| Class 2 Device Recall VITROS XT 7600 Integrated System |  |
Date Initiated by Firm | January 26, 2024 |
Date Posted | March 01, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-1233-2024 |
Recall Event ID |
93927 |
510(K)Number | K182063 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product | VITROS XT 7600 Integrated System, Catalog No. 6844461 |
Code Information |
VITROS System Software Versions 3.8.0 or 3.8.1
UDI-DI 10758750031610
Serial Numbers:
US:
J76000050 J76000878 J76001555
J76000108 J76000905 J76001558
J76000144 J76000909 J76001559
J76000148 J76000922 J76001565
J76000179 J76000930 J76001570
J76000192 J76000968 J76001577
J76000195 J76001106 J76001581
J76000199 J76001171 J76001583
J76000200 J76001172 J76001589
J76000206 J76001206 J76001592
J76000207 J76001207 J76001595
J76000208 J76001218 J76001596
J76000241 J76001230 J76001597
J76000298 J76001231 J76001599
J76000313 J76001232 J76001600
J76000317 J76001234 J76001611
J76000324 J76001235 J76001616
J76000333 J76001239 J76001617
J76000342 J76001260 J76001620
J76000352 J76001263 J76001638
J76000353 J76001270 J76001639
J76000420 J76001274 J76001643
J76000422 J76001292 J76001644
J76000456 J76001293 J76001647
J76000459 J76001294 J76001648
J76000460 J76001331 J76001650
J76000468 J76001340 J76001652
J76000495 J76001351 J76001659
J76000502 J76001357 J76001660
J76000524 J76001370 J76001666
J76000525 J76001376 J76001667
J76000537 J76001383 J76001668
J76000561 J76001384 J76001669
J76000609 J76001385 J76001671
J76000709 J76001392 J76001674
J76000743 J76001404 J76001675
J76000747 J76001406 J76001678
J76000783 J76001516 J76001679
J76000790 J76001517 J76001687
J76000803 J76001518 J76001692
J76000805 J76001519 J76001700
J76000806 J76001520 J76001703
J76000817 J76001535 J76001705
J76000862 J76001540 J76001711
J76000866 J76001549 J76001714
J76000868 J76001552 J76001748
J76000871 J76001553 J76001879
J76000873
OUS:
J76000106 J76000976
J76000118 J76000983
J76000211 J76001001
J76000253 J76001027
J76000257 J76001035
J76000265 J76001070
J76000274 J76001074
J76000276 J76001095
J76000278 J76001111
J76000280 J76001121
J76000281 J76001130
J76000284 J76001141
J76000288 J76001142
J76000289 J76001144
J76000290 J76001152
J76000292 J76001155
J76000293 J76001159
J76000370 J76001174
J76000436 J76001183
J76000446 J76001189
J76000447 J76001272
J76000475 J76001273
J76000620 J76001285
J76000628 J76001309
J76000760 J76001314
J76000782 J76001328
J76000807 J76001368
J76000813 J76001377
J76000838 J76001378
J76000845 J76001379
J76000855 J76001396
J76000856 J76001399
J76000888 J76001407
J76000898 J76001408
J76000901 J76001452
J76000973 J76001457
J76001484
J76001547
J76001562
J76001563
J76001673
J76001761
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Dr Rochester NY 14626-5101
|
For Additional Information Contact | Mr. Joe Falvo 585-453-3000 |
Manufacturer Reason for Recall | A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported. |
FDA Determined Cause 2 | Software Design Change |
Action | On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules.
The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024.
If you have further questions, please contact Global Services Organization at 1-800-421-3311. |
Quantity in Commerce | 142 US; 78 OUS |
Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil,
Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico,
Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
|
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|