• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ApexPro Telemetry System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall ApexPro Telemetry System see related information
Date Initiated by Firm November 12, 2019
Create Date December 20, 2019
Recall Status1 Open3, Classified
Recall Number Z-0725-2020
Recall Event ID 84372
510(K)Number K000779  K021325  K032369  K080251  K033365  
Product Classification Detector and alarm, arrhythmia - Product Code DSI
Product ApexPro Telemetry System - Product Usage: The ApexPro Telemetry System is intended for use under the direct supervision of a licensed healthcare practitioner. The system is designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The system processes this physiological data to detect various ECG arrhythmia events and select physiological parameter limit violations. The ApexPro Telemetry System is intended to be installed in the hospital or clinical environment in order to provide clinicians with patient physiological data, while allowing for patient mobility. These systems are typically deployed in sub acute care areas in hospitals or clinical sites where patient mobility can enhance the effectiveness of the medical procedures administered.
Code Information Serial Numbers: Mfg. Lot or Serial # UDI # SEE09020009GA 00840682109260SEE09020009GA SEE09020010GA 00840682109260SEE09020010GA SEE09020011GA 00840682109260SEE09020011GA SEE09020012GA 00840682109260SEE09020012GA SEE09020013GA 00840682109260SEE09020013GA SEE09020015GA 00840682109260SEE09020015GA SEE09020017GA 00840682109260SEE09020017GA SEE09020020GA 00840682109260SEE09020020GA SEE09020022GA 00840682109260SEE09020022GA SEE09020023GA 00840682109260SEE09020023GA SEE09020024GA 00840682109260SEE09020024GA SEE09020027GA 00840682109260SEE09020027GA SEE09020028GA 00840682109260SEE09020028GA SEE09020029GA 00840682109260SEE09020029GA SEE09020030GA 00840682109260SEE09020030GA SEE09020031GA 00840682109260SEE09020031GA SEE09020032GA 00840682109260SEE09020032GA SEE09020033GA 00840682109260SEE09020033GA SEE09020034GA 00840682109260SEE09020034GA SEE09020035GA 00840682109260SEE09020035GA SEE09020036GA 00840682109260SEE09020036GA SEE09020038GA 00840682109260SEE09020038GA SEE09020040GA 00840682109260SEE09020040GA SEE09020042GA 00840682109260SEE09020042GA SEE09020043GA 00840682109260SEE09020043GA SEE09040045GA 00840682109260SEE09040045GA SEE09040046GA 00840682109260SEE09040046GA SEE09040047GA 00840682109260SEE09040047GA SEE09040048GA 00840682109260SEE09040048GA SEE09040050GA 00840682109260SEE09040050GA SEE09040051GA 00840682109260SEE09040051GA SEE09040052GA 00840682109260SEE09040052GA SEE09040053GA 00840682109260SEE09040053GA SEE09040054GA 00840682109260SEE09040054GA SEE09040055GA 00840682109260SEE09040055GA SEE09040058GA 00840682109260SEE09040058GA SEE09040060GA 00840682109260SEE09040060GA SEE09040061GA 00840682109260SEE09040061GA SEE09040062GA 00840682109260SEE09040062GA SEE09040063GA 00840682109260SEE09040063GA SEE09040064GA 00840682109260SEE09040064GA SEE09040066GA 00840682109260SEE09040066GA SEE09040068GA 00840682109260SEE09040068GA SEE09040069GA 00840682109260SEE09040069GA SEE09040071GA 00840682109260SEE09040071GA SEE09040072GA 00840682109260SEE09040072GA SEE09040075GA 00840682109260SEE09040075GA SEE09040076GA 00840682109260SEE09040076GA SEE09040077GA 00840682109260SEE09040077GA SEE09040078GA 00840682109260SEE09040078GA SEE09040079GA 00840682109260SEE09040079GA SEE09100082GA 00840682109260SEE09100082GA SEE09100083GA 00840682109260SEE09100083GA SEE09100084GA 00840682109260SEE09100084GA SEE09100085GA 00840682109260SEE09100085GA SEE09100086GA 00840682109260SEE09100086GA SEE09100087GA 00840682109260SEE09100087GA SEE09100089GA 00840682109260SEE09100089GA SEE09100090GA 00840682109260SEE09100090GA SEE09100091GA 00840682109260SEE09100091GA SEE09100092GA 00840682109260SEE09100092GA SEE09100093GA 00840682109260SEE09100093GA SEE09100094GA 00840682109260SEE09100094GA SEE09100095GA 00840682109260SEE09100095GA SEE09100096GA 00840682109260SEE09100096GA SEE09100097GA 00840682109260SEE09100097GA SEE09100098GA 00840682109260SEE09100098GA SEE09100099GA 00840682109260SEE09100099GA SEE09100100GA 00840682109260SEE09100100GA SEE09100101GA 00840682109260SEE09100101GA SEE09100102GA 00840682109260SEE09100102GA SEE09100103GA 00840682109260SEE09100103GA SEE09100104GA 00840682109260SEE09100104GA SEE09100105GA 00840682109260SEE09100105GA SEE09100106GA 00840682109260SEE09100106GA SEE09100107GA 00840682109260SEE09100107GA SEE09100108GA 00840682109260SEE09100108GA SEE09100109GA 00840682109260SEE09100109GA SEE09100110GA 00840682109260SEE09100110GA SEE09100111GA 00840682109260SEE09100111GA SEE09100112GA 00840682109260SEE09100112GA SEE09100114GA 00840682109260SEE09100114GA SEE09100115GA 00840682109260SEE09100115GA SEE09100116GA 00840682109260SEE09100116GA SEE09100117GA 00840682109260SEE09100117GA SEE09100118GA 00840682109260SEE09100118GA SEE09100119GA 00840682109260SEE09100119GA SEE09100120GA 00840682109260SEE09100120GA SEE09100121GA 00840682109260SEE09100121GA SEE09100123GA 00840682109260SEE09100123GA SEE09100124GA 00840682109260SEE09100124GA SEE09100125GA 00840682109260SEE09100125GA SEE09100126GA 00840682109260SEE09100126GA SEE09100127GA 00840682109260SEE09100127GA SEE09100128GA 00840682109260SEE09100128GA SEE09100129GA 00840682109260SEE09100129GA SEE09100130GA 00840682109260SEE09100130GA SEE09100131GA 00840682109260SEE09100131GA SEE09100132GA 00840682109260SEE09100132GA SEE09100133GA 00840682109260SEE09100133GA SEE09100134GA 00840682109260SEE09100134GA SEE09100135GA 00840682109260SEE09100135GA SEE09100137GA 00840682109260SEE09100137GA SEE09100138GA 00840682109260SEE09100138GA SEE09100139GA 00840682109260SEE09100139GA SEE09100140GA 00840682109260SEE09100140GA SEE09100147GA 00840682109260SEE09100147GA SEE09100148GA 00840682109260SEE09100148GA SEE09100150GA 00840682109260SEE09100150GA SEE09100152GA 00840682109260SEE09100152GA SEE09110156GA 00840682109260SEE09110156GA SEE09130158GA 00840682109260SEE09130158GA SEE09130159GA 00840682109260SEE09130159GA SEE09130160GA 00840682109260SEE09130160GA SEE09130162GA 00840682109260SEE09130162GA SEE09130165GA 00840682109260SEE09130165GA SEE09130166GA 00840682109260SEE09130166GA SEE09130170GA 00840682109260SEE09130170GA SEE09130171GA 00840682109260SEE09130171GA SEE09130172GA 00840682109260SEE09130172GA SEE09140176GA 00840682109260SEE09140176GA SEE09140177GA 00840682109260SEE09140177GA SEE09140178GA 00840682109260SEE09140178GA SEE09140180GA 00840682109260SEE09140180GA SEE09140181GA 00840682109260SEE09140181GA SEE09140182GA 00840682109260SEE09140182GA SEE09140183GA 00840682109260SEE09140183GA SEE09140184GA 00840682109260SEE09140184GA SEE09140185GA 00840682109260SEE09140185GA SEE09140186GA 00840682109260SEE09140186GA SEE09140187GA 00840682109260SEE09140187GA SEE09140189GA 00840682109260SEE09140189GA SEE09140190GA 00840682109260SEE09140190GA SEE09140191GA 00840682109260SEE09140191GA SEE09140193GA 00840682109260SEE09140193GA SEE09140194GA 00840682109260SEE09140194GA SEE09140196GA 00840682109260SEE09140196GA SEE09150200GA 00840682109260SEE09150200GA SEE09150202GA 00840682109260SEE09150202GA SEE09150203GA 00840682109260SEE09150203GA SEE09150204GA 00840682109260SEE09150204GA SEE09150205GA 00840682109260SEE09150205GA SEE09150206GA 00840682109260SEE09150206GA SEE09150207GA 00840682109260SEE09150207GA SEE09150210GA 00840682109260SEE09150210GA SEE09150218GA 00840682109260SEE09150218GA SEE09150219GA 00840682109260SEE09150219GA SEE09150220GA 00840682109260SEE09150220GA SEE09170221GA 00840682109260SEE09170221GA SEE09170222GA 00840682109260SEE09170222GA SEE09170223GA 00840682109260SEE09170223GA SEE09190226GA 00840682109260SEE09190226GA SEE09190227GA 00840682109260SEE09190227GA SEE09190228GA 00840682109260SEE09190228GA SEE09190229GA 00840682109260SEE09190229GA SEE09190230GA 00840682109260SEE09190230GA SEE09190231GA 00840682109260SEE09190231GA SEE09190232GA 00840682109260SEE09190232GA SEE09190233GA 00840682109260SEE09190233GA SEE09190235GA 00840682109260SEE09190235GA SEE09190236GA 00840682109260SEE09190236GA SEE09190238GA 00840682109260SEE09190238GA SEE09190239GA 00840682109260SEE09190239GA SEE09190240GA 00840682109260SEE09190240GA SEE09190241GA 00840682109260SEE09190241GA SEE09190242GA 00840682109260SEE09190242GA SEE09190243GA 00840682109260SEE09190243GA SEE09190244GA 00840682109260SEE09190244GA SEE09190246GA 00840682109260SEE09190246GA SEE09230251GA 00840682109260SEE09230251GA SEE09230252GA 00840682109260SEE09230252GA SEE09230253GA 00840682109260SEE09230253GA SEE09230254GA 00840682109260SEE09230254GA SEE09230255GA 00840682109260SEE09230255GA SEE09230256GA 00840682109260SEE09230256GA SEE09230257GA 00840682109260SEE09230257GA SEE09230258GA 00840682109260SEE09230258GA SEE09230259GA 00840682109260SEE09230259GA SEE09230260GA 00840682109260SEE09230260GA SEE09230261GA 00840682109260SEE09230261GA SEE09230262GA 00840682109260SEE09230262GA SEE09230263GA 00840682109260SEE09230263GA SEE09230264GA 00840682109260SEE09230264GA SEE09230265GA 00840682109260SEE09230265GA SEE09230266GA 00840682109260SEE09230266GA SEE09230267GA 00840682109260SEE09230267GA SEE09230268GA 00840682109260SEE09230268GA SEE09230273GA 00840682109260SEE09230273GA SEE09230279GA 00840682109260SEE09230279GA SEE09230281GA 00840682109260SEE09230281GA SEE09230282GA 00840682109260SEE09230282GA SEE09230284GA 00840682109260SEE09230284GA SEE09230286GA 00840682109260SEE09230286GA SEE09230287GA 00840682109260SEE09230287GA SEE09240290GA 00840682109260SEE09240290GA SEE09240291GA 00840682109260SEE09240291GA SEE09240293GA 00840682109260SEE09240293GA SEE09240294GA 00840682109260SEE09240294GA SEE09240295GA 00840682109260SEE09240295GA SEE09240296GA 00840682109260SEE09240296GA SEE09240297GA 00840682109260SEE09240297GA SEE09240299GA 00840682109260SEE09240299GA SEE09240300GA 00840682109260SEE09240300GA SEE09240301GA 00840682109260SEE09240301GA SEE09240302GA 00840682109260SEE09240302GA SEE09240303GA 00840682109260SEE09240303GA SEE09250304GA 00840682109260SEE09250304GA SEE09250305GA 00840682109260SEE09250305GA SEE09250306GA 00840682109260SEE09250306GA SEE09250307GA 00840682109260SEE09250307GA SEE09250308GA 00840682109260SEE09250308GA SEE09250309GA 00840682109260SEE09250309GA SEE09250310GA 00840682109260SEE09250310GA SEE09250311GA 00840682109260SEE09250311GA SEE09250312GA 00840682109260SEE09250312GA SEE09250313GA 00840682109260SEE09250313GA SEE09250314GA 00840682109260SEE09250314GA SEE09250315GA 00840682109260SEE09250315GA SEE09250316GA 00840682109260SEE09250316GA SEE09250317GA 00840682109260SEE09250317GA SEE09250318GA 00840682109260SEE09250318GA SEE09250319GA 00840682109260SEE09250319GA SEE09250320GA 00840682109260SEE09250320GA SEE09250321GA 00840682109260SEE09250321GA SEE09250322GA 00840682109260SEE09250322GA SEE09250323GA 00840682109260SEE09250323GA SEE09250324GA 00840682109260SEE09250324GA SEE09250325GA 00840682109260SEE09250325GA SEE09250326GA 00840682109260SEE09250326GA SEE09250327GA 00840682109260SEE09250327GA SEE09250328GA 00840682109260SEE09250328GA SEE09250329GA 00840682109260SEE09250329GA SEE09250330GA 00840682109260SEE09250330GA SEE09250331GA 00840682109260SEE09250331GA SEE09250332GA 00840682109260SEE09250332GA SEE09250333GA 00840682109260SEE09250333GA SEE09260334GA 00840682109260SEE09260334GA SEE09260335GA 00840682109260SEE09260335GA SEE09260336GA 00840682109260SEE09260336GA SEE09260337GA 00840682109260SEE09260337GA SEE09260338GA 00840682109260SEE09260338GA SEE09260339GA 00840682109260SEE09260339GA SEE09260340GA 00840682109260SEE09260340GA SEE09260341GA 00840682109260SEE09260341GA SEE09260342GA 00840682109260SEE09260342GA SEE09260343GA 00840682109260SEE09260343GA SEE09260344GA 00840682109260SEE09260344GA SEE09260345GA 00840682109260SEE09260345GA SEE09300346GA 00840682109260SEE09300346GA SEE09300348GA 00840682109260SEE09300348GA SEE09300349GA 00840682109260SEE09300349GA SEE09300350GA 00840682109260SEE09300350GA SEE09300351GA 00840682109260SEE09300351GA SEE09300352GA 00840682109260SEE09300352GA SEE09300353GA 00840682109260SEE09300353GA SEE09300354GA 00840682109260SEE09300354GA SEE09300355GA 00840682109260SEE09300355GA SEE09300356GA 00840682109260SEE09300356GA SEE09300357GA 00840682109260SEE09300357GA SEE09310358GA 00840682109260SEE09310358GA SEE09310360GA 00840682109260SEE09310360GA SEE09310362GA 00840682109260SEE09310362GA SEE09310364GA 00840682109260SEE09310364GA SEE09310365GA 00840682109260SEE09310365GA SEE09310366GA 00840682109260SEE09310366GA SEE09310367GA 00840682109260SEE09310367GA SEE09310368GA 00840682109260SEE09310368GA SEE09310369GA 00840682109260SEE09310369GA SEE09310370GA 00840682109260SEE09310370GA SEE09310371GA 00840682109260SEE09310371GA SEE09310372GA 00840682109260SEE09310372GA SEE09310373GA 00840682109260SEE09310373GA SEE09310374GA 00840682109260SEE09310374GA SEE09310375GA 00840682109260SEE09310375GA SEE09310376GA 00840682109260SEE09310376GA SEE09310377GA 00840682109260SEE09310377GA SEE09310378GA 00840682109260SEE09310378GA SEE09310379GA 00840682109260SEE09310379GA SEE09310380GA 00840682109260SEE09310380GA SEE09310381GA 00840682109260SEE09310381GA SEE09310382GA 00840682109260SEE09310382GA SEE09310383GA 00840682109260SEE09310383GA SEE09310384GA 00840682109260SEE09310384GA SEE09310387GA 00840682109260SEE09310387GA SEE09310389GA 00840682109260SEE09310389GA SEE09310390GA 00840682109260SEE09310390GA SEE09310391GA 00840682109260SEE09310391GA SEE09310392GA 00840682109260SEE09310392GA SEE09310393GA 00840682109260SEE09310393GA SEE09310394GA 00840682109260SEE09310394GA SEE09310396GA 00840682109260SEE09310396GA SEE09310397GA 00840682109260SEE09310397GA SEE09310398GA 00840682109260SEE09310398GA SEE09310400GA 00840682109260SEE09310400GA SEE09310401GA 00840682109260SEE09310401GA SEE09310402GA 00840682109260SEE09310402GA SEE09310403GA 00840682109260SEE09310403GA SEE09310404GA 00840682109260SEE09310404GA SEE09310405GA 00840682109260SEE09310405GA SEE09310406GA 00840682109260SEE09310406GA SEE09330408GA 00840682109260SEE09330408GA SEE09330409GA 00840682109260SEE09330409GA SEE09330410GA 00840682109260SEE09330410GA SEE09330411GA 00840682109260SEE09330411GA SEE09330412GA 00840682109260SEE09330412GA SEE09330413GA 00840682109260SEE09330413GA SEE09330416GA 00840682109260SEE09330416GA SEE09330417GA 00840682109260SEE09330417GA SEE09330421GA 00840682109260SEE09330421GA SEE09330422GA 00840682109260SEE09330422GA SEE09330423GA 00840682109260SEE09330423GA SEE09330425GA 00840682109260SEE09330425GA SEE09330427GA 00840682109260SEE09330427GA SEE09330428GA 00840682109260SEE09330428GA SEE09340429GA 00840682109260SEE09340429GA SEE09340430GA 00840682109260SEE09340430GA SEE09340431GA 00840682109260SEE09340431GA SEE09340432GA 00840682109260SEE09340432GA SEE09340433GA 00840682109260SEE09340433GA SEE09340434GA 00840682109260SEE09340434GA SEE09340435GA 00840682109260SEE09340435GA SEE09340437GA 00840682109260SEE09340437GA SEE09340438GA 00840682109260SEE09340438GA SEE09340439GA 00840682109260SEE09340439GA SEE09340440GA 00840682109260SEE09340440GA SEE09340441GA 00840682109260SEE09340441GA SEE09340442GA 00840682109260SEE09340442GA SEE09340443GA 00840682109260SEE09340443GA SEE09340444GA 00840682109260SEE09340444GA SEE09340445GA 00840682109260SEE09340445GA SEE09340446GA 00840682109260SEE09340446GA SEE09340447GA 00840682109260SEE09340447GA SEE09340448GA 00840682109260SEE09340448GA SEE09340449GA 00840682109260SEE09340449GA SEE09340450GA 00840682109260SEE09340450GA SEE09340452GA 00840682109260SEE09340452GA SEE09340453GA 00840682109260SEE09340453GA SEE09340454GA 00840682109260SEE09340454GA SEE09340455GA 00840682109260SEE09340455GA SEE09340456GA 00840682109260SEE09340456GA SEE09350457GA 00840682109260SEE09350457GA SEE09350458GA 00840682109260SEE09350458GA SEE09350460GA 00840682109260SEE09350460GA SEE09350461GA 00840682109260SEE09350461GA SEE09350462GA 00840682109260SEE09350462GA SEE09350463GA 00840682109260SEE09350463GA SEE09350466GA 00840682109260SEE09350466GA SEE09350467GA 00840682109260SEE09350467GA SEE09350468GA 00840682109260SEE09350468GA SEE09360471GA 00840682109260SEE09360471GA SEE09360472GA 00840682109260SEE09360472GA SEE09360473GA 00840682109260SEE09360473GA SEE09360479GA 00840682109260SEE09360479GA SEE09360481GA 00840682109260SEE09360481GA SEE09360482GA 00840682109260SEE09360482GA SEE09360483GA 00840682109260SEE09360483GA SEE09360486GA 00840682109260SEE09360486GA SEE09360487GA 00840682109260SEE09360487GA SEE09360488GA 00840682109260SEE09360488GA SEE09360489GA 00840682109260SEE09360489GA SEE09360490GA 00840682109260SEE09360490GA SEE09360491GA 00840682109260SEE09360491GA SEE09360492GA 00840682109260SEE09360492GA SEE09370493GA 00840682109260SEE09370493GA SEE09370494GA 00840682109260SEE09370494GA SEE09370495GA 00840682109260SEE09370495GA SEE09370496GA 00840682109260SEE09370496GA SEE09370497GA 00840682109260SEE09370497GA SEE09370498GA 00840682109260SEE09370498GA SEE09370499GA 00840682109260SEE09370499GA SEE09370500GA 00840682109260SEE09370500GA SEE09370501GA 00840682109260SEE09370501GA SEE09370502GA 00840682109260SEE09370502GA SEE09370503GA 00840682109260SEE09370503GA SEE09370504GA 00840682109260SEE09370504GA SEE09370505GA 00840682109260SEE09370505GA SEE09370506GA 00840682109260SEE09370506GA SEE09370507GA 00840682109260SEE09370507GA SEE09370508GA 00840682109260SEE09370508GA SEE09370509GA 00840682109260SEE09370509GA SEE09370510GA 00840682109260SEE09370510GA SEE09380515GA 00840682109260SEE09380515GA SEE09380516GA 00840682109260SEE09380516GA SEE09380517GA 00840682109260SEE09380517GA SEE09380518GA 00840682109260SEE09380518GA SEE09380519GA 00840682109260SEE09380519GA SEE09380522GA 00840682109260SEE09380522GA SEE09380523GA 00840682109260SEE09380523GA SEE09380525GA 00840682109260SEE09380525GA SEE09380526GA 00840682109260SEE09380526GA SEE09380527GA 00840682109260SEE09380527GA SEE09380528GA 00840682109260SEE09380528GA SEE09380529GA 00840682109260SEE09380529GA SEE09380530GA 00840682109260SEE09380530GA SEE09380531GA 00840682109260SEE09380531GA SEE09380532GA 00840682109260SEE09380532GA SEE09390534GA 00840682109260SEE09390534GA SEE09390535GA 00840682109260SEE09390535GA SEE09390536GA 00840682109260SEE09390536GA SEE09390537GA 00840682109260SEE09390537GA SEE09390538GA 00840682109260SEE09390538GA SEE09390539GA 00840682109260SEE09390539GA SEE09390540GA 00840682109260SEE09390540GA SEE09390541GA 00840682109260SEE09390541GA SEE09390542GA 00840682109260SEE09390542GA SEE09390543GA 00840682109260SEE09390543GA SEE09390544GA 00840682109260SEE09390544GA SEE09390545GA 00840682109260SEE09390545GA SEE09390547GA 00840682109260SEE09390547GA SEE09390548GA 00840682109260SEE09390548GA SEE09400551GA 00840682109260SEE09400551GA SEE09410552GA 00840682109260SEE09410552GA SEE09410553GA 00840682109260SEE09410553GA SEE09410554GA 00840682109260SEE09410554GA SEE09410555GA 00840682109260SEE09410555GA SEE09410556GA 00840682109260SEE09410556GA SEE09410557GA 00840682109260SEE09410557GA SEE09410558GA 00840682109260SEE09410558GA SEE09410559GA 00840682109260SEE09410559GA SEE09410560GA 00840682109260SEE09410560GA SEE09410561GA 00840682109260SEE09410561GA SEE09410562GA 00840682109260SEE09410562GA SEE09410563GA 00840682109260SEE09410563GA SEE09410564GA 00840682109260SEE09410564GA SEE09410565GA 00840682109260SEE09410565GA SEE09410566GA 00840682109260SEE09410566GA SEE09410567GA 00840682109260SEE09410567GA SEE09410568GA 00840682109260SEE09410568GA SEE09410569GA 00840682109260SEE09410569GA SEE09410571GA 00840682109260SEE09410571GA SEE09410572GA 00840682109260SEE09410572GA SEE09410573GA 00840682109260SEE09410573GA SEE09410574GA 00840682109260SEE09410574GA SEE09410575GA 00840682109260SEE09410575GA SEE09410576GA 00840682109260SEE09410576GA SEE09410577GA 00840682109260SEE09410577GA SEE09410578GA 00840682109260SEE09410578GA SEE09410579GA 00840682109260SEE09410579GA SEE09410580GA 00840682109260SEE09410580GA SEE09410581GA 00840682109260SEE09410581GA SEE09410582GA 00840682109260SEE09410582GA SEE09410583GA 00840682109260SEE09410583GA SEE09410584GA 00840682109260SEE09410584GA SEE09410586GA 00840682109260SEE09410586GA SEE09410587GA 00840682109260SEE09410587GA SEE09410590GA 00840682109260SEE09410590GA SEE09420592GA 00840682109260SEE09420592GA SEE09420593GA 00840682109260SEE09420593GA SEE09420594GA 00840682109260SEE09420594GA SEE09420595GA 00840682109260SEE09420595GA SEE09420597GA 00840682109260SEE09420597GA SEE09420598GA 00840682109260SEE09420598GA SEE09430599GA 00840682109260SEE09430599GA SEE09430601GA 00840682109260SEE09430601GA SEE09430602GA 00840682109260SEE09430602GA SEE09430604GA 00840682109260SEE09430604GA SEE09430605GA 00840682109260SEE09430605GA SEE09430606GA 00840682109260SEE09430606GA SEE09430607GA 00840682109260SEE09430607GA SEE09430608GA 00840682109260SEE09430608GA SEE09430609GA 00840682109260SEE09430609GA SEE09430611GA 00840682109260SEE09430611GA SEE09430613GA 00840682109260SEE09430613GA SEE09430615GA 00840682109260SEE09430615GA SEE09430616GA 00840682109260SEE09430616GA SEE09430617GA 00840682109260SEE09430617GA SEE09430618GA 00840682109260SEE09430618GA SEE09430619GA 00840682109260SEE09430619GA SEE09430620GA 00840682109260SEE09430620GA SEE09430621GA 00840682109260SEE09430621GA SEE09430622GA 00840682109260SEE09430622GA SEE09450623GA 00840682109260SEE09450623GA SEE09450624GA 00840682109260SEE09450624GA SEE09450625GA 00840682109260SEE09450625GA SEE09450626GA 00840682109260SEE09450626GA SEE09450627GA 00840682109260SEE09450627GA SEE09450628GA 00840682109260SEE09450628GA SEE09450629GA 00840682109260SEE09450629GA SEE09450630GA 00840682109260SEE09450630GA SEE09460632GA 00840682109260SEE09460632GA SEE09460633GA 00840682109260SEE09460633GA SEE09460634GA 00840682109260SEE09460634GA SEE09460635GA 00840682109260SEE09460635GA SEE09460636GA 00840682109260SEE09460636GA SEE09460637GA 00840682109260SEE09460637GA SEE09460638GA 00840682109260SEE09460638GA SEE09460639GA 00840682109260SEE09460639GA SEE09460640GA 00840682109260SEE09460640GA SEE09460641GA 00840682109260SEE09460641GA SEE09460643GA 00840682109260SEE09460643GA SEE09460644GA 00840682109260SEE09460644GA SEE09460646GA 00840682109260SEE09460646GA SEE09460648GA 00840682109260SEE09460648GA SEE09460649GA 00840682109260SEE09460649GA SEE09460650GA 00840682109260SEE09460650GA SEE09460651GA 00840682109260SEE09460651GA SEE09460652GA 00840682109260SEE09460652GA SEE09460654GA 00840682109260SEE09460654GA SEE09470656GA 00840682109260SEE09470656GA SEE09470657GA 00840682109260SEE09470657GA SEE09470663GA 00840682109260SEE09470663GA SEE09470665GA 00840682109260SEE09470665GA SEE09470668GA 00840682109260SEE09470668GA SEE09470669GA 00840682109260SEE09470669GA SEE09470670GA 00840682109260SEE09470670GA SEE09470671GA 00840682109260SEE09470671GA SEE09470672GA 00840682109260SEE09470672GA SEE09470673GA 00840682109260SEE09470673GA SEE09470674GA 00840682109260SEE09470674GA SEE09470676GA 00840682109260SEE09470676GA SEE09470677GA 00840682109260SEE09470677GA SEE09470680GA 00840682109260SEE09470680GA SEE09470681GA 00840682109260SEE09470681GA SEE09470682GA 00840682109260SEE09470682GA SEE09470683GA 00840682109260SEE09470683GA SEE09470684GA 00840682109260SEE09470684GA SEE09470685GA 00840682109260SEE09470685GA SEE09470686GA 00840682109260SEE09470686GA SEE09470687GA 00840682109260SEE09470687GA SEE09470688GA 00840682109260SEE09470688GA SEE09470689GA 00840682109260SEE09470689GA SEE09470690GA 00840682109260SEE09470690GA SEE09480692GA 00840682109260SEE09480692GA SEE09480693GA 00840682109260SEE09480693GA SEE09480694GA 00840682109260SEE09480694GA SEE09480695GA 00840682109260SEE09480695GA SEE09480721GA 00840682109260SEE09480721GA SEE09480722GA 00840682109260SEE09480722GA SEE09480723GA 00840682109260SEE09480723GA SEE09480724GA 00840682109260SEE09480724GA SEE09480725GA 00840682109260SEE09480725GA SEE09480726GA 00840682109260SEE09480726GA SEE09480727GA 00840682109260SEE09480727GA SEE09480729GA 00840682109260SEE09480729GA SEE09480730GA 00840682109260SEE09480730GA SEE09480731GA 00840682109260SEE09480731GA SEE09480737GA 00840682109260SEE09480737GA SEE09480738GA 00840682109260SEE09480738GA SEE09480740GA 00840682109260SEE09480740GA SEE09480741GA 00840682109260SEE09480741GA SEE09480742GA 00840682109260SEE09480742GA SEE09480743GA 00840682109260SEE09480743GA SEE09480744GA 00840682109260SEE09480744GA SEE09480745GA 00840682109260SEE09480745GA SEE09480746GA 00840682109260SEE09480746GA SEE09480747GA 00840682109260SEE09480747GA SEE09480748GA 00840682109260SEE09480748GA SEE09490750GA 00840682109260SEE09490750GA SEE09490752GA 00840682109260SEE09490752GA SEE09490754GA 00840682109260SEE09490754GA SEE09490755GA 00840682109260SEE09490755GA SEE09490756GA 00840682109260SEE09490756GA SEE09490757GA 00840682109260SEE09490757GA SEE09490758GA 00840682109260SEE09490758GA SEE09490759GA 00840682109260SEE09490759GA SEE09490760GA 00840682109260SEE09490760GA SEE09490761GA 00840682109260SEE09490761GA SEE09510763GA 00840682109260SEE09510763GA SEE09510764GA 00840682109260SEE09510764GA SEE09510765GA 00840682109260SEE09510765GA SEE09510766GA 00840682109260SEE09510766GA SEE09510767GA 00840682109260SEE09510767GA SEE09510768GA 00840682109260SEE09510768GA SEE09510769GA 00840682109260SEE09510769GA SEE09510770GA 00840682109260SEE09510770GA SEE09510772GA 00840682109260SEE09510772GA SEE09510773GA 00840682109260SEE09510773GA SEE09510776GA 00840682109260SEE09510776GA SEE09510777GA 00840682109260SEE09510777GA SEE09510778GA 00840682109260SEE09510778GA SEE09510780GA 00840682109260SEE09510780GA SEE09510782GA 00840682109260SEE09510782GA SEE09510783GA 00840682109260SEE09510783GA SEE09510785GA 00840682109260SEE09510785GA SEE09510787GA 00840682109260SEE09510787GA SEE09510788GA 00840682109260SEE09510788GA SEE09510789GA 00840682109260SEE09510789GA SEE09510790GA 00840682109260SEE09510790GA SEE09510792GA 00840682109260SEE09510792GA SEE09510793GA 00840682109260SEE09510793GA SEE09510794GA 00840682109260SEE09510794GA SEE09510795GA 00840682109260SEE09510795GA SEE09510796GA 00840682109260SEE09510796GA SEE09510798GA 00840682109260SEE09510798GA SEE09510801GA 00840682109260SEE09510801GA SEE09510802GA 00840682109260SEE09510802GA SEE09510803GA 00840682109260SEE09510803GA SEE09510804GA 00840682109260SEE09510804GA SEE09510805GA 00840682109260SEE09510805GA SEE09510806GA 00840682109260SEE09510806GA SEE09510807GA 00840682109260SEE09510807GA SEE09510808GA 00840682109260SEE09510808GA SEE09510809GA 00840682109260SEE09510809GA SEE09510811GA 00840682109260SEE09510811GA SEE09510812GA 00840682109260SEE09510812GA SEE09510814GA 00840682109260SEE09510814GA SEE09510816GA 00840682109260SEE09510816GA SEE09510817GA 00840682109260SEE09510817GA SEE09510818GA 00840682109260SEE09510818GA SEE09510819GA 00840682109260SEE09510819GA SEE09510820GA 00840682109260SEE09510820GA SEE09510821GA 00840682109260SEE09510821GA SEE09510822GA 00840682109260SEE09510822GA SEE09510823GA 00840682109260SEE09510823GA SEE09510824GA 00840682109260SEE09510824GA SEE09510825GA 00840682109260SEE09510825GA SEE09510828GA 00840682109260SEE09510828GA SEE09510829GA 00840682109260SEE09510829GA SEE09510830GA 00840682109260SEE09510830GA SEE09510831GA 00840682109260SEE09510831GA SEE09510832GA 00840682109260SEE09510832GA SEE09510833GA 00840682109260SEE09510833GA SEE09510834GA 00840682109260SEE09510834GA SEE09530836GA 00840682109260SEE09530836GA SEE09530837GA 00840682109260SEE09530837GA SEE09530838GA 00840682109260SEE09530838GA SEE09530839GA 00840682109260SEE09530839GA SEE09530840GA 00840682109260SEE09530840GA SEE09530842GA 00840682109260SEE09530842GA SEE09530843GA 00840682109260SEE09530843GA SEE09530844GA 00840682109260SEE09530844GA SEE09530845GA 00840682109260SEE09530845GA SEE09530846GA 00840682109260SEE09530846GA SEE09530847GA 00840682109260SEE09530847GA SEE09530848GA 00840682109260SEE09530848GA SEE09530849GA 00840682109260SEE09530849GA SEE09530850GA 00840682109260SEE09530850GA SEE09530852GA 00840682109260SEE09530852GA SEE09530853GA 00840682109260SEE09530853GA SEE09530854GA 00840682109260SEE09530854GA SEE09530855GA 00840682109260SEE09530855GA SEE09530856GA 00840682109260SEE09530856GA SEE09530857GA 00840682109260SEE09530857GA SEE09530858GA 00840682109260SEE09530858GA SEE09530859GA 00840682109260SEE09530859GA SEE09530860GA 00840682109260SEE09530860GA SEE09530861GA 00840682109260SEE09530861GA SEE09530862GA 00840682109260SEE09530862GA SEE09530863GA 00840682109260SEE09530863GA SEE09530864GA 00840682109260SEE09530864GA SEE09530865GA 00840682109260SEE09530865GA SEE09530866GA 00840682109260SEE09530866GA SEE09530867GA 00840682109260SEE09530867GA SEE09530869GA 00840682109260SEE09530869GA SEE09530870GA 00840682109260SEE09530870GA SEE09530871GA 00840682109260SEE09530871GA SEE09530872GA 00840682109260SEE09530872GA SEE09530873GA 00840682109260SEE09530873GA SEE09530874GA 00840682109260SEE09530874GA SEE09530875GA 00840682109260SEE09530875GA SEE09530876GA 00840682109260SEE09530876GA SEE09530878GA 00840682109260SEE09530878GA SEE09530880GA 00840682109260SEE09530880GA SEE09530881GA 00840682109260SEE09530881GA SEE09530882GA 00840682109260SEE09530882GA SEE09530884GA 00840682109260SEE09530884GA SEE09530886GA 00840682109260SEE09530886GA SEE09530887GA 00840682109260SEE09530887GA SEE09530888GA 00840682109260SEE09530888GA SEE09530889GA 00840682109260SEE09530889GA SEE09530893GA 00840682109260SEE09530893GA SEE09530896GA 00840682109260SEE09530896GA SEE09530899GA 00840682109260SEE09530899GA SEE09530900GA 00840682109260SEE09530900GA SEE09530901GA 00840682109260SEE09530901GA SEE09530903GA 00840682109260SEE09530903GA SEE09530904GA 00840682109260SEE09530904GA SEE09530905GA 00840682109260SEE09530905GA SEE09530906GA 00840682109260SEE09530906GA SEE09530907GA 00840682109260SEE09530907GA SEE09530908GA 00840682109260SEE09530908GA SEE09530910GA 00840682109260SEE09530910GA SEE09530911GA 00840682109260SEE09530911GA SEE09530912GA 00840682109260SEE09530912GA SEE09530914GA 00840682109260SEE09530914GA SEE09530915GA 00840682109260SEE09530915GA SEE10080925GA 00840682109260SEE10080925GA SEE10080927GA 00840682109260SEE10080927GA SEE10080928GA 00840682109260SEE10080928GA SEE10080931GA 00840682109260SEE10080931GA SEE10090946GA 00840682109260SEE10090946GA SEE10090948GA 00840682109260SEE10090948GA SEE10090951GA 00840682109260SEE10090951GA SEE10090955GA 00840682109260SEE10090955GA SEE10090957GA 00840682109260SEE10090957GA SEE10090962GA 00840682109260SEE10090962GA SEE10100969GA 00840682109260SEE10100969GA SEE10100972GA 00840682109260SEE10100972GA SEE10100973GA 00840682109260SEE10100973GA SEE10100975GA 00840682109260SEE10100975GA SEE10100976GA 00840682109260SEE10100976GA SEE10100977GA 00840682109260SEE10100977GA SEE10100978GA 00840682109260SEE10100978GA SEE10100979GA 00840682109260SEE10100979GA SEE10100980GA 00840682109260SEE10100980GA SEE10100982GA 00840682109260SEE10100982GA SEE10100983GA 00840682109260SEE10100983GA SEE10120001GA 00840682109260SEE10120001GA SEE10120002GA 00840682109260SEE10120002GA SEE10120003GA 00840682109260SEE10120003GA SEE10120004GA 00840682109260SEE10120004GA SEE10120005GA 00840682109260SEE10120005GA SEE10120006GA 00840682109260SEE10120006GA SEE10120007GA 00840682109260SEE10120007GA SEE10120008GA 00840682109260SEE10120008GA SEE10120010GA 00840682109260SEE10120010GA SEE10120011GA 00840682109260SEE10120011GA SEE10120012GA 00840682109260SEE10120012GA SEE10120013GA 00840682109260SEE10120013GA SEE10120014GA 00840682109260SEE10120014GA SEE10120015GA 00840682109260SEE10120015GA SEE10120016GA 00840682109260SEE10120016GA SEE10120017GA 00840682109260SEE10120017GA SEE10120018GA 00840682109260SEE10120018GA SEE10120019GA 00840682109260SEE10120019GA SEE10120020GA 00840682109260SEE10120020GA SEE10120021GA 00840682109260SEE10120021GA SEE10120022GA 00840682109260SEE10120022GA SEE10120023GA 00840682109260SEE10120023GA SEE10120024GA 00840682109260SEE10120024GA SEE10120026GA 00840682109260SEE10120026GA SEE10120027GA 00840682109260SEE10120027GA SEE10120028GA 00840682109260SEE10120028GA SEE10120029GA 00840682109260SEE10120029GA SEE10120030GA 00840682109260SEE10120030GA SEE10120032GA 00840682109260SEE10120032GA SEE10120033GA 00840682109260SEE10120033GA SEE10120034GA 00840682109260SEE10120034GA SEE10120988GA 00840682109260SEE10120988GA SEE10120989GA 00840682109260SEE10120989GA SEE10120991GA 00840682109260SEE10120991GA SEE10120992GA 00840682109260SEE10120992GA SEE10120993GA 00840682109260SEE10120993GA SEE10120994GA 00840682109260SEE10120994GA SEE10120995GA 00840682109260SEE10120995GA SEE10120997GA 00840682109260SEE10120997GA SEE10120999GA 00840682109260SEE10120999GA SEE10121000GA 00840682109260SEE10121000GA SEE10121001GA 00840682109260SEE10121001GA SEE10121002GA 00840682109260SEE10121002GA SEE10121003GA 00840682109260SEE10121003GA SEE10121005GA 00840682109260SEE10121005GA SEE10121006GA 00840682109260SEE10121006GA SEE10121007GA 00840682109260SEE10121007GA SEE10121008GA 00840682109260SEE10121008GA SEE10121009GA 00840682109260SEE10121009GA SEE10121010GA 00840682109260SEE10121010GA SEE10121011GA 00840682109260SEE10121011GA SEE10121012GA 00840682109260SEE10121012GA SEE10121013GA 00840682109260SEE10121013GA SEE10121016GA 00840682109260SEE10121016GA SEE10121018GA 00840682109260SEE10121018GA SEE10121019GA 00840682109260SEE10121019GA SEE10121020GA 00840682109260SEE10121020GA SEE10121021GA 00840682109260SEE10121021GA SEE10121022GA 00840682109260SEE10121022GA SEE10121024GA 00840682109260SEE10121024GA SEE10121025GA 00840682109260SEE10121025GA SEE10121026GA 00840682109260SEE10121026GA SEE10121027GA 00840682109260SEE10121027GA SEE10121028GA 00840682109260SEE10121028GA SEE10121029GA 00840682109260SEE10121029GA SEE10130036GA 00840682109260SEE10130036GA SEE10130038GA 00840682109260SEE10130038GA SEE10130039GA 00840682109260SEE10130039GA SEE10130040GA 00840682109260SEE10130040GA SEE10130042GA 00840682109260SEE10130042GA SEE10130043GA 00840682109260SEE10130043GA SEE10130044GA 00840682109260SEE10130044GA SEE10130045GA 00840682109260SEE10130045GA SEE10130046GA 00840682109260SEE10130046GA SEE10130048GA 00840682109260SEE10130048GA SEE10130049GA 00840682109260SEE10130049GA SEE10130050GA 00840682109260SEE10130050GA SEE10130051GA 00840682109260SEE10130051GA SEE10130053GA 00840682109260SEE10130053GA SEE10130055GA 00840682109260SEE10130055GA SEE10130056GA 00840682109260SEE10130056GA SEE10130059GA 00840682109260SEE10130059GA SEE10130060GA 00840682109260SEE10130060GA SEE10130061GA 00840682109260SEE10130061GA SEE10130062GA 00840682109260SEE10130062GA SEE10130063GA 00840682109260SEE10130063GA SEE10130064GA 00840682109260SEE10130064GA SEE10130065GA 00840682109260SEE10130065GA SEE10130067GA 00840682109260SEE10130067GA SEE10130068GA 00840682109260SEE10130068GA SEE10130069GA 00840682109260SEE10130069GA SEE10130070GA 00840682109260SEE10130070GA SEE10130071GA 00840682109260SEE10130071GA SEE10130073GA 00840682109260SEE10130073GA SEE10130074GA 00840682109260SEE10130074GA SEE10130075GA 00840682109260SEE10130075GA SEE10130076GA 00840682109260SEE10130076GA SEE10250079GA 00840682109260SEE10250079GA SEE10250080GA 00840682109260SEE10250080GA SEE10250081GA 00840682109260SEE10250081GA SEE10250082GA 00840682109260SEE10250082GA SEE10250083GA 00840682109260SEE10250083GA SEE10250084GA 00840682109260SEE10250084GA SEE10250085GA 00840682109260SEE10250085GA SEE10250088GA 00840682109260SEE10250088GA SEE10250089GA 00840682109260SEE10250089GA SEE10250090GA 00840682109260SEE10250090GA SEE10250091GA 00840682109260SEE10250091GA SEE10250092GA 00840682109260SEE10250092GA SEE10250093GA 00840682109260SEE10250093GA SEE10250094GA 00840682109260SEE10250094GA SEE10250095GA 00840682109260SEE10250095GA SEE10250096GA 00840682109260SEE10250096GA SEE10250097GA 00840682109260SEE10250097GA SEE10250098GA 00840682109260SEE10250098GA SEE10250099GA 00840682109260SEE10250099GA SEE10250100GA 00840682109260SEE10250100GA SEE10250101GA 00840682109260SEE10250101GA SEE10250102GA 00840682109260SEE10250102GA SEE10250104GA 00840682109260SEE10250104GA SEE10250105GA 00840682109260SEE10250105GA SEE10250106GA 00840682109260SEE10250106GA SEE10250107GA 00840682109260SEE10250107GA SEE10250108GA 00840682109260SEE10250108GA SEE10250109GA 00840682109260SEE10250109GA SEE10250110GA 00840682109260SEE10250110GA SEE10250111GA 00840682109260SEE10250111GA SEE10250112GA 00840682109260SEE10250112GA SEE10250113GA 00840682109260SEE10250113GA SEE10250114GA 00840682109260SEE10250114GA SEE10250115GA 00840682109260SEE10250115GA SEE10250116GA 00840682109260SEE10250116GA SEE10250117GA 00840682109260SEE10250117GA SEE10250118GA 00840682109260SEE10250118GA SEE10250120GA 00840682109260SEE10250120GA SEE10250121GA 00840682109260SEE10250121GA SEE10250122GA 00840682109260SEE10250122GA SEE10250124GA 00840682109260SEE10250124GA SEE10250125GA 00840682109260SEE10250125GA SEE10250126GA 00840682109260SEE10250126GA SEE10250127GA 00840682109260SEE10250127GA SEE10250129GA 00840682109260SEE10250129GA SEE10250130GA 00840682109260SEE10250130GA SEE10250133GA 00840682109260SEE10250133GA SEE10250134GA 00840682109260SEE10250134GA SEE10250136GA 00840682109260SEE10250136GA SEE10250139GA 00840682109260SEE10250139GA SEE10250140GA 00840682109260SEE10250140GA SEE10250142GA 00840682109260SEE10250142GA SEE10250143GA 00840682109260SEE10250143GA SEE10250144GA 00840682109260SEE10250144GA SEE10250145GA 00840682109260SEE10250145GA SEE10250148GA 00840682109260SEE10250148GA SEE10250151GA 00840682109260SEE10250151GA SEE10250153GA 00840682109260SEE10250153GA SEE10250154GA 00840682109260SEE10250154GA SEE10250156GA 00840682109260SEE10250156GA SEE10250159GA 00840682109260SEE10250159GA SEE10250160GA 00840682109260SEE10250160GA SEE10250161GA 00840682109260SEE10250161GA SEE10250167GA 00840682109260SEE10250167GA SEE10250168GA 00840682109260SEE10250168GA SEE10250169GA 00840682109260SEE10250169GA SEE10250172GA 00840682109260SEE10250172GA SEE10250184GA 00840682109260SEE10250184GA SEE10250187GA 00840682109260SEE10250187GA SEE10260194GA 00840682109260SEE10260194GA SEE10260195GA 00840682109260SEE10260195GA SEE10260196GA 00840682109260SEE10260196GA SEE10260197GA 00840682109260SEE10260197GA SEE10260198GA 00840682109260SEE10260198GA SEE10260200GA 00840682109260SEE10260200GA SEE10260202GA 00840682109260SEE10260202GA SEE10260203GA 00840682109260SEE10260203GA SEE10260205GA 00840682109260SEE10260205GA SEE10260209GA 00840682109260SEE10260209GA SEE10260211GA 00840682109260SEE10260211GA SEE10260214GA 00840682109260SEE10260214GA SEE10260215GA 00840682109260SEE10260215GA SEE10260216GA 00840682109260SEE10260216GA SEE10260217GA 00840682109260SEE10260217GA SEE10260218GA 00840682109260SEE10260218GA SEE10260219GA 00840682109260SEE10260219GA SEE10260220GA 00840682109260SEE10260220GA SEE10260221GA 00840682109260SEE10260221GA SEE10260222GA 00840682109260SEE10260222GA SEE10260223GA 00840682109260SEE10260223GA SEE10260224GA 00840682109260SEE10260224GA SEE10260225GA 00840682109260SEE10260225GA SEE10260226GA 00840682109260SEE10260226GA SEE10260227GA 00840682109260SEE10260227GA SEE10260228GA 00840682109260SEE10260228GA SEE10260229GA 00840682109260SEE10260229GA SEE10260230GA 00840682109260SEE10260230GA SEE10260231GA 00840682109260SEE10260231GA SEE10260232GA 00840682109260SEE10260232GA SEE10260233GA 00840682109260SEE10260233GA SEE10260234GA 00840682109260SEE10260234GA SEE10260235GA 00840682109260SEE10260235GA SEE10260236GA 00840682109260SEE10260236GA SEE10260237GA 00840682109260SEE10260237GA SEE10260238GA 00840682109260SEE10260238GA SEE10260239GA 00840682109260SEE10260239GA SEE10270240GA 00840682109260SEE10270240GA SEE10270241GA 00840682109260SEE10270241GA SEE10270242GA 00840682109260SEE10270242GA SEE10270243GA 00840682109260SEE10270243GA SEE10270244GA 00840682109260SEE10270244GA SEE10270245GA 00840682109260SEE10270245GA SEE10270246GA 00840682109260SEE10270246GA SEE10270247GA 00840682109260SEE10270247GA SEE10270248GA 00840682109260SEE10270248GA SEE10270249GA 00840682109260SEE10270249GA SEE10270251GA 00840682109260SEE10270251GA SEE10270252GA 00840682109260SEE10270252GA SEE10270253GA 00840682109260SEE10270253GA SEE10270254GA 00840682109260SEE10270254GA SEE10270256GA 00840682109260SEE10270256GA SEE10270257GA 00840682109260SEE10270257GA SEE10270259GA 00840682109260SEE10270259GA SEE10270260GA 00840682109260SEE10270260GA SEE10270261GA 00840682109260SEE10270261GA SEE10270263GA 00840682109260SEE10270263GA SEE10270272GA 00840682109260SEE10270272GA SEE10270273GA 00840682109260SEE10270273GA SEE10270274GA 00840682109260SEE10270274GA SEE10270275GA 00840682109260SEE10270275GA SEE10310278GA 00840682109260SEE10310278GA SEE10310279GA 00840682109260SEE10310279GA SEE10310281GA 00840682109260SEE10310281GA SEE10310283GA 00840682109260SEE10310283GA SEE10310284GA 00840682109260SEE10310284GA SEE10310286GA 00840682109260SEE10310286GA SEE10390289GA 00840682109260SEE10390289GA SEE10390290GA 00840682109260SEE10390290GA SEE10390291GA 00840682109260SEE10390291GA SEE10390293GA 00840682109260SEE10390293GA SEE10390294GA 00840682109260SEE10390294GA SEE10390296GA 00840682109260SEE10390296GA SEE10390297GA 00840682109260SEE10390297GA SEE10390298GA 00840682109260SEE10390298GA SEE10390299GA 00840682109260SEE10390299GA SEE10390305GA 00840682109260SEE10390305GA SEE10390306GA 00840682109260SEE10390306GA SEE10390307GA 00840682109260SEE10390307GA SEE10390308GA 00840682109260SEE10390308GA SEE10390309GA 00840682109260SEE10390309GA SEE10390310GA 00840682109260SEE10390310GA SEE10390311GA 00840682109260SEE10390311GA SEE10390312GA 00840682109260SEE10390312GA SEE10390313GA 00840682109260SEE10390313GA SEE10390314GA 00840682109260SEE10390314GA SEE10390315GA 00840682109260SEE10390315GA SEE10390316GA 00840682109260SEE10390316GA SEE10390317GA 00840682109260SEE10390317GA SEE10390318GA 00840682109260SEE10390318GA SEE10390319GA 00840682109260SEE10390319GA SEE10390320GA 00840682109260SEE10390320GA SEE10390321GA 00840682109260SEE10390321GA SEE10390322GA 00840682109260SEE10390322GA SEE10390323GA 00840682109260SEE10390323GA SEE10390324GA 00840682109260SEE10390324GA SEE10390325GA 00840682109260SEE10390325GA SEE10390326GA 00840682109260SEE10390326GA SEE10390327GA 00840682109260SEE10390327GA SEE10390328GA 00840682109260SEE10390328GA SEE10390329GA 00840682109260SEE10390329GA SEE10390330GA 00840682109260SEE10390330GA SEE10390331GA 00840682109260SEE10390331GA SEE10390332GA 00840682109260SEE10390332GA SEE10390333GA 00840682109260SEE10390333GA SEE10390334GA 00840682109260SEE10390334GA SEE10390335GA 00840682109260SEE10390335GA SEE10390336GA 00840682109260SEE10390336GA SEE10390337GA 00840682109260SEE10390337GA SEE10390338GA 00840682109260SEE10390338GA SEE10390339GA 00840682109260SEE10390339GA SEE10390340GA 00840682109260SEE10390340GA SEE10390341GA 00840682109260SEE10390341GA SEE10390342GA 00840682109260SEE10390342GA SEE10390343GA 00840682109260SEE10390343GA SEE10390344GA 00840682109260SEE10390344GA SEE10390345GA 00840682109260SEE10390345GA SEE10390346GA 00840682109260SEE10390346GA SEE10390347GA 00840682109260SEE10390347GA SEE10390348GA 00840682109260SEE10390348GA SEE10390349GA 00840682109260SEE10390349GA SEE10390350GA 00840682109260SEE10390350GA SEE10390351GA 00840682109260SEE10390351GA SEE10390352GA 00840682109260SEE10390352GA SEE10390353GA 00840682109260SEE10390353GA SEE10390354GA 00840682109260SEE10390354GA SEE10390355GA 00840682109260SEE10390355GA SEE10390356GA 00840682109260SEE10390356GA SEE10390357GA 00840682109260SEE10390357GA SEE10390358GA 00840682109260SEE10390358GA SEE10390359GA 00840682109260SEE10390359GA SEE10390360GA 00840682109260SEE10390360GA SEE10390361GA 00840682109260SEE10390361GA SEE10400363GA 00840682109260SEE10400363GA SEE10400364GA 00840682109260SEE10400364GA SEE10400365GA 00840682109260SEE10400365GA SEE10400366GA 00840682109260SEE10400366GA SEE10400369GA 00840682109260SEE10400369GA SEE10400370GA 00840682109260SEE10400370GA SEE10400371GA 00840682109260SEE10400371GA SEE10400372GA 00840682109260SEE10400372GA SEE10400373GA 00840682109260SEE10400373GA SEE10400374GA 00840682109260SEE10400374GA SEE10400375GA 00840682109260SEE10400375GA SEE10400376GA 00840682109260SEE10400376GA SEE10400377GA 00840682109260SEE10400377GA SEE10400381GA 00840682109260SEE10400381GA SEE10400382GA 00840682109260SEE10400382GA SEE10400383GA 00840682109260SEE10400383GA SEE10400384GA 00840682109260SEE10400384GA SEE10400385GA 00840682109260SEE10400385GA SEE10400386GA 00840682109260SEE10400386GA SEE10400387GA 00840682109260SEE10400387GA SEE10400392GA 00840682109260SEE10400392GA SEE10400393GA 00840682109260SEE10400393GA SEE10400394GA 00840682109260SEE10400394GA SEE10400395GA 00840682109260SEE10400395GA SEE10400396GA 00840682109260SEE10400396GA SEE10400397GA 00840682109260SEE10400397GA SEE10400398GA 00840682109260SEE10400398GA SEE10400399GA 00840682109260SEE10400399GA SEE10400401GA 00840682109260SEE10400401GA SEE10400403GA 00840682109260SEE10400403GA SEE10400404GA 00840682109260SEE10400404GA SEE10400405GA 00840682109260SEE10400405GA SEE10400406GA 00840682109260SEE10400406GA SEE10400407GA 00840682109260SEE10400407GA SEE10400408GA 00840682109260SEE10400408GA SEE10400410GA 00840682109260SEE10400410GA SEE10400411GA 00840682109260SEE10400411GA SEE10400412GA 00840682109260SEE10400412GA SEE10400413GA 00840682109260SEE10400413GA SEE10400416GA 00840682109260SEE10400416GA SEE10400417GA 00840682109260SEE10400417GA SEE10400418GA 00840682109260SEE10400418GA SEE10400419GA 00840682109260SEE10400419GA SEE10400421GA 00840682109260SEE10400421GA SEE10400422GA 00840682109260SEE10400422GA SEE10400423GA 00840682109260SEE10400423GA SEE10400424GA 00840682109260SEE10400424GA SEE10400426GA 00840682109260SEE10400426GA SEE10400427GA 00840682109260SEE10400427GA SEE10400429GA 00840682109260SEE10400429GA SEE10400431GA 00840682109260SEE10400431GA SEE10400432GA 00840682109260SEE10400432GA SEE10400433GA 00840682109260SEE10400433GA SEE10400434GA 00840682109260SEE10400434GA SEE10400435GA 00840682109260SEE10400435GA SEE10400438GA 00840682109260SEE10400438GA SEE10400439GA 00840682109260SEE10400439GA SEE10400440GA 00840682109260SEE10400440GA SEE10400441GA 00840682109260SEE10400441GA SEE10400442GA 00840682109260SEE10400442GA SEE10400443GA 00840682109260SEE10400443GA SEE10400444GA 00840682109260SEE10400444GA SEE10400445GA 00840682109260SEE10400445GA SEE10400446GA 00840682109260SEE10400446GA SEE10400451GA 00840682109260SEE10400451GA SEE10400452GA 00840682109260SEE10400452GA SEE10400453GA 00840682109260SEE10400453GA SEE10400454GA 00840682109260SEE10400454GA SEE10400455GA 00840682109260SEE10400455GA SEE10400456GA 00840682109260SEE10400456GA SEE10400457GA 00840682109260SEE10400457GA SEE10400458GA 00840682109260SEE10400458GA SEE10400459GA 00840682109260SEE10400459GA SEE10400460GA 00840682109260SEE10400460GA SEE10400461GA 00840682109260SEE10400461GA SEE10400462GA 00840682109260SEE10400462GA SEE10400463GA 00840682109260SEE10400463GA SEE10410464GA 00840682109260SEE10410464GA SEE10410465GA 00840682109260SEE10410465GA SEE10410466GA 00840682109260SEE10410466GA SEE10410467GA 00840682109260SEE10410467GA SEE10410468GA 00840682109260SEE10410468GA SEE10410469GA 00840682109260SEE10410469GA SEE10410470GA 00840682109260SEE10410470GA SEE10410472GA 00840682109260SEE10410472GA SEE10410473GA 00840682109260SEE10410473GA SEE10410477GA 00840682109260SEE10410477GA SEE10410479GA 00840682109260SEE10410479GA SEE10410480GA 00840682109260SEE10410480GA SEE10410481GA 00840682109260SEE10410481GA SEE10410482GA 00840682109260SEE10410482GA SEE10410483GA 00840682109260SEE10410483GA SEE10410485GA 00840682109260SEE10410485GA SEE10410486GA 00840682109260SEE10410486GA SEE10410487GA 00840682109260SEE10410487GA SEE10410489GA 00840682109260SEE10410489GA SEE10410490GA 00840682109260SEE10410490GA SEE10410491GA 00840682109260SEE10410491GA SEE10410492GA 00840682109260SEE10410492GA SEE10410494GA 00840682109260SEE10410494GA SEE10410495GA 00840682109260SEE10410495GA SEE10410496GA 00840682109260SEE10410496GA SEE10410497GA 00840682109260SEE10410497GA SEE10410498GA 00840682109260SEE10410498GA SEE10410499GA 00840682109260SEE10410499GA SEE10410500GA 00840682109260SEE10410500GA SEE10410501GA 00840682109260SEE10410501GA SEE10410502GA 00840682109260SEE10410502GA SEE10410503GA 00840682109260SEE10410503GA SEE10410506GA 00840682109260SEE10410506GA SEE10410507GA 00840682109260SEE10410507GA SEE10410509GA 00840682109260SEE10410509GA SEE10410510GA 00840682109260SEE10410510GA SEE10410511GA 00840682109260SEE10410511GA SEE10410512GA 00840682109260SEE10410512GA SEE10410513GA 00840682109260SEE10410513GA SEE10410514GA 00840682109260SEE10410514GA SEE10420517GA 00840682109260SEE10420517GA SEE10420519GA 00840682109260SEE10420519GA SEE10420520GA 00840682109260SEE10420520GA SEE10420522GA 00840682109260SEE10420522GA SEE10420525GA 00840682109260SEE10420525GA SEE10420531GA 00840682109260SEE10420531GA SEE10420532GA 00840682109260SEE10420532GA SEE10420533GA 00840682109260SEE10420533GA SEE10420534GA 00840682109260SEE10420534GA SEE10420535GA 00840682109260SEE10420535GA SEE10420536GA 00840682109260SEE10420536GA SEE10420537GA 00840682109260SEE10420537GA SEE10420538GA 00840682109260SEE10420538GA SEE10420540GA 00840682109260SEE10420540GA SEE10420547GA 00840682109260SEE10420547GA SEE10420548GA 00840682109260SEE10420548GA SEE10420549GA 00840682109260SEE10420549GA SEE10420550GA 00840682109260SEE10420550GA SEE10420551GA 00840682109260SEE10420551GA SEE10420553GA 00840682109260SEE10420553GA SEE10420557GA 00840682109260SEE10420557GA SEE10420558GA 00840682109260SEE10420558GA SEE10420559GA 00840682109260SEE10420559GA SEE10420560GA 00840682109260SEE10420560GA SEE10420561GA 00840682109260SEE10420561GA SEE10420562GA 00840682109260SEE10420562GA SEE10420563GA 00840682109260SEE10420563GA SEE10420564GA 00840682109260SEE10420564GA SEE10420565GA 00840682109260SEE10420565GA SEE10420566GA 00840682109260SEE10420566GA SEE10420567GA 00840682109260SEE10420567GA SEE10420569GA 00840682109260SEE10420569GA SEE10420570GA 00840682109260SEE10420570GA SEE10420573GA 00840682109260SEE10420573GA SEE10420574GA 00840682109260SEE10420574GA SEE10420575GA 00840682109260SEE10420575GA SEE10420576GA 00840682109260SEE10420576GA SEE10420577GA 00840682109260SEE10420577GA SEE10420579GA 00840682109260SEE10420579GA SEE10420580GA 00840682109260SEE10420580GA SEE10420582GA 00840682109260SEE10420582GA SEE10420583GA 00840682109260SEE10420583GA SEE10420585GA 00840682109260SEE10420585GA SEE10420586GA 00840682109260SEE10420586GA SEE10420588GA 00840682109260SEE10420588GA SEE10420589GA 00840682109260SEE10420589GA SEE10420590GA 00840682109260SEE10420590GA SEE10420593GA 00840682109260SEE10420593GA SEE10420595GA 00840682109260SEE10420595GA SEE10420596GA 00840682109260SEE10420596GA SEE10420597GA 00840682109260SEE10420597GA SEE10420598GA 00840682109260SEE10420598GA SEE10420599GA 00840682109260SEE10420599GA SEE10420600GA 00840682109260SEE10420600GA SEE10420602GA 00840682109260SEE10420602GA SEE10430603GA 00840682109260SEE10430603GA SEE10430604GA 00840682109260SEE10430604GA SEE10430605GA 00840682109260SEE10430605GA SEE10430606GA 00840682109260SEE10430606GA SEE10430607GA 00840682109260SEE10430607GA SEE10430608GA 00840682109260SEE10430608GA SEE10430609GA 00840682109260SEE10430609GA SEE10430610GA 00840682109260SEE10430610GA SEE10430611GA 00840682109260SEE10430611GA SEE10430612GA 00840682109260SEE10430612GA SEE10430614GA 00840682109260SEE10430614GA SEE10430616GA 00840682109260SEE10430616GA SEE10430617GA 00840682109260SEE10430617GA SEE10430620GA 00840682109260SEE10430620GA SEE10430621GA 00840682109260SEE10430621GA SEE10430622GA 00840682109260SEE10430622GA SEE10430623GA 00840682109260SEE10430623GA SEE10430625GA 00840682109260SEE10430625GA SEE10430626GA 00840682109260SEE10430626GA SEE10430627GA 00840682109260SEE10430627GA SEE10430628GA 00840682109260SEE10430628GA SEE10430629GA 00840682109260SEE10430629GA SEE10430630GA 00840682109260SEE10430630GA SEE10430631GA 00840682109260SEE10430631GA SEE10430632GA 00840682109260SEE10430632GA SEE10430633GA 00840682109260SEE10430633GA SEE10430634GA 00840682109260SEE10430634GA SEE10430635GA 00840682109260SEE10430635GA SEE10430638GA 00840682109260SEE10430638GA SEE10430639GA 00840682109260SEE10430639GA SEE10430640GA 00840682109260SEE10430640GA SEE10430641GA 00840682109260SEE10430641GA SEE10430642GA 00840682109260SEE10430642GA SEE10430644GA 00840682109260SEE10430644GA SEE10430645GA 00840682109260SEE10430645GA SEE10430646GA 00840682109260SEE10430646GA SEE10430647GA 00840682109260SEE10430647GA SEE10430648GA 00840682109260SEE10430648GA SEE10430649GA 00840682109260SEE10430649GA SEE10430650GA 00840682109260SEE10430650GA SEE10430651GA 00840682109260SEE10430651GA SEE10430652GA 00840682109260SEE10430652GA SEE10430653GA 00840682109260SEE10430653GA SEE10450656GA 00840682109260SEE10450656GA SEE10450657GA 00840682109260SEE10450657GA SEE10450658GA 00840682109260SEE10450658GA SEE10450659GA 00840682109260SEE10450659GA SEE10450660GA 00840682109260SEE10450660GA SEE10450662GA 00840682109260SEE10450662GA SEE10450663GA 00840682109260SEE10450663GA SEE10450664GA 00840682109260SEE10450664GA SEE10450666GA 00840682109260SEE10450666GA SEE10450667GA 00840682109260SEE10450667GA SEE10450669GA 00840682109260SEE10450669GA SEE10450670GA 00840682109260SEE10450670GA SEE10450671GA 00840682109260SEE10450671GA SEE10450672GA 00840682109260SEE10450672GA SEE10450674GA 00840682109260SEE10450674GA SEE10450675GA 00840682109260SEE10450675GA SEE10450676GA 00840682109260SEE10450676GA SEE10450677GA 00840682109260SEE10450677GA SEE10450678GA 00840682109260SEE10450678GA SEE10450680GA 00840682109260SEE10450680GA SEE10450681GA 00840682109260SEE10450681GA SEE10450682GA 00840682109260SEE10450682GA SEE10450683GA 00840682109260SEE10450683GA SEE10450684GA 00840682109260SEE10450684GA SEE10450686GA 00840682109260SEE10450686GA SEE10450688GA 00840682109260SEE10450688GA SEE10450689GA 00840682109260SEE10450689GA SEE10460690GA 00840682109260SEE10460690GA SEE10460691GA 00840682109260SEE10460691GA SEE10460692GA 00840682109260SEE10460692GA SEE10460693GA 00840682109260SEE10460693GA SEE10460694GA 00840682109260SEE10460694GA SEE10460695GA 00840682109260SEE10460695GA SEE10460696GA 00840682109260SEE10460696GA SEE10460697GA 00840682109260SEE10460697GA SEE10460698GA 00840682109260SEE10460698GA SEE10460700GA 00840682109260SEE10460700GA SEE10460701GA 00840682109260SEE10460701GA SEE10460702GA 00840682109260SEE10460702GA SEE10460703GA 00840682109260SEE10460703GA SEE10460704GA 00840682109260SEE10460704GA SEE10460705GA 00840682109260SEE10460705GA SEE10460706GA 00840682109260SEE10460706GA SEE10460707GA 00840682109260SEE10460707GA SEE10460708GA 00840682109260SEE10460708GA SEE10460709GA 00840682109260SEE10460709GA SEE10460710GA 00840682109260SEE10460710GA SEE10460711GA 00840682109260SEE10460711GA SEE10460712GA 00840682109260SEE10460712GA SEE10460713GA 00840682109260SEE10460713GA SEE10460714GA 00840682109260SEE10460714GA SEE10460715GA 00840682109260SEE10460715GA SEE10460716GA 00840682109260SEE10460716GA SEE10460718GA 00840682109260SEE10460718GA SEE10460719GA 00840682109260SEE10460719GA SEE10460722GA 00840682109260SEE10460722GA SEE10460723GA 00840682109260SEE10460723GA SEE10460724GA 00840682109260SEE10460724GA SEE10460725GA 00840682109260SEE10460725GA SEE10460726GA 00840682109260SEE10460726GA SEE10460727GA 00840682109260SEE10460727GA SEE10460728GA 00840682109260SEE10460728GA SEE10460729GA 00840682109260SEE10460729GA SEE10460730GA 00840682109260SEE10460730GA SEE10460731GA 00840682109260SEE10460731GA SEE10460733GA 00840682109260SEE10460733GA SEE10460734GA 00840682109260SEE10460734GA SEE10460735GA 00840682109260SEE10460735GA SEE10460736GA 00840682109260SEE10460736GA SEE10460737GA 00840682109260SEE10460737GA SEE10460738GA 00840682109260SEE10460738GA SEE10460739GA 00840682109260SEE10460739GA SEE10460740GA 00840682109260SEE10460740GA SEE10460741GA 00840682109260SEE10460741GA SEE10460742GA 00840682109260SEE10460742GA SEE10460744GA 00840682109260SEE10460744GA SEE10460746GA 00840682109260SEE10460746GA SEE10460747GA 00840682109260SEE10460747GA SEE10460748GA 00840682109260SEE10460748GA SEE10460751GA 00840682109260SEE10460751GA SEE10460752GA 00840682109260SEE10460752GA SEE10460753GA 00840682109260SEE10460753GA SEE10460754GA 00840682109260SEE10460754GA SEE10460755GA 00840682109260SEE10460755GA SEE10460756GA 00840682109260SEE10460756GA SEE10460758GA 00840682109260SEE10460758GA SEE10460759GA 00840682109260SEE10460759GA SEE10460760GA 00840682109260SEE10460760GA SEE10460761GA 00840682109260SEE10460761GA SEE10460762GA 00840682109260SEE10460762GA SEE10460763GA 00840682109260SEE10460763GA SEE10460764GA 00840682109260SEE10460764GA SEE10460765GA 00840682109260SEE10460765GA SEE10460766GA 00840682109260SEE10460766GA SEE10460767GA 00840682109260SEE10460767GA SEE10460769GA 00840682109260SEE10460769GA SEE10460771GA 00840682109260SEE10460771GA SEE10460772GA 00840682109260SEE10460772GA SEE10460773GA 00840682109260SEE10460773GA SEE10460774GA 00840682109260SEE10460774GA SEE10460775GA 00840682109260SEE10460775GA SEE10470776GA 00840682109260SEE10470776GA SEE10470777GA 00840682109260SEE10470777GA SEE10470778GA 00840682109260SEE10470778GA SEE10470779GA 00840682109260SEE10470779GA SEE10470780GA 00840682109260SEE10470780GA SEE10470781GA 00840682109260SEE10470781GA SEE10470784GA 00840682109260SEE10470784GA SEE10470786GA 00840682109260SEE10470786GA SEE10470787GA 00840682109260SEE10470787GA SEE10470789GA 00840682109260SEE10470789GA SEE10470793GA 00840682109260SEE10470793GA SEE10470795GA 00840682109260SEE10470795GA SEE10470796GA 00840682109260SEE10470796GA SEE10470797GA 00840682109260SEE10470797GA SEE10470798GA 00840682109260SEE10470798GA SEE10470799GA 00840682109260SEE10470799GA SEE10470803GA 00840682109260SEE10470803GA SEE10470804GA 00840682109260SEE10470804GA SEE10470805GA 00840682109260SEE10470805GA SEE10470806GA 00840682109260SEE10470806GA SEE10510809GA 00840682109260SEE10510809GA SEE10510810GA 00840682109260SEE10510810GA SEE10510811GA 00840682109260SEE10510811GA SEE10510812GA 00840682109260SEE10510812GA SEE10510813GA 00840682109260SEE10510813GA SEE10510814GA 00840682109260SEE10510814GA SEE10530815GA 00840682109260SEE10530815GA SEE10530816GA 00840682109260SEE10530816GA SEE10530817GA 00840682109260SEE10530817GA SEE10530818GA 00840682109260SEE10530818GA SEE10530819GA 00840682109260SEE10530819GA SEE10530820GA 00840682109260SEE10530820GA SEE10530821GA 00840682109260SEE10530821GA SEE10530822GA 00840682109260SEE10530822GA SEE10530823GA 00840682109260SEE10530823GA SEE10530824GA 00840682109260SEE10530824GA SEE10530825GA 00840682109260SEE10530825GA SEE10530826GA 00840682109260SEE10530826GA SEE10530827GA 00840682109260SEE10530827GA SEE10530829GA 00840682109260SEE10530829GA SEE10530830GA 00840682109260SEE10530830GA SEE10530831GA 00840682109260SEE10530831GA SEE10530832GA 00840682109260SEE10530832GA SEE10530833GA 00840682109260SEE10530833GA SEE10530834GA 00840682109260SEE10530834GA SEE10530835GA 00840682109260SEE10530835GA SEE10530836GA 00840682109260SEE10530836GA SEE10530837GA 00840682109260SEE10530837GA SEE10530838GA 00840682109260SEE10530838GA SEE10530839GA 00840682109260SEE10530839GA SEE10530841GA 00840682109260SEE10530841GA SEE10530842GA 00840682109260SEE10530842GA SEE10530843GA 00840682109260SEE10530843GA SEE10530844GA 00840682109260SEE10530844GA SEE10530845GA 00840682109260SEE10530845GA SEE10530846GA 00840682109260SEE10530846GA SEE10530847GA 00840682109260SEE10530847GA SEE10530848GA 00840682109260SEE10530848GA SEE10530849GA 00840682109260SEE10530849GA SEE10530850GA 00840682109260SEE10530850GA SEE10530851GA 00840682109260SEE10530851GA SEE10530852GA 00840682109260SEE10530852GA SEE10530854GA 00840682109260SEE10530854GA SEE10530856GA 00840682109260SEE10530856GA SEE10530857GA 00840682109260SEE10530857GA SEE10530858GA 00840682109260SEE10530858GA SEE10530859GA 00840682109260SEE10530859GA SEE10530860GA 00840682109260SEE10530860GA SEE10530861GA 00840682109260SEE10530861GA SEE10530864GA 00840682109260SEE10530864GA SEE10530867GA 00840682109260SEE10530867GA SEE10530868GA 00840682109260SEE10530868GA SEE10530869GA 00840682109260SEE10530869GA SEE10530870GA 00840682109260SEE10530870GA SEE10530871GA 00840682109260SEE10530871GA SEE10530872GA 00840682109260SEE10530872GA SEE10530873GA 00840682109260SEE10530873GA SEE11030879GA 00840682109260SEE11030879GA SEE11030880GA 00840682109260SEE11030880GA SEE11030881GA 00840682109260SEE11030881GA SEE11030882GA 00840682109260SEE11030882GA SEE11030883GA 00840682109260SEE11030883GA SEE11030884GA 00840682109260SEE11030884GA SEE11030885GA 00840682109260SEE11030885GA SEE11030886GA 00840682109260SEE11030886GA SEE11030887GA 00840682109260SEE11030887GA SEE11030888GA 00840682109260SEE11030888GA SEE11030889GA 00840682109260SEE11030889GA SEE11030890GA 00840682109260SEE11030890GA SEE11030891GA 00840682109260SEE11030891GA SEE11030892GA 00840682109260SEE11030892GA SEE11030893GA 00840682109260SEE11030893GA SEE11030894GA 00840682109260SEE11030894GA SEE11030895GA 00840682109260SEE11030895GA SEE11040898GA 00840682109260SEE11040898GA SEE11040899GA 00840682109260SEE11040899GA SEE11040900GA 00840682109260SEE11040900GA SEE11040901GA 00840682109260SEE11040901GA SEE11040902GA 00840682109260SEE11040902GA SEE11040904GA 00840682109260SEE11040904GA SEE11040905GA 00840682109260SEE11040905GA SEE11040906GA 00840682109260SEE11040906GA SEE11040907GA 00840682109260SEE11040907GA SEE11040908GA 00840682109260SEE11040908GA SEE11040909GA 00840682109260SEE11040909GA SEE11040910GA 00840682109260SEE11040910GA SEE11040912GA 00840682109260SEE11040912GA SEE11040913GA 00840682109260SEE11040913GA SEE11040914GA 00840682109260SEE11040914GA SEE11040915GA 00840682109260SEE11040915GA SEE11040916GA 00840682109260SEE11040916GA SEE11040917GA 00840682109260SEE11040917GA SEE11040918GA 00840682109260SEE11040918GA SEE11050920GA 00840682109260SEE11050920GA SEE11050921GA 00840682109260SEE11050921GA SEE11050922GA 00840682109260SEE11050922GA SEE11050923GA 00840682109260SEE11050923GA SEE11050924GA 00840682109260SEE11050924GA SEE11050925GA 00840682109260SEE11050925GA SEE11050926GA 00840682109260SEE11050926GA SEE11050928GA 00840682109260SEE11050928GA SEE11050929GA 00840682109260SEE11050929GA SEE11050930GA 00840682109260SEE11050930GA SEE11050931GA 00840682109260SEE11050931GA SEE11050932GA 00840682109260SEE11050932GA SEE11050933GA 00840682109260SEE11050933GA SEE11050934GA 00840682109260SEE11050934GA SEE11050938GA 00840682109260SEE11050938GA SEE11050939GA 00840682109260SEE11050939GA SEE11050940GA 00840682109260SEE11050940GA SEE11050941GA 00840682109260SEE11050941GA SEE11050943GA 00840682109260SEE11050943GA SEE11050944GA 00840682109260SEE11050944GA SEE11050945GA 00840682109260SEE11050945GA SEE11050972GA 00840682109260SEE11050972GA SEE11050973GA 00840682109260SEE11050973GA SEE11050974GA 00840682109260SEE11050974GA SEE11050975GA 00840682109260SEE11050975GA SEE11050976GA 00840682109260SEE11050976GA SEE11060947GA 00840682109260SEE11060947GA SEE11060948GA 00840682109260SEE11060948GA SEE11060949GA 00840682109260SEE11060949GA SEE11060951GA 00840682109260SEE11060951GA SEE11060952GA 00840682109260SEE11060952GA SEE11060953GA 00840682109260SEE11060953GA SEE11060954GA 00840682109260SEE11060954GA SEE11060955GA 00840682109260SEE11060955GA SEE11060956GA 00840682109260SEE11060956GA SEE11060957GA 00840682109260SEE11060957GA SEE11060958GA 00840682109260SEE11060958GA SEE11060959GA 00840682109260SEE11060959GA SEE11060962GA 00840682109260SEE11060962GA SEE11060963GA 00840682109260SEE11060963GA SEE11060964GA 00840682109260SEE11060964GA SEE11060977GA 00840682109260SEE11060977GA SEE11060978GA 00840682109260SEE11060978GA SEE11060979GA 00840682109260SEE11060979GA SEE11060980GA 00840682109260SEE11060980GA SEE11060981GA 00840682109260SEE11060981GA SEE11060982GA 00840682109260SEE11060982GA SEE11060983GA 00840682109260SEE11060983GA SEE11060984GA 00840682109260SEE11060984GA SEE11060985GA 00840682109260SEE11060985GA SEE11060986GA 00840682109260SEE11060986GA SEE11060987GA 00840682109260SEE11060987GA SEE11060988GA 00840682109260SEE11060988GA SEE11060989GA 00840682109260SEE11060989GA SEE11060990GA 00840682109260SEE11060990GA SEE11060991GA 00840682109260SEE11060991GA SEE11060992GA 00840682109260SEE11060992GA SEE11060993GA 00840682109260SEE11060993GA SEE11060994GA 00840682109260SEE11060994GA SEE11060995GA 00840682109260SEE11060995GA SEE11060996GA 00840682109260SEE11060996GA SEE11060997GA 00840682109260SEE11060997GA SEE11060998GA 00840682109260SEE11060998GA SEE11060999GA 00840682109260SEE11060999GA SEE11061000GA 00840682109260SEE11061000GA SEE11061001GA 00840682109260SEE11061001GA SEE11061002GA 00840682109260SEE11061002GA SEE11061003GA 00840682109260SEE11061003GA SEE11061004GA 00840682109260SEE11061004GA SEE11071005GA 00840682109260SEE11071005GA SEE11071006GA 00840682109260SEE11071006GA SEE11071007GA 00840682109260SEE11071007GA SEE11071008GA 00840682109260SEE11071008GA SEE11071010GA 00840682109260SEE11071010GA SEE11071011GA 00840682109260SEE11071011GA SEE11071012GA 00840682109260SEE11071012GA SEE11071014GA 00840682109260SEE11071014GA SEE11071015GA 00840682109260SEE11071015GA SEE11071016GA 00840682109260SEE11071016GA SEE11071017GA 00840682109260SEE11071017GA SEE11071018GA 00840682109260SEE11071018GA SEE11071019GA 00840682109260SEE11071019GA SEE11071020GA 00840682109260SEE11071020GA SEE11071021GA 00840682109260SEE11071021GA SEE11071022GA 00840682109260SEE11071022GA SEE11071023GA 00840682109260SEE11071023GA SEE11071024GA 00840682109260SEE11071024GA SEE11071025GA 00840682109260SEE11071025GA SEE11071026GA 00840682109260SEE11071026GA SEE11071027GA 00840682109260SEE11071027GA SEE11071028GA 00840682109260SEE11071028GA SEE11071029GA 00840682109260SEE11071029GA SEE11071030GA 00840682109260SEE11071030GA SEE11071031GA 00840682109260SEE11071031GA SEE11071032GA 00840682109260SEE11071032GA SEE11071033GA 00840682109260SEE11071033GA SEE11071034GA 00840682109260SEE11071034GA SEE11071035GA 00840682109260SEE11071035GA SEE11071036GA 00840682109260SEE11071036GA SEE11071037GA 00840682109260SEE11071037GA SEE11071038GA 00840682109260SEE11071038GA SEE11071039GA 00840682109260SEE11071039GA SEE11071040GA 00840682109260SEE11071040GA SEE11071041GA 00840682109260SEE11071041GA SEE11071042GA 00840682109260SEE11071042GA SEE11071043GA 00840682109260SEE11071043GA SEE11071044GA 00840682109260SEE11071044GA SEE11071045GA 00840682109260SEE11071045GA SEE11071046GA 00840682109260SEE11071046GA SEE11071047GA 00840682109260SEE11071047GA SEE11071049GA 00840682109260SEE11071049GA SEE11071050GA 00840682109260SEE11071050GA SEE11071051GA 00840682109260SEE11071051GA SEE11071052GA 00840682109260SEE11071052GA SEE11071054GA 00840682109260SEE11071054GA SEE11071055GA 00840682109260SEE11071055GA SEE11071056GA 00840682109260SEE11071056GA SEE11071057GA 00840682109260SEE11071057GA SEE11071058GA 00840682109260SEE11071058GA SEE11071059GA 00840682109260SEE11071059GA SEE11071060GA 00840682109260SEE11071060GA SEE11071063GA 00840682109260SEE11071063GA SEE11071064GA 00840682109260SEE11071064GA SEE11071065GA 00840682109260SEE11071065GA SEE11071072GA 00840682109260SEE11071072GA SEE11071073GA 00840682109260SEE11071073GA SEE11071077GA 00840682109260SEE11071077GA SEE11071079GA 00840682109260SEE11071079GA SEE11071080GA 00840682109260SEE11071080GA SEE11071081GA 00840682109260SEE11071081GA SEE11071082GA 00840682109260SEE11071082GA SEE11071084GA 00840682109260SEE11071084GA SEE11071085GA 00840682109260SEE11071085GA SEE11071086GA 00840682109260SEE11071086GA SEE11071087GA 00840682109260SEE11071087GA SEE11071088GA 00840682109260SEE11071088GA SEE11071090GA 00840682109260SEE11071090GA SEE11071092GA 00840682109260SEE11071092GA SEE11071094GA 00840682109260SEE11071094GA SEE11071095GA 00840682109260SEE11071095GA SEE11071096GA 00840682109260SEE11071096GA SEE11071097GA 00840682109260SEE11071097GA SEE11071099GA 00840682109260SEE11071099GA SEE11071132GA 00840682109260SEE11071132GA SEE11071133GA 00840682109260SEE11071133GA SEE11071135GA 00840682109260SEE11071135GA SEE11071136GA 00840682109260SEE11071136GA SEE11071138GA 00840682109260SEE11071138GA SEE11071142GA 00840682109260SEE11071142GA SEE11071144GA 00840682109260SEE11071144GA SEE11071145GA 00840682109260SEE11071145GA SEE11071146GA 00840682109260SEE11071146GA SEE11071147GA 00840682109260SEE11071147GA SEE11071148GA 00840682109260SEE11071148GA SEE11071149GA 00840682109260SEE11071149GA SEE11071151GA 00840682109260SEE11071151GA SEE11071152GA 00840682109260SEE11071152GA SEE11071154GA 00840682109260SEE11071154GA SEE11071155GA 00840682109260SEE11071155GA SEE11081156GA 00840682109260SEE11081156GA SEE11081157GA 00840682109260SEE11081157GA SEE11081158GA 00840682109260SEE11081158GA SEE11081159GA 00840682109260SEE11081159GA SEE11081160GA 00840682109260SEE11081160GA SEE11081164GA 00840682109260SEE11081164GA SEE11081165GA 00840682109260SEE11081165GA SEE11081166GA 00840682109260SEE11081166GA SEE11081167GA 00840682109260SEE11081167GA SEE11081168GA 00840682109260SEE11081168GA SEE11081169GA 00840682109260SEE11081169GA SEE11081170GA 00840682109260SEE11081170GA SEE11081171GA 00840682109260SEE11081171GA SEE11081172GA 00840682109260SEE11081172GA SEE11081173GA 00840682109260SEE11081173GA SEE11081174GA 00840682109260SEE11081174GA SEE11081175GA 00840682109260SEE11081175GA SEE11081176GA 00840682109260SEE11081176GA SEE11081179GA 00840682109260SEE11081179GA SEE11081181GA 00840682109260SEE11081181GA SEE11081182GA 00840682109260SEE11081182GA SEE11081183GA 00840682109260SEE11081183GA SEE11081184GA 00840682109260SEE11081184GA SEE11081186GA 00840682109260SEE11081186GA SEE11081187GA 00840682109260SEE11081187GA SEE11081188GA 00840682109260SEE11081188GA SEE11081190GA 00840682109260SEE11081190GA SEE11081193GA 00840682109260SEE11081193GA SEE11081194GA 00840682109260SEE11081194GA SEE11081195GA 00840682109260SEE11081195GA SEE11081196GA 00840682109260SEE11081196GA SEE11081198GA 00840682109260SEE11081198GA SEE11081199GA 00840682109260SEE11081199GA SEE11081200GA 00840682109260SEE11081200GA SEE11081201GA 00840682109260SEE11081201GA SEE11081202GA 00840682109260SEE11081202GA SEE11081203GA 00840682109260SEE11081203GA SEE11081205GA 00840682109260SEE11081205GA SEE11081206GA 00840682109260SEE11081206GA SEE11081208GA 00840682109260SEE11081208GA SEE11081209GA 00840682109260SEE11081209GA SEE11081210GA 00840682109260SEE11081210GA SEE11081211GA 00840682109260SEE11081211GA SEE11091212GA 00840682109260SEE11091212GA SEE11101213GA 00840682109260SEE11101213GA SEE11101214GA 00840682109260SEE11101214GA SEE11101215GA 00840682109260SEE11101215GA SEE11101216GA 00840682109260SEE11101216GA SEE11101217GA 00840682109260SEE11101217GA SEE11101218GA 00840682109260SEE11101218GA SEE11101220GA 00840682109260SEE11101220GA SEE11101221GA 00840682109260SEE11101221GA SEE11101222GA 00840682109260SEE11101222GA SEE11101223GA 00840682109260SEE11101223GA SEE11101224GA 00840682109260SEE11101224GA SEE11101225GA 00840682109260SEE11101225GA SEE11111227GA 00840682109260SEE11111227GA SEE11111228GA 00840682109260SEE11111228GA SEE11111229GA 00840682109260SEE11111229GA SEE11111230GA 00840682109260SEE11111230GA SEE11111231GA 00840682109260SEE11111231GA SEE11111233GA 00840682109260SEE11111233GA SEE11111235GA 00840682109260SEE11111235GA SEE11111236GA 00840682109260SEE11111236GA SEE11111237GA 00840682109260SEE11111237GA SEE11111238GA 00840682109260SEE11111238GA SEE11111239GA 00840682109260SEE11111239GA SEE11111240GA 00840682109260SEE11111240GA SEE11111241GA 00840682109260SEE11111241GA SEE11111242GA 00840682109260SEE11111242GA SEE11111243GA 00840682109260SEE11111243GA SEE11111244GA 00840682109260SEE11111244GA SEE11111246GA 00840682109260SEE11111246GA SEE11111247GA 00840682109260SEE11111247GA SEE11111248GA 00840682109260SEE11111248GA SEE11141250GA 00840682109260SEE11141250GA SEE11141253GA 00840682109260SEE11141253GA SEE11141254GA 00840682109260SEE11141254GA SEE11141255GA 00840682109260SEE11141255GA SEE11141256GA 00840682109260SEE11141256GA SEE11141257GA 00840682109260SEE11141257GA SEE11141258GA 00840682109260SEE11141258GA SEE11141259GA 00840682109260SEE11141259GA SEE11141260GA 00840682109260SEE11141260GA SEE11141261GA 00840682109260SEE11141261GA SEE11141262GA 00840682109260SEE11141262GA SEE11141263GA 00840682109260SEE11141263GA SEE11141265GA 00840682109260SEE11141265GA SEE11141266GA 00840682109260SEE11141266GA SEE11141267GA 00840682109260SEE11141267GA SEE11141272GA 00840682109260SEE11141272GA SEE11141273GA 00840682109260SEE11141273GA SEE11141274GA 00840682109260SEE11141274GA SEE11141275GA 00840682109260SEE11141275GA SEE11141276GA 00840682109260SEE11141276GA SEE11141277GA 00840682109260SEE11141277GA SEE11141278GA 00840682109260SEE11141278GA SEE11141279GA 00840682109260SEE11141279GA SEE11141280GA 00840682109260SEE11141280GA SEE11141281GA 00840682109260SEE11141281GA SEE11141282GA 00840682109260SEE11141282GA SEE11141284GA 00840682109260SEE11141284GA SEE11141285GA 00840682109260SEE11141285GA SEE11141286GA 00840682109260SEE11141286GA SEE11141287GA 00840682109260SEE11141287GA SEE11141288GA 00840682109260SEE11141288GA SEE11141289GA 00840682109260SEE11141289GA SEE11141291GA 00840682109260SEE11141291GA SEE11141292GA 00840682109260SEE11141292GA SEE11141293GA 00840682109260SEE11141293GA SEE11141295GA 00840682109260SEE11141295GA SEE11151297GA 00840682109260SEE11151297GA SEE11151298GA 00840682109260SEE11151298GA SEE11151301GA 00840682109260SEE11151301GA SEE11151302GA 00840682109260SEE11151302GA SEE11151304GA 00840682109260SEE11151304GA SEE11151305GA 00840682109260SEE11151305GA SEE11151306GA 00840682109260SEE11151306GA SEE11151307GA 00840682109260SEE11151307GA SEE11151308GA 00840682109260SEE11151308GA SEE11151309GA 00840682109260SEE11151309GA SEE11151311GA 00840682109260SEE11151311GA SEE11151312GA 00840682109260SEE11151312GA SEE11151313GA 00840682109260SEE11151313GA SEE11151314GA 00840682109260SEE11151314GA SEE11151315GA 00840682109260SEE11151315GA SEE11151316GA 00840682109260SEE11151316GA SEE11161317GA 00840682109260SEE11161317GA SEE11161318GA 00840682109260SEE11161318GA SEE11161319GA 00840682109260SEE11161319GA SEE11161320GA 00840682109260SEE11161320GA SEE11161321GA 00840682109260SEE11161321GA SEE11161322GA 00840682109260SEE11161322GA SEE11161323GA 00840682109260SEE11161323GA SEE11161324GA 00840682109260SEE11161324GA SEE11161325GA 00840682109260SEE11161325GA SEE11161326GA 00840682109260SEE11161326GA SEE11161327GA 00840682109260SEE11161327GA SEE11161329GA 00840682109260SEE11161329GA SEE11161331GA 00840682109260SEE11161331GA SEE11161333GA 00840682109260SEE11161333GA SEE11161337GA 00840682109260SEE11161337GA SEE11161338GA 00840682109260SEE11161338GA SEE11161339GA 00840682109260SEE11161339GA SEE11161340GA 00840682109260SEE11161340GA SEE11161341GA 00840682109260SEE11161341GA SEE11161342GA 00840682109260SEE11161342GA SEE11161344GA 00840682109260SEE11161344GA SEE11161346GA 00840682109260SEE11161346GA SEE11161347GA 00840682109260SEE11161347GA SEE11171353GA 00840682109260SEE11171353GA SEE11171355GA 00840682109260SEE11171355GA SEE11171356GA 00840682109260SEE11171356GA SEE11171357GA 00840682109260SEE11171357GA SEE11171358GA 00840682109260SEE11171358GA SEE11171359GA 00840682109260SEE11171359GA SEE11171360GA 00840682109260SEE11171360GA SEE11171361GA 00840682109260SEE11171361GA SEE11171362GA 00840682109260SEE11171362GA SEE11171363GA 00840682109260SEE11171363GA SEE11171365GA 00840682109260SEE11171365GA SEE11171366GA 00840682109260SEE11171366GA SEE11171367GA 00840682109260SEE11171367GA SEE11171368GA 00840682109260SEE11171368GA SEE11171369GA 00840682109260SEE11171369GA SEE11171370GA 00840682109260SEE11171370GA SEE11171371GA 00840682109260SEE11171371GA SEE11171372GA 00840682109260SEE11171372GA SEE11171374GA 00840682109260SEE11171374GA SEE11171375GA 00840682109260SEE11171375GA SEE11171376GA 00840682109260SEE11171376GA SEE11171377GA 00840682109260SEE11171377GA SEE11171378GA 00840682109260SEE11171378GA SEE11171381GA 00840682109260SEE11171381GA SEE11171382GA 00840682109260SEE11171382GA SEE11171383GA 00840682109260SEE11171383GA SEE11171385GA 00840682109260SEE11171385GA SEE11171386GA 00840682109260SEE11171386GA SEE11171387GA 00840682109260SEE11171387GA SEE11181388GA 00840682109260SEE11181388GA SEE11181389GA 00840682109260SEE11181389GA SEE11181390GA 00840682109260SEE11181390GA SEE11181391GA 00840682109260SEE11181391GA SEE11181393GA 00840682109260SEE11181393GA SEE11181394GA 00840682109260SEE11181394GA SEE11181395GA 00840682109260SEE11181395GA SEE11181396GA 00840682109260SEE11181396GA SEE11181397GA 00840682109260SEE11181397GA SEE11181398GA 00840682109260SEE11181398GA SEE11181399GA 00840682109260SEE11181399GA SEE11181400GA 00840682109260SEE11181400GA SEE11181401GA 00840682109260SEE11181401GA SEE11181403GA 00840682109260SEE11181403GA SEE11181404GA 00840682109260SEE11181404GA SEE11181405GA 00840682109260SEE11181405GA SEE11181406GA 00840682109260SEE11181406GA SEE11181407GA 00840682109260SEE11181407GA SEE11181408GA 00840682109260SEE11181408GA SEE11181409GA 00840682109260SEE11181409GA SEE11181410GA 00840682109260SEE11181410GA SEE11181411GA 00840682109260SEE11181411GA SEE11181413GA 00840682109260SEE11181413GA SEE11181414GA 00840682109260SEE11181414GA SEE11181415GA 00840682109260SEE11181415GA SEE11181416GA 00840682109260SEE11181416GA SEE11181417GA 00840682109260SEE11181417GA SEE11181418GA 00840682109260SEE11181418GA SEE11181419GA 00840682109260SEE11181419GA SEE11181421GA 00840682109260SEE11181421GA SEE11191424GA 00840682109260SEE11191424GA SEE11191425GA 00840682109260SEE11191425GA SEE11191426GA 00840682109260SEE11191426GA SEE11191427GA 00840682109260SEE11191427GA SEE11191429GA 00840682109260SEE11191429GA SEE11191430GA 00840682109260SEE11191430GA SEE11191431GA 00840682109260SEE11191431GA SEE11191432GA 00840682109260SEE11191432GA SEE11191433GA 00840682109260SEE11191433GA SEE11191434GA 00840682109260SEE11191434GA SEE11191435GA 00840682109260SEE11191435GA SEE11191436GA 00840682109260SEE11191436GA SEE11191437GA 00840682109260SEE11191437GA SEE11191438GA 00840682109260SEE11191438GA SEE11191440GA 00840682109260SEE11191440GA SEE11191441GA 00840682109260SEE11191441GA SEE11201442GA 00840682109260SEE11201442GA SEE11201443GA 00840682109260SEE11201443GA SEE11201444GA 00840682109260SEE11201444GA SEE11201445GA 00840682109260SEE11201445GA SEE11201446GA 00840682109260SEE11201446GA SEE11201447GA 00840682109260SEE11201447GA SEE11201448GA 00840682109260SEE11201448GA SEE11201449GA 00840682109260SEE11201449GA SEE11201451GA 00840682109260SEE11201451GA SEE11201452GA 00840682109260SEE11201452GA SEE11201453GA 00840682109260SEE11201453GA SEE11201454GA 00840682109260SEE11201454GA SEE11201455GA 00840682109260SEE11201455GA SEE11201456GA 00840682109260SEE11201456GA SEE11201457GA 00840682109260SEE11201457GA SEE11201459GA 00840682109260SEE11201459GA SEE11201460GA 00840682109260SEE11201460GA SEE11201462GA 00840682109260SEE11201462GA SEE11201463GA 00840682109260SEE11201463GA SEE11201464GA 00840682109260SEE11201464GA SEE11201465GA 00840682109260SEE11201465GA SEE11201466GA 00840682109260SEE11201466GA SEE11201467GA 00840682109260SEE11201467GA SEE11201468GA 00840682109260SEE11201468GA SEE11201469GA 00840682109260SEE11201469GA SEE11201470GA 00840682109260SEE11201470GA SEE11201471GA 00840682109260SEE11201471GA SEE11201472GA 00840682109260SEE11201472GA SEE11211473GA 00840682109260SEE11211473GA SEE11211474GA 00840682109260SEE11211474GA SEE11211475GA 00840682109260SEE11211475GA SEE11211477GA 00840682109260SEE11211477GA SEE11211478GA 00840682109260SEE11211478GA SEE11211479GA 00840682109260SEE11211479GA SEE11211480GA 00840682109260SEE11211480GA SEE11211481GA 00840682109260SEE11211481GA SEE11211483GA 00840682109260SEE11211483GA SEE11211484GA 00840682109260SEE11211484GA SEE11211485GA 00840682109260SEE11211485GA SEE11211486GA 00840682109260SEE11211486GA SEE11211487GA 00840682109260SEE11211487GA SEE11211488GA 00840682109260SEE11211488GA SEE11211489GA 00840682109260SEE11211489GA SEE11211490GA 00840682109260SEE11211490GA SEE11211491GA 00840682109260SEE11211491GA SEE11211492GA 00840682109260SEE11211492GA SEE11211493GA 00840682109260SEE11211493GA SEE11211494GA 00840682109260SEE11211494GA SEE11211495GA 00840682109260SEE11211495GA SEE11211496GA 00840682109260SEE11211496GA SEE11211497GA 00840682109260SEE11211497GA SEE11211498GA 00840682109260SEE11211498GA SEE11211499GA 00840682109260SEE11211499GA SEE11211500GA 00840682109260SEE11211500GA SEE11211501GA 00840682109260SEE11211501GA SEE11211502GA 00840682109260SEE11211502GA SEE11211503GA 00840682109260SEE11211503GA SEE11211504GA 00840682109260SEE11211504GA SEE11211505GA 00840682109260SEE11211505GA SEE11211506GA 00840682109260SEE11211506GA SEE11211508GA 00840682109260SEE11211508GA SEE11211509GA 00840682109260SEE11211509GA SEE11211510GA 00840682109260SEE11211510GA SEE11211511GA 00840682109260SEE11211511GA SEE11211512GA 00840682109260SEE11211512GA SEE11211513GA 00840682109260SEE11211513GA SEE11211514GA 00840682109260SEE11211514GA SEE11211515GA 00840682109260SEE11211515GA SEE11211517GA 00840682109260SEE11211517GA SEE11211518GA 00840682109260SEE11211518GA SEE11211520GA 00840682109260SEE11211520GA SEE11211521GA 00840682109260SEE11211521GA SEE11211522GA 00840682109260SEE11211522GA SEE11211523GA 00840682109260SEE11211523GA SEE11231525GA 00840682109260SEE11231525GA SEE11231526GA 00840682109260SEE11231526GA SEE11231527GA 00840682109260SEE11231527GA SEE11231529GA 00840682109260SEE11231529GA SEE11231530GA 00840682109260SEE11231530GA SEE11231531GA 00840682109260SEE11231531GA SEE11231532GA 00840682109260SEE11231532GA SEE11231533GA 00840682109260SEE11231533GA SEE11231534GA 00840682109260SEE11231534GA SEE11231535GA 00840682109260SEE11231535GA SEE11231536GA 00840682109260SEE11231536GA SEE11231537GA 00840682109260SEE11231537GA SEE11231538GA 00840682109260SEE11231538GA SEE11231539GA 00840682109260SEE11231539GA SEE11231540GA 00840682109260SEE11231540GA SEE11231541GA 00840682109260SEE11231541GA SEE11231542GA 00840682109260SEE11231542GA SEE11231543GA 00840682109260SEE11231543GA SEE11231544GA 00840682109260SEE11231544GA SEE11231545GA 00840682109260SEE11231545GA SEE11231546GA 00840682109260SEE11231546GA SEE11241547GA 00840682109260SEE11241547GA SEE11241548GA 00840682109260SEE11241548GA SEE11241550GA 00840682109260SEE11241550GA SEE11241551GA 00840682109260SEE11241551GA SEE11301560GA 00840682109260SEE11301560GA SEE11321562GA 00840682109260SEE11321562GA SEE11321563GA 00840682109260SEE11321563GA SEE11321564GA 00840682109260SEE11321564GA SEE11321567GA 00840682109260SEE11321567GA SEE11321568GA 00840682109260SEE11321568GA SEE11321569GA 00840682109260SEE11321569GA SEE11321570GA 00840682109260SEE11321570GA SEE11321571GA 00840682109260SEE11321571GA SEE11321572GA 00840682109260SEE11321572GA SEE11321573GA 00840682109260SEE11321573GA SEE11321574GA 00840682109260SEE11321574GA SEE11321575GA 00840682109260SEE11321575GA SEE11321576GA 00840682109260SEE11321576GA SEE11321578GA 00840682109260SEE11321578GA SEE11321579GA 00840682109260SEE11321579GA SEE11321580GA 00840682109260SEE11321580GA SEE11321581GA 00840682109260SEE11321581GA SEE11321582GA 00840682109260SEE11321582GA SEE11321583GA 00840682109260SEE11321583GA SEE11321584GA 00840682109260SEE11321584GA SEE11321585GA 00840682109260SEE11321585GA SEE11321586GA 00840682109260SEE11321586GA SEE11321587GA 00840682109260SEE11321587GA SEE11331590GA 00840682109260SEE11331590GA SEE11331591GA 00840682109260SEE11331591GA SEE11331593GA 00840682109260SEE11331593GA SEE11331595GA 00840682109260SEE11331595GA SEE11331596GA 00840682109260SEE11331596GA SEE11331597GA 00840682109260SEE11331597GA SEE11331598GA 00840682109260SEE11331598GA SEE11331599GA 00840682109260SEE11331599GA SEE11331601GA 00840682109260SEE11331601GA SEE11341602GA 00840682109260SEE11341602GA SEE11341604GA 00840682109260SEE11341604GA SEE11341605GA 00840682109260SEE11341605GA SEE11341606GA 00840682109260SEE11341606GA SEE11341607GA 00840682109260SEE11341607GA SEE11341608GA 00840682109260SEE11341608GA SEE11341609GA 00840682109260SEE11341609GA SEE11341610GA 00840682109260SEE11341610GA SEE11341611GA 00840682109260SEE11341611GA SEE11341612GA 00840682109260SEE11341612GA SEE11341613GA 00840682109260SEE11341613GA SEE11341614GA 00840682109260SEE11341614GA SEE11341615GA 00840682109260SEE11341615GA SEE11341616GA 00840682109260SEE11341616GA SEE11341617GA 00840682109260SEE11341617GA SEE11341618GA 00840682109260SEE11341618GA SEE11341620GA 00840682109260SEE11341620GA SEE11341621GA 00840682109260SEE11341621GA SEE11341622GA 00840682109260SEE11341622GA SEE11341623GA 00840682109260SEE11341623GA SEE11341625GA 00840682109260SEE11341625GA SEE11341627GA 00840682109260SEE11341627GA SEE11341629GA 00840682109260SEE11341629GA SEE11341630GA 00840682109260SEE11341630GA SEE11341631GA 00840682109260SEE11341631GA SEE11341633GA 00840682109260SEE11341633GA SEE11341635GA 00840682109260SEE11341635GA SEE11341636GA 00840682109260SEE11341636GA SEE11341637GA 00840682109260SEE11341637GA SEE11341638GA 00840682109260SEE11341638GA SEE11341639GA 00840682109260SEE11341639GA SEE11341641GA 00840682109260SEE11341641GA SEE11341644GA 00840682109260SEE11341644GA SEE11341645GA 00840682109260SEE11341645GA SEE11341647GA 00840682109260SEE11341647GA SEE11341648GA 00840682109260SEE11341648GA SEE11341649GA 00840682109260SEE11341649GA SEE11341650GA 00840682109260SEE11341650GA SEE11341651GA 00840682109260SEE11341651GA SEE11341652GA 00840682109260SEE11341652GA SEE11341653GA 00840682109260SEE11341653GA SEE11341654GA 00840682109260SEE11341654GA SEE11341655GA 00840682109260SEE11341655GA SEE11341658GA 00840682109260SEE11341658GA SEE11341659GA 00840682109260SEE11341659GA SEE11341660GA 00840682109260SEE11341660GA SEE11341661GA 00840682109260SEE11341661GA SEE11341662GA 00840682109260SEE11341662GA SEE11341663GA 00840682109260SEE11341663GA SEE11341665GA 00840682109260SEE11341665GA SEE11351667GA 00840682109260SEE11351667GA SEE11351668GA 00840682109260SEE11351668GA SEE11361670GA 00840682109260SEE11361670GA SEE11361672GA 00840682109260SEE11361672GA SEE11361674GA 00840682109260SEE11361674GA SEE11361676GA 00840682109260SEE11361676GA SEE11361677GA 00840682109260SEE11361677GA SEE11361679GA 00840682109260SEE11361679GA SEE11361681GA 00840682109260SEE11361681GA SEE11361684GA 00840682109260SEE11361684GA SEE11361685GA 00840682109260SEE11361685GA SEE11361686GA 00840682109260SEE11361686GA SEE11361689GA 00840682109260SEE11361689GA SEE11361690GA 00840682109260SEE11361690GA SEE11361691GA 00840682109260SEE11361691GA SEE11361692GA 00840682109260SEE11361692GA SEE11361693GA 00840682109260SEE11361693GA SEE11361694GA 00840682109260SEE11361694GA SEE11361698GA 00840682109260SEE11361698GA SEE11361701GA 00840682109260SEE11361701GA SEE11361703GA 00840682109260SEE11361703GA SEE11361704GA 00840682109260SEE11361704GA SEE11361705GA 00840682109260SEE11361705GA SEE11361706GA 00840682109260SEE11361706GA SEE11361707GA 00840682109260SEE11361707GA SEE11361708GA 00840682109260SEE11361708GA SEE11361709GA 00840682109260SEE11361709GA SEE11361710GA 00840682109260SEE11361710GA SEE11361711GA 00840682109260SEE11361711GA SEE11361713GA 00840682109260SEE11361713GA SEE11361714GA 00840682109260SEE11361714GA SEE11361715GA 00840682109260SEE11361715GA SEE11361716GA 00840682109260SEE11361716GA SEE11361717GA 00840682109260SEE11361717GA SEE11361718GA 00840682109260SEE11361718GA SEE11361719GA 00840682109260SEE11361719GA SEE11361720GA 00840682109260SEE11361720GA SEE11361721GA 00840682109260SEE11361721GA SEE11361723GA 00840682109260SEE11361723GA SEE11361724GA 00840682109260SEE11361724GA SEE11361725GA 00840682109260SEE11361725GA SEE11361727GA 00840682109260SEE11361727GA SEE11361728GA 00840682109260SEE11361728GA SEE11361730GA 00840682109260SEE11361730GA SEE11361731GA 00840682109260SEE11361731GA SEE11361732GA 00840682109260SEE11361732GA SEE11361733GA 00840682109260SEE11361733GA SEE11361735GA 00840682109260SEE11361735GA SEE11361736GA 00840682109260SEE11361736GA SEE11361737GA 00840682109260SEE11361737GA SEE11361738GA 00840682109260SEE11361738GA SEE11361739GA 00840682109260SEE11361739GA SEE11361740GA 00840682109260SEE11361740GA SEE11361741GA 00840682109260SEE11361741GA SEE11361742GA 00840682109260SEE11361742GA SEE11361743GA 00840682109260SEE11361743GA SEE11361744GA 00840682109260SEE11361744GA SEE11361745GA 00840682109260SEE11361745GA SEE11371746GA 00840682109260SEE11371746GA SEE11371748GA 00840682109260SEE11371748GA SEE11371749GA 00840682109260SEE11371749GA SEE11371750GA 00840682109260SEE11371750GA SEE11371751GA 00840682109260SEE11371751GA SEE11371752GA 00840682109260SEE11371752GA SEE11371753GA 00840682109260SEE11371753GA SEE11371754GA 00840682109260SEE11371754GA SEE11371755GA 00840682109260SEE11371755GA SEE11371756GA 00840682109260SEE11371756GA SEE11371757GA 00840682109260SEE11371757GA SEE11371758GA 00840682109260SEE11371758GA SEE11371759GA 00840682109260SEE11371759GA SEE11371760GA 00840682109260SEE11371760GA SEE11371761GA 00840682109260SEE11371761GA SEE11371762GA 00840682109260SEE11371762GA SEE11371763GA 00840682109260SEE11371763GA SEE11371764GA 00840682109260SEE11371764GA SEE11371765GA 00840682109260SEE11371765GA SEE11371766GA 00840682109260SEE11371766GA SEE11371768GA 00840682109260SEE11371768GA SEE11371769GA 00840682109260SEE11371769GA SEE11371770GA 00840682109260SEE11371770GA SEE11371771GA 00840682109260SEE11371771GA SEE11371772GA 00840682109260SEE11371772GA SEE11371773GA 00840682109260SEE11371773GA SEE11371774GA 00840682109260SEE11371774GA SEE11371775GA 00840682109260SEE11371775GA SEE11371776GA 00840682109260SEE11371776GA SEE11371777GA 00840682109260SEE11371777GA SEE11371778GA 00840682109260SEE11371778GA SEE11371779GA 00840682109260SEE11371779GA SEE11371780GA 00840682109260SEE11371780GA SEE11371781GA 00840682109260SEE11371781GA SEE11371782GA 00840682109260SEE11371782GA SEE11371783GA 00840682109260SEE11371783GA SEE11371784GA 00840682109260SEE11371784GA SEE11371785GA 00840682109260SEE11371785GA SEE11371786GA 00840682109260SEE11371786GA SEE11371788GA 00840682109260SEE11371788GA SEE11371791GA 00840682109260SEE11371791GA SEE11371792GA 00840682109260SEE11371792GA SEE11371793GA 00840682109260SEE11371793GA SEE11371795GA 00840682109260SEE11371795GA SEE11371796GA 00840682109260SEE11371796GA SEE11371797GA 00840682109260SEE11371797GA SEE11371799GA 00840682109260SEE11371799GA SEE11371800GA 00840682109260SEE11371800GA SEE11371801GA 00840682109260SEE11371801GA SEE11371802GA 00840682109260SEE11371802GA SEE11371803GA 00840682109260SEE11371803GA SEE11371804GA 00840682109260SEE11371804GA SEE11371805GA 00840682109260SEE11371805GA SEE11371806GA 00840682109260SEE11371806GA SEE11371807GA 00840682109260SEE11371807GA SEE11371809GA 00840682109260SEE11371809GA SEE11371811GA 00840682109260SEE11371811GA SEE11371812GA 00840682109260SEE11371812GA SEE11371813GA 00840682109260SEE11371813GA SEE11381814GA 00840682109260SEE11381814GA SEE11381815GA 00840682109260SEE11381815GA SEE11381816GA 00840682109260SEE11381816GA SEE11381817GA 00840682109260SEE11381817GA SEE11381818GA 00840682109260SEE11381818GA SEE11381819GA 00840682109260SEE11381819GA SEE11381820GA 00840682109260SEE11381820GA SEE11381821GA 00840682109260SEE11381821GA SEE11381822GA 00840682109260SEE11381822GA SEE11381823GA 00840682109260SEE11381823GA SEE11381824GA 00840682109260SEE11381824GA SEE11381825GA 00840682109260SEE11381825GA SEE11381826GA 00840682109260SEE11381826GA SEE11381827GA 00840682109260SEE11381827GA SEE11381828GA 00840682109260SEE11381828GA SEE11381829GA 00840682109260SEE11381829GA SEE11381830GA 00840682109260SEE11381830GA SEE11381831GA 00840682109260SEE11381831GA SEE11391832GA 00840682109260SEE11391832GA SEE11391833GA 00840682109260SEE11391833GA SEE11391834GA 00840682109260SEE11391834GA SEE11391836GA 00840682109260SEE11391836GA SEE11391837GA 00840682109260SEE11391837GA SEE11391838GA 00840682109260SEE11391838GA SEE11391839GA 00840682109260SEE11391839GA SEE11511862GA 00840682109260SEE11511862GA SEE11511863GA 00840682109260SEE11511863GA SEE11511865GA 00840682109260SEE11511865GA SEE11511866GA 00840682109260SEE11511866GA SEE11511867GA 00840682109260SEE11511867GA SEE11511868GA 00840682109260SEE11511868GA SEE11511869GA 00840682109260SEE11511869GA SEE11511870GA 00840682109260SEE11511870GA SEE11511872GA 00840682109260SEE11511872GA SEE11511873GA 00840682109260SEE11511873GA SEE11511874GA 00840682109260SEE11511874GA SEE11511875GA 00840682109260SEE11511875GA SEE11511876GA 00840682109260SEE11511876GA SEE11511877GA 00840682109260SEE11511877GA SEE11511878GA 00840682109260SEE11511878GA SEE11511879GA 00840682109260SEE11511879GA SEE11511880GA 00840682109260SEE11511880GA SEE11511881GA 00840682109260SEE11511881GA SEE11511882GA 00840682109260SEE11511882GA SEE11511883GA 00840682109260SEE11511883GA SEE11511885GA 00840682109260SEE11511885GA SEE11511886GA 00840682109260SEE11511886GA SEE11511887GA 00840682109260SEE11511887GA SEE11521888GA 00840682109260SEE11521888GA SEE11521889GA 00840682109260SEE11521889GA SEE11521890GA 00840682109260SEE11521890GA SEE11521891GA 00840682109260SEE11521891GA SEE11521892GA 00840682109260SEE11521892GA SEE12011893GA 00840682109260SEE12011893GA SEE12011894GA 00840682109260SEE12011894GA SEE12011895GA 00840682109260SEE12011895GA SEE12011896GA 00840682109260SEE12011896GA SEE12011897GA 00840682109260SEE12011897GA SEE12011898GA 00840682109260SEE12011898GA SEE12011899GA 00840682109260SEE12011899GA SEE12011900GA 00840682109260SEE12011900GA SEE12021901GA 00840682109260SEE12021901GA SEE12021902GA 00840682109260SEE12021902GA SEE12021903GA 00840682109260SEE12021903GA SEE12021904GA 00840682109260SEE12021904GA SEE12021905GA 00840682109260SEE12021905GA SEE12021906GA 00840682109260SEE12021906GA SEE12021907GA 00840682109260SEE12021907GA SEE12021908GA 00840682109260SEE12021908GA SEE12021909GA 00840682109260SEE12021909GA SEE12021910GA 00840682109260SEE12021910GA SEE12021911GA 00840682109260SEE12021911GA SEE12021912GA 00840682109260SEE12021912GA SEE12021913GA 00840682109260SEE12021913GA SEE12021914GA 00840682109260SEE12021914GA SEE12021915GA 00840682109260SEE12021915GA SEE12021916GA 00840682109260SEE12021916GA SEE12021917GA 00840682109260SEE12021917GA SEE12021918GA 00840682109260SEE12021918GA SEE12021919GA 00840682109260SEE12021919GA SEE12021920GA 00840682109260SEE12021920GA SEE12021921GA 00840682109260SEE12021921GA SEE12021922GA 00840682109260SEE12021922GA SEE12021923GA 00840682109260SEE12021923GA SEE12021924GA 00840682109260SEE12021924GA SEE12031925GA 00840682109260SEE12031925GA SEE12031926GA 00840682109260SEE12031926GA SEE12031927GA 00840682109260SEE12031927GA SEE12031928GA 00840682109260SEE12031928GA SEE12031929GA 00840682109260SEE12031929GA SEE12031930GA 00840682109260SEE12031930GA SEE12031931GA 00840682109260SEE12031931GA SEE12031932GA 00840682109260SEE12031932GA SEE12031933GA 00840682109260SEE12031933GA SEE12031934GA 00840682109260SEE12031934GA SEE12031935GA 00840682109260SEE12031935GA SEE12031936GA 00840682109260SEE12031936GA SEE12031937GA 00840682109260SEE12031937GA SEE12031938GA 00840682109260SEE12031938GA SEE12031940GA 00840682109260SEE12031940GA SEE12031941GA 00840682109260SEE12031941GA SEE12031943GA 00840682109260SEE12031943GA SEE12031944GA 00840682109260SEE12031944GA SEE12041946GA 00840682109260SEE12041946GA SEE12041947GA 00840682109260SEE12041947GA SEE12041948GA 00840682109260SEE12041948GA SEE12041949GA 00840682109260SEE12041949GA SEE12041951GA 00840682109260SEE12041951GA SEE12041952GA 00840682109260SEE12041952GA SEE12041953GA 00840682109260SEE12041953GA SEE12041954GA 00840682109260SEE12041954GA SEE12041955GA 00840682109260SEE12041955GA SEE12041958GA 00840682109260SEE12041958GA SEE12041959GA 00840682109260SEE12041959GA SEE12041960GA 00840682109260SEE12041960GA SEE12041961GA 00840682109260SEE12041961GA SEE12041962GA 00840682109260SEE12041962GA SEE12041963GA 00840682109260SEE12041963GA SEE12041964GA 00840682109260SEE12041964GA SEE12041965GA 00840682109260SEE12041965GA SEE12041966GA 00840682109260SEE12041966GA SEE12041967GA 00840682109260SEE12041967GA SEE12041968GA 00840682109260SEE12041968GA SEE12041969GA 00840682109260SEE12041969GA SEE12041970GA 00840682109260SEE12041970GA SEE12041971GA 00840682109260SEE12041971GA SEE12041972GA 00840682109260SEE12041972GA SEE12041973GA 00840682109260SEE12041973GA SEE12041974GA 00840682109260SEE12041974GA SEE12041975GA 00840682109260SEE12041975GA SEE12041976GA 00840682109260SEE12041976GA SEE12041977GA 00840682109260SEE12041977GA SEE12051978GA 00840682109260SEE12051978GA SEE12051979GA 00840682109260SEE12051979GA SEE12051980GA 00840682109260SEE12051980GA SEE12051981GA 00840682109260SEE12051981GA SEE12051982GA 00840682109260SEE12051982GA SEE12051983GA 00840682109260SEE12051983GA SEE12051984GA 00840682109260SEE12051984GA SEE12051985GA 00840682109260SEE12051985GA SEE12051988GA 00840682109260SEE12051988GA SEE12051989GA 00840682109260SEE12051989GA SEE12051993GA 00840682109260SEE12051993GA SEE12051994GA 00840682109260SEE12051994GA SEE12051995GA 00840682109260SEE12051995GA SEE12061996GA 00840682109260SEE12061996GA SEE12061997GA 00840682109260SEE12061997GA SEE12061998GA 00840682109260SEE12061998GA SEE12061999GA 00840682109260SEE12061999GA SEE12062000GA 00840682109260SEE12062000GA SEE12062001GA 00840682109260SEE12062001GA SEE12062002GA 00840682109260SEE12062002GA SEE12062003GA 00840682109260SEE12062003GA SEE12062005GA 00840682109260SEE12062005GA SEE12062010GA 00840682109260SEE12062010GA SEE12062011GA 00840682109260SEE12062011GA SEE12062013GA 00840682109260SEE12062013GA SEE12082016GA 00840682109260SEE12082016GA SEE12082017GA 00840682109260SEE12082017GA SEE12082018GA 00840682109260SEE12082018GA SEE12082019GA 00840682109260SEE12082019GA SEE12082020GA 00840682109260SEE12082020GA SEE12082021GA 00840682109260SEE12082021GA SEE12082022GA 00840682109260SEE12082022GA SEE12082023GA 00840682109260SEE12082023GA SEE12082024GA 00840682109260SEE12082024GA SEE12082026GA 00840682109260SEE12082026GA SEE12082027GA 00840682109260SEE12082027GA SEE12082029GA 00840682109260SEE12082029GA SEE12082030GA 00840682109260SEE12082030GA SEE12082033GA 00840682109260SEE12082033GA SEE12082035GA 00840682109260SEE12082035GA SEE12082036GA 00840682109260SEE12082036GA SEE12082037GA 00840682109260SEE12082037GA SEE12082038GA 00840682109260SEE12082038GA SEE12082039GA 00840682109260SEE12082039GA SEE12082040GA 00840682109260SEE12082040GA SEE12092041GA 00840682109260SEE12092041GA SEE12092042GA 00840682109260SEE12092042GA SEE12092043GA 00840682109260SEE12092043GA SEE12092046GA 00840682109260SEE12092046GA SEE12092048GA 00840682109260SEE12092048GA SEE12092050GA 00840682109260SEE12092050GA SEE12092051GA 00840682109260SEE12092051GA SEE12092053GA 00840682109260SEE12092053GA SEE12092054GA 00840682109260SEE12092054GA SEE12092055GA 00840682109260SEE12092055GA SEE12092056GA 00840682109260SEE12092056GA SEE12092058GA 00840682109260SEE12092058GA SEE12092059GA 00840682109260SEE12092059GA SEE12092060GA 00840682109260SEE12092060GA SEE12092061GA 00840682109260SEE12092061GA SEE12092062GA 00840682109260SEE12092062GA SEE12092063GA 00840682109260SEE12092063GA SEE12092064GA 00840682109260SEE12092064GA SEE12092065GA 00840682109260SEE12092065GA SEE12092067GA 00840682109260SEE12092067GA SEE12092068GA 00840682109260SEE12092068GA SEE12092069GA 00840682109260SEE12092069GA SEE12092070GA 00840682109260SEE12092070GA SEE12092071GA 00840682109260SEE12092071GA SEE12092072GA 00840682109260SEE12092072GA SEE12092073GA 00840682109260SEE12092073GA SEE12092074GA 00840682109260SEE12092074GA SEE12092075GA 00840682109260SEE12092075GA SEE12092076GA 00840682109260SEE12092076GA SEE12092077GA 00840682109260SEE12092077GA SEE12092078GA 00840682109260SEE12092078GA SEE12092079GA 00840682109260SEE12092079GA SEE12092080GA 00840682109260SEE12092080GA SEE12092081GA 00840682109260SEE12092081GA SEE12092083GA 00840682109260SEE12092083GA SEE12092084GA 00840682109260SEE12092084GA SEE12092085GA 00840682109260SEE12092085GA SEE12102086GA 00840682109260SEE12102086GA SEE12102087GA 00840682109260SEE12102087GA SEE12102088GA 00840682109260SEE12102088GA SEE12102090GA 00840682109260SEE12102090GA SEE12102092GA 00840682109260SEE12102092GA SEE12102093GA 00840682109260SEE12102093GA SEE12102095GA 00840682109260SEE12102095GA SEE12102097GA 00840682109260SEE12102097GA SEE12102098GA 00840682109260SEE12102098GA SEE12102099GA 00840682109260SEE12102099GA SEE12102100GA 00840682109260SEE12102100GA SEE12102101GA 00840682109260SEE12102101GA SEE12102102GA 00840682109260SEE12102102GA SEE12102103GA 00840682109260SEE12102103GA SEE12102105GA 00840682109260SEE12102105GA SEE12102106GA 00840682109260SEE12102106GA SEE12112109GA 00840682109260SEE12112109GA SEE12112110GA 00840682109260SEE12112110GA SEE12112111GA 00840682109260SEE12112111GA SEE12112116GA 00840682109260SEE12112116GA SEE12112119GA 00840682109260SEE12112119GA SEE12112121GA 00840682109260SEE12112121GA SEE12112122GA 00840682109260SEE12112122GA SEE12122127GA 00840682109260SEE12122127GA SEE12122129GA 00840682109260SEE12122129GA SEE12122130GA 00840682109260SEE12122130GA SEE12122132GA 00840682109260SEE12122132GA SEE12122133GA 00840682109260SEE12122133GA SEE12122134GA 00840682109260SEE12122134GA SEE12122135GA 00840682109260SEE12122135GA SEE12122136GA 00840682109260SEE12122136GA SEE12122137GA 00840682109260SEE12122137GA SEE12122138GA 00840682109260SEE12122138GA SEE12122141GA 00840682109260SEE12122141GA SEE12122142GA 00840682109260SEE12122142GA SEE12122144GA 00840682109260SEE12122144GA SEE12122145GA 00840682109260SEE12122145GA SEE12122146GA 00840682109260SEE12122146GA SEE12122147GA 00840682109260SEE12122147GA SEE12122148GA 00840682109260SEE12122148GA SEE12122149GA 00840682109260SEE12122149GA SEE12122150GA 00840682109260SEE12122150GA SEE12122151GA 00840682109260SEE12122151GA SEE12122153GA 00840682109260SEE12122153GA SEE12122154GA 00840682109260SEE12122154GA SEE12122155GA 00840682109260SEE12122155GA SEE12122156GA 00840682109260SEE12122156GA SEE12122157GA 00840682109260SEE12122157GA SEE12122158GA 00840682109260SEE12122158GA SEE12122159GA 00840682109260SEE12122159GA SEE12122161GA 00840682109260SEE12122161GA SEE12122162GA 00840682109260SEE12122162GA SEE12122163GA 00840682109260SEE12122163GA SEE12122165GA 00840682109260SEE12122165GA SEE12122166GA 00840682109260SEE12122166GA SEE12122167GA 00840682109260SEE12122167GA SEE12162170GA 00840682109260SEE12162170GA SEE12162171GA 00840682109260SEE12162171GA SEE12162172GA 00840682109260SEE12162172GA SEE12162174GA 00840682109260SEE12162174GA SEE12162175GA 00840682109260SEE12162175GA SEE12162176GA 00840682109260SEE12162176GA SEE12162178GA 00840682109260SEE12162178GA SEE12162179GA 00840682109260SEE12162179GA SEE12162180GA 00840682109260SEE12162180GA SEE12162183GA 00840682109260SEE12162183GA SEE12162184GA 00840682109260SEE12162184GA SEE12172188GA 00840682109260SEE12172188GA SEE12172189GA 00840682109260SEE12172189GA SEE12172191GA 00840682109260SEE12172191GA SEE12172193GA 00840682109260SEE12172193GA SEE12172194GA 00840682109260SEE12172194GA SEE12172195GA 00840682109260SEE12172195GA SEE12172196GA 00840682109260SEE12172196GA SEE12172198GA 00840682109260SEE12172198GA SEE12172199GA 00840682109260SEE12172199GA SEE12172200GA 00840682109260SEE12172200GA SEE12172201GA 00840682109260SEE12172201GA SEE12172204GA 00840682109260SEE12172204GA SEE12182205GA 00840682109260SEE12182205GA SEE12182206GA 00840682109260SEE12182206GA SEE12182207GA 00840682109260SEE12182207GA SEE12182208GA 00840682109260SEE12182208GA SEE12182209GA 00840682109260SEE12182209GA SEE12182210GA 00840682109260SEE12182210GA SEE12182211GA 00840682109260SEE12182211GA SEE12182213GA 00840682109260SEE12182213GA SEE12182214GA 00840682109260SEE12182214GA SEE12182215GA 00840682109260SEE12182215GA SEE12182216GA 00840682109260SEE12182216GA SEE12182217GA 00840682109260SEE12182217GA SEE12182218GA 00840682109260SEE12182218GA SEE12182219GA 00840682109260SEE12182219GA SEE12182220GA 00840682109260SEE12182220GA SEE12182221GA 00840682109260SEE12182221GA SEE12182222GA 00840682109260SEE12182222GA SEE12182223GA 00840682109260SEE12182223GA SEE12182224GA 00840682109260SEE12182224GA SEE12182226GA 00840682109260SEE12182226GA SEE12182227GA 00840682109260SEE12182227GA SEE12182228GA 00840682109260SEE12182228GA SEE12182229GA 00840682109260SEE12182229GA SEE12182230GA 00840682109260SEE12182230GA SEE12182231GA 00840682109260SEE12182231GA SEE12182232GA 00840682109260SEE12182232GA SEE12182233GA 00840682109260SEE12182233GA SEE12232252GA 00840682109260SEE12232252GA SEE12232253GA 00840682109260SEE12232253GA SEE12232254GA 00840682109260SEE12232254GA SEE12232255GA 00840682109260SEE12232255GA SEE12232256GA 00840682109260SEE12232256GA SEE12232257GA 00840682109260SEE12232257GA SEE12232258GA 00840682109260SEE12232258GA SEE12232259GA 00840682109260SEE12232259GA SEE12232261GA 00840682109260SEE12232261GA SEE12232262GA 00840682109260SEE12232262GA SEE12232263GA 00840682109260SEE12232263GA SEE12232265GA 00840682109260SEE12232265GA SEE12232266GA 00840682109260SEE12232266GA SEE12242267GA 00840682109260SEE12242267GA SEE12242268GA 00840682109260SEE12242268GA SEE12242269GA 00840682109260SEE12242269GA SEE12242270GA 00840682109260SEE12242270GA SEE12242275GA 00840682109260SEE12242275GA SEE12242276GA 00840682109260SEE12242276GA SEE12242277GA 00840682109260SEE12242277GA SEE12242278GA 00840682109260SEE12242278GA SEE12242279GA 00840682109260SEE12242279GA SEE12242280GA 00840682109260SEE12242280GA SEE12242281GA 00840682109260SEE12242281GA SEE12242283GA 00840682109260SEE12242283GA SEE12242285GA 00840682109260SEE12242285GA SEE12242286GA 00840682109260SEE12242286GA SEE12242287GA 00840682109260SEE12242287GA SEE12242288GA 00840682109260SEE12242288GA SEE12242289GA 00840682109260SEE12242289GA SEE12242291GA 00840682109260SEE12242291GA SEE12242292GA 00840682109260SEE12242292GA SEE12242293GA 00840682109260SEE12242293GA SEE12252296GA 00840682109260SEE12252296GA SEE12252297GA 00840682109260SEE12252297GA SEE12252298GA 00840682109260SEE12252298GA SEE12252299GA 00840682109260SEE12252299GA SEE12252300GA 00840682109260SEE12252300GA SEE12252301GA 00840682109260SEE12252301GA SEE12252302GA 00840682109260SEE12252302GA SEE12252303GA 00840682109260SEE12252303GA SEE12252305GA 00840682109260SEE12252305GA SEE12252306GA 00840682109260SEE12252306GA SEE12252307GA 00840682109260SEE12252307GA SEE12252308GA 00840682109260SEE12252308GA SEE12252309GA 00840682109260SEE12252309GA SEE12252310GA 00840682109260SEE12252310GA SEE12252311GA 00840682109260SEE12252311GA SEE12252312GA 00840682109260SEE12252312GA SEE12252313GA 00840682109260SEE12252313GA SEE12252314GA 00840682109260SEE12252314GA SEE12252315GA 00840682109260SEE12252315GA SEE12252316GA 00840682109260SEE12252316GA SEE12252317GA 00840682109260SEE12252317GA SEE12252318GA 00840682109260SEE12252318GA SEE12252320GA 00840682109260SEE12252320GA SEE12252322GA 00840682109260SEE12252322GA SEE12252323GA 00840682109260SEE12252323GA SEE12252325GA 00840682109260SEE12252325GA SEE12262326GA 00840682109260SEE12262326GA SEE12262327GA 00840682109260SEE12262327GA SEE12262328GA 00840682109260SEE12262328GA SEE12262329GA 00840682109260SEE12262329GA SEE12262330GA 00840682109260SEE12262330GA SEE12262331GA 00840682109260SEE12262331GA SEE12262332GA 00840682109260SEE12262332GA SEE12262333GA 00840682109260SEE12262333GA SEE12262334GA 00840682109260SEE12262334GA SEE12262335GA 00840682109260SEE12262335GA SEE12262336GA 00840682109260SEE12262336GA SEE12262337GA 00840682109260SEE12262337GA SEE12262341GA 00840682109260SEE12262341GA SEE12262342GA 00840682109260SEE12262342GA SEE12262343GA 00840682109260SEE12262343GA SEE12262344GA 00840682109260SEE12262344GA SEE12262345GA 00840682109260SEE12262345GA SEE12262346GA 00840682109260SEE12262346GA SEE12262347GA 00840682109260SEE12262347GA SEE12282348GA 00840682109260SEE12282348GA SEE12282350GA 00840682109260SEE12282350GA SEE12282351GA 00840682109260SEE12282351GA SEE12282352GA 00840682109260SEE12282352GA SEE12282353GA 00840682109260SEE12282353GA SEE12282354GA 00840682109260SEE12282354GA SEE12282356GA 00840682109260SEE12282356GA SEE12282357GA 00840682109260SEE12282357GA SEE12282358GA 00840682109260SEE12282358GA SEE12282360GA 00840682109260SEE12282360GA SEE12292361GA 00840682109260SEE12292361GA SEE12292362GA 00840682109260SEE12292362GA SEE12292363GA 00840682109260SEE12292363GA SEE12292364GA 00840682109260SEE12292364GA SEE12292365GA 00840682109260SEE12292365GA SEE12292367GA 00840682109260SEE12292367GA SEE12302368GA 00840682109260SEE12302368GA SEE12302369GA 00840682109260SEE12302369GA SEE12302370GA 00840682109260SEE12302370GA SEE12302371GA 00840682109260SEE12302371GA SEE12302372GA 00840682109260SEE12302372GA SEE12302373GA 00840682109260SEE12302373GA SEE12302374GA 00840682109260SEE12302374GA SEE12302375GA 00840682109260SEE12302375GA SEE12302376GA 00840682109260SEE12302376GA SEE12302377GA 00840682109260SEE12302377GA SEE12302378GA 00840682109260SEE12302378GA SEE12302379GA 00840682109260SEE12302379GA SEE12302381GA 00840682109260SEE12302381GA SEE12302383GA 00840682109260SEE12302383GA SEE12302384GA 00840682109260SEE12302384GA SEE12302385GA 00840682109260SEE12302385GA SEE12302386GA 00840682109260SEE12302386GA SEE12302388GA 00840682109260SEE12302388GA SEE12302389GA 00840682109260SEE12302389GA SEE12302390GA 00840682109260SEE12302390GA SEE12302391GA 00840682109260SEE12302391GA SEE12302392GA 00840682109260SEE12302392GA SEE12302393GA 00840682109260SEE12302393GA SEE12302394GA 00840682109260SEE12302394GA SEE12302396GA 00840682109260SEE12302396GA SEE12312397GA 00840682109260SEE12312397GA SEE12312398GA 00840682109260SEE12312398GA SEE12312399GA 00840682109260SEE12312399GA SEE12312400GA 00840682109260SEE12312400GA SEE12312401GA 00840682109260SEE12312401GA SEE12312402GA 00840682109260SEE12312402GA SEE12312403GA 00840682109260SEE12312403GA SEE12312404GA 00840682109260SEE12312404GA SEE12312405GA 00840682109260SEE12312405GA SEE12312406GA 00840682109260SEE12312406GA SEE12322407GA 00840682109260SEE12322407GA SEE12322408GA 00840682109260SEE12322408GA SEE12322409GA 00840682109260SEE12322409GA SEE12322410GA 00840682109260SEE12322410GA SEE12322413GA 00840682109260SEE12322413GA SEE12322414GA 00840682109260SEE12322414GA SEE12322415GA 00840682109260SEE12322415GA SEE12322416GA 00840682109260SEE12322416GA SEE12322417GA 00840682109260SEE12322417GA SEE12322418GA 00840682109260SEE12322418GA SEE12322419GA 00840682109260SEE12322419GA SEE12322420GA 00840682109260SEE12322420GA SEE12322421GA 00840682109260SEE12322421GA SEE12322422GA 00840682109260SEE12322422GA SEE12322423GA 00840682109260SEE12322423GA SEE12322424GA 00840682109260SEE12322424GA SEE12322425GA 00840682109260SEE12322425GA SEE12322427GA 00840682109260SEE12322427GA SEE12322428GA 00840682109260SEE12322428GA SEE12322429GA 00840682109260SEE12322429GA SEE12322430GA 00840682109260SEE12322430GA SEE12322431GA 00840682109260SEE12322431GA SEE12322432GA 00840682109260SEE12322432GA SEE12322433GA 00840682109260SEE12322433GA SEE12322434GA 00840682109260SEE12322434GA SEE12322435GA 00840682109260SEE12322435GA SEE12322436GA 00840682109260SEE12322436GA SEE12322437GA 00840682109260SEE12322437GA SEE12322438GA 00840682109260SEE12322438GA SEE12322439GA 00840682109260SEE12322439GA SEE12332441GA 00840682109260SEE12332441GA SEE12332443GA 00840682109260SEE12332443GA SEE12332444GA 00840682109260SEE12332444GA SEE12332445GA 00840682109260SEE12332445GA SEE12340001GA 00840682109260SEE12340001GA SEE12340002GA 00840682109260SEE12340002GA SEE12340003GA 00840682109260SEE12340003GA SEE12340004GA 00840682109260SEE12340004GA SEE12340005GA 00840682109260SEE12340005GA SEE12340007GA 00840682109260SEE12340007GA SEE12340008GA 00840682109260SEE12340008GA SEE12340009GA 00840682109260SEE12340009GA SEE12340011GA 00840682109260SEE12340011GA SEE12340012GA 00840682109260SEE12340012GA SEE12340013GA 00840682109260SEE12340013GA SEE12340014GA 00840682109260SEE12340014GA SEE12340015GA 00840682109260SEE12340015GA SEE12340016GA 00840682109260SEE12340016GA SEE12340017GA 00840682109260SEE12340017GA SEE12340018GA 00840682109260SEE12340018GA SEE12340019GA 00840682109260SEE12340019GA SEE12340020GA 00840682109260SEE12340020GA SEE12342446GA 00840682109260SEE12342446GA SEE12342447GA 00840682109260SEE12342447GA SEE12342449GA 00840682109260SEE12342449GA SEE12342450GA 00840682109260SEE12342450GA SEE12342451GA 00840682109260SEE12342451GA SEE12342452GA 00840682109260SEE12342452GA SEE12342453GA 00840682109260SEE12342453GA SEE12342454GA 00840682109260SEE12342454GA SEE12342455GA 00840682109260SEE12342455GA SEE12350021GA 00840682109260SEE12350021GA SEE12350022GA 00840682109260SEE12350022GA SEE12350024GA 00840682109260SEE12350024GA SEE12350025GA 00840682109260SEE12350025GA SEE12350026GA 00840682109260SEE12350026GA SEE12350027GA 00840682109260SEE12350027GA SEE12350028GA 00840682109260SEE12350028GA SEE12350029GA 00840682109260SEE12350029GA SEE12350030GA 00840682109260SEE12350030GA SEE12360032GA 00840682109260SEE12360032GA SEE12360033GA 00840682109260SEE12360033GA SEE12360035GA 00840682109260SEE12360035GA SEE12360036GA 00840682109260SEE12360036GA SEE12360037GA 00840682109260SEE12360037GA SEE12360040GA 00840682109260SEE12360040GA SEE12370044GA 00840682109260SEE12370044GA SEE12370046GA 00840682109260SEE12370046GA SEE12370047GA 00840682109260SEE12370047GA SEE12370048GA 00840682109260SEE12370048GA SEE12370049GA 00840682109260SEE12370049GA SEE12370050GA 00840682109260SEE12370050GA SEE12370051GA 00840682109260SEE12370051GA SEE12370052GA 00840682109260SEE12370052GA SEE12370054GA 00840682109260SEE12370054GA SEE12370055GA 00840682109260SEE12370055GA SEE12370056GA 00840682109260SEE12370056GA SEE12370057GA 00840682109260SEE12370057GA SEE12370058GA 00840682109260SEE12370058GA SEE12370059GA 00840682109260SEE12370059GA SEE12370061GA 00840682109260SEE12370061GA SEE12370063GA 00840682109260SEE12370063GA SEE12370064GA 00840682109260SEE12370064GA SEE12370066GA 00840682109260SEE12370066GA SEE12370067GA 00840682109260SEE12370067GA SEE12370068GA 00840682109260SEE12370068GA SEE12390071GA 00840682109260SEE12390071GA SEE12390072GA 00840682109260SEE12390072GA SEE12390073GA 00840682109260SEE12390073GA SEE12390074GA 00840682109260SEE12390074GA SEE12390075GA 00840682109260SEE12390075GA SEE12390077GA 00840682109260SEE12390077GA SEE12390078GA 00840682109260SEE12390078GA SEE12390079GA 00840682109260SEE12390079GA SEE12390080GA 00840682109260SEE12390080GA SEE12390081GA 00840682109260SEE12390081GA SEE12390082GA 00840682109260SEE12390082GA SEE12390083GA 00840682109260SEE12390083GA SEE12390084GA 00840682109260SEE12390084GA SEE12390085GA 00840682109260SEE12390085GA SEE12390086GA 00840682109260SEE12390086GA SEE12390087GA 00840682109260SEE12390087GA SEE12390088GA 00840682109260SEE12390088GA SEE12390089GA 00840682109260SEE12390089GA SEE12390090GA 00840682109260SEE12390090GA SEE12400091GA 00840682109260SEE12400091GA SEE12400092GA 00840682109260SEE12400092GA SEE12400093GA 00840682109260SEE12400093GA SEE12400094GA 00840682109260SEE12400094GA SEE12400095GA 00840682109260SEE12400095GA SEE12400096GA 00840682109260SEE12400096GA SEE12400098GA 00840682109260SEE12400098GA SEE12400100GA 00840682109260SEE12400100GA SEE12400101GA 00840682109260SEE12400101GA SEE12400102GA 00840682109260SEE12400102GA SEE12400103GA 00840682109260SEE12400103GA SEE12400104GA 00840682109260SEE12400104GA SEE12400105GA 00840682109260SEE12400105GA SEE12400106GA 00840682109260SEE12400106GA SEE12400107GA 00840682109260SEE12400107GA SEE12400108GA 00840682109260SEE12400108GA SEE12400109GA 00840682109260SEE12400109GA SEE12400110GA 00840682109260SEE12400110GA SEE12400111GA 00840682109260SEE12400111GA SEE12400112GA 00840682109260SEE12400112GA SEE12400113GA 00840682109260SEE12400113GA SEE12400114GA 00840682109260SEE12400114GA SEE12400115GA 00840682109260SEE12400115GA SEE12400116GA 00840682109260SEE12400116GA SEE12400117GA 00840682109260SEE12400117GA SEE12400118GA 00840682109260SEE12400118GA SEE12400119GA 00840682109260SEE12400119GA SEE12410121GA 00840682109260SEE12410121GA SEE12410122GA 00840682109260SEE12410122GA SEE12410123GA 00840682109260SEE12410123GA SEE12410124GA 00840682109260SEE12410124GA SEE12410125GA 00840682109260SEE12410125GA SEE12410126GA 00840682109260SEE12410126GA SEE12410127GA 00840682109260SEE12410127GA SEE12410128GA 00840682109260SEE12410128GA SEE12410129GA 00840682109260SEE12410129GA SEE12410130GA 00840682109260SEE12410130GA SEE12410131GA 00840682109260SEE12410131GA SEE12410133GA 00840682109260SEE12410133GA SEE12410134GA 00840682109260SEE12410134GA SEE12410135GA 00840682109260SEE12410135GA SEE12410136GA 00840682109260SEE12410136GA SEE12410137GA 00840682109260SEE12410137GA SEE12410138GA 00840682109260SEE12410138GA SEE12410139GA 00840682109260SEE12410139GA SEE12410140GA 00840682109260SEE12410140GA SEE12410141GA 00840682109260SEE12410141GA SEE12410142GA 00840682109260SEE12410142GA SEE12410143GA 00840682109260SEE12410143GA SEE12410144GA 00840682109260SEE12410144GA SEE12410145GA 00840682109260SEE12410145GA SEE12410146GA 00840682109260SEE12410146GA SEE12410147GA 00840682109260SEE12410147GA SEE12410148GA 00840682109260SEE12410148GA SEE12410149GA 00840682109260SEE12410149GA SEE12420150GA 00840682109260SEE12420150GA SEE12420151GA 00840682109260SEE12420151GA SEE12420152GA 00840682109260SEE12420152GA SEE12420153GA 00840682109260SEE12420153GA SEE12420154GA 00840682109260SEE12420154GA SEE12420155GA 00840682109260SEE12420155GA SEE12420156GA 00840682109260SEE12420156GA SEE12420157GA 00840682109260SEE12420157GA SEE12420159GA 00840682109260SEE12420159GA SEE12420161GA 00840682109260SEE12420161GA SEE12420166GA 00840682109260SEE12420166GA SEE12420167GA 00840682109260SEE12420167GA SEE12420168GA 00840682109260SEE12420168GA SEE12420169GA 00840682109260SEE12420169GA SEE12420172GA 00840682109260SEE12420172GA SEE12420173GA 00840682109260SEE12420173GA SEE12420174GA 00840682109260SEE12420174GA SEE12430177GA 00840682109260SEE12430177GA SEE12430181GA 00840682109260SEE12430181GA SEE12430183GA 00840682109260SEE12430183GA SEE12430184GA 00840682109260SEE12430184GA SEE12430185GA 00840682109260SEE12430185GA SEE12430187GA 00840682109260SEE12430187GA SEE12430188GA 00840682109260SEE12430188GA SEE12430189GA 00840682109260SEE12430189GA SEE12430190GA 00840682109260SEE12430190GA SEE12430191GA 00840682109260SEE12430191GA SEE12430192GA 00840682109260SEE12430192GA SEE12430193GA 00840682109260SEE12430193GA SEE12430194GA 00840682109260SEE12430194GA SEE12430195GA 00840682109260SEE12430195GA SEE12430197GA 00840682109260SEE12430197GA SEE12430198GA 00840682109260SEE12430198GA SEE12430200GA 00840682109260SEE12430200GA SEE12430202GA 00840682109260SEE12430202GA SEE12430203GA 00840682109260SEE12430203GA SEE12430204GA 00840682109260SEE12430204GA SEE12430206GA 00840682109260SEE12430206GA SEE12430207GA 00840682109260SEE12430207GA SEE12430208GA 00840682109260SEE12430208GA SEE12430209GA 00840682109260SEE12430209GA SEE12430210GA 00840682109260SEE12430210GA SEE12430211GA 00840682109260SEE12430211GA SEE12430212GA 00840682109260SEE12430212GA SEE12430213GA 00840682109260SEE12430213GA SEE12430214GA 00840682109260SEE12430214GA SEE12430215GA 00840682109260SEE12430215GA SEE12430216GA 00840682109260SEE12430216GA SEE12430217GA 00840682109260SEE12430217GA SEE12430218GA 00840682109260SEE12430218GA SEE12430219GA 00840682109260SEE12430219GA SEE12430220GA 00840682109260SEE12430220GA SEE12430221GA 00840682109260SEE12430221GA SEE12430222GA 00840682109260SEE12430222GA SEE12430223GA 00840682109260SEE12430223GA SEE12430225GA 00840682109260SEE12430225GA SEE12430226GA 00840682109260SEE12430226GA SEE12430228GA 00840682109260SEE12430228GA SEE12430229GA 00840682109260SEE12430229GA SEE12440231GA 00840682109260SEE12440231GA SEE12440234GA 00840682109260SEE12440234GA SEE12440236GA 00840682109260SEE12440236GA SEE12440237GA 00840682109260SEE12440237GA SEE12440238GA 00840682109260SEE12440238GA SEE12440239GA 00840682109260SEE12440239GA SEE12440240GA 00840682109260SEE12440240GA SEE12490241GA 00840682109260SEE12490241GA SEE12490242GA 00840682109260SEE12490242GA SEE12490245GA 00840682109260SEE12490245GA SEE12490246GA 00840682109260SEE12490246GA SEE12490247GA 00840682109260SEE12490247GA SEE12490248GA 00840682109260SEE12490248GA SEE12490249GA 00840682109260SEE12490249GA SEE12490250GA 00840682109260SEE12490250GA SEE12490251GA 00840682109260SEE12490251GA SEE12490252GA 00840682109260SEE12490252GA SEE12490253GA 00840682109260SEE12490253GA SEE12490255GA 00840682109260SEE12490255GA SEE12490256GA 00840682109260SEE12490256GA SEE12490257GA 00840682109260SEE12490257GA SEE12490258GA 00840682109260SEE12490258GA SEE12490259GA 00840682109260SEE12490259GA SEE12490260GA 00840682109260SEE12490260GA SEE12490261GA 00840682109260SEE12490261GA SEE12490262GA 00840682109260SEE12490262GA SEE12490263GA 00840682109260SEE12490263GA SEE12490264GA 00840682109260SEE12490264GA SEE12490268GA 00840682109260SEE12490268GA SEE12490269GA 00840682109260SEE12490269GA SEE12490270GA 00840682109260SEE12490270GA SEE12490271GA 00840682109260SEE12490271GA SEE12490272GA 00840682109260SEE12490272GA SEE12490273GA 00840682109260SEE12490273GA SEE12490274GA 00840682109260SEE12490274GA SEE12490275GA 00840682109260SEE12490275GA SEE12490277GA 00840682109260SEE12490277GA SEE12490278GA 00840682109260SEE12490278GA SEE12490279GA 00840682109260SEE12490279GA SEE12490280GA 00840682109260SEE12490280GA SEE12490282GA 00840682109260SEE12490282GA SEE12490284GA 00840682109260SEE12490284GA SEE12490286GA 00840682109260SEE12490286GA SEE12490288GA 00840682109260SEE12490288GA SEE12490289GA 00840682109260SEE12490289GA SEE12490290GA 00840682109260SEE12490290GA SEE12490291GA 00840682109260SEE12490291GA SEE12490292GA 00840682109260SEE12490292GA SEE12490293GA 00840682109260SEE12490293GA SEE12490294GA 00840682109260SEE12490294GA SEE12490295GA 00840682109260SEE12490295GA SEE12490297GA 00840682109260SEE12490297GA SEE12490298GA 00840682109260SEE12490298GA SEE12490300GA 00840682109260SEE12490300GA SEE12510301GA 00840682109260SEE12510301GA SEE12510302GA 00840682109260SEE12510302GA SEE12510303GA 00840682109260SEE12510303GA SEE12510304GA 00840682109260SEE12510304GA SEE12510306GA 00840682109260SEE12510306GA SEE12510307GA 00840682109260SEE12510307GA SEE12510308GA 00840682109260SEE12510308GA SEE12510309GA 00840682109260SEE12510309GA SEE12510310GA 00840682109260SEE12510310GA SEE12510311GA 00840682109260SEE12510311GA SEE12510313GA 00840682109260SEE12510313GA SEE12510314GA 00840682109260SEE12510314GA SEE12510315GA 00840682109260SEE12510315GA SEE12510316GA 00840682109260SEE12510316GA SEE12510317GA 00840682109260SEE12510317GA SEE12510318GA 00840682109260SEE12510318GA SEE12510319GA 00840682109260SEE12510319GA SEE12510320GA 00840682109260SEE12510320GA SEE12510325GA 00840682109260SEE12510325GA SEE12510326GA 00840682109260SEE12510326GA SEE12510327GA 00840682109260SEE12510327GA SEE12510328GA 00840682109260SEE12510328GA SEE12510329GA 00840682109260SEE12510329GA SEE12510330GA 00840682109260SEE12510330GA SEE12510331GA 00840682109260SEE12510331GA SEE12510332GA 00840682109260SEE12510332GA SEE12510333GA 00840682109260SEE12510333GA SEE12510334GA 00840682109260SEE12510334GA SEE12520335GA 00840682109260SEE12520335GA SEE12520336GA 00840682109260SEE12520336GA SEE12520337GA 00840682109260SEE12520337GA SEE12520338GA 00840682109260SEE12520338GA SEE12520339GA 00840682109260SEE12520339GA SEE12520340GA 00840682109260SEE12520340GA SEE12520341GA 00840682109260SEE12520341GA SEE12520342GA 00840682109260SEE12520342GA SEE12520343GA 00840682109260SEE12520343GA SEE12520344GA 00840682109260SEE12520344GA SEE12520345GA 00840682109260SEE12520345GA SEE12520346GA 00840682109260SEE12520346GA SEE12520347GA 00840682109260SEE12520347GA SEE12520348GA 00840682109260SEE12520348GA SEE12520349GA 00840682109260SEE12520349GA SEE12520350GA 00840682109260SEE12520350GA SEE12520351GA 00840682109260SEE12520351GA SEE12520352GA 00840682109260SEE12520352GA SEE12520353GA 00840682109260SEE12520353GA SEE12520354GA 00840682109260SEE12520354GA SEE12520355GA 00840682109260SEE12520355GA SEE12520356GA 00840682109260SEE12520356GA SEE12520357GA 00840682109260SEE12520357GA SEE12520359GA 00840682109260SEE12520359GA SEE12520360GA 00840682109260SEE12520360GA SEE12520361GA 00840682109260SEE12520361GA SEE12520362GA 00840682109260SEE12520362GA SEE12520363GA 00840682109260SEE12520363GA SEE12520364GA 00840682109260SEE12520364GA SEE12530365GA 00840682109260SEE12530365GA SEE12530366GA 00840682109260SEE12530366GA SEE12530367GA 00840682109260SEE12530367GA SEE12530368GA 00840682109260SEE12530368GA SEE12530369GA 00840682109260SEE12530369GA SEE12530370GA 00840682109260SEE12530370GA SEE12530371GA 00840682109260SEE12530371GA SEE12530372GA 00840682109260SEE12530372GA SEE12530373GA 00840682109260SEE12530373GA SEE12530374GA 00840682109260SEE12530374GA SEE13000375GA 00840682109260SEE13000375GA SEE13000376GA 00840682109260SEE13000376GA SEE13000377GA 00840682109260SEE13000377GA SEE13000378GA 00840682109260SEE13000378GA SEE13000379GA 00840682109260SEE13000379GA SEE13000380GA 00840682109260SEE13000380GA SEE13000381GA 00840682109260SEE13000381GA SEE13000382GA 00840682109260SEE13000382GA SEE13000383GA 00840682109260SEE13000383GA SEE13000384GA 00840682109260SEE13000384GA SEE13000385GA 00840682109260SEE13000385GA SEE13000386GA 00840682109260SEE13000386GA SEE13000387GA 00840682109260SEE13000387GA SEE13000388GA 00840682109260SEE13000388GA SEE13000390GA 00840682109260SEE13000390GA SEE13000391GA 00840682109260SEE13000391GA SEE13000393GA 00840682109260SEE13000393GA SEE13010395GA 00840682109260SEE13010395GA SEE13010396GA 00840682109260SEE13010396GA SEE13010397GA 00840682109260SEE13010397GA SEE13010398GA 00840682109260SEE13010398GA SEE13010401GA 00840682109260SEE13010401GA SEE13010403GA 00840682109260SEE13010403GA SEE13010404GA 00840682109260SEE13010404GA SEE13010405GA 00840682109260SEE13010405GA SEE13010406GA 00840682109260SEE13010406GA SEE13020407GA 00840682109260SEE13020407GA SEE13020410GA 00840682109260SEE13020410GA SEE13020411GA 00840682109260SEE13020411GA SEE13020412GA 00840682109260SEE13020412GA SEE13020413GA 00840682109260SEE13020413GA SEE13020414GA 00840682109260SEE13020414GA SEE13040415GA 00840682109260SEE13040415GA SEE13040417GA 00840682109260SEE13040417GA SEE13040418GA 00840682109260SEE13040418GA SEE13040419GA 00840682109260SEE13040419GA SEE13040420GA 00840682109260SEE13040420GA SEE13040426GA 00840682109260SEE13040426GA SEE13040427GA 00840682109260SEE13040427GA SEE13040428GA 00840682109260SEE13040428GA SEE13050430GA 00840682109260SEE13050430GA SEE13050431GA 00840682109260SEE13050431GA SEE13050432GA 00840682109260SEE13050432GA SEE13050433GA 00840682109260SEE13050433GA SEE13050434GA 00840682109260SEE13050434GA SEE13050435GA 00840682109260SEE13050435GA SEE13050436GA 00840682109260SEE13050436GA SEE13050437GA 00840682109260SEE13050437GA SEE13050438GA 00840682109260SEE13050438GA SEE13050439GA 00840682109260SEE13050439GA SEE13050441GA 00840682109260SEE13050441GA SEE13050442GA 00840682109260SEE13050442GA SEE13050443GA 00840682109260SEE13050443GA SEE13060444GA 00840682109260SEE13060444GA SEE13060445GA 00840682109260SEE13060445GA SEE13060447GA 00840682109260SEE13060447GA SEE13060448GA 00840682109260SEE13060448GA SEE13060449GA 00840682109260SEE13060449GA SEE13060450GA 00840682109260SEE13060450GA SEE13060451GA 00840682109260SEE13060451GA SEE13060452GA 00840682109260SEE13060452GA SEE13060453GA 00840682109260SEE13060453GA SEE13060454GA 00840682109260SEE13060454GA SEE13100477GA 00840682109260SEE13100477GA SEE13100480GA 00840682109260SEE13100480GA SEE13100483GA 00840682109260SEE13100483GA SEE13100484GA 00840682109260SEE13100484GA SEE13100489GA 00840682109260SEE13100489GA SEE13100490GA 00840682109260SEE13100490GA SEE13110493GA 00840682109260SEE13110493GA SEE13110494GA 00840682109260SEE13110494GA SEE13110495GA 00840682109260SEE13110495GA SEE13110496GA 00840682109260SEE13110496GA SEE13110497GA 00840682109260SEE13110497GA SEE13110498GA 00840682109260SEE13110498GA SEE13110499GA 00840682109260SEE13110499GA SEE13110500GA 00840682109260SEE13110500GA SEE13110501GA 00840682109260SEE13110501GA SEE13110502GA 00840682109260SEE13110502GA SEE13110503GA 00840682109260SEE13110503GA SEE13110504GA 00840682109260SEE13110504GA SEE13110506GA 00840682109260SEE13110506GA SEE13110507GA 00840682109260SEE13110507GA SEE13110508GA 00840682109260SEE13110508GA SEE13110509GA 00840682109260SEE13110509GA SEE13110510GA 00840682109260SEE13110510GA SEE13110511GA 00840682109260SEE13110511GA SEE13110512GA 00840682109260SEE13110512GA SEE13110513GA 00840682109260SEE13110513GA SEE13110514GA 00840682109260SEE13110514GA SEE13110515GA 00840682109260SEE13110515GA SEE13110516GA 00840682109260SEE13110516GA SEE13110517GA 00840682109260SEE13110517GA SEE13110518GA 00840682109260SEE13110518GA SEE13110519GA 00840682109260SEE13110519GA SEE13110520GA 00840682109260SEE13110520GA SEE13110521GA 00840682109260SEE13110521GA SEE13110522GA 00840682109260SEE13110522GA SEE13110523GA 00840682109260SEE13110523GA SEE13110524GA 00840682109260SEE13110524GA SEE13110525GA 00840682109260SEE13110525GA SEE13110526GA 00840682109260SEE13110526GA SEE13110527GA 00840682109260SEE13110527GA SEE13110528GA 00840682109260SEE13110528GA SEE13110529GA 00840682109260SEE13110529GA SEE13110530GA 00840682109260SEE13110530GA SEE13110531GA 00840682109260SEE13110531GA SEE13110532GA 00840682109260SEE13110532GA SEE13110533GA 00840682109260SEE13110533GA SEE13110534GA 00840682109260SEE13110534GA SEE13110535GA 00840682109260SEE13110535GA SEE13110536GA 00840682109260SEE13110536GA SEE13110537GA 00840682109260SEE13110537GA SEE13110538GA 00840682109260SEE13110538GA SEE13110539GA 00840682109260SEE13110539GA SEE13110540GA 00840682109260SEE13110540GA SEE13110541GA 00840682109260SEE13110541GA SEE13110542GA 00840682109260SEE13110542GA SEE13110543GA 00840682109260SEE13110543GA SEE13110544GA 00840682109260SEE13110544GA SEE13110545GA 00840682109260SEE13110545GA SEE13110546GA 00840682109260SEE13110546GA SEE13110547GA 00840682109260SEE13110547GA SEE13110549GA 00840682109260SEE13110549GA SEE13110550GA 00840682109260SEE13110550GA SEE13110551GA 00840682109260SEE13110551GA SEE13110552GA 00840682109260SEE13110552GA SEE13110553GA 00840682109260SEE13110553GA SEE13110554GA 00840682109260SEE13110554GA SEE13110555GA 00840682109260SEE13110555GA SEE13110556GA 00840682109260SEE13110556GA SEE13110557GA 00840682109260SEE13110557GA SEE13110558GA 00840682109260SEE13110558GA SEE13110559GA 00840682109260SEE13110559GA SEE13110560GA 00840682109260SEE13110560GA SEE13110561GA 00840682109260SEE13110561GA SEE13110562GA 00840682109260SEE13110562GA SEE13110563GA 00840682109260SEE13110563GA SEE13110564GA 00840682109260SEE13110564GA SEE13110565GA 00840682109260SEE13110565GA SEE13110566GA 00840682109260SEE13110566GA SEE13120572GA 00840682109260SEE13120572GA SEE13120573GA 00840682109260SEE13120573GA SEE13120574GA 00840682109260SEE13120574GA SEE13120576GA 00840682109260SEE13120576GA SEE13120577GA 00840682109260SEE13120577GA SEE13120578GA 00840682109260SEE13120578GA SEE13120579GA 00840682109260SEE13120579GA SEE13120580GA 00840682109260SEE13120580GA SEE13120581GA 00840682109260SEE13120581GA SEE13120582GA 00840682109260SEE13120582GA SEE13120583GA 00840682109260SEE13120583GA SEE13120584GA 00840682109260SEE13120584GA SEE13120586GA 00840682109260SEE13120586GA SEE13120587GA 00840682109260SEE13120587GA SEE13120588GA 00840682109260SEE13120588GA SEE13120589GA 00840682109260SEE13120589GA SEE13120590GA 00840682109260SEE13120590GA SEE13120591GA 00840682109260SEE13120591GA SEE13120593GA 00840682109260SEE13120593GA SEE13120594GA 00840682109260SEE13120594GA SEE13120595GA 00840682109260SEE13120595GA SEE13120597GA 00840682109260SEE13120597GA SEE13120599GA 00840682109260SEE13120599GA SEE13120600GA 00840682109260SEE13120600GA SEE13120602GA 00840682109260SEE13120602GA SEE13120603GA 00840682109260SEE13120603GA SEE13120604GA 00840682109260SEE13120604GA SEE13120606GA 00840682109260SEE13120606GA SEE13120607GA 00840682109260SEE13120607GA SEE13120608GA 00840682109260SEE13120608GA SEE13120609GA 00840682109260SEE13120609GA SEE13120610GA 00840682109260SEE13120610GA SEE13120611GA 00840682109260SEE13120611GA SEE13120613GA 00840682109260SEE13120613GA SEE13120614GA 00840682109260SEE13120614GA SEE13130615GA 00840682109260SEE13130615GA SEE13130616GA 00840682109260SEE13130616GA SEE13130617GA 00840682109260SEE13130617GA SEE13130619GA 00840682109260SEE13130619GA SEE13130620GA 00840682109260SEE13130620GA SEE13130621GA 00840682109260SEE13130621GA SEE13130622GA 00840682109260SEE13130622GA SEE13130623GA 00840682109260SEE13130623GA SEE13130624GA 00840682109260SEE13130624GA SEE13130625GA 00840682109260SEE13130625GA SEE13130626GA 00840682109260SEE13130626GA SEE13130627GA 00840682109260SEE13130627GA SEE13130628GA 00840682109260SEE13130628GA SEE13130629GA 00840682109260SEE13130629GA SEE13130630GA 00840682109260SEE13130630GA SEE13130632GA 00840682109260SEE13130632GA SEE13130633GA 00840682109260SEE13130633GA SEE13130635GA 00840682109260SEE13130635GA SEE13130636GA 00840682109260SEE13130636GA SEE13130638GA 00840682109260SEE13130638GA SEE13130639GA 00840682109260SEE13130639GA SEE13130640GA 00840682109260SEE13130640GA SEE13130641GA 00840682109260SEE13130641GA SEE13130642GA 00840682109260SEE13130642GA SEE13130643GA 00840682109260SEE13130643GA SEE13130644GA 00840682109260SEE13130644GA SEE13130645GA 00840682109260SEE13130645GA SEE13130647GA 00840682109260SEE13130647GA SEE13130648GA 00840682109260SEE13130648GA SEE13130649GA 00840682109260SEE13130649GA SEE13140651GA 00840682109260SEE13140651GA SEE13140652GA 00840682109260SEE13140652GA SEE13140653GA 00840682109260SEE13140653GA SEE13140654GA 00840682109260SEE13140654GA SEE13140656GA 00840682109260SEE13140656GA SEE13140657GA 00840682109260SEE13140657GA SEE13140658GA 00840682109260SEE13140658GA SEE13150659GA 00840682109260SEE13150659GA SEE13150660GA 00840682109260SEE13150660GA SEE13150661GA 00840682109260SEE13150661GA SEE13150662GA 00840682109260SEE13150662GA SEE13150663GA 00840682109260SEE13150663GA SEE13150664GA 00840682109260SEE13150664GA SEE13150666GA 00840682109260SEE13150666GA SEE13150667GA 00840682109260SEE13150667GA SEE13150668GA 00840682109260SEE13150668GA SEE13150669GA 00840682109260SEE13150669GA SEE13150670GA 00840682109260SEE13150670GA SEE13150671GA 00840682109260SEE13150671GA SEE13150674GA 00840682109260SEE13150674GA SEE13150682GA 00840682109260SEE13150682GA SEE13150683GA 00840682109260SEE13150683GA SEE13150684GA 00840682109260SEE13150684GA SEE13150685GA 00840682109260SEE13150685GA SEE13150686GA 00840682109260SEE13150686GA SEE13150689GA 00840682109260SEE13150689GA SEE13150690GA 00840682109260SEE13150690GA SEE13150693GA 00840682109260SEE13150693GA SEE13150694GA 00840682109260SEE13150694GA SEE13150695GA 00840682109260SEE13150695GA SEE13150696GA 00840682109260SEE13150696GA SEE13150697GA 00840682109260SEE13150697GA SEE13150698GA 00840682109260SEE13150698GA SEE13150699GA 00840682109260SEE13150699GA SEE13150700GA 00840682109260SEE13150700GA SEE13160711GA 00840682109260SEE13160711GA SEE13160712GA 00840682109260SEE13160712GA SEE13160713GA 00840682109260SEE13160713GA SEE13160714GA 00840682109260SEE13160714GA SEE13160715GA 00840682109260SEE13160715GA SEE13160716GA 00840682109260SEE13160716GA SEE13160717GA 00840682109260SEE13160717GA SEE13160718GA 00840682109260SEE13160718GA SEE13160719GA 00840682109260SEE13160719GA SEE13160720GA 00840682109260SEE13160720GA SEE13180722GA 00840682109260SEE13180722GA SEE13180724GA 00840682109260SEE13180724GA SEE13180728GA 00840682109260SEE13180728GA SEE13180729GA 00840682109260SEE13180729GA SEE13180733GA 00840682109260SEE13180733GA SEE13180735GA 00840682109260SEE13180735GA SEE13180739GA 00840682109260SEE13180739GA SEE13180741GA 00840682109260SEE13180741GA SEE13180742GA 00840682109260SEE13180742GA SEE13180743GA 00840682109260SEE13180743GA SEE13180744GA 00840682109260SEE13180744GA SEE13180745GA 00840682109260SEE13180745GA SEE13180746GA 00840682109260SEE13180746GA SEE13180747GA 00840682109260SEE13180747GA SEE13180748GA 00840682109260SEE13180748GA SEE13180749GA 00840682109260SEE13180749GA SEE13180750GA 00840682109260SEE13180750GA SEE13180751GA 00840682109260SEE13180751GA SEE13180752GA 00840682109260SEE13180752GA SEE13180753GA 00840682109260SEE13180753GA SEE13180754GA 00840682109260SEE13180754GA SEE13180755GA 00840682109260SEE13180755GA SEE13180756GA 00840682109260SEE13180756GA SEE13180757GA 00840682109260SEE13180757GA SEE13180758GA 00840682109260SEE13180758GA SEE13180759GA 00840682109260SEE13180759GA SEE13180760GA 00840682109260SEE13180760GA SEE13180761GA 00840682109260SEE13180761GA SEE13180762GA 00840682109260SEE13180762GA SEE13180763GA 00840682109260SEE13180763GA SEE13180764GA 00840682109260SEE13180764GA SEE13180765GA 00840682109260SEE13180765GA SEE13180767GA 00840682109260SEE13180767GA SEE13180769GA 00840682109260SEE13180769GA SEE13180770GA 00840682109260SEE13180770GA SEE13180771GA 00840682109260SEE13180771GA SEE13180772GA 00840682109260SEE13180772GA SEE13180773GA 00840682109260SEE13180773GA SEE13180774GA 00840682109260SEE13180774GA SEE13180776GA 00840682109260SEE13180776GA SEE13180777GA 00840682109260SEE13180777GA SEE13180779GA 00840682109260SEE13180779GA SEE13180782GA 00840682109260SEE13180782GA SEE13180783GA 00840682109260SEE13180783GA SEE13180784GA 00840682109260SEE13180784GA SEE13180785GA 00840682109260SEE13180785GA SEE13180786GA 00840682109260SEE13180786GA SEE13180787GA 00840682109260SEE13180787GA SEE13180788GA 00840682109260SEE13180788GA SEE13190792GA 00840682109260SEE13190792GA SEE13190793GA 00840682109260SEE13190793GA SEE13190794GA 00840682109260SEE13190794GA SEE13190795GA 00840682109260SEE13190795GA SEE13190796GA 00840682109260SEE13190796GA SEE13190797GA 00840682109260SEE13190797GA SEE13190798GA 00840682109260SEE13190798GA SEE13190799GA 00840682109260SEE13190799GA SEE13190800GA 00840682109260SEE13190800GA SEE13190801GA 00840682109260SEE13190801GA SEE13190802GA 00840682109260SEE13190802GA SEE13190803GA 00840682109260SEE13190803GA SEE13190804GA 00840682109260SEE13190804GA SEE13190805GA 00840682109260SEE13190805GA SEE13190806GA 00840682109260SEE13190806GA SEE13190808GA 00840682109260SEE13190808GA SEE13190809GA 00840682109260SEE13190809GA SEE13190811GA 00840682109260SEE13190811GA SEE13220822GA 00840682109260SEE13220822GA SEE13220823GA 00840682109260SEE13220823GA SEE13220824GA 00840682109260SEE13220824GA SEE13220825GA 00840682109260SEE13220825GA SEE13220826GA 00840682109260SEE13220826GA SEE13220827GA 00840682109260SEE13220827GA SEE13220828GA 00840682109260SEE13220828GA SEE13220829GA 00840682109260SEE13220829GA SEE13220830GA 00840682109260SEE13220830GA SEE13220831GA 00840682109260SEE13220831GA SEE13220832GA 00840682109260SEE13220832GA SEE13220833GA 00840682109260SEE13220833GA SEE13220834GA 00840682109260SEE13220834GA SEE13220835GA 00840682109260SEE13220835GA SEE13220836GA 00840682109260SEE13220836GA SEE13220837GA 00840682109260SEE13220837GA SEE13220839GA 00840682109260SEE13220839GA SEE13220842GA 00840682109260SEE13220842GA SEE13220843GA 00840682109260SEE13220843GA SEE13220844GA 00840682109260SEE13220844GA SEE13220845GA 00840682109260SEE13220845GA SEE13220846GA 00840682109260SEE13220846GA SEE13220847GA 00840682109260SEE13220847GA SEE13220848GA 00840682109260SEE13220848GA SEE13220849GA 00840682109260SEE13220849GA SEE13220850GA 00840682109260SEE13220850GA SEE13220851GA 00840682109260SEE13220851GA SEE13240852GA 00840682109260SEE13240852GA SEE13240853GA 00840682109260SEE13240853GA SEE13240854GA 00840682109260SEE13240854GA SEE13240855GA 00840682109260SEE13240855GA SEE13240857GA 00840682109260SEE13240857GA SEE13240858GA 00840682109260SEE13240858GA SEE13240859GA 00840682109260SEE13240859GA SEE13240860GA 00840682109260SEE13240860GA SEE13240861GA 00840682109260SEE13240861GA SEE13240862GA 00840682109260SEE13240862GA SEE13240864GA 00840682109260SEE13240864GA SEE13240865GA 00840682109260SEE13240865GA SEE13240866GA 00840682109260SEE13240866GA SEE13240867GA 00840682109260SEE13240867GA SEE13240868GA 00840682109260SEE13240868GA SEE13240869GA 00840682109260SEE13240869GA SEE13240870GA 00840682109260SEE13240870GA SEE13240871GA 00840682109260SEE13240871GA SEE13270872GA 00840682109260SEE13270872GA SEE13270873GA 00840682109260SEE13270873GA SEE13270874GA 00840682109260SEE13270874GA SEE13270875GA 00840682109260SEE13270875GA SEE13270876GA 00840682109260SEE13270876GA SEE13270877GA 00840682109260SEE13270877GA SEE13270878GA 00840682109260SEE13270878GA SEE13270879GA 00840682109260SEE13270879GA SEE13270880GA 00840682109260SEE13270880GA SEE13270881GA 00840682109260SEE13270881GA SEE13270882GA 00840682109260SEE13270882GA SEE13270884GA 00840682109260SEE13270884GA SEE13270885GA 00840682109260SEE13270885GA SEE13270886GA 00840682109260SEE13270886GA SEE13270887GA 00840682109260SEE13270887GA SEE13270888GA 00840682109260SEE13270888GA SEE13270891GA 00840682109260SEE13270891GA SEE13270892GA 00840682109260SEE13270892GA SEE13270893GA 00840682109260SEE13270893GA SEE13270894GA 00840682109260SEE13270894GA SEE13270895GA 00840682109260SEE13270895GA SEE13270896GA 00840682109260SEE13270896GA SEE13270897GA 00840682109260SEE13270897GA SEE13270899GA 00840682109260SEE13270899GA SEE13270900GA 00840682109260SEE13270900GA SEE13270901GA 00840682109260SEE13270901GA SEE13270902GA 00840682109260SEE13270902GA SEE13280903GA 00840682109260SEE13280903GA SEE13280904GA 00840682109260SEE13280904GA SEE13280905GA 00840682109260SEE13280905GA SEE13280906GA 00840682109260SEE13280906GA SEE13280907GA 00840682109260SEE13280907GA SEE13280908GA 00840682109260SEE13280908GA SEE13280909GA 00840682109260SEE13280909GA SEE13280910GA 00840682109260SEE13280910GA SEE13280911GA 00840682109260SEE13280911GA SEE13280913GA 00840682109260SEE13280913GA SEE13280914GA 00840682109260SEE13280914GA SEE13280915GA 00840682109260SEE13280915GA SEE13280916GA 00840682109260SEE13280916GA SEE13280918GA 00840682109260SEE13280918GA SEE13280920GA 00840682109260SEE13280920GA SEE13280922GA 00840682109260SEE13280922GA SEE13280923GA 00840682109260SEE13280923GA SEE13280924GA 00840682109260SEE13280924GA SEE13280925GA 00840682109260SEE13280925GA SEE13280926GA 00840682109260SEE13280926GA SEE13280927GA 00840682109260SEE13280927GA SEE13280928GA 00840682109260SEE13280928GA SEE13280929GA 00840682109260SEE13280929GA SEE13280930GA 00840682109260SEE13280930GA SEE13280931GA 00840682109260SEE13280931GA SEE13280932GA 00840682109260SEE13280932GA SEE13280933GA 00840682109260SEE13280933GA SEE13280935GA 00840682109260SEE13280935GA SEE13280936GA 00840682109260SEE13280936GA SEE13280937GA 00840682109260SEE13280937GA SEE13280938GA 00840682109260SEE13280938GA SEE13280939GA 00840682109260SEE13280939GA SEE13280940GA 00840682109260SEE13280940GA SEE13280941GA 00840682109260SEE13280941GA SEE13280942GA 00840682109260SEE13280942GA SEE13280943GA 00840682109260SEE13280943GA SEE13280944GA 00840682109260SEE13280944GA SEE13280946GA 00840682109260SEE13280946GA SEE13280948GA 00840682109260SEE13280948GA SEE13280949GA 00840682109260SEE13280949GA SEE13280950GA 00840682109260SEE13280950GA SEE13290951GA 00840682109260SEE13290951GA SEE13290952GA 00840682109260SEE13290952GA SEE13290953GA 00840682109260SEE13290953GA SEE13290954GA 00840682109260SEE13290954GA SEE13290955GA 00840682109260SEE13290955GA SEE13290956GA 00840682109260SEE13290956GA SEE13290957GA 00840682109260SEE13290957GA SEE13290958GA 00840682109260SEE13290958GA SEE13290959GA 00840682109260SEE13290959GA SEE13290960GA 00840682109260SEE13290960GA SEE13290961GA 00840682109260SEE13290961GA SEE13290962GA 00840682109260SEE13290962GA SEE13290963GA 00840682109260SEE13290963GA SEE13290964GA 00840682109260SEE13290964GA SEE13290965GA 00840682109260SEE13290965GA SEE13290966GA 00840682109260SEE13290966GA SEE13290968GA 00840682109260SEE13290968GA SEE13290969GA 00840682109260SEE13290969GA SEE13290970GA 00840682109260SEE13290970GA SEE13290971GA 00840682109260SEE13290971GA SEE13290972GA 00840682109260SEE13290972GA SEE13290973GA 00840682109260SEE13290973GA SEE13290974GA 00840682109260SEE13290974GA SEE13290975GA 00840682109260SEE13290975GA SEE13290976GA 00840682109260SEE13290976GA SEE13290977GA 00840682109260SEE13290977GA SEE13290978GA 00840682109260SEE13290978GA SEE13290979GA 00840682109260SEE13290979GA SEE13290980GA 00840682109260SEE13290980GA SEE13290981GA 00840682109260SEE13290981GA SEE13290982GA 00840682109260SEE13290982GA SEE13290984GA 00840682109260SEE13290984GA SEE13290985GA 00840682109260SEE13290985GA SEE13290987GA 00840682109260SEE13290987GA SEE13290988GA 00840682109260SEE13290988GA SEE13290989GA 00840682109260SEE13290989GA SEE13290990GA 00840682109260SEE13290990GA SEE13300991GA 00840682109260SEE13300991GA SEE13311001GA 00840682109260SEE13311001GA SEE13311002GA 00840682109260SEE13311002GA SEE13311003GA 00840682109260SEE13311003GA SEE13311004GA 00840682109260SEE13311004GA SEE13311005GA 00840682109260SEE13311005GA SEE13311006GA 00840682109260SEE13311006GA SEE13311007GA 00840682109260SEE13311007GA SEE13311008GA 00840682109260SEE13311008GA SEE13311009GA 00840682109260SEE13311009GA SEE13351010GA 00840682109260SEE13351010GA SEE13351011GA 00840682109260SEE13351011GA SEE13351012GA 00840682109260SEE13351012GA SEE13351013GA 00840682109260SEE13351013GA SEE13351014GA 00840682109260SEE13351014GA SEE13351015GA 00840682109260SEE13351015GA SEE13351016GA 00840682109260SEE13351016GA SEE13351017GA 00840682109260SEE13351017GA SEE13351019GA 00840682109260SEE13351019GA SEE13351020GA 00840682109260SEE13351020GA SEE13351021GA 00840682109260SEE13351021GA SEE13351022GA 00840682109260SEE13351022GA SEE13351023GA 00840682109260SEE13351023GA SEE13351024GA 00840682109260SEE13351024GA SEE13351025GA 00840682109260SEE13351025GA SEE13351026GA 00840682109260SEE13351026GA SEE13351027GA 00840682109260SEE13351027GA SEE13351028GA 00840682109260SEE13351028GA SEE13351029GA 00840682109260SEE13351029GA SEE13351030GA 00840682109260SEE13351030GA SEE13351032GA 00840682109260SEE13351032GA SEE13351033GA 00840682109260SEE13351033GA SEE13351034GA 00840682109260SEE13351034GA SEE13351036GA 00840682109260SEE13351036GA SEE13351037GA 00840682109260SEE13351037GA SEE13351038GA 00840682109260SEE13351038GA SEE13351039GA 00840682109260SEE13351039GA SEE13351041GA 00840682109260SEE13351041GA SEE13351042GA 00840682109260SEE13351042GA SEE13351044GA 00840682109260SEE13351044GA SEE13351046GA 00840682109260SEE13351046GA SEE13351047GA 00840682109260SEE13351047GA SEE13351049GA 00840682109260SEE13351049GA SEE13351050GA 00840682109260SEE13351050GA SEE13351051GA 00840682109260SEE13351051GA SEE13351052GA 00840682109260SEE13351052GA SEE13351053GA 00840682109260SEE13351053GA SEE13361055GA 00840682109260SEE13361055GA SEE13361056GA 00840682109260SEE13361056GA SEE13361057GA 00840682109260SEE13361057GA SEE13361058GA 00840682109260SEE13361058GA SEE13361059GA 00840682109260SEE13361059GA SEE13361060GA 00840682109260SEE13361060GA SEE13361061GA 00840682109260SEE13361061GA SEE13361063GA 00840682109260SEE13361063GA SEE13361064GA 00840682109260SEE13361064GA SEE13361065GA 00840682109260SEE13361065GA SEE13361066GA 00840682109260SEE13361066GA SEE13361068GA 00840682109260SEE13361068GA SEE13361069GA 00840682109260SEE13361069GA SEE13361070GA 00840682109260SEE13361070GA SEE13361071GA 00840682109260SEE13361071GA SEE13361072GA 00840682109260SEE13361072GA SEE13361073GA 00840682109260SEE13361073GA SEE13361074GA 00840682109260SEE13361074GA SEE13361075GA 00840682109260SEE13361075GA SEE13361077GA 00840682109260SEE13361077GA SEE13361078GA 00840682109260SEE13361078GA SEE13361080GA 00840682109260SEE13361080GA SEE13371081GA 00840682109260SEE13371081GA SEE13371082GA 00840682109260SEE13371082GA SEE13371083GA 00840682109260SEE13371083GA SEE13371084GA 00840682109260SEE13371084GA SEE13371085GA 00840682109260SEE13371085GA SEE13371086GA 00840682109260SEE13371086GA SEE13371087GA 00840682109260SEE13371087GA SEE13371088GA 00840682109260SEE13371088GA SEE13371090GA 00840682109260SEE13371090GA SEE13371091GA 00840682109260SEE13371091GA SEE13371092GA 00840682109260SEE13371092GA SEE13371093GA 00840682109260SEE13371093GA SEE13371094GA 00840682109260SEE13371094GA SEE13371095GA 00840682109260SEE13371095GA SEE13371097GA 00840682109260SEE13371097GA SEE13371098GA 00840682109260SEE13371098GA SEE13371099GA 00840682109260SEE13371099GA SEE13371100GA 00840682109260SEE13371100GA SEE13371101GA 00840682109260SEE13371101GA SEE13371102GA 00840682109260SEE13371102GA SEE13371103GA 00840682109260SEE13371103GA SEE13371104GA 00840682109260SEE13371104GA SEE13371105GA 00840682109260SEE13371105GA SEE13371106GA 00840682109260SEE13371106GA SEE13371107GA 00840682109260SEE13371107GA SEE13371108GA 00840682109260SEE13371108GA SEE13371110GA 00840682109260SEE13371110GA SEE13371111GA 00840682109260SEE13371111GA SEE13371113GA 00840682109260SEE13371113GA SEE13371114GA 00840682109260SEE13371114GA SEE13371116GA 00840682109260SEE13371116GA SEE13371117GA 00840682109260SEE13371117GA SEE13371118GA 00840682109260SEE13371118GA SEE13371120GA 00840682109260SEE13371120GA SEE13381121GA 00840682109260SEE13381121GA SEE13381122GA 00840682109260SEE13381122GA SEE13381123GA 00840682109260SEE13381123GA SEE13381124GA 00840682109260SEE13381124GA SEE13381125GA 00840682109260SEE13381125GA SEE13381126GA 00840682109260SEE13381126GA SEE13381127GA 00840682109260SEE13381127GA SEE13381128GA 00840682109260SEE13381128GA SEE13381129GA 00840682109260SEE13381129GA SEE13391130GA 00840682109260SEE13391130GA SEE13391131GA 00840682109260SEE13391131GA SEE13391132GA 00840682109260SEE13391132GA SEE13391133GA 00840682109260SEE13391133GA SEE13391134GA 00840682109260SEE13391134GA SEE13391135GA 00840682109260SEE13391135GA SEE13391136GA 00840682109260SEE13391136GA SEE13391137GA 00840682109260SEE13391137GA SEE13391138GA 00840682109260SEE13391138GA SEE13391139GA 00840682109260SEE13391139GA SEE13391140GA 00840682109260SEE13391140GA SEE13401141GA 00840682109260SEE13401141GA SEE13401142GA 00840682109260SEE13401142GA SEE13401143GA 00840682109260SEE13401143GA SEE13401144GA 00840682109260SEE13401144GA SEE13401145GA 00840682109260SEE13401145GA SEE13401150GA 00840682109260SEE13401150GA SEE13401152GA 00840682109260SEE13401152GA SEE13401153GA 00840682109260SEE13401153GA SEE13401154GA 00840682109260SEE13401154GA SEE13401155GA 00840682109260SEE13401155GA SEE13401156GA 00840682109260SEE13401156GA SEE13401157GA 00840682109260SEE13401157GA SEE13401158GA 00840682109260SEE13401158GA SEE13401159GA 00840682109260SEE13401159GA SEE13401160GA 00840682109260SEE13401160GA SEE13401161GA 00840682109260SEE13401161GA SEE13401162GA 00840682109260SEE13401162GA SEE13401164GA 00840682109260SEE13401164GA SEE13401165GA 00840682109260SEE13401165GA SEE13401166GA 00840682109260SEE13401166GA SEE13401167GA 00840682109260SEE13401167GA SEE13401168GA 00840682109260SEE13401168GA SEE13401169GA 00840682109260SEE13401169GA SEE13401170GA 00840682109260SEE13401170GA SEE13401171GA 00840682109260SEE13401171GA SEE13401174GA 00840682109260SEE13401174GA SEE13401175GA 00840682109260SEE13401175GA SEE13401176GA 00840682109260SEE13401176GA SEE13401177GA 00840682109260SEE13401177GA SEE13401179GA 00840682109260SEE13401179GA SEE13401180GA 00840682109260SEE13401180GA SEE13401181GA 00840682109260SEE13401181GA SEE13401182GA 00840682109260SEE13401182GA SEE13401183GA 00840682109260SEE13401183GA SEE13401185GA 00840682109260SEE13401185GA SEE13411189GA 00840682109260SEE13411189GA SEE13411190GA 00840682109260SEE13411190GA SEE13411191GA 00840682109260SEE13411191GA SEE13411192GA 00840682109260SEE13411192GA SEE13411193GA 00840682109260SEE13411193GA SEE13411194GA 00840682109260SEE13411194GA SEE13411195GA 00840682109260SEE13411195GA SEE13411196GA 00840682109260SEE13411196GA SEE13411197GA 00840682109260SEE13411197GA SEE13411198GA 00840682109260SEE13411198GA SEE13411199GA 00840682109260SEE13411199GA SEE13411200GA 00840682109260SEE13411200GA SEE13411201GA 00840682109260SEE13411201GA SEE13411204GA 00840682109260SEE13411204GA SEE13411205GA 00840682109260SEE13411205GA SEE13411206GA 00840682109260SEE13411206GA SEE13411207GA 00840682109260SEE13411207GA SEE13411208GA 00840682109260SEE13411208GA SEE13411210GA 00840682109260SEE13411210GA SEE13411212GA 00840682109260SEE13411212GA SEE13411213GA 00840682109260SEE13411213GA SEE13411214GA 00840682109260SEE13411214GA SEE13411215GA 00840682109260SEE13411215GA SEE13411216GA 00840682109260SEE13411216GA SEE13411217GA 00840682109260SEE13411217GA SEE13411218GA 00840682109260SEE13411218GA SEE13411219GA 00840682109260SEE13411219GA SEE13411220GA 00840682109260SEE13411220GA SEE13411221GA 00840682109260SEE13411221GA SEE13411222GA 00840682109260SEE13411222GA SEE13411223GA 00840682109260SEE13411223GA SEE13421224GA 00840682109260SEE13421224GA SEE13421225GA 00840682109260SEE13421225GA SEE13421226GA 00840682109260SEE13421226GA SEE13421227GA 00840682109260SEE13421227GA SEE13421228GA 00840682109260SEE13421228GA SEE13421229GA 00840682109260SEE13421229GA SEE13421230GA 00840682109260SEE13421230GA SEE13421231GA 00840682109260SEE13421231GA SEE13421232GA 00840682109260SEE13421232GA SEE13421233GA 00840682109260SEE13421233GA SEE13421234GA 00840682109260SEE13421234GA SEE13421235GA 00840682109260SEE13421235GA SEE13421236GA 00840682109260SEE13421236GA SEE13421237GA 00840682109260SEE13421237GA SEE13421238GA 00840682109260SEE13421238GA SEE13421239GA 00840682109260SEE13421239GA SEE13421240GA 00840682109260SEE13421240GA SEE13421241GA 00840682109260SEE13421241GA SEE13421242GA 00840682109260SEE13421242GA SEE13421243GA 00840682109260SEE13421243GA SEE13421244GA 00840682109260SEE13421244GA SEE13421245GA 00840682109260SEE13421245GA SEE13421246GA 00840682109260SEE13421246GA SEE13421247GA 00840682109260SEE13421247GA SEE13421248GA 00840682109260SEE13421248GA SEE13421249GA 00840682109260SEE13421249GA SEE13421250GA 00840682109260SEE13421250GA SEE13421251GA 00840682109260SEE13421251GA SEE13421252GA 00840682109260SEE13421252GA SEE13421253GA 00840682109260SEE13421253GA SEE13431257GA 00840682109260SEE13431257GA SEE13431258GA 00840682109260SEE13431258GA SEE13431259GA 00840682109260SEE13431259GA SEE13431260GA 00840682109260SEE13431260GA SEE13431261GA 00840682109260SEE13431261GA SEE13431262GA 00840682109260SEE13431262GA SEE13431263GA 00840682109260SEE13431263GA SEE13431264GA 00840682109260SEE13431264GA SEE13431265GA 00840682109260SEE13431265GA SEE13431266GA 00840682109260SEE13431266GA SEE13431267GA 00840682109260SEE13431267GA SEE13431268GA 00840682109260SEE13431268GA SEE13431269GA 00840682109260SEE13431269GA SEE13431270GA 00840682109260SEE13431270GA SEE13431271GA 00840682109260SEE13431271GA SEE13431273GA 00840682109260SEE13431273GA SEE13431274GA 00840682109260SEE13431274GA SEE13431275GA 00840682109260SEE13431275GA SEE13431276GA 00840682109260SEE13431276GA SEE13431277GA 00840682109260SEE13431277GA SEE13431278GA 00840682109260SEE13431278GA SEE13431279GA 00840682109260SEE13431279GA SEE13431280GA 00840682109260SEE13431280GA SEE13431281GA 00840682109260SEE13431281GA SEE13431282GA 00840682109260SEE13431282GA SEE13431284GA 00840682109260SEE13431284GA SEE13431285GA 00840682109260SEE13431285GA SEE13431286GA 00840682109260SEE13431286GA SEE13431288GA 00840682109260SEE13431288GA SEE13431289GA 00840682109260SEE13431289GA SEE13431290GA 00840682109260SEE13431290GA SEE13431291GA 00840682109260SEE13431291GA SEE13431292GA 00840682109260SEE13431292GA SEE13431294GA 00840682109260SEE13431294GA SEE13431295GA 00840682109260SEE13431295GA SEE13431297GA 00840682109260SEE13431297GA SEE13431299GA 00840682109260SEE13431299GA SEE13431301GA 00840682109260SEE13431301GA SEE13431302GA 00840682109260SEE13431302GA SEE13431303GA 00840682109260SEE13431303GA SEE13431304GA 00840682109260SEE13431304GA SEE13441305GA 00840682109260SEE13441305GA SEE13441306GA 00840682109260SEE13441306GA SEE13441307GA 00840682109260SEE13441307GA SEE13451308GA 00840682109260SEE13451308GA SEE13451309GA 00840682109260SEE13451309GA SEE13451310GA 00840682109260SEE13451310GA SEE13451312GA 00840682109260SEE13451312GA SEE13451313GA 00840682109260SEE13451313GA SEE13451315GA 00840682109260SEE13451315GA SEE13451316GA 00840682109260SEE13451316GA SEE13451317GA 00840682109260SEE13451317GA SEE13451318GA 00840682109260SEE13451318GA SEE13451319GA 00840682109260SEE13451319GA SEE13451320GA 00840682109260SEE13451320GA SEE13451322GA 00840682109260SEE13451322GA SEE13451325GA 00840682109260SEE13451325GA SEE13451326GA 00840682109260SEE13451326GA SEE13451327GA 00840682109260SEE13451327GA SEE13451328GA 00840682109260SEE13451328GA SEE13451329GA 00840682109260SEE13451329GA SEE13451330GA 00840682109260SEE13451330GA SEE13451331GA 00840682109260SEE13451331GA SEE13451332GA 00840682109260SEE13451332GA SEE13451333GA 00840682109260SEE13451333GA SEE13451335GA 00840682109260SEE13451335GA SEE13451336GA 00840682109260SEE13451336GA SEE13451337GA 00840682109260SEE13451337GA SEE13451338GA 00840682109260SEE13451338GA SEE13451339GA 00840682109260SEE13451339GA SEE13451340GA 00840682109260SEE13451340GA SEE13451341GA 00840682109260SEE13451341GA SEE13451343GA 00840682109260SEE13451343GA SEE13451344GA 00840682109260SEE13451344GA SEE13451345GA 00840682109260SEE13451345GA SEE13451346GA 00840682109260SEE13451346GA SEE13451347GA 00840682109260SEE13451347GA SEE13451349GA 00840682109260SEE13451349GA SEE13451350GA 00840682109260SEE13451350GA SEE13451351GA 00840682109260SEE13451351GA SEE13451352GA 00840682109260SEE13451352GA SEE13451354GA 00840682109260SEE13451354GA SEE13451355GA 00840682109260SEE13451355GA SEE13451356GA 00840682109260SEE13451356GA SEE13451357GA 00840682109260SEE13451357GA SEE13451358GA 00840682109260SEE13451358GA SEE13451359GA 00840682109260SEE13451359GA SEE13451360GA 00840682109260SEE13451360GA SEE13451361GA 00840682109260SEE13451361GA SEE13451362GA 00840682109260SEE13451362GA SEE13451363GA 00840682109260SEE13451363GA SEE13451364GA 00840682109260SEE13451364GA SEE13451365GA 00840682109260SEE13451365GA SEE13451366GA 00840682109260SEE13451366GA SEE13451367GA 00840682109260SEE13451367GA SEE13451368GA 00840682109260SEE13451368GA SEE13451375GA 00840682109260SEE13451375GA SEE13451376GA 00840682109260SEE13451376GA SEE13451817GA 00840682109260SEE13451817GA SEE13481388GA 00840682109260SEE13481388GA SEE13481389GA 00840682109260SEE13481389GA SEE13481390GA 00840682109260SEE13481390GA SEE13481391GA 00840682109260SEE13481391GA SEE13481392GA 00840682109260SEE13481392GA SEE13481393GA 00840682109260SEE13481393GA SEE13481394GA 00840682109260SEE13481394GA SEE13481395GA 00840682109260SEE13481395GA SEE13481396GA 00840682109260SEE13481396GA SEE13481397GA 00840682109260SEE13481397GA SEE13481398GA 00840682109260SEE13481398GA SEE13481399GA 00840682109260SEE13481399GA SEE13481403GA 00840682109260SEE13481403GA SEE13481406GA 00840682109260SEE13481406GA SEE13481408GA 00840682109260SEE13481408GA SEE13481409GA 00840682109260SEE13481409GA SEE13481410GA 00840682109260SEE13481410GA SEE13481411GA 00840682109260SEE13481411GA SEE13481412GA 00840682109260SEE13481412GA SEE13481413GA 00840682109260SEE13481413GA SEE13481414GA 00840682109260SEE13481414GA SEE13481415GA 00840682109260SEE13481415GA SEE13481419GA 00840682109260SEE13481419GA SEE13481420GA 00840682109260SEE13481420GA SEE13491424GA 00840682109260SEE13491424GA SEE13491425GA 00840682109260SEE13491425GA SEE13491426GA 00840682109260SEE13491426GA SEE13491427GA 00840682109260SEE13491427GA SEE13491428GA 00840682109260SEE13491428GA SEE13491430GA 00840682109260SEE13491430GA SEE13491431GA 00840682109260SEE13491431GA SEE13491432GA 00840682109260SEE13491432GA SEE13491435GA 00840682109260SEE13491435GA SEE13491438GA 00840682109260SEE13491438GA SEE13491439GA 00840682109260SEE13491439GA SEE13491440GA 00840682109260SEE13491440GA SEE13491441GA 00840682109260SEE13491441GA SEE13491442GA 00840682109260SEE13491442GA SEE13491443GA 00840682109260SEE13491443GA SEE13491444GA 00840682109260SEE13491444GA SEE13491446GA 00840682109260SEE13491446GA SEE13491447GA 00840682109260SEE13491447GA SEE13491448GA 00840682109260SEE13491448GA SEE13491449GA 00840682109260SEE13491449GA SEE13491450GA 00840682109260SEE13491450GA SEE13491451GA 00840682109260SEE13491451GA SEE13501453GA 00840682109260SEE13501453GA SEE13501454GA 00840682109260SEE13501454GA SEE13501455GA 00840682109260SEE13501455GA SEE13501456GA 00840682109260SEE13501456GA SEE13501457GA 00840682109260SEE13501457GA SEE13501458GA 00840682109260SEE13501458GA SEE13501459GA 00840682109260SEE13501459GA SEE13501460GA 00840682109260SEE13501460GA SEE13501461GA 00840682109260SEE13501461GA SEE13501462GA 00840682109260SEE13501462GA SEE13501463GA 00840682109260SEE13501463GA SEE13501465GA 00840682109260SEE13501465GA SEE13501466GA 00840682109260SEE13501466GA SEE13501467GA 00840682109260SEE13501467GA SEE13501468GA 00840682109260SEE13501468GA SEE13501469GA 00840682109260SEE13501469GA SEE13501470GA 00840682109260SEE13501470GA SEE13501471GA 00840682109260SEE13501471GA SEE13501472GA 00840682109260SEE13501472GA SEE13501473GA 00840682109260SEE13501473GA SEE13501474GA 00840682109260SEE13501474GA SEE13501475GA 00840682109260SEE13501475GA SEE13501476GA 00840682109260SEE13501476GA SEE13501477GA 00840682109260SEE13501477GA SEE13501479GA 00840682109260SEE13501479GA SEE13501480GA 00840682109260SEE13501480GA SEE13501481GA 00840682109260SEE13501481GA SEE13501482GA 00840682109260SEE13501482GA SEE13501483GA 00840682109260SEE13501483GA SEE13501484GA 00840682109260SEE13501484GA SEE13501485GA 00840682109260SEE13501485GA SEE13501486GA 00840682109260SEE13501486GA SEE13501487GA 00840682109260SEE13501487GA SEE13501488GA 00840682109260SEE13501488GA SEE13501489GA 00840682109260SEE13501489GA SEE13501490GA 00840682109260SEE13501490GA SEE13501491GA 00840682109260SEE13501491GA SEE13501492GA 00840682109260SEE13501492GA SEE13501493GA 00840682109260SEE13501493GA SEE13501494GA 00840682109260SEE13501494GA SEE13501495GA 00840682109260SEE13501495GA SEE13501496GA 00840682109260SEE13501496GA SEE13501497GA 00840682109260SEE13501497GA SEE13501498GA 00840682109260SEE13501498GA SEE13501499GA 00840682109260SEE13501499GA SEE13501500GA 00840682109260SEE13501500GA SEE13501501GA 00840682109260SEE13501501GA SEE14021503GA 00840682109260SEE14021503GA SEE14021504GA 00840682109260SEE14021504GA SEE14021505GA 00840682109260SEE14021505GA SEE14021506GA 00840682109260SEE14021506GA SEE14021507GA 00840682109260SEE14021507GA SEE14021509GA 00840682109260SEE14021509GA SEE14021511GA 00840682109260SEE14021511GA SEE14021512GA 00840682109260SEE14021512GA SEE14021513GA 00840682109260SEE14021513GA SEE14021514GA 00840682109260SEE14021514GA SEE14021515GA 00840682109260SEE14021515GA SEE14021516GA 00840682109260SEE14021516GA SEE14021518GA 00840682109260SEE14021518GA SEE14021519GA 00840682109260SEE14021519GA SEE14021520GA 00840682109260SEE14021520GA SEE14021522GA 00840682109260SEE14021522GA SEE14021523GA 00840682109260SEE14021523GA SEE14021524GA 00840682109260SEE14021524GA SEE14021525GA 00840682109260SEE14021525GA SEE14021526GA 00840682109260SEE14021526GA SEE14021527GA 00840682109260SEE14021527GA SEE14021528GA 00840682109260SEE14021528GA SEE14021529GA 00840682109260SEE14021529GA SEE14021530GA 00840682109260SEE14021530GA SEE14021531GA 00840682109260SEE14021531GA SEE14021532GA 00840682109260SEE14021532GA SEE14021533GA 00840682109260SEE14021533GA SEE14021534GA 00840682109260SEE14021534GA SEE14021535GA 00840682109260SEE14021535GA SEE14021536GA 00840682109260SEE14021536GA SEE14021537GA 00840682109260SEE14021537GA SEE14021539GA 00840682109260SEE14021539GA SEE14021540GA 00840682109260SEE14021540GA SEE14021541GA 00840682109260SEE14021541GA SEE14021543GA 00840682109260SEE14021543GA SEE14021546GA 00840682109260SEE14021546GA SEE14021547GA 00840682109260SEE14021547GA SEE14021548GA 00840682109260SEE14021548GA SEE14021550GA 00840682109260SEE14021550GA SEE14021551GA 00840682109260SEE14021551GA SEE14021552GA 00840682109260SEE14021552GA SEE14021553GA 00840682109260SEE14021553GA SEE14021554GA 00840682109260SEE14021554GA SEE14021555GA 00840682109260SEE14021555GA SEE14021557GA 00840682109260SEE14021557GA SEE14021558GA 00840682109260SEE14021558GA SEE14021559GA 00840682109260SEE14021559GA SEE14021560GA 00840682109260SEE14021560GA SEE14021561GA 00840682109260SEE14021561GA SEE14021564GA 00840682109260SEE14021564GA SEE14021565GA 00840682109260SEE14021565GA SEE14021566GA 00840682109260SEE14021566GA SEE14021567GA 00840682109260SEE14021567GA SEE14021568GA 00840682109260SEE14021568GA SEE14021569GA 00840682109260SEE14021569GA SEE14021570GA 00840682109260SEE14021570GA SEE14021572GA 00840682109260SEE14021572GA SEE14021573GA 00840682109260SEE14021573GA SEE14021574GA 00840682109260SEE14021574GA SEE14021575GA 00840682109260SEE14021575GA SEE14021576GA 00840682109260SEE14021576GA SEE14021577GA 00840682109260SEE14021577GA SEE14021579GA 00840682109260SEE14021579GA SEE14021580GA 00840682109260SEE14021580GA SEE14021581GA 00840682109260SEE14021581GA SEE14021582GA 00840682109260SEE14021582GA SEE14021583GA 00840682109260SEE14021583GA SEE14021584GA 00840682109260SEE14021584GA SEE14021585GA 00840682109260SEE14021585GA SEE14031586GA 00840682109260SEE14031586GA SEE14031587GA 00840682109260SEE14031587GA SEE14031588GA 00840682109260SEE14031588GA SEE14031589GA 00840682109260SEE14031589GA SEE14031590GA 00840682109260SEE14031590GA SEE14031591GA 00840682109260SEE14031591GA SEE14031592GA 00840682109260SEE14031592GA SEE14031594GA 00840682109260SEE14031594GA SEE14031596GA 00840682109260SEE14031596GA SEE14031597GA 00840682109260SEE14031597GA SEE14031598GA 00840682109260SEE14031598GA SEE14031600GA 00840682109260SEE14031600GA SEE14031601GA 00840682109260SEE14031601GA SEE14031602GA 00840682109260SEE14031602GA SEE14031603GA 00840682109260SEE14031603GA SEE14031604GA 00840682109260SEE14031604GA SEE14031605GA 00840682109260SEE14031605GA SEE14031606GA 00840682109260SEE14031606GA SEE14031607GA 00840682109260SEE14031607GA SEE14031608GA 00840682109260SEE14031608GA SEE14031609GA 00840682109260SEE14031609GA SEE14031610GA 00840682109260SEE14031610GA SEE14031611GA 00840682109260SEE14031611GA SEE14031612GA 00840682109260SEE14031612GA SEE14031613GA 00840682109260SEE14031613GA SEE14031614GA 00840682109260SEE14031614GA SEE14031615GA 00840682109260SEE14031615GA SEE14031616GA 00840682109260SEE14031616GA SEE14031617GA 00840682109260SEE14031617GA SEE14031618GA 00840682109260SEE14031618GA SEE14031620GA 00840682109260SEE14031620GA SEE14031621GA 00840682109260SEE14031621GA SEE14031622GA 00840682109260SEE14031622GA SEE14031623GA 00840682109260SEE14031623GA SEE14031625GA 00840682109260SEE14031625GA SEE14031626GA 00840682109260SEE14031626GA SEE14031627GA 00840682109260SEE14031627GA SEE14031628GA 00840682109260SEE14031628GA SEE14031629GA 00840682109260SEE14031629GA SEE14031630GA 00840682109260SEE14031630GA SEE14031631GA 00840682109260SEE14031631GA SEE14031632GA 00840682109260SEE14031632GA SEE14031634GA 00840682109260SEE14031634GA SEE14031635GA 00840682109260SEE14031635GA SEE14031636GA 00840682109260SEE14031636GA SEE14031637GA 00840682109260SEE14031637GA SEE14031638GA 00840682109260SEE14031638GA SEE14031639GA 00840682109260SEE14031639GA SEE14031640GA 00840682109260SEE14031640GA SEE14031641GA 00840682109260SEE14031641GA SEE14031642GA 00840682109260SEE14031642GA SEE14031643GA 00840682109260SEE14031643GA SEE14031644GA 00840682109260SEE14031644GA SEE14031647GA 00840682109260SEE14031647GA SEE14031648GA 00840682109260SEE14031648GA SEE14031649GA 00840682109260SEE14031649GA SEE14031650GA 00840682109260SEE14031650GA SEE14031651GA 00840682109260SEE14031651GA SEE14031652GA 00840682109260SEE14031652GA SEE14031653GA 00840682109260SEE14031653GA SEE14031654GA 00840682109260SEE14031654GA SEE14031656GA 00840682109260SEE14031656GA SEE14031657GA 00840682109260SEE14031657GA SEE14031658GA 00840682109260SEE14031658GA SEE14031661GA 00840682109260SEE14031661GA SEE14031662GA 00840682109260SEE14031662GA SEE14031663GA 00840682109260SEE14031663GA SEE14031664GA 00840682109260SEE14031664GA SEE14031666GA 00840682109260SEE14031666GA SEE14031667GA 00840682109260SEE14031667GA SEE14031668GA 00840682109260SEE14031668GA SEE14031669GA 00840682109260SEE14031669GA SEE14031670GA 00840682109260SEE14031670GA SEE14031671GA 00840682109260SEE14031671GA SEE14031672GA 00840682109260SEE14031672GA SEE14031673GA 00840682109260SEE14031673GA SEE14031674GA 00840682109260SEE14031674GA SEE14031675GA 00840682109260SEE14031675GA SEE14031676GA 00840682109260SEE14031676GA SEE14031677GA 00840682109260SEE14031677GA SEE14031678GA 00840682109260SEE14031678GA SEE14031679GA 00840682109260SEE14031679GA SEE14031680GA 00840682109260SEE14031680GA SEE14031681GA 00840682109260SEE14031681GA SEE14031682GA 00840682109260SEE14031682GA SEE14031683GA 00840682109260SEE14031683GA SEE14031684GA 00840682109260SEE14031684GA SEE14031685GA 00840682109260SEE14031685GA SEE14031686GA 00840682109260SEE14031686GA SEE14031688GA 00840682109260SEE14031688GA SEE14031689GA 00840682109260SEE14031689GA SEE14031690GA 00840682109260SEE14031690GA SEE14031691GA 00840682109260SEE14031691GA SEE14031692GA 00840682109260SEE14031692GA SEE14031693GA 00840682109260SEE14031693GA SEE14031694GA 00840682109260SEE14031694GA SEE14031695GA 00840682109260SEE14031695GA SEE14031696GA 00840682109260SEE14031696GA SEE14031697GA 00840682109260SEE14031697GA SEE14031698GA 00840682109260SEE14031698GA SEE14031699GA 00840682109260SEE14031699GA SEE14031700GA 00840682109260SEE14031700GA SEE14031701GA 00840682109260SEE14031701GA SEE14031702GA 00840682109260SEE14031702GA SEE14031703GA 00840682109260SEE14031703GA SEE14031704GA 00840682109260SEE14031704GA SEE14031705GA 00840682109260SEE14031705GA SEE14031707GA 00840682109260SEE14031707GA SEE14031708GA 00840682109260SEE14031708GA SEE14031709GA 00840682109260SEE14031709GA SEE14031710GA 00840682109260SEE14031710GA SEE14031711GA 00840682109260SEE14031711GA SEE14031713GA 00840682109260SEE14031713GA SEE14031715GA 00840682109260SEE14031715GA SEE14031716GA 00840682109260SEE14031716GA SEE14031719GA 00840682109260SEE14031719GA SEE14031720GA 00840682109260SEE14031720GA SEE14031721GA 00840682109260SEE14031721GA SEE14031722GA 00840682109260SEE14031722GA SEE14031723GA 00840682109260SEE14031723GA SEE14031724GA 00840682109260SEE14031724GA SEE14031725GA 00840682109260SEE14031725GA SEE14041726GA 00840682109260SEE14041726GA SEE14041727GA 00840682109260SEE14041727GA SEE14041730GA 00840682109260SEE14041730GA SEE14041731GA 00840682109260SEE14041731GA SEE14041732GA 00840682109260SEE14041732GA SEE14041733GA 00840682109260SEE14041733GA SEE14041735GA 00840682109260SEE14041735GA SEE14041737GA 00840682109260SEE14041737GA SEE14041738GA 00840682109260SEE14041738GA SEE14041739GA 00840682109260SEE14041739GA SEE14041740GA 00840682109260SEE14041740GA SEE14041742GA 00840682109260SEE14041742GA SEE14041743GA 00840682109260SEE14041743GA SEE14041744GA 00840682109260SEE14041744GA SEE14041746GA 00840682109260SEE14041746GA SEE14041747GA 00840682109260SEE14041747GA SEE14041748GA 00840682109260SEE14041748GA SEE14041749GA 00840682109260SEE14041749GA SEE14041751GA 00840682109260SEE14041751GA SEE14041752GA 00840682109260SEE14041752GA SEE14041753GA 00840682109260SEE14041753GA SEE14041754GA 00840682109260SEE14041754GA SEE14041755GA 00840682109260SEE14041755GA SEE14041756GA 00840682109260SEE14041756GA SEE14041757GA 00840682109260SEE14041757GA SEE14041758GA 00840682109260SEE14041758GA SEE14041760GA 00840682109260SEE14041760GA SEE14041761GA 00840682109260SEE14041761GA SEE14041762GA 00840682109260SEE14041762GA SEE14041763GA 00840682109260SEE14041763GA SEE14041764GA 00840682109260SEE14041764GA SEE14041765GA 00840682109260SEE14041765GA SEE14081767GA 00840682109260SEE14081767GA SEE14081768GA 00840682109260SEE14081768GA SEE14081771GA 00840682109260SEE14081771GA SEE14081772GA 00840682109260SEE14081772GA SEE14081773GA 00840682109260SEE14081773GA SEE14081775GA 00840682109260SEE14081775GA SEE14101777GA 00840682109260SEE14101777GA SEE14101780GA 00840682109260SEE14101780GA SEE14101781GA 00840682109260SEE14101781GA SEE14101782GA 00840682109260SEE14101782GA SEE14101784GA 00840682109260SEE14101784GA SEE14101785GA 00840682109260SEE14101785GA SEE14101786GA 00840682109260SEE14101786GA SEE14101787GA 00840682109260SEE14101787GA SEE14101790GA 00840682109260SEE14101790GA SEE14101791GA 00840682109260SEE14101791GA SEE14101793GA 00840682109260SEE14101793GA SEE14101794GA 00840682109260SEE14101794GA SEE14101795GA 00840682109260SEE14101795GA SEE14101797GA 00840682109260SEE14101797GA SEE14111799GA 00840682109260SEE14111799GA SEE14111800GA 00840682109260SEE14111800GA SEE14111802GA 00840682109260SEE14111802GA SEE14111803GA 00840682109260SEE14111803GA SEE14111804GA 00840682109260SEE14111804GA SEE14111805GA 00840682109260SEE14111805GA SEE14111806GA 00840682109260SEE14111806GA SEE14151808GA 00840682109260SEE14151808GA SEE14151809GA 00840682109260SEE14151809GA SEE14151810GA 00840682109260SEE14151810GA SEE14151811GA 00840682109260SEE14151811GA SEE14151812GA 00840682109260SEE14151812GA SEE14151813GA 00840682109260SEE14151813GA SEE14151814GA 00840682109260SEE14151814GA SEE14151815GA 00840682109260SEE14151815GA SEE14151816GA 00840682109260SEE14151816GA SEE14151817GA 00840682109260SEE14151817GA SEE14151818GA 00840682109260SEE14151818GA SEE14151821GA 00840682109260SEE14151821GA SEE14151823GA 00840682109260SEE14151823GA SEE14151825GA 00840682109260SEE14151825GA SEE14151831GA 00840682109260SEE14151831GA SEE14151833GA 00840682109260SEE14151833GA SEE14151834GA 00840682109260SEE14151834GA SEE14151835GA 00840682109260SEE14151835GA SEE14151836GA 00840682109260SEE14151836GA SEE14151837GA 00840682109260SEE14151837GA SEE14151838GA 00840682109260SEE14151838GA SEE14151839GA 00840682109260SEE14151839GA SEE14151840GA 00840682109260SEE14151840GA SEE14151841GA 00840682109260SEE14151841GA SEE14151842GA 00840682109260SEE14151842GA SEE14151844GA 00840682109260SEE14151844GA SEE14151845GA 00840682109260SEE14151845GA SEE14151846GA 00840682109260SEE14151846GA SEE14151847GA 00840682109260SEE14151847GA SEE14151848GA 00840682109260SEE14151848GA SEE14151849GA 00840682109260SEE14151849GA SEE14151850GA 00840682109260SEE14151850GA SEE14151851GA 00840682109260SEE14151851GA SEE14151852GA 00840682109260SEE14151852GA SEE14151860GA 00840682109260SEE14151860GA SEE14151861GA 00840682109260SEE14151861GA SEE14151862GA 00840682109260SEE14151862GA SEE14151863GA 00840682109260SEE14151863GA SEE14151866GA 00840682109260SEE14151866GA SEE14151867GA 00840682109260SEE14151867GA SEE14151868GA 00840682109260SEE14151868GA SEE14151869GA 00840682109260SEE14151869GA SEE14151871GA 00840682109260SEE14151871GA SEE14151881GA 00840682109260SEE14151881GA SEE14151883GA 00840682109260SEE14151883GA SEE14161884GA 00840682109260SEE14161884GA SEE14161885GA 00840682109260SEE14161885GA SEE14161887GA 00840682109260SEE14161887GA SEE14161888GA 00840682109260SEE14161888GA SEE14161889GA 00840682109260SEE14161889GA SEE14161890GA 00840682109260SEE14161890GA SEE14161891GA 00840682109260SEE14161891GA SEE14161892GA 00840682109260SEE14161892GA SEE14161893GA 00840682109260SEE14161893GA SEE14161895GA 00840682109260SEE14161895GA SEE14161896GA 00840682109260SEE14161896GA SEE14161897GA 00840682109260SEE14161897GA SEE14161898GA 00840682109260SEE14161898GA SEE14161899GA 00840682109260SEE14161899GA SEE14161900GA 00840682109260SEE14161900GA SEE14161901GA 00840682109260SEE14161901GA SEE14161902GA 00840682109260SEE14161902GA SEE14161903GA 00840682109260SEE14161903GA SEE14161904GA 00840682109260SEE14161904GA SEE14161906GA 00840682109260SEE14161906GA SEE14161907GA 00840682109260SEE14161907GA SEE14161908GA 00840682109260SEE14161908GA SEE14161909GA 00840682109260SEE14161909GA SEE14161910GA 00840682109260SEE14161910GA SEE14161912GA 00840682109260SEE14161912GA SEE14161913GA 00840682109260SEE14161913GA SEE14161914GA 00840682109260SEE14161914GA SEE14171915GA 00840682109260SEE14171915GA SEE14171916GA 00840682109260SEE14171916GA SEE14171917GA 00840682109260SEE14171917GA SEE14171919GA 00840682109260SEE14171919GA SEE14171923GA 00840682109260SEE14171923GA SEE14171924GA 00840682109260SEE14171924GA SEE14171926GA 00840682109260SEE14171926GA SEE14171927GA 00840682109260SEE14171927GA SEE14171929GA 00840682109260SEE14171929GA SEE14171930GA 00840682109260SEE14171930GA SEE14171931GA 00840682109260SEE14171931GA SEE14171932GA 00840682109260SEE14171932GA SEE14171933GA 00840682109260SEE14171933GA SEE14171934GA 00840682109260SEE14171934GA SEE14191935GA 00840682109260SEE14191935GA SEE14191936GA 00840682109260SEE14191936GA SEE14191937GA 00840682109260SEE14191937GA SEE14191938GA 00840682109260SEE14191938GA SEE14191939GA 00840682109260SEE14191939GA SEE14191941GA 00840682109260SEE14191941GA SEE14191942GA 00840682109260SEE14191942GA SEE14191943GA 00840682109260SEE14191943GA SEE14191944GA 00840682109260SEE14191944GA SEE14191945GA 00840682109260SEE14191945GA SEE14191946GA 00840682109260SEE14191946GA SEE14191947GA 00840682109260SEE14191947GA SEE14191948GA 00840682109260SEE14191948GA SEE14191949GA 00840682109260SEE14191949GA SEE14191950GA 00840682109260SEE14191950GA SEE14201951GA 00840682109260SEE14201951GA SEE14201952GA 00840682109260SEE14201952GA SEE14201953GA 00840682109260SEE14201953GA SEE14201954GA 00840682109260SEE14201954GA SEE14201955GA 00840682109260SEE14201955GA SEE14201957GA 00840682109260SEE14201957GA SEE14201958GA 00840682109260SEE14201958GA SEE14201959GA 00840682109260SEE14201959GA SEE14201960GA 00840682109260SEE14201960GA SEE14201961GA 00840682109260SEE14201961GA SEE14211965GA 00840682109260SEE14211965GA SEE14211966GA 00840682109260SEE14211966GA SEE14211967GA 00840682109260SEE14211967GA SEE14211968GA 00840682109260SEE14211968GA SEE14211969GA 00840682109260SEE14211969GA SEE14211970GA 00840682109260SEE14211970GA SEE14211971GA 00840682109260SEE14211971GA SEE14211972GA 00840682109260SEE14211972GA SEE14211973GA 00840682109260SEE14211973GA SEE14211974GA 00840682109260SEE14211974GA SEE14211975GA 00840682109260SEE14211975GA SEE14211977GA 00840682109260SEE14211977GA SEE14211978GA 00840682109260SEE14211978GA SEE14211979GA 00840682109260SEE14211979GA SEE14211981GA 00840682109260SEE14211981GA SEE14211982GA 00840682109260SEE14211982GA SEE14211983GA 00840682109260SEE14211983GA SEE14211986GA 00840682109260SEE14211986GA SEE14211987GA 00840682109260SEE14211987GA SEE14221988GA 00840682109260SEE14221988GA SEE14221992GA 00840682109260SEE14221992GA SEE14221993GA 00840682109260SEE14221993GA SEE14221996GA 00840682109260SEE14221996GA SEE14221997GA 00840682109260SEE14221997GA SEE14221998GA 00840682109260SEE14221998GA SEE14221999GA 00840682109260SEE14221999GA SEE14222000GA 00840682109260SEE14222000GA SEE14222001GA 00840682109260SEE14222001GA SEE14222003GA 00840682109260SEE14222003GA SEE14222006GA 00840682109260SEE14222006GA SEE14222009GA 00840682109260SEE14222009GA SEE14222010GA 00840682109260SEE14222010GA SEE14222014GA 00840682109260SEE14222014GA SEE14222018GA 00840682109260SEE14222018GA SEE14222019GA 00840682109260SEE14222019GA SEE14222021GA 00840682109260SEE14222021GA SEE14222022GA 00840682109260SEE14222022GA SEE14222023GA 00840682109260SEE14222023GA SEE14222024GA 00840682109260SEE14222024GA SEE14222025GA 00840682109260SEE14222025GA SEE14222026GA 00840682109260SEE14222026GA SEE14222027GA 00840682109260SEE14222027GA SEE14222028GA 00840682109260SEE14222028GA SEE14222029GA 00840682109260SEE14222029GA SEE14222032GA 00840682109260SEE14222032GA SEE14222033GA 00840682109260SEE14222033GA SEE14222034GA 00840682109260SEE14222034GA SEE14222036GA 00840682109260SEE14222036GA SEE14222037GA 00840682109260SEE14222037GA SEE14222038GA 00840682109260SEE14222038GA SEE14222039GA 00840682109260SEE14222039GA SEE14222040GA 00840682109260SEE14222040GA SEE14222042GA 00840682109260SEE14222042GA SEE14222043GA 00840682109260SEE14222043GA SEE14222044GA 00840682109260SEE14222044GA SEE14222047GA 00840682109260SEE14222047GA SEE14222048GA 00840682109260SEE14222048GA SEE14222049GA 00840682109260SEE14222049GA SEE14222050GA 00840682109260SEE14222050GA SEE14222051GA 00840682109260SEE14222051GA SEE14222052GA 00840682109260SEE14222052GA SEE14222053GA 00840682109260SEE14222053GA SEE14222054GA 00840682109260SEE14222054GA SEE14222055GA 00840682109260SEE14222055GA SEE14222056GA 00840682109260SEE14222056GA SEE14222057GA 00840682109260SEE14222057GA SEE14222058GA 00840682109260SEE14222058GA SEE14222059GA 00840682109260SEE14222059GA SEE14222060GA 00840682109260SEE14222060GA SEE14222062GA 00840682109260SEE14222062GA SEE14222063GA 00840682109260SEE14222063GA SEE14222064GA 00840682109260SEE14222064GA SEE14222065GA 00840682109260SEE14222065GA SEE14222066GA 00840682109260SEE14222066GA SEE14222067GA 00840682109260SEE14222067GA SEE14222068GA 00840682109260SEE14222068GA SEE14222069GA 00840682109260SEE14222069GA SEE14222070GA 00840682109260SEE14222070GA SEE14222071GA 00840682109260SEE14222071GA SEE14222072GA 00840682109260SEE14222072GA SEE14222073GA 00840682109260SEE14222073GA SEE14222074GA 00840682109260SEE14222074GA SEE14222075GA 00840682109260SEE14222075GA SEE14222076GA 00840682109260SEE14222076GA SEE14222077GA 00840682109260SEE14222077GA SEE14242079GA 00840682109260SEE14242079GA SEE14242080GA 00840682109260SEE14242080GA SEE14242081GA 00840682109260SEE14242081GA SEE14242082GA 00840682109260SEE14242082GA SEE14242083GA 00840682109260SEE14242083GA SEE14242084GA 00840682109260SEE14242084GA SEE14242085GA 00840682109260SEE14242085GA SEE14242086GA 00840682109260SEE14242086GA SEE14242087GA 00840682109260SEE14242087GA SEE14242088GA 00840682109260SEE14242088GA SEE14242089GA 00840682109260SEE14242089GA SEE14242090GA 00840682109260SEE14242090GA SEE14242091GA 00840682109260SEE14242091GA SEE14242092GA 00840682109260SEE14242092GA SEE14242093GA 00840682109260SEE14242093GA SEE14242094GA 00840682109260SEE14242094GA SEE14242095GA 00840682109260SEE14242095GA SEE14242096GA 00840682109260SEE14242096GA SEE14242098GA 00840682109260SEE14242098GA SEE14242099GA 00840682109260SEE14242099GA SEE14242100GA 00840682109260SEE14242100GA SEE14242102GA 00840682109260SEE14242102GA SEE14242104GA 00840682109260SEE14242104GA SEE14242105GA 00840682109260SEE14242105GA SEE14242106GA 00840682109260SEE14242106GA SEE14242107GA 00840682109260SEE14242107GA SEE14242108GA 00840682109260SEE14242108GA SEE14242109GA 00840682109260SEE14242109GA SEE14242111GA 00840682109260SEE14242111GA SEE14242112GA 00840682109260SEE14242112GA SEE14242113GA 00840682109260SEE14242113GA SEE14242114GA 00840682109260SEE14242114GA SEE14242115GA 00840682109260SEE14242115GA SEE14242116GA 00840682109260SEE14242116GA SEE14242117GA 00840682109260SEE14242117GA SEE14242118GA 00840682109260SEE14242118GA SEE14242119GA 00840682109260SEE14242119GA SEE14242120GA 00840682109260SEE14242120GA SEE14252123GA 00840682109260SEE14252123GA SEE14252124GA 00840682109260SEE14252124GA SEE14252125GA 00840682109260SEE14252125GA SEE14252128GA 00840682109260SEE14252128GA SEE14252130GA 00840682109260SEE14252130GA SEE14252131GA 00840682109260SEE14252131GA SEE14252133GA 00840682109260SEE14252133GA SEE14252135GA 00840682109260SEE14252135GA SEE14272136GA 00840682109260SEE14272136GA SEE14272139GA 00840682109260SEE14272139GA SEE14272140GA 00840682109260SEE14272140GA SEE14272143GA 00840682109260SEE14272143GA SEE14272144GA 00840682109260SEE14272144GA SEE14272145GA 00840682109260SEE14272145GA SEE14272146GA 00840682109260SEE14272146GA SEE14272147GA 00840682109260SEE14272147GA SEE14272149GA 00840682109260SEE14272149GA SEE14272150GA 00840682109260SEE14272150GA SEE14272151GA 00840682109260SEE14272151GA SEE14272152GA 00840682109260SEE14272152GA SEE14272154GA 00840682109260SEE14272154GA SEE14272155GA 00840682109260SEE14272155GA SEE14272160GA 00840682109260SEE14272160GA SEE14272161GA 00840682109260SEE14272161GA SEE14272162GA 00840682109260SEE14272162GA SEE14272163GA 00840682109260SEE14272163GA SEE14272164GA 00840682109260SEE14272164GA SEE14272165GA 00840682109260SEE14272165GA SEE14272168GA 00840682109260SEE14272168GA SEE14272169GA 00840682109260SEE14272169GA SEE14272170GA 00840682109260SEE14272170GA SEE14272171GA 00840682109260SEE14272171GA SEE14272172GA 00840682109260SEE14272172GA SEE14272173GA 00840682109260SEE14272173GA SEE14272174GA 00840682109260SEE14272174GA SEE14272175GA 00840682109260SEE14272175GA SEE14272176GA 00840682109260SEE14272176GA SEE14272199GA 00840682109260SEE14272199GA SEE14272200GA 00840682109260SEE14272200GA SEE14272201GA 00840682109260SEE14272201GA SEE14272202GA 00840682109260SEE14272202GA SEE14272203GA 00840682109260SEE14272203GA SEE14272204GA 00840682109260SEE14272204GA SEE14272205GA 00840682109260SEE14272205GA SEE14272207GA 00840682109260SEE14272207GA SEE14272208GA 00840682109260SEE14272208GA SEE14272209GA 00840682109260SEE14272209GA SEE14272211GA 00840682109260SEE14272211GA SEE14272213GA 00840682109260SEE14272213GA SEE14272214GA 00840682109260SEE14272214GA SEE14272215GA 00840682109260SEE14272215GA SEE14272216GA 00840682109260SEE14272216GA SEE14272217GA 00840682109260SEE14272217GA SEE14272218GA 00840682109260SEE14272218GA SEE14272219GA 00840682109260SEE14272219GA SEE14272220GA 00840682109260SEE14272220GA SEE14272221GA 00840682109260SEE14272221GA SEE14272222GA 00840682109260SEE14272222GA SEE14272223GA 00840682109260SEE14272223GA SEE14272224GA 00840682109260SEE14272224GA SEE14272225GA 00840682109260SEE14272225GA SEE14272226GA 00840682109260SEE14272226GA SEE14272227GA 00840682109260SEE14272227GA SEE14272228GA 00840682109260SEE14272228GA SEE14272229GA 00840682109260SEE14272229GA SEE14292230GA 00840682109260SEE14292230GA SEE14292231GA 00840682109260SEE14292231GA SEE14292232GA 00840682109260SEE14292232GA SEE14292233GA 00840682109260SEE14292233GA SEE14292234GA 00840682109260SEE14292234GA SEE14292235GA 00840682109260SEE14292235GA SEE14292236GA 00840682109260SEE14292236GA SEE14292237GA 00840682109260SEE14292237GA SEE14292239GA 00840682109260SEE14292239GA SEE14292240GA 00840682109260SEE14292240GA SEE14292241GA 00840682109260SEE14292241GA SEE14292243GA 00840682109260SEE14292243GA SEE14292244GA 00840682109260SEE14292244GA SEE14292245GA 00840682109260SEE14292245GA SEE14302255GA 00840682109260SEE14302255GA SEE14302257GA 00840682109260SEE14302257GA SEE14302258GA 00840682109260SEE14302258GA SEE14302259GA 00840682109260SEE14302259GA SEE14302261GA 00840682109260SEE14302261GA SEE14302262GA 00840682109260SEE14302262GA SEE14302264GA 00840682109260SEE14302264GA SEE14302265GA 00840682109260SEE14302265GA SEE14302266GA 00840682109260SEE14302266GA SEE14302267GA 00840682109260SEE14302267GA SEE14302268GA 00840682109260SEE14302268GA SEE14302269GA 00840682109260SEE14302269GA SEE14302270GA 00840682109260SEE14302270GA SEE14302271GA 00840682109260SEE14302271GA SEE14302272GA 00840682109260SEE14302272GA SEE14302273GA 00840682109260SEE14302273GA SEE14302274GA 00840682109260SEE14302274GA SEE14302275GA 00840682109260SEE14302275GA SEE14302276GA 00840682109260SEE14302276GA SEE14302277GA 00840682109260SEE14302277GA SEE14302278GA 00840682109260SEE14302278GA SEE14302279GA 00840682109260SEE14302279GA SEE14302280GA 00840682109260SEE14302280GA SEE14302281GA 00840682109260SEE14302281GA SEE14302282GA 00840682109260SEE14302282GA SEE14302283GA 00840682109260SEE14302283GA SEE14302284GA 00840682109260SEE14302284GA SEE14302285GA 00840682109260SEE14302285GA SEE14302286GA 00840682109260SEE14302286GA SEE14302287GA 00840682109260SEE14302287GA SEE14312288GA 00840682109260SEE14312288GA SEE14312289GA 00840682109260SEE14312289GA SEE14312290GA 00840682109260SEE14312290GA SEE14312291GA 00840682109260SEE14312291GA SEE14312292GA 00840682109260SEE14312292GA SEE14312293GA 00840682109260SEE14312293GA SEE14312294GA 00840682109260SEE14312294GA SEE14312295GA 00840682109260SEE14312295GA SEE14312296GA 00840682109260SEE14312296GA SEE14312297GA 00840682109260SEE14312297GA SEE14312299GA 00840682109260SEE14312299GA SEE14312300GA 00840682109260SEE14312300GA SEE14312301GA 00840682109260SEE14312301GA SEE14312302GA 00840682109260SEE14312302GA SEE14312303GA 00840682109260SEE14312303GA SEE14312304GA 00840682109260SEE14312304GA SEE14312305GA 00840682109260SEE14312305GA SEE14312306GA 00840682109260SEE14312306GA SEE14312307GA 00840682109260SEE14312307GA SEE14322308GA 00840682109260SEE14322308GA SEE14322309GA 00840682109260SEE14322309GA SEE14322310GA 00840682109260SEE14322310GA SEE14322312GA 00840682109260SEE14322312GA SEE14322313GA 00840682109260SEE14322313GA SEE14322314GA 00840682109260SEE14322314GA SEE14322315GA 00840682109260SEE14322315GA SEE14322316GA 00840682109260SEE14322316GA SEE14322317GA 00840682109260SEE14322317GA SEE14322318GA 00840682109260SEE14322318GA SEE14322319GA 00840682109260SEE14322319GA SEE14322320GA 00840682109260SEE14322320GA SEE14322321GA 00840682109260SEE14322321GA SEE14322322GA 00840682109260SEE14322322GA SEE14322323GA 00840682109260SEE14322323GA SEE14322324GA 00840682109260SEE14322324GA SEE14322326GA 00840682109260SEE14322326GA SEE14322327GA 00840682109260SEE14322327GA SEE14322328GA 00840682109260SEE14322328GA SEE14322329GA 00840682109260SEE14322329GA SEE14322330GA 00840682109260SEE14322330GA SEE14322331GA 00840682109260SEE14322331GA SEE14322332GA 00840682109260SEE14322332GA SEE14322333GA 00840682109260SEE14322333GA SEE14322334GA 00840682109260SEE14322334GA SEE14322335GA 00840682109260SEE14322335GA SEE14322336GA 00840682109260SEE14322336GA SEE14322337GA 00840682109260SEE14322337GA SEE14332338GA 00840682109260SEE14332338GA SEE14332340GA 00840682109260SEE14332340GA SEE14332343GA 00840682109260SEE14332343GA SEE14332344GA 00840682109260SEE14332344GA SEE14362348GA 00840682109260SEE14362348GA SEE14362349GA 00840682109260SEE14362349GA SEE14362350GA 00840682109260SEE14362350GA SEE14362351GA 00840682109260SEE14362351GA SEE14362352GA 00840682109260SEE14362352GA SEE14362353GA 00840682109260SEE14362353GA SEE14362354GA 00840682109260SEE14362354GA SEE14362355GA 00840682109260SEE14362355GA SEE14362356GA 00840682109260SEE14362356GA SEE14362357GA 00840682109260SEE14362357GA SEE14362358GA 00840682109260SEE14362358GA SEE14362359GA 00840682109260SEE14362359GA SEE14362360GA 00840682109260SEE14362360GA SEE14362361GA 00840682109260SEE14362361GA SEE14362362GA 00840682109260SEE14362362GA SEE14362363GA 00840682109260SEE14362363GA SEE14362364GA 00840682109260SEE14362364GA SEE14362365GA 00840682109260SEE14362365GA SEE14362366GA 00840682109260SEE14362366GA SEE14362367GA 00840682109260SEE14362367GA SEE14362368GA 00840682109260SEE14362368GA SEE14362369GA 00840682109260SEE14362369GA SEE14362370GA 00840682109260SEE14362370GA SEE14362371GA 00840682109260SEE14362371GA SEE14362372GA 00840682109260SEE14362372GA SEE14362373GA 00840682109260SEE14362373GA SEE14362374GA 00840682109260SEE14362374GA SEE14362375GA 00840682109260SEE14362375GA SEE14362376GA 00840682109260SEE14362376GA SEE14362377GA 00840682109260SEE14362377GA SEE14362379GA 00840682109260SEE14362379GA SEE14362380GA 00840682109260SEE14362380GA SEE14362381GA 00840682109260SEE14362381GA SEE14362382GA 00840682109260SEE14362382GA SEE14362383GA 00840682109260SEE14362383GA SEE14362384GA 00840682109260SEE14362384GA SEE14362385GA 00840682109260SEE14362385GA SEE14362386GA 00840682109260SEE14362386GA SEE14362387GA 00840682109260SEE14362387GA SEE14372388GA 00840682109260SEE14372388GA SEE14372389GA 00840682109260SEE14372389GA SEE14372390GA 00840682109260SEE14372390GA SEE14372391GA 00840682109260SEE14372391GA SEE14372392GA 00840682109260SEE14372392GA SEE14372393GA 00840682109260SEE14372393GA SEE14372394GA 00840682109260SEE14372394GA SEE14372395GA 00840682109260SEE14372395GA SEE14372396GA 00840682109260SEE14372396GA SEE14372397GA 00840682109260SEE14372397GA SEE14372398GA 00840682109260SEE14372398GA SEE14372399GA 00840682109260SEE14372399GA SEE14372400GA 00840682109260SEE14372400GA SEE14372401GA 00840682109260SEE14372401GA SEE14372402GA 00840682109260SEE14372402GA SEE14372403GA 00840682109260SEE14372403GA SEE14372405GA 00840682109260SEE14372405GA SEE14372406GA 00840682109260SEE14372406GA SEE14372407GA 00840682109260SEE14372407GA SEE14372409GA 00840682109260SEE14372409GA SEE14372410GA 00840682109260SEE14372410GA SEE14372411GA 00840682109260SEE14372411GA SEE14372413GA 00840682109260SEE14372413GA SEE14372414GA 00840682109260SEE14372414GA SEE14372415GA 00840682109260SEE14372415GA SEE14372419GA 00840682109260SEE14372419GA SEE14372420GA 00840682109260SEE14372420GA SEE14372422GA 00840682109260SEE14372422GA SEE14372423GA 00840682109260SEE14372423GA SEE14372424GA 00840682109260SEE14372424GA SEE14372425GA 00840682109260SEE14372425GA SEE14372426GA 00840682109260SEE14372426GA SEE14372427GA 00840682109260SEE14372427GA SEE14372428GA 00840682109260SEE14372428GA SEE14372429GA 00840682109260SEE14372429GA SEE14372430GA 00840682109260SEE14372430GA SEE14372432GA 00840682109260SEE14372432GA SEE14372433GA 00840682109260SEE14372433GA SEE14372434GA 00840682109260SEE14372434GA SEE14372435GA 00840682109260SEE14372435GA SEE14372436GA 00840682109260SEE14372436GA SEE14372437GA 00840682109260SEE14372437GA SEE14372438GA 00840682109260SEE14372438GA SEE14372439GA 00840682109260SEE14372439GA SEE14372440GA 00840682109260SEE14372440GA SEE14372441GA 00840682109260SEE14372441GA SEE14372442GA 00840682109260SEE14372442GA SEE14382453GA 00840682109260SEE14382453GA SEE14382454GA 00840682109260SEE14382454GA SEE14382456GA 00840682109260SEE14382456GA SEE14382457GA 00840682109260SEE14382457GA SEE14382458GA 00840682109260SEE14382458GA SEE14382459GA 00840682109260SEE14382459GA SEE14382460GA 00840682109260SEE14382460GA SEE14382461GA 00840682109260SEE14382461GA SEE14382462GA 00840682109260SEE14382462GA SEE14382463GA 00840682109260SEE14382463GA SEE14382464GA 00840682109260SEE14382464GA SEE14382465GA 00840682109260SEE14382465GA SEE14382466GA 00840682109260SEE14382466GA SEE14382467GA 00840682109260SEE14382467GA SEE14382469GA 00840682109260SEE14382469GA SEE14382470GA 00840682109260SEE14382470GA SEE14382471GA 00840682109260SEE14382471GA SEE14382472GA 00840682109260SEE14382472GA SEE14402473GA 00840682109260SEE14402473GA SEE14402474GA 00840682109260SEE14402474GA SEE14402475GA 00840682109260SEE14402475GA SEE14402476GA 00840682109260SEE14402476GA SEE14402477GA 00840682109260SEE14402477GA SEE14402478GA 00840682109260SEE14402478GA SEE14402479GA 00840682109260SEE14402479GA SEE14402480GA 00840682109260SEE14402480GA SEE14402481GA 00840682109260SEE14402481GA SEE14402482GA 00840682109260SEE14402482GA SEE14402483GA 00840682109260SEE14402483GA SEE14402484GA 00840682109260SEE14402484GA SEE14402485GA 00840682109260SEE14402485GA SEE14402486GA 00840682109260SEE14402486GA SEE14402487GA 00840682109260SEE14402487GA SEE14402488GA 00840682109260SEE14402488GA SEE14402489GA 00840682109260SEE14402489GA SEE14402490GA 00840682109260SEE14402490GA SEE14402491GA 00840682109260SEE14402491GA SEE14402492GA 00840682109260SEE14402492GA SEE14412494GA 00840682109260SEE14412494GA SEE14412495GA 00840682109260SEE14412495GA SEE14412496GA 00840682109260SEE14412496GA SEE14412497GA 00840682109260SEE14412497GA SEE14412498GA 00840682109260SEE14412498GA SEE14412499GA 00840682109260SEE14412499GA SEE14412500GA 00840682109260SEE14412500GA SEE14412501GA 00840682109260SEE14412501GA SEE14412502GA 00840682109260SEE14412502GA SEE14412503GA 00840682109260SEE14412503GA SEE14412505GA 00840682109260SEE14412505GA SEE14412506GA 00840682109260SEE14412506GA SEE14412508GA 00840682109260SEE14412508GA SEE14412509GA 00840682109260SEE14412509GA SEE14412510GA 00840682109260SEE14412510GA SEE14412511GA 00840682109260SEE14412511GA SEE14412512GA 00840682109260SEE14412512GA SEE14412513GA 00840682109260SEE14412513GA SEE14422514GA 00840682109260SEE14422514GA SEE14422515GA 00840682109260SEE14422515GA SEE14422516GA 00840682109260SEE14422516GA SEE14422517GA 00840682109260SEE14422517GA SEE14422518GA 00840682109260SEE14422518GA SEE14422520GA 00840682109260SEE14422520GA SEE14422521GA 00840682109260SEE14422521GA SEE14422522GA 00840682109260SEE14422522GA SEE14422523GA 00840682109260SEE14422523GA SEE14422524GA 00840682109260SEE14422524GA SEE14422525GA 00840682109260SEE14422525GA SEE14422526GA 00840682109260SEE14422526GA SEE14422527GA 00840682109260SEE14422527GA SEE14422528GA 00840682109260SEE14422528GA SEE14422529GA 00840682109260SEE14422529GA SEE14422530GA 00840682109260SEE14422530GA SEE14422531GA 00840682109260SEE14422531GA SEE14422532GA 00840682109260SEE14422532GA SEE14422533GA 00840682109260SEE14422533GA SEE14422534GA 00840682109260SEE14422534GA SEE14422535GA 00840682109260SEE14422535GA SEE14422536GA 00840682109260SEE14422536GA SEE14422537GA 00840682109260SEE14422537GA SEE14422538GA 00840682109260SEE14422538GA SEE14422539GA 00840682109260SEE14422539GA SEE14422540GA 00840682109260SEE14422540GA SEE14422541GA 00840682109260SEE14422541GA SEE14422542GA 00840682109260SEE14422542GA SEE14422543GA 00840682109260SEE14422543GA SEE14422545GA 00840682109260SEE14422545GA SEE14422546GA 00840682109260SEE14422546GA SEE14422547GA 00840682109260SEE14422547GA SEE14422548GA 00840682109260SEE14422548GA SEE14422549GA 00840682109260SEE14422549GA SEE14422550GA 00840682109260SEE14422550GA SEE14422551GA 00840682109260SEE14422551GA SEE14422552GA 00840682109260SEE14422552GA SEE14422553GA 00840682109260SEE14422553GA SEE14422555GA 00840682109260SEE14422555GA SEE14422556GA 00840682109260SEE14422556GA SEE14422557GA 00840682109260SEE14422557GA SEE14422558GA 00840682109260SEE14422558GA SEE14422560GA 00840682109260SEE14422560GA SEE14422561GA 00840682109260SEE14422561GA SEE14422562GA 00840682109260SEE14422562GA SEE14422564GA 00840682109260SEE14422564GA SEE14422565GA 00840682109260SEE14422565GA SEE14422566GA 00840682109260SEE14422566GA SEE14422567GA 00840682109260SEE14422567GA SEE14422568GA 00840682109260SEE14422568GA SEE14422569GA 00840682109260SEE14422569GA SEE14422570GA 00840682109260SEE14422570GA SEE14422571GA 00840682109260SEE14422571GA SEE14422572GA 00840682109260SEE14422572GA SEE14422573GA 00840682109260SEE14422573GA SEE14432574GA 00840682109260SEE14432574GA SEE14432575GA 00840682109260SEE14432575GA SEE14432578GA 00840682109260SEE14432578GA SEE14432579GA 00840682109260SEE14432579GA SEE14432580GA 00840682109260SEE14432580GA SEE14432583GA 00840682109260SEE14432583GA SEE14432584GA 00840682109260SEE14432584GA SEE14432585GA 00840682109260SEE14432585GA SEE14432586GA 00840682109260SEE14432586GA SEE14432587GA 00840682109260SEE14432587GA SEE14432588GA 00840682109260SEE14432588GA SEE14432589GA 00840682109260SEE14432589GA SEE14432590GA 00840682109260SEE14432590GA SEE14432591GA 00840682109260SEE14432591GA SEE14432592GA 00840682109260SEE14432592GA SEE14432594GA 00840682109260SEE14432594GA SEE14432596GA 00840682109260SEE14432596GA SEE14432597GA 00840682109260SEE14432597GA SEE14432598GA 00840682109260SEE14432598GA SEE14432599GA 00840682109260SEE14432599GA SEE14432600GA 00840682109260SEE14432600GA SEE14432601GA 00840682109260SEE14432601GA SEE14432602GA 00840682109260SEE14432602GA SEE14432603GA 00840682109260SEE14432603GA SEE14432604GA 00840682109260SEE14432604GA SEE14432605GA 00840682109260SEE14432605GA SEE14432606GA 00840682109260SEE14432606GA SEE14432607GA 00840682109260SEE14432607GA SEE14432608GA 00840682109260SEE14432608GA SEE14432609GA 00840682109260SEE14432609GA SEE14432610GA 00840682109260SEE14432610GA SEE14432611GA 00840682109260SEE14432611GA SEE14432612GA 00840682109260SEE14432612GA SEE14432613GA 00840682109260SEE14432613GA SEE14432614GA 00840682109260SEE14432614GA SEE14452615GA 00840682109260SEE14452615GA SEE14452616GA 00840682109260SEE14452616GA SEE14452617GA 00840682109260SEE14452617GA SEE14452618GA 00840682109260SEE14452618GA SEE14452620GA 00840682109260SEE14452620GA SEE14452621GA 00840682109260SEE14452621GA SEE14452622GA 00840682109260SEE14452622GA SEE14452623GA 00840682109260SEE14452623GA SEE14452624GA 00840682109260SEE14452624GA SEE14452625GA 00840682109260SEE14452625GA SEE14452629GA 00840682109260SEE14452629GA SEE14452630GA 00840682109260SEE14452630GA SEE14452631GA 00840682109260SEE14452631GA SEE14452632GA 00840682109260SEE14452632GA SEE14452633GA 00840682109260SEE14452633GA SEE14452634GA 00840682109260SEE14452634GA SEE14452635GA 00840682109260SEE14452635GA SEE14452636GA 00840682109260SEE14452636GA SEE14452638GA 00840682109260SEE14452638GA SEE14452639GA 00840682109260SEE14452639GA SEE14452640GA 00840682109260SEE14452640GA SEE14452641GA 00840682109260SEE14452641GA SEE14452642GA 00840682109260SEE14452642GA SEE14462644GA 00840682109260SEE14462644GA SEE14462645GA 00840682109260SEE14462645GA SEE14462646GA 00840682109260SEE14462646GA SEE14462649GA 00840682109260SEE14462649GA SEE14462652GA 00840682109260SEE14462652GA SEE14462654GA 00840682109260SEE14462654GA SEE14462655GA 00840682109260SEE14462655GA SEE14462656GA 00840682109260SEE14462656GA SEE14462657GA 00840682109260SEE14462657GA SEE14462659GA 00840682109260SEE14462659GA SEE14462662GA 00840682109260SEE14462662GA SEE14462663GA 00840682109260SEE14462663GA SEE14462665GA 00840682109260SEE14462665GA SEE14462666GA 00840682109260SEE14462666GA SEE14462667GA 00840682109260SEE14462667GA SEE14462668GA 00840682109260SEE14462668GA SEE14462669GA 00840682109260SEE14462669GA SEE14462670GA 00840682109260SEE14462670GA SEE14462671GA 00840682109260SEE14462671GA SEE14472672GA 00840682109260SEE14472672GA SEE14472673GA 00840682109260SEE14472673GA SEE14472674GA 00840682109260SEE14472674GA SEE14472675GA 00840682109260SEE14472675GA SEE14472676GA 00840682109260SEE14472676GA SEE14472679GA 00840682109260SEE14472679GA SEE14472680GA 00840682109260SEE14472680GA SEE14472681GA 00840682109260SEE14472681GA SEE14482682GA 00840682109260SEE14482682GA SEE14482683GA 00840682109260SEE14482683GA SEE14482684GA 00840682109260SEE14482684GA SEE14482685GA 00840682109260SEE14482685GA SEE14482686GA 00840682109260SEE14482686GA SEE14482687GA 00840682109260SEE14482687GA SEE14482688GA 00840682109260SEE14482688GA SEE14482689GA 00840682109260SEE14482689GA SEE14482690GA 00840682109260SEE14482690GA SEE14482691GA 00840682109260SEE14482691GA SEE14482692GA 00840682109260SEE14482692GA SEE14482693GA 00840682109260SEE14482693GA SEE14482694GA 00840682109260SEE14482694GA SEE14482695GA 00840682109260SEE14482695GA SEE14482696GA 00840682109260SEE14482696GA SEE14482697GA 00840682109260SEE14482697GA SEE14482698GA 00840682109260SEE14482698GA SEE14482699GA 00840682109260SEE14482699GA SEE14482700GA 00840682109260SEE14482700GA SEE14482701GA 00840682109260SEE14482701GA SEE14482702GA 00840682109260SEE14482702GA SEE14482703GA 00840682109260SEE14482703GA SEE14482705GA 00840682109260SEE14482705GA SEE14482707GA 00840682109260SEE14482707GA SEE14482708GA 00840682109260SEE14482708GA SEE14482709GA 00840682109260SEE14482709GA SEE14482710GA 00840682109260SEE14482710GA SEE14482711GA 00840682109260SEE14482711GA SEE14482712GA 00840682109260SEE14482712GA SEE14482713GA 00840682109260SEE14482713GA SEE14482714GA 00840682109260SEE14482714GA SEE14482716GA 00840682109260SEE14482716GA SEE14482719GA 00840682109260SEE14482719GA SEE14482721GA 00840682109260SEE14482721GA SEE14492722GA 00840682109260SEE14492722GA SEE14492724GA 00840682109260SEE14492724GA SEE14492725GA 00840682109260SEE14492725GA SEE14492726GA 00840682109260SEE14492726GA SEE14492727GA 00840682109260SEE14492727GA SEE14492728GA 00840682109260SEE14492728GA SEE14492729GA 00840682109260SEE14492729GA SEE14492730GA 00840682109260SEE14492730GA SEE14492731GA 00840682109260SEE14492731GA SEE14492742GA 00840682109260SEE14492742GA SEE14492743GA 00840682109260SEE14492743GA SEE14492744GA 00840682109260SEE14492744GA SEE14492745GA 00840682109260SEE14492745GA SEE14492746GA 00840682109260SEE14492746GA SEE14492747GA 00840682109260SEE14492747GA SEE14492748GA 00840682109260SEE14492748GA SEE14492749GA 00840682109260SEE14492749GA SEE14492750GA 00840682109260SEE14492750GA SEE14492751GA 00840682109260SEE14492751GA SEE14492752GA 00840682109260SEE14492752GA SEE14492753GA 00840682109260SEE14492753GA SEE14492754GA 00840682109260SEE14492754GA SEE14492755GA 00840682109260SEE14492755GA SEE14492756GA 00840682109260SEE14492756GA SEE14492757GA 00840682109260SEE14492757GA SEE14492758GA 00840682109260SEE14492758GA SEE14492759GA 00840682109260SEE14492759GA SEE14492760GA 00840682109260SEE14492760GA SEE14502762GA 00840682109260SEE14502762GA SEE14502763GA 00840682109260SEE14502763GA SEE14502764GA 00840682109260SEE14502764GA SEE14502765GA 00840682109260SEE14502765GA SEE14502766GA 00840682109260SEE14502766GA SEE14502767GA 00840682109260SEE14502767GA SEE14502768GA 00840682109260SEE14502768GA SEE14502769GA 00840682109260SEE14502769GA SEE14502770GA 00840682109260SEE14502770GA SEE14502771GA 00840682109260SEE14502771GA SEE14502772GA 00840682109260SEE14502772GA SEE14502773GA 00840682109260SEE14502773GA SEE14502774GA 00840682109260SEE14502774GA SEE14502775GA 00840682109260SEE14502775GA SEE14502776GA 00840682109260SEE14502776GA SEE14502777GA 00840682109260SEE14502777GA SEE14502778GA 00840682109260SEE14502778GA SEE14502779GA 00840682109260SEE14502779GA SEE14502780GA 00840682109260SEE14502780GA SEE14502781GA 00840682109260SEE14502781GA SEE14502782GA 00840682109260SEE14502782GA SEE14502783GA 00840682109260SEE14502783GA SEE14502784GA 00840682109260SEE14502784GA SEE14502785GA 00840682109260SEE14502785GA SEE14502786GA 00840682109260SEE14502786GA SEE14502787GA 00840682109260SEE14502787GA SEE14502788GA 00840682109260SEE14502788GA SEE14502789GA 00840682109260SEE14502789GA SEE14502790GA 00840682109260SEE14502790GA SEE14502791GA 00840682109260SEE14502791GA SEE14502792GA 00840682109260SEE14502792GA SEE14502793GA 00840682109260SEE14502793GA SEE14502794GA 00840682109260SEE14502794GA SEE14502795GA 00840682109260SEE14502795GA SEE14502796GA 00840682109260SEE14502796GA SEE14502797GA 00840682109260SEE14502797GA SEE14502798GA 00840682109260SEE14502798GA SEE14502799GA 00840682109260SEE14502799GA SEE14502800GA 00840682109260SEE14502800GA SEE14502801GA 00840682109260SEE14502801GA SEE14502802GA 00840682109260SEE14502802GA SEE14502803GA 00840682109260SEE14502803GA SEE14502804GA 00840682109260SEE14502804GA SEE14502805GA 00840682109260SEE14502805GA SEE14502806GA 00840682109260SEE14502806GA SEE14502807GA 00840682109260SEE14502807GA SEE14502808GA 00840682109260SEE14502808GA SEE14502809GA 00840682109260SEE14502809GA SEE14502810GA 00840682109260SEE14502810GA SEE14502811GA 00840682109260SEE14502811GA SEE14502812GA 00840682109260SEE14502812GA SEE14502813GA 00840682109260SEE14502813GA SEE14502814GA 00840682109260SEE14502814GA SEE14502815GA 00840682109260SEE14502815GA SEE14502816GA 00840682109260SEE14502816GA SEE14502817GA 00840682109260SEE14502817GA SEE14502818GA 00840682109260SEE14502818GA SEE14512819GA 00840682109260SEE14512819GA SEE14512821GA 00840682109260SEE14512821GA SEE14512822GA 00840682109260SEE14512822GA SEE14512823GA 00840682109260SEE14512823GA SEE14512825GA 00840682109260SEE14512825GA SEE14512826GA 00840682109260SEE14512826GA SEE14512827GA 00840682109260SEE14512827GA SEE14512828GA 00840682109260SEE14512828GA SEE14512829GA 00840682109260SEE14512829GA SEE14512830GA 00840682109260SEE14512830GA SEE14512831GA 00840682109260SEE14512831GA SEE14512832GA 00840682109260SEE14512832GA SEE14512833GA 00840682109260SEE14512833GA SEE14512834GA 00840682109260SEE14512834GA SEE14512835GA 00840682109260SEE14512835GA SEE14512836GA 00840682109260SEE14512836GA SEE14512837GA 00840682109260SEE14512837GA SEE14512838GA 00840682109260SEE14512838GA SEE15012839GA 00840682109260SEE15012839GA SEE15012840GA 00840682109260SEE15012840GA SEE15012841GA 00840682109260SEE15012841GA SEE15012842GA 00840682109260SEE15012842GA SEE15012843GA 00840682109260SEE15012843GA SEE15012845GA 00840682109260SEE15012845GA SEE15012846GA 00840682109260SEE15012846GA SEE15012847GA 00840682109260SEE15012847GA SEE15012848GA 00840682109260SEE15012848GA SEE15012849GA 00840682109260SEE15012849GA SEE15012851GA 00840682109260SEE15012851GA SEE15012852GA 00840682109260SEE15012852GA SEE15012853GA 00840682109260SEE15012853GA SEE15012854GA 00840682109260SEE15012854GA SEE15012855GA 00840682109260SEE15012855GA SEE15012856GA 00840682109260SEE15012856GA SEE15012858GA 00840682109260SEE15012858GA SEE15022859GA 00840682109260SEE15022859GA SEE15022860GA 00840682109260SEE15022860GA SEE15022861GA 00840682109260SEE15022861GA SEE15022862GA 00840682109260SEE15022862GA SEE15022863GA 00840682109260SEE15022863GA SEE15022864GA 00840682109260SEE15022864GA SEE15022865GA 00840682109260SEE15022865GA SEE15022866GA 00840682109260SEE15022866GA SEE15022867GA 00840682109260SEE15022867GA SEE15022868GA 00840682109260SEE15022868GA SEE15032870GA 00840682109260SEE15032870GA SEE15032872GA 00840682109260SEE15032872GA SEE15032873GA 00840682109260SEE15032873GA SEE15032874GA 00840682109260SEE15032874GA SEE15032875GA 00840682109260SEE15032875GA SEE15032876GA 00840682109260SEE15032876GA SEE15032877GA 00840682109260SEE15032877GA SEE15032878GA 00840682109260SEE15032878GA SEE15032880GA 00840682109260SEE15032880GA SEE15032881GA 00840682109260SEE15032881GA SEE15032882GA 00840682109260SEE15032882GA SEE15032883GA 00840682109260SEE15032883GA SEE15032885GA 00840682109260SEE15032885GA SEE15032888GA 00840682109260SEE15032888GA SEE15032889GA 00840682109260SEE15032889GA SEE15032890GA 00840682109260SEE15032890GA SEE15032891GA 00840682109260SEE15032891GA SEE15032892GA 00840682109260SEE15032892GA SEE15032893GA 00840682109260SEE15032893GA SEE15032894GA 00840682109260SEE15032894GA SEE15032895GA 00840682109260SEE15032895GA SEE15032896GA 00840682109260SEE15032896GA SEE15032897GA 00840682109260SEE15032897GA SEE15032898GA 00840682109260SEE15032898GA SEE15032899GA 00840682109260SEE15032899GA SEE15042900GA 00840682109260SEE15042900GA SEE15042901GA 00840682109260SEE15042901GA SEE15042902GA 00840682109260SEE15042902GA SEE15042903GA 00840682109260SEE15042903GA SEE15042904GA 00840682109260SEE15042904GA SEE15042905GA 00840682109260SEE15042905GA SEE15042906GA 00840682109260SEE15042906GA SEE15042907GA 00840682109260SEE15042907GA SEE15052908GA 00840682109260SEE15052908GA SEE15052909GA 00840682109260SEE15052909GA SEE15052910GA 00840682109260SEE15052910GA SEE15052911GA 00840682109260SEE15052911GA SEE15052913GA 00840682109260SEE15052913GA SEE15052914GA 00840682109260SEE15052914GA SEE15052915GA 00840682109260SEE15052915GA SEE15052916GA 00840682109260SEE15052916GA SEE15052917GA 00840682109260SEE15052917GA SEE15052923GA 00840682109260SEE15052923GA SEE15052924GA 00840682109260SEE15052924GA SEE15052925GA 00840682109260SEE15052925GA SEE15052926GA 00840682109260SEE15052926GA SEE15052928GA 00840682109260SEE15052928GA SEE15052929GA 00840682109260SEE15052929GA SEE15052930GA 00840682109260SEE15052930GA SEE15052931GA 00840682109260SEE15052931GA SEE15052932GA 00840682109260SEE15052932GA SEE15052933GA 00840682109260SEE15052933GA SEE15052934GA 00840682109260SEE15052934GA SEE15052935GA 00840682109260SEE15052935GA SEE15052936GA 00840682109260SEE15052936GA SEE15052937GA 00840682109260SEE15052937GA SEE15052938GA 00840682109260SEE15052938GA SEE15052941GA 00840682109260SEE15052941GA SEE15052942GA 00840682109260SEE15052942GA SEE15052944GA 00840682109260SEE15052944GA SEE15052945GA 00840682109260SEE15052945GA SEE15052947GA 00840682109260SEE15052947GA SEE15052948GA 00840682109260SEE15052948GA SEE15052949GA 00840682109260SEE15052949GA SEE15052951GA 00840682109260SEE15052951GA SEE15052952GA 00840682109260SEE15052952GA SEE15053021GA 00840682109260SEE15053021GA SEE15062963GA 00840682109260SEE15062963GA SEE15062966GA 00840682109260SEE15062966GA SEE15062967GA 00840682109260SEE15062967GA SEE15062968GA 00840682109260SEE15062968GA SEE15062969GA 00840682109260SEE15062969GA SEE15062970GA 00840682109260SEE15062970GA SEE15062971GA 00840682109260SEE15062971GA SEE15062973GA 00840682109260SEE15062973GA SEE15062974GA 00840682109260SEE15062974GA SEE15062975GA 00840682109260SEE15062975GA SEE15062976GA 00840682109260SEE15062976GA SEE15062979GA 00840682109260SEE15062979GA SEE15062980GA 00840682109260SEE15062980GA SEE15062981GA 00840682109260SEE15062981GA SEE15062982GA 00840682109260SEE15062982GA SEE15062984GA 00840682109260SEE15062984GA SEE15062985GA 00840682109260SEE15062985GA SEE15062986GA 00840682109260SEE15062986GA SEE15062987GA 00840682109260SEE15062987GA SEE15062988GA 00840682109260SEE15062988GA SEE15062989GA 00840682109260SEE15062989GA SEE15062990GA 00840682109260SEE15062990GA SEE15062991GA 00840682109260SEE15062991GA SEE15062993GA 00840682109260SEE15062993GA SEE15062994GA 00840682109260SEE15062994GA SEE15062995GA 00840682109260SEE15062995GA SEE15062996GA 00840682109260SEE15062996GA SEE15062997GA 00840682109260SEE15062997GA SEE15062998GA 00840682109260SEE15062998GA SEE15062999GA 00840682109260SEE15062999GA SEE15063000GA 00840682109260SEE15063000GA SEE15063001GA 00840682109260SEE15063001GA SEE15063003GA 00840682109260SEE15063003GA SEE15063004GA 00840682109260SEE15063004GA SEE15063005GA 00840682109260SEE15063005GA SEE15063007GA 00840682109260SEE15063007GA SEE15063008GA 00840682109260SEE15063008GA SEE15063009GA 00840682109260SEE15063009GA SEE15063010GA 00840682109260SEE15063010GA SEE15063011GA 00840682109260SEE15063011GA SEE15063012GA 00840682109260SEE15063012GA SEE15063013GA 00840682109260SEE15063013GA SEE15063014GA 00840682109260SEE15063014GA SEE15063015GA 00840682109260SEE15063015GA SEE15063016GA 00840682109260SEE15063016GA SEE15063017GA 00840682109260SEE15063017GA SEE15063018GA 00840682109260SEE15063018GA SEE15063019GA 00840682109260SEE15063019GA SEE15063020GA 00840682109260SEE15063020GA SEE15063021GA 00840682109260SEE15063021GA SEE15063022GA 00840682109260SEE15063022GA SEE15073023GA 00840682109260SEE15073023GA SEE15093024GA 00840682109260SEE15093024GA SEE15093025GA 00840682109260SEE15093025GA SEE15093026GA 00840682109260SEE15093026GA SEE15093027GA 00840682109260SEE15093027GA SEE15093028GA 00840682109260SEE15093028GA SEE15093030GA 00840682109260SEE15093030GA SEE15093031GA 00840682109260SEE15093031GA SEE15093032GA 00840682109260SEE15093032GA SEE15093033GA 00840682109260SEE15093033GA SEE15093034GA 00840682109260SEE15093034GA SEE15093035GA 00840682109260SEE15093035GA SEE15093036GA 00840682109260SEE15093036GA SEE15093037GA 00840682109260SEE15093037GA SEE15093039GA 00840682109260SEE15093039GA SEE15093040GA 00840682109260SEE15093040GA SEE15093041GA 00840682109260SEE15093041GA SEE15093042GA 00840682109260SEE15093042GA SEE15093043GA 00840682109260SEE15093043GA SEE15093044GA 00840682109260SEE15093044GA SEE15093046GA 00840682109260SEE15093046GA SEE15093047GA 00840682109260SEE15093047GA SEE15093050GA 00840682109260SEE15093050GA SEE15093052GA 00840682109260SEE15093052GA SEE15093059GA 00840682109260SEE15093059GA SEE15093062GA 00840682109260SEE15093062GA SEE15093063GA 00840682109260SEE15093063GA SEE15093064GA 00840682109260SEE15093064GA SEE15093065GA 00840682109260SEE15093065GA SEE15093066GA 00840682109260SEE15093066GA SEE15093067GA 00840682109260SEE15093067GA SEE15093068GA 00840682109260SEE15093068GA SEE15093069GA 00840682109260SEE15093069GA SEE15093070GA 00840682109260SEE15093070GA SEE15093071GA 00840682109260SEE15093071GA SEE15093072GA 00840682109260SEE15093072GA SEE15093073GA 00840682109260SEE15093073GA SEE15093074GA 00840682109260SEE15093074GA SEE15093075GA 00840682109260SEE15093075GA SEE15103076GA 00840682109260SEE15103076GA SEE15103077GA 00840682109260SEE15103077GA SEE15103078GA 00840682109260SEE15103078GA SEE15103079GA 00840682109260SEE15103079GA SEE15103080GA 00840682109260SEE15103080GA SEE15103081GA 00840682109260SEE15103081GA SEE15103082GA 00840682109260SEE15103082GA SEE15103083GA 00840682109260SEE15103083GA SEE15103086GA 00840682109260SEE15103086GA SEE15103087GA 00840682109260SEE15103087GA SEE15103088GA 00840682109260SEE15103088GA SEE15103089GA 00840682109260SEE15103089GA SEE15103090GA 00840682109260SEE15103090GA SEE15103091GA 00840682109260SEE15103091GA SEE15103092GA 00840682109260SEE15103092GA SEE15103093GA 00840682109260SEE15103093GA SEE15103094GA 00840682109260SEE15103094GA SEE15103095GA 00840682109260SEE15103095GA SEE15103096GA 00840682109260SEE15103096GA SEE15103097GA 00840682109260SEE15103097GA SEE15103098GA 00840682109260SEE15103098GA SEE15103099GA 00840682109260SEE15103099GA SEE15103100GA 00840682109260SEE15103100GA SEE15103101GA 00840682109260SEE15103101GA SEE15103102GA 00840682109260SEE15103102GA SEE15103103GA 00840682109260SEE15103103GA SEE15103104GA 00840682109260SEE15103104GA SEE15103105GA 00840682109260SEE15103105GA SEE15103106GA 00840682109260SEE15103106GA SEE15103107GA 00840682109260SEE15103107GA SEE15103108GA 00840682109260SEE15103108GA SEE15103110GA 00840682109260SEE15103110GA SEE15103111GA 00840682109260SEE15103111GA SEE15103112GA 00840682109260SEE15103112GA SEE15103113GA 00840682109260SEE15103113GA SEE15103114GA 00840682109260SEE15103114GA SEE15103115GA 00840682109260SEE15103115GA SEE15113116GA 00840682109260SEE15113116GA SEE15113118GA 00840682109260SEE15113118GA SEE15113119GA 00840682109260SEE15113119GA SEE15113120GA 00840682109260SEE15113120GA SEE15113121GA 00840682109260SEE15113121GA SEE15113122GA 00840682109260SEE15113122GA SEE15113123GA 00840682109260SEE15113123GA SEE15113124GA 00840682109260SEE15113124GA SEE15113125GA 00840682109260SEE15113125GA SEE15113127GA 00840682109260SEE15113127GA SEE15113128GA 00840682109260SEE15113128GA SEE15113129GA 00840682109260SEE15113129GA SEE15113130GA 00840682109260SEE15113130GA SEE15113131GA 00840682109260SEE15113131GA SEE15113132GA 00840682109260SEE15113132GA SEE15113133GA 00840682109260SEE15113133GA SEE15113134GA 00840682109260SEE15113134GA SEE15113135GA 00840682109260SEE15113135GA SEE15113136GA 00840682109260SEE15113136GA SEE15113137GA 00840682109260SEE15113137GA SEE15113138GA 00840682109260SEE15113138GA SEE15113139GA 00840682109260SEE15113139GA SEE15113140GA 00840682109260SEE15113140GA SEE15113141GA 00840682109260SEE15113141GA SEE15113142GA 00840682109260SEE15113142GA SEE15113143GA 00840682109260SEE15113143GA SEE15113144GA 00840682109260SEE15113144GA SEE15113145GA 00840682109260SEE15113145GA SEE15113146GA 00840682109260SEE15113146GA SEE15113148GA 00840682109260SEE15113148GA SEE15113150GA 00840682109260SEE15113150GA SEE15113151GA 00840682109260SEE15113151GA SEE15113152GA 00840682109260SEE15113152GA SEE15113153GA 00840682109260SEE15113153GA SEE15113154GA 00840682109260SEE15113154GA SEE15113156GA 00840682109260SEE15113156GA SEE15113157GA 00840682109260SEE15113157GA SEE15113158GA 00840682109260SEE15113158GA SEE15113160GA 00840682109260SEE15113160GA SEE15113162GA 00840682109260SEE15113162GA SEE15113164GA 00840682109260SEE15113164GA SEE15113165GA 00840682109260SEE15113165GA SEE15113166GA 00840682109260SEE15113166GA SEE15113168GA 00840682109260SEE15113168GA SEE15113169GA 00840682109260SEE15113169GA SEE15113170GA 00840682109260SEE15113170GA SEE15113172GA 00840682109260SEE15113172GA SEE15113173GA 00840682109260SEE15113173GA SEE15113175GA 00840682109260SEE15113175GA SEE15113176GA 00840682109260SEE15113176GA SEE15113178GA 00840682109260SEE15113178GA SEE15113179GA 00840682109260SEE15113179GA SEE15113180GA 00840682109260SEE15113180GA SEE15113181GA 00840682109260SEE15113181GA SEE15113182GA 00840682109260SEE15113182GA SEE15113184GA 00840682109260SEE15113184GA SEE15113185GA 00840682109260SEE15113185GA SEE15120553SA 00840682109260SEE15120553SA SEE15120554SA 00840682109260SEE15120554SA SEE15120555SA 00840682109260SEE15120555SA SEE15120556SA 00840682109260SEE15120556SA SEE15120557SA 00840682109260SEE15120557SA SEE15120558SA 00840682109260SEE15120558SA SEE15153186GA 00840682109260SEE15153186GA SEE15153187GA 00840682109260SEE15153187GA SEE15153188GA 00840682109260SEE15153188GA SEE15153190GA 00840682109260SEE15153190GA SEE15153191GA 00840682109260SEE15153191GA SEE15153192GA 00840682109260SEE15153192GA SEE15153193GA 00840682109260SEE15153193GA SEE15153194GA 00840682109260SEE15153194GA SEE15153195GA 00840682109260SEE15153195GA SEE15153196GA 00840682109260SEE15153196GA SEE15153197GA 00840682109260SEE15153197GA SEE15153198GA 00840682109260SEE15153198GA SEE15153201GA 00840682109260SEE15153201GA SEE15153202GA 00840682109260SEE15153202GA SEE15153205GA 00840682109260SEE15153205GA SEE15173206GA 00840682109260SEE15173206GA SEE15173207GA 00840682109260SEE15173207GA SEE15173208GA 00840682109260SEE15173208GA SEE15173209GA 00840682109260SEE15173209GA SEE15173210GA 00840682109260SEE15173210GA SEE15173211GA 00840682109260SEE15173211GA SEE15173212GA 00840682109260SEE15173212GA SEE15173213GA 00840682109260SEE15173213GA SEE15173214GA 00840682109260SEE15173214GA SEE15173215GA 00840682109260SEE15173215GA SEE15173216GA 00840682109260SEE15173216GA SEE15173217GA 00840682109260SEE15173217GA SEE15173220GA 00840682109260SEE15173220GA SEE15173221GA 00840682109260SEE15173221GA SEE15173222GA 00840682109260SEE15173222GA SEE15173223GA 00840682109260SEE15173223GA SEE15173224GA 00840682109260SEE15173224GA SEE15173225GA 00840682109260SEE15173225GA SEE15173226GA 00840682109260SEE15173226GA SEE15173227GA 00840682109260SEE15173227GA SEE15173228GA 00840682109260SEE15173228GA SEE15173229GA 00840682109260SEE15173229GA SEE15173236GA 00840682109260SEE15173236GA SEE15173238GA 00840682109260SEE15173238GA SEE15173241GA 00840682109260SEE15173241GA SEE15173242GA 00840682109260SEE15173242GA SEE15173243GA 00840682109260SEE15173243GA SEE15173246GA 00840682109260SEE15173246GA SEE15183247GA 00840682109260SEE15183247GA SEE15183248GA 00840682109260SEE15183248GA SEE15183249GA 00840682109260SEE15183249GA SEE15183252GA 00840682109260SEE15183252GA SEE15183254GA 00840682109260SEE15183254GA SEE15183255GA 00840682109260SEE15183255GA SEE15183257GA 00840682109260SEE15183257GA SEE15183260GA 00840682109260SEE15183260GA SEE15183261GA 00840682109260SEE15183261GA SEE15193268GA 00840682109260SEE15193268GA SEE15193271GA 00840682109260SEE15193271GA SEE15193276GA 00840682109260SEE15193276GA SEE15193279GA 00840682109260SEE15193279GA SEE15193286GA 00840682109260SEE15193286GA SEE15193287GA 00840682109260SEE15193287GA SEE15193288GA 00840682109260SEE15193288GA SEE15193289GA 00840682109260SEE15193289GA SEE15193290GA 00840682109260SEE15193290GA SEE15193291GA 00840682109260SEE15193291GA SEE15193292GA 00840682109260SEE15193292GA SEE15193295GA 00840682109260SEE15193295GA SEE15193297GA 00840682109260SEE15193297GA SEE15193302GA 00840682109260SEE15193302GA SEE15193303GA 00840682109260SEE15193303GA SEE15193305GA 00840682109260SEE15193305GA SEE15210698SA 00840682109260SEE15210698SA SEE15210699SA 00840682109260SEE15210699SA SEE15210700SA 00840682109260SEE15210700SA SEE15210701SA 00840682109260SEE15210701SA SEE15210702SA 00840682109260SEE15210702SA SEE15210703SA 00840682109260SEE15210703SA SEE15210705SA 00840682109260SEE15210705SA SEE15220768SA 00840682109260SEE15220768SA SEE15220769SA 00840682109260SEE15220769SA SEE15220770SA 00840682109260SEE15220770SA SEE15220771SA 00840682109260SEE15220771SA SEE15220772SA 00840682109260SEE15220772SA SEE15220773SA 00840682109260SEE15220773SA SEE15220774SA 00840682109260SEE15220774SA SEE15220776SA 00840682109260SEE15220776SA SEE15220777SA 00840682109260SEE15220777SA SEE15220778SA 00840682109260SEE15220778SA SEE15220779SA 00840682109260SEE15220779SA SEE15220780SA 00840682109260SEE15220780SA SEE15220781SA 00840682109260SEE15220781SA SEE15220782SA 00840682109260SEE15220782SA SEE15220783SA 00840682109260SEE15220783SA SEE15220784SA 00840682109260SEE15220784SA SEE15220785SA 00840682109260SEE15220785SA SEE15220791SA 00840682109260SEE15220791SA SEE15220795SA 00840682109260SEE15220795SA SEE15220796SA 00840682109260SEE15220796SA SEE15220797SA 00840682109260SEE15220797SA SEE15220798SA 00840682109260SEE15220798SA SEE15230837SA 00840682109260SEE15230837SA SEE15230839SA 00840682109260SEE15230839SA SEE15240989SA 00840682109260SEE15240989SA SEE15240990SA 00840682109260SEE15240990SA SEE15240991SA 00840682109260SEE15240991SA SEE15241002SA 00840682109260SEE15241002SA SEE15241003SA 00840682109260SEE15241003SA SEE15241004SA 00840682109260SEE15241004SA SEE15241005SA 00840682109260SEE15241005SA SEE15241006SA 00840682109260SEE15241006SA SEE15241007SA 00840682109260SEE15241007SA SEE15251188SA 00840682109260SEE15251188SA SEE15251241SA 00840682109260SEE15251241SA SEE15251286SA 00840682109260SEE15251286SA SEE15251287SA 00840682109260SEE15251287SA SEE15251288SA 00840682109260SEE15251288SA SEE15251289SA 00840682109260SEE15251289SA SEE15251290SA 00840682109260SEE15251290SA SEE15251291SA 00840682109260SEE15251291SA SEE15251298SA 00840682109260SEE15251298SA SEE15251299SA 00840682109260SEE15251299SA SEE15251300SA 00840682109260SEE15251300SA SEE15251301SA 00840682109260SEE15251301SA SEE15251302SA 00840682109260SEE15251302SA SEE15251303SA 00840682109260SEE15251303SA SEE15261392SA 00840682109260SEE15261392SA SEE15261393SA 00840682109260SEE15261393SA SEE15261394SA 00840682109260SEE15261394SA SEE15261395SA 00840682109260SEE15261395SA SEE15261396SA 00840682109260SEE15261396SA SEE15261397SA 00840682109260SEE15261397SA SEE15261398SA 00840682109260SEE15261398SA SEE15261399SA 00840682109260SEE15261399SA SEE15261400SA 00840682109260SEE15261400SA SEE15261401SA 00840682109260SEE15261401SA SEE15261402SA 00840682109260SEE15261402SA SEE15261403SA 00840682109260SEE15261403SA SEE15261404SA 00840682109260SEE15261404SA SEE15261405SA 00840682109260SEE15261405SA SEE15261406SA 00840682109260SEE15261406SA SEE15261407SA 00840682109260SEE15261407SA SEE15261408SA 00840682109260SEE15261408SA SEE15261409SA 00840682109260SEE15261409SA SEE15261511SA 00840682109260SEE15261511SA SEE15261530SA 00840682109260SEE15261530SA SEE15261531SA 00840682109260SEE15261531SA SEE15261532SA 00840682109260SEE15261532SA SEE15261533SA 00840682109260SEE15261533SA SEE15261534SA 00840682109260SEE15261534SA SEE15261536SA 00840682109260SEE15261536SA SEE15261537SA 00840682109260SEE15261537SA SEE15261538SA 00840682109260SEE15261538SA SEE15261539SA 00840682109260SEE15261539SA SEE15261540SA 00840682109260SEE15261540SA SEE15261541SA 00840682109260SEE15261541SA SEE15261542SA 00840682109260SEE15261542SA SEE15261545SA 00840682109260SEE15261545SA SEE15261546SA 00840682109260SEE15261546SA SEE15261547SA 00840682109260SEE15261547SA SEE15261548SA 00840682109260SEE15261548SA SEE15261549SA 00840682109260SEE15261549SA SEE15261550SA 00840682109260SEE15261550SA SEE15261551SA 00840682109260SEE15261551SA SEE15261552SA 00840682109260SEE15261552SA SEE15261553SA 00840682109260SEE15261553SA SEE15261555SA 00840682109260SEE15261555SA SEE15261556SA 00840682109260SEE15261556SA SEE15261557SA 00840682109260SEE15261557SA SEE15261558SA 00840682109260SEE15261558SA SEE15261579SA 00840682109260SEE15261579SA SEE15261581SA 00840682109260SEE15261581SA SEE15281638SA 00840682109260SEE15281638SA SEE15281639SA 00840682109260SEE15281639SA SEE15281640SA 00840682109260SEE15281640SA SEE15281641SA 00840682109260SEE15281641SA SEE15281648SA 00840682109260SEE15281648SA SEE15281659SA 00840682109260SEE15281659SA SEE15291745SA 00840682109260SEE15291745SA SEE15291766SA 00840682109260SEE15291766SA SEE15291767SA 00840682109260SEE15291767SA SEE15291799SA 00840682109260SEE15291799SA SEE15291869SA 00840682109260SEE15291869SA SEE15301887SA 00840682109260SEE15301887SA SEE15301888SA 00840682109260SEE15301888SA SEE15301889SA 00840682109260SEE15301889SA SEE15301890SA 00840682109260SEE15301890SA SEE15301891SA 00840682109260SEE15301891SA SEE15301928SA 00840682109260SEE15301928SA SEE15301929SA 00840682109260SEE15301929SA SEE15301930SA 00840682109260SEE15301930SA SEE15301931SA 00840682109260SEE15301931SA SEE15301932SA 00840682109260SEE15301932SA SEE15301933SA 00840682109260SEE15301933SA SEE15301948SA 00840682109260SEE15301948SA SEE15312074SA 00840682109260SEE15312074SA SEE15312075SA 00840682109260SEE15312075SA SEE15312076SA 00840682109260SEE15312076SA SEE15312077SA 00840682109260SEE15312077SA SEE15312078SA 00840682109260SEE15312078SA SEE15312079SA 00840682109260SEE15312079SA SEE15312080SA 00840682109260SEE15312080SA SEE15312081SA 00840682109260SEE15312081SA SEE15312082SA 00840682109260SEE15312082SA SEE15312083SA 00840682109260SEE15312083SA SEE15312084SA 00840682109260SEE15312084SA SEE15312085SA 00840682109260SEE15312085SA SEE15312086SA 00840682109260SEE15312086SA SEE15312099SA 00840682109260SEE15312099SA SEE15312100SA 00840682109260SEE15312100SA SEE15312120SA 00840682109260SEE15312120SA SEE15312121SA 00840682109260SEE15312121SA SEE15312127SA 00840682109260SEE15312127SA SEE15312131SA 00840682109260SEE15312131SA SEE15312135SA 00840682109260SEE15312135SA SEE15312141SA 00840682109260SEE15312141SA SEE15322281SA 00840682109260SEE15322281SA SEE15322282SA 00840682109260SEE15322282SA SEE15322283SA 00840682109260SEE15322283SA SEE15322284SA 00840682109260SEE15322284SA SEE15322285SA 00840682109260SEE15322285SA SEE15332291SA 00840682109260SEE15332291SA SEE15332292SA 00840682109260SEE15332292SA SEE15332293SA 00840682109260SEE15332293SA SEE15332294SA 00840682109260SEE15332294SA SEE15332295SA 00840682109260SEE15332295SA SEE15332306SA 00840682109260SEE15332306SA SEE15332307SA 00840682109260SEE15332307SA SEE15332308SA 00840682109260SEE15332308SA SEE15332309SA 00840682109260SEE15332309SA SEE15332310SA 00840682109260SEE15332310SA SEE15332311SA 00840682109260SEE15332311SA SEE15332312SA 00840682109260SEE15332312SA SEE15332313SA 00840682109260SEE15332313SA SEE15332314SA 00840682109260SEE15332314SA SEE15332315SA 00840682109260SEE15332315SA SEE15332339SA 00840682109260SEE15332339SA SEE15332340SA 00840682109260SEE15332340SA SEE15332341SA 00840682109260SEE15332341SA SEE15332342SA 00840682109260SEE15332342SA SEE15332398SA 00840682109260SEE15332398SA SEE15332399SA 00840682109260SEE15332399SA SEE15332400SA 00840682109260SEE15332400SA SEE15332401SA 00840682109260SEE15332401SA SEE15342446SA 00840682109260SEE15342446SA SEE15342480SA 00840682109260SEE15342480SA SEE15342482SA 00840682109260SEE15342482SA SEE15342483SA 00840682109260SEE15342483SA SEE15352537SA 00840682109260SEE15352537SA SEE15352538SA 00840682109260SEE15352538SA SEE15352545SA 00840682109260SEE15352545SA SEE15352546SA 00840682109260SEE15352546SA SEE15352553SA 00840682109260SEE15352553SA SEE15352554SA 00840682109260SEE15352554SA SEE15352559SA 00840682109260SEE15352559SA SEE15352560SA 00840682109260SEE15352560SA SEE15352563SA 00840682109260SEE15352563SA SEE15352564SA 00840682109260SEE15352564SA SEE15352667SA 00840682109260SEE15352667SA SEE15352668SA 00840682109260SEE15352668SA SEE15352669SA 00840682109260SEE15352669SA SEE15352670SA 00840682109260SEE15352670SA SEE15352677SA 00840682109260SEE15352677SA SEE15352678SA 00840682109260SEE15352678SA SEE15352698SA 00840682109260SEE15352698SA SEE15352700SA 00840682109260SEE15352700SA SEE15352786SA 00840682109260SEE15352786SA SEE15352787SA 00840682109260SEE15352787SA SEE15352788SA 00840682109260SEE15352788SA SEE15352789SA 00840682109260SEE15352789SA SEE15352790SA 00840682109260SEE15352790SA SEE15352791SA 00840682109260SEE15352791SA SEE15352792SA 00840682109260SEE15352792SA SEE15352793SA 00840682109260SEE15352793SA SEE15352794SA 00840682109260SEE15352794SA SEE15352795SA 00840682109260SEE15352795SA SEE15352796SA 00840682109260SEE15352796SA SEE15352844SA 00840682109260SEE15352844SA SEE15352845SA 00840682109260SEE15352845SA SEE15352846SA 00840682109260SEE15352846SA SEE15353099SA 00840682109260SEE15353099SA SEE15353100SA 00840682109260SEE15353100SA SEE15353119SA 00840682109260SEE15353119SA SEE15363354SA 00840682109260SEE15363354SA SEE15363355SA 00840682109260SEE15363355SA SEE15373422SA 00840682109260SEE15373422SA SEE15373437SA 00840682109260SEE15373437SA SEE15373462SA 00840682109260SEE15373462SA SEE15373463SA 00840682109260SEE15373463SA SEE15373464SA 00840682109260SEE15373464SA SEE15373465SA 00840682109260SEE15373465SA SEE15373466SA 00840682109260SEE15373466SA SEE15373467SA 00840682109260SEE15373467SA SEE15373531SA 00840682109260SEE15373531SA SEE15373532SA 00840682109260SEE15373532SA SEE15373617SA 00840682109260SEE15373617SA SEE15383635SA 00840682109260SEE15383635SA SEE15383636SA 00840682109260SEE15383636SA SEE15383637SA 00840682109260SEE15383637SA SEE15383769SA 00840682109260SEE15383769SA SEE15383875SA 00840682109260SEE15383875SA SEE15383876SA 00840682109260SEE15383876SA SEE15394277SA 00840682109260SEE15394277SA SEE15394284SA 00840682109260SEE15394284SA SEE15394375SA 00840682109260SEE15394375SA SEE15394376SA 00840682109260SEE15394376SA SEE15394426SA 00840682109260SEE15394426SA SEE15394716SA 00840682109260SEE15394716SA SEE15394717SA 00840682109260SEE15394717SA SEE15394718SA 00840682109260SEE15394718SA SEE15394719SA 00840682109260SEE15394719SA SEE15394729SA 00840682109260SEE15394729SA SEE15394730SA 00840682109260SEE15394730SA SEE15394732SA 00840682109260SEE15394732SA SEE15394741SA 00840682109260SEE15394741SA SEE15394742SA 00840682109260SEE15394742SA SEE15394743SA 00840682109260SEE15394743SA SEE15405016SA 00840682109260SEE15405016SA SEE15405017SA 00840682109260SEE15405017SA SEE15405018SA 00840682109260SEE15405018SA SEE15405019SA 00840682109260SEE15405019SA SEE15405060SA 00840682109260SEE15405060SA SEE15405088SA 00840682109260SEE15405088SA SEE15415093SA 00840682109260SEE15415093SA SEE15415094SA 00840682109260SEE15415094SA SEE15415110SA 00840682109260SEE15415110SA SEE15415111SA 00840682109260SEE15415111SA SEE15415112SA 00840682109260SEE15415112SA SEE15415113SA 00840682109260SEE15415113SA SEE15415114SA 00840682109260SEE15415114SA SEE15415115SA 00840682109260SEE15415115SA SEE15415116SA 00840682109260SEE15415116SA SEE15415117SA 00840682109260SEE15415117SA SEE15415118SA 00840682109260SEE15415118SA SEE15415164SA 00840682109260SEE15415164SA SEE15415171SA 00840682109260SEE15415171SA SEE15415177SA 00840682109260SEE15415177SA SEE15415183SA 00840682109260SEE15415183SA SEE15415184SA 00840682109260SEE15415184SA SEE15415185SA 00840682109260SEE15415185SA SEE15415186SA 00840682109260SEE15415186SA SEE15415188SA 00840682109260SEE15415188SA SEE15415193SA 00840682109260SEE15415193SA SEE15415194SA 00840682109260SEE15415194SA SEE15415195SA 00840682109260SEE15415195SA SEE15425741SA 00840682109260SEE15425741SA SEE15425780SA 00840682109260SEE15425780SA SEE15425781SA 00840682109260SEE15425781SA SEE15425782SA 00840682109260SEE15425782SA SEE15425786SA 00840682109260SEE15425786SA SEE15425787SA 00840682109260SEE15425787SA SEE15425788SA 00840682109260SEE15425788SA SEE15425789SA 00840682109260SEE15425789SA SEE15425790SA 00840682109260SEE15425790SA SEE15425791SA 00840682109260SEE15425791SA SEE15425792SA 00840682109260SEE15425792SA SEE15425793SA 00840682109260SEE15425793SA SEE15425794SA 00840682109260SEE15425794SA SEE15425795SA 00840682109260SEE15425795SA SEE15425796SA 00840682109260SEE15425796SA SEE15425797SA 00840682109260SEE15425797SA SEE15425798SA 00840682109260SEE15425798SA SEE15425799SA 00840682109260SEE15425799SA SEE15425800SA 00840682109260SEE15425800SA SEE15425801SA 00840682109260SEE15425801SA SEE15435919SA 00840682109260SEE15435919SA SEE15435920SA 00840682109260SEE15435920SA SEE15435921SA 00840682109260SEE15435921SA SEE15435922SA 00840682109260SEE15435922SA SEE15435927SA 00840682109260SEE15435927SA SEE15435928SA 00840682109260SEE15435928SA SEE15435933SA 00840682109260SEE15435933SA SEE15435934SA 00840682109260SEE15435934SA SEE15446117SA 00840682109260SEE15446117SA SEE15446119SA 00840682109260SEE15446119SA SEE15446121SA 00840682109260SEE15446121SA SEE15446123SA 00840682109260SEE15446123SA SEE15446124SA 00840682109260SEE15446124SA SEE15446169SA 00840682109260SEE15446169SA SEE15446170SA 00840682109260SEE15446170SA SEE15446199SA 00840682109260SEE15446199SA SEE15446200SA 00840682109260SEE15446200SA SEE15446201SA 00840682109260SEE15446201SA SEE15446202SA 00840682109260SEE15446202SA SEE15446203SA 00840682109260SEE15446203SA SEE15446204SA 00840682109260SEE15446204SA SEE15446213SA 00840682109260SEE15446213SA SEE15446214SA 00840682109260SEE15446214SA SEE15446215SA 00840682109260SEE15446215SA SEE15446216SA 00840682109260SEE15446216SA SEE15446220SA 00840682109260SEE15446220SA SEE15446221SA 00840682109260SEE15446221SA SEE15446222SA 00840682109260SEE15446222SA SEE15446223SA 00840682109260SEE15446223SA SEE15446224SA 00840682109260SEE15446224SA SEE15446225SA 00840682109260SEE15446225SA SEE15446226SA 00840682109260SEE15446226SA SEE15446228SA 00840682109260SEE15446228SA SEE15446229SA 00840682109260SEE15446229SA SEE15446230SA 00840682109260SEE15446230SA SEE15446231SA 00840682109260SEE15446231SA SEE15446232SA 00840682109260SEE15446232SA SEE15446233SA 00840682109260SEE15446233SA SEE15446234SA 00840682109260SEE15446234SA SEE15446235SA 00840682109260SEE15446235SA SEE15446236SA 00840682109260SEE15446236SA SEE15446238SA 00840682109260SEE15446238SA SEE15446239SA 00840682109260SEE15446239SA SEE15446240SA 00840682109260SEE15446240SA SEE15446241SA 00840682109260SEE15446241SA SEE15446242SA 00840682109260SEE15446242SA SEE15446243SA 00840682109260SEE15446243SA SEE15446244SA 00840682109260SEE15446244SA SEE15446245SA 00840682109260SEE15446245SA SEE15446254SA 00840682109260SEE15446254SA SEE15446255SA 00840682109260SEE15446255SA SEE15446256SA 00840682109260SEE15446256SA SEE15446260SA 00840682109260SEE15446260SA SEE15446261SA 00840682109260SEE15446261SA SEE15446262SA 00840682109260SEE15446262SA SEE15446331SA 00840682109260SEE15446331SA SEE15446332SA 00840682109260SEE15446332SA SEE15446333SA 00840682109260SEE15446333SA SEE15446334SA 00840682109260SEE15446334SA SEE15446335SA 00840682109260SEE15446335SA SEE15446336SA 00840682109260SEE15446336SA SEE15456708SA 00840682109260SEE15456708SA SEE15456710SA 00840682109260SEE15456710SA SEE15456712SA 00840682109260SEE15456712SA SEE15456714SA 00840682109260SEE15456714SA SEE15456716SA 00840682109260SEE15456716SA SEE15456718SA 00840682109260SEE15456718SA SEE15456720SA 00840682109260SEE15456720SA SEE15456722SA 00840682109260SEE15456722SA SEE15456723SA 00840682109260SEE15456723SA SEE15456724SA 00840682109260SEE15456724SA SEE15456725SA 00840682109260SEE15456725SA SEE15456727SA 00840682109260SEE15456727SA SEE15456728SA 00840682109260SEE15456728SA SEE15456729SA 00840682109260SEE15456729SA SEE15456730SA 00840682109260SEE15456730SA SEE15466917SA 00840682109260SEE15466917SA SEE15466918SA 00840682109260SEE15466918SA SEE15466919SA 00840682109260SEE15466919SA SEE15466921SA 00840682109260SEE15466921SA SEE15466922SA 00840682109260SEE15466922SA SEE15466923SA 00840682109260SEE15466923SA SEE15466924SA 00840682109260SEE15466924SA SEE15466925SA 00840682109260SEE15466925SA SEE15466926SA 00840682109260SEE15466926SA SEE15466927SA 00840682109260SEE15466927SA SEE15466928SA 00840682109260SEE15466928SA SEE15466929SA 00840682109260SEE15466929SA SEE15466930SA 00840682109260SEE15466930SA SEE15466931SA 00840682109260SEE15466931SA SEE15466932SA 00840682109260SEE15466932SA SEE15466933SA 00840682109260SEE15466933SA SEE15466934SA 00840682109260SEE15466934SA SEE15466935SA 00840682109260SEE15466935SA SEE15466936SA 00840682109260SEE15466936SA SEE15467025SA 00840682109260SEE15467025SA SEE15467026SA 00840682109260SEE15467026SA SEE15467027SA 00840682109260SEE15467027SA SEE15467028SA 00840682109260SEE15467028SA SEE15477106SA 00840682109260SEE15477106SA SEE15477107SA 00840682109260SEE15477107SA SEE15477108SA 00840682109260SEE15477108SA SEE15477109SA 00840682109260SEE15477109SA SEE15477110SA 00840682109260SEE15477110SA SEE15477111SA 00840682109260SEE15477111SA SEE15477112SA 00840682109260SEE15477112SA SEE15477113SA 00840682109260SEE15477113SA SEE15477115SA 00840682109260SEE15477115SA SEE15477175SA 00840682109260SEE15477175SA SEE15477181SA 00840682109260SEE15477181SA SEE15477182SA 00840682109260SEE15477182SA SEE15477183SA 00840682109260SEE15477183SA SEE15477184SA 00840682109260SEE15477184SA SEE15487295SA 00840682109260SEE15487295SA SEE15497498SA 00840682109260SEE15497498SA SEE15497500SA 00840682109260SEE15497500SA SEE15497501SA 00840682109260SEE15497501SA SEE15497510SA 00840682109260SEE15497510SA SEE15497511SA 00840682109260SEE15497511SA SEE15497512SA 00840682109260SEE15497512SA SEE15497513SA 00840682109260SEE15497513SA SEE15497514SA 00840682109260SEE15497514SA SEE15497515SA 00840682109260SEE15497515SA SEE15497516SA 00840682109260SEE15497516SA SEE15497517SA 00840682109260SEE15497517SA SEE15497518SA 00840682109260SEE15497518SA SEE15497519SA 00840682109260SEE15497519SA SEE15497520SA 00840682109260SEE15497520SA SEE15497521SA 00840682109260SEE15497521SA SEE15497522SA 00840682109260SEE15497522SA SEE15497523SA 00840682109260SEE15497523SA SEE15497524SA 00840682109260SEE15497524SA SEE15497525SA 00840682109260SEE15497525SA SEE15497526SA 00840682109260SEE15497526SA SEE15497527SA 00840682109260SEE15497527SA SEE15497528SA 00840682109260SEE15497528SA SEE15497529SA 00840682109260SEE15497529SA SEE15497530SA 00840682109260SEE15497530SA SEE15497531SA 00840682109260SEE15497531SA SEE15497532SA 00840682109260SEE15497532SA SEE15497533SA 00840682109260SEE15497533SA SEE15497534SA 00840682109260SEE15497534SA SEE15497535SA 00840682109260SEE15497535SA SEE15497579SA 00840682109260SEE15497579SA SEE15497580SA 00840682109260SEE15497580SA SEE15497581SA 00840682109260SEE15497581SA SEE15497582SA 00840682109260SEE15497582SA SEE15497583SA 00840682109260SEE15497583SA SEE15497584SA 00840682109260SEE15497584SA SEE15497585SA 00840682109260SEE15497585SA SEE15497586SA 00840682109260SEE15497586SA SEE15497587SA 00840682109260SEE15497587SA SEE15497588SA 00840682109260SEE15497588SA SEE15507658SA 00840682109260SEE15507658SA SEE15507659SA 00840682109260SEE15507659SA SEE15507660SA 00840682109260SEE15507660SA SEE15507661SA 00840682109260SEE15507661SA SEE15507662SA 00840682109260SEE15507662SA SEE15507663SA 00840682109260SEE15507663SA SEE15507700SA 00840682109260SEE15507700SA SEE15507701SA 00840682109260SEE15507701SA SEE15507702SA 00840682109260SEE15507702SA SEE15507703SA 00840682109260SEE15507703SA SEE15507704SA 00840682109260SEE15507704SA SEE15507710SA 00840682109260SEE15507710SA SEE15507711SA 00840682109260SEE15507711SA SEE15517769SA 00840682109260SEE15517769SA SEE15517770SA 00840682109260SEE15517770SA SEE15517771SA 00840682109260SEE15517771SA SEE15517772SA 00840682109260SEE15517772SA SEE15517773SA 00840682109260SEE15517773SA SEE15517774SA 00840682109260SEE15517774SA SEE15517775SA 00840682109260SEE15517775SA SEE15517776SA 00840682109260SEE15517776SA SEE15517777SA 00840682109260SEE15517777SA SEE15517782SA 00840682109260SEE15517782SA SEE15517817SA 00840682109260SEE15517817SA SEE15517818SA 00840682109260SEE15517818SA SEE15517819SA 00840682109260SEE15517819SA SEE15517820SA 00840682109260SEE15517820SA SEE15517821SA 00840682109260SEE15517821SA SEE15517822SA 00840682109260SEE15517822SA SEE15517823SA 00840682109260SEE15517823SA SEE15517824SA 00840682109260SEE15517824SA SEE15517888SA 00840682109260SEE15517888SA SEE15517889SA 00840682109260SEE15517889SA SEE15518040SA 00840682109260SEE15518040SA SEE15518041SA 00840682109260SEE15518041SA SEE15518042SA 00840682109260SEE15518042SA SEE15518043SA 00840682109260SEE15518043SA SEE15518044SA 00840682109260SEE15518044SA SEE15518045SA 00840682109260SEE15518045SA SEE15518046SA 00840682109260SEE15518046SA SEE15518047SA 00840682109260SEE15518047SA SEE15528071SA 00840682109260SEE15528071SA SEE15528072SA 00840682109260SEE15528072SA SEE15528073SA 00840682109260SEE15528073SA SEE15528074SA 00840682109260SEE15528074SA SEE15528204SA 00840682109260SEE15528204SA SEE15528205SA 00840682109260SEE15528205SA SEE15528206SA 00840682109260SEE15528206SA SEE15528207SA 00840682109260SEE15528207SA SEE15528212SA 00840682109260SEE15528212SA SEE15528213SA 00840682109260SEE15528213SA SEE15528214SA 00840682109260SEE15528214SA SEE15528215SA 00840682109260SEE15528215SA SEE15528216SA 00840682109260SEE15528216SA SEE15528218SA 00840682109260SEE15528218SA SEE15538297SA 00840682109260SEE15538297SA SEE16028317SA 00840682109260SEE16028317SA SEE16028319SA 00840682109260SEE16028319SA SEE16028320SA 00840682109260SEE16028320SA SEE16028321SA 00840682109260SEE16028321SA SEE16028367SA 00840682109260SEE16028367SA SEE16028368SA 00840682109260SEE16028368SA SEE16028369SA 00840682109260SEE16028369SA SEE16028376SA 00840682109260SEE16028376SA SEE16038425SA 00840682109260SEE16038425SA SEE16038440SA 00840682109260SEE16038440SA SEE16038441SA 00840682109260SEE16038441SA SEE16038442SA 00840682109260SEE16038442SA SEE16038553SA 00840682109260SEE16038553SA SEE16038554SA 00840682109260SEE16038554SA SEE16038555SA 00840682109260SEE16038555SA SEE16038556SA 00840682109260SEE16038556SA SEE16038557SA 00840682109260SEE16038557SA SEE16038562SA 00840682109260SEE16038562SA SEE16058655SA 00840682109260SEE16058655SA SEE16058656SA 00840682109260SEE16058656SA SEE16058657SA 00840682109260SEE16058657SA SEE16058658SA 00840682109260SEE16058658SA SEE16058659SA 00840682109260SEE16058659SA SEE16058660SA 00840682109260SEE16058660SA SEE16058665SA 00840682109260SEE16058665SA SEE16058666SA 00840682109260SEE16058666SA SEE16058667SA 00840682109260SEE16058667SA SEE16058668SA 00840682109260SEE16058668SA SEE16058669SA 00840682109260SEE16058669SA SEE16058670SA 00840682109260SEE16058670SA SEE16058671SA 00840682109260SEE16058671SA SEE16058672SA 00840682109260SEE16058672SA SEE16058673SA 00840682109260SEE16058673SA SEE16058674SA 00840682109260SEE16058674SA SEE16058675SA 00840682109260SEE16058675SA SEE16058676SA 00840682109260SEE16058676SA SEE16058677SA 00840682109260SEE16058677SA SEE16058678SA 00840682109260SEE16058678SA SEE16058679SA 00840682109260SEE16058679SA SEE16058680SA 00840682109260SEE16058680SA SEE16058699SA 00840682109260SEE16058699SA SEE16068741SA 00840682109260SEE16068741SA SEE16068742SA 00840682109260SEE16068742SA SEE16068743SA 00840682109260SEE16068743SA SEE16068744SA 00840682109260SEE16068744SA SEE16068745SA 00840682109260SEE16068745SA SEE16068746SA 00840682109260SEE16068746SA SEE16068747SA 00840682109260SEE16068747SA SEE16068748SA 00840682109260SEE16068748SA SEE16068749SA 00840682109260SEE16068749SA SEE16068832SA 00840682109260SEE16068832SA SEE16078863SA 00840682109260SEE16078863SA SEE16078864SA 00840682109260SEE16078864SA SEE16078865SA 00840682109260SEE16078865SA SEE16078894SA 00840682109260SEE16078894SA SEE16078896SA 00840682109260SEE16078896SA SEE16078897SA 00840682109260SEE16078897SA SEE16078898SA 00840682109260SEE16078898SA SEE16078920SA 00840682109260SEE16078920SA SEE16078924SA 00840682109260SEE16078924SA SEE16078926SA 00840682109260SEE16078926SA SEE16078927SA 00840682109260SEE16078927SA SEE16078929SA 00840682109260SEE16078929SA SEE16078930SA 00840682109260SEE16078930SA SEE16078932SA 00840682109260SEE16078932SA SEE16079148SA 00840682109260SEE16079148SA SEE16089172SA 00840682109260SEE16089172SA SEE16089173SA 00840682109260SEE16089173SA SEE16089175SA 00840682109260SEE16089175SA SEE16089176SA 00840682109260SEE16089176SA SEE16089177SA 00840682109260SEE16089177SA SEE16089178SA 00840682109260SEE16089178SA SEE16089180SA 00840682109260SEE16089180SA SEE16089268SA 00840682109260SEE16089268SA SEE16089277SA 00840682109260SEE16089277SA SEE16089278SA 00840682109260SEE16089278SA SEE16089279SA 00840682109260SEE16089279SA SEE16099454SA 00840682109260SEE16099454SA SEE16099455SA 00840682109260SEE16099455SA SEE16099456SA 00840682109260SEE16099456SA SEE16099458SA 00840682109260SEE16099458SA SEE16099459SA 00840682109260SEE16099459SA SEE16099500SA 00840682109260SEE16099500SA SEE16109578SA 00840682109260SEE16109578SA SEE16109579SA 00840682109260SEE16109579SA SEE16109580SA 00840682109260SEE16109580SA SEE16109581SA 00840682109260SEE16109581SA SEE16109582SA 00840682109260SEE16109582SA SEE16109583SA 00840682109260SEE16109583SA SEE16109584SA 00840682109260SEE16109584SA SEE16109585SA 00840682109260SEE16109585SA SEE16109586SA 00840682109260SEE16109586SA SEE16109587SA 00840682109260SEE16109587SA SEE16109588SA 00840682109260SEE16109588SA SEE16109589SA 00840682109260SEE16109589SA SEE16109590SA 00840682109260SEE16109590SA SEE16109591SA 00840682109260SEE16109591SA SEE16109592SA 00840682109260SEE16109592SA SEE16109593SA 00840682109260SEE16109593SA SEE16109594SA 00840682109260SEE16109594SA SEE16109595SA 00840682109260SEE16109595SA SEE16109690SA 00840682109260SEE16109690SA SEE16109691SA 00840682109260SEE16109691SA SEE16109692SA 00840682109260SEE16109692SA SEE16109693SA 00840682109260SEE16109693SA SEE16109694SA 00840682109260SEE16109694SA SEE16109695SA 00840682109260SEE16109695SA SEE16109699SA 00840682109260SEE16109699SA SEE16109700SA 00840682109260SEE16109700SA SEE16109701SA 00840682109260SEE16109701SA SEE16109702SA 00840682109260SEE16109702SA SEE16109703SA 00840682109260SEE16109703SA SEE16109704SA 00840682109260SEE16109704SA SEE16109705SA 00840682109260SEE16109705SA SEE16119767SA 00840682109260SEE16119767SA SEE16119768SA 00840682109260SEE16119768SA SEE16119844SA 00840682109260SEE16119844SA SEE16119845SA 00840682109260SEE16119845SA SEE16119846SA 00840682109260SEE16119846SA SEE16119850SA 00840682109260SEE16119850SA SEE16119851SA 00840682109260SEE16119851SA SEE16119897SA 00840682109260SEE16119897SA SEE16119898SA 00840682109260SEE16119898SA SEE16119899SA 00840682109260SEE16119899SA SEE16119901SA 00840682109260SEE16119901SA SEE16119903SA 00840682109260SEE16119903SA SEE16119904SA 00840682109260SEE16119904SA SEE16119905SA 00840682109260SEE16119905SA SEE16119906SA 00840682109260SEE16119906SA SEE16119907SA 00840682109260SEE16119907SA SEE16119908SA 00840682109260SEE16119908SA SEE16119909SA 00840682109260SEE16119909SA SEE16119910SA 00840682109260SEE16119910SA SEE16119912SA 00840682109260SEE16119912SA SEE16119913SA 00840682109260SEE16119913SA SEE16119914SA 00840682109260SEE16119914SA SEE16119915SA 00840682109260SEE16119915SA SEE16119916SA 00840682109260SEE16119916SA SEE16119917SA 00840682109260SEE16119917SA SEE16119918SA 00840682109260SEE16119918SA SEE16119920SA 00840682109260SEE16119920SA SEE16119921SA 00840682109260SEE16119921SA SEE16119922SA 00840682109260SEE16119922SA SEE16119923SA 00840682109260SEE16119923SA SEE16119924SA 00840682109260SEE16119924SA SEE16119926SA 00840682109260SEE16119926SA SEE16119927SA 00840682109260SEE16119927SA SEE16119928SA 00840682109260SEE16119928SA SEE16120020SA 00840682109260SEE16120020SA SEE16120021SA 00840682109260SEE16120021SA SEE16120076SA 00840682109260SEE16120076SA SEE16120077SA 00840682109260SEE16120077SA SEE16120170SA 00840682109260SEE16120170SA SEE16120171SA 00840682109260SEE16120171SA SEE16130303SA 00840682109260SEE16130303SA SEE16130304SA 00840682109260SEE16130304SA SEE16130305SA 00840682109260SEE16130305SA SEE16130306SA 00840682109260SEE16130306SA SEE16130320SA 00840682109260SEE16130320SA SEE16130321SA 00840682109260SEE16130321SA SEE16130322SA 00840682109260SEE16130322SA SEE16130323SA 00840682109260SEE16130323SA SEE16130324SA 00840682109260SEE16130324SA SEE16140328SA 00840682109260SEE16140328SA SEE16140329SA 00840682109260SEE16140329SA SEE16140330SA 00840682109260SEE16140330SA SEE16140432SA 00840682109260SEE16140432SA SEE16140433SA 00840682109260SEE16140433SA SEE16140434SA 00840682109260SEE16140434SA SEE16140435SA 00840682109260SEE16140435SA SEE16140436SA 00840682109260SEE16140436SA SEE16150492SA 00840682109260SEE16150492SA SEE16150493SA 00840682109260SEE16150493SA SEE16150494SA 00840682109260SEE16150494SA SEE16150496SA 00840682109260SEE16150496SA SEE16150497SA 00840682109260SEE16150497SA SEE16150498SA 00840682109260SEE16150498SA SEE16150499SA 00840682109260SEE16150499SA SEE16150500SA 00840682109260SEE16150500SA SEE16150501SA 00840682109260SEE16150501SA SEE16150502SA 00840682109260SEE16150502SA SEE16150503SA 00840682109260SEE16150503SA SEE16150524SA 00840682109260SEE16150524SA SEE16150525SA 00840682109260SEE16150525SA SEE16150526SA 00840682109260SEE16150526SA SEE16150527SA 00840682109260SEE16150527SA SEE16150528SA 00840682109260SEE16150528SA SEE16150529SA 00840682109260SEE16150529SA SEE16150530SA 00840682109260SEE16150530SA SEE16150531SA 00840682109260SEE16150531SA SEE16150532SA 00840682109260SEE16150532SA SEE16150533SA 00840682109260SEE16150533SA SEE16150540SA 00840682109260SEE16150540SA SEE16150541SA 00840682109260SEE16150541SA SEE16150542SA 00840682109260SEE16150542SA SEE16150543SA 00840682109260SEE16150543SA SEE16150544SA 00840682109260SEE16150544SA SEE16150545SA 00840682109260SEE16150545SA SEE16150547SA 00840682109260SEE16150547SA SEE16150554SA 00840682109260SEE16150554SA SEE16150555SA 00840682109260SEE16150555SA SEE16150556SA 00840682109260SEE16150556SA SEE16150557SA 00840682109260SEE16150557SA SEE16150568SA 00840682109260SEE16150568SA SEE16150569SA 00840682109260SEE16150569SA SEE16160725SA 00840682109260SEE16160725SA SEE16160726SA 00840682109260SEE16160726SA SEE16160727SA 00840682109260SEE16160727SA SEE16160734SA 00840682109260SEE16160734SA SEE16160735SA 00840682109260SEE16160735SA SEE16160736SA 00840682109260SEE16160736SA SEE16160737SA 00840682109260SEE16160737SA SEE16180898SA 00840682109260SEE16180898SA SEE16180899SA 00840682109260SEE16180899SA SEE16180900SA 00840682109260SEE16180900SA SEE16180907SA 00840682109260SEE16180907SA SEE16180908SA 00840682109260SEE16180908SA SEE16180909SA 00840682109260SEE16180909SA SEE16180910SA 00840682109260SEE16180910SA SEE16180911SA 00840682109260SEE16180911SA SEE16180912SA 00840682109260SEE16180912SA SEE16180913SA 00840682109260SEE16180913SA SEE16180924SA 00840682109260SEE16180924SA SEE16180925SA 00840682109260SEE16180925SA SEE16190968SA 00840682109260SEE16190968SA SEE16190970SA 00840682109260SEE16190970SA SEE16191058SA 00840682109260SEE16191058SA SEE16191059SA 00840682109260SEE16191059SA SEE16191060SA 00840682109260SEE16191060SA SEE16191061SA 00840682109260SEE16191061SA SEE16191062SA 00840682109260SEE16191062SA SEE16191063SA 00840682109260SEE16191063SA SEE16191064SA 00840682109260SEE16191064SA SEE16191065SA 00840682109260SEE16191065SA SEE16201177SA 00840682109260SEE16201177SA SEE16201178SA 00840682109260SEE16201178SA SEE16201179SA 00840682109260SEE16201179SA SEE16201180SA 00840682109260SEE16201180SA SEE16201181SA 00840682109260SEE16201181SA SEE16201182SA 00840682109260SEE16201182SA SEE16201367SA 00840682109260SEE16201367SA SEE16201368SA 00840682109260SEE16201368SA SEE16201405SA 00840682109260SEE16201405SA SEE16201406SA 00840682109260SEE16201406SA SEE16201407SA 00840682109260SEE16201407SA SEE16201408SA 00840682109260SEE16201408SA SEE16211409SA 00840682109260SEE16211409SA SEE16211410SA 00840682109260SEE16211410SA SEE16211411SA 00840682109260SEE16211411SA SEE16211412SA 00840682109260SEE16211412SA SEE16211420SA 00840682109260SEE16211420SA SEE16211507SA 00840682109260SEE16211507SA SEE16211508SA 00840682109260SEE16211508SA SEE16211509SA 00840682109260SEE16211509SA SEE16211510SA 00840682109260SEE16211510SA SEE16211511SA 00840682109260SEE16211511SA SEE16211512SA 00840682109260SEE16211512SA SEE16211513SA 00840682109260SEE16211513SA SEE16221553SA 00840682109260SEE16221553SA SEE16221554SA 00840682109260SEE16221554SA SEE16221555SA 00840682109260SEE16221555SA SEE16221556SA 00840682109260SEE16221556SA SEE16221557SA 00840682109260SEE16221557SA SEE16221558SA 00840682109260SEE16221558SA SEE16221559SA 00840682109260SEE16221559SA SEE16221560SA 00840682109260SEE16221560SA SEE16221585SA 00840682109260SEE16221585SA SEE16221587SA 00840682109260SEE16221587SA SEE16221588SA 00840682109260SEE16221588SA SEE16221589SA 00840682109260SEE16221589SA SEE16221590SA 00840682109260SEE16221590SA SEE16221591SA 00840682109260SEE16221591SA SEE16221592SA 00840682109260SEE16221592SA SEE16221692SA 00840682109260SEE16221692SA SEE16221693SA 00840682109260SEE16221693SA SEE16221696SA 00840682109260SEE16221696SA SEE16221697SA 00840682109260SEE16221697SA SEE16221698SA 00840682109260SEE16221698SA SEE16221699SA 00840682109260SEE16221699SA SEE16233391GA 00840682109260SEE16233391GA SEE16241789SA 00840682109260SEE16241789SA SEE16241790SA 00840682109260SEE16241790SA SEE16241791SA 00840682109260SEE16241791SA SEE16241792SA 00840682109260SEE16241792SA SEE16241793SA 00840682109260SEE16241793SA SEE16241794SA 00840682109260SEE16241794SA SEE16241795SA 00840682109260SEE16241795SA SEE16241796SA 00840682109260SEE16241796SA SEE16241797SA 00840682109260SEE16241797SA SEE16241816SA 00840682109260SEE16241816SA SEE16241823SA 00840682109260SEE16241823SA SEE16241824SA 00840682109260SEE16241824SA SEE16241825SA 00840682109260SEE16241825SA SEE16241826SA 00840682109260SEE16241826SA SEE16241827SA 00840682109260SEE16241827SA SEE16241828SA 00840682109260SEE16241828SA SEE16241829SA 00840682109260SEE16241829SA SEE16241830SA 00840682109260SEE16241830SA SEE16241831SA 00840682109260SEE16241831SA SEE16241832SA 00840682109260SEE16241832SA SEE16241833SA 00840682109260SEE16241833SA SEE16241835SA 00840682109260SEE16241835SA SEE16241842SA 00840682109260SEE16241842SA SEE16241843SA 00840682109260SEE16241843SA SEE16241845SA 00840682109260SEE16241845SA SEE16241847SA 00840682109260SEE16241847SA SEE16241848SA 00840682109260SEE16241848SA SEE16241849SA 00840682109260SEE16241849SA SEE16241850SA 00840682109260SEE16241850SA SEE16241853SA 00840682109260SEE16241853SA SEE16241854SA 00840682109260SEE16241854SA SEE16241856SA 00840682109260SEE16241856SA SEE16241857SA 00840682109260SEE16241857SA SEE16241858SA 00840682109260SEE16241858SA SEE16241859SA 00840682109260SEE16241859SA SEE16251979SA 00840682109260SEE16251979SA SEE16251981SA 00840682109260SEE16251981SA SEE16251982SA 00840682109260SEE16251982SA SEE16251983SA 00840682109260SEE16251983SA SEE16251984SA 00840682109260SEE16251984SA SEE16252013SA 00840682109260SEE16252013SA SEE16252014SA 00840682109260SEE16252014SA SEE16252015SA 00840682109260SEE16252015SA SEE16252017SA 00840682109260SEE16252017SA SEE16252018SA 00840682109260SEE16252018SA SEE16252019SA 00840682109260SEE16252019SA SEE16252022SA 00840682109260SEE16252022SA SEE16252023SA 00840682109260SEE16252023SA SEE16252024SA 00840682109260SEE16252024SA SEE16252025SA 00840682109260SEE16252025SA SEE16252026SA 00840682109260SEE16252026SA SEE16252037SA 00840682109260SEE16252037SA SEE16260002SA 00840682109260SEE16260002SA SEE16262084SA 00840682109260SEE16262084SA SEE16262132SA 00840682109260SEE16262132SA SEE16262133SA 00840682109260SEE16262133SA SEE16262134SA 00840682109260SEE16262134SA SEE16262136SA 00840682109260SEE16262136SA SEE16262137SA 00840682109260SEE16262137SA SEE16262138SA 00840682109260SEE16262138SA SEE16262140SA 00840682109260SEE16262140SA SEE16262142SA 00840682109260SEE16262142SA SEE16262145SA 00840682109260SEE16262145SA SEE16262146SA 00840682109260SEE16262146SA SEE16262147SA 00840682109260SEE16262147SA SEE16262148SA 00840682109260SEE16262148SA SEE16262149SA 00840682109260SEE16262149SA SEE16262150SA 00840682109260SEE16262150SA SEE16262151SA 00840682109260SEE16262151SA SEE16262161SA 00840682109260SEE16262161SA SEE16262162SA 00840682109260SEE16262162SA SEE16262163SA 00840682109260SEE16262163SA SEE16262164SA 00840682109260SEE16262164SA SEE16262165SA 00840682109260SEE16262165SA SEE16262180SA 00840682109260SEE16262180SA SEE16262204SA 00840682109260SEE16262204SA SEE16262252SA 00840682109260SEE16262252SA SEE16272269SA 00840682109260SEE16272269SA SEE16272270SA 00840682109260SEE16272270SA SEE16272271SA 00840682109260SEE16272271SA SEE16272272SA 00840682109260SEE16272272SA SEE16272328SA 00840682109260SEE16272328SA SEE16272331SA 00840682109260SEE16272331SA SEE16272332SA 00840682109260SEE16272332SA SEE16272419SA 00840682109260SEE16272419SA SEE16272585SA 00840682109260SEE16272585SA SEE16272586SA 00840682109260SEE16272586SA SEE16272587SA 00840682109260SEE16272587SA SEE16272588SA 00840682109260SEE16272588SA SEE16272594SA 00840682109260SEE16272594SA SEE16272595SA 00840682109260SEE16272595SA SEE16272597SA 00840682109260SEE16272597SA SEE16272598SA 00840682109260SEE16272598SA SEE16272599SA 00840682109260SEE16272599SA SEE16272600SA 00840682109260SEE16272600SA SEE16272618SA 00840682109260SEE16272618SA SEE16272620SA 00840682109260SEE16272620SA SEE16282651SA 00840682109260SEE16282651SA SEE16282652SA 00840682109260SEE16282652SA SEE16282653SA 00840682109260SEE16282653SA SEE16282654SA 00840682109260SEE16282654SA SEE16282707SA 00840682109260SEE16282707SA SEE16282785SA 00840682109260SEE16282785SA SEE16282788SA 00840682109260SEE16282788SA SEE16282790SA 00840682109260SEE16282790SA SEE16282791SA 00840682109260SEE16282791SA SEE16282862SA 00840682109260SEE16282862SA SEE16292869SA 00840682109260SEE16292869SA SEE16292871SA 00840682109260SEE16292871SA SEE16292877SA 00840682109260SEE16292877SA SEE16293172SA 00840682109260SEE16293172SA SEE16293173SA 00840682109260SEE16293173SA SEE16293175SA 00840682109260SEE16293175SA SEE16293176SA 00840682109260SEE16293176SA SEE16293179SA 00840682109260SEE16293179SA SEE16293180SA 00840682109260SEE16293180SA SEE16293181SA 00840682109260SEE16293181SA SEE16303212SA 00840682109260SEE16303212SA SEE16303213SA 00840682109260SEE16303213SA SEE16303219SA 00840682109260SEE16303219SA SEE16303220SA 00840682109260SEE16303220SA SEE16303224SA 00840682109260SEE16303224SA SEE16303405GA 00840682109260SEE16303405GA SEE16303409GA 00840682109260SEE16303409GA SEE16313375SA 00840682109260SEE16313375SA SEE16313376SA 00840682109260SEE16313376SA SEE16313377SA 00840682109260SEE16313377SA SEE16313378SA 00840682109260SEE16313378SA SEE16313379SA 00840682109260SEE16313379SA SEE16313380SA 00840682109260SEE16313380SA SEE16324000SA 00840682109260SEE16324000SA SEE16324001SA 00840682109260SEE16324001SA SEE16324002SA 00840682109260SEE16324002SA SEE16324003SA 00840682109260SEE16324003SA SEE16324005SA 00840682109260SEE16324005SA SEE16324006SA 00840682109260SEE16324006SA SEE16324026SA 00840682109260SEE16324026SA SEE16324027SA 00840682109260SEE16324027SA SEE16324031SA 00840682109260SEE16324031SA SEE16324032SA 00840682109260SEE16324032SA SEE16324033SA 00840682109260SEE16324033SA SEE16334082SA 00840682109260SEE16334082SA SEE16334083SA 00840682109260SEE16334083SA SEE16334084SA 00840682109260SEE16334084SA SEE16334088SA 00840682109260SEE16334088SA SEE16334089SA 00840682109260SEE16334089SA SEE16334159SA 00840682109260SEE16334159SA SEE16334160SA 00840682109260SEE16334160SA SEE16334161SA 00840682109260SEE16334161SA SEE16334162SA 00840682109260SEE16334162SA SEE16334163SA 00840682109260SEE16334163SA SEE16334164SA 00840682109260SEE16334164SA SEE16334165SA 00840682109260SEE16334165SA SEE16334166SA 00840682109260SEE16334166SA SEE16334167SA 00840682109260SEE16334167SA SEE16334168SA 00840682109260SEE16334168SA SEE16334170SA 00840682109260SEE16334170SA SEE16334171SA 00840682109260SEE16334171SA SEE16334172SA 00840682109260SEE16334172SA SEE16334173SA 00840682109260SEE16334173SA SEE16334174SA 00840682109260SEE16334174SA SEE16334175SA 00840682109260SEE16334175SA SEE16334176SA 00840682109260SEE16334176SA SEE16334177SA 00840682109260SEE16334177SA SEE16334178SA 00840682109260SEE16334178SA SEE16334179SA 00840682109260SEE16334179SA SEE16334180SA 00840682109260SEE16334180SA SEE16334181SA 00840682109260SEE16334181SA SEE16334182SA 00840682109260SEE16334182SA SEE16334183SA 00840682109260SEE16334183SA SEE16334184SA 00840682109260SEE16334184SA SEE16334185SA 00840682109260SEE16334185SA SEE16334186SA 00840682109260SEE16334186SA SEE16334187SA 00840682109260SEE16334187SA SEE16334188SA 00840682109260SEE16334188SA SEE16334199SA 00840682109260SEE16334199SA SEE16334200SA 00840682109260SEE16334200SA SEE16334201SA 00840682109260SEE16334201SA SEE16334202SA 00840682109260SEE16334202SA SEE16334203SA 00840682109260SEE16334203SA SEE16334204SA 00840682109260SEE16334204SA SEE16334205SA 00840682109260SEE16334205SA SEE16334206SA 00840682109260SEE16334206SA SEE16334207SA 00840682109260SEE16334207SA SEE16334208SA 00840682109260SEE16334208SA SEE16334209SA 00840682109260SEE16334209SA SEE16334210SA 00840682109260SEE16334210SA SEE16344281SA 00840682109260SEE16344281SA SEE16344282SA 00840682109260SEE16344282SA SEE16344283SA 00840682109260SEE16344283SA SEE16344323SA 00840682109260SEE16344323SA SEE16344325SA 00840682109260SEE16344325SA SEE16344326SA 00840682109260SEE16344326SA SEE16344327SA 00840682109260SEE16344327SA SEE16344328SA 00840682109260SEE16344328SA SEE16354344SA 00840682109260SEE16354344SA SEE16354345SA 00840682109260SEE16354345SA SEE16354346SA 00840682109260SEE16354346SA SEE16360011SA 00840682109260SEE16360011SA SEE16360012SA 00840682109260SEE16360012SA SEE16360013SA 00840682109260SEE16360013SA SEE16360014SA 00840682109260SEE16360014SA SEE16360015SA 00840682109260SEE16360015SA SEE16360016SA 00840682109260SEE16360016SA SEE16360017SA 00840682109260SEE16360017SA SEE16360018SA 00840682109260SEE16360018SA SEE16360019SA 00840682109260SEE16360019SA SEE16360020SA 00840682109260SEE16360020SA SEE16360021SA 00840682109260SEE16360021SA SEE16360022SA 00840682109260SEE16360022SA SEE16360023SA 00840682109260SEE16360023SA SEE16360025SA 00840682109260SEE16360025SA SEE16360026SA 00840682109260SEE16360026SA SEE16364405SA 00840682109260SEE16364405SA SEE16364406SA 00840682109260SEE16364406SA SEE16364407SA 00840682109260SEE16364407SA SEE16364408SA 00840682109260SEE16364408SA SEE16364409SA 00840682109260SEE16364409SA SEE16364410SA 00840682109260SEE16364410SA SEE16364411SA 00840682109260SEE16364411SA SEE16364412SA 00840682109260SEE16364412SA SEE16364413SA 00840682109260SEE16364413SA SEE16364414SA 00840682109260SEE16364414SA SEE16364415SA 00840682109260SEE16364415SA SEE16364416SA 00840682109260SEE16364416SA SEE16364417SA 00840682109260SEE16364417SA SEE16364418SA 00840682109260SEE16364418SA SEE16364419SA 00840682109260SEE16364419SA SEE16364420SA 00840682109260SEE16364420SA SEE16364421SA 00840682109260SEE16364421SA SEE16370002SA 00840682109260SEE16370002SA SEE16370003SA 00840682109260SEE16370003SA SEE16370004SA 00840682109260SEE16370004SA SEE16370005SA 00840682109260SEE16370005SA SEE16370006SA 00840682109260SEE16370006SA SEE16380005SA 00840682109260SEE16380005SA SEE16380007SA 00840682109260SEE16380007SA SEE16380008SA 00840682109260SEE16380008SA SEE16380009SA 00840682109260SEE16380009SA SEE16380010SA 00840682109260SEE16380010SA SEE16380011SA 00840682109260SEE16380011SA SEE16380012SA 00840682109260SEE16380012SA SEE16380013SA 00840682109260SEE16380013SA SEE16380015SA 00840682109260SEE16380015SA SEE16380016SA 00840682109260SEE16380016SA SEE16380017SA 00840682109260SEE16380017SA SEE16380018SA 00840682109260SEE16380018SA SEE16380019SA 00840682109260SEE16380019SA SEE16380021SA 00840682109260SEE16380021SA SEE16380022SA 00840682109260SEE16380022SA SEE16380023SA 00840682109260SEE16380023SA SEE16380024SA 00840682109260SEE16380024SA SEE16380025SA 00840682109260SEE16380025SA SEE16380026SA 00840682109260SEE16380026SA SEE16380027SA 00840682109260SEE16380027SA SEE16380028SA 00840682109260SEE16380028SA SEE16380029SA 00840682109260SEE16380029SA SEE16380030SA 00840682109260SEE16380030SA SEE16380031SA 00840682109260SEE16380031SA SEE16380032SA 00840682109260SEE16380032SA SEE16380033SA 00840682109260SEE16380033SA SEE16383412GA 00840682109260SEE16383412GA SEE16383417GA 00840682109260SEE16383417GA SEE16383418GA 00840682109260SEE16383418GA SEE16390001SA 00840682109260SEE16390001SA SEE16390002SA 00840682109260SEE16390002SA SEE16390003SA 00840682109260SEE16390003SA SEE16390005SA 00840682109260SEE16390005SA SEE16390008SA 00840682109260SEE16390008SA SEE16390010SA 00840682109260SEE16390010SA SEE16390012SA 00840682109260SEE16390012SA SEE16390013SA 00840682109260SEE16390013SA SEE16390014SA 00840682109260SEE16390014SA SEE16390017SA 00840682109260SEE16390017SA SEE16390018SA 00840682109260SEE16390018SA SEE16390019SA 00840682109260SEE16390019SA SEE16390020SA 00840682109260SEE16390020SA SEE16390021SA 00840682109260SEE16390021SA SEE16390023SA 00840682109260SEE16390023SA SEE16390024SA 00840682109260SEE16390024SA SEE16400001SA 00840682109260SEE16400001SA SEE16410002SA 00840682109260SEE16410002SA SEE16410003SA 00840682109260SEE16410003SA SEE16410004SA 00840682109260SEE16410004SA SEE16410005SA 00840682109260SEE16410005SA SEE16410006SA 00840682109260SEE16410006SA SEE16410007SA 00840682109260SEE16410007SA SEE16410008SA 00840682109260SEE16410008SA SEE16410009SA 00840682109260SEE16410009SA SEE16410010SA 00840682109260SEE16410010SA SEE16410011SA 00840682109260SEE16410011SA SEE16410012SA 00840682109260SEE16410012SA SEE16410014SA 00840682109260SEE16410014SA SEE16410015SA 00840682109260SEE16410015SA SEE16410016SA 00840682109260SEE16410016SA SEE16420041SA 00840682109260SEE16420041SA SEE16420042SA 00840682109260SEE16420042SA SEE16420043SA 00840682109260SEE16420043SA SEE16420044SA 00840682109260SEE16420044SA SEE16420045SA 00840682109260SEE16420045SA SEE16420046SA 00840682109260SEE16420046SA SEE16420047SA 00840682109260SEE16420047SA SEE16420048SA 00840682109260SEE16420048SA SEE16420049SA 00840682109260SEE16420049SA SEE16420050SA 00840682109260SEE16420050SA SEE16420061SA 00840682109260SEE16420061SA SEE16420062SA 00840682109260SEE16420062SA SEE16420063SA 00840682109260SEE16420063SA SEE16420064SA 00840682109260SEE16420064SA SEE16420066SA 00840682109260SEE16420066SA SEE16420067SA 00840682109260SEE16420067SA SEE16420068SA 00840682109260SEE16420068SA SEE16420069SA 00840682109260SEE16420069SA SEE16420070SA 00840682109260SEE16420070SA SEE16420071SA 00840682109260SEE16420071SA SEE16420072SA 00840682109260SEE16420072SA SEE16420073SA 00840682109260SEE16420073SA SEE16420074SA 00840682109260SEE16420074SA SEE16420076SA 00840682109260SEE16420076SA SEE16420077SA 00840682109260SEE16420077SA SEE16420078SA 00840682109260SEE16420078SA SEE16420079SA 00840682109260SEE16420079SA SEE16420080SA 00840682109260SEE16420080SA SEE16430003SA 00840682109260SEE16430003SA SEE16430004SA 00840682109260SEE16430004SA SEE16430005SA 00840682109260SEE16430005SA SEE16430007SA 00840682109260SEE16430007SA SEE16430008SA 00840682109260SEE16430008SA SEE16430010SA 00840682109260SEE16430010SA SEE16430011SA 00840682109260SEE16430011SA SEE16430012SA 00840682109260SEE16430012SA SEE16430014SA 00840682109260SEE16430014SA SEE16430015SA 00840682109260SEE16430015SA SEE16430017SA 00840682109260SEE16430017SA SEE16430018SA 00840682109260SEE16430018SA SEE16430019SA 00840682109260SEE16430019SA SEE16430020SA 00840682109260SEE16430020SA SEE16430021SA 00840682109260SEE16430021SA SEE16430022SA 00840682109260SEE16430022SA SEE16430023SA 00840682109260SEE16430023SA SEE16430024SA 00840682109260SEE16430024SA SEE16430026SA 00840682109260SEE16430026SA SEE16430027SA 00840682109260SEE16430027SA SEE16430028SA 00840682109260SEE16430028SA SEE16430029SA 00840682109260SEE16430029SA SEE16430031SA 00840682109260SEE16430031SA SEE16430032SA 00840682109260SEE16430032SA SEE16430033SA 00840682109260SEE16430033SA SEE16430034SA 00840682109260SEE16430034SA SEE16430035SA 00840682109260SEE16430035SA SEE16430036SA 00840682109260SEE16430036SA SEE16430037SA 00840682109260SEE16430037SA SEE16430038SA 00840682109260SEE16430038SA SEE16430040SA 00840682109260SEE16430040SA SEE16430041SA 00840682109260SEE16430041SA SEE16430042SA 00840682109260SEE16430042SA SEE16430043SA 00840682109260SEE16430043SA SEE16430044SA 00840682109260SEE16430044SA SEE16430046SA 00840682109260SEE16430046SA SEE16430047SA 00840682109260SEE16430047SA SEE16430048SA 00840682109260SEE16430048SA SEE16430049SA 00840682109260SEE16430049SA SEE16430052SA 00840682109260SEE16430052SA SEE16430057SA 00840682109260SEE16430057SA SEE16430060SA 00840682109260SEE16430060SA SEE16430061SA 00840682109260SEE16430061SA SEE16430062SA 00840682109260SEE16430062SA SEE16430064SA 00840682109260SEE16430064SA SEE16430065SA 00840682109260SEE16430065SA SEE16440001SA 00840682109260SEE16440001SA SEE16440002SA 00840682109260SEE16440002SA SEE16440004SA 00840682109260SEE16440004SA SEE16440005SA 00840682109260SEE16440005SA SEE16440006SA 00840682109260SEE16440006SA SEE16440007SA 00840682109260SEE16440007SA SEE16450002SA 00840682109260SEE16450002SA SEE16450004SA 00840682109260SEE16450004SA SEE16450005SA 00840682109260SEE16450005SA SEE16450006SA 00840682109260SEE16450006SA SEE16450007SA 00840682109260SEE16450007SA SEE16450008SA 00840682109260SEE16450008SA SEE16450009SA 00840682109260SEE16450009SA SEE16450010SA 00840682109260SEE16450010SA SEE16450012SA 00840682109260SEE16450012SA SEE16450013SA 00840682109260SEE16450013SA SEE16450014SA 00840682109260SEE16450014SA SEE16450015SA 00840682109260SEE16450015SA SEE16450016SA 00840682109260SEE16450016SA SEE14031717GA 00840682109260SEE14031717GA SEE16333410GA 00840682109260SEE16333410GA SEE09100122GA 00840682109260SEE09100122GA SEE09100143GA 00840682109260SEE09100143GA SEE09100145GA 00840682109260SEE09100145GA SEE09190237GA 00840682109260SEE09190237GA SEE09310361GA 00840682109260SEE09310361GA SEE09310385GA 00840682109260SEE09310385GA SEE09310386GA 00840682109260SEE09310386GA SEE09310388GA 00840682109260SEE09310388GA SEE09330414GA 00840682109260SEE09330414GA SEE09410589GA 00840682109260SEE09410589GA SEE09480728GA 00840682109260SEE09480728GA SEE09530895GA 00840682109260SEE09530895GA SEE09530913GA 00840682109260SEE09530913GA SEE10090953GA 00840682109260SEE10090953GA SEE10130047GA 00840682109260SEE10130047GA SEE10400400GA 00840682109260SEE10400400GA SEE10410471GA 00840682109260SEE10410471GA SEE10420523GA 00840682109260SEE10420523GA SEE10420578GA 00840682109260SEE10420578GA SEE10470800GA 00840682109260SEE10470800GA SEE11071134GA 00840682109260SEE11071134GA SEE11191428GA 00840682109260SEE11191428GA SEE11341656SA 00840682109260SEE11341656SA SEE11371787GA 00840682109260SEE11371787GA SEE12082032GA 00840682109260SEE12082032GA SEE12082034GA 00840682109260SEE12082034GA SEE12092052GA 00840682109260SEE12092052GA SEE12112108GA 00840682109260SEE12112108GA SEE12122140GA 00840682109260SEE12122140GA SEE12162182GA 00840682109260SEE12162182GA SEE12360034GA 00840682109260SEE12360034GA SEE12370060GA 00840682109260SEE12370060GA SEE12400097GA 00840682109260SEE12400097GA SEE12430201GA 00840682109260SEE12430201GA SEE12430205GA 00840682109260SEE12430205GA SEE12490243GA 00840682109260SEE12490243GA SEE12490287GA 00840682109260SEE12490287GA SEE13000394GA 00840682109260SEE13000394GA SEE13140655GA 00840682109260SEE13140655GA SEE13180775GA 00840682109260SEE13180775GA SEE13180780GA 00840682109260SEE13180780GA SEE13180781GA 00840682109260SEE13180781GA SEE13270898GA 00840682109260SEE13270898GA SEE13280912GA 00840682109260SEE13280912GA SEE13290967GA 00840682109260SEE13290967GA SEE13290983GA 00840682109260SEE13290983GA SEE13290986GA 00840682109260SEE13290986GA SEE13351018GA 00840682109260SEE13351018GA SEE13351031GA 00840682109260SEE13351031GA SEE13371096GA 00840682109260SEE13371096GA SEE13451334GA 00840682109260SEE13451334GA SEE13481407GA 00840682109260SEE13481407GA SEE13501464GA 00840682109260SEE13501464GA SEE14021544GA 00840682109260SEE14021544GA SEE14021545GA 00840682109260SEE14021545GA SEE14021549GA 00840682109260SEE14021549GA SEE14021562GA 00840682109260SEE14021562GA SEE14101783GA 00840682109260SEE14101783GA SEE14151819GA 00840682109260SEE14151819GA SEE14151843GA 00840682109260SEE14151843GA SEE14161894GA 00840682109260SEE14161894GA SEE14222002GA 00840682109260SEE14222002GA SEE14222020GA 00840682109260SEE14222020GA SEE14242103GA 00840682109260SEE14242103GA SEE14252134GA 00840682109260SEE14252134GA SEE14452627GA 00840682109260SEE14452627GA SEE14462651GA 00840682109260SEE14462651GA SEE14462658GA 00840682109260SEE14462658GA SEE14502761GA 00840682109260SEE14502761GA SEE15012850GA 00840682109260SEE15012850GA SEE15153203GA 00840682109260SEE15153203GA SEE15183256GA 00840682109260SEE15183256GA SEE15193272GA 00840682109260SEE15193272GA SEE15193281GA 00840682109260SEE15193281GA SEE15373309GA 00840682109260SEE15373309GA SEE15413316GA 00840682109260SEE15413316GA SEE15423319GA 00840682109260SEE15423319GA SEE15423320GA 00840682109260SEE15423320GA SEE15423321GA 00840682109260SEE15423321GA SEE15423323GA 00840682109260SEE15423323GA SEE15423328GA 00840682109260SEE15423328GA SEE15423329GA 00840682109260SEE15423329GA SEE15423330GA 00840682109260SEE15423330GA SEE16123366GA 00840682109260SEE16123366GA SEE16123367GA 00840682109260SEE16123367GA SEE16173370GA 00840682109260SEE16173370GA SEE16173375GA 00840682109260SEE16173375GA SEE16173377GA 00840682109260SEE16173377GA SEE16183378GA 00840682109260SEE16183378GA SEE16183379GA 00840682109260SEE16183379GA SEE16213390GA 00840682109260SEE16213390GA SEE16272628SA 00840682109260SEE16272628SA SEE16282787SA 00840682109260SEE16282787SA SEE16282789SA 00840682109260SEE16282789SA SEE16303407GA 00840682109260SEE16303407GA SEE16390004SA 00840682109260SEE16390004SA SEE16420065SA 00840682109260SEE16420065SA SEE16420075SA 00840682109260SEE16420075SA SEE16430050SA 00840682109260SEE16430050SA SEE16450003SA 00840682109260SEE16450003SA SEE09140179GA 00840682109260SEE09140179GA SEE09230269GA 00840682109260SEE09230269GA SEE09470678GA 00840682109260SEE09470678GA SEE09480691GA 00840682109260SEE09480691GA SEE09530851GA 00840682109260SEE09530851GA SEE10250087GA 00840682109260SEE10250087GA SEE14031619GA 00840682109260SEE14031619GA SEE14302263GA 00840682109260SEE14302263GA SEE14472677GA 00840682109260SEE14472677GA SEE16440003SA 00840682109260SEE16440003SA SEE09190225GA 00840682109260SEE09190225GA SEE09230272GA 00840682109260SEE09230272GA SEE10250174GA 00840682109260SEE10250174GA SEE14151822GA 00840682109260SEE14151822GA SEE14322325GA 00840682109260SEE14322325GA SEE14332341GA 00840682109260SEE14332341GA SEE14422559GA 00840682109260SEE14422559GA SEE15312143SA 00840682109260SEE15312143SA SEE16173369GA 00840682109260SEE16173369GA SEE16450001SA 00840682109260SEE16450001SA SEE13270890GA 00840682109260SEE13270890GA SEE14021542GA 00840682109260SEE14021542GA SEE09310363GA 00840682109260SEE09310363GA SEE09330418GA 00840682109260SEE09330418GA SEE09340436GA 00840682109260SEE09340436GA SEE09460631GA 00840682109260SEE09460631GA SEE09460653GA 00840682109260SEE09460653GA SEE09470660GA 00840682109260SEE09470660GA SEE09470662GA 00840682109260SEE09470662GA SEE09470664GA 00840682109260SEE09470664GA SEE09470675GA 00840682109260SEE09470675GA SEE09470679GA 00840682109260SEE09470679GA SEE09510813GA 00840682109260SEE09510813GA SEE09530897GA 00840682109260SEE09530897GA SEE09530898GA 00840682109260SEE09530898GA SEE10090947GA 00840682109260SEE10090947GA SEE10120998GA 00840682109260SEE10120998GA SEE10130072GA 00840682109260SEE10130072GA SEE10250103GA 00840682109260SEE10250103GA SEE10260199GA 00840682109260SEE10260199GA SEE10270267GA 00840682109260SEE10270267GA SEE10400388GA 00840682109260SEE10400388GA SEE10400389GA 00840682109260SEE10400389GA SEE10400390GA 00840682109260SEE10400390GA SEE10400436GA 00840682109260SEE10400436GA SEE10400437GA 00840682109260SEE10400437GA SEE10420601GA 00840682109260SEE10420601GA SEE10430613GA 00840682109260SEE10430613GA SEE10430619GA 00840682109260SEE10430619GA SEE10430624GA 00840682109260SEE10430624GA SEE10430636GA 00840682109260SEE10430636GA SEE10430643GA 00840682109260SEE10430643GA SEE10430654GA 00840682109260SEE10430654GA SEE10460721GA 00840682109260SEE10460721GA SEE10460770GA 00840682109260SEE10460770GA SEE10530840GA 00840682109260SEE10530840GA SEE11030897GA 00840682109260SEE11030897GA SEE11050919GA 00840682109260SEE11050919GA SEE11050927GA 00840682109260SEE11050927GA SEE11050935GA 00840682109260SEE11050935GA SEE11050936GA 00840682109260SEE11050936GA SEE11050937GA 00840682109260SEE11050937GA SEE11071048GA 00840682109260SEE11071048GA SEE11071071GA 00840682109260SEE11071071GA SEE11071137GA 00840682109260SEE11071137GA SEE11081177GA 00840682109260SEE11081177GA SEE11081178GA 00840682109260SEE11081178GA SEE11081185GA 00840682109260SEE11081185GA SEE11081191GA 00840682109260SEE11081191GA SEE11081197GA 00840682109260SEE11081197GA SEE11151303GA 00840682109260SEE11151303GA SEE11151310GA 00840682109260SEE11151310GA SEE11161328GA 00840682109260SEE11161328GA SEE11171380GA 00840682109260SEE11171380GA SEE11201450GA 00840682109260SEE11201450GA SEE11231528GA 00840682109260SEE11231528GA SEE11341640GA 00840682109260SEE11341640GA SEE11361726GA 00840682109260SEE11361726GA SEE11371767GA 00840682109260SEE11371767GA SEE12031942GA 00840682109260SEE12031942GA SEE12092045GA 00840682109260SEE12092045GA SEE12092047GA 00840682109260SEE12092047GA SEE12092049GA 00840682109260SEE12092049GA SEE12092082GA 00840682109260SEE12092082GA SEE12162181GA 00840682109260SEE12162181GA SEE12172197GA 00840682109260SEE12172197GA SEE12252321GA 00840682109260SEE12252321GA SEE12282359GA 00840682109260SEE12282359GA SEE12302380GA 00840682109260SEE12302380GA SEE12302395GA 00840682109260SEE12302395GA SEE12340006GA 00840682109260SEE12340006GA SEE12342448GA 00840682109260SEE12342448GA SEE12370070GA 00840682109260SEE12370070GA SEE12420160GA 00840682109260SEE12420160GA SEE12420162GA 00840682109260SEE12420162GA SEE12420165GA 00840682109260SEE12420165GA SEE12420170GA 00840682109260SEE12420170GA SEE12440232GA 00840682109260SEE12440232GA SEE12490276GA 00840682109260SEE12490276GA SEE12490283GA 00840682109260SEE12490283GA SEE12510305GA 00840682109260SEE12510305GA SEE13000389GA 00840682109260SEE13000389GA SEE13020408GA 00840682109260SEE13020408GA SEE13110505GA 00840682109260SEE13110505GA SEE13150665GA 00840682109260SEE13150665GA SEE13150673GA 00840682109260SEE13150673GA SEE13150675GA 00840682109260SEE13150675GA SEE13150676GA 00840682109260SEE13150676GA SEE13150691GA 00840682109260SEE13150691GA SEE13190791GA 00840682109260SEE13190791GA SEE13401172GA 00840682109260SEE13401172GA SEE13401178GA 00840682109260SEE13401178GA SEE13431272GA 00840682109260SEE13431272GA SEE13431287GA 00840682109260SEE13431287GA SEE13451311GA 00840682109260SEE13451311GA SEE14031624GA 00840682109260SEE14031624GA SEE14031655GA 00840682109260SEE14031655GA SEE14081774GA 00840682109260SEE14081774GA SEE14101779GA 00840682109260SEE14101779GA SEE14382468GA 00840682109260SEE14382468GA SEE14432577GA 00840682109260SEE14432577GA SEE14452628GA 00840682109260SEE14452628GA SEE14482706GA 00840682109260SEE14482706GA SEE14482715GA 00840682109260SEE14482715GA SEE14482717GA 00840682109260SEE14482717GA SEE14482718GA 00840682109260SEE14482718GA SEE15052940GA 00840682109260SEE15052940GA SEE15113183GA 00840682109260SEE15113183GA SEE15373311GA 00840682109260SEE15373311GA SEE15473339GA 00840682109260SEE15473339GA SEE16053354GA 00840682109260SEE16053354GA SEE16053355GA 00840682109260SEE16053355GA SEE16292870SA 00840682109260SEE16292870SA SEE16313483SA 00840682109260SEE16313483SA SEE16313484SA 00840682109260SEE16313484SA SEE16324029SA 00840682109260SEE16324029SA SEE16430001SA 00840682109260SEE16430001SA SEE16430006SA 00840682109260SEE16430006SA SEE16430025SA 00840682109260SEE16430025SA SEE16430030SA 00840682109260SEE16430030SA SEE16430055SA 00840682109260SEE16430055SA SEE09470655GA 00840682109260SEE09470655GA SEE09510771GA 00840682109260SEE09510771GA SEE09510781GA 00840682109260SEE09510781GA SEE10120009GA 00840682109260SEE10120009GA SEE10120031GA 00840682109260SEE10120031GA SEE10121017GA 00840682109260SEE10121017GA SEE10400379GA 00840682109260SEE10400379GA SEE10410488GA 00840682109260SEE10410488GA SEE10460745GA 00840682109260SEE10460745GA SEE10460750GA 00840682109260SEE10460750GA SEE10470785GA 00840682109260SEE10470785GA SEE11111232GA 00840682109260SEE11111232GA SEE11161332GA 00840682109260SEE11161332GA SEE11331592GA 00840682109260SEE11331592GA SEE11361669GA 00840682109260SEE11361669GA SEE11361700GA 00840682109260SEE11361700GA SEE11371808GA 00840682109260SEE11371808GA SEE12051992GA 00840682109260SEE12051992GA SEE12082028GA 00840682109260SEE12082028GA SEE12332440GA 00840682109260SEE12332440GA SEE12430182GA 00840682109260SEE12430182GA SEE12430199GA 00840682109260SEE12430199GA SEE13050440GA 00840682109260SEE13050440GA SEE13100478GA 00840682109260SEE13100478GA SEE13150672GA 00840682109260SEE13150672GA SEE13150677GA 00840682109260SEE13150677GA SEE13150679GA 00840682109260SEE13150679GA SEE13180721GA 00840682109260SEE13180721GA SEE13180727GA 00840682109260SEE13180727GA SEE13180730GA 00840682109260SEE13180730GA SEE13180731GA 00840682109260SEE13180731GA SEE13180732GA 00840682109260SEE13180732GA SEE13180734GA 00840682109260SEE13180734GA SEE13180736GA 00840682109260SEE13180736GA SEE13180737GA 00840682109260SEE13180737GA SEE13180738GA 00840682109260SEE13180738GA SEE13180740GA 00840682109260SEE13180740GA SEE16233398GA 00840682109260SEE16233398GA SEE16383414GA 00840682109260SEE16383414GA SEE16390022SA 00840682109260SEE16390022SA SEE09230283GA 00840682109260SEE09230283GA SEE09390533GA 00840682109260SEE09390533GA SEE09460645GA 00840682109260SEE09460645GA SEE15062983GA 00840682109260SEE15062983GA SEE15173234GA 00840682109260SEE15173234GA SEE15183263GA 00840682109260SEE15183263GA SEE15193293GA 00840682109260SEE15193293GA SEE15312130SA 00840682109260SEE15312130SA SEE15383313GA 00840682109260SEE15383313GA SEE15413318GA 00840682109260SEE15413318GA SEE16252020SA 00840682109260SEE16252020SA SEE12252319GA 00840682109260SEE12252319GA SEE12252324GA 00840682109260SEE12252324GA SEE14031718GA 00840682109260SEE14031718GA SEE14021538GA 00840682109260SEE14021538GA SEE14021556GA 00840682109260SEE14021556GA SEE09230248GA 00840682109260SEE09230248GA SEE09230249GA 00840682109260SEE09230249GA SEE10121023GA 00840682109260SEE10121023GA SEE13361062GA 00840682109260SEE13361062GA SEE15093038GA 00840682109260SEE15093038GA SEE15312128SA 00840682109260SEE15312128SA SEE15373307GA 00840682109260SEE15373307GA SEE15373308GA 00840682109260SEE15373308GA SEE15373310GA 00840682109260SEE15373310GA SEE16272621SA 00840682109260SEE16272621SA SEE16313485SA 00840682109260SEE16313485SA SEE16390015SA 00840682109260SEE16390015SA SEE09100088GA 00840682109260SEE09100088GA SEE09130164GA 00840682109260SEE09130164GA SEE09130167GA 00840682109260SEE09130167GA SEE09140173GA 00840682109260SEE09140173GA SEE09230270GA 00840682109260SEE09230270GA SEE09230280GA 00840682109260SEE09230280GA SEE09330407GA 00840682109260SEE09330407GA SEE09330415GA 00840682109260SEE09330415GA SEE09330420GA 00840682109260SEE09330420GA SEE09330426GA 00840682109260SEE09330426GA SEE09350459GA 00840682109260SEE09350459GA SEE09350464GA 00840682109260SEE09350464GA SEE09350465GA 00840682109260SEE09350465GA SEE09360474GA 00840682109260SEE09360474GA SEE09370512GA 00840682109260SEE09370512GA SEE09380514GA 00840682109260SEE09380514GA SEE09390549GA 00840682109260SEE09390549GA SEE09390550GA 00840682109260SEE09390550GA SEE09410588GA 00840682109260SEE09410588GA SEE09420596GA 00840682109260SEE09420596GA SEE09430612GA 00840682109260SEE09430612GA SEE09510800GA 00840682109260SEE09510800GA SEE09510827GA 00840682109260SEE09510827GA SEE09530841GA 00840682109260SEE09530841GA SEE09530877GA 00840682109260SEE09530877GA SEE10100970GA 00840682109260SEE10100970GA SEE10120990GA 00840682109260SEE10120990GA SEE10130035GA 00840682109260SEE10130035GA SEE10130037GA 00840682109260SEE10130037GA SEE10250166GA 00840682109260SEE10250166GA SEE10250181GA 00840682109260SEE10250181GA SEE10270250GA 00840682109260SEE10270250GA SEE10310280GA 00840682109260SEE10310280GA SEE10310285GA 00840682109260SEE10310285GA SEE10400380GA 00840682109260SEE10400380GA SEE10400391GA 00840682109260SEE10400391GA SEE10400409GA 00840682109260SEE10400409GA SEE10400428GA 00840682109260SEE10400428GA SEE10420554GA 00840682109260SEE10420554GA SEE10450668GA 00840682109260SEE10450668GA SEE10450673GA 00840682109260SEE10450673GA SEE10530853GA 00840682109260SEE10530853GA SEE11060961GA 00840682109260SEE11060961GA SEE11141264GA 00840682109260SEE11141264GA SEE11161345GA 00840682109260SEE11161345GA SEE11171364GA 00840682109260SEE11171364GA SEE11181422GA 00840682109260SEE11181422GA SEE11191423GA 00840682109260SEE11191423GA SEE11200793GA 00840682109260SEE11200793GA SEE11211476GA 00840682109260SEE11211476GA SEE11211507GA 00840682109260SEE11211507GA SEE11241549GA 00840682109260SEE11241549GA SEE11301554GA 00840682109260SEE11301554GA SEE11301556GA 00840682109260SEE11301556GA SEE11301557GA 00840682109260SEE11301557GA SEE11331600GA 00840682109260SEE11331600GA SEE11341624GA 00840682109260SEE11341624GA SEE11341657GA 00840682109260SEE11341657GA SEE11351666GA 00840682109260SEE11351666GA SEE11361695GA 00840682109260SEE11361695GA SEE11361696GA 00840682109260SEE11361696GA SEE11361699GA 00840682109260SEE11361699GA SEE11361712GA 00840682109260SEE11361712GA SEE12031939GA 00840682109260SEE12031939GA SEE12102096GA 00840682109260SEE12102096GA SEE12112113GA 00840682109260SEE12112113GA SEE12172185GA 00840682109260SEE12172185GA SEE12172187GA 00840682109260SEE12172187GA SEE12182225GA 00840682109260SEE12182225GA SEE12242284GA 00840682109260SEE12242284GA SEE12242295GA 00840682109260SEE12242295GA SEE12322412GA 00840682109260SEE12322412GA SEE12370065GA 00840682109260SEE12370065GA SEE12370069GA 00840682109260SEE12370069GA SEE12420158GA 00840682109260SEE12420158GA SEE12430230GA 00840682109260SEE12430230GA SEE12490265GA 00840682109260SEE12490265GA SEE12520358GA 00840682109260SEE12520358GA SEE12520385GA 00840682109260SEE12520385GA SEE13220841GA 00840682109260SEE13220841GA SEE13270889GA 00840682109260SEE13270889GA SEE13280917GA 00840682109260SEE13280917GA SEE13371109GA 00840682109260SEE13371109GA SEE13371112GA 00840682109260SEE13371112GA SEE13371119GA 00840682109260SEE13371119GA SEE14021510GA 00840682109260SEE14021510GA SEE14021521GA 00840682109260SEE14021521GA SEE14031687GA 00840682109260SEE14031687GA SEE14101778GA 00840682109260SEE14101778GA SEE14171928GA 00840682109260SEE14171928GA SEE14211985GA 00840682109260SEE14211985GA SEE14222004GA 00840682109260SEE14222004GA SEE14222041GA 00840682109260SEE14222041GA SEE14272142GA 00840682109260SEE14272142GA SEE14362378GA 00840682109260SEE14362378GA SEE14372404GA 00840682109260SEE14372404GA SEE14412493GA 00840682109260SEE14412493GA SEE14432593GA 00840682109260SEE14432593GA SEE14492723GA 00840682109260SEE14492723GA SEE14512824GA 00840682109260SEE14512824GA SEE15012857GA 00840682109260SEE15012857GA SEE15032869GA 00840682109260SEE15032869GA SEE15032886GA 00840682109260SEE15032886GA SEE15113159GA 00840682109260SEE15113159GA SEE15153200GA 00840682109260SEE15153200GA SEE15153204GA 00840682109260SEE15153204GA SEE15193270GA 00840682109260SEE15193270GA SEE15193280GA 00840682109260SEE15193280GA SEE15193296GA 00840682109260SEE15193296GA SEE15193300GA 00840682109260SEE15193300GA SEE15373306GA 00840682109260SEE15373306GA SEE15423322GA 00840682109260SEE15423322GA SEE15423324GA 00840682109260SEE15423324GA SEE15423327GA 00840682109260SEE15423327GA SEE15473333GA 00840682109260SEE15473333GA SEE15473336GA 00840682109260SEE15473336GA SEE15473338GA 00840682109260SEE15473338GA SEE15473340GA 00840682109260SEE15473340GA SEE16043341GA 00840682109260SEE16043341GA SEE16053344GA 00840682109260SEE16053344GA SEE16053358GA 00840682109260SEE16053358GA SEE16173376GA 00840682109260SEE16173376GA SEE16252012SA 00840682109260SEE16252012SA SEE16252016SA 00840682109260SEE16252016SA SEE16370001SA 00840682109260SEE16370001SA SEE16430009SA 00840682109260SEE16430009SA SEE16430013SA 00840682109260SEE16430013SA SEE16430039SA 00840682109260SEE16430039SA SEE16430058SA 00840682109260SEE16430058SA SEE16430059SA 00840682109260SEE16430059SA SEE16430063SA 00840682109260SEE16430063SA SEE09040070GA 00840682109260SEE09040070GA SEE09040073GA 00840682109260SEE09040073GA SEE09100153GA 00840682109260SEE09100153GA SEE09110155GA 00840682109260SEE09110155GA SEE09330424GA 00840682109260SEE09330424GA SEE09370511GA 00840682109260SEE09370511GA SEE09380520GA 00840682109260SEE09380520GA SEE09380521GA 00840682109260SEE09380521GA SEE09380524GA 00840682109260SEE09380524GA SEE09470658GA 00840682109260SEE09470658GA SEE09470659GA 00840682109260SEE09470659GA SEE09470667GA 00840682109260SEE09470667GA SEE09510835GA 00840682109260SEE09510835GA SEE10080926GA 00840682109260SEE10080926GA SEE10090945GA 00840682109260SEE10090945GA SEE10090958GA 00840682109260SEE10090958GA SEE10090968GA 00840682109260SEE10090968GA SEE10100974GA 00840682109260SEE10100974GA SEE10250123GA 00840682109260SEE10250123GA SEE10250131GA 00840682109260SEE10250131GA SEE10250165GA 00840682109260SEE10250165GA SEE10250171GA 00840682109260SEE10250171GA SEE10250178GA 00840682109260SEE10250178GA SEE10250189GA 00840682109260SEE10250189GA SEE10250190GA 00840682109260SEE10250190GA SEE10270255GA 00840682109260SEE10270255GA SEE10450679GA 00840682109260SEE10450679GA SEE11030896GA 00840682109260SEE11030896GA SEE11060946GA 00840682109260SEE11060946GA SEE11111226GA 00840682109260SEE11111226GA SEE11201461GA 00840682109260SEE11201461GA SEE11361680GA 00840682109260SEE11361680GA SEE12031945GA 00840682109260SEE12031945GA SEE12051990GA 00840682109260SEE12051990GA SEE12162173GA 00840682109260SEE12162173GA SEE12192239GA 00840682109260SEE12192239GA SEE12242290GA 00840682109260SEE12242290GA SEE12282349GA 00840682109260SEE12282349GA SEE12302387GA 00840682109260SEE12302387GA SEE12430227GA 00840682109260SEE12430227GA SEE12490254GA 00840682109260SEE12490254GA SEE12490281GA 00840682109260SEE12490281GA SEE12490285GA 00840682109260SEE12490285GA SEE13120571GA 00840682109260SEE13120571GA SEE13120585GA 00840682109260SEE13120585GA SEE13180778GA 00840682109260SEE13180778GA SEE13190807GA 00840682109260SEE13190807GA SEE13240863GA 00840682109260SEE13240863GA SEE13361076GA 00840682109260SEE13361076GA SEE13371089GA 00840682109260SEE13371089GA SEE13401173GA 00840682109260SEE13401173GA SEE13431296GA 00840682109260SEE13431296GA SEE13451369GA 00840682109260SEE13451369GA SEE13451374GA 00840682109260SEE13451374GA SEE13481401GA 00840682109260SEE13481401GA SEE13481417GA 00840682109260SEE13481417GA SEE13481418GA 00840682109260SEE13481418GA SEE14031593GA 00840682109260SEE14031593GA SEE14031660GA 00840682109260SEE14031660GA SEE14031706GA 00840682109260SEE14031706GA SEE14041729GA 00840682109260SEE14041729GA SEE14041745GA 00840682109260SEE14041745GA SEE14041750GA 00840682109260SEE14041750GA SEE14101792GA 00840682109260SEE14101792GA SEE14101796GA 00840682109260SEE14101796GA SEE14151872GA 00840682109260SEE14151872GA SEE14151873GA 00840682109260SEE14151873GA SEE14161886GA 00840682109260SEE14161886GA SEE14171921GA 00840682109260SEE14171921GA SEE14222011GA 00840682109260SEE14222011GA SEE14222013GA 00840682109260SEE14222013GA SEE14272210GA 00840682109260SEE14272210GA SEE14332345GA 00840682109260SEE14332345GA SEE14432581GA 00840682109260SEE14432581GA SEE15052912GA 00840682109260SEE15052912GA SEE15052927GA 00840682109260SEE15052927GA SEE15052943GA 00840682109260SEE15052943GA SEE15062977GA 00840682109260SEE15062977GA SEE15113117GA 00840682109260SEE15113117GA SEE15113171GA 00840682109260SEE15113171GA SEE15153199GA 00840682109260SEE15153199GA SEE15173231GA 00840682109260SEE15173231GA SEE15173244GA 00840682109260SEE15173244GA SEE15173245GA 00840682109260SEE15173245GA SEE15183264GA 00840682109260SEE15183264GA SEE15193277GA 00840682109260SEE15193277GA SEE15193278GA 00840682109260SEE15193278GA SEE15193283GA 00840682109260SEE15193283GA SEE15413317GA 00840682109260SEE15413317GA SEE15423326GA 00840682109260SEE15423326GA SEE15473332GA 00840682109260SEE15473332GA SEE15473334GA 00840682109260SEE15473334GA SEE16053356GA 00840682109260SEE16053356GA SEE16173373GA 00840682109260SEE16173373GA SEE16183382GA 00840682109260SEE16183382GA SEE16183387GA 00840682109260SEE16183387GA SEE16241851SA 00840682109260SEE16241851SA SEE16241852SA 00840682109260SEE16241852SA SEE16272420SA 00840682109260SEE16272420SA SEE16272583SA 00840682109260SEE16272583SA SEE16313482SA 00840682109260SEE16313482SA SEE16410013SA 00840682109260SEE16410013SA SEE16410017SA 00840682109260SEE16410017SA SEE09100080GA 00840682109260SEE09100080GA SEE15103109GA 00840682109260SEE15103109GA SEE16241846SA 00840682109260SEE16241846SA SEE16390007SA 00840682109260SEE16390007SA SEE16430051SA 00840682109260SEE16430051SA SEE09510791GA 00840682109260SEE09510791GA SEE15062978GA 00840682109260SEE15062978GA SEE15052939GA 00840682109260SEE15052939GA SEE15113177GA 00840682109260SEE15113177GA SEE15193273GA 00840682109260SEE15193273GA SEE15193294GA 00840682109260SEE15193294GA SEE16390011SA 00840682109260SEE16390011SA SEE09190234GA 00840682109260SEE09190234GA SEE09190248GA 00840682109260SEE09190248GA SEE09230277GA 00840682109260SEE09230277GA SEE09430614GA 00840682109260SEE09430614GA SEE10250179GA 00840682109260SEE10250179GA SEE14272137GA 00840682109260SEE14272137GA SEE14272138GA 00840682109260SEE14272138GA SEE14272141GA 00840682109260SEE14272141GA SEE14272153GA 00840682109260SEE14272153GA SEE14272156GA 00840682109260SEE14272156GA SEE14272157GA 00840682109260SEE14272157GA SEE14272158GA 00840682109260SEE14272158GA SEE14272159GA 00840682109260SEE14272159GA SEE14272166GA 00840682109260SEE14272166GA SEE14272167GA 00840682109260SEE14272167GA SEE12490299GA 00840682109260SEE12490299GA SEE14332339GA 00840682109260SEE14332339GA SEE11071100GA 00840682109260SEE11071100GA SEE12510312GA 00840682109260SEE12510312GA SEE13361079GA 00840682109260SEE13361079GA SEE09020014GA 00840682109260SEE09020014GA SEE09040049GA 00840682109260SEE09040049GA SEE09340451GA 00840682109260SEE09340451GA SEE13491429GA 00840682109260SEE13491429GA SEE13491433GA 00840682109260SEE13491433GA SEE13491434GA 00840682109260SEE13491434GA SEE13491445GA 00840682109260SEE13491445GA SEE09110154GA 00840682109260SEE09110154GA SEE09230285GA 00840682109260SEE09230285GA SEE09300347GA 00840682109260SEE09300347GA SEE09310399GA 00840682109260SEE09310399GA SEE09380513GA 00840682109260SEE09380513GA SEE09510784GA 00840682109260SEE09510784GA SEE09530879GA 00840682109260SEE09530879GA SEE09530885GA 00840682109260SEE09530885GA SEE09530894GA 00840682109260SEE09530894GA SEE09530902GA 00840682109260SEE09530902GA SEE09530909GA 00840682109260SEE09530909GA SEE10090961GA 00840682109260SEE10090961GA SEE10090964GA 00840682109260SEE10090964GA SEE10120025GA 00840682109260SEE10120025GA SEE10130041GA 00840682109260SEE10130041GA SEE10130052GA 00840682109260SEE10130052GA SEE10130054GA 00840682109260SEE10130054GA SEE10130058GA 00840682109260SEE10130058GA SEE10270258GA 00840682109260SEE10270258GA SEE10420591GA 00840682109260SEE10420591GA SEE13451321GA 00840682109260SEE13451321GA SEE13451353GA 00840682109260SEE13451353GA SEE13481402GA 00840682109260SEE13481402GA SEE14081770GA 00840682109260SEE14081770GA SEE14101798GA 00840682109260SEE14101798GA SEE14222017GA 00840682109260SEE14222017GA SEE14222030GA 00840682109260SEE14222030GA SEE14222061GA 00840682109260SEE14222061GA SEE14322311GA 00840682109260SEE14322311GA SEE14422563GA 00840682109260SEE14422563GA SEE15183250GA 00840682109260SEE15183250GA SEE15183251GA 00840682109260SEE15183251GA SEE15183258GA 00840682109260SEE15183258GA SEE15193285GA 00840682109260SEE15193285GA SEE15312142SA 00840682109260SEE15312142SA SEE15423331GA 00840682109260SEE15423331GA SEE15473335GA 00840682109260SEE15473335GA SEE16053359GA 00840682109260SEE16053359GA SEE16123364GA 00840682109260SEE16123364GA SEE16173368GA 00840682109260SEE16173368GA SEE16173374GA 00840682109260SEE16173374GA SEE16183384GA 00840682109260SEE16183384GA SEE16262152SA 00840682109260SEE16262152SA SEE16272593SA 00840682109260SEE16272593SA SEE16272596SA 00840682109260SEE16272596SA SEE16272627SA 00840682109260SEE16272627SA SEE16390006SA 00840682109260SEE16390006SA SEE09360470GA 00840682109260SEE09360470GA SEE09360475GA 00840682109260SEE09360475GA SEE09360476GA 00840682109260SEE09360476GA SEE09360478GA 00840682109260SEE09360478GA SEE09430600GA 00840682109260SEE09430600GA SEE09430603GA 00840682109260SEE09430603GA SEE13481416GA 00840682109260SEE13481416GA SEE14021508GA 00840682109260SEE14021508GA SEE15423325GA 00840682109260SEE15423325GA SEE16068806SA 00840682109260SEE16068806SA SEE09480739GA 00840682109260SEE09480739GA SEE09480749GA 00840682109260SEE09480749GA SEE09490751GA 00840682109260SEE09490751GA SEE09490753GA 00840682109260SEE09490753GA SEE09040065GA 00840682109260SEE09040065GA SEE09360484GA 00840682109260SEE09360484GA SEE14432582GA 00840682109260SEE14432582GA SEE14432595GA 00840682109260SEE14432595GA SEE14332342GA 00840682109260SEE14332342GA SEE11161335GA 00840682109260SEE11161335GA SEE14332346GA 00840682109260SEE14332346GA SEE15113161GA 00840682109260SEE15113161GA SEE14222005GA 00840682109260SEE14222005GA SEE14222007GA 00840682109260SEE14222007GA SEE14222015GA 00840682109260SEE14222015GA SEE14222031GA 00840682109260SEE14222031GA SEE15173239GA 00840682109260SEE15173239GA SEE15193274GA 00840682109260SEE15193274GA SEE15193275GA 00840682109260SEE15193275GA SEE16303248SA 00840682109260SEE16303248SA SEE16183381GA 00840682109260SEE16183381GA SEE09530890GA 00840682109260SEE09530890GA SEE10121014GA 00840682109260SEE10121014GA SEE11141294GA 00840682109260SEE11141294GA SEE13451342GA 00840682109260SEE13451342GA SEE10090965GA 00840682109260SEE10090965GA SEE10250078GA 00840682109260SEE10250078GA SEE10260201GA 00840682109260SEE10260201GA SEE10400414GA 00840682109260SEE10400414GA SEE10400415GA 00840682109260SEE10400415GA SEE11040911GA 00840682109260SEE11040911GA SEE12082031GA 00840682109260SEE12082031GA SEE12282355GA 00840682109260SEE12282355GA SEE12360039GA 00840682109260SEE12360039GA SEE13220838GA 00840682109260SEE13220838GA SEE13351043GA 00840682109260SEE13351043GA SEE13351045GA 00840682109260SEE13351045GA SEE13411211GA 00840682109260SEE13411211GA SEE13481387GA 00840682109260SEE13481387GA SEE14031599GA 00840682109260SEE14031599GA SEE16043343GA 00840682109260SEE16043343GA SEE16053346GA 00840682109260SEE16053346GA SEE16123363GA 00840682109260SEE16123363GA SEE16170870SA 00840682109260SEE16170870SA SEE16293174SA 00840682109260SEE16293174SA SEE09420591GA 00840682109260SEE09420591GA SEE09470661GA 00840682109260SEE09470661GA SEE11211516GA 00840682109260SEE11211516GA SEE11341634GA 00840682109260SEE11341634GA SEE11361688GA 00840682109260SEE11361688GA SEE15153189GA 00840682109260SEE15153189GA SEE13351048GA 00840682109260SEE13351048GA SEE15173230GA 00840682109260SEE15173230GA SEE10460743GA 00840682109260SEE10460743GA SEE13120598GA 00840682109260SEE13120598GA SEE14021517GA 00840682109260SEE14021517GA SEE14031659GA 00840682109260SEE14031659GA SEE14041759GA 00840682109260SEE14041759GA SEE15032871GA 00840682109260SEE15032871GA SEE15062965GA 00840682109260SEE15062965GA SEE15062972GA 00840682109260SEE15062972GA SEE15063002GA 00840682109260SEE15063002GA SEE15183262GA 00840682109260SEE15183262GA SEE16123365GA 00840682109260SEE16123365GA SEE16233396GA 00840682109260SEE16233396GA SEE16383413GA 00840682109260SEE16383413GA SEE16430045SA 00840682109260SEE16430045SA SEE15193299GA 00840682109260SEE15193299GA SEE09100141GA 00840682109260SEE09100141GA SEE09100144GA 00840682109260SEE09100144GA SEE09100146GA 00840682109260SEE09100146GA SEE09100149GA 00840682109260SEE09100149GA SEE09140197GA 00840682109260SEE09140197GA SEE13501478GA 00840682109260SEE13501478GA SEE15093054GA 00840682109260SEE15093054GA SEE15093055GA 00840682109260SEE15093055GA SEE15093056GA 00840682109260SEE15093056GA SEE15093060GA 00840682109260SEE15093060GA SEE15093061GA 00840682109260SEE15093061GA SEE15261529SA 00840682109260SEE15261529SA SEE15261535SA 00840682109260SEE15261535SA SEE16324028SA 00840682109260SEE16324028SA SEE16324030SA 00840682109260SEE16324030SA SEE16303406GA 00840682109260SEE16303406GA SEE16303408GA 00840682109260SEE16303408GA SEE16333411GA 00840682109260SEE16333411GA SEE16383415GA 00840682109260SEE16383415GA SEE16383416GA 00840682109260SEE16383416GA SEE16450011SA 00840682109260SEE16450011SA SEE09510786GA 00840682109260SEE09510786GA SEE09510799GA 00840682109260SEE09510799GA SEE10100971GA 00840682109260SEE10100971GA SEE10100981GA 00840682109260SEE10100981GA SEE10250188GA 00840682109260SEE10250188GA SEE10260210GA 00840682109260SEE10260210GA SEE11181402GA 00840682109260SEE11181402GA SEE15173237GA 00840682109260SEE15173237GA SEE15373312GA 00840682109260SEE15373312GA SEE16233392GA 00840682109260SEE16233392GA SEE09230274GA 00840682109260SEE09230274GA SEE09360480GA 00840682109260SEE09360480GA SEE10130066GA 00840682109260SEE10130066GA SEE11201458GA 00840682109260SEE11201458GA SEE11361722GA 00840682109260SEE11361722GA SEE13190810GA 00840682109260SEE13190810GA SEE09040044GA 00840682109260SEE09040044GA SEE09140174GA 00840682109260SEE09140174GA SEE09140175GA 00840682109260SEE09140175GA SEE09150198GA 00840682109260SEE09150198GA SEE09310395GA 00840682109260SEE09310395GA SEE09360477GA 00840682109260SEE09360477GA SEE09360485GA 00840682109260SEE09360485GA SEE09390546GA 00840682109260SEE09390546GA SEE09460642GA 00840682109260SEE09460642GA SEE09510774GA 00840682109260SEE09510774GA SEE10090950GA 00840682109260SEE10090950GA SEE10250183GA 00840682109260SEE10250183GA SEE10420516GA 00840682109260SEE10420516GA SEE10420546GA 00840682109260SEE10420546GA SEE10420552GA 00840682109260SEE10420552GA SEE10420555GA 00840682109260SEE10420555GA SEE10530862GA 00840682109260SEE10530862GA SEE10530863GA 00840682109260SEE10530863GA SEE11071068GA 00840682109260SEE11071068GA SEE11081162GA 00840682109260SEE11081162GA SEE11361675GA 00840682109260SEE11361675GA SEE12041950GA 00840682109260SEE12041950GA SEE12232251GA 00840682109260SEE12232251GA SEE12322426GA 00840682109260SEE12322426GA SEE12430196GA 00840682109260SEE12430196GA SEE12440235GA 00840682109260SEE12440235GA SEE13180723GA 00840682109260SEE13180723GA SEE13280921GA 00840682109260SEE13280921GA SEE13371115GA 00840682109260SEE13371115GA SEE13481404GA 00840682109260SEE13481404GA SEE14021502GA 00840682109260SEE14021502GA SEE14222045GA 00840682109260SEE14222045GA SEE14252122GA 00840682109260SEE14252122GA SEE14422554GA 00840682109260SEE14422554GA SEE14512820GA 00840682109260SEE14512820GA SEE15062964GA 00840682109260SEE15062964GA SEE15093051GA 00840682109260SEE15093051GA SEE15093058GA 00840682109260SEE15093058GA SEE15173240GA 00840682109260SEE15173240GA SEE15183253GA 00840682109260SEE15183253GA SEE15193266GA 00840682109260SEE15193266GA SEE15413314GA 00840682109260SEE15413314GA SEE15473337GA 00840682109260SEE15473337GA SEE16053357GA 00840682109260SEE16053357GA SEE16173371GA 00840682109260SEE16173371GA SEE16183383GA 00840682109260SEE16183383GA SEE16183385GA 00840682109260SEE16183385GA SEE16183386GA 00840682109260SEE16183386GA SEE16241855SA 00840682109260SEE16241855SA SEE16251980SA 00840682109260SEE16251980SA SEE16303221SA 00840682109260SEE16303221SA SEE16313481SA 00840682109260SEE16313481SA SEE09100142GA 00840682109260SEE09100142GA SEE09100151GA 00840682109260SEE09100151GA SEE09130161GA 00840682109260SEE09130161GA SEE09410585GA 00840682109260SEE09410585GA SEE10090967GA 00840682109260SEE10090967GA SEE10250128GA 00840682109260SEE10250128GA SEE10250152GA 00840682109260SEE10250152GA SEE10250164GA 00840682109260SEE10250164GA SEE10250170GA 00840682109260SEE10250170GA SEE10250180GA 00840682109260SEE10250180GA SEE11071143GA 00840682109260SEE11071143GA SEE13351035GA 00840682109260SEE13351035GA SEE13351054GA 00840682109260SEE13351054GA SEE14292238GA 00840682109260SEE14292238GA SEE14292246GA 00840682109260SEE14292246GA SEE14302260GA 00840682109260SEE14302260GA SEE15012844GA 00840682109260SEE15012844GA SEE15173218GA 00840682109260SEE15173218GA SEE16053353GA 00840682109260SEE16053353GA SEE16410001SA 00840682109260SEE16410001SA SEE09020025GA 00840682109260SEE09020025GA SEE09530892GA 00840682109260SEE09530892GA SEE16380006SA 00840682109260SEE16380006SA SEE09510810GA 00840682109260SEE09510810GA SEE12112114GA 00840682109260SEE12112114GA SEE14021563GA 00840682109260SEE14021563GA SEE14222035GA 00840682109260SEE14222035GA SEE10130057GA 00840682109260SEE10130057GA SEE12041957GA 00840682109260SEE12041957GA SEE13411203GA 00840682109260SEE13411203GA SEE14292242GA 00840682109260SEE14292242GA SEE14382455GA 00840682109260SEE14382455GA SEE14452626GA 00840682109260SEE14452626GA SEE15113163GA 00840682109260SEE15113163GA SEE15113167GA 00840682109260SEE15113167GA SEE15193282GA 00840682109260SEE15193282GA SEE16390009SA 00840682109260SEE16390009SA SEE09230271GA 00840682109260SEE09230271GA SEE09360469GA 00840682109260SEE09360469GA SEE09530868GA 00840682109260SEE09530868GA SEE09530883GA 00840682109260SEE09530883GA SEE13180768GA 00840682109260SEE13180768GA SEE14272206GA 00840682109260SEE14272206GA SEE14462661GA 00840682109260SEE14462661GA SEE15183265GA 00840682109260SEE15183265GA SEE15193298GA 00840682109260SEE15193298GA SEE10420594GA 00840682109260SEE10420594GA SEE13431298GA 00840682109260SEE13431298GA SEE13501452GA 00840682109260SEE13501452GA SEE14312298GA 00840682109260SEE14312298GA SEE15113147GA 00840682109260SEE15113147GA SEE15113155GA 00840682109260SEE15113155GA SEE16053348GA 00840682109260SEE16053348GA SEE16053361GA 00840682109260SEE16053361GA SEE16053362GA 00840682109260SEE16053362GA SEE16173372GA 00840682109260SEE16173372GA SEE16213389GA 00840682109260SEE16213389GA SEE16233393GA 00840682109260SEE16233393GA SEE16233394GA 00840682109260SEE16233394GA SEE16233395GA 00840682109260SEE16233395GA SEE16233397GA 00840682109260SEE16233397GA SEE16233399GA 00840682109260SEE16233399GA SEE16344324SA 00840682109260SEE16344324SA SEE08510008GA 00840682109260SEE08510008GA SEE09020016GA 00840682109260SEE09020016GA SEE09020018GA 00840682109260SEE09020018GA SEE09020039GA 00840682109260SEE09020039GA SEE09040056GA 00840682109260SEE09040056GA SEE09040059GA 00840682109260SEE09040059GA SEE09040067GA 00840682109260SEE09040067GA SEE09100081GA 00840682109260SEE09100081GA SEE09100113GA 00840682109260SEE09100113GA SEE09190224GA 00840682109260SEE09190224GA SEE09230250GA 00840682109260SEE09230250GA SEE09230275GA 00840682109260SEE09230275GA SEE09240298GA 00840682109260SEE09240298GA SEE09330419GA 00840682109260SEE09330419GA SEE09430610GA 00840682109260SEE09430610GA SEE09460647GA 00840682109260SEE09460647GA SEE09470666GA 00840682109260SEE09470666GA SEE09510775GA 00840682109260SEE09510775GA SEE09530891GA 00840682109260SEE09530891GA SEE10090954GA 00840682109260SEE10090954GA SEE10090963GA 00840682109260SEE10090963GA SEE10090966GA 00840682109260SEE10090966GA SEE10120996GA 00840682109260SEE10120996GA SEE10121004GA 00840682109260SEE10121004GA SEE10121015GA 00840682109260SEE10121015GA SEE10250146GA 00840682109260SEE10250146GA SEE10250186GA 00840682109260SEE10250186GA SEE10260191GA 00840682109260SEE10260191GA SEE10400430GA 00840682109260SEE10400430GA SEE10410484GA 00840682109260SEE10410484GA SEE10420584GA 00840682109260SEE10420584GA SEE10420592GA 00840682109260SEE10420592GA SEE11071066GA 00840682109260SEE11071066GA SEE11071083GA 00840682109260SEE11071083GA SEE11071089GA 00840682109260SEE11071089GA SEE11371747GA 00840682109260SEE11371747GA SEE11371794GA 00840682109260SEE11371794GA SEE12062008GA 00840682109260SEE12062008GA SEE12062012GA 00840682109260SEE12062012GA SEE12062014GA 00840682109260SEE12062014GA SEE12092066GA 00840682109260SEE12092066GA SEE12420171GA 00840682109260SEE12420171GA SEE13000392GA 00840682109260SEE13000392GA SEE13020409GA 00840682109260SEE13020409GA SEE13130631GA 00840682109260SEE13130631GA SEE13280947GA 00840682109260SEE13280947GA SEE13411202GA 00840682109260SEE13411202GA SEE13411209GA 00840682109260SEE13411209GA SEE13431283GA 00840682109260SEE13431283GA SEE13431293GA 00840682109260SEE13431293GA SEE13481400GA 00840682109260SEE13481400GA SEE14021578GA 00840682109260SEE14021578GA SEE14031714GA 00840682109260SEE14031714GA SEE14041736GA 00840682109260SEE14041736GA SEE14081776GA 00840682109260SEE14081776GA SEE14332347GA 00840682109260SEE14332347GA SEE14452637GA 00840682109260SEE14452637GA SEE14462643GA 00840682109260SEE14462643GA SEE14462648GA 00840682109260SEE14462648GA SEE14462660GA 00840682109260SEE14462660GA SEE14462664GA 00840682109260SEE14462664GA SEE15093029GA 00840682109260SEE15093029GA SEE15093045GA 00840682109260SEE15093045GA SEE15093048GA 00840682109260SEE15093048GA SEE15093049GA 00840682109260SEE15093049GA SEE15093053GA 00840682109260SEE15093053GA SEE15093057GA 00840682109260SEE15093057GA SEE15113149GA 00840682109260SEE15113149GA SEE15173232GA 00840682109260SEE15173232GA SEE15173233GA 00840682109260SEE15173233GA SEE15173235GA 00840682109260SEE15173235GA SEE15183259GA 00840682109260SEE15183259GA SEE15193267GA 00840682109260SEE15193267GA SEE15193269GA 00840682109260SEE15193269GA SEE15193304GA 00840682109260SEE15193304GA SEE15312144SA 00840682109260SEE15312144SA SEE15413315GA 00840682109260SEE15413315GA SEE16043342GA 00840682109260SEE16043342GA SEE16053345GA 00840682109260SEE16053345GA SEE16053347GA 00840682109260SEE16053347GA SEE16053349GA 00840682109260SEE16053349GA SEE16053350GA 00840682109260SEE16053350GA SEE16053360GA 00840682109260SEE16053360GA SEE16183380GA 00840682109260SEE16183380GA SEE16213388GA 00840682109260SEE16213388GA SEE16272336SA 00840682109260SEE16272336SA SEE16272625SA 00840682109260SEE16272625SA SEE16292913SA 00840682109260SEE16292913SA SEE16293171SA 00840682109260SEE16293171SA SEE16380014SA 00840682109260SEE16380014SA SEE16390025SA 00840682109260SEE16390025SA SEE11071061GA 00840682109260SEE11071061GA SEE13270883GA 00840682109260SEE13270883GA SEE13361067GA 00840682109260SEE13361067GA SEE13401151GA 00840682109260SEE13401151GA SEE13401163GA 00840682109260SEE13401163GA SEE13431300GA 00840682109260SEE13431300GA SEE13451314GA 00840682109260SEE13451314GA SEE13451323GA 00840682109260SEE13451323GA SEE13451324GA 00840682109260SEE13451324GA SEE13451348GA 00840682109260SEE13451348GA SEE13451370GA 00840682109260SEE13451370GA SEE13451371GA 00840682109260SEE13451371GA SEE13451372GA 00840682109260SEE13451372GA SEE13451373GA 00840682109260SEE13451373GA SEE13481405GA 00840682109260SEE13481405GA SEE14031595GA 00840682109260SEE14031595GA SEE14031633GA 00840682109260SEE14031633GA SEE14031665GA 00840682109260SEE14031665GA SEE14081769GA 00840682109260SEE14081769GA SEE14101788GA 00840682109260SEE14101788GA SEE14101789GA 00840682109260SEE14101789GA SEE14111801GA 00840682109260SEE14111801GA SEE14161905GA 00840682109260SEE14161905GA SEE14302256GA 00840682109260SEE14302256GA SEE14412507GA 00840682109260SEE14412507GA SEE14422519GA 00840682109260SEE14422519GA SEE14432576GA 00840682109260SEE14432576GA SEE14452619GA 00840682109260SEE14452619GA SEE14462647GA 00840682109260SEE14462647GA SEE14462650GA 00840682109260SEE14462650GA SEE14462653GA 00840682109260SEE14462653GA SEE14472678GA 00840682109260SEE14472678GA SEE14482704GA 00840682109260SEE14482704GA SEE14482720GA 00840682109260SEE14482720GA SEE15032884GA 00840682109260SEE15032884GA SEE15032887GA 00840682109260SEE15032887GA SEE15052946GA 00840682109260SEE15052946GA SEE15052950GA 00840682109260SEE15052950GA SEE15062992GA 00840682109260SEE15062992GA SEE15063006GA 00840682109260SEE15063006GA SEE15113174GA 00840682109260SEE15113174GA SEE15173219GA 00840682109260SEE15173219GA SEE15193284GA 00840682109260SEE15193284GA SEE15193301GA 00840682109260SEE15193301GA  Serial numbers without UDIs: SAH04290001GA SAH04290002GA SAH04290003GA SAH04290004GA SAH04290005GA SAH04290007GA SAH04290009GA SAH04290010GA SAH04290012GA SAH04290013GA SAH04290014GA SAH04290018GA SAH04290022GA SAH04290023GA SAH04290024GA SAH04290025GA SAH04290026GA SAH04290027GA SAH04290028GA SAH04290030GA SAH04340031GA SAH04340033GA SAH04340034GA SAH04340035GA SAH04340036GA SAH04340038GA SAH04350039GA SAH04350040GA SAH04350041GA SAH04350042GA SAH04350043GA SAH04350044GA SAH04350045GA SAH04350047GA SAH04350048GA SAH04350050GA SAH04350051GA SAH04350052GA SAH04350053GA SAH04350054GA SAH04350055GA SAH04350056GA SAH04350057GA SAH04350058GA SAH04350060GA SAH04360059GA SAH04360060GA SAH04360061GA SAH04360062GA SAH04360063GA SAH04360064GA SAH04360065GA SAH04360066GA SAH04360067GA SAH04360068GA SAH04360069GA SAH04360070GA SAH04360071GA SAH04360072GA SAH04360073GA SAH04360074GA SAH04360075GA SAH04360076GA SAH04360077GA SAH04360078GA SAH04360079GA SAH04360080GA SAH04370081GA SAH04380082GA SAH04380083GA SAH04380084GA SAH04380085GA SAH04380086GA SAH04380087GA SAH04380088GA SAH04380089GA SAH04380090GA SAH04380091GA SAH04380092GA SAH04380093GA SAH04380094GA SAH04380095GA SAH04380096GA SAH04380097GA SAH04380098GA SAH04380099GA SAH04380100GA SAH04380101GA SAH04380102GA SAH04380103GA SAH04380104GA SAH04380105GA SAH04380106GA SAH04380107GA SAH04390108GA SAH04390109GA SAH04390110GA SAH04390111GA SAH04390112GA SAH04390113GA SAH04390114GA SAH04390115GA SAH04390116GA SAH04390117GA SAH04390118GA SAH04390119GA SAH04390120GA SAH04390121GA SAH04390122GA SAH04390123GA SAH04390124GA SAH04390125GA SAH04390126GA SAH04390127GA SAH04390128GA SAH04390129GA SAH04390130GA SAH04390131GA SAH04390132GA SAH04390133GA SAH04390134GA SAH04390135GA SAH04390136GA SAH04390137GA SAH04390138GA SAH04390139GA SAH04390140GA SAH04390141GA SAH04390142GA SAH04390143GA SAH04390144GA SAH04390145GA SAH04390146GA SAH04390147GA SAH04390148GA SAH04390149GA SAH04390150GA SAH04390151GA SAH04390152GA SAH04390153GA SAH04390154GA SAH04390155GA SAH04390156GA SAH04390157GA SAH04390158GA SAH04390159GA SAH04400160GA SAH04400161GA SAH04400162GA SAH04400163GA SAH04400164GA SAH04400165GA SAH04400166GA SAH04400167GA SAH04400168GA SAH04400169GA SAH04400170GA SAH04400171GA SAH04400172GA SAH04400173GA SAH04400174GA SAH04400175GA SAH04400176GA SAH04400177GA SAH04400178GA SAH04400180GA SAH04400181GA SAH04400182GA SAH04400183GA SAH04400184GA SAH04410185GA SAH04410186GA SAH04410189GA SAH04410190GA SAH04410191GA SAH04410192GA SAH04410193GA SAH04410194GA SAH04410195GA SAH04410196GA SAH04410197GA SAH04410198GA SAH04410199GA SAH04410200GA SAH04410201GA SAH04410202GA SAH04410204GA SAH04410205GA SAH04410206GA SAH04410207GA SAH04410208GA SAH04410209GA SAH04420210GA SAH04420211GA SAH04420212GA SAH04420213GA SAH04420214GA SAH04420215GA SAH04420216GA SAH04420217GA SAH04420218GA SAH04420219GA SAH04420220GA SAH04420221GA SAH04420222GA SAH04420223GA SAH04420224GA SAH04420225GA SAH04420226GA SAH04420227GA SAH04420228GA SAH04420229GA SAH04420230GA SAH04420231GA SAH04420232GA SAH04420233GA SAH04420234GA SAH04430235GA SAH04430236GA SAH04430237GA SAH04430238GA SAH04430239GA SAH04430240GA SAH04430241GA SAH04430242GA SAH04430243GA SAH04430246GA SAH04430247GA SAH04430248GA SAH04430249GA SAH04430250GA SAH04430251GA SAH04430252GA SAH04430253GA SAH04430255GA SAH04430256GA SAH04430257GA SAH04430258GA SAH04450260GA SAH04450262GA SAH04450263GA SAH04450265GA SAH04450266GA SAH04450267GA SAH04450268GA SAH04450269GA SAH04450270GA SAH04450271GA SAH04450272GA SAH04450273GA SAH04450274GA SAH04450275GA SAH04450276GA SAH04450277GA SAH04450278GA SAH04450279GA SAH04450280GA SAH04450281GA SAH04450282GA SAH04450283GA SAH04450284GA SAH04450285GA SAH04460284GA SAH04460285GA SAH04460286GA SAH04460287GA SAH04460288GA SAH04460289GA SAH04460290GA SAH04460291GA SAH04460292GA SAH04460293GA SAH04460294GA SAH04460295GA SAH04460297GA SAH04460298GA SAH04460300GA SAH04460301GA SAH04460304GA SAH04460306GA SAH04460307GA SAH04460308GA SAH04470309GA SAH04470310GA SAH04470311GA SAH04470312GA SAH04470313GA SAH04470314GA SAH04470316GA SAH04470317GA SAH04470318GA SAH04470319GA SAH04470320GA SAH04470322GA SAH04470324GA SAH04470325GA SAH04470326GA SAH04470327GA SAH04470328GA SAH04470329GA SAH04470332GA SAH04470333GA SAH04480335GA SAH04480336GA SAH04480337GA SAH04480340GA SAH04480341GA SAH04480342GA SAH04480343GA SAH04480345GA SAH04480346GA SAH04480347GA SAH04480348GA SAH04480349GA SAH04480350GA SAH04480353GA SAH04480355GA SAH04480356GA SAH04480357GA SAH04490360GA SAH04490361GA SAH04490362GA SAH04490365GA SAH04490367GA SAH04490368GA SAH04490372GA SAH04490373GA SAH04490374GA SAH04490375GA SAH04490378GA SAH04490379GA SAH04490380GA SAH04490381GA SAH04490382GA SAH04490384GA SAH04500385GA SAH04500386GA SAH04500387GA SAH04500389GA SAH04500390GA SAH04500391GA SAH04500393SP SAH04500395GA SAH04500397GA SAH04500398GA SAH04500399GA SAH04500400GA SAH04500401GA SAH04500402GA SAH04500403GA SAH04500404GA SAH04500405GA SAH04500406GA SAH04500407GA SAH04500408GA SAH04500409GA SAH04500410GA SAH04500411GA SAH04500412GA SAH04500413GA SAH04500414GA SAH04500415GA SAH04500416GA SAH04500417GA SAH04500418GA SAH04500419GA SAH04500420GA SAH04500421GA SAH04510423GA SAH04510425GA SAH04510426GA SAH04510427GA SAH04510429GA SAH04510430GA SAH04510431GA SAH04510432GA SAH04510433GA SAH04510434GA SAH04510435GA SAH04510436GA SAH04510437GA SAH04510438GA SAH04510440GA SAH04510441GA SAH04510442GA SAH04510443GA SAH04510444GA SAH04510445GA SAH04510446GA SAH04510447GA SAH04510449GA SAH04510450GA SAH04510451GA SAH04510452GA SAH04510453GA SAH04510454GA SAH04510455GA SAH04510456GA SAH04510457GA SAH04510458GA SAH04510459GA SAH04520460GA SAH04520463GA SAH04520464GA SAH04520465GA SAH04520466GA SAH04520467GA SAH04520470GA SAH04520471GA SAH04520472GA SAH04520473GA SAH04520474GA SAH04520475GA SAH04520476GA SAH04520477GA SAH04520478GA SAH04520479GA SAH04520480GA SAH04520481GA SAH04520482GA SAH04520483GA SAH04520484GA SAH04530485GA SAH04530486GA SAH04530487GA SAH04530488GA SAH04530490GA SAH04530491GA SAH04530492GA SAH04530493GA SAH04530494GA SAH04530495GA SAH04530496GA SAH04530498GA SAH04530499GA SAH04530500GA SAH04530501GA SAH04530502GA SAH04530504GA SAH04530505GA SAH04530506GA SAH04530507GA SAH04530508GA SAH04530509GA SAH05010510GA SAH05010511GA SAH05010512GA SAH05010513GA SAH05010514GA SAH05010515GA SAH05010517GA SAH05010518GA SAH05010519GA SAH05010520GA SAH05010521GA SAH05010522GA SAH05010523GA SAH05010524GA SAH05010525GA SAH05010526GA SAH05010527GA SAH05010528GA SAH05010530GA SAH05010531GA SAH05010532GA SAH05010533GA SAH05010534GA SAH05010536GA SAH05010537GA SAH05010538GA SAH05010539GA SAH05010540GA SAH05010541GA SAH05010542GA SAH05010543GA SAH05010544GA SAH05010545GA SAH05010546GA SAH05010547GA SAH05010549GA SAH05010550GA SAH05010551GA SAH05010553GA SAH05010554GA SAH05010555GA SAH05010556GA SAH05010557GA SAH05010559GA SAH05010561GA SAH05010562GA SAH05010563GA SAH05010564GA SAH05040565GA SAH05040567GA SAH05040568GA SAH05040569GA SAH05040570GA SAH05040572GA SAH05040573GA SAH05040577GA SAH05040578GA SAH05040579GA SAH05040582GA SAH05040583GA SAH05040585GA SAH05040586GA SAH05040587GA SAH05040588GA SAH05040589GA SAH05040590GA SAH05040592GA SAH05040593GA SAH05040594GA SAH05070595GA SAH05070596GA SAH05070597GA SAH05070598GA SAH05070599GA SAH05070600GA SAH05070601GA SAH05070602GA SAH05070603GA SAH05070604GA SAH05070605GA SAH05070606GA SAH05070607GA SAH05070608GA SAH05070609GA SAH05070612GA SAH05070613GA SAH05070614GA SAH05070615GA SAH05070617GA SAH05070618GA SAH05070619GA SAH05070620GA SAH05070622GA SAH05070623GA SAH05070624GA SAH05080625GA SAH05080626GA SAH05080627GA SAH05080628GA SAH05080629GA SAH05080630GA SAH05080631GA SAH05080632GA SAH05080633GA SAH05080634GA SAH05080635GA SAH05080636GA SAH05080637GA SAH05080638GA SAH05080639GA SAH05080640GA SAH05080641GA SAH05080642GA SAH05080643GA SAH05080645GA SAH05080646GA SAH05080647GA SAH05080648GA SAH05080649GA SAH05080650GA SAH05080655GA SAH05080656GA SAH05080657GA SAH05080658GA SAH05080659GA SAH05080660GA SAH05080661GA SAH05080662GA SAH05080664GA SAH05080665GA SAH05080667GA SAH05080668GA SAH05080670GA SAH05080671GA SAH05080672GA SAH05080673GA SAH05080675GA SAH05080678GA SAH05080680GA SAH05080682GA SAH05080684GA SAH05110685GA SAH05110686GA SAH05110687GA SAH05110688GA SAH05110689GA SAH05110692GA SAH05110694GA SAH05110695GA SAH05110696GA SAH05110697GA SAH05110698GA SAH05110699GA SAH05110701GA SAH05110702GA SAH05110703GA SAH05110705GA SAH05110707GA SAH05110710GA SAH05110711GA SAH05110712GA SAH05110713GA SAH05133397GA SAH05140716GA SAH05140717GA SAH05140718GA SAH05140720GA SAH05140721GA SAH05140722GA SAH05140723GA SAH05140725GA SAH05140726GA SAH05140727GA SAH05140728GA SAH05140729GA SAH05140730GA SAH05140731GA SAH05140732GA SAH05140733GA SAH05140735GA SAH05140738GA SAH05140739GA SAH05140740GA SAH05140742GA SAH05140743GA SAH05140744GA SAH05150775GA SAH05150776GA SAH05150778GA SAH05150780GA SAH05150781GA SAH05150782GA SAH05150784GA SAH05150785GA SAH05150786GA SAH05150788GA SAH05150789GA SAH05150793GA SAH05150794GA SAH05150795GA SAH05150796GA SAH05150797GA SAH05150798GA SAH05150799GA SAH05150800GA SAH05150801GA SAH05150802GA SAH05150803GA SAH05150804GA SAH05160806GA SAH05160808GA SAH05160809GA SAH05160810GA SAH05160811GA SAH05160812GA SAH05160813GA SAH05160815GA SAH05160817GA SAH05160818GA SAH05160819GA SAH05160820GA SAH05160821GA SAH05160822GA SAH05160823GA SAH05160824GA SAH05160826GA SAH05160827GA SAH05160828GA SAH05160831GA SAH05220835GA SAH05220836GA SAH05220837GA SAH05220841GA SAH05220842GA SAH05220843GA SAH05220844GA SAH05220846GA SAH05220847GA SAH05220848GA SAH05220850GA SAH05220851GA SAH05220853GA SAH05220854GA SAH05220855GA SAH05220856GA SAH05220858GA SAH05220859GA SAH05220861GA SAH05220862GA SAH05220863GA SAH05220864GA SAH05240865GA SAH05240866GA SAH05240867GA SAH05240868GA SAH05240869GA SAH05240870GA SAH05240871GA SAH05240872GA SAH05240873GA SAH05240875GA SAH05240876GA SAH05240877GA SAH05240878GA SAH05240880GA SAH05240881GA SAH05240885GA SAH05240886GA SAH05240887GA SAH05240888GA SAH05240889GA SAH05240890GA SAH05240891GA SAH05240892GA SAH05240893GA SAH05250894GA SAH05250895GA SAH05250896GA SAH05250897GA SAH05250898GA SAH05250899GA SAH05250900GA SAH05250901GA SAH05250902GA SAH05250903GA SAH05260904GA SAH05260905GA SAH05260906GA SAH05260908GA SAH05260909GA SAH05260911GA SAH05260912GA SAH05260913GA SAH05260914GA SAH05260917GA SAH05260919GA SAH05260920GA SAH05260921GA SAH05260922GA SAH05260923GA SAH05280924GA SAH05280926GA SAH05280927GA SAH05280928GA SAH05280929GA SAH05280930GA SAH05280931GA SAH05280932GA SAH05280933GA SAH05280934GA SAH05280935GA SAH05280936GA SAH05280938GA SAH05280939GA SAH05280940GA SAH05280941GA SAH05280942GA SAH05280943GA SAH05280944GA SAH05280945GA SAH05280946GA SAH05280948GA SAH05280949GA SAH05280950GA SAH05280951GA SAH05280952GA SAH05280953GA SAH05290956GA SAH05290957GA SAH05290958GA SAH05290959GA SAH05290960GA SAH05290961GA SAH05290963GA SAH05290964GA SAH05290965GA SAH05290966GA SAH05290967GA SAH05290968GA SAH05290969GA SAH05290970GA SAH05290971GA SAH05290972GA SAH05290974GA SAH05290976GA SAH05290977GA SAH05290978GA SAH05290979GA SAH05290981GA SAH05290982GA SAH05290983GA SAH05300984GA SAH05320985GA SAH05320986GA SAH05320987GA SAH05320990GA SAH05320991GA SAH05320993GA SAH05320994GA SAH05320997GA SAH05320998GA SAH05321000GA SAH05321001GA SAH05321002GA SAH05321003GA SAH05321004GA SAH05321005GA SAH05321006GA SAH05321007GA SAH05321008GA SAH05321009GA SAH05321010GA SAH05321011GA SAH05321012GA SAH05321013GA SAH05321014GA SAH05331015GA SAH05351016GA SAH05351017GA SAH05351018GA SAH05351019GA SAH05351020GA SAH05351021GA SAH05351022GA SAH05351023GA SAH05351024GA SAH05351025GA SAH05351026GA SAH05351027GA SAH05351028GA SAH05351029GA SAH05351030GA SAH05351031GA SAH05351032GA SAH05351033GA SAH05351034GA SAH05351035GA SAH05351037GA SAH05351038GA SAH05351039GA SAH05351040GA SAH05351042GA SAH05351043GA SAH05351044GA SAH05351045GA SAH05381046GA SAH05381047GA SAH05381048GA SAH05381049GA SAH05381050GA SAH05381051GA SAH05381052GA SAH05381053GA SAH05381054GA SAH05381055GA SAH05381056GA SAH05381057GA SAH05381058GA SAH05381059GA SAH05381060GA SAH05381061GA SAH05381062GA SAH05381063GA SAH05381064GA SAH05381065GA SAH05381066GA SAH05381067GA SAH05381068GA SAH05381070GA SAH05381071GA SAH05381072GA SAH05381073GA SAH05381074GA SAH05391075GA SAH05391076GA SAH05391077GA SAH05391078GA SAH05391079GA SAH05391080GA SAH05391081GA SAH05391082GA SAH05391083GA SAH05391084GA SAH05391086GA SAH05391087GA SAH05391089GA SAH05391090GA SAH05391091GA SAH05391092GA SAH05391093GA SAH05391094GA SAH05391095GA SAH05391096GA SAH05391097GA SAH05391098GA SAH05391099GA SAH05391100GA SAH05391101GA SAH05391102GA SAH05391103GA SAH05411105GA SAH05411106GA SAH05411107GA SAH05411108GA SAH05411110GA SAH05411111GA SAH05411114GA SAH05411115GA SAH05421118GA SAH05421119GA SAH05421120GA SAH05421121GA SAH05421122GA SAH05421123GA SAH05421124GA SAH05421127GA SAH05421128GA SAH05421129GA SAH05431130GA SAH05431132GA SAH05431133GA SAH05431134GA SAH05431135GA SAH05431136GA SAH05431137GA SAH05431138GA SAH05431139GA SAH05431140GA SAH05431141GA SAH05431142GA SAH05431143GA SAH05431144GA SAH05431145GA SAH05431146GA SAH05431147GA SAH05431148GA SAH05431149GA SAH05431150GA SAH05431151GA SAH05431152GA SAH05431153GA SAH05431154GA SAH05431155GA SAH05431156GA SAH05431157GA SAH05431158GA SAH05431159GA SAH05431160GA SAH05431162GA SAH05431163GA SAH05431164GA SAH05437361GA SAH05441166GA SAH05441167GA SAH05441169GA SAH05441170GA SAH05441171GA SAH05441172GA SAH05441173GA SAH05441174GA SAH05441176GA SAH05441177GA SAH05441178GA SAH05441179GA SAH05441180GA SAH05441181GA SAH05441182GA SAH05441183GA SAH05441184GA SAH05441185GA SAH05441186GA SAH05441187GA SAH05441188GA SAH05441189GA SAH05441190GA SAH05441191GA SAH05441192GA SAH05441193GA SAH05441194GA SAH05451195GA SAH05451196GA SAH05451197GA SAH05451198GA SAH05451199GA SAH05451200GA SAH05451201GA SAH05451202GA SAH05451204GA SAH05451205GA SAH05451206GA SAH05451207GA SAH05451208GA SAH05451209GA SAH05451210GA SAH05451211GA SAH05451213GA SAH05451214GA SAH05451215GA SAH05451216GA SAH05451217GA SAH05451218GA SAH05451219GA SAH05451220GA SAH05451222GA SAH05461223GA SAH05461224GA SAH05461225GA SAH05461226GA SAH05461227GA SAH05461228GA SAH05461229GA SAH05461230GA SAH05461231GA SAH05461232GA SAH05461233GA SAH05461234GA SAH05471235GA SAH05471236GA SAH05471237GA SAH05471239GA SAH05471240GA SAH05471241GA SAH05471242GA SAH05471244GA SAH05471247GA SAH05471248GA SAH05471250GA SAH05471251GA SAH05471252GA SAH05471253GA SAH05471254GA SAH05471255GA SAH05471256GA SAH05471257GA SAH05471259GA SAH05471260GA SAH05471261GA SAH05471262GA SAH05471263GA SAH05471264GA SAH05471265GA SAH05471266GA SAH05471267GA SAH05471271GA SAH05471272GA SAH05471273GA SAH05471274GA SAH05471275GA SAH05471276GA SAH05471277GA SAH05471278GA SAH05471279GA SAH05471280GA SAH05471281GA SAH05471282GA SAH05481285GA SAH05481286GA SAH05481287GA SAH05481288GA SAH05481289GA SAH05481291GA SAH05481292GA SAH05481293GA SAH05481294GA SAH05481297GA SAH05481298GA SAH05481299GA SAH05481302GA SAH05481303GA SAH05481304GA SAH05481305GA SAH05481306GA SAH05481307GA SAH05481308GA SAH05481309GA SAH05481310GA SAH05481311GA SAH05481312GA SAH05481313GA SAH05481314GA SAH05491315GA SAH05491317GA SAH05491319GA SAH05491320GA SAH05491321GA SAH05491322GA SAH05491323GA SAH05491324GA SAH05491325GA SAH05491326GA SAH05491327GA SAH05491328GA SAH05491329GA SAH05491330GA SAH05491332GA SAH05491333GA SAH05491334GA SAH05491335GA SAH05491336GA SAH05491338GA SAH05491339GA SAH05491340GA SAH05491341GA SAH05491342GA SAH05491343GA SAH05491344GA SAH05491345GA SAH05491346GA SAH05491347GA SAH05491348GA SAH05491349GA SAH05491351GA SAH05491352GA SAH05491353GA SAH05491354GA SAH05491356GA SAH05491358GA SAH05491359GA SAH05491360GA SAH05491363GA SAH05491364GA SAH05501365GA SAH05501366GA SAH05501367GA SAH05501368GA SAH05501369GA SAH05501370GA SAH05501371GA SAH05501372GA SAH05501373GA SAH05501374GA SAH05501375GA SAH05501376GA SAH05501377GA SAH05501379GA SAH05501380GA SAH05501381GA SAH05501382GA SAH05501383GA SAH05501384GA SAH05501385GA SAH05501386GA SAH05501387GA SAH05501388GA SAH05501389GA SAH05511390GA SAH05511391GA SAH05511394GA SAH05511395GA SAH05511396GA SAH05511397GA SAH05511399GA SAH05511400GA SAH05511401GA SAH05511402GA SAH05511403GA SAH05511404GA SAH05521405GA SAH05521406GA SAH05521407GA SAH05521408GA SAH05521410GA SAH05521411GA SAH05521412GA SAH05521413GA SAH05521414GA SAH05521415GA SAH05521416GA SAH05521417GA SAH05521418GA SAH05521419GA SAH05521420GA SAH05521421GA SAH05521422GA SAH05521423GA SAH05521424GA SAH05521425GA SAH06011426GA SAH06011427GA SAH06011428GA SAH06011429GA SAH06011430GA SAH06011431GA SAH06011432GA SAH06011433GA SAH06021435GA SAH06021436GA SAH06021437GA SAH06021438GA SAH06021439GA SAH06021440GA SAH06021441GA SAH06021443GA SAH06021444GA SAH06021445GA SAH06021446GA SAH06021447GA SAH06021449GA SAH06021450GA SAH06021451GA SAH06021452GA SAH06021453GA SAH06021454GA SAH06021455GA SAH06021456GA SAH06021457GA SAH06021458GA SAH06021459GA SAH06021460GA SAH06021461GA SAH06021462GA SAH06021463GA SAH06041465GA SAH06041466GA SAH06041467GA SAH06041468GA SAH06041469GA SAH06041470GA SAH06041471GA SAH06041473GA SAH06041475GA SAH06041476GA SAH06041477GA SAH06041478GA SAH06041479GA SAH06041480GA SAH06041481GA SAH06041482GA SAH06041483GA SAH06041485GA SAH06041486GA SAH06041487GA SAH06041490GA SAH06041491GA SAH06041492GA SAH06041493GA SAH06051494GA SAH06051496GA SAH06051497GA SAH06051498GA SAH06051499GA SAH06051500GA SAH06051501GA SAH06051502GA SAH06051503GA SAH06051504GA SAH06051505GA SAH06051506GA SAH06051507GA SAH06051508GA SAH06051509GA SAH06051510GA SAH06051511GA SAH06051512GA SAH06051513GA SAH06051514GA SAH06051515GA SAH06051516GA SAH06051517GA SAH06051518GA SAH06051519GA SAH06051520GA SAH06051521GA SAH06061522GA SAH06061523GA SAH06061525GA SAH06061527GA SAH06061528GA SAH06061529GA SAH06061530GA SAH06061531GA SAH06061532GA SAH06061533GA SAH06061534GA SAH06061535GA SAH06071536GA SAH06071537GA SAH06071538GA SAH06071539GA SAH06071540GA SAH06071541GA SAH06071542GA SAH06071543GA SAH06071545GA SAH06071546GA SAH06071547GA SAH06071550GA SAH06071552GA SAH06081553GA SAH06081554GA SAH06081556GA SAH06081560GA SAH06081562GA SAH06081563GA SAH06081564GA SAH06081565GA SAH06081566GA SAH06081568GA SAH06081569GA SAH06081570GA SAH06081571GA SAH06081572GA SAH06081573GA SAH06081575GA SAH06081576GA SAH06081579GA SAH06081580GA SAH06081593GA SAH06091582GA SAH06091583GA SAH06091584GA SAH06091588GA SAH06091589GA SAH06091590GA SAH06091591GA SAH06091592GA SAH06091593GA SAH06091594GA SAH06091595GA SAH06091596GA SAH06091597GA SAH06091598GA SAH06091599GA SAH06091600GA SAH06091601GA SAH06091603GA SAH06091604GA SAH06091605GA SAH06091606GA SAH06091607GA SAH06091608GA SAH06091609GA SAH06091610GA SAH06091611GA SAH06091612GA SAH06091614GA SAH06091615GA SAH06091616GA SAH06091617GA SAH06091619GA SAH06091620GA SAH06091621GA SAH06091622GA SAH06091623GA SAH06091624GA SAH06091625GA SAH06091626GA SAH06091627GA SAH06091628GA SAH06091629GA SAH06091630GA SAH06091631GA SAH06091632GA SAH06101633GA SAH06101634GA SAH06101635GA SAH06101636GA SAH06101637GA SAH06101638GA SAH06101639GA SAH06101640GA SAH06101641GA SAH06101642GA SAH06101643GA SAH06101645GA SAH06101646GA SAH06101647GA SAH06101648GA SAH06101649GA SAH06101650GA SAH06101652GA SAH06101653GA SAH06101655GA SAH06101656GA SAH06101657GA SAH06101658GA SAH06101659GA SAH06101660GA SAH06101661GA SAH06101662GA SAH06101663GA SAH06101665GA SAH06101666GA SAH06101667GA SAH06101668GA SAH06101669GA SAH06101670GA SAH06101671GA SAH06101672GA SAH06101673GA SAH06101674GA SAH06101675GA SAH06101676GA SAH06101677GA SAH06101678GA SAH06101679GA SAH06101680GA SAH06101681GA SAH06101682GA SAH06101683GA SAH06101684GA SAH06101685GA SAH06101686GA SAH06101687GA SAH06101688GA SAH06101689GA SAH06101690GA SAH06101691GA SAH06101692GA SAH06101693GA SAH06101694GA SAH06121695GA SAH06121696GA SAH06121698GA SAH06121699GA SAH06121700GA SAH06121701GA SAH06121702GA SAH06121703GA SAH06121704GA SAH06121705GA SAH06121706GA SAH06121707GA SAH06121708GA SAH06121709GA SAH06121710GA SAH06121712GA SAH06121713GA SAH06121714GA SAH06121715GA SAH06121717GA SAH06121718GA SAH06121720GA SAH06121721GA SAH06121722GA SAH06121723GA SAH06121725GA SAH06121726GA SAH06121727GA SAH06121728GA SAH06121729GA SAH06121730GA SAH06121731GA SAH06121732GA SAH06121733GA SAH06141734GA SAH06141736GA SAH06141737GA SAH06141738GA SAH06141739GA SAH06141740GA SAH06141742GA SAH06141743GA SAH06141744GA SAH06141745GA SAH06141746GA SAH06141747GA SAH06141748GA SAH06141749GA SAH06141751GA SAH06141752GA SAH06141753GA SAH06141754GA SAH06141755GA SAH06141756GA SAH06141757GA SAH06141758GA SAH06141759GA SAH06141761GA SAH06141762GA SAH06141763GA SAH06151764GA SAH06151765GA SAH06151766GA SAH06151767GA SAH06151768GA SAH06151771GA SAH06151772GA SAH06171776GA SAH06171777GA SAH06171778GA SAH06171779GA SAH06171780GA SAH06171782GA SAH06171783GA SAH06171784GA SAH06171786GA SAH06171788GA SAH06171789GA SAH06171790GA SAH06171792GA SAH06171793GA SAH06171794GA SAH06171795GA SAH06171796GA SAH06171797GA SAH06171798GA SAH06171799GA SAH06171801GA SAH06171802GA SAH06171803GA SAH06171805GA SAH06171806GA SAH06171807GA SAH06171808GA SAH06171809GA SAH06171810GA SAH06171811GA SAH06171813GA SAH06181814GA SAH06181815GA SAH06181816GA SAH06181817GA SAH06181818GA SAH06181819GA SAH06181820GA SAH06181822GA SAH06181823GA SAH06181825GA SAH06181827GA SAH06181828GA SAH06181829GA SAH06181830GA SAH06181831GA SAH06181832GA SAH06181833GA SAH06181834GA SAH06181835GA SAH06181837GA SAH06181838GA SAH06181839GA SAH06181840GA SAH06181841GA SAH06181842GA SAH06181843GA SAH06181844GA SAH06181845GA SAH06181846GA SAH06181847GA SAH06181848GA SAH06181849GA SAH06181850GA SAH06181851GA SAH06181852GA SAH06181853GA SAH06191854GA SAH06191856GA SAH06191857GA SAH06191858GA SAH06191862GA SAH06191863GA SAH06191864GA SAH06191865GA SAH06191866GA SAH06191867GA SAH06191868GA SAH06191869GA SAH06191871GA SAH06191872GA SAH06191873GA SAH06241874GA SAH06241876GA SAH06241877GA SAH06241878GA SAH06241879GA SAH06241880GA SAH06241882GA SAH06241883GA SAH06241884GA SAH06241885GA SAH06241886GA SAH06241887GA SAH06241888GA SAH06241889GA SAH06241891GA SAH06241892GA SAH06241893GA SAH06241895GA SAH06241897GA SAH06241899GA SAH06241900GA SAH06241901GA SAH06241904GA SAH06241905GA SAH06241906GA SAH06241907GA SAH06241908GA SAH06241909GA SAH06241910GA SAH06241911GA SAH06241912GA SAH06241913GA SAH06241914GA SAH06241916GA SAH06241917GA SAH06241918GA SAH06241920GA SAH06241921GA SAH06241922GA SAH06241923GA SAH06241924GA SAH06241925GA SAH06241926GA SAH06241928GA SAH06241930GA SAH06241931GA SAH06241932GA SAH06241933GA SAH06241934GA SAH06241935GA SAH06241936GA SAH06241937GA SAH06241938GA SAH06241939GA SAH06241940GA SAH06241941GA SAH06241949GA SAH06241951GA SAH06241952GA SAH06241954GA SAH06241955GA SAH06241956GA SAH06241957GA SAH06241958GA SAH06241959GA SAH06241960GA SAH06241962GA SAH06241963GA SAH06241965GA SAH06241966GA SAH06241967GA SAH06241969GA SAH06241971GA SAH06291976GA SAH06291977GA SAH06291978GA SAH06291979GA SAH06291980GA SAH06291981GA SAH06291982GA SAH06291983GA SAH06291985GA SAH06291986GA SAH06291987GA SAH06291988GA SAH06291989GA SAH06291990GA SAH06291991GA SAH06291992GA SAH06291993GA SAH06291997GA SAH06291998GA SAH06291999GA SAH06292000GA SAH06292001GA SAH06292002GA SAH06292004GA SAH06292005GA SAH06292006GA SAH06292007GA SAH06292008GA SAH06292009GA SAH06292010GA SAH06292012GA SAH06292013GA SAH06332014GA SAH06332015GA SAH06332016GA SAH06332017GA SAH06332018GA SAH06332019GA SAH06332020GA SAH06332021GA SAH06332022GA SAH06332023GA SAH06332025GA SAH06332026GA SAH06332027GA SAH06332029GA SAH06332031GA SAH06332032GA SAH06332033GA SAH06332034GA SAH06332036GA SAH06332037GA SAH06332038GA SAH06332039GA SAH06332040GA SAH06332041GA SAH06332042GA SAH06332043GA SAH06342044GA SAH06342045GA SAH06342046GA SAH06342047GA SAH06342048GA SAH06342049GA SAH06342051GA SAH06342052GA SAH06342053GA SAH06342054GA SAH06342055GA SAH06342056GA SAH06342057GA SAH06342058GA SAH06342059GA SAH06342060GA SAH06342061GA SAH06342062GA SAH06342063GA SAH06382064GA SAH06382065GA SAH06382067GA SAH06382068GA SAH06382069GA SAH06382070GA SAH06382072GA SAH06382073GA SAH06382074GA SAH06382075GA SAH06382076GA SAH06382077GA SAH06382078GA SAH06382079GA SAH06382080GA SAH06382081GA SAH06382082GA SAH06382083GA SAH06382084GA SAH06382085GA SAH06382086GA SAH06382087GA SAH06382088GA SAH06382090GA SAH06382091GA SAH06382092GA SAH06392094GA SAH06392095GA SAH06392096GA SAH06392097GA SAH06392098GA SAH06392099GA SAH06392100GA SAH06392101GA SAH06392102GA SAH06392103GA SAH06402104GA SAH06402105GA SAH06402106GA SAH06402107GA SAH06402108GA SAH06402109GA SAH06402110GA SAH06402111GA SAH06402112GA SAH06402113GA SAH06402114GA SAH06402116GA SAH06402117GA SAH06402118GA SAH06402119GA SAH06402120GA SAH06402121GA SAH06402122GA SAH06402123GA SAH06412124GA SAH06412125GA SAH06412126GA SAH06412127GA SAH06412128GA SAH06412129GA SAH06412130GA SAH06412131GA SAH06412132GA SAH06412133GA SAH06412134GA SAH06412135GA SAH06412136GA SAH06412137GA SAH06412138GA SAH06412139GA SAH06412141GA SAH06412142GA SAH06412143GA SAH06412144GA SAH06412145GA SAH06412146GA SAH06412147GA SAH06412148GA SAH06412150GA SAH06412151GA SAH06412152GA SAH06412153GA SAH06422154GA SAH06422155GA SAH06422156GA SAH06422157GA SAH06422158GA SAH06422159GA SAH06422160GA SAH06422162GA SAH06422163GA SAH06422164GA SAH06422165GA SAH06422166GA SAH06422167GA SAH06422170GA SAH06422171GA SAH06422172GA SAH06422174GA SAH06422175GA SAH06422176GA SAH06422177GA SAH06422178GA SAH06422179GA SAH06422180GA SAH06422182GA SAH06432184GA SAH06432186GA SAH06432187GA SAH06432188GA SAH06432189GA SAH06432190GA SAH06432191GA SAH06432192GA SAH06432193GA SAH06432195GA SAH06432196GA SAH06432197GA SAH06432199GA SAH06432201GA SAH06432202GA SAH06432203GA SAH06432204GA SAH06432205GA SAH06432206GA SAH06432207GA SAH06432209GA SAH06432210GA SAH06432211GA SAH06432212GA SAH06442214GA SAH06442215GA SAH06442216GA SAH06442217GA SAH06442218GA SAH06442219GA SAH06442220GA SAH06442222GA SAH06442223GA SAH06442224GA SAH06442226GA SAH06442228GA SAH06442229GA SAH06442230GA SAH06442231GA SAH06442232GA SAH06442233GA SAH06442234GA SAH06442235GA SAH06442236GA SAH06442238GA SAH06442239GA SAH06442240GA SAH06442241GA SAH06442242GA SAH06442243GA SAH06452244GA SAH06452245GA SAH06452246GA SAH06452247GA SAH06452248GA SAH06452249GA SAH06452250GA SAH06452251GA SAH06452252GA SAH06452253GA SAH06452256GA SAH06452257GA SAH06452258GA SAH06452259GA SAH06452261GA SAH06452262GA SAH06452263GA SAH06452264GA SAH06452266GA SAH06452268GA SAH06452269GA SAH06452270GA SAH06452271GA SAH06452272GA SAH06452273GA SAH06462275GA SAH06462276GA SAH06462277GA SAH06462278GA SAH06462279GA SAH06462280GA SAH06462281GA SAH06462285GA SAH06462286GA SAH06462287GA SAH06462289GA SAH06462290GA SAH06462291GA SAH06462292GA SAH06462293GA SAH06462294GA SAH06462295GA SAH06462296GA SAH06462297GA SAH06462298GA SAH06462299GA SAH06462300GA SAH06462301GA SAH06462302GA SAH06462303GA SAH06462304GA SAH06462305GA SAH06462306GA SAH06462307GA SAH06462308GA SAH06462309GA SAH06462310GA SAH06462311GA SAH06462312GA SAH06462313GA SAH06482314GA SAH06482315GA SAH06482319GA SAH06482321GA SAH06482322GA SAH06482323GA SAH06482324GA SAH06482325GA SAH06482326GA SAH06482327GA SAH06482329GA SAH06482338GA SAH06482339GA SAH06482340GA SAH06482341GA SAH06502346GA SAH06502347GA SAH06502352GA SAH06502353GA SAH06502354GA SAH06502355GA SAH06502356GA SAH06502357GA SAH06502359GA SAH06502361GA SAH06502362GA SAH06502363GA SAH06502364GA SAH06502365GA SAH06502366GA SAH06502367GA SAH06502368GA SAH06502369GA SAH06502370GA SAH06502371GA SAH06522372GA SAH06522373GA SAH06522374GA SAH06522375GA SAH06522376GA SAH06522377GA SAH06522378GA SAH06522379GA SAH06522380GA SAH06522381GA SAH06522382GA SAH06522383GA SAH06522384GA SAH06522386GA SAH06522387GA SAH06522389GA SAH06522390GA SAH06522391GA SAH06522393GA SAH06522394GA SAH06522395GA SAH06522396GA SAH06522397GA SAH06522398GA SAH06522400GA SAH06522401GA SAH06522402GA SAH06522404GA SAH06522405GA SAH06522406GA SAH06522407GA SAH06522408GA SAH06522409GA SAH06522410GA SAH06522411GA SAH06522412GA SAH06522413GA SAH06522414GA SAH06522415GA SAH06522417GA SAH06522418GA SAH06522420GA SAH06522421GA SAH06522422GA SAH06522423GA SAH06522424GA SAH06522425GA SAH06522427GA SAH06522428GA SAH06522429GA SAH06522430GA SAH06522431GA SAH07012432GA SAH07012434GA SAH07012435GA SAH07012436GA SAH07012437GA SAH07012438GA SAH07012439GA SAH07012440GA SAH07012441GA SAH07012442GA SAH07012445GA SAH07012448GA SAH07012450GA SAH07012451GA SAH07022452GA SAH07022453GA SAH07022454GA SAH07022457GA SAH07022459GA SAH07022460GA SAH07022462GA SAH07022463GA SAH07022465GA SAH07022467GA SAH07022468GA SAH07022469GA SAH07022470GA SAH07022472GA SAH07022473GA SAH07022474GA SAH07022475GA SAH07022476GA SAH07022477GA SAH07022478GA SAH07022479GA SAH07022480GA SAH07022481GA SAH07022482GA SAH07022483GA SAH07022484GA SAH07022485GA SAH07022486GA SAH07022487GA SAH07022488GA SAH07022489GA SAH07022490GA SAH07022491GA SAH07052494GA SAH07052495GA SAH07052496GA SAH07052497GA SAH07052498GA SAH07052499GA SAH07052500GA SAH07052501GA SAH07052502GA SAH07052503GA SAH07052504GA SAH07052505GA SAH07052506GA SAH07052507GA SAH07052508GA SAH07052509GA SAH07052510GA SAH07052511GA SAH07052512GA SAH07052515GA SAH07052516GA SAH07052518GA SAH07052520GA SAH07052522GA SAH07052523GA SAH07052531GA SAH07052535GA SAH07052543GA SAH07052551GA SAH07062524GA SAH07062525GA SAH07062526GA SAH07062527GA SAH07062528GA SAH07062529GA SAH07062530GA SAH07062532GA SAH07062534GA SAH07062537GA SAH07062540GA SAH07062541GA SAH07062542GA SAH07062544GA SAH07062545GA SAH07062548GA SAH07062550GA SAH07062552GA SAH07062553GA SAH07072554GA SAH07072555GA SAH07072556GA SAH07072557GA SAH07072558GA SAH07072559GA SAH07072560GA SAH07072561GA SAH07072562GA SAH07072563GA SAH07072566GA SAH07072567GA SAH07072569GA SAH07072570GA SAH07072571GA SAH07072573GA SAH07072574GA SAH07072576GA SAH07072577GA SAH07072578GA SAH07072579GA SAH07072580GA SAH07072581GA SAH07072582GA SAH07102583GA SAH07102584GA SAH07102585GA SAH07102586GA SAH07102587GA SAH07102588GA SAH07102589GA SAH07102590GA SAH07102591GA SAH07102592GA SAH07102593GA SAH07102594GA SAH07102595GA SAH07102596GA SAH07102597GA SAH07102598GA SAH07102599GA SAH07102600GA SAH07102601GA SAH07102602GA SAH07102603GA SAH07102604GA SAH07102607GA SAH07102608GA SAH07102609GA SAH07102610GA SAH07111313GA SAH07142612GA SAH07142614GA SAH07142615GA SAH07142616GA SAH07142617GA SAH07142619GA SAH07142620GA SAH07142621GA SAH07142622GA SAH07142623GA SAH07142624GA SAH07142625GA SAH07142626GA SAH07142627GA SAH07142628GA SAH07142631GA SAH07142632GA SAH07142633GA SAH07142635GA SAH07142636GA SAH07142637GA SAH07142639GA SAH07142640GA SAH07142641GA SAH07152642GA SAH07152644GA SAH07152646GA SAH07152647GA SAH07152648GA SAH07152649GA SAH07152650GA SAH07152651GA SAH07152656GA SAH07152658GA SAH07152660GA SAH07152663GA SAH07152664GA SAH07152666GA SAH07152667GA SAH07152668GA SAH07152671GA SAH07172673GA SAH07172674GA SAH07172675GA SAH07172676GA SAH07172678GA SAH07172679GA SAH07172680GA SAH07172681GA SAH07172682GA SAH07172683GA SAH07172686GA SAH07172687GA SAH07172688GA SAH07172689GA SAH07172690GA SAH07172692GA SAH07172693GA SAH07172694GA SAH07172695GA SAH07172696GA SAH07172697GA SAH07172698GA SAH07172699GA SAH07172700GA SAH07172701GA SAH07202702GA SAH07202703GA SAH07202704GA SAH07202707GA SAH07202708GA SAH07202709GA SAH07202710GA SAH07202711GA SAH07202713GA SAH07202714GA SAH07202715GA SAH07202716GA SAH07202717GA SAH07202721GA SAH07202722GA SAH07202723GA SAH07202724GA SAH07202725GA SAH07202726GA SAH07202727GA SAH07202728GA SAH07202729GA SAH07202730GA SAH07202731GA SAH07212733GA SAH07212734GA SAH07212735GA SAH07212736GA SAH07212737GA SAH07212738GA SAH07212739GA SAH07212740GA SAH07212741GA SAH07212742GA SAH07212743GA SAH07212744GA SAH07212745GA SAH07212746GA SAH07212747GA SAH07212748GA SAH07212749GA SAH07212750GA SAH07212751GA SAH07212752GA SAH07212753GA SAH07212754GA SAH07212755GA SAH07212756GA SAH07212757GA SAH07212758GA SAH07212759GA SAH07212760GA SAH07212761GA SAH07212762GA SAH07232763GA SAH07232764GA SAH07232765GA SAH07232766GA SAH07232767GA SAH07232770GA SAH07232771GA SAH07232772GA SAH07232774GA SAH07232775GA SAH07232776GA SAH07232777GA SAH07232778GA SAH07232779GA SAH07232780GA SAH07232781GA SAH07232782GA SAH07232783GA SAH07232784GA SAH07232785GA SAH07232786GA SAH07232787GA SAH07232788GA SAH07232789GA SAH07232790GA SAH07232791GA SAH07282792GA SAH07282793GA SAH07282794GA SAH07282795GA SAH07282796GA SAH07282798GA SAH07282800GA SAH07282801GA SAH07282802GA SAH07282803GA SAH07282804GA SAH07282805GA SAH07282806GA SAH07282808GA SAH07282809GA SAH07282810GA SAH07282811GA SAH07282812GA SAH07282813GA SAH07282814GA SAH07282815GA SAH07282816GA SAH07282817GA SAH07282819GA SAH07282820GA SAH07282821GA SAH07292822GA SAH07292823GA SAH07292824GA SAH07292825GA SAH07292827GA SAH07292828GA SAH07292829GA SAH07292830GA SAH07292831GA SAH07292832GA SAH07292836GA SAH07292837GA SAH07292838GA SAH07292839GA SAH07292841GA SAH07292842GA SAH07292843GA SAH07292844GA SAH07292847GA SAH07292848GA SAH07292850GA SAH07292851GA SAH07302862GA SAH07302865GA SAH07302866GA SAH07302867GA SAH07302868GA SAH07302869GA SAH07302870GA SAH07302871GA SAH07302872GA SAH07302873GA SAH07302875GA SAH07302876GA SAH07302877GA SAH07302879GA SAH07302880GA SAH07302881GA SAH07312882GA SAH07312885GA SAH07312886GA SAH07312887GA SAH07312891GA SAH07312892GA SAH07312893GA SAH07312894GA SAH07312896GA SAH07312897GA SAH07312898GA SAH07312899GA SAH07312900GA SAH07312901GA SAH07312902GA SAH07312903GA SAH07312904GA SAH07312905GA SAH07312907GA SAH07312908GA SAH07312909GA SAH07312911GA SAH07342912GA SAH07342913GA SAH07342914GA SAH07342916GA SAH07342917GA SAH07342918GA SAH07342919GA SAH07342921GA SAH07342922GA SAH07342923GA SAH07342924GA SAH07342925GA SAH07342926GA SAH07342927GA SAH07342928GA SAH07342929GA SAH07342930GA SAH07342931GA SAH07342932GA SAH07342933GA SAH07342934GA SAH07342935GA SAH07342936GA SAH07342937GA SAH07342938GA SAH07342939GA SAH07342940GA SAH07342941GA SAH07372942GA SAH07372943GA SAH07372944GA SAH07372945GA SAH07372946GA SAH07372947GA SAH07372948GA SAH07372949GA SAH07372951GA SAH07372952GA SAH07372953GA SAH07372954GA SAH07372955GA SAH07372956GA SAH07372957GA SAH07372958GA SAH07372959GA SAH07372960GA SAH07372962GA SAH07372964GA SAH07372966GA SAH07372968GA SAH07372970GA SAH07372972GA SAH07372973GA SAH07372974GA SAH07372975GA SAH07372976GA SAH07372977GA SAH07372978GA SAH07372980GA SAH07372981GA SAH07372982GA SAH07372983GA SAH07372984GA SAH07372985GA SAH07372986GA SAH07372987GA SAH07372988GA SAH07372989GA SAH07372990GA SAH07372991GA SAH07372992GA SAH07372993GA SAH07372994GA SAH07372995GA SAH07372996GA SAH07372997GA SAH07372998GA SAH07372999GA SAH07373000GA SAH07403001GA SAH07403002GA SAH07403003GA SAH07403005GA SAH07403012GA SAH07403013GA SAH07403014GA SAH07403015GA SAH07403016GA SAH07403017GA SAH07403018GA SAH07403019GA SAH07403020GA SAH07403021GA SAH07403022GA SAH07403023GA SAH07403024GA SAH07403025GA SAH07403027GA SAH07403028GA SAH07403029GA SAH07403031GA SAH07403032GA SAH07403033GA SAH07403035GA SAH07403036GA SAH07403037GA SAH07403038GA SAH07403039GA SAH07403040GA SAH07403041GA SAH07403043GA SAH07403044GA SAH07403046GA SAH07403047GA SAH07403048GA SAH07403049GA SAH07403050GA SAH07403052GA SAH07403053GA SAH07403054GA SAH07403055GA SAH07403056GA SAH07403057GA SAH07403059GA SAH07403061GA SAH07403062GA SAH07403063GA SAH07403064GA SAH07423066GA SAH07423067GA SAH07423069GA SAH07423070GA SAH07423071GA SAH07423075GA SAH07423076GA SAH07423077GA SAH07423078GA SAH07423079GA SAH07423080GA SAH07423081GA SAH07423082GA SAH07423085GA SAH07423087GA SAH07423088GA SAH07423089GA SAH07423090GA SAH07423091GA SAH07423092GA SAH07423093GA SAH07423094GA SAH07423095GA SAH07423096GA SAH07423097GA SAH07423098GA SAH07423099GA SAH07423100GA SAH07423102GA SAH07423103GA SAH07423104GA SAH07423105GA SAH07423106GA SAH07423107GA SAH07423108GA SAH07423111GA SAH07423113GA SAH07423115GA SAH07423117GA SAH07423118GA SAH07423119GA SAH07423120GA SAH07423123GA SAH07443125GA SAH07443127GA SAH07443129GA SAH07443131GA SAH07443132GA SAH07443133GA SAH07443134GA SAH07443135GA SAH07443136GA SAH07443138GA SAH07443139GA SAH07443140GA SAH07443142GA SAH07443143GA SAH07443144GA SAH07443145GA SAH07443146GA SAH07443148GA SAH07443149GA SAH07443150GA SAH07443154GA SAH07463155GA SAH07463156GA SAH07463157GA SAH07463158GA SAH07463159GA SAH07463161GA SAH07463162GA SAH07463164GA SAH07463165GA SAH07463166GA SAH07463167GA SAH07463168GA SAH07463169GA SAH07463170GA SAH07463171GA SAH07463172GA SAH07463173GA SAH07463174GA SAH07463175GA SAH07463176GA SAH07463177GA SAH07463178GA SAH07463179GA SAH07463180GA SAH07463181GA SAH07463183GA SAH07463184GA SAH07483185GA SAH07483186GA SAH07483187GA SAH07483188GA SAH07483189GA SAH07483190GA SAH07483191GA SAH07483192GA SAH07483193GA SAH07483194GA SAH07483195GA SAH07483196GA SAH07483197GA SAH07483198GA SAH07483199GA SAH07483200GA SAH07483201GA SAH07483202GA SAH07483203GA SAH07483204GA SAH07483205GA SAH07483206GA SAH07483207GA SAH07483208GA SAH07483209GA SAH07483210GA SAH07483211GA SAH07483212GA SAH07483213GA SAH07483214GA SAH07483215GA SAH07483216GA SAH07483217GA SAH07483218GA SAH07483219GA SAH07483220GA SAH07483221GA SAH07483222GA SAH07483223GA SAH07483224GA SAH07483225GA SAH07483226GA SAH07483227GA SAH07483228GA SAH07483229GA SAH07483230GA SAH07483231GA SAH07483232GA SAH07483233GA SAH07483234GA SAH07493235GA SAH07493236GA SAH07493237GA SAH07493238GA SAH07493239GA SAH07493240GA SAH07493241GA SAH07493242GA SAH07493243GA SAH07493244GA SAH07493245GA SAH07493246GA SAH07493248GA SAH07493249GA SAH07493251GA SAH07493252GA SAH07493253GA SAH07493254GA SAH07493255GA SAH07493256GA SAH07493257GA SAH07493258GA SAH07493259GA SAH07493260GA SAH07493261GA SAH07493262GA SAH07493263GA SAH07493264GA SAH07493265GA SAH07493266GA SAH07493267GA SAH07493268GA SAH07493269GA SAH07493271GA SAH07493273GA SAH07493275GA SAH07493276GA SAH07493277GA SAH07493278GA SAH07493279GA SAH07493280GA SAH07493281GA SAH07493282GA SAH07493283GA SAH07493285GA SAH07493286GA SAH07493287GA SAH07493288GA SAH07493289GA SAH07493290GA SAH07493291GA SAH07493292GA SAH07493293GA SAH07493294GA SAH07493295GA SAH07493296GA SAH07493297GA SAH07493298GA SAH07493299GA SAH07493300GA SAH07493301GA SAH07493302GA SAH07493303GA SAH07493304GA SAH07493305GA SAH07493306GA SAH07493307GA SAH07493308GA SAH07493309GA SAH07493310GA SAH07493311GA SAH07493312GA SAH07493313GA SAH07493314GA SAH07503316GA SAH07503317GA SAH07503318GA SAH07503320GA SAH08023321GA SAH08023322GA SAH08023323GA SAH08023324GA SAH08023325GA SAH08023326GA SAH08023327GA SAH08023328GA SAH08023329GA SAH08023331GA SAH08023333GA SAH08023334GA SAH08023335GA SAH08023337GA SAH08023338GA SAH08023339GA SAH08023341GA SAH08023342GA SAH08023343GA SAH08023344GA SAH08023345GA SAH08023349GA SAH08023350GA SAH08023351GA SAH08023352GA SAH08023353GA SAH08023354GA SAH08023356GA SAH08023358GA SAH08023359GA SAH08023360GA SAH08023361GA SAH08023362GA SAH08023363GA SAH08023366GA SAH08023367GA SAH08023368GA SAH08023369GA SAH08023370GA SAH08023371GA SAH08023372GA SAH08023373GA SAH08023374GA SAH08033375GA SAH08033377GA SAH08033378GA SAH08033379GA SAH08033381GA SAH08033383GA SAH08033384GA SAH08033386GA SAH08033388GA SAH08033389GA SAH08033390GA SAH08033391GA SAH08033392GA SAH08033393GA SAH08133395GA SAH08133396GA SAH08133397GA SAH08133399GA SAH08133400GA SAH08133401GA SAH08133402GA SAH08133403GA SAH08133405GA SAH08133406GA SAH08133407GA SAH08133408GA SAH08133409GA SAH08133410GA SAH08133411GA SAH08133412GA SAH08133414GA SAH08133415GA SAH08133416GA SAH08133417GA SAH08133418GA SAH08133419GA SAH08133420GA SAH08133421GA SAH08133422GA SAH08133423GA SAH08133424GA SAH08143425GA SAH08143426GA SAH08143427GA SAH08143428GA SAH08143430GA SAH08143431GA SAH08143432GA SAH08143433GA SAH08143434GA SAH08143435GA SAH08143436GA SAH08143437GA SAH08143439GA SAH08143440GA SAH08143441GA SAH08143442GA SAH08143443GA SAH08143444GA SAH08153445GA SAH08153446GA SAH08153447GA SAH08153448GA SAH08153449GA SAH08153451GA SAH08153452GA SAH08153453GA SAH08153454GA SAH08153456GA SAH08153457GA SAH08153458GA SAH08153461GA SAH08153463GA SAH08153465GA SAH08153466GA SAH08153467GA SAH08153468GA SAH08153469GA SAH08153470GA SAH08153471GA SAH08153472GA SAH08153473GA SAH08173475GA SAH08173477GA SAH08173478GA SAH08173479GA SAH08173480GA SAH08173481GA SAH08173482GA SAH08173483GA SAH08173484GA SAH08173485GA SAH08173486GA SAH08173488GA SAH08173489GA SAH08173490GA SAH08173494GA SAH08173496GA SAH08173497GA SAH08173501GA SAH08173504GA SAH08203505GA SAH08203506GA SAH08203508GA SAH08203509GA SAH08203510GA SAH08203511GA SAH08203512GA SAH08203513GA SAH08203514GA SAH08203515GA SAH08203517GA SAH08203518GA SAH08203519GA SAH08203521GA SAH08203522GA SAH08203523GA SAH08203524GA SAH08203525GA SAH08203526GA SAH08203527GA SAH08203528GA SAH08203529GA SAH08203530GA SAH08203531GA SAH08203532GA SAH08203533GA SAH08203534GA SAH08203535GA SAH08203537GA SAH08203538GA SAH08203539GA SAH08203540GA SAH08203541GA SAH08203543GA SAH08213549GA SAH08213552GA SAH08213554GA SAH08213555GA SAH08213556GA SAH08213557GA SAH08213558GA SAH08213559GA SAH08213560GA SAH08213561GA SAH08213562GA SAH08213563GA SAH08213564GA SAH08213566GA SAH08213567GA SAH08213568GA SAH08213570GA SAH08213573GA SAH08213575GA SAH08213576GA SAH08213577GA SAH08213578GA SAH08213580GA SAH08213581GA SAH08213582GA SAH08213583GA SAH08253585GA SAH08253586GA SAH08253587GA SAH08253588GA SAH08253589GA SAH08253590GA SAH08253591GA SAH08253592GA SAH08253593GA SAH08253594GA SAH08253595GA SAH08253596GA SAH08253597GA SAH08253598GA SAH08253599GA SAH08253600GA SAH08253601GA SAH08253602GA SAH08253603GA SAH08253604GA SAH08253605GA SAH08253606GA SAH08253607GA SAH08253608GA SAH08253609GA SAH08253610GA SAH08253612GA SAH08253613GA SAH08253614GA SAH08253615GA SAH08253616GA SAH08253617GA SAH08253618GA SAH08253619GA SAH08253621GA SAH08253622GA SAH08253623GA SAH08253624GA SAH08283625GA SAH08283626GA SAH08283627GA SAH08283628GA SAH08283629GA SAH08283630GA SAH08283633GA SAH08283634GA SAH08283635GA SAH08283636GA SAH08283637GA SAH08283639GA SAH08283640GA SAH08283641GA SAH08283642GA SAH08283643GA SAH08283644GA SAH08293648GA SAH08293649GA SAH08293651GA SAH08293652GA SAH08293653GA SAH08293654GA SAH08293655GA SAH08293656GA SAH08293657GA SAH08293658GA SAH08293659GA SAH08293663GA SAH08293664GA SAH08303665GA SAH08303666GA SAH08303667GA SAH08303668GA SAH08303671GA SAH08303672GA SAH08303673GA SAH08303674GA SAH08303675GA SAH08303676GA SAH08303677GA SAH08303679GA SAH08303680GA SAH08303681GA SAH08303682GA SAH08303683GA SAH08303684GA SAH08313688GA SAH08313690GA SAH08313691GA SAH08313692GA SAH08313693GA SAH08313694GA SAH08313695GA SAH08313696GA SAH08313698GA SAH08313699GA SAH08313700GA SAH08313701GA SAH08313704GA SAH08343705GA SAH08343706GA SAH08343707GA SAH08343708GA SAH08343709GA SAH08343710GA SAH08343712GA SAH08343713GA SAH08343714GA SAH08343715GA SAH08343716GA SAH08343717GA SAH08343718GA SAH08343720GA SAH08343721GA SAH08343722GA SAH08343723GA SAH08343724GA SAH08353725GA SAH08353726GA SAH08353727GA SAH08353729GA SAH08353730GA SAH08353731GA SAH08353732GA SAH08353733GA SAH08353734GA SAH08353735GA SAH08353736GA SAH08353737GA SAH08353738GA SAH08353739GA SAH08353740GA SAH08353741GA SAH08353742GA SAH08353743GA SAH08353744GA SAH08363745GA SAH08363746GA SAH08363747GA SAH08363748GA SAH08363750GA SAH08363751GA SAH08363752GA SAH08363753GA SAH08363754GA SAH08363755GA SAH08363756GA SAH08363757GA SAH08363759GA SAH08363760GA SAH08363761GA SAH08363762GA SAH08363763GA SAH08363764GA SAH08373767GA SAH08373769GA SAH08373770GA SAH08373771GA SAH08373773GA SAH08373775GA SAH08373776GA SAH08373777GA SAH08373779GA SAH08373780GA SAH08373781GA SAH08373782GA SAH08373783GA SAH08373784GA SAH08373785GA SAH08373787GA SAH08373788GA SAH08373789GA SAH08373791GA SAH08373792GA SAH08373793GA SAH08373794GA SAH08373795GA SAH08373796GA SAH08373797GA SAH08373798GA SAH08373799GA SAH08373800GA SAH08373801GA SAH08373802GA SAH08373803GA SAH08373804GA SAH08373805GA SAH08373806GA SAH08373807GA SAH08373808GA SAH08373809GA SAH08373810GA SAH08373811GA SAH08373812GA SAH08373813GA SAH08373814GA SAH08373815GA SAH08373816GA SAH08373817GA SAH08373818GA SAH08373819GA SAH08373820GA SAH08373821GA SAH08373824GA SAH08383825GA SAH08383826GA SAH08383827GA SAH08383828GA SAH08383829GA SAH08383830GA SAH08383831GA SAH08383832GA SAH08383833GA SAH08383834GA SAH08383835GA SAH08383836GA SAH08383837GA SAH08383838GA SAH08383839GA SAH08383840GA SAH08383841GA SAH08383842GA SAH08383843GA SAH08383844GA SAH08403845GA SAH08403846GA SAH08403847GA SAH08403848GA SAH08403849GA SAH08403850GA SAH08403851GA SAH08403852GA SAH08403853GA SAH08403854GA SAH08403855GA SAH08403856GA SAH08403857GA SAH08403858GA SAH08403859GA SAH08403860GA SAH08403861GA SAH08403862GA SAH08403863GA SAH08403864GA SAH08403866GA SAH08403867GA SAH08403868GA SAH08403869GA SAH08403870GA SAH08403871GA SAH08403872GA SAH08403873GA SAH08403874GA SAH08403875GA SAH08403876GA SAH08403877GA SAH08403878GA SAH08403879GA SAH08403880GA SAH08403881GA SAH08403882GA SAH08403883GA SAH08403884GA SAH08423888GA SAH08423889GA SAH08423890GA SAH08423891GA SAH08423892GA SAH08423893GA SAH08423894GA SAH08423895GA SAH08423896GA SAH08423897GA SAH08423898GA SAH08423899GA SAH08423900GA SAH08423901GA SAH08423902GA SAH08423903GA SAH08423904GA SAH08423905GA SAH08423906GA SAH08423907GA SAH08423908GA SAH08423910GA SAH08423911GA SAH08423912GA SAH08423913GA SAH08423914GA SAH08423915GA SAH08423916GA SAH08423917GA SAH08423918GA SAH08423920GA SAH08423922GA SAH08423923GA SAH08443924GA SAH08443925GA SAH08443926GA SAH08443927GA SAH08443929GA SAH08443930GA SAH08443933GA SAH08443934GA SAH08443935GA SAH08443936GA SAH08443937GA SAH08443938GA SAH08443939GA SAH08443940GA SAH08443941GA SAH08443942GA SAH08443943GA SAH08443944GA SAH08443945GA SAH08443946GA SAH08443947GA SAH08443948GA SAH08443949GA SAH08443950GA SAH08443952GA SAH08443953GA SAH08443954GA SAH08443955GA SAH08443956GA SAH08443957GA SAH08443958GA SAH08443960GA SAH08443961GA SAH08443962GA SAH08443963GA SAH08473964GA SAH08473965GA SAH08473966GA SAH08473967GA SAH08473968GA SAH08473969GA SAH08473970GA SAH08473971GA SAH08473972GA SAH08473973GA SAH08473974GA SAH08473976GA SAH08473977GA SAH08473978GA SAH08473979GA SAH08473980GA SAH08473981GA SAH08473982GA SAH08473983GA SAH08473984GA SAH08473985GA SAH08473987GA SAH08473988GA SAH08473989GA SAH08473990GA SAH08473992GA SAH08473993GA SAH08473994GA SAH08473995GA SAH08473996GA SAH08473998GA SAH08473999GA SAH08474000GA SAH08474001GA SAH08474002GA SAH08474003GA SAH08484004GA SAH08494005GA SAH08494006GA SAH08494007GA SAH08494008GA SAH08494009GA SAH08494010GA SAH08494011GA SAH08494012GA SAH08494013GA SAH08494014GA SAH08494015GA SAH08494016GA SAH08494017GA SAH08494018GA SAH08494019GA SAH08494020GA SAH08494021GA SAH08494022GA SAH08494023GA SAH08494024GA SAH08494025GA SAH08494026GA SAH08494027GA SAH08494028GA SAH08494029GA SAH08494030GA SAH08494031GA SAH08494032GA SAH08494034GA SAH08494035GA SAH08494036GA SAH08494037GA SAH08494038GA SAH08494040GA SAH08494041GA SAH08494042GA SAH08494043GA SAH08494045GA SAH08494046GA SAH08494047GA SAH08494048GA SAH08494049GA SAH08494051GA SAH08494053GA SAH08494054GA SAH08494056GA SAH08494057GA SAH08494058GA SAH08494059GA SAH08494060GA SAH08494062GA SAH08494063GA SAH08504066GA SAH08504067GA SAH08504068GA SAH08504071GA SAH08504072GA SAH08504074GA SAH08504075GA SAH08504076GA SAH08504077GA SAH08504078GA SAH08504081GA SAH08504082GA SAH08504083GA SAH08504084GA SAH08514085GA SAH08514086GA SAH08514087GA SAH08514088GA SAH08514090GA SAH08514091GA SAH08514092GA SAH08514093GA SAH08514094GA SAH08514096GA SAH08514097GA SAH08514098GA SAH08514099GA SAH08514100GA SAH08514102GA SAH08514103GA SAH08514104GA SAH08514107GA SAH08514110GA SAH08514111GA SAH08514112GA SAH08514113GA SAH08514114GA SAH08524115GA SAH08524116GA SAH08524117GA SAH08524118GA SAH08524119GA SAH08524120GA SAH08524121GA SAH08524122GA SAH08524123GA SAH08524124GA SAH08524125GA SAH08524126GA SAH08524127GA SAH08524128GA SAH08524129GA SAH08524130GA SAH08524131GA SAH08524132GA SAH08524133GA SAH08524134GA SAH08524135GA SAH08524136GA SAH08524137GA SAH08524138GA SAH08524139GA SAH08524140GA SAH08524141GA SAH08524142GA SAH08524144GA SAH09014145GA SAH09014146GA SAH09014147GA SAH09014148GA SAH09014149GA SAH09014150GA SAH09014151GA SAH09014152GA SAH09014153GA SAH09014154GA SAH09024156GA SAH09024157GA SAH09024158GA SAH09024159GA SAH09024160GA SAH09024161GA SAH09024162GA SAH09024164GA SAH09024165GA SAH09024166GA SAH09024167GA SAH09024168GA SAH09024169GA SAH09024170GA SAH09024171GA SAH09024172GA SAH09024173GA SAH09024174GA SAH09024175GA SAH09024176GA SAH09024177GA SAH09024178GA SAH09024179GA SAH09024180GA SAH09024181GA SAH09024182GA SAH09024183GA SAH09024184GA SAH09024186GA SAH09024188GA SAH09024189GA SAH09024190GA SAH09024191GA SAH09024192GA SAH09024193GA SAH09024194GA SAH09024196GA SAH09024197GA SAH09024200GA SAH09024201GA SAH09024202GA SAH09024203GA SAH09024204GA SAH09024206GA SAH04430259GA SAH04490364GA SAH04500392GA SAH05010529GA SAH05010560GA SAH05040581GA SAH05140719GA SAH05140724GA SAH05150777GA SAH05150779GA SAH05160834GA SAH05220838GA SAH05220840GA SAH05220845GA SAH05260907GA SAH05260915GA SAH05391088GA SAH05481295GA SAH05481300GA SAH05481301GA SAH05491361GA SAH05491362GA SAH05501378GA SAH06091587GA SAH06101654GA SAH06171791GA SAH06181821GA SAH06241898GA SAH06332035GA SAH06412140GA SAH06452255GA SAH06482330GA SAH06522416GA SAH06522426GA SAH07022464GA SAH07062549GA SAH07152652GA SAH07152654GA SAH07152657GA SAH07232769GA SAH07232773GA SAH07302864GA SAH07423122GA SAH07423124GA SAH07443128GA SAH07443130GA SAH07443141GA SAH07443152GA SAH08023365GA SAH08143429GA SAH08153460GA SAH08153462GA SAH08153464GA SAH08173500GA SAH08173502GA SAH08213584GA SAH08293660GA SAH08293661GA SAH08423885GA SAH08423887GA SAH08443932GA SAH08494039GA SAH08494044GA SAH07072572GA SAH07282818GA SAH07372971GA SAH08213547GA SAH08514108GA SAH08514109GA SAH05110700GA SAH04450264GA SAH04460296GA SAH04470330GA SAH04480344GA SAH04480352GA SAH04480354GA SAH04480358GA SAH04490369GA SAH04500394GA SAH04520468GA SAH05010552GA SAH05040574GA SAH05040580GA SAH05070610GA SAH05080644GA SAH05080652GA SAH05080653GA SAH05080683GA SAH05110708GA SAH05110709GA SAH05280947GA SAH05471243GA SAH05471245GA SAH05471249GA SAH05481296GA SAH06121716GA SAH06171800GA SAH06191855GA SAH06241881GA SAH06241896GA SAH06241902GA SAH06292003GA SAH06422173GA SAH06422183GA SAH06432213GA SAH06502344GA SAH06502345GA SAH06502348GA SAH06502349GA SAH06502350GA SAH06502351GA SAH06522392GA SAH06522399GA SAH07022458GA SAH07052521GA SAH07062546GA SAH07102611GA SAH07142613GA SAH07152670GA SAH07202718GA SAH07292834GA SAH07372979GA SAH07423116GA SAH07443137GA SAH07493247GA SAH07493250GA SAH07503315GA SAH07503319GA SAH08023340GA SAH08023346GA SAH08023348GA SAH08033376GA SAH08033387GA SAH08133404GA SAH08143438GA SAH08173476GA SAH08173503GA SAH08203507GA SAH08203520GA SAH08303670GA SAH08313687GA SAH08443928GA SAH09024185GA SAH09024198GA SAH08504073GA SAH06442237GA SAH06452267GA SAH08514089GA SAH08423886GA SAH05080681GA SAH07463160GA SAH08293645GA SAH08313689GA SAH08373765GA SAH08373766GA SAH08373768GA SAH08373772GA SAH08494050GA SAH08494064GA SAH04410188GA SAH04410203GA SAH04430244GA SAH04460305GA SAH05070611GA SAH05110690GA SAH05110704GA SAH05160814GA SAH05160816GA SAH05160829GA SAH05160832GA SAH05160833GA SAH05220849GA SAH05220860GA SAH05280925GA SAH05280937GA SAH05320988GA SAH05320995GA SAH05391104GA SAH05411112GA SAH05421125GA SAH05431131GA SAH05431161GA SAH05441165GA SAH05441168GA SAH05441175GA SAH05471238GA SAH05471268GA SAH05471270GA SAH05491331GA SAH06041474GA SAH06071551GA SAH06101664GA SAH06121719GA SAH06181824GA SAH06241968GA SAH06432200GA SAH06452265GA SAH06482317GA SAH06482334GA SAH06522388GA SAH06522403GA SAH07012444GA SAH07012449GA SAH07022466GA SAH07292840GA SAH07372961GA SAH07403004GA SAH07403006GA SAH07403009GA SAH07423065GA SAH07423072GA SAH07463182GA SAH07493272GA SAH08033394GA SAH08153450GA SAH08213565GA SAH08343711GA SAH08373790GA SAH08423909GA SAH08423921GA SAH08473975GA SAH08473986GA SAH08514105GA SAH08514106GA SAH09024205GA SAH09024207GA SAH04340032GA SAH04350046GA SAH05290955GA SAH05320989GA SAH05491318GA SAH06241944GA SAH06241947GA SAH06291994GA SAH06291995GA SAH06382066GA SAH07152662GA SAH07172684GS SAH07202732GA SAH07312889GA SAH07342915GA SAH08213550GA SAH08504065GA SAH08524143GA SAH09024199GA SAH05351036GA SAH05411113GA SAH07423086GA SAH04480351GA SAH07302863GA SAH05240874GA SAH06241961GA SAH07443151GA SAH08023347GA SAH05320992GA SAH08293662GA SAH04500388GA SAH08203542GA SAH08213545GA SAH08363749GA SAH08403865GA SAH09024155GA SAH07403034GA SAH07493270GA SAH08023330GA SAH08033385GA SAH07292826GA SAH07302878GA SAH07372963GA SAH07372969GA SAH07403042GA SAH07443147GA SAH07443153GA SAH07493284GA SAH08173495GA SAH08173499GA SAH08213548GA SAH08213569GA SAH08213572GA SAH08213574GA SAH08253611GA SAH08283632GA SAH08313703GA SAH08353728GA SAH08363758GA SAH08494052GA SAH08504070GA SAH06011434GA SAH06041472GA SAH06141735GA SAH06241964GA SAH06241973GA SAH06292011GA SAH06452260GA SAH07152669GA SAH07172685GA SAH07172691GA SAH07202706GA SAH07292833GA SAH07292845GA SAH07292846GA SAH07292852GA SAH07292853GA SAH07292854GA SAH07292855GA SAH07292856GA SAH07292857GA SAH07292858GA SAH07292859GA SAH07292860GA SAH07292861GA SAH07312883GA SAH07312910GA SAH07403026GA SAH08213553GA SAH08293646GA SAH08303678GA SAH08313686GA SAH07423114GA SAH07403008GA SAH07423110GA SAH08173492GA SAH08173493GA SAH08173498GA SAH08373786GA SAH08293647GA SAH08293650GA SAH08303669GA SAH08313685GA SAH08313697GA SAH08313702GA SAH08443931GA SAH04430254GA SAH07403010GA SAH07423068GA SAH07423073GA SAH07423074GA SAH07423083GA SAH07423084GA SAH08423919GA SAH08504069GA SAH07423121GA SAH07312888GA SAH07312895GA SAH07493274GA SAH08203516GA SAH08373774GA SAH04490359GA SAH05451212GA SAH06021448GA SAH06071549GA SAH06101644GA SAH06241919GA SAH06291974GA SAH06291975GA SAH07072564GA SAH07282807GA SAH07292835GA SAH07403058GA SAH05471284GA SAH07142638GA SAH07172672GA SAH07403030GA SAH08213546GA SAH08473991GA SAH08473997GA SAH06081561GA SAH08213551GA SAH06482331GA SAH06482343GA SAH06502360GA SAH08494055GA SAH08494061GA SAH07292849GA SAH07423101GA SAH07312884GA SAH07312890GA SAH07423109GA SAH06363221GA SAH06462283GA SAH06482333GA SAH04340037GA SAH04410187GA SAH04510439GA SAH04520462GA SAH05140736GA SAH05240879GA SAH05240884GA SAH05290962GA SAH05451203GA SAH05491316GA SAH06041488GA SAH06141741GA SAH06141760GA SAH06181836GA SAH06241972GA SAH06382071GA SAH06412149GA SAH06432208GA SAH06502358GA SAH07152645GA SAH07282797GA SAH07403011GA SAH07403045GA SAH07423112GA SAH07463163GA SAH08023336GA SAH08133413GA SAH08203536GA SAH08373822GA SAH08443959GA SAH08504080GA SAH09024195GA SAH07062533GA SAH04290011GA SAH04490370GA SAH05080666GA SAH05150792GA SAH05491350GA SAH07302874GA SAH07372965GA SAH08213579GA SAH04530489GA SAH05521409GA SAH06051495GA SAH06171787GA SAH06171804GA SAH06462274GA SAH06482337GA SAH09024187GA SAH05080676GA SAH06241903GA SAH08153474GA SAH08173491GA SAH08504079GA SAH04470331GA SAH04480334GA SAH04480338GA SAH04480339GA SAH05010548GA sah05010558ga SAH05080663GA SAH05140734GA SAH05160830GA SAH05260918GA SAH05290973GA sah05292295ga SAH05320996GA SAH05481290GA SAH06061524GA SAH06061526GA SAH06071544GA SAH06081557GA SAH06081574GA SAH06121697GA SAH06332028GA SAH06422168GA SAH06522385GA SAH06522419GA SAH07012447GA SAH07022456GA SAH07022471GA SAH07052513GA SAH07052547GA SAH07062536GA SAH07072575GA SAH07102605GA SAH07102606GA SAH07312906GA SAH07342920GA SAH07372950GA SAH07403051GA SAH07403060GA SAH08023332GA SAH08023355GA SAH08033362GA SAH08033380GA SAH08033382GA SAH08133398GA SAH08203544GA SAH08213571GA SAH08283631GA SAH08373778GA SAH08373823GA SAH08494033GA SAH08514095GA SAH08514101GA SAH09024163GA SAH04460299GA SAH04460302GA SAH04460303GA SAH05160805GA SAH07372967GA SAH07403007GA 
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact GE Healthcare Service
800-437-1171
Manufacturer Reason
for Recall
When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
FDA Determined
Cause 2
Software in the Use Environment
Action The firm disseminated the notices by mail on 11/12/2019. The letter identified the security vulnerability of certain GE Central Stations and ApexPro Telemetry servers and provided additional best practices, ensuring the following: 1. MC and IX Networks are isolated; 2. MC and IX Router/Firewalls block incoming traffic, as applicable; 3. Restricted physical access to Central Stations, Telemetry Servers, MC network and IX network; 4. Default passwords are changed as applicable; and 5. Password management best practices are adhered to Ensuring the networks are properly configured and isolated protects against these potential concerns and mitigates the risk.
Quantity in Commerce 11347 devices
Distribution Worldwide distribution.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = DSI and Original Applicant = GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
510(K)s with Product Code = DSI and Original Applicant = GENERAL ELECTRIC MEDICAL SYSTEMS INFORMATION TECHN
-
-