| Class 2 Device Recall VITROS 5600 Immunodiagnostic System |  |
Date Initiated by Firm | January 26, 2024 |
Date Posted | March 01, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-1231-2024 |
Recall Event ID |
93927 |
510(K)Number | K081543 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product | VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished) |
Code Information |
VITROS System Software Versions 3.8.0 or 3.8.1
Regular UDI-DI 10758750002740
Refurbished UD-DI 10758750007110
Serial Numbers:
US:
J56000116 J56003369
J56000317 J56003373
J56001292 J56003377
J56001523 J56003433
J56001629 J56003477
J56001885 J56003478
J56001886 J56003484
J56001910 J56003496
J56002041 J56003541
J56002159 J56003570
J56002198 J56003578
J56002199 J56003581
J56002201 J56003583
J56002249 J56003632
J56002301 J56003646
J56002308 J56003655
J56002461 J56003668
J56002463 J56003669
J56002466 J56003670
J56002474 J56003672
J56002477 J56003725
J56002500 J56003730
J56002591 J56003752
J56002603 J56003759
J56002607 J56003774
J56002608 J56003911
J56002615 J56004068
J56002698 J56004137
J56002703 J56004174
J56002704 J56004240
J56002722 J56004325
J56002769 J56004326
J56002895 J56004485
J56003035 J56004491
J56003045 J56004498
J56003076 J56004506
J56003113 J56004512
J56003199 J56004549
J56003220 J56004551
J56003287 J56004571
J56003299 J56004577
J56003317 J56004587
J56003330 J56004589
J56003334 J56004604
J56003342 J56004606
J56003356
OUS:
J56000190 J56001929
J56000246 J56001944
J56000418 J56001979
J56000425 J56002028
J56000468 J56002064
J56000512 J56002085
J56000533 J56002099
J56000665 J56002116
J56000671 J56002218
J56000674 J56002258
J56000680 J56002300
J56000719 J56002410
J56000809 J56002460
J56000831 J56002518
J56000836 J56002615
J56000910 J56002625
J56000929 J56002637
J56000993 J56002784
J56000994 J56002821
J56001015 J56002862
J56001051 J56002913
J56001097 J56002926
J56001111 J56002929
J56001114 J56002966
J56001140 J56003061
J56001208 J56003162
J56001210 J56003179
J56001237 J56003260
J56001260 J56003285
J56001320 J56003401
J56001355 J56003451
J56001378 J56003464
J56001389 J56003508
J56001469 J56003523
J56001509 J56003538
J56001591 J56003919
J56001631 J56003951
J56001689 J56004078
J56001731 J56004082
J56001733 J56004308
J56001765 J56004309
J56001780 J56004346
J56001793 J56004355
J56001826 J56004414
J56001844 J56004415
J56001853 J56004531
J56001925 J56004695
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Dr Rochester NY 14626-5101
|
For Additional Information Contact | Mr. Joe Falvo 585-453-3000 |
Manufacturer Reason for Recall | A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported. |
FDA Determined Cause 2 | Software Design Change |
Action | On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules.
The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024.
If you have further questions, please contact Global Services Organization at 1-800-421-3311. |
Quantity in Commerce | 91 US; 94 OUS |
Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil,
Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico,
Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
|
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|