| | Class 2 Device Recall OSSTM Compress & Mini Compress Anchor Plugs |  |
| Date Initiated by Firm | October 09, 2023 |
| Date Posted | November 21, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0371-2024 |
| Recall Event ID |
93323 |
| 510(K)Number | K062998 |
| Product Classification |
Prosthesis, hip, semi-constrained (metal uncemented acetabular component) - Product Code KWA
|
| Product | Compress Device Segmental Anchor Plug, 12 MM -For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
Item Number: 178402 |
| Code Information |
Lot Numbers/UDI
240750 (01)00880304259713(17)270112(10)240750
671330 (01)00880304259713(17)271114(10)671330
671420 (01)00880304259713(17)270831(10)671420
671660 (01)00880304259713(17)270901(10)671660
752730 (01)00880304259713(17)270802(10)752730
752760 (01)00880304259713(17)270816(10)752760
798010 (01)00880304259713(17)270120(10)798010
921660 (01)00880304259713(17)270724(10)921660
930670 (01)00880304259713(17)270727(10)930670
930680 (01)00880304259713(17)270802(10)930680
965330 (01)00880304259713(17)271112(10)965330 |
| FEI Number |
1825034
|
Recalling Firm/ Manufacturer |
Biomet, Inc. 56 E Bell Dr Warsaw IN 46582-6989
|
| For Additional Information Contact | 411 Technical Services 574-371-3071 |
Manufacturer Reason for Recall | Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery |
FDA Determined Cause 2 | Device Design |
| Action | Zimmer Biomet issued Urgent Medical Device Recall letter on 10/09/23 to Distributors and Risk Managers via email/courier . Letter states reason for recall, health risk and action to take:
1. Review this notification and ensure that affected personnel are aware of the contents.
2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility.
3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation.
4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility.
5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation.
6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com |
| Distribution | Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA,
CANADA, CHILE, JAPAN, NETHERLANDS, NEW ZEALAND.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KWA
|
|
|
|