• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Winged Sponges

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Winged Spongessee related information
Date Initiated by FirmFebruary 10, 2020
Create DateMarch 20, 2020
Recall Status1 Terminated 3 on February 13, 2023
Recall NumberZ-1545-2020
Recall Event ID 85022
Product Classification Skin prep tray - Product Code OJU
ProductMedline Winged Sponges Presaturated W/PVP REF DYND70289 DYND70289H Expiration 06/30/2022.
Code Information REF DYND70289 Expiration 06/30/2022
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries Inc
Three Lakes Drive
Northfield IL 60093
For Additional Information ContactBenjamin Gora
866-359-1704
Manufacturer Reason
for Recall
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.
FDA Determined
Cause 2
Packaging
ActionMedline notified customers on February 18, 2020 via "Urgent Recall Immediate Action required" letter. The consignees were instructed to complete the response form and return the product. The recall letter identified the affected product and lot numbers. Reason for the correction was also noted in the recall letter.
Quantity in Commerce117,874 units
DistributionUS Nationwide: MA, LA, IL, MO, MI, MN, NJ, KS, GA, NY, PA, SD, CT, TN, MD, OH, VA, FL, TX, KY, CO, IN, AL, CA, MT, OR, AK, AL, DC, IA, ME, NV, NE, SC, WV, WY, RI, NC, AZ, UT, DE and WI.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-