| | Class 2 Device Recall IV Administration Set |  |
| Date Initiated by Firm | July 06, 2020 |
| Create Date | October 22, 2020 |
| Recall Status1 |
Terminated 3 on May 11, 2023 |
| Recall Number | Z-0252-2021 |
| Recall Event ID |
86054 |
| 510(K)Number | K974589 |
| Product Classification |
Set, administration, intravascular - Product Code FPA
|
| Product | 120 IN (305cm)APPX 15.4ml, 10 DROP, 2 Clave STOPCOCK, NanoClave T-Connector, 2 CHECK VALVES. 1 unit per pouch, 50 pouches per case.
UDI:(01)10887709040141(17)250101(30)25(10)4573652.
The SmartValve is a single use, sterile, non-pyrogenic pressure activated check valve system intended for use as an accessory to Intravascular administration set. |
| Code Information |
Lot Number:4573652. |
Recalling Firm/ Manufacturer |
ICU Medical, Inc. 951 Calle Amanecer San Clemente CA 92673-6212
|
| For Additional Information Contact | 844-654-7780 |
Manufacturer Reason for Recall | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
FDA Determined Cause 2 | Process control |
| Action | On July 6, 2020, the firm, icumedical, sent an "URGENT: MEDICAL DEVICE RECALL" notification of the market action to direct and indirect consignees of record via UPS. Stericycle, Inc. was contracted to send letters to consignees.
The following is included in the "Required Actions for Users" section of the medical device recall notification:
1) Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility.
2) Inform potential users of the product in your organization of this notification and complete the response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product.
3) Return affected product using the return label provided with the letter. Call 1-866-203-5570 (M F, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, you will be credited for any product returned. You will only receive credit for product that you return. NOTE: Credits for product purchased through distributor will be credited by the distributor.
4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product I Table 1 sections of this notification and ask them to call 1-866-203-5570 (M-F, 8am-5pm ET) to obtain a response form.
Follow up Actions:
Contact Customer Service using the information provided below for assistance reordering replacement product and for further inquiries:
Global Complaint Management -1-844-654-7780 or ProductCompla |
| Quantity in Commerce | Total of all products (Listed #1 thru 101) = 304735 units |
| Distribution | Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FPA
|
|
|
|