| | Class 2 Device Recall BD BACTEC FX40 Instrument |  |
| Date Initiated by Firm | October 31, 2024 |
| Date Posted | December 03, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0594-2025 |
| Recall Event ID |
95636 |
| 510(K)Number | K915796 |
| Product Classification |
System, blood culturing - Product Code MDB
|
| Product | BD BACTEC FX40 Instrument
Version or Model: 442296
Catalog Number: 442296 |
| Code Information |
Catalog No: 442296; UDI-DI: 00382904422967; Serial Numbers: FF7385, FF7425, FF2727, FF1261, FF2717, FF1262, FF7254, FF7365, FF7419, FF7374, FF6762, FF2759, FF2756, FF7380, FF9548, FF9544, FF7367, FF2745, FF7615, FF2789, FF1371, FF7208, FF7481, FF7294, FF7225, FF7258, FF2760, FF9091, FF7875, FF6519, FF6835, FF2838, FF9860, FF7293, FF9668, FF9469, FF4440, FF9484, FF4506, FF9796, FF9602, FF2835, FF7238, FF2836, FF6984, FF9637, FF9689, FF9690, FF9700, FF9706, FF9743, FFA832, FFA846, FF9264, FF9734, FF6896, FF9735, FF9737, FF9800, FF2844, FF9633, FF0169, FF2954, FF1445, FF9838, FF3217, FF9642, FF9842, FF9339, FF1308, FF1314, FF1595, FF1341, FF1592, FF1375, FF9676, FF2875, FF2881, FF1448, FF9613, FF9614, FF1631, FF9271, FF9290, FF9753, FF9198, FF2324, FF9288, FF2914, FF5094, FF 2887, FF2978, FF9284, FF3960, FF2611, FF2979, FF4517, FF9278, FF4505, FF9279, FF9285, FF1414, FF9292, FF9297, FF9646, FF1417, FF5188, FF1460, FF2899, FF2921, FF1387, FF5098, FF9255, FF2922, FF9386, FF1391, FF1388, FFA256, FF9528, FF5178, FF9342, FF0756, FF6635, FF9447, FF6632, FF9639, FF6871, FF9641, FF2335, FF4464, FF9644, FF9458, FF9460, FFA366, FF5353, FF9479, FF9514, FF9522, FF7693, FF9594, FF7536, FF7687, FF7755, FF0877, FF0249, FF7510, FF7754, FF1837, FF7531, FF3484, FF3480, FF0988, FF3002, FF1501, FF0215, FF3385, FF3413, FF3384, FF1531, FF9015, FF3377, FF7547, FF7550, FF9022, FF9244, FF4570, FF7534, FF3218, FF3435, FF6636, FF7542, FF6885, FF3440, FF7182, FF9682, FF4627, FF7544, FF9684, FF7593, FF6663, FFA134, FF7538, FFA114, FF7555, FF7570, FF9000, FF7621, FF3549, FF7619, FF7622, FF9995, FF7694, FF7838, FFA052, FFA057, FF7887, FFA074, FF7688, FF7721, FF7719, FF7708, FF7711, FFA276, FF7845, FF4291, FF7842, FF7843, FF3553, FF7799, FF6390, FF6252, FF7095, FF3693, FF7765, FF3696, FF7771, FF3760, FF7775, FF3494, FF3516, FF9180, FF3589, FF9197, FFA616, FF3762, FF3899, FF9171, FF9182, FF3891, FF9188, FFA619, FF9190, FF4004, FF9124, FF4257, FF7770, FF4258, FF7773, FF4427, FF9193, FF4323, FFA744, FF2087, FFA973, FF2971, FF5701, FF7804, FF2171, FF2253, FF2874, FF2610, FF2100, FF2377, FF2431, FF9200, FF2151, FF9270, FF3868, FF3302, FF9273, FF2853, FF9275, FF2937, FF5770, FF2876, FF5092, FF9728, FF0248, FF9741, FF0607, FF1054, FF1107, FF3009, FF1404, FF1381, FF1585, FF0695, FF1998, FF3606, FF0171, FF0977, FF2024, FF1111, FF2025, ff6889, FF0598, FF1284, FF2379, FF9063, FF2219, FF6631, FF9053, FF2477, FF2972, FF2478, FF2141, FF7878, FF3318, FF3321, FF5300, FF0046, FF1997, FF2641, FF3609, FF3611, FF6728, FF0008, FF2697, FF0657, FF7444, FF2689, FF2685, FF0272, FF2776, FF9895, FF2782, FF9900, FF2783, FF6933, FF9729, FF0013, FF0265, FF9874, FF9404, FF3124, FF3120, FF2535, FF9877, FF9863, FF9866, FF9873, FF9876, FF3415, FF9882, FF9883, FF9894, FF9896, FF3740, FF2074, FF5192, FF0979, FF0652, FF1021, FF3819, FF3885, FF1871, FF0480, FF4164, FF9903, FF4370, FF4375, FF4460, FF4116, FF0057, FF4125, FF5656, FF4865, FF5723, FF4763, FFA039, FF4938, FF1833, FF1601, FF5696, FF1832, FFA733, FF0036, FF5461, FF5749, FF3190, FF9923, FF0047, FFA001, FFA002, FFA009, FF3716, FF0112, FF0478, FF0124, FF3768, FFA069, FF3763, FF3765, FFA040, FFA045, FF7369, FF6780, FF0978, FF6956, FF0175, FF3780, FF3801, FFA550, FFA552, FF1379, FF0291, FF0278, FF0346, FF0364, FF2788, FF0483, FF3794, FF0384, FF2941, FF0394, FF2966, FF0445, FF3122, FF0454, FF0453, FF0482, FF0486, FF3833, FF2189, FF2190, FF3391, FF3389, FF3394, FF0509, FF3839, FF0535, FF3047, FF0569, FF3043, FF3834, FF3050, FF3053, FF0011, FF0623, FF0122, FF3554, FF0627, FF4002, FF3608, FF4003, FF3519, FF0539, FF0770, FF1140, FF2426, FF0545, FF0440, FF0719, FF2590, FF0732, FF2593, FF2679, FF2684, FF1342, FF0168, FF1340, FF1349, FF1350, FF3040, FF1351, FF1352, FF0748, FF3123, FF0869, FF0253, FF4089, FF3237, FF0764, FF3380, FF1582, FF4096, FF0910, FF4100, FF0922, FF3228, FF4103, FF3442, FF3443, FF1455, FF3444, FF1579, FF4091, FF1892, FF4102, FF1891, FF4108, FF2071, FF3612, FF0833, FF1956, FF4107, FF0834, FF1955, FF4128, FF2206, FF0932, FF2212, FF5492, FF2110, FF3616, FF3619, FF2326, FF3620, FF2350, FF4000, FF0083, FF0320, FF1609, FF5613, FF4195, FF3623, FF0931, FF3799, FF0957, FF3927, FF1027, FF3790, FF5697, FF1608, FF6746, FF3558, FF3931, FF0651, FF0831, FF9631, FF5628, FF0237, FF3929, FF3976, FF0412, FF1113, FF1079, FF1083, FF1086, FF9021, FF0634, FF7852, FF0316, FF0639, FF7853, FF1184, FF4130, FF3982, FF0317, FF4272, FF4126, FF0325, FF0024, FF0321, FF4274, FFA006, FF4328, FFA087, FFA044, FF1224, FFA046, FF0324, FF4332, FFA049, FF4475, FFA081, FF0323, FF4416, FFA083, FFA084, FF4479, FF4481, FF0403, FF4431, FF4432, FFA230, FF9975, FF9773, FF1742, FF9981, FF4429, FF4434, FF6581, FFA194, FF4433, FF4209, FFA423, FF2985, FF4203, FF2616, FF1333, FF1305, FF1358, FF4204, FF2138, FF4539, FF2139, FF3520, FF6888, FF4543, FF9120, FFA754, FF1736, FF9049, FF1394, FF9001, FF1420, FF2989, FF1525, FF2936, FF1406, FF4568, FF4569, FF9380, FF9396, FF4572, FF4577, FF4245, FF4578, FF4868, FF3326, FF4275, FF2563, FF4870, FF4970, FF5025, FF5030, FF5466, FF1465, FF5068, FF4316, FF1399, FF4326, FF5071, FF1495, FF5265, FF4169, FF5161, FF1776, FF4349, FF2154, FF5348, FF5292, FF3838, FF3845, FF4391, FF5195, FF5198, FF1498, FF1533, FF5301, FF5545, FF5574, FF5737, FF5263, FF5279, FF5777, FF5721, FF4367, FF4392, FF5076, FF4371, FF4372, FF4415, FF5313, FF5151, FF5261, FF5262, FF1728, FF5763, FF6071, FF4390, FF1564, FF5347, FF6011, FF1853, FF2080, FF6341, FF2079, FF5396, FF4420, FF6564, FF5448, FF5480, FF2908, FF5504, FF1909, FF5508, FF1856, FF3092, FF1855, FF3103, FF5526, FF1854, FF2865, FF5652, FF6140, FF0211, FF1884, FF1946, FF2052, FF9414, FF5674, FF4494, FF5679, FF1878, FF2008, FF2009, FF0850, FF4497, FF1984, FF4504, FF2010, FF2007, FF0389, FF3096, FF3759, FF3468, FF1873, FF0897, FF4524, FF3948, FF3950, FF2220, FF1179, FF2994, FF1206, FF2210, FF6393, FF1671, FF1673, FF1134, FF7300, FF1135, FF2372, FF9616, FF2388, FF9467, FF1225, FF1237, FF2385, FF2345, FF2348, FF9623, FF7352, FF9889, FF0970, FF2400, FF7794, FF2404, FF2405, FF1073, FF1075, FF1076, FF1085, FF1090, FF1091, FF4679, FF1092, FFA119, FF1093, FF1096, FF0443, FF1402, FF0139, FF0138, FF0137, FF0437, FF0442, FF4715, FF1487, FF0135, FF0158, FF0155, FF0861, FF0444, FF2530, FF0095, FF2529, FF0448, FF0436, FF1979, FF1689, FF1696, FF1819, FF1707, FF1970, FF1749, FF1750, FF7488, FF1786, FF1784, FF6964, FF1787, FF7576, FF2792, FF1789, FF1792, FF2716, FF7493, FF7527, FF2746, FF3039, FFA498, FF2740, FF2919, FF2751, FF2221, FF3049, FF2222, FF7656, FF2444, FF2959, FF2773, FF0131, FF3240, FF7567, FF3797, FF3989, FF4055, FF4058, FF4114, FF2728, FF7675, FF4115, FF3216, FF2561, FF4171, FF2562, FF4315, FF2565, FF4173, FF2604, FF4767, FF2606, FF2597, FF2566, FF5051, FF2567, FF5052, FF2576, FF0078, FF5055, FF2572, FF5595, FF0397, FF5424, FF2580, FF6944, FF5226, FF4734, FF2663, FF6942, FF2666, FF2668, FF6073, FF6082, FF3227, FF7030, FF7033, FF6987, FF4748, FF4753, FF0074, FF1538, FF3676, FF2837, FF2011, FF4769, FF7015, FF2044, FF2065, FF4780, FF2069, FF4781, FF4787, FF0004, FF0030, FF0032, FF0065, FF0179, FF4796, FF7154, FF7155, FF4792, FF0595, FF1153, FF1094, FF1356, FF1357, FF1294, FF5260, FF1292, FF1288, FF7226, FF4897, FF2078, FF7241, FF0810, FF5432, FF7242, FF0813, FF2315, FF6613, FF6239, FF6886, FF7005, FF6240, FF7312, FF6247, FF7309, FF4816, FF6256, FF7189, FF7230, FF6080, FF6129, FF6403, FF7305, FF6585, FFA693, FF6583, FFA709, FF6285, FF0371, FF6192, FF6193, FF6195, FF6196, FF6198, FF6200, FF4835, FF0946, FF6205, FF6206, FF6212, FF6214, FF6288, FF0052, FF6349, FF6376, FF6379, FF6422, FF6424, FF6404, FF6405, FF6415, FF5924, FF6417, FF0338, FF6418, FF6429, FF0327, FF6430, FF0995, FF6727, FF6413, FF6425, FF0326, FF6716, FF7310, FF0395, FF0328, FF0330, FF2310, FF7164, FF7501, FF0335, FF7524, FF7120, FF7126, FF3364, FF7124, FF3877, FF4305, FF6750, FF6745, FF0166, FF9156, FF4926, FF9002, FF9062, FF9068, FF9085, FF0250, FF1183, FF9146, FF9067, FF9077, FF1230, FF4892, FF9220, FF9223, FF9149, FF4907, FF9153, FF4904, FF9649, FF4910, FF9216, FF5050, FF1405, FF9228, FF9400, FF5522, FF9497, FF9213, FF5606, FF9250, FF9607, FF5608, FF9402, FF9403, FF9408, FF9410, FF9411, FF5515, FF9413, FF0713, FF9559, FF5659, FF5655, FF5174, FF5172, FF5391, FF5579, FF4987, FF0393, FF7741, FF1040, FF9140, FF9144, FF0386, FF9141, FF9163, FF1016, Proto-003, FFPP02, FF2391, FF0092, FF1273, FF1763, FF5095, FF5117, FF1129, FF0619, FF9961, FF9966, FF1167, FF5148, FFA465, FFA633, FFA912, FFA913, FF0790, FFA917, FF0971, FF1757, FFA921, FF3840, FF0498, FF5233, FF1425, FF9300, FF0876, FF0410, FF9319, FF0884, FF5245, FF0885, FF5248, FF1361, FF1699, FF1663, FF0839, FF1326, FF0840, FF0841, FF0842, FF5242, FF0847, FF0845, FF1511, FF1769, FF1029, FF2056, FF2159, FF2146, FF2061, FF2063, FF2062, FF2147, FF1137, FF1896, FF6739, FF6742, FF1147, FF0561, FF0791, FF0792, FF1222, FF2425, FF1186, FF2725, FF0905, FF2736, FF0906, FF2732, FF0959, FF5632, FF0625, FF0587, FF1648, FF9364, FF1651, FF3490, FF3717, FF0499, FF3585, FF1760, FF4425, FF3678, FF1158, FF4426, FF5675, FF4101, FF9420, FF4255, FF4775, FF6567, FF4776, FF6571, FF4814, FF5731, FF4735, FF7422, FF7421, FF6753, FF4937, FF9499, FF6947, FF7061, FF5154, FF4860, FF9554, FF2805, FF2884, FF3077, FF1960, FF5246, FF5475, FF5747, FF5693, FF6293, FF1977, FF2038, FF2090, FF2093, FF3052, FF6518, FF1561, FF9209, FF5717, FF9425, FF6134, FF9730, FF7796, FF2357, FFA063, FF2192, FF6189, FF7760, FF7697, FF1888, FF6309, FF2592, FF6340, FF6359, FF0999, FF0506, FF1876, FF3792, FF7504, FF7505, FF5805, FF5796, FF5802, FF3420, FF7295, FF5797, FF3417, FF3816, FF3807, FF5771, FF2832, FF6313, FF6339, FF6345, FF3028, FF3906, FF0205, FF9014, FF0206, FF6370, FF0285, FF6792, FF9239, FF6834, FF6839, FF5806, FF6726, FF0368, FF0377, FF5426, FF0710, FF6729, FF5403, FF0380, FF5406, FF0381, FF0711, FF5411, FF6758, FF6675, FF0475, FF6864, FF1330, FF6767, FF6751, FF6771, FF0219, FF6989, FF0785, FF6733, FF0899, FF1009, FF1166, FF9253, FF1325, FF1003, FF0297, FF7137, FF7145, FF7153, FF7085, FF7041, FF0300, FF7298, FF7198, FF1449, FF1451, FF1661, FF1006, FF7039, FF1289, FF1008, FF7032, FF9388, FF6611, FF5795, FF5785, FF6378, FF7354, FF1739, FF1717, FF7356, FF1429, FF0715, FF7359, FF9440, FF7361, FF9438, FF1013, FF9435, FF0222, FF5827, FF7144, FF1654, FF7491, FF5807, FF7496, FF5809, FF0110, FF7499, FF1259, FF0230, FF9492, FF1264, FF9493, FF9501, FF3172, FF3173, FF1266, FF5916, FF1764, FF1766, FF3066, FF9502, FF3143, FF0617, FF2263, FF0089, FF9636, FF1267, FF0088, FF2503, FF9530, FF2505, FF2676, FF2677, FF5863, FF2764, FF2766, FF5875, FF2769, FF7784, FF9780, FF7829, FF9783, FF9784, FF3145, FF5860, FF5876, FF0502, FF3282, FF7832, FF9758, FF3283, FF1061, FF7839, FF3288, FF5879, FF5880, FF7844, FF5388, FF1064, FF5892, FF5893, FF9092, FF0512, FF9093, FF9094, FF0520, FF9095, FF9111, FF5043, FF0302, FF5045, FF0525, FF9099, FF5046, FF9105, FF9109, FF9114, FF5220, FF5048, FF5218, FF1515, FF5219, FF0303, FF1516, FF0530, FF0305, FF0531, FF9786, FF9491, FF1521, FFA042, FF0533, FFA117, FF6793, FF0758, FF4321, FF5678, FF6799, FF5680, FF0759, FF7274, FF5682, FF7276, FF6027, FF0761, FF7279, FF6329, FFA105, FF7509, FF7549, FF7575, FF7581, FF6063, FF0763, FF7605, FF0241, FF7616, FF7714, FF0773, FF7716, FF7718, FF6121, FF3684, FF7724, FF1587, FF1121, FF1122, FF9534, FF1123, FF9535, FF9332, FF9552, FF9553, FF9513, FF9678, FF3004, FF2329, FF9686, FF9957, FF9958, FF6184, FF9963, FF6171, FF6012, FF6179, FF6276, FF6574, FF6536, FFA139, FF6238, FF0948, FF6270, FF0953, FF0635, FFA055, FF1614, FF5966, FFA171, FF1348, FF1422, FF1403, FF1570, FF1624, FF6303, FF1857, FF2086, FF2208, FFA195, FF7577, FF9880, FF0186, FF9899, FF9954, FF0197, FF9685, FF9618, FF6332, FF9980, FF9905, FF1368, FF1416, FF0331, FF2371, FF0334, FF0434, FF7647, FF0333, FFA257, FF9907, FF6331, FF5106, FF9952, FF5115, FF0776, FF9955, FF0778, FF6347, FF0340, FF9964, FFA297, FFA298, FF6377, FF6381, FF1421, FF0342, FF1494, FF0343, FF0401, FF1897, FFA096, FF7429, FF0707, FF0264, FFA622, FF0269, FF6414, FF2763, FF6409, FF1216, FF1220, FF1226, FFA133, FF1145, FF0601, FFA303.FF1001, FF0600, FF1000, FF1285, FF0606, FF1479, FF1457, FFA227, FFA299, FF1625, FF1626, FF1627, FF1628, FF1623, FF1899, FFA470, FFA029, FF1850, FF1980, FF0947, FFA526, FF0965, FFA527, FF0366, FF1868, FF2248, FF2261, FFA500, FF2331, FF6555, FF6615, FF7583, FF6958, FF7584, FF7611, FF6587, FFA349, FF6605, FF1676, FFA056, FF0273, FF7349, FF0279, FF7358, FF1678, FFA370, FFA035, FF9621, FF9624, FFA833, FF5911, FFA473, FF9598, FF1686, FF6664, FFA838, FFA384, FF1691, FF3293, FF9777, FFA344, FFA383, FFA839, FF6768, FFA048, FF3731, FF2639, FF3164, FF3065, FF3698, FF3736, FF4516, FF3732, FF1655, FF5591, FF1656, FF3993, FF3879, FF5225, FFA435, FF4121, FF4122, FF4307, FF4188, FF1708, FF4340, FF4363, FF4364, FF3197, FF3080, FFA440, FF1670, FFA439, FF5898, FF4445, FF5906, FF4521, FF4530, FF2509, FF4708, FF2595, FF4723, FF6335, FF4733, FF2645, FF4746, FF2651, FF2656, FF0743, FF5156, FF5160, FF0747, FF5227, FF3741, FF6589, FF5600, FF2412, FF4093, FF5624, FF4094, FF4446, FF4187, FF4362, FF4361, FF4658, FF5642, FF4197, FF5598, FF5599, FF4845, FF5673, FF5146, FF5305, FF5138, FF5705, FF5291, FF5839, FF5398, FF5372, FF5983, FF5982, FF5896, FF5386, FF0870, FF5775, FF6086, FF6085, FF5913, FF5909, FF5686, FF6019, FF5331, FFA482, FF6504, FF6459, FF6266, FF7142, FF6269, FF7151, FF7149, FF6338, FF0123, FF0277, FF0307, FF0487, FF0541, FF0767, FF0851, FF0873, FF1067, FF1154, FF0217, FF1149, FF1117, FF1726, FF0628, FF1428, FF2218, FF2334, FF2403, FF6548, FF6386, FF6389, FF6899, FF9029, FF9031, FF5778, FF7526, FF6634, FF6662, FF7580, FF9756, FF9691, FF7652, FFA944, FF7179, FF5707, FF6001, FF6003, FF0806, FF5997, FFA699, FF7373, FF9628, FF0993, FFA525, FF9661, FFA627, FF9341, FF1695, FF2296, FF9226, FF9429, FF9448, FF9303, FF0145, FF9431, FF3704, FF3705, FF7710, FF9472, FF7707, FF9390, FF7759, FF3437, FF7761, FF3432, FF3433, FF9113, FF3454, FF9115, FF2571, FF2575, FFA706, FF3455, FF0820, FF7767, FF9225, FF9241, FFA708, FFA523, FF3711, FF9757, FF3713, FF3714, FF3715, FF3919, FF3926, FF9778, FF3603, FF3940, FF3952, FF4109, FF4137, FFA928, FF3593, FF3599, FF4217, FF4246, FF3944, FFA742, FF3946, FF3965, FFA965, FF4252, FF2742, FF2780, FF2312, FF2929, FF2721, FF2846, FF2849, FF4591, FF2946, FF2965, FF4585, FF4586, FF4587, FF2634, FF4575, FF2799, FF4406, FF3169, FF3200, FF3163, FF3178, FF4670, FF4672, FF2429, FF4673, FF3242, FF4600, FF3303, FF3306, FF4610, FF3307, FF4664, FF3309, FF3324, FF3328, FFA547, FF3335, FF3336, FF3342, FF3344, FF3347, FF3339, FF4250, FF3404, FF2636, FF4820, FF4659, FF4822, FF2707, FF4661, FF4826, FF2712, FF4667, FF4840, FF4841, FF2950, FF4863, FFA713, FF3118, FF4872, FFA779, FF3315, FF4875, FFA782, FF4882, FF4883, FF3369, FF4884, FF3398, FF4887, FF3400, FF4890, FF3419, FF4891, FF3645, FF4908, FF4895, FF3648, FF4915, FF4925, FF3707, FF4917, FF4056, FF0151, FF5813, FF4117, FF4606, FF4118, FFA977, FF4601, FF5001, FF5985, FF3025, FF4397, FF5042, FF0150, FF5995, FF4387, FF4919, FF6265, FF4351, FF4920, FF6268, FF2748, FF4452, FF3655, FF4466, FF6494, FF5134, FF4642, FF4111, FF4186, FF4513, FF4474, FF5804, FF4852, FF4851, FF4649, FF4652, FF4717, FF5358, FF1022, FF4831, FF4858, FF4855, FF3472, FF5067, FF1017, FF3626, FF5079, FF3697, FF3640, FF5066, FF3646, FF6544, FF3651, FF6593, FF5587, FF5351, FF5344, FF5346, FF5716, FF3493, FF5596, FF5742, FF5350, FF0793, FF0054, FF5828, FF5826, FF0919, FF0332, FF5399, FF0818, FF0982, FF5177, FF0989, FF6006, FF5186, FF1031, FF4990, FF0233, FF6208, FF5103, FF6135, FF1080, FF5184, FF1256, FF1257, FF6565, FF0823, FF1395, FF0311, FF5235, FF1509, FF1077, FF1666, FF1588, FF1667, FF5187, FF5236, FF3013, FF1233, FF1748, FF7138, FF1251, FF5328, FF1255, FF7028, FF2393, FF2394, FF1385, FF2618, FF5333, FF1746, FF9409, FF9032, FF9051, FF5867, FF9056, FF5870, FF9108, FF9232, FF9233, FF9234, FF9066, FF1803, FF1806, FF1810, FFA678, FF0212, FF6177, FFA703, FF0214, FF6232, FF0182, FF1844, FF2019, FF6222, FF1374, FF6225, FF2013, FF1849, FF6230, FF0256, FF6231, FF0723, FF3729, FF1908, FF6483, FF1023, FF0727, FF2306, FF0731, FF6445.FF0907, FF2389, FF6625, FF7562, FF1192, FF7627, FF1191, FF7629, FF7657, FF1948, FF7625, FF2016, FF7628, FF7689, FF1198, FF2018, FF3015, FF0741, FF1204, FF1210, FF7263, FF1207, FF3074, FF3079, FF7108, FF2960, FF1108, FF2993, FF2997, FF7192, FF2998, FF7234, FF0336, FF7245, FF7303, FF7434, FF0779, FF3534, FF7379, FF1807, FF1242, FF3539, FF7389, FF0784, FF9632, FF3540, FF3541, FF0576, FF9305, FF3550, FF9345, FF3561, FF9383, FF3563, FF9351, FF3564, FF9343, FF3567, FF0188, FF0181, FF3571, FF9461, FF3573, FF9464, FF3574, FF9474, FF1310, FF3576, FF1313, FF3652, FF0665, FF9525, FF7751, FF1070, FF1319, FF1069, FF3938, FF0671, FF7766, FF9247, FF1825, FF9251, FF9252, FF0686, FF7865, FF9764, FF9767, FF1105, FF9785, FF0804, FF0805, FF4396, FFA125, FF0807, FFA127, FF0811, FFA136, FFA138, FF1397, FF1842, FF1438, FF9627, FF1439, FF9374, FF1442, FF9375, FF1443, FF1583, FFA031, FF9999, FFA108, FF1504, FFA034, FFA426, FF0416, FFA429, FF1859, FF0425, FF0426, FF1864, FF0420, FFA368, FF1508, FFA367, FF1514, FFA425, FF6582, FFA400, FF6492, FFA404, FF1692, FF5755, FF5451, FFA505, FFA535, FFA536, FF1877, FF1902, FF5791, FFA614, FFA716, FFA717, FFA705, FF1788, FF1793, FF1860, FF1861, FF1865, FF1866, FF1797, FF1925, FF1930, FF1932, FF3061, FF4082, FF1904, FF4175, FF1935, FF4201, FF1936, FF4312, FF0102, FF0101, FF0071, FF4534, FF4435, FF2804, FF2814, FF2131, FF2132, FF2893, FF2149, FF2300, FF0704, FF2558, FF2573, FF2581, FF7559, FF7568, FF1774, FF2229, FF6509, FF0706, FF6515, FF6592, FF2239, FF0645, FF2275, FF2277, FF1344, FF6522, FF1961, FF0717, FF6599, FF1965, FF2615, FF6596, FF2613, FF6809, FF2282, FF0653, FF6811, FF2287, FF2284, FF0797, FF0714, FF6545, FF2341, FF2623, FF1983, FF2381, FF2027, FF2028, FF2029, FF2030, FF2392, FF2045, FF2395, FF2046, FF2048, FF2724, FF1773, FF1775, FF7088, FF7089, FF7086, FF1194, FF1799, FF3273, FF2112, FF2113, FF3486, FF3130, FF3277, FF9476, FF3185, FF2121, FF2815, FF2122, FF3205, FF3280, FF2215, FF3024, FF6468, FF3471, FF2396, FF2397, FF2703, FF3478, FF3459, FF3487, FF3204, FF3453, FF3458, FF9508, FF9509, FF7454, FF3464, FF9306, FF3479, FF0726, FF3062, FF3463, FF3465, FF3524, FF9436, FF3543, FF3577, FF9444, FF3578, FF9445, FF3580, FF9488, FF3462, FF1734, FF3466, FF2265, FF3635, FF3505, FF9489, FF1880, FF3641, FF3636, FF2269, FF3583, FF3278, FF2730, FF7459, FF2050, FF2733, FF2812, FF9985, FF7448, FF7453, FF3117, FF3836, FF3823, FF3828, FF3835, FF7461, FF3829, FF2274, FF7849, FF3887, FF2272, FF3896, FF3913, FF7835, FF7837, FF0428, FF9536, FF3837, FF9543, FF9539, FF3842, FF3874, FF3996, FF3892, FF4011, FF4036, FF4042, FF4044, FF4046, FF4053, FF4073, FF4075, FF4013, FFA446, FF5299, FF4932, FF9977, FFA008, FFA235, FF9909, FF4045, FF2540, FFA231, FF4038, FF2338, FF4144, FF4151, FF4153, FF2864, FF4182, FF3488, FF4020, FF2868, FF9915, FF4145, FF4247, FF4265, FF9819, FF4266, FF2542, FF3739, FFA269, FF2134, FF4267, FF2243, FF2544, FFA258, FF4047, FF3658, FFA396, FF4039, FF4027, FF2546, FF7786, FF4076, FF4149, FF3863, FFA609, FF4152, FFA268, FF2399, FF4713, FF2351, FFA738, FF2352, FF3895, FFA323, FF4342, FF2948, FFA453, FF2658, FF3005, FF2918, FF2439, FF3281, FF2924, FF3298, FF1975, FFA459, FF2537, FF2359, FF2932, FF3201, FF1573, FF2358, FF4236, FF2928, FF2057, FF2933, FF2953, FF2362, FFA681, FF2534, FF2366, FFA686, FF5609, FF2958, 151957-0, FF5829, FF4239, FF4628, FF3869, FF4177, FF4510, FF2547, FF4259, FF4439, FF4148, FF4727, FF4277, FF2549, FFA775, FF4264, FF4242, FF1985, FF2180, FF4223, FFA593, FF2188, FF4206, FFA625, FF2183, FF4237, FF2551, FF4224, FF4726, FF2186, FF4243, FF4263, FF2552, FF4939, FF2553, FF4228, FF3482, FF4214, FF4412, FF4418, FF5713, FF4281, FF2213, FF4288, FF4280, FF4408, FF4278, FF3187, FF2224, FF2824, FF5614, FF2827, FF3097, FF4409, FF3102, FF6065, FF6282, FF4618, FF3109, FF4620, FF3110, FF4560, FF4630, FF4565, FF4634, FF4574, FF4622, FF4632, FF4336, FF4284, FF4559, FF4702, FF2231, FF2040, FF2622, FF4557, FF2041, FF4699, FF4567, FF4629, FF2042, FF4563, FF3133, FF3407, FF4688, FF2043, FF2246, FF2251, FF4700, FF0062, FF4230, FF0067, FF4624, FF4219, FF1303, FF2301, FF2806, FF2489, FF4704, FF4626, FF1559, FF4686, FF0293, FF1545, FF4696, FF1737, FF3502, FF3162, FF1954, FF4562, FF2209, FF4229, FF2075, FF4692, FF3955, FF2343, FF4739, FF3184, FF4705, FF4742, FF3167, FF7321, FF9454, FF2322, FF4561, FF4980, FF9475, FF2323, FF4981, FF9450, FF2325, FF4959, FF5317, FF9449, FF4736, FF7330, FF4899, FF3231, FF5074, FF3214, FFA107, FF9763, FF6317, FF6236, FF4889, FF5124, FF5244, FF3238, FF2363, FF5387, FF1895, FF5253, FF2518, FF2519, FF5250, FF2521, FF5255, FF5435, FF1881, FF1887, FF5256, FF4810, FF2520, FF3059, FF5546, FF3188, FF3144, FF5429, FF5431, FF5257, FF1921, FF5290, FF1916, FF5447, FF4615, FF5485, FF2120, FF4616, FF4929, FF1924, FF2445, FF2466, FF4902, FF2126, FF2538, FF4741, FF3287, FF1169, FF2501, FF4930, FF3290, FF4934, FF1942, FF5072, FF1304, FF4697, FF2128, FF5020, FF5024, FF2487, FF3353, FF1605, FF3774, FF5126, FF5023, FF2647, FF5207, FF2411, FF5209, FF2649, FF5210, FF2652, FF2409, FF5293, FF4936, FF5471, FF2650, FF0565, FF5470, FF2654, FF5410, FF5436, FF5427, FF2177, FF5481, FF1926, FF2176, FF5360, FF2731, FF5384, FF2709, FF5444, FF5216, FF3274, FFPP16, FF5525, FF1316, FF2687, FF5472, FF5494, FF2702, FF5473, FF5363, FF5469, FF5496, FF5251, FF5468, FF5408, FF2473, FF5497, FF4690, FF3426, FF3427, FF5933, FF5938, FF5929, FF3723, FF5947, FF3720, FF2289, FF0176, FF3276, FF3811, FF0189, FF0193, FF6024, FF4080, FF5851, FF5953, FF5871, FF4308, FF6218, FF4309, FF5936, FF4311, FF6096, FF5861, FF2492, FF6093, FF5859, FF2504, FF6168, FF6223, FF5877, FF6244, FF6330, FF6119, FF6125, FF6117, FF6122, FF6115, FF6114, FF6128, FF5004, FF6772, FF6204, FF6217, FF5128, FF5131, FF6132, FF5168, FF6319, FF5163, FF6320, FF5167, FF6823, FF6637, FF6321, FF6324, FF6825, FF6643, FF5685, FF6443, FF5684, FF6447, FF6441, FF5321, FF6373, FF6365, FF2920, FF6374, FF6362, FF5597, FF6838, FF5780, FF6558, FF6506, FF6744, FF6706, FF6738, FF6737, FF7050, FF7111, FF7255, FF7270, FF2863, FF7272, FF7195, FF7316, FF7329, FF6216, FF7406, FF6708, FF7265, FF7266, FF7267, FF7269, FF7327, FF7326, FF7169, FF7187, FF7412, FF7407, FF6460, FF6469, FF6814, FF2934, FF6817, FF2940, FF2930, FF1278, FF1282, FF6820, FF6818, FF6520, FF6812, FF7641, FF6815, FF6816, FF6821, FF6826, FF3022, FF2905, FF0006, FF6646, FF6648, FF6654, FF6829, FF7659, FF6658, FF6842, FF7661, FF6653, FF6813, FF6844, FF7677, FF7681, FF6883, FF6808, FF6867, FF2340, FF6916, FF2375, FF6920, FF6915, FF6913, FF6917, FF7011, FF7013, FF6852, FF6851, FF6940, FF2416, FF7717, FF6473, FF7665, FF6909, FF6921, FF6924, FF6923, FF6927, FF6930, FF0007, FF4922, FF2447, FF7004, FF7772, FF6926, FF6561, FF6974, FF7780, FF7008, FF7012, FF7676, FF6992, FF7023, FF6928, FF2462, FF7014, FF6938, FF6380, FF7042, FF7001, FF7680, FF7052, FF2479, FF6929, FF2463, FF7043, FF7678, FF7044, FF7691, FF7062, FF0354, FF7021, FF7060, FF7163, FF6874, FF6953, FF9358, FF0973, FF9362, FF9043, FF9152, FF9354, FF9659, FF3365, FF9673, FF0375, FF9574, FF9583, FF9578, FF9660, FF9482, FF9568, FF9666, FF9585, FF9586, FF6976, FF7485, FF6994, FF3752, FF7730, FF3753, FF7128, FF7725, FF7828, FF7623, FF7734, FF3754, FF9046, FF3773, FF9044, FF9045, FF3772, FF7869, FF3771, FF9047, FF7737, FF3791, FF6991, FF9035, FF3789, FF7736, FF9187, FF9204, FF7817, FF9130, FF2511, FF9201, FF9205, FF2427, FF9206, FF7720, FF9315, FF9316, FF9320, FF2437, FF9322, FF2469, FF7860, FF9033, FF9040, FF9132, FF9090, FF9088, FF9147, FF0974, FF9036, FF1517, FF9145, FF9150, FF9122, FF9133, FF9041, FF9143, FF9083, FF9136, FF2704, FF9487, FF2705, FF9370, FF2750, FF9369, FF7415, FF2690, FF7411, FF2692, FF7410, FF7404, FF7388, FF7396, FF2637, FF7176, FF7393, FF2775, FF7482, FF7405, FF7476, FF2785, FF7417, FF2786, FF7477, FF2791, FF7598, FF7171, FF7101, FF7635, FF7106, FF7597, FF7607, FF7633, FF7125, FF7739, FF2802, FF7127, FF7746, FF2803, FF7727, FF9655, FF2833, FF9481, FF9584, FF3215, FF9801, FF0568, FF9587, FF7591, FF3387, FF3395, FF9371, FF0945, FF9575, FF1919, FF0673, FF9569, FF2935, FF3350, FF0641, FF9804, FF3381, FF0640, FFA094, FF3371, FF0663, FF7159, FFA099, FF1814, FF3316, FF0662, FF7170, FF9565, FF1711, FF3311, FF0571, FF1915, FF0642, FF1820, FF9802, FF3405, FF0660, FF0722, FFA150, FF9807, FFA102, FF9718, FF9726, FF0721, FFA142, FF0730, FFA151, FFA208, FFA211, FF9982, FF0771, FFA092, FF9857, FF9813, FF7282, FF7173, FF9572, FF7284, FF9567, FF9582, FF9806, FF7286, FF9570, FF7288, FF9808, FF7289, FF9716, FF1636, FF0893, FF7290, FF3418, FF0694, FF7292, FF9580, FFA060, FFA147, FF1078, FFA152, FFA209, FFA213, FFA252, FF9822, FFA287, FFA295, FF7319, FFA248, FF1336, FFA288, FF9878, FF1377, FFA296, FF0201, FF1360, FFA246, FF1367, FFA247, FF3512, FF3547, FF3370, FF3595, FFA325, FF3356, FFA326, FF3372, FF7331, FFA356, FF3383, FF7360, FFA407, FF3449, FF3665, FFA427, FF3508, FF3666, FF3533, FFA321, FF3354, FFA313, FF3373, FF1576, FFA311, FFA347, FF1643, FF3667, FF7432, FFA317, FF7433, FFA329, FF3670, FF2103, FF7435, FFA319, FF2106, FFA572, FFA065, FFA318, FFA485, FF3681, FF3357, FF3682, FFA464, FF3450, FF1911, FFA492, FF3531, FF3685, FFA320, FF3686, FF1872, FFA378, FF3545, FFA493, FFA487, FF2098, FFA491, FFA456, FF3687, FFA669, FF3617, FF7489, FF7512, FF2107, FF3601, FFA372, FF3830, FF6845, FFA636, FF3457, FF1830, FF3613, FF7560, FF1182, FF4395, FF3624, FF3631, FF7436, FF7178, FF7646, FF3855, FF3634, FF7558, FF7556, FF3517, FF0360, FF3629, FF6734, FF0044, FF7231, FF7246, FF1771, FF3518, FF7302, FF3529, FF7337, FF7377, FF2097, FF5323, FF0143, FF9629, FF3625, FF3755, FF9619, FF5298, FF2234, FF9662, FF2244, FF3628, FF3438, FF5463, FF3785, FF3800, FF9891, FF3802, FF3905, FF3470, FF3469, FF3796, FFA244, FF9515, FF5119, FF4052, FF3903, FF9518, FF4086, FF3908, FF9523, FF3914, FF4268, FF4253, FF3909, FF3956, FF4299, FF4300, FF3793, FF9010, FF9018, FF3910, FF4057, FF4290, FF4553, FF4081, FF7769, FF0254, FFA076, FFA088, FF1500, FFA090, FF3897, FF0697, FF4012, FF3968, FF3954, FF0933, FF3093, FF5288, FFA522, FF4896, FF4905, FFA116, FF5085, FF4161, FF5368, FF5704, FF5920, FF5823, FF5367, FF5377, FF6402, FF6436, FF6438, FFA131, FFA274, FF6537, FF5308, FF6419, FF4344, FF3975, FF6936, FF9862, FF6764, FF3964, FF9870, FF3978, FF1499, FF9071, FF1025, FF9127, FF9157, FF4376, FF4383, FF4385, FF3036, FF4026, FF3044, FF4030, FF3048, FF7553, FF7643, FF4382, FF5511, FF4022, FF4384, FF7824, FF7826, FF9721, FFA061, FF4032, FF6787, FF2943, FF6794, FF4457, FF2945, FF2939, FF2962, FF2983, FF4031, FF9433, FF3054, FF9714, FF3068, FF9723, FFA059, FF4465, FF3880, FF3149, FFA240, FF3151, FF4519, FFA468, FF2433, FF3089, FF4455, FF3087, FF2482, FF2483, FF4060, FF2493, FF2476, FF3254, FF9453, FF3761, FF9455, FF3871, FF3111, FF4531, FF2638, FF4522, FF4069, FF2578, FF3115, FF3114, FF4070, FF4318, FF2585, FF4324, FF2586, FF3876, FF2778, FF9470, FF4135, FF4179, FF3229, FF4141, FF4196, FF0950, FF4178, FF4765, FF4317, FF4191, FF3409, FF4322, FF3153, FF3139, FF4319, FF3445, FF3446, FF9541, FF4762, FF3653, FF4320, FF3659, FF4758, FF3974, FF3936, FF3961, FF3135, FF9596, FF5462, FF4403, FF4413, FF4811, FF3994, FF4849, FF4067, FF4422, FF4848, FF3951, FF4092, FF9669, FF9680, FF4441, FF3192, FF3923, FF3148, FF4190, FF3155, FF3259, FF4473, FF9710, FF9712, FF9717, FF4437, FF4725, FF4472, FF4724, FF4476, FF4946, FF4720, FF4722, FF3165, FF4731, FF4359, FF4495, FF5061, FF4471, FF5059, FF2867, FF5084, FF5086, FF0221, FF3063, FF3064, FF3150, FF3366, FF3160, FF4358, FF4529, FF4353, FF4356, FF5643, FF4523, FF5645, FF4528, FF9781, FF9766, FF5670, FF4526, FF5709, FF4544, FF4547, FF0086, FF4545, FF5394, FF4548, FF4542, FF4581, FF5787, FF5397, FF4580, FF4613, FF9821, FF3286, FF7877, FF3295, FF4582, FF4588, FF5925, FF5926, FF5976, FF4589, FF4607, FF6005, FF4583, FF3330, FF4608, FF4619, FF4643, FF7703, FF7726, FF6088, FF5647, FF6142, FF4654, FF7782, FF6219, FF7758, FF2523, FF2526, FF4660, FF2525, FF4655, FF4749, FF6336, FF6453, FF9016, FF7818, FF4754, FF4751, FF2527, FF7833, FF7836, FF7859, FF4770, FF0750, FF4745, FF5714, FF4804, FF4801, FF4809, FF4744, FF5654, FF4802, FF9058, FF4824, FF5342, FF5349, FF5306, FF4836, FF5356, FF4829, FF0457, FF9175, FF4834, FF4843, FF4873, FF2625, FF2627, FF1886, FF5743, FF5746, FF2628, FF9215, FF9219, FF4906, FF0452, FF5649, FF2667, FF4954, FF4961, FF5593, FF0511, FF4962, FF0510, FF4956, FF4958, FF4957, FF5700, FF4997, FF9941, FF4960, FF3495, FF5834, FF5835, FF5894, FF5895, FF3497, FF4912, FF0128, FF0245, FF4911, FF4913, FF0547, FF4996, FF0643, FF0243, FF5671, FF0526, FF9818, FF0314, FF9971, FF0609, FF0612, FF0566, FF5101, FF5951, FF0821, FF5110, FF9817, FF9839, FF0903, FF4989, FF1037, FF0676, FF1047, FF1053, FF3528, FF4998, FF1146, FF3525, FF5113, FF5116, FF0146, FF0680, FF9960, FF0679, FF1272, FF1276, FF1331, FF6013, FF1346, FF6394, FF6395, FF6396, FF9956, FF1669, FF0076, FF1724, FF0936, FF9996, FF6060, FF1950, FF1992, FF1995, FF3861, FF5272, FF1974, FF5414, FF1991, FF0555, FF0556, FF0562, FF2059, FF5283, FF2072, FF5378, FF5625, FF2201, FF9936, FF5988, FF2108, FF2109, FF3932, FF1953, FF3918, FF2369, FF3921, FF2367, FF5416, FF2441, FF5312, FF6087, FF6416, FF5454, FF6604, FF7617, FF7618, FFA203, FFA202, FF6255, FF0597, FF0604, FF6271, FF6272, FF0075, FF6300, FF5622, FF5657, FF5658, FF5439, FF5413, FF0690, FFA301, FFA444, FF5010, FF2674, FF5111, FF5143, FF5130, FF5112, FF2682, FF5114, FF6591, FF5145, FF5109, FF5197, FF2729, FFA280, FF2543, FF2238, FF1181, FF5002, FF1213, FF2273, FF2278, FF2279, FF2280, FF2281, FF1201, FF5139, FF1209, FF2619, FF2620, FF5122, FF2617, FF0162, FF5140, FFA556, FF2288, FF2286, FF1180, FF2298, FF5105, FF2290, FF2292, FFA374, FF0107, FFA560, FF1362, FF2701, FF5269, FF2713, FF2734, FF5273, FF2726, FF5738, FF5740, FFA592, FF5535, FF1616, FF1795, FF3125, FF3206, FF3500, FF5720, FF3182, FF3270, FF2822, FF3105, FF5726, FF5736, FF5665, FF3461, FF5733, FF3476, FF3202, FFA646, FF3523, FF3596, FF3597, FF5523, FF3485, FFA659, FF5644, FF5534, FF5537, FF5621, FF3506, FF5518, FF3515, FF3664, FF3642, FF5631, FF3649, FF5636, FF3637, FF3639, FF3654, FF3503, FF5822, FF3824, FF3832, FF3841, FF3827, FF5438, FF5528, FF3826, FF3820, FF5703, FF3822, FF3825, FF3890, FF3898, FF5695, FF3881, FF3882, FF5544, FF3893, FF5930, FF3821, FF3843, FFA767, FF2276, FFA755, FF1104, FF3886, FF3995, FFA781, FF4001, FF4005, FFA829, FF4008, FF4009, FF4010, FF4018, FF4023, FF5959, FF4037, FF5905, FF4043, FF4054, FF4072, FF3990, FF5961, FF5536, FF5380, FF5882, FF1101, FF5994, FF3003, FF3107, FF5986, FF4028, FF5952, FF4150, FF4154, FF4006, FF4147, FF4254, FF2241, FF2242, FF2130, FF4077, FF4079, FF5852, FF4335, FF3894, FF5960, FF5866, FF5846, FF2735, FF4337, FF4338, FF4339, FF5856, FF5853, FF3888, FF5971, FFA731, FFA915, FF5970, FF4227, FF6040, FF6035, FF6046, FF6054, FF3878, FF4238, FF4083, FF5854, FF4084, FFA911, FF4233, FFA916, FF4211, FF4221, FF4215, FF4220, FF4218, FF6033, FF6039, FF6052, FF4226, FF4040, FF4212, FF6030, FF4231, FF6021, FF4269, FF6044, FF4216, FF6032, FF6034, FF2891, FF4205, FF4208, FF3452, FF4235, FF4210, FF4285, FF4404, FF4405, FF4283, FF6036, FF2828, FF4222, FF4410, FF6031, FF4240, FF6107, FF4614, FF4279, FF4225, FF6111, FF4287, FF6116, FF4282, FF4213, FF4631, FF4633, FF4621, FF3346, FF6159, FF4625, FF3399, FF4635, FF3388, FF4558, FF3226, FF3481, FF4289, FF4564, FF6050, FF4556, FF4576, FF4701, FF6045, FF6053, FF4706, FF6120, FF4286, FFA990, FFA994, FF5869, FF4694, FF5969, FF4695, FF4691, FF5972, FF4159, FF6079, FF4555, FF6094, FF4232, FF6099, FF4234, FF4684, FF1632, FF1782, FF1638, FF6078, FF1645, FF6095, FF1647, FF1652, FF1630, FF5975, FF4707, FF4571, FF6069, FF4698, FF6068, FF6144, FF4931, FF4743, FF6058, FF4685, FF6260, FF4764, FF3510, FF3513, FF3530, FF3376, FF3396, FF6156, FF3501, FF4687, FF3509, FF4812, FF3514, FF3511, FF4900, FF4737, FF3447, FF3538, FF6263, FF6318, FF6630, FF5247, FF6262, FF5338, FF0294, FF5336, FF5032, FF5362, FF3532, FF5204, FF3448, FF5205, FF3456, FF0290, FF3000, FF5206, FF5252, FF3602, FF6261, FF5379, FF3604, FF5412, FF5940, FF5254, FF3627, FF4757, FF2599, FF5029, FF3630, FF4761, FF3355, FF2600, FF4991, FF5031, FF3618, FF5069, FF4689, FF3521, FF4901, FF3522, FF3632, FF5967, FF4752, FF3633, FF1532, FF2021, FF6185, FF5208, FF5213, FF5211, FF5214, FF4964, FF4898, FF5217, FF5440, FF5467, FF5428, FF0140, FF2981, FF6412, FF5430, FF6316, FF5442, FF5441, FF0203, FF4444, FF2125, FF4270, FF5479, FF6158, FF5249, FF3959, FF6186, FF6187, FF6188, FF5474, FF6367, FF5381, FF6368, FF1196, FF5458, FF5097, FF5490, FF6182, FF5459, FF6183, FF5493, FF3701, FF6406, FF5750, FF3702, FF5093, FF6505, FF5294, FF3757, FF5181, FF5200, FF0575, FF2821, FF5495, FF3691, FF5339, FF3700, FF3798, FF5175, FF5766, FF3094, FF5524, FF5759, FF5934, FF5932, FF6499, FF5935, FF3803, FF6508, FF5848, FF5864, FF6371, FF5948, FF3758, FF6479, FF5946, FF3795, FF3912, FF5928, FF5482, FF6018, FF1635, FF6075, FF6215, FF3900, FF6076, FF3907, FF6237, FF5858, FF3949, FF3969, FF6145, FF6154, FF5850, FF5258, FF3911, FF6465, FF6169, FF5865, FF6124, FF6092, FF3966, FF3980, FF6210, FF6366, FF6113, FF3963, FF6118, FF3970, FF3973, FF6131, FF6325, FF6486, FF6243, FF3962, FF6490, FF4894, FF6323, FF6628, FF5374, FF5280, FF5282, FF6439, FF5383, FF1755, FF6529, FF6534, FF6655, FF6711, FF6446, FF5314, FF6400, FF6629, FF6401, FF6510, FF6435, FF6516, FF6497, FF6388, FF6489, FF5316, FF6701, FF6524, FF6488, FF6464, FF6501, FF6513, FF3477, FF3902, FF6512, FF6757, FF6741, FF7250, FF3883, FF7121, FF7474, FF7053, FF7548, FF7175, FF7473, FF7391, FF7162, FF7471, FF4034, FF3884, FF7114, FF7262, FF7259, FF7328, FF7260, FF1917, FF7275, FF1910, FF7318, FF1912, FF7256, FF4033, FF2137, FF7472, FF6306, FF7196, FF4071, FF6467, FF7530, FF7403, FF6423, FF7479, FF6700, FF6797, FF7324, FF6702, FF7186, FF6233, FF6176, FF7190, FF6801, FF6496, FF6824, FF6819, FF6822, FF6798, FF6517, FF6649, FF4140, FF6463, FF6850, FF3875, FF6873, FF6846, FF6759, FF6657, FF4066, FF6832, FF6639, FF9059, FF6640, FF6837, FF6828, FF9701, FF6652, FF6761, FF6843, FF9054, FF6638, FF6879, FF6830, FF6878, FF9696, FF6659, FF5847, FF6919, FF4065, FF5390, FF6869, FF6908, FF9694, FF6647, FF6642, FF6922, FF6912, FF6914, FF6918, FF6995, FF3180, FF6712, FF6795, FF9852, FF6713, FF6791, FF6934, FF6806, FF6931, FF6932, FF6910, FF7016, FF9794, FF6925, FF6502, FF7027, FF4193, FF4198, FF6986, FF6384, FF6935, FF3268, FF7002, FF6717, FF7010, FF4194, FF6847, FF7051, FF2977, FF7167, FF2321, FF3647, FF6982, FF7193, FF3971, FF7313, FF7079, FF7113, FF7073, FF7074, FF7054, FF4424, FF7047, FF3972, FF4423, FF7040, FF4366, FF7025, FF7110, FF7112, FF7158, FF7177, FF7055, FF3984, FF7067, FF7261, FF7071, FF7075, FF9365, FF4599, FF9366, FF4928, FF7066, FF9563, FF5102, FF9683, FF5816, FF9600, FF5819, FF9657, FF9671, FF9589, FF7064, FF9571, FF4421, FF9576, FF5949, FF9665, FF7107, FF9667, FF4448, FF6074, FF4411, FF5445, FF9663, FF6364, FF0744, FF6167, FF4357, FF7139, FF7059, FF7130, FF7729, FF6907, FF7402, FF7486, FF1685, FF6905, FF9037, FF9081, FF7827, FF4442, FF6891, FF7814, FF6902, FF7819, FF4355, FF7222, FF7732, FF4496, FF7224, FF7606, FF7871, FF4354, FF7206, FF9087, FF7821, FF4401, FF7217, FF7823, FF7735, FF4360, FF2003, FF7742, FF2305, FF9034, FF7740, FF2497, FF7752, FF7248, FF7874, FF7253, FF7846, FF9192, FF9356, FF4546, FF7805, FF7285, FF7301, FF9208, FF9126, FF7862, FF9138, FF4623, FF9119, FF9139, FF9082, FF9203, FF9350, FF9084, FF9129, FF9131, FF9202, FF9485, FF2319, FF9494, FF9495, FF9151, FF9357, FF9486, FF3241, FF9135, FF2316, FF7345, FF9042, FF9352, FF9360, FF9211, FF9348, FF9089, FF7401, FF4617, FF7408, FF4644, FF3248, FF7430, FF3253, FFA772, FF7445, FFA771, FF7251, FF7325, FF7387, FF7386, FF4651, FF7467, FF4646, FF7470, FF3853, FF7174, FF3925, FF7400, FF3930, FF4656, FF7116, FF4584, FF2214, FF7636, FF4590, FF7667, FF7122, FF7631, FF4755, FF2866, FF7165, FF7390, FF2926, FF7500, FF3246, FF7600, FF7397, FF3260, FF3261, FF7731, FF3255, FF7398, FF3263, FF7483, FF7409, FF7416, FF3935, FF7634, FF7143, FF3924, FF7484, FF7166, FF5028, FF5035, FF9353, FF4768, FF5501, FF9137, FF9816, FF9656, FF5507, FF9654, FF9809, FF2938, FFA093, FF3673, FF2947, FFA098, FF4790, FFA100, FF4797, FF3262, FF9820, FF4803, FF9830, FF9797, FF9844, FF4134, FF4180, FF9724, FF4800, FF4189, FF4798, FFA091, FF4799, FF4181, FF4825, FF4334, FF9805, FFA143, FFA212, FF4489, FF9811, FF4331, FFA101, FF4491, FF4486, FF4830, FF9984, FF4832, FF9992, FF9581, FF9372, FF9558, FF4579, FF4592, FF9590, FF4593, FF9591, FF9592, FF7057, FF9828, FF9827, FF4596, FF9562, FF4785, FF2402, FF9579, FF2456, FF9848, FF9886, FF9986, FF9987, FF9993, FFA064, FF4789, FFA097, FFA207, FF2506, FFA089, FFA245, FF9815, FFA113, FF4982, FFA239, FFA251, FFA294, FF4979, FFA286, FF4945, FF4968, FF1089, FF4944, FF4903, FFA359, FF5572, FFA406, FF5573, FF2507, FFA416, FF4965, FF2508, FFA417, FF4969, FF5561, FFA418, FF5563, FFA428, FF4943, FF5552, FF4973, FF3317, FFA312, FF4951, FF0647, FFA328, FF4950, FF4971, FFA327, FFA316, FF4967, FFA315, FF4947, FFA314, FF5556, FF5555, FFA495, FF5557, FFA448, FFA449, FFA496, FFA486, FFA639, FF5567, FFA651, FFA467, FFA462, FFA457, FFA660, FFA818, FF4986, FF3688, FFA637, FFA262, FF5558, FF3750, FF5547, FF5551, FF5548, FF3751, FF0271, FF5014, FF5553, FF4966, FF5560, FF1664, FF4999, FF5549, FF5000, FF3776, FF1081, FF1363, FF5003, FF4974, FF5012, FF5089, FF5054, FF5034, FF4975, FF1790, FF1999, FF6363, FF5038, FF2099, FF6372, FF2435, FF5539, FF2194, FF5758, FF5285, FF2457, FF5757, FF5271, FF6051, FF6224, FF5287, FF5543, FF5415, FF5538, FF5238, FF0549, FF1659, FF3475, FF2768, FF2765, FF3669, FF0586, FF4861, FF0275, FF2810, FF0548, FF5153, FF2693, FF5266, FF0550, FF5329, FF0574, FF0581, FF2673, FF0584, FF4260, FF5284, FF2774, FF0251, FF4294, FF5315, FF0257, FF2795, FF4293, FF5542, FF0259, FF5540, FF4417, FF0267, FF4298, FF0263, FF2831, FF4301, FF4419, FF0689, FF4428, FF4550, FF4552, FF2843, FF4554, FF2696, FF2142, FF2852, FF5527, FF3026, FF5541, FF0868, FF5638, FF5409, FF2148, FF5434, FF5405, FF5286, FF5914, FF0856, FF6437, FF6531, FF6650, FF0867, FF0872, FF5531, FF6542, FF5640, FF0458, FF6946, FF9159, FF0880, FF0882, FF9227, FF0886, FF3313, FF5011, FF3411, FF3236, FF3338, FF5202, FF5201, FF7529, FF5107, FF3299, FF7630, FF0503, FF3340, FF3341, FF7868, FF5276, FF9715, FF5199, FFA058, FFA103, FFA135, FF3416, FFA242, FFA243, FF0500, FF3656, FFA331, FF3657, FFA466, FF5274, FF1469, FF5015, FF3663, FF1244, FF1245, FF1252, FF1720, FF5275, FF1458, FF1462, FF1463, FF1770, FF5135, FF1823, FF3746, FF1762, FF3692, FF1829, FF3805, FF5127, FF5141, FF1824, FF5142, FF5307, FF5310, FF1822, FF3854, FF3860, FF0461, FF5278, FF5277, FF5764, FF5270, FF6661, FF5725, FF6666, FF6672, FF6680, FF3852, FF6681, FF6682, FF6683, FF5324, FF6684, FF6692, FF5663, FF6694, FF5668, FF6689, FF5464, FF6690, FF5667, FF6687, FF6691, FF6697, FF6686, FF6705, FF6696, FF5641, FF5520, FF3865, FF5630, FF0460, FF3915, FF3916, FF5629, FF5533, FF5529, FF4061, FF3872, FF5530, FF5453, FF5450, FF0372, FF5437, FF0370, FF5820, FF9118, FF9004, FF9017, FF9079, FF5702, FF5817, FFA154, FFA166, FFA164, FF5692, FFA167, FF5694, FFA201, FF5637, FFA200, FFA191, FFA192, FF5943, FFA160, FF5945, FFA161, FFA169, FFA178, FF5760, FFA189, FF5899, FFA184, FFA224, FFA225, FFA226, FF4166, FF5886, FF4306, FF5901, FF4304, FF5903, FF5937, FF4381, FF4386, FF5955, FF5956, FF4400, FF5958, FF4399, FF5904, FF5910, FF4377, FF9922, FF5964, FF5532, FF9921, FF5639, FFA174, FF5881, FFA156, FF5888, FF5993, FFA219, FFA358, FFA436, FF5990, FF5991, FFA393, FF4456, FFA649, FF4469, FFA784, FFA630, FF4525, FF2817, FF4454, FF5849, FFA793, FFA991, FF0557, FF4638, FF4537, FF4014, FF4099, FF4436, FF6049, FF2490, FF6042, FF6756, FF9789, FF6048, FF1530, FF0315, FF0213, FF0209, FF6023, FF4295, FF6618, FF6047, FF9751, FF9792, FF6043, FF9704, FF5842, FF6017, FF9116, FF3223, FFA843, FF9859, FF6963, FF6022, FF6038, FF6112, FF4710, FF4711, FF6057, FF3219, FF6161, FF3431, FF6061, FF6041, FF2035, FF2036, FF2448, FF6064, FF6097, FF6098, FF6103, FF4846, FF6104, FF6106, FF6109, FF2378, FF6110, FF5491, FF6100, FF5425, FF2964, FF2907, FF6066, FF3527, FF3032, FF4518, FF6157, FF2446, FF4503, FF4443, FF6149, FF6155, FF5182, FF4949, FF5005, FF4566, FF5176, FF6152, FF5164, FF6360, FF5170, FF6361, FF6875, FF2967, FF6884, FF5062, FF5063, FF5421, FF5064, FF5963, FF6895, FF6872, FF5080, FF6355, FF6160, FF6259, FF5965, FF5974, FF5123, FF5125, FF6408, FF5058, FF6327, FF6322, FF7881, FF6407, FF6500, FF5973, FF6328, FF5968, FF6305, FF5592, FF6462, FF6472, FF6352, FF6392, FF6440, FF5619, FF5620, FF6387, FF6523, FF6532, FF0053, FF6710, FF6470, FF6805, FF6800, FF2612, FF7478, FF6258, FF7466, FF6264, FF7551, FF3101, FF4499, FF6868, FF7468, FF5653, FF6709, FF6803, FF6730, FF6607, FF6610, FF6720, FF6665, FF6667, FF6668, FF6827, FF6722, FF6714, FF7850, FF6715, FF5718, FF5748, FF6015, FF6901, FF5729, FF6983, FF6988, FF5650, FF7078, FF7072, FF6981, FF7096, FF7081, FF5825, FF7098, FFA343, FF7076, FF7070, FF5788, FF7056, FF7077, FF7097, FF7103, FF7102, FF5889, FF5890, FF5891, FF0246, FF7134, FF5857, FF5845, FF7105, FF6880, FF7087, FF7117, FF1240, FF7136, FF1063, FF6906, FF0700, FF5908, FF7184, FF7218, FF5902, FF5927, FF7205, FF7215, FF7213, FF7219, FF7211, FF7209, FF7247, FF7249, FF7252, FF5999, FF2888, FF7216, FF7346, FF2386, FF7347, FF2374, FF7350, FF2621, FF7322, FF5418, FF5419, FF7323, FF7332, FF7334, FF1407, FF1526, ff6887, ff6890, FF9702, FF2089, FF7336, FF2092, FF7342, FF6102, FF2088, FF6026, FF2082, FF5149, FF2252, FF7351, FF9988, FF7341, FF7339, FF7343, FF7348, FF7420, FF7427, FF7441, FF7344, FF0341, FF7363, FF0344, FF7366, FF0477, FF7370, FF7381, FF7382, FF6454, FF4273, FF6455, FF6457, FF4088, FF0027, FF2536, FF7480, FF6302, FF2254, FF3029, FF3679, FF1994, FF6391, FF7428, FF6248, FF7475, FF3491, FF3537, FF3582, FF9480, FF3584, FF3889, FF9601, FF3985, FF3983, FF3986, FF4021, FF4024, FF9635, FF9688, FF4261, FF9693, FF9698, FF5231, FF9705, FF9740, FF9736, FF5259, FF5267, FF5918, FF9738, FF3294, FF9742, FF9793, FF2153, FF9798, FF2438, FF2694, FF3301, FF3360, FF9833, FF9836, FF1835, FF2068, FF1221, FF2443, FF9183, FF1831, FF0702, FF9177, FF0474, FF1175, FF0968, FF9191, FF6784, FF1390, FF1281, FF1444, FF1297, FF0969, FF0984, FF1301, FF0985, FF1048, FF1431, FF1435, FF9287, FF2587, FF9274, FF9196, FF9280, FF9194, FF9277, FF9286, FF6997, FF7020, FF9276, FF5885, FF9477, FF7613, FF9727, FF2598, FF3778, FF9640, FF2603, FF9643, FF9648, FF9533, FF9526, FF9550, FF9478, FF9551, FF9711, FF9720, FF0025, FF7539, FF7540, FF7535, FF7537, FF7692, FF7514, FF7532, FF7437, FF7438, FF7464, FF0692, FF7543, FF7588, FF7592, FF0687, FF7533, FF1052, FF7541, FF7557, FF7571, FF7573, FF7594, FF7620, FF7624, FF7614, FF7705, FF7709, FF7712, FF7841, FF1540, FF1541, FF7695, FF7840, FF7792, FF9189, FF9186, FF7774, FF9185, FF0119, FF9179, FF5423, FF7793, FF5884, FF7885, FF0236, FF0043, FF9269, FF0649, FF9281, FF9283, FF7800, FF9195, FF0415, FF0413, FF6580, FF2140, FF2986, FF4514, FF9731, FF9744, FF9845, FF9865, FF9867, FF9868, FF9872, FF9875, FF9879, FF6911, FF9064, FF6633, FF5352, FF1883, FF1923, FF0045, FF0082, FF9925, FF9902, FF9917, FF9919, FF9938, FF9824, FFA005, FFA023, FF1777, FFA068, FFA036, FFA037, FFA551, FF2790, FF2710, FF2787, FF2900, FF0084, FF3038, FF2187, FF2185, FF2902, FF3067, FF3078, FF3082, FF4482, FF3045, FF3034, FF3605, FF3557, FF2909, FF3083, FF3084, FF3088, FF3091, FF3100, FF3161, FF0051, FF5345, FF2360, FF2813, FF2361, FF2834, FF2376, FF2872, FF2428, FF2528, FF0041, FF0042, FF2532, FF5801, FF2533, FF2591, FF2596, FF2678, FF2680, FF2681, FF5800, FF5900, FF6070, FF0055, FF3041, FF0056, FF6277, FF0063, FF6334, FF3234, FF3378, FF3441, FF0021, FF0035, FF6344, FF6342, FF3382, FF3555, FF3556, FF3375, FF3610, FF3614, FF0033, FF3615, FF0421, FF0864, FF0113, FF0114, FF0121, FF1176, FF1178, FF7133, FF3977, FF3979, FF1644, FF1703, FF0199, FF9106, FF3987, FF4097, FF0390, FF9620, FF3981, FF4095, FF0392, FF4468, FF4329, FF4477, FF4480, FF4414, FF0411, FF4478, FF0455, FF0449, FF4484, FF1163, FF2798, FF3057, FF4430, FF4538, FF4540, FF4450, FF2626, FF2686, FF0507, FF4541, FF1885, FF3474, FF0431, FF1976, FF4327, FF4485, FF2583, FF0620, FF2718, FF4573, FF2719, FF4782, FF4784, FF4791, FF4794, FF4795, FF2722, FF2974, FF0630, FF0887, FF0890, FF1839, FF2737, FF2951, FF4876, FF2743, FF4877, FF2165, FF5026, FF0740, FF0513, FF0736, FF0544, FF2555, FF1232, FF2556, FF2557, FF1234, FF2560, FF2564, FF5073, FF2605, FF0749, FF2607, FF2609, FF5264, FF2594, FF2568, FF5166, FF0987, FF0994, FF2662, FF1311, FF0801, FF2669, FF2670, FF2714, FF2715, FF1087, FF1088, FF0701, FF0909, FF0359, FF0362, FF1554, FF5077, FF0924, FF1693, FF1590, FF1593, FF1738, FF1772, FF5078, FF5150, FF0942, FF2055, FF5155, FF5735, FF0085, FF0018, FF5722, FF0009, FF0077, FF5728, FF0010, FF5732, FF0225, FF0668, FF5289, FF0941, FF0958, FF1014, FF5745, FF1042, FF5753, FF1045, FF5744, FF5364, FF5452, FF1170, FF5498, FF1279, FF5505, FF5499, FF1231, FF0149, FFA111, FF5602, FF1335, FF1019, FF1334, FF5618, FF5610, FF3544, FF3675, FF3694, FF5611, FF5719, FF6139, FF1130, FF5616, FF1413, FF2873, FF0347, FF2311, FF1466, FF1467, FF6432, FF5268, FF5708, FF5088, FF6249, FF6314, FF6245, FF5075, FF6292, FF6308, FF6191, FF6123, FF6194, FF6197, FF6199, FF6201, FF6202, FF1496, FF6126, FF1497, FF6203, FF6235, FF6350, FF6353, FF6354, FF6358, FF6383, FF6421, FF6426, FF6428, FF6427, FF1539, FF6433, FF6721, FF6718, FF6719, FF6723, FF6566, FF1544, FF6579, FF7123, FF7180, FF7521, FF7522, FF7525, FF7552, FF7022, FF7024, FF7045, FF1543, FF7119, FF7118, FF6749, FF6503, FF9069, FF9075, FF9086, FF9123, FF9162, FF1563, Proto-002, FFPP09 , FF9222, FF9314, FF1580, FF9397, FF1735, FF9074, FF9212, FF9248, FF9325, FF9328, FF9347, FF9405, FF9406, FF9556, FF9560, FF9561, FF9564, FF7523, FF1550, FF1057, FF9080, FF1062, FF9158, FF9160, FF1565, FF9072, FF9076, FF9142, FF1907, FF0096, FF0136, FF1890, FF0157, FF5782, FF0152, FF0097, FF0837, FF0490, FF0447, FF1996, FF0094, FF2000, FF0111, FF2060, FF0093, FF0091, FF0153, FF0451, FFA577, FF0154, FFA578, FFA631, FFA914, FFA918, FF7806, FFA919, FFA638, FF3992, FF1202, FF1205, FF0280, FF0276, FF1190, FF1171, FF0921, FF0551, FF0553, FF0554, FF1982, FF1353, FF0693, FF0696, FF0699, FF1380, FF0836, FF2051, FF0838, FF1552, FF4016, FF3945, FF0843, FF0844, FF1752, FF1785, FF2198, FF6577, FF1945, FF6606, FF1967, FF2169, FF6804, FF6754, FF6949, FF6950, FF2723, FF6960, FF6957, FF7306, FF7520, FF7639, FF2076, FF3362, FF3680, FF4041, FF4048, FF4049, FF4050, FF4051, FF4244, FF3128, FF4292, FF3129, FF2271, FF4241, FF3212, FF2342, FF2387, FF4778, FF2285, FF4815, FF4827, FF4738, FF4774, FF5169, FF6420, FF3594, FF1813, FF3361, FF3846, FF1879, FF5476, FF5922, FF2384, FF2383, FF5912, FF2344, FF5919, FF2184, FF2382, FF5915, FF2155, FF6848, FF6541, FF9359, FF9326, FF2337, FF1668, FF2418, FF2398, FF3011, FF9363, FFA145, FFA238, FF2191, FFA249, FFA431, FFA795, FF3815, FF3817, FF3818, FF0894, FF3788, FF3814, FF3422, FF0846, FF0848, FF0788, FF3410, FF0745, FF1162, FF1208, FF1679, FF1660, FF3904, FF3901, FF0286, FF7654, FF0287, FF0208, FF0289, FF0288, FF7655, FF1662, FF0379, FF0712, FF7683, FF7684, FF6755, FF7696, FF0218, FF1657, FF1658, FF0296, FF1712, FF6945, FF6671, FF1004, FF1450, FF0039, FF1440, FF6972, FF7026, FF1036, FF1007, FF0058, FF1010, FF1005, FF1761, FF7031, FF0960, FF0962, FF0787, FF0090, FF0720, FF0621, FF0497, FF0902, FF1263, FF7194, FF7197, FF7236, FF0963, FF0613, FF1265, FF0615, FF0895, FF0901, FF7243, FF9974, FF0616, FF0542, FF1268, FF1328, FF7315, FF1164, FF7093, FF1011, FF7203, FF1269, FF7317, FF1571, FF1056, FF0964, FF1139, FF1012, FF1296, FF1118, FF0517, FF0560, FF7372, FF1068, FF0521, FF0972, FF1071, FF0524, FF0622, FF0522, FF0591, FF0588, FF0527, FF0304, FF0794, FF0308, FF1518, FF0310, FF1520, FF0532, FF1759, FF9324, FF0534, FF9519, FF2470, FF1028, FF3727, FF1527, FF3726, FF0762, FF3767, FF2582, FF1524, FF1329, FF3430, FF3412, FF4062, FF4064, FF1747, FF9638, FF6299, FF9652, FF1781, FF9498, FF1809, FF1817, FF1811, FF0129, FF0765, FF0766, FF1034, FF0242, FF1039, FF1821, FF0768, FF1963, FF0769, FF1968, FF1488, FF1969, FF1489, FF0774, FF0929, FF2039, FF2207, FF2091, FF7578, FF1119, FF1577, FF6136, FF1120, FF9787, FF0563, FF1370, FF2440, FF6357, FF9754, FF1124, FF1126, FF6343, FF1125, FF6346, FF1127, FF1128, FF1493, FF7502, FF1491, FF7503, FF1131, FF7506, FF7507, FF0943, FF2102, FF3195, FF9301, FF0944, FF9296, FF9298, FF0949, FF0951, FF6369, FF1591, FF6853, FF6863, FF1615, FF6996, FF6725, FF6882, FF1523, FF9335, FF9231, FF5099, FF1382, FF1621, FF9003, FF1620, FF9006, FF7297, FF3749, FF0462, FFA472, FF0703, FFA645, FFA688, FF0709, FFA590, FF0708, FF9418, FF9430, FF0997, FF9424, FF1002, FF7355, FF9245, FF9421, FF7357, FF0608, FF7487, FF3392, FF9510, FF4132, FF4133, FF9240, FF9505, FF5215, FF0367, FF9532, FF7781, FF9610, FF7783, FF7785, FF9609, FF9608, FF7830, FF7831, FF7702, FF7848, FF7698, FF9025, FF5589, FF9096, FF0385, FF9097, FF9098, FF9100, FF9102, FF9103, FF9110, FF9121, FF9125, FF0387, FF9168, FF9490, FF1688, FF7278, FF1355, FF1159, FF1687, FF7271, FF7277, FF9243, FF9242, FF7280, FF9249, FF7508, FF7141, FF7582, FF7148, FF7749, FF9254, FF7723, FF9670, FF9679, FF9681, FF9832, FFA054, FF7181, FF7183, FF1700, FF7528, FF9176, FF9962, FF9834, FF1702, FF1704, FF1706, FF9826, FF9869, FF9803, FF9825, FF9871, FF9887, FFA524, FF9904, FFA520, FF9890, FFA620, FF9906, FFA683, FF9779, FF9970, FFA095, FFA106, FFA175, FF6854, FF6856, FFA214, FF7704, FFA217, FFA220, FFA300, FFA302, FFA304, FFA043, FF0115, FFA305, FF2531, FFA104, FFA306, FFA307, FFA308, FFA309, FFA310, FFA348, FF2691, FF2695, FF2699, FF2700, FFA352, FFA725, FFA748, FFA836, FFA837, FFA993, FFA124, FF2857, FF2858, FF4817, FF4818, FF4819, FF4823, FF4837, FF4839, FF4842, FF4856, FF4871, FF4874, FF4880, FF4881, FF4885, FF4886, FF2496, FF4893, FF5772, FF2640, FF2664, FF2671, FF5992, FF2495, FF2744, FF2747, FF2842, FF3181, FF3076, FF6576, FF3689, FF3690, FF3735, FF3730, FF9782, FFA511, FF3813, FF3668, FF3779, FF3786, FF3812, FF2895, FF3483, FF5322, FF3622, FF5402, FF3644, FF5584, FF3106, FF3991, FF4029, FF4120, FF4119, FF4143, FF4165, FX0830, FF4146, FF4163, FF0981, FF0986, FF4343, FF0322, FF1199, FF1200, FFA254, FF0564, FF0567, FF0610, FF0626, FF4350, FF1260, FF4369, FF1378, FF4389, FF1432, FF1112, FF4461, FF1753, FF1253, FF1254, FF4637, FF4636, FF2797, FF4527, FF4729, FF4657, FF4850, FF4847, FF5082, FF5159, FF5147, FF5162, FF5228, FF5223, FF5585, FF1133, FF1802, FF1808, FF5617, FF5676, FF1812, FF5371, FF5762, FF1846, FF1847, FF5687, FF6083, FF6062, FF0912, FF6267, FF0914, FF1189, FF1193, FF1195, FF0495, FF1197, FF0734, FF0738, FF1211, FF1815, FF0918, FF0920, FF1818, FF0629, FF1214, FF5979, FFA053, FF1235, FF1816, FF0782, FF1243, FF0577, FF0306, FF1990, FF6669, FF6670, FF7650, FF0190, FF7414, FF6747, FF7036, FF7109, FF1315, FF1026, FF7442, FF1033, FF7443, FF0666, FF7371, FF1318, FFA460, FF1065, FF1321, FF1322, FF1324, FF1072, FF0688, FF9221, FF1074, FF9224, FF0691, FF1084, FF9329, FF9331, FF9333, FF9229, FF9238, FF1102, FF6535, FF1110, FF0234, FF7750, FF0231, FF0232, FF1132, FF7854, FF7856, FF0808, FF1364, FF0822, FF7273, FF1396, FF7815, FF7857, FF1398, FF9117, FF1400, FF1401, FF1408, FF7808, FF7810, FF1430, FF1434, FF1569, FF1581, FFA442, FF1594, FF0423, FF0427, FF1613, FF1617, FF1618, FF1513, FF1680, FF1681, FF1758, FF1794, FF1863, FF1862, FF1798, FF1796, FF1800, FF1801, FF1928, FF1931, FF1929, FF1933, FF1938, FFA615, FF1937, FF0108, FF0103, FF0079, FF0099, FF0080, FF0105, FF0081, FF0070, FF0106, FFA443, FF2811, FF2816, FF2823, FF2124, FF2129, FF2127, FFA644, FF2133, FF2150, FF2152, FFA712, FFA762, FFA765, FFA689, FF5500, FF4110, FF4112, FF4378, FF4379, FF4438, FF4853, FF4867, FF5603, FF5677, FF5691, FF0391, FF0661, FF7494, FF7812, FF6399, FFA783, FF7747, FF7763, FF9107, FF9030, FF9235, FF9060, FF9061, FFA707, FFA272, FFA704, FFA516, FF2020, FF2023, FF2014, FFA940, FF2017, FF3014, FF3551, FFA967, FF3535, FF3536, FF2754, FF2755, FF3677, FF2757, FF3560, FF2839, FF3562, FF3565, FF2847, FF3566, FF3570, FF2917, FF3572, FF2104, FF2105, FF3177, FF3304, FF3305, FF3937, FF3308, FF3323, FF3327, FF3332, FF3337, FF3333, FF3334, FF3343, FF4085, FF3348, FF4087, FF3312, FF4090, FF3770, FF2711, FF3179, FF3119, FF3403, FF4185, FF3244, FF4183, FF3368, FF4184, FF3397, FF3401, FF3402, FF3406, FF3421, FF3349, FF3661, FF3660, FF3708, FF4059, FF3870, FF4341, FF4398, FF4388, FF4352, FF4639, FF4647, FF5136, FF4648, FF2266, FF4520, FF6273, FF6491, FF4653, FF4718, FF4838, FF2267, FF4862, FF6594, FF5582, FF5457, FF5604, FF2270, FF5605, FF2851, FF5623, FF5646, FF2861, FF5672, FF5699, FF5712, FF5711, FF5761, FF5583, FF5730, FF5980, FF5981, FF5998, FF6007, FF6141, FF6283, FF6540, FF6556, FF6563, FF6568, FF2336, FF2869, FF2541, FF2545, FF2349, FF2889, FF2356, FF2353, FF2354, FF2355, FF2925, FF1972, FF1973, FF0173, FF1971, FF0646, FF0183, FF2931, FF0187, FF0192, FF0196, FF1306, FF1309, FT1343, FF1952, FF2005, FF2006, FF2548, FF1947, FF2550, FF2308, FF2307, FF0644, FF0796, FF7671, FF7672, FF7673, FF7674, FF7626, FF7685, FF7744, FF6735, FF2031, FF2032, FF2049, FF7413, FF2111, FF7418, FF2114, FF7426, FF3099, FF2116, FF9344, FF2118, FF9382, FF2119, FF9337, FF3108, FF9446, FF9462, FF9463, FF9465, FF7682, FF9466, FF3112, FF9304, FF9521, FF9527, FF9606, FF9434, FF2033, FF9257, FFA126, FFA128, FF9256, FF9617, FF9377, FFA027, FF3134, FFA030, FF3138, FFA337, FFA338, FFA339, FF2474, FFA193, FF3141, FFA433, FFA438, FF9770, FF2488, FF9768, FF2475, FF9840, FF2486, FFA364, FFA363, FF9983, FFA361, FFA398, FFA409, FFA413, FFA504, FFA484, FFA539, FFA437, FFA479, FFA481, FF3183, FFA476, FFA377, FFA519, FFA613, FFA521, FFA855, FFA654, FFA735, FFA655, FFA938, FF3221, FF3225, FF3235, FF2514, FF2515, FF3738, FF2516, FF2517, FF1044, FF2910, FF1874, FF2949, FF1889, FF1894, FF2460, FF1918, FF5365, FF4512, FF3256, FF4511, FF2123, FF1920, FF5830, FF2117, FF1922, FF1927, FF3291, FF1934, FF1940, FF3345, FF3351, FF2630, FF2635, FF2170, FF2644, FF2646, FF2648, FF2633, FF2659, FF2675, FF2642, FF2653, FF2661, FF2672, FF2172, FF2178, FF2173, FF2175, FF2179, FF0185, FF0195, FF0159, FF0160, FF1547, FF0202, FF0204, FF9545, FF9542, FF9540, FF9538, FF9537, FF9524, FF9293, FF9289, FF7807, FF7492, FF9626, FF9373, FFA051, FF9976, FFA236, FF7715, FFA121, FFA271, FFA342, FF2330, FFA397, FFA395, FFA391, FFA255, FFA658, FFA736, FFA028, FFA953, FF2781, FF2850, FF5790, FF2657, FF2982, FF2434, FF2458, FF2955, FF2956, FF3310, FF3808, FF5385, FF4078, FF3742, FF3126, FF6253, FF1277, FF5361, FF9384, FF0020, FF0216, FF0484, FF0681, FF0682, FF0724, FF0728, FF0815, FF0814, FF0599, FF0991, FF1302, FF1337, FF1492, FF1506, FF1535, FF1534, FF1606, FF1725, FF1840, FF0365, FF0373, FF2332, FF7257, FF9496, FF9452, FF7699, FF7690, FF7700, FF7497, FF7816, FF9027, FF3056, FF3058, FF3189, FF3191, FF3152, FF3193, FF3147, FF3140, FF2467, FF2500, FF2499, FF2513, FF2498, FF0618, FF0573, FF0753, FF0757, FF0433, FF0572, FF0654, FF0683, FF0655, FF0752, FF3425, FF0685, FF3671, FF3672, FF3722, FF3719, FF3724, FF3725, FF3721, FF3728, FF3831, FF0669, FF3777, FF0670, FF1115, FF4167, FF4310, FF1359, FF4453, FF1468, FF1560, FF1642, FF1548, FF5044, FF1914, FF2053, FF2202, FF1867, FF7563, FF0488, FF1528, FF7376, FF7378, FF1783, FF2145, FF9516, FF9517, FF1566, FF1568, FF9531, FF9007, FF2144, FF9008, FF9009, FF9019, FF9020, FF9011, FF3284, FF9266, FF0772, FFA115, FFA264, FF0570, FF0638, FF1599, FFA975, FF2904, FF2901, FF2942, FF3007, FF3199, FF3156, FF2459, FF9593, FF2480, FF2481, FF9755, FF2471, FF3245, FF7561, FF3747, FF3748, FF9236, FF2512, FF2574, FF2588, FF2752, FF2758, FF2761, FF2779, FF7876, FF3232, FF3233, FF9893, FF9924, FF3526, FF3650, FF9968, FF3942, FF3943, FFA075, FF4074, FFA353, FF4345, FFA958, FF4719, FF4721, FF4716, FF4773, FF4806, FF4807, FF4942, FF5108, FF2796, FF3158, FF3159, FF3081, FF3075, FF3090, FF5664, FF5662, FF5669, FF5666, FF5295, FF3116, FF5783, FF5789, FF5793, FF5792, FF3136, FF3132, FF5833, FF5821, FF5977, FF3157, FF3127, FF3137, FF5296, FF5194, FF3146, FF3194, FF3168, FF1300, FF1227, FF3250, FF3166, FF3230, FF3239, FF3271, FF3272, FF3296, FF0026, FF0127, FF0172, FF0408, FF3352, FF0904, FF0874, FF1051, FF1275, FF1332, FF2524, FF1383, FF2522, FF1427, FF1490, FF1562, FF1607, FF1602, FF1722, FF1723, FF1858, FF1951, FF2058, FF2081, FF2196, FF1851, FF2054, FF2370, FF2368, FF1869, FF2415, FF2419, FF2584, FF5843, FF5840, FF2631, FF7200, FF7210, FF7239, FF7240, FF7314, FF6965, FF9615, FF9625, FF2665, FF9338, FF9752, FF6398, FF3498, FF3496, FF3390, FF9334, FF9385, FF9387, FF9340, FF9389, FF9456, FF9457, FF9459, FF9512, FF9520, FF9595, FF3917, FF7645, FF3922, FF7375, FF7801, FFPP03, FF7809, FF3953, FF3967, FF9672, FF9687, FF9677, FF9774, FF2683, FFA132, FFA112, FFA250, FF9998, FFA259, FFA412, FFA411, FFA530, FFA617, FFA618, FFA690, FFA621, FFA623, FF2855, FF2879, FFA941, FFA943, FF2880, FFA951, FFA750, FF2870, FF2871, FF2906, FF2913, FF3060, FF0725, FF0244, FF3319, FF3320, FF2643, FF2753, FF2784, FF3121, FF3072, FF4017, FF4373, FF4374, FF2961, FF4750, FF4948, FF4952, FF4953, FF3012, FF3035, FF3046, FF3023, FF6281, FF3386, FF3379, FF3548, FF0226, FF0432, FF0439, FF0954, FF3552, FF1287, FF1415, FF1536, FF3586, FF3579, FF3581, FF1665, FF1893, FF1958, FF1957, FF0675, FF2168, FF2217, FF2156, FF2216, FF2328, FF1148, FF6207, FF4192, FF6740, FF7431, FF7658, FF7649, FF9346, FF3621, FF9432, FF7670, FF7669, FF9771, FFA047, FFA077, FFA078, FFA085, FFA086, FF9468, FFA277, FFA278, FFA033, FFA434, FFA421, FFA422, FF9979, FFA607, FF3718, FFA410, FFA293, FF3699, FFA375, FF3017, FF3695, FF3018, FF3712, FF3769, FF3764, FF3766, FF2688, FF3784, FF3325, FF3781, FF3810, FF4170, FF4168, FF4199, FF4162, FF4200, FF3393, FF3850, FF4714, FF3849, FF4747, FF4854, FF4972, FF3844, FF3851, FF3847, FF4007, FF5460, FF6307, FF4104, FF4106, FF4105, FF4123, FF4124, FF4127, FF5503, FF5588, FF5607, FF0351, FF0349, FF5627, FF1015, FF1223, FF1989, FF4129, FF2223, FF2339, FF7565, FF6857, FF6858, FF7185, FF7199, FF7299, FF9237, FF6213, FF4302, FF4325, FF1271, FFA120, FF3030, FF3033, FF2442, FF3408, FF3988, FF4015, FF4449, FF4019, FF4314, FF4368, FF4172, FF4458, FF4808, FF4766, FF4772, FF5056, FF5053, FF5594, FF5395, FF4447, FF5400, FF5401, FF6009, FF6008, FF6067, FF6072, FF6127, FF6133, FF6275, FF6485, FF0003, FF0038, FF4502, FF0049, FF0048, FF0352, FF0491, FF0481, FF0005, FF1097, FF1095, FF1165, FF1161, FF1290, FF4501, FF1901, FF1959, FF1962, FF2077, FF0489, FF0350, FF7235, FF7307, FFA420, FFA512, FFA648, FF4549, FF4551, FF4645, FF6081, FF5389, FF5510, FF5512, FF5586, FF5590, FF5612, FF4678, FF4680, FF5513, FF5516, FF5626, FF5661, FF5173, FF4683, FF4728, FF4681, FF6211, FF6209, FF6014, FF4682, FF4732, FF4740, FF4786, FF4805, FF4833, FF4869, FF4921, FF4924, FF4914, FF4909, FF4927, FF5021, FF5022, FF6673, FF5060, FF6660, FF6948, FF7035, FF6975, FF6893, FF7080, FF7156, FF7037, FF7084, FF6898, FF5120, FF7335, FF5683, FF6612, FF9441, FF6623, FF6620, FF9311, FF3174, FF5303, FF3743, FF3744, FF3745, FF5369, FF5635, FF2422, FF2417, FF2770, FF2772, FF3154, FF3142, FF3210, FF3176, FF3285, FF3289, FF3775, FF3873, FF3997, FF3999, FF4174, FF4155, FF4156, FF5009, FF5016, FF5710, FF5688, FF5689, FF5366, FF5681, FF5803, FF5799, FF5810, FF5774, FF5781, FF5794, FF5382, FF5404, FF0022, FF0198, FF0292, FF0399, FF0438, FF5486, FF5489, FF0795, FF0037, FF0069, FF1049, FF5784, FF5786, FF1217, FF1286, FF1470, FF1472, FF1476, FF1478, FF1480, FF1481, FF1483, FF1484, FF5814, FF5815, FF5868, FF1986, FF1900, FF1987, FF5873, FF1988, FF5878, FF2262, FF2347, FF6616, FF7612, FF7640, FF6586, FF7082, FF7202, FF7353, FF7384, FF6004, FF6010, FF9597, FF9769, FF9775, FF9776, FF7813, FFA050, FF9978, FFA062, FF6146, FF6153, FFA702, FF6241, FF2887, FF3021, FF3070, FF3674, FF3588, FF2589, FF2629, FF2632, FF3367, FF6290, FF3733, FF6296, FF3737, FF6301, FF6304, FF4113, FF3683, FF4451, FF4532, FF4640, FF4536, FF4712, FF4844, FF4160, FF5651, FF5304, FF5222, FF5984, FF5987, FF6289, FF6356, FF6351, FF0130, FF0229, FF0348, FF0357, FF0358, FF0361, FF6250, FF0363, FF0540, FF0594, FF0593, FF0684, FF0835, FF0992, FF6458, FF0892, FF6487, FF1274, FF1185, FF1537, FF1586, FF1567, FF2414, FF6539, FF6382, FF7701, FF6478, FF9772, FFA270, FFA682, FFA945, FFA947, FF4380, FF5706, FF5831, FF6002, FF6000, FF6551, FF6554, FF6557, FF3710, FF3706, FF3439, FF6609, FF3436, FF3856, FF3857, FF3862, FF5354, FF5355, FF6763, FF6766, FF3939, FF3941, FF3600, FF3590, FF3957, FF1639, FF1629, FF1633, FF3329, FF3591, FF1649, FF1646, FF3592, FF4136, FF1650, FF4138, FF1653, FF3598, FF4142, FF4248, FF4249, FF4251, FF4402, FF4597, FF4669, FF4663, FF4674, FF4675, FF4666, FF4828, FF4662, FF4665, FF4611, FF4793, FF4788, FF4609, FF4783, FF4857, FF5041, FF4994, FF4995, FF4916, FF4918, FF5633, FF5422, FF5420, FF5326, FF5190, FF4993, FF5104, FF5234, FF5281, FF5320, FF5334, FF5239, FF5942, FF5230, FF5944, FF5874, FF6055, FF6059, FF6056, FF6148, FF6178, FF6150, FF6173, FF6228, FF6348, FF6442, FF6590, FF2577, FF2293, FF2559, FF2570, FF2579, FF7569, FF6507, FF6598, FF6597, FF6547, FF6807, FF6810, FF6546, FF6602, FF7091, FF7092, FF6603, FF9416, FF9423, FF9426, FF9417, FF9419, FF9651, FF6475, FF9178, FF9259, FF6877, FF7161, FF9443, FF9451, FF9471, FF7455, FF4365, FF7447, FF7462, FF7463, FF7458, FF7465, FF7460, FF7637, FF7644, FF7638, FF2957, FF9912, FF9913, FF9914, FF9910, FFA109, FFA340, FFA341, FFA345, FFA458, FFA749, FFA774, FFA626, FFA624, FFA778, FF3213, FF2655, FF3848, FF0132, FF3559, FF3958, FF4035, FF4025, FF4703, FF5096, FF5297, FF5370, FF5950, FF6511, FF0739, FF0898, FF0913, FF0417, FF0464, FF0466, FF0494, FF0446, FF0450, FF0473, FF0674, FF0552, FF0624, FF0729, FF1215, FF0164, FF0061, FF1187, FF1392, FF0170, FF1423, FF1280, FF1258, FF1365, FF1393, FF1339, FF1446, FF1327, FF1307, FF1312, FF1447, FF1369, FF1473, FF0194, FF1453, FF1474, FF1477, FF1386, FF1485, FF1510, FF1502, FF0228, FF0255, FF1611, FF1610, FF0283, FF0319, FF0398, FF0407, FF0405, FF0174, FF0177, FF1743, FF0284, FF1744, FF0072, FF1745, FF1768, FF0396, FF0238, FF0337, FF0382, FF0339, FF0383, FF0167, FF1690, FF0261, FF1050, FF1058, FF0165, FF0266, FF0073, FF1572, FF0956, FF1622, FF1619, FF1603, FF1589, FF1672, FF0775, FF0746, FF0735, FF0879, FF0875, FF0888, FF0983, FF0996, FF0812, FF0940, FF0926, FF0802, FF0862, FF1020, FF0915, FF1030, FF0786, FF0783, FF0896, FF0832, FF0828, FF0824, FF1109, FF1100, FF1099, FF1779, FF1804, FF1805, FF1838, FF2199, FF2197, FF1905, FF1964, FF1966, FF1981, FF1778, FF2002, FF2101, FF2143, FF2181, FF2195, FF2249, FF2257, FF2259, FF2256, FF2258, FF2302, FF2365, FF2364, FF2410, FF2436, FF2455, FF2472, FF2494, FF6769, FF6774, FF6773, FF6775, FF6777, FF6788, FF6840, FF6841, FF6855, FF6904, FF7664, FF7764, FF7768, FF7787, FF6985, FF6990, FF7063, FF7058, FF7099, FF7132, FF7146, FF7150, FF7157, FF7204, FF7017, FF7228, FF7281, FF7283, FF7320, FF7333, FF7364, FF7392, FF7394, FF7395, FF7439, FF7440, FF7490, FF7511, FF7566, FF7564, FF7585, FF7587, FF7642, FF7648, FF7686, FF9812, FF9810, FF9814, FF9885, FF9428, FF9427, FF9473, FF9622, FF9630, FF9650, FF9653, FF9664, FF9674, FF9697, FF9707, FF9765, FF9791, FF9823, FF7866, FF7728, FF7777, FF9055, FF9057, FF9050, FF9169, FF9207, FF9218, FF9214, FFA199, FFA198, FFA237, FFA282, FFA279, FFA281, FFA351, FFA373, FF9940, FF9953, FF9973, FF7660, FF9994, FFA025, FFA022, FFA130, FFA187, FFA380, FFA414, FFA415, FFA445, FFA469, FFA532, FFA557, FFA576, FFA632, FFA647, FFA653, FFA740, FFA768, FFA764, FFA769, FFA770, FFA780, FFA831, FFA842, FFA942, FFA955, FFB002, FFA992, FFB029, FFB058, FF2373, FF2990, FF2601, FF2973, FF2976, FF3866, FF4509, FF6311, FF4271, FF5091, FF5179, FF4507, FF5196, FF4508, FF5340, FF5144, FF5779, FF5844, FF5768, FF2988, FF3095, FF2882, FF2991, FF3098, FF6527, FF2022, FF6498, FF6310, FF6704, FF6175, FF6627, FF4515, FF7879, FF9695, FF9065, FF7888, FF7882, FF9745, FF9703, FF9052, FF9749, FF9750, FF3209, FF2317, FF2320, FF3267, FF4131, FF4459, FF4730, FF5243, FF5754, FF5751, FF5837, FF5752, FF6286, FF0313, FF2314, FF7847, FF3243, FF2226, FF2313, FF3251, FF3249, FF3252, FF3859, FF3858, FF3933, FF2318, FF3257, FF3265, FF5033, FF5027, FF5018, FF5037, FF4976, FF4983, FF5502, FF4176, FF3258, FF3920, FF4333, FF3264, FF4487, FF4488, FF4492, FF4595, FF4594, FF4604, FF4779, FF4977, FF4978, FF5570, FF5036, FF5571, FF5565, FF5569, FF5562, FF5580, FF5566, FF5581, FF5550, FF5554, FF5575, FF5040, FF5049, FF5090, FF5564, FF5577, FF5765, FF5576, FF6450, FF6456, FF6451, FF0578, FF0585, FF0252, FF0258, FF0260, FF0262, FF0672, FF0419, FF0422, FF0424, FF0863, FF0866, FF0865, FF0871, FF0883, FF0467, FF0468, FF0470, FF0881, FF0117, FF0118, FF1247, FF1248, FF1249, FF1454, FF1459, FF1721, FF1727, FF1751, FF1138, FF1714, FF1765, FF1716, FF1719, FF1827, FF1826, FF1836, FF2211, FF6674, FF6676, FF6677, FF6678, FF6679, FF6688, FF6693, FF6698, FF6695, FF6685, FF9166, FF9078, FF9024, FFA148, FFA153, FF9989, FFA162, FFA172, FFA159, FFA176, FFA179, FFA182, FFA185, FFA186, FFA221, FFA181, FFA216, FFA177, FFA232, FFA234, FFA229, FFA158, FFA163, FFA168, FFA170, FFA180, FFA183, FFA190, FFA196, FFA197, FFA233, FFA222, FFA223, FF9920, FFA157, FFA149, FFA188, FFA218, FFA390, FFA394, FFA389, FFA385, FFA794, FFA574, FFA575, FFA430, FFA432, FF1512. |
| FEI Number |
1119779
|
Recalling Firm/ Manufacturer |
Becton Dickinson & Co. 7 Loveton Cir Sparks MD 21152-9212
|
| For Additional Information Contact | North American Regional Complaint Center 410-316-4000 |
Manufacturer Reason for Recall | BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data. |
FDA Determined Cause 2 | Software design |
| Action | On October 31, 2024 URGENT" MEDICAL DEVICE PRODUCT CORRECTION letters were sent to consignees. IT, Safety and Security Actions to be Taken:
BD encourages customers to follow best practices for maintaining strong security measures to protect hospital networks and medical devices including:
"Ensure access to potentially vulnerable devices is limited to authorized personnel
"Inform authorized users of issue, and ensure all relevant passwords are tightly controlled
"Monitor and log network traffic attempting to reach medical device managementenvironments for suspicious activity
"Where possible, isolate affected devices in a secure VLAN or behind firewalls with restrictedaccess that only permits communication with trusted hosts in other networks when needed
"Impacted devices do not require use of RDP ports and these should be disabled or blockedif enabled
"Ensure permissions on file shares are appropriately established and enforced, and monitorand log access for evidence of suspicious activity
"Disconnect devices from the network if connectivity is not necessary
Next Steps to be Taken by Customer:
1.Ensure the contents of this notification are read and understood.
2.Share and post this customer letter within your facility network including IT and safety andsecurity department. Forward to any customers to whom you may have distributed theproduct(s) to ensure awareness.
3.Complete the attached Customer Response Form and return it to the BD contact noted on theForm so that BD may acknowledge your receipt of this notification per FDA requirements.
4.Report any adverse health consequences experienced with the use of these products to BD.Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via:
Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088)
Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787
If you require further assistance, please contact: Technical Service
+1 888 451 5514
Technical Questi |
| Quantity in Commerce | 8,085 units (838 US, 7,247 OUS) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MDB
|
|
|
|