| | Class 2 Device Recall Vivid S6 |  |
| Date Initiated by Firm | November 10, 2022 |
| Date Posted | December 22, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0753-2023 |
| Recall Event ID |
91139 |
| 510(K)Number | K071985 K092079 K102393 K121063 |
| Product Classification |
System, imaging, pulsed doppler, ultrasonic - Product Code IYN
|
| Product | Vivid S6 ultrasound.
Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and
pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology
(including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative
(abdominal, thoracic, and vascular). |
| Code Information |
Model Number (System ID): H45021LD (VS6R0610, VS60368, VS6253, VS6229, VS60870, VS60838, VS627, VS671, VS6332, VS60647, VS629, VS60084, VS6302, VS674, VS6423, VS6411, VS6R0161, VS6634, VS682, VS60841, VS694, VS6341, VS6R0202, VS67633, VS5956VS5, BG4575US01, 604557VS6A, 819732VS6, 403541VS6A, 905523VS6, 416431VS6A, 905884VS6A, 647938VS6A, 705674VS6B, 705674VS6A, 604588VS6B, 604588VS6A, 416597VS6, HR4052US04, HR4179US01, CZ6302US01, CZ6267US01, CZ4031US21, CZ4024US08, CZ6112US02, CZ4031US18, CZ6256US01, CZ6240US01, USM4474502, 145865U002, 120008U030, 2195878U01, 142540U001, 2090670U02, 105827U235, 1926826U02, 818670U341, 800830U451, H1039US19, H2015US20, H4044US04, H1383US33, H2115US05, H4850US01, B5343402, LV4046US01, LT4001US05, LT4000US28, LT4070US01, LT4000US26, LT4000US25, ALR0440, AMS1047, AMS1046, GRN01412, GRN01411, GRN01410, IJM0240, LDN0170, UTR2141, PL2263US01, PL1736US03, PL1623US01, PL2217US01, PL1107US02, PL1006US06, PL1893US02, RO4159US02, RO4341US01, RO4334US02, RU1591US10, RU1106US22, RU1295US14, RU1295US15, RU1018US12, RU1695US02, RU3298US01, RU3299US01, RU1653US02, RU3431US01, RU1553US03, SK4206US02, SI4077US02, SI4077US03, SI4077US04, MPX76304, ME597504, A020US07, P31025UL05, 26011ULT04, 27005ULT06, 45009ULT01, 48015ULT01); H45021LJ (VS600103, VS61226, VS6329VS6, 082438070075, 082438070090, 082438160133, 082438180028, 082438090132, 082438090133, 082438090134, 082438220039, 082438290035, 082438010038, 082438150038, 082438160106, 082438180032, 082438180048, 082438270034, 082438270035, 082438270033, 082438070063, 082438290036, 082438290040, 082438220027, 082438210053, 082438160154, 082438130074, 082438130065, 082438130071, 082438130079, 082438140075, 082438070083, 082438050033, 082438140088, 082438210057, 082438310058, 082438310064, 082438280007, 082438290044, 082438160119, 082438160166, 110256U011, 2258373U01, 083038821002811, 083038841101714, 083038065102678, 083038105104829, 083038005104569, 083038005746413, UX9054, UXA012, UX9053, UXA028, UXA100, UX9098, UX8014, UXA009, UXA042, UX9090, UXA073, UXA024, UXA033, UX8032, UX9088, UX8034, UX8026, UX9077, UX9085, UX9057, UXA029, UX9092, UXA017, UXA001, UX9064, UX9066, UX8021, UX9079, UXA002, UX8005, UX8019, UX8008, UX9078, UX8044, UX9062, UX8036, UX8037, UX8049, UXA020, UX9068, UXA047, UX9087, UX9052, UX9051, UXA072, UX8018, UXA048, UX9060, UXA075, UX9081, UX8035, UX9086, UXA045, UX8003, UX8040, UX8007, UX9063, UX8004, UX8041, UX9061, UX9056, UX8025, UX9084, 0843384096, PL1232US03, PER13818, RU7430US03, RU3297US01, 0856380121, 34197ULT01, VS6578VS6); H45031EB (VS61302, VS6918, VS60793, VS61205, VS62007, 6061, VS61308, VS61866, VS61128, VS6R0695, VS61502, VS61498, 5864, 6120, VS61126, VS61420, VS60960, VS60361, VS60344, VS61334, VS61137, VS61830, VS61179, VS61740, VS61095, VS61331, VS6R0359, VS61213, VS6MB1686, VS6MB1339, VS6R1049, VS61115, VS60965, VS60988, VS61140, VS61244, VS61315, VS61345, VS61839, VS61231, VS61474, VS60994, VS61344, VS61033, VS60980, VS61941, VS61988, VS61968, VS60972, VS6666, VS61121, VS61116, VS61630, VS61629, VS60983, VS60999, VS61199, VS61098, VS60267, VS62160, VS61131, VS61953, VS61134, VS61012, VS61811, VS61211, VS61249, VS61592, VS61596, VS61449, VS61109, VS61318, VS61989, VS61100, VS61197, VS61441, VS61432, VS61283, VS61380, VS61317, VS61603, VS61585, VS61826, VS61661, VS61635, VS61939, VS61293, VS61007, VS61014, VS61437, VS61426, VS61374, VS61645, VS61299, VS61608, VS61364, VS61384, VS61371, VS6MB1248, VS63929VS6, 091136DEMO2467, 0911367150, AP5187US04, AP8694US01, AP0514US02, AH5287US10, AP0341US01, AP6694US01, AH0132US31, AH0132US32, AH0132US33, AP7099US01, AH0101US76, AH5848US61, AH4660US28, AH5848US57, 140024US18, 080053US03, 010028US02, 030344US02, 686348US01, 220166US14, 110037US13, 210002US44, 090009US02, 030018US07, 300013US06, 030029US08, 030029US10, 2847741, VS68335VS6, VS63428VS6, VS63748VS6, VS62261VS6, VS61716VS6, VS61568VS6, VS63483VS6, VS63225VS6, VS68332VS6, VS61721VS6, VS63147VS6, VS63677VS6, VS63676VS6, VS63514VS6.VS63018VS6, 87197, VS61963VS6, VS63115VS6, VS63125VS6, VS61590VS6, VS65149VS6, VS63609VS6, VS62386VS6, VS61560VS6, VS63098VS6, VS62201VS6, VS63119VS6, VS61253VS6, VS69174VS6, VS62008VS6, VS63215VS6, VS67791VS6, VS63735VS6, VS63188VS6, VS63626VS6, VS63789VS6, VS61713VS6, VS61582VS6, VS61703VS6, VS63264VS6, VS63207VS6, VS63262VS6, VS62242VS6, VS60613VS6, VS67660VS6, VS63263VS6, VS63121VS6-1, VS63674VS6, BG4553US01, 705435VS6A, 403719VS6B, 705869VS6A, 416748VS6A, 905472VS6A, 905828VS6A, 514843AVS6B, 514281VS6A, 514876VS6A, 514340VS6A, 514845VS6A, 519372VS6A, 519426VS6A, 905735VS6A, 905688VS6A, 705494VS6, 519254VS6B, 416921VS6, 416422VS6A, 416586VS6A, 519254VS6A, 519250ECHOSW3, 519258VS6A, 204787VS6A, 519973VS6A, VS64441VS6, VS65095VS6, VS63642VS6, VS64388VS6, VS63219VS6, VS63901VS6, VS63996VS6, VS67452VS6, VS62086VS6, VS63492VS6, VS63967VS6, VS64011VS6, VS63965VS6, VS64114VS6, VS61451VS6, VS61463VS6, VS61166VS6, VS62013VS6, VS63353VS6, VS62155VS6, HR4208US01, HR4053US02, CZ4017US39, CZ6402US01, CZ4037US03, CZ6099US02, CZ6355US01, CZ6307US02, CZ6362US01, CZ6403US01, DK1083US45, DK1179US03, DK1098US04, DK1007US12, DK1106US12, DK1042US18, DK1264US01, FI1313US12, FI1349US01, FI1128US25, FI1128US29, FI1014US07, FI1014US08, FI1018US06, USM4162506, USM4478101, USM4478403, U98381329, US70725812, US87682305, US87682306, U201476602, USM4480410, UO55319606, US88069801, USM4023013, USM2266802, USM2173204, U187868406, U195829531, US26664717, US26664716, USM1293303, U206685902, US55554106, US48270615, US48270616, US48270617, USM4048013, USM4048011, USN4165416, USN4191201, U233125702, US77608102, USM4488007, UM75732304, UM75732303, U200487603, U227680301, U232975501, US38390604, U189783117, U176608402, US63277702, USM4840903, US29081203, USM18301, USN4644002, U80004501, U159630008, USM4212606, U236072901, US4484804, US70837703, US94752261, US94752334, US94752590, 2140532U01, 110910U040, 110636U006, 801954U004, 139946U016, 4561426U01, 145228U010, 121221U007, 6491323U01, 4843462U01, 110173U017, 110783U033, 149682U002, 148035U007, 148035U006, 1738060U02, 803274U003, 2426437U01, 2208430U02, 110513U032, 3534055U01, 1924556U14, 149033U007, 145133U010, 1695829U10, 1078247U06, 101829U003, 143288U003, 100371U012, 2204991U02, 100371U011, 109881U006, 100009U013, 112014U008, 104132U008, 2266273U01, 119882U004, 139931U031, 802714U006, 2321817U01, 109529U066, 145884U004, 110047U011, 117675U003, 110810U009, 6113585U01, 110055U007, 4565911U01, 389839U002, 4205247U01, 106989U033, 100018U017, 802786U100, 2002142U02, 100019U025, 2569214U03, 1020149U04, 1528754U02, 109970U013, 4812644U01, 112099U006, 6032292U03, 112099U005, 112099U007, 6032292U04, 120158U013, 110305U031, 120019U020, 800163U003, 1607646U04, 1607646U05, 104089U009, 104025U001, 142470U020, 3344113U13, 122008U013, 145684U006, 110143U009, 2249243U02, 142541U049, 109734U002, 2416273U01, 146221U006, 1305892U04, 1305892U05, 146584U002, 145771U027, 1496753U03, 150097U002, 2168118U01, 110939U009, 1823354U02, 1418517U05, 104658U004, 110604U036, 113128U042, 113128U046, 109341U007, 2139981U08, 2139981U09, 147992U003, 2139981U07, 120609U004, 2139981U02, 2139981U10, 4537138U01, 148208U004, 1937574U06, 2009258U02, 138780U006, 104040U006, 104080U010, 109537U008, 2372101U01, 4820609U02, 2431481U01, 110560U028, 104592U003, 104044U004, 138036U035, 1930392U23, 1930392U22, 109672U006, 109097U005, 1557297U09, 1447625U06, 120041U025, 104046U002, 110355U011, 2496472U01, 2193178U01, 110127U011, 110128U011, 141092U015, 103301U703, 103301U565, 120050U058, 2189160U01, 2217872U01, 110203U010, 109045U021, 2047099U01, 120056U019, 146403U005, H5085US01, H14052US02, H2080US29, H66071US01, H2032US13, H5088US01, H2035US08, H5211US01, H4193US02, H4015US02, H1427US02, H2071US06, H2065US09, H5290US01, H1890US02, VS63790VS6, IS1002US01, IS1001US06, 083038821012210, 083038821462120, 083038101030610, 083038600305710, 083038201040010, 083038041032110, 083038041032111, 083038600312210, 083038600312211, 083038600313310, IE1038US02, IE1018US24, IL4039US02, D5367802, A5360466, A5360465, A5360464, A5811813, C5294638, A5107777, C5294639, B5810912, A5109830, A5142840, B5803122, A5197937, A5201325, M5368924, A5182305, D5373007, A170463802, A5129818, A5178035, A5110285, F5803102, D5111630, D5111629, A5111253, D5343101, A5123638, A5123639, A5264205, A5806324, A224857334, A5293309, A5996402, A6382201, A5103526, A5103527, B5177232, A5116065, A5127610, A5815421, A5112610, A5806211, VS67222VS6, VS61654VS6, UXA186, UXA105, UX9142, UX9182, UX9212, UX9158, UX9106, UXA152, UX9122, UXA127, UX9215, UXA257, UX9209, UX9216, UX9255, UX9203, UX9220, UX9141, UX9146, UX9144, UX9265, UX9266, UXA133, UX9151, UXA170, UX9183, UX9131, UX9268, UXA167, UXA137, UXA191, UXA189, UX9173, UX9275, UX9276, UX9109, UX9219, UX9233, UXA069, UX9150, UX9222, UX9221, UXA270, UX9289, UX9171, UX9157, UX9272, UX9273, UX9124, UX9135, UX9256, UX9282, UX9165, UX9172, UX9211, UX9238, UX9123, UX9148, UXA030, UX9143, UX9129, UX9274, UX9234, UX9235, UX9271, UX9130, UXA132, UXA239, UXA103, UX9254, UX9121, UX9218, UX9065, UX9210, UX9207, UX9180, UX9136, UX9267, UXA104, UX9116, UX9283, UX9253, UX9232, KA2799US03, 0850US7964, 0850US7136, LT4001US10, VS68434VS6, VS63164VS6, VS61571VS6, 2947277, VS62138VS6, VS61566VS6, VS61419VS6, VS62346VS6, VS63442VS6, VS63452VS6, 2931753, VS63390VS6, 2921428, VS62005, VS61833VS6, 2924570, MO1002US01, AMS03405, DRK0148, END1740, GOE0143, HLN0444, HLN0447, HLN0445, HLN0446, LDN01402, LLS0144, SLD0141, UTR0544, VRS0240, ZWD0145, 091136DEMO2344, NO4146US09, NO4015US20, NO4751US01, NO4031US17, VS63042VS6, VS62402VS6, VS63015VS6, VS64353VS6, VS64539VS6, VS64409VS6, PL0173US02, PL7294US01, PL1934US06, PL4019US04, PL6252US04, PL0054US03, PL1023US47, PL1310US04, PL1811US02, PL7501US01, PL1853US02, PL2430US01, PL2272US02, PL0428US01, PL1960US01, PL2618US01, PL6580US02, PL3169US04, PL0707US01, PL1029US09, PL2054US02, PL0694US03, PL6239US04, PL1105US01, PL7626US01, PL1068US08, PL6207US03, PL4359US07, PL6488US04, PL7615US01, PL6028US07, PL6588US04, PL4436US03, PL4359US08, PL4359US10, PL6588US08, PL6588US07, 1210VS6, PL0260US03, PL8068US01, PL7491US01, PL0635US01, PL1843US03, PL1840US04, PL2031US02, PL6271US05, PL7490US01, PL6992US02, PL6804US05, PL1006US48, 600042US30, PL6682US02, PL7475US02, PL7303US01, PER10961, PER00724, PMR64001, PER05608, PER78001, PER25612, VS62216VS6, VS61169VS6, VS64003VS6, VS61254VS6, VS65098VS6, RO4458US01, RO5467US07, RO4207US03, RO5180US01, RO4376US02, RO4429US03, RU2076US10, RU3081US02, RU6671US01, RU3256US02, RU3098US03, RU4355US03, RU6703US01, RU5140US04, RU8124US05, RU3513US01, RU3519US01, RU8474US01, RU3515US01, RU3516US01, RU3517US01, RU2133US01, RU3850US02, RU6391US02, RU1070US08, RU2741US16, RU2741US17, RU1749US02, RU2332US01, RU3460US01, RU7617US01, RU1954US01, RU3459US01, YU4156US03, YU4243US02, 0847384158, 0847384444, 0847384084, 0847384159, SK4116US02, SK4080US02, SI4175US01, ZA2301US01, FEG37715, DEF11021, UEW06414, NPV79331, DPN88617, MP111606, UPW17007, CEZ00311, BD002US03, N008US02, T002US16, 262020US03, 393258US02, 409279US02, 793040US04, 0856380214, 0856380186, P06029UL06, 38016ULT06, P52030UL03, 1208907552, IE1560US11, 1115011632, 1115982562, 00002ULT19, 10238MUM02, 1133385312, 11441ULT01, 170461166, 1193821132, 00160ULT17, 00934ULT03, 1193666802, 1-76439882, 00026ULT13, IE1169US03, 1-93931812, 1209506692, 00488ULT15, 1201477872, VS64038VS6, VS63671VS6, VS63479VS6); H45031EC (0911365723, 0911367183, 0911365626, 083038807316615, 083038807318915, 083038003374914, 083038201014414, 083038004461312, 0850US7469, 0920369432, 681020US04); H45031TC (VS63905, 6273, VS62236, VS62198, VS62026, VS61466, VS61513, VS61525, VS61995, 7070, VS62115, 6274, 6368, VS62187, VS63594, VS61893, VS62080, 6641, VS63020, VS62204, VS63007, VS63494, VS62199, VS6MB2325, VS62167, VS63556, VS62023, VS64071, VS62079, VS63652, VS63982, VS62360, VS6R0944, VS6R1805, VS63584, VS6MB3155, VS63159, VS6MB0982, VS63307, VS63903, VS63653, VS63456, VS62312, VS62196, VS63078, VS63645, VS63361.VS63191, VS6R1136, VS63186, VS63327, VS63911, VS63237, VS62205, VS62232, VS63029, VS62176, VS62244, VS63968, VS63160, VS62153, VS63440, VS6R0349, VS63245, VS62378, VS62135, VS62179, VS63167, VS62190, VS63669, VS63618, VS62374, VS63508, VS63258, VS63229, VS63239, VS61974, VS62293, VS63338, VS63063, VS63065, VS62087, VS63600, VS6R0378, VS6R2365, VS62197, VS62148, AL4024US01, AH4001US61, AH1301US02, AH0132US38, AP9036US02, AP6946US01, AH4660US35, AP7103US01, AH4316US07, AP6504US02, AP7147US01, AH0129US22, 220078US28, 030020US14, 090095US13, 210001US33, 210002US58, 110901US05, 060801US10, 110094US15, 030030US05, BA4231US02, 604875VS6A, 519770VS6A, 403719VS6A, 705743VS6D, 604946VS6A, 780428VS6A, 780428VS6B, 905522VS6A, 519685VS6A, 905276VS6A, 905820VS6A, 905615VS6, 705743VS6B, 705743VS6C, 604278VS6A, 519520VS6A, 250635VS6A, 519842VS6A, 416748VIVS6A, 204235VS6A, 204235VS6B, HR4214US02, HR4202US01, HR4215US01, CY1028US01, CY1221US01, CZ6028US06, CZ4012US34, 600055US10, CZ6422US01, CZ4066US15, CZ4066US08, CZ6521US01, CZ6085US02, CZ4113US08, CZ6009US20, CZ6818US01, CZ4031US24, CZ6652US01, CZ4010US13, CZ4010US11, CZ6651US01, CZ6079US03, DK1047US96, DK1059US24, DK1059US25, DK1059US26, DK1059US29, FI1015US84, FI1033US09, U530898701, USM4476512, U259299602, U259299601, U227818306, U242207901, USM4143707, US80002509, U151511801, USM4185701, US78034305, U201475803, U444306001, U75547801, USM4167406, US26490010, USM4168805, US60575611, US60575609, US87682309, US87682311, US87682310, U127366503, U254516210, U254516204, USM4172305, U460234403, USM4173306, M4173333, U167726504, US26665106, U261779203, USM4399309, U243852101, U129353003, USM4172904, USA4463302, U159998203, U227973503, US26187812, US88657316, U245056101, US27508702, USM5485201, USM5485203, USM5156801, U847344801, USM4144310, USM4207607, USM4207608, U415804908, U128328003, U238565501, US11917202, US87989204, USM1101002, USM4459904, U209843602, U143228202, US29012305, US29011703, U200487606, U794737601, U356534301, USM5460616, US2733405, U210275402, U185855803, USM4197616, USM4197614, U185855804, USM4197618, U143436806, US26411008, USM4008404, U170816502, U523564601, U267287801, USM4484110, USM4491201, USM3186720, US49087201, U180243001, U206713808, U206713806, RE1080US02, USM4202902, USM4201901, US68578705, USM4205804, USM4206611, U100324705, U215636803, USM9205701, US26657007, USM4496205, U187429002, U257020401, UME0614501, USM2422807, USM2422806, USM4496912, USM4496913, US78045604, U123093102, US62353107, U267601401, U155611502, U4217609, USM4487803, 1906476U26, 142404U003, 110655U002, 145800U001, 804348U001, 117062U017, 2409161U01, 2556677U02, 147535U001, 3295435U01, 6774319U01, 110012U044, 1500451U02, 1767561U27, 110012U041, 802156U005, 397776U003, 112010U015, 148127U007, 142456U010, 146135U004, TP4447US12, 1976756U03, 2594442U01, 389502U008, 1892918U11, 802714U008, 1892918U12, 2559562U01, 122024U037, 120111U023, 112026U008, 149106U006, 145884U005, 138806U010, 122038U026, 145223U012, 802786U110, 100019U029, 802786U107, 121623U007, 138142U011, 2551720U01, 142784U007, 145676U016, 150191U004, 142470U021, 112092U007, 122033U038, 110549U007, 1212012U06, 112037U055, 119988U005, 2480932U01, 2030745U02, 104035U015, 148005U003, 110626U004, 110604U045, 102547U002, 109506U001, 2571676U01, 2772606U01, 110238U006, 149570U003, 146667U011, 138251U017, 149704U003, 110868U013, 109097U010, 120044U017, 110605U011, 107088U006, 2333539U01, 800432U003, 141092U021, 141092U018, 2222813U09, 146464U009, 110235U011, H2052US20, H4426US02, H6302US01, H4017US02, IS1010US04, 083038000004811, IL4746US08, IL4746US07, IL4301US04, IL4662US01, IL4764US15, IL4764US14, IL4764US16, IL4285US07, IL4285US06, A5625803, A6397701, A5146639, A5936904, A5325629, A6144201, A5304829, A5936604, B5298502, A6128802, KA1104US02, KA3067US02, KO1126US01, LV4041US02, LV4069US01, LV4005US04, LV4029US03, LV4023US02, LT4016US04, LT4016US05, LT4005US01, LT4098US01, LT4096US03, LT4096US01, LT4102US01, LT4093US01, LT4095US01, LT4001US14.LT4001US15, LT4097US02, LT4097US01, LT4026US04, LT4026US05, LT4036US04, LT4090US01, LT4100US01, LT4100US02, LT4035US01, LT4091US01, LT4087US01, LT4099US01, LT4094US01, LT4101US01, LT4086US01, LT4092US01, LT4000US43, LT4088US01, LT4048US04, 130145US03, MD1038US02, BOZ0142, BVW0145, GOU01401, GRN01424, HDD0144, RTD09401, UTR0149, NO4054US26, PL6691US06, PL2487US03, PL7661US01, PL6770US04, PL7198US02, PL1808US01, PL8715US01, PL4192US08, PL1602US01, PL1751US07, PL7810US01, PL7632US01, PL2783US01, PL8567US01, PL7115US02, PL3185US02, PL7112US02, PL4358US05, PL6108US08, PL1268US06, PL0555US01, PL7822US01, PL6396US03, PL6804US08, PL7114US06, PL4346US12, PL1025US02, PER00226, RO4540US02, RO4167US03, RO4922US01, RO5201US01, RO4228US02, RO4437US01, RO4177US10, RO4502US01, RO4334US06, RO4372US03, RO4500US03, RO5188US01, RO5325US01, RU1000US12, RU1000US11, RU1000US10, RU1284US15, RU7198US01, RU1146US05, YU4060US05, YU4127US05, SK4037US04, SK4583US01, SK4025US02, SK4243US02, SK4063US02, SK4347US01, SK4577US01, SI4077US08, SI4131US02, SI4158US01, SI4054US06, SI4036US07, UPW36601, DEN00742, FEG89700, MP842015, NE741446, FPG83301, O012US01, E001US53, C001US08, 791050US02, 222461US01, UA4190US02); H45031TD (VS6R3548, VS6R1981, AH5047US14, AP6415US02, AH1017US08, AH6647US02, AH0132US36, AH5407US14, AP6105US02, AP7661US01, AH0116US06, AH0500US15, AH4307US37, AP0881US02, AP7147US02, 110806US04, 030241US02, 110040US11, 030091US06, 100032US07, 110011US03, 681393US01, VS67837VS6, VS67929VS6, VS67450VS6, VS67317VS6, VS67809VS6, VS67784VS6, VS67947VS6, VS67693VS6, VS65029VS6, VS65007VS6, VS64158VS6, VS67442VS6, VS64128VS6, VS65191VS6, VS67852VS6, VS64488VS6, VS64119VS6, VS67746VS6, VS67022VS6, VS67025VS6, VS67925VS6, VS67549VS6, VS67061VS6, VS67647VS6, VS67005VS6, VS67004VS6, VS61516VS6, VS65174VS6, VS67027VS6, VS64455VS6, VS67019VS6, VS67819VS6, VS64453VS6, VS67461VS6, VS67447VS6, VS67690VS6, VS67445VS6, VS67457VS6, VS67661VS6, VS67718VS6, VS67833VS6, VS67467VS6, VS67657VS6, VS67094VS6, VS67308VS6, VS65188VS6, VS67313VS6, VS67705VS6, VS67181VS6, VS67719VS6, VS67284VS6, VS67309VS6, VS67276VS6, VS65183VS6, VS67939VS6, VS67315VS6, VS67444VS6, VS67928VS6, VS67721VS6, BF1022US01, 519432VS6A, VS67096VS6, VS67012VS6, VS65169VS6, VS67010VS6, CZ6398US01, CZ6112US03, DK1047U187, DK1047US95, DK1047U113, DK1014US06, DK1588US01, DK1042US33, DK1058US28, DK1059US23, VS65184VS6, FI1032US05, FI2016US01, FI1160US01, FI1175US01, US26234121, U410948301, 120002U007, 121221U005, 110906U021, 110173U020, 110886U003, 110767U010, 110783U034, 4657094U01, 1009902U10, 2396353U01, 1767561U29, 1924556U12, 145424U053, 110012U043, 109397U004, 1767561U44, 138808U040, 145835U004, 2363285U01, 145750U049, 397776U004, 145100U026, 109818U008, 109762U026, 1438183U02, 145963U009, 111007U021, 148214U004, 7654460U01, 103135U033, 1961127U05, 2503742U01, 110040U004, 104007U035, 103584U004, 149106U008, 802509U003, 145120U004, 110810U012, 110055U009, EK00943U02, 1886792U02, 802786U112, 149495U005, 802786U108, 1837410U36, 1837410U34, 802786U109, 145760U012, 6032292U01, 120158U016, 144545U004, 2585794U01, 125085U005, 112091U015, 1200595U02, 106018U002, 145525U011, 122033U037, 122033U036, 122033U035, 804388U002, 2710949U01, 125096U006, 125096U005, 4784887U01, 145840U006, 142541U052, 120139U036, 110079U013, 1466557U03, 803253U004, 147180U011, 101154U016, 2584288U02, 120032U047, 110094U006, 2139981U13, 1891694U04, 2520328U02, 804104U003, 104077U020, 1874554U04, 145730U009, 800574U005, 210256U059, EK00335U01, 3860901U01, 2589170U01, 117981U004, 110826U009, 120044U014, 120044U013, 110570U008, 103608U002, 119387U003, 106027U004, 2306386U02, 141092U019, 141092U017, 1799978U01, 145780U004, 103558U006, 138818U008, 112058U016, 112058U015, 2352345U01, 803425U005, 120052U014, 2017571U14, 2017571U15, 109874U004, 100902U002, H6660US01, H5612US01, H6855US01, VS65156VS6, HU1028US05, HU1074US03, IE1068US12, IE1068US11, IE1077US05, B139923601, A5156938, A5768731, D5111633, A6147301, A5697713, B5316818, A5188555, A5114726, A6025603, A5410314, A5268207, A5183652, A179023805, B5418913, A6126701, A6157301, A5728815, A5728817, A5127612, 419296US01, VS67820VS6, VS67448VS6, VS67014VS6, VS65047VS6, VS64586VS6, DTC0148, GRN01415, HLS0141, RTD1341, TLB02402, NO4064US15, NO4229US02, NO4366US02, NO4070US25, NO4004US34, NO4035US20, PL2938US01, PL7234US02, PL7198US01, PL7126US01, PL1764US02, PL3827US01, PL2618US02, PL0501US01, PL6992US01, PL0462US04, PER12816, PES17200, PER12704, PER06634, PPR59201, RU2700US02, RU1116US03, RU3993US01, RU1086US10, RU0306US01, RU3514US06, RU7520US03, 600100US03, RU1146US07, SI4055US11, SI4071US02, FPG93031, FEB33844, MEX68979, MEX689141, X018US01, X001US09, O048US04, O009US36, F001US17, BD001US03, E001US52, BD004US02, A007US40, 795021US01, 252020US03, 676020US03, 707100US01, 573010US03, 412491US02, 00004ULT13, 00440ULT82, 10181ULT19, 10181ULT18, 20308ULT03, 20418ULT02, 01041ULT13, VS67413VS6); H45031XR (US94752344, 11959AU103, 1400010536); H45041D (VS67095VS6, VS63864VS6, AH4001US53, AP5120US03, AP9156US01, AH0127US89, AP7197US01, AH6871US01, AH1908US37, AP7289US01, 110028US17, 030020US15, 100012US21, 150019US10, 030091US05, 030070US07, 080041US08, VS65187VS6, VS62274VS6, VS65023VS6, VS67941VS6, VS68086VS6, VS65022VS6, VS67950VS6, VS63536VS6, VS63697VS6, VS67100VS6, VS68099VS6, VS67473VS6, VS68100VS6, VS67954VS6, VS67717VS6, VS67884VS6, VS65132VS6, VS67944VS6, VS67882VS6, VS65164VS6, VS62012, VS67069VS6, VS67716VS6, VS64619VS6, VS67747VS6, VS67783VS6, VS67039VS6, VS67792VS6, VS67744VS6, VS67314VS6, VS67953VS6, VS67938VS6, VS63303VS6, VS63039VS6, VS65193VS6, VS67883VS6, VS67097VS6, VS67715VS6, VS64612VS6, VS67079VS6, VS62271VS6, VS67896VS6, VS67932VS6, VS67881VS6, VS67957VS6, VS64026VS6, VS68165VS6, VS68385VS6, VS63413VS6, VS63362VS6, VS68388VS6, VS67455VS6, VS64383VS6, VS63938VS6, VS63505VS6, VS67927VS6, VS63499VS6, VS64449VS6, VS65185VS6, VS62146VS6, VS67951VS6, VS68068VS6, VS67836VS6, VS68994VS6, VS67307VS6, VS63837VS6, VS65043VS6, VS67460VS6, VS68277VS6, VS65178VS6, VS67081VS6, VS63678VS6, VS67029VS6, VS63140VS6, VS62389VS6, VS62270VS6, VS64005VS6, VS63498VS6, 82438240013, 82438210086, 82438190047, 82438050044, 82438250059, VS63387VS6, VS64352VS6, VS64110VS6, VS63858VS6, VS64608VS6, VS63758VS6, VS64197VS6, VS64207VS6, 600055US13, VS65120VS6, VS6339VS6, FI1030US22, FI1003U131, 112001U008, 145082U022, 145082U023, 145297U030, 1501742U08, 2347285U01, 121566U003, EK00319U01, 2692422U01, 145465U011, 2266273U02, 138294U023, 104007U036, 149106U007, 800625U007, 2512330U01, 802786U068, 1930450U04, 5020808U10, 1837410U33, 110060U007, 122033U033, 122008U049, 122033U031, 122033U034, 142491U002, 120139U035, 1466557U02, 105042U009, 142400U012, 149670U035, 2600374U01, 142453U039, 150077U003, 2360565U01, 145410U003, 803213U004, 2628334U01, 2646696U01, 802501U014, 2112894U03, 2570020U02, 2646088U01, 120545U012, 4093051U01, 141092U020, 103301U714, 108003U007, 2566790U01, 2552578U01, 110235U010, 147583U003, 102529U003, H2080US31, H5724US01, VS63733VS6, HU1120US01, 083038201016212, 083038201024113, IE1068US15, B5128502, B5481110, A6096301, A5190707, A5209708, E5172904, A5348011, A5123662, A5422104, A5109319, UX9361, UXA286, UX9550, UX9547, UX9460, UX9566, UX9433, UX9399, UX9423, UX9356, UX9391, UX9523, UX9404, UX9369, UXA331, UX9401, UX9304, UX9187, UX9390, UX9548, UX9479, UXA168, UX9519, UX9466, UXA264, UX9392, UX9480, UX9297, UX9520, UX9343, UX9469, UX9350, UX9409, UX9503, UX9367, UX9478, UX9368, UX9353, UX9321, UX9452, UX9354, UX9549, UXA403, UX9436, UX9437, UX9461, UX9337, UX9565, UXB069, UX9513, UX9311, UXA341, UX9380, UXA237, UX9298, UX9336, UX9454, UX9340, UX9400, UXA346, UX9471, UX9285, UX9525, UX9363, UX9333, UXA242, UXA208, UX9517, UXA236, UXA340, UX9467, UX9545, UX9425, UXA330, UXA287, UX9508, UX9379, UX9258, UX9410, UXA312, UX9393, UXA259, UX9564, UX9070, UX9470, UXA329, UX9468, UX9316, UX9110, UX9115, UX9299, UX9422, UX9414, UX9567, UX9335, UX9362, UX9309, UXA342, UX9516, UX9518, UX9462, UX9421, 0850US7990, VS64017VS6, VS61967VS6, VS63031VS6, VS63131VS6, VS63345VS6, VS61969VS6, VS69324VS6, VS6CMV, DHG09402, GRN01427, STK0148, VS61792VS6, VS61489VS6, NO4015US25, NO4035US18, NO4088US30, NO4072US02, VS63896VS6, VS63875VS6, VS62395VS6, VS63893VS6, VS67001VS6, VS63005VS6, VS63524VS6, PL2085US02, PL0319US01, PL1862US02, PL2675US01, PL2837US01, PL2089US05, PL1025US01, PER34930, VS61710VS6, VS62094, VS63139VS6, RU1558US01, RU1558US02, RU1875US03, RU3776US01, RU6048US02, 6073UL0064, MP324909, O003US05, 502010US17, VS64482VS6, 00222ULT10, VS67006VS6, VS64491VS6); H45041DZ
(140024US15, 190073US05, 220078US20, 030295US04, 100012US17, 747722US02, 550020US04, 030077US01, 060801US06, 030800US15, 030028US04, VS63948VS6, VS63912VS6, VS64001VS6, VS64069VS6, VS63821VS6, VS64081VS6, VS63940VS6, VS63004VS6, VS63956VS6, VS67325VS6, VS63749VS6, VS64059VS6, VS63819VS6, VS67945VS6, VS67738VS6, VS63801VS6, VS63817VS6, VS63829VS6, VS63788VS6, VS63711VS6, VS63812VS6, VS63784VS6, VS63832VS6, DK1561US01, DK1558US01, DK1008US01, DK1066US47, DK1314US06, DK1314US07, FI1307US02, HU1294US03, HU1389US01, IL4642US01, IL4129US01, C5124632, B5114288, A5261504, A5812268, A5812266, A205391201, A5812264, A5812273, D5803102, G5803102, A5404704, B5373007, A51152183, A51152176, 080087US05, VS60640VS6, VS62009VS6, NO4070US19, NO4018US20, NO4060US11, NO4198US10, NO4047US03, NO4005US69, NO4477US04, NO4068US58, NO4068US57, NO4068US68, NO4068US59, PL6599US02, PL6605US02, PL4739US03, PL6360US02, PL6804US03, PL6804US04, PL6804US02, SI4010US02, NE209949, DEN00723, YE643316, DEN00921, FEG37609, 437918US13, 260010US18, 569841US01, 00150ULT07, 00631ULT07, 20458ULT01, 00004ULT10, 00231ULT10, 20080ULT10, 01050ULT07, 00314ULT04, 00314ULT06, 00432ULT27, 00398ULT01); H45041FT (VS68885, VS64484, VS67567, VS61745, VS6MB7466, VS68883, VS68842, VS6MB003647, VS68810, VS67137, VS64103, VS6MB8656, VS6MB4573, VS68868, VS67821, VS65159, VS67392, VS67495, VS68050, VS68839, VS67872, VS65073, VS67892, VS68853, VS6R8268, VS68844, VS67854, VS67152, VS64202, VS67439, VS67296, VS67869, VS68806, VS6R774, VS64055, VS68090, VS6R0142, VS5R0718, VS68543, VS64146, VS6MB0791, VS67016, VS65180, VS67194, VS6R1998, VS6R2104, VS68858, VS67691, VS68275, VS68995, VS67855, VS68686, VS68308, AL4097US01, AL4104US01, BA4117US13, VS63947VS6, VS67629VS6, VS65189VS6, VS63961VS6, 604875VS6C, 604875VS6B, 403943VS6A, 403235VIV, 250737VS6, 905848VS6, 514747VS6, 705674VS6C, 416481VS6, 416340VS6, SN3976VS6, 8769VS6, CZ4012US37, CZ4047US11, CZ6323US10, DK1167US76, DK1167U114, DK1167U117, DK1047U141, USM4159709, U165309003, U123311602, U774132501, U553863901, U191824202, USM4548511, US55554118, 7487VS6, US29081314, USM9194401, U415804987, U415804978, U415804998, U415804984, U550684701, UN95957101, US26572603, U602013301, USS4440401, U489209001, US67238103, 3702VS6, US68233305, U502238501, US93011506, US88022321, U472797501, UMM4440401, U399665502, U226791503, U149087205, US26670008, 404420U712, 181034U113, 181034U240, 181034U188, 181034U149, 181034U211, 181034U119, 181034U179, 181034U081, 1810348U95, 1810348U96, 181034U120, 181034U148, 181034U028, 1810348U42, 181034U239, 210272U014, 210256U935, 210272U062, 210272U061, 210272U043, 210256U980, 105827U453, 1926829U05, 1926829U06, A106333808, A471376301, KO1148US01, 043066US01, PL6378US05, PL8034US01, ZA1025US22, ZA1025US21, A328US01); H45041FU (VS68990, VS68964, VS68538, VS67493, VS6MB2025, VS67121, VS67868, VS64157, VS67860, VS67075, VS64499, VS67117, VS6MB1992, VS64536, VS68721, VS6MB7673, VS68342, VS68274, VS68507, VS68884, VS64195, VS67199, VS68866, VS67848, VS67024, VS67877, VS68284, VS67611, VS68048, VS6289TEE, VS68387, VS68819, VS68762, VS68485, VS68673, VS67166, 604631VS6A, 905497VS6, 905602VS6, 416462VS6, 905547VS6, 250370VS6); H45041JW (0911363389); H45041PS (0911369283, 0911363838, 0911363968, 0911365672, 0911363939, 0911363943, 0911363952, 0911373547, 0911373548, 0911373609, 0911367296, 0911373617, 0911365792, 0911373564, 0911373594, 082438220047, 082438110060, 082438040210, 082438040209, 082438030082, 082438070140, 082438190055, 082438190057, 082438110069, 082438190054, 082438080064, 082438030106, 082438230090, 082438230105, 082438230106, 082438230114, 082438230111, 082438220045, 082438220046, 082438220056, 082438220049, 082438230112, 082438160246, 082438100220, 082438210107, 082438160304, 082438250047, 082438030113, 082438170037, 082438140134, 082438070169, 082438090215, 082438190050, 082438120395, 082438120432, 082438100229, 082437290214, 082438020042, 082438080043, 082438200028, 082438120454, 082438100240, 082438290059, 082438160219, 082438090265, 082438090234, 082437230588, 082438230093, 082438120414, 082437230587, 082438250048, 082438120415, 082438150070, 082438240012, 082438150061, 082438150086, 082438130088, 082438130092, 082438130097, 082438130098, 082438130105, 082438130106, 082438130107, 082438070144, 082438110054, 082438110087, 082438120397, 082438240011, 082438100207, 082438210089, 082438310088, 082438140110, 082438090209, 082438080047, 082438090247, 082438090268, 082438080046, 082438300021, 082438300024, 082438070109, 082438070127, 082438290060, 082438090287, 082438210090, 082438090210, 82438220053, 82438050045, 82438050048, 82438250058, U151072403, US94752361, US94752432, US94752597, US94752617, US94752631, 109045U027, 0828370363, 0828380232, 0828380263, 0828380266, 083038021046910, 083038200370112, 083038201018212, 083038281083815, 083038201170712, 083038601133311, 083038100058613, 083038281182512, 083038600312511, 083038600312512, 083038860516811, 083038600107913, 083038800150612, 083038201058511, 083038841134511, 083037005638413, 083037005638414, 083038001022411, 083038840108213, 083038860513611, 083038820007711, 083038001019612, 083038860154912, 083038801184112, 083038600002812, 083038020011811, 083038200058512, 083038807449113, 083038801094913, 083038801028512, 083038861061111, 083038800507312, UX9296, UXA447, UX9313, UX9381, UX9514, UX9413, UX9315, UX9382, UX9288, UX9441, UXA314, UX9322, UX9351, UX9295, UXA334, UX9412, UX9451, UX9435, UX9305, UX8011, UX9347, UX9506, UX9434, UX9355, UX9370, UXA240, UX9402, UXA438, UX9546, UX9515, UXA306, UX9507, UX9569, UX9453, UX9307, UX9352, 0850370021, 0920369469, 0920369539, 0843384100, PL1023US21, PL1023US22, PL0331US01, PL1777US02, PL2089US04, RU5734US01, RU6150US02, RU7214US01, RU3674US01, RU7367US01, RU7367US02, RU7368US01, RU7390US01, RU5488US03, RU3835US01, RU8618US01, RU3531US04, RU3836US01, RU1579US02, RU7877US01, RU7801US01, RU1461US02, RU1284US14, RU8267US01, RU1808US03, RU3851US01, RU3942US01, RU1047US08, RU2952US03, RU2952US06, RU7520US06, RU6484US03, RU5552US03, RU3257US04, RU1138US05, RU1382US03, RU3832US01, RU5455US02, RU7302US01, RU6048US03, RU6048US05, RU3593US04, RU5899US03, RU7365US02, RU1999US26, RU3843US01, 0847384177, 0847384439, 0847384442, 0847384456, 0856380222, 0856380226, 0856380231, 0856380247, 0856380254, 0856380255, 0856380207, 1648922002, 1647668482, 1647796242, 1890119046, 1647668602, 1706082462, 1601809375, 1890330112, 1648128832, 1889297662, 1878659932, 1566873842, 1650964522, 1647973382, 1844849302, 1592077982, 1509809174, 1796700592, 1648862082, 1648722512, 1648722516, 1634986662, 1560315461, 1515948463, 1525949672, 1642597652, 1582219582); H45041QG (VS68297, VS64072, VS68295, VS67043, VS68057, VS68937, VS67542, VS67863, VS65139, VS64475, VS67008, VS64524, VS67692, VS68110, VS64631, VS68843, VS6R8230, VS67573, VS68857, VS67771, VS6R0625, VS67674, VS62326, VS64129, VS68867, VS68850, VS68750, VS68892, VS6MB0482, VS68377, AL4101US01, AL4043US03, AL4032US01, AO1015US07, AP7830US01, AH0305US22, AP2305US03, AP8600US01, AP0790US01, AH4983US08, AH4001US70, AP6796US01, AH1301US05, AP5094US04, AH4323US80, AH1915US08, AP7489US01, AP5213US03, AH0132US42, AH0132US47, AP7094US01, AP0394US03, AP6285US02, AH2310US18, AH7157US02, AP6502US02, AH5090US07, AH5090US08, AP4515US03, AH5689US33, AH0101U101, AH4660US38, AP6056US02, AH0101U104, AP8439US01, 391885US02, AP8805US01, AP6557US02, AP8381US01, AH0101U105, AP8332US04, AH5848US76, AP7103US03, AH4316US09, AH4316US10, AP7147US03, AH0101U103, AH5848US77, AP6498US02, AH5848US80, AH0101U108, AP8708US04, AP5097US04, AP9017US01, 110069US03, 220176US01, 090070US17, 190059US16, 030020US18, 030082US18, 030082US17, 190190US13, 010120US23, 210001US35, 110065US38, 220913US02, 110040US13, 563899US01, 210002US62, 210002US63, 190747US16, 210013US20, 928880US01, 190747US18, 180043US08, 491838US04, 110901US12, 030032US03, 030032US02, 110094US20, 110094US16, 190145US14, 190171US12, 130015US04, BA4080US28, BA4068US03, BA4080US22, BA4209US01, BA4210US01, BA4211US01, BA4208US01, BG4521US02, BG4572US06, BG4600US01, BG4551US02, BG4655US01, 905848VS6A, 506862VS6, 877718VS6, 905534VS6, 416462VS6A, HR4020US03, HR4049US01, HR4131US03, HR4008US14, HR4008US13, CY1157US01, CZ6598US01, CZ6644US01, CZ6010US03, CZ6149US03, CZ4015US44, CZ6028US08, CZ4015US42, CZ4015US43, CZ4015US32, CZ6708US01, CZ4012US16, CZ4047US12, CZ6572US01, CZ6461US05, CZ6645US01, CZ6461US06, CZ4137US06, CZ4019US09, CZ4113US10, CZ4113US13, CZ4091US02, CZ6009US23, CZ4013US09, CZ6877US01, CZ6586US01, CZ6003US03, CZ6018US15, CZ4013US07, CZ6161US06, CZ6669US01, CZ6026US31, CZ6796US01, CZ6217US07, CZ6217US06, LV4060US01, CZ6192US02, CZ6100US04, CZ6522US02, CZ4033US03, CZ6574US01, CZ6694US02, CZ6535US03, CZ6150US02, DK1083US58, DK1047U105, DK1047U109, DK1047U118, DK1047U136, DK1047U122, DK1004US13, DK1288US03, DK1185US02, DK1030US50, DK1031US55, DK1079US25, DK1079US15, DK1386US01, DK1043US38, DK1108US24, DK1108US20, DK1424US05, DK1000US27, DK1048US05, DK1009US17, DK1059US43, EG6004US02, EE4024US14, EE4019US22, EE4089US02, FI1126US04, FI1012US06, FI1479US01, FI1019US13, FI1030US11, FI1155US03, FI1015US53, FI1021US16, FI1021US17, FI1025US20, FI1025US21, FI1289US03, FI1272US01, FI1003US53, FI1003US50, FI1003US54, FI1003US51, FI2004US01, 9160VS6, U456706901, U409199701, US67214611, U256826203, U256826204, U454285201, U323016001, USM4476519, US67483403, USM4161613, U471419101, USM4017506, US29089405, U460069801, U287345802, USM4163014, USM4163020, USM4163015, US73018016, U104546807, U104546803, U383482101, U280465001, U204016002, US06756001, US48271606, USM4478214, USM4478213, USM4478212, USM4478211, US78034311, US78034308, USN4172701, US61709105, USM4166010, USM4166006, U312751203, U312751202, U401871501, US66717107, U239027718, US75547811, US75547809, USM1607005, US75547810, US75547807, U218322803, U98381328, U98381323, USM4168204, 4326VS6, U362597102, US24297401, USM2722422, USN4630503, US70725813, U151303606, US15001708, U205963804, U456133001, US37008601, U175750004, U471441701, U384031301, U383485101, US15000406, U175562105, U101513302, US80000806, US80000805, U464196001, USN4234207, USN4234209, USN4234206, USN4234211, USN4234201, U227981502, U168571105, U216620812, U380258701, US88057105, US88057104, USM4173009, US26671808, USM1582901, U363935301, U330984501, US69059310, US42330204, US62664119, UM80004501, U200411604, U200411603, USM4173510, U350520901, U355459501, US4175802, U475531301, U203073002, U121951004, U243684703, MQ1078US03, MQ1078US3, U228270304, U252056202, U214652606, U152870606, USM4176503, USM4482609, USM4001302, US59715903, USM4463923, USM4463921, U169514802, USM4463924, U192589802, U257401014, U416545301, U123378703, USM0025902, U207711802, GF1006US04, USM4181474, US10001306, US10001305, U130810402, USM4180805, USM4485905, USM4548304, USM4548503, 8563VS6, US41820009, USM3168007, U349933501, USO4185501, US62967909, USM4186005, USN4464301, USM4464303, U202618802, U262146001, U179037602, US15003104, U253891202, USM9957405, U370680701, UM15001502, U633546201, U98381312, U98381307, US55554114, US55554117, U98381306, US55554112, U98381309, USM4024506, U98381304, USM4024503, U98381308, U98381310, U98381311, U190255013, U190255011, U190255014, U190255015, U190255016, U127219603, US64559805, U129771605, USM2851009, USM2851003, USM2851004, USM2851002, U356607301, US67682903, US29087605, US57401004, U129771606, U123467902, U150668903, U417418101, USM5156803, U269652601, U136818212, USM4488016, U255961602, U101610803, US29353001, US2746002, US51831605, USM4490405, UM75732310, U203665505, U203665506, USM2264705, USM2605505, USM2838403, U180393008, US78043909, US20002727, U383283701, U396387701, U210216902, U463266901, U394474501, USN9190114, U222593302, U396293701, U395131301, U338030501, USM7412801, US2733410, U106033031, 8264VS6, US26372302, U289468801, USA6331001, U402959901, USM4464418, USM4464419, USM4464420, U393994101, US55000309, U255588604, U343886701, USM3186731, MDP8233306, US25013104, U473695801, 8930VS6, RE1000US05, RE1000US06, USM4201816, USM4201812, USM4201815, USM4202906, U197370203, U384026301, RE1121US02, UM88634205, UM88634201, UM88634204, US78033006, US32510706, US29082903, U455314301, USM2603501, U188216002, USM4401504, US96526603, U189783107, U415370101, US61165104, U485493101, US61120601, U233125302, U355259101, U209966803, UM29081201, US67011603, U460616001, U346906901, US88029302, US62899103, US63494203, U195415203, U356710501, US78039406, US78039407, US78039408, US26492214, U256516802, USM4216609, USN4216705, U143036103, U4217613, US27225803, U347563101, USM4475005, 110636U013, 143994U009, 145781U004, 138880U005, 2583611U01, 802608U002, 121221U011, 145196U009, 4561435U01, 4745014U01, 110016U019, 110783U058, 109545U004, 1889948U01, 3854043U01, 404420U118, 404420U032, 4044201U11, 145424U072, 145424U086, 2042064U02, 110012U047, 145133U011, 150299U004, 4203315U01, 1924556U09, 110012U048, 110459U046, 149826U012, 4203316U01, 2396353U02, 3841231U01, 3624743U01, 1009902U17, 4025637U01, 393264U003, 148251U006, 104001U008, 1810336U03, 110266U023, 104001U009, 1078247U13, 801867U005, 145843U005, 114278U002, 145100U040, 109818U011, 4118609U01, 127089U005, 142456U011, 111007U031, 111007U032, 111007U033, 147976U012, 1464242U08, 151299U006, 148214U005, 3754995U01, 2266273U05, 141219U008, 1513970U03, 1513970U08, 804365U003, 393407U003, EK01259U01, 2261281U03, 110593U013, 1892918U20, 3464659U01, 122024U031, 104007U041, 110003U011, 103584U006, 112026U012, 148379U010, 142495U001, 1917922U02, 120013U006, 2613970U02, 1567892U10, 2206962U02, 110814U004, 110396U016, 138032U005, 2836309U02, 4015231U02, 110055U035, 147371U008, 104012U018, 147371U009, 141114U055, 141114U044, 4495291U01, 2553247U02, 1904946U02, 148004U003, 148004U007, 151112U005, 144470U004, 110320U006, 1886792U01, 2095096U02, 106989U038, 145287U020, 106989U037, 802786U165, 2592276U02, 2795540U02, 802786U150, 109970U020, 802786U141, 5020808U23, 802786U149, 1946300U08, 802786U151, 121623U008, 1845808U02, 1442850U18, 1882626U37, 141814U005, 120019U025, 120019U027, 120019U031, 120019U028, 120017U022, 118192U004, 118192U005, 3152028U01, 800805U003, 800748U003, 150733U004, 104019U012, 1064810U01, 1607646U10, 1607646U06, 210502U008, 1563124U02, 1563124U03, 1882484U03, 106590U005, 106590U007, 142470U023, 145948U019, 120064U111, 120064U107, 120064U112, 147995U008, 3126192U01, 122008U038, 1211404U02, 122008U020, 802902U006, 149407U031, 104026U015, 104027U004, 125096U007, TP3007U007, 2471220U03, 3066456U02, 2392493U01, 4168181U01, 4011620U02, 141818U004, 112038U004, 121613U008, 803457U002, 120597U004, 2213271U02, 142541U069, 1212012U08, 149615U018, 149615U014, 142541U066, 3859581U01, 120139U049, 110079U018, 141831U003, 110251U010, 146568U002, 145091U013, 1654304U03, 803312U004, 1556157U03, 112093U009, 1305892U12, 1305892U08, 1305892U10, 145550U027, 145550U025, 145628U011, 4053445U01, 138560U010, 2247681U03, 110089U010, 1496753U05, 1847670U05, 1847670U07, 114248U010, 4053365U01, 3825243U01, 103302U071, 101154U023, 103302U082, 101154U022, 3949183U01, 149670U031, 149670U034, 149670U030, 104138U007, 2853952U02, 140468U007, 145567U010, 145567U009, 1826730U02, 151176U009, 3650585U01, 110604U050, 110107U010, 103556U002, 110094U007, 803260U005, 138330U017, 803260U006, 120609U005, 803260U008, 3273145U01, 803630U004, 141824U005, 120035U027, 149843U009, 149016U002, 109034U011, EK00778U01, 6600341U01, 112801U009, 104077U028, 110112U022, 2646696U02, 800587U006, 110690U018, 143021U004, 1713464U02, 1836179U03, 146667U062, 110240U010, 3943751U01, 138036U048, 145101U005, 138036U043, 110868U016, 2628982U02, 210256U674, 210256U617, 210256U253, 210256U637, 110253U003, 3441087U01, 3441087U02, 138607U011, 145726U017, 120108U013, 145505U032, 2395757U01, 3627367U02, 2522416U02, 120044U021, 3106892U02, 3106892U01, 2631812U01, 1695977U02, 110120U005, 104046U010, 1974635U01, 2377892U01, 145272U014, 2770284U01, 119387U001, 104066U020, 104066U021, 104066U016, 105827U430, 150354U002, 138004U003, 138004U004, 110569U001, 113025U007, 113025U006, 801972U004, 104555U009, 104555U008, 103301U831, 141695U002, 119041U007, 119041U008, 149429U611, 109725U035, 109725U024, 109725U026, 109725U036, 9002434U01, 112089U003, 3105780U01, 3337517U01, 109044U007, 1715002U06, 112600U010, 120052U020, 146403U006, 1771187U02, 138209U004, H2052US22, H6144US01, H1801US03, H4445US02, H2094US31, H4762US02, H22401US05, HU1366US02, HU1113US09, HU1113US05, HU1416US01, IE1068US22, IL4746US10, IL4270US11, IL4740US02, IL4652US17, IL4086US01, IL4758US17, IL4123US01, IL4014US11, IL4787US05, A6136002, A5969102, A5672216, A5251142, A6092102, A391516701, B5352005, A5810830, A6082202, A6227701, A5551109, A129176302, A277032605, A5551404, A101427104, A5998402, A6011902, C5316830, A5320161, A6151001, A5379611, A6238001, A5529130, D5109626, D5109612, D5109625, B5109627, B5109628, A6185902, A5123654, A6288201, A6152602, A5744308, A5930403, A5295820, A5967103, A5293312, A5926905, A5926903, A187751002, A6232502, A6120501, E5120704, KA2924US01, KA2738US03, KA3318US01, KA3134US01, KA3136US01, KA3139US01, KA1190US03, KA3034US02, KA2837US02, KA1179US01, KA4131US01, KO1024US01, KO1141US01, LV4071US01, LV4070US01, LV4004US08, LV4058US03, LV4058US01, LV4011US03, LY1155US06, LT4026US09, LT4104US01, LT4105US02, LT4000US48, LT4000US46, 130326US05, 043066US03, AMS03411, AMS03484, BLR0241, DHG09412, DVT01412, DRK01402, LDN0179, LDN0181, LDN01410, NMG02427, ZTP0146, NO4344US01, NO4663US01, NO4011U135, NO4245US02, NO4145US06, NO4004US59, NO4144US07, NO4014US53, NO4110US05, NO4005US96, NO4296US12, NO4005US90, NO4005U106, NO4016US85, NO4005US98, NO4302US11, NO4418US03, NO4009US54, PL2922US01, PL1455US06, PL8039US01, PL8065US02, PL8355US01, PL8036US01, PL7794US03, PL8222US03, PL0312US01, PL8852US01, PL8287US01, PL2117US02, PL4247US03, PL8283US01, PL1736US05, PL3415US06, PL3756US01, PL8395US01, PL7149US02, PL7186US01, PL6120US03, PL8404US01, PL0546US02, PL1602US23, PL3168US01, PL3807US01, PL1027US08, PL1027US07, PL3052US02, PL3166US01, PL3473US02, PL8805US01, PL2055US04, PL6580US03, PL3549US01, PL8062US01, PL4739US05, PL1096US04, PL1111US06, PL7991US01, PL3893US01, PL8352US01, PL2879US01, PL8282US01, PL1891US03, PL7444US02, PL6207US04, PL4624US06, PL7936US02, PL6177US03, PL7660U219, PL3098US01, PL8176US01, PL8658US01, PL3234US02, PL8362US02, PL7918US01, PL3363US01, PL3732US01, PL1827US04, PL2648US02, PL1962US02, PL7994US01, PL8424US01, PL6804US18, PL6804US13, PL6804US19, PL6804US15, PL6804US14, PL6804US17, PL7894US01, PL1985US03, PL1828US02, PL1313US02, PL1006US24, PL1006US47, PL1006US32, PL1006US23, PL8792US01, PL6682US08, PL8229US04, PL7953US01, PL7929US02, PL8360US01, PL7909US01, PL8445US01, PL7545US04, PL7947US01, PPR110107, PPS19303, PER09111, PER03524, PPR49804, PER40309, PER00235, PPR97401, PPS19810, RO4458US04, RO4925US01, RO4569US02, RO5026US01, RO4808US01, RO4693US01, RO4668US01, RO4483US03, RO4162US95, RO4186US09, RO4483US05, RO4926US01, RO4823US06, RO4943US03, RO4793US01, RO4892US02, RO4187US17, RO4207US06, RO4187US16, RO4078US05, RO4030US06, RO4660US01, RO4772US03, RO4923US01, RO4924US01, RU7781US01, RU9255US02, RU8528US01, RU3199US02, RU7794US01, RU3322US05, 600148US10, RU1081US03, RU7613US02, RU8330US01, RU1835US03, RU1232US03, RU8660US01, RU3338US02, RU1514US03, RU7243US03, RU1475US14, RU1187US05, RU9625US02, RU8533US01, RU8687US01, RU6024US03, RU1018US20, RU7853US01, RU8468US01, RU9701US01, RU1047US04, RU8132US01, RU2340US01, RU9933US01, RU9222US02, YU4259US01, YU4140US05, YU4184US02, SK4110US02, SK4028US02, SK4004US03, SK4481US01, SK4076US02, SK4425US01, SK4079US02, SK4190US02, SK4479US01, SK4537US01, SK4288US02, SI4077US12, SI4056US19, SI4056US12, SI4056US17, SI4054US08, YP652201, YP170A24, CE677627, ME2083101, MPX84720, NP739731, MPX77014, MP473920, DEL71942, DPB22828, FP721A16, FEG20216, CPZ47400, MP737809, MEX61744, MPX39919, MEX61746, FPG23616, FPG30540, W001US15, X006US12, F001US36, F001US21, F001US34, S001US26, S001US22, E001US72, A017US12, A007US42, A007US58, A016US02, A007US57, G007US01, 850010US06, 262020US06, 652060US26, 200139US01, 475200US42, 241987US01, 356010US02, 129716US01, 200295US01, 679352US01, 679352US02, 121512US10, 434658US02, 836020US07, 138107US01, UA4296US01, AE1641US03, 00001ULT31, 00112ULT15, GB1089US01, 10223ULT02, 00481ULT09, 00149ULT40, 00684ULT13, 00018ULT26, 00289ULT06, 10627ULT07, 77266ULT01, 00329ULT23, 00329ULT19, 00139ULT08, 00023ULT36, 00060ULT78, 00232ULT16, 10448ULT49, 10448ULT51, 00116ULT07, 00345ULT13, 00345ULT14, 00312ULT36, 00312ULT35, 00265ULT30, 00082ULT11, 10592US001, 00276ULT01, UZ1056US01); H45041RK (VS6R8813, VS6R8846, VE9R008256, VE9R008267, VS6R8408, VS6R7862, VS6R8041, VS6R7451, VS6R7703, VS6R0422, VS6R3276, VS68627VS6, VS68830VS6, VS68645VS6, VS68827VS6, VS68828VS6, VS67991VS6, VS68651VS6, VS67548VS6, VS68236VS6, VS67570VS6, VS65099VS6, 83098751445, 83103239569, 83108012776, 83112348619, 83122035037, 83137508806, 83137821032, 0911373795, AP6038US02, AH6647US03, AP6924US02, AH0132US46, AP7738US01, AH5526US11, AP7762US01, AH5810US17, AH5810US15, AP7777US01, AP0872US02, AH5806US22, AH4510US09, AP7475US01, AP7483US03, AH5846US42, AH4660US40, AP6072US03, AH0101US97, 140024US25, 140024US26, 653317US01, 090070US13, 090070US16, 110065US28, 030005US07, 220166US18, 210002US61, 210013US22, 110045US12, 871757US01, 110046US08, 100039US04, 681394US01, VS68318VS6, VS69195VS6, VS69229VS6, VS69009VS6, VS69249VS6, VS68221VS6, VS8117VS6, VS69180VS6, VS69152VS6, VS69288VS6, VS68319VS6, VS69012VS6, VS68764VS6, VS68519VS6, VS68799VS6, VS68517VS6, VS69188VS6, VS68763VS6, VS68852VS6, VS68181VS6, VS68364VS6, VS68832VS6, VS68062VS6, VS68829VS6, VS68317VS6, VS68421VS6, VS68024VS6, VS68811VS6, VS63506VS6, VS67720VS6, VS63504VS6, VS68031VS6, VS63541VS6, VS68514VS6, VS68513VS6, VS68361VS6, VS68365VS6, VS69193VS6, VS68800VS6, VS68115VS6, VS68368VS6, VS67316VS6, VS68826VS6, VS69121VS6, VS68981VS6, VS69122VS6, VS68367VS6, VS68512VS6, VS68033VS6, VS68316VS6, VS68034VS6, VS68431VS6, VS69212VS6, VS69169VS6, VS68106VS6, VS65173VS6, VS68051VS6, VS69075VS6, VS68359VS6, VS67267VS6, VS68801VS6, VS67427VS6, VS68054VS6, VS65026VS6, VS68276VS6, VS67967VS6, VS65024VS6, VS68002VS6, VS68026VS6, VS68046VS6, VS68114VS6, VS68047VS6, VS69112VS6, VS68111VS6, VS68082VS6, VS65175VS6, VS68824VS6, VS68040VS6, VS68426VS6, VS65000VS6, VS68437VS6, VS69027VS6, VS67318VS6, VS68315VS6, VS68720VS6, VS68360VS6, VS67993VS6, VS68515VS6, VS6388VS6, VS68797VS6, VS68320VS6, VS68333VS6, VS68362VS6, VS68104VS6, VS63465VS6, VS68314VS6, VS69181VS6, VS67975VS6, VS68378VS6, VS69147VS6, VS68425VS6, VS68438VS6, VS68030VS6, VS68414VS6, VS63561VS6, VS69192VS6, VS68861VS6, VS68486VS6, VS68845VS6, VS68445VS6, VS68329VS6, VS68354VS6, VS69279VS6, VS69226VS6, VS64097VS6, VS68812VS6, VS68006VS6, VS68510VS6, VS68330VS6, VS68809VS6, VS68508VS6, VS68372VS6, VS69198VS6, VS68855VS6, VS69252VS6, VS68357VS6, VS63399VS6, VS69200VS6, VS68060VS6, VS68988VS6, VS68963VS6, VS67056VS6, VS69312VS6, VS67213VS6, VS68860VS6, VS68511VS6, VS68321VS6, VS69267VS6, VS69197VS6, VS69168VS6, VS69286VS6, VS68083VS6, VS68410VS6, BG4609US04, BG4033US02, BG4679US02, VS67937VS6, KS68779VS6, VS68756VS6, VS67285VS6, KS68965VS6, VS68056VS6, VS67762VS6, VS67986VS6, VS69183VS6, VS67247VS6, VS69063VS6, VS67753VS6, VS67482VS6, VS68967VS6, VS68986VS6, VS67729VS6, VS67145VS6, VS67978VS6, VS68561VS6, VS69069VS6, VS69307VS6, VS68831VS6, VS67992VS6, VS67789VS6, VS69061VS6, VS68189VS6, VS68227VS6, VS67825VS6, VS67983VS6.082438060080, 082438040237, 082438040249, 082438040279, 082438040277, 082438030126, 082438160376, 082438190073, 082438190091, 082438220069, 082438160329, 082438090322, 082438160366, 082438160347, 082438270074, 082438190077, 082438210143, 082438230157, 082438100343, 082438250063, 082438110106, 082438160358, 082438120506, 082438150120, 082438240025, 082438100291, 082438170046, 082438170042, 082438130124, 082438130157, 082438160391, 082438120446, 082438050062, 082438210139, 082438150116, 082438150117, 082438210116, 082438110082, 082438090296, 082438090297, 082438290081, 082438290090, 082438160381, 082438160342, 082438240026, 82438130147, VS67806VS6, VS67727VS6, VS68935VS6, VS69231VS6, VS68972VS6, VS67739VS6, VS67678VS6, VS68738VS6, VS69242VS6, VS68979VS6, VS69006VS6, VS64462VS6, VS69239VS6, VS67590VS6, VS69013VS6, VS67741VS6, VS68571VS6, VS68760VS6, VS68206VS6, VS69128VS6, VS68590VS6, VS68389VS6, VS68888VS6, VS67211VS6, VS67465VS6, VS67550VS6, VS68752VS6, DK1047U184, DK1047U121, DK1271US02, DK1018US30, DK1330US01, DK1042US38, DK1042US34, DK1296US11, VS68358VS6, VS68355VS6, VS67940VS6, VS68547VS6, VS67878VS6, FI1063US18, FI1313US13, FI1075US52, FI1109US03, FI1003US43, FI1003US52, FI1003US78, FI1250US01, U200205202, U242974304, U242974303, U146124409, U382506301, U151909205, U150899904, USM4490803, U226287703, U198324010, U198324011, U253439201, US94752308, US94752385, US94752599, U116384901, 138977U005, 143994U008, 804142U010, 110387U025, 110182U018, 803632U003, 104106U002, 6044328U01, 110016U018, 145800U004, 3446313U02, 110783U055, 117062U019, 117062U018, 3902093U01, 145219U029, 4044201U07, 404420U273, 404420U119, 110996U006, 3111906U02, 110012U056, 145424U081, 110012U054, 110009U011, 3438697U02, 4741984U01, 148127U008, 120085U004, 110536U013, 103185U663, 145856U003, 145135U024, 1881872U03, 2561480U02, 1464242U09, 112032U004, 6135636U02, 103135U046, 103135U055, 1513970U07, 1513970U16, 6860752U01, 2955286U02, 2261281U02, 4030987U01, 122024U036, 110040U010, 122024U028, 122024U032, 122024U030, 7485350U01, 142410U004, 800625U014, 804426U004, 148004U004, 1773661U02, 110320U005, 145041U002, 145287U021, 802786U140, 802786U147, 103130U005, 5020808U22, 109970U023, 4540415U01, 3792593U01, 121623U009, 107006U018, 4716353U01, 120017U013, 6644453U01, 210502U007, 3635915U01, 106590U004, 148273U005, 120064U113, 120064U110, 120075U012, 147995U009, TP1103U027, 6419780U01, 122008U031, 122008U037, TP3007U006, 104026U012, 1702988U02, 3066456U03, 3066456U01, 138058U009, 110076U010, 8276339U02, 149615U017, 142541U070, 142541U071, 110295U009, 103041U007, 112037U077, 1305892U09, 114248U009, 114248U011, 104451U930, 110899U006, 110939U014, 4158391U01, 101154U019, 103302U079, 142781U002, 142453U046, 140468U008, 4040833U01, 2622220U02, 110604U049, 113128U058, 113128U059, 2358027U02, 3595039U01, 109341U009, 145461U007, 120035U030, 1874450U02, 140141U019, 104077U027, 2215719U03, 110596U026, 1874554U05, 110560U044, 802891U005, 138251U022, 125127U011, 138251U023, 5176316U01, 3040458U01, 800234U003, 142377U007, 1777146U02, 3867691U01, 1270657U02, 1543156U04, 104066U017, 2333539U02, 4873462U01, 2279761U02, 150149U003, 2108250U92, 2108250U50, 2222813U02, 1915916U06, 4186057U01, 2552578U02, 145358U019, 148085U004, 3896619U01, 119436U075, 110742U008, 1209532U05, H5991US02, H2020US04, H6908US01, 0828380401, 0828380427, 0828380444, 0828380446, HU1945US04, HU1551US01, 083038865005613, 083038240305213, 083038248452613, 083038061080113, 083038060070013, 083037000128614, 083037203338414, 083038241088514, 083038241093714, 083038840123613, 083038860520814, 083038860528514, 083037001012614, 083037600026114, 083038000109813, 083038000305014, 083038000320813, 083038001032214, 083038020081314, 083038021022314, 083038021041514, 083038021101215, 083038060097613, 083038060104013, 083038060129414, 083038061011714, 083038061032914, 083038101052913, 083038101064614, 083038200060514, 083038200094713, 083038201016614, 083038201058413, 083038201061914, 083038240128714, 083038240505714.083038241018214, 083038600021414, 083038600025514, 083038600026214, 083038600026314, 083038600306714, 083038800334315, 083038800409114, 083038801063514, 083038820012615, 083038820013915, 083038821096915, 083038841091515, 083038865017414, 083036821031813, 083037821031813, 083038821031613, 083038821031813, 083038821043613, 083037021094915, 083037288363615, 083038000061916, 083038020151816, 083038020308115, 083038021037516, 083038040111016, 083038040131916, 083038060097417, 083038062102616, 083038100098816, 083038101023715, 083038108010215, 083038108031715, 083038200006515, 083038200094417, 083038200106917, 083038200302116, 083038201083416, 083038202127316, 083038209012315, 083038240059216, 083038240074015, 083038240113217, 083038240126317, 083038241026115, 083038241084416, 083038242091915, 083038280016216, 083038600015015, 083038600028616, 083038600039216, 083038600339116, 083038600541116, 083038601072316, 083038602124816, 083038800003815, 083038800051916, 083038800111516, 083038800141916, 083038801334317, 083038802087415, 083038802122716, 083038802123416, 083038804724115, 083038805000315, 083038805005715, 083038806345416, 083038821000415, 083038821006815, 083038823643416, 083038825233515, 083038840075715, 083038860068815, 083037800506314, 083037840099014, 083038000121914, 083038001020014, 083038005833214, 083038100106314, 083038240097514, 083038280114913, 083038602770514, 083038800046414, 083038820008914, 083038820011913, 083038821020313, 083038827209614, 083038827446614, 083038840114313, 083038821015913, 083038001550512, 083038061016313, 083038601093513, 083038061055812, 083038801073913, 083038801019313, 083037201085514, 083038200114213, 083038600327114, 083038820012513, 083038101038813, 083038820002413, 083037600107913, 083038860510213, 083038801096813, 083038061018713, 083038101075713, 083038021024313, 083037021069713, 083038021010813, 083038601065512, 083038600308213, 083038600141913, 083038020104513, 083038000109113, 083038821052013, 083038821073713, 083038001024813, 083038280142613, 083038040064113, 083038821073613, 083038060142813, 083038241019613, 083038201024013, 083038860507413, 083037801048414, 083038241046313, 083038860536313, IE1024US22, IE1018US20, IE1199US01, A6232601, A6249001, A251050201, A5235905, A251398007, A58122108, A174760401, B5245002, A6082204, A5645607, A5103351, A179667116, A140377602, A6208401, A6135202, A137307103, F5672201, A5691703, B5346380, A6203301, A5926902, A5926904, A6272901, A6200801, A5391803, A6198501, UX9637, UX9713, UXA885, UXA593, UX9765, UXA873, UX9902, UXA815, UX9639, UX9728, UX9620, UX9703, UX9579, UX9590, UX9683, UX9788, UX9914, UX9664, UX9781, UX9622, UX9847, UXA522, UXA777, UXA638, UX9851, UX9812, UX9619, UX9383, UXA904, UXA854, UX9890, UX9879, UX9190, UX9886, UXA502, UX9727, UXA651, UXA876, UXA524, UX9702, UX9819, UX9705, UX9594, UX9921, UX9725, UX9804, UX9883, UX9739, UXA450, UXA055, UX9625, UXA867, UX9682, UX9644, UX9591, UXA855, UX9820, UXA697, UX9849, UX9729, UXA551, UXA711, UX9919, UXA825, UX9844, UXA872, UX9845, UX9846, UX9853, UX9913, UX9704, UX9763, UX9149, UX9719, UX9824, UX9600, UXA905, UX9776, UX9663, UXA603, UX9864, UX9669, UX9681, UX9623, UX9665, UX9640, UX9889, UX9731, UX9730, UX9834, UX9833, UXA761, UX8050, UX9714, UX9668, UX9871, UX9870, UX9147, UX9592, UX9901, UX9758, UXA653, UXA449, UX9737, UX9878, UX9917, UX9882, UXA500, UX9294, UX9810, UXA595, UX9839, UXA654, UXA701, UX9848, UX9884, UX9058, UX9915, UX9918, UXA911, UX9762, UX9578, UX9641, UX9903, UXA448, UX9626, UX9910, UX9850, UXA725, UX9852, UXA696, UXA798, UX9601, UX9875, UX9712, UX9754, UX9624, UX9880, UX9616, UX9618, UX9676, UX9617, UX9710, UXA803, UX9757, UX9814, UX9832, UXA446, UXA747, UX9670, UX9837, UX9621, UX9802, UX9835, UX9811, UX9597, UX9907, UX9598, UX9813, UX9891, UX9836, UX9783, UX9764, UX9797, UX9877, UX9908, UX9909, UX9599, UX9838, UX9755, UX9920, UX9906, UXA756, UX9786, UXA571, UX9760, UX9675, UX9856, UX9857, UX9596, UX9602, KA3077US03, 0850370249, 0850370993, 0850371085, 0850371367, 0850371408, 0850371560, 0850371759, 0850371827, 0850371845, 0850371979, 0850372142, 0850372186, 0850372325, 0850372351, 0850372450, 0850372493, 0850372813, 0850373089, 0850373383, 0850373698, 0850375115, VS69215VS6, VS67842VS6, VS68987VS6, VS69287VS6, VS69275VS6, VS69289VS6, VS69294VS6, VS67902VS6, VS68823VS6, VS68822VS6, VS67658VS6, VS69300VS6, VS69244VS6, VS69230VS6, VS69290VS6, VS69251VS6, VS69309VS6, VS68975VS6, VS67472VS6, VS69285VS6, VS67468VS6, 083037000150512, ALK01408, APL02423, APL02422, KMP0141, LDN01425, ZWL0445, 85108155152, 85114254507, 85135271069, NO4011U126, NO4004US43, NO4018US37, NO4033US15, NO4014US49, NO4302US14, NO4246US05, NO4009US51, NO4009US74, NO4772US01, VS68310VS6, VS69119VS6, VS67364VS6, 0843384126, PL1023US29, PL2768US01, PL2266US02, PL3567US01, PL1602US21, PL3555US01, PL3220US02, PL3670US01, PL0447US01, P00011US01, PL3365US01, PL0359US01, PL3224US02, PL3607US01, PL0397US04, PL3366US01, PL1658US01, PL1006US34, PL1006US25, PL1006US46, PL0462US02, PL0141US03, PL0543US02, PPS64700, PPR85301, PPS19808, VS63869VS6, VS67670VS6, VS67175VS6, RO1326US02, RU8932US01, RU8490US01, RU2721US02, RU7782US01, RU7544US04, RU7786US01, RU7796US01, RU3846US02, RU7783US01, RU2551US01, RU9527US01, RU7746US01, RU7746US12, RU6248US03, RU8196US01, RU8221US03, RU5488US05, RU5488US06, RU5488US07, RU6340US07, RU7793US01, RU8711US01, RU7866US01, RU8466US01, RU9501US01, RU5150US04, RU7790US01, RU2552US03, RU7547US03, RU7785US01, RU7548US02, RU4948US02, RU1586US02, RU4267US03, RU3078US01, RU1558US04, RU1558US05, RU7947US01, RU7800US01, RU1085US06, RU1914US04, RU7298US07, RU1225US06, RU1295US24, RU1295US25, RU1757US05, RU4313US06, RU4313US07, RU8267US04, RU9303US01, RU2126US04, RU5440US05, RU5440US06, RU3639US02, RU7730US04, RU9516US01, RU6790US03, RU8147US01, RU3191US02, RU7575US02, RU7789US01, RU7791US01, RU8505US01, RU7179US01, RU8308US02, RU9755US01, RU9881US01, RU1047US06, RU8311US01, RU9651US01, RU7792US01, RU7803US01, RU9457US02, RU9457US03, RU5552US05, RU2772US01, RU3015US06, RU8585US01, RU8790US01, RU3468US39, RU9232US05, RU3670US05, RU2345US01, RU1896US05, RU2353US02, RU6284US03, RU8096US01, RU8668US01, RU8911US01, RU7604US02, RU8714US01, RU7219US02, RU7734US01, RU8462US01, RU4489US08, RU2468US03, RU4815US03, RU2010US02, RU8415US01, RU6633US02, RU2826US02, 6073UL0159, 6073UL0161, 6073UL0160, 6073UL0188, DEL89988, FE816173, DEN00828, NEV44767, FPG36300, DPB15871, DPN88633, MP842019, 083038802676816, S005US04, M007US21, F001US35, S001US25, S001US23, S001US24, A130US01, C001US41, C001US42, 547020US05, 957979US01, 412080US02, 560040US04, 475430US09, 505090US04, 0853371513, 0853371556, 0853371557, 0853380334, 0853380344, 0853380349, 0853380355, 0853380363, 0853380386, 0853380387, 0853380388, 1573UL0344, 0856380257, 0856380262, 0856380275, 0856380284, 0856380296, 0856380302, 0856380320, 0856380334, 0856380336, 0856380348, 1573UL0047, 0856380289, 0856380290, 0856380291, 0856380292, 0856380326, 0856380327, 0856380328, 0856380335, 198276112, 198318982, 21136ULT02, 21097US035, 00288ULT11, 00288ULT08, 190631732, 00091ULT10, 00091ULT11, 05217ULT11, 00058ULT40, 175963302, 186807042, 00468ULT10, 159217532, 00014ULT16, 21096ULT01, 194585602, 00393ULT13, 1923334402, 10310ULT02, 01240ULT01, 21122ULT10, 156868742, 175897372, 175985252, 132200334, 196713062, 182755257, 110367972, 110385362, 100100972, 188867222, 00223ULT15, 190884912, 10447ULT11, 169772352, 00361ULT05, 159560902, 198356372, 191795502, 00442ULT06, 1949154402, VS67867VS6, VS64547VS6, VS68754VS6, 5973UL0878, 5973UL0877, 5973UL0885, 5973UL0879, 5973UL0880, 5973UL0881, 5973UL0001, 5973UL0080, 5973UL0002, 5973UL0129, 5973UL0126, 5973UL0127, 5973UL0128, 5973UL0286, 5973UL0289, 0859384105, 0859384109) |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | GE Healthcare Service 800-437-1171 |
Manufacturer Reason for Recall | Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire. |
FDA Determined Cause 2 | Device Design |
| Action | GE Healthcare notified consignees on about 11/10/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION. For U.S. customers, letters were sent via FedEx. Consignees were instructed to follow updated safety instructions provided in the appendix included with the notification letter, place the appendix with the product labeling, and replace the batter every 2 years or if the battery is not capable of powering the system for more than 30 minutes. In response, GE Healthcare requested consignees return the response form. |
| Quantity in Commerce | 5,145 units |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, AZERBAIJAN, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, CHINA Colombia, CONGO, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, CZECH REPUBLIC, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, HONG KONG, Hungary, ICELAND, INDIA, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Kuwait, KYRGYZSTAN, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, LUXEMBOURG, Macedonia, MALAYSIA, Mali, Mauritania, Mauritius, Mexico, MICRONESIA, Moldova, Morocco, Mozambique, Namibia, Nepal, Netherlands, NEW ZEALAND, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, SYRIA, TAIWAN, TANZANIA, THAILAND, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = IYN 510(K)s with Product Code = IYN 510(K)s with Product Code = IYN
|
|
|
|