|
Class 2 Device Recall Philips |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
May 03, 2024 |
Date Posted |
May 17, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1839-2024 |
Recall Event ID |
94508 |
510(K)Number |
K193215
|
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product |
Achieva 3.0T Model Number (REF): (1) 781345; (2) 781344; (3) 781278; (4) 781277; (5) 781177 |
Code Information |
UDI-DI:
(1) 781345 N/A;
(2) 781344 N/A;
(3) 781278 (01)00884838004139(21);
(4) 781277 N/A;
(5) 781177 N/A
Serial Numbers:
24079
34231
34256
34289
34220
34308
34139
34200
34213
34255
34258
34276
34301
34306
34020
34052
34082
34113
34120
34123
34230
34234
34235
34239
34240
34247
34260
34264
34270
34272
34274
34285
34287
34288
34297
34298
34299
34309
34312
34056
34282
34023
34144
34228
24115
34001
34095
34138
34143
34165
34269
34293
34300
34305
34115
34225
34253
34038
34047
34077
34126
34135
34160
34206
34254
34268
34275
34118
34237
34271
34273
34291
34242
34261
34290
34033
34104
34281
34086
34106
34125
34159
34241
34263
34286
34294
34295
34311
34022
34043
34208
34109
34130
34203
34207
34304
34074
34028
34029
34030
34034
34065
34066
34071
34073
34085
34201
34011
38295
38074
38191
38247
38254
38276
38326
38346
38355
38363
38285
38043
38073
38087
38092
38098
38108
38124
38140
38152
38157
38158
38171
38181
38185
38189
38190
38194
38196
38206
38209
38210
38212
38215
38216
38227
38231
38233
38236
38238
38240
38245
38248
38249
38252
38255
38257
38258
38260
38264
38266
38267
38269
38270
38271
38279
38282
38284
38287
38292
38293
38298
38299
38303
38304
38307
38314
38318
38319
38320
38321
38324
38327
38328
38329
38330
38332
38335
38338
38339
38340
38343
38344
38347
38349
38351
38358
38360
38361
38367
38368
38370
38372
38401
38402
38405
38407
38408
38410
38412
38414
38415
38420
38421
38423
38424
38425
38426
38427
38429
38430
38431
38432
38433
38436
38437
38438
38439
38441
38442
38443
38445
38447
38448
38450
38453
38454
38456
38457
38458
38128
38288
38082
38174
38253
38044
38049
38079
38139
38141
38151
38167
38170
38188
38220
38228
38127
38145
38009
38034
38162
38251
38336
38354
38359
38055
38056
38278
38207
38021
38031
38054
38059
38061
38086
38095
38104
38115
38133
38137
38169
38186
38187
38193
38223
38237
38262
38273
38280
38313
38272
38322
38323
38337
38345
38200
38268
38090
38172
38265
38046
38150
38197
38217
38225
38261
38067
38130
38022
38047
38093
38241
38060
38138
38017
38064
38069
38112
38113
38143
38144
38154
38160
38205
38229
38234
38246
38256
38274
38291
38296
38297
38315
38331
38099
38003
38011
38078
38006
38057
38072
38085
38283
38302
38004
38016
38018
38033
38091
38103
38173
38341
38232
382320
73005
73008
73009
73002
73003
73004
73006
73010
74027
74034
74035
74041
74009
74004
74039
74047
74052
74001
74002
74010
74020
74026
74033
74046
74032
74005
74007
74021
74029
74030
74044
74048
74037
|
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
For Additional Information Contact |
Customer Care Solution Center 800-722-9377
|
Manufacturer Reason for Recall |
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's
technical room, resulting in Smoke and/or fire
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Philips issued Urgent Medical Device Correction notifications to customers on 22-Apr-2024 via FedEx . Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take:
If a smoke/fire alarm is detected:
a. Immediately stop scanning and evacuate the patient and staff from the examination room.
b. If a developing fire is detected, adhere to established hospital fire emergency procedures, which may include switching off power to the complete system and/or removing the magnet field by using the Emergency Magnet Off button.
c. Do not attempt to continue scanning.
d. Immediately contact Philips Service.
" Ensure all users are aware of facility specific Emergency Procedures as outlined in Chapter 2: Safety in the Instructions for Use Emergency procedures The User is required to establish emergency procedures for the following situations:
" A medical emergency
" A fire
" An emergency that requires immediate removal of the magnetic field
" The release of helium gas into the examination room
Philips MRI systems have an Emergency Table Stop button in case there is an emergency
during tabletop movement.
" Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue.
" Post this notice near the affected MR system(s) for ease of reference.
Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt via email to: pd.cnr@philips.com. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken.
5. The actions planned by Philips to correct the problem
Philips will contact you to schedule time for a Field Service Engineer (FSE) to visit your site to inspect the g-MDU connections in the technical room and apply the proper torque to the connection if necessary.(reference FCO78100582). Philips plans t |
Quantity in Commerce |
386 units |
Distribution |
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada,
China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan,
Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia,
Spain, Sweden, Switzerland, Thailand, United Kingdom.
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = LNH and Original Applicant = Philips Medical Systems Nederland BV
|
|
|
|