| | Class 2 Device Recall Sterile Procedural Trays |  |
| Date Initiated by Firm | November 17, 2022 |
| Date Posted | February 06, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1078-2023 |
| Recall Event ID |
91386 |
| Product Classification |
Kit, surgical instrument, disposable - Product Code KDD
|
| Product | Sterile Procedural Trays, labeled as the following:
a. ST. ANNE'S MAJOR BASIN SET UP
b. SINGLE BASIN SET
c. DOUBLE BASIN SET
d. LAPAROTOMY BASIN
e. ORTHOPEDIC CUSTOM BASIN SET
f. OISC BASIN TRAY
g. BASIN PACK
h. BASIN SET
i. SURGI START PACK
j. PK, GEN-SET UP - MIDWEST
k. SINGLE BASIN PACK-LF
l. AGHM MINOR SINGLE PACK
m. MINOR PACK
n. SINGLE BASIN PACK
o. MAJOR BASIN PACK
p. BASIC SET UP-LF
q. BASIC SET-UP-LF
r. BASIN PACK
s. MINOR BASIN
t. AMC MAJOR BASIN
u. AMC MAJOR BASIN |
| Code Information |
a. ST. ANNE'S MAJOR BASIN SET UP, Item Number: DYNJ24628M, Case UDI/GTIN: 40195327109135, Unit UDI/GTIN: 10195327109134, Lot Number: 22JLA323,22ILA590;
b. SINGLE BASIN SET, Item Number: DYNJ35592K, Case UDI/GTIN: 40195327110827, Unit UDI/GTIN: 10195327110826, Lot Number: 22JBW819,22IBK206;
c. DOUBLE BASIN SET, Item Number: DYNJ35596G, Case UDI/GTIN: 40195327110810, Unit UDI/GTIN: 10195327110819, Lot Number: 22JBH675,22HBY465;
d. LAPAROTOMY BASIN, Item Number: DYNJ41354B, Case UDI/GTIN: 40195327246311, Unit UDI/GTIN: 10195327246310, Lot Number: 22JBE052;
e. ORTHOPEDIC CUSTOM BASIN SET, Item Number: DYNJ47991J, Case UDI/GTIN: 40195327230600, Unit UDI/GTIN: 10195327230609, Lot Number: 22JBX225,22JBH710,22HBW917,22HBV171;
f. OISC BASIN TRAY, Item Number: DYNJ48354I, Case UDI/GTIN: 40195327154302, Unit UDI/GTIN: 10195327154301, Lot Number: 22IMF862;
g. BASIN PACK, Item Number: DYNJ50123D, Case UDI/GTIN: 40195327201990, Unit UDI/GTIN: 10195327201999, Lot Number: 22HBU014;
h. BASIN SET, Item Number: DYNJ50906F, Case UDI/GTIN: 40195327140145, Unit UDI/GTIN: 10195327140144, Lot Number: 22JMB641,22HMD087;
i. SURGI START PACK, Item Number: DYNJ52541D, Case UDI/GTIN: 40195327164509, Unit UDI/GTIN: 10195327164508, Lot Number: 22JBH137;
j. PK, GEN-SET UP - MIDWEST, Item Number: DYNJ57608B, Case UDI/GTIN: 40195327253296, Unit UDI/GTIN: 10195327253295, Lot Number: 22JMB780;
k. SINGLE BASIN PACK-LF, Item Number: DYNJ58389D, Case UDI/GTIN: 40195327144198, Unit UDI/GTIN: 10195327144197, Lot Number: 22GBW998;
l. AGHM MINOR SINGLE PACK, Item Number: DYNJ62766B, Case UDI/GTIN: 40195327116775, Unit UDI/GTIN: 10195327116774, Lot Number: 22JMC641,22HMD203;
m. MINOR PACK, Item Number: DYNJ63961B, Case UDI/GTIN: 40195327146970, Unit UDI/GTIN: 10195327146979, Lot Number: 22IDA519;
n. SINGLE BASIN PACK, Item Number: DYNJ66012A, Case UDI/GTIN: 40195327221455, Unit UDI/GTIN: 10195327221454, Lot Number: 22JLA612,22ILA886,22HLA208;
o. MAJOR BASIN PACK, Item Number: DYNJ69406A, Case UDI/GTIN: 40195327112333, Unit UDI/GTIN: 10195327112332, Lot Number: 22JMH995,22JMA882;
p. BASIC SET UP-LF, Item Number: DYNJ81646, Case UDI/GTIN: 40195327130269, Unit UDI/GTIN: 10195327130268, Lot Number: 22IBH244;
q. BASIC SET-UP-LF, Item Number: DYNJ82390, Case UDI/GTIN: 40195327189540, Unit UDI/GTIN: 10195327189549, Lot Number: 22JBV631,22IBC646;
r. BASIN PACK, Item Number: DYNJ82969, Case UDI/GTIN: 40195327218325, Unit UDI/GTIN: 10195327218324, Lot Number: 22JBN923;
s. MINOR BASIN, Item Number: DYNJ907808B, Case UDI/GTIN: 40195327244430, Unit UDI/GTIN: 10195327244439, Lot Number: 22JBU478,22JBA257,22IBV340,22IBS662;
t. AMC MAJOR BASIN, Item Number: GSP158AA, Case UDI/GTIN: 40195327143511, Unit UDI/GTIN: 10195327143510, Lot Number: 22HBT951;
u. AMC MAJOR BASIN, Item Number: GSP158AB, Case UDI/GTIN: 40195327252022, Unit UDI/GTIN: 10195327252021, Lot Number: 22JBD534 |
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Karin Johnson 886-359-1704 |
Manufacturer Reason for Recall | Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging. |
FDA Determined Cause 2 | Device Design |
| Action | Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed. |
| Quantity in Commerce | 2.542 cases (6,739 units) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|