| | Class 1 Device Recall GE Healthcare CARESTATION |  |
| Date Initiated by Firm | March 21, 2025 |
| Date Posted | April 29, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1625-2025 |
| Recall Event ID |
96634 |
| Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
| Product | GE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002. Anesthesia system. |
| Code Information |
GTIN 00840682124546, Serial Numbers: SM623510024WB, SM623510028WB, SM623510029WB, SM623510030WB, SM623510031WB, SM623510032WB, SM623510025WB, SM623510026WB, SM623510027WB, SM624390041WB, SM624390042WB, SM623510033WB, SM623510034WB, SM624220006WB, SM624220008WB, SM623160001WB, SM623160002WB, SM624120009WB, SM624120010WB, SM624510014WB, SM624510015WB, SM624510016WB, SM624250055WB, SM624250056WB, SM624120013WB, SM624120015WB, SM624320001WB, SM623250007WB, SM623250011WB, SM624250057WB, SM624250058WB, SM624250059WB, SM624250060WB, SM624250061WB, SM624320002WB, SM624370013WB, SM624370014WB, SM624370015WB, SM624370016WB, SM624370017WB, SM624370018WB, SM623460225WB, SM623460226WB, SM623460227WB, SM623460228WB, SM623460229WB, SM623460230WB, SM623460231WB, SM624120007WB, SM624120008WB, SM623250005WB, SM623250006WB, SM623460223WB, SM623460224WB, SM623390147WB, SM624120011WB, SM624120012WB, SM624120014WB, SM623250001WB, SM623250002WB, SM623250003WB, SM623250004WB, SM623390148WB, SM623390149WB, SM623390150WB, SM623390151WB, SM623390152WB, SM624120001WB, SM624120002WB, SM624120003WB, SM624120004WB, SM624120005WB, SM624120006WB, SM624120016WB, SM624120017WB, SM624120018WB, SM623140001WB, SM623140002WB, SM623140003WB, SM623230040WB, SM623230041WB, SM623230042WB, SM623100004WB, SM623100005WB, SM623100006WB, SM623100007WB, SM623100008WB, SM623370090WB, SM623370092WB, SM623370094WB, SM623370095WB, SM623020002WB, SM623240002WB, SM623240003WB, SM623240004WB, SM623240005WB, SM623500007WB, SM623500008WB, SM623500009WB, SM623500010WB, SM623500011WB, SM623500012WB, SM623240007WB, SM623240008WB, SM623240009WB, SM623240010WB, SM623240011WB, SM623240012WB, SM623240013WB, SM623240014WB, SM623370091WB, SM623370093WB, SM623010006WB, SM623110011WB, SM623110012WB, SM623240006WB, SM623240001WB, SM624380007WB, SM624380008WB, SM624380009WB, SM624380010WB, SM624380011WB, SM623470244WB, SM623470245WB, SM623440188WB, SM623440189WB, SM623440190WB, SM623430161WB, SM623430162WB, SM623430163WB, SM623430164WB, SM623430165WB, SM624110040WB, SM624110041WB, SM624110042WB, SM624110043WB, SM624110044WB, SM624110045WB, SM624110046WB, SM623290015WB, SM623290016WB, SM623290017WB, SM624020001WB, SM624020002WB, SM624100029WB, SM624100030WB, SM624100031WB, SM624100032WB, SM624100033WB, SM623270009WB, SM623470248WB, SM623470249WB, SM623170001WB, SM623170002WB, SM623440191WB, SM624020003WB, SM624100034WB, SM624100035WB, SM624100036WB, SM624100037WB, SM624100038WB, SM624100039WB, SM624250043WB, SM624250044WB, SM624250045WB, SM624250046WB, SM624250047WB, SM624250048WB, SM624250049WB, SM624250050WB, SM624250051WB, SM624250052WB, SM624250053WB, SM624250054WB, SM624370019WB, SM624370020WB, SM624370021WB, SM624370022WB, SM624370023WB, SM624370024WB, SM624370025WB, SM624370026WB, SM624370027WB, SM623150008WB, SM623470247WB, SM623290001WB, SM624020004WB, SM623290013WB, SM623130001WB, SM625020001WB, SM625020002WB, SM623270004WB, SM623270005WB, SM623270006WB, SM623270007WB, SM623270008WB, SM623290014WB, SM623470246WB, SM623260032WB, SM623260033WB, SM623070005WB, SM624010001WB, SM624160001WB, SM624160002WB, SM624160003WB, SM624160004WB, SM624160005WB, SM624270001WB, SM624270002WB, SM624340002WB, SM624340003WB, SM623130002WB, SM623130003WB, SM623430180WB, SM623430181WB, SM624170001WB, SM624340004WB, SM624340005WB, SM624340006WB, SM624340007WB, SM624340008WB, SM624340009WB, SM623430168WB, SM623430169WB, SM623430170WB, SM623430171WB, SM623430172WB, SM623430173WB, SM623290019WB, SM623290020WB, SM624050009WB, SM624110062WB, SM624110063WB, SM623400153WB, SM624240001WB, SM623270001WB, SM624270003WB, SM623490007WB, SM623180001WB, SM623180002WB, SM623180003WB, SM623180004WB, SM623360046WB, SM623360048WB, SM623360049WB, SM623360053WB, SM624120029WB, SM623020001WB, SM623090027WB, SM623090028WB, SM623460214WB, SM623360044WB, SM623360047WB, SM623360050WB, SM623360051WB, SM623360057WB, SM623360058WB, SM623470251WB, SM623470252WB, SM623470253WB, SM623470254WB, SM623470255WB, SM623470256WB, SM624110014WB, SM624110015WB, SM624110016WB, SM624110017WB, SM624110018WB, SM624110021WB, SM624110022WB, SM624110023WB, SM623150005WB, SM623330001WB, SM624110019WB, SM624110020WB, SM623220009WB, SM623150001WB, SM623460212WB, SM623460213WB, SM623490004WB, SM623490005WB, SM623490006WB, SM623220007WB, SM623220008WB, SM624370028WB, SM624370029WB, SM624370030WB, SM624370031WB, SM624370032WB, SM624370033WB, SM624370034WB, SM624370035WB, SM624370036WB, SM624370037WB, SM624370038WB, SM624370039WB, SM624370040WB, SM624370041WB, SM624370042WB, SM624250066WB, SM624250067WB, SM624250068WB, SM624250069WB, SM623360045WB, SM623360052WB, SM623360056WB, SM623360054WB, SM623360055WB, SM623340001WB, SM623340002WB, SM623340003WB, SM623340004WB, SM623340005WB, SM623340006WB, SM623340007WB, SM623110017WB, SM623110018WB, SM623110019WB, SM623110020WB, SM623110021WB, SM623110022WB, SM623110023WB, SM623260030WB, SM623260031WB, SM623260026WB, SM623260027WB, SM623260028WB, SM623260029WB, SM623460215WB, SM623460216WB, SM623350027WB, SM623350028WB, SM623360059WB, SM623360060WB, SM623360061WB, SM623360062WB, SM623360063WB, SM623490003WB, SM623350029WB, SM623350030WB, SM623460232WB, SM623460233WB, SM623460234WB, SM623460235WB, SM623460236WB, SM623460237WB, SM623460219WB, SM623460220WB, SM623460221WB, SM623350033WB, SM623460206WB, SM624120030WB, SM624490012WB, SM624490014WB, SM624490015WB, SM624490016WB, SM623350037WB, SM623350038WB, SM623460217WB, SM623460218WB, SM623460222WB, SM623500013WB, SM623500014WB, SM623500015WB, SM623500016WB, SM623500017WB, SM623500018WB, SM624120036WB, SM624120037WB, SM624120038WB, SM624120039WB, SM624120040WB, SM624120041WB, SM624120042WB, SM624120043WB, SM624120044WB, SM624120045WB, SM624120046WB, SM624490011WB, SM623460201WB, SM623460202WB, SM623460203WB, SM623460204WB, SM623460205WB, SM624120031WB, SM624120032WB, SM624120033WB, SM624120034WB, SM624120035WB, SM624490013WB, SM623360066WB, SM623360067WB, SM623360068WB, SM624090007WB, SM624090008WB, SM624090009WB, SM624090010WB, SM624090011WB, SM624090012WB, SM624120047WB, SM624240003WB, SM624240004WB, SM624240005WB, SM623250013WB, SM623250014WB, SM623250015WB, SM623250016WB, SM623250017WB, SM624330019WB, SM624330020WB, SM623210003WB, SM623210004WB, SM623210005WB, SM623210006WB, SM623510048WB, SM623510049WB, SM623510050WB, SM623510051WB, SM623510052WB, SM623510053WB, SM623510054WB, SM623510055WB, SM623520001WB, SM623520002WB, SM623520003WB, SM623520004WB, SM623520005WB, SM623520006WB, SM623440187WB, SM624120049WB, SM624120050WB, SM624120051WB, SM624120052WB, SM624380027WB, SM624380028WB, SM624090013WB, SM624090014WB, SM624380029WB, SM624380030WB, SM624380031WB, SM624380032WB, SM624380033WB, SM624380034WB, SM624090015WB, SM624090016WB, SM624090017WB, SM624090018WB, SM624220021WB, SM623360069WB, SM623360070WB, SM623440186WB, SM624120048WB, SM623360064WB, SM623360065WB, SM624200001WB, SM624200002WB, SM624200003WB, SM624200004WB, SM624200005WB, SM624200006WB, SM624200007WB, SM623250012WB, SM623350034WB, SM623350035WB, SM623350036WB, SM624510003WB, SM623510040WB, SM623510041WB, SM623510042WB, SM623510043WB, SM623510044WB, SM623510045WB, SM623510046WB, SM623510047WB, SM624510001WB, SM624510002WB, SM624510004WB, SM624490018WB, SM624490023WB, SM624490025WB, SM624490026WB, SM623230017WB, SM623230015WB, SM623230018WB, SM623370121WB, SM623370122WB, SM623370123WB, SM623370124WB, SM623370125WB, SM624490019WB, SM624490020WB, SM624490021WB, SM624490024WB, SM624380015WB, SM624380016WB, SM624380017WB, SM624490017WB, SM624490022WB, SM623370109WB, SM623370110WB, SM623370111WB, SM623370112WB, SM623370113WB, SM624380014WB, SM623470243WB, SM623360075WB, SM624110009WB, SM624110010WB, SM623370120WB, SM623230013WB, SM623230014WB, SM623230016WB, SM623270002WB, SM623270003WB, SM624110011WB, SM624110012WB, SM624120064WB, SM624120065WB, SM623360071WB, SM624350001WB, SM624350002WB, SM624380019WB, SM624380020WB, SM624380021WB, SM624380022WB, SM624380023WB, SM624380024WB, SM624380025WB, SM624380026WB, SM624380012WB, SM624380013WB, SM623290018WB, SM623290021WB, SM624380018WB, SM624110051WB, SM624250086WB, SM624250087WB, SM624250088WB, SM624500014WB, SM624500015WB, SM624500016WB, SM623360073WB, SM623360074WB, SM623360072WB, SM624350003WB, SM624350004WB, SM624350005WB, SM624350006WB, SM623240052WB, SM623500004WB, SM623500005WB, SM623500006WB, SM623370096WB, SM623370088WB, SM623370089WB, SM623240043WB, SM623240053WB, SM623110013WB, SM623110014WB, SM623110015WB, SM623110016WB, SM623250059WB, SM623250060WB, SM624490009WB, SM624490010WB, SM623220006WB, SM623240036WB, SM624250002WB, SM624250003WB, SM624250004WB, SM624340001WB, SM623090001WB, SM623090013WB, SM623090014WB, SM623180007WB, SM623180008WB, SM623180009WB, SM623160009WB, SM623220001WB, SM623220002WB, SM623220003WB, SM623250008WB, SM623250009WB, SM623250010WB, SM623240035WB, SM623240037WB, SM623240038WB, SM623240041WB, SM623510020WB, SM623510021WB, SM623510022WB, SM623510023WB, SM624110047WB, SM624110048WB, SM624110049WB, SM624110050WB, SM624110053WB, SM623240015WB, SM623240016WB, SM624110052WB, SM623240039WB, SM623240040WB, SM624250031WB, SM624250032WB, SM624250033WB, SM624250034WB, SM624250035WB, SM624100024WB, SM623370082WB, SM623370083WB, SM623370084WB, SM623370085WB, SM623370086WB, SM623370087WB, SM623220005WB, SM623240047WB, SM623240050WB, SM623010002WB, SM623010003WB, SM623240044WB, SM623240045WB, SM623240046WB, SM623240048WB, SM623240049WB, SM623240051WB, SM623060003WB, SM624500008WB, SM624500009WB, SM624500010WB, SM624500011WB, SM624500012WB, SM624500013WB, SM624110055WB, SM624110056WB, SM624110057WB, SM624110058WB, SM624110059WB, SM624110060WB, SM624110061WB, SM624240002WB, SM624110054WB, SM623110024WB, SM623110025WB, SM623110026WB, SM623110027WB, SM623110028WB, SM624250014WB, SM624250015WB, SM624250016WB, SM624250017WB, SM624250018WB, SM624510023WB, SM624510024WB, SM624510025WB, SM624510026WB, SM624510027WB, SM623370101WB, SM623370102WB, SM623370103WB, SM623120004WB, SM624250009WB, SM624250010WB, SM624250011WB, SM624250012WB, SM624250013WB, SM623120003WB, SM623130017WB, SM623130018WB, SM623130019WB, SM623130020WB, SM623130021WB, SM623370100WB, SM623340008WB, SM623350011WB, SM623370078WB, SM623370079WB, SM623370080WB, SM623370081WB, SM623350014WB, SM623350009WB, SM623350010WB, SM623350013WB, SM624250006WB, SM624250008WB, SM624370009WB, SM624370010WB, SM624370011WB, SM624370012WB, SM624510011WB, SM624510012WB, SM624510013WB, SM623350012WB, SM624250007WB, SM623060006WB, SM623060007WB, SM623060008WB, SM623060009WB, SM623060010WB, SM623060011WB, SM623370076WB, SM624250005WB, SM623370077WB, SM624480007WB, SM624050011WB, SM624050012WB, SM623240023WB, SM623240024WB, SM623240025WB, SM623240026WB, SM623240027WB, SM623240028WB, SM623240029WB, SM623240030WB, SM624120063WB, SM624120078WB, SM624120079WB, SM624220001WB, SM624220002WB, SM624220003WB, SM624220004WB, SM624220005WB, SM624220007WB, SM624220009WB, SM624220010WB, SM623210001WB, SM623210002WB, SM623370117WB, SM623370118WB, SM623370119WB, SM624240008WB, SM624240009WB, SM624510005WB, SM624510006WB, SM624510007WB, SM624510008WB, SM624510009WB, SM624510010WB, SM624250065WB, SM625040003WB, SM625040004WB, SM625040005WB, SM625040006WB, SM625040007WB, SM623250054WB, SM623250055WB, SM623080001WB, SM623200008WB, SM624240016WB, SM624240017WB, SM624520001WB, SM624520002WB, SM624520003WB, SM624520004WB, SM624520005WB, SM624520006WB, SM623230039WB, SM624240012WB, SM624240013WB, SM624240014WB, SM624240015WB, SM623090002WB, SM623090003WB, SM623090004WB, SM623090006WB, SM623090007WB, SM623090008WB, SM623090009WB, SM623090010WB, SM623090012WB, SM623090025WB, SM623090026WB, SM624120072WB, SM623180010WB, SM624050013WB, SM623470238WB, SM623470239WB, SM623470240WB, SM623470241WB, SM624120076WB, SM624480008WB, SM623090005WB, SM624230006WB, SM624230007WB, SM623250021WB, SM623250022WB, SM624120073WB, SM624120074WB, SM624120075WB, SM624120077WB, SM623250020WB, SM623490008WB, SM624050010WB, SM623090024WB, SM623240034WB, SM624490001WB, SM624490002WB, SM624490003WB, SM624490004WB, SM624490005WB, SM624490006WB, SM624490007WB, SM624490008WB, SM625040001WB, SM625040002WB, SM623250024WB, SM623250025WB, SM623090021WB, SM623090022WB, SM623090023WB, SM623240031WB, SM623240032WB, SM623240033WB, SM623370114WB, SM623370115WB, SM623370116WB, SM624240010WB, SM624240011WB, SM624250062WB, SM624250063WB, SM624250064WB, SM623510009WB, SM623510012WB, SM623010001WB, SM623370105WB, SM623370106WB, SM623250018WB, SM623510005WB, SM623510007WB, SM623510010WB, SM623510011WB, SM623510013WB, SM623510014WB, SM623510015WB, SM624140001WB, SM624140002WB, SM624140003WB, SM624140004WB, SM624140005WB, SM624140006WB, SM623250019WB, SM623370108WB, SM623510008WB, SM623510016WB, SM624380001WB, SM624380002WB, SM624380003WB, SM624380004WB, SM624380005WB, SM624380006WB, SM623370104WB, SM623370107WB, SM623510006WB, SM623310001WB, SM623370097WB, SM623370098WB, SM623370099WB, SM623230007WB, SM623230010WB, SM623230012WB, SM624370007WB, SM623010004WB, SM623010005WB, SM623230008WB, SM623230009WB, SM623230011WB, SM623510001WB, SM624250026WB, SM624250027WB, SM624250019WB, SM624250020WB, SM624250021WB, SM624250022WB, SM624250023WB, SM624250024WB, SM624250025WB, SM624250028WB, SM624250029WB, SM624250030WB, SM623450194WB, SM623180005WB, SM623180006WB, SM624100025WB, SM624100027WB, SM624370008WB, SM624100026WB, SM624370006WB, SM623510002WB, SM623510003WB, SM623170003WB, SM623330002WB, SM624230001WB, SM624230002WB, SM624230003WB, SM624230004WB, SM624230005WB, SM623060004WB, SM623450192WB, SM623450193WB, SM623150006WB, SM623150007WB, SM624360001WB, SM624360002WB, SM624360003WB, SM623050001WB, SM623050002WB, SM623050003WB, SM623050004WB, SM623050005WB, SM623410154WB, SM623520007WB, SM623520008WB, SM624500017WB, SM624500018WB, SM624500019WB, SM623110001WB, SM623110002WB, SM623110003WB, SM623110004WB, SM623110005WB, SM623110006WB, SM623110007WB, SM623110008WB, SM623110009WB, SM623110010WB, SM623250056WB, SM623250057WB, SM623250058WB, SM624330024WB, SM624330025WB, SM624330026WB, SM624220022WB, SM624330031WB, SM625020003WB, SM624110003WB, SM624110004WB, SM624110005WB, SM624110006WB, SM624110007WB, SM624110008WB, SM624030001WB, SM624030002WB, SM624330032WB, SM624330033WB, SM624330034WB, SM624330035WB, SM624330036WB, SM624330037WB, SM624290002WB, SM624290003WB, SM623350015WB, SM623350016WB, SM623350017WB, SM623350018WB, SM623350019WB, SM623350020WB, SM623350021WB, SM623350022WB, SM623350023WB, SM623350024WB, SM623350025WB, SM623350026WB, SM624110001WB, SM624210003WB, SM624210004WB, SM624210005WB, SM624480003WB, SM624480004WB, SM624480005WB, SM624480006WB, SM623250023WB, SM624100015WB, SM624120019WB, SM624120020WB, SM624120021WB, SM624120022WB, SM624120023WB, SM624120024WB, SM624120025WB, SM624120026WB, SM624120027WB, SM624120028WB, SM624250001WB, SM624260007WB, SM624260008WB, SM624260009WB, SM624260010WB, SM624260011WB, SM624260012WB, SM624260015WB, SM623510017WB, SM623510018WB, SM623510019WB, SM624100016WB, SM624100017WB, SM624100021WB, SM624100022WB, SM624100023WB, SM624330027WB, SM624330028WB, SM624330029WB, SM624330030WB, SM624110033WB, SM624110034WB, SM624330021WB, SM624330022WB, SM624330023WB, SM623230043WB, SM624210006WB, SM624210007WB, SM624210008WB, SM624210009WB, SM623090032WB, SM623350031WB, SM623350032WB, SM624110031WB, SM624110035WB, SM624110036WB, SM624110037WB, SM624110038WB, SM624110039WB, SM624210010WB, SM624210011WB, SM623190001WB, SM623280010WB, SM624090019WB, SM624090020WB, SM624110002WB, SM624110032WB, SM623090029WB, SM623090030WB, SM623090031WB, SM623100002WB, SM623100003WB, SM624480001WB, SM625040008WB, SM624100018WB, SM624100019WB, SM624100020WB, SM624110030WB, SM624480002WB, SM625040009WB, SM625040010WB, SM625040011WB, SM625040012WB, SM625040013WB, SM623480001WB, SM623480002WB, SM623100001WB, SM624210001WB, SM624210002WB, SM624190001WB, SM623470242WB, SM624080001WB, SM624110013WB, SM623370126WB, SM623370127WB, SM623370128WB, SM623370129WB, SM623370130WB, SM623370131WB, SM623370132WB, SM623370133WB, SM624370002WB, SM624370003WB, SM623060002WB, SM623130029WB, SM623130034WB, SM623130036WB, SM623130038WB, SM623240042WB, SM625020004WB, SM623130022WB, SM623130023WB, SM623130024WB, SM623130025WB, SM623130026WB, SM623130027WB, SM623130028WB, SM623130030WB, SM623130031WB, SM623130032WB, SM623130033WB, SM623130035WB, SM623130037WB, SM623130039WB, SM623130040WB, SM623130041WB, SM623060001WB, SM624370001WB, SM623430182WB, SM623430183WB, SM623430184WB, SM623430185WB, SM624510017WB, SM624510018WB, SM624510019WB, SM624510020WB, SM624510021WB, SM624510022WB, SM623430174WB, SM623430175WB, SM623430176WB, SM623430177WB, SM623430178WB, SM623430179WB, SM624050001WB, SM624050005WB, SM624050007WB, SM624250036WB, SM624250037WB, SM624250038WB, SM624250039WB, SM624250040WB, SM624250041WB, SM624250042WB, SM624500003WB, SM624500004WB, SM624500005WB, SM624500006WB, SM624500007WB, SM623110031WB, SM624050002WB, SM624050003WB, SM624050004WB, SM624050006WB, SM624050008WB, SM623120015WB, SM623120016WB, SM623220004WB, SM623230029WB, SM623230030WB, SM623230031WB, SM623230032WB, SM623230033WB, SM623230034WB, SM623230035WB, SM623230036WB, SM623230037WB, SM623230038WB, SM623230001WB, SM623230002WB, SM623230003WB, SM623230004WB, SM623230005WB, SM623230006WB, SM623230019WB, SM623230020WB, SM623230021WB, SM623230022WB, SM623230023WB, SM623230024WB, SM623230025WB, SM623230026WB, SM623230027WB, SM623230028WB, SM623110029WB, SM623110030WB, SM623380134WB, SM623380135WB, SM623380136WB, SM623380137WB, SM623380138WB, SM623490001WB, SM623200009WB, SM623200010WB, SM623160003WB, SM623160004WB, SM624240006WB, SM624240007WB, SM624220011WB, SM624220012WB, SM624220013WB, SM624220014WB, SM624220015WB, SM624220016WB, SM624220017WB, SM624220018WB, SM624220019WB, SM624220020WB, SM623500001WB, SM623500002WB, SM623500003WB, SM624290001WB, SM623480003WB, SM623480004WB, SM623240020WB, SM624340010WB, SM624340011WB, SM624340014WB, SM624340012WB, SM624340013WB, SM624110025WB, SM624110026WB, SM624110029WB, SM623240017WB, SM623240018WB, SM623240019WB, SM623240021WB, SM623240022WB, SM623390146WB, SM624110027WB, SM624510028WB, SM624510029WB, SM624510030WB, SM624510031WB, SM624510032WB, SM624110024WB, SM624110028WB, SM624120053WB, SM624120055WB, SM624120059WB, SM624120060WB, SM624120061WB, SM623380145WB, SM623380139WB, SM623380140WB, SM623380141WB, SM623380143WB, SM623510035WB, SM623510036WB, SM623510037WB, SM624280011WB, SM624280012WB, SM624280013WB, SM624280014WB, SM624280015WB, SM624280016WB, SM624280017WB, SM624280018WB, SM624280019WB, SM624280020WB, SM624280021WB, SM624280022WB, SM624280023WB, SM624280024WB, SM624280025WB, SM624280026WB, SM624450001WB, SM624450002WB, SM624450003WB, SM624280001WB, SM624280002WB, SM624280003WB, SM624280004WB, SM624280005WB, SM624280006WB, SM624280007WB, SM624280008WB, SM624280009WB, SM624280010WB, SM623510038WB, SM623510039WB, SM623180011WB, SM623180012WB, SM624370004WB, SM624370005WB, SM624430001WB, SM623510004WB, SM623120002WB, SM624120054WB, SM624120056WB, SM624120057WB, SM624120058WB, SM624120062WB, SM623380142WB, SM623380144WB, SM624250077WB, SM624250078WB, SM624250079WB, SM624250080WB, SM624250081WB, SM624250082WB, SM624250083WB, SM624250084WB, SM624250085WB, SM623200005WB, SM624390001WB, SM624390002WB, SM624390003WB, SM624390004WB, SM624390005WB, SM624390006WB, SM624390007WB, SM624390008WB, SM624390009WB, SM624390010WB, SM624390011WB, SM624390012WB, SM624390017WB, SM624390018WB, SM624390019WB, SM624390020WB, SM624390022WB, SM624390023WB, SM624390024WB, SM624390025WB, SM624390026WB, SM624390027WB, SM624390028WB, SM624390029WB, SM624390030WB, SM624390031WB, SM624390032WB, SM624390033WB, SM624390034WB, SM624390035WB, SM624390036WB, SM624390037WB, SM624390038WB, SM624390039WB, SM624390040WB, SM624500001WB, SM624500002WB, SM623450200WB, SM623360043WB, SM624260022WB, SM624260032WB, SM623120012WB, SM623200002WB, SM623200004WB, SM623200007WB, SM624100011WB, SM624260013WB, SM624260019WB, SM624260016WB, SM624260017WB, SM624260018WB, SM624260020WB, SM624260021WB, SM624260023WB, SM624260024WB, SM624260025WB, SM624260026WB, SM624260027WB, SM624260028WB, SM624260029WB, SM624260030WB, SM624260031WB, SM624330001WB, SM624330002WB, SM624330003WB, SM624330004WB, SM624330005WB, SM624330006WB, SM624330007WB, SM624330008WB, SM624330009WB, SM624330010WB, SM624330011WB, SM624330012WB, SM624330013WB, SM624330014WB, SM624330015WB, SM624330016WB, SM624330017WB, SM624330018WB, SM624390013WB, SM624390014WB, SM624390015WB, SM624390016WB, SM623120010WB, SM623430155WB, SM624390021WB, SM624250093WB, SM624250094WB, SM624250095WB, SM624250096WB, SM624250097WB, SM624250099WB, SM624250100WB, SM624250101WB, SM624250103WB, SM624250105WB, SM624520007WB, SM624520008WB, SM624520009WB, SM624520010WB, SM624520011WB, SM624520012WB, SM624520013WB, SM624520014WB, SM624520015WB, SM624520016WB, SM624520017WB, SM624520018WB, SM624520019WB, SM624520020WB, SM624520021WB, SM624520022WB, SM624520023WB, SM624520024WB, SM624520025WB, SM624520026WB, SM624520027WB, SM624520028WB, SM624520029WB, SM624520030WB, SM624520031WB, SM624520032WB, SM624520033WB, SM624520034WB, SM624520035WB, SM624520036WB, SM624520037WB, SM624520038WB, SM624520039WB, SM624520040WB, SM624520041WB, SM624520042WB, SM624520043WB, SM624520044WB, SM624520045WB, SM624520046WB, SM623360039WB, SM624250070WB, SM624250071WB, SM624250072WB, SM624250073WB, SM624260006WB, SM623200006WB, SM624250074WB, SM624100005WB, SM624100002WB, SM624100004WB, SM624100006WB, SM623200001WB, SM623460207WB, SM623460208WB, SM623460209WB, SM623460210WB, SM623460211WB, SM623470250WB, SM623360042WB, SM624100012WB, SM624100013WB, SM624100014WB, SM624250089WB, SM624250090WB, SM623200003WB, SM624100007WB, SM623360040WB, SM623360041WB, SM624250092WB, SM624250098WB, SM624250104WB, SM623120006WB, SM623120008WB, SM623120009WB, SM623120014WB, SM624250102WB, SM623150002WB, SM623150003WB, SM623450195WB, SM623450196WB, SM623450197WB, SM623450198WB, SM623450199WB, SM623120013WB, SM623430166WB, SM623430167WB, SM624100003WB, SM624250076WB, SM624260014WB, SM623120005WB, SM623120007WB, SM623120011WB, SM624100008WB, SM624100009WB, SM624260001WB, SM624260002WB, SM624260003WB, SM624260004WB, SM624260005WB, SM623430158WB, SM623430159WB, SM623430160WB, SM624470001WB, SM624100010WB, SM624250091WB, SM624250075WB, SM623430156WB, SM623430157WB, SM623490002WB. |
| FEI Number |
3003513129
|
Recalling Firm/ Manufacturer |
GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No. 19 Changjiang Road; Xin Wuxi China
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation. |
FDA Determined Cause 2 | Device Design |
| Action | GE Healthcare issued an URGENT MEDICAL DEVICE notice to its consignees on 03/21/2025. The notice explained the safety issue with the device, potential risk, and requested the following actions:
The anesthesia system may continue to be used following these instructions:
Perform the Ventilation Screening Test specified in Appendix 1 for each affected Carestation system.
If the Carestation system passes the Ventilation Screening Test, you can continue to use the device in accordance with the instructions in the User Reference Manual (URM).
If the Carestation system fails the Ventilation Screening Test and must be used prior to the system being corrected by GE HealthCare, follow the instructions below:
1. Use the device in accordance with the instructions in the URM with these changes:
i. Use only Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) mode to mechanically ventilate a patient.
ii. Do not use Volume Control Ventilation (VCV) mode to mechanically ventilate a patient.
NOTE: Manual mode of the anesthesia system can be used to provide manual ventilation or allow spontaneous ventilation of the patient.
2. Users are not to use VCV mode on the device until the device has been corrected by GE HealthCare.
Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions, and retain this document for your record.
GE HealthCare will correct all devices that fail the Ventilation Screening Test at no cost. A GE HealthCare representative will contact you to arrange for the correction.
Questions: contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Notice sent on 04/21/25 with updated instructions. Change: Attach a 2 Liter reservoir bag (original) to Attach a reservoir bag (updated)in step 5 of the Ventilation Screening Test in Appendix 1 of the letter. |
| Quantity in Commerce | 1483 units |
| Distribution | Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|