Date Initiated by Firm | November 04, 2021 |
Date Posted | December 09, 2021 |
Recall Status1 |
Terminated 3 on July 09, 2024 |
Recall Number | Z-0347-2022 |
Recall Event ID |
88987 |
510(K)Number | K193056 |
Product Classification |
Ventilator, continuous, facility use - Product Code CBK
|
Product | Puritan Bennett 980 Series Ventilator, 980X1ENDIUU |
Code Information |
GTIN/UDI - 10884521171541
Serial Numbers:
35B1700300
35B1700306
35B1700307
35B1700362
35B1700372
35B1700384
35B1700387
35B1700388
35B1700396
35B1700397
35B1700403
35B1700413
35B1700416
35B1700418
35B1700426
35B1700431
35B1700434
35B1700440
35B1700445
35B1700451
35B1700452
35B1700453
35B1700459
35B1700461
35B1700464
35B1700468
35B1700472
35B1700477
35B1700478
35B1700479
35B1700480
35B1700484
35B1700485
35B1700494
35B1700495
35B1700496
35B1700501
35B1700502
35B1700503
35B1700504
35B1700534
35B1700574
35B1700579
35B1700604
35B1700609
35B1700638
35B1700655
35B1700664
35B1700816
35B1700919
35B1700994
35B1701001
35B1701418
35B1701698
35B1702832
35B1703438
35B1800033
35B1800721
35B1800811
|
Recalling Firm/ Manufacturer |
Covidien Llc 15 Hampshire St Mansfield MA 02048-1113
|
For Additional Information Contact | Technical Service 508-261-8000 |
Manufacturer Reason for Recall | Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use. |
FDA Determined Cause 2 | Reprocessing Controls |
Action | On November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use.
Actions you should take
Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies.
Notify all personnel in all care environments in which the affected PB980 series ventilators are used about this medical device correction.
If your facility has distributed affected PB980 series ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients.
Complete the attached form and return it as directed to confirm your receipt and understanding of this information.
If you are aware of any incidents related to this issue, please contact our Technical Support Department immediately at 1-800-255-6774, option 4, and then option 1 to provide information regarding those events so regulatory reporting obligations can be fulfilled.
Work with the Medtronic Technical Support Department if you require assistance finding alternative ventilation devices.
Beginning November 5, 2021, Medtronic Technical Service will contact all affected customers to schedule a service appointment for all potentially affected ventilators.
If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774. |
Quantity in Commerce | 278 total |
Distribution | Worldwide distribution - US Nationwide. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = CBK
|