| | Class 2 Device Recall Exactech Equinoxe |  |
| Date Initiated by Firm | March 06, 2024 |
| Date Posted | March 29, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1422-2024 |
| Recall Event ID |
94092 |
| 510(K)Number | K063569 |
| Product Classification |
shoulder prosthesis, reverse configuration - Product Code PHX
|
| Product | Exactech Equinoxe REVERSE SHOULDER,42mm Humeral Liner, a) +0mm, Item number 320-42-00, b) +2.5mm, Item Number 320-42-03; Shoulder Arthroplasty |
| Code Information |
a) Item number 320-42-00, UDI/DI 10885862086693, Serial Numbers: 0982062, 1064609, 1093925, 1000132, 1000124, 1064624, 1000137, 1064629, 1064643, 1064639, 1093902, 1064617, 1093913, 1093909, 1093905, 1000121, 1000115, 1064610, 0982056, 1000123, 0982059, 1000128, 0982053, 0982073, 0982064, 0982066, 0982076, 0982063, 0982071, 0982065, 1064636, 0982074, 1064642, 0982075, 1000136, 1000135, 1000131, 1093931, 1000130, 0982070, 1000117, 1000113, 1064634, 1000129, 1064623, 1093910, 1000116, 0982055, 1064594, 1064600, 1064614, 1064638, 1064641, 1000126, 1000122, 0982058, 1064633, 1093912, 1093939, 1093907, 1093922, 1093923, 1093928, 1093929, 1100342, 1064611, 1093918, 1100320, 1100347, 1100359, 1064602, 1064640, 1064606, 1064631, 1064607, 1064604, 1100357, 1100361, 1100317, 1093924, 1100349, 1100327, 1000114, 1064625, 1100345, 1100353, 1000125, 1064598, 1093908, 1064632, 1064612, 1064595, 1000138, 1093919, 1000119, 1000120, 1064596, 1100331, 1100336, 1064635, 1093935, 1100324, 1064622, 1093901, 1100323, 0982068, 0982061, 1100341, 1100334, 1064627, 1100363, 1093917, 1100322, 1100364, 0982060, 1064613, 1064608, 1100346, 1064621, 1093911, 0982052, 1093920, 1100329, 1100365, 1100328, 1064603, 1064616, 1093934, 1100360, 1100355, 1100366, 1100326, 1100344, 1064597, 1100335, 0982072, 1064630, 1064593, 1064618, 1093921, 1100337, 1100338, 1093914, 1100358, 1100348, 1100362, 1100319, 1100321, 1100330, 1093916, 1100332, 1100354, 1000134, 1093930, 1093927, 1290049, 1216151, 1358417, 1216160, 1216119, 1216208, 1216205, 1216183, 1216189, 1290022, 1290006, 1289971, 1365488, 1290020, 1216125, 1365487, 1216194, 1216130, 1216106, 1290058, 1365459, 1358383, 1216177, 1289977, 1358410, 1216128, 1358425, 1216105, 1216140, 1216116, 1216108, 1216182, 1216173, 1290025, 1216132, 1216193, 1290029, 1290059, 1216120, 1216136, 1365468, 1289973, 1216188, 1289964, 1216186, 1216190, 1216152, 1289968, 1216143, 1216201, 1216170, 1216104, 1290007, 1216133, 1216175, 1216204, 1290026, 1216157, 1216169, 1358413, 1216192, 1216185, 1216155, 1216198, 1290056, 1216115, 1216181, 1216167, 1290027, 1216146, 1216162, 1289990, 1290053, 1290054, 1216114, 1216144, 1216141, 1289982, 1358418, 1290008, 1290034, 1216118, 1216110, 1216111, 1216122, 1216124, 1216129, 1216159, 1216165, 1216166, 1216107, 1216134, 1216135, 1216164, 1216206, 1216168, 1216171, 1216176, 1216179, 1216149, 1289996, 1289965, 1289980, 1289988, 1289993, 1289999, 1290001, 1290002, 1290003, 1290011, 1290023, 1290024, 1290031, 1290033, 1290041, 1290043, 1290045, 1290047, 1290050, 1290051, 1290052, 1216203, 1093926, 1100325, 1064637, 1093903, 1100333, 1100340, 1100316, 1100343, 1093938, 1064619, 1064599, 1064626, 1093933, 1064628, 0982054, 1100356, 1064615, 1100339, 1528611, 1216163, 1358377, 1365497, 1216138, 1216154, 1289966, 1438100, 1358385, 1430227, 1358395, 1430270, 1482896, 1430220, 1216156, 1528610, 1430234, 1365499, 1289995, 1545563, 1358382, 1552360, 1459738, 1358407, 1358396, 1552324, 1290004, 1438090, 1430282, 1445734, 1445728, 1552337, 1358372, 1365480, 1290017, 1438104, 1482876, 1358398, 1482898, 1528612, 1482892, 1476773, 1459728, 1482880, 1476772, 1289983, 1365486, 1365470, 1358391, 1482908, 1358401, 1459737, 1482890, 1358420, 1358406, 1290021, 1430278, 1365458, 1438089, 1459744, 1482909, 1482887, 1438101, 1552340, 1476780, 1216102, 1445732, 1290005, 1358423, 1290016, 1545573, 1365495, 1430202, 1216158, 1216202, 1365496, 1552339, 1476788, 1365461, 1438108, 1365489, 1482895, 1459734, 1535735, 1358393, 1290000, 1290044, 1290012, 1216195, 1358381, 1358380, 1365464, 1365456, 1552341, 1430222, 1430221, 1216126, 1365482, 1289978, 1535738, 1430276, 1438097, 1289969, 1535746, 1535749, 1430243, 1358419, 1482873, 1290032, 1482891, 1290042, 1438102, 1365500, 1482882, 1365492, 1482883, 1545564, 1545562, 1528615, 1535748, 1545575, 1289987, 1430275, 1528614, 1358411, 1365473, 1365502, 1430246, 1430250, 1358388, 1358374, 1289985, 1430271, 1365475, 1289998, 1365460, 1289984, 1290028, 1358424, 1438093, 1445729, 1438105, 1430237, 1438092, 1430231, 1535733, 1358402, 1216172, 1365498, 1358386, 1365469, 1358397, 1482885, 1459747, 1365481, 1545566, 1430256, 1430218, 1459745, 1476783, 1430219, 1482899, 1430230, 1216161, 1430289, 1430236, 1482877, 1438094, 1430215, 1216174, 1552354, 1482901, 1545579, 1528599, 1528616, 1290030, 1290037, 1365472, 1365484, 1365479, 1430251, 1430224, 1358404, 1216200, 1290010, 1365476, 1289986, 1290039, 1289967, 1216145, 1365467, 1445733, 1358403, 1476785, 1290055, 1438095, 1358379, 1430203, 1438103, 1358405, 1430216, 1290015, 1216147, 1358373, 1552361, 1482881, 1445731, 1535737, 1535734, 1365454, 1365466, 1358394, 1289972, 1482888, 1482875, 1552325, 1528590, 1430226, 1365463, 1430283, 1430228, 1535736, 1445735, 1445730, 1476771, 1430212, 1528585, 1528586, 1528598, 1216123, 1289989, 1289981, 1365491, 1552342, 1594945, 1594953, 1459742, 1482902, 1459740, 1482907, 1289975, 1430213, 1528587, 1289992, 1358399, 1430241, 1552359, 1216112, 1290040, 1365485, 1552353, 1459746, 1438099, 1476781, 1216187, 1476770, 1459741, 1476784, 1528604, 1216196, 1365462, 1438096, 1482897, 1358421, 1216184, 1430238, 1430259, 1528617, 1482903, 1552334, 1430258, 1430287, 1552343, 1528581, 1430205, 1430239, 1358390, 1482871, 1545569, 1430207, 1594947, 1093936, 1100352, 1093915, 1482870, 1216191, 1358375, 1216127, 1365501, 1365494, 1365471, 1216153, 1535731, 1430214, 1528588, 1594958, 1430210, 1476769, 1552331, 1528592, 1535741, 1552348, 1604466, 1528582, 1290038, 1290018, 1358408, 1459733, 1528607, 1594954, 1528580, 1594956, 1482889, 1604453, 1604442, 1604431, 1289970, 1216148, 1535747, 1552326, 1358415, 1365465, 1358378, 1535739, 1430206, 1552336, 1604426, 1438098, 1289979, 1430249, 1528578, 1594959, 1430281, 1430232, 1552344, 1528584, 1545574, 1216113, 1459731, 1552333, 1430288, 1476774, 1528605, 1545565, 1604469, 1552350, 1289997, 1552349, 1430217, 1289944, 1430280, 1476778, 1430242, 1430245, 1552362, 1482893, 1430223, 1289974, 1482879, 1545568, 1358412, 1430277, 1528608, 1604430, 1604461, 1365478, 1545567, 1594955, 1535730, 1545577, 1445736, 1604435, 1552358, 1535745, 1594946, 1430225, 1528603, 1545572, 1430201, 1535743, 1552363, 1290048, 1290035, 1430229, 1482874, 1430254, 1528589, 1604441, 1594952, 1604460, 1430209, 1438106, 1528579, 1604462, 1476776, 1552338, 1594948, 1604434, 1476786, 1545581, 1358400, 1545580, 1552335, 1604455, 1545571, 1430204, 1216178, 1216103, 1430252, 1216207, 1216109, 1604433, 1430247, 1430244, 1430240, 1365457, 1604447, 1604464, 1604446, 1604436, 1604450, 1430255, 1430272, 1604456, 1358387, 1358392, 1365493, 1430248, 1216131, 1289963, 1594944, 1604439, 1604451, 1290014, 1594951, 1535740, 1552355, 1552356, 1594949, 1358416, 1552346, 1604428, 1604457, 1430284, 1604437, 1604468, 1290009, 1528583, 1482906, 1528596, 1528600, 1216117, 1528597, 1528606, 1476779, 1528594, 1290013, 1528613, 1430233, 1482884, 1552328, 1604440, 1604425, 1552327, 1545570, 1430208, 1604429, 1594957, 1604438, 1604444, 1604459, 1604452, 1476787, 1604443, 1552357, 1289994, 1216137, 1784905, 1765521, 1765524, 1673240, 1673241, 1648560, 1730661, 1634875, 1648542, 1634853, 1648535, 1648541, 1648564, 1673224, 1673249, 1730637, 1766107, 1648567, 1765519, 1841527, 1841515, 1730665, 1730656, 1887013, 1673204, 1765512, 1841499, 1887033, 1634864, 1730635, 1634863, 1634862, 1648536, 1648529, 1765518, 1673214, 1798007, 1801902, 1887035, 1915757, 1766104, 1915760, 1917969, 1648569, 1634877, 1634871, 1730663, 1801881, 1784929, 1887004, 1766098, 1887006, 1784920, 1841507, 1887029, 1765523, 1730655, 1887039, 1673210, 1648546, 1634857, 1673208, 1648555, 1673242, 1801886, 1730677, 1784930, 1798012, 1730630, 1730676, 1634858, 1648543, 1648562, 1673251, 1784924, 1887003, 1730646, 1917981, 1841503, 1730672, 1801883, 1841508, 1673252, 1730660, 1801894, 1648523, 1673244, 1673227, 1634860, 1648554, 1648521, 1730678, 1784927, 1766087, 1766085, 1634861, 1648559, 1648528, 1673226, 1730645, 1730679, 1730675, 1766105, 1874408, 1648556, 1648532, 1673239, 1765511, 1886996, 1874361, 1648552, 1673237, 1765508, 1673238, 1673248, 1673232, 1801925, 1801912, 1766100, 1648547, 1887008, 1648566, 1648533, 1673245, 1730659, 1648573, 1673247, 1841548, 1917965, 1841504, 1730636, 1766108, 1673217, 1673219, 1730649, 1730627, 1798009, 1915754, 1673246, 1673250, 1648545, 1634852, 1648551, 1673215, 1673220, 1730669, 1765509, 1634855, 1801884, 1648565, 1648522, 1841530, 1648525, 1673243, 1648572, 1730664, 1648531, 1634867, 1648548, 1730647, 1765515, 1765499, 1765504, 1730631, 1673229, 1766088, 1798001, 1801897, 1673225, 1798002, 1801890, 1765506, 1730641, 1841540, 1841519, 1841529, 1887005, 1887032, 1798008, 1673234, 1648549, 1634869, 1634865, 1648527, 1648575, 1648539, 1801926, 1801891, 1765516, 1765517, 1648568, 1841497, 1673231, 1801907, 1648574, 1730667, 1730671, 1841532, 1886999, 1634874, 1648544, 1798662, 1801904, 1801922, 1841525, 1766092, 1801896, 1874376, 1874386, 1874396, 1874401, 1730640, 1784912, 1784921, 1784934, 1801898, 1801906, 1801908, 1801909, 1801911, 1801913, 1801915, 1801924, 1766089, 1766099, 1887000, 1887001, 1887002, 1887019, 1887020, 1887024, 1887025, 1887027, 1730673, 1784914, 1784922, 1784933, 1634866, 1648530, 1648570, 1730628, 1801876, 1634859, 1730666, 1801919, 1797997, 1797998, 1766097, 1730651, 1648526, 1765514, 1648571, 1673205, 1801874, 1673203, 1765502, 1648524, 1648557, 1784906, 1730654, 1648563, 1801888, 1874400, 1648550, 1841524, 1673202, 1648534, 1730633, 1730632, 1841536, 1801905, 1730674, 1841502, 1841517, 1841526, 1801877, 1798003, 1784917, 1784913, 1784909, 1874381, 1730680, 1841511, 1100351, 1100318, 1365477, 1216139, 1430285, 1430211, 1358389, 1358409, 1535732, 1604465, 1430273, 1216180, 1216142, 1604458, 1459735, 1358384, 1604449, 1459729, 1430235, 1604454, 1482900, 1535744, 1290057, 1545576, 1482904, 1438091, 1594950, 1289976, 1552345, 1552352, 1604467, 1604448, 1604463, 1552347, 1604432, 2042063, 1784904, 2090017, 1784925, 1915745, 1915750, 1798010, 2042310, 1841505, 1841538, 1917955, 2042083, 1915764, 1915755, 1982541, 1887028, 1841516, 1952385, 1766083, 1673216, 1841535, 1917958, 1917957, 1784907, 1841528, 1952372, 1966086, 1966078, 1886995, 1980795, 1986239, 1841531, 2042241, 2042283, 2079430, 2079391, 2079390, 2079432, 1801903, 2079440, 1966079, 1766103, 2012182, 1915746, 2090020, 1874370, 1952368, 1874395, 2042309, 2090002, 2042060, 1766096, 1766095, 1784923, 1801917, 1730643, 2042258, 1980799, 2090006, 2042281, 2079416, 2089995, 2043325, 2042302, 2090011, 1917946, 1887031, 1986244, 1966082, 2012198, 2042072, 2079387, 2079396, 1966069, 1952367, 1966089, 1952363, 1887040, 1982529, 2042280, 1917949, 1801910, 1887010, 1986235, 1730652, 2012188, 2042237, 2090000, 2042270, 2079392, 2079413, 2079395, 1874357, 1766094, 1966081, 2042087, 2042313, 1801899, 1801895, 2079397, 1634856, 2079411, 1874385, 2042263, 2042259, 1798000, 1952386, 1952387, 1986243, 2090010, 2042311, 1887007, 1798014, 1801900, 1966077, 1841523, 2042244, 1982532, 1952352, 1797999, 2079427, 2042240, 2079431, 1887034, 1887021, 1766090, 2079419, 1801880, 1917975, 2042252, 2042085, 1952371, 2042285, 2090016, 2043313, 1730653, 2042257, 2042266, 2042086, 1986238, 1887043, 1730642, 1841533, 1841543, 1915767, 1952366, 1966084, 1874390, 2042238, 2042286, 2079408, 2079426, 2079436, 2042277, 1874369, 1798005, 1887009, 1673228, 1801885, 1915753, 2042228, 1980798, 2042233, 2090014, 2089999, 2042068, 1980804, 1917967, 2090005, 1915751, 1915748, 2079428, 2042074, 2012191, 2079407, 1917951, 1917970, 2042282, 1915749, 1915752, 1798006, 1980803, 2079439, 1730670, 1801892, 1915758, 2042253, 2042242, 1952373, 2079398, 2079425, 2042062, 1673218, 1784932, 1765520, 1887042, 1952388, 1915756, 1952380, 2042239, 2042316, 2012190, 2043320, 2042297, 2042075, 1766091, 2089990, 1982545, 1887012, 1982544, 1982546, 1874384, 1874362, 2042076, 1887014, 2079393, 2079421, 2089992, 1887018, 1980796, 1673213, 1765522, 1798011, 1952350, 1730668, 1980794, 2090008, 1730650, 1841541, 1917974, 1784903, 1801887, 2042230, 1952346, 1917948, 2042232.1980797, 1917960, 2042078, 1841542, 1841544, 1887037, 1952370, 1673222, 1966087, 1801901, 1952362, 1841495, 1917978, 1917979, 2012193, 1730644, 2042234, 1980790, 1730638, 2042301, 1887038, 1766084, 2042307, 2042287, 2042271, 1730658, 1874379, 2042294, 2042088, 2042296, 2089998, 2042265, 2079388, 2042254, 2042080, 2090027, 1874404, 1673230, 1634876, 1874359, 1874406, 1952369, 1982531, 1952379, 2079401, 2079417, 1980787, 2042066, 1952374, 1887017, 1673211, 2012199, 2079423, 1887011, 2090007, 1917959, 1841522, 1966070, 1673221, 1673212, 1917953, 2042236, 1966092, 2042274, 2042288, 1887023, 2079409, 2090004, 1874405, 1952381, 1952345, 1986236, 1673254, 2012197, 1966080, 1887026, 1952360, 1952376, 1766109, 1952382, 1841547, 1673253, 2042292, 2012184, 1982539, 2042303, 2079404, 1841496, 1841510, 1841537, 1874393, 1874402, 1917962, 1952358, 1980788, 1986241, 1986247, 1917956, 1915765, 1801889, 1917971, 2042300, 2042291, 1952389, 1966076, 1917968, 1952383, 1952384, 1952364, 1917983, 2012186, 2012200, 2012204, 1952353, 1952359, 2042077, 2043315, 2043321, 2043323, 2043327, 1917961, 1917973, 2042273, 2089989, 2079420, 2079422, 2089996, 2089997, 1952356, 1986229, 1986245, 2042293, 2042314, 2012187, 2042229, 2042231, 1841521, 1765503, 1766101, 1874391, 1730648, 1986234, 1874366, 1801882, 1648561, 1841514, 1874389, 2012189, 1634870, 1874375, 1648553, 1673236, 1730634, 1841512, 1980800, 2012192, 1952342, 1874358, 1730662, 1952361, 1765513, 1952378, 1784910, 1915762, 1766093, 1980791, 2042276, 2042081, 1648537, 1966091, 1874407, 2042306, 1966075, 1784931, 2079403, 1874388, 1874377, 1986231, 1986232, 1966083, 1874374, 1874367, 1874364, 1874368, 1874378, 1841518, 1874382, 1980785, 1980786, 1966074, 2042065, 2042279, 1966073, 1765501, 1841520, 1874383, 1874398, 2090023, 1730681, 1801923, 1887016, 1982538, 1874365, 1915763, 1980793, 1980801, 1648558, 1766102, 2012183, 1952349, 1986237, 1982537, 2090001, 2043317, 1841546, 1915768, 1980802, 1980789, 2012185, 2042284, 1766106, 1874394, 2104344, 2104353, 2104368, 2104350, 2149773, 2149738, 2149780, 2157599, 2149743, 2157611, 2168614, 2168585, 2157586, 2168757, 2104348, 2104369, 2149744, 2104365, 2149778, 2104347, 2149737, 2168606, 2104364, 2104346, 2149782, 2168613, 2168587, 2168753, 2149767, 2104373, 2104345, 2157581, 2104360, 2149757, 2120959, 2104356, 2104336, 2149756, 2149763, 2157602, 2149769, 2149758, 2168612, 2104337, 2149779, 2120967, 2104363, 2104370, 2157608, 2149764, 2104341, 2104354, 2104358, 2104361, 2157591, 2168609, 2168594, 2104357, 2157614, 2157607, 2149775, 2120966, 2157618, 2157595, 2120956, 2104334, 2120961, 2120963, 2120964, 2120960, 2120965, 2120968, 2120969, 2168761, 2149781, 2104351, 2104349, 2157588, 2104372, 2157613, 2194699, 2149741, 2149749, 1594943, 1476775, 1358422, 1430257, 1365483, 1476777, 1430274, 1459736, 1216197, 1358414, 1459730, 1430279, 1528591, 1438107, 1365455, 1290019, 1528601, 1528609, 1552332, 2042079, 2079400, 1917952, 1986233, 1784918, 2042250, 1886997, 2079434, 1886994, 1765498, 2042262, 2089993, 2090019, 1986240, 1874380, 2042067, 2012201, 2042261, 1801893, 2090025, 2042256, 1917954, 2042289, 1915761, 2043324, 1801878, 2042082, 1887022, 1952377, 1874409, 1634851, 2012181, 1874387, 2042315, 1915769, 1917966, 1966088, 2042245, 2090003, 1801920, 1730639, 1634868, 1784916, 2090022, 2012196, 2079429, 1730657, 2042071, 1673206, 1982534, 2012195, 1874392, 1634854, 2079410, 2090013, 1766086, 1887036, 2043318, 1874371, 1982533, 1841506, 2042272, 2042269, 2042267, 2079437, 1801879, 2043316, 2042061, 1886998, 2042264, 1952348, 1982542, 2090012, 1917947, 1917950, 2079414, 1801875, 1801921, 1952357, 2042059, 2042246, 1917964, 1841509, 2089991, 2043322, 2012180, 1982527, 2012202, 1917980, 2079435, 1952347, 2042299, 2090015, 2042084, 2042298, 2042069, 1982543, 1966085, 1765510, 1841539, 1841545, 1952390, 1982528, 1801918, 2042308, 2043326, 2012203, 2090026, 1798004, 1765505, 1982535, 1952351, 2079405, 2090018, 2079438, 1673209, 2012194, 2042295, 2090024, 1986242, 2079402, 1874403, 2042312, 2042073, 2079418, 2042317, 1966071, 1982530, 2079424, 1986228, 2042305, 1966090, 1952375, 1917984, 2079415, 1952344, 2194674, 2274489, 2327521, 2417960, 2418202, 2478389, 2157605, 2252529, 2327514, 2390671, 2362688, 2157610, 2194705, 2274528, 2362718, 2168599, 2149771, 2291459, 2252543, 2274521, 2194678, 2362716, 2418209, 2362692, 2417943, 2226915, 2291452, 2194696, 2244350, 2327534, 2327522, 2327527, 2157619, 2291469, 2362682, 2226929, 2244338, 2252501, 2291455, 2417962, 2362704, 2149736, 2229671, 2390644, 2291476, 2390648, 2417942, 2525022, 2149748, 2194713, 2149734, 2194689, 2226932, 2194722, 2226928, 2226921, 2252528, 2229692, 2274488, 2274506, 2244340, 2327495, 2157592, 2149759, 2362705, 2252507, 2362720, 2362714, 2418215, 2390663, 2418243, 2417985, 2194706, 2320090, 2471719, 2471700, 2417953, 2417945, 2417941, 2471693, 2471706, 2517167, 2537472, 2517171, 2244339, 2229693, 2168764, 2418204, 2252508, 2194695, 2168592, 2274514, 2274490, 2327493, 2332446, 2274503, 2517184, 2291447, 2149774, 2226927, 2149739, 2327530, 2417956, 2332424, 2194672, 2327496, 2252552, 2362695, 2362732, 2390673, 2291443, 2244343, 2194700, 2291475, 2320103, 2327517, 2417970, 2471720, 2517182, 2194681, 2274508, 2291466, 2332401, 2390668, 2390680, 2194680, 2244335, 2229683, 2362697, 2418201, 2362699, 2168598, 2168749, 2291453, 2332438, 2320094, 2274496, 2390654, 2390646, 2417968, 2157616, 2168763, 2418248, 2517165, 2478388, 2194709, 2252514, 2332410, 2362715, 2332445, 2362702, 2417949, 2471711, 2417984, 2417954, 2244349, 2291480, 2517172, 2537435, 2471690, 2362706, 2226922, 2157582, 2252550, 2252534, 2229673, 2274499, 2252536, 2332432, 2418233, 2517164, 2418211, 2517169, 2418255, 2390670, 2418232, 2327494, 2168611, 2525027, 2291456, 2252512, 2362713, 2471684, 2417944, 2157606, 2418254, 2274520, 2418249, 2157617, 2168597, 2229689, 2320093, 2332435, 2252526, 2332409, 2168595, 2362729, 2327533, 2332430, 2390690, 2390687, 2390650, 2390657, 2362708, 2418231, 2274532, 2390649, 2291479, 2149750, 2149740, 2157620, 2194715, 2229677, 2252545, 2252525, 2274507, 2149765, 2229685, 2274536, 2362719, 2390677, 2390639, 2104359, 2274497, 2417979, 2327532, 2168603, 2194716, 2168605, 2418218, 2327528, 2362685, 2417958, 2104338, 2194691, 2149733, 2168772, 2274501, 2194719, 2418245, 2390686, 2104342, 2194701, 2291470, 2320089, 2320097, 2327519, 2291441, 2362698, 2418237, 2327500, 2478386, 2418236, 2537434, 2417974, 2471689, 2471716, 2168752, 2168601, 2244344, 2274504, 2168589, 2332408, 2274502, 2418225, 2327523, 2194703, 2274518, 2471714, 2194682, 2168754, 2332416, 2194726, 2362700, 2320106, 2517174, 2252535, 2327529, 2291474, 2418234, 2327509, 2332418, 2120958, 2120957, 2194702, 2226931, 2252518, 2194712, 2149776, 2274517, 2418216, 2194708, 2332447, 2362725, 2194721, 2120962, 2229676, 2252523, 2390647, 2417987, 2478397, 2252506, 2244347, 2362728, 2229684, 2390675, 2362690, 2226935, 2537439, 2320108, 2320100, 2320088, 2194693, 2252521, 2417972, 2478400, 2417938, 2149752, 2194720, 2149751, 2104352, 2229682, 2226939, 2104339, 2252510, 2252537, 2149766, 2291436, 2194694, 2362721, 2274537, 2252544, 2226938, 2332414, 2417951, 2418205, 2417969, 2291461, 2417961, 2417957, 2327518, 2478384, 2471717, 2471721, 2252502, 2417973, 2471702, 2320099, 2417976, 2525043, 2517189, 2517188, 2537459, 2320098, 2471715, 2194686, 2291457, 2471703, 2104340, 2274509, 2229697, 2332398, 2229680, 2417955, 2226920, 2418210, 2274505, 2390666, 2537423, 2157609, 2194675, 2194679, 2226933, 2226916, 2252515, 2320104, 2149754, 2274500, 2194673, 2229672, 2104366, 2274523, 2168762, 2244341, 2194717, 2194711, 2362707, 2252551, 2390641, 2417966, 2478395, 2517163, 2537466, 2418240, 2471710, 2417963, 2252540, 2157601, 2320107, 2471692, 2229678, 2168608, 2229688, 2390665, 2149777, 2252527, 2168750, 2471688, 2252511, 2478380, 2229675, 2226930, 2252499, 2252500, 2194690, 2168765, 2157603, 2252546, 2291462, 2291445, 2320105, 2327511, 2327510, 2327515, 2390656, 2362709, 2390682, 2390683, 2362694, 2390655, 2417936, 2417946, 2418207, 2471730, 2537442, 2362689, 2362693, 2252505, 2274530, 2274531, 2291438, 2291448, 2291454, 2362687, 2362701, 2168586, 2168600, 2274492, 2274498, 2252517, 2252547, 2291463, 2291477, 2332403, 2332419, 2332421, 2332444, 2291435, 2291446, 2291465, 2390676, 2418247, 2390674, 2390678, 2390679, 2390688, 2390692, 2418220, 2418230, 2418251, 2390691, 2418208, 2418203, 2418212, 2418226, 2418241, 2471707, 2471708, 2471731, 2525031, 2525038, 2226918, 2244337, 2244346, 2274486, 2291440, 2168768, 2226936, 2229687, 2274524, 2327524, 2418206, 2517170, 2104367, 2149760, 2291482, 2157590, 2168588, 2418222, 2390685, 2418221, 2327507, 2320101, 2390658, 2157598, 2168593, 2149747, 2471701, 2418253, 2417971, 2157596, 2168590, 2149742, 2226934, 2149761, 2274534, 2194684, 2168607, 2418250, 2157583, 2327492, 2327505, 2194698, 2327516, 2274533, 2252504, 2327506, 2104335, 2194723, 2252520, 2332412, 2194725, 2194692, 2157587, 2194704, 2229694, 2291444, 2194714, 2291460, 2332428, 2332420, 2332425, 2332406, 2194724, 2327503, 2332434, 2332411, 2252519, 2332436, 2332427, 2252516, 2274510, 2332417, 2244351, 2244333, 2332437, 2332399, 2332402, 2104343, 2194677, 2157615, 2149745, 2226926, 2332407, 2157597, 2327498, 2291467, 2390653, 2471729, 2157584, 2291439, 2362717, 2390661, 2417986, 2274513, 2417983, 2390660, 2274485, 2149768, 2168767, 2274526, 2362727, 2471725, 2525053, 2560108, 2560112, 1459732, 1430253, 1604445, 1528593, 1482894, 1430286, 1552329, 1552351, 1365474, 2042268, 1917945, 2042064, 2079399, 1874399, 1797996, 1917977, 1982540, 2043314, 1765500, 2042290, 1634872, 1986230, 1952354, 1673235, 1952343, 1874372, 1784919, 2042070, 2090021, 1841513, 2042260, 2089994, 1874360, 1784926, 2042249, 1841534, 1874373, 1917963, 2390667, 2537469, 2471695, 2610124, 2417975, 2418228, 2327531, 2362683, 2610096, 2252539, 2362711, 2390672, 2537430, 2610151, 2537464, 2252532, 2537468, 2537454, 2478399, 2610058, 2610057, 2610155, 2610068, 2610055, 2610153, 2525026, 2417990, 2471691, 2525040, 2274515, 2537449, 2610054, 2610129, 2537427, 2252524, 2320096, 2471718, 2537448, 2610078, 2610148, 2478390, 2525039, 2168610, 2537462, 2418224, 2471705, 2194707, 2610083, 2517185, 2332441, 2537453, 2610086, 2610141, 2537450, 2610138, 2537470, 2327520, 2537441, 2610154, 2478383, 2229696, 2471697, 2610140, 2149755, 2168766, 2525046, 2471698, 2157604, 2417952, 2517187, 2525047, 2610064, 2362681, 2517180, 2471723, 2252553, 2525050, 2610158, 2390640, 2610075, 2610160, 2610067, 2471712, 2610126, 2517166, 2478381, 2252513, 2517175, 2537424, 2537452, 2525028, 2525041, 2252522, 2291468, 2362723, 2517160, 2229679, 2517161, 2478392, 2244334, 2525032, 2525030, 2478391, 2417959, 2478393, 2610136, 2390659, 2229690, 2525037, 2226923, 2390684, 2229695, 2104355, 2327526, 2252548, 2610070, 2168604, 2478385, 2149770, 2610157, 2418244, 2327525, 2610143, 2525025, 2229669, 2390681, 2252531, 2537425, 2418217, 2525034, 2610062, 2471713, 2390662, 2417940, 2610135, 2525023, 2291458, 2168755, 2327513, 2537440, 2417950, 2478396, 2525033, 2610087, 2327508, 2417980, 2471722, 2417947, 2610084, 2418229, 2525051, 2610128, 2517177, 2471709, 2418246, 2291483, 2610080, 2610074, 2252538, 2610134, 2610059, 2610131, 2471724, 2525029, 2332423, 2471686, 2610156, 2610056, 2274525, 2537463, 2332440, 2478387, 2537436, 2517181, 2537438, 2610077, 2478398, 2610146, 2194676, 2610065, 2537465, 2517159, 2332422, 2525049, 2229670, 2417988, 2417977, 2418242, 2610061, 2610079, 2537458, 2525024, 2610088, 2252541, 2226924, 2525048, 2418238, 2537444, 2390669, 2537431, 2517173, 2229681, 2417964, 2252533, 2418219, 2610149, 2610072, 2274493, 2610130, 2104362, 2327501, 2362703, 2610142, 2149772, 2525036, 2418214, 2610133, 2417978, 2417939, 2390651, 2274484, 2537461, 2537428, 2525054, 2610063, 2327502, 2252503, 2537445, 2194687, 2362712, 2471682, 2537456, 2471694, 2229668, 2537432, 2537422, 2327499, 2610115, 2610123, 2610120, 2610090, 2610093, 2610110, 2610117, 2610097, 2610109, 2610098, 2610119, 2610100, 2610122, 2252542, 2274491, 2332404, 2362726, 2390689, 2332426, 2418252, 2157594, 2168748, 2417948, 2610094, 2610092, 2517186, 2149753, 2194685, 2104371, 2517190, 2517178, 2610060, 2610111, 2418239, 2362724, 2274512, 2194718, 2149762, 2291478, 2327497, 2390643, 2291449, 2362710, 2471728, 2471727, 2537475, 2537433, 2149735, 2332413, 2332431, 2327512, 2320095, 2332415, 2332400, 2332433, 2332442, 2537467, 2194697, 2537447, 2610113, 2537457, 2332443, 2332439, 2332405, 2320092, 2274516, 2418213, 2362730, 2168760, 2610103, 2610108, 2274535, 2610114, 2610099, 2610091, 2274527, 2157589, 2823814, 2560115, 2574312, 2574306, 2647274, 2560110, 2715410, 2715352, 2635141, 2560133, 2647235, 2711700, 2574285, 2737678, 2711662, 2788548, 2659864, 2635154, 2791814, 2823797, 2823806, 2832163, 2559704, 2574191, 2564875, 2574214, 2574291, 2564880, 2564881, 2635151, 2635185, 2635165, 2647249, 2659858, 2659856, 2635145, 2647240, 2659862, 2647277, 2647271, 2711686, 2715336, 2715327, 2715308, 2659834, 2737663, 2715329, 2763734, 2788585, 2763731, 2788589, 2715469, 2788590, 2832159, 2788558, 2823821, 2832139, 2791831, 2870066, 2823828, 2870030, 2870072, 2791837, 2659867, 2574265, 2564879, 2659832, 2715452, 2715455, 2806805, 2564882, 2574260, 2763720, 2823826, 2564853, 2711709, 2715324, 2715316, 2806799, 2711685, 2874011, 2737676, 2635168, 2574276, 2564843, 2564852, 2715453, 2763721, 2559723, 2647223, 2635162, 2737683, 2711696, 2832126, 2806793, 2647228, 2635164, 2647239, 2635159, 2715445, 2737675, 2574250, 2737645, 2715310, 2647265, 2832138, 2788553, 2564884, 2715326, 2635176, 2564849, 2659841, 2564851, 2715463, 2659843, 2788565, 2715347, 2574269, 2788581, 2788575, 2870047, 2715458, 2711706, 2791835, 2574193, 2737634, 2711682, 2647238, 2635146, 2647268, 2635153, 2737635, 2788547, 2806818, 2715428, 2763686, 2574287, 2647236, 2715414, 2559720, 2564844, 2635188, 2791834, 2659851, 2763706, 2737669, 2560124, 2806815, 2791836, 2832145, 2823819, 2832165, 2564854, 2574254, 2560128, 2659835, 2659842, 2647267, 2635172, 2647272, 2737654, 2647220, 2715344, 2806816, 2806836, 2823785, 2806819, 2737648, 2647262, 2659863, 2647242, 2564861, 2560127, 2635149, 2564850, 2763690, 2870046, 2559701, 2635147, 2564895, 2647255, 2715429, 2870064, 2737679, 2788588, 2635158, 2763710, 2823811, 2832157, 2659833, 2806801, 2763735, 2647254, 2560122, 2737659, 2564873, 2659840, 2659865, 2737667, 2564866, 2763688, 2806832, 2806830, 2806817, 2832169, 2832125, 2564886, 2564888, 2560123, 2763711, 2564860, 2564859, 2564887, 2659829, 2832123, 2564845, 2564870, 2635180, 2574208, 2715454, 2564858, 2737671, 2823790, 2763712, 2763715, 2788583, 2806795, 2806814, 2564892, 2715430, 2647222, 2870043, 2715411, 2874012, 2870053, 2737677, 2574200, 2574215, 2564891, 2647234, 2737668, 2564874, 2574271, 2715338, 2574272, 2559706, 2659847, 2659839, 2574264, 2711708, 2788566, 2791821, 2823791, 2564894, 2564889, 2559700, 2737657, 2647232, 2559702, 2715457, 2659855, 2715447, 2832160, 2763691, 2574281, 2574251, 2564872, 2574256, 2564868, 2715339, 2788556, 2559709, 2791809, 2823803, 2560132, 2715332, 2806791, 2715436, 2806840, 2823793, 2832135, 2806824, 2791829, 2870032, 2870059, 2737650, 2832143, 2791844, 2560121, 2647224, 2737660, 2711690, 2715465, 2635187, 2647247, 2788554, 2788555, 2806812, 2715419, 2574202, 2737647, 2737661, 2737644, 2791843, 2564863, 2715340, 2574249, 2711711, 2763722, 2806827, 2823802, 2823799, 2715467, 2788552, 2635186, 2715351, 2647279, 2870068, 2647261, 2560125, 2715323, 2715451, 2832168, 2832141, 2635175, 2806825, 2574267, 2559705, 2659868, 2564871, 2823825, 2635160, 2806839, 2560111, 2635167, 2560131, 2635142, 2564883, 2635156, 2564876, 2659854, 2659823, 2647270, 2635143, 2647269, 2737673, 2715425, 2711687, 2737649, 2715325, 2763687, 2823804, 2737674, 2788577, 2832161, 2737636, 2832128, 2823798, 2763692, 2823801, 2788584, 2832171, 2870071, 2832153, 2763726, 2560120, 2832162, 2715433, 2791813, 2647231, 2788579, 2832144, 2823823, 2715426, 2559712, 2788572, 2635182, 2870033, 2559710, 2560129, 2559721, 2711688, 2806835, 2560113, 2715439, 2564885, 2763725, 2659853, 2647244, 2737646, 2574248, 2560118, 2635173, 2737643, 2574255, 2560130, 2559717, 2559714, 2659849, 2647259, 2715319, 2574207, 2715328, 2659857, 2763694, 2806808, 2574212, 2574268, 2635155, 2711689, 2711694, 2870051, 2559719, 2559718, 2806813, 2788580, 2635166, 2711697, 2737656, 2737655, 2647258, 2574263, 2715471, 2715460, 2832132, 2635152, 2715315, 2647266, 2659859, 2823815, 2715432, 2715441, 2715348, 2763704, 2832136, 2715349, 2560105, 2560106, 2560109, 2560114, 2560117, 2560104, 2560107, 2574307, 2574322, 2635144, 2659824, 2711672, 2715412, 2715417, 2788545, 2788549, 2788587, 2788564, 2788569, 2715353, 2832127, 2574301, 2715416, 2715418, 2791820, 2791827, 2635150, 2635178, 2635189, 2647226, 2647257, 2647273, 2659836, 2574196, 2574197, 2574201, 2574206, 2574284, 2635157, 2635161, 2635171, 2647278, 2574192, 2647243, 2659852, 2711670, 2711692, 2711698, 2711699, 2715420, 2574198, 2574278, 2574305, 2574326, 2574317, 2659860, 2715346, 2647253, 2574318, 2574319, 2715342, 2873979, 2711668, 2574323, 2574295, 2635170, 2647227, 2659870, 2711701, 2832149, 2574298, 2574274, 2574290, 2574309, 2659830, 2659831, 2763700, 2763705, 2763717, 2574282, 2574289, 2574302, 2574316, 2659826, 2659846, 2763689, 2763723, 2635169, 2574313, 2574297, 2574311, 2635190, 2574253, 2574210, 2659828, 2574315, 2560126, 2711703, 2647245, 2574304, 2574286, 2574195, 2574283, 2574189, 2574288, 2647246, 2574280, 2823816, 2560116, 2659844, 2711676, 2659871, 2715422, 2737639, 2574293, 2574325, 2574321, 2574275, 2711705, 1535742, 1528602, 2042278, 1966072, 1874410, 1874363, 1917976, 1784911, 2079433, 1952365, 1915747, 1648538, 1673233, 1982536, 1765507, 1952355, 2042251, 2043319, 2079406, 2042255, 2079386, 2089988, 2418227, 2610118, 2610116, 2157600, 2478394, 2291450, 2244352, 2226925, 2417989, 2517183, 2537426, 2610137, 2168756, 2418235, 2244331, 2610069, 2417937, 2274487, 2537437, 2610066, 2362696, 2226917, 2471699, 2610125, 2471726, 2478401, 2610081, 2291442, 2418223, 2610053, 2168602, 2537474, 2610076, 2194710, 2537451, 2244348, 2610073, 2320102, 2610071, 2537460, 2390664, 2537473, 2471704, 2362731, 2525042, 2610139, 2610159, 2417967, 2244336, 2610132, 2525052, 2537443, 2610145, 2226937, 2610085, 2610127, 2525044, 2157585, 2417982, 2327504, 2244342, 2537455, 2252530, 2274495, 2149746, 2291481, 2537429, 2537446, 2291437, 2362686, 2537471, 2610107, 2610095, 2320091, 2291434, 2291473, 2610102, 2244345, 2229691, 2274494, 2417981, 2574308, 2635177, 2763724, 3589337, 2897530, 2763716, 2937966, 2659845, 2806792, 2574194, 3724473, 2791822, 2915232, 3681033, 2574328, 2715421, 2897539, 2791856, 2715449, 2635148, 3589349, 3724505, 2915225, 3661964, 2870037, 2763729, 2931677, 2931672, 3553809, 2915196, 3553819, 3665015, 3661944, 3687251, 3665058, 2870060, 2870040, 2897535, 2915203, 2715437, 2915214, 2915222, 2715468, 2921051, 2921069, 2966189, 2966194, 2966221, 2931663, 2966208, 2921045, 3553803, 3553800, 2574261, 2647256, 3553812, 3589315, 3589292, 3606809, 3606770, 3589343, 3589273, 3589262, 3606792, 3553795, 3661933, 3681046, 3665014, 2788546, 3665054, 3724464, 3661970, 3606767, 3724459, 3724462, 3606781, 3724494, 3724479, 3687257, 3687236, 3740748, 3740785, 3661897, 3740740, 3661895, 3661901, 3775094, 3775114, 3790758, 3775132, 2832134, 2966220, 2915208, 2938004, 3589316, 3553813, 2715466, 3687266, 2931655, 2897528, 2763732, 2806838, 3665044, 3665031, 3665026, 3661945, 2870061, 2791817, 2931666, 2937960, 2921037, 3606812, 3665052, 3589299, 3724466, 3687265, 3589342, 2931687, 3681028, 2715333, 2897514, 2915195, 2915233, 2979656, 2937974, 2931667, 2870034, 3553848, 3553843, 3589285, 3589255, 3606815, 3606830, 3661938, 3553799, 3705753, 3661942, 2966195, 3553860, 2966205, 3606825, 3661947, 2832166, 2870038, 3589264, 2870027, 3553783, 2715312, 3589252, 2870036, 3681044, 2931697, 3705756, 3754773, 3740747, 2564877, 2763707, 3553778, 3606814, 3606771, 3665011, 3687231, 3754754, 2874014, 2937957, 2897523, 2711707, 3661949, 3606772, 3740752, 3606822, 3681043, 2832151, 3589241, 2931685, 3681039, 2647233, 2647225, 3606816, 3661950, 2870041, 2711684, 3553781, 3553772, 2659827, 2931680, 3553796, 2931654, 2832129, 2711671, 2915218, 2931657, 2938003, 2966222, 2937995, 2931674, 3606806, 3553807, 3665018, 3724460, 3553797, 2966228, 3606786, 2915189, 3681006, 2931676, 3606761, 2564856, 2806809, 3553830, 2931698, 2937971, 2715424, 2897504, 2897501, 2897492, 2870073, 2915194, 2921050, 2921036, 2921042, 2966210, 2931671, 3553827, 3553771, 3589256, 3589278, 2715335, 3589341, 3589325, 3589321, 3589313, 3665028, 3705767, 3589307, 3740753, 3754780, 3754770, 2791846, 2788563, 2763708, 2715456, 3589254, 2870048, 2711663, 2937981, 3665051, 2791832, 2921058, 2931691, 2711677, 3661869, 3724471, 3775110, 2915229, 2921024, 2938006, 2921033, 2966188, 2937985, 3589286, 3553861, 3589324, 2966223, 3589335, 3553821, 2737662, 3681029, 2921057, 3705755, 2659838, 2715337, 2873999, 2931656, 2823786, 2806811, 2921034, 2979651, 3606807, 2921025, 3606754, 3589322, 3606789, 3665048, 3681021, 3589253, 3606773, 3606821, 3665034, 3705768, 3705776, 3705765, 3724469, 3705751, 2659850, 3687254, 3754764, 3705791, 2870052, 2559699, 2559703, 2559722, 2921031, 2921028, 3553855, 2979652, 2564865, 2921029, 3589277, 2647229, 3705792, 3589281, 3681045, 3589344, 2897507, 2870044, 2737637, 3589284, 2931693, 3661925, 2715415, 2874007, 2897497, 2832152, 2897534, 2921044, 2574259, 2921053, 2791848, 2921061, 2763733, 2915219, 2921046, 2921039, 2806823, 2806804, 2937970, 2937982, 2937959, 2966198, 2937958, 2966212, 2979655, 2979661, 2931658, 2915188, 3553817, 3553790, 3553780, 2897513, 3606817, 3589259, 3606820, 3589334, 3606780, 2931694, 3606764, 3553853, 3606774, 3589333, 3606829, 3589310, 3681013, 3665049, 3606748, 3661900, 3665021, 3740761, 3661892, 3665023, 3754776, 3553849, 3754772, 3775106, 3687245, 2791826, 2715435, 3553815, 3681018, 2870069, 3687228, 3724501, 2559716, 3589293, 3687229, 3553841, 3775112, 2788591, 3589320, 2937969, 3740769, 2806803, 3606795, 2763728, 2915224, 2921027, 2711680, 2979665, 3775118, 2832172, 3589247, 2806822, 2832148, 3665043, 3754752, 3553801, 2715423, 3705748, 2931670, 2564869, 3606827, 2931673, 2711664, 2659825, 3606785, 2897498, 2574257, 2897524, 2564846, 2870026, 2966184, 3553773, 2931686, 3589263, 3606788, 3589258, 3606747, 2788560, 3665020, 3606755, 2711678, 3724485, 2737641, 2788550, 2915217, 2874008, 2823805, 3553852, 3589260, 3589280, 2870025, 2737681, 2791847, 2832158, 2870065, 2788557, 2966183, 2832154, 2832124, 2966191, 2870039, 2966215, 2931695, 2966181, 2870058, 2788573, 2921065, 2979654, 3553838, 2966179, 2937979, 3553844, 3553856, 3589250, 3589257, 3606791, 3606810, 3606793, 3553770, 3606798, 3589326, 3589300, 3606768, 3553834, 3589268, 3661968, 3553779, 3553788, 3661930, 3589330, 3661951, 3589283, 3705775, 3705771, 3724481, 3661890, 3775121, 3661956, 3661963, 3775093, 3553806, 3775128, 3775130, 3661926, 3661907, 3661905, 3661908, 2823808, 2763718, 2788571, 3754785, 2870057, 2921040, 2823800, 2931678, 2937961, 2966217, 2763727, 2559713, 3724476, 3740776, 2791830, 2915236, 2921070, 2931699, 3606799, 2832147, 2564864, 2711666, 3589331, 2560119, 3687270, 3661885, 3681000, 2870070, 2915209, 2870045, 2715331, 2966227, 2966185, 2979644, 3705754, 2921064, 3606749, 2931700, 3665057, 3724492, 3754746, 2564857, 3589336, 2559707, 2870050, 2832133, 2737638, 2564878, 2647252, 2806821, 2915226, 2966219, 3606777, 3589301, 2737658, 2966201, 3661957, 3665045, 3665035, 3681025, 3724497, 3705795, 3661896, 2806806, 2559711, 2564847, 2806794, 3589276, 3606804, 2763696, 2870029, 2931669, 3553802, 2937968, 3606794, 3665012, 3589279, 3705774, 3740760, 2715446, 2979647, 2921067, 2979645, 3606783, 3606808, 2921049, 2937963, 3606819, 3589306, 3606746, 3606796, 2870042, 2938000, 3681030, 2647248, 3606826, 2791849, 3589302, 3589339, 3681008, 2788578, 2874005, 2937977, 2979653, 2938002, 3724503, 2564893, 2921047, 3589303, 3553825, 2574262, 2647276, 2897529, 2931664, 2915227, 2647251, 2659848, 2574266, 3606766, 3606757, 2979662, 2788551, 2966213, 3589275, 3589304, 3705797, 3589291, 3740767, 2715443, 2647237, 2832137, 2832142, 3705762, 3705784, 2979641, 2806820, 2915205, 2915191, 2937976, 3553791, 2823788, 2921060, 3687241, 3665009, 3754743, 3681005, 2763730, 3705796, 2737680, 3661935, 2897508, 2574270, 2921038, 3661955, 2823822, 2832164, 2832130, 2788574, 3705793, 2763695, 3661953, 3790751, 2966209, 2715309, 2791838, 2931692, 3681004, 2791816, 3553857, 2897526, 2915228, 2931661, 3754756, 3754755, 2874013, 2715462, 2737652, 2791858, 3661924, 3589295, 2737640, 2873996, 2897500, 2915204, 2897517, 2915202, 2921035, 2921030, 2897525, 2931689, 2966207, 2897495, 2737642, 2937973, 2931681, 3553787, 2870063, 3606787, 2763701, 3606803, 2938005, 3606769, 3589314, 3589249, 2874003, 3589244, 2763699, 3665029, 2915213, 3661937, 3589319, 2806807, 2874006, 3705770, 2979659, 3724491, 3724474, 3560181, 3665037, 3754766, 3665053, 3740738, 3661940, 3589245, 3553832, 3661906, 3553798, 3661912, 3775115, 3661965, 3775103, 3775127, 3661889, 3775090, 3775085, 3775096, 3687230, 2870062, 2737664, 3553831, 3606775, 2788562, 3661958, 2921071, 2574258, 2715321, 2870056, 3553774, 2915223, 2806833, 2870035, 2874000, 2832156, 2737672, 3687259, 3665056, 2711691, 3606778, 2832150, 3606800, 3665022, 2737653, 3589288, 2897520, 2788582, 2921062, 2659869, 3553836, 2737666, 3589311, 3560180, 2791819, 2806837, 2832167, 2897494, 2937993, 2897516, 2921048, 2966200, 2979646, 3589272, 2715408, 2823795, 3705783, 3553840, 2931683, 2870031, 2823824, 2763693, 3589298, 2564848, 3606831, 2647275, 3661948, 2574216, 3754740, 2873994, 3606776, 2966226, 2564862, 2915197, 3724498, 2715311, 2635181, 2870074, 2874015, 2763709, 2966206, 2791839, 2931675, 2806826, 2937987, 2921063, 3687256, 2574211, 2921068, 3553868, 2823789, 2715322, 3553782, 3553816, 2931688, 2791815, 2870067, 2979650, 2937999, 2921056, 2921041, 2873978, 2711695, 2870055, 2921055, 3589243, 3606753, 3661867, 3687268, 3724486, 3665013, 3589340, 3589323, 2715434, 2921023, 3553828, 3553793, 2715461, 2823796, 2897503, 2791854, 2791842, 2915198, 2915210, 2915212, 2915221, 2915230, 2915235, 2937997, 2966182, 3553794, 3553824, 3553850, 3553865, 3553835, 3553866, 3724482, 3724508, 3687224, 3687233, 3687235, 3687238, 3740786, 3740778, 2873967, 2873968, 2874002, 2574204, 2574320, 2873974, 2873990, 2873991, 2873992, 2873973, 2873981, 2873971, 2873970, 2873972, 2873980, 2873975, 2873977, 2874001, 2715448, 2715345, 2715356, 3606751, 2715314, 2715318, 2715357, 2874009, 3661879, 2873976, 3661931, 2979667, 2574205, 2711702, 2574252, 2791851, 2937962, 2915215, 2711674, 2832140, 2966202, 2937991, 3553823, 3681019, 2806797, 2823820, 2823830, 2823831, 2897493, 2897506, 3553864, 2791855, 2791812, 2937983, 2966186, 2979664, 3589287, 3705773, 3705781, 3740765, 3606765, 3740772, 2574273, 2791818, 2788568, 2915201, 2937986, 3606759, 2647221, 2564867, 2715409, 2832146, 2574314, 2897531, 2711681, 2737670, 2715341, 2823832, 2791825, 2823834, 2791810, 2915199, 2979669, 2791850, 2979649, 2574199, 2574279, 2979648, 2915200, 2897515, 2897540, 2574188, 2715334, 2788544, 2711704, 2897502, 2897499, 3553845, 2897510, 2897537, 2574303, 3553859, 2897527, 2574203, 3681007, 2711673, 3589265, 2937992, 2711710, 2715413, 2823817, 2715313, 2715330, 2763702, 2791840, 3589246, 2574209, 2791823, 2897519, 2937998, 2715355, 2791841, 2897521, 2897505, 2791852, 2937964, 2938001, 2715442, 2791857, 2806796, 2791811, 2931662, 2791824, 3553785, 3681014, 3681032, 3724500, 3724504, 3740773, 2715450, 2574296, 2823833, 2791853, 3681010, 3665042, 3705761, 3724507, 3553784, 3589282, 3553810, 3754750, 2574327, 2042248, 1784928, 1841500, 1887030, 2610104, 2471696, 2610106, 2291472, 2168751, 2537476, 2274529, 2390645, 2517179, 2194683, 2610152, 2274522, 2471683, 2168769, 2517176, 2274511, 2390652, 2168596, 2525035, 2362684, 2274519, 2362691, 2168770, 2157593, 2229674, 2610101, 2471685, 2168758, 2332429, 2291471, 2291451, 2291464, 3754741, 3790792, 3740784, 2915216, 3589248, 3740764, 3681022, 3687246, 2823812, 3687263, 2897533, 2915192, 3681031, 3790769, 3740750, 3724477, 3687272, 3740781, 2915211, 3754742, 3687264, 3790766, 3754763, 3754739, 3681003, 3790793, 2715438, 2635183, 2574294, 2931668, 2659822, 3775084, 3790759, 2873982, 2788576, 2788559, 3790796, 2635174, 3705750, 3681011, 2931679, 3754779, 2937989, 3754759, 3665039, 3665025, 2931696, 3681040, 2873969, 2806810, 3790768, 3775097, 3606762, 3687271, 3661876, 2832170, 3790781, 3589269, 2874004, 2897536, 3606784, 3740744, 3661870, 2806800, 3740758, 3661904, 3775104, 3665016, 3754737, 3681017, 3724463, 3665019, 3705782, 2806802, 2966193, 3790783, 2966204, 2788586, 3661903, 3754744, 3754786, 3724468, 3606828, 3687248, 3661911, 3680998, 3589327, 3606763, 3687262, 2915206, 3754761, 3553822, 2979642, 3606801, 3754760, 3589261, 2806829, 2763703, 3661891, 3740780, 3724470, 2870054, 3661884, 3665032, 3681020, 3681027, 3661865, 3687253, 3606779, 3661962, 3665046, 3687242, 3790757, 3665017, 3740742, 3754775, 3740746, 3705764, 2874010, 3665027, 3705794, 3665047, 3661927, 3661969, 3740762, 2966197, 3705785, 2873989, 3790791, 3661899, 2966192, 3553818, 3661887, 2931684, 3606758, 3790797, 3740754, 2715427, 2715470, 3681001, 3661914, 3790780, 3606782, 2711679, 3790767, 3790778, 3606752, 3687255, 3754748, 2921054, 3775123, 3665024, 3589318, 3553789, 3661868, 3661888, 3705749, 3661966, 2915231, 3661872, 2966214, 3665050, 3705777, 3661909, 3661893, 3661873, 2788561, 3661913, 3775105, 3606823, 2737682, 3705779, 2921043, 3665038, 3775111, 2832155, 2966218, 3681002, 3687239, 2559708, 3724487, 3665041, 3775098, 2873993, 3687244, 3687237, 3775091, 3553776, 2937967, 3661923, 2635163, 3687247, 3790753, 3740756, 3661939, 2647230, 3740755, 3553792, 3589274, 3775131, 3661894, 3589266, 3775117, 3754762, 3790794, 3740766, 3754771, 3661961, 3589242, 3606811, 2711665, 3661886, 3775119, 3775124, 3754749, 3705789, 2897491, 3790788, 3754783, 3775122, 3553826, 3724461, 2715459, 3589309, 2715317, 3661932, 3606824, 2659861, 3754753, 2659837, 3775109, 3790798, 2874016, 3775120, 2737651, 3665033, 3665030, 3687240, 3606818, 3740739, 3740745, 3790771, 3754774, 3681042, 3775092, 3754758, 3553863, 3687225, 2635184, 3681015, 3790786, 3589289, 3790784, 3681026, 3775108, 3661902, 3661943, 3754747, 2873983, 2873984, 3790755, 3790775, 3754738, 3687249, 3661929, 3661922, 3740783, 3687260, 2715464, 3687252, 3665036, 3790787, 3775099, 3589297, 2931690, 2931682, 2823809, 2979660, 3681047, 3775087, 3681038, 2915234, 3775095, 3775100, 3661952, 3661960, 3790790, 3740741, 3775089, 3661898, 3661882, 3754782, 2763719, 3661954, 3740782, 3754751, 3661959, 3724467, 2966180, 3665040, 3687227, 2823827, 3687232, 3775083, 3553808, 2921026, 3705757, 3589271, 3681012, 3724495, 2873997, 2937972, 2915190, 2897538, 2559715, 3665055, 3740777, 3790752, 3661967, 3724489, 2873987, 2873985, 2737665, 3790779, 3705788, 2873986, 3705786, 3705790, 2937988, 3724480, 2635179, 3775107, 3705787, 3754781, 3790760, 2937965, 3754765, 3724490, 3790754, 3790763, 3790782, 3790799, 3740763, 3740770, 3740775, 3775126, 3754778, 2823794, 2897496, 3754768, 2966199, 2823829, 3606760, 3606797, 3775125, 3775113, 2806798, 3553862, 2870028, 2897509, 2931701, 3740737, 2979668, 3790773, 2574277, 3740779, 3724483, 3790777, 3790774, 2979666, 3553858, 3790761, 3553811, 2897518, 2823813, 3775088, 3754767, 3724493, 2715343, 3589338, 2715440, 3589348, 3553851, 2788542, 2931665, 3681037, 3724499, 3790789, 3681034, 2937980, 3553846, 2897512, 3681016, 2711667, 3589345, 2574300, 2711669, 2788543, 3681035, 3790762, 3553847, 2931702, 3754777, 3775129, 3724496, 2574292, 3740768, 3740759, 3740751, 3553775, 3740757, 3941574, 3941611, 3972923, 4028292, 3941599, 3814926, 3941615, 3941604, 3839592, 3925401, 3843194, 4008410, 4008400, 4054867, 3839512, 3925458, 3925119, 3972960, 4008407, 4073929, 4077262, 4073919, 3814933, 3814922, 3817236, 3817259, 3837327, 3843175, 3843191, 3817235, 3817281, 3839543, 3817279, 3839544, 3839562, 3839546, 3892084, 3892081, 3843185, 3892073, 3925395, 3917019, 3941578, 3925407, 3972939, 3925441, 3916999, 3974947, 3925063, 3980791, 3974937, 3974926, 3972945, 3985537, 3925438, 4039561, 3925454, 4068229, 4028296, 4087236, 4077215, 4077226, 4077214, 4077231, 3980789, 3974943, 4108427, 4068207, 4083692, 4083705, 4087224, 4151133, 4157661, 4157639, 4073926, 4157647, 4087222, 4157681, 4167240, 3892067, 3839602, 3837334, 3925425, 3980758, 4051324, 3941621, 4108421, 3843172, 3814959, 3839556, 3925412, 3814929, 3837323, 3843190, 3839577, 3839534, 3892050, 3814948, 3917018, 3925115, 3980744, 3980710, 3980730, 3985536, 4068203, 3814953, 3839561, 3916997, 3917014, 3925131, 3839510, 4008392, 3817273, 3892082, 3843202, 3941628, 3925442, 3925406, 3980704, 4008412, 3925443, 4054872, 3917009, 4083689, 4083700, 4108447, 4051325, 3837315, 3843177, 3980748, 3814955, 3839598, 3814925, 3925109, 3980717, 3814939, 3814958, 3839569, 3917001, 3941598, 3839539, 3980705, 4028298, 4068216, 4068202, 3925455, 3814927, 3837326, 3837335, 3843187, 3839551, 3892063, 3843204, 3925104, 3972952, 4039555, 3925108, 4008406, 3814957, 3817260, 3941596, 3817240, 4054879, 4008415, 4008409, 4008402, 4054869, 4073930, 3814934, 4008383, 4151101, 4073924, 4008428, 3839603, 4028300, 4083679, 4167235, 3839547, 3839515, 3974938, 4028294, 4028303, 4157673, 4008404, 3980722, 3980753, 3980718, 3925057, 4008413, 4087242, 3925106, 3980778, 4008417, 3817283, 3917023, 3814930, 3837333, 3814952, 3843207, 4083683, 3817270, 3817250, 3839530, 3843208, 3839563, 3839542, 3917002, 3925403, 3980752, 4008385, 3972957, 3925421, 4073918, 3839597, 4073905, 3814964, 3814951, 3843179, 3837318, 3843182, 3839523, 3839588, 3839585, 3837330, 3892087, 3925436, 3925398, 3941581, 3972961, 3980708, 3972920, 3892060, 3985540, 4028297, 4008429, 3972942, 4068188, 4008381, 4087217, 4087194, 3917000, 4108441, 4108429, 4151123, 4087226, 4087214, 4157672, 4174122, 3916996, 3837320, 3839560, 4008414, 3814943, 4051317, 4051299, 3972958, 3843183, 3925427, 3925457, 3985534, 3941609, 3843196, 3925392, 4087193, 4051311, 4054866, 3892080, 3839526, 3925134, 3925415, 4068223, 3843199, 3839519, 3917005, 3814966, 4054881, 3917022, 3925420, 3925113, 3839595, 4039560, 3892086, 3843178, 3839594, 3839545, 3814923, 3843205, 3817262, 3892069, 3839548, 3980732, 3925402, 3980742, 3925099, 3925064, 3941597, 4054883, 4054884, 3980706, 3941618, 4083704, 3892083, 3817275, 4151135, 3839507, 3843203, 4008398, 3980703, 3892049, 4151130, 4039552, 3837321, 3941576, 3843189, 3917011, 3925430, 3925410, 3839533, 3972962, 3972954, 3980763, 3941606, 4068184, 4083694, 4151117, 4054874, 3814936, 3839559, 4008388, 3980733, 3837328, 4087228, 3843210, 3892055, 3814938, 3925405, 4068225, 3837311, 3817247, 3972935, 3980751, 3941610, 4008389, 4054868, 4039563, 4039553, 3980750, 4087198, 4151093, 4151097, 4068194, 4157676, 4178078, 3980756, 3892071, 3980743, 3839531, 3817245, 3839520, 3837322, 3843211, 3839596, 3843165, 3892018, 3892058, 3843212, 3843184, 3925422, 3925429, 3925447, 3980776, 4008390, 3925448, 3985539, 3985541, 3814946, 3892009, 4039543, 4039545, 4077220, 4083699, 4083678, 4087207, 4039546, 4054882, 4087213, 4077238, 4087208, 3941625, 3980725, 3925404, 4157654, 3917010, 3839573, 3839555, 3839600, 3843169, 3925396, 3925118, 4008396, 3925105, 3839525, 3925116, 3817272, 4083696, 3839538, 3839521, 3892059, 3980713, 3925431, 4087235, 3814971, 3925418, 4068200, 3972922, 3843188, 3892062, 3843171, 3917008, 3974951, 3980749, 4008387, 3941601, 4054871, 4157645, 4077225, 3839582, 3980775, 3843168, 3839571, 3980765, 3980740, 4087230, 4157664, 4008427, 3843167, 3972947, 3925440, 3916995, 4077213, 4108435, 4151087, 3814969, 3892051, 3817264, 4054880, 4077240, 4083698, 3839570, 3839549, 4157656, 3814941, 3985535, 3972924, 3974948, 3817277, 3839576, 3892005, 3837312, 3925411, 3980727, 3892006, 4077224, 3837325, 3925434, 4054877, 4039558, 4054875, 3980762, 4083690, 3814935, 3839604, 3972941, 3972930, 3980746, 3972951, 4054886, 4073916, 3817248.3916998, 3839593, 4073906, 3839564, 4108439, 4167241, 4051308, 3839572, 3925121, 3814961, 3839580, 4073913, 4108445, 3839527, 3892014, 3972928, 4073915, 4054870, 4083702, 4087218, 4008411, 3814962, 3839514, 3814963, 3839529, 3839579, 3843201, 3817255, 3843186, 3839522, 3839506, 3839528, 3892048, 3925394, 3925107, 3892064, 3917006, 3925135, 3892042, 3925449, 3925103, 3980712, 3980747, 3843206, 3925399, 3980716, 3925450, 3925397, 3925393, 4028316, 4028299, 3925419, 3941592, 3980738, 3925069, 4008397, 4039565, 3980719, 4054865, 4054887, 4073932, 4077222, 4039548, 4068193, 4039562, 4087202, 4108423, 4108433, 4051320, 4167233, 4151110, 3941591, 4167221, 3980723, 3814945, 3817241, 4028319, 4077254, 4028315, 4083682, 3925426, 4008422, 3817246, 3980728, 4073927, 3925130, 3892010, 3925122, 3892017, 3925096, 3980745, 3972925, 3980714, 3980735, 3980757, 4083684, 4083695, 4157652, 3817263, 3837319, 3839558, 3917015, 3972914, 3925451, 3980760, 3972936, 3972931, 4157644, 3817265, 3892088, 3839540, 3941627, 3892047, 3892046, 4008420, 4054863, 3839511, 3925123, 3817242, 3941612, 3980786, 3974927, 3839552, 3925423, 4077219, 3972938, 3837316, 3817243, 3941577, 3839591, 4077237, 3839537, 4008423, 3817251, 3843174, 3843170, 3925127, 3925424, 3985538, 4039551, 4051313, 4108436, 4039556, 3925111, 3837324, 4151134, 3817249, 3817256, 3837314, 3837329, 3837331, 3839508, 3839557, 3839565, 3917020, 3917021, 3925039, 3925042, 3925047, 3925061, 3925079, 3925084, 3925087, 3925098, 3925117, 3925132, 3941605, 3941613, 3949387, 3925040, 3925092, 3972944, 3980739, 3980755, 3980759, 3980782, 3980788, 3980798, 3980711, 4008382, 4028304, 4028305, 4028310, 4028311, 4028320, 4151124, 4151092, 4151129, 3925102, 3925124, 3925048, 3925052, 3925062, 3925065, 3925059, 3925068, 3925071, 3925082, 3925417, 3925125, 3972963, 3974929, 3974930, 3974931, 3974933, 3974940, 3974941, 3974944, 3974946, 3925101, 3941584, 3941608, 3974932, 3974936, 3974925, 3925055, 3925041, 3925050, 3941607, 3814947, 3814967, 3814942, 3892053, 3817238, 3817237, 3941585, 3925043, 3817266, 3925086, 3980773, 3972955, 3974934, 3980795, 3814949, 3972969, 3941602, 4039559, 4087206, 4157670, 3817254, 3925075, 3892012, 3892016, 3892041, 3892043, 3892044, 3892045, 3892061, 3892065, 3892066, 3892072, 3892074, 3892075, 3892076, 3892079, 3892085, 3925081, 3925114, 3941575, 3941603, 3974935, 3974942, 4151096, 4077239, 3817239, 3925089, 3972926, 3839541, 3814950, 3817280, 3817267, 3839554, 3892057, 3925091, 3980772, 4087233, 3941614, 4087210, 3925093, 3843198, 3814954, 3817282, 3843209, 3925076, 3892008, 3892056, 3817284, 3839518, 4028314, 4068199, 3814956, 3814931, 3814970, 3925037, 3925083, 3980766, 3925088, 3980741, 3814960, 3839550, 3843195, 3925044, 3817271, 3925120, 3925045, 3980770, 3980701, 4028306, 4068195, 3817244, 3839601, 3980793, 3972934, 3980799, 3817253, 2517158, 2252549, 2226919, 2610150, 2517162, 2610082, 2610105, 2229686, 2168771, 2244332, 2610121, 3705758, 3724484, 3790772, 2979663, 2823818, 3790770, 2823792, 3661875, 3687273, 3553869, 3661910, 2763697, 3553814, 3705769, 2788570, 3754784, 2715431, 3681041, 3705752, 3665010, 3775101, 3705763, 3661921, 3606790, 3661866, 2715444, 2937984, 2897532, 2966211, 3724502, 3589296, 3589267, 3661934, 2715354, 3553839, 3589251, 3589328, 3705780, 2921052, 3589290, 2823810, 3606813, 3661928, 3553804, 3790800, 3589317, 3589294, 3661946, 2791828, 2897522, 2966216, 2937978, 2966203, 3687269, 3553837, 2979657, 2921032, 3681009, 2647264, 3606756, 3724475, 2715350, 2763714, 3775086, 2873988, 2921066, 2966187, 3754745, 2937975, 3661936, 3740749, 3553829, 3754769, 2921059, 2564855, 3681024, 3606805, 2937994, 2574213, 2574299, 2937996, 3740771, 3724506, 2979658, 3790765, 3553842, 2647241, 2966190, 2806828, 3553867, 3705760, 2791833, 3724478, 3553777, 2931703, 3687234, 3553833, 3681023, 3589347, 3553820, 3553786, 3839574, 4157674, 4151105, 3941579, 4039541, 4467218, 4214443, 4253810, 4326994, 4214409, 4264669, 4419008, 4486031, 4494412, 4467279, 4467282, 4087209, 4419029, 3980796, 4281259, 4375775, 4415692, 3980781, 4375800, 4467257, 4253807, 3972918, 3972946, 4039547, 4253822, 4253832, 4419006, 3980797, 4157649, 4157677, 4253790, 4254808, 4419012, 4415684, 4253831, 4254775, 3925097, 4087240, 3980792, 4151126, 3974945, 4281261, 4253801, 4419031, 4415708, 4419016, 4419032, 4467205, 4264681, 4467245, 4418998, 4419022, 4494429, 4157643, 4191706, 4191742, 4327014, 4415727, 4467232, 4068226, 4443558, 3980769, 4068228, 4157651, 4077241, 4077244, 4068191, 4191711, 4214434, 4191747, 4191702, 4214400, 4214428, 4300775, 4264690, 4314023, 4314009, 4326983, 4314033, 4341655, 4375784, 4327004, 4403150, 4415694, 4415646, 4415698, 4420967, 4415749, 4420948, 4467195, 4467190, 4467201, 4467199, 4415716, 4467307, 4313994, 4314044, 4467286, 4415654, 4494432, 4253803, 4151116, 4108426, 4191714, 4191736, 4253802, 4253804, 4314027, 4314065, 4327026, 4326998, 4327024, 4375797, 4415647, 4415705, 4420949, 4419033, 4419021, 4419003, 4314059, 4151115, 3839553, 4191709, 4254773, 4254800, 4327008, 4467208, 4151107, 4191730, 4151091, 4253791, 3839581, 4327034, 4494435, 4073910, 4191715, 4467303, 4108443, 4375785, 4415706, 4420952, 4467254, 4443575, 4313990, 3892052, 4420972, 4167234, 4214396, 4264668, 4281260, 4313979, 4028293, 4341672, 4327039, 4375771, 4327012, 4415669, 4447827, 4467272, 4415688, 4415740, 4157648, 4361310, 4077227, 4008418, 4327028, 3925100, 4157680, 4327011, 3917007, 4068218, 3980777, 3925453, 4314034, 4051307, 4341678, 3839586, 4167230, 4253793, 4467305, 4483524, 4253823, 4327031, 4361330, 4157671, 4314047, 4415754, 4068227, 3980761, 4443611, 4375805, 4051312, 4254798, 3941586, 4151119, 4467192, 3980774, 4281253, 4327006, 4419014, 4087196, 4108438, 4157678, 4214438, 4281238, 4300772, 4300784, 4300776, 4327009, 4415723, 4314045, 4415736, 4467238, 4467297, 4453426, 4483529, 4483518, 4073923, 3972915, 4214414, 4254802, 4314036, 4314004, 4191708, 4443609, 4375762, 4087197, 4314041, 4314002, 4375774, 4341693, 4375757, 4467300, 4483520, 4494448, 4314030, 4281242, 4314054, 4214431, 4341685, 4254807, 4167244, 4191724, 4167226, 4174120, 4264687, 4314029, 4253796, 4415732, 4483526, 4253820, 4068201, 3972921, 4174130, 3843181, 4314001, 4326965, 4191704, 4039550, 4361332, 3817274, 4403148, 4403158, 4419034, 4327020, 4157658, 4313989, 4483523, 4486030, 4419011, 4039567, 4326964, 3941590, 4039554, 4415747, 4174124, 4157638, 4375782, 3925456, 4314061, 4300773, 4327003, 4361322, 4415724, 4375752, 4415746, 4087232, 3839575, 4191728, 4253783, 4214439, 4281245, 4087241, 4375760, 4415690, 4453422, 4467309, 4254785, 4415675, 4314058, 4483533, 4453419, 4494422, 4108440, 3839578, 3941580, 3941622, 3843193, 4214430, 4151090, 4264686, 3817276, 4087211, 4254791, 4300769, 4327030, 4214440, 4327001, 4375806, 4375787, 4375779, 4443593, 4419045, 4467290, 4483532, 4326981, 4494439, 4313993, 4028318, 4214420, 4214395, 4341660, 4403146, 4415650, 4415728, 4443612, 4083687, 4151088, 4415755, 4467187, 4214398, 4375780, 4327013, 4264689, 4253814, 4077221, 4327018, 4253798, 4453431, 4068205, 4077257, 3839513, 4254786, 4191735, 4108428, 4341700, 4073921, 4415679, 4167227, 4108430, 4157637, 3843166, 4073907, 4375761, 4419039, 4077228, 4178087, 3843176, 4151106, 4068186, 3985532, 4254761, 4254803, 4281262, 4264679, 4281255, 4314052, 4083691, 3843180, 4264665, 4327023, 4327038, 4191703, 4327035, 4341666, 4314012, 4375768, 4341662, 4420966, 4420969, 4453427, 4494450, 4415682, 4467283, 4494436, 4483527, 4167223, 4191725, 4361321, 4157646, 4453413, 4467301, 3941588, 3972948, 3972964, 4453416, 3972949, 4300780, 4415738, 4443582, 4008424, 4214418, 4167236, 3980709, 4494427, 3925428, 4191727, 4327002, 4415745, 4443601, 4157660, 4191723, 4375793, 4253811, 4467262, 4375763, 4415652, 4326996, 4415734, 4443571, 4467180, 4341696, 4326985, 4467277, 4068221, 4326988, 4375754, 4486054, 4443562, 3941587, 4326967, 4467311, 4419013, 4341659, 4108446, 4087239, 4254795, 4253785, 4151125, 4326977, 4361334, 3980780, 4314028, 4415744, 4108434, 4375767, 4068224, 4108425, 4264670, 4264685, 4281257, 4151108, 4361323, 4341677, 4361312, 4157653, 4051319, 4375783, 3972950, 4415718, 3839532, 4083688, 4483534, 4467186, 4281248, 4178094, 4314013, 4415687, 4254787, 4341653, 4008393, 4191710, 4253812, 4326972, 4327021, 4068212, 4415683, 4068198, 4443567, 4073931, 4415696, 4313999, 4467193, 4068183, 4191719, 4253784, 4108424, 4214417, 4361316, 4418999, 4341682, 4494402, 4068204, 4051314, 3980734, 4327033, 4361318, 4191716, 4191718, 3843164, 4083685, 4341691, 4443578, 4326963, 4300794, 4077247, 4178077, 4077251, 4191721, 4174125, 4191739, 4214423, 4253794, 4300786, 4326987, 4341705, 4253795, 4403152, 4467234, 4467224, 4191744, 3980715, 4483538, 4494416, 4467264, 3839536, 4264680, 4073909, 4087220, 4039564, 4083686, 4254804, 4253816, 4314005, 4467280, 4157655, 4087215, 3839524, 3892078, 3839509, 3941624, 4174127, 3925433, 4077252, 4214422, 4174118, 4087227, 4174121, 4314053, 4264675, 4264664, 4087247, 4281258, 4314007, 4326973, 4326993, 4313975, 4326980, 4375769, 3925070, 4375801, 4327007, 4253797, 4253792, 4415676, 4361320, 4415695, 4191749, 4313981, 4443568, 4415731, 4467211, 4415741, 4375770, 4453423, 4453424, 4253782, 4415702, 3892013, 4178085, 4419001, 4494420, 4467182, 3837332, 4483530, 4494437, 4494404, 4494425, 4300774, 4494434, 4483517, 4419023, 4300792, 4443583, 4443592, 4443579, 4327037, 4483536, 4077233, 4178074, 4415668, 3892054, 4178082, 4375795, 3917016, 4314021, 4361311, 3980737, 4077249, 3925437, 4403154, 4415691, 4443574, 4314014, 4486040, 4467203, 4008425, 3814937, 4467294, 3843173, 4264674, 4151132, 4281250, 4313976, 4420964, 4361307, 4077246, 4087225, 4214405, 4253800, 4214416, 4253789, 4254792, 4420955, 4483515, 4415680, 3980721, 4375772, 4443596, 3941626, 4254809, 4415735, 4008399, 4083701, 4326999, 4077250, 4415710, 4415661, 4467185, 4157659, 3980724, 4254757, 4443591, 4418996, 4494451, 4151094, 4254776, 3925432, 4415715, 4191741, 4174137, 4157650, 4167222, 4191720, 4087219, 4300787, 4254763, 4314060, 4313974, 4326997, 4375789, 4313997, 4415721, 4415664, 4420951, 4453430, 4419027, 4415720, 4486029, 4467229, 4483522, 4453429, 4178093, 4326974, 4191743, 4264688, 4151095, 4253799, 3817278, 4375766, 4028295, 4375759, 3941589, 4467250, 4157642, 3980700, 3980720, 4361309, 4254759, 3814968, 4264676, 4051310, 4254797, 4178072, 4313986, 4254801, 4341683, 4467194, 4178076, 4174138, 4264682, 4151109, 4253815, 4341670, 4254799, 4191734, 4077216, 4167237, 4467273, 4254762, 4077242, 4281256, 4314000, 4314015, 4327043, 4341674, 4327019, 4341657, 4341667, 4415681, 4039566, 4361329, 4157663, 4073925, 4327022, 4068197, 4087223, 4157640, 4191713, 4191722, 4191737, 4068214, 4313980, 4253817, 4051304, 4486034, 4375781, 4415651, 3974928, 4178091, 4415700, 4214401, 4483516, 4486042, 4327015, 4254783, 4415714, 4314035, 4151127, 4467207, 4415666, 4073917, 4087212, 4314057, 4068196, 4415750, 4341664, 4327044, 4341668, 4077218, 4314038, 4073908, 4361313, 4415704, 4419035, 4077259, 3843163, 4254766, 4214429, 4008403, 4341703, 3892011, 4178070, 4415712, 4419046, 4214433, 4214446, 4068209, 4068206, 4157675, 4467304, 4178075, 4108448, 4467221, 4108444, 4375758, 3837317, 4178088, 4039549, 4178084, 4214445, 4326962, 3837313, 3925129, 4087192, 4254768, 4327017, 4415733, 4415719, 4375765, 4314020, 4341702, 4077248, 4051309, 4254777, 4313983, 4151098, 4178069, 4051300, 4157662, 4151131, 4191705, 4167224, 3839516, 4191732, 4191700, 4264677, 4174129, 4174133, 4281252, 4264683, 4214442, 4313973, 4326970, 4077245, 4167220, 4326952, 4314064, 4361314, 4214415, 4281254, 4375756, 4300791, 4281237, 4361326, 4314016, 4327042, 4341689, 4415743, 4419042, 4327000, 4341695, 4483521, 4494431, 4415663, 4467209, 4483539, 4467220, 4494407, 4415674, 4077260, 4083693, 4254793, 4167238, 4087244, 4174128, 4494419, 3892089, 3980729, 4191748, 4420962, 4415713, 4453425, 4443603, 4494423, 4083703, 3917017, 4214407, 4108422, 4467242, 4083680, 4403151, 4375764, 4073911, 4008416, 4087200, 4415722, 4253824, 4214447, 3941617, 4494445, 4300782, 4300793, 4068210, 4314069, 4077223, 4077212, 4254767, 4083681, 4191699, 4214444, 4314032, 4300783, 4300789, 4327029, 4326959, 4467233, 4254784, 4419040, 4087201, 3980754, 4415703, 4008391, 3980702, 4151099, 4300777, 4157641, 4068192, 4151102, 4178071, 4314019, 4157668, 4443606, 4068222, 4361327, 4157665, 4420970, 4420960, 4419024, 4341704, 4467316, 4467312, 4494446, 4341690, 4494406, 4494449, 4087221, 4151112, 4341661, 4214413, 4051322, 4051323, 4415665, 4039557, 4415748, 4415660, 4467197, 4494447, 4467191, 4486032, 4087231, 4494441, 4051315, 4300778, 4341699, 4467219, 4326956, 4157682, 4214441, 4191712, 4254771, 4214426, 4327016, 4361324, 4443588, 4341671, 4341675, 4174117, 4174132, 4028321, 4174142, 4254764, 4178086, 4415737, 4467302, 4214437, 4415739, 3980726, 4415685, 3972937, 4467276, 4486048, 4453417, 4486055, 4415709, 4151122, 4077265, 4068234, 3817252, 3814928, 4077258, 4254796, 4341684, 4077232, 4253813, 4326960, 4326961, 4467215, 4467225, 4214402, 4214406, 4214410, 4214411, 4214419, 4214432, 4254769, 4254788, 4281240, 4281246, 4314040, 4314068, 4313995, 4314011, 4314042, 4341688, 4443607, 4443613, 4447828, 4447829, 4419047, 4467252, 4467270, 3974949, 3974950, 4375777, 4415659, 4419048, 4443580, 4467288, 4415678, 4415725, 4467184, 4467189, 4467230, 4467259, 4443595, 4443598, 4415672, 4415701, 4467214, 4494403, 4494418, 4494415, 4494398, 4494410, 4467236, 4467244, 4467269, 4467281, 4467251, 3925056, 4051306, 4415707, 3925060, 3980785, 3925067, 4068182, 4151120, 4214435, 4254780, 3817269, 4264667, 4039568, 4214436, 4281241, 3980790, 3980768, 3941594, 3925058, 4314039, 4326975, 4341697, 4068187, 4314017, 4415656, 4264666, 4467248, 4467266, 4420965, 4254805, 3925095, 4415662, 4420954, 4264671, 4415699, 4415673, 4486039, 3925110, 4467315, 4254758, 4403145, 4418995, 4174139, 4327040, 4253805, 4151114, 4174126, 4174141, 4178081, 4178083, 4174119, 4361315, 4077255, 4157667, 3980767, 3925051, 3980783, 4415677, 4264678, 4403156, 4314024, 4214449, 3925077, 4415689, 4254781, 4415667, 4178090, 3925445, 4054878, 4467202, 4486038, 4008395, 4073920, 4281244, 4420963, 3972932, 3925094, 3925409, 4313985, 3972943, 3941620, 4254806, 4028313, 3941583, 3925074, 4087205, 4077229, 4264684, 4314025, 4313992, 3980771, 4326984, 4419030, 4077253, 4375794, 3980787, 4087237, 4167219, 4264672, 4281236, 4326979, 3941619, 4028307, 4214399, 4254770, 3925053, 4254810, 4254790, 4403159, 4418997, 4077236, 3839589, 3972916, 4375791, 3814932, 4403153, 4191746, 4028301, 4151100, 4178079, 4341680, 3925046, 4313987, 4467217, 3925080, 4068189, 4281263, 4214412, 4151103, 4314022, 4068211, 3972966, 4039544, 4151118, 3980794, 4313996, 4419009, 3972968, 3817258, 3925408, 4008408, 4008384, 4151121, 3892068, 4326968, 4375798, 4174143, 4191738, 4467255, 3972940, 4108432, 4008430, 4028309, 4375786, 3972965, 3925036, 4314026, 4326992, 4314055, 4077256, 4167232, 4151128, 3972956, 4254782, 4214404, 3941616, 4167228, 4253825, 4264691, 4191740, 4253829, 4341706, 4313972, 4419004, 3941595, 4077235, 4157683, 4191745, 4178073, 4254812, 4254779, 4214424, 4264673, 4403160, 4415753, 4415729, 4028317, 4676690, 4676715, 4676724, 4676731, 4676713, 2362722, 2252509, 3724472, 3680999, 3606802, 3553854, 3661871, 2647263, 3687267, 2564890, 3687226, 3661874, 3705766, 3705759, 2873998, 2823787, 2791845, 3661878, 3661883, 2979643, 2763713, 2763698, 2647250, 2574310, 3790776, 2574324, 2966196, 3589270, 2832131, 3775116, 3775102, 2897511, 3925066, 4420947, 4253788, 4419044, 3892090, 3839605, 4253826, 4415726, 4039542, 4314008, 4326989, 4494400, 4420959, 4419050, 3925090, 4415730, 4253809, 4415693, 4253818, 4486033, 4068181, 4443565, 4494444, 4467228, 3839590, 4326954, 3917013, 4314050, 4494413, 4254811, 4443570, 4174140, 4281243, 4087245, 4494442, 3843192, 4486047, 4443602, 4174144, 4314066, 4443600, 4453414, 4443581, 4447830, 4341694, 4494428, 4253821, 4068190, 4494453, 4467287, 4403149, 4483519, 4453412, 4467289, 4157636, 4087238, 4028302, 4483535, 4375776, 4314049, 4443608, 4467295, 4467196, 4447826, 4054889, 4008419, 4467241, 3814924, 4467204, 3925413, 4068215, 4419026, 4419018, 4167243, 4341676, 4300795, 4494433, 4008405, 4191726, 4486041, 4341687, 4300781, 4494443, 4443566, 4375799, 4191698, 4087243, 3917004, 4486036, 4327005, 4326990, 4443590, 4467256, 4157669, 4341701, 4313984, 4174135, 4443594, 4068233, 4326991, 4403157, 4467246, 4083697, 4313970, 4443569, 4443589, 4467267, 4443597, 4077234, 3925444, 3925414, 4054864, 4008421, 4486050, 3839587, 4443586, 4403147, 4443587, 4467293, 4483513, 4419000, 3814940, 4068231, 4073914, 4467263, 4467278, 4447825, 4467284, 4419010, 3925435, 3925133, 4494417, 4054873, 4467239, 3925126, 4254772, 4174123, 4314031, 4326953, 4313971, 4254794, 4415711, 4073928, 4443573, 4375790, 4483525, 4486045, 4157679, 4415649, 4361333, 4420968, 4341651, 4054885, 4054888, 4341686, 4467292, 4361308, 4419038, 4419049, 4326976, 4443561, 4361331, 4326986, 4467223, 4467216, 4087229, 4254760, 4077263, 4467188, 4453421, 4281251, 4214421, 4178080, 4443585, 4467275, 4054876, 3972919, 4415671, 4415697, 4214408, 4073912, 4494430, 4341698, 4191729, 4443576, 4191731, 4375773, 4300785, 4281247, 4313991, 4443599, 4443577, 4314051, 4415657, 4375753, 4341656, 4253828, 4327041, 4314056, 4051321, 4494438, 4467306, 4178089, 4486035, 4467268, 4467314, 4467181, 4051302, 4087234, 4327010, 4467313, 3972927, 4467274, 4178092, 4326950, 4443560, 4313978, 4467227, 4341673, 4419005, 4300790, 4483514, 3892015, 4051316, 4486056, 4467206, 4443604, 4341669, 4467249, 4419028, 4341681, 3892007, 3892070, 4467222, 4326958, 4253819, 4361319, 4486037, 4174131, 4281249, 4467210, 4419041, 4254774, 4486053, 4326995, 4486043, 4443605, 4467271, 3817261, 4326978, 4375802, 4375796, 3925128, 4167239, 4415653, 3892077, 4415670, 4415655, 3925054, 4375788, 4467226, 4419025, 4415648, 4419007, 4253830, 4467243, 3925400, 4443584, 4314010, 4415686, 4443559, 4467258, 4314043, 4467183, 4151104, 4467235, 3980779, 4151113, 4313982, 4327025, 3972929, 3925078, 3972953, 4467237, 4326951, 4467213, 4467212, 4467308, 4419017, 4713691, 4808757, 4961704, 4779104, 4859176, 4755263, 4773161, 4975001, 4676714, 4676712, 5045542, 4810549, 5045544, 4978155, 4755274, 4859190, 4827301, 4916622, 4676627, 4763526, 4839766, 4916609, 4810552, 4713676, 4713683, 4859183, 4729573, 5045515, 4992037, 4978154, 4804890, 4958965, 4992085, 4676692, 4773150, 5095079, 5118726, 4676615, 4676648, 4676638, 4727842, 4676695, 4729577, 4727817, 4729555, 4690585, 4773164, 4779090, 4755283, 4779085, 4713692, 4773143, 4713700, 4787130, 4839754, 4713703, 4859180, 4890060, 4796396, 4713684, 4892119, 4892124, 4916623, 4892132, 4869534, 4933647, 4975003, 4975014, 4992091, 4859184, 5034820, 4907985, 5045508, 5053373, 5045540, 5045533, 5034849, 5122012, 5096733, 5141703, 5034852, 5141699, 5151834, 5161827, 5151824, 4925229, 4907976, 4676642, 4690576, 4690589, 4713661, 4763514, 4890041, 4834119, 4848919, 4961692, 5015601, 4890068, 4859182, 4676736, 4820121, 4866433, 4848946, 4827271, 4808766, 4810542, 4892134, 4934100, 4934077, 4848907, 4892117, 4934081, 5095064, 5045504, 4729556, 4763502, 4779106, 4779086, 4773149, 4907973, 4952477, 4961721, 4961683, 5015585, 4974999, 5161857, 4727843, 5118715, 4676749, 4676635, 4713697, 4676705, 4763521, 4755282, 4763515, 4779092, 4787083, 4787150, 4787101, 4820008, 4690568, 4907984, 4933659, 5118701, 4992039, 4933593, 5025325, 4958980, 5015587, 4713698, 4978140, 5034824, 4690599, 4676702, 4804901, 4727846, 4676626, 4763525, 4755269, 4699920, 4755267, 4779107, 4869540, 4727822, 4848910, 4848940, 4848909, 4866435, 4763507, 4907986, 4713652, 4676723, 5137381, 4676661, 4727835, 4729574, 4676609, 4713664, 4834106, 4892127, 4787099, 5015614, 4866458, 5045527, 5076081, 4992021, 5034837, 4773162, 4773172, 5025334, 4676662, 4676636, 4827269, 4834097, 4892142, 5045513, 5082589, 4676722, 4727805, 4729572, 4690562, 4819996, 4773128, 4804894, 4892118, 4755279, 4961688, 5045532, 5015566, 4961716, 5082587, 4992082, 5151827, 5122022, 4713668, 4958962, 4961732, 4690615, 4804912, 4727814, 4796373, 4866444, 4676734, 4934087, 5053353, 4810562, 4827275, 5111574, 4961709, 4804892, 4916616, 4907989, 4907980, 5034818, 4925217, 5118703, 4992090, 4690610, 4713701, 4820122, 4676631, 4729534, 4890064, 4869541, 5053356, 4992032, 4974988, 5096730, 5025318, 5025328, 4869554, 4779088, 4827250, 5045541, 4961714, 4961724, 5025324, 5137376, 5151823, 4727812, 4819992, 4848942, 5034817, 4974977, 5053357, 4869542, 4676640, 4727806, 4727832, 4713685, 4676693, 4834098, 4787151, 4796398, 4820108, 4834103, 4890065, 4978143, 4961690, 5015610, 5082573, 5122055, 5034815, 5122060, 5015595, 5161845, 5122014, 4676700, 4676606, 4690587, 4676639, 4755268, 4729580, 4773159, 4676645, 4699925, 4713682, 4804899, 4892141, 4839765, 4907982, 4933601, 4933605, 4892139, 4978151, 4975015, 4975020, 4934094, 5015577, 4992083, 4992028, 4787103, 5076062, 5082586, 4992097, 5095073, 5053364, 5161858, 5151813, 4890067, 4839768, 4804903, 4961722, 4961689, 4866437, 4933672, 5082584, 4933669, 5096735, 5025322, 5111578, 5015576, 5172639, 4773170, 4834123, 4787090, 4869550, 4848952, 4961734, 5151835, 4933649, 4773141, 4713695, 4773125, 4796392, 4859187, 5076082, 5025335, 4925232, 5141712, 5015594, 4729554, 4827300, 4958969, 4787140, 5076058, 5118700, 5111575, 4975026, 4676739, 4787155, 5122072, 4773122, 5053377, 4869539, 5096745, 5053378, 5032686, 5118723, 5045546, 4820003, 4787092, 4827264, 4907977, 4978137, 4859171, 4676743, 4676699, 4676738, 4676697, 4690603, 4727821, 4699916, 4729557, 4773135, 4773153, 4779093, 4713670, 4839745, 4834104, 4773127, 4820105, 4819994, 4820000, 4773152, 4820015, 4827270, 4787104, 4820017, 4787100, 4890044, 4890042, 4890046, 4848958, 4952471, 4916612, 4952456, 4975010, 4974989, 4992095, 4961725, 5015617, 5015600, 5034844, 5045502, 5034831, 5015604, 5045538, 5076069, 5045512, 4787139, 4848931, 5025341, 5053363, 5118702, 5111562, 5015567, 5137366, 5122013, 5151822, 5141694, 5137693, 5137718, 4729568, 4690570, 4779103, 4690590, 4690598, 4804915, 4834116, 4916605, 4866439, 4866449, 5034823, 5053369, 4729537, 4787143, 4676740, 4755277, 4690601, 4796390, 4848906, 4699932, 4810557, 4787078, 4796395, 4820001, 4796372, 4992071, 5015584, 4848924, 4866434, 4763509, 4848941, 4907981, 4676737, 4729560, 4763499, 4866447, 4729544, 4827313, 5015602, 4992031, 5111585, 5111582, 4978134, 5141707, 4869545, 4961735, 4916624, 4961733, 4820006, 4690608, 4729553, 4890043, 4961711, 4975013, 4890054, 4961736, 4690600, 4690571, 4907992, 5015572, 5015588, 5034830, 4952465, 4727825, 4892136, 4779098, 4975027, 5045514, 4690561, 5082596, 4713672, 5045520, 4763520, 4729578, 4676629, 4727833, 4676664, 4763519, 4810556, 4820012, 4992087, 4676688, 4804916, 4839763, 4992070, 4974994, 4866441, 4727834, 5096729, 4892123, 5121999, 4727813, 4869552, 4804891, 4975028, 5015573, 5141693, 4676735, 4727829, 4808752, 4839744, 4890057, 4952479, 5137716, 4773157, 4713659, 4773139, 4690582, 4796397, 4819993, 4890061, 4848935, 4952464, 4958983, 4952474, 4992092, 5045530, 5082583, 4974991, 5161835, 4907988, 4961699, 4755273, 4978157, 4961693, 5122009, 5032687, 5095076, 4787082, 4933604, 4978142, 4961706, 5015608, 5111565, 4729558, 4763501, 4676744, 4755278, 4755285, 4779094, 4699922, 4787154, 4779111, 4848923, 4934089, 4779096, 4961694, 4992086, 4779109, 5034835, 5122062, 4727828, 4676687, 4676655, 4676742, 4676632, 4676633, 4676730, 4699931, 4676746, 4676728, 4690605, 4727849, 4676745, 4676625, 4676727, 4690566, 4690609, 4690565, 4676619, 4729570, 4727841, 4727824, 4690572, 4727803, 4690581, 4727816, 4727809, 4763524, 4763516, 4773132, 4779095, 4690580, 4763518, 4727815, 4773121, 4713660, 4787156, 4810541, 4787105, 4839753, 4839758, 4787145, 4810547, 4839748, 4820004, 4773146, 4820016, 4820014, 4787142, 4827294, 4820130, 4787147, 4755265, 4916626, 4848961, 4952476, 4834107, 4952461, 4952466, 4827317, 4961708, 4992018, 5015596, 4992017, 4992080, 5034806, 5034839, 5034828, 4848921, 5034847, 5032701, 4796389, 4787141, 4839749, 5045535, 5034846, 4848908, 5096754, 5076063, 5096738, 5025329, 5096737, 5095068, 5122066, 4827302, 5118706, 5137708, 5137391, 5137390, 4796383, 4796376, 4676696, 4699934, 4713687, 4713677, 4820125, 4848913, 4961681, 5082577, 4727836, 4676634, 4713667, 4808767, 4975029, 5025321, 4773124, 4958977, 4820113, 4820106, 4961717, 4866436, 5045534, 5015564, 5137368, 4958985, 4676733, 4796379, 4934088, 4773138, 5096736, 5122065, 4763504, 4729542, 4755266, 5034853, 4713680, 5025332, 4974981, 5034821, 5151832, 5034822, 5034856, 4727837, 4848959, 5034838, 5034832, 4773147, 5015568, 4975019, 4729548, 5122000, 4848954, 4848951, 4848948, 4787137, 4729551, 4933656, 4729535, 4787089, 4755288, 4834122, 4834112, 4729527, 4676621, 4727826, 4755275, 4690597, 4810545, 4827308, 4907993, 4952472, 4787146, 4961707, 4958970, 4961684, 4992088, 5025337, 5025339, 4848937, 5096752, 4676710, 4907996, 4773148, 4859173, 4934084, 4961715, 4804895, 4810554, 4819997, 5082588, 5032684, 4958986, 4992081, 4974995, 5141713, 4796385, 4892116, 4869557, 4952457, 4975007, 4690596, 5053354, 5137377, 4839757, 5082571, 5096732, 5141686, 4713681, 4834109, 4907998, 5032692, 4978141, 4676689, 5053366, 4676725, 4676646, 4699939, 4713655, 4676660, 4975016, 4992093, 5045529, 5015570, 4978138, 4727853, 4676663, 4869546, 4892125, 4892120, 4848930, 5082597, 4907975, 4676747, 4676649, 4690584, 4729559, 4787091, 4779101, 4890062, 4952455, 5015565, 4848932, 5034845, 4866456, 5034834, 4974986, 4827256, 4892128, 4869535, 4827254, 4699924, 5111583, 4713656, 4713671, 4787096, 4796381, 4820127, 4827297, 4907991, 4848960, 4978158, 4676605, 4713690, 4975006, 4676653, 4820120, 4992076, 5076075, 4755270, 4676750, 4834101, 4933646, 4890047, 4676658, 4890048, 4934076, 4933666, 4975011, 5082591, 4834114, 4690612, 4699942, 4773137, 4925219, 5034804, 5015607, 4690619, 4755281, 5015603, 4676726, 4727818, 4773130, 4916621, 4787095, 5096734, 4763511, 5118714, 4779099, 5015615, 4958967, 4961718, 4975030, 4676643, 4892121, 4808751, 4727840, 4848950, 4796388, 4808750, 5096727, 4839769, 4907974, 4859194, 4974993, 5155951, 4676651, 4773133, 4787138, 4796380, 5015586, 5161868, 4676618, 4676656, 4676637, 4699918, 4729571, 4729575, 4755280, 4727847, 4729563, 4713650, 4755284, 4773123, 4787152, 4690564, 4729564, 4690575, 4729562, 4796387, 4796394, 4727845, 4787098, 4820009, 4834124, 4779089, 4834121, 4892140, 4892130, 4866454, 4820007, 4869559, 4839743, 4974996, 4839746, 4952463, 4869551, 4827312, 4839761, 4916617, 4952468, 4827291, 4933611, 5045545, 5034850, 5015583, 5015578, 5076076, 4729566, 4869536, 5082581, 5082582, 5025336, 4859188, 4933603, 4907978, 5095071, 5122053, 5141711, 5095072, 5137711, 4975021, 4974982, 4907994, 5053352, 4839742, 4676698, 4763510, 4729539, 4729528, 4787149, 4839750, 4779087, 4961685, 4975022, 5045524, 5053372, 4827307, 4934099, 4869549, 4808753, 4729545, 4834117, 5095070, 5034825, 4933607, 4810555, 4773154, 4934101, 4779105, 4676616, 4763512, 4892122, 4729525, 4779097, 4779091, 4804909, 4820018, 5141705, 4676650, 4676659, 4690574, 4713648, 4763500, 4773136, 4773134, 4796377, 4869538, 4713665, 4676691, 4727819, 5015574, 4773142, 4796371, 4699929, 4952470, 4848936, 4820104, 4796378, 5025333, 4690577, 4690607, 4713651, 4890049, 4961703, 4676652, 4676613, 4773126, 4848943, 4827316, 4975009, 4866438, 5034811, 5032706, 4961710, 5045525, 5095056, 4690573, 4779100, 4834099, 5076072, 4676622, 4729567, 4907999, 4925238, 4848929, 4808765, 4848945, 5118705, 4755286, 4992020, 4925222, 5161854, 4727807, 4729561, 4713662, 4787080, 4892131, 4859196, 5034829, 5082570, 5111584, 5161844, 4676623, 4676607, 4810543, 4820010, 4974998, 5122003, 5053361, 4820119, 4676654, 4676608, 4961727, 4992078, 4961713, 5141688, 4890053, 4890058, 4804893, 5034842, 4961719, 5045500, 4961691, 5045516, 5034843, 5076065, 4952458, 4676630, 4925240, 5025320, 4866432, 4699917, 4699930, 4690592, 4690613, 4690593, 4690602, 4713679, 4727844, 4727854, 4773156, 4787084, 4787093, 4787134, 4787136, 4787153, 4796375, 4713693, 4827268, 4834102, 4834111, 4859172, 4827255, 4827274, 4827299, 4916610, 4916620, 4916630, 4934075, 4934078, 4934091, 4934093, 4934095, 4934097, 4958973, 4958982, 4992025, 4992015, 4992022, 4992030, 4848911, 4848914, 4848917, 4676703, 4676709, 4729536, 4713673, 4773158, 4773173, 4699919, 4773169, 4713653, 4713657, 4773168, 4839751, 4839755, 4839767, 4859186, 4859195, 4820115, 4820116, 4866459, 4820109, 4820110, 4820117, 4820118, 4892126, 4933597, 4933598, 4933615, 4992014, 4992023, 4975018, 5015597, 5015618, 4958974, 4978147, 5025317, 5034814, 5045522, 5025316, 5118711, 5118713, 5118721, 5137699, 5155952, 5155965, 5155943, 5155953, 5155954, 5155957, 5155960, 5155962, 5155963, 5155970, 5161823, 5161842, 4808754, 4676751, 4729533, 4676729, 4729529, 4787135, 4690616, 4773165, 4796382, 4713674, 4729579, 4810540, 4755287, 4810548, 4676741, 4808758, 4810546, 4808759, 4804902, 4810551, 4866452, 4804889, 4916608, 4713689, 4827253, 4808768, 5034809, 4992016, 4690620, 5053371, 5082574, 4690614, 4773174, 4804913, 4974976, 5034827, 4916613, 4933616, 4676711, 4796386, 5076079, 5122067, 4763508, 5015569, 4676647, 4727856, 4713696, 4787097, 4827272, 4827252, 5076083, 5161830, 4934090, 4934092, 5032710, 4773176, 4773151, 4787131, 4810561, 4820005, 4916607, 4916618, 4916619, 4933599, 4933606, 4933613, 4690618, 4925230, 4763517, 4827258, 4727838, 4958972, 4952462, 4699923, 5122011, 4787148, 4820107, 4820124, 4699940, 4690588, 4827267, 4755290, 5015580, 4690586, 4729526, 4713675, 4820111, 4820128, 4925235, 4819999, 4820123, 4933594, 4925225, 4892138, 4933592, 4890050, 4848938, 5076070, 5076061, 4925218, 4827277, 4810544, 4859178, 5034851, 4699941, 4787133, 4859181, 4827310, 5034819, 4925221, 5082594, 4713678, 4699928, 4820112, 4690617, 4690578, 4958981, 4820126, 4690595, 5015605, 4690611, 4810538, 4808775, 4773163, 5015582, 4699933, 4933608, 4827298, 4992040, 4808756, 5025340, 5034836, 4727808, 4699927, 4690567, 4827259, 4773155, 5015590, 5015571, 5034841, 5032705, 4859189, 4808773, 4808769, 4890059, 4933617, 4808771, 4934098, 4808762, 4859193, 4992019, 4925220, 5034855, 4727839, 4961695, 4934074, 4866448, 4810550, 4676707, 5076068, 4729552, 4810559, 4729546, 4779102, 4827305, 4727810, 4727802, 4804896, 4713666, 4729543, 4934096, 4727830, 4676706, 4866442, 4834100, 4755276, 4787132, 4727804, 4992033, 4961723, 4992096, 4690604, 4690594, 4808749, 4804911, 4808763, 4804907, 4787085, 4866451, 4907983, 4869547, 4925224, 4978136, 5015579, 5015606, 5015593, 4804898, 4933652, 5118712, 5015612, 4699938, 4727848, 4773171, 4773167, 4804900, 4787079, 4827296, 4839752, 5015598, 5025319, 5032707, 5082592, 5096749, 4933648, 4974975, 4978139, 4958961, 5032689, 4934086, 5032683, 5032702, 1841501, 3754757, 2915193, 3661880, 3661881, 3681036, 2921072, 2806831, 3661941, 2873995, 3724465, 3790764, 3687261, 4494401, 4420956, 4453428, 4108431, 4467291, 4253806, 4467296, 3985533, 3972959, 4467310, 3925112, 3917003, 4419020, 4494452, 4403155, 4467247, 4341663, 4483528, 4419043, 3814944, 4326966, 4191701, 4483537, 4419019, 4494409, 4068219, 4341658, 4300779, 3980731, 4068232, 4108442, 4486051, 4415658, 4087195, 4300770, 3814965, 4327027, 4077261, 4361325, 4051326, 4167242, 4375755, 4167231, 4191717, 3839517, 4453420, 4467298, 4326971, 4494421, 4191707, 4467253, 4327036, 4326957, 3839599, 4341654, 3925416, 4467231, 4214427, 4077211, 3817268, 4420971, 3980736, 3972933, 4494411, 4191733, 4068217, 4375803, 4051318, 4419015, 4087204, 4467285, 4314063, 4375778, 4214397, 4494424, 4073922, 4300788, 4300771, 4467265, 4415751, 4008401, 4420958, 4314018, 4028308, 4420953, 4375792, 3941623, 3925085, 4375804, 4313998, 4314006, 3817257, 4467260, 4151089, 5045505, 5137710, 5215390, 5076060, 5122001, 4834113, 5161832, 5161816, 5210157, 5224649, 5161833, 5189505, 5111566, 5161839, 5161817, 5155966, 5258347, 5189489, 5053362, 4925244, 5258359, 4992038, 5224657, 5111561, 5053374, 5161843, 4810558, 4992036, 5189497, 4890052, 5258361, 4676704, 5161819, 5015591, 5034854, 4834118, 4808760, 5155956, 4992072, 5161838, 5122023, 5202943, 5161825, 5210156, 5258381, 5199960, 4992089, 4992073, 5161853, 5232841, 5118724, 5202950, 4978152, 5189514, 5172640, 5189490, 5045539, 5199963, 5202975, 5189503, 5225534, 5252027, 5225685, 5232840, 5202941, 5172629, 5161834, 5202936, 5225525, 5225698, 5225687, 4848920, 5161826, 5225688, 5032690, 5252006, 5252081, 5243152, 5032696, 4961728, 5215378, 5202947, 4925239, 5210153, 5189531, 5243163, 5202946, 5225689, 5032691, 5137378, 5161861, 5199940, 4796393, 5121998, 5202991, 5215374, 5137384, 5189521, 5232832, 5034826, 5155948, 4848956, 5252026, 4676610, 5122021, 5189533, 4729540, 5252070, 4933610, 4810537, 4933663, 4992027, 5076078, 5202962, 5199946, 5095078, 4676624, 5161852, 5232857, 4933614, 5232820, 4961701, 5161862, 5151836, 5243168, 5232856, 4763505, 5258348, 4866450, 5252008, 4961682, 5202990, 4933664, 5151831, 4933591, 4804910, 4992029, 4699943, 5199944, 5053368, 5258355, 5258388, 5232844, 5252086, 4834110, 4907995, 5195115, 5258398, 5161851, 5195119, 4676628, 5118719, 5095069, 5202937, 5252061, 5258387, 5232823, 4848918, 5151814, 5141696, 5076074, 5232822, 5243157, 5258383, 5151819, 5202976, 5151818, 5155946, 5210154, 4958979, 5225515, 4907972, 5161821, 5172651, 5202964, 5224653, 5224663, 5151830, 5202958, 5111580, 5111563, 4727852, 4796384, 5082580, 5195131, 4690606, 5232847, 4804904, 4961731, 5258401, 5225705, 5161871, 5137387, 5161828, 5215366, 5137375, 5225694, 5032693, 5172653, 5224668, 4933651, 5189520, 5199938, 5189536, 5224651, 5243158, 5225526, 5252067, 5252078, 4763513, 5161840, 5215373, 5045506, 5202951, 5122064, 5118725, 5232842, 4974984, 5258385, 5161855, 5195107, 5252015, 5161866, 4808764, 5096728, 5195116, 5189530, 5210150, 5076056, 5195128, 5202942, 5224673, 5199937, 5224655, 4848955, 5199956, 5199951, 5137704, 5172654, 5172630, 5202944, 5172633, 5210143, 5045543, 5252030, 5232818, 5258395, 4974980, 4859185, 5195110, 4729550, 5202968, 5032688, 5172646, 5195126, 4933600, 5202948, 5199955, 5243156, 5224654, 5243162, 5225530, 4907987, 5224656, 5258378, 4933596, 5202978, 5224661, 5224659, 5252018, 5122069, 4827251, 5224674, 4975000, 5224671, 5122070, 4866455, 5161863, 5095062, 5137694, 5189537, 5225690, 5034810, 4787086, 4763503, 4933661, 5189524, 5258382, 5172648, 5243172, 4890056, 5232848, 5172652, 5141709, 5122017, 5195118, 5252007, 5225681, 5053355, 5095082, 5137702, 5025342, 4978133, 4848933, 4787144, 5202961, 5252025, 4975012, 5202989, 5232846, 5202938, 5032709, 5045501, 5225686, 4676620, 5225684, 5252014, 5034840, 5151811, 5195112, 5195114, 5161836, 4848939, 5122018, 4820013, 5210152, 5161859, 5252020, 5252013, 5225693, 4755272, 4961696, 5232855, 5258389, 5034808, 5151812, 5137374, 5137370, 5195133, 5137393, 5045536, 5161867, 5111571, 5225513, 5137392, 4934079, 5232858, 5232835, 5224664, 4690591, 4848926, 5095066, 4676748, 5189519, 5224660, 4796374, 5258372, 5082585, 5258390, 5082576, 5137369, 4890066, 4961686, 5137373, 4773175, 5195129, 4978149, 5141704, 5096741, 5224652, 5232843, 5137701, 5137389, 5252072, 5122054, 5122059, 4787081, 5137382, 4869553, 4713669, 5258397, 5252016, 5243174, 4690583, 4961700, 5195124, 5096739, 5122073, 5199947, 5137371, 5015613, 4952469, 5161850, 5225538, 5202983, 5224666, 5232834, 5202965, 5224669, 4974990, 4916628, 5258399, 5258351, 4773145, 5258374, 5258380, 5095075, 5258366, 5258384, 5252076, 4958966, 5189516, 4869555, 4810536, 5076064, 4916606, 5137379, 5111564, 5252009, 5137388, 4992075, 4978156, 4952453, 5096746, 4727857, 5189538, 5189541, 5252012, 5252074, 4690563, 5195125, 5015592, 4848915, 4713699, 5096731, 5151829, 5258364, 5232815, 5258393, 5111576, 5141692, 5225695, 5025338, 5202969, 5243150, 5172631, 4992077, 5161869, 5252069, 5141687, 5202970, 4961687, 4961702, 5199945, 5210155, 4961705, 5225528, 5202988, 5232816, 5252031, 5258400, 4961712, 5225516, 5172649, 4787102, 5141706, 5172642, 5172637, 5202952, 5210147, 5232833, 5202945, 5232808, 5172644, 5161847, 5172645, 5202940, 4916629, 5243153, 5243165, 5252071, 5232830, 5252079, 5252022, 4974978, 5111559.5225521, 5172643, 5137706, 5252032, 5252075, 4713654, 5045531, 5225704, 4975005, 5224648, 4952473, 4916611, 5189526, 4676716, 5215385, 5258367, 5232806, 5121997, 5215367, 5232854, 5225519, 5258377, 5243151, 5095081, 5095065, 5225518, 5151815, 4952460, 4892137, 5202984, 5172635, 5202971, 4834105, 5252068, 5189518, 5095060, 4676732, 5202986, 4763523, 5232852, 4866443, 5189515, 5202987, 5252063, 5161831, 5122007, 5032708, 5161822, 5141702, 4827263, 5195132, 4933673, 5045526, 5161870, 5095061, 5225691, 5189535, 4713658, 5053358, 4933655, 5172647, 5155945, 5199939, 4975008, 5032698, 5096742, 5225706, 5111586, 5195120, 5111567, 5258375, 4892143, 5137385, 5137714, 4978146, 5122004, 5161856, 5189532, 5137719, 4819995, 5095057, 5199957, 4727851, 4713694, 5195130, 4978131, 5141689, 5015599, 5122058, 4992041, 5195123, 5045528, 4975017, 5224670, 5151828, 5195127, 5258391, 4848944, 5252066, 5199943, 5151817, 5225537, 5202954, 4992084, 4975024, 4892129, 4729569, 4839762, 5111572, 5202985, 5045517, 4933653, 5199952, 5155944, 5045509, 5161849, 5161846, 5189527, 5095067, 4827265, 4933654, 4787094, 5225520, 5202979, 5045537, 5095058, 5252080, 5215389, 4773129, 5252065, 5252064, 5151826, 5215391, 5053365, 4974985, 4848934, 4961698, 5172638, 5232819, 5232845, 5151810, 5141690, 5202982, 5258357, 5225699, 5195117, 5252017, 4713663, 4713688, 5199954, 5210149, 4804908, 5045510, 4848947, 5141695, 5141708, 5258376, 5172655, 4992034, 5122068, 4925233, 5225708, 4974987, 5122063, 4975025, 5076066, 5118710, 5195134, 5137709, 4974979, 5076071, 5199941, 5199959, 5195113, 5243148, 5258379, 5258396, 5258365, 5232825, 4779084, 5151833, 4676641, 4713686, 5243171, 5172634, 4819991, 4958978, 4907979, 4848953, 5243169, 5195108, 4978148, 5141691, 5137698, 4839756, 5045503, 5118717, 5189539, 5122015, 4933660, 5122019, 5225517, 5199948, 5224665, 5225527, 5141697, 5202974, 4827292, 4933662, 4834120, 4820002, 5232839, 5122024, 4699937, 5034812, 5151821, 5202973, 5225522, 4804897, 5118727, 5202980, 4773140, 4933668, 5202981, 5199958, 5225700, 5202977, 4779108, 5199961, 5225512, 5141710, 5215371, 5161864, 4958959, 5161860, 5225523, 5199949, 5082590, 5076080, 5141698, 4958984, 5210145, 5053360, 5258392, 5232859, 5034816, 4763522, 5082579, 5199936, 5215377, 5215379, 5215384, 5215386, 5215388, 5215392, 5172632, 5172650, 5232810, 5232817, 5232805, 5232824, 5232838, 5252010, 5232809, 4952475, 4975002, 5122079, 4958968, 4974997, 5032694, 5189504, 5025326, 4974992, 5155967, 5122077, 5215382, 5155969, 5189506, 5189511, 4808761, 5215368, 4713702, 5215387, 5252021, 5225703, 4810539, 5189500, 5155968, 5189487, 5258373, 5258358, 4827262, 5258363, 5122010, 5232814, 5252073, 4729541, 5210151, 5252024, 5252029, 5243155, 5243159, 5243164, 5243167, 5243170, 5258360, 5258370, 5258350, 4975023, 5076067, 4827266, 5258371, 4925241, 5232828, 5225533, 5202960, 5258369, 4755271, 5025327, 5034807, 4808755, 5096743, 5161820, 5161841, 4869556, 5195111, 5034803, 4978150, 5189493, 5232850, 5232821, 5122074, 4676611, 5225707, 5096748, 5215372, 5189499, 4933590, 5215369, 5232836, 5189496, 5189492, 4869548, 4869544, 5210170, 4866445, 5189513, 5202939, 4820129, 4834108, 4834115, 4859175, 5202956, 5202949, 4925226, 4978153, 5189507, 4961729, 4934082, 5232849, 5258349, 5258352, 5232837, 4925237, 5095077, 5215375, 5034805, 5210146, 5210158, 4827257, 5258362, 5082595, 5210162, 5210161, 4827315, 5045519, 5210165, 5053370, 5232826, 5252023, 5252077, 5189512, 5096750, 5189508, 4859179, 4676614, 5137692, 4859174, 5225696, 5258386, 5243166, 4916627, 4755289, 5199942, 4676617, 5161837, 5232829, 5122076, 5232812, 4934085, 5210144, 4729538, 4961726, 5034848, 5258368, 5045511, 4978132, 5155961, 4787088, 5137367, 4958963, 5096744, 5034857, 5045523, 5095083, 4925227, 4933667, 5258354, 5118709, 4859192, 5189501, 5155958, 4773131, 5025331, 5258356, 4958976, 5155950, 4992026, 5161829, 5122006, 4810563, 4925231, 4961730, 5025323, 5225529, 5118720, 5095059, 4729532, 4933602, 4925236, 5122016, 5076059, 4779110, 4958964, 5095074, 5155964, 4808770, 5032703, 5155947, 4808774, 5189495, 5189502, 5189494, 5210159, 5210160, 4839764, 5137707, 5155959, 5189509, 5210163, 5210166, 5210167, 5137703, 5210164, 5252028, 5096753, 5118716, 4958960, 5195121, 4992024, 5232807, 4253827, 3980784, 4467261, 4051301, 4494426, 4415645, 4008426, 4314067, 3925439, 4151086, 3980764, 4327032, 4167225, 4494414, 4494405, 4419037, 4467299, 4068185, 4415752, 4486046, 4443610, 3941582, 4341652, 4341665, 5111569, 5161818, 4773166, 5082593, 4827290, 5015563, 4933595, 5122005, 5111568, 5118708, 5151820, 4727850, 4869558, 5053376, 5032695, 5118704, 5258402, 5118707, 4729530, 5243154, 5189525, 5053375, 5045507, 4729547, 5225683, 4978135, 5252082, 4859198, 4848916, 5232811, 4699935, 5232851, 4676694, 5225535, 5076073, 5076077, 5151816, 5215383, 5151825, 5202959, 4787157, 5189523, 5122008, 5243149, 5202953, 4713649, 5172656, 5252088, 5252085, 4916614, 5224650, 4839759, 5137386, 5118718, 5243173, 5053359, 4934083, 5096740, 5082572, 4866453, 4690569, 5202963, 5202957, 5122002, 4827304, 5224662, 4978144, 4961720, 5195109, 5015611, 5252083, 4773144, 5141700, 5232813, 5137383, 5189528, 4827309, 5210148, 5025330, 5215370, 5199962, 5161824, 4961697, 4890045, 5095063, 4974983, 5045518, 5111577, 5232827, 5137697, 5137713, 4978145, 4992074, 5224672, 5189529, 5161865, 5141701, 4810560, 5034833, 5172641, 5252011, 5252087, 5189491, 5225682, 5215376, 5225701, 5199953, 5195122, 5243160, 5082578, 5015575, 5210169, 5111573, 5096747, 5252019, 5189486, 5202955, 4925223, 5210168, 4820011, 4839747, 4808776, 5032697, 5032685, 5252005, 5224658, 5225514, 5225702, 4992035, 5034813, 4933609, 5025315, 5122080, 4925242, 4866457, 5015589, 4727855, 4808772, 5258353, 5096751, 5118722, 4934080, 4890055, 5034802, 4907997, 5880137, 6118828, 5946631, 6077641, 6230390, 6077606, 6281900, 6050742, 5880115, 5924639, 5899028, 6072888, 6234482, 6207657, 6245535, 5924656, 5880485, 6077892, 6110735, 6100649, 5924658, 5897695, 6245527, 6142494, 6202192, 6245563, 6229094, 6142496, 6077874, 5880505, 5897701, 6077616, 6077612, 6142512, 6100642, 5880136, 5899012, 6245533, 6245508, 6077893, 5880139, 5880119, 5924650, 5899016, 6077890, 6072869, 5897757, 5902910, 5897747, 5924657, 6050689, 6077877, 6225193, 6234479, 6255728, 5924654, 6072867, 5899109, 6245555, 6043119, 6145124, 6202249, 6215927, 6202242, 6118819, 5897691, 5880146, 5985225, 5988843, 5931483, 6026665, 5897675, 6050711, 6050732, 6077618, 6050716, 6077650, 6090478, 6113535, 6145135, 6145116, 6185663, 6145123, 6185686, 6185661, 6212537, 6202220, 6225170, 6230404, 5836440, 6212536, 5836441, 5899120, 5884878, 5897677, 5898224, 5899026, 5902920, 5897641, 5931491, 5988838, 5994813, 6014744, 6014748, 6050714, 6043118, 6013893, 6050676, 6079014, 6050701, 6077645, 6077894, 6090312, 6090473, 6090457, 6050724, 6145131, 6176392, 6177068, 6176391, 6185664, 6118831, 6185659, 6014760, 6185681, 6202250, 6212534, 6202245, 6230418, 6254055, 6215915, 6245553, 6254046, 6255727, 6225182, 6090311, 6077914, 6207668, 6215932, 5951315, 6017266, 5924655, 5924651, 6043113, 6043123, 6077901, 6072885, 6090306, 6138892, 6159257, 5897665, 6142751, 5985217, 6090292, 6177069, 6159252, 6138910, 6202230, 6296395, 6287426, 5946629, 6072889, 5897664, 5988831, 6050692, 6050691, 6050751, 6050669, 6072874, 6077904, 6079020, 6090314, 6090323, 6118835, 6090293, 6077880, 6118838, 6138908, 6145129, 6100643, 6050699, 6100650, 6185706, 6225169, 6185665, 6230406, 6230392, 6202244, 6230416, 6215945, 6254016, 6255640, 6050726, 6110741, 5899032, 5988850, 6026689, 6110738, 6176395, 6276323, 6236438, 6276344, 6276322, 6013897, 6090475, 6079120, 6177048, 6177046, 6185668, 6229098, 6156282, 6026578, 6026585, 5897678, 6090301, 6145134, 5897720, 6100638, 6229076, 5898203, 5902923, 5836417, 5899048, 5931474, 5994784, 5899062, 6014751, 6118830, 6090279, 6212535, 6230391, 6225190, 6255715, 6185689, 5897748, 5994822, 6026604, 5897728, 5884880, 5931481, 5880118, 5951299, 6077886, 6090322, 6276503, 6138918, 5897712, 6090480, 6138916, 6043121, 6255904, 5931488, 5899101, 5902909, 6185655, 5897736, 6077916, 6017258, 5836434, 5836427, 5897722, 5931495, 5994826, 5931480, 5897645, 6079122, 6185699, 6079027, 6245503, 6296397, 6176386, 6260459, 6276328, 6285234, 6079131, 6142508, 6050698, 6177054, 6145125, 6215916, 6207672, 5836424, 5924643, 5931478, 6026663, 6017250, 5899027, 6050721, 6026583, 6118825, 5897745, 5899051, 5897632, 6014755, 5897671, 6026668, 6245491, 6255734, 5994791, 6090288, 6050708, 6043105, 6212539, 6245489, 6185662, 6138893, 5988845, 5994830, 5902919, 6138894, 6254013, 6254054, 5899042, 6043122, 6017253, 5899105, 5931493, 5994803, 6017270, 6050697, 6118818, 6079029, 6185701, 6225206, 6236410, 6287212, 6229069, 6255888, 6311380, 5898205, 5897754, 6234503, 6072806, 6185671, 5897749, 5994792, 6077599, 6255894, 6230402, 5985224, 6079134, 5897660, 5951293, 5898209, 6077923, 6090477, 5897659, 6138902, 6118817, 6177053, 6090454, 6185660, 6234476, 6229097, 6215933, 6245549, 5836422, 5897625, 5994823, 6026674, 6014767, 6043115, 5897740, 6077885, 6014753, 6090290, 6079026, 6090298, 6043109, 6177067, 6100630, 6185656, 6113538, 6110751, 6207681, 6212540, 6145142, 6234481, 6236447, 6229086, 6215911, 6234487, 6255646, 6287417, 6255649, 6245495, 6285219, 6299602, 5897718, 6177076, 5951313, 6207676, 5899033, 5994794, 6050688, 6050750, 6077924, 6077636, 6072802, 6159240, 6245544, 6072811, 6254007, 6215902, 6215940, 5994808, 6276498, 6296412, 6287423, 5994810, 5946643, 6014746, 5836415, 6017265, 5836420, 6145117, 6234496, 6185678, 6230400, 6236414, 6145132, 6236409, 6118824, 6026688, 6026675, 6177061, 5897655, 5924640, 5899119, 5897761, 5897730, 5924634, 5836438, 5898223, 6079124, 5931494, 6230405, 6276339, 6100644, 5924648, 6118844, 6281889, 6026669, 5897721, 5931489, 6072873, 6050735, 6050715, 6118826, 6176374, 6185683, 6234478, 6215912, 6254040, 6159453, 6287429, 5902916, 6304041, 6050749, 6043120, 6079018, 6090280, 6156284, 5897667, 5897758, 5899030, 6311361, 5946628, 5946647, 6072881, 6145141, 5988839, 6077898, 6072788, 6202226, 6142511, 6254030, 5898204, 5897715, 6225168, 5897739, 5880125, 5898199, 6013907, 6185674, 6013881, 6202248, 6207671, 6225211, 6236456, 6215947, 6185679, 5899056, 5897731, 5897714, 6090470, 6090297, 6072890, 6110734, 6138903, 6177066, 6177059, 6138906, 6202219, 6236429, 6138909, 6245496, 6276330, 6287218, 5946633, 6043099, 6090282, 5899037, 5951294, 6026679, 5897704, 5884881, 5931497, 5985206, 6118833, 6090459, 6177040, 6287419, 5946623, 6026579, 6142495, 6177057, 6202227, 6236416, 6077620, 6118834, 5884883, 6077897, 6079013, 6079008, 6079015, 5924647, 5924652, 5897699, 6077891, 5924653, 5924637, 6077611, 6090318, 5994828, 6118840, 6145119, 6207674, 6118827, 6225178, 6215905, 6229080, 6225173, 6254031, 6254018, 6177073, 6013882, 5924649, 6185651, 5897662, 5880493, 5897668, 5897719, 6215899, 6159251, 5880156, 6013895, 6113537, 6110747, 5899018, 6077603, 5988836, 6202204, 6245536, 6159243, 5994795, 6234474, 5994832, 6142507, 5898207, 6050727, 5946645, 6225186, 5897698, 5924641, 5951304, 5951301, 6017267, 6072797, 6077652, 5899058, 5880129, 6026582, 6207670, 6050694, 6079006, 6014763, 5899031, 6079130, 6050743, 6110753, 6013901, 6229092, 6245538, 5899054, 6090275, 6156281, 5880165, 6177042, 6176371, 6296389, 5899014, 6026593, 6090455, 6234486, 6090476, 5951310, 5931479, 6079019, 6176378, 6142506, 5897680, 6185696, 5988832, 6100651, 6215936, 5899022, 5902932, 6017248, 6176387, 6245510, 5899059, 6090299, 6100648, 6177052, 6176369, 6225197, 6260460, 6110752, 5899121, 5897717, 5897711, 5880491, 5897656, 5898219, 5946621, 5946644, 5946637, 5897672, 6043116, 6079030, 6077911, 6077910, 6110728, 6043098, 6090274, 6090285, 5946639, 6077624, 6077607, 5902928, 6118843, 6090481, 6156283, 6090462, 6176390, 6145118, 6185667, 6185685, 6142499, 6138905, 6159242, 6202200, 6229070, 6225174, 6236445, 5994836, 6215942, 6225181, 6215913, 6236444, 6215935, 6142505, 6254044, 6255903, 6254050, 6296386, 6296408, 6254047, 6276502, 6276343, 6254032, 5836430, 5897661, 5880123, 5988837, 6043107, 6072791, 5899029, 6050722, 6110743, 6177043, 5897658, 6207658, 6236454, 5988854, 5899097, 5880494, 5897735, 5951307, 5924645, 5946638, 6050672, 6090289, 6156280, 6202211, 6072796, 6254025, 5897669, 6207665, 6311379, 5897646, 5899013, 6013899, 5994831, 6177063, 6176372, 6236455, 6185684, 6225185, 6229064, 5897654, 6142502, 6090474, 5988834, 5898212, 6050673, 6245486, 6285231, 5884867, 5902929, 6072800, 6026684, 5988830, 6245537, 5931498, 5897760, 6276342, 5902930, 6245551, 6207667, 6100639, 6072803, 5899043, 6043104, 6118823, 5836418, 5899023, 5899104, 6013900, 6026667, 6026664, 6043125, 6090319, 6013884, 6142752, 6177049, 6185666, 6185705, 6050681, 6202213, 6207655, 6225171, 6229063, 6299595, 6285243, 6236411, 6202201, 6287416, 6090460, 6118820, 5951308, 5898208, 5899057, 5897723, 5897652, 5899111, 5951311, 5946624, 5931500, 5988833, 6014762, 5931477, 6077625, 6159440, 6077608, 5899044, 6234497, 6229079, 6245490, 6254051, 6245554, 6260451, 6299582, 6245488, 6026680, 6079109, 6142748, 6177064, 6281906, 5899096, 5897643, 5899035, 6077614, 6159439, 6276331, 5897727, 5897726, 5897755, 5899106, 5951296, 5951298, 6026670, 5897753, 6276497, 6276332, 5951319, 5985207, 6090284, 6077654, 6159445, 6017256, 5897644, 6017261, 6090320, 6255895, 6215922, 5899020, 5899019, 6026580, 6100646, 6050731, 6110736, 6110745, 6142509, 6245525, 6276333, 6276508, 6050674, 5931487, 5924646, 5931484, 6050690, 6077642, 6017268, 6142749, 6177071, 6177055, 5898221, 6215937, 6255721, 6255887, 6276486, 6260457, 6207677, 6276326, 6304044, 5884872, 6072814, 6079107, 6077633, 6255716, 6013903, 6013898, 6138897, 5880492, 6050728, 6234490, 6234499, 6245524, 5924638, 5897637, 5994827, 6017260, 6229093, 6245517, 6215941, 6255637, 6177038, 5899050, 5897627, 5899063, 5899123, 6118837, 5931482, 6142753, 6100637, 6110740, 6050710, 6090465, 6225192, 6229067, 6255717, 6229062, 6254017, 6255885, 6236437, 6255902, 6260462, 6255642, 6255889, 6245560, 6281895, 6260465, 6311360, 6255724, 6287412, 6260464, 6138896, 6285224, 5897697, 5897762, 6077895, 6185702, 6236407, 6050741, 5899049, 5899114, 5994829, 6229091, 6236431, 6013892, 6276337, 6202233, 6013904, 5988853, 6159461, 6079115, 5836433, 5899034, 6185704, 5898225, 5902911, 6050729, 6079111, 6090302, 6254027, 6050693, 6077613, 6050678, 5994835, 5897635, 6100627, 6177075, 5899024, 5897734, 6072884, 6050696, 6177065, 6176388, 5924636, 5931485, 6072863, 5897743, 6225209, 6276491, 6281898, 5899052, 5985205, 5985219, 5897732, 6013889, 6017246, 6043112, 6079110, 6050679, 6090281, 6090291, 6090464, 6050718, 6100629, 6118839, 6176389, 6043101, 6185692, 6202215, 6230407, 6255738, 6026594, 6255650, 6260446, 6254053, 6285229, 6285223, 5924642, 6185688, 5899046, 5946640, 5988840, 6090326, 6276493, 6026677, 5946620, 6050707, 6296393, 5880159, 6079118, 5897648, 6142497, 5988829, 5899099, 6215903, 6072872, 5897725, 6185693, 5897676, 6014747, 6276325, 5924633, 5899103, 5899021, 5951314, 5994833, 5899015, 6014752, 6014743, 6050733, 6077646, 6138907, 6013906, 6077617, 5897629, 6077922, 6110732, 6229081, 6110744, 6296410, 6077644, 5902924, 6043114, 6281897, 5897663, 5899098, 5897683, 5897666, 5985213, 6026686, 6026602, 6043108, 5988846, 6090305, 5931492, 6072870, 6079125, 6077619, 6110739, 6050740, 6138901, 6177050, 6090300, 6225208, 5951295, 6159444, 6156279, 6245522, 6245520, 6245494, 6245506, 6255898, 6225210, 6254035, 6255896, 6229095, 6215948, 6145130, 6050682, 6245531, 6185682, 6255730, 5931486, 6072812, 6225175, 6245557, 6255661, 6177044, 6138911, 6050709, 6245507, 5898215, 6207679, 6026687, 5884882, 6026597, 6026600, 6090456, 6229088, 6245514, 6255718, 6014759, 6013890, 6090461, 6159262, 5902921, 6026596, 6050706, 6142504, 6299603, 5897733, 5897750, 6077605, 6281911, 5897653, 5898210, 5899036, 5988828, 6014749, 5951320, 6100633, 6014745, 6077623, 6118842, 6185670, 6236424, 6145115, 6276320, 6177045, 6299580, 6299607, 6260463, 5931501, 5988852, 6100624, 6090328, 6145137, 5836429, 6215949, 5880488, 5985221, 6026584, 6013888, 6245504, 6255878, 5897696, 5865265, 5931490, 6017249, 6079126, 6079132, 6090307, 6072804, 6118829, 6156286, 6090458, 6177078, 6234477, 6254021, 6225203, 6255884, 6260468, 5899113, 6230411, 5884877, 5899017, 5897674, 6077640, 5985222, 6026682, 5884868, 5897673, 5902927, 5985211, 5985218, 6072793, 6090472, 6050671, 5931475, 6245543, 6100634, 6050748, 5897709, 6014750, 5988849, 6050712, 6079024, 6043124, 6050730, 5946632, 6142498, 5880487, 5897764, 6050705, 5994821, 6072871, 6072795, 6072882, 6026586, 6026581, 6159441, 6177062, 6145127, 6236435, 6299588, 6072787, 6072807, 6100645, 6138895, 6090277, 5902931, 5951306, 5951317, 5985212, 6014757, 6100626, 6100635, 6159442, 6159455, 6159457, 6159459, 6176373, 6176375, 6176377, 6176383, 6185654, 6202196, 6202198, 6202197, 6202235, 6234484, 6234492, 6234495, 6234498, 6260444, 6260447, 6260449, 6260455, 6260456, 6254020, 6254028, 6276495, 6276499, 6276501, 6276504, 6276505, 5880148, 5884859, 5880157, 5880167, 5897681, 5897687, 5897689, 5897694, 5897688, 5897737, 5897751, 5897759, 5899118, 5897710, 5897657, 5902908, 5902913, 5902915, 5988847, 5946626, 5946630, 5946635, 5946642, 5946646, 5994799, 6050704, 6077629, 6077637, 6077648, 6077653, 6077906, 6090310, 6090324, 6118821, 6118832, 6118836, 6118841, 6176370, 6176382, 6225172, 6225176, 6225183, 6225188, 6225191, 6225194, 6225205, 6177047, 6177060, 6177070, 6255882, 6296387, 6296388, 6296394, 6296399, 6296406, 6296398, 6296407, 5880116, 5880158, 5880161, 5880134, 5880135, 5880507, 5897746, 5836421, 5884862, 5884863, 5884864, 5884865, 5880163, 5884873, 5884856, 5884871, 5880155, 5884874, 5880143, 5884858, 5880117, 5836431, 5880151, 5880506, 5884857, 5836439, 5880490, 5988842, 5985226, 5897686, 5897626, 6079009, 6079117, 5884860, 5880166, 5988835, 5880147, 5836435, 5884875, 5897630, 6050734, 5897631, 6077882, 5902907, 5902912, 6281914, 6026676, 5899039, 5897684, 5897690, 5897763, 5897651, 5902926, 6013905, 6072794, 6072798, 6185698, 6185672, 6207673, 6207680, 6287193, 5880126, 5836437, 5899065, 6072876, 6072879, 6100641, 6142493, 6142500, 6142503, 6142515, 6142516, 6142517, 6236419, 6255713, 6255720, 6276490, 6276496, 5897647, 6090313, 6077621, 6077626, 6245556, 5899061, 5988851, 5985223, 5985228, 5994798, 5994801, 6212538, 6236423, 5994785, 5898217, 5836419, 5994839, 6017247, 5897692, 6245540, 5880149, 5951303, 6079021, 6177058, 6072864, 6090295, 6017259, 6202239, 6285222, 6207660, 6202225, 6013902, 5985204, 5994787, 6090286, 6110733, 6202238, 5902925, 6110748, 6100625, 6176376, 5898214, 5898201, 6177041, 5994807, 6077900, 6013896, 5880486, 6245541, 6026672, 6090276, 5988848, 5897707, 5899108, 5902918, 5899038, 5902917, 5988827, 6077918, 5994837, 6072880, 6202241, 5985210, 6245515, 6202199, 6079119, 6110730, 6159258, 6236452, 6255662, 5898218, 6079128, 6077876, 6177074, 6079121, 6202202, 5985209, 6090296, 6159246, 6072790, 6100631, 6255733, 6079114, 6017264, 6090278, 6185691, 5951318, 6017271, 5897752, 5899040, 5902905, 6176380, 5897685, 6185677, 6202203, 6236440, 6245519, 5898220, 6236417, 5880138, 5897700, 5898206, 6236425, 6090327, 6202221, 5897729, 5994814, 5902914, 6077635, 5897706, 5994812, 6072808, 6185700, 6207664, 6245539, 5880120, 6072809, 5951297, 5898226, 5899060, 5985203, 5985208, 6110746, 6176381, 5988844, 6245542, 6202246, 6287202, 6202231, 5898202, 6202234, 6230415, 6230413, 6281899, 5880489, 5898213, 6077649, 6077634, 6090321, 6202237, 6090287, 6245530, 6245564, 6254011, 6287205, 6287203, 5899122, 5880150, 5880483, 5985227, 5880131, 5880484, 5880130, 5899117, 5880162, 6077917, 5880133, 5880121, 5880124, 5880113, 5880127, 5880497, 5880504, 6230389, 5880144, 5880164, 5880482, 5880502, 5880132, 6077902, 6245550, 6090294, 6079025, 5897693, 5897708, 5880145, 5897705, 5898200, 5899047, 6207656, 6207663, 6110754, 6255739, 6245523, 6100647, 5899045, 5880114, 5994819, 5994820, 5994790, 5994834, 6079127, 6072805, 6077628, 6072878, 6072789, 4420961, 4087199, 4314048, 4108437, 4839760, 5243161, 5045521, 4975004, 5015581, 5189522, 4916625, 5076057, 5137715, 5243175, 4933658, 4933671, 5032704, 5199950, 5137700, 5225536, 4729565, 4848912, 4952467, 4827295, 5189534, 5189517, 5015609, 5161848, 4827306, 5111579, 4827314, 4890063, 5032700, 5189488, 4699921, 5053351, 5225692, 5155949, 6255652, 6401358, 6357323, 6488734, S112584, 6185680, 6255647, 6281904, 6371909, 6413313, 6452198, 6488782, 6255651, 6506984, 6245526, 6506986, 6507030, 6540862, 6202247, 6281891, 6281908, 6401361, 6452174, 6452150, 6387481, S112585, 6230417, 6202193, 6245528, 6452148, 6142513, 6488787, 6254008, 6245532, 6072866, 6387489, 6281901, 6507003, 6255643, 6236443, 5924632, 6229084, 6202195, 6452119, 6454905, 6207675, 6077872, 6255641, 6534575, 6281893, 6077878, 6077602, 6371922, 6413297, 6401349, 6413255, 6452104, 6387483, 6050744, 6287210, 6401357, 6452190, 6142490, 6340505, 6281902, 6357349, 6401239, 6371324, 6452138, 6401228, 6207678, 6401334, 5899055, 6452103, 6215925, 6515508, 6371630, 6345466, 6215921, 6255659, 6245487, 6236420, 6401244, 6413238, 6245561, 6452139, 6454922, 6511530, 6515491, 6452182, 6371606, S112567, 6505824, 6505842, 6540871, 6540835, 6540861, S112572, S112998, S116143, 6285218, 6285217, 6202243, 6540864, S116101, 6079108, 6026599, 6357346, 6255892, 6354163, 6354142, 6276345, 6299594, 6357356, 6371321, 6401332, 6401328, 6413289, 6413338, 6387500, 6505840, 6507023, 6357764, 6387512, S113082, S112574, S116106, 6296392, 6357770, 6371608, 6413314, 6452191, 6454902, 6511498, 6454918, 5946627, 6276319, 6215944, 6401272, 6413270, 6505846, 6413246, 6505801, 6515518, 6340507, 6255660, 6340501, 6207662, 6357751, 6357355, 6371326, 6371317, 6357364, 6401263, 6401261, 6413275, 6357369, 6413271, 6413265, 6413340, 6413329, 6413330, 6488757, 6488736, 6488783, 6488737, 6452142, 6452137, 6452140, 6511516, 6534588, 6534589, 6507027, 6515517, S112586, S113000, S112995, 6281910, 6340489, 6255658, 6254045, 6401346, 6401320, 6413269, 6454914, 6540863, 6229085, 6345481, 6506982, 6534566, 6507009, 6050670, 5931476, 6255890, 6540891, 6413317, 6534605, 6079031, 6077919, S113005, 6340504, 6515520, 6340493, 6345454, 6401266, 6540858, 6401359, 6079017, 6255726, S112902, 6488784, 6357344, 6215938, 6481864, 6534568, 5946625, 6534599, 6287414, 6215929, 6299601, 6454925, 6215909, 6488773, 6299590, 6357308, 6515482, 6488769, 6215939, 6413233, 6387496, 6507012, 6540877, 6215924, 6090309, 6481856, 6454904, 6511492, 6113536, 6202229, 6050677, 6371916, 6296400, 6255719, 6077925, 6511512, 6515519, 6511493, 6090283, 6225200, 6511501, 6345471, 6050687, 6260453, 6304057, 6357845, 6371311, 5994816, 6354144, 6345465, 6371931, 6401330, 6401306, 6401267, 6454916, 6511490, 6371611, 6413321, 6452100, 6505815, 6013887, 6540874, 6511502, 6534573, 6540898, 6481861, 6285221, 6245480, 6077601, 6345479, 6299598, 6077638, 6079112, 6229060, 6357298, 6185694, 6255655, 6371624, 6287211, 6260466, 6345433, 6371327, 6357747, 6452143, 6145128, 5994786, 6026591, 6345477, 6304039, 6371605, 6299587, 6505829, 6285220, 6452126, 6090467, 6506990, 6413287, 6534554, 6299599, 6311364, 6215920, 6371923, 6452141, 6452114, 6050752, 6413232, 6236432, 6357840, 6454927, 6488747, 6454912, 6413244, 6401274, 6401251, S112870, 5994805, 6346479, 6338280, 6156285, 6276327, 6505799, 6515523, 6345473, 6511535, 6540904, S116128, 6387521, 6413276, 6488765, 6534563, 6357753, 6142756, 6234475, 6185653, 6507005, 6413332, 6354153, 6255879, 6345472, 6488746, 6505802, 6511527, 6202209, 6515526, 6276506, 6387490, 6413296, 6276483, 6026666, 6481848, 6245500, 6540834, 6540806, 6511541, 6138912, 6185675, 6202228, 6371316, 6215926, 6505819, 6540813, 6515499, 6515483, 6357362, 6452107, 6276489, 6357351, 6413236, 6401279, 6452133, 6413274, 6287196, 6488760, 6540842, 6540878, 6050745, S116099, 6346486, 6401339, 6454907, 6234480, 6413302, 6488771, 6287413, 6311366, 6285237, 6357741, 6345446, 6371612, 6229096, 6357750, 6371906, 6505837, 6488750, 6540916, 6287427, 6245512, 6413234, 6185676, 6346476, 6511504, 6505838, 6079116, 6452128, 6338282, 6304060, 6357720, 6387488, 6387526, 6401331, 6413253, 6357728, 6413349, 6413284, 6413337, 6488778, 6506999, 6505818, 6481850, 5994788, 6540893, 6357309, 6026587, 6346498, 6077604, 6413310, 6202236, 6159437, 6287424, 6304040, 6506988, 6177039, 6413341, 6340506, 6255638, 6401268, 6357368, 6452135, 6304043, 6401345, 6454919, 6540830, 6505817, 6540853, 6413254, 6511538, 6413273, 6401329, S116105, 6287431, 6090471, 6340496, 6177051, 6387510, 6357752, 6357762, 6345480, 6287422, 6215900, 6077887, 6260452, 6285241, 6401368, 6296409, 6387522, 6488781, 6387476, 6505794, 6540831, 6488762, 6506993, 6401348, 6215930, 6077879, 6413298, 6413279, 6540865, 6340500, 6159253, 5946636, 6401284, 6345449, 6185690, 6255725, 6110750, 6013885, 6505830, 6296390, 6387485, 6296404, 6254049, 6304053, 6299583, 6401227, 6401254, 6387509, 6346478, 6534570, 6540807, 6452164, 6452168, S113055, 6202208, 6311367, 6452109, 6138904, 6371625, 6357363, 6357350, 6285240, 6255723, 6225207, 6401278, 6401250, 6371917, 6401269, 6413261, 6401281, 6413243, 6454910, 6454906, 6540814, 6357361, 6454911, 6255891, 6413306, 6488767, 6357366, 6540876, 6276340, 6481855, 6371318, 6357827, 6159458, 6401353, 6452171, 6371631, 6454928, 6507026, 6515510, 6345459, 6488772, 5897634, 5884879, 6515525, 6387518, 6401256, 6488776, 6534594, 6540841, 6255893, 6311371, 6401252, 6371607, 6515496, 6287435, 6401246, 6229087, 6413371, 6229083, 6254048, 6346500, 6254036, 6387529, 6357370, 6413242, 6515500, 6511542, 6481845, 6338293, 6079011, 6452101, 6515490, 6017262, 6050717, 6506977, 6534596, 6511491, 6488761, 6413292, S113054, 6285233, 6401336, 6299581, 6371616, 6296413, 6254057, 6540906, 6234483, 6255900, 5897744, 6371907, 6345457, 6387482, 6413259, 6340492, 6454901, 6452134, 6401238, 6540880, 6401240, S116140, 6072801, 6387477, 6014768, 6357738, 6311377, 6540816, 6311370, 6260458, 6413345, 6505828, 6357312, 6371312, S112550, 6357844, 6207669, 6488741, 6413362, 6254043, 6515480, 6511505, 6245484, 6515505, 6229068, 6234493, 6234485, 6413252, 6245518, 6176384, 6276488, 6357843, 6413262, 6401360, 6077609, 6185687, 6534562, 6357347, 6511510, 6534549, 6481849, 6488748, 6345444, 6371619, 6371926, 6236421, 6413339, 6276341, 6340497, 6142750, 6401355, 6488745, 6026592, 6401356, 6340486, 6357829, 6245548, 6077615, 6413333, 6481865, 6357760, 6296405, 6215918, 6371303, 6043100, 6534601, 6534595, 5898222, 6371925, 6387498, S116139, 6215923, 5899110, 6413263, 6452157, 6454903, 6296411, 6340487, 6338303, 6357718, 6401366, 6401364, 6185658, 6299597, 6118822, 6340485, 6345461, 6345484, 6345437, 6345452, 6345468, 6340483, 6311372, 6357367, 6357359, 6357739, 6340484, 6371609, 6401264, 6340491, 6371627, 6090466, 6401333, 6413319, 6371306, 6401294, 6340503, 6371305, 6452201, 6506996, 6511522, 6540885, 6371301, 6540866, 6506987, S113039, 6540827, S113048, S112878, 6077627, 6245534, 6245562, 6488752, 6488779, 6285235, 6014758, 6236448, 6540867, 6311378, 6371315, 6340488, 6299585, 6260448, 6454923, 6515488, 6481852, 6507006, 6287199, 6100628, 6540847, 6515494, 6043111, 6260469, 6354149, 6413325, 6234505, 6345478, 6185673, 6026590, 6507022, 6534552, 6413268, 6357358, 6245516, 6229078, 6506985, 6505822, 6260470, 6276492, 6540837, 6340502, 6354150, 6202251, 5897716, 5897713, 6229071, 6299589, 6515492, 6176394, 5836428, 6413281, 6245505, 6387480, 5951300, 6276336, 6287418, 6357841, 6138917, 6371626, 6401230, 6311357, 6413358, 6452187, 6488744, 6505827, 6488785, 6534603, 6452197, 6540848, 6285225, 6401365, 6540839, 6159263, S113021, S112573, S113003, S113061, 6159260, 6304059, 6387524, 6304034, 6401273, 6413305, 6260471, 6401262, 6505796, 6515531, S112879, 6338289, 6338307, 6245485, 6452160, S113049, 6346494, 6357754, 6215917, 6401352, 6401280, 6505834, 6511524, 6506997, 6540903, 6488764, 6540917, S112575, 5994797, 6357821, 6413315, 6452127, 6505804, 6357348, 6511526, 6505847, S112564, 6236418, 6215943, 6413303, 6236449, 6245552, 6413304, 6515486, 6304042, 6225179, 6304047, 6159249, 6304038, 6229077, 6357743, 6371928, 6371924, 6387501, 6387499, 6401255, 6077651, 5899025, 6357846, 6401300, 6452118, 6357314, 6338292, 6338283, 6401277, 6452144, 6452098, 6338294, 6540875, 6254023, 6413250, S112582, S116144, 6371908, 6202224, 6255644, 6145138, 6304054, 6534569, 6285236, 6159264, 6371629, 6413363, 6413336, 6505831, 6540899, 6534555, 6401308, 5898211, 6254024, 6229075, 6506979, 6401354, 6287436, 6245493, 6017252, 6338291, 6245558, 6357724, 6276338, 6413299, S112581, 6340494, 6338296, 6346481, 6354158, 6354162, 6357758, 6413239, 6338288, 6452131, 6357828, 6507007, 6534561, 6534581, 6507000, 6540833, 6540910, 6540887, 6534583, 6413350, 6481846, S112562, S112896, 6304048, S112877, S113015, S116109, 6345438, 6185697, 6145122, 6371302, S113029, 6215904, 6413309, 6245502, 6050746, 6505803, 6340490, 6387475, 6345439, 6281890, 6236430, 6287430, 6401283, 6371613, 6534564, 6488770, 6505812, 6505792, 6452176, 6254056, 6401319, 6077600, 6050684, 6079129, 6387511, 6413312, 6413256, 6534559, 6245511, 6340498, 6340482, 6255880, 6452149, 6043103, 6159248, 6255897, 6345474, 6202218, 6413266, 6515514, 6488774, 6276507, 6185657, 6346495, 6245482, 6229082, S112563, S113052, S112890, S113089, 5897702, 6050719, 6354141, 6413260, 5994800, S112886, 6245498, 6299591, 6013894, 6296396, 6357297, 6345451, 6357746, 6387493, 6413286, 6401289, 6413316, 6452153, 6255881, 6507018, 6515484, 6488755, 6311381, 6507002, 6540840, S112872, 6515513, S113045, S113063, S116096, S116133, 6413295, 6090317, 6507004, 6202232, 5924635, 6299605, 6345463, 6345458, 6357834, 6371910, 6346484, 6285230, 6371915, 6357371, 6371921, 6357365, 6338285, 6413264, 6540913, 6540914, S112568, S112900, 6338304, 6299584, 6387516, 6540825, 6540859, 6401325, 5994817, 5897642, 5897633, 6488751, 6079023, 6515530, 6276329, 6254041, 6540873, 6401242, 6338298, 6338284, 6338301, 6413331, 6225167, 6413326, 6506991, 6515532, 6515535, 6452163, 6357313, 6079033, 6387486, 6287208, 6285228, 6338302, 6285239, 6255731, 6338305, 6254042, 6311359, 6276318, 6338297, 6311356, 6401324, 6401321, 6338290, 6401326, 6354152, 6452158, 6452105, 6357765, 6401245, 6413277, 6387491, 6506998, 6534600, 6515495, 6511489, 6488788, 6311376, 6454908, 6413367, 6254058, 6371325, 6345448, S112885, 6506995, S112577, S112871, 6050695, S112996, S113016, 6202212, 5836432, 6401363, 6254014, 6110737, 6401370, 6338281, 6387494, 6413280, 6387478, 6287433, 6346496, 6452121, 6488749, 6413301, 6354151, S116092, 6159265, 6159259, 6299586, 6215907, 6245481, 6338295, 6138898, 6225189, 6357315, 6371617, 6345440, 6387507, 6488743, 6296402, 6304056, 6255735, 6507024, 6229073, 6236422, 6488733, 6401297, 6401322, 6345460, 6255729, 6413293, 6452129, 6505813, 6507015, 6452130, 6515507, 6515509, S112873, 6304050, 6287411, 6338287, 6338286, 6357847, 6371300, 6371920, 6401237, 6413365, 6413290, 6401337, 6413327, 6413322, 6505835, 6505826, 6506994, 6507017, 6506978, 6488775, 6507020, 6540832, S116112, 6540882, 6311355, 6287428, 6145133, 6229072, 6276494, 6511507, 6507013, S113056, 6276334, 6255722, 6507029, 6177056, 6234500, 6050738, 6202217, 6357822, 6413283, 6488786, 6299600, 6357729, S112551, 6304045, 6345467, S112892, 6371323, 6413318, 6413361, 6488766, 6534584, 6299596, 6215906, 6304035, 6357835, 6371319, 6159462, 6357826, 6371604, S113038, 6371905, 6505844, 6534604, 6488758, 6452112, S116094, S112904, 6215910, 6234488, 6357848, 6145121, 6387515, 6488763, S112884, 6276485, 6452145, 6534577, 6511534, 6225202, 6296403, 6357744, 6142501, 6236451, 6481854, 6371610, 6413368, 6452177, 6452200, 6511518, 6505808, 6540907, 6357306, 6357310, 6357311, 6357317, 6357320, 6357322, 6357749, 6357756, 6357766, 6357767, 6413291, 6413311, 6413354, 6413355, 6413357, 6413364, 6413366, 6413369, 6515504, 6515506, 6515511, 6534560, 6534572, 6534580, 6185703, 6357296, 6413257, 6401265, 6401340, 6401335, 6452169, 6452170, 6452179, 6452183, 6452188, 6013891, 6346488, 6452199, 6452204, 6506989, 6387479, 6401350, 6401351, 6507001, 6506976, S113041, S112571, S113008, S113009, S113010, S113011, S113012, S113013, S113019, S113020, S113022, S113023, S113025, S113026, S113027, S113028, S113030, S113031, S113032, S113033, S113034, S113035, S113036, S113037, S116116, 6401231, 6401253, 6413282, 6357726, 6357732, 6507011, 6507025, 6202214, 6413285, 6050686, 6515501, 5880509, 6287192, 6452116, 5994809, 6299592, 6346487, 6357301, 6236434, 6515497, 6100632, 6287219, 6452161, 6357761, 6245565, 6304055, 6452178, 6540845, 6452111, 6077920, 6452113, 6511517, 6511523, 6452159, 6452122, 6511529, 6413288, 6413353, 6540854, 6230395, 6285242, 6413344, 6515524, 6371911, 6287437, 6346489, 6452154, 6401372, 6371913, 6357823, 6401299, 6452165, 6452123, 6540857, 6452175, 6299604, 6357318, 6072810, 6511519, 6236442, 6511499, 6255657, 6345434, 6043110, 6159452, 6026671, 6043102, 6145120, 6287213, 6357838, 6072877, 6387492, 6287198, 6287209, 6401292, 6401291, 6452156, 6255737, 6540817, 6481869, S113057, 6505823, 6452166, 6452117, 6507010, 6159255, 6287217, 6245483, 6077632, 6413356, 6079123, 6287200, 6017257, 6311368, 6357733, 6357721, 6202240, 6281888, 6401276, 6357723, 6285238, 6357832, 6413267, 5994806, 6287206, 6401293, 6304052, 6452185, 6357715, 6387523, S112903, 6026673, 6452172, 6401275, 6515479, 6413352, 6511528, S113069, 6401302, 6511496, 6540883, 6452203, 6505798, 6357357, 6357839, 6505814, 6540879, 6357321, 6255905, 6255645, 6401367, 6357719, 6540815, 6413258, 6401347, 6346497, 6540890, 6534548, 5951316, S113058, 6230401, 6507021, 6236453, 6281913, 6387505, 6413360, 6285227, 6017244, 6287434, 6346491, 5994796, 6215914, 6304046, 6357833, 6540808, S113072, 6387497, 6285232, 6357316, 6230394, 6245497, 6511503, 6255736, 6387487, 6401287, 6481868, 6304051, 6505805, S113074, 6357824, 6371912, 6401296, 6413307, 6511543, 6540869, 6534565, 6452124, 6185652, 6357722, 6401232, 6511515, 6287421, S113006, 6281905, 6254038, 6230403, 6072883, 6255636, 6371929, 6452115, 6401286, 6511531, 6507019, 6540819, 6452162, 6281912, 6311373, 6357763, 6371307, 6202216, 6401258, 6413351, 6255656, 6357836, S112553, 6401369, S113059, 6287216, 6304037, 6371918, 6413347, 6413348, 6481847, 6511508, 6540821, 6507014, 6534593, S113066, 6452195, 6481851, 6077912, 6245559, 6452202, 6287215, 6079007, 6511536, 6540826, 6511511, 6540810, 6401236, 6540836, 5897738, 5880499, 6357727, 6357737, 6345445, 6254033, 6110755, 6540884, 6077883, 6077888, 6345453, 5880112, 6225177, 5985216, 6534590, 6276335, 6230410, 6255883, 6345483, 6110731, 5880500, 5899115, 5899112, 6230419, 6254019, 6387520, 6345482, 6254039, 6371320, 6534550, S112554, 6452194, 6481843, 6245521, 6481860, S113073, 6287204, 6215908, 6254015, 6511521, 6540818, 6540812, 6511497, S113076, 6287197, 6077896, 6357299, 6357830, 6413346, 6413370, 6505820, 6534607, 6505816, 6584067, S119886, 6476990, 6590867, S109552, 6604054, 6458445, 6458438, 6590866, 6458306, 6476984, 6471537, 6458369, S122293, S109565, 6590841, S109729, S120359, 6604065, 6604039, 6656503, S109831, S120471, S122288, 6428153, 6458368, 6477079, 6458440, S109560, 6648733, S120373, S109840, S109702, 6428121, 6438750, 6477014, 6458371, 6561202, 6561191, 6575855, S109118, S109735, S120348, S122287, 6458426, 6544002, 6656493, 6477023, S109566, 6575845, 6584064, S109835, 6458446, 6428157, 6544006, 6590842, 6604067, 6471529, 6467537, 6477009, 6477008, 6477063, 6590878, 6477070, 6561180, 6476997, 6575842, 6604066, 6604050, 6604064, 6561195, 6458428, 6561203, S109717, 6584062, 6544007, S120464, 6604047, S119896, S120487, 6458319, S120335, 6563339, 6458430, 6471525, 6590845, 6477004, 6428118, 6477030, 6561192, 6656486, 6584065, 6648715, 6438763, 6458380, S109722, 6543988, 6458321, 6470149, 6604073, 6584051, 6648712, S109549, S109734, 6458379, 6458393, 6648717, 6590876, 6428143, 6470164, 6471527, 6561197, 6590862, S109730, S122269, S109704, 6561193, 6458434, 6477003, 6584053, 6471538, 6471530, 6428146, 6477013, 6563348, 6458316, 6458382, 6458328, 6656494, 6604091, 6428135, 6428128, 6471520, 6428140, S109829, 6477017, 6590874, 6476993, 6575846, 6458367, 6471524, 6438754, 6575859.6561185, S109823, S122280, 6438768, 6470163, 6477067, 6543998, S109115, 6648714, 6544001, 6604075, S109807, 6458310, 6428142, 6476988, 6471521, 6458372, 6470161, 6590856, 6476979, S109820, 6561189, S120377, 6438761, 6471542, 6476994, 6471531, 6563350, 6590871, 6563340, S122264, 6467522, 6458370, 6543999, 6590843, 6458443, 6477075, 6590852, 6604052, 6656489, S109713, 6590885, S109844, 6458315, 6604086, S120475, S120454, 6477024, 6458374, S109574, S109141, 6458383, 6470155, S122237, 6648739, 6561183, 6543996, 6428131, 6471543, 6561177, 6561198, 6428155, 6458447, 6438759, 6458390, 6438749, 6458373, 6544000, 6575854, 6584087, 6590863, 6458311, 6563352, S109547, 6428133, 6477064, 6477022, 6470160, S109568, 6604089, 6458325, S120498, 6438747, 6648720, 6467516, 6471528, 6458435, 6563343, 6575839, 6458431, 6458427, 6543987, 6543997, 6438764, 6575861, 6584070, 6584069, 6590881, 6428145, 6458444, 6590847, 6584076, 6471534, 6471519, 6561182, 6544003, S120469, S126297, S109571, 6428134, 6563359, S109819, S119887, S109133, 6476989, 6438762, 6590835, S122268, 6477012, 6561194, S109732, S109559, 6590858, S109572, 6428111, 6428125, 6477000, 6477073, 6476995, 6476999, 6656491, 6575850, 6575848, S109126, 6438766, S109125, S109709, S109827, S122281, S120369, 6477077, 6428120, 6470166, 6590855, 6477018, 6470145, S109127, 6428138, S122252, 6470162, 6476986, 6477071, 6590851, 6477080, 6561179, 6561178, S122272, 6471526, 6428154, 6575858, 6428152, 6563337, 6648741, S109124, 6590886, 6575830, S120486, 6471532, 6458381, 6467521, 6575860, 6590846, 6604038, 6584095, 6584066, 6604063, 6575856, S120350, 6477001, 6604088, 6471540, 6428136, 6544004, S109128, 6544005, 6428150, 6470147, S109738, 6584055, S109842, S120349, S120323, 6458322, 6476998, S109577, S122238, 6477066, S109833, 6470167, 6458318, 6656490, S109712, 6584082, 6470146, 6476983, 6477062, 6604057, 6656497, S109832, 6656487, S120376, S120324, S120354, 6470156, 6470153, 6438765, 6590860, 6648737, S109718, 6470159, 6438758, S109548, 6428103, 6438773, 6477028, 6477076, 6477069, 6584091, S109554, S109561, 6458387, 6575865, 6428104, S109562, S109117, 6477005, 6575832, 6590873, 6477068, 6561184, 6476985, S109131, 6458437, S109558, 6584068, 6575840, 6648719, 6428156, 6467541, 6458442, 6477015, 6458366, 6458375, 6476976, 6458449, 6477027, 6471522, 6476981, 6575828, 6648728, 6575833, 6584084, S109541, 6590849, 6584057, S109544, S119892, 6471523, S109556, S130343, S109711, S120451, 6561181, S122251, S120453, S120470, 6476991, S109821, 6590869, S109828, 6590882, 6428151, 6575866, S120446, 6428112, 6470151, 6428117, 6458441, 6476980, 6458330, 6458329, 6648738, 6561201, 6575847, 6428102, 6604090, 6477002, 6543995, 6575831, 6575826, 6575836, 6584086, 6584090, 6584092, 6584094, 6648721, 6648722, 6648723, 6648724, 6648726, 6648730, 6648731, 6648734, 6648736, S122253, S122258, S122267, 6428105, 6428113, 6458320, 6590879, 6604053, 6604081, S109543, 6604041, S120344, S130346, 6458376, 6458377, 6458378, 6458384, 6458386, 6458389, 6428122, 6590884, 6648725, 6590857, S109824, 6590831, 6590839, 6477010, 6477026, S109724, 6428126, 6477019, S109120, 6590854, 6590837, 6563344, 6656501, 6575837, 6467534, 6477084, 6467532, 6467531, 6467540, 6471539, 6458432, 6467520, 6467525, S109122, 6458422, 6563364, 6590859, 6604059, 6458424, 6428116, 6471517, 6563362, 6590877, 6428127, 6471536, S109845, 6467526, S109837, 6543994, 6477029, 6477088, 6477087, 6477083, 6458423, 6563355, 6467533, 6648713, 6590872, 6604044, 6590864, 6467530, 6590836, 6604051, 6438769, 6428129, 6543989, 6563360, S126289, 6476987, 6438772, 6458323, 6604074, 6604056, 6428108, 6477021, 6477016, 6590880, 6563347, S109838, 6590875, S109726, 6477086, 6476992, 6584075, 6584073, 6471516, 6590838, S109825, 6575841, 6648735, 6476996, 6458436, 6428107, 6438752, 6477020, 6477025, S109846, 6467529, 6477011, 6477007, 6467535, 4453418, 4486052, 4415717, 4859191, 5122078, 5202972, 4727820, 4727831, 4933657, 4859177, 5224667, 4727823, 4676657, 4890051, 5151837, 5215381, 4952480, 5111570, 5172636, 5202966, 5111581, 4892135, 4925234, 4729549, 4952478, 5225531, S113083, 6401373, 5946641, S116120, S113080, 6245545, 6229074, 6229089, S116091, S116089, 6236439, 6340499, 6287194, S112888, 6229065, S113065, 6026577, S112557, 6077639, 6534574, S113070, 6540870, 5924644, S116093, 5994793, 6287438, 6454921, 6345435, 6050723, S112580, S116100, 6387504, 6159436, 6304036, 6354160, 6346483, 6013886, 5897636, 6229061, 6505825, S113004, S113047, 5836423, 6159267, 6540911, 6207661, 6413342, 6540886, S113064, 6287432, S116102, 6540895, 6511539, S113062, 6534586, 6507008, S112887, 6515493, 6142514, 6215919, S113088, 6354157, 6138915, 6401362, 6072875, 6090469, 6260467, 5899102, 6505845, 6014764, S112883, 6534598, S113050, 6481844, 5897724, 6401247, 5899053, 6159446, S113002, 6014766, 6159443, S113051, 6138900, 6371615, 6540838, S116129, S113046, S112566, 6534602, 6452110, 6401301, 6534576, S113081, S112569, S112558, 6452209, 6014770, S116141, S116136, 6488754, 6454917, S113053, 6452102, 6185695, 6255639, 6488753, 6540908, 6515528, 6534558, 6159456, 6515485, 6215898, 6281915, S116107, 6413251, 6236408, 6540850, S113085, 6138913, 6511506, 6540915, S112891, 6357745, 6142747, 6202194, S116146, S112555, S113018, 6159261, 5836426, 6452184, 6515522, 6354161, 6515534, 6387495, 6515487, 6371622, S116088, S116110, S116111, 6540901, 6304049, 6413334, 6159448, 6159447, 6540851, 6079016, 6540860, S113040, S116098, 5931499, 6159451, 5897670, 6511537, S112556, 6413308, 5994815, 6354143, S116123, S113042, 6371914, 6050683, S112578, S112994, 6534597, 6401259, 6511540, 6142491, 6540909, 5899011, 6357353, 6357352, 5880152, 6077631, 6534579, 6401288, S112889, S116108, 6236450, 6138899, 6255714, 5994825, 6276484, 6481863, 6540855, 6515516, 5899100, 6354159, 6401260, 5946622, 6345442, S116138, 6540902, 5899041, 6452208, 6401270, 6079012, 6079028, S112560, S113024, S113017, S113014, 6452167, 6050725, 5897756, 6511533, 6236415, 6505843, S112880, S113087, 6254012, S116113, 6371919, 6338306, 6452205, 6506983, 6452120, 6401338, 6413328, 6346493, 6371322, 6285216, S116137, S116095, 6287415, 6026588, 6506992, 6296401, 6488777, 6100640, 6371313, 6371310, 6413343, 6357354, S116130, 6534553, 6287425, 6534557, 6515527, 6311374, 6534578, 6345470, 6276324, 6077630, 6505809, 6013883, 6296391, 5884866, 5994789, S112997, 6371628, 6145136, 6276321, 6401303, S112893, S116125, 6515533, S113001, 6481858, 6505841, S112895, S112897, 6159254, 6014765, 5985220, 6230409, 6505836, 6354148, 6345462, 6515498, 6159266, 6401307, 6452151, 6176385, 6234494, 6540872, 6079010, 6515536, S113043, 6540912, 6452152, S113090, S116114, 6540809, 6245501, 6234504, 6138914, 6505839, 6207682, 6515512, 6452136, S116097, 6371932, 6401241, S112583, S112570, 6371614, 6371309, 6387506, 6245546, 6050703, S112882, 6540823, 6511509, 6515529, 6276500, 6050685, 6357360, 6185669, 6230397, 6515515, 6387514, S116145, 6050680, 6401249, 6357734, 6357307, 6515537, 6413300, 6505810, 6540828, 6077622, 6515481, S112901, 6236412, 6050675, 6110742, 6202222, 6090325, 6090315, S116131, 6452181, 6534587, 6534585, 6540822, 6311362, 6401371, 6254037, 6515489, S116126, 6401257, S113007, 6507028, S113077, 6452155, 6540852, 6255648, 6176379, 6110749, 6287207, S116121, 6540820, S113060, S116142, 6511494, S116132, 6285226, S112876, 6540844, 6540900, 6236441, S116135, 6534592, 6357300, 6540846, 6511525, S116127, 6505821, 6540881, 6387508, 6345475, 6345441, 6215931, 5880154, 6254034, 6357735, 5880508, 5880122, 6254029, 6357740, 5880496, 5880153, 6254010, 6488735, 6452108, 6357717, 6540894, 6345476, 6345456, 6077909, 6413278, 6357716, 6452189, 6481862, 6481857, S112875, 6481853, 6476977, S109553, S130366, S202956, S219308, S219270, S246718, S130359, S270206, S270224, S246701, S277894, S278824, 6458308, S119912, S202932, 6563338, S296488, S202957, S259127, S278781, S221577, S221562, 6590850, 6563357, S109539, S122291, S240400, 6656498, S213247, S246959, S246706, S246715, S120374, S237017, S229339, S229328, S280614, S202965, S130407, S119894, S237018, S280425, S280671, S271709, S219278, S278826, S130564, S219309, S246956, S296489, S280426, S246702, S270226, S286373, S119897, S109705, S213241, S282023, 6590868, S229411, S280459, S259145, 6604048, S130419, S130585, S130413, S109728, S219279, S203172, S246961, S269239, S219261, S274910, S273355, S277888, S278819, S286369, S289632, S296524, S270214, S289628, S296776, S280644, S130424, S130594, S130376, S130450, S213521, S130399, S229436, S219254, S130430, S248193, S248207, S269155, S269167, S270177, S277880, S280608, S246965, S269215, S289609, S280628, S282225, S280621, S246626, S280627, S280419, S213222, S223096, S120450, S202908, S246955, S270112, S213520, S130418, 6584052, S120347, S130371, S211091, S219511, S246954, S120346, S237016, S266238, S269220, S266248, S269237, S270195, S259281, S282043, S289598, S280427, S280414, S294541, 6428148, S246654, S130382, S219491, S282033, S289332, S289604, S229415, S219287, S246727, S282025, S119891, S130560, S221559, S109706, S280673, S120488, S120439, S120444, S211087, S130381, S203194, S120490, S130377, S219506, S219504, S130372, S219532, S219493, S240387, S223098, S219533, S223093, S246709, S229432, S248085, S229319, S246630, S248205, S248188, S266252, S122265, S259130, S259283, S248177, S269181, S259131, S282037, S270181, S289623, S289618, S289616, S280405, S294496, S296544, S296543, S294464, S130548, S109809, S202934, S213237, S223094, S229011, S223145, S246960, S246653, S229388, S259140, S270229, 6590865, S270189, S278820, S246947, S294474, S282218, S296551, S280605, S269173, 6604043, S280413, S283767, S120468, S219548, S269236, S130428, S281894, S289607, S130453, S203209, 6656499, S130402, S219544, S219776, S219760, S223113, S229338, S246710, S223091, S248213, S259301, S270196, S269184, S270218, S259302, S282216, 6470150, S294544, S122254, 6584050, S271745, S240386, S280613, S109723, S122290, S219498, S213234, S120452, S229322, S229336, S280423, S202940, S213245, S248148, S278809, S248206, S213254, S109134, S126281, S270170, S219538, S266237, 6458332, S219522, S120447, S120457, 6575834, S253909, S259146, S213217, S223149, S283780, S221567, S202912, S213244, S223126, S229417, S246713, S248138, S120362, S270209, S278782, S119901, S278784, S296495, S278791, S280635, S296552, S122275, S294511, S130385, S284211, S122273, S246966, S246944, S278854, S294559, S130569, S221553, S213529, S203167, 6656502, S221539, S270188, 6458309, S203157, S122248, S229329, S259266, S229313, S289343, S259278, S280659, S266246, S278823, S281872, S219272, 6561188, S269190, S219280, S269166, S219288, S240401, S109578, S202946, 6584079, S130389, S202942, S120443, S221525, S240385, S240399, S246964, S259270, S259148, S202945, S246946, S269238, S202906, S280594, S259161, S282029, S271740, S130396, S253927, S280422, S278849, S283768, S289627, S294558, S120463, S219264, S219546, S248084, 6428144, S283762, S237026, S282217, S130596, S229384, S229327, S246730, S280587, S202952, S109563, S296538, S283776, S248151, S266232, S253918, S246942, S284189, S130553, S120476, S219311, S130571, S248204, S219770, S266218, S248179, S286366, 6477081, S273350, S289593, S278779, S294552, S213253, S219530, S213242, S229407, S280646, S277881, S278842, S219260, S294553, S286367, S281868, S296491, 6575829, S130350, S202936, S213509, S219249, S223144, S283787, S130545, 6590848, S280412, S219496, S283770, S229321, S122245, S253921, S282028, S122270, S130456, 6604045, S130572, S246650, S248195, S246721, S296498, S280630, S281914, S284194, S246635, S223108, 6575844, S130598, S130601, S120321, S219310, S130452, S219306, S229435, S223125, S229434, S120322, S259282, S266250, S250668, S270227, S259280, S266220, S250660, S270171, S271721, S277891, S281869, S219549, S280650, S294501, S246646, S250675, S278806, S229409, S280447, S283785, S240391, S203203, S211086, S280415, S130354, S213246, S130600, S253925, S219540, S271724, S130602, S270199, S289617, S280664, S280666, S223122, S130565, S202938, S219524, S223099, S259273, S281890, S296778, S219523, S109731, S211093, S130398, S203193, S250695, S219307, S219518, S248183, S120345, S277876, S271758, S109573, S130558, S202922, S130387, S202933, S219262, S240398, S246624, S246733, S269172, S271737, S269242, S109703, S286472, S296794, S213540, S130373, S202907, S229016, S221532, S248189, S259133, S270216, S229349, S269216, S289606, S271687, S281893, S280616, S294487, S248190, S274912, S278783, S130361, S120361, S130567, S221568, S130364, S213510, S219545, S213537, S203202, S221528, S228999, S229326, S250685, S229404, S248122, S219267, S259265, S219777, S237025, S250672, S266249, S271743, S282036, S280643, S294475, S250691, S294557, S284196, S283774, S294462, S270194, S277887, S213533, S284212, S221569, S223115, S240388, S294554, S130390, S270113, S253924, S130417, S266244, S281898, S289631, S203200, S109814, S203206, S213544, S219509, S213543, S109839, 6590870, S250659, S250665, S253913, S266219, S266230, S284202, S286358, S289603, S278800, S266236, S229363, S221579, S109715, S219534, S280662, S246640, S219259, S246734, S248201, S271715, S280636, S277886, 6428110, S130403, S219771, S120360, S223109, S109143, S296786, 6575863, S130355, S213251, 6584058, S221550, S122241, S246963, S203189, S120462, S246712, S248196, S248214, S278816, S278829, S280612, S122239, S120455, S213243, 6656504, S130573, S223121, S213248, S223130, S109546, S223139, 6428132, S219276, S229331, S246717, S253926, S248182, S248211, S269245, S270203, S269170, S259152, S229356, S281887, S296519, S289342, S229428, 6575864, S248076, 6656496, S248080, S277877, S270163, S259135, S248185, 6458314, S122240, S130552, S126270, S122250, S202960, S126267, S223092, S213519, S219292, S229360, S229340, 6561186, S270215, S248155, S270105, S271692, S282238, S270106, S246625, S282214, S259158, S269174, 6604080, S266227, S266253, S280611, 6561200, S278836, S269214, S280599, S280442, S280631, 6458331, S130588, S270225, S294533, S130575, 6604040, S270178, S122242, S130586, S270180, S246722, S120333, S248215, S270202, S109557, S271717, S271733, S246951, S280651, S294495, S219298, S130550, S130388, S122256, S130409, S109114, S229335, S250683, S248071, S266263, S246656, 6438756, S248197, S269219, S271701, 6477072, S271746, S294549, 6604058, S229337, S246651, S289595, S282040, S296528, S213538, 6471533, S202913, S120358, S126272, 6584081, S211095, S109139, S282022, S280660, S248125, S250693, S280450, S228998, S130421, S130411, S130566, S269232, 6584060, S219527, S253920, S109550, S130580, S237023, S223107, S122286, S248068, S271693, S270129, S281901, S278804, S278850, S282231, S120367, S271690, 6458312, 6648740, S294524, S130584, S259139, S237030, S296516, 6438751, S130349, S219263, S246636, S223103, S130546, S280623, S246629, S296747, S294478, S284200, S296732, S282031, S120368, S122277, 6471535, S109570, S219536, S213545, S253923, S271747, S283766, S280598, S280409, S274906, S120378, S202950, S296500, S219303, S213240, S246962, S280649, S221540, S213539, S266261, S248150, S130351, 6477074, S130401, S221560, 6563345, S213530, S221580, S246729, S250674, S259269, S219269, S271716, S237021, S278805, S286365, S296518, S296729, S278793, 6458385, S229047, S202935, S219505, S219551, 6428119, S223100, S130603, S223133, S248180, S229357, S253915, S223119, S266259, S270166, S281876, S286364, S270192, S284209, S280610, S266270, S280657, S296540, S246619, S294482, S296492, S294525, S237034, 6604071, S120459, S259128, S246652, S294508, S229044, S120458, S221546, 6476978, S120496, S120325, S202947, S219517, S213218, S246958, S223090, S266276, S259160, S271759, S280606, S281895, S219281, S120356, S229007, S266231, S219547, S130423, 6575835, S219537, S130437, S223146, S229052, S248192, S229437, S280655, S269221, S281886, S109135, S122249, S130408, 6575843, S109545, S280443, S296780, S281885, S296549, S219494, S130368, S248120, S259124, S282235, S284213, S221548, S296525, S284197, S286354, S278855, S283772, S289605, S122255, S130443, S219492, S122278, S122295, S203176, S221552, S211097, S221556, S223095, S223101, S223102, S229395, S223097, S229362, S221545, S237029, S248077, 6428124, S229366, S246707, S246953, S246723, S250680, S283796, S283763, S246704, 6604061, S271700, 6584071, S278856, S278810, S296506, S283769, S237019, S281874, S296532, S213526, S213531, S109830, S223131, S119888, S119883, S259263, S202955, S130420, S219521, S122244, 6561176, S246699, S221555, S259276, S282044, S109818, S109542, S248209, S277883, 6458326, S219490, S280607, S271694, S296507, S280672, S221571, S294461, 6428137, S219265, S294469, S282038, S296766, S120499, S130341, S246940, 6563346, S219302, S229056, S219489, S119905, S120331, 6584056, S259126, S130556, S246731, S126268, S289614, S109116, S119893, S221570, S223104, S250688, S266254, S270162, S248095, S289615, S286363, S284185, S109132, 6561187, S130383, S130597, S219541, S246736, S126279, S130455, S203171, S120478, S202920, S211088, S130393, S223106, S219297, S229393, S246735, S248199, S250667, S250687, S237020, S259123, S219300, S282027, S296785, S266262, S281879, S296537, S229311, S229043, 6590832, S282241, S273352, S289608, S240395, S122285, S280431, S250684, S246737, 6604068, S219301, S219503, S223118, S246627, S259129, S269175, S286362, S286481, 6458307, S120365, S120370, S221547, S229018, S223147, S229021, S229410, S246939, S259291, S266266, S281875, S282035, S271744, S120479, S120341, S130449, S109119, S202909, S219277, S223132, S229307, S250682, S246732, S266228, S280617, S283789, S296499, S120375, S219512, S130587, S266234, S271742, S203164, S280592, S130561, S229334, S278796, S269189, S294531, S269188, 6604084, S269226, S202951, S122289, S211081, S246631, S259136, S266268, S266247, S269218, S280648, S278785, S286467, S294465, S294556, S266225, S294555, 6584059, S286479, S296763, S109834, S130352, S203158, S130555, S219767, S229053, S246719, S219246, S237022, S248067, S248128, S250673, S229308, S270208, S269162, S271757, S296531, S296513, S266239, S283783, S296788, S109555, S109736, 6575853, S280403, S130378, S280434, 6656485, S223105, S246655, S278792, S126280, S202921, S246948, S130427, 6604085, S294502, S130562, S122266, S120445, S130445, S223134, S130551, S213512, S219520, S229346, S219510, S237031, S246638, S259275, S211082, S246941, S248145, S248141, S248202, S246632, S223142, S109737, S266223, S248210, S248194, S270217, S269165, S280603, S282240, S269157, S278780, S269156, S278817, S283764, S278852, S278812, 6590853, S284190, 6471518, S270212, S250679, S289596, S119902, S280407, S248129, S296509, S284201, S122261, S246943, S259284, S109720, S266241, S286478, S269234, S280584, S278815, S109826, S269223, S223110, S280591, 6477006, S130397, S130439, S229020, S280595, S240396, S282041, S130451, S120440, S219296, S269177, S221573, S229426, S223120, S246726, S219255, S269158, S246725, S130559, S109551, S130414, S130415, S278786, S259151, 6428139, S109569, S109575, S246952, S248178, S270219, S270108, S259274, S269222, S122276, S283761, S120492, S213252, S246728, S259287, S120474, S213250, S130447, 6458391, S229045, S122274, S250678, S122243, S221530, S280609, S273349, S219305, S294476, S248081, S130363, S120353, S202941, S202937, S237024, S120371, S281911, S253922, S294504, S246957, S120327, S250681, S281878, S130454, S130345, S229424, S221542, S250686, S229392, S280590, S219252, S221544, S120355, S240393, S122260, 6648727, S221566, S240397, S219514, S259122, S248123, S270204, S266265, S270193, S281877, S130442, S120351, S248154, S278851, S281873, S130549, S280629, S269161, S130440, S270130, S122279, S120352, S130431, S122257, S130425, S130576, S109130, S130547, S109138, S122282, S130570, S203160, S203159, S203174, S219535, S221538, S219304, S219273, S229341, S203166, S229008, S109137, S246634, S248136, S203175, S250669, S259134, S259144, S248137, S270223, S270183, S248121, S250670, S270167, S284191, S271725, S280618, S271703, S278790, S284204, S289619, S283794, S120337, S203162, S213523, S229332, S248127, S270190, S270151, S294551, S213518, S120372, S221533, S229015, S253908, S248086, S266243, S120340, S211083, S120456, S278813, S278802, S296756, S289622, S246705, S246724, S266226, S269248, S289630, S294497, S281882, S289629, S202918, S280601, 6561196, S122292, S271696, S280661, S213238, S296760, S270213, S296790, S120465, S277893, S269217, S270128, S289594, S229414, S282045, S120379, S119904, S130360, S211084, S213522, S221537, 6584054, S221541, S223114, S229325, S237028, S259159, S248187, S213249, S250692, S126269, S203163, S219553, S229361, S130410, S120357, S246628, S270158, S219513, 6428130, S278811, S271697, 6561190, S269230, S122262, S296745, S213535, S219245, S246637, S266217, S296530, 6428115, S219516, S223124, S246700, S211092, 6458433, S120343, S219508, S219550, S229386, S246633, S229394, S248118, S130392, S246945, S219542, S266233, S278801, S266229, S277879, S266255, S270175, S280593, S280604, S296490, S283784, S296535, S221549, S221543, S281888, S296746, S240389, S246645, S269231, S211085, S221565, S246621, S229330, 6604076, S229385, S289345, S296775, S109567, S248198, S109576, 6438767, S130554, S202910, S202919, S202939, S202961, S211089, S213507, S213511, S219299, S229382, S229383, S229389, S229405, S229412, S253916, S259157, S269164, S269229, S270116, S270118, S271727, S271736, S277895, S269163, S269182, S281892, S296497, S280633, S282222, S282230, S282236, S281891, S283777, S283786, S270200, S229344, S229347, S271730, S270117, S271711, S280626, S283778, S130577, S130579, S130599, S130441, S266260, S284215, S284216, S284217, S284181, S284182, S284183, S284184, S284219, S284220, S296547, S296554, S294529, S294530, S130578, S248132, S270187, S270201, S280652, S280653, S213229, S213534, S213536, S219497, S221529, S221531, S221558, S246657, S203161, S203165, S229023, S229040, S229048, S278838, S278843, S219531, 6656492, S259149, S248131, S284186, S269235, S130583, S246967, S246968, S229399, S281905, S120460, S120330, S271754, S219529, S219525, S281883, S109836, S213515, S219773, S219285, S259268, S271706, S270221, S271751, S296761, S296783, S286371, S284193, S219774, S213525, S202962, S271734, S271689, S296752, S259300, S229316, S271714, S271748, S248082, S282212, S286466, S219495, S120363, S229030, S229017, S229402, S270207, S266224, S270210, S119899, S286464, S278833, S294481, S130404, S219258, S221535, S259293, S280458, S280418, S269224, S126294, S270211, S281910, 6575827, 6563351, S223123, 6477089, S229358, S229038, S259155, S273354, S269244, S119907, S271728, S126298, S202926, S213517, S229025, S266222, S229323, 6563363, S283791, S229420, S229034, S266240, S119895, 6467515, S213528, S213256, S211094, S229353, S130434, S109817, S120442, 6604087, S130379, S219501, S213508, S248078, S213514, S266272, S130386, S229032, S229037, S270185, S229027, S274915, S246644, S270114, S266275, S269179, S266271, S278853, S289602, S282051, S282227, S286477, S281880, S286470, S126273, S229324, S248200, S120332, S126278, S223129, S130436, S223111, S202916, S219271, S120441, S120320, S286465, S282229, S271756, S219275, S119906, S229351, S280620, 6467518, S229035, S280460, S120482, S130391, S130435, S223112, S202917, S203170, 6458313, S221574, S229010, S248079, S246642, S266256, S270184, S213231, S282228, S271755, S289339, S120319, S246708, S213513, S122294, 6656495, S266273, 6604079, 6648716, S286359, S119884, S229364, S248090, S229028, S120461, S126288, S130563, S213220, S271712, S273356, S229354, S281903, S294485, S229039, S120473, S271731, S248153, S120366, S219284, S213527, S294528, S296734, S120494, S278827, 6563361, S223135, S213516, S120485, S213236, S219775, S120364, S248208, S273359, S278788, S278832, S280437, S211096, S270222, S259143, S270220, S280452, S294470, S120481, S229397, S219257, S271702, S296757, S229350, S281900, S289601, S280439, S130433, S228997, S120329, S286475, S278834, S271753, S271719, 6590834, S246641, S248156, S219519, S266274, S130446, S248133, S229400, S229396, S202923, S130557, S130406, S248186, S259285, S274914, S273358, S271749, S120338, S126276, S126302, S130592, S219499, S202915, S120449, S202914, S221564, S223141, S229012, S219283, S248083, S280596, S120328, S277892, S282226, S271720, 6563358, S126290, S120483, S126274, S130369, S213219, S213239, S213524, S229003, S248089, S259142, S126299, S271729, S271732, S274909, S274905, S274911, S109123, S122247, S278828, S130367, S229054, 6458333, S219500, S126277, S126275, S211098, S126283, S248094, S280619, S229014, S271707, S286480, S296759, S219526, S126284, S248212, S229318, S271705, S294490, S278835, S130544, 6458317, 6590861, S202949, S203182, S296774, S109145, S130370, S203173, S248139, S271723, 6656488, S130342, S202948, S203180, S130347, S203192, S203156, 6584080, S130353, S266267, S203205, S280669, S269154, S109142, S203178, S250663, S130348, 6575867, S130344, S203215, S203208, S203196, S203183, S294484, S203201, S202959, S282232, S130356, S280656, S270173, S246950, S248117, S266269, S203212, S280634, 6590833, 6476982, S126285, S229406, S229408, S120334, S269240, S229029, S271718, S120493, S229031, S229000, S120472, S229004, S126303, S296784, S109140, S202925, S219761, S213255, S270168, S109725, S280441, S281904, 6467538, S273362, S213221, S280445, S126266, 6590883, S119903, S119908, S229401, S202927, S229416, S229418, S259156, S274907, S229355, S248144, S278822, S283795, S286357, 5189540, 5215380, 4958971, 5137695, 5137696, 5122020, 5137380, 5111560, 5224675, 5252084, 4819998, 5155955, 6540824, 6245529, 6507016, 6357757, 6255654, 6281903, 6072887, 6145126, 6077643, 6254052, 6515502, 6540889, 6354139, S112576, 6245492, 6050747, 6452132, 6346490, 6357837, 6346473, 6401290, 6260461, 6481859, 6299606, 6260450, 6345436, 5899116, 6072813, 6077903, 6354156, 6079113, S116115, 6145140, 6357345, 6077908, 6260454, 6354155, 6236446, 6345455, 6534556, 6452206, 6540843, 6357319, S112898, 5931496, 6159460, 6100636, 6357842, 6014754, 6230396, 6346499, 6090316, 6413323, 6255653, 6014769, 6452186, 6245547, 6245513, 6077905, 6176396, 6401243, 6413359, 6304058, 6401309, 6354140, 6481870, 6345447, 6225199, 6540888, 6207659, 6371930, S112579, 5897640, 6276482, 6357742, 6401229, 6354147, 6159450, 6357831, 6013908, 5899064, 6452207, 6234489, 6540896, 6346474, 6050720, 6540829, 6142754, 5836416, 6488780, 6346482, S112561, 6338300, 6481866, 6505806, 6159256, S112881, 6511514, 6177072, 6017263, 6234501, 6371304, 6159241, 6346492, 6090273, 6229090, 6452196, 6511488, 6017255, S116119, 6311369, S112552, 6413240, 6142492, 6454909, 6371618, S116090, 5897741, 5836425, 6077647, S112559, S116134, 6026683, 6540892, 6505832, 6177077, 6357748, 6454915, 6254022, 6387517, 6387503, 6077881, 6215901, 5880498, 5880501, 6215934, 6345450, 6345443, 6230412, 5899107, 6357730, 5880495, 6079022, S286356, S248069, S229041, S248074, S246647, S280602, S270197, S296750, S284187, S294477, S294472, S294546, S130438, S202954, S294479, S280453, S246711, S250677, S296795, S248216, S296767, S278818, S253917, S278825, 6575838, S250661, S219256, S296487, S286370, S284199, S296527, S294505, S266245, S296781, S280421, 6470154, S273351, S109144, S237032, S269185, S284195, S281909, S280625, S229055, S229320, S283773, S284214, S278795, S289624, S294526, S280624, S246703, 6438753, S282239, 6458425, S229049, S296754, S248184, S283781, S280637, S280408, S283782, S229425, S294545, S294540, S286355, S229345, S109841, S270198, S296758, S278799, S229019, S259298, S278846, 6467519, S266235, S120489, S277878, S266221, S246720, S296771, S296769, S296545, S294498, S281881, S223143, S269211, S270107, S294550, S119882, S284207, S248140, 6648729, 6477078, S202911, S273360, S280667, S253914, S280647, S250664, S296737, S248152, S130574, S278814, S294473, S219552, S283788, S296744, S250671, 6467523, S266251, S130429, S109129, 6584085, S294483, S294463, S296733, S280654, S278807, S269176, S248147, S269186, S296770, S296789, S289621, S296791, S296730, S282039, S280449, S280665, S219251, S296486, S286473, 6458392, S296772, S296520, S296512, 6438757, 6543991, S223127, S130416, S122271, S283765, S269159, S294503, S280585, S278803, 6477031, S248191, 6604036, S294548, S296728, S271739, S277884, S296517, S282221, S294538, S269171, S294537, S281902, S250696, S296515, S296511, S296764, S259271, S240384, S213542, S294539, S120326, S213532, S219769, S281896, S259147, S281897, S266264, S283775, S229430, S259299, S296522, S281871, 6458439, S280586, S130444, S289331, S284192, S259150, S294506, S284188, S283779, S259272, S221572, S296796, S281870, S250676, S271704, S278789, S282024, S284208, 6543990, S109540, 6561199, S109811, S280632, S278837, S130400, S278840, S109147, S237035, S120495, S219515, S229433, S296510, S278848, S270228, S120342, S296505, S280404, S296768, S282220, S294532, S296792, S219555, S270179, S119898, S296777, S280639, S280424, S130412, S130374, S289620, S269168, S296793, S296773, S296514, S282042, S296782, 6477065, S221576, S289341, 6438760, S294547, S223117, S289599, S281899, S120448, S269178, S289335, S270164, S130426, S296779, S259292, S278857, S280448, S280641, S282046, 6584061, S284210, S213232, S130405, S282224, S296533, S282219, S280446, S280663, S283790, S280451, S280430, S221554, S296762, S202924, S280428, S280588, S280429, S246622, S284203, S281889, 6458327, S286468, S130422, S282047, S296755, S246623, S130595, S202953, S277882, S280658, S223136, S271713, S269213, S270169, S271726, S294499, S280411, S296494, S130394, 6543992, S271735, S296765, S283771, S280406, S277885, S274917, S223137, S120491, S240390, 6467517, S221578, S259290, S229005, S294527, S269233, S213541, S237027, S269241, S246643, S221551, S282050, S248142, S271688, S280622, S259141, S237033, 6604055, S269180, S284218, S229022, S259132, S269192, S286474, S274916, S294521, S294512, S229036, S294534, S286482, S289600, S274908, S281906, S296751, S213235, S280436, 6477085, S120477, S296787, S296748, S229002, S281884, S281908, S294514, S229026, S294480, S286476, S273353, S280435, 6604078, S280456, S289597, S229398, S294536, S294535, S296735, S229024, S289337, S280420, S221563, S248143, S282048, S229431, S269227, S223138, S213230, S296741, S296740, S270205, S269247, S278797, S284206, S294513, S294515, S126282, 6648718, S280597, S296736, S296739, S269246, S221527, S280444, S266242, S221561, S120480, S278787, S282049, S280416, S296738, S269243, S273357, S278821, S294493, S269183, S296742, S246639, S286361, S294468, S126293, S221536, S286360, S280455, S229009, S229429, S278831, S219766, S294467, S294466, S266258, S203197, S202943, S259153, S109813, S282237, S286368, S202964, S203214, S280432, S250662, S203207, S294494, S294517, S271695, S203195, S294518, S109808, S277889, S203213, S278847, S203188, S130357, S271698, S109812, S109716, S202963, S202929, 6584083, S202931, S250657, S250658, S202930, S294519, S203204, S203184, S246949, S294492, S289625, S286372, S294491, S280461, S294522, S283793, S280438, S109121, S294488, S278830, S294516, 6287201, 6077875, 6452146, 6452099, 6540897, 6142510, 6540905, 6505795, 6354154, 6079133, 6371927, 6387525, 5994838, 6072868, 6452192, 5946634, 6401285, 6515521, 6413294, 6346475, 5994811, 6534551, 6401271, 6534567, 6452106, 6254026, 6401323, 6159245, 6050702, 6345464, 6072799, 6090479, 6245499, 5902906, 6276509, 6357825, 6207666, 6159449, 6311358, 6050700, 6357759, S116122, 6354146, 6454926, 6540868, 6401341, 6401295, 5897628, 6413320, 6311363, 6413272, S113075, S113079, S113067, S112894, 6159438, 6043117, S112899, S113078, 6401342, 6299593, S113071, 6357769, 5880503, 6017251, 6357731, 6387528, 6387527, 6230408, 6481867, 6563341, S294489, S259288, S248146, S296749, S250666, S270191, S120336, 6590840, S122246, S248203, S253912, S281913, S296493, S259137, S259125, S270127, 6604042, S248070, 6604070, S248135, S270140, S219282, S296508, 6584078, 6458388, S259138, S280645, 6604062, S202905, S248130, S246620, S130380, S282032, S203169, S109733, S229042, S259289, S229427, S282030, S229310, S213223, S246714, 6458448, S250689, S120484, S271738, S109843, S283792, S282215, S259267, S277890, S280670, S278798, 6656500, S126271, S269169, S229348, 6458324, 6604049, S219294, S120380, S120339, S281912, 6575862, S229051, 6604046, S229342, S130395, S269225, S280410, S259264, 6604060, S202958, S221526, 6575852, 6428149, S282034, S229387, S109822, S219268, S282026, S248181, S294543, S202928, S248124, S270186, S271741, S274904, S294523, S219539, S229013, S271722, S213228, S213227, 6477082, S213226, S213224, S213233, S280457, S280454, S219502, S271752, S229033, S126292, S296743, S269228, S119910, S126286, S274913, S259154, S219274, S286483, S294520, S119889, S126291, S289336, S203199, S203191, 6575851, S203198, S270172, S203190, S271710, S289626, S271699, S266257, S294510, S203211, S278845, S269249, S109146, S202944, S122283, S130365, 6245566, S113044, 6236436, 6505797, 6260445, 6159244, 6050713, 5994802, 6176393, 6401304, 6387502, S113068, S271691, S280589, S223140, 6428109, 6590844, S229006, S229046, S130432, S126287, S122284, S109148, S119911, S120497, S119909, S278844, S203187;
b) Item Number 320-42-03, UDI/DI 10885862086709, Serial Numbers: 1000105, 1092802, 1000108, 1042940, 0982241, 0982238, 0982261, 0982253, 0982252, 0982255, 1000101, 0982257, 0982258, 0982248, 1042974, 0982247, 1000097, 1000100, 0982246, 0982242, 1000104, 1000106, 1000109, 1042956, 1042979, 1042980, 1042982, 1092800, 1042944, 1042945, 1042965, 1042966, 1042985, 1042943, 1092810, 1042964, 1092821, 1000099, 1042967, 1042976, 1092819, 1092786, 1092822, 1042958, 1042969, 0982260, 1042946, 1042948, 1042942, 1000096, 1000102, 1092808, 0982245, 1092792, 0982256, 1042971, 1042988, 1092806, 1042978, 1092804, 1042961, 1000095, 1000094, 1092805, 1000110, 1092799, 1092801, 1092825, 1092832, 1092830, 1042984, 1092828, 1042949, 1042954, 1042953, 1042957, 1042963, 1042970, 1042939, 1220908, 1252332, 1252306, 1252309, 1368433, 1252341, 1252331, 1220910, 1220928, 1252339, 1252316, 1220914, 1220893, 1220923, 1220917, 1220911, 1252301, 1252312, 1220918, 1220903, 1252325, 1220915, 1357625, 1220891, 1220894, 1252296, 1220922, 1220916, 1357616, 1252303, 1220887, 1220900, 1220895, 1357653, 1362230, 1362238, 1252330, 1368449, 1220925, 1252295, 1252297, 1252298, 1252313, 1252318, 1252319, 1252333, 1252335, 1252336, 1042973, 1042975, 1042950, 1042986, 1042977, 1042972, 1042955, 1042952, 1042947, 1000112, 1042983, 0982254, 1042962, 0982262, 0982240, 0982250, 0982244, 1357628, 1530550, 1252323, 1443851, 1357649, 1428628, 1476720, 1443854, 1443857, 1545972, 1545971, 1499328, 1362239, 1220919, 1443845, 1530547, 1435767, 1368441, 1428624, 1357646, 1476717, 1357636, 1443838, 1476721, 1435763, 1357620, 1435749, 1357619, 1368443, 1545975, 1362245, 1499301, 1545978, 1252315, 1252302, 1252320, 1252307, 1428642, 1499321, 1357637, 1357654, 1368431, 1220886, 1252304, 1362248, 1362237, 1220920, 1428639, 1220912, 1357622, 1357642, 1357667, 1252329, 1368437, 1443852, 1252334, 1220892, 1220889, 1252294, 1368446, 1406251, 1362229, 1368436, 1357629, 1406253, 1428626, 1428638, 1252342, 1428641, 1357640, 1589268, 1220904, 1357639, 1357644, 1499297, 1443841, 1589589, 1443844, 1499309, 1499311, 1362235, 1357655, 1406247, 1368439, 1220932, 1362247, 1357621, 1443856, 1252300, 1499307, 1545970, 1368432, 1435739, 1362228, 1499300, 1428632, 1435746, 1406248, 1362251, 1252326, 1368445, 1362236, 1362232, 1435766, 1499327, 1252337, 1368435, 1476708, 1357659, 1368430, 1220933, 1499315, 1368454, 1362240, 1435745, 1476712, 1435743, 1499299, 1443848, 1476727, 1499319, 1530540, 1530549, 1357634, 1428634, 1428635, 1435756, 1499305, 1499306, 1357664, 1499294, 1499323, 1357663, 1368444, 1357643, 1252310, 1368452, 1499317, 1220930, 1220883, 1406249, 1428621, 1220931, 1252328, 1368434, 1368440, 1357652, 1545960, 1530534, 0982243, 1042987, 0982259, 1042938, 0982251, 1000107, 1042959, 1499316, 1357641, 1357635, 1357660, 1357669, 1428625, 1499295, 1545973, 1443846, 1362227, 1530539, 1476716, 1589269, 1252291, 1357668, 1589254, 1589579, 1589261, 1545976, 1589578, 1357631, 1476723, 1357662, 1357618, 1476722, 1530544, 1589252, 1220898, 1530548, 1443843, 1545974, 1499318, 1589263, 1362233, 1357626, 1435753, 1530532, 1499325, 1589272, 1443850, 1428633, 1357656, 1252324, 1589271, 1589599, 1220902, 1443839, 1362243, 1589580, 1545962, 1362244, 1357671, 1476715, 1428627, 1545964, 1435759, 1252314, 1589245, 1589267, 1252293, 1476724, 1589258, 1589592, 1220899, 1406250, 1589605, 1589274, 1545961, 1589591, 1589584, 1252321, 1589603, 1589602, 1428640, 1443840, 1428623, 1252340, 1435741, 1362250, 1589273, 1589597, 1530529, 1589596, 1589266, 1545979, 1589586, 1589585, 1589257, 1357613, 1435757, 1589246, 1589248, 1252343, 1499322, 1530531, 1589582, 1545968, 1499296, 1252292, 1357627, 1589270, 1530533, 1368453, 1499324, 1589255, 1435742, 1357615, 1499310, 1368448, 1220909, 1220934, 1589260, 1435752, 1368450, 1368447, 1499314, 1252308, 1530535, 1530536, 1499290, 1435744, 1499293, 1589265, 1357648, 1499302, 1428637, 1530541, 1545969, 1499292, 1589590, 1589595, 1865330, 1651704, 1734558, 1734578, 1770386, 1795009, 1734575, 1651665, 1651689, 1865346, 1865319, 1865347, 1651694, 1770374, 1865323, 1770383, 1865332, 1734567, 1651683, 1770372, 1840289, 1865349, 1651676, 1734569, 1651667, 1865340, 1734561, 1651687, 1719418, 1840290, 1865344, 1770381, 1651673, 1734570, 1734559, 1795030, 1651690, 1734565, 1651666, 1840316, 1651671, 1770373, 1651697, 1772081, 1651672, 1772087, 1795338, 1734548, 1795014, 1840312, 1840317, 1651679, 1865318, 1840309, 1734573, 1795344, 1651681, 1770376, 1795001, 1651674, 1651688, 1770384, 1734579, 1770370, 1840292, 1840296, 1772078, 1772091, 1772092, 1772096, 1770368, 1770369, 1770380, 1795012, 1795026, 1795029, 1795031, 1795032, 1865307, 1865317, 1734555, 1772072, 1795329, 1795330, 1795334, 1795335, 1734552, 1734576, 1734549, 1840304, 1651692, 1795025, 1770375, 1840329, 1840333, 1772074, 1651680, 1651691, 1770367, 1840307, 1795011, 1840310, 1840323, 1000111, 1042951, 1042941, 1220921, 1220896, 1362242, 1589259, 1545963, 1530537, 1220906, 1252338, 1476719, 1476714, 1530546, 1435738, 1499303, 1252311, 1220888, 1357614, 1357617, 1476709, 1428616, 1443842, 1589601, 1589264, 1443847, 1476711, 1357658, 1435765, 1589604, 1406252, 1589251, 1499320, 1443853, 1252322, 1530553, 1368442, 1428617, 1589262, 1428636, 1589249, 1252327, 1530545, 1362241, 1357624, 1443855, 1530554, 1476725, 1589594, 1589253, 1545967, 1530551, 1589593, 1589581, 1435755, 1435751, 1530542, 1865355, 1840308, 1918244, 1734551, 1865336, 1971824, 1865321, 1918259, 1971817, 1734577, 1734553, 1918252, 1963328, 1963333, 1971806, 1865315, 1795015, 1971822, 1840306, 1918251, 1795004, 1963342, 2068299, 1840288, 1918250, 2068281, 1971805, 1795348, 1971825, 1865352, 1865314, 1865306, 1795022, 1772075, 1865324, 1918241, 1963321, 1865339, 1918245, 1971835, 2008662, 2008657, 2036881, 1918256, 1918254, 1651685, 1840319, 1734562, 2068288, 1971812, 2036889, 1840281, 2036892, 2008654, 1772079, 1918242, 1865342, 1963329, 2036874, 1971833, 2008661, 2036883, 1772097, 1963323, 1772086, 1651677, 1963341, 1865345, 1865312, 1971829, 1865341, 1651696, 1770378, 1795019, 2036890, 1795343, 1840286, 1795336, 1795339, 1734566, 2008656, 2068302, 1734563, 1865320, 1840293, 1865334, 1840318, 1651668, 1918248, 1770385, 1772090, 1918253, 1651695, 1963332, 1795005, 1963324, 1840297, 1651678, 1840300, 1651675, 1865338, 1865356, 1963322, 2008659, 2036882, 1865351, 2068293, 2068303, 1865327, 1651702, 2036878, 1772085, 2036875, 1865325, 2036886, 1918247, 2068274, 2068277, 1971799, 2008664, 2036880, 1971804, 1770388, 1795342, 1840332, 1651699, 1963336, 1795332, 2036893, 1865313, 1734550, 1971828, 1971834, 1963334, 1865350, 2068278, 1865343, 1865354, 2091679, 2152129, 2152104, 2152125, 2091686, 2091653, 2152136, 2091692, 2152138, 2180562, 2091694, 2152105, 2152108, 2091685, 2091701, 2152110, 2091657, 2152143, 2091669, 2152145, 2152126, 2091662, 2091690, 2091695, 2091654, 2152106, 1406246, 1357645, 1357661, 1362234, 1435747, 1220927, 1220897, 1428618, 1220890, 1357647, 1499289, 1428629, 1499291, 1357651, 1357633, 1357650, 1435750, 1220901, 1362249, 1545977, 1589587, 1545966, 1428643, 1545965, 1435748, 1589250, 1589247, 1840313, 1734554, 2068275, 1770387, 1971836, 1772095, 2068301, 1772088, 1651693, 1963327, 2008652, 2068276, 1795033, 1840328, 1971820, 2036891, 1865322, 1840282, 1971808, 2068284, 1795020, 2068286, 2036888, 2068292, 1971838, 2008660, 1734557, 1651698, 2008648, 2036884, 2008649, 1734560, 2008640, 2036879, 1918249, 1795337, 1772080, 1865348, 1963325, 1865308, 2036877, 2068290, 1772083, 1918260, 2008655, 2036885, 1865309, 1971810, 1971813, 1795027, 1963331, 1963343, 1734564, 2068283, 1795341, 1795008, 1795007, 1840283, 1772098, 2008643, 1865305, 1772093, 1918243, 1772084, 1651670, 2068279, 2008651, 1971803, 1963335, 1734571, 1865331, 1795035, 1840315, 2068280, 2068289, 2068305, 1719422, 1963337, 1651700, 1772094, 1840330, 1840320, 2068296, 1840326, 2008644, 2068298, 1963339, 2284679, 2224548, 2224546, 2241348, 2472129, 2426333, 2241333, 2180543, 2152133, 2241360, 2351678, 2180548, 2152148, 2342612, 2224555, 2284650, 2351687, 2180547, 2426317, 2091659, 2284638, 2180558, 2152120, 2224545, 2091687, 2241355, 2284657, 2180556, 2284640, 2426288, 2351659, 2426309, 2472140, 2091681, 2091683, 2180560, 2224554, 2351699, 2091674, 2091665, 2152115, 2180551, 2180555, 2426336, 2284639, 2091652, 2426331, 2091688, 2284664, 2152130, 2472164, 2224556, 2241364, 2284670, 2224531, 2224530, 2284637, 2351684, 2426308, 2472135, 2513029, 2241335, 2351702, 2342616, 2241352, 2351669, 2152122, 2180557, 2284658, 2091704, 2224539, 2224551, 2152117, 2284649, 2091702, 2351664, 2351696, 2426302, 2426334, 2152146, 2284663, 2284644, 2513013, 2152134, 2091678, 2180550, 2152127, 2224529, 2152141, 2091696, 2224559, 2224558, 2091668, 2152135, 2342613, 2180559, 2091680, 2152144, 2224547, 2091698, 2241356, 2284630, 2241343, 2241332, 2426326, 2284680, 2426322, 2284661, 2426306, 2351705, 2426323, 2180553, 2152116, 2241344, 2152118, 2180545, 2284666, 2426310, 2284660, 2152102, 2091673, 2241366, 2284635, 2241331, 2091691, 2284641, 2351679, 2180546, 2091651, 2091693, 2152111, 2224533, 2284678, 2224550, 2241347, 2342615, 2426316, 2351676, 2152100, 2152121, 2241349, 2241361, 2241365, 2284648, 2284655, 2284654, 2426315, 2351673, 2351682, 2351693, 2284643, 2342624, 2351675, 2351658, 2426303, 2426311, 2426328, 2426296, 2426305, 2351661, 2351681, 2342625, 2426321, 2470267, 2470268, 2513015, 2513022, 2091670, 2551444, 2551450, 2472168, 2351680, 2224538, 2152139, 2241346, 2342614, 2284682, 2351695, 2241363, 2180552, 2091675, 2224540, 2152132, 2152107, 2152112, 2152140, 2472166, 2152131, 2152147, 2241341, 2351668, 1368451, 1362231, 1435740, 1499304, 1428620, 1435764, 1476718, 1589600, 1589598, 1357657, 1589583, 1499298, 1220929, 1435760, 1368438, 1357666, 1499308, 1499312, 1840321, 1770377, 1840280, 1772077, 1971818, 1971826, 1795028, 1734572, 2008663, 1840287, 1840303, 1971827, 1651701, 1971811, 2008646, 1795034, 1651684, 1840311, 1840295, 1865333, 2008645, 1734556, 1918255, 1840284, 2008642, 1734568, 1840291, 1795345, 1795331, 2068291, 2068282, 2068285, 1772076, 2008650, 1772089, 1770371, 1918258, 2342617, 2284668, 2587249, 2472157, 2587262, 2550700, 2472163, 2587256, 2587250, 2587268, 2513010, 2426285, 2426307, 2351703, 2342622, 2472159, 2351694, 2513023, 2590737, 2551447, 2472169, 2224560, 2351688, 2610481, 2284659, 2513011, 2590724, 2550708, 2091689, 2470271, 2351707, 2224552, 2342618, 2091700, 2550698, 2284675, 2241342, 2590734, 2472128, 2551439, 2551436, 2091664, 2472145, 2472161, 2152119, 2470270, 2610450, 2551455, 2284652, 2241359, 2513014, 2472170, 2550692, 2472139, 2284669, 2091677, 2224553, 2284653, 2590733, 2472143, 2610443, 2351698, 2284656, 2513016, 2351691, 2426327, 2550697, 2550710, 2610461, 2590718, 2610458, 2224537, 2351685, 2610479, 2472156, 2091656, 2472165, 2610483, 2224528, 2472131, 2152113, 2284651, 2513024, 2426299, 2551440, 2551460, 2513026, 2152128, 2284671, 2551461, 2610469, 2426325, 2472144, 2513009, 2241358, 2513027, 2351663, 2513033, 2513006, 2590721, 2351660, 2426301, 2550713, 2550699, 2342619, 2241345, 2426282, 2284634, 2351686, 2351692, 2284631, 2091655, 2610456, 2426293, 2284647, 2224532, 2472167, 2610467, 2610463, 2426298, 2472130, 2610444, 2513031, 2284681, 2610482, 2426287, 2610470, 2550691, 2551464, 2180544, 2472152, 2241340, 2470273, 2472151, 2426335, 2351690, 2284676, 2513008, 2284672, 2152142, 2241336, 2590720, 2610454, 2513035, 2284667, 2590729, 2284683, 2513020, 2284677, 2152114, 2284642, 2513028, 2152109, 2551443, 2610459, 2513037, 2426318, 2426324, 2513036, 2550694, 2351666, 2426290, 2091661, 2426320, 2513034, 2550690, 2472154, 2587258, 2551442, 2551463, 2551441, 2610434, 2610436, 2610441, 2610447, 2342611, 2351665, 2284673, 2426304, 2610451, 2587265, 2472147, 2587251, 2587260, 2587261, 2587263, 2426332, 2472141, 2224543, 2610452, 2091697, 2551438, 2091667, 2091684, 2241351, 2610476, 2513007, 2610446, 2551453, 2610474, 2587271, 2788609, 2720692, 2748131, 2637878, 2647315, 2720700, 2720684, 2720699, 2788607, 2647298, 2860547, 2647307, 2726072, 2748155, 2720709, 2637897, 2637862, 2647295, 2748128, 2726054, 2860579, 2748117, 2647296, 2647301, 2720697, 2788625, 2720710, 2637882, 2637863, 2637866, 2637864, 2788614, 2823867, 2637875, 2720714, 2748134, 2748123, 2788612, 2637870, 2720673, 2748122, 2860561, 2637899, 2720683, 2647287, 2637887, 2748127, 2637898, 2637884, 2748156, 2720690, 2637865, 2748160, 2647330, 2788599, 2647286, 2823871, 2647312, 2860563, 2647320, 2748144, 2726066, 2823851, 2823883, 2823875, 2647290, 2788629, 2720708, 2823840, 2647302, 2637871, 2788620, 2788623, 2647322, 2748146, 2637869, 2637868, 2647324, 2637906, 2637880, 2637883, 2637889, 2637895, 2647283, 2647323, 2647329, 2647331, 2637873, 2637881, 2637886, 2637891, 2720696, 2720716, 2647319, 2647325, 2647288, 2647299, 2823872, 2823868, 2647313, 2720670, 1589577, 1499313, 1865353, 1772082, 1971807, 1971819, 1971814, 2068308, 2068300, 1770382, 1971832, 1651686, 2008658, 1865335, 1840331, 1734574, 1918246, 1840302, 1795021, 2036876, 1865310, 2068294, 1865316, 1840314, 1795010, 1651682, 1719417, 1772099, 2036887, 1795018, 1840301, 1971837, 2587269, 2610437, 2472136, 2426291, 2284633, 2551459, 2284645, 2342626, 2590715, 2351704, 2351706, 2550706, 2610453, 2241354, 2426314, 2426313, 2610466, 2224549, 2152137, 2470269, 2513030, 2426330, 2241338, 2342609, 2472155, 2224535, 2426329, 2091705, 2224536, 2426295, 2224542, 2610440, 2550704, 2513025, 2284662, 2550705, 2610472, 2610478, 2342623, 2590719, 2472133, 2610455, 2590717, 2551449, 2590731, 2241337, 2426284, 2610460, 2472153, 2180542, 2426300, 2610480, 2610475, 2152101, 2472132, 2513038, 2610484, 2551437, 2426289, 2426319, 2610471, 2513019, 2472146, 2590732, 2472148, 2550711, 2551452, 2426297, 2610473, 2551457, 2610439, 2610457, 2587253, 2351672, 2472149, 2470272, 2550712, 2587255, 2180554, 2587259, 2610468, 2152103, 2587254, 2152124, 2351697, 2513021, 2550695, 2550703, 2351700, 2587264, 2587252, 2788635, 2931743, 2788593, 2637876, 2823869, 2823865, 2905865, 2931717, 2637896, 2928438, 2905852, 2860537, 2860555, 2860564, 2860578, 2720702, 3667987, 2823862, 3740822, 3633440, 2860545, 2823855, 2928432, 2931724, 2726052, 3534741, 2720687, 2720715, 2788626, 2823859, 2720686, 2931722, 2905877, 3633442, 3667992, 2928446, 3633454, 2647311, 2637902, 2905869, 2931740, 2905867, 3633441, 2928458, 2726059, 2860577, 3579096, 3534734, 2637907, 2823835, 3667965, 2931746, 2931737, 3645395, 2860560, 2726048, 2905857, 3645400, 2823838, 3579074, 3534730, 2931751, 3534710, 2823845, 3667960, 2748157, 3613498, 2720705, 2647289, 2905854, 3633457, 3667961, 3775167, 2726063, 2823836, 3633445, 3579064, 3645406, 2860546, 3534733, 3633448, 3633460, 2905890, 2748137, 2928436, 2928463, 3645397, 2720703, 2928477, 2931736, 3534717, 3579052, 3775147, 2860534, 2647328, 2931719, 2928430, 3534721, 3613502, 3633439, 3579061, 2905874, 3667954, 2931745, 2647285, 3645396, 3534732, 2748143, 2748133, 2637909, 2928441, 2860562, 2748153, 3645413, 2905872, 2748126, 3579071, 2748158, 3579088, 2931729, 2748139, 2905889, 3534728, 2720688, 2860538, 2788608, 2860535, 2905883, 2823846, 2905858, 3534703, 3613500, 3667979, 2647293, 3645392, 2905860, 2905897, 3579076, 3740818, 2726069, 2823860, 2788618, 2720701, 2928476, 2905896, 2931752, 2928433, 3534718, 2788621, 2647321, 3667997, 3667980, 2931725, 3579056, 3775150, 2905848, 2788631, 2647297, 2726060, 2928475, 2788632, 3633453, 2720671, 2647291, 3667990, 2928445, 3667993, 2726065, 2647294, 2823884, 2931735, 2905888, 2788604, 3579087, 2931730, 3645390, 3579078, 3579069, 3579067, 2823879, 3579090, 2931713, 3667948, 3645403, 3534711, 2748118, 2905882, 2860549, 2905868, 2788628, 2720682, 2788605, 2788602, 2823861, 3534696, 2720706, 2788606, 2860574, 2823866, 3534699, 3667966, 2928459, 2931710, 2748154, 3740809, 2931723, 3645409, 2637901, 2931706, 2726074, 2905859, 2647335, 3667994, 2720689, 2726056, 2748132, 2905891, 2726064, 3534714, 2928444, 3534742, 2788603, 3534738, 3667953, 2788630, 2823856, 2928451, 2823842, 2748138, 2726051, 2637894, 2931708, 2637903, 3667975, 2647303, 3534727, 2905892, 2905886, 3579098, 3579079, 3579077, 2931707, 2647300, 2788610, 2726057, 2748164, 2905861, 2928461, 3633451, 2748159, 3534723, 3633444, 2931739, 3534719, 3534739, 3633438, 3633452, 2931715, 2823839, 2637861, 2823881, 3633446, 2788596, 2637872, 3667978, 3667950, 3740830, 2788601, 2823849, 2720675, 2720680, 2860531, 2720712, 2860544, 2647308, 2647316, 2720694, 2720718, 2860556, 2860559, 2860571, 2928447, 3740805, 3740826, 3534725, 3740835, 2647309, 2931734, 3534740, 2931705, 2931753, 2928435, 2931718, 2647310, 2931741, 2748149, 2905881, 2720681, 2720717, 2720713, 2647314, 2647332, 2788597, 3579063, 3667991, 3740814, 2788627, 2931749, 1865311, 1918257, 1963338, 1971801, 1795006, 2008647, 1971802, 2068306, 1840327, 2068304, 1795333, 2068307, 1770379, 1719420, 2342629, 2472137, 2550702, 2513012, 2550709, 2241357, 2550707, 2241350, 2550701, 2590722, 2513032, 2590738, 2241367, 2590723, 2551458, 2551456, 2590727, 2091658, 2241334, 2091666, 2426294, 2351683, 2091703, 2091672, 2224541, 2351662, 2550693, 2513018, 2351667, 2610445, 2610464, 2590730, 2590726, 2472162, 2152123, 2610433, 2342620, 2587247, 2610432, 2587266, 2351671, 2587248, 2587272, 2351677, 2351689, 2241339, 2284632, 2342610, 2472150, 2637905, 3579084, 3740798, 3775161, 3740836, 3740791, 3740790, 3579057, 2905875, 2928470, 3534731, 2823863, 3775181, 3667964, 3579058, 2637879, 2860541, 2928462, 3775164, 3775165, 2931731, 2637904, 2823837, 2720678, 3740817, 2860530, 3645414, 3534704, 3633443, 3740823, 2726062, 2748115, 3667995, 2788611, 3775145, 2748121, 3740795, 3579065, 2931709, 3633455, 3633458, 2788600, 3667967, 3667996, 3775178, 2905850, 3775168, 3645405, 3775153, 2931733, 2748140, 3775133, 2726070, 3579051, 3740789, 3579089, 3740804, 2823858, 3775141, 2823870, 3667949, 3613501, 3579081, 2928443, 3645393, 2823874, 3775173, 2726067, 3579082, 2905879, 2647327, 3775154, 2647292, 2860557, 2823844, 2720691, 2931732, 3740806, 2647337, 3579053, 3775148, 3579085, 2788633, 3775143, 3667955, 3534702, 3775180, 2726068, 3645391, 2928460, 3775162, 2931748, 3645412, 3534713, 3740792, 3667984, 2823850, 3775172, 2748150, 2931750, 3775142, 3534701, 2905895, 3775135, 3667957, 3645417, 3633456, 3579086, 2748130, 3645416, 2931714, 3534707, 2931738, 3775170, 3534724, 2720711, 2931727, 3775158, 3579075, 2860532, 3579066, 3534695, 2647282, 3645402, 2860542, 2905866, 2720707, 3667974, 3667963, 3775151, 3740808, 2748120, 3740819, 3667969, 2905887, 3534700, 2720676, 2928454, 2860536, 2860551, 3645388, 2931721, 3534698, 3534705, 2748162, 3645410, 3645404, 3579054, 3645401, 3579092, 3740793, 3775157, 2726073, 3775156, 3667973, 2860539, 3667983, 3579091, 2928442, 2647317, 2905855, 2748148, 3645389, 2860566, 2860572, 2860573, 3740828, 3775144, 3740829, 3740799, 2928440, 2788595, 3775137, 3775152, 3775166, 3740800, 3740832, 3775176, 3775149, 2720704, 2647318, 2637888, 2647284, 3740797, 2928472, 3740834, 2647333, 3740788, 2720693, 2928456, 3579049, 3740810, 3740787, 2928466, 3579059, 3740796, 2931744, 3740824, 2788622, 3740801, 3740803, 2928455, 3868590, 3957559, 3977145, 4026832, 4026845, 4054477, 3974923, 3868585, 3974906, 3977162, 3977156, 3912108, 4026837, 4043736, 3912123, 4082828, 4054473, 4091855, 4176515, 4108405, 4108400, 3921170, 3817106, 3977150, 3921186, 3921167, 3912138, 3817103, 3912119, 4026836, 3817099, 4026838, 3817122, 3817125, 3817131, 3868583, 3868549, 3921168, 3868563, 3977228, 3817100, 4043774, 4054474, 4040594, 4082821, 3817144, 3817102, 3912136, 3868530, 3817118, 3957563, 4054460, 3974912, 3912099, 3817117, 3817098, 3868595, 3912117, 3912142, 3817129, 3977135, 4040587, 4026826, 3974911, 4043723, 4026847, 4068664, 3912098, 3921188, 3817120, 3817105, 3868579, 3868591, 3912102, 3912118, 3868553, 3977163, 3817124, 4068679, 3817116, 3921198, 4082836, 4043761, 3868552, 3817139, 3868551, 3868548, 3817113, 3912120, 3977099, 3868564, 3868571, 3957570, 3974915, 4040604, 3921196, 3977077, 3912109, 4082838, 3957573, 3868565, 3912110, 3977120, 4026824, 4043731, 4082822, 3817132, 3868586, 3868573, 3912121, 3921193, 3921190, 3974910, 3868541, 3977224, 3977079, 3977165, 3977102, 4068689, 3912122, 4054469, 4026843, 4054465, 3868561, 4043744, 4068687, 4054470, 4082839, 3868587, 3868529, 3957571, 3868569, 3817110, 3957556, 3868535, 3868578, 3977076, 3925808, 3912114, 3868557, 3912128, 3817126, 3912137, 3868554, 3977127, 4040598, 4054464, 3957557, 3957578, 4043726, 4043746, 4026840, 4040588, 3868556, 3817119, 4040601, 4043752, 4068680, 3957558, 4091838, 3957574, 3868545, 3977113, 4040614, 3912103, 4068672, 3921194, 3957560, 3974902, 3957562, 3977108, 3977103, 4040609, 4043755, 4091850, 4043758, 4068688, 4043762, 3868592, 3868575, 4108407, 3868593, 4091858, 3817128, 3868580, 3974899, 3957566, 3977104, 3977105, 4026828, 3868567, 3868542, 3817134, 3912143, 3974919, 3817109, 3912113, 3912097, 3868546, 3912112, 3868582, 3817123, 3974916, 3868544, 3868577, 3921180, 3921181, 3925790, 3925792, 3936219, 3977124, 3944742, 3944746, 3944749, 3944751, 3944741, 3921182, 3925789, 3921171, 3944750, 3912131, 3944745, 3944752, 3944744, 3925815, 3925803, 3925813, 3912132, 3912130, 3912144, 3912146, 3912147, 3912148, 3921179, 3944729, 3944730, 3944734, 3944738, 3944740, 3944743, 3925809, 3925810, 3944748, 3936218, 3977221, 3925794, 3925799, 3817140, 3817137, 3925798, 3921197, 3977141, 3977149, 3817097, 2284636, 2610435, 2472138, 2590716, 2590735, 2590736, 2610438, 2284674, 2224557, 2342627, 2610448, 2610431, 2587270, 2472158, 2905878, 3740794, 2720669, 2928465, 3740816, 2905870, 3667988, 3645407, 2860558, 3645411, 2637908, 3667972, 3579094, 3534735, 2788634, 3633437, 3579080, 2860540, 3667956, 3775160, 2860565, 2905853, 3775139, 3579097, 3775136, 3775140, 3613504, 3613503, 2748129, 2860569, 3534729, 2928469, 3667971, 2720672, 3579073, 3534726, 2931726, 2905885, 2637885, 3667968, 3579055, 2928434, 2748119, 3534716, 2823864, 2726058, 3534737, 2726055, 2726046, 2720685, 2905862, 2748152, 2788615, 2720677, 2726061, 3667959, 3645398, 2860533, 3667958, 2823880, 3534697, 3775182, 2720698, 3645394, 2823882, 3645415, 2637860, 3667982, 3775177, 3534706, 2928471, 2637890, 2726049, 2905871, 3633449, 2905851, 2860567, 2860568, 3775159, 3775163, 3740833, 3740815, 2928452, 2823877, 2928468, 3613497, 2928431, 3534720, 3740807, 3740820, 3775169, 2931704, 2931711, 2637893, 3534708, 3740813, 3921177, 3977143, 3921184, 3974924, 3977151, 3974900, 4026827, 4040613, 3921183, 3868562, 4336180, 4421045, 4176484, 4091843, 4176501, 4176482, 4336157, 4495822, 4483544, 4453402, 4082824, 3977106, 4176470, 4281186, 4336172, 4336175, 3944737, 4421046, 4420983, 4368643, 3912116, 4040590, 4336197, 4368635, 4421006, 4421023, 4043772, 4254846, 4040610, 4108409, 4421017, 4108414, 3817138, 4254860, 4254847, 4040605, 4254827, 4281225, 4043765, 4281220, 4421039, 4495836, 4040608, 4176480, 4068668, 4054471, 4176498, 4483564, 4108406, 4483553, 4336186, 4254852, 4026851, 4420992, 4176472, 4336179, 4043749, 4254854, 4368642, 4254819, 4043729, 4176479, 4176506, 4068665, 4391146, 4336159, 4469647, 3974917, 4336165, 4176495, 4368657, 3868574, 4281227, 3817114, 3977160, 4043751, 4254829, 4043770, 3977133, 4176512, 4368636, 4421029, 4495835, 4043739, 4421019, 3912104, 3944731, 3977110, 4281190, 4176466, 4176500, 4091849, 3817130, 4176469, 4368656, 4254859, 4176494, 4176497, 4082831, 4176489, 4043725, 4469642, 4421036, 4469626, 4176475, 4420978, 4336183, 3868584, 4254843, 4108397, 4176476, 4281233, 4091853, 4368661, 4391155, 4108416, 3868559, 4483559, 4254851, 4281207, 3912140, 4421055, 4421027, 4368662, 4082841, 4091841, 4054467, 4281235, 4026834, 4108396, 4281232, 4040596, 4091856, 4254815, 4495820, 4483554, 3977115, 4108393, 4054485, 4336206, 4108419, 4421009, 4421010, 4026830, 4368654, 3957564, 4054479, 4336195, 4453383, 4026844, 4082832, 4176461, 4495814, 4082826, 4368641, 4336178, 4495826, 4054475, 4281184, 4108411, 4026846, 4068675, 3977118, 4043759, 4043753, 4254863, 4254848, 4336193, 4421049, 4420996, 4043734, 4054482, 4368646, 4176516, 4254818, 4054463, 4254832, 4281226, 4254835, 4040611, 4483561, 4495819, 4421013, 3977101, 4068667, 4068676, 3944728, 4421051, 4108403, 4068682, 3912096, 3977159, 3921189, 4082829, 4176471, 4176486, 4281191, 4281218, 4254858, 4281187, 4281214, 4368658, 4254845, 4420984, 4336199, 4281210, 4082835, 4469621, 4421031, 4495811, 4421050, 4483558, 4420994, 4483549, 3957561, 4421052, 4176508, 4254862, 4368653, 4043748, 4281203, 4368639, 4336156, 4483547, 3912127, 4091837, 4254840, 3977078, 4082830, 4453388, 4108412, 4254813, 3974901, 4469620, 3977073, 4176477, 3868547, 4040591, 4082840, 4091842, 4054468, 4254817, 3977167, 4421022, 4421012, 4495829, 4469634, 3912095, 3912124, 4336208, 4054481, 3912111, 4176473, 3974903, 4176481, 4254844, 4281215, 4391142, 4421021, 4336171, 4254855, 4495837, 4453410, 4421054, 4254816, 4478137, 3817108, 3817142, 4254834, 3977096, 3977139, 3977168, 4254820, 4281205, 4281193, 4336192, 4281222, 4336201, 4368637, 4108402, 4478121, 4054462, 4026841, 4176483, 4421032, 3957579, 4391144, 4043728, 4054484, 3868539, 4254853, 4108420, 4043724, 4469638, 4281224, 3977107, 4453387, 4254831, 4281231, 4478133, 4176490, 4108410, 4054478, 4254826, 4108394, 4082842, 4176491, 4176496, 4391138, 4254865, 4453396, 4368651, 4421056, 3817141, 4082823, 4254836, 4108399, 4368655, 3977148, 4043764, 4091854, 4176488, 4176493, 3817127, 4254866, 3944747, 4054472, 4281202, 4469637, 4281200, 4254856, 4453407, 4082834, 4068669, 4108395, 4176463, 4068673, 4176505, 4391141, 4254833, 4281201, 4420977, 4054483, 4043722, 4281230, 3977144, 4336169, 3977089, 3868588, 4453404, 4421041, 4176511, 4054466, 4091839, 4091851, 3977122, 4421028, 4421024, 4043769, 3977119, 4281181, 4281199, 4054476, 3977125, 4368640, 4176464, 4391137, 4281212, 4420979, 4391139, 4495834, 4176509, 4091848, 4421011, 4421005, 4391145, 4040600, 4043721, 4483550, 4420986, 4420997, 4421001, 4421004, 4453386, 4420987, 3977226, 3977084, 4026835, 4026825, 4040595, 3936217, 4469635, 3977121, 4068670, 3977153, 3977085, 4420976, 4068683, 4040612, 4068690, 3977161, 3925801, 3974907, 4281217, 3977074, 4043730, 3977164, 3977095, 3817136, 3977157, 3817111, 3925806, 4420989, 4495827, 3925804, 3936220, 3977071, 3817096, 4420981, 3977072, 3868594, 3925812, 3925793, 4478130, 3868540, 3921176, 3974918, 3977154, 4026831, 3977097, 3977222, 4469640, 4483555, 3977088, 2551445, 2590725, 2472134, 2426286, 2610465, 3667951, 2905873, 2928474, 2637874, 2647306, 2647305, 3740831, 3667981, 2905863, 2647304, 3645408, 3667985, 2726071, 2637867, 3534743, 2823841, 3645399, 2928453, 2860576, 2823878, 2788624, 3740825, 3667976, 3667989, 3667952, 3613496, 2928467, 2647281, 2905876, 2647334, 3667977, 3534715, 3667962, 2860543, 2748125, 3579095, 2788613, 2928437, 2748124, 2788619, 3775155, 2748145, 2928464, 2928450, 2928473, 2931728, 2928457, 3775179, 3667970, 3740821, 2931716, 3534722, 2905864, 3868536, 4281198, 3977114, 4421033, 4469643, 3944732, 4453392, 3817133, 4176478, 4420982, 4336168, 4082837, 4254837, 3977170, 3912141, 4421038, 3977223, 4091836, 3957567, 4176510, 4040597, 4478136, 3912125, 3912094, 4421007, 4469622, 4483562, 4043733, 4091847, 3912100, 4483545, 4176474, 4368644, 4176504, 4281192, 4483546, 4043741, 3974908, 3817143, 4281234, 4254857, 4421008, 4281219, 4483565, 4336166, 4336177, 3868532, 4469639, 4082820, 4043775, 4483556, 4421014, 3957575, 4469625, 4254823, 3977092, 3957583, 3957572, 4336164, 4368645, 4336200, 4043740, 4281229, 4421030, 3868572, 4469648, 4254839, 4368648, 4281206, 4368659, 3868543, 4478134, 4368649, 4254864, 4043747, 4176513, 4176462, 4068678, 4254821, 4043763, 4420993, 4420980, 3868581, 3868531, 4420999, 4453394, 4420985, 4453408, 4453409, 4336198, 3817121, 3977134, 4082819, 4421040, 4421037, 4391149, 4453391, 4420991, 4281180, 4281188, 4469629, 4421000, 3912101, 4254842, 3977227, 4495830, 4091859, 4254838, 3974913, 3977229, 4336202, 4336190, 4091846, 4082843, 4043757, 4421020, 4336182, 3944733, 4026839, 4336209, 4483541, 3912135, 4420998, 4495823, 4043756, 4420988, 4176502, 4453393, 4336173, 4421015, 4453397, 3921173, 4176499, 4281185, 4336185, 4421016, 4040599, 4026848, 4254850, 3868534, 4281213, 4495813, 4082833, 4108408, 4495817, 4453385, 4281182, 4040589, 4495818, 4336170, 4336194, 4176514, 4453389, 4453401, 4421026, 4483551, 3977146, 3977131, 3925814, 3936214, 3977081, 4421043, 3977112, 3925800, 3925795, 3925802, 4068677, 4068681, 4336204, 3921187, 4483566, 3977225, 3912133, 4281228, 3868538, 4281196, 4453395, 3925791, 4728184, 4788133, 4807987, 4848463, 5091781, 5009636, 4690410, 4852140, 4690413, 4827197, 4834134, 4848457, 4956371, 4934038, 4709551, 4827180, 4872617, 4967943, 4978251, 4852141, 4956357, 4872615, 4922760, 4770355, 4834136, 4728153, 4728151, 4755304, 4755312, 4872614, 4884095, 4848465, 4934037, 4807988, 4872627, 4834152, 4690424, 5009639, 4808020, 4709559, 4755314, 4728173, 4788134, 4934062, 5041600, 4690415, 4755308, 4848467, 4728161, 4852135, 5024971, 5041599, 4709549, 4808028, 4690425, 5041605, 5139497, 4834127, 4827178, 4934060, 4848466, 4852148, 4827189, 5009641, 4934052, 4884101, 4690423, 4978255, 4956360, 4834130, 4808012, 4834143, 4852138, 4755303, 4956358, 4728178, 4808005, 4834150, 4770344, 4834147, 4967945, 5009623, 5024963, 4728175, 4755299, 4872612, 4922752, 4872616, 4967960, 4808007, 4808001, 4934042, 4956370, 4967959, 4934018, 4827191, 5009626, 4690417, 4978248, 5139532, 4770356, 5091774, 5024960, 5009631, 4834149, 4755297, 4922754, 4827176, 5091793, 4770342, 4770357, 5139539, 4728172, 4788139, 4770358, 4728152, 4884097, 5034549, 4967951, 5091792, 5034551, 4884088, 4884107, 4728137, 4956377, 4709553, 4852142, 4852128, 4956354, 4690414, 5009640, 5034543, 4967942, 5009645, 4834129, 4967935, 4967934, 4728141, 4852137, 4978247, 5024961, 5091800, 4934043, 4755310, 4709539, 4690411, 4728162, 4834137, 4934067, 4956356, 4848464, 4728183, 4728181, 4709550, 4728170, 4755313, 4808008, 4788150, 4755295, 4834140, 4852149, 4834131, 4852139, 4884102, 4755305, 4807989, 4934039, 5024980, 4808029, 4978237, 5041612, 5041601, 4978246, 4978231, 4922742, 4807999, 5053333, 5053339, 4808013, 4690420, 5024972, 5009624, 5091799, 4978235, 4978232, 5024975, 4872623, 4978240, 5034550, 5034554, 4922739, 5091775, 4807979, 4922759, 4884086, 5102805, 4922757, 5102802, 5102799, 5102803, 5139486, 5139499, 5139485, 4884096, 5139487, 5139490, 4956355, 5151803, 5151793, 5152343, 5151792, 5151801, 5151797, 4884104, 4884094, 4978250, 4872610, 4709536, 5009619, 4690409, 4770361, 4728160, 4967946, 4852143, 4770359, 4884098, 4770348, 4934026, 4884081, 4728169, 4788148, 4978229, 4788145, 4709545, 4728158, 4956369, 4967948, 4872633, 4934032, 5041616, 5009637, 5034561, 4872626, 4808019, 4848460, 5139543, 4807996, 4788129, 4834138, 4852144, 5034556, 4807995, 5034544, 5091786, 5034552, 4934025, 5041607, 5034558, 4788143, 4827183, 5034545, 4852152, 5053330, 4755300, 4827195, 4807985, 4709543, 5034541, 4728176, 4709548, 4728174, 4852146, 4872620, 4834128, 4956380, 4934034, 5024966, 4728154, 4728142, 4709540, 4848474, 4807986, 4755292, 5152341, 4709547, 4807991, 5041606, 4728150, 5139489, 4884099, 5041608, 4709555, 4934041, 5034557, 4978239, 4934022, 4690416, 4755301, 4709556, 4755307, 4770363, 4808022, 4808010, 4827173, 4956365, 4884100, 4978256, 4884108, 4922758, 4967953, 5009634, 4770349, 4690412, 4872635, 4808003, 4709552, 4690406, 4728163, 4967939, 4934028, 4755294, 5009628, 4690426, 4709544, 4884089, 4807997, 4922740, 4922746, 4922735, 4934054, 4690405, 4728180, 4770367, 4770362, 4807974, 4827194, 4834145, 4852147, 4808002, 4808026, 4788124, 4808014, 5009643, 4922737, 4922738, 4922733, 4978244, 5009620, 5009625, 5102800, 5139537, 5151785, 5139495, 5139541, 5139558, 4690428, 4709558, 4690431, 4808006, 4770350, 4788146, 4834144, 4872622, 5053336, 4770354, 5009632, 4848462, 4956375, 5091773, 4978241, 5053349, 4788135, 4808011, 4807981, 4808018, 4934049, 4934065, 4770343, 4978234, 4922734, 4788136, 4922749, 4872621, 4848479, 4884105, 4690421, 4770368, 4755302, 4827171, 4834146, 4852154, 4709554, 4827174, 4827184, 4852136, 5009627, 5009630, 5034560, 4872630, 5009644, 2587257, 3740802, 2823848, 3579062, 3579072, 3534709, 2726047, 2928439, 2931742, 2860552, 2748141, 3633450, 3633436, 2823857, 2905893, 3633447, 3613499, 2823873, 3534712, 3534736, 2788636, 3579060, 2748163, 2788594, 3740811, 4420974, 4040593, 4026833, 4043743, 4026850, 3921185, 4040592, 4068691, 3912145, 3868589, 4368638, 4495824, 4040606, 4108418, 3817101, 4176485, 3817104, 4421034, 4421048, 4091840, 4469627, 4281223, 3868537, 3921172, 3868533, 3977075, 4176503, 3974898, 3868568, 3977111, 3912115, 4483548, 4281216, 4391140, 4336176, 3974905, 4478131, 4176487, 3977117, 4082827, 4453406, 4054480, 4453384, 4108398, 4453400, 4336163, 4453403, 4453405, 4336161, 4420995, 4495816, 4254825, 3912107, 4368650, 4043750, 4091844, 4336181, 4336167, 4176468, 4254830, 4254824, 3957580, 3921169, 4176507, 4082818, 4453398, 4478124, 3912126, 3977155, 4336191, 4043742, 4421002, 3944736, 4043766, 3936215, 4336162, 3977094, 4495821, 3957582, 4026829, 3936216, 3912106, 4469633, 3977147, 3944735, 4469645, 3974909, 3817112, 3977109, 4281189, 4043737, 3977158, 3921195, 3974914, 3925811, 4755311, 5230020, 4807992, 4728185, 5152337, 5230045, 5224645, 5152351, 5091780, 5178213, 4834133, 5041615, 5178216, 5230053, 5230066, 5260581, 4967958, 5024957, 5152331, 5041611, 5139549, 4934040, 4827177, 5194505, 5009646, 5091777, 5194513, 5178227, 4934023, 4956378, 4956363, 4690422, 4956367, 4956359, 5194506, 4728138, 5041618, 5053337, 5009622, 5224637, 5194509, 5151789, 5230137, 5178231, 5230056, 4934072, 5230024, 5230043, 4934071, 5009642, 5224643, 4956374, 5152326, 5230074, 4934050, 5041592, 5024984, 5139500, 5194512, 5009638, 4934019, 5232734, 4807984, 4967938, 4807977, 4934033, 5139557, 4728182, 5091776, 5178211, 4934046, 5230059, 5034564, 5041619, 5091794, 5178210, 5230150, 5230145, 5230140, 5178214, 5260597, 5151809, 4807994, 4788144, 5151790, 5230060, 4967955, 5091784, 4934056, 5230033, 5178236, 4872625, 5102790, 5178229, 5224623, 5041597, 5024981, 5230146, 4934036, 5053331, 4827190, 5230044, 5230073, 5230052, 4770364, 5041604, 4956373, 4728167, 5139533, 5232728, 5024959, 4967961, 4967944, 4934035, 5230029, 5194520, 5053343, 5151808, 4934053, 5053332, 5194530, 5053345, 5041593, 5230133, 5224639, 4788147, 5224629, 4934058, 5178212, 4956372, 5152338, 4770352, 5260576, 4978249, 5151782, 4827182, 5053329, 5102785, 5139484, 5024968, 4709538, 5139482, 4728143, 5041614, 4956362, 5041617, 4934070, 5152340, 5024970, 5194517, 5194518, 5041609, 5194531, 5194515, 5224641, 5224622, 5194519, 4956379, 5230138, 5232712, 5232724, 5230030, 5224647, 5232725, 5230068, 5230039, 5230065, 5230069, 5102793, 5224642, 4872632, 5224625, 5151791, 5260592, 4807976, 5151784, 4967941, 4978233, 5230049, 5151786, 5232722, 5041595, 5024964, 5230051, 4755309, 5009633, 5091787, 5041610, 5230075, 4884085, 5178218, 5230149, 5053334, 5024965, 4922744, 5230040, 5230042, 5152342, 4788127, 5194524, 5178221, 5194526, 5230034, 5152345, 5230031, 4967947.5230021, 5230139, 4827186, 5151783, 5151805, 4807993, 5230071, 5194529, 5260588, 5091779, 4884092, 5091790, 5091778, 5139552, 4884090, 5151807, 5232720, 5230134, 5230035, 5178234, 5232723, 5230152, 4852151, 4808000, 5260578, 5041594, 5024958, 4807978, 5178225, 4872636, 5178219, 4967954, 5139536, 5139540, 5139553, 5230058, 5139534, 4967952, 5194510, 5178220, 4728140, 5230127, 4848459, 4884082, 5151799, 5152349, 5232714, 5053341, 5230130, 5152335, 5152334, 5139498, 5230026, 5053338, 4934064, 4834148, 5232721, 5230041, 5230070, 5151787, 4728155, 5260603, 4956376, 5139483, 5151798, 4922745, 4834135, 4827193, 5034562, 5178235, 5024979, 4770365, 5053324, 5230128, 5230057, 5053325, 4978245, 5091782, 5139492, 5194507, 5224620, 4934057, 5224626, 5152332, 5224640, 5178232, 5139551, 4728148, 4788130, 5230129, 5024976, 4967950, 5230136, 4848472, 5232730, 5024962, 4978253, 5139554, 5139556, 5194525, 4808023, 5230064, 4934055, 4956364, 5230038, 5230061, 5194521, 4978242, 4827196, 4709542, 4872611, 4852145, 5024974, 5151806, 4852133, 4967956, 5194528, 5224634, 5224633, 5230142, 5230054, 5152327, 5152330, 5230050, 5024978, 5232718, 5232733, 5024973, 5230032, 4788137, 4690408, 4807975, 4755317, 4922750, 4848481, 5224632, 5230135, 5194511, 5230153, 5232711, 5232713, 5232717, 5232727, 5260600, 5102797, 5224621, 5178233, 4934061, 4770366, 5034568, 4922755, 4808016, 5260590, 4788125, 5260599, 5102778, 4827187, 4978230, 5260587, 5260589, 4934051, 5102804, 4872637, 4884091, 5224638, 4709537, 4834126, 5102796, 5102781, 5102787, 4884083, 4922741, 5139547, 5034542, 5224627, 5224635, 4755298, 5178230, 4978236, 4852127, 5224646, 5102788, 4848455, 3579050, 3775174, 4336188, 4026842, 4469623, 4420973, 4469644, 3977128, 4082825, 4495828, 4495831, 4281195, 4495815, 3977093, 4391153, 3974921, 4420975, 4091852, 3957577, 3977100, 4068685, 3868560, 4495838, 4043720, 3912105, 4091857, 3817135, 4391143, 4421053, 4043768, 4421042, 4469641, 4368660, 4040607, 4453390, 4043767, 4054461, 4336174, 4281183, 3977083, 3977080, 3912139, 4483543, 3977090, 3921178, 3977152, 3921175, 3921191, 3868566, 4281194, 4848458, 4807998, 5034555, 5034553, 5230046, 5194508, 4728147, 5139488, 5091783, 4834141, 4755318, 4709541, 5091796, 4967957, 5260577, 5102791, 4967937, 5139491, 4788149, 5260580, 5232715, 4728171, 5139548, 4728168, 4934030, 4956368, 4690419, 4922736, 5178215, 5024983, 4808021, 5230132, 4852150, 5230025, 5230131, 5053344, 5041602, 4728144, 5041603, 5139542, 5139538, 5232719, 5152329, 5230147, 4922743, 4852131, 5230141, 5091785, 5194533, 5224624, 5224631, 4848476, 5151796, 5152333, 5139493, 4967949, 5053326, 4922756, 5230023, 5041598, 5194514, 5178209, 5260582, 4934068, 4755291, 4852130, 5178222, 5178224, 5178217, 4755296, 4967940, 5034559, 4956361, 4728159, 5232726, 5139494, 5232732, 4807982, 4755306, 5041613, 4728139, 5102783, 5009621, 4827188, 5224636, 4834142, 4808025, 4728145, 4770346, 4872634, 4848477, 4808015, 4848480, 5260584, 5009635, 4884106, 4848456, 5102779, 5102780, 5091798, 5102784, 4922751, 5224628, 5102801, 6139315, 5964419, 6090570, 5992441, 6078176, 6139339, 6176367, 6176355, 5902850, 5992436, 6044270, 6050808, 5964413, 5985302, 6078218, 5905808, 6078173, 6050782, 6193820, 5905796, 6139328, 6176347, 5884951, 5899144, 5884949, 5905794, 5868050, 5905797, 5953953, 5985312, 6044273, 5992442, 6090577, 6138888, 6139329, 6110789, 6139069, 6139091, 5992514, 5898148, 6287308, 5899128, 6110797, 6176278, 6050798, 6050793, 6110796, 6139336, 6139074, 5899135, 6044279, 6176275, 6078227, 6090901, 5964410, 6100607, 6050788, 6090589, 5899151, 6050784, 6139330, 6090898, 5992449, 6253934, 6050764, 5898163, 6044266, 5992506, 5868061, 6090882, 5992456, 6078209, 6100601, 5884941, 5868039, 6176342, 6050804, 6225275, 6090900, 5899139, 6100610, 6193808, 5985307, 6050766, 6100623, 5898159, 6078213, 6276633, 6193816, 5884930, 6090569, 6176279, 6014723, 5964412, 6078165, 6225270, 5836388, 5964414, 5985297, 5992518, 6044258, 5985293, 6014738, 6078169, 6044275, 6078229, 6090567, 6090896, 6078224, 5836407, 6090890, 6193806, 6176284, 6253943, 6176345, 6090909, 6253980, 6253995, 5902854, 6176356, 6090885, 6078226, 5992440, 5992450, 5898143, 5953938, 6050781, 5992448, 5902842, 6100615, 5868051, 5919208, 6014715, 6078211, 6044262, 6176289, 6138887, 6176274, 5898145, 6176286, 6225285, 5953958, 6078222, 6176269, 6044274, 6050801, 6078175, 6090893, 5868054, 6110786, 6078223, 6176365, 6260434, 6044282, 5836411, 6100620, 5953952, 6110787, 5868042, 5919196, 6050776, 6050805, 6078159, 6090887, 5898152, 6139324, 5992437, 6253941, 6253928, 6225277, 6159357, 6283411, 5898162, 6100604, 6050770, 6110801, 5964396, 6265551, 6265544, 5964398, 6014731, 5836402, 6138884, 6225279, 6193803, 6044281, 5902839, 6260438, 6283400, 5898166, 5898154, 5905802, 5992446, 5836389, 5905789, 6014739, 6014724, 6050794, 6014730, 6078154, 6090889, 5992455, 6110793, 6100600, 6100608, 6138889, 6193825, 6193810, 6193821, 6193807, 6139331, 6193813, 6139083, 6090883, 6138870, 6254005, 6193800, 6254001, 6253953, 5992443, 6260416, 6265557, 6260422, 6287334, 6253958, 6078210, 6283405, 6253965, 6283407, 6283402, 6253976, 6276630, 6276626, 6276616, 6276623, 6139319, 6299543, 6253954, 6299547, 6299549, 6078216, 5992435, 5919203, 5899147, 6100597, 5898156, 6176346, 5953943, 6176349, 6090576, 5964416, 5868043, 5884942, 5898158, 5992503, 5992454, 6090894, 6100611, 6090587, 6139316, 6193804, 6253987, 6044259, 5953941, 6176277, 6260417, 5902845, 5899148, 5899129, 5902853, 5992515, 6014733, 6044257, 5898155, 6050763, 6078151, 6050773, 6110803, 6078177, 6139341, 6139089, 6139090, 6176361, 6193801, 6176268, 5992516, 6193799, 5902855, 6090578, 5964399, 5898168, 5898167, 5919211, 5985306, 5985309, 5992522, 6044277, 6090907, 6090908, 6110799, 6139073, 6260427, 5905805, 5992504, 5992452, 6139067, 5964395, 5953931, 6090572, 5964415, 6050772, 6078225, 6138872, 6253992, 6253931, 6253948, 5992499, 5985305, 6050806, 6078206, 6044276, 6090583, 6090892, 6090565, 5868055, 6176290, 6176288, 6225269, 6014742, 6253938, 6276637, 6283401, 6139085, 5898151, 6139335, 6254002, 6090888, 6090903, 6253930, 5992520, 5898150, 6050765, 6139323, 6078215, 6253946, 5905782, 5905784, 6014717, 6090586, 6090566, 6050792, 6044280, 6138864, 5898153, 5905788, 5898170, 6090571, 6044261, 6078231, 6139338, 6090906, 6100603, 6139321, 6139071, 6253983, 5836406, 5884927, 5899138, 5953932, 6193817, 6050768, 6090902, 6014726, 5898144, 5899127, 6287327, 5899140, 5992508, 6014716, 6078157, 5953933, 6138885, 6100596, 6253998, 6078232, 6050803, 6260419, 5953954, 6014728, 6014719, 6254004, 6319615, 5836394, 6193819, 6225284, 5992432, 6090564, 5905795, 6253989, 6265554, 6225265, 6253963, 6260435, 6287310, 5902843, 5884929, 5899126, 6139084, 6139076, 6100622, 5953944, 5919197, 6014729, 6050785, 5905793, 5884937, 6090895, 5902841, 5836414, 5899143, 5964409, 5953946, 6193811, 6283417, 5836403, 5905781, 5905791, 5905807, 5919219, 5899131, 5905783, 5905801, 5905806, 5953956, 5964394, 5964402, 5964406, 5964408, 5919194, 5919195, 6050797, 6078161, 6078170, 6139340, 6176265, 6176271, 6176273, 6176282, 6176283, 6176287, 6110785, 6176363, 6193805, 6265560, 6276618, 6276627, 5884940, 5868034, 5868036, 5868038, 5868040, 5868058, 5884924, 5836390, 5884938, 5905798, 5836391, 5919209, 5953950, 5884934, 5884935, 5898164, 5836398, 5884933, 5902856, 5953955, 5992510, 6044256, 6050756, 6139070, 6139086, 6139093, 5836393, 5905792, 6100614, 6159346, 6159351, 6159355, 6159356, 6159358, 6159364, 6159368, 6159372, 6260421, 6260430, 5884932, 5992445, 5992500, 6078167, 6078217, 6138879, 6100598, 5953934, 5884947, 5919214, 5919217, 6078155, 6139327, 5985294, 6253939, 5868053, 5919207, 6193814, 6139081, 6090584, 5884946, 4421018, 4483567, 4043760, 4054459, 3957576, 4421003, 4068686, 4254814, 4254822, 4043745, 4254861, 4336203, 4421035, 4368652, 4336187, 4281221, 4483552, 3977136, 4483560, 4495812, 5232708, 4852132, 4728177, 5230027, 4690418, 4934021, 5034563, 4807990, 5194516, 5034566, 5034547, 4934073, 4848475, 5152347, 5009629, 4788138, 5139545, 4808004, 4852134, 4967936, 5053348, 5232731, 4934066, 4808009, 5053350, 5152350, 5152348, 5139501, 5091789, 5053327, 5024977, 5230062, 5230047, 4934045, 4690407, 4755316, 4728164, 5151795, 5260598, 5151794, 5034548, 5091791, 5178228, 4807983, 5260596, 5139544, 5232729, 5194504, 4884087, 4755315, 4788128, 5151802, 5024982, 5178223, 4827185, 5152336, 5139550, 5034546, 4872624, 4872619, 4788131, 4827172, 5230067, 4728157, 6138871, 6283425, 6488033, 6078158, 6176359, 6520956, 6387585, 6253977, 6176343, 6253971, 6050787, 6283419, 6283413, 6078152, 5905790, 6507155, S115986, 6299544, 5919210, 6283406, 6276642, 6299542, 6265568, 6253942, 6371653, 6387549, 6371643, 6387566, 6385221, 6487989, 6520962, 6534665, 6541231, 5884948, S116006, 6534654, 6253929, 6387581, 6159352, 6541290, 6482590, 6541268, 5905804, 6534657, 6541299, 6371634, 6139080, 6488025, 6176354, 6176285, 6283398, 6371639, S116014, 6387582, 6253952, 6541249, 6482608, 6534664, 6371644, 6371641, 6138865, 6253991, 6276617, 6225280, 6139317, 6488022, 6541257, 6387573, 6319612, 5868047, 6371640, 6253926, 6253969, 6299533, 5899149, 6253986, 6387601, 6487996, 6387542, 6276634, 6387543, 6225272, 6225286, 6253970, 6260418, 6507164, 6265556, 6319617, 6488030, 6225268, 5899130, 6299559, 5905800, 5868056, 6193802, 6139068, 6100618, 6387536, 6541334, 6287328, 6541307, 6371646, 6254003, 6078207, 5899134, 6347778, 5868044, 6260420, 6225271, 6260426, 6387553, 6541288, 6487985, 6276643, 6541329, 6371650, S116017, 6488035, 6176358, 6176341, 6265564, 6265545, 6253996, 6265570, 6265571, 6387600, 6387603, 6520951, 6541243, 6387589, S115998, S115995, 5964411, 5899124, 6385228, 6482601, 5899137, 6299553, 6265566, 6347781, 6078163, 6253936, 6176364, 6110784, 6078220, 6014735, 6225282, 6265547, 6487983, 6100617, 6265562, 5953949, 6283416, 6387541, 6347777, 6014741, 6090899, 5953935, 6090591, 5985310, 5992511, 6541326, 6110798, 6487999, 6319619, 6520952, 6541304, 6260428, 5992505, 6299538, 6287318, 6225287, 6385214, 6371652, 6260436, 6078171, 6488034, 6488006, 6176344, 6283423, 6371657, 6287333, 6299539, 6260439, 6050760, 6253974, 6349589, 6139092, 6347774, 6387592, 6488021, 5992433, 6534659, 6078221, 6541267, 6507173, 6050796, 6299532, 6260442, 6044272, 6541276, 6193812, 6371656, 6541269, 6090574, 5905799, 6520953, 5902846, 6385213, 6078164, 6176264, 6541336, 6283399, 6253951, 6078219, 6283408, 6100612, 6385212, 6253993, 6488027, 6541279, 6225267, 6276628, 6534662, 6299535, 6299558, 6299546, 6349605, 6371651, 6371647, 6371645, 6385234, 6385227, 6387596, 6387594, 6387533, 6387560, 6387570, 6387561, 6287316, 6507158, 6520961, 6159361, 6387555, S115974, S116001, S116029, S115975, S115968, 6287323, 6090905, 6176272, 6139078, 6253957, 6159347, 6193824, 6283409, 6253960, 6265553, 6488024, 6265546, 6110792, 5992502, 6014721, 6387568, 6541239, 5985290, 6050754, 6319613, 6287319, 6387564, 6090579, 6276624, 6534649, 6534666, 6014722, S115984, 6387610, 6265561, 6349586, 5953936, 6385235, 6387558, 6487998, 6253964, 6253978, 6534673, 6541328, 6534667, 6534670, 6541242, 6159362, 6541259, 6541236, 6260443, 6520959, 6541248, 6287331, 6100613, 6387578, 6276620, 6276639, 5992521, 6110802, 6276621, 6488001, 6387545, 6299550, 6488014, S116012, 6193822, 5964401, 6044264, 6488004, 6100616, 6541297, 6050755, 6260432, 6253937, 6387609, 6487982, 6541264, 6541331, 5899150, S115987, S115985, 6287324, 6349599, 6507162, 6283422, 6276619, 6371635, 6253950, 6044267, 6387576, 6387565, 6487990, 6507176, 6507159, 6507174, 6534660, 6541237, 6541275, 6050802, S116033, 5953948, 6193798, 6014718, 5985303, 6050791, 6139088, 6283415, 6276638, 6387547, 6100602, 6100606, 6349581, 5992512, 6371648, 6225276, 6260425, 6299541, 6347789, 6385229, 6387583, 5902852, 6193823, 6276632, 6253945, 6488002, 6387534, 6139066, 6139322, 6225278, 5898169, 6225274, 6287326, 6159353, 6488011, 6534672, 6254006, 5902840, 5898157, 5992444, 6482593, 6541238, 6287329, 6487991, 6287315, 6276635, 6371658, 6520958, 6253988, 5902834, 6299540, 6385232, 6347790, 6299551, 6371632, 6276636, 6387591, 6507177, 6534650, 6487992, 6534661, 6541235, 6299554, 5953951, 6044265, 6349603, 6371659, 6260441, 6287314, 6541311, 6482610, 6385220, 6319618, 6253982, 6507175, 6482586, 6541252, 6276640, 6265548, 6260429, 6100621, 6347785, 6349591, 6507179, 6520960, 6534674, 6534656, 6541308, 6159360, 6253956, S116009, 6159370, 6507168, 6159371, 6482594, 6287320, 6371633, 6385225, 6541321, 6520955, 6541324, 6541323, 6176360, 6253933, 5884950, 6265549, 6541265, 6253962, 6385210, 6265552, 6541318, 6287312, 5953942, 5836392, 6534653, 6541256, 6541271, 6050753, 6507152, 6507156, 6541245, 6387598, 6387607, 6387569, 6488003, 6541301, 6541333, S116024, S115973, 6387532, 6387590, 6387562, 6487997, 6488028, 6488005, 5992501, 5836396, 6176353, 6276629, 6387588, 6110791, 6193809, 5919215, 6078160, 5868048, 6387539, 5992525, 5899146, 6541300, 5919200, 6541316, 6387606, 5953947, 6265550, 6487981, 6299556, 6287322, 6387605, 6507153, 6507166, 6276641, 6253940, 6138880, 6253985, 6541260, 5902851, 6488009, 6541274, 6253999, 5919213, 6387538, 6482587, 5992526, 6176351, 5868037, 6482591, 5884939, 6507154, 6482603, 6176276, 6110788, 5868060, 6349583, 6541315, 6349585, 6349598, 6225283, 6387584, 6139320, 6265567, 6138882, 5899142, 6100619, 6541280, 6253944, 6458757, 6650259, 6458756, 6475829, 6590753, S109617, 6590731, 6420944, 6421084, 6563382, 6475820, 6421161, 6590730, 6467506, 6475828, 6650263, 6421497, 6421519, 6563398, 6650253, 6563372, 6421090, 6470270, S109612, S109773, S109776, 6421510, 6420936, 6429031, S109600, S109395, 6421030, S109764, 6467499, 6470277, 6429013, 6475806, 6420876, 6421059, 6420883, 6420874, 6590739, S109396, S109582, 6421085, 6458769, 6563368, 6421052, 6429011, 6420935, 6421051, 6590755, 6420932, 6421078, 6467496, 6420940, 6420939, 6475817, 6420933, 6475809, 6420934, 6590712, 6421043, S109603, S109593, S109393, 6420877, S109420, 6420882, 6467504, 6420891, S109410, S109411, 6458775, 6420937, 6421505, 6563418, 6420892, 6650272, 6650276, S109771, 6475808, 6420887, 6590705, 6421096, 6421170, 6421147, 6590704, 6421022, 6421089, 6421100, 6650251, 6429018, 6420946, 6470263, 6475803, 6475818, 6421172, 6563407, 6421508, 6475825, 6563400, 6563411, 6590718, 6421037, 6467512, 6475813, 6421517, 6590754, 6458777, 6421099, 6429012, 6421516, S120506, 6563401, 6420899, 6475815, S109607, 6421019, 6421028, 6421058, 6421033, 6421045, 6470264, 6458762, 6421062, 6470279, 6475822, 6429017, 6590708, 6421046, 6590729, S109391, S109762, 6470276, 6429024, 6421167, S130340, S109579, 6429027, 6467486, 6563420, 6467511, 6421035, S109759, S109758, S109424, S119777, 6420949, 6421499, 6467500, 6590722, 6590751, 6650264, S109399, 6590720, S119771, 6590713, 6650273, S109587, S120508, 6420953, 6650255, 6475811, S109591, 6650249, 6421169, 6458776, S109602, 6421094, 6467510, 6421023, 6467493, 6563383, 6563404, 6590740, S109755, 6421092, 6429010, 6421055, 6421041, 6421026, 6563395, 6590726, 6590743, S109595, S109609, 6590717, 6650271, 6650266, S109403, 6420952, S109596, S120505, S120503, S119779, 6421157, S109753, 6458763, 6563397, 6470282, S109774, 6458782, 6470261, 6467494, 6429028, 6563373, 6590749, 6563380, 6590748, 6421093, 6421075, 6458778, 6467503, 6421071, 6475827, 6590744, S119761, S109754, 6590711, S120509, 6421103, 6590702, S109394, 6421021, 6421068, 6421077, 6420943, 6421501, 6429035, 6429029, 6470284, S109601, S109751, 6421081, S109597, 6470278, 6421025, 6429015, 6420900, 6467490, 6590733, 6590738, 6475816, 6563391, 6475830, 6470269, 6421061, 6429009, 6421158, 6421086, 6421166, 6467501, 6563419, 6421153, S109611, 6421024, S109604, S119764, S109406, 6470262, 6470267, 6470268, 6470272, 6470281, 6470283, 6421152, 6421154, 6420881, 6421079, 6590710, 6421101, 6563402, 6421044, 6458760, 6458764, S109413, 6563392, 6470260, 6420884, 6429022, 6563417, 6421042, S109763, 6590728, 6421036, 6590719, 6421072, 6458759, 6421513, 6590724, 6420898, 5102789, 4956366, 4848461, 4834139, 4872613, 4788142, 5091795, 5053328, 4788140, 5230022, 5260602, 5053335, 4852129, 4978238, 5102795, 5260579, 5230036, 5139559, 4834151, 4834132, 5230144, 4872631, 5034565, 4978243, 5053347, 5139496, 4934047, 4848473, 5260591, 5178226, 5139555, 4872629, 4884084, 4728166, 6078174, 6260423, 6139334, 6534663, 6387599, 6044260, 6541292, 6139332, 6176357, 6265563, S116036, 6507171, S115976, 6078168, 6385223, 6541262, S115979, 6488020, 6050767, 5905787, 6541303, 6541305, 6138869, S115977, 6347772, 6534658, 6138875, 5992439, 6349597, 5919201, 6265555, 6253955, 6507170, 6050775, 6050795, 6534671, 6090891, 6541229, 6044268, 6541228, 6387575, S115978, 6347773, S115997, 6541241, 6488016, 6260424, 6078212, 6371655, 6371638, 5985288, 6541254, 6487995, 6541240, 5898147, 6488023, 6385215, 6488026, S116015, 6159367, 6541255, 6299552, S115996, 6541330, S116003, S116034, 6541234, 5902836, 6482596, 5992453, S115999, 5985295, 6138863, 5919212, 6253990, S116030, 5992513, 6276622, S116000, 6159365, 6283414, 5953940, 6541246, S116038, 6541244, 6541285, S115992, 6541287, 6534668, 6534652, 6487980, 6507178, 6050786, 5902835, 6050799, 6487987, 6050757, 6507169, 6225266, 6541338, 6482598, 6283410, S115981, 5899132, 6541314, 6159348, 6319614, 6541266, 6541281, 6385217, 6319611, 6387559, 6138868, 6299536, 6387580, S116028, 6283424, 6488018, 6253997, 6507163, S116005, 5964407, 6541298, S116004, 6534651, S116002, 6534655, 6014734, 6385222, 6488010, 6541251, 6090575, 6349596, 6482600, 6385231, 5899145, 5899125, S116035, 6387548, 6541337, 6253994, 5902847, 6488007, S116026, S115982, 6176366, 6541309, 6139077, 5836408, 6488032, S116037, 5992524, S116025, 6347775, 6287317, S116013, 6541253, 6159373, 6347769, 6347770, 5868046, 6520954, S116007, 6110808, 6139087, 6487986, 6347791, 6488017, 6541250, 6541322, 6283403, 6078166, 5836410, 6541294, 6507161, 6139326, 5992519, 6287332, S115991, 6541282, 6090588, 5919216, 6487988, 6541302, 5919199, 6387537, 6253949, 6541293, 6541258, 5992431, 6487993, 5919204, 5884943, 6541247, 6541270, S115988, 6541261, 6482604, 6299557, S115970, 6482602, 5919206, 6349587, 6349593, 6138867, 5868059, 5868041, 5868035, 6283412, 6650257, S259703, S109404, S233617, S250154, S215717, 6420894, S250024, S250141, S120544, S202792, 6590715, S218677, S109768, S282188, S109425, S282408, 6650269, 6467492, S282398, S259700, S130317, S130337, S250114, 6590727, S222944, S296393, S218687, S282393, S109746, S202807, S130330, S211004, S215729, 6590736, S219838, S250137, S109590, S215747, S259713, S259710, S269359, S269021, S289393, S248162, S119773, S237096, S259702, S211005, 6421151, S222952, S237091, S109408, S202815, S222953, S229514, S250145, S229511, S250151, S250026, S269142, S269354, S296394, S271847, S119772, S130318, S211009, S215752, S248169, S250046, S269385, S109605, S296405, S282066, S281936, S215707, S250044, S296687, 6429023, S271830, 6421156, S215742, S215706, S282389, S282055, S282194, S250144, S259207, S296402, S282199, 6467513, S250147, S120557, 6470274, 6563378, S202773, S215754, S218673, S215730, S130339, S248167, S269346, 6421165, S229510, 6590714, S229463, S215709, S277978, S268991, S109750, 6421074, 6420880, S259203, 6467498, S219841, S202787, 6421082, S202777, S120558, S120520, 6475807, S119748, S269006, S269034, S282197, S120510, S218699, S229440, 6650260, S271837, S109592, S202808, S229451, 6421155, S289407, S250034, S130329, 6650274, S202772, S284053, S248158, S250054, S250135, 6470273, S250125, S202768, S259698, S269138, S269042, S277990, S289383, S269338, S229442, S229498, S268995, S218670, 6650275, S269112, S278005, S296397, S222950, S109426, S215740, S120519, S218704, S269364, S269122, S282201, 6563403, S229500, S296410, S229453, S259714, 6420878, S259711, S269118, S109402, S211012, S218707, S218679, S229476, S250120, S211021, S219831, S237104, 6563367, S269002, S281949, S296370, S229454, S109745, S202796, S119753, 6563386, S211011, S219832, S202783, S215748, S215705, S269028, S211025, S269010, S269124, S120537, S269366, S296369, 6420886, S109765, S281939, 6650250, S250128, S250030, S282182, S284029, S202765, S202791, S215725, S259217, S277981, S277994, S282211, S269014, S218706, S271818, S284025, S218689, S215713, S202816, S229484, S229506, 6467502, S269102, S269101, S250036, S277997, S250061, S250060, S250031, S289404, S289394, S229513, S202818, S259712, S202789, S269352, S237102, S130315, S229505, S282187, S222951, S109421, S109423, S202753, S202813, S229493, S109405, S109584, S233609, S120529, S233621, S109740, S259695, 6475826, S269100, S282210, S284048, S282382, S289386, 6420941, S215738, S250033, S259218, S218682, S109400, S259701, S248174, S130319, S218672, S215712, 6421070, S229512, 6563389, S250155, S250156, S269376, S282061, S284054, S282079, S229443, S109409, 6421496, 6421164, S119778, S109428, S237097, S278006, S271839, S250121, S278007, S284058, S296661, 6467497, 6429033, S259208, S202758, S277988, S218692, S120525, S215755, 6590709, S250116, S268994, S269334, S229449, S296379, S268992, 6420956, S120526, S120511, S259705, S229507, 6590735, S269038, S277976, S269368, 6475821, S202805, S229458, S269043, S250119, S296392, S109606, S218669, S282404, S281951, S120515, S284051, S278014, 6421091, S109594, S130312, S109742, S202755, S202819, S120541, S211018, S215726, S215753, S211027, S211006, S109769, S219835, S219834, S202767, S229446, S229489, S222948, S229450, S219833, S229482, S233608, S233619, S229494, S229465, S250138, S269032, S269374, S202759, S269025, S271842, S250037, S250041, S250139, S233620, S237086, S284042, 6421054, S120507, S202812, S282385, S229485, S259697, 6421034, 6458758, 6420895, S202776, S268996, S120501, S202775, S250053, S109583, S277985, S289403, S282189, 6563413, S120502, S202794, S218690, S248160, S250032, S269033, S282205, S282078, S229497, S281942, S284021, 6467508, S202814, S229471, S229486, S250055, S229439, S229477, S269377, S269107, S250134, S271845, S278012, S271828, S282067, 6420873, S130314, S211020, S215731, S215733, 6420958, S248166, S248173, S109585, S250148, S237101, S259206, S250150, S269096, S271829, S282198, S282384, S237094, S296404, S282208, S237105, S269331, 6420954, S296665, S119757, S271844, 6420942, S277995, S130328, S109613, 6420950, S250051, S250025, S289392, S282388, S202802, S289379, S215756, S269350, S271840, S296406, S289381, S211022, S250029, S109775, S120521, S250043, S289380, S130321, S229515, S296680, S202790, S130311, S130324, S211008, S202782, S218695, S215711, S229472, S233610, S250021, S202788, S269126, S269012, S250131, S271827, S278000, S278009, S289409, S277989, S277980, S281944, 6563394, S296686, S296679, S233618, S250042, 6421504, S250163, 6458779, S269030, S269020, S202798, S211016, S211026, S233614, S250149, S268999, S269343, S269113, S271822, S202770, S282195, S119780, S202806, S215757, S109416, S120536, S250035, S269347, S222960, S109752, S211015, S211023, S202761, S218693, S215710, S229455, S248176, S250158, S269128, S215739, 6429034, S271832, S269340, S277998, S282410, 6421056, S269017, S282394, S269332, 6421494, S215723, 6590732, S269117, S250039, S119768, S109761, S211017, S109615, S271812, S218703, 6421064, 6470285, S250064, S130327, 6650252, S215708, S250146, S281952, S109398, S250123, S282185, S229447, S284033, 6421173, S229495, 6421060, 6563371, S109747, 6421171, 6650254, S215734, S215727, S215716, S229487, S222947, S229466, S229480, S259220, S250162, S219842, S269116, 6421057, S211013, S269019, S229502, S296403, S296389, S211024, 6420947, 6650268, S130316, S119775, S202754, S120531, S229438, S237088, S202762, S202774, S218668, S269026, S271816, S282077, S269339, S281935, S282059, S269153, S215704, S119767, 6420890, S250047, S259219, S282403, S130338, S120559, S250157, S269381, S237093, S281946, S248168, S282060, S282206, S109397, S119759, S109419, S120545, S202786, S211007, S222945, S211030, S250017, S269365, S259706, 6421039, S269356, S281945, S282052, S296642, S202781, S202784, S269384, S120523, S282063, 6421503, 6420893, 6421500, 6563399, S130320, S282057, S237089, S215759, S250040, S282183, S120534, S269111, S277979, S248164, S269362, S269125, S237100, S202757, S130333, S130334, S269146, S269152, S269105, S237099, S269355, S284046, S284047, S289414, S215743, S215745, S229483, S229491, S250022, 6421097, S109757, S282209, S271834, S284022, S269145, S271841, 6563409, 6429026, S218698, S120530, S120542, S269143, S229488, 6421066, S248159, S218701, S215744, S218691, S109412, S120513, S237095, S215728, S250126, S269027, S282207, S284040, S269015, S119770, S202760, S119754, S119765, S269039, S269005, S268989, 6650267, S282406, S284060, 6563381, S218702, S120518, 6590703, S269130, S296681, S211010, S120533, S271833, S218684, S282391, S282401, S218675, S248172, S218700, S218685, S215746, S269022, S109581, S250160, S202799, S269029, S120548, 6421098, 6590706, S109610, S120540, S229474, 6590742, S119758, S269044, S259209, S109614, S229492, S120528, 5260593, 4934069, 5194522, 5053342, 5232735, 5053346, 5260586, 5041596, 5230126, 5152344, 5102782, 5102786, 5053323, 4956353, 5260594, 5053340, 5224644, S116016, 5884931, S115972, 6139333, S115969, 6078162, 6044271, 6078172, 6283421, 6541335, 5985300, 6541272, 5919205, 6176266, S115971, 6050761, 6347784, 6347776, 5836387, 6349580, 6541230, 6507167, 6287321, 6371654, 6050777, 6265565, 6347771, 6520957, 6265569, 6078233, 6090897, 6090886, 6507165, 5992447, 6488029, 5964418, 6487994, 6385233, 6110794, 6090904, 6050769, 6371636, 5985304, S116023, 5985289, 6139325, 6050783, 6260437, 5899136, 5905785, 6253927, 6385218, 6347786, 6078230, S116008, 6387552, 6387530, S116032, 6488015, 6347787, 6541312, 5902844, 6159349, 6050774, S116027, 6138890, S116018, 6138881, 5985313, 5985298, 6139079, 6044283, S115983, 5992434, 6176368, 5898161, 5902838, 6090582, 6299548, 6387556, 6387574, 6349594, 6482595, 6385216, S115994, 5868057, 6541320, 5899133, 6488013, 6050771, 6254000, 5902830, 6014727, 6347783, 5985314, 6371642, 6159359, 6159369, 6299545, 5902833, 5902837, 6488008, 6253932, 6050762, 5964420, 6482589, 6225273, 6100599, 6387540, 5902848, 6287325, 6482592, 6482584, 6541339, 6387550, 6488000, 6349582, 6253961, 5884926, 6482585, 5884936, S109616, S229504, S250023, S296654, 6590757, S296645, S296639, 6590741, 6421148, S296643, S250117, S296677, S281941, S296381, 6421518, S109744, 6467491, S271824, S269003, S296382, 6429020, S233611, S296367, S269001, S284032, S296644, S289397, S120516, S271836, S250115, 6420885, S259696, S120504, 6650270, S202779, S277993, S250129, S282076, S281940, 6458767, S282073, 6421506, S278010, 6563376, 6467487, S250049, S229478, S296684, S278002, S289399, S269018, S282193, S289387, S233622, S296662, S296640, S218705, S269000, S269037, S202800, S282190, S269349, S296683, S120517, 6650256, S296395, 6475812, S219837, 6458771, S289408, S289412, S296387, S259213, 6421102, S269147, S296408, S282056, S278003, S248161, S271846, S271838, S250052, S296399, S211029, S109749, S229448, S233604, S296657, 6420938, S284052, S296653, S289396, S281948, S296646, S296667, 6420957, 6563377, 6467507, 6467488, S248157, 6421088, 6421029, S233615, S269369, 6421160, S296670, S259707, S250143, S259216, S250020, S250050, S109401, 6590721, 6420945, S281950, S269329, S282202, S282080, S284031, S269357, S120522, S296391, S289398, S296672, S282054, S109390, S109741, S289390, 6421521, S296363, S250132, S277983, S282196, S282184, 6429025, S289389, S296664, S289413, S289388, S296682, S284039, S296400, S289411, 6421520, 6421038, S215722, S215750, S269135, S269328, S278008, S289384, 6421063, S296659, S282395, S269333, S250059, S229473, S250161, S269345, S250159, S296372, S289400, S229441, S282402, 6458761, 6420951, S202797, S289406, 6421053, S250133, S202785, S269121, S259708, S229508, S202810, S109739, S296674, S296647, S271819, S282186, S282064, S269330, S120514, S296641, 6421168, 6421515, S268987, S250045, 6590750, S282071, S269119, S269380, 6420897, S289382, S222954, S282058, S259699, S269351, S250056, S130331, S296650, S271825, 6421149, S202817, S296398, 6590745, S271843, S222959, 6420931, S229475, S296671, S289395, S296685, S296675, S284043, S296373, 6429019, S269383, S284038, S296648, S269342, S281937, S109586, S271817, S202780, S202766, S289391, S281943, S229459, S269327, S282200, S296396, S284049, S289415, S282069, S109743, 6563416, S271820, 6590734, S233616, S269035, S218671, S250019, S269144, S269137, S284034, S269040, S229481, S219840, S269031, S119756, S202769, 6421065, 6563412, S269023, S284059, S296378, S229469, 6590723, S222949, S296361, S282204, 6421069, S271823, S130336, S278001, S119750, S277999, S119752, S222942, S296362, 6563396, S109414, S120539, S296386, S229499, S282068, S229479, S218694, S284041, S296673, S271810, S222956, S259210, S119762, S219828, S120527, S271811, S120550, S120555, S120551, S284028, S259205, S259215, S120546, S284027, S120554, S120547, 6421502, S222958, S278004, S229456, S250018, S120532, S271814, S229457, S119766, S229501, 6488019, 6283404, 6176263, S115980, 6176348, 6349604, 6541289, 6090581, 6387546, 6139075, 5898146, 6176362, 6110804, 5992438, 6387577, 6541319, 6050807, 6541296, 6541310, 6090568, 5899141, 6090580, 6541313, 6385208, 5985308, 6050790, 6176267, 6541317, 5985292, 6110809, S116020, 6371649, 6100605, 5953937, 6385211, 6387571, 5902849, 6139072, 6387572, 6287330, 6387551, 6253973, 6050758, 6176270, 5836405, 6110807, 6110800, 5868045, 6090573, 6139082, 5992457, 6299537, 6253984, 6078214, 6253925, 6541306, 6176350, 5953945, 6507160, 6225281, 6482605, 6387587, 5868052, 6349602, 5985301, 6349578, 6260431, 6349592, 6349584, 6349579, 5985299, 6488031, S116011, S282399, 6429032, S237087, S284057, 6458781, S284036, 6458772, 6563415, S229445, 6467489, S296365, S296651, S269382, S229464, S211014, 6563366, S119774, S277986, S130332, S282400, 6590707, 6475819, S282065, S218683, 6421163, 6590746, S282407, 6421067, S269360, 6563374, S250027, S282062, 6420948, 6421159, S202778, S120556, S109427, S289401, 6590752, S269341, S269129, 6563388, S202820, 6429016, S296663, 6590747, S248171, S269375, 6563405, 6429021, S269371, 6475823, S250038, S282387, S271831, S109389, 6563365, S296676, S237090, S296668, S229460, S250058, S278013, S269367, S277991, S269120, S289385, 6421514, 6421509, 6421095, 6590725, 6421150, S268998, S229496, 6421027, 6421073, 6563390, S269011, S119769, 6420889, S282053, S218680, S269103, 6590756, S282397, S282072, S109418, S282405, S229509, S259709, 6421498, S215721, 6421040, S248165, S296669, S222957, S109417, 6458765, S219830, 6421050, S269098, 6563369, 6650262, 6563393, 6590716, S269131, S218681, 6421080, S296401, S282386, S269024, S282203, 6650265, S109766, S271815, S130323, S282409, S269151, S269136, S259214, S296374, S259204, S296375, S269110, S284026, S120552, S296364, S109580, S248175, S120549, S202803, S259221, 6421507, S219827, S119781, 5985296, 5992458, 6541291, 6110805, 6429014, S282074, S284030, 6420888, S229470, S250122, S284023, S109756, S120524, 6563408;
|
| FEI Number |
1038671
|
Recalling Firm/ Manufacturer |
Exactech, Inc. 2320 Nw 66th Ct Gainesville FL 32653-1630
|
| For Additional Information Contact | Thien Doan 352-377-1140 |
Manufacturer Reason for Recall | The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken:
"" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required.
" IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing.
" EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession.
" COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com.
The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device.
Contact your local Exactech Representative if you have any questions regarding this communication. |
| Quantity in Commerce | 24303 units |
| Distribution | Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = PHX
|
|
|
|