| | Class 2 Device Recall Drill Tapered RP 4.3x10 mm SinglePat |  |
| Date Initiated by Firm | March 14, 2019 |
| Date Posted | February 19, 2020 |
| Recall Status1 |
Terminated 3 on February 15, 2024 |
| Recall Number | Z-1308-2020 |
| Recall Event ID |
84450 |
| 510(K)Number | K990846 |
| Product Classification |
Implant, endosseous, root-form - Product Code DZI
|
| Product | Drill Tapered RP 4.3x10 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants. |
| Code Information |
(01)07332747065633(10)73433 (01)07332747065633(10)73495 Catalog# 37080 |
| FEI Number |
1000125598
|
Recalling Firm/ Manufacturer |
Nobel Biocare Usa Llc 22715 Savi Ranch Pkwy Yorba Linda CA 92887-4609
|
| For Additional Information Contact | 714-282-4800 |
Manufacturer Reason for Recall | Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life. |
FDA Determined Cause 2 | Packaging process control |
| Action | On 3/27/2018 the firm issued a recall notice by email to affected consignees. The recall notification informed the following actions to be taken by the consignees:
1. Identify affected devices as detailed in the Customer Reply Form.
2. Do not use affected devices.
3. Send back the Customer Reply Form and the affected devices.
The firm will replace affected devices free of charge.
This letter needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred.
We kindly ask you to follow instructions provided in this Recall Letter and return the attached Customer Reply Form to the firm via email or fax at 714- 282-5073. |
| Quantity in Commerce | 346 units |
| Distribution | Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DZI
|
|
|
|