8GU0075G, K8GU0076G, K9GU1383G, C0GU2080G, C0GU2081G, C0GU2082G, M8GU0334G, M8GU0335G, C1GU3315G, M9GU1732G, D0GU2110G, K8GU0046G, B9GU0543G, B9GU0544G, D0GU2119G, D0GU2120G, A0GU1855G, A0GU1856G, A0GU1857G, A0GU1858G, A0GU1859G, A0GU1860G, C0GU2020G, F9GU0947G, B9GU0470G, B9GU0471G, M8GU0230G, C9GU0550G, C9GU0554G, E9GU0886G, B0GU1955G, B0GU1972G, K8GU0053G, K8GU0054G, K8GU0055G, F9GU0954G, H9GU1107G, D0GU2143G, D9GU0718G, D9GU0762G, D9GU7186G, L9GU1627G, L0GU3040G, B0GU1943G, B0GU1944G, H0GU2671G, K0GU2934G, K0GU2935G, K0GU2978G, A9GU0369G, A9GU0374G, L9GU1544G, L9GU1574G, L9GU1575G, L9GU1576G, L9GU1615G, L9GU1616G, B1GU3251G, G0GU2612G, B9GU0520G, C9GU0558G, C9GU0559G, C9GU0567G, G0GU2596G, G0GU2597G, F9GU0928G, F9GU0933G, G9GU1040G, M9GU1724G, D0GU2163G, B0GU1904G, B0GU1905G, B9GU0462G, B9GU0463G, H0GU2688G, H0GU2689G, H0GU2690G, J1GU3406G, J9GU1281G, K8GU0034G, K8GU0035G, K8GU0036G, K8GU0037G, M9GU1731G, B9GU0448G, E9GU0827G, M9GU1702G, M9GU1703G, F9GU0968G, C9GU0645G, F9GU0978G, H9GU1226G, A9GU0421G, B9GU0509G, B9GU0513G, F0GU2436G, F0GU2437G, G0GU2544G, G0GU2586G, H9GU1104G, H9GU1132G, H9GU1208G, J9GU1277G, J9GU1301G, K9GU1503G, D9GU0704G, D9GU0740G, E9GU0820G, A0GU1766G, F0GU2406G, M9GU1648G, B0GU1917G, F0GU2349G, F0GU2365G, F0GU2366G, M9GU1681G, M9GU1682G, B9GU0480G, B9GU0484G, B9GU0508G, B9GU0526G, C9GU0549G, C9GU0553G, D9GU0776G, D9GU0777G, D9GU0778G, L0GU3014G, H0GU2649G, H9GU1106G, J0GU2873G, D9GU0706G, E9GU0841G, A0GU1745G, A0GU1746G, A0GU1747G, A0GU1851G, A0GU1852G, A0GU1853G, B1GU3221G, M9GU1692G, M9GU1693G, K8GU0027G, J0GU2852G, J0GU2831G, A0GU1895G, C9GU0552G, J0GU2715G, J0GU2783G, J0GU2785G, J0GU2T84G, M8GU0326G, M8GU0327G, J0GU2782G, J0GU2784G, M9GU1635G, C9GU0590G, D9GU0683G, D9GU0684G, D9GU0705G, G0GU2523G, G9GU1032G, G9GU1033G, J0GU2843G, H0GU2675G, L9GU1532G, E9GU0858G, L8GU0168G, L8GU0169G, J9GU1258G, C0GU2030G, M0GU3105G, A9GU0403G, A9GU0407G, A9GU0408G, A9GU0412G, B0GU1939G, A0GU1808G, A0GU1832G, A0GU1833G, A0GU1834G, C1GU3323G, C9GU0636G, C9GU0637G, G9GU1063G, J0GU2834G, J9GU1314G, L8GU0171G, L9GU1555G, L9GU1556G, M0GU3115G, M0GU3116G, M0GU3117G, M0GU3136G, M8GU0325G, K9GU1437G, K9GU1439G, A9GU0402G, C1GU3320G, E9GU0839G, H0GU2632G, J9GU1379G, B9GU0493G, B9GU0494G, J0GU2863G, J0GU2864G, J9GU1282G, J9GU1283G, J9GU1284G, J9GU1285G, J9GU1289G, J9GU1290G, J9GU1291G, J9GU1292G, J9GU1293G, J9GU1294G, J9GU1295G, J9GU1296G, J9GU1297G, K8GU0074G, K8GU0080G, K8GU0081G, K8GU0094G, K8GU0096G, K8GU0097G, K8GU0100G, K8GU0101G, K8GU0102G, K8GU0104G, K8GU0105G, K8GU0108G, K8GU0109G, K8GU0110G, K8GU0111G, K8GU0112G, K8GU0114G, K8GU0115G, K8GU0117G, K8GU0118G, L8GU0123G, L8GU0127G, L8GU0128G, L8GU0129G, L8GU0130G, L8GU0140G, L8GU0142G, L8GU0143G, L8GU0144G, L8GU0145G, L8GU0148G, L8GU0149G, L8GU0150G, L8GU0151G, L8GU0153G, L8GU0154G, L8GU0157G, L8GU0206G, L9GU1586G, M8GU0305G, M8GU0306G, M8GU0307G, M8GU0308G, M8GU0309G, M8GU0310G, M8GU0311G, M8GU0314G, M8GU0315G, M8GU0316G, M8GU0317G, M8GU0318G, M8GU0319G, M8GU0320G, M9GU1712G, A0GU1802G, C9GU0618G, G0GU2531G, G0GU2534G, G0GU2536G, G0GU2537G, G0GU2566G, G0GU2600G, H0GU2621G, H0GU2623G, H0GU2624G, H0GU2629G, H0GU2633G, J9GU1355G, L8GU0199G, L1GU0154G, L8GU0155G, B9GU0445G, B9GU0435G, B9GU0441G, G0GU2483G, H0GU2618G, C1GU3322G, M0GU3146G, E9GU0826G, G0GU2538G, G0GU2539G, G0GU2540G, G0GU2541G, K0GU2982G, M8GU0331G, G9GU1088G, F1GU3399G, J0GU2886G, A1GU3199G, C0GU2037G, C0GU2038G, C0GU2039G, C1GU3316G, C9GU0576G, D0GU2134G, E0GU2314G, M0GU3102G, M0GU3110G, M0GU3131G, M0GU3132G, M0GU3133G, A1GU3167G, A1GU3168G, B1GU3253G, C0GU2036G, C0GU2042G, C0GU2062G, C0GU2063G, C0GU2064G, C1GU3326G, C1GU3327G, C1GU3332G, D0GU2189G, D9GU0770G, D9GU0771G, D9GU0772G, D9GU0773G, D9GU0780G, D9GU0781G, D9GU0782G, D9GU0783G, D9GU0784G, D9GU0785G, E0GU2343G, F9GU0917G, G0GU2490G, G0GU2491G, K0GU2923G, K0GU2924G, K0GU2933G, K9GU1464G, K9GU1465G, K9GU1466G, L8GU0174G, L8GU0175G, L8GU0176G, L8GU0177G, L8GU0178G, L8GU0179G, L8GU0181G, M9GU1666G, M9GU1708G, G9GU1062G, G9GU1065G, G9GU1067G, G9GU1081G, G9GU1083G, G9GU1084G, G9GU1085G, G9GU1086G, C0GU2065G, G0GU2484G, A0GU1867G, A0GU1868G, D9GU0713G, G0GU2526G, G0GU2527G, G0GU2528G, G0GU2529G, G0GU2530G, H0GU2686G, H0GU2687G, L0GU2989G, D0GU2140G, D0GU2141G, D0GU2142G, B0GU1949G, B1GU3210G, B1GU3211G, B1GU3212G, B1GU3213G, B1GU3214G, B1GU3215G, B1GU3229G, B1GU3231G, B1GU3233G, B1GU3238G, B1GU3239G, B1GU3240G, B1GU3241G, B1GU3243G, B1GU3246G, B1GU3247G, B9GU0483G, B9GU0514G, B9GU0515G, B9GU0516G, B9GU0521G, B9GU0540G, B9GU0541G, B9GU0542G, B9GU0547G, B9GU0548G, C9GU0575G, D9GU0763G, E0GU2260G, E0GU2347G, E1GU3352G, E1GU3353G, E1GU3354G, E1GU3355G, E1GU3356G, E1GU3357G, E1GU3358G, E9GU0834G, F0GU2468G, F9GU0939G, F9GU0940G, F9GU0942G, F9GU0943G, G0GU2603G, H0GU2638G, J0GU2847G, J0GU2848G, J0GU2849G, J0GU2855G, J0GU2856G, J0GU2859G, J0GU2860G, J0GU2865G, J0GU2866G, J0GU2868G, J0GU2869G, J0GU2870G, J0GU2871G, J0GU2872G, K8GU0032G, K8GU0033G, K8GU0040G, K8GU0041G, K8GU0042G, K8GU0043G, K8GU0044G, M8GU0364G, L8GU0162G, L8GU0164G, A0GU1842G, A0GU1846G, A0GU1869G, A0GU1888G, A0GU1887G, C0GU2043G, C0GU2044G, C0GU2077G, L0GU3011G, L9GU1510G, G0GU2511G, B9GU0450G, C9GU0569G, D1GU3343G, K8GU0038G, L0GU3049G, L0GU3053G, A0GU1770G, A0GU1771G, A0GU1871G, A9GU0409G, A9GU0410G, E9GU0887G, E9GU0888G, E9GU0889G, E9GU1001G, F9GU0957G, F9GU0980G, G0GU2585G, H9GU1149G, H9GU1150G, K0GU2977G, L0GU2998G, L9GU1612G, M0GU3089G, J9GU1332G, J9GU1333G, K9GU1472G, K9GU1473G, F0GU2461G, M8GU0251G, D9GU0753G, F9GU0945G, F9GU0946G, K8GU0056G, A0GU1829G, D0GU2137G, J0GU2753G, F9GU0948G, F9GU0977G, K9GU1449G, E9GU0817G, E9GU0818G, A9GU0389G, C9GU0653G, D9GU0726G, D9GU0727G, D9GU0728G, D9GU0729G, D9GU0730G, D9GU0731G, D9GU0732G, D9GU0733G, D9GU0734G, D9GU0735G, D9GU0736G, D9GU0737G, D9GU0789G, E9GU0880G, E9GU1880G, L8GU0158G, L8GU0159G, L8GU0163G, L8GU0213G, L8GU0214G, L8GU0215G, L8GU0216G, L8GU0218G, L8GU0219G, M8GU0253G, M8GU0254G, M8GU0255G, M8GU0256G, M8GU0257G, M8GU0258G, M8GU0259G, M8GU0260G, M8GU0261G, M8GU0262G, M8GU0263G, J9GU1302G, E9GU0867G, L9GU1617G, D9GU0689G, E0GU2197G, K8GU0060G, K8GU0061G, K8GU0062G, K8GU0068G, K8GU0069G, B0GU1983G, A9GU0399G, B0GU1900G, B1GU3252G, B9GU0533G, B9GU0545G, B9GU0546G, F0GU2359G, F0GU2360G, F9GU0959G, F9GU1002G, G0GU2501G, J0GU2776G, K8GU0077G, K8GU0078G, J0GU2825G, J0GU2888G, B9GU0446G, F0GU2431G, L8GU0124G, L8GU0125G, L8GU0152G, M8GU0337G, M8GU0349G, C1GU3334G, H0GU2697G, C9GU0640G, C9GU0641G, C9GU0647G, C9GU0650G, C9GU0651G, C9GU0652G, D9GU0688G, E0GU2240G, F9GU0999G, K8GU0098G, L8GU0207G, L8GU0208G, L8GU0209G, L8GU0210G, M8GU0252G, J0GU2853G, L9GU1547GR, E9GU0801G, J0GU2818G, J0GU2846G, J0GU2851G, L0GU3051G, B1GU3222G, K9GU1405G, H0GU2674G, D0GU2127G, G0GU2545G, G0GU2546G, G0GU2547G, G0GU2548G, G0GU2549G, G0GU2550G, G0GU2551G, G0GU2552G, G0GU2553G, G0GU2554G, G0GU2555G, G0GU2556G, G0GU557G, B1GU3223G, A0GU1749G, C9GU0595G, C9GU0596G, D9GU0692G, D9GU0693G, D9GU0694G, K8GU0058G, K8GU0059G, C0GU2054G, C0GU2056G, C0GU2057G, C0GU2058G, C1GU3281G, M0GU3104G, A0GU1806G, A0GU1807G, E9GU0844G, E0GU2201G, E0GU2202G, E0GU2203G, E0GU2305G, E0GU2306G, E0GU2307G, F0GU2458G, F0GU2459G, F0GU2460G, F0GU2462G, F0GU2463G, F0GU2464G, F0GU2465G, F9GU0989G, F9GU0990G, F9GU0991G, H9GU1091G, H9GU1092G, H9GU1094G, H9GU1095G, K9GU1384G, K9GU1407G, K9GU1408G, K9GU1409G, L9GU1094G, M9GU1721G, D0GU2111G, D0GU2112G, D0GU2113G, D0GU2114G, D0GU2118G, B9GU0472G, B9GU0473G, B9GU0474G, B9GU0475G, B9GU0476G, B9GU0477G, B9GU0478G, K0GU2980G, C0GU2074G, C9GU0574G, D0GU2184G, D0GU2185G, A0GU1881G, A0GU1882G, B9GU0530G, B9GU0532G, D8GU0295G, F9GU0927G, F9GU0934G, F9GU0935G, H9GU1152G, L9GU1596G, L9GU1597G, L9GU1598G, L9GU1599G, E0GU2204G, C1GU3335G, D0GU2129G, E8GU0806G, E9GU0809G, E9GU0822G, E9GU0823G, F1GU3387G, G0GU2499G, G0GU2567G, G0GU2568G, G0GU2579G, G9GU1035G, K9GU1404G, G9GU1073G, G9GU1074G, K8GU0089G, K8GU0090G, L0GU2986G, L0GU2988G, K9GU1471G, C0GU2021G, C0GU2088G, L0GU3000G, L0GU3001G, L0GU3002G, E0GU2269G, E0GU2270G, E0GU2235G, E0GU2236G, M9GU1699G, M9GU1700G, A9GU0368G, M0GU3074G, G9GU1009G, J0GU2746G, M8GU0282G, M8GU0283G, M8GU0288G, B1GU3276G, G0GU2506G, C0GU2055G, F0GU2466G, C0GU2052G, C0GU2070G, C0GU2071G, B0GU1961G, B0GU1964G, K9GU1416G, D9GU0779G, A0GU1768G, B0GU1948G, L9GU1588G, L9GU1589G, L9GU1590G, L9GU1591G, L9GU1592G, L9GU1593G, A0GU1862G, B0GU1938G, E9GU0793G, E9GU0824G, G0GU2602G, M9GU1673G, H9GU1134G, A0GU1779G, H9GU1246G, J9GU1324G, J9GU1342G, J9GU1325G, G0GU2521G, G0GU2563G, G0GU2564G, G0GU2565G, G0GU2584G, G0GU2598G, J0GU2726G, F0GU2441G, B0GU1957G, B1GU3230G, E0GU2341G, F0GU2405G, G0GU2482G, B1GU3236G, E9GU0813G, A0GU1822G, H0GU2713G, M0GU3099G, M9GU1687G, M9GU1688G, G0GU2592G, G0GU2593G, L9GU1629G, J9GU1361G, M0GU3119G, J0GU2805G, J9GU1360G, C0GU2076G, F0GU2471G, F0GU2472G, F0GU2422G, F0GU2423G, F0GU2444G, G0GU2509G, J0GU2813G, L0GU3060G, C9GU0594G, G0GU2518G, J0GU2804G, J0GU2764G, F0GU2400G, D0GU2105G, J0GU2803G, A0GU1787G, F0GU2419G, F0GU2415G, M0GU3130G, H9GU1200G, J0GU2763G, J0GU2765G, E1GU3360G, G0GU2594G, G0GU2599G, J0GU2762G, K0GU2942G, G0GU2591G, J0GU2766G, B0GU1935G, F0GU2410G, J0GU2737G, M0GU3129G, G0GU2519G, C0GU2005G, K0GU2974G, J0GU2767G, J0GU2769G, F0GU2413G, J0GU2768G, F0GU2414G, F0GU2412G.J0GU2786G, A9GU373G, C1GU3317G, G0GU2497G, J0GU2752G, F9GU0937G, H0GU2714G, H9GU1224G, F0GU2371G, H9GU1119G, K9GU1419G, G9GU1042G, G0GU2590G, M9GU1683G, G0GU2496G, G0GU2588G, A0GU1820G, H0GU2663G, H9GU1118G, K0GU29230G, F9GU0976G, H0GU2707G, H0GU2708G, D0GU2147G, B0GU1995G, H0GU2628G, E0GU2333G, E0GU2334G, B1GU3237G, H0GU2679, H0GU2679G, H0GU2680G, H9GU1120G, B0GU2193G, D0GU2193G, A1GU3191G, A9GU0423, C9GU0571G, D0GU2102G, D0GU2139G, E0GU2335G, H0GU1818G, J9GUI319G, D0GU2101G, A1GU3190G, K0GU2936G, F9GU0938G, G8GU1023G, G9GU1023G, J0GU2750G, F0GU2385G, B0GU1991G, B0GU1993G, K0GU2919G, B1GU3235G, B0GU1995, F0GU2386G, C9GU0570G, H9GU1236G, H9GU1237G, L9GU1516G, M9GU1685G, H9GU1144G, A0GU1765G, D0GU2151G, D0GU2152G, D0GU2166G, D0GU2167G, D0GU2168G, C0GU2014G, B0GU1996G, C1GU3297G, A0GU1776G, A0GU1828G, A0GU4128G, D9GU0695G, H0GU2626G, H0GU2627G, A0GU1777G, J9GU1337G, J9GU1338G, L0GU2985G, M8GU0238G, B0GU1931G, B1GU3217G, C1GU3295G, E0GU2214G, F0GU1486G, F9GU0124G, F9GU0912G, G0GU2558G, G0GU2559G, M0GU3080G, G9GU1012G, C0GU2045G, C0GU2027G, M8GU0352G, C9GU0557G, C9GU0564G, G0GU2562G, D0GU2148G, L8GU0119G, L8GU0120G, D9GU0687G, K0GU2807G, K0GU2907G, B9GU0465G, G0GU2606G, M0GU3077G, D0GU2115G, G0GU2508G, L0GU3054G, L9GU1521G, M8GU0344G, F9GU0993G, K9GU1479G, K9GU1483G, K9GU1484G, K9GU1485G, K9GU1486G, K9GU1488G, K9GU1490G, K9GU1491G, L9GU1512G, L9GU1513G, L9GU1515G, L8GU0138G, B0GU1998G, C0GU2001G, C0GU2004G, C0GU2008G, H0GU2642G, E0GU2279G, D9GU0723G, D9GU0724G, K9GU1493G, K9GU1489G, M9GU1696G, C1GU3296G, J0GU2861G, L9GU1504G, K9GU1463G, C9GU0600G, C9GU0601G, C9GU0602G, C9GU0603G, C9GU0604G, C9GU0605G, C9GU0606G, D9GU0708G, E9GU0795G, M0GU3085G, L0GU3044G, L0GU3045G, K8GU0009G, K8GU0010G, K8GU0011G, K8GU0012G, K8GU0013G, K8GU0014G, C0GU2099G, L9GU1572G, H0GU2643G, L8GU0139G, D9GU0765G, K0GU2964G, L0GU3042G, L0GU3043G, A1GU3161G, L8GU0132G, M8GU0361G, A1GU3164G, A1GU3195G, B1GU3226G, J9GU12366, A0GU2641G, M0GU3082G, B9GU0466G, L9GU1571G, M8GU0360G, K0GU2963G, K8GU0039G, F9GU0804G, L9GU1533G, F0GU2388G, B0GU1984G, B0GU1985G, C0GU2013G, J9GU1311G, J9GU1312G, J9GU1313G, B9GU0504G, D9GU0764G, K0GU2961G, G0GU2561G, J9GU1340G, K8GU0015G, A0GU1890G, M9GU1719G, E0GU2280G, L0GU2995G, J0GU2862G, L9GU1573G, M0GU3083G, D0GU2103G, D9GU0714G, C9GU0582G, F9GU0932G, K0GU2906G, D0GU2149G, C0GU2009G, A1GU3162G, G0GU2577G, G0GU2578G, H0GU2639G, H0GU2640G, H0GU2644G, H0GU2645G, G0GU2007G, L9GU1524G, E0GU2281G, C0GU2096G, D0GU2116G, F0GU2356G, A0GU1763G, G9GU1011G, H9GU1111G, A1GU3150G, A1GU3159G, D0GU2124G, M0GU3084G, D0GU2150G, H9GU1145G, B1GU3263G, A1GU3160G, A1GU3194G, L9GU1506G, L0GU3024G, A1GU3179G, A1GU3192G, B1GU3225G, A0GU1762G, D1GU3339G, A0GU1864G, A0GU1883G, J0GU2893G, L9GU1565G, M9GU1639G, M9GU1640G, M9GU1660G, M9GU1674G, K0GU2971G, L0GU3028G, J1GU3410G, J1GU3409G, F1GU3402G, E1GU3364G, F1GU3378G, J9GU1270G, H9GU1185G, J1GU3412G, G0GU2510G, A0GU1792G, H0GU2681G, B0GU1962G, K0GU2932G, A0GU1774G, E1GU3370G, E9GU0883G, L0GU3039G, B1GU3228G, L9GU1526G, B0GU1946G, J0GU2773G, E0GU2268G, J0GU2770G, J0GU2771G, J0GU2772G, J0GU2778G, J0GU2895G, K0GU2897G, K0GU2898G, K0GU2896G, J9GU1346G, H0GU2664G, L0GU3050G, G9GU1045G, L9GU1610G, B1GU3275G, H0GU2691G, L0GU3061G, A0GU1811G, A0GU1812G, A1GU3165G, D0GU2117G, A9GU0370G, F9GU0949G, K0GU2981G, C0GU2026G, B9GU0479G, B9GU0482G, A1GU3208G, A0GU17556, A0GU1755G, C0GU2094G, M0GU3091G, A1GU3203G, A1GU3176G, L8GU0121G, K9GU1457G, C1GU3311G, C1GU3285G, J9GU1275G, A9GU0429G, C0GU2025G, H9GU1180G, B0GU1958G, H9GU1154G, D9GU0700G, A0GU1880G, K0GU2948G, K0GU2949G, A1GU3207G, K0GU2947G, K0GU2950G, F0GU2390, F0GU2390G, A0GU1754G, H9GU1155G, D1GU3350G, D1GU3351G, A0GU1800G, F0GU2369G, M0GU3139G, M0GU3140G, M0GU3141G, M0GU3142G, J9GU1326G, L0GU3058G, L0GU3059G, E1GU3369G, C9GU0619G, E9GU0796G, A9GU0428G, A0GU1892G, A0GU1893G, K0GU2926G, D1GU3342G, K9GU1406G, J0GU2725G, M0GU1733G, B1GU3267G, J0GU2730G, J0GU2781G, L0GU3048G, F0GU2368G, F0GU2375G, F0GU2376G, D9GU0685G, D9GU685G, M8GU0303G, J0GU2826G, K9GU1410G, H0GU2711G, L8GU0203G, B9GU0492G, L0GU3071G, H0GU2659G, L0GU3164G, G0GU2535G, M0GU3154G;
4. Product Code: SAH, UDI Number, Serial Numbers: SAH06432207GA, SAH06442214GA, SAH06442222GA, SAH06181833GA, SAH06241925GA, SAH04350057GA, SAH08423888GA, SAH08423889GA, SAH08423891GA, SAH08423893GA, SAH08423894GA, SAH08423899GA, SAH08423904GA, SAH07052497GA, SAH05160826GA, SAH07052500GA, SAH05110701GA, SAH05140732GA, SAH05140738GA, SAH05140739GA, SAH05140744GA, SAH06392094GA, SAH06392095GA, SAH06392096GA, SAH06392097GA, SAH06392098GA, SAH06392099GA, SAH06392100GA, SAH06392101GA, SAH06392102GA, SAH06392103GA, SAH05491340GA, SAH05491341GA, SAH05491342GA, SAH05491345GA, SAH05491346GA, SAH05501387GA, SAH05501388GA, SAH06141763GA, SAH04420229GA, SAH04470318GA, SAH04470322GA, SAH05250895GA, SAH06432188GA, SAH06432191GA, SAH05431146GA, SAH05010520GA, SAH05010523GA, SAH05160823GA, SAH08023359GA, SAH05150795GA, SAH05150796GA, SAH05150798GA, SAH05150800GA, SAH05150801GA, SAH05150802GA, SAH05150803GA, SAH05160812GA, SAH05160819GA, SAH05431151GA, SAH05431152GA, SAH05431153GA, SAH06021460GA, SAH06452257GA, SAH06462294GA, SAH06462296GA, SAH06462300GA, SAH06462306GA, SAH06462307GA, SAH06462310GA, SAH06462312GA, SAH04460304GA, SAH07403012GA, SAH07403013GA, SAH07403014GA, SAH07403016GA, SAH07403040GA, SAH07403043GA, SAH07403059GA, SAH04420221GA, SAH04420233GA, SAH05220863GA, SAH05290966GA, SAH08383841GA, SAH08403876GA, SAH05300984GA, SAH07142623GA, SAH06141736GA, SAH06141738GA, SAH06141740GA, SAH06141742GA, SAH06141743GA, SAH08303681GA, SAH08313695GA, SAH05040568GA, SAH05511395GA, SAH08023361GA, SAH08443929GA, SAH05010515GA, SAH04360061GA, SAH04360065GA, SAH04360068GA, SAH04410185GA, SAH04410189GA, SAH04410190GA, SAH04410192GA, SAH04410194GA, SAH04410197GA, SAH04410198GA, SAH04410201GA, SAH04410202GA, SAH04470314GA, SAH04490379GA, SAH05250899GA, SAH05250900GA, SAH05250902GA, SAH06522422GA, SAH06522423GA, SAH06522424GA, SAH07212733GA, SAH07212740GA, SAH07212742GA, SAH07212749GA, SAH07212750GA, SAH07212751GA, SAH07372991GA, SAH07372992GA, SAH07403035GA, SAH07493255GA, SAH06051517GA, SAH06061525GA, SAH06071545GA, SAH06071546GA, SAH06071547GA, SAH07142624GA, SAH08203506GA, SAH08203528GA, SAH08203532GA, SAH08213559GA, SAH08253589GA, SAH08253594GA, SAH08253595GA, SAH08253596GA, SAH08253598GA, SAH08253599GA, SAH08253601GA, SAH08253602GA, SAH08253603GA, SAH08253604GA, SAH08373771GA, SAH08423892GA, SAH08423898GA, SAH04490380GA, SAH06191871GA, SAH04530498GA, SAH05010512GA, SAH05010521GA, SAH05260922GA, SAH05290956GA, SAH08373815GA, SAH05250897GA, SAH07423092GA, SAH07423093GA, SAH04500399GA, SAH04500404GA, SAH04500405GA, SAH04500408GA, SAH05080643GA, SAH05260914GA, SAH05260920GA, SAH05260921GA, SAH07372943GA, SAH07372946GA, SAH07372951GA, SAH07372954GA, SAH07372958GA, SAH07372960GA, SAH05040565GA, SAH07202702GA, SAH08283634GA, SAH08293663GA, SAH08313688GA, SAH08313694GA, SAH05140729GA, SAH07062545GA, SAH06021455GA, SAH08494058GA, SAH08514087GA, SAH04490361GA, SAH09024176GA, SAH05471247GA, SAH05471248GA, SAH05471250GA, SAH05471279GA, SAH05471282GA, SAH07292841GA, SAH07493281GA, SAH08373775GA, SAH08373780GA, SAH05321006GA, SAH05511397GA, SAH07172678GA, SAH07172679GA, SAH07172680GA, SAH07172681GA, SAH07202727GA, SAH07202728GA, SAH07202729GA, SAH08353729GA, SAH08383838GA, SAH08423923GA, SAH05070619GA, SAH05070624GA, SAH05491359GA, SAH08143430GA, SAH08143437GA, SAH06191864GA, SAH06191868GA, SAH06191869GA, SAH07312885GA, SAH08253617GA, SAH08283639GA, SAH08283641GA, SAH08283643GA, SAH08293659GA, SAH08313693GA, SAH04520471GA, SAH04530487GA, SAH05441184GA, SAH07172673GA, SAH07172674GA, SAH07202723GA, SAH07202724GA, SAH07202725GA, SAH07232763GA, SAH07232764GA, SAH07232765GA, SAH07232766GA, SAH07232767GA, SAH07232771GA, SAH07232772GA, SAH07232776GA, SAH07232777GA, SAH07232778GA, SAH07232779GA, SAH07232780GA, SAH07232781GA, SAH07232782GA, SAH07232783GA, SAH07232786GA, SAH07232787GA, SAH07232788GA, SAH07232791GA, SAH07292832GA, SAH04510440GA, SAH05280935GA, SAH07372953GA, SAH07372999GA, SAH04410204GA, SAH08443940GA, SAH06171782GA, SAH07142639GA, SAH07102585GA, SAH07102586GA, SAH06382064GA, SAH06382068GA, SAH06382069GA, SAH06382072GA, SAH05010530GA, SAH05290958GA, SAH06412135GA, SAH06412141GA, SAH04410186GA, SAH04410196GA, SAH04420231GA, SAH04430242GA, SAH04450265GA, SAH04450276GA, SAH04460301GA, SAH04500393SP, SAH07052505GA, SAH07142633GA, SAH07152651GA, SAH07202721GA, SAH06452250GA, SAH07072554GA, SAH07072561GA, SAH07172676GA, SAH05010517GA, SAH05010537GA, SAH05010555GA, SAH07403047GA, SAH07443143GA, SAH08343707GA, SAH08353736GA, SAH08353737GA, SAH08353739GA, SAH08353743GA, SAH08363745GA, SAH08363747GA, SAH08363753GA, SAH08524133GA, SAH06382065GA, SAH07372998GA, SAH07403027GA, SAH08023331GA, SAH08023333GA, SAH05160818GA, SAH05521405GA, SAH05521413GA, SAH08133423GA, SAH08173477GA, SAH06522393GA, SAH07212753GA, SAH07172687GA, SAH07172688GA, SAH07172689GA, SAH07172690GA, SAH07292836GA, SAH07292839GA, SAH05220843GA, SAH05010538GA, SAH05150776GA, SAH05150778GA, SAH05150799GA, SAH05160806GA, SAH06502356GA, SAH06502357GA, SAH07012432GA, SAH07012434GA, SAH07012439GA, SAH07012450GA, SAH07012451GA, SAH07302877GA, SAH07302879GA, SAH07302881GA, SAH07443145GA, SAH06332016GA, SAH06332022GA, SAH08143425GA, SAH08173482GA, SAH05381046GA, SAH05381065GA, SAH04410193GA, SAH04500385GA, SAH04500386GA, SAH04500387GA, SAH04500389GA, SAH04500390GA, SAH04500391GA, SAH06241958GA, SAH06241963GA, SAH06241906GA, SAH07342937GA, SAH07372945GA, SAH07372959GA, SAH07372994GA, SAH07372995GA, SAH07372996GA, SAH07372997GA, SAH04450260GA, SAH04450266GA, SAH04450275GA, SAH06291978GA, SAH06462297GA, SAH06462313GA, SAH07292823GA, SAH07292824GA, SAH07292825GA, SAH07292828GA, SAH07443129GA, SAH07443135GA, SAH08423903GA, SAH04510436GA, SAH04510437GA, SAH05110710GA, SAH05140733GA, SAH05160810GA, SAH06181828GA, SAH06181850GA, SAH06191854GA, SAH06382079GA, SAH06382090GA, SAH06382091GA, SAH06382092GA, SAH06402120GA, SAH07423090GA, SAH08423914GA, SAH06382085GA, SAH07052504GA, SAH07142625GA, SAH05321004GA, SAH06241937GA, SAH06241938GA, SAH06241939GA, SAH08143427GA, SAH08143428GA, SAH08143439GA, SAH05451205GA, SAH05451207GA, SAH05451211GA, SAH05451213GA, SAH05451220GA, SAH06412134GA, SAH07152656GA, SAH07102599GA, SAH08203525GA, SAH08203527GA, SAH08203530GA, SAH08203531GA, SAH08213549GA, SAH08494007GA, SAH08494008GA, SAH06091593GA, SAH07152663GA, SAH07152664GA, SAH05491351GA, SAH06241966GA, SAH07372948GA, SAH08033386GA, SAH08203521GA, SAH08353735GA, SAH08353738GA, SAH08353740GA, SAH08353741GA, SAH08353742GA, SAH08353744GA, SAH08363748GA, SAH08363757GA, SAH08363760GA, SAH08363763GA, SAH08373769GA, SAH08383827GA, SAH08383835GA, SAH08383836GA, SAH08383843GA, SAH08494057GA, SAH09024157GA, SAH09024158GA, SAH09024170GA, SAH09024186GA, SAH04360070GA, SAH05431160GA, SAH05521407GA, SAH08423901GA, SAH05150797GA, SAH08213555GA, SAH08213560GA, SAH06462302GA, SAH06462303GA, SAH04430250GA, SAH05381067GA, SAH05471276GA, SAH05481298GA, SAH05481299GA, SAH05481302GA, SAH05481303GA, SAH05481308GA, SAH05481313GA, SAH05491320GA, SAH05491321GA, SAH05491322GA, SAH05491323GA, SAH05491324GA, SAH05491325GA, SAH05491326GA, SAH05491327GA, SAH05491328GA, SAH05491329GA, SAH05491330GA, SAH05491332GA, SAH05491333GA, SAH05491334GA, SAH05491335GA, SAH05491336GA, SAH06021445GA, SAH06021446GA, SAH06181822GA, SAH07052522GA, SAH08213561GA, SAH08343720GA, SAH08343721GA, SAH07493273GA, SAH05290977GA, SAH05290981GA, SAH05290983GA, SAH05320998GA, SAH06442215GA, SAH06442218GA, SAH06442220GA, SAH06442223GA, SAH06502354GA, SAH06502355GA, SAH06502362GA, SAH06502363GA, SAH06502364GA, SAH06502365GA, SAH06502366GA, SAH06502367GA, SAH06502368GA, SAH06502369GA, SAH06502370GA, SAH06502371GA, SAH07052494GA, SAH07052495GA, SAH07052499GA, SAH07052502GA, SAH07052520GA, SAH07062525GA, SAH08373767GA, SAH08373796GA, SAH08373799GA, SAH05140735GA, SAH05220835GA, SAH05240893GA, SAH05260908GA, SAH05260913GA, SAH05280924GA, SAH05280926GA, SAH05280927GA, SAH05280928GA, SAH05280930GA, SAH05280933GA, SAH05280943GA, SAH05290959GA, SAH05320993GA, SAH06041482GA, SAH07372968GA, SAH07372970GA, SAH08473982GA, SAH06141737GA, SAH07483193GA, SAH07483222GA, SAH08023366GA, SAH08023367GA, SAH08023371GA, SAH04360069GA, SAH04400181GA, SAH04400183GA, SAH04410205GA, SAH04420210GA, SAH04420211GA, SAH04420212GA, SAH04420213GA, SAH04420214GA, SAH04420215GA, SAH04420216GA, SAH04420217GA, SAH04420218GA, SAH04420219GA, SAH04480337GA, SAH04480356GA, SAH04510434GA, SAH04510456GA, SAH04530495GA, SAH04530499GA, SAH04530508GA, SAH04530509GA, SAH05010510GA, SAH05010511GA, SAH05010514GA, SAH05010518GA, SAH05010526GA, SAH05010527GA, SAH05010531GA, SAH05010532GA, SAH05010534GA, SAH05010542GA, SAH05010544GA, SAH05010551GA, SAH05010562GA, SAH05040569GA, SAH05070597GA, SAH05070602GA, SAH05070603GA, SAH05070604GA, SAH05070605GA, SAH05070606GA, SAH05070617GA, SAH05080636GA, SAH05080640GA, SAH05080642GA, SAH05140717GA, SAH05140727GA, SAH05140730GA, SAH05140743GA, SAH05240873GA, SAH05391075GA, SAH05391089GA, SAH05391093GA, SAH05421120GA, SAH05431130GA, SAH05471237GA, SAH05471244GA, SAH06011433GA, SAH06081565GA, SAH06091592GA, SAH06091609GA, SAH06091611GA, SAH06091620GA, SAH06101638GA, SAH06101643GA, SAH06101650GA, SAH06121705GA, SAH06121708GA, SAH06141745GA, SAH06141751GA, SAH06171792GA, SAH06171798GA, SAH07052518GA, SAH07102583GA, SAH07282811GA, SAH08033375GA, SAH08133405GA, SAH08133409GA, SAH08143426GA, SAH08143431GA, SAH08143434GA, SAH08143435GA, SAH08143440GA, SAH08143441GA, SAH08143443GA, SAH08153456GA, SAH08153457GA, SAH08153458GA, SAH08153461GA, SAH08153463GA, SAH08213566GA, SAH08213578GA, SAH08373782GA, SAH08443933GA, SAH08443938GA, SAH08443957GA, SAH08443961GA, SAH08473968GA, SAH08473973GA, SAH08473979GA, SAH08473983GA, SAH08473985GA, SAH08473994GA, SAH08473999GA, SAH08524129GA, SAH08524132GA, SAH08524134GA, SAH08133414GA, SAH08253605GA, SAH08253607GA, SAH08253608GA, SAH08253612GA, SAH08473972GA, SAH08443943GA, SAH08443956GA, SAH05260923GA, SAH06041477GA, SAH07493277GA, SAH07493298GA, SAH07493302GA, SAH07493304GA, SAH04360064GA, SAH04400161GA, SAH04400166GA, SAH04400169GA, SAH04400182GA, SAH04400184GA, SAH06041465GA, SAH06422162GA, SAH06462285GA, SAH06462290GA, SAH06462291GA, SAH07282815GA, SAH07403021GA, SAH05290961GA, SAH06522425GA, SAH04510432GA, SAH05240880GA, SAH05290971GA, SAH05441167GA, SAH05441169GA, SAH05451196GA, SAH05451198GA, SAH05451202GA, SAH05451206GA, SAH05461223GA, SAH05481307GA, SAH06422178GA, SAH06482319GA, SAH06482321GA, SAH06482322GA, SAH05481305GA, SAH05481311GA, SAH05481312GA, SAH07102595GA, SAH07102596GA, SAH07493295GA, SAH07493296GA, SAH07493297GA, SAH07493300GA, SAH07493305GA, SAH07493306GA, SAH07493307GA, SAH07493308GA, SAH07493310GA, SAH07493311GA, SAH04470310GA, SAH05441172GA, SAH08213573GA, SAH08313691GA, SAH05320997GA, SAH05321000GA, SAH05321001GA, SAH05321002GA, SAH06291986GA, SAH06291993GA, SAH05441179GA, SAH06522379GA, SAH07493276GA, SAH05441170GA, SAH05441189GA, SAH05441190GA, SAH05441191GA, SAH05441192GA, SAH05441194GA, SAH06051515GA, SAH07282813GA, SAH07372986GA, SAH07372988GA, SAH04380107GA, SAH04450279GA, SAH04460297GA, SAH05391076GA, SAH06342058GA, SAH07152667GA, SAH07212754GA, SAH07212755GA, SAH08253622GA, SAH07403022GA, SAH07423071GA, SAH07493268GA, SAH08023335GA, SAH08023343GA, SAH06342049GA, SAH07022470GA, SAH04520466GA, SAH04520467GA, SAH07052498GA, SAH07052503GA, SAH07212737GA, SAH07212738GA, SAH07212747GA, SAH05070600GA, SAH05070612GA, SAH05391077GA, SAH05391078GA, SAH05391079GA, SAH05391080GA, SAH05481285GA, SAH05481286GA, SAH05481288GA, SAH05481289GA, SAH05481291GA, SAH05481293GA, SAH05481294GA, SAH08293657GA, SAH04430246GA, SAH04430247GA, SAH07342936GA, SAH06241879GA, SAH06241882GA, SAH06241884GA, SAH06241949GA, SAH07423108GA, SAH05351018GA, SAH05351021GA, SAH05351022GA, SAH05351029GA, SAH06081556GA, SAH07493275GA, SAH05133397GA, SAH08133395GA, SAH08133396GA, SAH08133397GA, SAH08133401GA, SAH08133402GA, SAH08133422GA, SAH05150793GA, SAH05220846GA, SAH05220851GA, SAH05240875GA, SAH05250894GA, SAH05250896GA, SAH05451197GA, SAH05451199GA, SAH05451201GA, SAH06041493GA, SAH06241911GA, SAH08373784GA, SAH06091628GA, SAH06091632GA, SAH06021450GA, SAH07212757GA, SAH05140740GA, SAH05431157GA, SAH06021438GA, SAH06021449GA, SAH06021452GA, SAH06021453GA, SAH06021459GA, SAH08203539GA, SAH08473990GA, SAH08473992GA, SAH08494053GA, SAH08494062GA, SAH06241885GA, SAH06241887GA, SAH06101642GA, SAH06151768GA, SAH06241897GA, SAH06292001GA, SAH06522378GA, SAH06522404GA, SAH06522405GA, SAH06522406GA, SAH08023338GA, SAH07052531GA, SAH06041487GA, SAH07142626GA, SAH04530505GA, SAH05040578GA, SAH07403005GA, SAH07483230GA, SAH07493269GA, SAH06011427GA, SAH06011432GA, SAH04520470GA, SAH04520472GA, SAH04520474GA, SAH06041492GA, SAH08173501GA, SAH06041471GA, SAH04500414GA, SAH04500418GA, SAH05080670GA, SAH07403052GA, SAH06291982GA, SAH06332018GA, SAH06382084GA, SAH08173496GA, SAH08363761GA, SAH07483218GA, SAH07483219GA, SAH07483229GA, SAH07483234GA, SAH08473964GA, SAH05471252GA, SAH06432184GA, SAH05220842GA, SAH05220844GA, SAH05220850GA, SAH07062530GA, SAH08253597GA, SAH08303684GA, SAH07282821GA, SAH07403057GA, SAH07403061GA, SAH07403064GA, SAH09024202GA, SAH05140731GA, SAH04470329GA, SAH04470333GA, SAH04480346GA, SAH04480347GA, SAH05080656GA, SAH05471241GA, SAH05471267GA, SAH05481310GA, SAH05491360GA, SAH05491363GA, SAH05501365GA, SAH05501366GA, SAH05501367GA, SAH05501368GA, SAH05501369GA, SAH05501370GA, SAH05501371GA, SAH05501372GA, SAH05501373GA, SAH05501374GA, SAH06051510GA, SAH0610141761GA, SAH06101633GA, SAH06101634GA, SAH06101636GA, SAH06101637GA, SAH06101639GA, SAH06101641GA, SAH06141749GA, SAH06141757GA, SAH06141761GA, SAH06191862GA, SAH06241930GA, SAH06241962GA, SAH06332027GA, SAH06332031GA, SAH06332032GA, SAH06332033GA, SAH06332034GA, SAH06332037GA, SAH06332038GA, SAH06332040GA, SAH06332041GA, SAH06332042GA, SAH06332043GA, SAH06342045GA, SAH06342048GA, SAH06342056GA, SAH06522387GA, SAH06522412GA, SAH06522421GA, SAH07202726GA, SAH07372944GA, SAH07372955GA, SAH07372956GA, SAH07403002GA, SAH07403017GA, SAH07403018GA, SAH07403032GA, SAH07403033GA, SAH08473996GA, SAH08494035GA, SAH08494036GA, SAH08494037GA, SAH08494038GA, SAH08494040GA, SAH08494041GA, SAH08494042GA, SAH08494043GA, SAH08524115GA, SAH08524116GA, SAH08524117GA, SAH08524118GA, SAH08524119GA, SAH08524120GA, SAH08524121GA, SAH08524122GA, SAH08524123GA, SAH09024161GA, SAH09024162GA, SAH09024167GA, SAH09024168GA, SAH09024171GA, SAH09024172GA, SAH04430252GA, SAH07212734GA, SAH07212739GA, SAH07212741GA, SAH06041473GA, SAH06051500GA, SAH06051505GA, SAH08373798GA, SAH05431140GA, SAH05431141GA, SAH05431142GA, SAH05431145GA, SAH05431147GA, SAH07012435GA, SAH06291988GA, SAH06402106GA, SAH06402109GA, SAH06402116GA, SAH06402117GA, SAH06402118GA, SAH06402119GA, SAH06402122GA, SAH06402123GA, SAH05110685GA, SAH05110688GA, SAH05110694GA, SAH05110695GA, SAH05140718GA, SAH05140722GA, SAH05381068GA, SAH06091584GA, SAH06101658GA, SAH06101681GA, SAH06191865GA, SAH06291976GA, SAH06342059GA, SAH05331015GA, SAH06412131GA, SAH06412138GA, SAH07172699GA, SAH07202731GA, SAH07403046GA, SAH07403049GA, SAH07403050GA, SAH07443140GA, SAH07443142GA, SAH05471235GA, SAH05471240GA, SAH05471272GA, SAH08023349GA, SAH08033390GA, SAH08153471GA, SAH08173485GA, SAH07423113GA, SAH04360063GA, SAH04360066GA, SAH04430235GA, SAH04430241GA, SAH05471262GA, SAH05471265GA, SAH08353731GA, SAH08423900GA, SAH08423902GA, SAH07202717GA, SAH06071550GA, SAH06051506GA, SAH06051507GA, SAH06051508GA, SAH06051509GA, SAH06051511GA, SAH06081568GA, SAH05491347GA, SAH05491348GA, SAH05491352GA, SAH05491353GA, SAH05491356GA, SAH05501376GA, SAH05501377GA, SAH06051514GA, SAH07493283GA, SAH05080675GA, SAH05110686GA, SAH05110687GA, SAH05110689GA, SAH05110692GA, SAH05080658GA, SAH05080662GA, SAH05080664GA, SAH05080680GA, SAH06141754GA, SAH06171790GA, SAH06171793GA, SAH06181814GA, SAH06181820GA, SAH08293654GA, SAH08524126GA, SAH08524127GA, SAH08524128GA, SAH08133412GA, SAH05080684GA, SAH07463176GA, SAH07463177GA, SAH07483188GA, SAH06241914GA, SAH06241917GA, SAH06241969GA, SAH04480349GA, SAH04480350GA, SAH04480355GA, SAH04490362GA, SAH05010545GA, SAH05451217GA, SAH06241926GA, SAH06241928GA, SAH06241932GA, SAH06292004GA, SAH06412148GA, SAH07062540GA, SAH07062548GA, SAH08443944GA, SAH05280941GA, SAH05280948GA, SAH06091617GA, SAH06091619GA, SAH06121729GA, SAH06151772GA, SAH07423085GA, SAH07423087GA, SAH07423095GA, SAH06021435GA, SAH05140716GA, SAH05140720GA, SAH06091614GA, SAH06091616GA, SAH06091621GA, SAH06091622GA, SAH06091627GA, SAH06121720GA, SAH06121726GA, SAH04480335GA, SAH05351020GA, SAH05481304GA, SAH07312892GA, SAH07312894GA, SAH07312908GA, SAH07372976GA, SAH06241886GA, SAH06241888GA, SAH06241935GA, SAH06241941GA, SAH07052543GA, SAH07102592GA, SAH07102593GA, SAH07102597GA, SAH07102600GA, SAH07102601GA, SAH07102608GA, SAH07423067GA, SAH07423077GA, SAH07493239GA, SAH07493244GA, SAH07493266GA, SAH08153472GA, SAH08173478GA, SAH08173479GA, SAH08203505GA, SAH08203515GA, SAH08203534GA, SAH08203538GA, SAH07463156GA, SAH07463159GA, SAH07463184GA, SAH06332014GA, SAH06332020GA, SAH06332021GA, SAH06332023GA, SAH06502346GA, SAH06502352GA, SAH06502353GA, SAH07312893GA, SAH06101690GA, SAH06181819GA, SAH06181830GA, SAH08494051GA, SAH07342914GA, SAH07342916GA, SAH07483226GA, SAH08494015GA, SAH08494016GA, SAH08494017GA, SAH08494018GA, SAH08494020GA, SAH08494021GA, SAH08494022GA, SAH08494029GA, SAH05351016GA, SAH05351017GA, SAH05351023GA, SAH05351026GA, SAH05351033GA, SAH05351034GA, SAH07012445GA, SAH07062552GA, SAH04520484GA, SAH04400172GA, SAH05431154GA, SAH05381072GA, SAH05381074GA, SAH05391081GA, SAH05391082GA, SAH07172701GA, SAH07202722GA, SAH04360073GA, SAH07022473GA, SAH07282806GA, SAH07282809GA, SAH07423069GA, SAH08023352GA, SAH05441171GA, SAH05441173GA, SAH05441188GA, SAH05441193GA, SAH06241895GA, SAH06241900GA, SAH06241956GA, SAH04480348GA, SAH06181823GA, SAH07493282GA, SAH06081573GA, SAH06181816GA, SAH06181827GA, SAH06181835GA, SAH06181840GA, SAH06181843GA, SAH06181848GA, SAH06482338GA, SAH07232770GA, SAH07232774GA, SAH07232775GA, SAH07423118GA, SAH07423123GA, SAH08253621GA, SAH08443949GA, SAH04520483GA, SAH05240887GA, SAH05471271GA, SAH05260904GA, SAH05260909GA, SAH05260911GA, SAH05260917GA, SAH07012438GA, SAH06171776GA, SAH06171777GA, SAH08504081GA, SAH04410209GA, SAH07022488GA, SAH07062528GA, SAH08343713GA, SAH08353725GA, SAH08353726GA, SAH08353727GA, SAH08353730GA, SAH08353732GA, SAH08353733GA, SAH07493253GA, SAH06241893GA, SAH06241901GA, SAH06241909GA, SAH06241910GA, SAH08383826GA, SAH08383829GA, SAH09024165GA, SAH09024178GA, SAH09024179GA, SAH09024184GA, SAH06382077GA, SAH06422164GA, SAH06422166GA, SAH06422167GA, SAH06422170GA, SAH06422172GA, SAH06432193GA, SAH06462277GA, SAH06462286GA, SAH06462289GA, SAH05070598GA, SAH05080682GA, SAH05110705GA, SAH05110707GA, SAH05110711GA, SAH05110713GA, SAH07052501GA, SAH08514100GA, SAH04470317GA, SAH04470319GA, SAH04520479GA, SAH05160828GA, SAH05351031GA, SAH05501375GA, SAH06021451GA, SAH06091605GA, SAH06091612GA, SAH06191856GA, SAH06191872GA, SAH06191873GA, SAH06452252GA, SAH07012436GA, SAH07012437GA, SAH07012440GA, SAH07012441GA, SAH07022452GA, SAH07022453GA, SAH07022459GA, SAH07022460GA, SAH07022474GA, SAH07022475GA, SAH07022477GA, SAH07022478GA, SAH07022483GA, SAH07022484GA, SAH07022489GA, SAH07022490GA, SAH07152668GA, SAH07403023GA, SAH07403024GA, SAH07403025GA, SAH07403029GA, SAH07403031GA, SAH07423066GA, SAH07493256GA, SAH07493261GA, SAH08343715GA, SAH08383828GA, SAH06412127GA, SAH05160831GA, SAH07403044GA, SAH08474001GA, SAH06382080GA, SAH06081563GA, SAH06081593GA, SAH06091590GA, SAH05491344GA, SAH06292005GA, SAH06292010GA, SAH05010519GA, SAH06442224GA, SAH06442228GA, SAH07012442GA, SAH07012448GA, SAH07282816GA, SAH07282819GA, SAH06442231GA, SAH05010525GA, SAH05040589GA, SAH05040594GA, SAH05441166GA, SAH07052509GA, SAH07493262GA, SAH08313696GA, SAH04510423GA, SAH04510435GA, SAH04510455GA, SAH05010540GA, SAH05010556GA, SAH05140742GA, SAH05160827GA, SAH05290965GA, SAH05501382GA, SAH05501384GA, SAH05521415GA, SAH05521420GA, SAH06091604GA, SAH06091626GA, SAH06091629GA, SAH06291985GA, SAH06292007GA, SAH06402105GA, SAH06402107GA, SAH06402110GA, SAH06402113GA, SAH06422175GA, SAH06422179GA, SAH06502361GA, SAH07142631GA, SAH07152644GA, SAH07152671GA, SAH07202704GA, SAH07202709GA, SAH07202710GA, SAH07292822GA, SAH07292837GA, SAH07292838GA, SAH07493248GA, SAH08373805GA, SAH08423896GA, SAH08443962GA, SAH08473974GA, SAH08473988GA, SAH08504071GA, SAH08514107GA, SAH07312887GA, SAH07342918GA, SAH07342921GA, SAH07342925GA, SAH07342933GA, SAH07342939GA, SAH07342940GA, SAH07342941GA, SAH08213568GA, SAH05080657GA, SAH05080661GA, SAH08494031GA, SAH08494032GA, SAH07022462GA, SAH07022467GA, SAH07022468GA, SAH04510444GA, SAH04510445GA, SAH04510452GA, SAH05521406GA, SAH08473989GA, SAH08494054GA, SAH05010513GA, SAH07493241GA, SAH08373812GA, SAH06241907GA, SAH06241908GA, SAH06241913GA, SAH08504083GA, SAH08524130GA, SAH08524131GA, SAH09014145GA, SAH07212758GA, SAH07212759GA, SAH07212761GA, SAH07292848GA, SAH07423076GA, SAH07483205GA, SAH06332017GA, SAH06332025GA, SAH06332026GA, SAH07493299GA, SAH05260906GA, SAH05260912GA, SAH05351027GA, SAH05351028GA, SAH05351042GA, SAH05351043GA, SAH05431144GA, SAH08494049GA, SAH05320987GA, SAH07403048GA, SAH04430237GA, SAH08343706GA, SAH08293655GA, SAH08443960GA, SAH06462298GA, SAH06462308GA, SAH08473981GA, SAH08494063GA, SAH07022454GA, SAH07022457GA, SAH07022463GA, SAH07022469GA, SAH08303680GA, SAH04420223GA, SAH05421122GA, SAH06101649GA, SAH06181815GA, SAH06181837GA, SAH06181838GA, SAH06181841GA, SAH06181842GA, SAH06181846GA, SAH04290012GA, SAH04290013GA, SAH04290023GA, SAH07372984GA, SAH07372985GA, SAH07342923GA, SAH07342930GA, SAH05220861GA, SAH06121718GA, SAH08303666GA, SAH07022491GA, SAH07052523GA, SAH07062532GA, SAH07463170GA, SAH07463180GA, SAH07483187GA, SAH07443154GA, SAH08343717GA, SAH05481306GA, SAH04490365GA, SAH05040586GA, SAH07423098GA, SAH07423099GA, SAH07423104GA, SAH07463173GA, SAH07403003GA, SAH07403020GA, SAH07463165GA, SAH05240878GA, SAH05280939GA, SAH05290970GA, SAH07503320GA, SAH06291990GA, SAH08133399GA, SAH08373814GA, SAH08373820GA, SAH08373824GA, SAH06081566GA, SAH06241959GA, SAH07372949GA, SAH07403039GA, SAH07423070GA, SAH07423081GA, SAH07493235GA, SAH07493236GA, SAH07493237GA, SAH07493240GA, SAH07493242GA, SAH07493288GA, SAH07493289GA, SAH07493290GA, SAH07493291GA, SAH07493292GA, SAH07493293GA, SAH07493294GA, SAH04400171GA, SAH04400173GA, SAH08494006GA, SAH08494009GA, SAH09024194GA, SAH06241916GA, SAH06241920GA, SAH06051504GA, SAH07022476GA, SAH07372962GA, SAH08033379GA, SAH06342063GA, SAH07483216GA, SAH07372942GA, SAH04360067GA, SAH06292013GA, SAH04450268GA, SAH04450269GA, SAH06241921GA, SAH07022465GA, SAH05391090GA, SAH06402108GA, SAH05471236GA, SAH05471275GA, SAH05240885GA, SAH06011428GA, SAH07483203GA, SAH06051519GA, SAH06061522GA, SAH06061523GA, SAH06522396GA, SAH06522420GA, SAH05391096GA, SAH05391099GA, SAH05391103GA, SAH05471277GA, SAH06462292GA, SAH06482324GA, SAH06482325GA, SAH06482327GA, SAH06482339GA, SAH06482340GA, SAH06482341GA, SAH05080625GA, SAH05080627GA, SAH05080628GA, SAH05080629GA, SAH05080631GA, SAH05080632GA, SAH05080633GA, SAH05080634GA, SAH06171783GA, SAH04430239GA, SAH06522400GA, SAH07022485GA, SAH05110697GA, SAH05471242GA, SAH06121702GA, SAH06241933GA, SAH07142627GA, SAH07202711GA, SAH07022481GA, SAH07372977GA, SAH07312896GA, SAH07423079GA, SAH04410206GA, SAH04410207GA, SAH04410208GA, SAH07342913GA, SAH07342926GA, SAH07342931GA, SAH05411108GA, SAH05431132GA, SAH05431133GA, SAH05431134GA, SAH05431136GA, SAH05431138GA, SAH05431139GA, SAH05471280GA, SAH06041485GA, SAH06041486GA, SAH07372952GA, SAH07372957GA, SAH07372978GA, SAH07142614GA, SAH07142637GA, SAH07142640GA, SAH06432197GA, SAH06432201GA, SAH06291979GA, SAH06291980GA, SAH06291983GA, SAH06291989GA, SAH07052551GA, SAH07102587GA, SAH07102588GA, SAH07102591GA, SAH05351045GA, SAH05381047GA, SAH05381048GA, SAH05381049GA, SAH05381050GA, SAH05381051GA, SAH05381052GA, SAH05381053GA, SAH05381054GA, SAH05381059GA, SAH05381062GA, SAH05381064GA, SAH05511402GA, SAH05511403GA, SAH05511404GA, SAH07493279GA, SAH07142632GA, SAH08443924GA, SAH08443947GA, SAH08443953GA, SAH08023356GA, SAH08173490GA, SAH08173494GA, SAH08213575GA, SAH08213577GA, SAH08213580GA, SAH08213581GA, SAH08213582GA, SAH08213583GA, SAH06432190GA, SAH08343708GA, SAH08363755GA, SAH08363756GA, SAH08363759GA, SAH08363762GA, SAH08363764GA, SAH08373770GA, SAH08373802GA, SAH08383844GA, SAH08403845GA, SAH08403846GA, SAH08403847GA, SAH08403848GA, SAH08403852GA, SAH08403853GA, SAH08403854GA, SAH08403856GA, SAH08403859GA, SAH08403860GA, SAH08403861GA, SAH08403863GA, SAH08443936GA, SAH08443939GA, SAH06021457GA, SAH06021461GA, SAH06021462GA, SAH06021463GA, SAH06051518GA, SAH07212736GA, SAH07212752GA, SAH08494005GA, SAH08494011GA, SAH08494012GA, SAH08494013GA, SAH08494014GA, SAH08524135GA, SAH08524136GA, SAH08524137GA, SAH08524138GA, SAH08524139GA, SAH08524140GA, SAH08524141GA, SAH08524142GA, SAH07403055GA, SAH04400168GA, SAH05421119GA, SAH05421124GA, SAH05040588GA, SAH04470328GA, SAH05040585GA, SAH05070614GA, SAH07062542GA, SAH07062544GA, SAH07062550GA, SAH07102598GA, SAH07372993GA, SAH07463172GA, SAH08443926GA, SAH08303673GA, SAH08403882GA, SAH08473980GA, SAH04520477GA, SAH08213556GA, SAH08213558GA, SAH08213563GA, SAH08213564GA, SAH08403870GA, SAH09024175GA, SAH09024204GA, SAH07282810GA, SAH07292829GA, SAH04520476GA, SAH05220853GA, SAH06091583GA, SAH06101663GA, SAH06101665GA, SAH06101666GA, SAH06101667GA, SAH06101668GA, SAH06101669GA, SAH06101670GA, SAH06101671GA, SAH06101672GA, SAH06101673GA, SAH06101674GA, SAH06101675GA, SAH06101676GA, SAH06101677GA, SAH06101678GA, SAH06101679GA, SAH06101682GA, SAH06101683GA, SAH06141746GA, SAH06141755GA, SAH06171794GA, SAH06171795GA, SAH06171799GA, SAH06171801GA, SAH06171802GA, SAH06171803GA, SAH06171806GA, SAH06171807GA, SAH06171808GA, SAH06171809GA, SAH06181817GA, SAH06181818GA, SAH06191863GA, SAH06462280GA, SAH06522431GA, SAH08023369GA, SAH08023374GA, SAH08203533GA, SAH08443963GA, SAH08473984GA, SAH06021436GA, SAH06021437GA, SAH06021439GA, SAH06021440GA, SAH06442229GA, SAH06442233GA, SAH08473969GA, SAH04350040GA, SAH04350054GA, SAH05150782GA, SAH05150788GA, SAH05150789GA, SAH05290974GA, SAH06101680GA, SAH06121710GA, SAH07372966GA, SAH07372973GA, SAH07372974GA, SAH07372975GA, SAH07372980GA, SAH07372983GA, SAH07463157GA, SAH08373795GA, SAH08373801GA, SAH08373804GA, SAH08373806GA, SAH08373808GA, SAH08373809GA, SAH08373813GA, SAH08373817GA, SAH08373819GA, SAH08403850GA, SAH08403851GA, SAH08403855GA, SAH08403857GA, SAH08403858GA, SAH08403864GA, SAH06382081GA, SAH05521416GA, SAH05521417GA, SAH05110696GA, SAH05150804GA, SAH05160813GA, SAH05220859GA, SAH05280929GA, SAH05280931GA, SAH05391086GA, SAH05421123GA, SAH05441181GA, SAH05481309GA, SAH05491315GA, SAH05511400GA, SAH05511401GA, SAH05521425GA, SAH06121706GA, SAH06121722GA, SAH06171810GA, SAH06171813GA, SAH06332015GA, SAH06332019GA, SAH06422171GA, SAH06522384GA, SAH06522418GA, SAH07282803GA, SAH07403054GA, SAH07443127GA, SAH08283644GA, SAH08303672GA, SAH08373803GA, SAH08403866GA, SAH04490373GA, SAH06101694GA, SAH05010528GA, SAH08023341GA, SAH08023345GA, SAH08023351GA, SAH08403880GA, SAH04520475GA, SAH05070599GA, SAH05080635GA, SAH05080637GA, SAH07072579GA, SAH07072581GA, SAH07072582GA, SAH07443148GA, SAH07463178GA, SAH07463179GA, SAH07463181GA, SAH07483190GA, SAH07483192GA, SAH05290967GA, SAH05290968GA, SAH05290978GA, SAH05381066GA, SAH05381070GA, SAH08153473GA, SAH05381055GA, SAH05381056GA, SAH05381057GA, SAH05381058GA, SAH05381060GA, SAH05381061GA, SAH05381063GA, SAH05010553GA, SAH06101653GA, SAH06171779GA, SAH06241891GA, SAH06522413GA, SAH07072567GA, SAH07072573GA, SAH07072574GA, SAH07072576GA, SAH07072577GA, SAH07072580GA, SAH08383825GA, SAH07072556GA, SAH07072557GA, SAH07072559GA, SAH07072562GA, SAH07072566GA, SAH08494019GA, SAH08494023GA, SAH08494024GA, SAH04360071GA, SAH04400162GA, SAH04400163GA, SAH04400164GA, SAH04400165GA, SAH04490374GA, SAH04520481GA, SAH04520482GA, SAH05220848GA, SAH05491317GA, SAH05491319GA, SAH05501381GA, SAH05521408GA, SAH05521410GA, SAH05521411GA, SAH05521412GA, SAH05521414GA, SAH05521418GA, SAH05521419GA, SAH05521421GA, SAH05521423GA, SAH05521424GA, SAH06121730GA, SAH06241924GA, SAH06412144GA, SAH06412145GA, SAH06412146GA, SAH06412147GA, SAH06412150GA, SAH06412151GA, SAH06412152GA, SAH06412153GA, SAH06422154GA, SAH06422155GA, SAH06422156GA, SAH06422157GA, SAH06422158GA, SAH06422159GA, SAH06422160GA, SAH07052511GA, SAH04410191GA, SAH04510449GA, SAH04520480GA, SAH05471278GA, SAH05491343GA, SAH04400160GA, SAH07052512GA, SAH05381071GA, SAH07052507GA, SAH05010546GA, SAH06241877GA, SAH06241912GA, SAH08383831GA, SAH08383832GA, SAH08383833GA, SAH08383834GA, SAH08383839GA, SAH08383840GA, SAH08383842GA, SAH08443948GA, SAH06121715GA, SAH06121732GA, SAH06121733GA, SAH08373777GA, SAH06061527GA, SAH06061529GA, SAH06061531GA, SAH06061533GA, SAH06061535GA, SAH05080672GA, SAH05491338GA, SAH05511399GA, SAH07463166GA, SAH07463169GA, SAH08213554GA, SAH08494025GA, SAH08494027GA, SAH08494034GA, SAH08494048GA, SAH07342932GA, SAH07342934GA, SAH07342935GA, SAH07342938GA, SAH06181834GA, SAH07372982GA, SAH07372987GA, SAH07373000GA, SAH08153468GA, SAH04500406GA, SAH05381073GA, SAH05080673GA, SAH08153466GA, SAH08153469GA, SAH07202730GA, SAH08443958GA, SAH05220837GA, SAH05240867GA, SAH05240868GA, SAH07022472GA, SAH07022479GA, SAH07292843GA, SAH05220836GA, SAH05240889GA, SAH05250898GA, SAH05280946GA, SAH05290963GA, SAH05290972GA, SAH05351035GA, SAH05351039GA, SAH05351040GA, SAH05351044GA, SAH05391101GA, SAH05451208GA, SAH08373787GA, SAH08373788GA, SAH05491339GA, SAH05010557GA, SAH07493249GA, SAH07493254GA, SAH04490382GA, SAH05431135GA, SAH07072569GA, SAH07072570GA, SAH05411107GA, SAH08473971GA, SAH05080678GA, SAH05110712GA, SAH06522430GA, SAH07142622GA, SAH07142628GA, SAH07142635GA, SAH07202707GA, SAH07202708GA, SAH07202715GA, SAH07202716GA, SAH07372981GA, SAH07483214GA, SAH07493280GA, SAH07493285GA, SAH07493286GA, SAH07493287GA, SAH08133406GA, SAH08133407GA, SAH08133408GA, SAH08133410GA, SAH08133411GA, SAH08133415GA, SAH08133417GA, SAH08133418GA, SAH08133419GA, SAH08133421GA, SAH08253588GA, SAH08253591GA, SAH08363754GA, SAH08373800GA, SAH08423895GA, SAH08423908GA, SAH08423912GA, SAH08443950GA, SAH08473966GA, SAH08473967GA, SAH08473976GA, SAH08473977GA, SAH08473978GA, SAH07282802GA, SAH07403036GA, SAH07403038GA, SAH08153465GA, SAH08203526GA, SAH04350051GA, SAH04350053GA, SAH04380091GA, SAH04390158GA, SAH04500410GA, SAH05010524GA, SAH05150780GA, SAH08033392GA, SAH08023368GA, SAH08033381GA, SAH08033388GA, SAH08143444GA, SAH08283626GA, SAH06121698GA, SAH06121703GA, SAH06171778GA, SAH06171780GA, SAH06171788GA, SAH06171789GA, SAH06171805GA, SAH07232784GA, SAH08313700GA, SAH08313701GA, SAH06482315GA, SAH06482323GA, SAH05481297GA, SAH07062524GA, SAH08033378GA, SAH04450285GA, SAH04460284GA, SAH04460286GA, SAH04460287GA, SAH04460288GA, SAH04460289GA, SAH04460290GA, SAH04460291GA, SAH04460292GA, SAH04460293GA, SAH04460294GA, SAH04460295GA, SAH04470309GA, SAH05220862GA, SAH06332029GA, SAH06342051GA, SAH08153445GA, SAH08153446GA, SAH08153447GA, SAH08153448GA, SAH08153449GA, SAH08153451GA, SAH08153452GA, SAH08153453GA, SAH08153454GA, SAH05431159GA, SAH06241892GA, SAH09024166GA, SAH05501386GA, SAH05511390GA, SAH05511391GA, SAH07463158GA, SAH04470324GA, SAH06021447GA, SAH06021456GA, SAH06181853GA, SAH06462281GA, SAH09014154GA, SAH09024160GA, SAH08474003GA, SAH06241899GA, SAH06241905GA, SAH06462278GA, SAH06462293GA, SAH07052506GA, SAH06011429GA, SAH07282792GA, SAH07282798GA, SAH07282801GA, SAH04510442GA, SAH05240876GA, SAH05240877GA, SAH05250903GA, SAH05290982GA, SAH05471261GA, SAH05481292GA, SAH06101655GA, SAH06121723GA, SAH07302862GA, SAH07302865GA, SAH07302867GA, SAH07302872GA, SAH07302873GA, SAH07302875GA, SAH07302876GA, SAH07302880GA, SAH07423091GA, SAH07423100GA, SAH07443125GA, SAH08023358GA, SAH08023360GA, SAH08023362GA, SAH08203517GA, SAH08203537GA, SAH08253600GA, SAH08253606GA, SAH08253610GA, SAH08253613GA, SAH08253615GA, SAH08253618GA, SAH08253619GA, SAH08253623GA, SAH08253624GA, SAH08283625GA, SAH08283627GA, SAH08283630GA, SAH08283633GA, SAH08283636GA, SAH08283637GA, SAH08283640GA, SAH08283642GA, SAH08293649GA, SAH08293651GA, SAH08293652GA, SAH08293653GA, SAH08293664GA, SAH08313690GA, SAH08313692GA, SAH08373776GA, SAH08373791GA, SAH08383830GA, SAH08403871GA, SAH04470332GA, SAH04520463GA, SAH05110698GA, SAH05110703GA, SAH05160808GA, SAH05260905GA, SAH05320991GA, SAH05501389GA, SAH06241876GA, SAH06241880GA, SAH06522382GA, SAH07423082GA, SAH07443146GA, SAH07463162GA, SAH07463168GA, SAH07483191GA, SAH07483194GA, SAH07483225GA, SAH07483227GA, SAH08494056GA, SAH08494059GA, SAH04340033GA, SAH04350039GA, SAH04350041GA, SAH04350042GA, SAH04350047GA, SAH04350050GA, SAH04360062GA, SAH04360072GA, SAH04360074GA, SAH04360075GA, SAH04360077GA, SAH04360078GA, SAH04360079GA, SAH04360080GA, SAH06432199GA, SAH04340036GA, SAH04480341GA, SAH06442216GA, SAH06442217GA, SAH06442219GA, SAH06452246GA, SAH06462275GA, SAH07443149GA, SAH08473965GA, SAH07493245GA, SAH07493246GA, SAH04520464GA, SAH04520465GA, SAH04530488GA, SAH05070596GA, SAH05070601GA, SAH05070609GA, SAH06041470GA, SAH07152642GA, SAH07152646GA, SAH07152648GA, SAH07152650GA, SAH05501383GA, SAH08473998GA, SAH07342919GA, SAH07312891GA, SAH07342912GA, SAH07342917GA, SAH07342922GA, SAH07342924GA, SAH07342928GA, SAH06241904GA, SAH05491349GA, SAH05491354GA, SAH05320986GA, SAH07493301GA, SAH07493303GA, SAH07503317GA, SAH05080646GA, SAH08474002GA, SAH05431137GA, SAH05431149GA, SAH05441177GA, SAH05441178GA, SAH05441182GA, SAH06041478GA, SAH08303675GA, SAH08303676GA, SAH05220856GA, SAH05240886GA, SAH05240890GA, SAH05240892GA, SAH07142636GA, SAH07202703GA, SAH07202714GA, SAH08293658GA, SAH04530492GA, SAH06021454GA, SAH06021458GA, SAH08423910GA, SAH08423915GA, SAH08423916GA, SAH08423917GA, SAH08423918GA, SAH08423920GA, SAH08443925GA, SAH06241951GA, SAH06241971GA, SAH08343724GA, SAH07282820GA, SAH04450270GA, SAH06041475GA, SAH06041479GA, SAH05280950GA, SAH06342057GA, SAH06291987GA, SAH04450273GA, SAH04450274GA, SAH04450277GA, SAH04450278GA, SAH05010549GA, SAH05431158GA, SAH05431162GA, SAH05431163GA, SAH05441176GA, SAH05441180GA, SAH05441183GA, SAH06241955GA, SAH06292002GA, SAH06292008GA, SAH07172692GA, SAH07172693GA, SAH08303674GA, SAH08443952GA, SAH08443954GA, SAH08443955GA, SAH06482314GA, SAH06181832GA, SAH07403028GA, SAH05220854GA, SAH05220855GA, SAH05220858GA, SAH05220864GA, SAH05240872GA, SAH05290964GA, SAH05431148GA, SAH08033383GA, SAH08033384GA, SAH06522389GA, SAH05080647GA, SAH05080648GA, SAH05080649GA, SAH05280938GA, SAH05280944GA, SAH05280945GA, SAH05280953GA, SAH05321005GA, SAH05321007GA, SAH05321008GA, SAH05321009GA, SAH05321010GA, SAH05321011GA, SAH05321013GA, SAH05321014GA, SAH07142616GA, SAH04530490GA, SAH06241918GA, SAH08173486GA, SAH08213576GA, SAH05080645GA, SAH05080650GA, SAH05150784GA, SAH07062527GA, SAH06241883GA, SAH05471281GA, SAH05481314GA, SAH06402114GA, SAH06412124GA, SAH06412128GA, SAH06412130GA, SAH06412133GA, SAH06412136GA, SAH06432211GA, SAH07403056GA, SAH07403062GA, SAH08143433GA, SAH05080638GA, SAH09024182GA, SAH09024183GA, SAH07493265GA, SAH05290979GA, SAH08423897GA, SAH05471264GA, SAH06121714GA, SAH06121721GA, SAH06181847GA, SAH06191858GA, SAH06422174GA, SAH06522417GA, SAH08524125GA, SAH09024177GA, SAH06071538GA, SAH06241954GA, SAH06292000GA, SAH06342060GA, SAH06422177GA, SAH06442234GA, SAH06482326GA, SAH07072558GA, SAH07072563GA, SAH07102610GA, SAH07312903GA, SAH07312907GA, SAH07493309GA, SAH07493314GA, SAH08143436GA, SAH08143442GA, SAH08173475GA, SAH08173484GA, SAH09024188GA, SAH09024189GA, SAH05140725GA, SAH05140726GA, SAH06241923GA, SAH04420222GA, SAH04420234GA, SAH04490367GA, SAH04490372GA, SAH04490375GA, SAH04490378GA, SAH04490381GA, SAH07312886GA, SAH05411111GA, SAH09014146GA, SAH09014147GA, SAH09014148GA, SAH09014149GA, SAH09014150GA, SAH09014151GA, SAH09014152GA, SAH09014153GA, SAH09024156GA, SAH04450271GA, SAH04450280GA, SAH04450281GA, SAH04450282GA, SAH04450283GA, SAH07212760GA, SAH05110702GA, SAH07282793GA, SAH07282794GA, SAH07282796GA, SAH07282800GA, SAH06081576GA, SAH05010522GA, SAH06101657GA, SAH06101662GA, SAH05240865GA, SAH05240866GA, SAH05250901GA, SAH04430251GA, SAH04480343GA, SAH05511394GA, SAH05511396GA, SAH08203508GA, SAH08203509GA, SAH08203510GA, SAH08203511GA, SAH08203512GA, SAH08203513GA, SAH08203514GA, SAH08203518GA, SAH08203519GA, SAH08203522GA, SAH08203523GA, SAH08203524GA, SAH08203535GA, SAH08203540GA, SAH09024169GA, SAH09024173GA, SAH09024174GA, SAH05080630GA, SAH05080639GA, SAH05150785GA, SAH04480336GA, SAH04530501GA, SAH04530506GA, SAH04530507GA, SAH07423115GA, SAH08443934GA, SAH05010533GA, SAH05010541GA, SAH05010550GA, SAH05010554GA, SAH05010561GA, SAH08373810GA, SAH08373816GA, SAH06051520GA, SAH06051521GA, SAH06442235GA, SAH06442236GA, SAH06442238GA, SAH06442239GA, SAH06442240GA, SAH06442241GA, SAH06442242GA, SAH06442243GA, SAH06452244GA, SAH06452245GA, SAH06452248GA, SAH06462279GA, SAH08373779GA, SAH08373818GA, SAH07292847GA, SAH05240870GA, SAH05080626GA, SAH06382067GA, SAH06382073GA, SAH06382074GA, SAH06382075GA, SAH06382076GA, SAH06382078GA, SAH06382082GA, SAH06382083GA, SAH07282814GA, SAH07403001GA, SAH07443144GA, SAH07463174GA, SAH08023339GA, SAH08023342GA, SAH08023350GA, SAH08023373GA, SAH07102589GA, SAH07102590GA, SAH07102594GA, SAH05040590GA, SAH06432212GA, SAH08494045GA, SAH08494046GA, SAH08494047GA, SAH06091595GA, SAH06091596GA, SAH06091597GA, SAH06091598GA, SAH06091599GA, SAH06091601GA, SAH06091607GA, SAH07372972GA, SAH07493251GA, SAH07493252GA, SAH06522376GA, SAH06522381GA, SAH06522394GA, SAH08253616GA, SAH07202713GA, SAH08373807GA, SAH08373821GA, SAH04350043GA, SAH04420220GA, SAH04420226GA, SAH04480340GA, SAH04480342GA, SAH05010539GA, SAH05040573GA, SAH05040577GA, SAH05040579GA, SAH05040582GA, SAH05040593GA, SAH05070615GA, SAH05070620GA, SAH05070622GA, SAH05140721GA, SAH05140723GA, SAH05421118GA, SAH06432189GA, SAH06432195GA, SAH07292842GA, SAH07292850GA, SAH08213562GA, SAH08213570GA, SAH08494010GA, SAH04400167GA, SAH05280934GA, SAH06071536GA, SAH05391087GA, SAH06081572GA, SAH06071542GA, SAH06071552GA, SAH06081560GA, SAH06081580GA, SAH06081575GA, SAH06081579GA, SAH06101656GA, SAH06101661GA, SAH06121700GA, SAH06141748GA, SAH06332039GA, SAH07142612GA, SAH07292844GA, SAH07403015GA, SAH06101648GA, SAH06121696GA, SAH06121712GA, SAH06171797GA, SAH08423905GA, SAH08423906GA, SAH09024164GA, SAH06041480GA, SAH06041483GA, SAH06462304GA, SAH07493271GA, SAH08033391GA, SAH08303677GA, SAH08373797GA, SAH06502359GA, SAH08443937GA, SAH08443942GA, SAH08443946GA, SAH04420225GA, SAH04420228GA, SAH04420230GA, SAH06522401GA, SAH07062534GA, SAH07062537GA, SAH07062541GA, SAH05150786GA, SAH05220841GA, SAH05471257GA, SAH06291997GA, SAH06291998GA, SAH06292009GA, SAH08343716GA, SAH08343718GA, SAH08443927GA, SAH08443930GA, SAH08443935GA, SAH08443941GA, SAH07142617GA, SAH05471251GA, SAH05471253GA, SAH05471254GA, SAH08283628GA, SAH08283629GA, SAH04500420GA, SAH06181844GA, SAH06181849GA, SAH06181851GA, SAH07212735GA, SAH07212743GA, SAH07212744GA, SAH07212745GA, SAH07212746GA, SAH07212748GA, SAH07342929GA, SAH08504066GA, SAH08504074GA, SAH08514102GA, SAH08514103GA, SAH08514104GA, SAH08514110GA, SAH08514111GA, SAH08524124GA, SAH04530500GA, SAH05040567GA, SAH05040570GA, SAH05040583GA, SAH05040587GA, SAH05040592GA, SAH05070607GA, SAH05080660GA, SAH05471260GA, SAH05471263GA, SAH05471266GA, SAH05501379GA, SAH06011430GA, SAH06011431GA, SAH06051494GA, SAH06051496GA, SAH06051497GA, SAH06051498GA, SAH06051499GA, SAH06241965GA, SAH06342061GA, SAH06342062GA, SAH06412142GA, SAH06422163GA, SAH07172682GA, SAH07423075GA, SAH07423080GA, SAH07423105GA, SAH07423107GA, SAH07483215GA, SAH07483217GA, SAH07483221GA, SAH07483223GA, SAH07483228GA, SAH07483231GA, SAH07483232GA, SAH07483233GA, SAH07493278GA, SAH08023372GA, SAH08253593GA, SAH08373811GA, SAH08473970GA, SAH04430240GA, SAH05160815GA, SAH05160817GA, SAH05160820GA, SAH05160821GA, SAH05160822GA, SAH06041481GA, SAH06061528GA, SAH06081569GA, SAH04430236GA, SAH04430243GA, SAH04430253GA, SAH05080655GA, SAH05080659GA, SAH05280952GA, SAH04430257GA, SAH07142641GA, SAH07172700GA, SAH04470311GA, SAH04480357GA, SAH04500413GA, SAH04500416GA, SAH04530491GA, SAH05260919GA, SAH05320985GA, SAH05320990GA, SAH05351024GA, SAH05411110GA, SAH05421128GA, SAH05421129GA, SAH05451215GA, SAH05451216GA, SAH05451222GA, SAH05461230GA, SAH05461234GA, SAH06041476GA, SAH06051513GA, SAH06091615GA, SAH06121704GA, SAH06151766GA, SAH06241874GA, SAH06241878GA, SAH06402121GA, SAH06412129GA, SAH06412132GA, SAH06432202GA, SAH06432209GA, SAH06522427GA, SAH06522428GA, SAH06522429GA, SAH09024159GA, SAH07072578GA, SAH04500421GA, SAH06171784GA, SAH06171811GA, SAH06181839GA, SAH06181852GA, SAH08213552GA, SAH06021443GA, SAH06021444GA, SAH06041466GA, SAH06041467GA, SAH06041468GA, SAH06041469GA, SAH07423111GA, SAH07483224GA, SAH07443150GA, SAH08023321GA, SAH08023322GA, SAH08023323GA, SAH08023324GA, SAH08023325GA, SAH08023326GA, SAH08023328GA, SAH08023329GA, SAH08023334GA, SAH04350055GA, SAH04410199GA, SAH04410200GA, SAH04510427GA, SAH04510431GA, SAH04510433GA, SAH04510457GA, SAH04530496GA, SAH04530502GA, SAH05080671GA, SAH05160811GA, SAH05240869GA, SAH05240871GA, SAH05320994GA, SAH05391100GA, SAH05411114GA, SAH05411115GA, SAH05471273GA, SAH05471274GA, SAH06071540GA, SAH06071541GA, SAH06071543GA, SAH06091603GA, SAH06091608GA, SAH06121695GA, SAH06121699GA, SAH06191867GA, SAH06342054GA, SAH06342055GA, SAH06432186GA, SAH06432187GA, SAH07232789GA, SAH07232790GA, SAH07312901GA, SAH07463183GA, SAH07483189GA, SAH08023370GA, SAH08173480GA, SAH08173481GA, SAH04470312GA, SAH04470313GA, SAH04500409GA, SAH04510426GA, SAH08133400GA, SAH08133403GA, SAH04480345GA, SAH06522414GA, SAH05280936GA, SAH06291999GA, SAH04530504GA, SAH06091606GA, SAH06422180GA, SAH06422165GA, SAH06462276GA, SAH04490368GA, SAH08504068GA, SAH08504075GA, SAH08504076GA, SAH08504077GA, SAH08504078GA, SAH08504084GA, SAH08514086GA, SAH08514090GA, SAH08514091GA, SAH08514092GA, SAH08514093GA, SAH08303667GA, SAH08443945GA, SAH08514097GA, SAH08514098GA, SAH08514099GA, SAH06452268GA, SAH06452269GA, SAH06452270GA, SAH06452271GA, SAH06452272GA, SAH06452273GA, SAH08023337GA, SAH08023354GA, SAH08253614GA, SAH08293656GA, SAH08173488GA, SAH08173489GA, SAH08173504GA, SAH05451204GA, SAH05451214GA, SAH05451218GA, SAH05451219GA, SAH06241967GA, SAH06452259GA, SAH06452263GA, SAH06452264GA, SAH04400170GA, SAH04400175GA, SAH04400177GA, SAH04400178GA, SAH04430249GA, SAH04430255GA, SAH04430256GA, SAH04480353GA, SAH05290976GA, SAH05431155GA, SAH05431156GA, SAH05437361GA, SAH06382086GA, SAH06382087GA, SAH06382088GA, SAH06422182GA, SAH06432206GA, SAH07102584GA, SAH08023353GA, SAH08504067GA, SAH08504072GA, SAH09024180GA, SAH09024181GA, SAH09024196GA, SAH09024197GA, SAH09024200GA, SAH09024201GA, SAH09024203GA, SAH07312902GA, SAH07312905GA, SAH07312911GA, SAH08343710GA, SAH06522373GA, SAH06522374GA, SAH06522377GA, SAH04350044GA, SAH04350048GA, SAH04420232GA, SAH04450272GA, SAH04530485GA, SAH04530486GA, SAH05010536GA, SAH05010543GA, SAH05010547GA, SAH05010563GA, SAH05010564GA, SAH05290957GA, SAH05290960GA, SAH05321012GA, SAH05351037GA, SAH05391084GA, SAH05391091GA, SAH05391094GA, SAH05391097GA, SAH05391098GA, SAH05431143GA, SAH05431150GA, SAH06091594GA, SAH06091600GA, SAH06091610GA, SAH06121701GA, SAH06121707GA, SAH06121709GA, SAH06121728GA, SAH06141734GA, SAH06141759GA, SAH06171786GA, SAH06171796GA, SAH06442230GA, SAH06522372GA, SAH06522375GA, SAH06522380GA, SAH06522402GA, SAH06522411GA, SAH07052510GA, SAH07072555GA, SAH07102609GA, SAH07152649GA, SAH07372947GA, SAH07372989GA, SAH07372990GA, SAH08303671GA, SAH08423890GA, SAH08473987GA, SAH08473993GA, SAH08484004GA, SAH08494030GA, SAH08504082GA, SAH08514085GA, SAH08514094GA, SAH08514096GA, SAH08514112GA, SAH08514113GA, SAH08514114GA, SAH08253609GA, SAH08303665GA, SAH08303682GA, SAH08313698GA, SAH08313699GA, SAH08313704GA, SAH07062526GA, SAH08143432GA, SAH06061534GA, SAH06071539GA, SAH06081554GA, SAH06081562GA, SAH06081570GA, SAH06091582GA, SAH06091589GA, SAH06091631GA, SAH06522383GA, SAH06522390GA, SAH06522415GA, SAH07403053GA, SAH07403063GA, SAH04350056GA, SAH04360076GA, SAH07022487GA, SAH07052515GA, SAH05070608GA, SAH05070613GA, SAH05070618GA, SAH06452262GA, SAH06452266GA, SAH07111313GA, SAH07152660GA, SAH04470325GA, SAH04470327GA, SAH04490360GA, SAH04490384GA, SAH06191866GA, SAH08203541GA, SAH08253585GA, SAH08253586GA, SAH08253587GA, SAH08253590GA, SAH08253592GA, SAH07372964GA, SAH07403037GA, SAH07403041GA, SAH07152647GA, SAH07172694GA, SAH07172695GA, SAH07172696GA, SAH07172697GA, SAH07172698GA, SAH07463164GA, SAH08173497GA, SAH08343714GA, SAH08343722GA, SAH06452258GA, SAH06452261GA, SAH06462299GA, SAH06462309GA, SAH06462311GA, SAH06502347GA, SAH08494028GA, SAH06081571GA, SAH06091623GA, SAH05321003GA, SAH08403875GA, SAH08423907GA, SAH08423911GA, SAH08423913GA, SAH08423922GA, SAH04500395GA, SAH04500397GA, SAH04500398GA, SAH04500400GA, SAH04500401GA, SAH04500402GA, SAH04500407GA, SAH04500411GA, SAH04500417GA, SAH04500419GA, SAH05240891GA, SAH07423117GA, SAH08023327GA, SAH08343705GA, SAH08343712GA, SAH08383837GA, SAH08403867GA, SAH08403868GA, SAH08403869GA, SAH08403872GA, SAH08403873GA, SAH08403874GA, SAH08403877GA, SAH08403878GA, SAH08403879GA, SAH08403881GA, SAH08403883GA, SAH08403884GA, SAH08494026GA, SAH04500415GA, SAH04510459GA, SAH04520473GA, SAH04520478GA, SAH06412126GA, SAH06181845GA, SAH06241934GA, SAH06241940GA, SAH06241952GA, SAH07052508GA, SAH07282808GA, SAH04350058GA, SAH04350060GA, SAH05421121GA, SAH05351025GA, SAH06081564GA, SAH06091630GA, SAH06101659GA, SAH06101660GA, SAH06121713GA, SAH06121725GA, SAH06291981GA, SAH06382070GA, SAH06522398GA, SAH07022482GA, SAH07062553GA, SAH07102602GA, SAH07102603GA, SAH07102604GA, SAH07102607GA, SAH07282804GA, SAH07292851GA, SAH07312882GA, SAH07493313GA, SAH05110699GA, SAH06101693GA, SAH06051501GA, SAH06051503GA, SAH08303683GA, SAH07342927GA, SAH04430248GA, SAH06522386GA, SAH06522391GA, SAH07312897GA, SAH07312909GA, SAH04410195GA, SAH05521422GA, SAH06402111GA, SAH07142619GA, SAH07142620GA, SAH07282817GA, SAH07403019GA, SAH07443131GA, SAH07443132GA, SAH07443133GA, SAH07443134GA, SAH07443136GA, SAH07443138GA, SAH07443139GA, SAH04420224GA, SAH04420227GA, SAH05280951GA, SAH05351019GA, SAH05431164GA, SAH08023344GA, SAH08133420GA, SAH05461225GA, SAH05461226GA, SAH05461227GA, SAH05461228GA, SAH05461229GA, SAH05461231GA, SAH05461232GA, SAH05461233GA, SAH05471239GA, SAH06061530GA, SAH08203529GA, SAH05240881GA, SAH05411105GA, SAH08213557GA, SAH06452256GA, SAH05140728GA, SAH05280940GA, SAH06041491GA, SAH06241889GA, SAH06291977GA, SAH04500412GA, SAH04510430GA, SAH07072571GA, SAH062422176GA, SAH06422176GA, SAH06011426GA, SAH07142615GA, SAH07142621GA, SAH07282795GA, SAH07302866GA, SAH07302868GA, SAH07302869GA, SAH07302870GA, SAH07302871GA, SAH05391095GA, SAH05391102GA, SAH06091624GA, SAH06091625GA, SAH07423119GA, SAH07493267GA, SAH07493238GA, SAH07493243GA, SAH06241960GA, SAH07463155GA, SAH07463161GA, SAH07463167GA, SAH07463171GA, SAH07463175GA, SAH08033389GA, SAH08033393GA, SAH05280942GA, SAH06181825GA, SAH06181829GA, SAH06181831GA, SAH07423094GA, SAH07423096GA, SAH07022480GA, SAH07022486GA, SAH07152658GA, SAH07152666GA, SAH07172686GA, SAH07212756GA, SAH05451209GA, SAH08023363GA, SAH08033377GA, SAH06151765GA, SAH08213567GA, SAH04290002GA, SAH04290003GA, SAH07312904GA, SAH04360059GA, SAH04460306GA, SAH04460307GA, SAH04530493GA, SAH04530494GA, SAH05451210GA, SAH05501385GA, SAH06432204GA, SAH06432205GA, SAH06432210GA, SAH06452247GA, SAH06452249GA, SAH06452251GA, SAH06452253GA, SAH06462287GA, SAH06462295GA, SAH06522395GA, SAH06522397GA, SAH06522407GA, SAH06522408GA, SAH06522409GA, SAH06522410GA, SAH08353734GA, SAH08363746GA, SAH08363750GA, SAH08363751GA, SAH08363752GA, SAH08373792GA, SAH07052496GA, SAH07232785GA, SAH05411106GA, SAH04290009GA, SAH07423106GA, SAH07423120GA, SAH08153467GA, SAH08153470GA, SAH08373773GA, SAH08403862GA, SAH04470326GA, SAH06241957GA, SAH06462305GA, SAH06482329GA, SAH06292012GA, SAH06332036GA, SAH06342044GA, SAH06342046GA, SAH06342047GA, SAH06342052GA, SAH06342053GA, SAH07212762GA, SAH07312899GA, SAH07483220GA, SAH05290969GA, SAH05391092GA, SAH04340035GA, SAH04350052GA, SAH06412125GA, SAH04500403GA, SAH07172683GA, SAH04290004GA, SAH04290005GA, SAH04290007GA, SAH04290010GA, SAH04290026GA, SAH07312898GA, SAH07312900GA, SAH08343723GA, SAH08343709GA, SAH04470316GA, SAH05280949GA, SAH05220847GA, SAH04510443GA, SAH04510446GA, SAH04510447GA, SAH04510450GA, SAH04510451GA, SAH04510453GA, SAH04510454GA, SAH08293648GA, SAH07052516GA, SAH06462301GA, SAH08203543GA, SAH08283635GA, SAH07493312GA, SAH07443141GA, SAH08293660GA, SAH08153460GA, SAH05140724GA, SAH07232769GA, SAH07232773GA, SAH06101654GA, SAH07423122GA, SAH07423124GA, SAH08023365GA, SAH08143429GA, SAH08153462GA, SAH08153464GA, SAH07443130GA, SAH05491361GA, SAH05491362GA, SAH06412140GA, SAH05391088GA, SAH08443932GA, SAH05040581GA, SAH05150777GA, SAH05150779GA, SAH05220838GA, SAH05220840GA, SAH05220845GA, SAH05140719GA, SAH06452255GA, SAH06091587GA, SAH05260915GA, SAH06241898GA, SAH05481295GA, SAH05481300GA, SAH05481301GA, SAH06482330GA, SAH08173500GA, SAH08173502GA, SAH04430259GA, SAH08423885GA, SAH08423887GA, SAH08494044GA, SAH07443128GA, SAH06332035GA, SAH07062549GA, SAH06522416GA, SAH07302864GA, SAH07152654GA, SAH07072572GA, SAH05110700GA, SAH07443126GA , SAH04450264GA, SAH07102611GA, SAH07423116GA, SAH09024198GA, SAH05080652GA, SAH05080653GA, SAH05070610GA, SAH08203520GA, SAH07372979GA, SAH04470330GA, SAH04500394GA, SAH06241896GA, SAH06241902GA, SAH06432213GA, SAH06522399GA, SAH07052521GA, SAH09024185GA, SAH07503319GA, SAH08033387GA, SAH06171800GA, SAH06191855GA, SAH05080644GA, SAH05471245GA, SAH05040574GA, SAH08303670GA, SAH05481296GA, SAH06241881GA, SAH05080683GA, SAH05110708GA, SAH05110709GA, SAH08173503GA, SAH05471249GA, SAH05280947GA, SAH06422173GA, SAH06522392GA, SAH07022458GA, SAH07152670GA, SAH07292834GA, SAH08143438GA, SAH08173476, SAH08173476GA, SAH04490369GA, SAH06422183GA, SAH06502348GA, SAH08133404GA, SAH07443137GA, SAH08443928GA, SAH08023340GA, SAH08023346GA, SAH08023348GA, SAH08203507GA, SAH04520468GA, SAH05010552GA, SAH05471243GA, SAH06121716GA, SAH06502344GA, SAH06502345GA, SAH06502349GA, SAH06502350GA, SAH06502351GA, SAH07062546GA, SAH07142613GA, SAH07202718GA, SAH07503315GA, SAH08033376GA, SAH05040580GA, SAH04460296GA, SAH04480352GA, SAH04480354GA, SAH04480358GA, SAH07493247GA, SAH07493250GA, SAH08313687GA, SAH06292003GA, SAH04480344GA, SAH05080681GA, SAH06081559GA, SAH07463160GA, SAH08293645GA, SAH08373765GA, SAH08373768GA, SAH08494050GA, SAH08494064GA, SAH07463182GA, SAH06482334GA, SAH08423909GA, SAH08473975GA, SAH07403004GA, SAH09024207GA, SAH08514105GA, SAH08514106GA, SAH07012449GA, SAH05320995GA, SAH05471268GA, SAH05471270GA, SAH05391104GA, SAH06522403GA, SAH04430244GA, SAH08373790GA, SAH05491331GA, SAH06522388GA, SAH06452265GA, SAH08423921GA, SAH05110704GA, SAH04410188GA, SAH08153450GA, SAH05280925GA, SAH06071551GA, SAH05110690GA, SAH05280937GA, SAH05220849GA, SAH07403009GA, SAH06121719GA, SAH07292840GA, SAH05070611GA, SAH05411112GA, SAH07022466GA, SAH07403006GA, SAH05160814GA, SAH07423065GA, SAH06432200GA, SAH05421125GA, SAH05160816GA, SAH06241968GA, SAH04410203GA, SAH05431131GA, SAH08343711GA, SAH04460305GA, SAH05471238GA, SAH08473986GA, SAH05160833GA, SAH5160832GA, SAH07012444GA, SAH06101664GA, SAH07493272GA, SAH07423072GA, SAH05320988GA, SAH07372961GA, SAH05220860GA, SAH05441168GA, SAH06241947GA, SAH06291994GA, SAH08213550GA, SAH05290955GA, SAH08504065GA, SAH07152662GA, SAH07172684GA, SAH07172684GS, SAH07202732GA, SAH05351036GA, SAH05411113GA, SAH06291996GA, SAH07052539GA, SAH07423086GA, SAH07302863GA, SAH04480351GA, SAH06241961GA, SAH05320992GA, SAH08293662GA, SAH08303678GA, SAH08313686GA, SAH06452260GA, SAH06141735GA, SAH07312910GA, SAH07172685GA, SAH06241964GA, SAH06292011GA, SAH07403026GA, SAH06241973GA, SAH08293646GA, SAH06011434GA, SAH07312883GA, SAH07292833GA, SAH07202706GA, SAH07423114GA, SAH07423110GA, SAH08173498GA, SAH07403008GA, SAH08173493GA, SAH08373786GA, SAH08443931GA, SAH07423083GA, SAH08504069GA, SAH07403010GA.SAH04430254GA, SAH08423919GA, SAH07423084GA, SAH07423068GA, SAH07423073GA, SAH07423074GA, SAH05451212GA, SAH06291975GA, SAH07292835GA, SAH07403058GA, SAH06241919GA, SAH07072564GA, SAH06071549GA, SAH06101644GA, SAH06291974GA, SAH08473991GA, SAH08473997GA, SAH06081561GA, SAH08213551GA, SAH08494061GA, SAH06482331GA, SAH06482343GA, SAH08494055GA, SAH06502360GA, SAH07423101GA, SAH06363221GA, SAH06462283GA, SAH06482333GA, SAH08203536GA, SAH05240879GA, SAH06181836GA, SAH09024195GA, SAH07403045GA, SAH07403011GA, SAH06412149GA, SAH06241972GA, SAH04340037GA, SAH04520462GA, SAH045220462GA, SAH06141741GA, SAH07152645GA, SAH0715643GA, SAH08504080GA, SAH07282797GA, SAH07463163GA, SAH08133413GA, SAH04510439GA, SAH04410187GA, SAH06432208GA, SAH05140736GA, SAH07423112GA, SAH05290962GA, SAH05240884GA, SAH06141760GA, SAH08373822GA, SAH06502358GA, SAH05451203GA, SAH08443959GA, SAH06382071GA, SAH06041488GA, SAH07062533GA, SAH07142618GA, SAH07202712GA, SAH07202719GA, SAH07202720GA, SAH08213579GA, SAH06171787GA, SAH04530489GA, SAH06051495GA, SAH09024187GA, SAH06462274GA, SAH05521409GA, SAH06171804GA, SAH06482337GA, SAH05080676GA, SAH06241903GA, SAH08173491GA, SAH08504079GA, SAH08153474GA, SAH04470331GA, SAH07372950GA, SAH08133398GA, SAH08033362GA, SAH08033382GA, SAH04480334GA, SAH08213571GA, SAH07072575GA, SAH06332028GA, SAH07403060GA, SAH06422168GA, SAH07062536GA, SAH07052513GA, SAH07102605GA, SAH07102606GA, SAH08033380GA, sah05010558ga, SAH07052547GA, SAH08373778GA, SAH08494033GA, SAH05010548GA, SAH08023355GA, SAH05140734GA, SAH04480339GA, SAH05290973GA, SAH06071544GA, SAH06522385GA, SAH07012447GA, SAH07022471GA, SAH06081557GA, SAH05080663GA, SAH0508633GA, SAH05160830GA, SAH07403051GA, SAH07022456GA, SAH08283631GA, SAH08373823GA, SAH052408826GA, sah05292295ga, SAH07312906GA, SAH06091602, SAH0702565GA, SAH05320996GA, SAH08514095GA, SAH08514101GA, SAH05481290GA, SAH05160805GA. UPDATE added SAH05080669GA, SAH05080651GA.
5. Product Code: SE2, UDI Number (Serial Numbers): 00840682109260SE217350041SA (SE217350041SA), 00840682109260SE213410181GA (SE213410181GA);
6. Product Code: SEE, UDI Number (Serial Numbers): Not Applicable (SEE11361693GA), Not Applicable (SEE11381821GA), Not Applicable (SEE11381824GA), Not Applicable (SEE12420169GA), Not Applicable (SEE15173206GA), Not Applicable (SEE15173214GA), Not Applicable (SEE09530903GA), Not Applicable (SEE10430627GA), Not Applicable (SEE16324000SA), Not Applicable (SEE16324032SA), Not Applicable (SEE12182218GA), Not Applicable (SEE12041973GA), Not Applicable (SEE12062000GA), Not Applicable (SEE12312406GA), Not Applicable (SEE12322413GA), Not Applicable (SEE12322420GA), Not Applicable (SEE12322422GA), Not Applicable (SEE12322423GA), Not Applicable (SEE12322425GA), Not Applicable (SEE12322428GA), Not Applicable (SEE12322429GA), Not Applicable (SEE12322430GA), Not Applicable (SEE12322431GA), Not Applicable (SEE12322435GA), Not Applicable (SEE12322436GA), Not Applicable (SEE12322437GA), Not Applicable (SEE12332441GA), Not Applicable (SEE12332444GA), Not Applicable (SEE12400093GA), Not Applicable (SEE12400095GA), Not Applicable (SEE13220834GA), Not Applicable (SEE15032878GA), Not Applicable (SEE15032895GA), Not Applicable (SEE15042900GA), Not Applicable (SEE15042902GA), Not Applicable (SEE15042903GA), Not Applicable (SEE15042904GA), Not Applicable (SEE15042905GA), Not Applicable (SEE15042907GA), Not Applicable (SEE15446336SA), Not Applicable (SEE15456714SA), Not Applicable (SEE15456725SA), Not Applicable (SEE15456729SA), Not Applicable (SEE15477106SA), Not Applicable (SEE15477108SA), Not Applicable (SEE15477109SA), Not Applicable (SEE15477113SA), Not Applicable (SEE15477115SA), Not Applicable (SEE16380005SA), Not Applicable (SEE16380009SA), Not Applicable (SEE16380010SA), Not Applicable (SEE16380011SA), Not Applicable (SEE16380012SA), Not Applicable (SEE16380013SA), Not Applicable (SEE16380016SA), Not Applicable (SEE16380017SA), Not Applicable (SEE16380018SA), Not Applicable (SEE16380023SA), 00840682109260SEE18500020SA (SEE18500020SA), 00840682109260SEE18500022SA (SEE18500022SA), 00840682109260SEE18500033SA (SEE18500033SA), 00840682109260SEE18500042SA (SEE18500042SA), Not Applicable (SEE11361705GA), Not Applicable (SEE11361711GA), Not Applicable (SEE11361714GA), Not Applicable (SEE11361720GA), Not Applicable (SEE11361739GA), 00840682109260SEE19120019SA (SEE19120019SA), 00840682109260SEE19120028SA (SEE19120028SA), 00840682109260SEE19150002SA (SEE19150002SA), 00840682109260SEE19160001SA (SEE19160001SA), 00840682109260SEE19160003SA (SEE19160003SA), 00840682109260SEE19160004SA (SEE19160004SA), 00840682109260SEE19160005SA (SEE19160005SA), 00840682109260SEE19160006SA (SEE19160006SA), 00840682109260SEE19160007SA (SEE19160007SA), 00840682109260SEE19160008SA (SEE19160008SA), 00840682109260SEE19160009SA (SEE19160009SA), 00840682109260SEE19160011SA (SEE19160011SA), 00840682109260SEE19160012SA (SEE19160012SA), 00840682109260SEE19160013SA (SEE19160013SA), Not Applicable (SEE15291745SA), Not Applicable (SEE15291767SA), Not Applicable (SEE15291799SA), Not Applicable (SEE12172193GA), Not Applicable (SEE11151313GA), Not Applicable (SEE11161324GA), Not Applicable (SEE16430004SA), Not Applicable (SEE16430015SA), Not Applicable (SEE16430017SA), Not Applicable (SEE16430018SA), Not Applicable (SEE16430019SA), Not Applicable (SEE16430022SA), Not Applicable (SEE16430026SA), Not Applicable (SEE16430027SA), Not Applicable (SEE16430029SA), 00840682109260SEE18290001SA (SEE18290001SA), 00840682109260SEE18040045SA (SEE18040045SA), 00840682109260SEE18040050SA (SEE18040050SA), Not Applicable (SEE12122151GA), Not Applicable (SEE12122153GA), Not Applicable (SEE12122155GA), Not Applicable (SEE12122158GA), Not Applicable (SEE14151862GA), Not Applicable (SEE14151883GA), Not Applicable (SEE14161910GA), Not Applicable (SEE14171917GA), Not Applicable (SEE14222043GA), Not Applicable (SEE14222044GA), Not Applicable (SEE14222049GA), Not Applicable (SEE14222060GA), Not Applicable (SEE14222062GA), Not Applicable (SEE14222065GA), Not Applicable (SEE14222074GA), Not Applicable (SEE14222077GA), Not Applicable (SEE14452624GA), Not Applicable (SEE15113176GA), Not Applicable (SEE15113182GA), Not Applicable (SEE09100090GA), Not Applicable (SEE09140180GA), Not Applicable (SEE09140182GA), Not Applicable (SEE09140185GA), Not Applicable (SEE09140189GA), Not Applicable (SEE09140193GA), Not Applicable (SEE09140196GA), Not Applicable (SEE11171359GA), Not Applicable (SEE11171381GA), Not Applicable (SEE11181393GA), Not Applicable (SEE11181396GA), Not Applicable (SEE11191425GA), Not Applicable (SEE15113151GA), Not Applicable (SEE16420072SA), Not Applicable (SEE16430023SA), Not Applicable (SEE11361706GA), Not Applicable (SEE11361737GA), Not Applicable (SEE15446244SA), Not Applicable (SEE15446256SA), Not Applicable (SEE10530847GA), 00840682109260SEE18040016SA (SEE18040016SA), 00840682109260SEE18040042SA (SEE18040042SA), 00840682109260SEE18040044SA (SEE18040044SA), 00840682109260SEE18040046SA (SEE18040046SA), Not Applicable (SEE12282348GA), Not Applicable (SEE12282351GA), Not Applicable (SEE12390077GA), Not Applicable (SEE12390079GA), Not Applicable (SEE12390084GA), Not Applicable (SEE15456727SA), Not Applicable (SEE11361709GA), Not Applicable (SEE11361724GA), Not Applicable (SEE11361730GA), Not Applicable (SEE13451325GA), Not Applicable (SEE13451326GA), Not Applicable (SEE13451327GA), Not Applicable (SEE11161339GA), Not Applicable (SEE11161341GA), Not Applicable (SEE11161344GA), Not Applicable (SEE11161346GA), Not Applicable (SEE11201467GA), Not Applicable (SEE11231531GA), Not Applicable (SEE11231537GA), Not Applicable (SEE11231540GA), Not Applicable (SEE15312100SA), Not Applicable (SEE15261403SA), Not Applicable (SEE14171915GA), Not Applicable (SEE14171926GA), Not Applicable (SEE14171927GA), Not Applicable (SEE12510304GA), Not Applicable (SEE12510316GA), Not Applicable (SEE12520340GA), Not Applicable (SEE12520345GA), Not Applicable (SEE12520348GA), Not Applicable (SEE13000390GA), Not Applicable (SEE13000391GA), Not Applicable (SEE11181394GA), Not Applicable (SEE12182215GA), Not Applicable (SEE12182223GA), Not Applicable (SEE12292367GA), Not Applicable (SEE13150668GA), Not Applicable (SEE09040063GA), Not Applicable (SEE16221554SA), Not Applicable (SEE16221560SA), Not Applicable (SEE16221585SA), Not Applicable (SEE16221587SA), Not Applicable (SEE16221588SA), Not Applicable (SEE16221591SA), Not Applicable (SEE16221592SA), Not Applicable (SEE12410122GA), Not Applicable (SEE12410130GA), Not Applicable (SEE12410143GA), Not Applicable (SEE12420150GA), Not Applicable (SEE12172201GA), Not Applicable (SEE12172204GA), 00840682109260SEE19220024SA (SEE19220024SA), 00840682109260SEE19220035SA (SEE19220035SA), 00840682109260SEE18370005SA (SEE18370005SA), Not Applicable (SEE11361710GA), Not Applicable (SEE11361718GA), Not Applicable (SEE14272163GA), Not Applicable (SEE14272165GA), Not Applicable (SEE14272169GA), Not Applicable (SEE14272176GA), Not Applicable (SEE14272202GA), Not Applicable (SEE14272204GA), Not Applicable (SEE14272217GA), Not Applicable (SEE14292230GA), Not Applicable (SEE14292234GA), Not Applicable (SEE14292235GA), Not Applicable (SEE14292236GA), Not Applicable (SEE15363354SA), Not Applicable (SEE15363355SA), Not Applicable (SEE13240862GA), Not Applicable (SEE13351012GA), Not Applicable (SEE13451329GA), Not Applicable (SEE13451332GA), 00840682109260SEE19100053SA (SEE19100053SA), Not Applicable (SEE14151815GA), Not Applicable (SEE09100130GA), Not Applicable (SEE09230279GA), Not Applicable (SEE10420570GA), Not Applicable (SEE10460712GA), Not Applicable (SEE12430189GA), Not Applicable (SEE12430195GA), Not Applicable (SEE12430197GA), Not Applicable (SEE12430211GA), Not Applicable (SEE12430219GA), Not Applicable (SEE09530899GA), 00840682109260SEE17450001SA (SEE17450001SA), 00840682109260SEE17450002SA (SEE17450002SA), Not Applicable (SEE10410502GA), Not Applicable (SEE10420551GA), Not Applicable (SEE11371755GA), Not Applicable (SEE11371763GA), Not Applicable (SEE15415110SA), Not Applicable (SEE15415111SA), Not Applicable (SEE15415114SA), Not Applicable (SEE15415115SA), Not Applicable (SEE15415116SA), Not Applicable (SEE15415117SA), Not Applicable (SEE15415164SA), Not Applicable (SEE11201470GA), Not Applicable (SEE16262162SA), Not Applicable (SEE16272328SA), Not Applicable (SEE13391130GA), Not Applicable (SEE15113121GA), Not Applicable (SEE15113134GA), Not Applicable (SEE14462659GA), Not Applicable (SEE14462669GA), Not Applicable (SEE14462670GA), Not Applicable (SEE14462671GA), Not Applicable (SEE10270274GA), Not Applicable (SEE10400374GA), Not Applicable (SEE10400442GA), Not Applicable (SEE10400443GA), Not Applicable (SEE10400459GA), Not Applicable (SEE10410480GA), Not Applicable (SEE10410483GA), Not Applicable (SEE10410507GA), Not Applicable (SEE10410511GA), Not Applicable (SEE15415093SA), 00840682109260SEE18290012SA (SEE18290012SA), 00840682109260SEE17180002SA (SEE17180002SA), 00840682109260SEE17350011SA (SEE17350011SA), Not Applicable (SEE11071010GA), Not Applicable (SEE11071016GA), Not Applicable (SEE12092067GA), Not Applicable (SEE12242267GA), Not Applicable (SEE12252297GA), Not Applicable (SEE12252306GA), Not Applicable (SEE12370056GA), Not Applicable (SEE13190793GA), Not Applicable (SEE13190794GA), Not Applicable (SEE13190795GA), Not Applicable (SEE13190796GA), Not Applicable (SEE13190797GA), Not Applicable (SEE13190799GA), Not Applicable (SEE13190800GA), Not Applicable (SEE13190801GA), Not Applicable (SEE13190802GA), Not Applicable (SEE15528205SA), Not Applicable (SEE15528213SA), Not Applicable (SEE15528214SA), Not Applicable (SEE15528215SA), Not Applicable (SEE15528216SA), Not Applicable (SEE16303219SA), 00840682109260SEE18040049SA (SEE18040049SA), 00840682109260SEE18290002SA (SEE18290002SA), 00840682109260SEE18290007SA (SEE18290007SA), 00840682109260SEE18290011SA (SEE18290011SA), 00840682109260SEE18290016SA (SEE18290016SA), 00840682109260SEE18390002SA (SEE18390002SA), 00840682109260SEE18390005SA (SEE18390005SA), 00840682109260SEE18390014SA (SEE18390014SA), 00840682109260SEE18390020SA (SEE18390020SA), 00840682109260SEE18400002SA (SEE18400002SA), 00840682109260SEE18400004SA (SEE18400004SA), 00840682109260SEE18400006SA (SEE18400006SA), 00840682109260SEE18400009SA (SEE18400009SA), 00840682109260SEE18400010SA (SEE18400010SA), 00840682109260SEE18400011SA (SEE18400011SA), 00840682109260SEE18400014SA (SEE18400014SA), 00840682109260SEE18400020SA (SEE18400020SA), 00840682109260SEE18470023SA (SEE18470023SA), 00840682109260SEE18510015SA (SEE18510015SA), 00840682109260SEE18510016SA (SEE18510016SA), 00840682109260SEE18510019SA (SEE18510019SA), 00840682109260SEE19110024SA (SEE19110024SA), Not Applicable (SEE15093065GA), Not Applicable (SEE15093069GA), Not Applicable (SEE15113127GA), Not Applicable (SEE15113133GA), Not Applicable (SEE11081200GA), Not Applicable (SEE11081202GA), Not Applicable (SEE11101214GA), Not Applicable (SEE11101223GA), Not Applicable (SEE11111227GA), Not Applicable (SEE11111229GA), Not Applicable (SEE11111230GA), Not Applicable (SEE11111235GA), Not Applicable (SEE11111242GA), Not Applicable (SEE11111243GA), Not Applicable (SEE12430212GA), Not Applicable (SEE12430214GA), Not Applicable (SEE13010401GA), Not Applicable (SEE13401145GA), Not Applicable (SEE13401174GA), Not Applicable (SEE13401176GA), Not Applicable (SEE13401180GA), Not Applicable (SEE13401185GA), Not Applicable (SEE14031603GA), Not Applicable (SEE15312074SA), Not Applicable (SEE15312075SA), Not Applicable (SEE15312076SA), Not Applicable (SEE15312077SA), Not Applicable (SEE15312078SA), Not Applicable (SEE15312079SA), Not Applicable (SEE15312081SA), Not Applicable (SEE15312082SA), Not Applicable (SEE15312083SA), Not Applicable (SEE15312084SA), Not Applicable (SEE15312085SA), Not Applicable (SEE15312086SA), 00840682109260SEE17130002SA (SEE17130002SA), 00840682109260SEE17130004SA (SEE17130004SA), 00840682109260SEE17130007SA (SEE17130007SA), Not Applicable (SEE10410509GA), Not Applicable (SEE14402490GA), 00840682109260SEE17260003SA (SEE17260003SA), 00840682109260SEE17260004SA (SEE17260004SA), Not Applicable (SEE12322438GA), Not Applicable (SEE16324006SA), Not Applicable (SEE14101787GA), Not Applicable (SEE14111804GA), Not Applicable (SEE10450677GA), Not Applicable (SEE10450683GA), Not Applicable (SEE15113116GA), Not Applicable (SEE15113130GA), Not Applicable (SEE15113136GA), Not Applicable (SEE15240991SA), Not Applicable (SEE09380531GA), Not Applicable (SEE09390548GA), Not Applicable (SEE16028367SA), 00840682109260SEE18400019SA (SEE18400019SA), Not Applicable (SEE12340005GA), Not Applicable (SEE12340013GA), Not Applicable (SEE12360037GA), Not Applicable (SEE12360040GA), Not Applicable (SEE15352668SA), Not Applicable (SEE15352678SA), Not Applicable (SEE15538297SA), Not Applicable (SEE16028320SA), Not Applicable (SEE16028321SA), 00840682109260SEE19220021SA (SEE19220021SA), Not Applicable (SEE09530855GA), Not Applicable (SEE09530857GA), Not Applicable (SEE09530860GA), Not Applicable (SEE11201442GA), Not Applicable (SEE11201443GA), Not Applicable (SEE11201454GA), Not Applicable (SEE11231544GA), Not Applicable (SEE15446199SA), Not Applicable (SEE15446213SA), 00840682109260SEE17420002SA (SEE17420002SA), 00840682109260SEE17500019SA (SEE17500019SA), 00840682109260SEE19130011SA (SEE19130011SA), Not Applicable (SEE13040415GA), Not Applicable (SEE12092062GA), Not Applicable (SEE12092064GA), Not Applicable (SEE12092065GA), Not Applicable (SEE12430183GA), Not Applicable (SEE12430193GA), Not Applicable (SEE12430204GA), Not Applicable (SEE15220797SA), Not Applicable (SEE15220798SA), Not Applicable (SEE13351022GA), Not Applicable (SEE13351036GA), Not Applicable (SEE13351042GA), Not Applicable (SEE13351046GA), Not Applicable (SEE13351049GA), 00840682109260SEE18030014SA (SEE18030014SA), 00840682109260SEE18030015SA (SEE18030015SA), 00840682109260SEE18030016SA (SEE18030016SA), 00840682109260SEE18030018SA (SEE18030018SA), 00840682109260SEE18030020SA (SEE18030020SA), 00840682109260SEE18410023SA (SEE18410023SA), Not Applicable (SEE10121006GA), Not Applicable (SEE10121024GA), Not Applicable (SEE10130076GA), Not Applicable (SEE11201451GA), Not Applicable (SEE11371792GA), Not Applicable (SEE11371812GA), Not Applicable (SEE12370059GA), Not Applicable (SEE12370064GA), Not Applicable (SEE12370067GA), Not Applicable (SEE13371087GA), Not Applicable (SEE13371090GA), Not Applicable (SEE13371103GA), Not Applicable (SEE12021906GA), Not Applicable (SEE12021910GA), Not Applicable (SEE14151823GA), Not Applicable (SEE14151833GA), Not Applicable (SEE14482688GA), Not Applicable (SEE15052909GA), Not Applicable (SEE15052937GA), Not Applicable (SEE16089178SA), Not Applicable (SEE15120553SA), Not Applicable (SEE15120554SA), Not Applicable (SEE15120555SA), Not Applicable (SEE15120557SA), Not Applicable (SEE15210700SA), Not Applicable (SEE15210701SA), Not Applicable (SEE13110516GA), Not Applicable (SEE09250308GA), Not Applicable (SEE09250309GA), Not Applicable (SEE09250315GA), Not Applicable (SEE09250326GA), Not Applicable (SEE10120011GA), Not Applicable (SEE10120018GA), Not Applicable (SEE10121018GA), Not Applicable (SEE10260239GA), Not Applicable (SEE10270256GA), Not Applicable (SEE14031589GA), Not Applicable (SEE14031608GA), Not Applicable (SEE14031610GA), Not Applicable (SEE14031611GA), Not Applicable (SEE14031616GA), Not Applicable (SEE14031627GA), Not Applicable (SEE15528071SA), 00840682109260SEE17020023SA (SEE17020023SA), 00840682109260SEE17020029SA (SEE17020029SA), 00840682109260SEE19170006SA (SEE19170006SA), 00840682109260SEE19170007SA (SEE19170007SA), 00840682109260SEE19170009SA (SEE19170009SA), 00840682109260SEE19170012SA (SEE19170012SA), 00840682109260SEE19170015SA (SEE19170015SA), 00840682109260SEE19170019SA (SEE19170019SA), 00840682109260SEE19170022SA (SEE19170022SA), 00840682109260SEE19170024SA (SEE19170024SA), 00840682109260SEE19170038SA (SEE19170038SA), 00840682109260SEE19170039SA (SEE19170039SA), 00840682109260SEE19170040SA (SEE19170040SA), 00840682109260SEE19170042SA (SEE19170042SA), Not Applicable (SEE09530914GA), Not Applicable (SEE13120573GA), Not Applicable (SEE13120576GA), 00840682109260SEE19220001SA (SEE19220001SA), 00840682109260SEE19220003SA (SEE19220003SA), Not Applicable (SEE11361742GA), Not Applicable (SEE13270894GA), Not Applicable (SEE13270899GA), Not Applicable (SEE13270901GA), Not Applicable (SEE13280913GA), Not Applicable (SEE13280918GA), 00840682109260SEE17220002SA (SEE17220002SA), 00840682109260SEE17220004SA (SEE17220004SA), 00840682109260SEE17220005SA (SEE17220005SA), Not Applicable (SEE09020029GA), Not Applicable (SEE09340440GA), Not Applicable (SEE11371766GA), Not Applicable (SEE11371779GA), Not Applicable (SEE12510306GA), Not Applicable (SEE12510307GA), Not Applicable (SEE12510309GA), Not Applicable (SEE12510314GA), Not Applicable (SEE12510315GA), Not Applicable (SEE12510318GA), Not Applicable (SEE12510319GA), Not Applicable (SEE12510320GA), Not Applicable (SEE12510326GA), Not Applicable (SEE12510327GA), Not Applicable (SEE12510328GA), Not Applicable (SEE12510329GA), Not Applicable (SEE12510330GA), Not Applicable (SEE12510331GA), Not Applicable (SEE12510333GA), Not Applicable (SEE12520339GA), Not Applicable (SEE12520343GA), Not Applicable (SEE13000384GA), Not Applicable (SEE13000388GA), Not Applicable (SEE13050442GA), Not Applicable (SEE14031670GA), Not Applicable (SEE14031676GA), Not Applicable (SEE14031680GA), Not Applicable (SEE14031681GA), Not Applicable (SEE14031682GA), Not Applicable (SEE14031688GA), Not Applicable (SEE15093066GA), Not Applicable (SEE15093068GA), Not Applicable (SEE15103103GA), Not Applicable (SEE15528212SA), Not Applicable (SEE09510793GA), Not Applicable (SEE09510802GA), Not Applicable (SEE09510808GA), Not Applicable (SEE09510819GA), 00840682109260SEE19180010SA (SEE19180010SA), Not Applicable (SEE10420534GA), Not Applicable (SEE10420563GA), Not Applicable (SEE10420574GA), Not Applicable (SEE10430651GA), Not Applicable (SEE10450684GA), Not Applicable (SEE10460696GA), Not Applicable (SEE10470784GA), Not Applicable (SEE11071084GA), Not Applicable (SEE11071144GA), Not Applicable (SEE11071155GA), Not Applicable (SEE11081190GA), Not Applicable (SEE09100083GA), Not Applicable (SEE09100132GA), Not Applicable (SEE09350463GA), Not Applicable (SEE09360471GA), Not Applicable (SEE11060979GA), Not Applicable (SEE11060998GA), Not Applicable (SEE11061004GA), Not Applicable (SEE11161329GA), Not Applicable (SEE11371749GA), Not Applicable (SEE16068746SA), Not Applicable (SEE16241835SA), 00840682109260SEE18110002SA (SEE18110002SA), Not Applicable (SEE11071096GA), Not Applicable (SEE11081187GA), Not Applicable (SEE11101216GA), Not Applicable (SEE11101225GA), Not Applicable (SEE11141250GA), Not Applicable (SEE11171356GA), Not Applicable (SEE12430181GA), Not Applicable (SEE13040427GA), Not Applicable (SEE13110494GA), 00840682109260SEE18290008SA (SEE18290008SA), 00840682109260SEE18290014SA (SEE18290014SA), Not Applicable (SEE15352670SA), Not Applicable (SEE10130039GA), Not Applicable (SEE10130044GA), Not Applicable (SEE10130046GA), Not Applicable (SEE10130048GA), Not Applicable (SEE10130050GA), Not Applicable (SEE11151311GA), Not Applicable (SEE15312141SA), 00840682109260SEE17340004SA (SEE17340004SA), 00840682109260SEE17340006SA (SEE17340006SA), Not Applicable (SEE14482697GA), Not Applicable (SEE09470676GA), Not Applicable (SEE09470685GA), Not Applicable (SEE09490758GA), Not Applicable (SEE10530857GA), Not Applicable (SEE11060992GA), Not Applicable (SEE11061000GA), Not Applicable (SEE11061003GA), Not Applicable (SEE11071088GA), Not Applicable (SEE11071136GA), Not Applicable (SEE11141289GA), Not Applicable (SEE13371093GA), Not Applicable (SEE13371101GA), Not Applicable (SEE15012847GA), 00840682109260SEE17400002SA (SEE17400002SA), 00840682109260SEE17400004SA (SEE17400004SA), Not Applicable (SEE09430620GA), Not Applicable (SEE10250134GA), Not Applicable (SEE10390349GA), Not Applicable (SEE10390355GA), Not Applicable (SEE13411219GA), Not Applicable (SEE13411223GA), Not Applicable (SEE11331598GA), Not Applicable (SEE11361704GA), 00840682109260SEE18160023SA (SEE18160023SA), 00840682109260SEE18160032SA (SEE18160032SA), 00840682109260SEE19120018SA (SEE19120018SA), 00840682109260SEE19160048SA (SEE19160048SA), 00840682109260SEE19170032SA (SEE19170032SA), 00840682109260SEE19170033SA (SEE19170033SA), 00840682109260SEE17260002SA (SEE17260002SA), Not Applicable (SEE11511877GA), 00840682109260SEE17130030SA (SEE17130030SA), 00840682109260SEE17130031SA (SEE17130031SA), 00840682109260SEE17130033SA (SEE17130033SA), 00840682109260SEE17130034SA (SEE17130034SA), 00840682109260SEE19180001SA (SEE19180001SA), 00840682109260SEE19190006SA (SEE19190006SA), Not Applicable (SEE11341614GA), Not Applicable (SEE11341616GA), Not Applicable (SEE11341622GA), Not Applicable (SEE11341627GA), Not Applicable (SEE11341629GA), Not Applicable (SEE11341631GA), Not Applicable (SEE11341633GA), Not Applicable (SEE11341638GA), Not Applicable (SEE11361670GA), Not Applicable (SEE14462657GA), 00840682109260SEE17190003SA (SEE17190003SA), 00840682109260SEE17190004SA (SEE17190004SA), 00840682109260SEE17190008SA (SEE17190008SA), 00840682109260SEE17190009SA (SEE17190009SA), 00840682109260SEE17190010SA (SEE17190010SA), 00840682109260SEE17390008SA (SEE17390008SA), Not Applicable (SEE10400387GA), Not Applicable (SEE10400404GA), Not Applicable (SEE10400432GA), Not Applicable (SEE10400434GA), Not Applicable (SEE11141253GA), Not Applicable (SEE14512831GA), Not Applicable (SEE14512834GA), Not Applicable (SEE10100973GA), Not Applicable (SEE10100976GA), Not Applicable (SEE10100977GA), Not Applicable (SEE10120989GA), Not Applicable (SEE11361708GA), Not Applicable (SEE12092043GA), Not Applicable (SEE12252301GA), Not Applicable (SEE13401179GA), Not Applicable (SEE14151808GA), Not Applicable (SEE14151809GA), Not Applicable (SEE15113168GA), Not Applicable (SEE15113169GA), Not Applicable (SEE15113172GA), Not Applicable (SEE15113184GA), Not Applicable (SEE15113185GA), Not Applicable (SEE15456724SA), Not Applicable (SEE12041951GA), Not Applicable (SEE12041966GA), Not Applicable (SEE14221988GA), Not Applicable (SEE14221993GA), Not Applicable (SEE14222000GA), Not Applicable (SEE15352698SA), Not Applicable (SEE15352789SA), Not Applicable (SEE15352791SA), Not Applicable (SEE15352794SA), Not Applicable (SEE15352844SA), Not Applicable (SEE15352845SA), 00840682109260SEE17190019SA (SEE17190019SA), 00840682109260SEE17190020SA (SEE17190020SA), Not Applicable (SEE09150220GA), Not Applicable (SEE09190239GA), Not Applicable (SEE09190243GA), Not Applicable (SEE09410584GA), Not Applicable (SEE09460646GA), Not Applicable (SEE09530861GA), Not Applicable (SEE10130043GA), Not Applicable (SEE10130055GA), Not Applicable (SEE10130073GA), Not Applicable (SEE10390341GA), Not Applicable (SEE10460718GA), Not Applicable (SEE10460735GA), Not Applicable (SEE12292365GA), Not Applicable (SEE12420155GA), Not Applicable (SEE14031605GA), Not Applicable (SEE14031607GA), Not Applicable (SEE14031613GA), Not Applicable (SEE14031614GA), Not Applicable (SEE14031620GA), Not Applicable (SEE14031622GA), Not Applicable (SEE14031628GA), Not Applicable (SEE14041733GA), Not Applicable (SEE16028319SA), Not Applicable (SEE15052908GA), Not Applicable (SEE15103112GA), Not Applicable (SEE15113145GA), Not Applicable (SEE15113146GA), Not Applicable (SEE15113152GA), 00840682109260SEE17330012SA (SEE17330012SA), 00840682109260SEE17330014SA (SEE17330014SA), 00840682109260SEE17330015SA (SEE17330015SA), 00840682109260SEE17330016SA (SEE17330016SA), Not Applicable (SEE11191434GA), Not Applicable (SEE11211473GA), Not Applicable (SEE11211484GA), Not Applicable (SEE11211495GA), Not Applicable (SEE11211506GA), Not Applicable (SEE11241551GA), Not Applicable (SEE12011897GA), Not Applicable (SEE12011900GA), Not Applicable (SEE16119845SA), Not Applicable (SEE16119846SA), 00840682109260SEE18320014SA (SEE18320014SA), 00840682109260SEE18320016SA (SEE18320016SA), 00840682109260SEE18320029SA (SEE18320029SA), 00840682109260SEE18320032SA (SEE18320032SA), 00840682109260SEE18500013SA (SEE18500013SA), 00840682109260SEE18500018SA (SEE18500018SA), 00840682109260SEE18500043SA (SEE18500043SA), Not Applicable (SEE11060977GA), Not Applicable (SEE11071073GA), Not Applicable (SEE11071081GA), Not Applicable (SEE11071085GA), Not Applicable (SEE11081159GA), Not Applicable (SEE11081160GA), Not Applicable (SEE11081171GA), Not Applicable (SEE11081176GA), Not Applicable (SEE11211481GA), Not Applicable (SEE11211491GA), Not Applicable (SEE11321570GA), Not Applicable (SEE11321575GA), Not Applicable (SEE11361716GA), Not Applicable (SEE15497524SA), Not Applicable (SEE15507661SA), Not Applicable (SEE13180776GA), Not Applicable (SEE14302280GA), Not Applicable (SEE14312288GA), Not Applicable (SEE14312290GA), Not Applicable (SEE14312291GA), Not Applicable (SEE14312292GA), Not Applicable (SEE14312296GA), Not Applicable (SEE13180769GA), Not Applicable (SEE13180783GA), Not Applicable (SEE13180787GA), Not Applicable (SEE13180788GA), Not Applicable (SEE13220823GA), Not Applicable (SEE13220824GA), Not Applicable (SEE13220825GA), Not Applicable (SEE13220830GA), Not Applicable (SEE13220835GA), 00840682109260SEE17520002SA (SEE17520002SA), 00840682109260SEE18100026SA (SEE18100026SA), 00840682109260SEE18100037SA (SEE18100037SA), 00840682109260SEE18100040SA (SEE18100040SA), 00840682109260SEE18100041SA (SEE18100041SA), 00840682109260SEE18100042SA (SEE18100042SA), 00840682109260SEE18100043SA (SEE18100043SA), 00840682109260SEE18100044SA (SEE18100044SA), Not Applicable (SEE09510829GA), Not Applicable (SEE12122127GA), Not Applicable (SEE12122129GA), Not Applicable (SEE14221992GA), Not Applicable (SEE14221996GA), Not Applicable (SEE14221997GA), Not Applicable (SEE14221999GA), Not Applicable (SEE14222001GA), Not Applicable (SEE14242083GA), Not Applicable (SEE14242090GA), Not Applicable (SEE14242091GA), Not Applicable (SEE14242095GA), Not Applicable (SEE14242099GA), Not Applicable (SEE14242102GA), Not Applicable (SEE14242106GA), Not Applicable (SEE14242112GA), Not Applicable (SEE14242115GA), Not Applicable (SEE14242116GA), Not Applicable (SEE14242120GA), Not Applicable (SEE14422527GA), Not Applicable (SEE14422532GA), Not Applicable (SEE15113137GA), 00840682109260SEE17310002SA (SEE17310002SA), 00840682109260SEE17380004SA (SEE17380004SA), 00840682109260SEE17380006SA (SEE17380006SA), 00840682109260SEE17380010SA (SEE17380010SA), Not Applicable (SEE15352669SA), Not Applicable (SEE15352786SA), Not Applicable (SEE15352787SA), Not Applicable (SEE15352788SA), Not Applicable (SEE13431297GA), Not Applicable (SEE16420047SA), Not Applicable (SEE16430014SA), Not Applicable (SEE16430031SA), Not Applicable (SEE10430604GA), Not Applicable (SEE10460744GA), Not Applicable (SEE11040899GA), Not Applicable (SEE11040900GA), Not Applicable (SEE11040901GA), Not Applicable (SEE13451315GA), Not Applicable (SEE13451336GA), Not Applicable (SEE13481388GA), Not Applicable (SEE13481406GA), Not Applicable (SEE13481408GA), Not Applicable (SEE16099455SA), Not Applicable (SEE16099456SA), Not Applicable (SEE16099459SA), Not Applicable (SEE15446223SA), Not Applicable (SEE15446225SA), Not Applicable (SEE15446241SA), Not Applicable (SEE15446242SA), Not Applicable (SEE15446243SA), Not Applicable (SEE12400096GA), Not Applicable (SEE12400098GA), Not Applicable (SEE12400102GA), Not Applicable (SEE12400108GA), Not Applicable (SEE12400111GA), Not Applicable (SEE12400113GA), Not Applicable (SEE12400115GA), Not Applicable (SEE14412499GA), Not Applicable (SEE14412505GA), Not Applicable (SEE14412512GA), Not Applicable (SEE14422514GA), Not Applicable (SEE14422522GA), Not Applicable (SEE14422523GA), Not Applicable (SEE16068806SA), 00840682109260SEE19110004SA (SEE19110004SA), 00840682109260SEE19120020SA (SEE19120020SA), 00840682109260SEE19120029SA (SEE19120029SA), Not Applicable (SEE11071148GA), Not Applicable (SEE12162175GA), Not Applicable (SEE11141256GA), Not Applicable (SEE11141261GA), Not Applicable (SEE11141262GA), Not Applicable (SEE11141267GA), Not Applicable (SEE11141275GA), Not Applicable (SEE11141291GA), Not Applicable (SEE11141292GA), Not Applicable (SEE12430213GA), Not Applicable (SEE12430217GA), Not Applicable (SEE12440239GA), Not Applicable (SEE13150670GA), Not Applicable (SEE14432578GA), Not Applicable (SEE14432599GA), Not Applicable (SEE14432600GA), Not Applicable (SEE14432601GA), Not Applicable (SEE14432607GA), Not Applicable (SEE14432608GA), Not Applicable (SEE14432609GA), Not Applicable (SEE14432614GA), Not Applicable (SEE15052942GA), Not Applicable (SEE15052947GA), Not Applicable (SEE15062971GA), Not Applicable (SEE15113173GA), Not Applicable (SEE16334181SA), 00840682109260SEE19110001SA (SEE19110001SA), Not Applicable (SEE10250144GA), Not Applicable (SEE10250148GA), Not Applicable (SEE09020034GA), Not Applicable (SEE13501467GA), Not Applicable (SEE13501474GA), 00840682109260SEE19220008SA (SEE19220008SA), 00840682109260SEE19220010SA (SEE19220010SA), 00840682109260SEE19220011SA (SEE19220011SA), 00840682109260SEE19220012SA (SEE19220012SA), 00840682109260SEE19220013SA (SEE19220013SA), 00840682109260SEE19220014SA (SEE19220014SA), 00840682109260SEE19220015SA (SEE19220015SA), 00840682109260SEE19220016SA (SEE19220016SA), 00840682109260SEE19220019SA (SEE19220019SA), 00840682109260SEE19220026SA (SEE19220026SA), Not Applicable (SEE09100106GA), Not Applicable (SEE11321586GA), Not Applicable (SEE11371805GA), Not Applicable (SEE11511886GA), Not Applicable (SEE13160711GA), Not Applicable (SEE13160712GA), Not Applicable (SEE13160713GA), Not Applicable (SEE13160715GA), Not Applicable (SEE13160716GA), Not Applicable (SEE13160717GA), Not Applicable (SEE13160720GA), Not Applicable (SEE16078920SA), Not Applicable (SEE16078926SA), Not Applicable (SEE16089175SA), Not Applicable (SEE16089176SA), Not Applicable (SEE13150684GA), Not Applicable (SEE13150690GA), 00840682109260SEE18170001SA (SEE18170001SA), 00840682109260SEE18170020SA (SEE18170020SA), Not Applicable (SEE14362381GA), Not Applicable (SEE16078927SA), Not Applicable (SEE16078929SA), Not Applicable (SEE16089172SA), Not Applicable (SEE16089180SA), Not Applicable (SEE16180925SA), Not Applicable (SEE16201177SA), Not Applicable (SEE16201179SA), Not Applicable (SEE16201180SA), Not Applicable (SEE16201182SA), Not Applicable (SEE16201367SA), Not Applicable (SEE16201368SA), Not Applicable (SEE16201408SA), Not Applicable (SEE16440006SA), 00840682109260SEE19100002SA (SEE19100002SA), Not Applicable (SEE16211409SA), Not Applicable (SEE16211510SA), Not Applicable (SEE09330408GA), Not Applicable (SEE11331595GA), Not Applicable (SEE11331599GA), Not Applicable (SEE15153190GA), Not Applicable (SEE15153191GA), Not Applicable (SEE15153192GA), Not Applicable (SEE15153193GA), Not Applicable (SEE15153195GA), Not Applicable (SEE14171929GA), Not Applicable (SEE14171930GA), Not Applicable (SEE14171931GA), Not Applicable (SEE14171932GA), Not Applicable (SEE14171934GA), Not Applicable (SEE14191935GA), Not Applicable (SEE14191936GA), Not Applicable (SEE14191937GA), Not Applicable (SEE14191938GA), Not Applicable (SEE14191939GA), Not Applicable (SEE14191944GA), Not Applicable (SEE14191946GA), Not Applicable (SEE15352793SA), Not Applicable (SEE16150568SA), Not Applicable (SEE16160726SA), Not Applicable (SEE16160727SA), Not Applicable (SEE11521892GA), 00840682109260SEE17340007SA (SEE17340007SA), 00840682109260SEE17340011SA (SEE17340011SA), 00840682109260SEE17340015SA (SEE17340015SA), 00840682109260SEE17340019SA (SEE17340019SA), Not Applicable (SEE09330410GA), Not Applicable (SEE12252307GA), Not Applicable (SEE09530881GA), Not Applicable (SEE09530887GA), Not Applicable (SEE10250142GA), Not Applicable (SEE14021504GA), Not Applicable (SEE16380025SA), Not Applicable (SEE16380026SA), Not Applicable (SEE16380027SA), Not Applicable (SEE16380028SA), Not Applicable (SEE16380029SA), Not Applicable (SEE16380030SA), Not Applicable (SEE16380032SA), Not Applicable (SEE16380033SA), 00840682109260SEE17120001SA (SEE17120001SA), 00840682109260SEE17120002SA (SEE17120002SA), 00840682109260SEE18190003SA (SEE18190003SA), 00840682109260SEE18190004SA (SEE18190004SA), 00840682109260SEE18190009SA (SEE18190009SA), 00840682109260SEE18190010SA (SEE18190010SA), 00840682109260SEE18190012SA (SEE18190012SA), 00840682109260SEE18190013SA (SEE18190013SA), 00840682109260SEE18190016SA (SEE18190016SA), 00840682109260SEE18190020SA (SEE18190020SA), Not Applicable (SEE10450656GA), Not Applicable (SEE10460694GA), Not Applicable (SEE10250105GA), Not Applicable (SEE12510311GA), Not Applicable (SEE12510317GA), Not Applicable (SEE13050434GA), Not Applicable (SEE11371753GA), Not Applicable (SEE12302388GA), Not Applicable (SEE12322415GA), Not Applicable (SEE12322432GA), Not Applicable (SEE12332445GA), Not Applicable (SEE14151839GA), Not Applicable (SEE14171923GA), Not Applicable (SEE14171924GA), Not Applicable (SEE16109589SA), Not Applicable (SEE16109593SA), Not Applicable (SEE16119897SA), Not Applicable (SEE16119899SA), Not Applicable (SEE16119904SA), Not Applicable (SEE16119905SA), Not Applicable (SEE16119906SA), Not Applicable (SEE16119907SA), Not Applicable (SEE16119909SA), Not Applicable (SEE16119912SA), Not Applicable (SEE16119913SA), Not Applicable (SEE16119915SA), Not Applicable (SEE16119916SA), Not Applicable (SEE16119917SA), Not Applicable (SEE16119918SA), Not Applicable (SEE16119920SA), Not Applicable (SEE16119922SA), Not Applicable (SEE16119924SA), Not Applicable (SEE16119926SA), Not Applicable (SEE16119927SA), Not Applicable (SEE16119928SA), Not Applicable (SEE16430036SA), Not Applicable (SEE16430061SA), Not Applicable (SEE16430064SA), Not Applicable (SEE16430065SA), 00840682109260SEE17090012SA (SEE17090012SA), 00840682109260SEE17190021SA (SEE17190021SA), 00840682109260SEE19160047SA (SEE19160047SA), 00840682109260SEE19160051SA (SEE19160051SA), Not Applicable (SEE15477182SA), 00840682109260SEE18060001SA (SEE18060001SA), 00840682109260SEE18060007SA (SEE18060007SA), Not Applicable (SEE09140186GA), Not Applicable (SEE15012842GA), Not Applicable (SEE15012852GA), Not Applicable (SEE15022864GA), Not Applicable (SEE14492746GA), Not Applicable (SEE15012851GA), Not Applicable (SEE15456722SA), Not Applicable (SEE15466919SA), Not Applicable (SEE15466926SA), Not Applicable (SEE15466928SA), Not Applicable (SEE15466932SA), Not Applicable (SEE13220851GA), Not Applicable (SEE13280905GA), 00840682109260SEE17250002SA (SEE17250002SA), 00840682109260SEE17250006SA (SEE17250006SA), 00840682109260SEE17250007SA (SEE17250007SA), 00840682109260SEE17250008SA (SEE17250008SA), 00840682109260SEE17250009SA (SEE17250009SA), Not Applicable (SEE10420583GA), Not Applicable (SEE10420588GA), Not Applicable (SEE10460697GA), Not Applicable (SEE10460698GA), Not Applicable (SEE10460714GA), Not Applicable (SEE11071090GA), Not Applicable (SEE11181400GA), Not Applicable (SEE11181403GA), Not Applicable (SEE11201445GA), Not Applicable (SEE11211478GA), Not Applicable (SEE14242098GA), Not Applicable (SEE14422534GA), Not Applicable (SEE14422536GA), Not Applicable (SEE14422537GA), Not Applicable (SEE14422538GA), Not Applicable (SEE14422540GA), Not Applicable (SEE14422545GA), Not Applicable (SEE14422546GA), Not Applicable (SEE14422548GA), Not Applicable (SEE14422549GA), Not Applicable (SEE14422550GA), Not Applicable (SEE14422553GA), Not Applicable (SEE14422555GA), Not Applicable (SEE14422556GA), Not Applicable (SEE14422558GA), Not Applicable (SEE14422560GA), Not Applicable (SEE14422561GA), Not Applicable (SEE14422564GA), Not Applicable (SEE14422566GA), Not Applicable (SEE14422567GA), Not Applicable (SEE14422568GA), Not Applicable (SEE14422570GA), Not Applicable (SEE14422571GA), Not Applicable (SEE14422572GA), Not Applicable (SEE14422573GA), Not Applicable (SEE14462645GA), Not Applicable (SEE15241006SA), Not Applicable (SEE15251241SA), Not Applicable (SEE16191063SA), 00840682109260SEE17170004SA (SEE17170004SA), 00840682109260SEE19090001SA (SEE19090001SA), 00840682109260SEE19090003SA (SEE19090003SA), 00840682109260SEE19090018SA (SEE19090018SA), Not Applicable (SEE13150667GA), Not Applicable (SEE15193297GA), Not Applicable (SEE15193302GA), Not Applicable (SEE16038555SA), Not Applicable (SEE16058656SA), Not Applicable (SEE16058657SA), Not Applicable (SEE16058658SA), Not Applicable (SEE16058659SA), Not Applicable (SEE16058665SA), Not Applicable (SEE16058666SA), Not Applicable (SEE16058667SA), Not Applicable (SEE16058668SA), Not Applicable (SEE16058670SA), Not Applicable (SEE16058671SA), Not Applicable (SEE16058672SA), Not Applicable (SEE16058673SA), Not Applicable (SEE16058674SA), Not Applicable (SEE16058675SA), Not Applicable (SEE16058676SA), Not Applicable (SEE16058677SA), Not Applicable (SEE16058678SA), Not Applicable (SEE16058679SA), Not Applicable (SEE16058680SA), Not Applicable (SEE16058699SA), Not Applicable (SEE16130306SA), Not Applicable (SEE16130320SA), Not Applicable (SEE16130322SA), 00840682109260SEE17090017SA (SEE17090017SA), Not Applicable (SEE12021920GA), Not Applicable (SEE12021921GA), Not Applicable (SEE12051988GA), Not Applicable (SEE14382469GA), Not Applicable (SEE14402489GA), Not Applicable (SEE14402491GA), Not Applicable (SEE14412495GA), Not Applicable (SEE14412496GA), Not Applicable (SEE14412497GA), Not Applicable (SEE14412502GA), Not Applicable (SEE14412509GA), Not Applicable (SEE14412510GA), Not Applicable (SEE14412513GA), Not Applicable (SEE16252013SA), Not Applicable (SEE16252019SA), Not Applicable (SEE16252025SA), 00840682109260SEE17110005SA (SEE17110005SA), 00840682109260SEE18100021SA (SEE18100021SA), 00840682109260SEE18100022SA (SEE18100022SA), 00840682109260SEE18100023SA (SEE18100023SA), 00840682109260SEE18100024SA (SEE18100024SA), 00840682109260SEE18100033SA (SEE18100033SA), Not Applicable (SEE12490247GA), Not Applicable (SEE12490257GA), Not Applicable (SEE13100477GA), Not Applicable (SEE13100484GA), Not Applicable (SEE13100489GA), Not Applicable (SEE13100490GA), Not Applicable (SEE13110495GA), Not Applicable (SEE13110496GA), Not Applicable (SEE13110503GA), 00840682109260SEE17070006SA (SEE17070006SA), 00840682109260SEE17070012SA (SEE17070012SA), 00840682109260SEE17070013SA (SEE17070013SA), 00840682109260SEE17070014SA (SEE17070014SA), 00840682109260SEE17070015SA (SEE17070015SA), 00840682109260SEE17070016SA (SEE17070016SA), 00840682109260SEE17070017SA (SEE17070017SA), 00840682109260SEE17070018SA (SEE17070018SA), 00840682109260SEE17070021SA (SEE17070021SA), 00840682109260SEE17070022SA (SEE17070022SA), 00840682109260SEE17070023SA (SEE17070023SA), 00840682109260SEE17070024SA (SEE17070024SA), 00840682109260SEE17070025SA (SEE17070025SA), 00840682109260SEE17070026SA (SEE17070026SA), 00840682109260SEE17350001SA (SEE17350001SA), 00840682109260SEE17350002SA (SEE17350002SA), 00840682109260SEE17350003SA (SEE17350003SA), Not Applicable (SEE12340012GA), Not Applicable (SEE12340016GA), Not Applicable (SEE12340019GA), Not Applicable (SEE12350021GA), Not Applicable (SEE12350024GA), Not Applicable (SEE12360033GA), Not Applicable (SEE12360035GA), Not Applicable (SEE12360036GA), Not Applicable (SEE14432598GA), Not Applicable (SEE14432602GA), Not Applicable (SEE15052917GA), Not Applicable (SEE15052923GA), Not Applicable (SEE15052926GA), Not Applicable (SEE15052932GA), 00840682109260SEE18170005SA (SEE18170005SA), 00840682109260SEE18170025SA (SEE18170025SA), Not Applicable (SEE10390290GA), Not Applicable (SEE10400451GA), Not Applicable (SEE13140657GA), Not Applicable (SEE13150661GA), Not Applicable (SEE13160714GA), Not Applicable (SEE13160718GA), Not Applicable (SEE13160719GA), Not Applicable (SEE13180742GA), Not Applicable (SEE13361074GA), Not Applicable (SEE13371086GA), Not Applicable (SEE13371100GA), Not Applicable (SEE15113139GA), 00840682109260SEE18370001SA (SEE18370001SA), Not Applicable (SEE14161890GA), Not Applicable (SEE14161891GA), Not Applicable (SEE14161892GA), Not Applicable (SEE14161895GA), Not Applicable (SEE14161900GA), Not Applicable (SEE14161901GA), Not Applicable (SEE14161912GA), Not Applicable (SEE10470804GA), Not Applicable (SEE10470805GA), Not Applicable (SEE10470806GA), Not Applicable (SEE10530817GA), Not Applicable (SEE10530825GA), Not Applicable (SEE09020017GA), Not Applicable (SEE09040052GA), Not Applicable (SEE09040053GA), Not Applicable (SEE09040076GA), Not Applicable (SEE15261393SA), Not Applicable (SEE15261394SA), Not Applicable (SEE15261395SA), Not Applicable (SEE15261396SA), Not Applicable (SEE15261397SA), Not Applicable (SEE15261398SA), Not Applicable (SEE15261399SA), Not Applicable (SEE15261401SA), Not Applicable (SEE15261402SA), Not Applicable (SEE15261404SA), Not Applicable (SEE15261405SA), Not Applicable (SEE15261406SA), Not Applicable (SEE15261407SA), Not Applicable (SEE15261511SA), Not Applicable (SEE16364410SA), Not Applicable (SEE16364416SA), 00840682109260SEE18380022SA (SEE18380022SA), Not Applicable (SEE10120001GA), Not Applicable (SEE10120010GA), Not Applicable (SEE10121019GA), Not Applicable (SEE13371120GA), Not Applicable (SEE12490249GA), Not Applicable (SEE12490252GA), Not Applicable (SEE12490258GA), Not Applicable (SEE09430602GA), Not Applicable (SEE15052933GA), Not Applicable (SEE15052935GA), Not Applicable (SEE15052936GA), Not Applicable (SEE15052941GA), Not Applicable (SEE15053021GA), Not Applicable (SEE15062993GA), Not Applicable (SEE15063003GA), Not Applicable (SEE15063005GA), Not Applicable (SEE15063008GA), Not Applicable (SEE15063010GA), Not Applicable (SEE15063015GA), Not Applicable (SEE15063017GA), Not Applicable (SEE15063019GA), Not Applicable (SEE15063020GA), Not Applicable (SEE15063021GA), 00840682109260SEE19220018SA (SEE19220018SA), Not Applicable (SEE11050920GA), Not Applicable (SEE11050932GA), Not Applicable (SEE11050938GA), Not Applicable (SEE11050943GA), Not Applicable (SEE11071082GA), 00840682109260SEE18290017SA (SEE18290017SA), Not Applicable (SEE09100116GA), Not Applicable (SEE12092076GA), Not Applicable (SEE12092077GA), Not Applicable (SEE12092078GA), Not Applicable (SEE12092079GA), Not Applicable (SEE12092080GA), Not Applicable (SEE12102086GA), Not Applicable (SEE12102087GA), Not Applicable (SEE12102088GA), Not Applicable (SEE12102090GA), Not Applicable (SEE12102093GA), Not Applicable (SEE12102095GA), Not Applicable (SEE12102097GA), Not Applicable (SEE12102098GA), Not Applicable (SEE13120606GA), Not Applicable (SEE13130615GA), Not Applicable (SEE13130626GA), Not Applicable (SEE13130628GA), Not Applicable (SEE09100091GA), Not Applicable (SEE09100096GA), Not Applicable (SEE09100114GA), Not Applicable (SEE09100118GA), Not Applicable (SEE09100119GA), Not Applicable (SEE09100120GA), Not Applicable (SEE09100128GA), Not Applicable (SEE09470657GA), Not Applicable (SEE14161884GA), Not Applicable (SEE14161885GA), Not Applicable (SEE14161887GA), Not Applicable (SEE14161888GA), Not Applicable (SEE14161889GA), Not Applicable (SEE14161896GA), Not Applicable (SEE14161899GA), Not Applicable (SEE14161907GA), Not Applicable (SEE14161909GA), Not Applicable (SEE14161913GA), Not Applicable (SEE14161914GA), Not Applicable (SEE15477110SA), Not Applicable (SEE15477112SA), Not Applicable (SEE15415185SA), Not Applicable (SEE15415188SA), Not Applicable (SEE15415195SA), Not Applicable (SEE15446335SA), Not Applicable (SEE15456723SA), Not Applicable (SEE10430629GA), Not Applicable (SEE10450666GA), Not Applicable (SEE10450672GA), Not Applicable (SEE10450681GA), Not Applicable (SEE10450689GA), Not Applicable (SEE10530824GA), Not Applicable (SEE10530836GA), Not Applicable (SEE10530837GA), Not Applicable (SEE10530838GA), Not Applicable (SEE10530839GA), Not Applicable (SEE10530843GA), Not Applicable (SEE10530844GA), Not Applicable (SEE10530845GA), Not Applicable (SEE10530846GA), Not Applicable (SEE10530848GA), Not Applicable (SEE11050933GA), Not Applicable (SEE11050934GA), Not Applicable (SEE15173207GA), Not Applicable (SEE15446123SA), Not Applicable (SEE09100147GA), Not Applicable (SEE10120032GA), Not Applicable (SEE10430603GA), Not Applicable (SEE10450671GA), Not Applicable (SEE10460729GA), Not Applicable (SEE10460753GA), Not Applicable (SEE10510813GA), Not Applicable (SEE10510814GA), Not Applicable (SEE11050941GA), Not Applicable (SEE11071036GA), Not Applicable (SEE11081186GA), Not Applicable (SEE11161320GA), Not Applicable (SEE11191432GA), Not Applicable (SEE13280923GA), Not Applicable (SEE13280927GA), Not Applicable (SEE13280932GA), Not Applicable (SEE13280941GA), Not Applicable (SEE13280946GA), Not Applicable (SEE14272161GA), Not Applicable (SEE14272200GA), Not Applicable (SEE14272203GA), Not Applicable (SEE14272205GA), Not Applicable (SEE14272207GA), Not Applicable (SEE14272219GA), Not Applicable (SEE14472675GA), Not Applicable (SEE16360016SA), Not Applicable (SEE10410499GA), Not Applicable (SEE10420540GA), Not Applicable (SEE11141288GA), Not Applicable (SEE12340009GA), Not Applicable (SEE12340011GA), Not Applicable (SEE12342451GA), Not Applicable (SEE12350029GA), Not Applicable (SEE12390085GA), Not Applicable (SEE12390087GA), Not Applicable (SEE12390088GA), Not Applicable (SEE12390089GA), Not Applicable (SEE12390090GA), Not Applicable (SEE13050439GA), Not Applicable (SEE13060445GA), Not Applicable (SEE13190798GA), Not Applicable (SEE13290951GA), Not Applicable (SEE13290953GA), Not Applicable (SEE13290955GA), Not Applicable (SEE13290958GA), Not Applicable (SEE15353099SA), Not Applicable (SEE15353100SA), Not Applicable (SEE15353119SA), Not Applicable (SEE16150524SA), Not Applicable (SEE16221590SA), Not Applicable (SEE11071037GA), Not Applicable (SEE11071043GA), Not Applicable (SEE13020407GA), Not Applicable (SEE14362385GA), Not Applicable (SEE10260222GA), Not Applicable (SEE10270251GA), Not Applicable (SEE13040419GA), 00840682109260SEE17060005SA (SEE17060005SA), Not Applicable (SEE12172188GA), Not Applicable (SEE16390001SA), Not Applicable (SEE16390005SA), Not Applicable (SEE09340450GA), Not Applicable (SEE09360491GA), Not Applicable (SEE09390537GA), Not Applicable (SEE12430185GA), Not Applicable (SEE12430187GA), Not Applicable (SEE12430203GA), Not Applicable (SEE12490241GA), Not Applicable (SEE12490242GA), Not Applicable (SEE12490253GA), Not Applicable (SEE12490255GA), Not Applicable (SEE12490284GA), Not Applicable (SEE12490286GA), Not Applicable (SEE12490291GA), Not Applicable (SEE12490292GA), Not Applicable (SEE12490297GA), Not Applicable (SEE13371095GA), Not Applicable (SEE13371097GA), Not Applicable (SEE13371098GA), Not Applicable (SEE13371104GA), Not Applicable (SEE13371110GA), Not Applicable (SEE13371111GA), Not Applicable (SEE13371117GA), Not Applicable (SEE13371118GA), 00840682109260SEE17180003SA (SEE17180003SA), Not Applicable (SEE12510301GA), Not Applicable (SEE13000387GA), Not Applicable (SEE13220849GA), Not Applicable (SEE13270880GA), Not Applicable (SEE13270884GA), Not Applicable (SEE13270891GA), Not Applicable (SEE13270895GA), Not Applicable (SEE13270900GA), Not Applicable (SEE13270902GA), Not Applicable (SEE13280906GA), Not Applicable (SEE09230255GA), Not Applicable (SEE09230259GA), Not Applicable (SEE09370499GA), Not Applicable (SEE09530854GA), Not Applicable (SEE10390289GA), 00840682109260SEE17120003SA (SEE17120003SA), Not Applicable (SEE14111800GA), Not Applicable (SEE10250160GA), Not Applicable (SEE10270242GA), Not Applicable (SEE10430628GA), Not Applicable (SEE10430638GA), Not Applicable (SEE10460765GA), Not Applicable (SEE10460771GA), Not Applicable (SEE10460774GA), Not Applicable (SEE10470777GA), Not Applicable (SEE14031587GA), 00840682109260SEE18510002SA (SEE18510002SA), 00840682109260SEE19130007SA (SEE19130007SA), 00840682109260SEE19150001SA (SEE19150001SA), Not Applicable (SEE10250140GA), Not Applicable (SEE10250167GA), Not Applicable (SEE10270248GA), Not Applicable (SEE10270259GA), Not Applicable (SEE10270261GA), Not Applicable (SEE09100104GA), Not Applicable (SEE10260236GA), Not Applicable (SEE10260238GA), Not Applicable (SEE10430614GA), Not Applicable (SEE10430630GA), Not Applicable (SEE10430645GA), Not Applicable (SEE10430653GA), Not Applicable (SEE11361703GA), Not Applicable (SEE12092046GA), Not Applicable (SEE12172198GA), Not Applicable (SEE12172200GA), Not Applicable (SEE12182226GA), Not Applicable (SEE12182227GA), Not Applicable (SEE12182230GA), Not Applicable (SEE12182232GA), Not Applicable (SEE12430190GA), Not Applicable (SEE13150660GA), Not Applicable (SEE13220832GA), Not Applicable (SEE13220839GA), Not Applicable (SEE15062974GA), Not Applicable (SEE15062976GA), Not Applicable (SEE15062987GA), Not Applicable (SEE15062990GA), Not Applicable (SEE15062994GA), Not Applicable (SEE15062995GA), Not Applicable (SEE15062996GA), Not Applicable (SEE15062998GA), Not Applicable (SEE15063011GA), Not Applicable (SEE15113132GA), Not Applicable (SEE15281638SA), Not Applicable (SEE15281639SA), Not Applicable (SEE15281640SA), Not Applicable (SEE15312099SA), Not Applicable (SEE15446220SA), Not Applicable (SEE15446221SA), Not Applicable (SEE15446222SA), Not Applicable (SEE15446224SA), Not Applicable (SEE15446230SA), Not Applicable (SEE15446235SA), Not Applicable (SEE15446236SA), Not Applicable (SEE15446238SA), Not Applicable (SEE15446239SA), Not Applicable (SEE16119850SA), Not Applicable (SEE16201406SA), Not Applicable (SEE16201407SA), 00840682109260SEE17460013SA (SEE17460013SA), 00840682109260SEE17460016SA (SEE17460016SA), 00840682109260SEE17460021SA (SEE17460021SA), 00840682109260SEE17460022SA (SEE17460022SA), Not Applicable (SEE13270886GA), Not Applicable (SEE13270887GA), Not Applicable (SEE13270892GA), Not Applicable (SEE13270893GA), Not Applicable (SEE16241792SA), Not Applicable (SEE16241796SA), Not Applicable (SEE09330409GA), Not Applicable (SEE09330411GA), Not Applicable (SEE09230273GA), Not Applicable (SEE09230284GA), Not Applicable (SEE09230286GA), Not Applicable (SEE09240295GA), Not Applicable (SEE09250317GA), Not Applicable (SEE09380515GA), Not Applicable (SEE09380517GA), Not Applicable (SEE09380522GA), Not Applicable (SEE09380525GA), Not Applicable (SEE10420558GA), Not Applicable (SEE10420564GA), Not Applicable (SEE10420577GA), Not Applicable (SEE10420585GA), Not Applicable (SEE10420595GA), Not Applicable (SEE10430620GA), Not Applicable (SEE10450660GA), Not Applicable (SEE10460692GA), Not Applicable (SEE11201447GA), Not Applicable (SEE11201455GA), Not Applicable (SEE11201456GA), Not Applicable (SEE11201464GA), Not Applicable (SEE15093059GA), Not Applicable (SEE15103080GA), Not Applicable (SEE15103086GA), Not Applicable (SEE15103087GA), Not Applicable (SEE15103093GA), Not Applicable (SEE15103095GA), 00840682109260SEE17460007SA (SEE17460007SA), 00840682109260SEE18420007SA (SEE18420007SA), 00840682109260SEE18420013SA (SEE18420013SA), 00840682109260SEE18420015SA (SEE18420015SA), 00840682109260SEE18420016SA (SEE18420016SA), 00840682109260SEE18420020SA (SEE18420020SA), 00840682109260SEE18420021SA (SEE18420021SA), Not Applicable (SEE09190240GA), Not Applicable (SEE10390299GA), Not Applicable (SEE13431285GA), Not Applicable (SEE10400422GA), Not Applicable (SEE10410468GA), Not Applicable (SEE10410481GA), Not Applicable (SEE10410513GA), Not Applicable (SEE10420533GA), Not Applicable (SEE10420553GA), Not Applicable (SEE14312307GA), Not Applicable (SEE15251188SA), Not Applicable (SEE16028317SA), Not Applicable (SEE13120595GA), Not Applicable (SEE13120597GA), Not Applicable (SEE13120600GA), Not Applicable (SEE13120608GA), Not Applicable (SEE13130617GA), Not Applicable (SEE13130619GA), Not Applicable (SEE13130624GA), Not Applicable (SEE13130627GA), Not Applicable (SEE13130629GA), Not Applicable (SEE13130632GA), Not Applicable (SEE13130633GA), Not Applicable (SEE13130635GA), Not Applicable (SEE13130645GA), Not Applicable (SEE13130648GA), Not Applicable (SEE15113119GA), Not Applicable (SEE15113122GA), 00840682109260SEE17490015SA (SEE17490015SA), 00840682109260SEE17490016SA (SEE17490016SA), 00840682109260SEE17490037SA (SEE17490037SA), 00840682109260SEE18470018SA (SEE18470018SA), 00840682109260SEE18470019SA (SEE18470019SA), 00840682109260SEE18470024SA (SEE18470024SA), 00840682109260SEE18500035SA (SEE18500035SA), 00840682109260SEE18500037SA (SEE18500037SA), Not Applicable (SEE13240854GA), Not Applicable (SEE13240859GA), Not Applicable (SEE13240861GA), Not Applicable (SEE13240864GA), Not Applicable (SEE13240865GA), Not Applicable (SEE13240870GA), Not Applicable (SEE12102100GA), Not Applicable (SEE09100092GA), Not Applicable (SEE09100100GA), Not Applicable (SEE12112119GA), Not Applicable (SEE12322416GA), Not Applicable (SEE12322419GA), Not Applicable (SEE15261547SA), Not Applicable (SEE15261548SA), Not Applicable (SEE15261549SA), Not Applicable (SEE15261550SA), Not Applicable (SEE15261551SA), Not Applicable (SEE15261552SA), Not Applicable (SEE15261553SA), Not Applicable (SEE15261555SA), Not Applicable (SEE15261556SA), Not Applicable (SEE15261557SA), Not Applicable (SEE15261558SA), Not Applicable (SEE16360021SA), Not Applicable (SEE16370002SA), Not Applicable (SEE16370004SA), Not Applicable (SEE10100983GA), Not Applicable (SEE10120008GA), Not Applicable (SEE10120991GA), Not Applicable (SEE10120992GA), Not Applicable (SEE10390319GA), Not Applicable (SEE10390328GA), Not Applicable (SEE13010406GA), Not Applicable (SEE15352667SA), 00840682109260SEE17070019SA (SEE17070019SA), 00840682109260SEE17070028SA (SEE17070028SA), 00840682109260SEE18450010SA (SEE18450010SA), 00840682109260SEE18450011SA (SEE18450011SA), 00840682109260SEE18450013SA (SEE18450013SA), Not Applicable (SEE11040917GA), Not Applicable (SEE11050921GA), Not Applicable (SEE11050923GA), Not Applicable (SEE11050924GA), Not Applicable (SEE11050925GA), Not Applicable (SEE11050926GA), Not Applicable (SEE11050928GA), Not Applicable (SEE16221555SA), Not Applicable (SEE12490245GA), Not Applicable (SEE16078863SA), Not Applicable (SEE12490248GA), Not Applicable (SEE12490251GA), Not Applicable (SEE12490300GA), Not Applicable (SEE13371088GA), Not Applicable (SEE13371094GA), Not Applicable (SEE13371099GA), Not Applicable (SEE13371102GA), Not Applicable (SEE13371106GA), Not Applicable (SEE13371107GA), Not Applicable (SEE13371114GA), Not Applicable (SEE13371116GA), Not Applicable (SEE15173227GA), 00840682109260SEE18240052SA (SEE18240052SA), 00840682109260SEE19120016SA (SEE19120016SA), 00840682109260SEE19120026SA (SEE19120026SA), 00840682109260SEE19120027SA (SEE19120027SA), 00840682109260SEE19170034SA (SEE19170034SA), Not Applicable (SEE13110530GA), Not Applicable (SEE13110532GA), Not Applicable (SEE13110534GA), Not Applicable (SEE13110535GA), Not Applicable (SEE13110541GA), Not Applicable (SEE13110543GA), Not Applicable (SEE13110545GA), Not Applicable (SEE16370003SA), Not Applicable (SEE16370005SA), 00840682109260SEE19180005SA (SEE19180005SA), 00840682109260SEE19190002SA (SEE19190002SA), Not Applicable (SEE11071080GA), Not Applicable (SEE11071135GA), Not Applicable (SEE11081175GA), Not Applicable (SEE11081196GA), Not Applicable (SEE12520357GA), 00840682109260SEE18040018SA (SEE18040018SA), 00840682109260SEE18040041SA (SEE18040041SA), 00840682109260SEE18040043SA (SEE18040043SA), 00840682109260SEE18040048SA (SEE18040048SA), Not Applicable (SEE14512822GA), Not Applicable (SEE14512826GA), Not Applicable (SEE14512829GA), Not Applicable (SEE15291766SA), Not Applicable (SEE16360012SA), Not Applicable (SEE16360013SA), Not Applicable (SEE11111244GA), Not Applicable (SEE11211501GA), 00840682109260SEE18320020SA (SEE18320020SA), Not Applicable (SEE09340449GA), Not Applicable (SEE09340453GA), Not Applicable (SEE09350462GA), Not Applicable (SEE14031625GA), Not Applicable (SEE14031626GA), Not Applicable (SEE14031629GA), Not Applicable (SEE14031637GA), Not Applicable (SEE14031648GA), Not Applicable (SEE14031651GA), Not Applicable (SEE14031652GA), Not Applicable (SEE14031653GA), Not Applicable (SEE14031654GA), Not Applicable (SEE14031656GA), Not Applicable (SEE09100133GA), Not Applicable (SEE10390314GA), Not Applicable (SEE10390340GA), Not Applicable (SEE10400426GA), Not Applicable (SEE10410486GA), Not Applicable (SEE12242279GA), Not Applicable (SEE12322408GA), Not Applicable (SEE13110509GA), Not Applicable (SEE13110512GA), Not Applicable (SEE13110517GA), Not Applicable (SEE13110525GA), Not Applicable (SEE13110526GA), Not Applicable (SEE13381123GA), Not Applicable (SEE13391137GA), Not Applicable (SEE13391138GA), Not Applicable (SEE13391139GA), Not Applicable (SEE13391140GA), Not Applicable (SEE15093075GA), Not Applicable (SEE15103076GA), Not Applicable (SEE15103078GA), Not Applicable (SEE15103079GA), Not Applicable (SEE15103081GA), Not Applicable (SEE15103082GA), Not Applicable (SEE15103083GA), Not Applicable (SEE15103088GA), Not Applicable (SEE15103089GA), Not Applicable (SEE15103090GA), Not Applicable (SEE15103091GA), Not Applicable (SEE15103092GA), Not Applicable (SEE15456710SA), 00840682109260SEE18100020SA (SEE18100020SA), 00840682109260SEE18160051SA (SEE18160051SA), 00840682109260SEE18160052SA (SEE18160052SA), 00840682109260SEE18160054SA (SEE18160054SA), 00840682109260SEE18160057SA (SEE18160057SA), 00840682109260SEE18160059SA (SEE18160059SA), 00840682109260SEE18510012SA (SEE18510012SA), Not Applicable (SEE11361745GA), Not Applicable (SEE11371764GA), Not Applicable (SEE16109588SA), Not Applicable (SEE16109592SA), Not Applicable (SEE16109595SA), Not Applicable (SEE16109699SA), Not Applicable (SEE16109704SA), 00840682109260SEE17080010SA (SEE17080010SA), Not Applicable (SEE12162171GA), Not Applicable (SEE14502781GA), Not Applicable (SEE14502782GA), Not Applicable (SEE14502783GA), Not Applicable (SEE14502784GA), Not Applicable (SEE11361727GA), Not Applicable (SEE11521889GA), Not Applicable (SEE11521891GA), Not Applicable (SEE12031931GA), Not Applicable (SEE12031933GA), Not Applicable (SEE12031937GA), Not Applicable (SEE10400370GA), Not Applicable (SEE10400373GA), Not Applicable (SEE10400435GA), Not Applicable (SEE10400441GA), Not Applicable (SEE14502788GA), Not Applicable (SEE14502792GA), Not Applicable (SEE14502794GA), Not Applicable (SEE16354345SA), Not Applicable (SEE12410121GA), Not Applicable (SEE12410123GA), Not Applicable (SEE12410124GA), Not Applicable (SEE12410125GA), Not Applicable (SEE12410127GA), Not Applicable (SEE12410128GA), Not Applicable (SEE12410129GA), Not Applicable (SEE12410145GA), Not Applicable (SEE12430202GA), Not Applicable (SEE12430208GA), Not Applicable (SEE12430210GA), Not Applicable (SEE13411189GA), Not Applicable (SEE13411190GA), Not Applicable (SEE13411191GA), Not Applicable (SEE13411192GA), Not Applicable (SEE13411193GA), Not Applicable (SEE15063004GA), 00840682109260SEE18080007SA (SEE18080007SA), Not Applicable (SEE09140191GA), Not Applicable (SEE09230262GA), Not Applicable (SEE10120017GA), Not Applicable (SEE11071039GA), Not Applicable (SEE11191436GA), Not Applicable (SEE11211480GA), Not Applicable (SEE11241547GA), Not Applicable (SEE13501458GA), Not Applicable (SEE13501465GA), Not Applicable (SEE13501469GA), Not Applicable (SEE14031644GA), Not Applicable (SEE14031671GA), Not Applicable (SEE14211978GA), Not Applicable (SEE14222026GA), Not Applicable (SEE14222032GA), Not Applicable (SEE14222034GA), Not Applicable (SEE14322312GA), Not Applicable (SEE14372420GA), Not Applicable (SEE14372432GA), Not Applicable (SEE14372437GA), Not Applicable (SEE14372441GA), Not Applicable (SEE14372442GA), Not Applicable (SEE14382456GA), Not Applicable (SEE14382457GA), Not Applicable (SEE14382465GA), Not Applicable (SEE14382467GA), Not Applicable (SEE14382471GA), Not Applicable (SEE15062982GA), Not Applicable (SEE15062985GA), Not Applicable (SEE15093024GA), Not Applicable (SEE15093025GA), Not Applicable (SEE15093026GA), Not Applicable (SEE15093027GA), Not Applicable (SEE15093028GA), Not Applicable (SEE15093030GA), Not Applicable (SEE15093031GA), Not Applicable (SEE15093033GA), Not Applicable (SEE15456728SA), Not Applicable (SEE15456730SA), Not Applicable (SEE15497530SA), Not Applicable (SEE15497579SA), Not Applicable (SEE15497580SA), Not Applicable (SEE15497581SA), Not Applicable (SEE15497583SA), Not Applicable (SEE15497584SA), Not Applicable (SEE15497585SA), Not Applicable (SEE15497586SA), Not Applicable (SEE15497587SA), Not Applicable (SEE15497588SA), Not Applicable (SEE10470787GA), Not Applicable (SEE12082030GA), Not Applicable (SEE12082035GA), Not Applicable (SEE12092048GA), Not Applicable (SEE16058669SA), Not Applicable (SEE11181388GA), Not Applicable (SEE11181390GA), Not Applicable (SEE11181398GA), Not Applicable (SEE11201444GA), Not Applicable (SEE11201453GA), Not Applicable (SEE11201463GA), Not Applicable (SEE11231541GA), Not Applicable (SEE11231542GA), Not Applicable (SEE15261392SA), Not Applicable (SEE15261408SA), Not Applicable (SEE16380008SA), 00840682109260SEE17490004SA (SEE17490004SA), 00840682109260SEE18500014SA (SEE18500014SA), Not Applicable (SEE09430609GA), Not Applicable (SEE09430611GA), Not Applicable (SEE09430613GA), Not Applicable (SEE09430616GA), Not Applicable (SEE09430618GA), Not Applicable (SEE09430621GA), Not Applicable (SEE11191433GA), Not Applicable (SEE11211494GA), Not Applicable (SEE10410495GA), Not Applicable (SEE12430200GA), Not Applicable (SEE15332311SA), Not Applicable (SEE15332312SA), Not Applicable (SEE15332313SA), Not Applicable (SEE15332314SA), Not Applicable (SEE15332315SA), Not Applicable (SEE15332339SA), Not Applicable (SEE15332341SA), Not Applicable (SEE11321574GA), Not Applicable (SEE12302383GA), Not Applicable (SEE12312398GA), Not Applicable (SEE12312401GA), Not Applicable (SEE12312404GA), Not Applicable (SEE15415094SA), Not Applicable (SEE15415193SA), Not Applicable (SEE15415194SA), Not Applicable (SEE12340007GA), Not Applicable (SEE12342449GA), Not Applicable (SEE12342452GA), Not Applicable (SEE16364405SA), Not Applicable (SEE09410573GA), Not Applicable (SEE09470670GA), Not Applicable (SEE09480740GA), Not Applicable (SEE10400401GA), 00840682109260SEE17150006SA (SEE17150006SA), 00840682109260SEE18240015SA (SEE18240015SA), 00840682109260SEE18240036SA (SEE18240036SA), 00840682109260SEE18240037SA (SEE18240037SA), 00840682109260SEE18240044SA (SEE18240044SA), 00840682109260SEE18240045SA (SEE18240045SA), 00840682109260SEE18240047SA (SEE18240047SA), 00840682109260SEE18240048SA (SEE18240048SA), Not Applicable (SEE11171382GA), Not Applicable (SEE16109578SA), Not Applicable (SEE16109591SA), Not Applicable (SEE16119767SA), Not Applicable (SEE16119923SA), Not Applicable (SEE16120076SA), Not Applicable (SEE16120077SA), Not Applicable (SEE16130304SA), Not Applicable (SEE16150547SA), Not Applicable (SEE16160735SA), Not Applicable (SEE16160737SA), 00840682109260SEE18170002SA (SEE18170002SA), 00840682109260SEE18170004SA (SEE18170004SA), 00840682109260SEE18170007SA (SEE18170007SA), 00840682109260SEE18170010SA (SEE18170010SA), 00840682109260SEE18170011SA (SEE18170011SA), 00840682109260SEE18170014SA (SEE18170014SA), 00840682109260SEE18170015SA (SEE18170015SA), 00840682109260SEE18170017SA (SEE18170017SA), 00840682109260SEE18170018SA (SEE18170018SA), 00840682109260SEE18170019SA (SEE18170019SA), 00840682109260SEE18170021SA (SEE18170021SA), 00840682109260SEE18170022SA (SEE18170022SA), 00840682109260SEE18170029SA (SEE18170029SA), 00840682109260SEE18170030SA (SEE18170030SA), 00840682109260SEE18170031SA (SEE18170031SA), 00840682109260SEE18170032SA (SEE18170032SA), 00840682109260SEE18170033SA (SEE18170033SA), 00840682109260SEE18280005SA (SEE18280005SA), 00840682109260SEE19180004SA (SEE19180004SA), 00840682109260SEE19180006SA (SEE19180006SA), 00840682109260SEE19180009SA (SEE19180009SA), 00840682109260SEE19180011SA (SEE19180011SA), 00840682109260SEE19180012SA (SEE19180012SA), 00840682109260SEE19190001SA (SEE19190001SA), 00840682109260SEE19190005SA (SEE19190005SA), 00840682109260SEE19190007SA (SEE19190007SA), 00840682109260SEE19190009SA (SEE19190009SA), 00840682109260SEE19190010SA (SEE19190010SA), Not Applicable (SEE09140177GA), Not Applicable (SEE09190242GA), Not Applicable (SEE14422515GA), Not Applicable (SEE14422516GA), Not Applicable (SEE14422517GA), Not Applicable (SEE14422518GA), Not Applicable (SEE14422520GA), Not Applicable (SEE14422521GA), Not Applicable (SEE16160736SA), 00840682109260SEE17490001SA (SEE17490001SA), 00840682109260SEE17490002SA (SEE17490002SA), 00840682109260SEE17490003SA (SEE17490003SA), 00840682109260SEE17490006SA (SEE17490006SA), 00840682109260SEE17490007SA (SEE17490007SA), 00840682109260SEE17490009SA (SEE17490009SA), 00840682109260SEE17490010SA (SEE17490010SA), 00840682109260SEE17490012SA (SEE17490012SA), Not Applicable (SEE11030881GA), Not Applicable (SEE11030883GA), Not Applicable (SEE11030888GA), Not Applicable (SEE13351010GA), Not Applicable (SEE13351023GA), Not Applicable (SEE14372403GA), Not Applicable (SEE10530841GA), Not Applicable (SEE12490256GA), Not Applicable (SEE12490260GA), Not Applicable (SEE12490273GA), Not Applicable (SEE12490274GA), Not Applicable (SEE12490278GA), Not Applicable (SEE13120572GA), Not Applicable (SEE13120574GA), Not Applicable (SEE13120579GA), Not Applicable (SEE13120581GA), Not Applicable (SEE13120582GA), Not Applicable (SEE13120583GA), Not Applicable (SEE13120584GA), Not Applicable (SEE13120599GA), Not Applicable (SEE13120604GA), Not Applicable (SEE14031720GA), Not Applicable (SEE12490295GA), Not Applicable (SEE13401181GA), Not Applicable (SEE13401182GA), Not Applicable (SEE15332309SA), Not Applicable (SEE15332310SA), Not Applicable (SEE09390542GA), Not Applicable (SEE09390547GA), Not Applicable (SEE09460636GA), Not Applicable (SEE09460649GA), Not Applicable (SEE11151312GA), Not Applicable (SEE15113166GA), Not Applicable (SEE15153197GA), Not Applicable (SEE16354346SA), Not Applicable (SEE16360011SA), Not Applicable (SEE16360015SA), 00840682109260SEE17110007SA (SEE17110007SA), Not Applicable (SEE12122137GA), Not Applicable (SEE12122138GA), Not Applicable (SEE12122146GA), Not Applicable (SEE15113138GA), 00840682109260SEE17360010SA (SEE17360010SA), 00840682109260SEE18500030SA (SEE18500030SA), Not Applicable (SEE09360482GA), Not Applicable (SEE14201952GA), Not Applicable (SEE14201953GA), Not Applicable (SEE14201958GA), Not Applicable (SEE14201959GA), Not Applicable (SEE14201960GA), Not Applicable (SEE14201961GA), 00840682109260SEE17150017SA (SEE17150017SA), Not Applicable (SEE11331596GA), 00840682109260SEE18450004SA (SEE18450004SA), Not Applicable (SEE14151816GA), Not Applicable (SEE14151817GA), Not Applicable (SEE14151818GA), Not Applicable (SEE13290964GA), Not Applicable (SEE13290971GA), Not Applicable (SEE15301928SA), Not Applicable (SEE11030890GA), Not Applicable (SEE09250311GA), Not Applicable (SEE09250313GA), Not Applicable (SEE09250314GA), Not Applicable (SEE09300351GA), Not Applicable (SEE09300353GA), Not Applicable (SEE09330416GA), Not Applicable (SEE09330423GA), Not Applicable (SEE09330425GA), Not Applicable (SEE13351050GA), Not Applicable (SEE14502809GA), Not Applicable (SEE14502810GA), Not Applicable (SEE14502811GA), Not Applicable (SEE14502812GA), Not Applicable (SEE14502813GA), Not Applicable (SEE14502814GA), Not Applicable (SEE14502815GA), Not Applicable (SEE14502816GA), Not Applicable (SEE14502817GA), Not Applicable (SEE14502818GA), Not Applicable (SEE15240990SA), Not Applicable (SEE16028369SA), 00840682109260SEE17430016SA (SEE17430016SA), Not Applicable (SEE11071142GA), Not Applicable (SEE12420174GA), Not Applicable (SEE14151837GA), Not Applicable (SEE14151867GA), 00840682109260SEE19220004SA (SEE19220004SA), Not Applicable (SEE09450629GA), Not Applicable (SEE09470673GA), Not Applicable (SEE15405016SA), Not Applicable (SEE15405019SA), 00840682109260SEE17130001SA (SEE17130001SA), 00840682109260SEE17130005SA (SEE17130005SA), 00840682109260SEE17130010SA (SEE17130010SA), 00840682109260SEE17130012SA (SEE17130012SA), 00840682109260SEE18060024SA (SEE18060024SA), 00840682109260SEE18330001SA (SEE18330001SA), 00840682109260SEE18330006SA (SEE18330006SA), Not Applicable (SEE11371752GA), Not Applicable (SEE11371759GA), Not Applicable (SEE11371797GA), Not Applicable (SEE11371804GA), Not Applicable (SEE12162179GA), Not Applicable (SEE14242096GA), 00840682109260SEE18410009SA (SEE18410009SA), 00840682109260SEE18410024SA (SEE18410024SA), Not Applicable (SEE13140651GA), Not Applicable (SEE13290963GA), Not Applicable (SEE13290985GA), Not Applicable (SEE11511862GA), Not Applicable (SEE11511863GA), Not Applicable (SEE11211477GA), Not Applicable (SEE12021903GA), Not Applicable (SEE12021905GA), Not Applicable (SEE12021912GA), Not Applicable (SEE12021915GA), Not Applicable (SEE12021916GA), Not Applicable (SEE12021917GA), Not Applicable (SEE12021919GA), Not Applicable (SEE12092041GA), Not Applicable (SEE12102105GA), Not Applicable (SEE12102106GA), Not Applicable (SEE14031609GA), 00840682109260SEE17130015SA (SEE17130015SA), 00840682109260SEE17130016SA (SEE17130016SA), 00840682109260SEE19130013SA (SEE19130013SA), 00840682109260SEE19130015SA (SEE19130015SA), 00840682109260SEE19130017SA (SEE19130017SA), 00840682109260SEE18320005SA (SEE18320005SA), 00840682109260SEE18330005SA (SEE18330005SA), Not Applicable (SEE12021901GA), Not Applicable (SEE12410126GA), Not Applicable (SEE12410136GA), Not Applicable (SEE12410141GA), Not Applicable (SEE15312120SA), Not Applicable (SEE15312131SA), Not Applicable (SEE16380015SA), Not Applicable (SEE16380019SA), Not Applicable (SEE09340455GA), Not Applicable (SEE09340456GA), Not Applicable (SEE09350461GA), Not Applicable (SEE09350468GA), Not Applicable (SEE13501455GA), Not Applicable (SEE13501471GA), Not Applicable (SEE13501482GA), 00840682109260SEE18370013SA (SEE18370013SA), 00840682109260SEE18380013SA (SEE18380013SA), Not Applicable (SEE09430605GA), Not Applicable (SEE11141259GA), Not Applicable (SEE14031663GA), Not Applicable (SEE15093034GA), Not Applicable (SEE15093039GA), Not Applicable (SEE11050973GA), Not Applicable (SEE11060953GA), Not Applicable (SEE11060955GA), Not Applicable (SEE11060956GA), Not Applicable (SEE11060959GA), Not Applicable (SEE11060964GA), Not Applicable (SEE11071045GA), Not Applicable (SEE11071145GA), Not Applicable (SEE11081168GA), Not Applicable (SEE13040417GA), Not Applicable (SEE13040428GA), Not Applicable (SEE15312121SA), Not Applicable (SEE12302368GA), Not Applicable (SEE12302369GA), Not Applicable (SEE12302370GA), Not Applicable (SEE12302371GA), Not Applicable (SEE12302372GA).Not Applicable (SEE12302374GA), Not Applicable (SEE12302375GA), Not Applicable (SEE12302378GA), Not Applicable (SEE12302385GA), Not Applicable (SEE09340431GA), Not Applicable (SEE09340433GA), Not Applicable (SEE09340435GA), Not Applicable (SEE11341645GA), Not Applicable (SEE11361698GA), Not Applicable (SEE11361740GA), Not Applicable (SEE09470663GA), Not Applicable (SEE11341653GA), Not Applicable (SEE13411204GA), Not Applicable (SEE13411206GA), Not Applicable (SEE13411208GA), Not Applicable (SEE09330428GA), Not Applicable (SEE12092070GA), Not Applicable (SEE12092071GA), Not Applicable (SEE12102099GA), Not Applicable (SEE12102102GA), Not Applicable (SEE12102103GA), Not Applicable (SEE09480738GA), Not Applicable (SEE09480746GA), Not Applicable (SEE09480747GA), 00840682109260SEE19220017SA (SEE19220017SA), Not Applicable (SEE15022865GA), Not Applicable (SEE16241847SA), Not Applicable (SEE16241856SA), 00840682109260SEE18330037SA (SEE18330037SA), Not Applicable (SEE15517888SA), Not Applicable (SEE15518040SA), Not Applicable (SEE15518041SA), Not Applicable (SEE15518042SA), Not Applicable (SEE15518045SA), Not Applicable (SEE12520335GA), Not Applicable (SEE13000386GA), Not Applicable (SEE10530849GA), Not Applicable (SEE10530859GA), Not Applicable (SEE10530864GA), Not Applicable (SEE10530871GA), Not Applicable (SEE11030880GA), Not Applicable (SEE11040898GA), Not Applicable (SEE10530827GA), Not Applicable (SEE10530831GA), Not Applicable (SEE10530832GA), Not Applicable (SEE10530833GA), Not Applicable (SEE10530834GA), Not Applicable (SEE10530842GA), Not Applicable (SEE12390082GA), Not Applicable (SEE13451333GA), 00840682109260SEE19100032SA (SEE19100032SA), Not Applicable (SEE09100089GA), Not Applicable (SEE09100131GA), Not Applicable (SEE09340441GA), Not Applicable (SEE09410576GA), Not Applicable (SEE10120021GA), Not Applicable (SEE10130074GA), Not Applicable (SEE10260230GA), Not Applicable (SEE10310283GA), Not Applicable (SEE10390325GA), Not Applicable (SEE10390326GA), Not Applicable (SEE10390337GA), Not Applicable (SEE10390353GA), Not Applicable (SEE10390359GA), Not Applicable (SEE10390360GA), Not Applicable (SEE11081173GA), Not Applicable (SEE11081174GA), Not Applicable (SEE11081179GA), Not Applicable (SEE11081182GA), Not Applicable (SEE11081199GA), Not Applicable (SEE11101222GA), Not Applicable (SEE11151315GA), Not Applicable (SEE11171358GA), Not Applicable (SEE11171361GA), Not Applicable (SEE11211512GA), Not Applicable (SEE14312301GA), Not Applicable (SEE14372424GA), Not Applicable (SEE14372439GA), Not Applicable (SEE14382460GA), Not Applicable (SEE14382462GA), Not Applicable (SEE14502785GA), Not Applicable (SEE14502786GA), Not Applicable (SEE14502789GA), Not Applicable (SEE14502790GA), Not Applicable (SEE14502799GA), Not Applicable (SEE14502800GA), Not Applicable (SEE14502802GA), Not Applicable (SEE14502803GA), Not Applicable (SEE14502808GA), Not Applicable (SEE15113158GA), Not Applicable (SEE15113160GA), Not Applicable (SEE15113162GA), Not Applicable (SEE15113164GA), Not Applicable (SEE15113165GA), Not Applicable (SEE15113178GA), Not Applicable (SEE15113179GA), Not Applicable (SEE15220769SA), Not Applicable (SEE15220770SA), Not Applicable (SEE15220773SA), Not Applicable (SEE15220774SA), Not Applicable (SEE15220776SA), Not Applicable (SEE15220778SA), Not Applicable (SEE15220780SA), Not Applicable (SEE15220782SA), Not Applicable (SEE15220783SA), Not Applicable (SEE15220784SA), Not Applicable (SEE15220785SA), Not Applicable (SEE15220791SA), Not Applicable (SEE15261531SA), Not Applicable (SEE15261538SA), Not Applicable (SEE15261539SA), Not Applicable (SEE16099500SA), Not Applicable (SEE16109587SA), Not Applicable (SEE16109701SA), Not Applicable (SEE16109705SA), Not Applicable (SEE16119914SA), Not Applicable (SEE16150555SA), 00840682109260SEE17230046SA (SEE17230046SA), 00840682109260SEE17230047SA (SEE17230047SA), 00840682109260SEE17230048SA (SEE17230048SA), 00840682109260SEE17230049SA (SEE17230049SA), 00840682109260SEE17230050SA (SEE17230050SA), 00840682109260SEE17230051SA (SEE17230051SA), 00840682109260SEE17230053SA (SEE17230053SA), 00840682109260SEE17230054SA (SEE17230054SA), 00840682109260SEE17230064SA (SEE17230064SA), 00840682109260SEE17230065SA (SEE17230065SA), 00840682109260SEE17230066SA (SEE17230066SA), 00840682109260SEE17230067SA (SEE17230067SA), 00840682109260SEE17230068SA (SEE17230068SA), 00840682109260SEE17230069SA (SEE17230069SA), 00840682109260SEE17230070SA (SEE17230070SA), 00840682109260SEE17230071SA (SEE17230071SA), 00840682109260SEE17230072SA (SEE17230072SA), 00840682109260SEE17230073SA (SEE17230073SA), 00840682109260SEE17430046SA (SEE17430046SA), 00840682109260SEE18010002SA (SEE18010002SA), 00840682109260SEE18010017SA (SEE18010017SA), 00840682109260SEE18010018SA (SEE18010018SA), 00840682109260SEE18010019SA (SEE18010019SA), 00840682109260SEE18010021SA (SEE18010021SA), 00840682109260SEE18010022SA (SEE18010022SA), 00840682109260SEE18170003SA (SEE18170003SA), 00840682109260SEE18190026SA (SEE18190026SA), 00840682109260SEE18190030SA (SEE18190030SA), 00840682109260SEE18190032SA (SEE18190032SA), 00840682109260SEE18190037SA (SEE18190037SA), 00840682109260SEE18190038SA (SEE18190038SA), Not Applicable (SEE09020011GA), Not Applicable (SEE09020033GA), Not Applicable (SEE09020038GA), Not Applicable (SEE09430604GA), Not Applicable (SEE11060991GA), Not Applicable (SEE11071017GA), Not Applicable (SEE11071033GA), Not Applicable (SEE11071050GA), Not Applicable (SEE11171371GA), 00840682109260SEE18170028SA (SEE18170028SA), 00840682109260SEE18190011SA (SEE18190011SA), 00840682109260SEE18190017SA (SEE18190017SA), 00840682109260SEE18190024SA (SEE18190024SA), 00840682109260SEE18250031SA (SEE18250031SA), 00840682109260SEE18260001SA (SEE18260001SA), Not Applicable (SEE09430622GA), Not Applicable (SEE14502787GA), Not Applicable (SEE14502796GA), Not Applicable (SEE14502801GA), Not Applicable (SEE14502805GA), Not Applicable (SEE14502807GA), Not Applicable (SEE09340430GA), Not Applicable (SEE09340434GA), Not Applicable (SEE09340437GA), Not Applicable (SEE09100129GA), Not Applicable (SEE10390297GA), Not Applicable (SEE11511882GA), 00840682109260SEE17250001SA (SEE17250001SA), 00840682109260SEE17250003SA (SEE17250003SA), 00840682109260SEE17250004SA (SEE17250004SA), 00840682109260SEE17250005SA (SEE17250005SA), 00840682109260SEE17250010SA (SEE17250010SA), 00840682109260SEE17250011SA (SEE17250011SA), 00840682109260SEE17250012SA (SEE17250012SA), 00840682109260SEE17250013SA (SEE17250013SA), 00840682109260SEE17250014SA (SEE17250014SA), 00840682109260SEE17250015SA (SEE17250015SA), 00840682109260SEE17250016SA (SEE17250016SA), 00840682109260SEE17250017SA (SEE17250017SA), 00840682109260SEE17250018SA (SEE17250018SA), 00840682109260SEE17250019SA (SEE17250019SA), Not Applicable (SEE14322314GA), Not Applicable (SEE15220768SA), Not Applicable (SEE15220771SA), Not Applicable (SEE15220772SA), Not Applicable (SEE15220777SA), Not Applicable (SEE15220779SA), Not Applicable (SEE15220781SA), Not Applicable (SEE10130064GA), Not Applicable (SEE10130075GA), Not Applicable (SEE12082029GA), Not Applicable (SEE12082036GA), Not Applicable (SEE12262327GA), Not Applicable (SEE12262347GA), Not Applicable (SEE14021569GA), Not Applicable (SEE13431288GA), Not Applicable (SEE13431289GA), Not Applicable (SEE10430646GA), Not Applicable (SEE13401183GA), Not Applicable (SEE09530849GA), Not Applicable (SEE09530858GA), Not Applicable (SEE09530875GA), Not Applicable (SEE12340003GA), Not Applicable (SEE12370058GA), Not Applicable (SEE12370066GA), Not Applicable (SEE09530874GA), Not Applicable (SEE09530900GA), Not Applicable (SEE10090957GA), Not Applicable (SEE10121000GA), Not Applicable (SEE10121002GA), Not Applicable (SEE10121013GA), Not Applicable (SEE10121022GA), Not Applicable (SEE14222006GA), Not Applicable (SEE14222009GA), Not Applicable (SEE14222019GA), Not Applicable (SEE09480723GA), Not Applicable (SEE09480730GA), Not Applicable (SEE09480742GA), Not Applicable (SEE09480745GA), Not Applicable (SEE09490752GA), Not Applicable (SEE09490754GA), Not Applicable (SEE10410490GA), Not Applicable (SEE13040420GA), Not Applicable (SEE15153194GA), 00840682109260SEE16078894SA (SEE16078894SA), 00840682109260SEE16078896SA (SEE16078896SA), 00840682109260SEE16078897SA (SEE16078897SA), 00840682109260SEE18040020SA (SEE18040020SA), 00840682109260SEE18040021SA (SEE18040021SA), Not Applicable (SEE13501486GA), Not Applicable (SEE13501487GA), Not Applicable (SEE13501489GA), Not Applicable (SEE13501490GA), Not Applicable (SEE13501494GA), Not Applicable (SEE13501495GA), Not Applicable (SEE13501496GA), Not Applicable (SEE13501497GA), Not Applicable (SEE13501501GA), Not Applicable (SEE15507702SA), 00840682109260SEE11511869GA (SEE11511869GA), 00840682109260SEE11511879GA (SEE11511879GA), 00840682109260SEE11521888GA (SEE11521888GA), Not Applicable (SEE12340001GA), Not Applicable (SEE12340008GA), 00840682109260SEE12340015GA (SEE12340015GA), Not Applicable (SEE12342446GA), 00840682109260SEE12490262GA (SEE12490262GA), 00840682109260SEE12520360GA (SEE12520360GA), 00840682109260SEE12520363GA (SEE12520363GA), 00840682109260SEE12520364GA (SEE12520364GA), Not Applicable (SEE14432613GA), Not Applicable (SEE14492742GA), 00840682109260SEE17430020SA (SEE17430020SA), Not Applicable (SEE11081165GA), 00840682109260SEE15230837SA (SEE15230837SA), 00840682109260SEE17350005SA (SEE17350005SA), 00840682109260SEE17350006SA (SEE17350006SA), 00840682109260SEE17350007SA (SEE17350007SA), 00840682109260SEE17150001SA (SEE17150001SA), 00840682109260SEE17150003SA (SEE17150003SA), 00840682109260SEE17150007SA (SEE17150007SA), 00840682109260SEE18050005SA (SEE18050005SA), 00840682109260SEE18050006SA (SEE18050006SA), 00840682109260SEE18050007SA (SEE18050007SA), 00840682109260SEE18050008SA (SEE18050008SA), 00840682109260SEE18050009SA (SEE18050009SA), 00840682109260SEE18050010SA (SEE18050010SA), 00840682109260SEE18050011SA (SEE18050011SA), Not Applicable (SEE10120029GA), Not Applicable (SEE11371757GA), Not Applicable (SEE11371800GA), Not Applicable (SEE13491442GA), Not Applicable (SEE13491443GA), Not Applicable (SEE13491444GA), Not Applicable (SEE13491446GA), Not Applicable (SEE13491450GA), Not Applicable (SEE13491451GA), Not Applicable (SEE10120023GA), 00840682109260SEE17510001SA (SEE17510001SA), 00840682109260SEE17510010SA (SEE17510010SA), Not Applicable (SEE10121011GA), Not Applicable (SEE10130065GA), Not Applicable (SEE15425793SA), Not Applicable (SEE15425796SA), Not Applicable (SEE15425798SA), Not Applicable (SEE15425799SA), Not Applicable (SEE15425800SA), Not Applicable (SEE09100148GA), Not Applicable (SEE09430607GA), Not Applicable (SEE10400365GA), Not Applicable (SEE10400375GA), Not Applicable (SEE10400385GA), Not Applicable (SEE10400397GA), Not Applicable (SEE12490282GA), Not Applicable (SEE13431259GA), Not Applicable (SEE13431260GA), Not Applicable (SEE13431268GA), Not Applicable (SEE13180784GA), Not Applicable (SEE09470674GA), Not Applicable (SEE12322421GA), Not Applicable (SEE14031672GA), Not Applicable (SEE15113123GA), Not Applicable (SEE15113141GA), Not Applicable (SEE15113156GA), Not Applicable (SEE13110527GA), Not Applicable (SEE13110547GA), Not Applicable (SEE13110551GA), Not Applicable (SEE13110552GA), Not Applicable (SEE13110553GA), Not Applicable (SEE13110555GA), Not Applicable (SEE13110557GA), Not Applicable (SEE13110558GA), Not Applicable (SEE13110559GA), Not Applicable (SEE13110560GA), Not Applicable (SEE13110564GA), Not Applicable (SEE13451317GA), Not Applicable (SEE13451817GA), Not Applicable (SEE13481390GA), Not Applicable (SEE13481391GA), Not Applicable (SEE13481392GA), Not Applicable (SEE13481395GA), Not Applicable (SEE13481396GA), 00840682109260SEE17510005SA (SEE17510005SA), Not Applicable (SEE13491439GA), Not Applicable (SEE13491441GA), Not Applicable (SEE15113120GA), Not Applicable (SEE15113142GA), Not Applicable (SEE15113144GA), Not Applicable (SEE15113148GA), 00840682109260SEE17150018SA (SEE17150018SA), Not Applicable (SEE13361073GA), Not Applicable (SEE13371081GA), Not Applicable (SEE13371091GA), Not Applicable (SEE14362357GA), Not Applicable (SEE14452621GA), Not Applicable (SEE15466917SA), Not Applicable (SEE15466918SA), Not Applicable (SEE15466923SA), Not Applicable (SEE15466924SA), Not Applicable (SEE15466925SA), Not Applicable (SEE15466934SA), Not Applicable (SEE15466935SA), Not Applicable (SEE09170221GA), 00840682109260SEE17130008SA (SEE17130008SA), Not Applicable (SEE14151851GA), Not Applicable (SEE13411198GA), Not Applicable (SEE13411212GA), Not Applicable (SEE13411214GA), 00840682109260SEE19090026SA (SEE19090026SA), 00840682109260SEE19090027SA (SEE19090027SA), Not Applicable (SEE13290960GA), Not Applicable (SEE15183254GA), Not Applicable (SEE15183255GA), Not Applicable (SEE14211983GA), Not Applicable (SEE14252123GA), Not Applicable (SEE14252124GA), Not Applicable (SEE14252125GA), Not Applicable (SEE14252128GA), Not Applicable (SEE14252130GA), Not Applicable (SEE14252131GA), Not Applicable (SEE14252133GA), Not Applicable (SEE14252135GA), Not Applicable (SEE14362353GA), Not Applicable (SEE09510798GA), Not Applicable (SEE10250081GA), Not Applicable (SEE10250091GA), Not Applicable (SEE10250095GA), Not Applicable (SEE10250104GA), Not Applicable (SEE10250110GA), Not Applicable (SEE10250122GA), Not Applicable (SEE10390309GA), Not Applicable (SEE11361725GA), Not Applicable (SEE14512832GA), Not Applicable (SEE14512837GA), Not Applicable (SEE14512838GA), Not Applicable (SEE15012839GA), Not Applicable (SEE15012840GA), Not Applicable (SEE15193279GA), Not Applicable (SEE12510310GA), 00840682109260SEE18370003SA (SEE18370003SA), 00840682109260SEE18370004SA (SEE18370004SA), Not Applicable (SEE09530848GA), Not Applicable (SEE09530852GA), Not Applicable (SEE09020028GA), Not Applicable (SEE10120993GA), Not Applicable (SEE10121005GA), Not Applicable (SEE10121007GA), Not Applicable (SEE10121008GA), Not Applicable (SEE11361713GA), Not Applicable (SEE11361721GA), Not Applicable (SEE11361723GA), Not Applicable (SEE12252316GA), Not Applicable (SEE12252322GA), Not Applicable (SEE12302386GA), Not Applicable (SEE12312397GA), Not Applicable (SEE12312399GA), Not Applicable (SEE12312405GA), Not Applicable (SEE13381124GA), Not Applicable (SEE13381128GA), 01008406821092601116101121SEE16410010SA (SEE13401141GA), Not Applicable (SEE13401154GA), Not Applicable (SEE13401158GA), Not Applicable (SEE13401165GA), Not Applicable (SEE13401168GA), Not Applicable (SEE13401169GA), Not Applicable (SEE13401170GA), Not Applicable (SEE13401171GA), Not Applicable (SEE13501472GA), Not Applicable (SEE14031594GA), Not Applicable (SEE14031604GA), Not Applicable (SEE14031606GA), Not Applicable (SEE14222018GA), Not Applicable (SEE15042906GA), Not Applicable (SEE15052910GA), Not Applicable (SEE15052911GA), Not Applicable (SEE15052924GA), Not Applicable (SEE15052931GA), 00840682109260SEE18160041SA (SEE18160041SA), 00840682109260SEE18160043SA (SEE18160043SA), 00840682109260SEE18160047SA (SEE18160047SA), 00840682109260SEE18160053SA (SEE18160053SA), 00840682109260SEE18160060SA (SEE18160060SA), 00840682109260SEE19120022SA (SEE19120022SA), 00840682109260SEE19120023SA (SEE19120023SA), 00840682109260SEE19120025SA (SEE19120025SA), 00840682109260SEE19120031SA (SEE19120031SA), 00840682109260SEE19120032SA (SEE19120032SA), 00840682109260SEE19120033SA (SEE19120033SA), 00840682109260SEE19120034SA (SEE19120034SA), 00840682109260SEE19130009SA (SEE19130009SA), 00840682109260SEE19130020SA (SEE19130020SA), 00840682109260SEE19150008SA (SEE19150008SA), Not Applicable (SEE15517774SA), Not Applicable (SEE15517821SA), Not Applicable (SEE15517824SA), Not Applicable (SEE11511881GA), Not Applicable (SEE14021519GA), 00840682109260SEE19220020SA (SEE19220020SA), 00840682109260SEE19220030SA (SEE19220030SA), 00840682109260SEE19220031SA (SEE19220031SA), 00840682109260SEE19220032SA (SEE19220032SA), 00840682109260SEE19220040SA (SEE19220040SA), Not Applicable (SEE14502791GA), Not Applicable (SEE14502793GA), Not Applicable (SEE14502797GA), Not Applicable (SEE14502798GA), Not Applicable (SEE14512819GA), Not Applicable (SEE14512823GA), Not Applicable (SEE14512827GA), Not Applicable (SEE16119768SA), Not Applicable (SEE13110502GA), Not Applicable (SEE13110524GA), Not Applicable (SEE14472676GA), Not Applicable (SEE14482685GA), Not Applicable (SEE14482687GA), Not Applicable (SEE14482691GA), Not Applicable (SEE14482692GA), Not Applicable (SEE14482693GA), Not Applicable (SEE14482694GA), Not Applicable (SEE14482696GA), Not Applicable (SEE14482699GA), Not Applicable (SEE14362379GA), Not Applicable (SEE14362380GA), Not Applicable (SEE14362383GA), Not Applicable (SEE14362386GA), Not Applicable (SEE14372388GA), Not Applicable (SEE14372392GA), Not Applicable (SEE14372393GA), Not Applicable (SEE14372394GA), 00840682109260SEE18060005SA (SEE18060005SA), 00840682109260SEE18060006SA (SEE18060006SA), 00840682109260SEE19130016SA (SEE19130016SA), Not Applicable (SEE09250307GA), Not Applicable (SEE15446332SA), Not Applicable (SEE12092056GA), Not Applicable (SEE13180762GA), Not Applicable (SEE09190226GA), Not Applicable (SEE11171374GA), Not Applicable (SEE11081195GA), 00840682109260SEE17150004SA (SEE17150004SA), Not Applicable (SEE16252018SA), Not Applicable (SEE10450670GA), Not Applicable (SEE13401161GA), Not Applicable (SEE12302389GA), Not Applicable (SEE12302390GA), Not Applicable (SEE15251303SA), Not Applicable (SEE12322434GA), Not Applicable (SEE15352700SA), Not Applicable (SEE10420589GA), Not Applicable (SEE12430223GA), Not Applicable (SEE12430225GA), Not Applicable (SEE12440234GA), Not Applicable (SEE12440237GA), Not Applicable (SEE12490261GA), Not Applicable (SEE13371113GA), Not Applicable (SEE14372438GA), Not Applicable (SEE10270244GA), Not Applicable (SEE16303409GA), Not Applicable (SEE10250168GA), 00840682109260SEE18110004SA (SEE18110004SA), Not Applicable (SEE14242082GA), Not Applicable (SEE14242084GA), Not Applicable (SEE14242093GA), Not Applicable (SEE10120006GA), Not Applicable (SEE10120012GA), Not Applicable (SEE10120016GA), Not Applicable (SEE10120034GA), Not Applicable (SEE10121003GA), Not Applicable (SEE14021559GA), Not Applicable (SEE14021574GA), Not Applicable (SEE14021575GA), Not Applicable (SEE14272145GA), Not Applicable (SEE14302274GA), Not Applicable (SEE14312295GA), Not Applicable (SEE14322320GA), Not Applicable (SEE14322335GA), Not Applicable (SEE16038425SA), Not Applicable (SEE14432579GA), 00840682109260SEE17080011SA (SEE17080011SA), Not Applicable (SEE12520337GA), Not Applicable (SEE12520352GA), Not Applicable (SEE12520356GA), Not Applicable (SEE12520361GA), Not Applicable (SEE12520362GA), Not Applicable (SEE14362382GA), Not Applicable (SEE14362387GA), 00840682109260SEE18470010SA (SEE18470010SA), 00840682109260SEE18470011SA (SEE18470011SA), 00840682109260SEE18470012SA (SEE18470012SA), 00840682109260SEE18470013SA (SEE18470013SA), 00840682109260SEE18470014SA (SEE18470014SA), 00840682109260SEE18470017SA (SEE18470017SA), 00840682109260SEE18470021SA (SEE18470021SA), 00840682109260SEE18470025SA (SEE18470025SA), 00840682109260SEE18470026SA (SEE18470026SA), 00840682109260SEE18470027SA (SEE18470027SA), Not Applicable (SEE12520355GA), Not Applicable (SEE14151813GA), Not Applicable (SEE14151881GA), 00840682109260SEE17480011SA (SEE17480011SA), Not Applicable (SEE16430003SA), Not Applicable (SEE14021514GA), Not Applicable (SEE14161893GA), Not Applicable (SEE15415171SA), Not Applicable (SEE14492729GA), Not Applicable (SEE14492743GA), 00840682109260SEE18040040SA (SEE18040040SA), Not Applicable (SEE13411215GA), Not Applicable (SEE10120019GA), 00840682109260SEE17050012SA (SEE17050012SA), 00840682109260SEE17050013SA (SEE17050013SA), 00840682109260SEE17050014SA (SEE17050014SA), Not Applicable (SEE15173226GA), Not Applicable (SEE09100150GA), Not Applicable (SEE11071007GA), Not Applicable (SEE11141258GA), Not Applicable (SEE11141266GA), Not Applicable (SEE15173228GA), Not Applicable (SEE15173229GA), Not Applicable (SEE15173241GA), Not Applicable (SEE15173242GA), Not Applicable (SEE15173246GA), Not Applicable (SEE16221553SA), Not Applicable (SEE16221557SA), 00840682109260SEE19100023SA (SEE19100023SA), 00840682109260SEE19100024SA (SEE19100024SA), 00840682109260SEE19100029SA (SEE19100029SA), 00840682109260SEE19100031SA (SEE19100031SA), 00840682109260SEE19100034SA (SEE19100034SA), 00840682109260SEE19100063SA (SEE19100063SA), 00840682109260SEE19120011SA (SEE19120011SA), 00840682109260SEE19120013SA (SEE19120013SA), 00840682109260SEE17250026SA (SEE17250026SA), 00840682109260SEE17250027SA (SEE17250027SA), 00840682109260SEE17250028SA (SEE17250028SA), 00840682109260SEE17270016SA (SEE17270016SA), Not Applicable (SEE13150696GA), Not Applicable (SEE13150698GA), Not Applicable (SEE13150699GA), Not Applicable (SEE13150700GA), Not Applicable (SEE15352677SA), Not Applicable (SEE16089278SA), Not Applicable (SEE16262252SA), Not Applicable (SEE13361077GA), Not Applicable (SEE13491435GA), Not Applicable (SEE13491438GA), Not Applicable (SEE14101777GA), Not Applicable (SEE14111799GA), Not Applicable (SEE14111805GA), Not Applicable (SEE12520338GA), Not Applicable (SEE12520344GA), 00840682109260SEE17040001SA (SEE17040001SA), 00840682109260SEE17180016SA (SEE17180016SA), Not Applicable (SEE11050976GA), Not Applicable (SEE10100972GA), Not Applicable (SEE11071012GA), Not Applicable (SEE11071022GA), Not Applicable (SEE11071095GA), Not Applicable (SEE14302273GA), Not Applicable (SEE11141272GA), Not Applicable (SEE10420557GA), Not Applicable (SEE10430606GA), Not Applicable (SEE10450659GA), Not Applicable (SEE10450676GA), Not Applicable (SEE14362350GA), Not Applicable (SEE14362351GA), Not Applicable (SEE14362374GA), Not Applicable (SEE15517823SA), Not Applicable (SEE13180767GA), Not Applicable (SEE15173212GA), Not Applicable (SEE12021914GA), Not Applicable (SEE12062005GA), Not Applicable (SEE12082024GA), 00840682109260SEE17100006SA (SEE17100006SA), 00840682109260SEE18510017SA (SEE18510017SA), 00840682109260SEE18510020SA (SEE18510020SA), 00840682109260SEE18510025SA (SEE18510025SA), 00840682109260SEE18510026SA (SEE18510026SA), 00840682109260SEE18510027SA (SEE18510027SA), 00840682109260SEE18510029SA (SEE18510029SA), 00840682109260SEE19010005SA (SEE19010005SA), 00840682109260SEE19010006SA (SEE19010006SA), 00840682109260SEE19010011SA (SEE19010011SA), 00840682109260SEE19010012SA (SEE19010012SA), 00840682109260SEE19010013SA (SEE19010013SA), 00840682109260SEE19010014SA (SEE19010014SA), Not Applicable (SEE13280904GA), Not Applicable (SEE14312299GA), Not Applicable (SEE14312304GA), Not Applicable (SEE14312305GA), Not Applicable (SEE14312306GA), Not Applicable (SEE14422551GA), Not Applicable (SEE11381815GA), 00840682109260SEE18420014SA (SEE18420014SA), 00840682109260SEE18420025SA (SEE18420025SA), 00840682109260SEE18420040SA (SEE18420040SA), 00840682109260SEE18420050SA (SEE18420050SA), Not Applicable (SEE14482690GA), Not Applicable (SEE11181397GA), Not Applicable (SEE11191424GA), Not Applicable (SEE11191426GA), Not Applicable (SEE11191438GA), 00840682109260SEE19190004SA (SEE19190004SA), Not Applicable (SEE16201405SA), Not Applicable (SEE16211410SA), Not Applicable (SEE16211412SA), Not Applicable (SEE16211420SA), Not Applicable (SEE16211507SA), Not Applicable (SEE16211508SA), Not Applicable (SEE16211509SA), Not Applicable (SEE16211511SA), Not Applicable (SEE16211512SA), 00840682109260SEE09510805GA (SEE09510805GA), 00840682109260SEE09510825GA (SEE09510825GA), 00840682109260SEE11081206GA (SEE11081206GA), 00840682109260SEE11081208GA (SEE11081208GA), 00840682109260SEE12041948GA (SEE12041948GA), 00840682109260SEE12092074GA (SEE12092074GA), 00840682109260SEE12122144GA (SEE12122144GA), 00840682109260SEE12162174GA (SEE12162174GA), 00840682109260SEE12162176GA (SEE12162176GA), 00840682109260SEE12172189GA (SEE12172189GA), 00840682109260SEE12172191GA (SEE12172191GA), 00840682109260SEE12172199GA (SEE12172199GA), 00840682109260SEE12182207GA (SEE12182207GA), 00840682109260SEE12182208GA (SEE12182208GA), 00840682109260SEE12182209GA (SEE12182209GA), 00840682109260SEE12182213GA (SEE12182213GA), 00840682109260SEE12182216GA (SEE12182216GA), 00840682109260SEE12182222GA (SEE12182222GA), 00840682109260SEE12182228GA (SEE12182228GA), 00840682109260SEE12182229GA (SEE12182229GA), 00840682109260SEE13501457GA (SEE13501457GA), 00840682109260SEE13501480GA (SEE13501480GA), 00840682109260SEE14041739GA (SEE14041739GA), 00840682109260SEE14041743GA (SEE14041743GA), Not Applicable (SEE15456708SA), Not Applicable (SEE15477181SA), 00840682109260SEE17520003SA (SEE17520003SA), Not Applicable (SEE14312297GA), Not Applicable (SEE14312300GA), Not Applicable (SEE14312302GA), Not Applicable (SEE14312303GA), Not Applicable (SEE16119851GA), Not Applicable (SEE16119851SA), Not Applicable (SEE10460691GA), Not Applicable (SEE10460702GA), Not Applicable (SEE10470786GA), Not Applicable (SEE10470795GA), Not Applicable (SEE13180757GA), Not Applicable (SEE13180759GA), Not Applicable (SEE12182214GA), Not Applicable (SEE12182220GA), 00840682109260SEE18010020SA (SEE18010020SA), Not Applicable (SEE09230263GA), Not Applicable (SEE09240291GA), Not Applicable (SEE14021511GA), Not Applicable (SEE14021530GA), Not Applicable (SEE14161897GA), Not Applicable (SEE14161898GA), Not Applicable (SEE14161902GA), Not Applicable (SEE14161903GA), Not Applicable (SEE14161904GA), Not Applicable (SEE14161906GA), Not Applicable (SEE14161908GA), Not Applicable (SEE13110538GA), Not Applicable (SEE13110542GA), Not Applicable (SEE13120580GA), Not Applicable (SEE13120587GA), Not Applicable (SEE13120589GA), Not Applicable (SEE13120590GA), Not Applicable (SEE13120603GA), Not Applicable (SEE16109700SA), Not Applicable (SEE16120021SA), Not Applicable (SEE16221556SA), 00840682109260SEE17090014SA (SEE17090014SA), 00840682109260SEE17090015SA (SEE17090015SA), Not Applicable (SEE15281648SA), Not Applicable (SEE14482682GA), Not Applicable (SEE14482686GA), Not Applicable (SEE10270241GA), Not Applicable (SEE10310281GA), Not Applicable (SEE14151835GA), Not Applicable (SEE14151840GA), Not Applicable (SEE14151845GA), Not Applicable (SEE14151846GA), Not Applicable (SEE14151852GA), Not Applicable (SEE14151869GA), Not Applicable (SEE14171916GA), Not Applicable (SEE15456716SA), Not Applicable (SEE15497518SA), Not Applicable (SEE15497519SA), Not Applicable (SEE15497527SA), Not Applicable (SEE15497582SA), Not Applicable (SEE15507660SA), Not Applicable (SEE16430049SA), Not Applicable (SEE16430057SA), Not Applicable (SEE16440001SA), Not Applicable (SEE16440005SA), 00840682109260SEE17080012SA (SEE17080012SA), 00840682109260SEE17090011SA (SEE17090011SA), 00840682109260SEE17090013SA (SEE17090013SA), 00840682109260SEE17090016SA (SEE17090016SA), 00840682109260SEE17090018SA (SEE17090018SA), 00840682109260SEE17090019SA (SEE17090019SA), 00840682109260SEE17190022SA (SEE17190022SA), 00840682109260SEE17090024SA (SEE17090024SA), Not Applicable (SEE11371746GA), 00840682109260SEE11371768GA (SEE11371768GA), Not Applicable (SEE11371770GA), Not Applicable (SEE11371771GA), Not Applicable (SEE11371772GA), Not Applicable (SEE11371773GA), Not Applicable (SEE11371776GA), Not Applicable (SEE11371777GA), Not Applicable (SEE11371778GA), Not Applicable (SEE11371780GA), Not Applicable (SEE11371783GA), Not Applicable (SEE12051993GA), Not Applicable (SEE12051994GA), Not Applicable (SEE12051995GA), Not Applicable (SEE15173213GA), 00840682109260SEE19110010SA (SEE19110010SA), 00840682109260SEE19110012SA (SEE19110012SA), 00840682109260SEE19110013SA (SEE19110013SA), 00840682109260SEE19110014SA (SEE19110014SA), 00840682109260SEE19110015SA (SEE19110015SA), 00840682109260SEE19110020SA (SEE19110020SA), 00840682109260SEE19110021SA (SEE19110021SA), Not Applicable (SEE10250126GA), Not Applicable (SEE10250151GA), Not Applicable (SEE10250153GA), Not Applicable (SEE10250154GA), Not Applicable (SEE15373617SA), Not Applicable (SEE14492744GA), Not Applicable (SEE10460747GA), Not Applicable (SEE10460752GA), Not Applicable (SEE11030887GA), 00840682109260SEE18170024SA (SEE18170024SA), Not Applicable (SEE10530872GA), Not Applicable (SEE110308789GA), Not Applicable (SEE11030879GA), Not Applicable (SEE15517782SA), Not Applicable (SEE09020009GA), Not Applicable (SEE09020013GA), Not Applicable (SEE09020023GA), Not Applicable (SEE09020024GA), Not Applicable (SEE09020040GA), Not Applicable (SEE09020042GA), Not Applicable (SEE09040045GA), Not Applicable (SEE09040051GA), Not Applicable (SEE09040054GA), Not Applicable (SEE09040055GA), Not Applicable (SEE09040058GA), Not Applicable (SEE09040060GA), Not Applicable (SEE09040061GA), Not Applicable (SEE09040064GA), Not Applicable (SEE09040069GA), Not Applicable (SEE09040071GA), Not Applicable (SEE09040072GA), Not Applicable (SEE09040075GA), Not Applicable (SEE09040077GA), Not Applicable (SEE09100087GA), Not Applicable (SEE09100095GA), Not Applicable (SEE09100105GA), Not Applicable (SEE09100108GA), Not Applicable (SEE09100109GA), Not Applicable (SEE09100134GA), Not Applicable (SEE09100139GA), Not Applicable (SEE09100140GA), Not Applicable (SEE09100152GA), Not Applicable (SEE09140183GA), Not Applicable (SEE09190244GA), Not Applicable (SEE09230253GA), Not Applicable (SEE09230266GA), Not Applicable (SEE09250310GA), Not Applicable (SEE09370494GA), Not Applicable (SEE09370502GA), Not Applicable (SEE09410552GA), Not Applicable (SEE09410561GA), Not Applicable (SEE09410563GA), Not Applicable (SEE09410564GA), Not Applicable (SEE09410565GA), Not Applicable (SEE09410566GA), Not Applicable (SEE09410567GA), Not Applicable (SEE09410574GA), Not Applicable (SEE09410577GA), Not Applicable (SEE09410578GA), Not Applicable (SEE09410579GA), Not Applicable (SEE09410587GA), Not Applicable (SEE10420525GA), Not Applicable (SEE10420535GA), Not Applicable (SEE10420547GA), Not Applicable (SEE10420597GA), Not Applicable (SEE10430607GA), Not Applicable (SEE10430609GA), Not Applicable (SEE10430611GA), Not Applicable (SEE10430617GA), Not Applicable (SEE10430621GA), Not Applicable (SEE10430635GA), Not Applicable (SEE10430642GA), Not Applicable (SEE10430644GA), Not Applicable (SEE10450682GA), Not Applicable (SEE10460693GA), Not Applicable (SEE10460700GA), Not Applicable (SEE10460728GA), Not Applicable (SEE10460758GA), Not Applicable (SEE10460775GA), Not Applicable (SEE10470776GA), Not Applicable (SEE10470779GA), Not Applicable (SEE10470780GA), Not Applicable (SEE10470781GA), Not Applicable (SEE10470797GA), Not Applicable (SEE12530365GA), Not Applicable (SEE13270881GA), Not Applicable (SEE13491440GA), Not Applicable (SEE14031597GA), Not Applicable (SEE14031598GA), Not Applicable (SEE14031601GA), Not Applicable (SEE14031602GA), Not Applicable (SEE14031693GA), Not Applicable (SEE14031694GA), Not Applicable (SEE14031699GA), Not Applicable (SEE14031701GA), Not Applicable (SEE14031709GA), Not Applicable (SEE14031711GA), Not Applicable (SEE14031713GA), Not Applicable (SEE14041762GA), Not Applicable (SEE14151868GA), Not Applicable (SEE14151871GA), Not Applicable (SEE16119903SA), Not Applicable (SEE16430033SA), Not Applicable (SEE16430048SA), 00840682109260SEE17430043SA (SEE17430043SA), 00840682109260SEE17430044SA (SEE17430044SA), 00840682109260SEE17490005SA (SEE17490005SA), 00840682109260SEE17490008SA (SEE17490008SA), 00840682109260SEE17490011SA (SEE17490011SA), 00840682109260SEE17490013SA (SEE17490013SA), 00840682109260SEE18040053SA (SEE18040053SA), 00840682109260SEE18290005SA (SEE18290005SA), 00840682109260SEE18320008SA (SEE18320008SA), Not Applicable (SEE11141295GA), Not Applicable (SEE14101784GA), Not Applicable (SEE14222028GA), Not Applicable (SEE14222037GA), Not Applicable (SEE15394376SA), Not Applicable (SEE16068748SA), Not Applicable (SEE16241826SA), Not Applicable (SEE16292869SA), 00840682109260SEE17430045SA (SEE17430045SA), Not Applicable (SEE16390014SA), Not Applicable (SEE11201448GA), Not Applicable (SEE14322313GA), Not Applicable (SEE14322316GA), Not Applicable (SEE09310360GA), Not Applicable (SEE09310378GA), Not Applicable (SEE14422525GA), Not Applicable (SEE14422529GA), Not Applicable (SEE15394730SA), Not Applicable (SEE15394732SA), Not Applicable (SEE15394741SA), Not Applicable (SEE15405018SA), 00840682109260SEE18420054SA (SEE18420054SA), Not Applicable (SEE10400383GA), Not Applicable (SEE10400396GA), Not Applicable (SEE10400429GA), Not Applicable (SEE10400454GA), Not Applicable (SEE10400458GA), Not Applicable (SEE10420569GA), Not Applicable (SEE11231543GA), 00840682109260SEE14041744GA (SEE14041744GA), Not Applicable (SEE14041753GA), Not Applicable (SEE14292241GA), Not Applicable (SEE14302265GA), Not Applicable (SEE14322324GA), Not Applicable (SEE14322328GA), Not Applicable (SEE14322337GA), 00840682109260SEE19240018SA (SEE19240018SA), Not Applicable (SEE09340446GA), Not Applicable (SEE12302384GA), Not Applicable (SEE12340020GA), Not Applicable (SEE13110493GA), Not Applicable (SEE13110540GA), Not Applicable (SEE13120602GA), Not Applicable (SEE13140658GA), Not Applicable (SEE14031612GA), Not Applicable (SEE09020012GA), Not Applicable (SEE09020031GA), Not Applicable (SEE10130042GA), Not Applicable (SEE10250094GA), Not Applicable (SEE10420522GA), Not Applicable (SEE10460764GA), Not Applicable (SEE11030886GA), Not Applicable (SEE11071014GA), Not Applicable (SEE11071042GA), Not Applicable (SEE11071149GA), Not Applicable (SEE11081183GA), Not Applicable (SEE11081184GA), Not Applicable (SEE11111236GA), Not Applicable (SEE11111238GA), Not Applicable (SEE11111246GA), Not Applicable (SEE11141278GA), Not Applicable (SEE11141286GA), Not Applicable (SEE11201462GA), Not Applicable (SEE11211474GA), Not Applicable (SEE11211479GA), Not Applicable (SEE11211485GA), Not Applicable (SEE11211490GA), Not Applicable (SEE11211499GA), Not Applicable (SEE11211502GA), Not Applicable (SEE11211522GA), Not Applicable (SEE11231530GA), Not Applicable (SEE11231545GA), Not Applicable (SEE11241548GA), Not Applicable (SEE11341658GA), Not Applicable (SEE11341662GA), Not Applicable (SEE11371750GA), Not Applicable (SEE11371791GA), Not Applicable (SEE11371811GA), Not Applicable (SEE11511873GA), Not Applicable (SEE11511880GA), Not Applicable (SEE12011896GA), Not Applicable (SEE12092042GA), Not Applicable (SEE12162172GA), Not Applicable (SEE12182205GA), Not Applicable (SEE12262334GA), Not Applicable (SEE12262342GA), Not Applicable (SEE12262346GA), Not Applicable (SEE12322427GA), Not Applicable (SEE12390074GA), Not Applicable (SEE12420173GA), Not Applicable (SEE12510313GA), Not Applicable (SEE13110497GA), Not Applicable (SEE13110499GA), Not Applicable (SEE13110501GA), Not Applicable (SEE13110506GA), Not Applicable (SEE13110518GA), Not Applicable (SEE13110522GA), Not Applicable (SEE13110523GA), Not Applicable (SEE13150659GA), Not Applicable (SEE13361078GA), Not Applicable (SEE13411205GA), Not Applicable (SEE14031675GA), Not Applicable (SEE14031700GA), Not Applicable (SEE14041749GA), Not Applicable (SEE14041752GA), Not Applicable (SEE14041754GA), Not Applicable (SEE14041755GA), Not Applicable (SEE14041764GA), Not Applicable (SEE14191950GA), Not Applicable (SEE14201951GA), Not Applicable (SEE14201954GA), Not Applicable (SEE14201955GA), Not Applicable (SEE14201957GA), Not Applicable (SEE14211965GA), Not Applicable (SEE14211966GA), Not Applicable (SEE14211967GA), Not Applicable (SEE14211968GA), Not Applicable (SEE14211969GA), Not Applicable (SEE14211970GA), Not Applicable (SEE14211971GA), Not Applicable (SEE14211973GA), Not Applicable (SEE14211974GA), Not Applicable (SEE14211975GA), Not Applicable (SEE14211977GA), Not Applicable (SEE14272146GA), Not Applicable (SEE14272164GA), Not Applicable (SEE14272168GA), Not Applicable (SEE14272170GA), Not Applicable (SEE14272175GA), Not Applicable (SEE14272224GA), Not Applicable (SEE14332343GA), Not Applicable (SEE14362369GA), Not Applicable (SEE15032882GA), Not Applicable (SEE15062980GA), Not Applicable (SEE15062986GA), Not Applicable (SEE15210703SA), Not Applicable (SEE15251288SA), Not Applicable (SEE15251289SA), Not Applicable (SEE15251290SA), Not Applicable (SEE15251291SA), Not Applicable (SEE16272332SA), Not Applicable (SEE16293173SA), Not Applicable (SEE16293179SA), Not Applicable (SEE16303212SA).Not Applicable (SEE16303220SA), Not Applicable (SEE16334183SA), Not Applicable (SEE16360018SA), Not Applicable (SEE16360019SA), Not Applicable (SEE16360022SA), Not Applicable (SEE16364417SA), 00840682109260SEE17070001SA (SEE17070001SA), 00840682109260SEE17070003SA (SEE17070003SA), 00840682109260SEE17260007SA (SEE17260007SA), 00840682109260SEE18320004SA (SEE18320004SA), 00840682109260SEE18370006SA (SEE18370006SA), 00840682109260SEE18390003SA (SEE18390003SA), 00840682109260SEE18420041SA (SEE18420041SA), 00840682109260SEE18420045SA (SEE18420045SA), 00840682109260SEE18470016SA (SEE18470016SA), 00840682109260SEE18470020SA (SEE18470020SA), 00840682109260SEE18510022SA (SEE18510022SA), Not Applicable (SEE12282354GA), Not Applicable (SEE12282356GA), Not Applicable (SEE13290984GA), Not Applicable (SEE13311004GA), Not Applicable (SEE13351011GA), Not Applicable (SEE13351013GA), Not Applicable (SEE13351016GA), Not Applicable (SEE13351017GA), Not Applicable (SEE13351024GA), Not Applicable (SEE13351027GA), Not Applicable (SEE13351028GA), Not Applicable (SEE13351029GA), Not Applicable (SEE13351032GA), Not Applicable (SEE13491428GA), Not Applicable (SEE11341655GA), Not Applicable (SEE11341659GA), Not Applicable (SEE11341661GA), Not Applicable (SEE11371758GA), Not Applicable (SEE11371793GA), Not Applicable (SEE12420159GA), Not Applicable (SEE12490250GA), Not Applicable (SEE14031630GA), Not Applicable (SEE16272270SA), Not Applicable (SEE16272272SA), 00840682109260SEE17490020SA (SEE17490020SA), 00840682109260SEE18290009SA (SEE18290009SA), Not Applicable (SEE16068832SA), Not Applicable (SEE09530904GA), Not Applicable (SEE09530838GA), Not Applicable (SEE09530850GA), Not Applicable (SEE09530908GA), Not Applicable (SEE10120007GA), Not Applicable (SEE10390296GA), Not Applicable (SEE10390312GA), Not Applicable (SEE10390313GA), Not Applicable (SEE13040418GA), Not Applicable (SEE16028368SA), Not Applicable (SEE16119908SA), Not Applicable (SEE16450015SA), Not Applicable (SEE16450016SA), 00840682109260SEE17270011SA (SEE17270011SA), Not Applicable (SEE15301890SA), Not Applicable (SEE16089268SA), Not Applicable (SEE16089277SA), Not Applicable (SEE16089279SA), Not Applicable (SEE09250318GA), Not Applicable (SEE11151316GA), Not Applicable (SEE11161319GA), Not Applicable (SEE11161321GA), Not Applicable (SEE11371762GA), Not Applicable (SEE12122130GA), Not Applicable (SEE12410134GA), Not Applicable (SEE12490259GA), Not Applicable (SEE12520354GA), Not Applicable (SEE13190804GA), Not Applicable (SEE13190806GA), Not Applicable (SEE14031679GA), Not Applicable (SEE14222042GA), Not Applicable (SEE09530839GA), Not Applicable (SEE09530878GA), Not Applicable (SEE09530882GA), Not Applicable (SEE09530906GA), Not Applicable (SEE10120999GA), Not Applicable (SEE10460738GA), Not Applicable (SEE11371760GA), Not Applicable (SEE11371761GA), Not Applicable (SEE14021584GA), Not Applicable (SEE14031662GA), Not Applicable (SEE14031668GA), Not Applicable (SEE14031673GA), 00840682109260SEE18410022SA (SEE18410022SA), Not Applicable (SEE14452634GA), Not Applicable (SEE14452636GA), Not Applicable (SEE14452639GA), Not Applicable (SEE14452641GA), Not Applicable (SEE16390003SA), Not Applicable (SEE16390010SA), Not Applicable (SEE15113118GA), Not Applicable (SEE15113140GA), Not Applicable (SEE09450623GA), Not Applicable (SEE09450627GA), Not Applicable (SEE09460632GA), Not Applicable (SEE09460633GA), Not Applicable (SEE09460634GA), Not Applicable (SEE09460635GA), Not Applicable (SEE09460637GA), Not Applicable (SEE15466921SA), Not Applicable (SEE09330421GA), Not Applicable (SEE10120020GA), Not Applicable (SEE10530829GA), Not Applicable (SEE15093067GA), Not Applicable (SEE15103096GA), Not Applicable (SEE15103100GA), Not Applicable (SEE15103102GA), Not Applicable (SEE15103104GA), Not Applicable (SEE15103108GA), Not Applicable (SEE15103110GA), Not Applicable (SEE10420559GA), Not Applicable (SEE10420600GA), Not Applicable (SEE10430608GA), Not Applicable (SEE10430612GA), Not Applicable (SEE10430631GA), Not Applicable (SEE10430632GA), Not Applicable (SEE10430648GA), Not Applicable (SEE10450674GA), Not Applicable (SEE10450688GA), Not Applicable (SEE11060990GA), Not Applicable (SEE11071046GA), Not Applicable (SEE11081170GA), Not Applicable (SEE11081194GA), Not Applicable (SEE11081201GA), Not Applicable (SEE11081209GA), Not Applicable (SEE11101220GA), Not Applicable (SEE11101221GA), Not Applicable (SEE12062011GA), Not Applicable (SEE13411218GA), Not Applicable (SEE13411220GA), Not Applicable (SEE13411221GA), Not Applicable (SEE13411222GA), 00840682109260SEE18500027SA (SEE18500027SA), 00840682109260SEE19120014SA (SEE19120014SA), Not Applicable (SEE12282358GA), Not Applicable (SEE13501481GA), Not Applicable (SEE13501484GA), Not Applicable (SEE13501488GA), Not Applicable (SEE13501491GA), Not Applicable (SEE13501492GA), Not Applicable (SEE13501499GA), Not Applicable (SEE13501500GA), Not Applicable (SEE14372419GA), Not Applicable (SEE15425780SA), Not Applicable (SEE15425797SA), Not Applicable (SEE16241824SA), Not Applicable (SEE16241842SA), Not Applicable (SEE16360023SA), Not Applicable (SEE16410003SA), Not Applicable (SEE09100098GA), Not Applicable (SEE12282353GA), Not Applicable (SEE09340452GA), Not Applicable (SEE15394277SA), Not Applicable (SEE14031657GA), Not Applicable (SEE14031658GA), Not Applicable (SEE13180735GA), Not Applicable (SEE11371807GA), Not Applicable (SEE12031927GA), Not Applicable (SEE12172196GA), Not Applicable (SEE13491430GA), Not Applicable (SEE13270874GA), Not Applicable (SEE09490750GA), Not Applicable (SEE09490759GA), Not Applicable (SEE12031941GA), Not Applicable (SEE11141257GA), Not Applicable (SEE09480741GA), Not Applicable (SEE09130170GA), Not Applicable (SEE09150207GA), Not Applicable (SEE09240299GA), Not Applicable (SEE09310367GA), Not Applicable (SEE09310389GA), Not Applicable (SEE09310390GA), Not Applicable (SEE09310391GA), Not Applicable (SEE09310392GA), Not Applicable (SEE09310393GA), Not Applicable (SEE09310394GA), Not Applicable (SEE09310396GA), Not Applicable (SEE09310397GA), Not Applicable (SEE09310400GA), Not Applicable (SEE09310401GA), Not Applicable (SEE09310402GA), Not Applicable (SEE09310403GA), Not Applicable (SEE09310404GA), Not Applicable (SEE09310405GA), Not Applicable (SEE09310406GA), Not Applicable (SEE10470778GA), Not Applicable (SEE10470798GA), Not Applicable (SEE13180765GA), Not Applicable (SEE13180774GA), Not Applicable (SEE14031705GA), 00840682109260SEE17270001SA (SEE17270001SA), 00840682109260SEE17270014SA (SEE17270014SA), 00840682109260SEE19010002SA (SEE19010002SA), 00840682109260SEE19010004SA (SEE19010004SA), 00840682109260SEE19010009SA (SEE19010009SA), 00840682109260SEE19090002SA (SEE19090002SA), 00840682109260SEE19090024SA (SEE19090024SA), Not Applicable (SEE14472680GA), Not Applicable (SEE15446228SA), Not Applicable (SEE15446229SA), Not Applicable (SEE15446231SA), Not Applicable (SEE15446232SA), Not Applicable (SEE15446233SA), Not Applicable (SEE15517773SA), Not Applicable (SEE11181406GA), Not Applicable (SEE11181408GA), Not Applicable (SEE11181409GA), Not Applicable (SEE11181411GA), Not Applicable (SEE11181416GA), Not Applicable (SEE11181418GA), Not Applicable (SEE11181419GA), Not Applicable (SEE11181421GA), Not Applicable (SEE16089173SA), 00840682109260SEE17300005SA (SEE17300005SA), 00840682109260SEE17300007SA (SEE17300007SA), 00840682109260SEE17300009SA (SEE17300009SA), 00840682109260SEE18190007SA (SEE18190007SA), 00840682109260SEE18190008SA (SEE18190008SA), 00840682109260SEE18190014SA (SEE18190014SA), 00840682109260SEE18190015SA (SEE18190015SA), 00840682109260SEE18190018SA (SEE18190018SA), 00840682109260SEE18190022SA (SEE18190022SA), 00840682109260SEE18190025SA (SEE18190025SA), 00840682109260SEE18190027SA (SEE18190027SA), 00840682109260SEE18190034SA (SEE18190034SA), 00840682109260SEE17100003SA (SEE17100003SA), 00840682109260SEE18080004SA (SEE18080004SA), 00840682109260SEE18100001SA (SEE18100001SA), 00840682109260SEE18100003SA (SEE18100003SA), 00840682109260SEE18100004SA (SEE18100004SA), 00840682109260SEE18100007SA (SEE18100007SA), 00840682109260SEE18100008SA (SEE18100008SA), 00840682109260SEE18100011SA (SEE18100011SA), 00840682109260SEE18100013SA (SEE18100013SA), 00840682109260SEE18100015SA (SEE18100015SA), 00840682109260SEE18100016SA (SEE18100016SA), 00840682109260SEE18100017SA (SEE18100017SA), 00840682109260SEE18100018SA (SEE18100018SA), 00840682109260SEE18100019SA (SEE18100019SA), Not Applicable (SEE11151309GA), Not Applicable (SEE13451313GA), Not Applicable (SEE14021524GA), Not Applicable (SEE14021527GA), Not Applicable (SEE14021528GA), Not Applicable (SEE14021529GA), Not Applicable (SEE14021533GA), Not Applicable (SEE14021534GA), Not Applicable (SEE14021536GA), Not Applicable (SEE14021539GA), Not Applicable (SEE14021540GA), Not Applicable (SEE14021541GA), Not Applicable (SEE14021543GA), Not Applicable (SEE14021546GA), Not Applicable (SEE14021547GA), Not Applicable (SEE14021548GA), Not Applicable (SEE14021550GA), Not Applicable (SEE14021557GA), Not Applicable (SEE14021560GA), Not Applicable (SEE14021565GA), Not Applicable (SEE14021567GA), Not Applicable (SEE14021572GA), Not Applicable (SEE14242089GA), Not Applicable (SEE16430043SA), Not Applicable (SEE16440002SA), Not Applicable (SEE12031932GA), Not Applicable (SEE12031936GA), Not Applicable (SEE12031938GA), Not Applicable (SEE12031940GA), Not Applicable (SEE14031722GA), Not Applicable (SEE15113170GA), Not Applicable (SEE15113181GA), 00840682109260SEE18370007SA (SEE18370007SA), 00840682109260SEE18410032SA (SEE18410032SA), 00840682109260SEE18410033SA (SEE18410033SA), 00840682109260SEE18420001SA (SEE18420001SA), 00840682109260SEE18420002SA (SEE18420002SA), 00840682109260SEE18420003SA (SEE18420003SA), 00840682109260SEE18420005SA (SEE18420005SA), 00840682109260SEE18420019SA (SEE18420019SA), 00840682109260SEE18420046SA (SEE18420046SA), Not Applicable (SEE09040047GA), Not Applicable (SEE09100115GA), Not Applicable (SEE09410582GA), Not Applicable (SEE10120003GA), Not Applicable (SEE10120994GA), Not Applicable (SEE10121001GA), Not Applicable (SEE10130071GA), Not Applicable (SEE14021505GA), Not Applicable (SEE14021513GA), Not Applicable (SEE14021522GA), Not Applicable (SEE14242094GA), Not Applicable (SEE14242107GA), Not Applicable (SEE14242117GA), Not Applicable (SEE14242118GA), Not Applicable (SEE14242119GA), 00840682109260SEE17480006SA (SEE17480006SA), Not Applicable (SEE10450680GA), Not Applicable (SEE15373462SA), Not Applicable (SEE09490756GA), Not Applicable (SEE09490760GA), Not Applicable (SEE09510768GA), Not Applicable (SEE09510782GA), Not Applicable (SEE12490275GA), 00840682109260SEE18160066SA (SEE18160066SA), Not Applicable (SEE12430184GA), Not Applicable (SEE10460731GA), Not Applicable (SEE16119898SA), Not Applicable (SEE10400427GA), 00840682109260SEE11381827GA (SEE11381827GA), Not Applicable (SEE15153187GA), Not Applicable (SEE13431264GA), Not Applicable (SEE11071040GA), 00840682109260SEE15251298SA (SEE15251298SA), 00840682109260SEE15251300SA (SEE15251300SA), 00840682109260SEE15251302SA (SEE15251302SA), 00840682109260SEE15261409SA (SEE15261409SA), 00840682109260SEE15261532SA (SEE15261532SA), 00840682109260SEE15261533SA (SEE15261533SA), Not Applicable (SEE15261534SA), 00840682109260SEE15261540SA (SEE15261540SA), 00840682109260SEE15261541SA (SEE15261541SA), 00840682109260SEE15261542SA (SEE15261542SA), 00840682109260SEE15261545SA (SEE15261545SA), 00840682109260SEE15261546SA (SEE15261546SA), Not Applicable (SEE12041946GA), Not Applicable (SEE14412494GA), Not Applicable (SEE14412500GA), Not Applicable (SEE14412508GA), Not Applicable (SEE14412511GA), Not Applicable (SEE11171377GA), Not Applicable (SEE10530860GA), Not Applicable (SEE10530873GA), Not Applicable (SEE11030882GA), Not Applicable (SEE11030891GA), Not Applicable (SEE12322417GA), Not Applicable (SEE12322424GA), Not Applicable (SEE12322439GA), Not Applicable (SEE10460709GA), Not Applicable (SEE12490246GA), Not Applicable (SEE12182217GA), Not Applicable (SEE14512828GA), Not Applicable (SEE15518043SA), 00840682109260SEE18360001SA (SEE18360001SA), 00840682109260SEE18360009SA (SEE18360009SA), 00840682109260SEE18360011SA (SEE18360011SA), 00840682109260SEE18360013SA (SEE18360013SA), 00840682109260SEE18360014SA (SEE18360014SA), 00840682109260SEE18360017SA (SEE18360017SA), Not Applicable (SEE11030884GA), Not Applicable (SEE11030885GA), Not Applicable (SEE11030889GA), Not Applicable (SEE11030892GA), Not Applicable (SEE11030893GA), Not Applicable (SEE11030894GA), Not Applicable (SEE11030895GA), Not Applicable (SEE12122148GA), Not Applicable (SEE12122150GA), Not Applicable (SEE13381122GA), Not Applicable (SEE13381125GA), Not Applicable (SEE13391134GA), Not Applicable (SEE09460651GA), Not Applicable (SEE12172194GA), Not Applicable (SEE09470656GA), Not Applicable (SEE09470680GA), Not Applicable (SEE09470681GA), Not Applicable (SEE09470682GA), Not Applicable (SEE09470684GA), Not Applicable (SEE09470688GA), Not Applicable (SEE09470689GA), Not Applicable (SEE09470690GA), Not Applicable (SEE09480692GA), Not Applicable (SEE09480693GA), Not Applicable (SEE09480694GA), Not Applicable (SEE09480695GA), Not Applicable (SEE09480721GA), Not Applicable (SEE09480722GA), Not Applicable (SEE09480724GA), Not Applicable (SEE09480725GA), Not Applicable (SEE09480726GA), Not Applicable (SEE09480727GA), Not Applicable (SEE09480731GA), Not Applicable (SEE15332399SA), Not Applicable (SEE15332400SA), 00840682109260SEE17190012SA (SEE17190012SA), 00840682109260SEE17190017SA (SEE17190017SA), 00840682109260SEE18330004SA (SEE18330004SA), 00840682109260SEE18470008SA (SEE18470008SA), 00840682109260SEE18470009SA (SEE18470009SA), 00840682109260SEE18470015SA (SEE18470015SA), 00840682109260SEE18470022SA (SEE18470022SA), Not Applicable (SEE10390318GA), Not Applicable (SEE11211503GA), Not Applicable (SEE11361684GA), Not Applicable (SEE11361690GA), Not Applicable (SEE12262341GA), Not Applicable (SEE13311006GA), Not Applicable (SEE14272162GA), Not Applicable (SEE14272171GA), Not Applicable (SEE14272208GA), Not Applicable (SEE14272209GA), Not Applicable (SEE14272211GA), Not Applicable (SEE14272215GA), Not Applicable (SEE14272216GA), Not Applicable (SEE14272220GA), Not Applicable (SEE14272221GA), Not Applicable (SEE14272222GA), Not Applicable (SEE14272223GA), Not Applicable (SEE14272225GA), Not Applicable (SEE14272226GA), Not Applicable (SEE14272227GA), Not Applicable (SEE14272228GA), Not Applicable (SEE14272229GA), Not Applicable (SEE14472679GA), Not Applicable (SEE15383769SA), 00840682109260SEE19010007SA (SEE19010007SA), 00840682109260SEE19220039SA (SEE19220039SA), Not Applicable (SEE13431270GA), Not Applicable (SEE13441307GA), Not Applicable (SEE14222014GA), Not Applicable (SEE14222021GA), Not Applicable (SEE14222022GA), Not Applicable (SEE14222023GA), Not Applicable (SEE14222024GA), Not Applicable (SEE14222025GA), Not Applicable (SEE14222058GA), Not Applicable (SEE14222066GA), Not Applicable (SEE14222069GA), Not Applicable (SEE14222071GA), 00840682109260SEE17090020SA (SEE17090020SA), Not Applicable (SEE09020020GA), Not Applicable (SEE11341663GA), 00840682109260SEE19100010SA (SEE19100010SA), 00840682109260SEE19100027SA (SEE19100027SA), 00840682109260SEE19100058SA (SEE19100058SA), 00840682109260SEE19190008SA (SEE19190008SA), 00840682109260SEE18420042SA (SEE18420042SA), 00840682109260SEE18430002SA (SEE18430002SA), 00840682109260SEE18430003SA (SEE18430003SA), 00840682109260SEE18430006SA (SEE18430006SA), 00840682109260SEE18430007SA (SEE18430007SA), 00840682109260SEE18430008SA (SEE18430008SA), 00840682109260SEE18430009SA (SEE18430009SA), 00840682109260SEE18450076SA (SEE18450076SA), 00840682109260SEE18400007SA (SEE18400007SA), 00840682109260SEE18420009SA (SEE18420009SA), 00840682109260SEE18420011SA (SEE18420011SA), 00840682109260SEE18420022SA (SEE18420022SA), 00840682109260SEE18420023SA (SEE18420023SA), 00840682109260SEE18420027SA (SEE18420027SA), 00840682109260SEE18420028SA (SEE18420028SA), 00840682109260SEE18420030SA (SEE18420030SA), 00840682109260SEE18420034SA (SEE18420034SA), 00840682109260SEE18420035SA (SEE18420035SA), 00840682109260SEE18420037SA (SEE18420037SA), 00840682109260SEE18420038SA (SEE18420038SA), 00840682109260SEE18420039SA (SEE18420039SA), 00840682109260SEE18420043SA (SEE18420043SA), 00840682109260SEE18420048SA (SEE18420048SA), 00840682109260SEE18420049SA (SEE18420049SA), 00840682109260SEE18420051SA (SEE18420051SA), 00840682109260SEE18420052SA (SEE18420052SA), Not Applicable (SEE13280949GA), Not Applicable (SEE13351044GA), Not Applicable (SEE13351051GA), Not Applicable (SEE13361055GA), Not Applicable (SEE13361056GA), Not Applicable (SEE13361059GA), Not Applicable (SEE13361060GA), Not Applicable (SEE13361063GA), Not Applicable (SEE13361064GA), Not Applicable (SEE13361068GA), Not Applicable (SEE13361071GA), Not Applicable (SEE13361072GA), 00840682109260SEE17110004SA (SEE17110004SA), 00840682109260SEE17180001SA (SEE17180001SA), Not Applicable (SEE12122135GA), Not Applicable (SEE12122141GA), Not Applicable (SEE12122159GA), Not Applicable (SEE12122161GA), Not Applicable (SEE12122166GA), 00840682109260SEE17250023SA (SEE17250023SA), 00840682109260SEE17250025SA (SEE17250025SA), Not Applicable (SEE15456712SA), Not Applicable (SEE15466922SA), Not Applicable (SEE15466927SA), 00840682109260SEE17210017SA (SEE17210017SA), Not Applicable (SEE09130159GA), Not Applicable (SEE09260338GA), Not Applicable (SEE09260339GA), Not Applicable (SEE10400424GA), Not Applicable (SEE10400433GA), Not Applicable (SEE10400452GA), Not Applicable (SEE10400456GA), Not Applicable (SEE10460773GA), Not Applicable (SEE10470793GA), Not Applicable (SEE14502795GA), Not Applicable (SEE10390307GA), Not Applicable (SEE10390317GA), Not Applicable (SEE10390320GA), Not Applicable (SEE10390330GA), Not Applicable (SEE10390331GA), Not Applicable (SEE10390342GA), Not Applicable (SEE16180907SA), Not Applicable (SEE16180908SA), Not Applicable (SEE16180911SA), Not Applicable (SEE16180912SA), Not Applicable (SEE16180913SA), Not Applicable (SEE14031586GA), Not Applicable (SEE14031590GA), Not Applicable (SEE14292231GA), Not Applicable (SEE14292232GA), Not Applicable (SEE14292233GA), Not Applicable (SEE14292239GA), Not Applicable (SEE14292240GA), Not Applicable (SEE14292244GA), Not Applicable (SEE14302257GA), Not Applicable (SEE14302258GA), Not Applicable (SEE15052928GA), Not Applicable (SEE15052930GA), Not Applicable (SEE15052938GA), Not Applicable (SEE15052948GA), Not Applicable (SEE15052949GA), Not Applicable (SEE15052951GA), Not Applicable (SEE15052952GA), Not Applicable (SEE15062963GA), Not Applicable (SEE15062966GA), Not Applicable (SEE15062967GA), Not Applicable (SEE15062968GA), Not Applicable (SEE15062969GA), Not Applicable (SEE13110500GA), 00840682109260SEE18170023SA (SEE18170023SA), 00840682109260SEE19220023SA (SEE19220023SA), Not Applicable (SEE10270260GA), Not Applicable (SEE09350458GA), Not Applicable (SEE10250089GA), Not Applicable (SEE10260202GA), Not Applicable (SEE10260203GA), Not Applicable (SEE10260214GA), Not Applicable (SEE10260215GA), Not Applicable (SEE10260220GA), Not Applicable (SEE10260223GA), Not Applicable (SEE10260225GA), Not Applicable (SEE10260226GA), Not Applicable (SEE10260227GA), Not Applicable (SEE10260228GA), Not Applicable (SEE10260231GA), Not Applicable (SEE16282654SA), Not Applicable (SEE16344281SA), Not Applicable (SEE16344328SA), Not Applicable (SEE16354344SA), Not Applicable (SEE16364411SA), Not Applicable (SEE16364412SA), Not Applicable (SEE16364413SA), Not Applicable (SEE16364414SA), Not Applicable (SEE16364415SA), Not Applicable (SEE16420064SA), Not Applicable (SEE11341665GA), Not Applicable (SEE13401175GA), Not Applicable (SEE16221589SA), Not Applicable (SEE15467028SA), Not Applicable (SEE09020015GA), Not Applicable (SEE09040046GA), Not Applicable (SEE16400001SA), Not Applicable (SEE16410005SA), Not Applicable (SEE16410006SA), Not Applicable (SEE16410012SA), Not Applicable (SEE15240989SA), 00840682109260SEE18160045SA (SEE18160045SA), Not Applicable (SEE14302267GA), Not Applicable (SEE14302276GA), 00840682109260SEE19170026SA (SEE19170026SA), Not Applicable (SEE16120020SA), Not Applicable (SEE12350026GA), Not Applicable (SEE10250124GA), Not Applicable (SEE16241829SA), Not Applicable (SEE11231529GA), Not Applicable (SEE14222033GA), Not Applicable (SEE14222038GA), Not Applicable (SEE14222048GA), Not Applicable (SEE14222052GA), Not Applicable (SEE14222053GA), Not Applicable (SEE14222070GA), 00840682109260SEE18250012SA (SEE18250012SA), 00840682109260SEE18250013SA (SEE18250013SA), 00840682109260SEE18250014SA (SEE18250014SA), Not Applicable (SEE09340432GA), Not Applicable (SEE09340438GA), Not Applicable (SEE09340442GA), Not Applicable (SEE09340443GA), Not Applicable (SEE14312294GA), Not Applicable (SEE15093063GA), Not Applicable (SEE15093064GA), Not Applicable (SEE15093071GA), Not Applicable (SEE15093072GA), Not Applicable (SEE12092068GA), Not Applicable (SEE12092069GA), Not Applicable (SEE12092073GA), Not Applicable (SEE12092075GA), Not Applicable (SEE12092081GA), Not Applicable (SEE12092084GA), Not Applicable (SEE10270275GA), Not Applicable (SEE11040907GA), Not Applicable (SEE14322321GA), Not Applicable (SEE14322323GA), Not Applicable (SEE14322329GA), Not Applicable (SEE14332344GA), Not Applicable (SEE14362348GA), Not Applicable (SEE14362349GA), Not Applicable (SEE14362354GA), Not Applicable (SEE14362355GA), Not Applicable (SEE14362356GA), Not Applicable (SEE14462654GA), Not Applicable (SEE14462655GA), Not Applicable (SEE15193276GA), Not Applicable (SEE16380022SA), Not Applicable (SEE14332338GA), Not Applicable (SEE14332340GA), Not Applicable (SEE15183257GA), 00840682109260SEE18160063SA (SEE18160063SA), Not Applicable (SEE13361057GA), Not Applicable (SEE13361061GA), Not Applicable (SEE12262344GA), Not Applicable (SEE09100123GA), Not Applicable (SEE09250333GA), Not Applicable (SEE10260229GA), Not Applicable (SEE10270249GA), Not Applicable (SEE10270252GA), Not Applicable (SEE10310278GA), Not Applicable (SEE10310284GA), Not Applicable (SEE13361080GA), Not Applicable (SEE14482719GA), Not Applicable (SEE14512835GA), Not Applicable (SEE14512836GA), Not Applicable (SEE15120556SA), Not Applicable (SEE15120558SA), Not Applicable (SEE15210699SA), Not Applicable (SEE15210702SA), Not Applicable (SEE15210705SA), Not Applicable (SEE09330422GA), Not Applicable (SEE11201469GA), Not Applicable (SEE13010403GA), Not Applicable (SEE14021520GA), Not Applicable (SEE14021523GA), Not Applicable (SEE14021525GA), Not Applicable (SEE14021526GA), Not Applicable (SEE14021532GA), Not Applicable (SEE14021535GA), Not Applicable (SEE14021552GA), Not Applicable (SEE14021577GA), Not Applicable (SEE14021582GA), Not Applicable (SEE14031690GA), Not Applicable (SEE15113125GA), Not Applicable (SEE15113129GA), Not Applicable (SEE15113135GA), Not Applicable (SEE16078864SA), Not Applicable (SEE16078865SA), Not Applicable (SEE12122142GA), Not Applicable (SEE12122167GA), Not Applicable (SEE14362384GA), Not Applicable (SEE11301560GA), Not Applicable (SEE11371790GA), Not Applicable (SEE09250306GA), Not Applicable (SEE14452633GA), Not Applicable (SEE15052914GA), Not Applicable (SEE14372389GA), Not Applicable (SEE11071011GA), Not Applicable (SEE11071047GA), Not Applicable (SEE13130616GA), Not Applicable (SEE13130623GA), Not Applicable (SEE13130630GA), Not Applicable (SEE10260219GA), 00840682109260SEE19100039SA (SEE19100039SA), 00840682109260SEE19100040SA (SEE19100040SA), 00840682109260SEE17480004SA (SEE17480004SA), 00840682109260SEE17480005SA (SEE17480005SA), 00840682109260SEE17480007SA (SEE17480007SA), 00840682109260SEE17480009SA (SEE17480009SA), 00840682109260SEE17480013SA (SEE17480013SA), Not Applicable (SEE16272597SA), Not Applicable (SEE09370495GA), Not Applicable (SEE09390535GA), Not Applicable (SEE09460641GA), Not Applicable (SEE11111248GA), Not Applicable (SEE13130649GA), Not Applicable (SEE15193286GA), Not Applicable (SEE15193287GA), Not Applicable (SEE15193288GA), Not Applicable (SEE15193289GA), Not Applicable (SEE15193290GA), Not Applicable (SEE15193291GA), Not Applicable (SEE15193292GA), Not Applicable (SEE15193295GA), Not Applicable (SEE15193305GA), Not Applicable (SEE15394426SA), Not Applicable (SEE15518046SA), Not Applicable (SEE15518047SA), Not Applicable (SEE15528072SA), Not Applicable (SEE15528204SA), Not Applicable (SEE15528206SA), Not Applicable (SEE15528207SA), Not Applicable (SEE11341636GA), 00840682109260SEE19220007SA (SEE19220007SA), Not Applicable (SEE10100969GA), Not Applicable (SEE10100979GA), Not Applicable (SEE10120988GA), Not Applicable (SEE13130636GA), Not Applicable (SEE10270253GA), Not Applicable (SEE13150694GA), Not Applicable (SEE13150697GA), Not Applicable (SEE13220822GA), Not Applicable (SEE13220837GA), Not Applicable (SEE14041737GA), 00840682109260SEE18240059SA (SEE18240059SA), Not Applicable (SEE09390539GA), Not Applicable (SEE09390541GA), Not Applicable (SEE09390543GA), Not Applicable (SEE09390545GA), Not Applicable (SEE09400551GA), 00840682109260SEE17320001SA (SEE17320001SA), 00840682109260SEE17320002SA (SEE17320002SA), 00840682109260SEE17320003SA (SEE17320003SA), 00840682109260SEE17320004SA (SEE17320004SA), 00840682109260SEE17320005SA (SEE17320005SA), 00840682109260SEE17320006SA (SEE17320006SA), Not Applicable (SEE14041740GA), Not Applicable (SEE09410583GA), 00840682109260SEE17090026SA (SEE17090026SA), 00840682109260SEE17090027SA (SEE17090027SA), 00840682109260SEE17360007SA (SEE17360007SA), 00840682109260SEE17460010SA (SEE17460010SA), 00840682109260SEE18160065SA (SEE18160065SA), 00840682109260SEE19240028SA (SEE19240028SA), 00840682109260SEE18190029SA (SEE18190029SA), 00840682109260SEE17020022SA (SEE17020022SA), 00840682109260SEE17400005SA (SEE17400005SA), Not Applicable (SEE14242087GA), Not Applicable (SEE09510828GA), Not Applicable (SEE11071152GA), Not Applicable (SEE11081157GA), Not Applicable (SEE11081164GA), Not Applicable (SEE16430034SA), Not Applicable (SEE09150218GA), Not Applicable (SEE10450667GA), Not Applicable (SEE10460711GA), Not Applicable (SEE11071029GA), Not Applicable (SEE12092083GA), Not Applicable (SEE09390536GA), Not Applicable (SEE10250121GA), Not Applicable (SEE10250139GA), Not Applicable (SEE10250169GA), Not Applicable (SEE10250172GA), Not Applicable (SEE15415177SA), Not Applicable (SEE15425781SA), Not Applicable (SEE15425791SA), 00840682109260SEE17240098SA (SEE17240098SA), 00840682109260SEE17240120SA (SEE17240120SA), 00840682109260SEE17240122SA (SEE17240122SA), 00840682109260SEE17500015SA (SEE17500015SA), 00840682109260SEE17500017SA (SEE17500017SA), 00840682109260SEE17500018SA (SEE17500018SA), 00840682109260SEE18100038SA (SEE18100038SA), 00840682109260SEE18100039SA (SEE18100039SA), 00840682109260SEE18170009SA (SEE18170009SA), 00840682109260SEE18170013SA (SEE18170013SA), 00840682109260SEE19170001SA (SEE19170001SA), 00840682109260SEE19170003SA (SEE19170003SA), 00840682109260SEE19170010SA (SEE19170010SA), 00840682109260SEE19170016SA (SEE19170016SA), 00840682109260SEE19170021SA (SEE19170021SA), 00840682109260SEE19170023SA (SEE19170023SA), Not Applicable (SEE11181399GA), Not Applicable (SEE11181414GA), Not Applicable (SEE12340018GA), 00840682109260SEE18270001SA (SEE18270001SA), Not Applicable (SEE09360490GA), Not Applicable (SEE12410133GA), 00840682109260SEE18260002SA (SEE18260002SA), 00840682109260SEE18260003SA (SEE18260003SA), 00840682109260SEE18260004SA (SEE18260004SA), 00840682109260SEE18260005SA (SEE18260005SA), 00840682109260SEE18260006SA (SEE18260006SA), 00840682109260SEE18260008SA (SEE18260008SA), 00840682109260SEE18260009SA (SEE18260009SA), 00840682109260SEE18260010SA (SEE18260010SA), 00840682109260SEE18260011SA (SEE18260011SA), 00840682109260SEE18260012SA (SEE18260012SA), Not Applicable (SEE11081198GA), 00840682109260SEE17130011SA (SEE17130011SA), 00840682109260SEE17130013SA (SEE17130013SA), 00840682109260SEE17210015SA (SEE17210015SA), 00840682109260SEE17210019SA (SEE17210019SA), 00840682109260SEE17260001SA (SEE17260001SA), 00840682109260SEE17380016SA (SEE17380016SA), 00840682109260SEE18100027SA (SEE18100027SA), 00840682109260SEE18100028SA (SEE18100028SA), 00840682109260SEE18280007SA (SEE18280007SA), 00840682109260SEE18290003SA (SEE18290003SA), 00840682109260SEE18350001SA (SEE18350001SA), 00840682109260SEE18350003SA (SEE18350003SA), 00840682109260SEE17180013SA (SEE17180013SA), 00840682109260SEE17180015SA (SEE17180015SA), 00840682109260SEE17180017SA (SEE17180017SA), 00840682109260SEE17200002SA (SEE17200002SA), 00840682109260SEE17410003SA (SEE17410003SA), 00840682109260SEE17460001SA (SEE17460001SA), 00840682109260SEE17460012SA (SEE17460012SA), Not Applicable (SEE09310365GA), Not Applicable (SEE09310374GA), Not Applicable (SEE09310381GA), Not Applicable (SEE09340448GA), Not Applicable (SEE09340454GA), Not Applicable (SEE09360472GA), Not Applicable (SEE09360487GA), Not Applicable (SEE09360489GA), Not Applicable (SEE09410581GA), Not Applicable (SEE09410590GA), Not Applicable (SEE09430608GA), Not Applicable (SEE09430615GA), Not Applicable (SEE09430617GA), Not Applicable (SEE09430619GA), Not Applicable (SEE09510818GA), Not Applicable (SEE09530843GA), Not Applicable (SEE13421227GA), Not Applicable (SEE13421228GA), Not Applicable (SEE13421231GA), Not Applicable (SEE13421233GA), Not Applicable (SEE13421234GA), Not Applicable (SEE13421237GA), Not Applicable (SEE13421242GA), Not Applicable (SEE13421244GA), Not Applicable (SEE12410131GA), Not Applicable (SEE09020032GA), Not Applicable (SEE09040048GA), Not Applicable (SEE09040050GA), Not Applicable (SEE09420595GA), Not Applicable (SEE10390310GA), Not Applicable (SEE15383635SA), Not Applicable (SEE15383636SA), Not Applicable (SEE15383637SA), Not Applicable (SEE15517822SA), 00840682109260SEE18030012SA (SEE18030012SA), 00840682109260SEE18040014SA (SEE18040014SA), 00840682109260SEE18040025SA (SEE18040025SA), 00840682109260SEE18500009SA (SEE18500009SA), 00840682109260SEE18510008SA (SEE18510008SA), 00840682109260SEE18510011SA (SEE18510011SA), 00840682109260SEE17440004SA (SEE17440004SA), 00840682109260SEE17440006SA (SEE17440006SA), 00840682109260SEE17440008SA (SEE17440008SA), Not Applicable (SEE10120028GA), Not Applicable (SEE10120033GA), Not Applicable (SEE10130036GA), Not Applicable (SEE10400384GA), Not Applicable (SEE10400437GA), Not Applicable (SEE10400453GA), Not Applicable (SEE10460725GA), Not Applicable (SEE12410146GA), Not Applicable (SEE15113157GA), Not Applicable (SEE15467027SA), Not Applicable (SEE15517777SA), 00840682109260SEE17020024SA (SEE17020024SA), 00840682109260SEE17020027SA (SEE17020027SA), Not Applicable (SEE12112111GA), Not Applicable (SEE14272160GA), Not Applicable (SEE15193303GA), 00840682109260SEE17460003SA (SEE17460003SA), 00840682109260SEE18100009SA (SEE18100009SA), 00840682109260SEE18100010SA (SEE18100010SA), 00840682109260SEE19130018SA (SEE19130018SA), 00840682109260SEE19150006SA (SEE19150006SA), Not Applicable (SEE11101215GA), 00840682109260SEE18060017SA (SEE18060017SA), Not Applicable (SEE10530858GA), Not Applicable (SEE10530868GA), Not Applicable (SEE14021551GA), Not Applicable (SEE12122147GA), Not Applicable (SEE12122154GA), Not Applicable (SEE12122156GA), Not Applicable (SEE15012858GA), 00840682109260SEE17250022SA (SEE17250022SA), 00840682109260SEE17250024SA (SEE17250024SA), 00840682109260SEE17390009SA (SEE17390009SA), Not Applicable (SEE11381819GA), Not Applicable (SEE12360032GA), Not Applicable (SEE09100102GA), Not Applicable (SEE13110508GA), Not Applicable (SEE15052916GA), Not Applicable (SEE16130324SA), 00840682109260SEE17480016SA (SEE17480016SA), 00840682109260SEE19100022SA (SEE19100022SA), Not Applicable (SEE12312400GA), Not Applicable (SEE09190231GA), Not Applicable (SEE09190241GA), Not Applicable (SEE09510767GA), Not Applicable (SEE09510770GA), Not Applicable (SEE09510780GA), Not Applicable (SEE11201449GA), Not Applicable (SEE12232252GA), Not Applicable (SEE12232254GA), Not Applicable (SEE12232255GA), Not Applicable (SEE12232256GA), Not Applicable (SEE12232262GA), Not Applicable (SEE12242275GA), Not Applicable (SEE12242277GA), Not Applicable (SEE12242278GA).Not Applicable (SEE12242280GA), Not Applicable (SEE12242283GA), Not Applicable (SEE12242285GA), Not Applicable (SEE12252296GA), Not Applicable (SEE12252298GA), Not Applicable (SEE12252299GA), Not Applicable (SEE12252300GA), Not Applicable (SEE12252302GA), Not Applicable (SEE12252303GA), Not Applicable (SEE12252305GA), Not Applicable (SEE12252308GA), Not Applicable (SEE12252309GA), Not Applicable (SEE12252310GA), Not Applicable (SEE12252312GA), Not Applicable (SEE12252313GA), Not Applicable (SEE12252314GA), Not Applicable (SEE12252315GA), Not Applicable (SEE12322418GA), Not Applicable (SEE13491447GA), Not Applicable (SEE13491448GA), Not Applicable (SEE15113124GA), Not Applicable (SEE15113150GA), Not Applicable (SEE16420061SA), Not Applicable (SEE16420063SA), Not Applicable (SEE16420066SA), Not Applicable (SEE16420067SA), Not Applicable (SEE16420068SA), Not Applicable (SEE16420070SA), Not Applicable (SEE16420073SA), Not Applicable (SEE16420076SA), Not Applicable (SEE16420078SA), Not Applicable (SEE16430042SA), 00840682109260SEE17150009SA (SEE17150009SA), 00840682109260SEE18240013SA (SEE18240013SA), 00840682109260SEE18240034SA (SEE18240034SA), 00840682109260SEE18240035SA (SEE18240035SA), 00840682109260SEE18240039SA (SEE18240039SA), 00840682109260SEE18240040SA (SEE18240040SA), 00840682109260SEE18240041SA (SEE18240041SA), 00840682109260SEE18240043SA (SEE18240043SA), 00840682109260SEE18240046SA (SEE18240046SA), 00840682109260SEE18240049SA (SEE18240049SA), 00840682109260SEE18240051SA (SEE18240051SA), Not Applicable (SEE14432594GA), Not Applicable (SEE13280935GA), Not Applicable (SEE13280943GA), Not Applicable (SEE13290954GA), Not Applicable (SEE13290956GA), Not Applicable (SEE13290957GA), Not Applicable (SEE13290959GA), Not Applicable (SEE13290974GA), Not Applicable (SEE15261400SA), Not Applicable (SEE12490270GA), Not Applicable (SEE12490279GA), Not Applicable (SEE16252014SA), Not Applicable (SEE16282652SA), Not Applicable (SEE16390002SA), Not Applicable (SEE16390012SA), Not Applicable (SEE14452640GA), Not Applicable (SEE14452642GA), Not Applicable (SEE12092051GA), Not Applicable (SEE13040426GA), Not Applicable (SEE16028376SA), 00840682109260SEE17330013SA (SEE17330013SA), 00840682109260SEE17070005SA (SEE17070005SA), 00840682109260SEE17070008SA (SEE17070008SA), 00840682109260SEE17070009SA (SEE17070009SA), 00840682109260SEE17070011SA (SEE17070011SA), Not Applicable (SEE10400372GA), Not Applicable (SEE10410467GA), Not Applicable (SEE10430650GA), Not Applicable (SEE10450663GA), Not Applicable (SEE11141287GA), Not Applicable (SEE11371809GA), Not Applicable (SEE12122165GA), Not Applicable (SEE13371092GA), Not Applicable (SEE14021566GA), Not Applicable (SEE14372411GA), Not Applicable (SEE14382466GA), Not Applicable (SEE14382472GA), Not Applicable (SEE15173211GA), Not Applicable (SEE15291869SA), Not Applicable (SEE15342483SA), Not Applicable (SEE16380021SA), 00840682109260SEE17190018SA (SEE17190018SA), 00840682109260SEE17460008SA (SEE17460008SA), 00840682109260SEE17460009SA (SEE17460009SA), 00840682109260SEE19170027SA (SEE19170027SA), Not Applicable (SEE10090955GA), Not Applicable (SEE10460759GA), Not Applicable (SEE10090946GA), Not Applicable (SEE15032892GA), Not Applicable (SEE11361707GA), Not Applicable (SEE13381129GA), Not Applicable (SEE13391131GA), Not Applicable (SEE14462662GA), Not Applicable (SEE14462663GA), Not Applicable (SEE16364408SA), Not Applicable (SEE16364409SA), Not Applicable (SEE10080931GA), Not Applicable (SEE10090951GA), Not Applicable (SEE16252022SA), Not Applicable (SEE16262133SA), Not Applicable (SEE16262136SA), Not Applicable (SEE16262147SA), Not Applicable (SEE16262148SA), Not Applicable (SEE16262150SA), Not Applicable (SEE16262163SA), Not Applicable (SEE16262165SA), Not Applicable (SEE16262180SA), Not Applicable (SEE16262204SA), Not Applicable (SEE16272331SA), Not Applicable (SEE16272419SA), Not Applicable (SEE16272585SA), Not Applicable (SEE16272586SA), Not Applicable (SEE16272587SA), Not Applicable (SEE16272599SA), Not Applicable (SEE16272600SA), Not Applicable (SEE16272618SA), Not Applicable (SEE16272620SA), 00840682109260SEE17190005SA (SEE17190005SA), 00840682109260SEE17190006SA (SEE17190006SA), 00840682109260SEE17190007SA (SEE17190007SA), 00840682109260SEE17190013SA (SEE17190013SA), 00840682109260SEE17190014SA (SEE17190014SA), 00840682109260SEE17190015SA (SEE17190015SA), 00840682109260SEE17190016SA (SEE17190016SA), Not Applicable (SEE15113154GA), Not Applicable (SEE14432596GA), Not Applicable (SEE14432597GA), 00840682109260SEE19120012SA (SEE19120012SA), Not Applicable (SEE13110528GA), Not Applicable (SEE13110550GA), Not Applicable (SEE13110554GA), Not Applicable (SEE13110556GA), Not Applicable (SEE13110561GA), Not Applicable (SEE13110563GA), 00840682109260SEE19100025SA (SEE19100025SA), 00840682109260SEE19100065SA (SEE19100065SA), Not Applicable (SEE13431274GA), Not Applicable (SEE13451312GA), Not Applicable (SEE13451320GA), Not Applicable (SEE13451331GA), Not Applicable (SEE11391837GA), Not Applicable (SEE12021909GA), Not Applicable (SEE12031926GA), Not Applicable (SEE12031930GA), Not Applicable (SEE12031935GA), Not Applicable (SEE12061997GA), Not Applicable (SEE13491424GA), Not Applicable (SEE14021553GA), Not Applicable (SEE14021564GA), 00840682109260SEE17220006SA (SEE17220006SA), 00840682109260SEE17380001SA (SEE17380001SA), 00840682109260SEE17380002SA (SEE17380002SA), 00840682109260SEE17380013SA (SEE17380013SA), Not Applicable (SEE09470668GA), Not Applicable (SEE09470672GA), Not Applicable (SEE10260195GA), Not Applicable (SEE11071032GA), Not Applicable (SEE11071060GA), Not Applicable (SEE11321562GA), Not Applicable (SEE11321563GA), Not Applicable (SEE11321568GA), Not Applicable (SEE11321569GA), Not Applicable (SEE11321571GA), Not Applicable (SEE11321572GA), Not Applicable (SEE11321584GA), Not Applicable (SEE11321585GA), Not Applicable (SEE11331601GA), Not Applicable (SEE13280907GA), Not Applicable (SEE13280908GA), Not Applicable (SEE13280909GA), Not Applicable (SEE13280910GA), Not Applicable (SEE13280911GA), Not Applicable (SEE13280914GA), Not Applicable (SEE13280915GA), Not Applicable (SEE13280916GA), Not Applicable (SEE13280920GA), Not Applicable (SEE13501460GA), Not Applicable (SEE13501463GA), Not Applicable (SEE13501475GA), Not Applicable (SEE13501479GA), Not Applicable (SEE16150531SA), Not Applicable (SEE16180899SA), Not Applicable (SEE16324003SA), Not Applicable (SEE16324005SA), Not Applicable (SEE10420548GA), Not Applicable (SEE10450657GA), Not Applicable (SEE10460710GA), Not Applicable (SEE11151304GA), 00840682109260SEE17490036SA (SEE17490036SA), Not Applicable (SEE11181395GA), Not Applicable (SEE11191435GA), Not Applicable (SEE14211982GA), Not Applicable (SEE14211986GA), Not Applicable (SEE15507710SA), Not Applicable (SEE15507711SA), Not Applicable (SEE16079148SA), 00840682109260SEE18070004SA (SEE18070004SA), 00840682109260SEE18080008SA (SEE18080008SA), 00840682109260SEE18080011SA (SEE18080011SA), 00840682109260SEE18080014SA (SEE18080014SA), 00840682109260SEE18510004SA (SEE18510004SA), Not Applicable (SEE13000385GA), Not Applicable (SEE12430191GA), Not Applicable (SEE12430192GA), Not Applicable (SEE12430221GA), Not Applicable (SEE12430222GA), Not Applicable (SEE12430226GA), Not Applicable (SEE12430228GA), Not Applicable (SEE12440236GA), Not Applicable (SEE12440240GA), 00840682109260SEE19110002SA (SEE19110002SA), 00840682109260SEE19110005SA (SEE19110005SA), 00840682109260SEE19110008SA (SEE19110008SA), 00840682109260SEE19110009SA (SEE19110009SA), 00840682109260SEE19110011SA (SEE19110011SA), 00840682109260SEE19110016SA (SEE19110016SA), 00840682109260SEE19110017SA (SEE19110017SA), 00840682109260SEE19110018SA (SEE19110018SA), 00840682109260SEE19110027SA (SEE19110027SA), 00840682109260SEE19110030SA (SEE19110030SA), 00840682109260SEE19110031SA (SEE19110031SA), 00840682109260SEE19110032SA (SEE19110032SA), 00840682109260SEE19110033SA (SEE19110033SA), 00840682109260SEE19110034SA (SEE19110034SA), 00840682109260SEE19110035SA (SEE19110035SA), 00840682109260SEE19110036SA (SEE19110036SA), 00840682109260SEE19110038SA (SEE19110038SA), 00840682109260SEE19110042SA (SEE19110042SA), 00840682109260SEE19120007SA (SEE19120007SA), 00840682109260SEE19120008SA (SEE19120008SA), 00840682109260SEE19120009SA (SEE19120009SA), 00840682109260SEE19120010SA (SEE19120010SA), Not Applicable (SEE10420567GA), Not Applicable (SEE10420590GA), Not Applicable (SEE10420596GA), Not Applicable (SEE10450669GA), Not Applicable (SEE11161340GA), Not Applicable (SEE11161342GA), Not Applicable (SEE11161347GA), Not Applicable (SEE11171386GA), Not Applicable (SEE11171387GA), Not Applicable (SEE11191440GA), Not Applicable (SEE16334208SA), Not Applicable (SEE12400119GA), Not Applicable (SEE12410137GA), Not Applicable (SEE12410139GA), Not Applicable (SEE16390017SA), Not Applicable (SEE14432583GA), Not Applicable (SEE14432586GA), Not Applicable (SEE14432589GA), 00840682109260SEE19170008SA (SEE19170008SA), Not Applicable (SEE14041747GA), Not Applicable (SEE14041748GA), Not Applicable (SEE14041751GA), Not Applicable (SEE14041757GA), Not Applicable (SEE14041758GA), Not Applicable (SEE14041760GA), Not Applicable (SEE14041761GA), Not Applicable (SEE14041763GA), Not Applicable (SEE14041765GA), Not Applicable (SEE15312127SA), Not Applicable (SEE15301929SA), Not Applicable (SEE15301930SA), Not Applicable (SEE15301931SA), Not Applicable (SEE15301932SA), Not Applicable (SEE15301933SA), Not Applicable (SEE15301948SA), Not Applicable (SEE14151836GA), Not Applicable (SEE14151838GA), Not Applicable (SEE14151844GA), Not Applicable (SEE14151860GA), Not Applicable (SEE14151861GA), Not Applicable (SEE14222010GA), Not Applicable (SEE13270879GA), Not Applicable (SEE14031677GA), Not Applicable (SEE14151842GA), Not Applicable (SEE14151849GA), Not Applicable (SEE15062979GA), Not Applicable (SEE15153196GA), Not Applicable (SEE16420080SA), Not Applicable (SEE16430010SA), Not Applicable (SEE16430011SA), Not Applicable (SEE16430041SA), Not Applicable (SEE16430044SA), Not Applicable (SEE16430046SA), 00840682109260SEE17060001SA (SEE17060001SA), 00840682109260SEE17060002SA (SEE17060002SA), 00840682109260SEE17060004SA (SEE17060004SA), 00840682109260SEE17380003SA (SEE17380003SA), 00840682109260SEE17380005SA (SEE17380005SA), 00840682109260SEE17380007SA (SEE17380007SA), 00840682109260SEE17380009SA (SEE17380009SA), 00840682109260SEE17380011SA (SEE17380011SA), 00840682109260SEE17380014SA (SEE17380014SA), 00840682109260SEE17380015SA (SEE17380015SA), 00840682109260SEE17380017SA (SEE17380017SA), 00840682109260SEE18190035SA (SEE18190035SA), 00840682109260SEE18450005SA (SEE18450005SA), 00840682109260SEE18450006SA (SEE18450006SA), 00840682109260SEE18450008SA (SEE18450008SA), 00840682109260SEE19100001SA (SEE19100001SA), 00840682109260SEE19100028SA (SEE19100028SA), Not Applicable (SEE09040078GA), Not Applicable (SEE11111237GA), Not Applicable (SEE16334082SA), Not Applicable (SEE16334083SA), Not Applicable (SEE16334084SA), Not Applicable (SEE16334088SA), Not Applicable (SEE16334159SA), Not Applicable (SEE16334160SA), Not Applicable (SEE16334161SA), Not Applicable (SEE16334163SA), Not Applicable (SEE16334164SA), Not Applicable (SEE16334165SA), Not Applicable (SEE16334166SA), Not Applicable (SEE16334167SA), Not Applicable (SEE16334168SA), Not Applicable (SEE16334170SA), Not Applicable (SEE16334171SA), Not Applicable (SEE16334172SA), Not Applicable (SEE16334173SA), Not Applicable (SEE16334174SA), Not Applicable (SEE16334175SA), Not Applicable (SEE16334176SA), Not Applicable (SEE16334177SA), Not Applicable (SEE16334178SA), Not Applicable (SEE16334179SA), Not Applicable (SEE16334180SA), Not Applicable (SEE16334184SA), Not Applicable (SEE16334186SA), Not Applicable (SEE16334187SA), Not Applicable (SEE16334188SA), Not Applicable (SEE16334199SA), Not Applicable (SEE16334200SA), Not Applicable (SEE16334201SA), Not Applicable (SEE16334202SA), Not Applicable (SEE16334203SA), Not Applicable (SEE16334204SA), Not Applicable (SEE16334205SA), Not Applicable (SEE16334206SA), Not Applicable (SEE16334207SA), Not Applicable (SEE16334209SA), Not Applicable (SEE16334210SA), Not Applicable (SEE11171367GA), Not Applicable (SEE10250118GA), Not Applicable (SEE10250184GA), Not Applicable (SEE10250187GA), Not Applicable (SEE13290962GA), Not Applicable (SEE10460733GA), Not Applicable (SEE10460737GA), Not Applicable (SEE10460755GA), Not Applicable (SEE10460760GA), Not Applicable (SEE11050931GA), Not Applicable (SEE10420586GA), Not Applicable (SEE10430616GA), Not Applicable (SEE10460708GA), Not Applicable (SEE10460723GA), Not Applicable (SEE16383412GA), Not Applicable (SEE13110498GA), Not Applicable (SEE13110504GA), Not Applicable (SEE13110507GA), Not Applicable (SEE13120588GA), Not Applicable (SEE11050922GA), Not Applicable (SEE13190803GA), Not Applicable (SEE13190808GA), Not Applicable (SEE13190809GA), Not Applicable (SEE14021531GA), Not Applicable (SEE14432580GA), Not Applicable (SEE14432584GA), Not Applicable (SEE14432585GA), Not Applicable (SEE14432592GA), Not Applicable (SEE16241816SA), 00840682109260SEE18510023SA (SEE18510023SA), Not Applicable (SEE15183247GA), Not Applicable (SEE15183249GA), 00840682109260SEE18220001SA (SEE18220001SA), 00840682109260SEE18220002SA (SEE18220002SA), 00840682109260SEE18220003SA (SEE18220003SA), 00840682109260SEE18220004SA (SEE18220004SA), 00840682109260SEE18220006SA (SEE18220006SA), 00840682109260SEE18220007SA (SEE18220007SA), 00840682109260SEE18220011SA (SEE18220011SA), 00840682109260SEE18220012SA (SEE18220012SA), 00840682109260SEE18220013SA (SEE18220013SA), Not Applicable (SEE14422547GA), Not Applicable (SEE14422557GA), Not Applicable (SEE14422562GA), Not Applicable (SEE14422565GA), Not Applicable (SEE14292243GA), Not Applicable (SEE16390021SA), Not Applicable (SEE16420062SA), Not Applicable (SEE16420069SA), Not Applicable (SEE16420071SA), Not Applicable (SEE16420079SA), Not Applicable (SEE16430012SA), Not Applicable (SEE16430020SA), Not Applicable (SEE16430040SA), Not Applicable (SEE16430047SA), Not Applicable (SEE10390324GA), Not Applicable (SEE10390347GA), Not Applicable (SEE11081158GA), Not Applicable (SEE11081166GA), Not Applicable (SEE11081181GA), Not Applicable (SEE12082020GA), Not Applicable (SEE12082022GA), Not Applicable (SEE13431257GA), Not Applicable (SEE13431262GA), Not Applicable (SEE13431279GA), Not Applicable (SEE13431286GA), Not Applicable (SEE13431292GA), Not Applicable (SEE13401156GA), Not Applicable (SEE09100111GA), Not Applicable (SEE13491426GA), Not Applicable (SEE13491427GA), Not Applicable (SEE13491431GA), Not Applicable (SEE13491432GA), Not Applicable (SEE15405088SA), Not Applicable (SEE11071049GA), Not Applicable (SEE13130639GA), Not Applicable (SEE13130640GA), Not Applicable (SEE13130642GA), Not Applicable (SEE13150689GA), Not Applicable (SEE13150695GA), Not Applicable (SEE10250159GA), Not Applicable (SEE10260197GA), Not Applicable (SEE12410135GA), Not Applicable (SEE12410140GA), Not Applicable (SEE12490268GA), Not Applicable (SEE15517889SA), 00840682109260SEE19120017SA (SEE19120017SA), 00840682109260SEE19120021SA (SEE19120021SA), 00840682109260SEE19120030SA (SEE19120030SA), 00840682109260SEE19170025SA (SEE19170025SA), 00840682109260SEE19170035SA (SEE19170035SA), 00840682109260SEE19170036SA (SEE19170036SA), 00840682109260SEE19170037SA (SEE19170037SA), 00840682109260SEE19170041SA (SEE19170041SA), Not Applicable (SEE13120591GA), Not Applicable (SEE16390019SA), Not Applicable (SEE11061002GA), Not Applicable (SEE13120594GA), Not Applicable (SEE11071058GA), Not Applicable (SEE11071063GA), Not Applicable (SEE11071138GA), Not Applicable (SEE13180758GA), Not Applicable (SEE13180760GA), Not Applicable (SEE13180770GA), Not Applicable (SEE13180771GA), Not Applicable (SEE13270888GA), Not Applicable (SEE13431269GA), Not Applicable (SEE14452623GA), Not Applicable (SEE14452635GA), Not Applicable (SEE10400444GA), Not Applicable (SEE10400460GA), Not Applicable (SEE11081203GA), Not Applicable (SEE11081205GA), Not Applicable (SEE11101213GA), Not Applicable (SEE11101217GA), Not Applicable (SEE12182219GA), Not Applicable (SEE12232253GA), Not Applicable (SEE12232257GA), Not Applicable (SEE12232258GA), Not Applicable (SEE12232259GA), Not Applicable (SEE12232261GA), Not Applicable (SEE12232263GA), Not Applicable (SEE12232265GA), Not Applicable (SEE12232266GA), Not Applicable (SEE12242268GA), Not Applicable (SEE12242269GA), Not Applicable (SEE12242270GA), Not Applicable (SEE12242276GA), Not Applicable (SEE12242281GA), Not Applicable (SEE12242286GA), Not Applicable (SEE12242288GA), Not Applicable (SEE12242291GA), Not Applicable (SEE12242293GA), Not Applicable (SEE12430229GA), Not Applicable (SEE12440231GA), Not Applicable (SEE12440238GA), Not Applicable (SEE12510308GA), Not Applicable (SEE12520346GA), Not Applicable (SEE12520359GA), Not Applicable (SEE12530366GA), Not Applicable (SEE12530369GA), Not Applicable (SEE12530372GA), Not Applicable (SEE13000376GA), Not Applicable (SEE13000377GA), Not Applicable (SEE13000378GA), Not Applicable (SEE13000393GA), Not Applicable (SEE13150683GA), Not Applicable (SEE13150685GA), Not Applicable (SEE13431271GA), Not Applicable (SEE13431273GA), Not Applicable (SEE13431290GA), Not Applicable (SEE13431294GA), Not Applicable (SEE13431299GA), Not Applicable (SEE13431303GA), Not Applicable (SEE13431304GA), Not Applicable (SEE13441305GA), Not Applicable (SEE13451337GA), Not Applicable (SEE13451341GA), Not Applicable (SEE13451350GA), Not Applicable (SEE14021570GA), Not Applicable (SEE14021580GA), Not Applicable (SEE14031698GA), Not Applicable (SEE14031707GA), Not Applicable (SEE14031710GA), Not Applicable (SEE14031716GA), Not Applicable (SEE14041738GA), Not Applicable (SEE14041742GA), Not Applicable (SEE14372390GA), Not Applicable (SEE14372391GA), Not Applicable (SEE14372395GA), Not Applicable (SEE14372396GA), Not Applicable (SEE14372397GA), Not Applicable (SEE14372398GA), Not Applicable (SEE14372399GA), Not Applicable (SEE14372400GA), Not Applicable (SEE14372401GA), Not Applicable (SEE14372406GA), Not Applicable (SEE14372407GA), Not Applicable (SEE14372440GA), Not Applicable (SEE15052925GA), Not Applicable (SEE15052944GA), Not Applicable (SEE15052945GA), Not Applicable (SEE15062970GA), Not Applicable (SEE15062973GA), Not Applicable (SEE15062975GA), Not Applicable (SEE15062981GA), Not Applicable (SEE15063014GA), Not Applicable (SEE15063022GA), Not Applicable (SEE15073023GA), Not Applicable (SEE15241002SA), Not Applicable (SEE15241003SA), Not Applicable (SEE15241005SA), Not Applicable (SEE15241007SA), Not Applicable (SEE15373437SA), Not Applicable (SEE16038440SA), Not Applicable (SEE16038441SA), Not Applicable (SEE16038442SA), Not Applicable (SEE16344283SA), Not Applicable (SEE16344323SA), 00840682109260SEE17070004SA (SEE17070004SA), 00840682109260SEE17070010SA (SEE17070010SA), 00840682109260SEE17200001SA (SEE17200001SA), 00840682109260SEE17200003SA (SEE17200003SA), 00840682109260SEE18080001SA (SEE18080001SA), 00840682109260SEE18080012SA (SEE18080012SA), 00840682109260SEE18240011SA (SEE18240011SA), 00840682109260SEE18240019SA (SEE18240019SA), 00840682109260SEE18510018SA (SEE18510018SA), Not Applicable (SEE13431291GA), Not Applicable (SEE10260232GA), Not Applicable (SEE10260237GA), Not Applicable (SEE10310286GA), Not Applicable (SEE16293172SA), Not Applicable (SEE16293175SA), Not Applicable (SEE16293176SA), Not Applicable (SEE16293181SA), 00840682109260SEE18240056SA (SEE18240056SA), Not Applicable (SEE12340002GA), Not Applicable (SEE14452617GA), Not Applicable (SEE14452620GA), Not Applicable (SEE16241825SA), Not Applicable (SEE16241830SA), Not Applicable (SEE16241843SA), 00840682109260SEE19190003SA (SEE19190003SA), Not Applicable (SEE15052915GA), Not Applicable (SEE15052929GA), Not Applicable (SEE15093032GA), Not Applicable (SEE14372434GA), Not Applicable (SEE10460742GA), Not Applicable (SEE15210698SA), Not Applicable (SEE16430060SA), 00840682109260SEE17480001SA (SEE17480001SA), 00840682109260SEE18240058SA (SEE18240058SA), Not Applicable (SEE11331590GA), Not Applicable (SEE13270878GA), Not Applicable (SEE15517770SA), Not Applicable (SEE15517775SA), Not Applicable (SEE15301891SA), 00840682109260SEE18070005SA (SEE18070005SA), Not Applicable (SEE09530847GA), Not Applicable (SEE09530853GA), Not Applicable (SEE09530867GA), Not Applicable (SEE09530869GA), Not Applicable (SEE13280936GA), Not Applicable (SEE13280948GA), Not Applicable (SEE13501454GA), Not Applicable (SEE13501468GA), Not Applicable (SEE13501477GA), Not Applicable (SEE14312293GA), Not Applicable (SEE16380007SA), 00840682109260SEE18390008SA (SEE18390008SA), 00840682109260SEE19220009SA (SEE19220009SA), 00840682109260SEE19180007SA (SEE19180007SA), Not Applicable (SEE12410149GA), Not Applicable (SEE11071024GA), Not Applicable (SEE14151847GA), Not Applicable (SEE11211492GA), Not Applicable (SEE11211504GA), Not Applicable (SEE11241550GA), Not Applicable (SEE16191062SA), Not Applicable (SEE09140176GA), Not Applicable (SEE09140184GA), Not Applicable (SEE12182210GA), Not Applicable (SEE13240853GA), Not Applicable (SEE13240855GA), Not Applicable (SEE14362358GA), Not Applicable (SEE14362375GA), Not Applicable (SEE10120022GA), Not Applicable (SEE10250116GA), Not Applicable (SEE10250133GA), Not Applicable (SEE10250143GA), Not Applicable (SEE11040916GA), Not Applicable (SEE11050939GA), Not Applicable (SEE13110549GA), 00840682109260SEE17070020SA (SEE17070020SA), Not Applicable (SEE10121020GA), Not Applicable (SEE10121028GA), Not Applicable (SEE11201472GA), Not Applicable (SEE09340444GA), Not Applicable (SEE09510832GA), Not Applicable (SEE09530896GA), Not Applicable (SEE09530907GA), Not Applicable (SEE09530911GA), Not Applicable (SEE10450675GA), Not Applicable (SEE11081193GA), Not Applicable (SEE11371785GA), Not Applicable (SEE11511866GA), Not Applicable (SEE13280924GA), Not Applicable (SEE13290981GA), Not Applicable (SEE13311001GA), Not Applicable (SEE14432587GA), Not Applicable (SEE14462656GA), Not Applicable (SEE14462665GA), Not Applicable (SEE14462666GA), Not Applicable (SEE15113153GA), Not Applicable (SEE15113175GA), Not Applicable (SEE15113180GA), 00840682109260SEE18370010SA (SEE18370010SA), 00840682109260SEE18410016SA (SEE18410016SA), 00840682109260SEE19090021SA (SEE19090021SA), 00840682109260SEE19090023SA (SEE19090023SA), 01008406821092601119061121SEE19240027SA (SEE19240027SA), Not Applicable (SEE11071035GA), Not Applicable (SEE11071056GA), Not Applicable (SEE11071086GA), Not Applicable (SEE11071092GA), Not Applicable (SEE11071094GA), Not Applicable (SEE11151298GA), Not Applicable (SEE13220843GA), Not Applicable (SEE16390023SA), Not Applicable (SEE16420077SA), Not Applicable (SEE16430038SA), 00840682109260SEE19240021SA (SEE19240021SA), 00840682109260SEE19240029SA (SEE19240029SA), Not Applicable (SEE10420575GA), Not Applicable (SEE11361685GA), Not Applicable (SEE12390080GA), Not Applicable (SEE12390081GA), Not Applicable (SEE09390534GA), Not Applicable (SEE09020036GA), Not Applicable (SEE09190228GA), Not Applicable (SEE09240302GA), Not Applicable (SEE09240303GA), Not Applicable (SEE09250304GA), Not Applicable (SEE09250305GA), Not Applicable (SEE09250316GA), Not Applicable (SEE09250320GA), Not Applicable (SEE09250321GA), Not Applicable (SEE09250323GA), Not Applicable (SEE09250325GA), Not Applicable (SEE09250327GA), Not Applicable (SEE09250329GA), Not Applicable (SEE09250330GA), Not Applicable (SEE09250331GA), Not Applicable (SEE09250332GA), Not Applicable (SEE09260334GA), Not Applicable (SEE09260335GA), Not Applicable (SEE09260336GA), Not Applicable (SEE09260342GA), Not Applicable (SEE09260343GA), Not Applicable (SEE09260344GA), Not Applicable (SEE09260345GA), Not Applicable (SEE09300346GA), Not Applicable (SEE09300348GA), Not Applicable (SEE09300349GA), Not Applicable (SEE09300350GA), Not Applicable (SEE09300354GA), Not Applicable (SEE09300355GA), Not Applicable (SEE10260234GA), Not Applicable (SEE10430649GA), Not Applicable (SEE10450686GA), Not Applicable (SEE11141279GA), Not Applicable (SEE11141282GA), Not Applicable (SEE11341625GA), Not Applicable (SEE11361701GA), Not Applicable (SEE12530371GA), Not Applicable (SEE12530373GA), Not Applicable (SEE13050431GA), Not Applicable (SEE13050432GA), Not Applicable (SEE13050433GA), Not Applicable (SEE13050435GA), Not Applicable (SEE13050436GA), Not Applicable (SEE13050437GA), Not Applicable (SEE13050441GA), Not Applicable (SEE13060444GA), Not Applicable (SEE13060447GA), Not Applicable (SEE13060448GA), Not Applicable (SEE13060449GA), Not Applicable (SEE13060451GA), Not Applicable (SEE13060452GA), Not Applicable (SEE13060453GA), Not Applicable (SEE13060454GA), Not Applicable (SEE13180741GA), Not Applicable (SEE13180772GA), Not Applicable (SEE13451344GA), Not Applicable (SEE14382453GA), Not Applicable (SEE14382458GA), Not Applicable (SEE14382461GA), Not Applicable (SEE14382464GA), Not Applicable (SEE14412498GA), Not Applicable (SEE14452629GA), 00840682109260SEE17210006SA (SEE17210006SA), 00840682109260SEE17210007SA (SEE17210007SA), 00840682109260SEE17210009SA (SEE17210009SA), 00840682109260SEE17210010SA (SEE17210010SA), 00840682109260SEE17220001SA (SEE17220001SA), 00840682109260SEE17220003SA (SEE17220003SA), 00840682109260SEE18100029SA (SEE18100029SA), 00840682109260SEE19150003SA (SEE19150003SA), 00840682109260SEE19160010SA (SEE19160010SA), Not Applicable (SEE14422543GA), 00840682109260SEE17130003SA (SEE17130003SA), 00840682109260SEE17130006SA (SEE17130006SA), Not Applicable (SEE13140652GA), Not Applicable (SEE13180739GA), Not Applicable (SEE16450005SA), Not Applicable (SEE16450006SA), Not Applicable (SEE11071051GA), Not Applicable (SEE11071097GA), Not Applicable (SEE11071154GA), Not Applicable (SEE12122149GA), Not Applicable (SEE09330417GA), Not Applicable (SEE09530836GA), Not Applicable (SEE09530873GA), Not Applicable (SEE10250102GA), Not Applicable (SEE10390311GA), Not Applicable (SEE10530851GA), Not Applicable (SEE11161333GA), Not Applicable (SEE11171355GA), Not Applicable (SEE11171357GA), Not Applicable (SEE11171362GA), Not Applicable (SEE11211486GA), Not Applicable (SEE11211518GA), Not Applicable (SEE11321587GA), Not Applicable (SEE11341630GA), Not Applicable (SEE11341635GA), Not Applicable (SEE11351667GA), Not Applicable (SEE12312402GA), Not Applicable (SEE13351014GA), Not Applicable (SEE13351025GA), Not Applicable (SEE13351030GA), Not Applicable (SEE13351053GA), Not Applicable (SEE13361066GA), Not Applicable (SEE13361070GA), Not Applicable (SEE13391135GA), Not Applicable (SEE13411197GA), Not Applicable (SEE14021503GA), Not Applicable (SEE14031689GA), Not Applicable (SEE14031719GA), Not Applicable (SEE14041726GA), Not Applicable (SEE15093042GA), Not Applicable (SEE15093043GA), Not Applicable (SEE15173225GA), Not Applicable (SEE15261581SA), 00840682109260SEE17180004SA (SEE17180004SA), 00840682109260SEE17230074SA (SEE17230074SA), 00840682109260SEE19090025SA (SEE19090025SA), Not Applicable (SEE12021902GA), Not Applicable (SEE13220850GA), Not Applicable (SEE13270873GA), Not Applicable (SEE14151848GA), Not Applicable (SEE15352537SA), Not Applicable (SEE15352538SA), Not Applicable (SEE15352546SA), Not Applicable (SEE15352553SA), Not Applicable (SEE15352554SA), Not Applicable (SEE15352559SA), Not Applicable (SEE15352560SA), Not Applicable (SEE15352563SA), Not Applicable (SEE15352564SA), Not Applicable (SEE14452622GA), Not Applicable (SEE14452632GA), 00840682109260SEE19100030SA (SEE19100030SA), 00840682109260SEE19100033SA (SEE19100033SA), Not Applicable (SEE10260198GA), Not Applicable (SEE10400363GA), Not Applicable (SEE10400369GA), Not Applicable (SEE10400381GA), Not Applicable (SEE10400382GA), Not Applicable (SEE11141255GA), Not Applicable (SEE11361732GA), Not Applicable (SEE11361735GA), Not Applicable (SEE13150663GA), Not Applicable (SEE13150666GA), Not Applicable (SEE13150669GA), 00840682109260SEE19110023SA (SEE19110023SA), 00840682109260SEE19110025SA (SEE19110025SA), 00840682109260SEE19110028SA (SEE19110028SA), 00840682109260SEE19110041SA (SEE19110041SA), 00840682109260SEE09510834GA (SEE09510834GA), 00840682109260SEE09530844GA (SEE09530844GA), 00840682109260SEE10100975GA (SEE10100975GA), 00840682109260SEE10100978GA (SEE10100978GA), 00840682109260SEE10120002GA (SEE10120002GA), Not Applicable (SEE10120004GA), 00840682109260SEE10120005GA (SEE10120005GA), 00840682109260SEE10120013GA (SEE10120013GA), 00840682109260SEE10120014GA (SEE10120014GA), 00840682109260SEE10120015GA (SEE10120015GA), 00840682109260SEE10120024GA (SEE10120024GA), 00840682109260SEE10120995GA (SEE10120995GA), 00840682109260SEE10120997GA (SEE10120997GA), 00840682109260SEE10121010GA (SEE10121010GA), 00840682109260SEE10121012GA (SEE10121012GA), 00840682109260SEE10121016GA (SEE10121016GA), 00840682109260SEE10130038GA (SEE10130038GA), 00840682109260SEE10430605GA (SEE10430605GA), 00840682109260SEE17060007SA (SEE17060007SA), 00840682109260SEE18350004SA (SEE18350004SA), 00840682109260SEE18350006SA (SEE18350006SA), 00840682109260SEE19110007SA (SEE19110007SA), 00840682109260SEE19110043SA (SEE19110043SA), 00840682109260SEE19110044SA (SEE19110044SA), Not Applicable (SEE10400457GA), Not Applicable (SEE10410465GA), Not Applicable (SEE10410466GA), Not Applicable (SEE10410496GA), Not Applicable (SEE10410497GA), Not Applicable (SEE10410500GA), Not Applicable (SEE10410501GA), Not Applicable (SEE10410510GA), Not Applicable (SEE10410512GA), Not Applicable (SEE11321581GA), Not Applicable (SEE11321582GA), Not Applicable (SEE12092053GA), Not Applicable (SEE12092060GA), Not Applicable (SEE12092061GA), Not Applicable (SEE12092063GA), Not Applicable (SEE13411194GA), Not Applicable (SEE13411196GA), Not Applicable (SEE13411199GA), Not Applicable (SEE13411200GA), Not Applicable (SEE13411201GA), Not Applicable (SEE13411210GA), Not Applicable (SEE13411213GA), Not Applicable (SEE13411216GA), Not Applicable (SEE13411217GA), Not Applicable (SEE14151811GA), Not Applicable (SEE14151814GA), Not Applicable (SEE14151825GA), Not Applicable (SEE14151831GA), Not Applicable (SEE14151850GA), 00840682109260SEE18240038SA (SEE18240038SA), 00840682109260SEE18240042SA (SEE18240042SA), Not Applicable (SEE13150671GA), Not Applicable (SEE10410469GA), Not Applicable (SEE10410482GA), 00840682109260SEE17100002SA (SEE17100002SA), Not Applicable (SEE16380024SA), Not Applicable (SEE10530854GA), Not Applicable (SEE10530856GA), Not Applicable (SEE10530861GA), Not Applicable (SEE10530869GA), Not Applicable (SEE13220844GA), Not Applicable (SEE09250324GA), Not Applicable (SEE09530893GA), Not Applicable (SEE09530912GA), Not Applicable (SEE13401142GA), 00840682109260SEE18160046SA (SEE18160046SA), 00840682109260SEE18160055SA (SEE18160055SA), Not Applicable (SEE11060997GA), Not Applicable (SEE11071005GA), Not Applicable (SEE11071006GA), Not Applicable (SEE10100982GA), Not Applicable (SEE10260216GA), Not Applicable (SEE10270246GA), Not Applicable (SEE11171383GA), Not Applicable (SEE12282357GA), Not Applicable (SEE15220796SA), Not Applicable (SEE15230839SA), 00840682109260SEE16520003SA (SEE16520003SA), 00840682109260SEE17450003SA (SEE17450003SA), 00840682109260SEE17450004SA (SEE17450004SA), 00840682109260SEE17450005SA (SEE17450005SA), 00840682109260SEE18100002SA (SEE18100002SA), 00840682109260SEE18100012SA (SEE18100012SA), Not Applicable (SEE11391832GA), Not Applicable (SEE11391838GA), Not Applicable (SEE11391839GA), 00840682109260SEE17270007SA (SEE17270007SA), Not Applicable (SEE09020027GA), Not Applicable (SEE15022868GA), Not Applicable (SEE15415183SA), 00840682109260SEE18050001SA (SEE18050001SA), 00840682109260SEE18050002SA (SEE18050002SA), Not Applicable (SEE12322414GA), Not Applicable (SEE13421253GA), Not Applicable (SEE13431275GA), Not Applicable (SEE13431301GA), Not Applicable (SEE09100138GA), Not Applicable (SEE09130171GA), Not Applicable (SEE09140187GA), Not Applicable (SEE09140190GA), Not Applicable (SEE09150200GA), Not Applicable (SEE09150202GA), Not Applicable (SEE09150205GA), Not Applicable (SEE09480743GA), Not Applicable (SEE10121021GA), Not Applicable (SEE10390335GA), Not Applicable (SEE10390336GA), Not Applicable (SEE10390338GA), Not Applicable (SEE10390339GA), Not Applicable (SEE10390345GA), Not Applicable (SEE10390346GA), Not Applicable (SEE10390351GA), Not Applicable (SEE10390352GA), Not Applicable (SEE10390354GA), Not Applicable (SEE10390357GA), Not Applicable (SEE10390358GA), Not Applicable (SEE10400371GA), Not Applicable (SEE10430623GA), Not Applicable (SEE10430652GA), Not Applicable (SEE10460705GA), Not Applicable (SEE11050972GA), Not Applicable (SEE11361672GA), Not Applicable (SEE12342453GA), Not Applicable (SEE14322315GA), Not Applicable (SEE14322318GA), Not Applicable (SEE14322327GA), Not Applicable (SEE14322332GA), Not Applicable (SEE14482683GA), Not Applicable (SEE14482702GA), Not Applicable (SEE14492722GA), Not Applicable (SEE14492724GA), Not Applicable (SEE14492725GA), Not Applicable (SEE14492726GA), Not Applicable (SEE14492727GA), Not Applicable (SEE14492728GA), Not Applicable (SEE14492730GA), Not Applicable (SEE14492731GA), Not Applicable (SEE14492745GA), Not Applicable (SEE14492747GA), Not Applicable (SEE14492748GA), Not Applicable (SEE14492749GA), Not Applicable (SEE14492750GA), Not Applicable (SEE14492751GA), Not Applicable (SEE14492752GA), Not Applicable (SEE14492753GA), Not Applicable (SEE14492754GA), Not Applicable (SEE14492755GA), Not Applicable (SEE14492756GA), Not Applicable (SEE14492757GA), Not Applicable (SEE14492758GA), Not Applicable (SEE14492759GA), Not Applicable (SEE14492760GA), Not Applicable (SEE14502762GA), Not Applicable (SEE14502763GA), Not Applicable (SEE14502764GA), Not Applicable (SEE14502765GA), Not Applicable (SEE14502766GA), Not Applicable (SEE14502767GA), Not Applicable (SEE14502768GA), Not Applicable (SEE14502769GA), Not Applicable (SEE14502770GA), Not Applicable (SEE14502771GA), Not Applicable (SEE14502772GA), Not Applicable (SEE14502773GA), Not Applicable (SEE14502774GA), Not Applicable (SEE14502775GA), Not Applicable (SEE14502776GA), Not Applicable (SEE14502777GA), Not Applicable (SEE14502778GA), Not Applicable (SEE14502779GA), Not Applicable (SEE14502780GA), 00840682109260SEE18160028SA (SEE18160028SA), Not Applicable (SEE09410560GA), Not Applicable (SEE09510769GA), Not Applicable (SEE09510776GA), Not Applicable (SEE09510783GA), Not Applicable (SEE09510801GA), Not Applicable (SEE09530863GA), Not Applicable (SEE11071034GA), Not Applicable (SEE11141273GA), Not Applicable (SEE11211514GA), Not Applicable (SEE11391833GA), Not Applicable (SEE11391834GA), Not Applicable (SEE11391836GA), Not Applicable (SEE12021907GA), Not Applicable (SEE12092058GA), Not Applicable (SEE12112121GA), Not Applicable (SEE12112122GA), Not Applicable (SEE12242289GA), Not Applicable (SEE12390071GA), Not Applicable (SEE12420168GA), Not Applicable (SEE13010405GA), Not Applicable (SEE13240857GA), Not Applicable (SEE13401153GA), Not Applicable (SEE13401157GA), Not Applicable (SEE13501453GA), Not Applicable (SEE13501470GA), Not Applicable (SEE14151812GA), Not Applicable (SEE14151841GA), Not Applicable (SEE14151863GA), Not Applicable (SEE14222039GA), Not Applicable (SEE14222050GA), Not Applicable (SEE14222057GA), Not Applicable (SEE14222059GA), Not Applicable (SEE14222067GA), Not Applicable (SEE14222072GA), Not Applicable (SEE14222073GA), Not Applicable (SEE14222075GA), Not Applicable (SEE15022862GA), 00840682109260SEE15032870GA (SEE15032870GA), 00840682109260SEE15032872GA (SEE15032872GA), 00840682109260SEE15032873GA (SEE15032873GA), 00840682109260SEE15032874GA (SEE15032874GA), 00840682109260SEE15032875GA (SEE15032875GA), Not Applicable (SEE15032877GA), Not Applicable (SEE15032881GA), 00840682109260SEE15032885GA (SEE15032885GA), 00840682109260SEE15032888GA (SEE15032888GA), 00840682109260SEE15032890GA (SEE15032890GA), Not Applicable (SEE15032893GA), 00840682109260SEE15301888SA (SEE15301888SA), Not Applicable (SEE15435919SA), Not Applicable (SEE15435920SA), Not Applicable (SEE15435921SA), Not Applicable (SEE15435922SA), Not Applicable (SEE15435927SA), Not Applicable (SEE15435928SA), Not Applicable (SEE15435933SA), Not Applicable (SEE15435934SA), Not Applicable (SEE16324001SA), Not Applicable (SEE16324026SA), Not Applicable (SEE16324027SA), Not Applicable (SEE16324031SA), Not Applicable (SEE16324033SA), Not Applicable (SEE16370006SA), 00840682109260SEE17270010SA (SEE17270010SA), 00840682109260SEE17340005SA (SEE17340005SA), 00840682109260SEE17340009SA (SEE17340009SA), 00840682109260SEE17340010SA (SEE17340010SA), 00840682109260SEE17340012SA (SEE17340012SA), 00840682109260SEE17340013SA (SEE17340013SA), 00840682109260SEE17340014SA (SEE17340014SA), 00840682109260SEE17460004SA (SEE17460004SA), 00840682109260SEE17510002SA (SEE17510002SA), 00840682109260SEE18030010SA (SEE18030010SA), 00840682109260SEE18040013SA (SEE18040013SA), Not Applicable (SEE12420167GA), Not Applicable (SEE10250083GA), Not Applicable (SEE10250085GA), Not Applicable (SEE10250098GA), Not Applicable (SEE10250100GA), Not Applicable (SEE10250101GA), Not Applicable (SEE10250111GA), Not Applicable (SEE10250145GA), Not Applicable (SEE10250156GA), Not Applicable (SEE10260233GA), Not Applicable (SEE15332401SA), Not Applicable (SEE15342480SA), Not Applicable (SEE15342482SA), Not Applicable (SEE15446117SA), Not Applicable (SEE15446119SA), Not Applicable (SEE15446121SA), Not Applicable (SEE15446124SA), Not Applicable (SEE15446169SA), Not Applicable (SEE15446170SA), Not Applicable (SEE15446201SA), Not Applicable (SEE15446203SA), Not Applicable (SEE15446204SA), Not Applicable (SEE15446215SA), Not Applicable (SEE15446216SA), 00840682109260SEE18190005SA (SEE18190005SA), 00840682109260SEE18190006SA (SEE18190006SA), 00840682109260SEE18190021SA (SEE18190021SA), 00840682109260SEE18240012SA (SEE18240012SA), 00840682109260SEE18240014SA (SEE18240014SA), 00840682109260SEE18240050SA (SEE18240050SA), Not Applicable (SEE13150686GA), Not Applicable (SEE09390540GA), Not Applicable (SEE09410554GA), Not Applicable (SEE09410556GA), Not Applicable (SEE09410557GA), Not Applicable (SEE09410558GA), Not Applicable (SEE09410559GA), Not Applicable (SEE09410575GA), Not Applicable (SEE14171933GA), Not Applicable (SEE14191941GA), Not Applicable (SEE14191942GA), Not Applicable (SEE14191943GA), Not Applicable (SEE14191945GA), Not Applicable (SEE14191947GA), Not Applicable (SEE14191948GA), Not Applicable (SEE14191949GA), Not Applicable (SEE09310372GA), Not Applicable (SEE09310379GA), Not Applicable (SEE11091212GA), Not Applicable (SEE11101218GA), Not Applicable (SEE11211475GA), Not Applicable (SEE11211508GA), Not Applicable (SEE14462667GA), Not Applicable (SEE14462668GA), 00840682109260SEE18160022SA (SEE18160022SA), 00840682109260SEE18160030SA (SEE18160030SA), Not Applicable (SEE09100084GA), Not Applicable (SEE09230282GA), Not Applicable (SEE10250125GA), Not Applicable (SEE10270257GA), Not Applicable (SEE09510812GA), Not Applicable (SEE12061996GA), Not Applicable (SEE14272172GA), Not Applicable (SEE14272173GA), Not Applicable (SEE14272213GA), Not Applicable (SEE14272214GA), Not Applicable (SEE10390348GA), Not Applicable (SEE10390350GA), Not Applicable (SEE10390361GA), Not Applicable (SEE10400364GA), Not Applicable (SEE10400366GA), Not Applicable (SEE11341606GA), Not Applicable (SEE11341609GA), Not Applicable (SEE11341612GA), Not Applicable (SEE11341615GA), Not Applicable (SEE11341620GA), Not Applicable (SEE11341660GA), Not Applicable (SEE11361676GA), Not Applicable (SEE11371765GA), Not Applicable (SEE12021908GA), Not Applicable (SEE12021911GA), Not Applicable (SEE12021922GA), Not Applicable (SEE12021923GA), Not Applicable (SEE12021924GA), Not Applicable (SEE12051989GA), Not Applicable (SEE16211513SA), Not Applicable (SEE11511883GA), Not Applicable (SEE11521890GA), Not Applicable (SEE13290980GA), Not Applicable (SEE13300991GA), Not Applicable (SEE13311002GA), Not Applicable (SEE13311005GA), Not Applicable (SEE13311008GA), Not Applicable (SEE09480748GA), Not Applicable (SEE13180733GA), Not Applicable (SEE13180756GA), Not Applicable (SEE13240852GA), Not Applicable (SEE13240858GA), Not Applicable (SEE13240860GA), Not Applicable (SEE13240866GA), Not Applicable (SEE13240869GA), Not Applicable (SEE14362365GA), Not Applicable (SEE12011898GA), Not Applicable (SEE16344325SA), 00840682109260SEE17050010SA (SEE17050010SA), Not Applicable (SEE16450012SA), Not Applicable (SEE16450014SA), Not Applicable (SEE09230265GA), Not Applicable (SEE09230268GA), Not Applicable (SEE09230281GA), Not Applicable (SEE09530837GA), Not Applicable (SEE09530859GA), Not Applicable (SEE11161338GA), Not Applicable (SEE11231532GA), Not Applicable (SEE11231536GA), Not Applicable (SEE10390298GA), Not Applicable (SEE13130620GA), Not Applicable (SEE13130622GA), Not Applicable (SEE12302379GA), Not Applicable (SEE12302391GA), Not Applicable (SEE12302396GA), Not Applicable (SEE15093035GA), Not Applicable (SEE15093036GA), Not Applicable (SEE15093041GA), Not Applicable (SEE15093044GA), 00840682109260SEE19090012SA (SEE19090012SA), Not Applicable (SEE13220842GA), Not Applicable (SEE13220845GA), Not Applicable (SEE13270875GA), Not Applicable (SEE13270876GA), Not Applicable (SEE13270882GA), Not Applicable (SEE13270885GA), Not Applicable (SEE11171369GA), Not Applicable (SEE11181410GA), Not Applicable (SEE11181413GA), Not Applicable (SEE11181415GA), Not Applicable (SEE11171365GA), Not Applicable (SEE09510787GA), Not Applicable (SEE11181401GA), Not Applicable (SEE13270896GA), Not Applicable (SEE13270897GA), Not Applicable (SEE15062999GA), Not Applicable (SEE15063001GA), Not Applicable (SEE15063007GA), Not Applicable (SEE12031929GA), Not Applicable (SEE12031934GA), Not Applicable (SEE09360486GA), Not Applicable (SEE09360488GA), Not Applicable (SEE12031928GA), Not Applicable (SEE12041975GA), Not Applicable (SEE12041976GA), Not Applicable (SEE12041977GA), Not Applicable (SEE12051978GA), Not Applicable (SEE12051979GA), Not Applicable (SEE12051980GA), Not Applicable (SEE12051981GA), Not Applicable (SEE12051982GA), Not Applicable (SEE12051983GA), Not Applicable (SEE12051984GA), Not Applicable (SEE12051985GA), Not Applicable (SEE14322308GA), Not Applicable (SEE14322310GA), Not Applicable (SEE14322317GA), Not Applicable (SEE15446200SA), Not Applicable (SEE15446214SA), 00840682109260SEE12092054GA (SEE12092054GA), 00840682109260SEE14081771GA (SEE14081771GA), 00840682109260SEE14081775GA (SEE14081775GA), 00840682109260SEE15032889GA (SEE15032889GA), Not Applicable (SEE16410002SA), Not Applicable (SEE16410004SA), Not Applicable (SEE16410009SA), Not Applicable (SEE16410014SA), Not Applicable (SEE16410016SA), Not Applicable (SEE10250084GA), Not Applicable (SEE10250099GA), Not Applicable (SEE10250136GA), Not Applicable (SEE15518044SA), Not Applicable (SEE15528074SA), Not Applicable (SEE11081188GA), Not Applicable (SEE15183261GA), 00840682109260SEE18320010SA (SEE18320010SA), 00840682109260SEE18320011SA (SEE18320011SA), Not Applicable (SEE12370048GA), Not Applicable (SEE12370057GA), Not Applicable (SEE09190238GA), Not Applicable (SEE09530840GA), Not Applicable (SEE09530865GA), Not Applicable (SEE12430206GA), Not Applicable (SEE12430209GA), Not Applicable (SEE16099454SA), Not Applicable (SEE12520336GA), Not Applicable (SEE12520351GA), Not Applicable (SEE12520353GA), Not Applicable (SEE13020413GA), Not Applicable (SEE13280922GA), Not Applicable (SEE13280925GA), Not Applicable (SEE13280926GA), Not Applicable (SEE13280928GA), Not Applicable (SEE13280930GA), Not Applicable (SEE13280931GA), Not Applicable (SEE16130303SA), Not Applicable (SEE16130305SA), Not Applicable (SEE16140328SA), Not Applicable (SEE16140330SA), Not Applicable (SEE16150532SA), Not Applicable (SEE16150540SA), Not Applicable (SEE16150541SA), Not Applicable (SEE16150542SA), Not Applicable (SEE16150543SA), Not Applicable (SEE16150544SA), Not Applicable (SEE16180898SA), Not Applicable (SEE16334182SA), Not Applicable (SEE16334185SA), Not Applicable (SEE14382454GA), Not Applicable (SEE16344326SA), Not Applicable (SEE15477175SA), Not Applicable (SEE15477184SA), Not Applicable (SEE15497498SA), Not Applicable (SEE15497500SA), Not Applicable (SEE15497501SA), Not Applicable (SEE15497510SA), Not Applicable (SEE15497511SA), Not Applicable (SEE15497512SA), Not Applicable (SEE15497513SA), Not Applicable (SEE15497514SA), Not Applicable (SEE15497515SA), Not Applicable (SEE15497516SA), Not Applicable (SEE15497517SA), Not Applicable (SEE15497520SA), Not Applicable (SEE15497521SA), Not Applicable (SEE15497522SA), Not Applicable (SEE15497523SA), Not Applicable (SEE15497525SA), Not Applicable (SEE15497526SA), Not Applicable (SEE15497528SA), Not Applicable (SEE15497529SA), Not Applicable (SEE15497531SA), Not Applicable (SEE15497532SA), Not Applicable (SEE15497533SA), Not Applicable (SEE15497534SA), Not Applicable (SEE15497535SA), Not Applicable (SEE15528073SA), Not Applicable (SEE10530852GA), Not Applicable (SEE10530870GA), Not Applicable (SEE11071055GA), Not Applicable (SEE11071147GA), Not Applicable (SEE11371748GA), Not Applicable (SEE11371756GA), Not Applicable (SEE11371795GA), Not Applicable (SEE15241004SA), 00840682109260SEE19110003SA (SEE19110003SA), 00840682109260SEE19110022SA (SEE19110022SA), 00840682109260SEE19110026SA (SEE19110026SA), 00840682109260SEE19110029SA (SEE19110029SA), 00840682109260SEE19110045SA (SEE19110045SA), 00840682109260SEE19110046SA (SEE19110046SA), Not Applicable (SEE11321564GA), Not Applicable (SEE11321567GA), Not Applicable (SEE11321573GA), Not Applicable (SEE16430024SA), Not Applicable (SEE13431295GA), Not Applicable (SEE11321579GA), Not Applicable (SEE10390323GA), Not Applicable (SEE11371788GA), Not Applicable (SEE11151301GA), Not Applicable (SEE11151302GA), Not Applicable (SEE11151305GA), Not Applicable (SEE11161327GA), Not Applicable (SEE12430188GA), Not Applicable (SEE14031684GA), Not Applicable (SEE14462644GA), Not Applicable (SEE14462646GA), Not Applicable (SEE14462649GA), Not Applicable (SEE14462652GA), Not Applicable (SEE15062989GA), Not Applicable (SEE15062991GA), Not Applicable (SEE15062997GA), Not Applicable (SEE15063009GA), Not Applicable (SEE15063012GA), Not Applicable (SEE15063016GA), Not Applicable (SEE16120171SA), Not Applicable (SEE09190232GA), Not Applicable (SEE09190235GA), Not Applicable (SEE09510830GA), Not Applicable (SEE09530842GA), Not Applicable (SEE11081156GA), Not Applicable (SEE14031664GA), Not Applicable (SEE15032883GA), Not Applicable (SEE15032899GA), 00840682109260SEE17410004SA (SEE17410004SA), Not Applicable (SEE13010395GA), Not Applicable (SEE10270254GA), Not Applicable (SEE12061999GA), Not Applicable (SEE12062001GA), Not Applicable (SEE12062010GA), Not Applicable (SEE12082018GA), Not Applicable (SEE12082019GA), Not Applicable (SEE12082021GA), Not Applicable (SEE12082023GA), Not Applicable (SEE12092055GA), Not Applicable (SEE12092059GA), Not Applicable (SEE12102101GA), Not Applicable (SEE13280903GA), Not Applicable (SEE16420074SA), Not Applicable (SEE13190792GA), Not Applicable (SEE14302281GA), Not Applicable (SEE15322281SA), Not Applicable (SEE15322282SA), Not Applicable (SEE15322283SA), Not Applicable (SEE15322284SA), Not Applicable (SEE15322285SA), Not Applicable (SEE15332291SA), Not Applicable (SEE15332292SA), Not Applicable (SEE15332293SA), Not Applicable (SEE15332294SA), Not Applicable (SEE15332295SA), 00840682109260SEE18420044SA (SEE18420044SA), 00840682109260SEE18420047SA (SEE18420047SA), 00840682109260SEE18420053SA (SEE18420053SA), 00840682109260SEE18500011SA (SEE18500011SA), 00840682109260SEE18500034SA (SEE18500034SA), 00840682109260SEE18510014SA (SEE18510014SA), Not Applicable (SEE12102092GA), Not Applicable (SEE13451308GA), Not Applicable (SEE13451322GA), Not Applicable (SEE12332443GA), 00840682109260SEE17510009SA (SEE17510009SA), 00840682109260SEE17400003SA (SEE17400003SA), 00840682109260SEE18080002SA (SEE18080002SA), 00840682109260SEE18080015SA (SEE18080015SA), Not Applicable (SEE10530867GA), Not Applicable (SEE11161317GA), Not Applicable (SEE11171385GA), Not Applicable (SEE13010404GA), Not Applicable (SEE13120613GA), Not Applicable (SEE13270877GA), 00840682109260SEE17500024SA (SEE17500024SA), Not Applicable (SEE13451335GA), Not Applicable (SEE13451345GA), Not Applicable (SEE13140653GA), Not Applicable (SEE13180722GA), Not Applicable (SEE13180743GA), Not Applicable (SEE13180744GA), Not Applicable (SEE13180745GA), Not Applicable (SEE13180746GA), Not Applicable (SEE13180747GA), Not Applicable (SEE13180748GA), Not Applicable (SEE13180749GA), Not Applicable (SEE13180750GA), Not Applicable (SEE13180754GA), Not Applicable (SEE13180755GA), Not Applicable (SEE13190811GA), Not Applicable (SEE15517819SA), Not Applicable (SEE15517820SA), 00840682109260SEE17480010SA (SEE17480010SA), 00840682109260SEE18060009SA (SEE18060009SA), 00840682109260SEE18060010SA (SEE18060010SA), 00840682109260SEE18060012SA (SEE18060012SA), 00840682109260SEE18060013SA (SEE18060013SA), 00840682109260SEE18060014SA (SEE18060014SA), 00840682109260SEE18060016SA (SEE18060016SA), 00840682109260SEE18060018SA (SEE18060018SA), 00840682109260SEE18080003SA (SEE18080003SA), 00840682109260SEE18080005SA (SEE18080005SA), 00840682109260SEE18080006SA (SEE18080006SA), 00840682109260SEE18080009SA (SEE18080009SA), 00840682109260SEE18080010SA (SEE18080010SA), 00840682109260SEE18080013SA (SEE18080013SA), 00840682109260SEE18350002SA (SEE18350002SA), 00840682109260SEE18400017SA (SEE18400017SA), Not Applicable (SEE14151821GA), Not Applicable (SEE14482689GA), Not Applicable (SEE14482701GA), Not Applicable (SEE14211981GA), Not Applicable (SEE14211987GA), Not Applicable (SEE15342446SA), Not Applicable (SEE09250319GA), Not Applicable (SEE09390544GA), Not Applicable (SEE11141277GA), Not Applicable (SEE11141280GA), Not Applicable (SEE11141285GA), Not Applicable (SEE11161326GA), Not Applicable (SEE11371754GA), Not Applicable (SEE11371796GA), Not Applicable (SEE11371801GA), Not Applicable (SEE11371802GA), Not Applicable (SEE11371803GA), Not Applicable (SEE12011893GA), Not Applicable (SEE12011899GA), Not Applicable (SEE16430062SA), 00840682109260SEE17090021SA (SEE17090021SA), 00840682109260SEE17090022SA (SEE17090022SA), 00840682109260SEE17090023SA (SEE17090023SA), 00840682109260SEE17090025SA (SEE17090025SA), 00840682109260SEE17090028SA (SEE17090028SA), Not Applicable (SEE14041730GA), Not Applicable (SEE09460654GA), Not Applicable (SEE09470671GA), Not Applicable (SEE13110566GA), Not Applicable (SEE14272143GA), Not Applicable (SEE14292237GA), Not Applicable (SEE15062988GA), Not Applicable (SEE15063018GA), Not Applicable (SEE15332306SA), Not Applicable (SEE15332308SA), Not Applicable (SEE15373422SA), Not Applicable (SEE16150492SA), Not Applicable (SEE16150497SA), Not Applicable (SEE16150498SA), Not Applicable (SEE16150500SA), Not Applicable (SEE16150501SA), Not Applicable (SEE16150503SA), Not Applicable (SEE16150525SA), Not Applicable (SEE16150526SA), Not Applicable (SEE16150527SA), Not Applicable (SEE16150529SA), Not Applicable (SEE16150530SA), Not Applicable (SEE16150533SA), Not Applicable (SEE16160734SA), 00840682109260SEE17250020SA (SEE17250020SA), 00840682109260SEE17270002SA (SEE17270002SA), 00840682109260SEE17270003SA (SEE17270003SA), 00840682109260SEE17270004SA (SEE17270004SA), 00840682109260SEE17270006SA (SEE17270006SA), 00840682109260SEE17270009SA (SEE17270009SA), 00840682109260SEE17270012SA (SEE17270012SA), 00840682109260SEE17490022SA (SEE17490022SA), 00840682109260SEE17490023SA (SEE17490023SA), 00840682109260SEE17490027SA (SEE17490027SA), 00840682109260SEE17490030SA (SEE17490030SA), 00840682109260SEE17490032SA (SEE17490032SA), 00840682109260SEE18040022SA (SEE18040022SA), 00840682109260SEE18040024SA (SEE18040024SA), Not Applicable (SEE11050974GA), Not Applicable (SEE11060947GA), Not Applicable (SEE12021913GA), 00840682109260SEE18500003SA (SEE18500003SA), 00840682109260SEE18500016SA (SEE18500016SA), 00840682109260SEE18500019SA (SEE18500019SA), Not Applicable (SEE16130321SA), Not Applicable (SEE16140436SA), Not Applicable (SEE16191058SA), Not Applicable (SEE16201178SA), Not Applicable (SEE16201181SA), Not Applicable (SEE16272598SA), Not Applicable (SEE13481389GA), Not Applicable (SEE13481393GA), Not Applicable (SEE13481394GA), Not Applicable (SEE13481397GA), Not Applicable (SEE13481398GA), Not Applicable (SEE13481399GA), Not Applicable (SEE10430633GA), 00840682109260SEE18270002SA (SEE18270002SA), 00840682109260SEE18270003SA (SEE18270003SA), Not Applicable (SEE16078924SA), Not Applicable (SEE16089177SA), 00840682109260SEE17500014SA (SEE17500014SA), 00840682109260SEE17500020SA (SEE17500020SA), 00840682109260SEE17500023SA (SEE17500023SA), 00840682109260SEE17510004SA (SEE17510004SA), 00840682109260SEE17510006SA (SEE17510006SA), Not Applicable (SEE14031708GA), Not Applicable (SEE14031715GA), Not Applicable (SEE14031723GA), Not Applicable (SEE12041947GA), Not Applicable (SEE12041952GA), Not Applicable (SEE12041954GA), Not Applicable (SEE12041955GA), Not Applicable (SEE14031592GA), Not Applicable (SEE12112109GA), Not Applicable (SEE12122157GA), Not Applicable (SEE12122162GA), Not Applicable (SEE12162170GA), Not Applicable (SEE13050438GA), Not Applicable (SEE13050443GA), Not Applicable (SEE13451310GA), Not Applicable (SEE13451318GA), Not Applicable (SEE09100103GA), 00840682109260SEE19240020SA (SEE19240020SA), Not Applicable (SEE09140194GA), Not Applicable (SEE09150204GA), Not Applicable (SEE09150210GA), Not Applicable (SEE10460748GA), Not Applicable (SEE10460751GA), Not Applicable (SEE15394716SA), Not Applicable (SEE15394717SA), Not Applicable (SEE15394718SA), Not Applicable (SEE15394719SA), Not Applicable (SEE15394742SA), Not Applicable (SEE15394743SA), Not Applicable (SEE09170223GA), Not Applicable (SEE09230252GA), Not Applicable (SEE09230257GA), Not Applicable (SEE09230261GA), Not Applicable (SEE09240294GA), Not Applicable (SEE09240296GA), Not Applicable (SEE09510796GA), Not Applicable (SEE09510804GA), Not Applicable (SEE10120027GA), Not Applicable (SEE10121025GA), Not Applicable (SEE10130045GA), Not Applicable (SEE10130069GA), Not Applicable (SEE11060948GA), Not Applicable (SEE11060949GA), Not Applicable (SEE11060951GA), Not Applicable (SEE11060954GA), Not Applicable (SEE11060980GA), Not Applicable (SEE11060981GA), Not Applicable (SEE11060982GA), Not Applicable (SEE11060983GA), Not Applicable (SEE11060984GA), Not Applicable (SEE11060985GA), Not Applicable (SEE11060986GA), Not Applicable (SEE11060987GA), Not Applicable (SEE11060989GA), Not Applicable (SEE11060994GA), Not Applicable (SEE11511872GA), Not Applicable (SEE11511876GA), Not Applicable (SEE11511885GA), Not Applicable (SEE11511887GA), Not Applicable (SEE12122132GA), Not Applicable (SEE12182211GA), Not Applicable (SEE12182224GA), Not Applicable (SEE12312403GA), Not Applicable (SEE13290982GA), Not Applicable (SEE14021554GA), Not Applicable (SEE14101786GA), Not Applicable (SEE14101793GA), Not Applicable (SEE14101794GA), Not Applicable (SEE14101795GA), Not Applicable (SEE14111802GA), Not Applicable (SEE14111803GA), Not Applicable (SEE14111806GA), Not Applicable (SEE14222003GA), Not Applicable (SEE15032891GA), Not Applicable (SEE15507658SA), Not Applicable (SEE15507662SA), 00840682109260SEE16241789SA (SEE16241789SA), 00840682109260SEE16241790SA (SEE16241790SA), 00840682109260SEE16241791SA (SEE16241791SA), 00840682109260SEE16241793SA (SEE16241793SA), 00840682109260SEE16241794SA (SEE16241794SA), 00840682109260SEE16241795SA (SEE16241795SA), 00840682109260SEE16241797SA (SEE16241797SA), Not Applicable (SEE16344282SA), Not Applicable (SEE16430028SA), 00840682109260SEE18060008SA (SEE18060008SA), 00840682109260SEE18060011SA (SEE18060011SA), 00840682109260SEE18060015SA (SEE18060015SA), 00840682109260SEE18160021SA (SEE18160021SA), 00840682109260SEE18160031SA (SEE18160031SA), 00840682109260SEE18160048SA (SEE18160048SA), 00840682109260SEE18160058SA (SEE18160058SA), 00840682109260SEE18160064SA (SEE18160064SA), 00840682109260SEE18160068SA (SEE18160068SA), 00840682109260SEE18160069SA (SEE18160069SA), Not Applicable (SEE09310398GA), Not Applicable (SEE09360492GA), Not Applicable (SEE09370498GA), Not Applicable (SEE09370501GA), Not Applicable (SEE09370503GA), Not Applicable (SEE09370504GA), Not Applicable (SEE09370506GA), Not Applicable (SEE09370509GA), Not Applicable (SEE09380519GA), Not Applicable (SEE09380528GA), Not Applicable (SEE09380529GA), Not Applicable (SEE09390538GA), Not Applicable (SEE10400461GA), Not Applicable (SEE10410473GA), Not Applicable (SEE10410487GA), Not Applicable (SEE10410489GA), Not Applicable (SEE10410491GA), Not Applicable (SEE10410494GA), Not Applicable (SEE10410514GA), Not Applicable (SEE10420536GA), Not Applicable (SEE10420537GA), Not Applicable (SEE10420538GA), Not Applicable (SEE10420550GA), Not Applicable (SEE12041953GA), Not Applicable (SEE12082037GA), Not Applicable (SEE12262330GA), Not Applicable (SEE12262331GA), Not Applicable (SEE12262332GA), Not Applicable (SEE12262333GA), Not Applicable (SEE12262335GA), Not Applicable (SEE12262336GA), Not Applicable (SEE12262337GA), Not Applicable (SEE12262343GA), Not Applicable (SEE12062002GA), Not Applicable (SEE13351019GA), Not Applicable (SEE14031661GA), 00840682109260SEE18460012SA (SEE18460012SA), Not Applicable (SEE11071059GA), Not Applicable (SEE13180753GA), Not Applicable (SEE13220847GA), Not Applicable (SEE13220848GA), Not Applicable (SEE13270872GA), Not Applicable (SEE14272201GA), Not Applicable (SEE16272271SA), Not Applicable (SEE16440004SA), 00840682109260SEE17430024SA (SEE17430024SA), 00840682109260SEE17220007SA (SEE17220007SA), Not Applicable (SEE10450678GA), Not Applicable (SEE16252024SA), Not Applicable (SEE16252037SA), Not Applicable (SEE16262084SA), Not Applicable (SEE10250127GA), Not Applicable (SEE13040422GA), Not Applicable (SEE13100457GA), Not Applicable (SEE13100458GA), Not Applicable (SEE13100459GA), Not Applicable (SEE13100460GA), Not Applicable (SEE13100461GA), Not Applicable (SEE13100462GA), Not Applicable (SEE13100463GA), Not Applicable (SEE13100464GA), Not Applicable (SEE13100465GA), Not Applicable (SEE13100466GA), Not Applicable (SEE13100479GA), Not Applicable (SEE13100480GA), Not Applicable (SEE13100481GA), Not Applicable (SEE13100482GA), Not Applicable (SEE13100483GA), Not Applicable (SEE13100485GA), Not Applicable (SEE13100486GA), Not Applicable (SEE13100487GA), Not Applicable (SEE13100488GA), Not Applicable (SEE13110491GA), Not Applicable (SEE13110492GA), Not Applicable (SEE14302282GA), Not Applicable (SEE15517776SA), Not Applicable (SEE15517818SA), Not Applicable (SEE10410470GA), Not Applicable (SEE10410485GA), 00840682109260SEE18030009SA (SEE18030009SA), 00840682109260SEE19110006SA (SEE19110006SA), 00840682109260SEE17350004SA (SEE17350004SA), Not Applicable (SEE14021579GA), Not Applicable (SEE14021581GA), Not Applicable (SEE14021583GA), Not Applicable (SEE14021585GA), Not Applicable (SEE14031588GA), Not Applicable (SEE14031591GA), 00840682109260SEE18100036SA (SEE18100036SA), Not Applicable (SEE11511865GA), Not Applicable (SEE11511868GA), Not Applicable (SEE11511870GA), Not Applicable (SEE11511878GA), Not Applicable (SEE09130172GA), Not Applicable (SEE09150206GA), Not Applicable (SEE10420593GA), Not Applicable (SEE10460715GA), Not Applicable (SEE12082017GA), Not Applicable (SEE15446202SA), Not Applicable (SEE15446234SA), Not Applicable (SEE15446245SA), Not Applicable (SEE15446255SA), Not Applicable (SEE15446260SA), Not Applicable (SEE15446262SA), Not Applicable (SEE15446331SA), Not Applicable (SEE15528218SA), 00840682109260SEE17150010SA (SEE17150010SA), 00840682109260SEE18500029SA (SEE18500029SA), 00840682109260SEE18500032SA (SEE18500032SA), 00840682109260SEE18500039SA (SEE18500039SA), 00840682109260SEE18500040SA (SEE18500040SA), 00840682109260SEE18500045SA (SEE18500045SA), Not Applicable (SEE16180910SA), Not Applicable (SEE16262145SA), Not Applicable (SEE12430207GA), 00840682109260SEE17290003SA (SEE17290003SA), 00840682109260SEE17290004SA (SEE17290004SA), 00840682109260SEE17290005SA (SEE17290005SA), 00840682109260SEE17290006SA (SEE17290006SA), 00840682109260SEE17290007SA (SEE17290007SA), 00840682109260SEE17300001SA (SEE17300001SA), 00840682109260SEE17300002SA (SEE17300002SA), 00840682109260SEE17300003SA (SEE17300003SA), 00840682109260SEE17300008SA (SEE17300008SA), 00840682109260SEE17300010SA (SEE17300010SA), 00840682109260SEE17300011SA (SEE17300011SA), 00840682109260SEE17300012SA (SEE17300012SA), 00840682109260SEE17300013SA (SEE17300013SA), 00840682109260SEE17300014SA (SEE17300014SA), Not Applicable (SEE13371108GA), 00840682109260SEE17490021SA (SEE17490021SA), Not Applicable (SEE13180777GA), Not Applicable (SEE14322309GA), 00840682109260SEE17430015SA (SEE17430015SA), 00840682109260SEE17430017SA (SEE17430017SA), 00840682109260SEE17430018SA (SEE17430018SA), 00840682109260SEE17430019SA (SEE17430019SA), 00840682109260SEE17430021SA (SEE17430021SA), 00840682109260SEE17430025SA (SEE17430025SA), 00840682109260SEE17430026SA (SEE17430026SA), 00840682109260SEE17430027SA (SEE17430027SA), 00840682109260SEE17430028SA (SEE17430028SA), 00840682109260SEE17430031SA (SEE17430031SA), 00840682109260SEE17430032SA (SEE17430032SA), 00840682109260SEE17430033SA (SEE17430033SA), 00840682109260SEE17430034SA (SEE17430034SA), 00840682109260SEE17430036SA (SEE17430036SA), 00840682109260SEE17430037SA (SEE17430037SA), 00840682109260SEE17430039SA (SEE17430039SA), 00840682109260SEE18160044SA (SEE18160044SA), 00840682109260SEE18160050SA (SEE18160050SA), 00840682109260SEE18160070SA (SEE18160070SA), Not Applicable (SEE12410138GA), Not Applicable (SEE12410144GA), Not Applicable (SEE12410147GA), Not Applicable (SEE12410148GA), Not Applicable (SEE09140181GA), Not Applicable (SEE09250322GA), Not Applicable (SEE09250328GA), Not Applicable (SEE09260337GA), Not Applicable (SEE16282785SA), Not Applicable (SEE16292871SA), 00840682109260SEE18160061SA (SEE18160061SA), 00840682109260SEE19180003SA (SEE19180003SA), Not Applicable (SEE09510764GA), Not Applicable (SEE09510803GA), Not Applicable (SEE10400403GA), Not Applicable (SEE10400440GA), Not Applicable (SEE10400455GA), Not Applicable (SEE10400462GA), Not Applicable (SEE10400463GA), Not Applicable (SEE10410472GA), Not Applicable (SEE10410479GA), Not Applicable (SEE10410492GA), Not Applicable (SEE10510811GA), Not Applicable (SEE11171376GA), Not Applicable (SEE11181391GA), Not Applicable (SEE11211509GA), Not Applicable (SEE11231525GA), Not Applicable (SEE12061998GA), Not Applicable (SEE15062984GA), Not Applicable (SEE15063013GA), 00840682109260SEE17160001SA (SEE17160001SA), 00840682109260SEE17160002SA (SEE17160002SA), 00840682109260SEE17440015SA (SEE17440015SA), 00840682109260SEE17440016SA (SEE17440016SA), 00840682109260SEE17440017SA (SEE17440017SA), 00840682109260SEE17440018SA (SEE17440018SA), 00840682109260SEE17440019SA (SEE17440019SA), Not Applicable (SEE10390356GA), 00840682109260SEE18450018SA (SEE18450018SA), 00840682109260SEE18450020SA (SEE18450020SA), 00840682109260SEE18450061SA (SEE18450061SA), 00840682109260SEE18450069SA (SEE18450069SA), 00840682109260SEE18450073SA (SEE18450073SA), 00840682109260SEE18460001SA (SEE18460001SA), 00840682109260SEE18460002SA (SEE18460002SA), Not Applicable (SEE09100099GA), Not Applicable (SEE09100112GA), Not Applicable (SEE09100137GA), Not Applicable (SEE09530846GA), Not Applicable (SEE10430626GA), 00840682109260SEE18370008SA (SEE18370008SA), 00840682109260SEE19160002SA (SEE19160002SA), 00840682109260SEE19170004SA (SEE19170004SA), Not Applicable (SEE11060963GA), Not Applicable (SEE11071146GA), 00840682109260SEE17380012SA (SEE17380012SA), Not Applicable (SEE10100980GA), 00840682109260SEE17050011SA (SEE17050011SA), Not Applicable (SEE15517769SA), Not Applicable (SEE14512825GA), Not Applicable (SEE11361743GA), Not Applicable (SEE11371775GA), Not Applicable (SEE11371806GA), Not Applicable (SEE11371813GA), Not Applicable (SEE11381816GA), Not Applicable (SEE11381817GA), Not Applicable (SEE11381818GA), Not Applicable (SEE11381820GA), Not Applicable (SEE11381822GA), Not Applicable (SEE11381823GA), Not Applicable (SEE11381825GA), Not Applicable (SEE11381828GA), Not Applicable (SEE11381830GA), Not Applicable (SEE11381831GA), Not Applicable (SEE12122163GA), Not Applicable (SEE13120586GA), Not Applicable (SEE13351037GA), Not Applicable (SEE13351038GA), Not Applicable (SEE13351039GA), Not Applicable (SEE13351041GA), Not Applicable (SEE13351047GA), Not Applicable (SEE13351052GA), Not Applicable (SEE14101790GA), Not Applicable (SEE14222027GA), Not Applicable (SEE14222047GA), Not Applicable (SEE14222054GA), Not Applicable (SEE14222055GA), Not Applicable (SEE16180909SA), Not Applicable (SEE16180924SA), Not Applicable (SEE16190968SA), Not Applicable (SEE16190970SA), Not Applicable (SEE16364421SA), 00840682109260SEE17150019SA (SEE17150019SA), 00840682109260SEE18330039SA (SEE18330039SA), 00840682109260SEE18330040SA (SEE18330040SA), 00840682109260SEE18370009SA (SEE18370009SA), 00840682109260SEE18370014SA (SEE18370014SA), 00840682109260SEE19100005SA (SEE19100005SA), 00840682109260SEE19100006SA (SEE19100006SA), 00840682109260SEE19100008SA (SEE19100008SA), 00840682109260SEE19100021SA (SEE19100021SA), Not Applicable (SEE16130323SA), Not Applicable (SEE16140329SA), Not Applicable (SEE09040066GA), Not Applicable (SEE10470796GA), Not Applicable (SEE13481403GA), Not Applicable (SEE13481410GA), Not Applicable (SEE13481413GA), Not Applicable (SEE13481419GA), 00840682109260SEE19120006SA (SEE19120006SA), Not Applicable (SEE10410503GA), Not Applicable (SEE14021507GA), Not Applicable (SEE14021509GA), Not Applicable (SEE14021515GA), Not Applicable (SEE14021516GA), Not Applicable (SEE15052934GA), 00840682109260SEE18510007SA (SEE18510007SA), Not Applicable (SEE13501456GA), Not Applicable (SEE13501483GA), Not Applicable (SEE13501485GA), Not Applicable (SEE13501493GA), Not Applicable (SEE13501498GA), Not Applicable (SEE09410553GA), Not Applicable (SEE12430216GA), Not Applicable (SEE12430220GA), Not Applicable (SEE15507704SA), Not Applicable (SEE14362367GA), Not Applicable (SEE14362372GA), Not Applicable (SEE14362373GA), Not Applicable (SEE14362377GA), Not Applicable (SEE13290969GA), Not Applicable (SEE13290972GA), Not Applicable (SEE13290975GA), Not Applicable (SEE14031600GA), Not Applicable (SEE16334089SA), 00840682109260SEE19100007SA (SEE19100007SA), 00840682109260SEE19100043SA (SEE19100043SA), 00840682109260SEE19100044SA (SEE19100044SA), 00840682109260SEE19100055SA (SEE19100055SA), 00840682109260SEE19100056SA (SEE19100056SA), 00840682109260SEE18030011SA (SEE18030011SA), Not Applicable (SEE09470669GA), Not Applicable (SEE10400439GA), Not Applicable (SEE10410498GA), Not Applicable (SEE10420520GA), Not Applicable (SEE10420532GA), Not Applicable (SEE10420562GA), Not Applicable (SEE10420580GA), Not Applicable (SEE10420582GA), Not Applicable (SEE10420602GA), Not Applicable (SEE10430641GA), Not Applicable (SEE10460722GA), Not Applicable (SEE10460726GA), Not Applicable (SEE10460739GA), Not Applicable (SEE10460763GA), Not Applicable (SEE10470799GA), Not Applicable (SEE10530815GA), Not Applicable (SEE10530816GA), Not Applicable (SEE10530818GA), Not Applicable (SEE10530819GA), Not Applicable (SEE10530820GA), Not Applicable (SEE10530821GA), Not Applicable (SEE10530822GA), Not Applicable (SEE10530823GA), Not Applicable (SEE10530826GA), Not Applicable (SEE10530830GA), Not Applicable (SEE10530835GA), Not Applicable (SEE12011895GA), Not Applicable (SEE15394284SA), Not Applicable (SEE15467026SA), 00840682109260SEE17440010SA (SEE17440010SA), 00840682109260SEE17440012SA (SEE17440012SA), 00840682109260SEE19170031SA (SEE19170031SA), 00840682109260SEE19180008SA (SEE19180008SA), 00840682109260SEE19220005SA (SEE19220005SA), Not Applicable (SEE09530862GA), Not Applicable (SEE13190805GA), Not Applicable (SEE15507701SA), Not Applicable (SEE12031925GA), Not Applicable (SEE14432610GA), Not Applicable (SEE16140432SA), Not Applicable (SEE16140434SA), Not Applicable (SEE12430194GA), Not Applicable (SEE12430198GA), Not Applicable (SEE12490271GA), Not Applicable (SEE16440007SA), Not Applicable (SEE10130040GA), Not Applicable (SEE10130049GA), Not Applicable (SEE10130051GA), Not Applicable (SEE10130056GA), Not Applicable (SEE10130059GA), Not Applicable (SEE10130061GA), Not Applicable (SEE10130062GA).Not Applicable (SEE10130063GA), Not Applicable (SEE10130067GA), Not Applicable (SEE10130068GA), 00840682109260SEE18100031SA (SEE18100031SA), 00840682109260SEE18160040SA (SEE18160040SA), 00840682109260SEE12390073GA (SEE12390073GA), Not Applicable (SEE13290965GA), Not Applicable (SEE13290966GA), Not Applicable (SEE13290968GA), Not Applicable (SEE13290976GA), Not Applicable (SEE13290979GA), Not Applicable (SEE13290987GA), Not Applicable (SEE13290988GA), Not Applicable (SEE13290989GA), Not Applicable (SEE13290990GA), Not Applicable (SEE13311003GA), Not Applicable (SEE13311007GA), Not Applicable (SEE13311009GA), Not Applicable (SEE13421224GA), Not Applicable (SEE13421225GA), Not Applicable (SEE13421226GA), Not Applicable (SEE13421229GA), Not Applicable (SEE13421230GA), Not Applicable (SEE13421236GA), Not Applicable (SEE13421240GA), Not Applicable (SEE13421241GA), Not Applicable (SEE13421243GA), Not Applicable (SEE13421251GA), Not Applicable (SEE13431261GA), Not Applicable (SEE13431263GA), Not Applicable (SEE13431265GA), Not Applicable (SEE13431267GA), Not Applicable (SEE13431277GA), Not Applicable (SEE13431278GA), Not Applicable (SEE13431280GA), Not Applicable (SEE13431281GA), Not Applicable (SEE13431282GA), Not Applicable (SEE13431284GA), Not Applicable (SEE14031704GA), 00840682109260SEE14031724GA (SEE14031724GA), 00840682109260SEE14031725GA (SEE14031725GA), Not Applicable (SEE14041735GA), Not Applicable (SEE15093037GA), Not Applicable (SEE15093040GA), Not Applicable (SEE15093050GA), 00840682109260SEE17250021SA (SEE17250021SA), 00840682109260SEE18510006SA (SEE18510006SA), Not Applicable (SEE09230256GA), Not Applicable (SEE09230258GA), Not Applicable (SEE09240297GA), Not Applicable (SEE09240301GA), Not Applicable (SEE09250312GA), Not Applicable (SEE11381826GA), Not Applicable (SEE15383876SA), Not Applicable (SEE09380516GA), Not Applicable (SEE09380523GA), Not Applicable (SEE09380527GA), Not Applicable (SEE09510790GA), Not Applicable (SEE15446254SA), Not Applicable (SEE15093074GA), Not Applicable (SEE15103077GA), Not Applicable (SEE15103094GA), Not Applicable (SEE16344327SA), Not Applicable (SEE10390329GA), 00840682109260SEE18500001SA (SEE18500001SA), 00840682109260SEE18500010SA (SEE18500010SA), 00840682109260SEE18500012SA (SEE18500012SA), 00840682109260SEE18500017SA (SEE18500017SA), 00840682109260SEE18500021SA (SEE18500021SA), Not Applicable (SEE14452631GA), Not Applicable (SEE12510302GA), Not Applicable (SEE15467025SA), Not Applicable (SEE16109691SA), Not Applicable (SEE16109693SA), Not Applicable (SEE16109694SA), Not Applicable (SEE16109695SA), Not Applicable (SEE14272136GA), Not Applicable (SEE14272140GA), Not Applicable (SEE14272144GA), Not Applicable (SEE11351668GA), Not Applicable (SEE11361674GA), Not Applicable (SEE11361677GA), Not Applicable (SEE11361679GA), Not Applicable (SEE11361691GA), Not Applicable (SEE11361694GA), Not Applicable (SEE12252311GA), 00840682109260SEE18060003SA (SEE18060003SA), Not Applicable (SEE09510817GA), Not Applicable (SEE09510824GA), Not Applicable (SEE10121026GA), Not Applicable (SEE10121027GA), Not Applicable (SEE10121029GA), Not Applicable (SEE10130070GA), Not Applicable (SEE10390305GA), Not Applicable (SEE10390306GA), Not Applicable (SEE10390334GA), Not Applicable (SEE15093062GA), Not Applicable (SEE15507659SA), Not Applicable (SEE15507663SA), Not Applicable (SEE16150528SA), Not Applicable (SEE16150545SA), Not Applicable (SEE16150554SA), Not Applicable (SEE16150556SA), Not Applicable (SEE16150557SA), Not Applicable (SEE16150569SA), 00840682109260SEE19220006SA (SEE19220006SA), Not Applicable (SEE14101782GA), Not Applicable (SEE11341641GA), Not Applicable (SEE11341644GA), Not Applicable (SEE11341647GA), Not Applicable (SEE11341648GA), Not Applicable (SEE11341649GA), Not Applicable (SEE11341650GA), Not Applicable (SEE11341652GA), Not Applicable (SEE11361715GA), Not Applicable (SEE11361717GA), Not Applicable (SEE11361719GA), Not Applicable (SEE11361728GA), Not Applicable (SEE11361731GA), Not Applicable (SEE13220828GA), Not Applicable (SEE13220829GA), Not Applicable (SEE13220831GA), Not Applicable (SEE13220833GA), Not Applicable (SEE13220836GA), Not Applicable (SEE14362352GA), Not Applicable (SEE14362359GA), Not Applicable (SEE14362360GA), Not Applicable (SEE14362361GA), Not Applicable (SEE14362362GA), Not Applicable (SEE14362363GA), Not Applicable (SEE14362364GA), Not Applicable (SEE14362368GA), Not Applicable (SEE14432590GA), Not Applicable (SEE15012845GA), Not Applicable (SEE16252023SA), Not Applicable (SEE16292877SA), Not Applicable (SEE16303213SA), 00840682109260SEE17010001SA (SEE17010001SA), 00840682109260SEE17010002SA (SEE17010002SA), 00840682109260SEE17010003SA (SEE17010003SA), 00840682109260SEE17010004SA (SEE17010004SA), 00840682109260SEE17010005SA (SEE17010005SA), 00840682109260SEE17010006SA (SEE17010006SA), 00840682109260SEE17010007SA (SEE17010007SA), 00840682109260SEE17010008SA (SEE17010008SA), 00840682109260SEE17010009SA (SEE17010009SA), 00840682109260SEE17020025SA (SEE17020025SA), 00840682109260SEE17260006SA (SEE17260006SA), 00840682109260SEE18500023SA (SEE18500023SA), 00840682109260SEE18500025SA (SEE18500025SA), 00840682109260SEE18500026SA (SEE18500026SA), 00840682109260SEE18500028SA (SEE18500028SA), 00840682109260SEE19160044SA (SEE19160044SA), 00840682109260SEE19160045SA (SEE19160045SA), Not Applicable (SEE16360020SA), Not Applicable (SEE12282350GA), Not Applicable (SEE12282360GA), Not Applicable (SEE12292361GA), Not Applicable (SEE12292364GA), Not Applicable (SEE15352545SA), Not Applicable (SEE12510332GA), Not Applicable (SEE12510334GA), Not Applicable (SEE12520341GA), Not Applicable (SEE12520350GA), Not Applicable (SEE12530370GA), Not Applicable (SEE13000383GA), Not Applicable (SEE13010398GA), Not Applicable (SEE13020411GA), Not Applicable (SEE13020412GA), Not Applicable (SEE13020414GA), Not Applicable (SEE16252015SA), Not Applicable (SEE13110511GA), Not Applicable (SEE13110519GA), Not Applicable (SEE15405017SA), Not Applicable (SEE15405060SA), Not Applicable (SEE16068747SA), Not Applicable (SEE09510820GA), Not Applicable (SEE12082033GA), Not Applicable (SEE13481409GA), Not Applicable (SEE13481411GA), Not Applicable (SEE13481412GA), Not Applicable (SEE13481414GA), Not Applicable (SEE13481420GA), 00840682109260SEE17220008SA (SEE17220008SA), 00840682109260SEE17220009SA (SEE17220009SA), 00840682109260SEE17220011SA (SEE17220011SA), 00840682109260SEE17110001SA (SEE17110001SA), 00840682109260SEE19170030SA (SEE19170030SA), Not Applicable (SEE10450662GA), Not Applicable (SEE10460703GA), Not Applicable (SEE13000380GA), Not Applicable (SEE13010396GA), Not Applicable (SEE13010397GA), Not Applicable (SEE10460727GA), Not Applicable (SEE10460746GA), Not Applicable (SEE10460767GA), Not Applicable (SEE11040902GA), Not Applicable (SEE11040904GA), Not Applicable (SEE11040906GA), Not Applicable (SEE11040912GA), Not Applicable (SEE11040913GA), Not Applicable (SEE11040918GA), Not Applicable (SEE12390086GA), Not Applicable (SEE12400100GA), Not Applicable (SEE14021512GA), Not Applicable (SEE14021518GA), Not Applicable (SEE13130621GA), Not Applicable (SEE13130638GA), Not Applicable (SEE11071031GA), Not Applicable (SEE11071099GA), Not Applicable (SEE12162183GA), Not Applicable (SEE12510303GA), Not Applicable (SEE12520349GA), Not Applicable (SEE13110510GA), Not Applicable (SEE13110529GA), Not Applicable (SEE13110531GA), Not Applicable (SEE13110533GA), Not Applicable (SEE13110536GA), Not Applicable (SEE13110537GA), Not Applicable (SEE13110539GA), Not Applicable (SEE13110544GA), Not Applicable (SEE13110546GA), Not Applicable (SEE13110562GA), Not Applicable (SEE13110565GA), Not Applicable (SEE13120577GA), Not Applicable (SEE13120578GA), Not Applicable (SEE14422528GA), Not Applicable (SEE14422530GA), Not Applicable (SEE14422531GA), Not Applicable (SEE14422533GA), Not Applicable (SEE16282651SA), Not Applicable (SEE16282653SA), Not Applicable (SEE16282788SA), Not Applicable (SEE16282790SA), Not Applicable (SEE16282791SA), 00840682109260SEE17130009SA (SEE17130009SA), 00840682109260SEE17130014SA (SEE17130014SA), 00840682109260SEE17130017SA (SEE17130017SA), 00840682109260SEE17130018SA (SEE17130018SA), 00840682109260SEE17130019SA (SEE17130019SA), 00840682109260SEE17430029SA (SEE17430029SA), 00840682109260SEE18400005SA (SEE18400005SA), 00840682109260SEE18400008SA (SEE18400008SA), 00840682109260SEE18400012SA (SEE18400012SA), 00840682109260SEE18400013SA (SEE18400013SA), 00840682109260SEE18400015SA (SEE18400015SA), 00840682109260SEE18400016SA (SEE18400016SA), Not Applicable (SEE09040079GA), Not Applicable (SEE09240300GA), Not Applicable (SEE09260341GA), Not Applicable (SEE09300357GA), Not Applicable (SEE09310358GA), Not Applicable (SEE09310362GA), Not Applicable (SEE09310364GA), Not Applicable (SEE09310366GA), Not Applicable (SEE09310369GA), Not Applicable (SEE09310370GA), Not Applicable (SEE09310373GA), Not Applicable (SEE09310377GA), Not Applicable (SEE09350466GA), Not Applicable (SEE09410569GA), Not Applicable (SEE09410572GA), Not Applicable (SEE09430601GA), Not Applicable (SEE09530884GA), Not Applicable (SEE10250082GA), Not Applicable (SEE10250088GA), Not Applicable (SEE10250092GA), Not Applicable (SEE10250096GA), Not Applicable (SEE11071015GA), Not Applicable (SEE11071025GA), Not Applicable (SEE11071027GA), Not Applicable (SEE11071028GA), Not Applicable (SEE11081210GA), Not Applicable (SEE11171363GA), Not Applicable (SEE11181404GA), Not Applicable (SEE11191429GA), Not Applicable (SEE11371782GA), Not Applicable (SEE15312135SA), Not Applicable (SEE15517771SA), Not Applicable (SEE16068741SA), Not Applicable (SEE16068742SA), Not Applicable (SEE16068743SA), Not Applicable (SEE16068745SA), Not Applicable (SEE16068749SA), Not Applicable (SEE16252026SA), Not Applicable (SEE16262132SA), Not Applicable (SEE16262134SA), Not Applicable (SEE16262137SA), Not Applicable (SEE16262138SA), Not Applicable (SEE16262140SA), Not Applicable (SEE16262142SA), Not Applicable (SEE16262146SA), Not Applicable (SEE16262149SA), Not Applicable (SEE16262151SA), Not Applicable (SEE16262161SA), Not Applicable (SEE16293180SA), Not Applicable (SEE16313375SA), Not Applicable (SEE16313376SA), Not Applicable (SEE16313377SA), Not Applicable (SEE16313379SA), Not Applicable (SEE16313380SA), Not Applicable (SEE16383418GA), Not Applicable (SEE16390013SA), 00840682109260SEE17330004SA (SEE17330004SA), 00840682109260SEE17330005SA (SEE17330005SA), 00840682109260SEE17330006SA (SEE17330006SA), 00840682109260SEE17330007SA (SEE17330007SA), 00840682109260SEE17330008SA (SEE17330008SA), 00840682109260SEE17330009SA (SEE17330009SA), 00840682109260SEE17330010SA (SEE17330010SA), 00840682109260SEE17360004SA (SEE17360004SA), 00840682109260SEE17360005SA (SEE17360005SA), 00840682109260SEE18050003SA (SEE18050003SA), 00840682109260SEE18050004SA (SEE18050004SA), 00840682109260SEE18250007SA (SEE18250007SA), 00840682109260SEE18250008SA (SEE18250008SA), 00840682109260SEE18250010SA (SEE18250010SA), 00840682109260SEE18250011SA (SEE18250011SA), 00840682109260SEE18250015SA (SEE18250015SA), 00840682109260SEE18250016SA (SEE18250016SA), 00840682109260SEE18250017SA (SEE18250017SA), 00840682109260SEE18250018SA (SEE18250018SA), 00840682109260SEE18250019SA (SEE18250019SA), 00840682109260SEE18250020SA (SEE18250020SA), 00840682109260SEE18250021SA (SEE18250021SA), 00840682109260SEE18250023SA (SEE18250023SA), 00840682109260SEE18250024SA (SEE18250024SA), 00840682109260SEE18250027SA (SEE18250027SA), 00840682109260SEE18250029SA (SEE18250029SA), 00840682109260SEE18250030SA (SEE18250030SA), 00840682109260SEE18250032SA (SEE18250032SA), 00840682109260SEE18250033SA (SEE18250033SA), 00840682109260SEE18250034SA (SEE18250034SA), Not Applicable (SEE15477111SA), Not Applicable (SEE15477183SA), 00840682109260SEE17180011SA (SEE17180011SA), 00840682109260SEE17180012SA (SEE17180012SA), 00840682109260SEE17180014SA (SEE17180014SA), Not Applicable (SEE13180785GA), Not Applicable (SEE13180786GA), Not Applicable (SEE09360483GA), Not Applicable (SEE09370505GA), Not Applicable (SEE09370507GA), Not Applicable (SEE12082016GA), Not Applicable (SEE13361058GA), Not Applicable (SEE13361065GA), Not Applicable (SEE14502804GA), Not Applicable (SEE14502806GA), Not Applicable (SEE11111228GA), Not Applicable (SEE11111233GA), Not Applicable (SEE11111239GA), Not Applicable (SEE11111247GA), Not Applicable (SEE15153186GA), Not Applicable (SEE15153188GA), Not Applicable (SEE15153198GA), Not Applicable (SEE15153201GA), Not Applicable (SEE15153202GA), Not Applicable (SEE15153205GA), Not Applicable (SEE16360025SA), 00840682109260SEE17080015SA (SEE17080015SA), 00840682109260SEE18450023SA (SEE18450023SA), 00840682109260SEE18450070SA (SEE18450070SA), 00840682109260SEE18450071SA (SEE18450071SA), Not Applicable (SEE16410007SA), Not Applicable (SEE16410015SA), 00840682109260SEE17010010SA (SEE17010010SA), Not Applicable (SEE12242287GA), Not Applicable (SEE12252317GA), Not Applicable (SEE12252318GA), Not Applicable (SEE12262326GA), Not Applicable (SEE12262345GA), Not Applicable (SEE13110515GA), Not Applicable (SEE13401143GA), Not Applicable (SEE13401150GA), Not Applicable (SEE14242092GA), Not Applicable (SEE14242100GA), Not Applicable (SEE14242108GA), Not Applicable (SEE14242113GA), Not Applicable (SEE14242114GA), Not Applicable (SEE15012848GA), Not Applicable (SEE15251299SA), Not Applicable (SEE15251301SA), Not Applicable (SEE16180900SA), 00840682109260SEE17270005SA (SEE17270005SA), 00840682109260SEE17270013SA (SEE17270013SA), 00840682109260SEE17270015SA (SEE17270015SA), 00840682109260SEE17410005SA (SEE17410005SA), 00840682109260SEE18040012SA (SEE18040012SA), 00840682109260SEE18100006SA (SEE18100006SA), 00840682109260SEE18100014SA (SEE18100014SA), 00840682109260SEE19090007SA (SEE19090007SA), Not Applicable (SEE10390291GA), Not Applicable (SEE10390343GA), Not Applicable (SEE10400392GA), Not Applicable (SEE10400406GA), Not Applicable (SEE10400421GA), Not Applicable (SEE11181389GA), Not Applicable (SEE11191441GA), Not Applicable (SEE15093070GA), Not Applicable (SEE15093073GA), 00840682109260SEE18040008SA (SEE18040008SA), 00840682109260SEE18040009SA (SEE18040009SA), 00840682109260SEE18040010SA (SEE18040010SA), 00840682109260SEE18040017SA (SEE18040017SA), 00840682109260SEE18040023SA (SEE18040023SA), 00840682109260SEE18040027SA (SEE18040027SA), 00840682109260SEE18040047SA (SEE18040047SA), Not Applicable (SEE15415184SA), Not Applicable (SEE15415186SA), Not Applicable (SEE15425741SA), Not Applicable (SEE15425782SA), Not Applicable (SEE09300352GA), Not Applicable (SEE09300356GA), Not Applicable (SEE10120026GA), Not Applicable (SEE10121009GA), Not Applicable (SEE10470803GA), Not Applicable (SEE10510810GA), Not Applicable (SEE11050940GA), Not Applicable (SEE11050944GA), Not Applicable (SEE11060996GA), Not Applicable (SEE11511867GA), Not Applicable (SEE11511874GA), Not Applicable (SEE13351020GA), Not Applicable (SEE13351021GA), Not Applicable (SEE13351033GA), Not Applicable (SEE13351034GA), Not Applicable (SEE13411195GA), Not Applicable (SEE13411207GA), Not Applicable (SEE14031666GA), Not Applicable (SEE14031692GA), Not Applicable (SEE14372405GA), Not Applicable (SEE14372410GA), Not Applicable (SEE14372414GA), Not Applicable (SEE14372415GA), Not Applicable (SEE14372422GA), Not Applicable (SEE14382459GA), Not Applicable (SEE14472673GA), Not Applicable (SEE15301887SA), Not Applicable (SEE15301889SA), Not Applicable (SEE15394375SA), Not Applicable (SEE16221692SA), Not Applicable (SEE16221693SA), Not Applicable (SEE16221697SA), Not Applicable (SEE16221698SA), Not Applicable (SEE16221699SA), 00840682109260SEE16520005SA (SEE16520005SA), 00840682109260SEE17350009SA (SEE17350009SA), 00840682109260SEE19240025SA (SEE19240025SA), 00840682109260SEE19240026SA (SEE19240026SA), Not Applicable (SEE09150203GA), Not Applicable (SEE09230287GA), Not Applicable (SEE09240290GA), Not Applicable (SEE12092085GA), Not Applicable (SEE12182206GA), Not Applicable (SEE09020035GA), Not Applicable (SEE09260340GA), Not Applicable (SEE10430639GA), Not Applicable (SEE10460706GA), Not Applicable (SEE12182231GA), Not Applicable (SEE14041732GA), Not Applicable (SEE15425786SA), Not Applicable (SEE15425787SA), Not Applicable (SEE15425788SA), Not Applicable (SEE15425789SA), Not Applicable (SEE15425790SA), Not Applicable (SEE15425792SA), Not Applicable (SEE15425794SA), Not Applicable (SEE15425795SA), Not Applicable (SEE15425801SA), Not Applicable (SEE09100086GA), Not Applicable (SEE09100110GA), Not Applicable (SEE09190227GA), Not Applicable (SEE09190229GA), Not Applicable (SEE09370510GA), Not Applicable (SEE09380518GA), Not Applicable (SEE09380526GA), Not Applicable (SEE09380530GA), Not Applicable (SEE09430599GA), Not Applicable (SEE09430606GA), Not Applicable (SEE09460650GA), Not Applicable (SEE09510811GA), Not Applicable (SEE09510822GA), Not Applicable (SEE09510833GA), Not Applicable (SEE09530866GA), Not Applicable (SEE10080925GA), Not Applicable (SEE10080927GA), Not Applicable (SEE10080928GA), Not Applicable (SEE10090948GA), Not Applicable (SEE10250106GA), Not Applicable (SEE10270272GA), Not Applicable (SEE10270273GA), Not Applicable (SEE10400376GA), Not Applicable (SEE10400377GA), Not Applicable (SEE10400399GA), Not Applicable (SEE10400416GA), Not Applicable (SEE10400417GA), Not Applicable (SEE10420560GA), Not Applicable (SEE11040905GA), Not Applicable (SEE11040908GA), Not Applicable (SEE11040909GA), Not Applicable (SEE11040910GA), Not Applicable (SEE11040914GA), Not Applicable (SEE11060957GA), Not Applicable (SEE11060962GA), 00840682109260SEE11171353GA (SEE11171353GA), Not Applicable (SEE11171368GA), Not Applicable (SEE11171372GA), Not Applicable (SEE11181407GA), Not Applicable (SEE11321578GA), Not Applicable (SEE11321583GA), Not Applicable (SEE12041949GA), Not Applicable (SEE12041958GA), Not Applicable (SEE12041959GA), Not Applicable (SEE12041960GA), Not Applicable (SEE12041961GA), Not Applicable (SEE12041962GA), Not Applicable (SEE12041963GA), Not Applicable (SEE12041964GA), Not Applicable (SEE12041965GA), Not Applicable (SEE12041967GA), Not Applicable (SEE12041968GA), Not Applicable (SEE12041969GA), Not Applicable (SEE12041970GA), Not Applicable (SEE12041971GA), Not Applicable (SEE12041972GA), Not Applicable (SEE12041974GA), Not Applicable (SEE12390083GA), Not Applicable (SEE12490277GA), Not Applicable (SEE12520347GA), Not Applicable (SEE12530367GA), Not Applicable (SEE12530368GA), Not Applicable (SEE13050430GA), Not Applicable (SEE13180728GA), Not Applicable (SEE13180729GA), Not Applicable (SEE13280939GA), Not Applicable (SEE13290961GA), Not Applicable (SEE13290970GA), Not Applicable (SEE13290978GA), Not Applicable (SEE13451346GA), Not Applicable (SEE13451349GA), Not Applicable (SEE14031667GA), Not Applicable (SEE14031669GA), Not Applicable (SEE14412503GA), Not Applicable (SEE14412506GA), Not Applicable (SEE14422524GA), Not Applicable (SEE14422526GA), Not Applicable (SEE14422539GA), Not Applicable (SEE14422552GA), Not Applicable (SEE14482695GA), Not Applicable (SEE15063000GA), Not Applicable (SEE15113143GA), Not Applicable (SEE15517772SA), Not Applicable (SEE15517817SA), Not Applicable (SEE16078930SA), Not Applicable (SEE16078932SA), Not Applicable (SEE16109579SA), Not Applicable (SEE16109580SA), Not Applicable (SEE16109583SA), Not Applicable (SEE16109584SA), Not Applicable (SEE16109594SA), Not Applicable (SEE16233391GA), Not Applicable (SEE16303405GA), 00840682109260SEE17460006SA (SEE17460006SA), 00840682109260SEE17460014SA (SEE17460014SA), 00840682109260SEE17460018SA (SEE17460018SA), 00840682109260SEE17490017SA (SEE17490017SA), 00840682109260SEE17490019SA (SEE17490019SA), 00840682109260SEE17490024SA (SEE17490024SA), 00840682109260SEE17490025SA (SEE17490025SA), 00840682109260SEE17490026SA (SEE17490026SA), 00840682109260SEE17490028SA (SEE17490028SA), 00840682109260SEE17490029SA (SEE17490029SA), 00840682109260SEE17490031SA (SEE17490031SA), 00840682109260SEE17490033SA (SEE17490033SA), 00840682109260SEE17490034SA (SEE17490034SA), 00840682109260SEE17490035SA (SEE17490035SA), 00840682109260SEE18160001SA (SEE18160001SA), 00840682109260SEE18160024SA (SEE18160024SA), 00840682109260SEE18160027SA (SEE18160027SA), 00840682109260SEE18330045SA (SEE18330045SA), 00840682109260SEE18330046SA (SEE18330046SA), 00840682109260SEE18330047SA (SEE18330047SA), 00840682109260SEE18330048SA (SEE18330048SA), 00840682109260SEE18370011SA (SEE18370011SA), 00840682109260SEE18370012SA (SEE18370012SA), 00840682109260SEE18450084SA (SEE18450084SA), 00840682109260SEE19090009SA (SEE19090009SA), 00840682109260SEE19090022SA (SEE19090022SA), 00840682109260SEE19130008SA (SEE19130008SA), 00840682109260SEE19170020SA (SEE19170020SA), 00840682109260SEE19370038SA (SEE19370038SA), 00840682109260SEE19370058SA (SEE19370058SA), 00840682109260SEE19370059SA (SEE19370059SA), 00840682109260SEE11060999GA (SEE11060999GA), 00840682109260SEE11071008GA (SEE11071008GA), 00840682109260SEE11071019GA (SEE11071019GA), 00840682109260SEE11071023GA (SEE11071023GA), 00840682109260SEE13481415GA (SEE13481415GA), 00840682109260SEE13491425GA (SEE13491425GA), 00840682109260SEE16058655SA (SEE16058655SA), 00840682109260SEE16058660SA (SEE16058660SA), Not Applicable (SEE13361069GA), Not Applicable (SEE13361075GA), Not Applicable (SEE13371083GA), Not Applicable (SEE13371085GA), Not Applicable (SEE15022859GA), Not Applicable (SEE16303224SA), Not Applicable (SEE16313378SA), Not Applicable (SEE16450002SA), Not Applicable (SEE16450004SA), Not Applicable (SEE16450009SA), Not Applicable (SEE12292362GA), Not Applicable (SEE12302373GA), Not Applicable (SEE10460740GA), Not Applicable (SEE10460761GA), Not Applicable (SEE10460766GA), Not Applicable (SEE10460772GA), Not Applicable (SEE10510809GA), Not Applicable (SEE10510812GA), Not Applicable (SEE11371774GA), Not Applicable (SEE12021904GA), Not Applicable (SEE12021918GA), Not Applicable (SEE12062013GA), Not Applicable (SEE14222051GA), Not Applicable (SEE14222063GA), Not Applicable (SEE14222064GA), Not Applicable (SEE14242104GA), Not Applicable (SEE14242105GA), 00840682109260SEE18240018SA (SEE18240018SA), Not Applicable (SEE09350457GA), Not Applicable (SEE09350460GA), Not Applicable (SEE09350467GA), Not Applicable (SEE09360473GA), Not Applicable (SEE11050975GA), Not Applicable (SEE11060958GA), Not Applicable (SEE11060978GA), Not Applicable (SEE14432588GA), Not Applicable (SEE14432591GA), Not Applicable (SEE14432603GA), Not Applicable (SEE14432604GA), Not Applicable (SEE14432605GA), Not Applicable (SEE14432606GA), Not Applicable (SEE15093047GA), Not Applicable (SEE15093052GA), Not Applicable (SEE15352795SA), Not Applicable (SEE15352796SA), Not Applicable (SEE15352846SA), 00840682109260SEE16520002SA (SEE16520002SA), 00840682109260SEE16520004SA (SEE16520004SA), Not Applicable (SEE10410506GA), Not Applicable (SEE10460762GA), Not Applicable (SEE10460769GA), 00840682109260SEE18500038SA (SEE18500038SA), Not Applicable (SEE12302376GA), Not Applicable (SEE12302381GA), Not Applicable (SEE12302392GA), Not Applicable (SEE12302393GA), Not Applicable (SEE12302394GA), Not Applicable (SEE14452615GA), Not Applicable (SEE14452616GA), Not Applicable (SEE14452630GA), Not Applicable (SEE14452638GA), Not Applicable (SEE15012854GA), Not Applicable (SEE15012855GA), Not Applicable (SEE11191437GA), Not Applicable (SEE11201459GA), Not Applicable (SEE11211489GA), Not Applicable (SEE11211505GA), Not Applicable (SEE11211511GA), Not Applicable (SEE11211513GA), Not Applicable (SEE11211517GA), Not Applicable (SEE11211523GA), Not Applicable (SEE11361681GA), Not Applicable (SEE11361686GA), Not Applicable (SEE11361692GA), Not Applicable (SEE12410142GA), Not Applicable (SEE11071030GA), Not Applicable (SEE11071052GA), Not Applicable (SEE11141263GA), Not Applicable (SEE11141265GA), Not Applicable (SEE11141276GA), Not Applicable (SEE11141284GA), Not Applicable (SEE15281641SA), Not Applicable (SEE09310368GA), Not Applicable (SEE09310371GA), Not Applicable (SEE09310375GA), Not Applicable (SEE09310382GA), Not Applicable (SEE09310384GA), Not Applicable (SEE09310387GA), Not Applicable (SEE09330412GA), Not Applicable (SEE09330413GA), Not Applicable (SEE09330427GA), Not Applicable (SEE14472674GA), Not Applicable (SEE09530845GA), Not Applicable (SEE13180751GA), Not Applicable (SEE13180752GA), Not Applicable (SEE13180761GA), Not Applicable (SEE13180764GA), Not Applicable (SEE13180773GA), Not Applicable (SEE13180779GA), Not Applicable (SEE13180782GA), Not Applicable (SEE14031721GA), Not Applicable (SEE14081767GA), Not Applicable (SEE14081768GA), Not Applicable (SEE14081772GA), Not Applicable (SEE14081773GA), Not Applicable (SEE14222040GA), Not Applicable (SEE14222056GA), Not Applicable (SEE14242079GA), Not Applicable (SEE14242080GA), Not Applicable (SEE14242085GA), Not Applicable (SEE14242086GA), Not Applicable (SEE14242088GA), Not Applicable (SEE14242109GA), Not Applicable (SEE14242111GA), Not Applicable (SEE11151308GA), Not Applicable (SEE11161325GA), Not Applicable (SEE13130625GA), Not Applicable (SEE13130641GA), Not Applicable (SEE13130643GA), Not Applicable (SEE13130644GA), Not Applicable (SEE13130647GA), Not Applicable (SEE13220826GA), 00840682109260SEE18420018SA (SEE18420018SA), Not Applicable (SEE14452618GA), Not Applicable (SEE14452625GA), 00840682109260SEE17390003SA (SEE17390003SA), 00840682109260SEE18450012SA (SEE18450012SA), 00840682109260SEE18450015SA (SEE18450015SA), Not Applicable (SEE09100085GA), Not Applicable (SEE09100126GA), Not Applicable (SEE09130160GA), Not Applicable (SEE09130165GA), Not Applicable (SEE11181417GA), Not Applicable (SEE13421232GA), Not Applicable (SEE13421235GA), Not Applicable (SEE13421238GA), Not Applicable (SEE13421246GA), Not Applicable (SEE13421247GA), Not Applicable (SEE14222029GA), Not Applicable (SEE14222068GA), Not Applicable (SEE14222076GA), Not Applicable (SEE15373463SA), Not Applicable (SEE15373464SA), Not Applicable (SEE15373465SA), Not Applicable (SEE15373466SA), Not Applicable (SEE15373467SA), Not Applicable (SEE15373531SA), Not Applicable (SEE15373532SA), 00840682109260SEE17460011SA (SEE17460011SA), 00840682109260SEE18360002SA (SEE18360002SA), 00840682109260SEE18360003SA (SEE18360003SA), 00840682109260SEE18360008SA (SEE18360008SA), Not Applicable (SEE14031617GA), Not Applicable (SEE14031623GA), Not Applicable (SEE14031631GA), Not Applicable (SEE14031634GA), Not Applicable (SEE14031642GA), Not Applicable (SEE15456718SA), Not Applicable (SEE15456720SA), Not Applicable (SEE15487295SA), Not Applicable (SEE13140654GA), Not Applicable (SEE13140656GA), Not Applicable (SEE13150664GA), Not Applicable (SEE13180724GA), Not Applicable (SEE16364406SA), Not Applicable (SEE16364407SA), 00840682109260SEE17440005SA (SEE17440005SA), 00840682109260SEE17440009SA (SEE17440009SA), 00840682109260SEE17440011SA (SEE17440011SA), 00840682109260SEE17440013SA (SEE17440013SA), 00840682109260SEE17520001SA (SEE17520001SA), Not Applicable (SEE09100097GA), Not Applicable (SEE09100117GA), Not Applicable (SEE09130158GA), Not Applicable (SEE09190230GA), Not Applicable (SEE09380532GA), Not Applicable (SEE09470686GA), Not Applicable (SEE09470687GA), Not Applicable (SEE09480737GA), Not Applicable (SEE09510763GA), Not Applicable (SEE09510765GA), Not Applicable (SEE09510766GA), Not Applicable (SEE09510772GA), Not Applicable (SEE09510773GA), Not Applicable (SEE09510777GA), Not Applicable (SEE09510778GA), Not Applicable (SEE09510788GA), Not Applicable (SEE09510795GA), Not Applicable (SEE09510806GA), Not Applicable (SEE09510807GA), Not Applicable (SEE09510809GA), Not Applicable (SEE10130053GA), Not Applicable (SEE10250090GA), Not Applicable (SEE10250093GA), Not Applicable (SEE10250108GA), Not Applicable (SEE10250112GA), Not Applicable (SEE10250120GA), Not Applicable (SEE10250161GA), Not Applicable (SEE10390293GA), Not Applicable (SEE10390294GA), Not Applicable (SEE10390308GA), Not Applicable (SEE10390316GA), Not Applicable (SEE10390321GA), Not Applicable (SEE10390322GA), Not Applicable (SEE10420573GA), Not Applicable (SEE10460756GA), Not Applicable (SEE11071054GA), Not Applicable (SEE11071064GA), Not Applicable (SEE11071132GA), Not Applicable (SEE11101224GA), Not Applicable (SEE11161318GA), Not Applicable (SEE11161323GA), Not Applicable (SEE11161331GA), Not Applicable (SEE11211498GA), Not Applicable (SEE11341602GA), Not Applicable (SEE11341604GA), Not Applicable (SEE11341605GA), Not Applicable (SEE11341607GA), Not Applicable (SEE11341608GA), Not Applicable (SEE11341610GA), Not Applicable (SEE11341611GA), Not Applicable (SEE11341613GA), Not Applicable (SEE11341617GA), Not Applicable (SEE11341618GA), Not Applicable (SEE11341621GA), Not Applicable (SEE11341623GA), Not Applicable (SEE11341637GA), Not Applicable (SEE11341639GA), Not Applicable (SEE11341654GA), Not Applicable (SEE11361689GA), Not Applicable (SEE11371786GA), Not Applicable (SEE11371799GA), Not Applicable (SEE11381829GA), Not Applicable (SEE12031943GA), Not Applicable (SEE12031944GA), Not Applicable (SEE12182221GA), Not Applicable (SEE12510325GA), Not Applicable (SEE13280929GA), Not Applicable (SEE13280933GA), Not Applicable (SEE13280940GA), Not Applicable (SEE13290952GA), Not Applicable (SEE13351015GA), Not Applicable (SEE13351026GA), Not Applicable (SEE13401152GA), Not Applicable (SEE13401166GA), Not Applicable (SEE13401167GA), Not Applicable (SEE13401177GA), Not Applicable (SEE14021576GA), Not Applicable (SEE14322336GA), Not Applicable (SEE14362366GA), Not Applicable (SEE14362370GA), Not Applicable (SEE14362371GA), Not Applicable (SEE14362376GA), Not Applicable (SEE14402473GA), Not Applicable (SEE14402474GA), Not Applicable (SEE14402475GA), Not Applicable (SEE14402476GA), Not Applicable (SEE14402477GA), Not Applicable (SEE14402478GA), Not Applicable (SEE14402479GA), Not Applicable (SEE14402480GA), Not Applicable (SEE14402482GA), Not Applicable (SEE14402483GA), Not Applicable (SEE14402487GA), Not Applicable (SEE14482684GA), Not Applicable (SEE14482700GA), Not Applicable (SEE14482703GA), Not Applicable (SEE14482705GA), Not Applicable (SEE14482707GA), Not Applicable (SEE14482708GA), Not Applicable (SEE14482709GA), Not Applicable (SEE14482710GA), Not Applicable (SEE14482711GA), Not Applicable (SEE14482712GA), Not Applicable (SEE14482713GA), Not Applicable (SEE14482714GA), Not Applicable (SEE14482716GA), Not Applicable (SEE14482721GA), Not Applicable (SEE15022861GA), Not Applicable (SEE15022867GA), Not Applicable (SEE15042901GA), Not Applicable (SEE16099458SA), Not Applicable (SEE16160725SA), Not Applicable (SEE16191065SA), Not Applicable (SEE16262164SA), Not Applicable (SEE16360014SA), Not Applicable (SEE16360017SA), Not Applicable (SEE16364418SA), Not Applicable (SEE16364420SA), Not Applicable (SEE16390024SA), Not Applicable (SEE16410011SA), Not Applicable (SEE16430032SA), Not Applicable (SEE16430035SA), Not Applicable (SEE16430037SA), Not Applicable (SEE16430052SA), 00840682109260SEE17020021SA (SEE17020021SA), 00840682109260SEE17310006SA (SEE17310006SA), 00840682109260SEE17310007SA (SEE17310007SA), 00840682109260SEE17310008SA (SEE17310008SA), 00840682109260SEE17310009SA (SEE17310009SA), 00840682109260SEE17310011SA (SEE17310011SA), 00840682109260SEE17310013SA (SEE17310013SA), 00840682109260SEE17310014SA (SEE17310014SA), 00840682109260SEE17310016SA (SEE17310016SA), 00840682109260SEE17310017SA (SEE17310017SA), 00840682109260SEE17340016SA (SEE17340016SA), 00840682109260SEE17380008SA (SEE17380008SA), 00840682109260SEE17410001SA (SEE17410001SA), 00840682109260SEE17410002SA (SEE17410002SA), 00840682109260SEE17430023SA (SEE17430023SA), 00840682109260SEE17430035SA (SEE17430035SA), 00840682109260SEE18170012SA (SEE18170012SA), 00840682109260SEE18450002SA (SEE18450002SA), 00840682109260SEE18450007SA (SEE18450007SA), 00840682109260SEE18450014SA (SEE18450014SA), 00840682109260SEE19100003SA (SEE19100003SA), 00840682109260SEE19100004SA (SEE19100004SA), 00840682109260SEE19100009SA (SEE19100009SA), 00840682109260SEE19100035SA (SEE19100035SA), 00840682109260SEE19220029SA (SEE19220029SA), 00840682109260SEE19220033SA (SEE19220033SA), Not Applicable (SEE13120593GA), Not Applicable (SEE13120607GA), Not Applicable (SEE13120609GA), Not Applicable (SEE13120610GA), Not Applicable (SEE13120611GA), Not Applicable (SEE13120614GA), Not Applicable (SEE16380031SA), Not Applicable (SEE16068744SA), Not Applicable (SEE10420517GA), Not Applicable (SEE10420519GA), Not Applicable (SEE09410586GA), Not Applicable (SEE11050930GA), Not Applicable (SEE11201452GA), Not Applicable (SEE11201465GA), Not Applicable (SEE11201468GA), Not Applicable (SEE11201471GA), Not Applicable (SEE14031596GA), Not Applicable (SEE14031691GA), Not Applicable (SEE14031695GA), Not Applicable (SEE14031696GA), Not Applicable (SEE14031697GA), Not Applicable (SEE14221998GA), Not Applicable (SEE15261530SA), Not Applicable (SEE15261537SA), Not Applicable (SEE16038553SA), Not Applicable (SEE16038554SA), Not Applicable (SEE16038556SA), Not Applicable (SEE16038557SA), Not Applicable (SEE16390018SA), Not Applicable (SEE16390020SA), 00840682109260SEE18100005SA (SEE18100005SA), 00840682109260SEE18360016SA (SEE18360016SA), Not Applicable (SEE11321576GA), Not Applicable (SEE11321580GA), Not Applicable (SEE11331591GA), Not Applicable (SEE11331593GA), Not Applicable (SEE11331597GA), Not Applicable (SEE09410568GA), Not Applicable (SEE09410571GA), Not Applicable (SEE09410580GA), Not Applicable (SEE14031641GA), Not Applicable (SEE14031674GA), Not Applicable (SEE14031683GA), Not Applicable (SEE14031686GA), Not Applicable (SEE14031703GA), Not Applicable (SEE16221558SA), Not Applicable (SEE16221559SA), Not Applicable (SEE16430005SA), Not Applicable (SEE16430007SA), Not Applicable (SEE16430021SA), 00840682109260SEE17480012SA (SEE17480012SA), 00840682109260SEE17480014SA (SEE17480014SA), 00840682109260SEE19130019SA (SEE19130019SA), 00840682109260SEE19150004SA (SEE19150004SA), 00840682109260SEE19150005SA (SEE19150005SA), 00840682109260SEE19170028SA (SEE19170028SA), 00840682109260SEE19170029SA (SEE19170029SA), Not Applicable (SEE10420531GA), Not Applicable (SEE12082026GA), Not Applicable (SEE12082027GA), Not Applicable (SEE12082038GA), Not Applicable (SEE12082039GA), Not Applicable (SEE12082040GA), Not Applicable (SEE12092050GA), Not Applicable (SEE12520342GA), 00840682109260SEE18330044SA (SEE18330044SA), 00840682109260SEE18330050SA (SEE18330050SA), Not Applicable (SEE11171360GA), Not Applicable (SEE11211515GA), Not Applicable (SEE15446240SA), Not Applicable (SEE15446334SA), Not Applicable (SEE16109582SA), Not Applicable (SEE16119901SA), Not Applicable (SEE16119921SA), Not Applicable (SEE11071026GA), Not Applicable (SEE09310380GA), Not Applicable (SEE09310383GA), Not Applicable (SEE10420598GA), Not Applicable (SEE10430640GA), Not Applicable (SEE10450658GA), Not Applicable (SEE10460704GA), Not Applicable (SEE10460707GA), Not Applicable (SEE10460716GA), Not Applicable (SEE10460730GA), Not Applicable (SEE10460736GA), Not Applicable (SEE10460754GA), Not Applicable (SEE11071020GA), Not Applicable (SEE11071021GA), Not Applicable (SEE11071057GA), Not Applicable (SEE12182233GA), Not Applicable (SEE14272218GA), Not Applicable (SEE15251286SA), Not Applicable (SEE15251287SA), Not Applicable (SEE09110156GA), Not Applicable (SEE09170222GA), Not Applicable (SEE09190233GA), Not Applicable (SEE14031632GA), Not Applicable (SEE14031635GA), Not Applicable (SEE14031636GA), Not Applicable (SEE14031638GA), Not Applicable (SEE14031639GA), Not Applicable (SEE14031643GA), Not Applicable (SEE14031647GA), Not Applicable (SEE14031649GA), Not Applicable (SEE14031650GA), Not Applicable (SEE14422542GA), Not Applicable (SEE15193268GA), Not Applicable (SEE15193271GA), Not Applicable (SEE13220827GA), 00840682109260SEE17430030SA (SEE17430030SA), Not Applicable (SEE12430215GA), Not Applicable (SEE12430218GA), Not Applicable (SEE16383417GA), Not Applicable (SEE14272174GA), Not Applicable (SEE14272199GA), Not Applicable (SEE13371084GA), Not Applicable (SEE09020043GA), Not Applicable (SEE09100094GA), Not Applicable (SEE09100101GA), Not Applicable (SEE09100124GA), Not Applicable (SEE09100125GA), Not Applicable (SEE14322319GA), Not Applicable (SEE14322326GA), Not Applicable (SEE14322330GA), Not Applicable (SEE14322331GA), Not Applicable (SEE14322333GA), Not Applicable (SEE14322334GA), 00840682109260SEE15415112SA (SEE15415112SA), 00840682109260SEE15415113SA (SEE15415113SA), 00840682109260SEE15415118SA (SEE15415118SA), 00840682109260SEE18170008SA (SEE18170008SA), Not Applicable (SEE10460695GA), 00840682109260SEE17500016SA (SEE17500016SA), Not Applicable (SEE09340429GA), Not Applicable (SEE09340439GA), Not Applicable (SEE09340445GA), Not Applicable (SEE09340447GA), Not Applicable (SEE11201457GA), Not Applicable (SEE11201466GA), Not Applicable (SEE12370044GA), Not Applicable (SEE12370046GA), Not Applicable (SEE12370047GA), Not Applicable (SEE12370050GA), Not Applicable (SEE12370051GA), Not Applicable (SEE12370052GA), Not Applicable (SEE12370054GA), Not Applicable (SEE12370055GA), Not Applicable (SEE14322322GA), Not Applicable (SEE15173209GA), Not Applicable (SEE15173210GA), Not Applicable (SEE15173216GA), Not Applicable (SEE15173217GA), Not Applicable (SEE15173220GA), Not Applicable (SEE15173221GA), Not Applicable (SEE15173222GA), Not Applicable (SEE15173223GA), Not Applicable (SEE15173224GA), Not Applicable (SEE16109692SA), SEE16410010SA (SEE16410010SA), 00840682109260SEE19090005SA (SEE19090005SA), 00840682109260SEE19090006SA (SEE19090006SA), 00840682109260SEE19090034SA (SEE19090034SA), Not Applicable (SEE16241827SA), Not Applicable (SEE16241828SA), Not Applicable (SEE16241831SA), Not Applicable (SEE16241832SA), Not Applicable (SEE16241833SA), Not Applicable (SEE16241845SA), Not Applicable (SEE16241848SA), Not Applicable (SEE16241849SA), Not Applicable (SEE16241850SA), Not Applicable (SEE16241853SA), Not Applicable (SEE16241854SA), Not Applicable (SEE16241857SA), Not Applicable (SEE16241858SA), Not Applicable (SEE16241859SA), Not Applicable (SEE16251979SA), Not Applicable (SEE16251981SA), Not Applicable (SEE16251982SA), Not Applicable (SEE16251983SA), Not Applicable (SEE16251984SA), Not Applicable (SEE16252017SA), Not Applicable (SEE11151306GA), Not Applicable (SEE11151307GA), Not Applicable (SEE11151314GA), Not Applicable (SEE11161322GA), Not Applicable (SEE15466929SA), Not Applicable (SEE15466930SA), Not Applicable (SEE15466931SA), Not Applicable (SEE15466933SA), Not Applicable (SEE15466936SA), 00840682109260SEE17100001SA (SEE17100001SA), Not Applicable (SEE13431276GA), Not Applicable (SEE13431302GA), Not Applicable (SEE13441306GA), Not Applicable (SEE13451368GA), Not Applicable (SEE13451375GA), Not Applicable (SEE14041727GA), 00840682109260SEE19100050SA (SEE19100050SA), Not Applicable (SEE09530888GA), Not Applicable (SEE10400386GA), Not Applicable (SEE10400393GA), Not Applicable (SEE10400394GA), Not Applicable (SEE10400395GA), Not Applicable (SEE10400398GA), Not Applicable (SEE10400405GA), Not Applicable (SEE10400407GA), Not Applicable (SEE10400408GA), Not Applicable (SEE10400410GA), Not Applicable (SEE10400411GA), Not Applicable (SEE10400412GA), Not Applicable (SEE10400413GA), Not Applicable (SEE10400418GA), Not Applicable (SEE10400423GA), Not Applicable (SEE10400431GA), Not Applicable (SEE11141254GA), Not Applicable (SEE11201460GA), Not Applicable (SEE11211483GA), Not Applicable (SEE11211487GA), Not Applicable (SEE11211488GA), Not Applicable (SEE11211493GA), Not Applicable (SEE11211496GA), Not Applicable (SEE11211510GA), Not Applicable (SEE11211521GA), Not Applicable (SEE11231526GA), Not Applicable (SEE11231533GA), Not Applicable (SEE11231534GA), Not Applicable (SEE11231535GA), Not Applicable (SEE11231538GA), Not Applicable (SEE11231539GA), Not Applicable (SEE11231546GA), Not Applicable (SEE11371781GA), Not Applicable (SEE11371784GA), Not Applicable (SEE12420151GA), Not Applicable (SEE12420152GA), Not Applicable (SEE12420153GA), Not Applicable (SEE12420154GA), Not Applicable (SEE12420156GA), Not Applicable (SEE12420157GA), Not Applicable (SEE12420166GA), Not Applicable (SEE12430177GA), Not Applicable (SEE13431258GA), Not Applicable (SEE13431266GA), Not Applicable (SEE14021555GA), Not Applicable (SEE14021558GA), Not Applicable (SEE14021561GA), Not Applicable (SEE14021568GA), Not Applicable (SEE14021573GA), Not Applicable (SEE14101780GA), Not Applicable (SEE14101781GA), Not Applicable (SEE14101785GA), Not Applicable (SEE14101791GA), Not Applicable (SEE14101797GA), Not Applicable (SEE15281659SA), Not Applicable (SEE15446261SA), Not Applicable (SEE15446333SA), Not Applicable (SEE12340004GA), Not Applicable (SEE12342447GA), Not Applicable (SEE12342450GA), Not Applicable (SEE12342454GA), Not Applicable (SEE09510831GA), Not Applicable (SEE09530856GA), Not Applicable (SEE09530864GA), Not Applicable (SEE09530880GA), Not Applicable (SEE09530886GA), Not Applicable (SEE09530889GA), Not Applicable (SEE09530905GA), Not Applicable (SEE09530910GA), Not Applicable (SEE10260196GA), Not Applicable (SEE11381814GA), Not Applicable (SEE09450624GA), Not Applicable (SEE09450625GA), Not Applicable (SEE09450626GA), Not Applicable (SEE09450628GA), Not Applicable (SEE09450630GA), Not Applicable (SEE09510816GA), Not Applicable (SEE12011894GA), Not Applicable (SEE13451367GA), Not Applicable (SEE13451376GA), Not Applicable (SEE14372413GA), Not Applicable (SEE14041746GA), 00840682109260SEE18070001SA (SEE18070001SA), Not Applicable (SEE11071038GA), Not Applicable (SEE11071041GA), Not Applicable (SEE11071044GA), Not Applicable (SEE11151297GA), 00840682109260SEE18320007SA (SEE18320007SA), 00840682109260SEE19100061SA (SEE19100061SA), Not Applicable (SEE11060952GA), Not Applicable (SEE11060988GA), Not Applicable (SEE14512830GA), Not Applicable (SEE14512833GA), Not Applicable (SEE15012841GA), Not Applicable (SEE15012843GA), Not Applicable (SEE15012846GA), Not Applicable (SEE15012849GA), Not Applicable (SEE15012853GA), Not Applicable (SEE15022860GA), Not Applicable (SEE15022863GA), Not Applicable (SEE15022866GA), 00840682109260SEE17390001SA (SEE17390001SA), 00840682109260SEE17390002SA (SEE17390002SA), 00840682109260SEE17390004SA (SEE17390004SA), 00840682109260SEE17390005SA (SEE17390005SA), 00840682109260SEE17390006SA (SEE17390006SA), 00840682109260SEE17390007SA (SEE17390007SA), 00840682109260SEE18160049SA (SEE18160049SA), 00840682109260SEE18160071SA (SEE18160071SA), 00840682109260SEE18170016SA (SEE18170016SA), 00840682109260SEE18190033SA (SEE18190033SA), 00840682109260SEE18330007SA (SEE18330007SA), 00840682109260SEE18330014SA (SEE18330014SA), 00840682109260SEE18330015SA (SEE18330015SA), 00840682109260SEE19220038SA (SEE19220038SA), Not Applicable (SEE09140178GA), Not Applicable (SEE09150219GA), Not Applicable (SEE09510821GA), Not Applicable (SEE11071072GA), Not Applicable (SEE11071077GA), Not Applicable (SEE11071079GA), Not Applicable (SEE11071133GA), Not Applicable (SEE11071151GA), Not Applicable (SEE11081169GA), Not Applicable (SEE11081172GA), Not Applicable (SEE11081211GA), Not Applicable (SEE11141293GA), Not Applicable (SEE12262328GA), Not Applicable (SEE12262329GA), Not Applicable (SEE12302377GA), Not Applicable (SEE12370049GA), Not Applicable (SEE12370061GA), Not Applicable (SEE12370063GA), Not Applicable (SEE12370068GA), Not Applicable (SEE12400101GA), Not Applicable (SEE12400103GA), Not Applicable (SEE12400104GA), Not Applicable (SEE12400105GA), Not Applicable (SEE12400107GA), Not Applicable (SEE12400109GA), Not Applicable (SEE12400110GA), Not Applicable (SEE12400112GA), Not Applicable (SEE12400114GA), Not Applicable (SEE12400116GA), Not Applicable (SEE12400117GA), Not Applicable (SEE12400118GA), Not Applicable (SEE13020410GA), Not Applicable (SEE13060450GA), Not Applicable (SEE13150693GA), Not Applicable (SEE14021537GA), Not Applicable (SEE15352790SA), Not Applicable (SEE15352792SA), Not Applicable (SEE16282707SA), Not Applicable (SEE16360026SA), Not Applicable (SEE16390008SA), Not Applicable (SEE16410008SA), 00840682109260SEE17100005SA (SEE17100005SA), 00840682109260SEE17370002SA (SEE17370002SA), 00840682109260SEE18170026SA (SEE18170026SA), 00840682109260SEE18450075SA (SEE18450075SA), 00840682109260SEE19010008SA (SEE19010008SA), 00840682109260SEE19130010SA (SEE19130010SA), Not Applicable (SEE13150662GA), Not Applicable (SEE13150682GA), Not Applicable (SEE13371082GA), Not Applicable (SEE13371105GA), 00840682109260SEE19090010SA (SEE19090010SA), 00840682109260SEE19090011SA (SEE19090011SA), 00840682109260SEE19090016SA (SEE19090016SA), 00840682109260SEE19090017SA (SEE19090017SA), 00840682109260SEE19090020SA (SEE19090020SA), 00840682109260SEE19090033SA (SEE19090033SA), Not Applicable (SEE14031618GA), Not Applicable (SEE15332340SA), Not Applicable (SEE15332342SA), Not Applicable (SEE11081167GA), Not Applicable (SEE13180763GA), Not Applicable (SEE09100135GA), Not Applicable (SEE15093046GA), Not Applicable (SEE14472672GA), Not Applicable (SEE09510789GA), Not Applicable (SEE12490263GA), Not Applicable (SEE15477107SA), Not Applicable (SEE10430625GA), Not Applicable (SEE10430634GA), Not Applicable (SEE10460741GA), Not Applicable (SEE15394729SA), Not Applicable (SEE09510785GA), Not Applicable (SEE11060993GA), Not Applicable (SEE11061001GA), Not Applicable (SEE12322407GA), Not Applicable (SEE12322410GA), Not Applicable (SEE15261536SA), Not Applicable (SEE16140435SA), Not Applicable (SEE16150494SA), Not Applicable (SEE16150496SA), Not Applicable (SEE16150499SA), Not Applicable (SEE16150502SA), Not Applicable (SEE16272269SA), 00840682109260SEE18320026SA (SEE18320026SA), 00840682109260SEE18320027SA (SEE18320027SA), 00840682109260SEE18320031SA (SEE18320031SA), Not Applicable (SEE16241823SA), Not Applicable (SEE12292363GA), Not Applicable (SEE12390072GA), Not Applicable (SEE12390075GA), Not Applicable (SEE15173208GA), Not Applicable (SEE15173215GA), Not Applicable (SEE16420041SA), Not Applicable (SEE16420042SA), Not Applicable (SEE16420043SA), Not Applicable (SEE16420044SA), Not Applicable (SEE16420045SA), Not Applicable (SEE16420046SA), Not Applicable (SEE16420048SA), Not Applicable (SEE16420049SA), Not Applicable (SEE16420050SA), 00840682109260SEE18240016SA (SEE18240016SA), 00840682109260SEE18320002SA (SEE18320002SA), 00840682109260SEE18320012SA (SEE18320012SA), 00840682109260SEE18320015SA (SEE18320015SA), 00840682109260SEE18320018SA (SEE18320018SA), 00840682109260SEE18320019SA (SEE18320019SA), 00840682109260SEE18320022SA (SEE18320022SA), 00840682109260SEE18320023SA (SEE18320023SA), 00840682109260SEE18330009SA (SEE18330009SA), 00840682109260SEE18330016SA (SEE18330016SA), Not Applicable (SEE10120030GA), Not Applicable (SEE14272147GA), Not Applicable (SEE14272149GA), Not Applicable (SEE14272150GA), Not Applicable (SEE14272151GA), Not Applicable (SEE14272152GA), Not Applicable (SEE14272154GA), Not Applicable (SEE14272155GA), Not Applicable (SEE16272588SA), Not Applicable (SEE16272594SA), Not Applicable (SEE16272595SA), Not Applicable (SEE09190236GA), Not Applicable (SEE09190246GA), Not Applicable (SEE13220846GA), Not Applicable (SEE16120170SA), Not Applicable (SEE14422541GA), Not Applicable (SEE13451309GA), Not Applicable (SEE09460639GA), Not Applicable (SEE09460640GA), Not Applicable (SEE09460643GA), Not Applicable (SEE09460644GA), Not Applicable (SEE09470665GA), Not Applicable (SEE12172195GA), Not Applicable (SEE12530374GA), Not Applicable (SEE13000375GA), Not Applicable (SEE13000379GA), Not Applicable (SEE13000381GA), Not Applicable (SEE13000382GA), Not Applicable (SEE14432574GA), Not Applicable (SEE11211497GA), Not Applicable (SEE14151834GA), Not Applicable (SEE14151866GA), Not Applicable (SEE12340017GA), Not Applicable (SEE12400091GA), Not Applicable (SEE12400092GA), 00840682109260SEE17180005SA (SEE17180005SA), Not Applicable (SEE14171919GA), Not Applicable (SEE12490264GA), Not Applicable (SEE12490269GA), Not Applicable (SEE12490272GA), Not Applicable (SEE12490280GA), Not Applicable (SEE16109590SA), Not Applicable (SEE16119910SA), Not Applicable (SEE16430008SA), Not Applicable (SEE11111240GA), Not Applicable (SEE09470677GA), Not Applicable (SEE09480744GA), Not Applicable (SEE09490755GA), Not Applicable (SEE15103105GA), Not Applicable (SEE15103113GA), Not Applicable (SEE15103115GA), Not Applicable (SEE14242081GA), 00840682109260SEE19240043SA (SEE19240043SA), Not Applicable (SEE10270245GA), Not Applicable (SEE14031678GA), Not Applicable (SEE14031685GA), Not Applicable (SEE14031702GA), 00840682109260SEE18190036SA (SEE18190036SA), 00840682109260SEE18190040SA (SEE18190040SA), Not Applicable (SEE11141260GA), Not Applicable (SEE10420579GA), Not Applicable (SEE10420576GA), Not Applicable (SEE10460719GA), Not Applicable (SEE10460734GA), Not Applicable (SEE16334162SA), Not Applicable (SEE09100127GA), Not Applicable (SEE09230254GA), Not Applicable (SEE09240293GA), Not Applicable (SEE10460724GA), Not Applicable (SEE14031621GA), Not Applicable (SEE10460690GA), Not Applicable (SEE14211979GA), Not Applicable (SEE15032876GA), Not Applicable (SEE09310376GA), Not Applicable (SEE10270263GA), Not Applicable (SEE10400419GA), Not Applicable (SEE10400438GA), Not Applicable (SEE10400446GA), Not Applicable (SEE10410477GA), Not Applicable (SEE10420561GA), Not Applicable (SEE10470789GA), Not Applicable (SEE11071018GA), Not Applicable (SEE11111231GA), Not Applicable (SEE11111241GA), Not Applicable (SEE11511875GA), Not Applicable (SEE12112110GA), Not Applicable (SEE12112116GA), Not Applicable (SEE13240868GA), Not Applicable (SEE13290973GA), Not Applicable (SEE15113128GA), Not Applicable (SEE15113131GA), Not Applicable (SEE15183260GA), Not Applicable (SEE16119844SA), Not Applicable (SEE16282862SA), Not Applicable (SEE15173236GA), Not Applicable (SEE15173243GA), Not Applicable (SEE14512821GA), Not Applicable (SEE09490761GA), Not Applicable (SEE11211520GA), Not Applicable (SEE11231527GA), Not Applicable (SEE13381121GA), Not Applicable (SEE13391133GA), Not Applicable (SEE13391136GA), Not Applicable (SEE16211411SA), Not Applicable (SEE09510792GA), Not Applicable (SEE09510823GA), Not Applicable (SEE12162178GA), Not Applicable (SEE12162180GA), Not Applicable (SEE12162184GA), Not Applicable (SEE13381126GA), Not Applicable (SEE13381127GA), Not Applicable (SEE13391132GA), Not Applicable (SEE13280938GA), Not Applicable (SEE13280942GA), Not Applicable (SEE13280944GA), Not Applicable (SEE13280950GA), Not Applicable (SEE09460648GA), Not Applicable (SEE11161337GA), Not Applicable (SEE10420566GA), Not Applicable (SEE10420599GA), Not Applicable (SEE10430610GA), Not Applicable (SEE10430622GA), Not Applicable (SEE10430647GA), Not Applicable (SEE10450664GA), Not Applicable (SEE11371769GA), Not Applicable (SEE12400094GA), Not Applicable (SEE13451316GA), Not Applicable (SEE13451319GA), Not Applicable (SEE13451330GA), Not Applicable (SEE13451338GA), Not Applicable (SEE13451339GA), Not Applicable (SEE13451340GA), Not Applicable (SEE13451343GA), Not Applicable (SEE13451351GA), Not Applicable (SEE13451352GA), Not Applicable (SEE13451354GA), Not Applicable (SEE13451355GA), Not Applicable (SEE13451356GA), Not Applicable (SEE13451357GA), Not Applicable (SEE13451358GA), Not Applicable (SEE13451359GA), Not Applicable (SEE13451360GA), Not Applicable (SEE13451361GA), Not Applicable (SEE13451362GA), Not Applicable (SEE13451363GA), Not Applicable (SEE13451364GA), Not Applicable (SEE13451365GA), Not Applicable (SEE13451366GA), Not Applicable (SEE14472681GA), 00840682109260SEE19100045SA (SEE19100045SA), 00840682109260SEE19100059SA (SEE19100059SA), Not Applicable (SEE09100093GA), Not Applicable (SEE09100121GA), Not Applicable (SEE12420161GA), Not Applicable (SEE12490288GA), Not Applicable (SEE12490289GA), Not Applicable (SEE12490290GA), Not Applicable (SEE12490293GA), Not Applicable (SEE12490294GA), Not Applicable (SEE12490298GA), Not Applicable (SEE14021506GA), Not Applicable (SEE10390315GA), Not Applicable (SEE10390327GA), Not Applicable (SEE10390332GA), Not Applicable (SEE10390333GA), Not Applicable (SEE12322409GA), Not Applicable (SEE13451328GA), Not Applicable (SEE13451347GA), Not Applicable (SEE13491449GA), Not Applicable (SEE13501462GA), Not Applicable (SEE13501473GA), Not Applicable (SEE13501476GA), Not Applicable (SEE14041756GA), Not Applicable (SEE15032898GA), Not Applicable (SEE15507700SA), Not Applicable (SEE15507703SA), Not Applicable (SEE16078898SA), Not Applicable (SEE12122133GA), Not Applicable (SEE09130162GA), Not Applicable (SEE09130166GA), Not Applicable (SEE09460652GA), Not Applicable (SEE09470683GA), Not Applicable (SEE09480729GA), Not Applicable (SEE09530870GA), Not Applicable (SEE10130060GA), Not Applicable (SEE10250079GA), Not Applicable (SEE10250080GA), Not Applicable (SEE10250097GA), Not Applicable (SEE10250109GA), Not Applicable (SEE10250113GA), Not Applicable (SEE10250114GA), Not Applicable (SEE10250115GA), Not Applicable (SEE10250117GA), Not Applicable (SEE10260194GA), Not Applicable (SEE10260205GA), Not Applicable (SEE10260211GA), Not Applicable (SEE10260235GA), Not Applicable (SEE10270240GA), Not Applicable (SEE10270243GA), Not Applicable (SEE10270247GA), Not Applicable (SEE10310279GA), Not Applicable (SEE10410464GA), Not Applicable (SEE11071065GA), Not Applicable (SEE11171366GA), Not Applicable (SEE11171370GA), Not Applicable (SEE11171375GA), Not Applicable (SEE11171378GA), Not Applicable (SEE11191427GA), Not Applicable (SEE11191430GA), Not Applicable (SEE11341651GA), Not Applicable (SEE11361733GA), Not Applicable (SEE11361736GA), Not Applicable (SEE11361738GA), Not Applicable (SEE11361741GA), Not Applicable (SEE11361744GA), Not Applicable (SEE12122136GA), Not Applicable (SEE12122145GA), Not Applicable (SEE12400106GA), Not Applicable (SEE12420172GA), Not Applicable (SEE13110513GA), Not Applicable (SEE13110514GA), Not Applicable (SEE13110520GA), Not Applicable (SEE13110521GA), Not Applicable (SEE13280937GA), Not Applicable (SEE13421239GA), Not Applicable (SEE13421245GA), Not Applicable (SEE13421248GA), Not Applicable (SEE13421249GA), Not Applicable (SEE13421250GA), Not Applicable (SEE13421252GA), Not Applicable (SEE13501459GA), Not Applicable (SEE13501461GA), Not Applicable (SEE13501466GA), Not Applicable (SEE14041731GA), Not Applicable (SEE14211972GA), Not Applicable (SEE14222036GA), Not Applicable (SEE14292245GA), Not Applicable (SEE14302255GA), Not Applicable (SEE14302259GA), Not Applicable (SEE14302261GA), Not Applicable (SEE14302262GA), Not Applicable (SEE14302264GA), Not Applicable (SEE14302266GA), Not Applicable (SEE14302268GA), Not Applicable (SEE14302269GA), Not Applicable (SEE14302270GA), Not Applicable (SEE14302271GA), Not Applicable (SEE14302272GA), Not Applicable (SEE14302275GA), Not Applicable (SEE14302277GA), Not Applicable (SEE14302278GA), Not Applicable (SEE14302279GA), Not Applicable (SEE14302283GA), Not Applicable (SEE14302284GA), Not Applicable (SEE14302285GA), Not Applicable (SEE14302286GA), Not Applicable (SEE14302287GA), Not Applicable (SEE14372409GA), Not Applicable (SEE14372425GA), Not Applicable (SEE14372427GA), Not Applicable (SEE14372429GA), Not Applicable (SEE14372436GA), Not Applicable (SEE14382463GA), Not Applicable (SEE14382470GA), Not Applicable (SEE14402481GA), Not Applicable (SEE14402484GA), Not Applicable (SEE14402485GA), Not Applicable (SEE14402486GA), Not Applicable (SEE14402488GA), Not Applicable (SEE14402492GA), Not Applicable (SEE14432575GA), Not Applicable (SEE15032880GA), Not Applicable (SEE15220795SA), Not Applicable (SEE15261579SA), Not Applicable (SEE15383875SA), Not Applicable (SEE16109585SA), Not Applicable (SEE16150493SA), Not Applicable (SEE16324002SA), 00840682109260SEE19090019SA (SEE19090019SA), 00840682109260SEE19130006SA (SEE19130006SA), Not Applicable (SEE12062003GA), Not Applicable (SEE16140433SA), 00840682109260SEE15103097GA (SEE15103097GA), 00840682109260SEE15103098GA (SEE15103098GA), 00840682109260SEE15103099GA (SEE15103099GA), 00840682109260SEE15103101GA (SEE15103101GA), 00840682109260SEE15103106GA (SEE15103106GA), 00840682109260SEE15103107GA (SEE15103107GA), 00840682109260SEE15103111GA (SEE15103111GA), 00840682109260SEE15103114GA (SEE15103114GA), Not Applicable (SEE16191064SA), 00840682109260SEE19130012SA (SEE19130012SA), 00840682109260SEE19130014SA (SEE19130014SA), Not Applicable (SEE12322433GA), Not Applicable (SEE12282352GA), Not Applicable (SEE15446226SA), Not Applicable (SEE16450008SA), Not Applicable (SEE16450010SA), Not Applicable (SEE12350027GA), Not Applicable (SEE15332307SA), 00840682109260SEE18330018SA (SEE18330018SA), Not Applicable (SEE09100107GA), Not Applicable (SEE13150674GA), Not Applicable (SEE14031615GA), Not Applicable (SEE14031640GA), Not Applicable (SEE11050945GA), Not Applicable (SEE10420549GA), Not Applicable (SEE14372402GA), Not Applicable (SEE10400445GA), Not Applicable (SEE15312080SA), Not Applicable (SEE11181405GA), Not Applicable (SEE14151810GA), Not Applicable (SEE15032896GA), Not Applicable (SEE10460701GA), Not Applicable (SEE11071087GA), 00840682109260SEE18160029SA (SEE18160029SA), Not Applicable (SEE10250130GA), Not Applicable (SEE09510794GA), Not Applicable (SEE09510814GA), 00840682109260SEE19220027SA (SEE19220027SA), Not Applicable (SEE09490757GA), Not Applicable (SEE11191431GA), Not Applicable (SEE12092072GA), Not Applicable (SEE09040062GA), Not Applicable (SEE09040068GA), Not Applicable (SEE12340014GA), Not Applicable (SEE12342455GA), Not Applicable (SEE12350022GA), Not Applicable (SEE12350025GA), Not Applicable (SEE12350028GA), Not Applicable (SEE12350030GA), Not Applicable (SEE13290977GA), Not Applicable (SEE14272139GA), Not Applicable (SEE15183248GA), Not Applicable (SEE15183252GA), 00840682109260SEE18320030SA (SEE18320030SA), Not Applicable (SEE11040915GA), Not Applicable (SEE10420565GA), Not Applicable (SEE16364419SA), Not Applicable (SEE10250107GA), Not Applicable (SEE10250129GA), Not Applicable (SEE10260200GA), Not Applicable (SEE10260217GA), Not Applicable (SEE10260218GA), Not Applicable (SEE10260221GA), Not Applicable (SEE15332398SA), Not Applicable (SEE16221696SA), 00840682109260SEE17510011SA (SEE17510011SA), 00840682109260SEE18510001SA (SEE18510001SA), 00840682109260SEE18510010SA (SEE18510010SA), Not Applicable (SEE16109690SA), Not Applicable (SEE16109702SA), Not Applicable (SEE14312289GA), Not Applicable (SEE13240867GA), Not Applicable (SEE13240871GA), Not Applicable (SEE14412501GA), Not Applicable (SEE12242292GA), Not Applicable (SEE12252320GA), Not Applicable (SEE12252323GA), Not Applicable (SEE12252325GA), Not Applicable (SEE15173238GA), 00840682109260SEE17210014SA (SEE17210014SA), 00840682109260SEE17290001SA (SEE17290001SA), 00840682109260SEE17290002SA (SEE17290002SA), Not Applicable (SEE12122134GA), Not Applicable (SEE10530850GA), Not Applicable (SEE12390078GA), Not Applicable (SEE11050929GA), Not Applicable (SEE11201446GA), Not Applicable (SEE11211500GA), Not Applicable (SEE09370496GA), Not Applicable (SEE09370500GA), Not Applicable (SEE09370508GA), Not Applicable (SEE14372423GA), Not Applicable (SEE14372426GA), Not Applicable (SEE14372428GA), Not Applicable (SEE14372430GA), Not Applicable (SEE14372433GA), Not Applicable (SEE14372435GA), Not Applicable (SEE14422535GA), Not Applicable (SEE14422569GA), Not Applicable (SEE14432611GA), Not Applicable (SEE14432612GA), Not Applicable (SEE11060995GA), Not Applicable (SEE11141274GA), Not Applicable (SEE11141281GA), Not Applicable (SEE11371751GA), Not Applicable (SEE14482698GA), Not Applicable (SEE10460713GA), Not Applicable (SEE14031717GA), Not Applicable (SEE19110019SA), Not Applicable (SEE19110040SA), Not Applicable (SEE16333410GA), 00840682109260SEE18200013SA (SEE18200013SA), Not Applicable (SEE13290967GA), Not Applicable (SEE15413316GA), 00840682109260SEE17150002SA (SEE17150002SA), Not Applicable (SEE13290983GA), Not Applicable (SEE13290986GA), Not Applicable (SEE13351018GA), Not Applicable (SEE14462651GA), Not Applicable (SEE14462658GA), Not Applicable (SEE16303407GA), Not Applicable (SEE13180775GA), Not Applicable (SEE13180780GA), Not Applicable (SEE13180781GA), Not Applicable (SEE15193282GA), Not Applicable (SEE16420075SA), 00840682109260SEE17210013SA (SEE17210013SA), 00840682109260SEE17310005SA (SEE17310005SA), 00840682109260SEE18500015SA (SEE18500015SA), 00840682109260SEE18160038SA (SEE18160038SA), 00840682109260SEE18160039SA (SEE18160039SA), 00840682109260SEE19170002SA (SEE19170002SA), Not Applicable (SEE15153203GA), Not Applicable (SEE15193272GA), 00840682109260SEE18510028SA (SEE18510028SA), Not Applicable (SEE12122140GA), 00840682109260SEE18160026SA (SEE18160026SA), 00840682109260SEE18160037SA (SEE18160037SA), Not Applicable (SEE14222002GA), 00840682109260SEE17070002SA (SEE17070002SA), 00840682109260SEE18460004SA (SEE18460004SA), 00840682109260SEE18460011SA (SEE18460011SA), Not Applicable (SEE14021562GA), Not Applicable (SEE10420523GA), Not Applicable (SEE12112108GA), Not Applicable (SEE11341656GA), Not Applicable (SEE11341656SA), Not Applicable (SEE10470800GA), Not Applicable (SEE11071134GA), Not Applicable (SEE11371787GA), Not Applicable (SEE11191428GA), Not Applicable (SEE09100122GA), Not Applicable (SEE09190237GA), Not Applicable (SEE12370060GA), Not Applicable (SEE12400097GA), Not Applicable (SEE14151819GA), Not Applicable (SEE10130047GA), Not Applicable (SEE12490243GA), Not Applicable (SEE14021544GA), Not Applicable (SEE12360034GA), Not Applicable (SEE13270898GA), Not Applicable (SEE13280912GA), Not Applicable (SEE16272628SA), Not Applicable (SEE16282787SA), Not Applicable (SEE16282789SA), Not Applicable (SEE15423320GA), Not Applicable (SEE14101783GA), Not Applicable (SEE15183256GA), 00840682109260SEE17150005SA (SEE17150005SA), Not Applicable (SEE09100143GA), Not Applicable (SEE13351031GA), Not Applicable (SEE12430201GA), 00840682109260SEE17210008SA (SEE17210008SA), Not Applicable (SEE13451334GA), 00840682109260SEE17410006SA (SEE17410006SA), Not Applicable (SEE15423321GA), Not Applicable (SEE15423319GA), Not Applicable (SEE15423323GA), Not Applicable (SEE14021545GA), Not Applicable (SEE14021549GA), Not Applicable (SEE14222020GA), Not Applicable (SEE16430050SA), Not Applicable (SEE13140655GA), 00840682109260SEE18160025SA (SEE18160025SA), 00840682109260SEE18160033SA (SEE18160033SA), Not Applicable (SEE15373309GA), Not Applicable (SEE16123366GA), Not Applicable (SEE16420065SA), Not Applicable (SEE14452627GA), Not Applicable (SEE16183379GA), Not Applicable (SEE09530895GA), Not Applicable (SEE09530913GA), Not Applicable (SEE14242103GA), Not Applicable (SEE14252134GA), 00840682109260SEE18160034SA (SEE18160034SA), 00840682109260SEE18220008SA (SEE18220008SA), 00840682109260SEE18220009SA (SEE18220009SA), Not Applicable (SEE14161894GA), Not Applicable (SEE16390004SA), 00840682109260SEE18510005SA (SEE18510005SA), Not Applicable (SEE12092052GA), 00840682109260SEE17160004SA (SEE17160004SA), Not Applicable (SEE13000394GA), Not Applicable (SEE12162182GA), Not Applicable (SEE10420578GA), 00840682109260SEE19220022SA (SEE19220022SA), Not Applicable (SEE15012850GA), Not Applicable (SEE12430205GA), Not Applicable (SEE12490287GA), Not Applicable (SEE13481407GA), Not Applicable (SEE13501464GA), Not Applicable (SEE10400400GA), Not Applicable (SEE16173377GA), Not Applicable (SEE16183378GA), Not Applicable (SEE14502761GA), Not Applicable (SEE13371096GA), Not Applicable (SEE15423328GA), Not Applicable (SEE15423329GA), Not Applicable (SEE15423330GA), Not Applicable (SEE16123367GA), Not Applicable (SEE16173375GA), Not Applicable (SEE09330414GA), Not Applicable (SEE16450003SA), Not Applicable (SEE16173370GA), Not Applicable (SEE15193281GA), Not Applicable (SEE09310361GA), Not Applicable (SEE09310386GA), Not Applicable (SEE09310388GA), Not Applicable (SEE10410471GA), Not Applicable (SEE12082032GA), Not Applicable (SEE12082034GA), Not Applicable (SEE16440003SA), Not Applicable (SEE09140179GA), Not Applicable (SEE13481418GA), Not Applicable (SEE14031619GA), Not Applicable (SEE14302263GA), Not Applicable (SEE14472677GA), Not Applicable (SEE15193277GA), Not Applicable (SEE15193283GA), 00840682109260SEE17060003SA (SEE17060003SA), 00840682109260SEE17400006SA (SEE17400006SA), 00840682109260SEE17500022SA (SEE17500022SA), 00840682109260SEE18190031SA (SEE18190031SA), 00840682109260SEE18380034SA (SEE18380034SA), Not Applicable (SEE16173372GA), 00840682109260SEE18070002SA (SEE18070002SA), 00840682109260SEE18190039SA (SEE18190039SA), 00840682109260SEE18200008SA (SEE18200008SA), 00840682109260SEE18200009SA (SEE18200009SA), 00840682109260SEE18200010SA (SEE18200010SA), 00840682109260SEE18200011SA (SEE18200011SA), 00840682109260SEE18200012SA (SEE18200012SA), Not Applicable (SEE10250174GA), Not Applicable (SEE09370511GA), Not Applicable (SEE12370045GA), Not Applicable (SEE12360038GA), Not Applicable (SEE11361682GA), Not Applicable (SEE11391835GA), Not Applicable (SEE13180790GA), Not Applicable (SEE09190225GA), Not Applicable (SEE11361673GA), Not Applicable (SEE11151300GA), Not Applicable (SEE11511871GA), Not Applicable (SEE12122128GA), Not Applicable (SEE12370062GA), Not Applicable (SEE12182234GA), Not Applicable (SEE13361076GA), Not Applicable (SEE13451369GA), Not Applicable (SEE13451374GA), Not Applicable (SEE14151822GA), Not Applicable (SEE14322325GA), Not Applicable (SEE14332341GA), Not Applicable (SEE14422559GA), Not Applicable (SEE15312143SA), Not Applicable (SEE15473332GA), Not Applicable (SEE15473334GA), Not Applicable (SEE16173369GA), Not Applicable (SEE16450001SA), 00840682109260SEE17110006SA (SEE17110006SA), Not Applicable (SEE18390015SA), 00840682109260SEE18420017SA (SEE18420017SA), Not Applicable (SEE11111245GA), Not Applicable (SEE12192236GA), Not Applicable (SEE13270890GA), Not Applicable (SEE14021542GA), 00840682109260SEE17420003SA (SEE17420003SA), Not Applicable (SEE09340436GA), Not Applicable (SEE10260199GA), Not Applicable (SEE10270267GA), Not Applicable (SEE10420601GA), Not Applicable (SEE10430613GA), Not Applicable (SEE10430643GA), Not Applicable (SEE10460721GA), Not Applicable (SEE10460770GA), Not Applicable (SEE11231528GA), Not Applicable (SEE13020408GA), 00840682109260SEE18170027SA (SEE18170027SA), Not Applicable (SEE12420165GA), Not Applicable (SEE14101779GA), Not Applicable (SEE11371767GA), Not Applicable (SEE14482704GA), Not Applicable (SEE10120998GA), Not Applicable (SEE10130072GA), Not Applicable (SEE12252321GA), Not Applicable (SEE12162181GA), Not Applicable (SEE12282359GA), Not Applicable (SEE13361067GA), Not Applicable (SEE14432576GA), Not Applicable (SEE14432577GA), Not Applicable (SEE14452628GA), Not Applicable (SEE15052946GA), 00840682109260SEE17020030SA (SEE17020030SA), Not Applicable (SEE11201450GA), Not Applicable (SEE11361726GA), Not Applicable (SEE12340006GA), Not Applicable (SEE12342448GA), Not Applicable (SEE13401151GA), Not Applicable (SEE13401172GA), 00840682109260SEE18320017SA (SEE18320017SA), 00840682109260SEE18030019SA (SEE18030019SA), Not Applicable (SEE15113183GA), Not Applicable (SEE13401163GA), Not Applicable (SEE13401178GA), Not Applicable (SEE09460631GA), Not Applicable (SEE10530840GA), Not Applicable (SEE13190791GA), Not Applicable (SEE13451311GA), Not Applicable (SEE13451323GA), Not Applicable (SEE13451324GA), Not Applicable (SEE14382468GA), 00840682109260SEE17450007SA (SEE17450007SA), 00840682109260SEE19120024SA (SEE19120024SA), Not Applicable (SEE12370070GA), Not Applicable (SEE13270883GA), 00840682109260SEE17350010SA (SEE17350010SA), Not Applicable (SEE12302395GA), Not Applicable (SEE09470660GA), Not Applicable (SEE09470662GA), Not Applicable (SEE09470675GA), Not Applicable (SEE09470679GA), Not Applicable (SEE12302380GA), Not Applicable (SEE10430619GA), Not Applicable (SEE10430624GA), Not Applicable (SEE10430636GA), Not Applicable (SEE14031624GA), Not Applicable (SEE14031633GA), Not Applicable (SEE14081769GA), Not Applicable (SEE14081774GA), 00840682109260SEE17080001SA (SEE17080001SA), 00840682109260SEE18280004SA (SEE18280004SA), 00840682109260SEE18290006SA (SEE18290006SA), 00840682109260SEE18290015SA (SEE18290015SA), Not Applicable (SEE16053354GA), Not Applicable (SEE16053355GA), Not Applicable (SEE15473339GA), Not Applicable (SEE09530898GA), Not Applicable (SEE15063006GA), Not Applicable (SEE12490276GA), Not Applicable (SEE15052940GA), Not Applicable (SEE11161328GA), Not Applicable (SEE12440232GA), Not Applicable (SEE09510813GA), Not Applicable (SEE11081177GA), Not Applicable (SEE11081197GA), Not Applicable (SEE15173219GA), Not Applicable (SEE15193301GA), Not Applicable (SEE16292870SA), Not Applicable (SEE10400388GA), Not Applicable (SEE10400390GA), Not Applicable (SEE10430654GA), Not Applicable (SEE11050919GA), Not Applicable (SEE11050935GA), Not Applicable (SEE11050936GA), Not Applicable (SEE11050937GA), Not Applicable (SEE11071071GA), Not Applicable (SEE11151303GA), Not Applicable (SEE11151310GA), Not Applicable (SEE12420170GA), Not Applicable (SEE12510305GA), Not Applicable (SEE13431272GA), Not Applicable (SEE13431300GA), Not Applicable (SEE12420160GA), Not Applicable (SEE12031942GA), 00840682109260SEE18320009SA (SEE18320009SA), Not Applicable (SEE09330418GA), Not Applicable (SEE14412507GA), Not Applicable (SEE14452619GA), Not Applicable (SEE14462650GA), Not Applicable (SEE15193284GA), 00840682109260SEE18330036SA (SEE18330036SA), 00840682109260SEE18330038SA (SEE18330038SA), 00840682109260SEE18330041SA (SEE18330041SA), 00840682109260SEE18330042SA (SEE18330042SA), 00840682109260SEE18330043SA (SEE18330043SA), Not Applicable (SEE15052950GA), Not Applicable (SEE15062992GA), Not Applicable (SEE16324029SA), Not Applicable (SEE14422519GA), Not Applicable (SEE11071137GA), Not Applicable (SEE11081185GA), Not Applicable (SEE11081178GA), Not Applicable (SEE12092045GA), Not Applicable (SEE12092047GA), Not Applicable (SEE12092049GA), Not Applicable (SEE12420162GA), Not Applicable (SEE12490283GA), Not Applicable (SEE13150665GA), Not Applicable (SEE13150675GA), Not Applicable (SEE14482706GA), Not Applicable (SEE14482715GA), Not Applicable (SEE14482717GA), Not Applicable (SEE14482718GA), Not Applicable (SEE14482720GA), 01008406821092601118020921SEE18060023SA (SEE18060023SA), Not Applicable (SEE13431287GA), Not Applicable (SEE11081191GA), Not Applicable (SEE11050927GA), Not Applicable (SEE10250103GA), Not Applicable (SEE11030897GA), Not Applicable (SEE11071061GA), Not Applicable (SEE13240856GA), Not Applicable (SEE15032884GA), Not Applicable (SEE15032887GA), Not Applicable (SEE15113174GA), Not Applicable (SEE14462647GA), Not Applicable (SEE14462653GA), Not Applicable (SEE14472678GA), Not Applicable (SEE10090947GA), Not Applicable (SEE09470664GA), Not Applicable (SEE09530897GA), Not Applicable (SEE10400436GA), Not Applicable (SEE11341640GA), Not Applicable (SEE13451314GA), Not Applicable (SEE13451348GA), Not Applicable (SEE13451370GA), Not Applicable (SEE13451371GA), Not Applicable (SEE13451372GA), Not Applicable (SEE13451373GA), Not Applicable (SEE14031595GA), Not Applicable (SEE14101788GA), Not Applicable (SEE14101789GA), Not Applicable (SEE14111801GA), Not Applicable (SEE14161905GA), Not Applicable (SEE16313483SA), Not Applicable (SEE16313484SA), Not Applicable (SEE16430006SA), 00840682109260SEE18220014SA (SEE18220014SA), 00840682109260SEE18220015SA (SEE18220015SA), Not Applicable (SEE13110505GA), Not Applicable (SEE13000389GA), Not Applicable (SEE09310363GA), Not Applicable (SEE11171380GA), Not Applicable (SEE12322411GA), Not Applicable (SEE13150676GA), Not Applicable (SEE13150673GA), Not Applicable (SEE13150691GA), Not Applicable (SEE14031655GA), Not Applicable (SEE14031665GA), Not Applicable (SEE12172197GA), Not Applicable (SEE16430001SA), Not Applicable (SEE16430025SA), Not Applicable (SEE14302256GA), Not Applicable (SEE12092082GA), Not Applicable (SEE09460653GA), Not Applicable (SEE13481405GA), Not Applicable (SEE11361669GA), Not Applicable (SEE12051992GA), Not Applicable (SEE13050440GA), Not Applicable (SEE13180738GA), Not Applicable (SEE11111232GA), Not Applicable (SEE09470655GA), Not Applicable (SEE11361700GA), Not Applicable (SEE11161332GA), Not Applicable (SEE10400379GA), Not Applicable (SEE12082028GA), Not Applicable (SEE13180740GA), Not Applicable (SEE09510771GA), Not Applicable (SEE10120009GA), Not Applicable (SEE10120031GA), Not Applicable (SEE10121017GA), Not Applicable (SEE11371808GA), Not Applicable (SEE12332440GA), Not Applicable (SEE13180731GA), Not Applicable (SEE13180732GA), Not Applicable (SEE13180736GA), Not Applicable (SEE10410488GA), Not Applicable (SEE10460750GA), Not Applicable (SEE10470785GA), Not Applicable (SEE13150672GA), Not Applicable (SEE13180737GA), Not Applicable (SEE12430199GA), Not Applicable (SEE12430182GA), Not Applicable (SEE13150679GA), Not Applicable (SEE11331592GA), Not Applicable (SEE10460745GA), Not Applicable (SEE13180730GA), Not Applicable (SEE13150677GA), Not Applicable (SEE13180721GA), Not Applicable (SEE13180727GA), Not Applicable (SEE13180734GA), Not Applicable (SEE13100478GA), Not Applicable (SEE16233398GA), Not Applicable (SEE16383414GA), Not Applicable (SEE16390022SA), Not Applicable (SEE18400001SA), Not Applicable (SEE09230283GA), Not Applicable (SEE09390533GA), Not Applicable (SEE09460645GA), Not Applicable (SEE10410504GA), Not Applicable (SEE10420581GA), Not Applicable (SEE10430615GA), Not Applicable (SEE10430618GA), Not Applicable (SEE10460717GA), Not Applicable (SEE10460732GA), Not Applicable (SEE10470782GA), Not Applicable (SEE10470783GA), Not Applicable (SEE10470792GA), Not Applicable (SEE10470794GA), Not Applicable (SEE11341664GA), Not Applicable (SEE11361734GA), Not Applicable (SEE12122152GA), Not Applicable (SEE12122160GA), Not Applicable (SEE12122164GA), Not Applicable (SEE13040424GA), Not Applicable (SEE13110548GA), Not Applicable (SEE13180789GA), Not Applicable (SEE13401173GA), Not Applicable (SEE14031593GA), 00840682109260SEE18320003SA (SEE18320003SA), Not Applicable (SEE18420008SA), Not Applicable (SEE15383313GA), Not Applicable (SEE15413318GA), 00840682109260SEE17340017SA (SEE17340017SA), 00840682109260SEE18190028SA (SEE18190028SA), 00840682109260SEE18320006SA (SEE18320006SA), 00840682109260SEE18380025SA (SEE18380025SA), 00840682109260SEE18380026SA (SEE18380026SA), 00840682109260SEE18380028SA (SEE18380028SA), 00840682109260SEE18380030SA (SEE18380030SA), Not Applicable (SEE15062983GA), Not Applicable (SEE15173234GA), Not Applicable (SEE15183263GA), Not Applicable (SEE15193293GA), Not Applicable (SEE15312130SA), Not Applicable (SEE16252020SA), 00840682109260SEE16520001SA (SEE16520001SA), 00840682109260SEE17360008SA (SEE17360008SA), 00840682109260SEE17460020SA (SEE17460020SA), Not Applicable (SEE12252319GA), Not Applicable (SEE12252324GA), Not Applicable (SEE14031718GA), Not Applicable (SEE12262340GA), Not Applicable (SEE14021538GA), Not Applicable (SEE14021556GA), Not Applicable (SEE11321589GA), Not Applicable (SEE13371089GA), Not Applicable (SEE16313485SA), Not Applicable (SEE15373307GA), Not Applicable (SEE15312128SA), Not Applicable (SEE15373308GA), 00840682109260SEE17210018SA (SEE17210018SA), Not Applicable (SEE15093038GA), Not Applicable (SEE15373310GA), Not Applicable (SEE13361062GA), Not Applicable (SEE15193270GA), Not Applicable (SEE15423322GA), Not Applicable (SEE15032886GA), Not Applicable (SEE16430039SA), Not Applicable (SEE09370512GA), Not Applicable (SEE11361712GA), 00840682109260SEE17460002SA (SEE17460002SA), 00840682109260SEE17460015SA (SEE17460015SA), 00840682109260SEE17460019SA (SEE17460019SA), Not Applicable (SEE18450022SA), 00840682109260SEE17480008SA (SEE17480008SA), Not Applicable (SEE09530877GA), Not Applicable (SEE11301554GA), Not Applicable (SEE12182225GA), Not Applicable (SEE14432593GA), Not Applicable (SEE10120990GA), 00840682109260SEE18410026SA (SEE18410026SA), Not Applicable (SEE11181422GA), Not Applicable (SEE10250166GA), Not Applicable (SEE16053344GA), 00840682109260SEE18160056SA (SEE18160056SA), 00840682109260SEE18290010SA (SEE18290010SA), 00840682109260SEE18290013SA (SEE18290013SA), 00840682109260SEE18410034SA (SEE18410034SA), 00840682109260SEE18420031SA (SEE18420031SA), 00840682109260SEE17510003SA (SEE17510003SA), Not Applicable (SEE12242295GA), Not Applicable (SEE14171928GA), 01008406821092601118101621SEE18420033SA (SEE18420033SA), Not Applicable (SEE11351666GA), 00840682109260SEE17330002SA (SEE17330002SA), Not Applicable (SEE15193296GA).Not Applicable (SEE12172185GA), Not Applicable (SEE11331600GA), Not Applicable (SEE11361699GA), Not Applicable (SEE15473333GA), Not Applicable (SEE11200793GA), Not Applicable (SEE14412493GA), Not Applicable (SEE10130035GA), Not Applicable (SEE15473338GA), Not Applicable (SEE09510827GA), 00840682109260SEE17270008SA (SEE17270008SA), Not Applicable (SEE16043341GA), Not Applicable (SEE12112113GA), Not Applicable (SEE14101778GA), Not Applicable (SEE16370001SA), Not Applicable (SEE10400380GA), Not Applicable (SEE09230280GA), Not Applicable (SEE14031687GA), 00840682109260SEE17240117SA (SEE17240117SA), Not Applicable (SEE11161345GA), Not Applicable (SEE14222041GA), Not Applicable (SEE13220841GA), Not Applicable (SEE13371112GA), Not Applicable (SEE10400391GA), Not Applicable (SEE09140173GA), Not Applicable (SEE15012857GA), Not Applicable (SEE15373306GA), Not Applicable (SEE09330426GA), Not Applicable (SEE093304426GA), Not Applicable (SEE10100970GA), Not Applicable (SEE14211985GA), Not Applicable (SEE13270889GA), Not Applicable (SEE10450668GA), Not Applicable (SEE11301557GA), Not Applicable (SEE15032869GA), Not Applicable (SEE09530841GA), Not Applicable (SEE090100088GA), Not Applicable (SEE09100088GA), Not Applicable (SEE09130167GA), Not Applicable (SEE10530853GA), Not Applicable (SEE10130037GA), Not Applicable (SEE11211476GA), Not Applicable (SEE09390549GA), Not Applicable (SEE09390550GA), Not Applicable (SEE14512824GA), Not Applicable (SEE15423324GA), Not Applicable (SEE15473336GA), 00840682109260SEE18100030SA (SEE18100030SA), 00840682109260SEE1810003SA (SEE1810003SA), Not Applicable (SEE14021510GA), Not Applicable (SEE14021521GA), Not Applicable (SEE09420596GA), 00840682109260SEE18320021SA (SEE18320021SA), Not Applicable (SEE16252012SA), Not Applicable (SEE16053358GA), 00840682109260SEE18410018SA (SEE18410018SA), Not Applicable (SEE09330415GA), Not Applicable (SEE15113159GA), 00840682109260SEE18420010SA (SEE18420010SA), 00840682109260SEE18420024SA (SEE18420024SA), Not Applicable (SEE09130164GA), Not Applicable (SEE11341624GA), Not Applicable (SEE12520358GA), Not Applicable (SEE12520385GA), Not Applicable (SEE14362378GA), 00840682109260SEE18290004SA (SEE18290004SA), Not Applicable (SEE16430063SA), Not Applicable (SEE11341657GA), Not Applicable (SEE09230270GA), Not Applicable (SEE13371119GA), 00840682109260SEE17430041SA (SEE17430041SA), 00840682109260SEE18240010SA (SEE18240010SA), Not Applicable (SEE10450673GA), Not Applicable (SEE14272142GA), Not Applicable (SEE12430186GA), Not Applicable (SEE12490267GA), Not Applicable (SEE12242284GA), Not Applicable (SEE11361696GA), Not Applicable (SEE11301556GA), Not Applicable (SEE12172187GA), 00840682109260SEE18410017SA (SEE18410017SA), 00840682109260SEE18410019SA (SEE18410019SA), Not Applicable (SEE11141264GA), Not Applicable (SEE15193300GA), 00840682109260SEE18410025SA (SEE18410025SA), 00840682109260SEE18410036SA (SEE18410036SA), 00840682109260SEE18420026SA (SEE18420026SA), 00840682109260SEE18420032SA (SEE18420032SA), Not Applicable (SEE11361695GA), Not Applicable (SEE15153204GA), Not Applicable (SEE09350459GA), Not Applicable (SEE09350464GA), Not Applicable (SEE09350465GA), Not Applicable (SEE09380514GA), Not Applicable (SEE13280917GA), 01008406821092601116102521SEE16430009SA (SEE16430009SA), 01008406821092601116102521SEE16430013SA (SEE16430013SA), 01008406821092601116102721SEE16430058SA (SEE16430058SA), 00840682109260SEE18420012SA (SEE18420012SA), Not Applicable (SEE11171364GA), Not Applicable (SEE09330420GA), Not Applicable (SEE09510800GA), Not Applicable (SEE18390001SA), Not Applicable (SEE10310285GA), Not Applicable (SEE11060961GA), Not Applicable (SEE14372404GA), Not Applicable (SEE16252016SA), Not Applicable (SEE09360474GA), Not Applicable (SEE12420158GA), Not Applicable (SEE12490265GA), Not Applicable (SEE15153200GA), Not Applicable (SEE15193280GA), Not Applicable (SEE15423327GA), 00840682109260SEE17210016SA (SEE17210016SA), Not Applicable (SEE120620007GA), Not Applicable (SEE09330407GA), Not Applicable (SEE12102096GA), Not Applicable (SEE12370069GA), Not Applicable (SEE14492723GA), Not Applicable (SEE12322412GA), Not Applicable (SEE10400428GA), Not Applicable (SEE10420554GA), 00840682109260SEE17390010SA (SEE17390010SA), Not Applicable (SEE10270250GA), Not Applicable (SEE16430059SA), Not Applicable (SEE13371109GA), Not Applicable (SEE11211507GA), Not Applicable (SEE14222004GA), Not Applicable (SEE12370065GA), Not Applicable (SEE10310280GA), 00840682109260SEE17460005SA (SEE17460005SA), Not Applicable (SEE16173376GA), 00840682109260SEE17310012SA (SEE17310012SA), Not Applicable (SEE11241549GA), Not Applicable (SEE12430230GA), Not Applicable (SEE11191423GA), Not Applicable (SEE10400409GA), 00840682109260SEE18410015SA (SEE18410015SA), 00840682109260SEE18450016SA (SEE18450016SA), Not Applicable (SEE15183264GA), Not Applicable (SEE09470659GA), Not Applicable (SEE09470667GA), Not Applicable (SEE15193278GA), 00840682109260SEE18380033SA (SEE18380033SA), 00840682109260SEE18380035SA (SEE18380035SA), Not Applicable (SEE12282349GA), 00840682109260SEE17190011SA (SEE17190011SA), 00840682109260SEE17040002SA (SEE17040002SA), 00840682109260SEE18390009SA (SEE18390009SA), 00840682109260SEE18390011SA (SEE18390011SA), Not Applicable (SEE16272583SA), Not Applicable (SEE09330424GA), Not Applicable (SEE10080926GA), Not Applicable (SEE11361680GA), Not Applicable (SEE09380524GA), Not Applicable (SEE15052912GA), 00840682109260SEE18220010SA (SEE18220010SA), 00840682109260SEE17430047SA (SEE17430047SA), Not Applicable (SEE16241851SA), 00840682109260SEE17480003SA (SEE17480003SA), 00840682109260SEE18040052SA (SEE18040052SA), 00840682109260SEE18040054SA (SEE18040054SA), Not Applicable (SEE10450679GA), Not Applicable (SEE12242290GA), 00840682109260SEE18380039SA (SEE18380039SA), Not Applicable (SEE15113117GA), Not Applicable (SEE12302387GA), Not Applicable (SEE12192239GA), Not Applicable (SEE13481401GA), Not Applicable (SEE14332345GA), 01008406821092601117080121SEE17310001SA (SEE17310001SA), 00840682109260SEE17370001SA (SEE17370001SA), Not Applicable (SEE16053356GA), Not Applicable (SEE11060946GA), Not Applicable (SEE11201461GA), Not Applicable (SEE12051990GA), Not Applicable (SEE14041745GA), Not Applicable (SEE14041750GA), 00840682109260SEE17260005SA (SEE17260005SA), 00840682109260SEE18380043SA (SEE18380043SA), 00840682109260SEE18410021SA (SEE18410021SA), Not Applicable (SEE16410017SA), Not Applicable (SEE16183387GA), Not Applicable (SEE16313482SA), Not Applicable (SEE15113171GA), 00840682109260SEE18240053SA (SEE18240053SA), 00840682109260SEE18240054SA (SEE18240054SA), 00840682109260SEE18240057SA (SEE18240057SA), Not Applicable (SEE13190807GA), Not Applicable (SEE13240863GA), Not Applicable (SEE15153199GA), Not Applicable (SEE16272420SA), Not Applicable (SEE14151873GA), Not Applicable (SEE10250123GA), 00840682109260SEE18380036SA (SEE18380036SA), 00840682109260SEE18390006SA (SEE18390006SA), 00840682109260SEE18390010SA (SEE18390010SA), Not Applicable (SEE15173231GA), Not Applicable (SEE15173244GA), Not Applicable (SEE18380040SA), Not Applicable (SEE18390004SA), 00840682109260SEE18010003SA (SEE18010003SA), Not Applicable (SEE13180778GA), Not Applicable (SEE14031660GA), 00840682109260SEE18330008SA (SEE18330008SA), Not Applicable (SEE09110155GA), Not Applicable (SEE14031706GA), Not Applicable (SEE15173245GA), 00840682109260SEE16410013SA (SEE16410013SA), Not Applicable (SEE14041729GA), Not Applicable (SEE12430227GA), Not Applicable (SEE13120571GA), Not Applicable (SEE13120585GA), Not Applicable (SEE14222011GA), Not Applicable (SEE14222013GA), 00840682109260SEE17100004SA (SEE17100004SA), Not Applicable (SEE15052927GA), Not Applicable (SEE15413317GA), Not Applicable (SEE16183382GA), 00840682109260SEE17460017SA (SEE17460017SA), Not Applicable (SEE14161886GA), 00840682109260SEE17310010SA (SEE17310010SA), 00840682109260SEE18320013SA (SEE18320013SA), 00840682109260SEE17130032SA (SEE17130032SA), Not Applicable (SEE12031945GA), Not Applicable (SEE16241852SA), 00840682109260SEE18380038SA (SEE18380038SA), Not Applicable (SEE11030896GA), Not Applicable (SEE10090945GA), Not Applicable (SEE10090958GA), Not Applicable (SEE10090968GA), Not Applicable (SEE11111226GA), Not Applicable (SEE14432581GA), Not Applicable (SEE15052943GA), Not Applicable (SEE12490254GA), Not Applicable (SEE12490281GA), Not Applicable (SEE12490285GA), Not Applicable (SEE14101792GA), Not Applicable (SEE14101796GA), Not Applicable (SEE14151872GA), Not Applicable (SEE14171921GA), Not Applicable (SEE12162173GA), Not Applicable (SEE18380041SA), Not Applicable (SEE18380042SA), 00840682109260SEE18030013SA (SEE18030013SA), Not Applicable (SEE18380044SA), Not Applicable (SEE16173373GA), Not Applicable (SEE09100080GA), Not Applicable (SEE13150678GA), Not Applicable (SEE15103109GA), Not Applicable (SEE16241846SA), Not Applicable (SEE16390007SA), Not Applicable (SEE16430051SA), 00840682109260SEE18260007SA (SEE18260007SA), 00840682109260SEE18420004SA (SEE18420004SA), Not Applicable (SEE15032894GA), Not Applicable (SEE15032897GA), 00840682109260SEE18500002SA (SEE18500002SA), 00840682109260SEE18500024SA (SEE18500024SA), 00840682109260SEE18500031SA (SEE18500031SA), 00840682109260SEE18500041SA (SEE18500041SA), Not Applicable (SEE09510791GA), Not Applicable (SEE15062978GA), Not Applicable (SEE15062977GA), Not Applicable (SEE15193273GA), Not Applicable (SEE15193294GA), Not Applicable (SEE16390011SA), Not Applicable (SEE18410035SA), Not Applicable (SEE15052939GA), Not Applicable (SEE15113177GA), Not Applicable (SEE09430614GA), Not Applicable (SEE12490299GA), Not Applicable (SEE14332339GA), Not Applicable (SEE13280947GA), Not Applicable (SEE15173232GA), Not Applicable (SEE11071100GA), Not Applicable (SEE12510312GA), Not Applicable (SEE13361079GA), Not Applicable (SEE15173232SA), 00840682109260SEE17430022SA (SEE17430022SA), 00840682109260SEE18360024SA (SEE18360024SA), Not Applicable (SEE13491429GA), Not Applicable (SEE13491433GA), Not Applicable (SEE13491434GA), Not Applicable (SEE13491445GA), 00840682109260SEE18330003SA (SEE18330003SA), 00840682109260SEE18220005SA (SEE18220005SA), Not Applicable (SEE16390006SA), 00840682109260SEE18420006SA (SEE18420006SA), Not Applicable (SEE11101219GA), Not Applicable (SEE11341643GA), Not Applicable (SEE09110154GA), Not Applicable (SEE10420591GA), Not Applicable (SEE10090961GA), Not Applicable (SEE12332442GA), Not Applicable (SEE16173374GA), Not Applicable (SEE16272593SA), Not Applicable (SEE16272596SA), 00840682109260SEE17500013SA (SEE17500013SA), Not Applicable (SEE10420539GA), Not Applicable (SEE13180766GA), 00840682109260SEE17360006SA (SEE17360006SA), 00840682109260SEE17430038SA (SEE17430038SA), 00840682109260SEE17430042SA (SEE17430042SA), 00840682109260SEE18450019SA (SEE18450019SA), Not Applicable (SEE12192238GA), 00840682109260SEE18320028SA (SEE18320028SA), Not Applicable (SEE16053359GA), 00840682109260SEE17310003SA (SEE17310003SA), 00840682109260SEE18240020SA (SEE18240020SA), Not Applicable (SEE12420163GA), Not Applicable (SEE09310399GA), Not Applicable (SEE13010402GA), Not Applicable (SEE09300347GA), Not Applicable (SEE09530885GA), Not Applicable (SEE09380513GA), Not Applicable (SEE11060950GA), Not Applicable (SEE11081161GA), 00840682109260SEE18320001SA (SEE18320001SA), Not Applicable (SEE13150692GA), Not Applicable (SEE10390295GA), Not Applicable (SEE09530879GA), Not Applicable (SEE09530894GA), Not Applicable (SEE10120025GA), Not Applicable (SEE10130041GA), Not Applicable (SEE10130054GA), Not Applicable (SEE10130058GA), Not Applicable (SEE12062009GA), Not Applicable (SEE13120612GA), Not Applicable (SEE16272627SA), Not Applicable (SEE11060960GA), Not Applicable (SEE09510784GA), 00840682109260SEE17440020SA (SEE17440020SA), 00840682109260SEE18100034SA (SEE18100034SA), Not Applicable (SEE11111249GA), Not Applicable (SEE11321577GA), Not Applicable (SEE13060446GA), Not Applicable (SEE10420518GA), Not Applicable (SEE10450665GA), Not Applicable (SEE15473335GA), Not Applicable (SEE12192235GA), 00840682109260SEE18240055SA (SEE18240055SA), 00840682109260SEE18450062SA (SEE18450062SA), Not Applicable (SEE13120592GA), 00840682109260SEE17500021SA (SEE17500021SA), Not Applicable (SEE09530902GA), Not Applicable (SEE15183258GA), 00840682109260SEE17050009SA (SEE17050009SA), Not Applicable (SEE15423331GA), Not Applicable (SEE15183251GA), Not Applicable (SEE15183250GA), Not Applicable (SEE16123364GA), Not Applicable (SEE10310288GA), Not Applicable (SEE12192237GA), Not Applicable (SEE11081180GA), Not Applicable (SEE09530909GA), Not Applicable (SEE09230285GA), Not Applicable (SEE11181420GA), 00840682109260SEE18060004SA (SEE18060004SA), 00840682109260SEE18070003SA (SEE18070003SA), 00840682109260SEE18250026SA (SEE18250026SA), Not Applicable (SEE15312142SA), Not Applicable (SEE10130052GA), Not Applicable (SEE16262152SA), Not Applicable (SEE12192240GA), Not Applicable (SEE12112120GA), Not Applicable (SEE13040421GA), Not Applicable (SEE10270258GA), Not Applicable (SEE13010399GA), Not Applicable (SEE16183384GA), Not Applicable (SEE16173368GA), Not Applicable (SEE10090964GA), Not Applicable (SEE12041956GA), Not Applicable (SEE10400402GA), Not Applicable (SEE10410508GA), Not Applicable (SEE10420521GA), Not Applicable (SEE12051986GA), Not Applicable (SEE15423325GA), Not Applicable (SEE09430603GA), Not Applicable (SEE10310287GA), Not Applicable (SEE11301555GA), Not Applicable (SEE14021508GA), Not Applicable (SEE13481416GA), Not Applicable (SEE11301553GA), Not Applicable (SEE10410505GA), Not Applicable (SEE12262338GA), Not Applicable (SEE12262339GA), Not Applicable (SEE13150687GA), Not Applicable (SEE09360476GA), Not Applicable (SEE10390292GA), 00840682109260SEE14462664GA (SEE14462664GA), Not Applicable (SEE15113147GA), Not Applicable (SEE15113155GA), Not Applicable (SEE09480749GA), Not Applicable (SEE09480739GA), Not Applicable (SEE09490753GA), Not Applicable (SEE09490751GA), Not Applicable (SEE16053348GA), Not Applicable (SEE16053361GA), Not Applicable (SEE16053362GA), Not Applicable (SEE16233393GA), Not Applicable (SEE16233395GA), Not Applicable (SEE16233397GA), Not Applicable (SEE16233399GA), Not Applicable (SEE16344324SA), Not Applicable (SEE17160003SA), Not Applicable (SEE14432582GA), Not Applicable (SEE16213389GA), 00840682109260SEE17300015SA (SEE17300015SA), 00840682109260SEE17300016SA (SEE17300016SA), 00840682109260SEE17310004SA (SEE17310004SA), 00840682109260SEE18200014SA (SEE18200014SA), 00840682109260SEE18510003SA (SEE18510003SA), 00840682109260SEE18510013SA (SEE18510013SA), Not Applicable (SEE14432595GA), Not Applicable (SEE14332342GA), 00840682109260SEE17020028SA (SEE17020028SA), 00840682109260SEE17070027SA (SEE17070027SA), Not Applicable (SEE11161335GA), Not Applicable (SEE14332346GA), Not Applicable (SEE15113161GA), 00840682109260SEE14222015GA (SEE14222015GA), Not Applicable (SEE15193274GA), Not Applicable (SEE15193275GA), Not Applicable (SEE16183381GA), Not Applicable (SEE11071062GA), Not Applicable (SEE11071069GA), Not Applicable (SEE11071091GA), Not Applicable (SEE11071093GA), Not Applicable (SEE11081163GA), Not Applicable (SEE11141294GA), Not Applicable (SEE13451342GA), 00840682109260SEE17450006SA (SEE17450006SA), 00840682109260SEE17110003SA (SEE17110003SA), 00840682109260SEE18380024SA (SEE18380024SA), 00840682109260SEE18380027SA (SEE18380027SA), 00840682109260SEE18380031SA (SEE18380031SA), 00840682109260SEE17150011SA (SEE17150011SA), Not Applicable (SEE11040911GA), Not Applicable (SEE13411211GA), Not Applicable (SEE10250078GA), Not Applicable (SEE13220838GA), Not Applicable (SEE16170870SA), Not Applicable (SEE16123363GA), Not Applicable (SEE10400389GA), Not Applicable (SEE12360039GA), Not Applicable (SEE12082031GA), Not Applicable (SEE16293174SA), Not Applicable (SEE16043343GA), Not Applicable (SEE10090965GA), Not Applicable (SEE10260201GA), Not Applicable (SEE10400414GA), Not Applicable (SEE10400415GA), Not Applicable (SEE13481387GA), Not Applicable (SEE14031599GA), Not Applicable (SEE12282355GA), Not Applicable (SEE13351043GA), Not Applicable (SEE13351045GA), 00840682109260SEE18170006SA (SEE18170006SA), Not Applicable (SEE16053346GA), Not Applicable (SEE11341634GA), Not Applicable (SEE13351048GA), Not Applicable (SEE15173230GA), Not Applicable (SEE15062972GA), 00840682109260SEE17220010SA (SEE17220010SA), Not Applicable (SEE15183262GA), Not Applicable (SEE15062965GA), Not Applicable (SEE16233396GA), Not Applicable (SEE16123365GA), Not Applicable (SEE16383413GA), 01008406821092601116102521SEE16430045SA (SEE16430045SA), 00840682109260SEE17050008SA (SEE17050008SA), 00840682109260SEE17230052SA (SEE17230052SA), Not Applicable (SEE14031659GA), Not Applicable (SEE14041759GA), Not Applicable (SEE15063002GA), Not Applicable (SEE14021517GA), Not Applicable (SEE13120598GA), 00840682109260SEE17490018SA (SEE17490018SA), Not Applicable (SEE10460743GA), Not Applicable (SEE10460699GA), Not Applicable (SEE12370053GA), Not Applicable (SEE09510835GA), Not Applicable (SEE11191439GA), Not Applicable (SEE15193299GA), Not Applicable (SEE15261529SA), Not Applicable (SEE15261535SA), Not Applicable (SEE15093054GA), Not Applicable (SEE15093055GA), Not Applicable (SEE15093056GA), Not Applicable (SEE15093060GA), Not Applicable (SEE15093061GA), Not Applicable (SEE13501478GA), Not Applicable (SEE16324028SA), Not Applicable (SEE16324030SA), 00840682109260SEE18190001SA (SEE18190001SA), 00840682109260SEE18190002SA (SEE18190002SA), 00840682109260SEE18190019SA (SEE18190019SA), 00840682109260SEE18190023SA (SEE18190023SA), Not Applicable (SEE09100149GA), Not Applicable (SEE10250131GA), Not Applicable (SEE10250171GA), Not Applicable (SEE10250190GA), Not Applicable (SEE10390344GA), Not Applicable (SEE10400367GA), Not Applicable (SEE10400368GA), Not Applicable (SEE12340010GA), Not Applicable (SEE12350023GA), Not Applicable (SEE16303406GA), Not Applicable (SEE16303408GA), Not Applicable (SEE16333411GA), Not Applicable (SEE16383415GA), Not Applicable (SEE16383416GA), Not Applicable (SEE16450011SA), 00840682109260SEE17330001SA (SEE17330001SA), Not Applicable (SEE10100971GA), Not Applicable (SEE10100981GA), Not Applicable (SEE10450661GA), Not Applicable (SEE11161343GA), Not Applicable (SEE12410132GA), Not Applicable (SEE09480720GA), Not Applicable (SEE09510786GA), Not Applicable (SEE09510799GA), Not Applicable (SEE09510826GA), Not Applicable (SEE10250188GA), Not Applicable (SEE10260210GA), Not Applicable (SEE11181402GA), Not Applicable (SEE12102089GA), Not Applicable (SEE12102091GA), Not Applicable (SEE12122131GA), Not Applicable (SEE12122139GA), Not Applicable (SEE12430178GA), Not Applicable (SEE15173237GA), Not Applicable (SEE15373312GA), Not Applicable (SEE16233392GA), 00840682109260SEE16233394GA (SEE16233394GA), 00840682109260SEE18430004SA (SEE18430004SA), 00840682109260SEE18430005SA (SEE18430005SA), 00840682109260SEE18450017SA (SEE18450017SA), Not Applicable (SEE10400420GA), Not Applicable (SEE09360480GA), Not Applicable (SEE10130066GA), Not Applicable (SEE11201458GA), Not Applicable (SEE11361722GA), Not Applicable (SEE13190810GA), 00840682109260SEE17340008SA (SEE17340008SA), Not Applicable (SEE18400003SA), 00840682109260SEE18400018SA (SEE18400018SA), 00840682109260SEE13481417GA (SEE13481417GA), 00840682109260SEE18380016SA (SEE18380016SA), Not Applicable (SEE09460642GA), Not Applicable (SEE16241855SA), Not Applicable (SEE16183385GA), 00840682109260SEE18380012SA (SEE18380012SA), 00840682109260SEE17510008SA (SEE17510008SA), 00840682109260SEE17430040SA (SEE17430040SA), Not Applicable (SEE09360485GA), Not Applicable (SEE12322426GA), 00840682109260SEE17480002SA (SEE17480002SA), Not Applicable (SEE10530862GA), Not Applicable (SEE10420516GA), Not Applicable (SEE10420546GA), Not Applicable (SEE10420552GA), Not Applicable (SEE10420555GA), Not Applicable (SEE13371115GA), 00840682109260SEE17350008SA (SEE17350008SA), Not Applicable (SEE09310395GA), Not Applicable (SEE15173240GA), 00840682109260SEE18160035SA (SEE18160035SA), Not Applicable (SEE14222045GA), Not Applicable (SEE14252122GA), 00840682109260SEE17150012SA (SEE17150012SA), Not Applicable (SEE12232251GA), Not Applicable (SEE15413314GA), Not Applicable (SEE11081162GA), Not Applicable (SEE10090950GA), 00840682109260SEE17240119SA (SEE17240119SA), 00840682109260SEE18380014SA (SEE18380014SA), Not Applicable (SEE12440235GA), Not Applicable (SEE12430196GA), Not Applicable (SEE14021502GA), Not Applicable (SEE15062964GA), Not Applicable (SEE16173371GA), 00840682109260SEE17200004SA (SEE17200004SA), 00840682109260SEE18010001SA (SEE18010001SA), 00840682109260SEE18110001SA (SEE18110001SA), 00840682109260SEE18110003SA (SEE18110003SA), Not Applicable (SEE09140175GA), Not Applicable (SEE09150198GA), Not Applicable (SEE09510774GA), Not Applicable (SEE11071068GA), Not Applicable (SEE12041950GA), 00840682109260SEE17480015SA (SEE17480015SA), Not Applicable (SEE10530863GA), Not Applicable (SEE16313481SA), 00840682109260SEE17330011SA (SEE17330011SA), 00840682109260SEE18380015SA (SEE18380015SA), Not Applicable (SEE18380021SA), Not Applicable (SEE15183253GA), Not Applicable (SEE09360477GA), 00840682109260SEE17440007SA (SEE17440007SA), 00840682109260SEE17150008SA (SEE17150008SA), 00840682109260SEE18040019SA (SEE18040019SA), 00840682109260SEE18040026SA (SEE18040026SA), Not Applicable (SEE18390013SA), Not Applicable (SEE18390016SA), 00840682109260SEE17510007SA (SEE17510007SA), Not Applicable (SEE16183383GA), Not Applicable (SEE11361675GA), Not Applicable (SEE16251980SA), 00840682109260SEE17440014SA (SEE17440014SA), Not Applicable (SEE13180723GA), Not Applicable (SEE14422554GA), Not Applicable (SEE15093058GA), Not Applicable (SEE13280921GA), Not Applicable (SEE13481404GA), Not Applicable (SEE14512820GA), Not Applicable (SEE15093051GA), Not Applicable (SEE16053357GA), Not Applicable (SEE16183386GA), 01008406821092601118101121SEE18410020SA (SEE18410020SA), Not Applicable (SEE10090967GA), Not Applicable (SEE13351054GA), Not Applicable (SEE14292238GA), Not Applicable (SEE14292246GA), Not Applicable (SEE14302260GA), Not Applicable (SEE09410585GA), Not Applicable (SEE15173218GA), Not Applicable (SEE10250170GA), Not Applicable (SEE10250128GA), Not Applicable (SEE10250164GA), Not Applicable (SEE11071143GA), Not Applicable (SEE13351035GA), Not Applicable (SEE10250180GA), Not Applicable (SEE16410001SA), Not Applicable (SEE16053353GA), Not Applicable (SEE10250152GA), Not Applicable (SEE13431296GA), Not Applicable (SEE15423326GA), Not Applicable (SEE09020025GA), Not Applicable (SEE09530892GA), Not Applicable (SEE11071013GA), Not Applicable (SEE11361678GA), Not Applicable (SEE11361683GA), Not Applicable (SEE12062006GA), Not Applicable (SEE16380006SA), 00840682109260SEE17080013SA (SEE17080013SA), 00840682109260SEE17150013SA (SEE17150013SA), 00840682109260SEE17300006SA (SEE17300006SA), 00840682109260SEE18390017SA (SEE18390017SA), 00840682109260SEE18390018SA (SEE18390018SA), 00840682109260SEE18390019SA (SEE18390019SA), Not Applicable (SEE14021563GA), Not Applicable (SEE09510810GA), Not Applicable (SEE12112114GA), Not Applicable (SEE14222035GA), Not Applicable (SEE14382455GA), 00840682109260SEE17390011SA (SEE17390011SA), Not Applicable (SEE16390009SA), Not Applicable (SEE15113163GA), Not Applicable (SEE15113167GA), Not Applicable (SEE14452626GA), Not Applicable (SEE10130057GA), Not Applicable (SEE12041957GA), Not Applicable (SEE13411203GA), Not Applicable (SEE14292242GA), 00840682109260SEE18360015SA (SEE18360015SA), Not Applicable (SEE12112117GA), Not Applicable (SEE11181392GA), Not Applicable (SEE14462661GA), Not Applicable (SEE12400120GA), Not Applicable (SEE13180768GA), Not Applicable (SEE14272206GA), Not Applicable (SEE15193298GA), Not Applicable (SEE15183265GA), 00840682109260SEE18330019SA (SEE18330019SA), Not Applicable (SEE10420594GA), Not Applicable (SEE13431298GA), 00840682109260SEE14312298GA (SEE14312298GA), Not Applicable (SEE16183380GA), 01008406821092601118090721SEE18360019SA (SEE18360019SA), Not Applicable (SEE10410484GA), 00840682109260SEE18360018SA (SEE18360018SA), 00840682109260SEE18500004SA (SEE18500004SA), 00840682109260SEE18360010SA (SEE18360010SA), Not Applicable (SEE13130631GA), Not Applicable (SEE15113149GA), Not Applicable (SEE15173233GA), Not Applicable (SEE15173235GA), Not Applicable (SEE15193269GA), Not Applicable (SEE16043342GA), 00840682109260SEE18030008SA (SEE18030008SA), 00840682109260SEE18030017SA (SEE18030017SA), 01008406821092601118110821SEE18450074SA (SEE18450074SA), Not Applicable (SEE11371794GA), Not Applicable (SEE10250146GA), 01008406821092601118090721SEE18360022SA (SEE18360022SA), Not Applicable (SEE09470666GA), Not Applicable (SEE10090966GA), Not Applicable (SEE13000392GA), Not Applicable (SEE15193267GA), 00840682109260SEE17020026SA (SEE17020026SA), Not Applicable (SEE14041736GA), Not Applicable (SEE09040056GA), 00840682109260SEE17400001SA (SEE17400001SA), Not Applicable (SEE09020018GA), Not Applicable (SEE09190224GA), Not Applicable (SEE10400430GA), Not Applicable (SEE09240298GA), 00840682109260SEE18360007SA (SEE18360007SA), 01008406821092601116052521SEE16213388GA (SEE16213388GA), Not Applicable (SEE12062008GA), Not Applicable (SEE09530891GA), Not Applicable (SEE13431283GA), Not Applicable (SEE13481400GA), 00840682109260SEE17300004SA (SEE17300004SA), Not Applicable (SEE09460647GA), 00840682109260SEE17060006SA (SEE17060006SA), 00840682109260SEE18360012SA (SEE18360012SA), Not Applicable (SEE11071089GA), Not Applicable (SEE11071066GA), Not Applicable (SEE09020016GA), Not Applicable (SEE09040067GA), Not Applicable (SEE09230250GA), Not Applicable (SEE9040073GA), Not Applicable (SEE10090963GA), Not Applicable (SEE10120996GA), Not Applicable (SEE10121004GA), Not Applicable (SEE10121015GA), Not Applicable (SEE10420584GA), Not Applicable (SEE11371747GA), Not Applicable (SEE12062012GA), Not Applicable (SEE12062014GA), Not Applicable (SEE12092066GA), Not Applicable (SEE12420171GA), Not Applicable (SEE14452637GA), Not Applicable (SEE14462643GA), Not Applicable (SEE14462660GA), 00840682109260SEE17170007SA (SEE17170007SA), Not Applicable (SEE14462648GA), Not Applicable (SEE11071083GA), 00840682109260SEE18360023SA (SEE18360023SA), Not Applicable (SEE15193304GA), Not Applicable (SEE09330419GA), Not Applicable (SEE09510775GA), 0100840682109260SEE18450064SA (SEE18450064SA), Not Applicable (SEE10250186GA), Not Applicable (SEE10420592GA), Not Applicable (SEE15093057GA), 00840682109260SEE18280006SA (SEE18280006SA), 00840682109260SEE18450066SA (SEE18450066SA), Not Applicable (SEE25093057GA), Not Applicable (SEE09230275C), Not Applicable (SEE09230275GA), Not Applicable (SEE10090954GA), 01008406821092601118090721SEE18360021SA (SEE18360021SA), Not Applicable (SEE09020039GA), Not Applicable (SEE09040059GA), Not Applicable (SEE15093029GA), Not Applicable (SEE15093045GA), Not Applicable (SEE15093048GA), Not Applicable (SEE15093049GA), Not Applicable (SEE15093053GA), 00840682109260SEE18100025SA (SEE18100025SA), 00840682109260SEE18420029SA (SEE18420029SA), Not Applicable (SEE13020409GA), Not Applicable (SEE09100081GA), Not Applicable (SEE13411202GA), Not Applicable (SEE13411209GA), Not Applicable (SEE13431293GA), Not Applicable (SEE10260191GA), Not Applicable (SEE10430655GA), 00840682109260SEE18100032SA (SEE18100032SA). UPDATE. added No applicable UDI (SEE12122168GA), Not Applicable (SEE09100113GA), Not Applicable (SEE08510008GA).
7. Product Code: SPV, No UDI Number; Serial Number: SPV18300540SA.
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | 800-437-1171 |
Manufacturer Reason for Recall | May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions. |
FDA Determined Cause 2 | Software design |
| Action | GE Healthcare notified customers on about 08/30/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were made aware of the below six (6) potential safety issues along with corresponding safety instructions.
Safety Issue #1: Telemetry patients monitored on a Central Station or in combination monitoring mode with a CARESCAPE B850/650/450, DASH 3000/4000/5000, or SOLAR 8000M/I bedside monitor under certain conditions can experience the following: ECG arrhythmias configured to MEDIUM (WARNING) or lower alarm priority occurring before and after a NO TELEM condition may not re-activate after the NO TELEM condition resolves. This issue may result in delayed treatment for a potentially life-threatening ECG arrhythmia event. There have been no reported injuries as a result of this issue. Safety instructions included that the monitor can continue being used under these instructions: 1. Configure the VTACH arrhythmia to HIGH (CRISIS) alarm priority.
2. Any ECG arrhythmia of clinical concern should be configured to HIGH (CRISIS) alarm priority. 3. Make sure that the ECG HR alarm is set to priority of MEDIUM (WARNING) or higher to ensure proper alarming. 4. Ensure that your ECG HR limits are appropriately configured for your patient population.
Safety Issue #2: Telemetry patients monitored on a Central Station or in combination monitoring mode with a CARESCAPE B850/650/450, DASH 3000/4000/5000 or SOLAR 8000M/I bedside monitor under certain conditions can experience the following: A NO TELEM INFORMATIONAL (MESSAGE) indication may remain displayed on the Central Station instead of a HIGH (CRISIS) LEADS FAIL or a MEDIUM (WARNING) NO TELEM alarm. This issue can result in delayed treatment for a potentially a life-threatening event, as a failure to notice prolonged loss of ECG monitoring. There have been no reported injuries as a result of this issue. Safety instructions included that the monitor can continue being used by following these instructions: For a N |
| Quantity in Commerce | 22,438 (17,971 US; 4,467 OUS) devices total |
| Distribution | Worldwide distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
The countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BERMUDA, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUAM, GUATEMALA, HONDURAS, HONG KONG, HUNGARY, INDIA, Indonesia, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, Malaysia, MEXICO, MOROCCO, Myanmar, NETHERLANDS, New Zealand, NORWAY, PANAMA, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, ROMANIA, Russia, SAUDI ARABIA, Singapore, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TAJIKISTAN, Thailand, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, Viet Nam, YUGOSLAVIA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = DSI 510(K)s with Product Code = DSI
|
|
|
|