|
Class 2 Device Recall Datascope |
|
Date Initiated by Firm |
July 27, 2020 |
Date Posted |
August 21, 2020 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-2843-2020 |
Recall Event ID |
85823 |
Product Classification |
System, balloon, intra-aortic and control - Product Code DSP
|
Product |
Datascope Intra-Aortic Ballon Catheters (IABs)-TRANS-RAY PLUS 7.5Fr 35cc IAB (Japan) Kit P/N: 0684-00-0607 - Product Usage: Indications For Use: Acute Coronary Syndrome Cardiac and Non-Cardiac Surgery Complications of Heart Failure.
|
Code Information |
Kit Lot Number: 3000047804 3000053322 3000053323 3000060107 3000060108 3000062255 3000067698 3000069727 3000071976 3000073186 3000073680 3000073681 3000074885 3000076710 3000076711 3000077752 3000077753 3000079103 3000079104 3000080873 3000080872 3000081671 3000084267 3000084966 3000084965 3000086799 3000088548 3000088371 3000090599 3000090600 3000090601 3000090602 3000090603 3000092664 3000092663 3000092665 3000096263 3000095467 3000096264 3000096265 3000096266 3000098176 3000098177 3000102408 3000102409 3000103469 3000104520 3000104615 3000106012 3000107758 3000107759 3000110626 3000113155 3000112276 3000113156 |
Recalling Firm/ Manufacturer |
Datascope Corporation 15 Law Dr Fairfield NJ 07004-3206
|
For Additional Information Contact |
SAME 973-244-6100
|
Manufacturer Reason for Recall |
Potential Endotoxin Contamination
|
FDA Determined Cause 2 |
Process change control |
Action |
Datascope/Getinge issued Urgent Medical Device Removal letter (customer specific) on 7/27/20, via FedEx with Signature Proof of Delivery. Letter states reason for recall, health risk and action to take: Records indicate that you received potentially affected IABs identifiable by the lot and serial numbers listed . Please complete the steps below:
Monitor patients for pyrogens reaction/humoral immune response/ coagulation and complement cascades/ inflammation. Monitor and treat any signs of inflammation according to your facilitys protocols and clinical judgment.
¿
Please examine your inventory immediately, remove and quarantine any unexpired affected IAB following the steps below:
Identify any unexpired IAB Kits referencing the part and lot numbers listed in Table 1.
The Kit part and lot numbers can be found on the Outer Shelf Carton Using steps below, Identify any IAB Yfittings with serial numbers listed in Table 1.
¿ Remove tape on one side of the Shelf Carton.
¿ Hinge the box open, keeping the other side taped.
¿ Do not remove any items from the carton. Instead, Lift the Accessories up and locate the YFitting through the clear Mylar overwrap.
¿ Read the serial number off the IAB YFitting ¿ If you find an IAB with a serial number listed in Table 1, immediately place the entire Kit in
quarantine for return and replacement or credit.
¿ If the affected serial number is not found, close the carton and reseal. Cut out and affix one of the enclosed adhesive labels to designate the kit has been checked for affected recall Serial Numbers and may be placed back in stock. 40 peeloff labels have been provided. Color copies may be made and affixed with tape if more than 40 are needed.
¿ If you have unexpired affected product to return, please contact Getinge Customer Service at 888 9GETUSA (8889438872) (option 2) between the hours of 9 AM and 6 PM Eastern Standard Time to request a return authorization number (RMA) and shipping instructions.
Pack th |
Quantity in Commerce |
12032 Kits |
Distribution |
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Austrailia Austria Belgium Brazil Canada Chile Colombia Czech Republic Denmark Finland France Germany Hong Kong India Ireland Italy Japan Korea Mexico Netherlands New Zealand Norway Paraguay Poland Portugal Russia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Kenya, South Africa, Israel, Mauritus, Nigeria, South Africa, Namibia, Uganda. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|