|
Class 2 Device Recall Philips |
|
Date Initiated by Firm |
May 03, 2024 |
Date Posted |
May 17, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1851-2024 |
Recall Event ID |
94508 |
510(K)Number |
K213583 K193215
|
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product |
Ingenia Ambition X- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782138; (2) 782109; (3) 781356
|
Code Information |
UDI-DI:
(1) 782138 (01)00884838108622(21);
(2) 782109 (01)00884838098350(21);
(3) 781356 (01)00884838090040(21);
Serial Number
47194
47040
47092
47091
47048
47195
47345
47361
47050
47209
47246
47247
47293
47254
47275
47019
47049
47200
47356
47357
47197
47270
47013
47014
47015
47053
47065
47082
47099
47268
47274
47000
47002
47003
47017
47020
47024
47035
47047
47056
47058
47074
47078
47089
47097
47192
47225
47337
47226
47227
47341
47307
47364
47365
47372
47284
47350
47057
47075
47081
47088
47174
47175
47188
47193
47288
47052
47059
47072
47085
47184
47187
47211
47252
47262
47304
47328
47342
47384
47064
47244
47219
47220
47298
47094
47329
47096
47207
47283
47190
47264
47309
47034
47202
47315
47332
47367
47385
47016
47269
47382
47352
47012
47021
47221
47276
47297
47313
47314
47271
47281
47001
47032
47041
47248
47255
47258
47301
47311
47272
47306
47009
47023
47030
47080
47083
47189
47215
47222
47230
47231
47242
47245
47286
47299
47302
47310
47312
47004
47006
47007
47011
47018
47022
47025
47026
47027
47028
47029
47033
47037
47038
47039
47043
47044
47045
47051
47054
47060
47061
47062
47063
47066
47067
47068
47070
47071
47084
47090
47093
47095
47098
47177
47183
47185
47196
47198
47199
47203
47204
47205
47206
47208
47210
47216
47233
47237
47251
47253
47260
47278
47279
47280
47282
47289
47290
47291
47292
47294
47303
47305
47316
47322
47323
47324
47325
47330
47331
47333
47335
47336
47338
47339
47340
47344
47346
47347
47348
47351
47353
47370
47371
47377
47378
47380
47387
47173
47077
47657
47866
47531
47540
47878
47612
47717
47518
47519
47523
47526
47552
47555
47573
47584
47839
47565
47644
47742
47585
47582
47885
47669
47748
47755
47507
47508
47505
47570
47610
47500
47527
47529
47545
47640
47725
47778
47855
47532
47533
47547
47548
47591
47831
47843
47882
47607
47583
47625
47626
47627
47628
47629
47630
47675
47676
47786
47509
47525
47765
47848
47517
47504
47579
47683
47688
47737
47784
47808
47810
47812
47511
47520
47536
47543
47553
47561
47562
47564
47634
47646
47671
47681
47694
47698
47706
47708
47751
47603
47623
47705
47578
47516
47841
47502
47716
47754
47782
47783
47649
47656
47798
47665
47557
47589
47834
47876
47550
47790
47513
47602
47528
47770
47804
47860
47883
47785
47791
47874
47549
47635
47512
47648
47674
47696
47747
47806
47807
47864
47535
47588
47614
47615
47617
47651
47652
47660
47661
47662
47663
47709
47710
47711
47718
47720
47721
47722
47723
47724
47727
47728
47729
47730
47731
47780
47781
47851
47884
47503
47560
47530
47546
47563
47680
47685
47697
47811
47845
47566
47597
47679
47695
47753
47577
47701
47738
47514
47521
47522
47539
47559
47594
47595
47596
47622
47632
47666
47667
47677
47678
47684
47792
47801
47534
47538
47541
47542
47554
47558
47567
47568
47569
47572
47574
47575
47576
47580
47590
47598
47599
47606
47608
47621
47642
47653
47654
47659
47668
47672
47682
47687
47692
47700
47702
47703
47726
47732
47733
47739
47740
47759
47761
47762
47763
47767
47771
47789
47794
47796
47799
47800
47802
47814
47815
47816
47820
47824
47835
47849
47859
47861
47862
47888
47609
49004
49000
49091
49019
49062
49015
49044
49045
49052
49053
49058
49071
49054
49013
49047
49001
49002
49003
49066
49008
49009
49010
49064
49097
49017
49057
|
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
For Additional Information Contact |
Customer Care Solution Center 800-722-9377
|
Manufacturer Reason for Recall |
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's
technical room, resulting in Smoke and/or fire
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Philips issued Urgent Medical Device Correction notifications to customers on 22-Apr-2024 via FedEx . Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take:
If a smoke/fire alarm is detected:
a. Immediately stop scanning and evacuate the patient and staff from the examination room.
b. If a developing fire is detected, adhere to established hospital fire emergency procedures, which may include switching off power to the complete system and/or removing the magnet field by using the Emergency Magnet Off button.
c. Do not attempt to continue scanning.
d. Immediately contact Philips Service.
" Ensure all users are aware of facility specific Emergency Procedures as outlined in Chapter 2: Safety in the Instructions for Use Emergency procedures The User is required to establish emergency procedures for the following situations:
" A medical emergency
" A fire
" An emergency that requires immediate removal of the magnetic field
" The release of helium gas into the examination room
Philips MRI systems have an Emergency Table Stop button in case there is an emergency
during tabletop movement.
" Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue.
" Post this notice near the affected MR system(s) for ease of reference.
Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt via email to: pd.cnr@philips.com. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken.
5. The actions planned by Philips to correct the problem
Philips will contact you to schedule time for a Field Service Engineer (FSE) to visit your site to inspect the g-MDU connections in the technical room and apply the proper torque to the connection if necessary.(reference FCO78100582). Philips plans t |
Quantity in Commerce |
509 units |
Distribution |
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada,
China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan,
Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia,
Spain, Sweden, Switzerland, Thailand, United Kingdom.
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = LNH and Original Applicant = Philips Medical Systems Nederland B.V. 510(K)s with Product Code = LNH and Original Applicant = Philips Medical Systems Nederland BV
|
|
|
|