| | Class 2 Device Recall ADVANTA VXT Graft |  |
| Date Initiated by Firm | May 03, 2024 |
| Date Posted | June 04, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1975-2024 |
| Recall Event ID |
94614 |
| 510(K)Number | K992960 |
| Product Classification |
Prosthesis, vascular graft, of 6mm and greater diameter - Product Code DSY
|
| Product | ADVANTA VXT, 6X50, 1GDS, NH, STR-TW. Single-Ended Slider GDS. Vascular graft |
| Code Information |
Product Code: 22175; UDI-DI: 00650862221756. |
| FEI Number |
3011175548
|
Recalling Firm/ Manufacturer |
Atrium Medical Corporation 40 Continental Blvd Merrimack NH 03054-4332
|
| For Additional Information Contact | Josephine McBride 973-709-6929 |
Manufacturer Reason for Recall | Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces. |
FDA Determined Cause 2 | Equipment maintenance |
| Action | Consignees were sent an URGENT MEDICAL DEVICE- REMOVAL notification by mail, dated 5/3/24. Consignees are instructed to examine their inventory immediately to determine if any recalled product is in their possession. Recalled product possesses the REF number on device labeling and has an expiration date prior to 24FEB2027. All consignees are asked to complete and return the provided response form by email to GDSgrafts2024@getinge.com or by fax to 1-866-560-4062. Consignees that are distributors are asked to forward the provided notification to their customers. Consignees are to post a copy of the recall notification in all inventory locations where affected devices are stored and ensure all appropriate personnel are aware. If consignees possess recalled devices, they are to contact their local Atrium/Getinge Customer Support Department at 888-943-8872 (option 2) to receive instructions for returning any affected devices. The recalling firm will provide replacement product within 5-8 weeks of return of devices. The recalling firm states that if there is a medical need to use the device and no alterative device is available, then the graft may be used per the provided instructions. Customers can contact Customer support with any questions at 888-943-8872 (option 2) from Monday through Friday, between 8:00 AM and 6:00 PM. |
| Quantity in Commerce | 53,308 units (US: 11, 236; OUS: 42,072) |
| Distribution | Worldwide distribution - US Nationwide including Puerto Rico, and the countries of AE, AR, AT, AU, BE, BG, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, ID, IE, IL, IQ, IT, JO, JP, KR, KW, LT, LU, LV, ME, MN, MU, MX, MY, NI, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, PY, RO, RU, SA, SE, SG, SI, SV, TH, TN, TR, TW, UA, VN, XK, YE, ZA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = DSY
|
|
|
|