|
Class 2 Device Recall Philips |
|
Date Initiated by Firm |
May 03, 2024 |
Date Posted |
May 17, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1855-2024 |
Recall Event ID |
94508 |
510(K)Number |
K063559
|
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product |
Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350
|
Code Information |
UDI-DI:
(1) 782152 N/A
(2) 782110 (01)00884838099364(21)
Serial Numbers:
Serial Number
37019
37067
37149
37188
37184
37034
37036
37058
37167
37173
37183
37195
37197
37009
37011
37013
37020
37027
37031
37050
37055
37080
37093
37100
37121
37127
37132
37137
37141
37143
37150
37157
37170
37180
37193
37194
37023
37044
37066
37071
37106
37124
37160
37175
37182
37192
37022
37068
37082
37129
37134
37145
37156
37171
37181
37187
19083
37161
37178
37088
37024
37098
37135
37139
37176
37153
37060
37116
37196
37015
37073
37109
37122
37133
37162
37185
37026
37096
37177
37002
37012
37017
37029
37039
37041
37047
37048
37052
37053
37056
37077
37083
37084
37086
37087
37095
37102
37104
37107
37110
37113
37120
37123
37125
37128
37130
37136
37142
37158
37159
37163
37165
37168
37189
37069
37174
37126
|
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
For Additional Information Contact |
Customer Care Solution Center 800-722-9377
|
Manufacturer Reason for Recall |
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's
technical room, resulting in Smoke and/or fire
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Philips issued Urgent Medical Device Correction notifications to customers on 22-Apr-2024 via FedEx . Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take:
If a smoke/fire alarm is detected:
a. Immediately stop scanning and evacuate the patient and staff from the examination room.
b. If a developing fire is detected, adhere to established hospital fire emergency procedures, which may include switching off power to the complete system and/or removing the magnet field by using the Emergency Magnet Off button.
c. Do not attempt to continue scanning.
d. Immediately contact Philips Service.
" Ensure all users are aware of facility specific Emergency Procedures as outlined in Chapter 2: Safety in the Instructions for Use Emergency procedures The User is required to establish emergency procedures for the following situations:
" A medical emergency
" A fire
" An emergency that requires immediate removal of the magnetic field
" The release of helium gas into the examination room
Philips MRI systems have an Emergency Table Stop button in case there is an emergency
during tabletop movement.
" Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue.
" Post this notice near the affected MR system(s) for ease of reference.
Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt via email to: pd.cnr@philips.com. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken.
5. The actions planned by Philips to correct the problem
Philips will contact you to schedule time for a Field Service Engineer (FSE) to visit your site to inspect the g-MDU connections in the technical room and apply the proper torque to the connection if necessary.(reference FCO78100582). Philips plans t |
Quantity in Commerce |
118 units |
Distribution |
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada,
China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan,
Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia,
Spain, Sweden, Switzerland, Thailand, United Kingdom.
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = LNH and Original Applicant = PHILIPS MEDICAL SYSTEMS
|
|
|
|