• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall GE Healthcare CARESTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall GE Healthcare CARESTATIONsee related information
Date Initiated by FirmMarch 21, 2025
Date PostedApril 29, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1635-2025
Recall Event ID 96634
510(K)NumberK213867 
Product Classification Gas-machine, anesthesia - Product Code BSZ
ProductGE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000
Code Information GTIN 00840682145596, Serial Numbers: SQS24500010MA, SQS24500011MA, SQS24500012MA, SQS24500013MA, SQS24500014MA, SQS24500015MA, SQS24500016MA, SQS24500017MA, SQS24130011MA, SQS24130012MA, SQS24130013MA, SQS24130014MA, SQS24130015MA, SQS24130016MA, SQS24130017MA, SQS24130018MA, SQS24170062MA, SQS24170063MA, SQS24170064MA, SQS24170065MA, SQS24170066MA, SQS24170079MA, SQS24170080MA, SQS24170084MA, SQS24170085MA, SQS24170086MA, SQS24170087MA, SQS24170088MA, SQS24190115MA, SQS24190116MA, SQS24190117MA, SQS24430007MA, SQS24430008MA, SQS24430009MA, SQS24430010MA, SQS24150018MA, SQS24390013MA, SQS24390014MA, SQS24460010MA, SQS23210004MA, SQS23210001MA, SQS23210002MA, SQS23210003MA, SQS25020021MA, SQS25020022MA, SQS25020023MA, SQS25020024MA, SQS25020025MA, SQS24480001MA, SQS24250003MA, SQS24160054MA, SQS24160055MA, SQS24160056MA, SQS24160057MA, SQS24180105MA, SQS24190120MA, SQS24190121MA, SQS24190122MA, SQS24190123MA, SQS24190124MA, SQS25020015MA, SQS24440002MA, SQS24440003MA, SQS24440004MA, SQS24440005MA, SQS24440006MA, SQS24440007MA, SQS24440008MA, SQS24440009MA, SQS24440010MA, SQS24440011MA, SQS24440012MA, SQS24440013MA, SQS24460001MA, SQS24500002MA, SQS24150022MA, SQS24150023MA, SQS24150024MA, SQS24150025MA, SQS24150026MA, SQS24150027MA, SQS24150028MA, SQS24150029MA, SQS24150030MA, SQS24150031MA, SQS24150032MA, SQS25030026MA, SQS23340006MA, SQS23340005MA, SQS23380004MA, SQS23380005MA, SQS24190118MA, SQS24190119MA, SQS24490033MA, SQS24490034MA, SQS24210005MA, SQS24210006MA, SQS24380019MA, SQS24370002MA, SQS24370003MA, SQS24370004MA, SQS24370005MA, SQS25030028MA, SQS25030029MA, SQS25030030MA, SQS24420022MA, SQS24420023MA, SQS24290048MA, SQS24290049MA, SQS24290050MA, SQS24290051MA, SQS24290052MA, SQS24290053MA, SQS24290054MA, SQS24290055MA, SQS24290056MA, SQS24290057MA, SQS24290058MA, SQS24290059MA, SQS24290060MA, SQS24290062MA, SQS24290063MA, SQS24290064MA, SQS24290065MA, SQS24290066MA, SQS24290067MA, SQS24290068MA, SQS24290069MA, SQS24290070MA, SQS24290071MA, SQS24290072MA, SQS24290073MA, SQS24290074MA, SQS24290075MA, SQS24290076MA, SQS24290077MA, SQS24290078MA, SQS24290079MA, SQS24480002MA, SQS24480003MA, SQS24480004MA, SQS24480005MA, SQS24480006MA, SQS23380003MA, SQS24320007MA, SQS24320009MA, SQS24510016MA, SQS24390007MA, SQS24390008MA, SQS24390009MA, SQS23360001MA, SQS23360002MA, SQS24040003MA, SQS24040004MA, SQS24060003MA, SQS24060004MA, SQS24060005MA, SQS24060006MA, SQS24060007MA, SQS24060008MA, SQS24060009MA, SQS24170060MA, SQS24170061MA, SQS24070007MA, SQS24180095MA, SQS24150016MA, SQS24380021MA, SQS24380022MA, SQS24380023MA, SQS24040012MA, SQS24070008MA, SQS24070009MA, SQS24070010MA, SQS24070011MA, SQS24120010MA, SQS24050080MA, SQS24050081MA, SQS24050082MA, SQS24050083MA, SQS24050084MA, SQS24050085MA, SQS24050086MA, SQS24050087MA, SQS24420001MA, SQS24420002MA, SQS24420003MA, SQS24420004MA, SQS24420005MA, SQS24420006MA, SQS24420007MA, SQS24420008MA, SQS24420009MA, SQS24420010MA, SQS24420011MA, SQS24420012MA, SQS24420013MA, SQS24420014MA, SQS24420015MA, SQS24420016MA, SQS24420017MA, SQS24420018MA, SQS24420019MA, SQS24420020MA, SQS24420021MA, SQS23430010MA, SQS24420024MA, SQS23380006MA, SQS23380007MA, SQS23490002MA, SQS23490003MA, SQS23490004MA, SQS23490005MA, SQS24060024MA, SQS23440024MA, SQS24460008MA, SQS24460009MA, SQS24460011MA, SQS24380002MA, SQS24380003MA, SQS24380004MA, SQS24390005MA, SQS24390006MA, SQS24210007MA, SQS24210008MA, SQS24070002MA, SQS24070003MA, SQS24090006MA, SQS24090007MA, SQS24040005MA, SQS24040006MA, SQS24040007MA, SQS24040008MA, SQS24040009MA, SQS24040010MA, SQS24400007MA, SQS24400008MA, SQS24400009MA, SQS24400010MA, SQS24330001MA, SQS24330002MA, SQS24330003MA, SQS24330004MA, SQS24330005MA, SQS24330006MA, SQS25030027MA, SQS23440023MA, SQS24010021MA, SQS24010022MA, SQS24010023MA, SQS24010024MA, SQS24010025MA, SQS24010026MA, SQS24010027MA, SQS24150033MA, SQS24150034MA, SQS24150035MA, SQS24150036MA, SQS24150037MA, SQS24150038MA, SQS24150039MA, SQS24150040MA, SQS24320006MA, SQS23500001MA, SQS23500004MA, SQS24320008MA, SQS24370006MA, SQS24380020MA, SQS24170059MA, SQS23010001MA, SQS23080033MA, SQS24110014MA, SQS24400001MA, SQS24110013MA, SQS24320010MA, SQS24320011MA, SQS24070004MA, SQS24070005MA, SQS24070006MA, SQS24360001MA, SQS24360002MA, SQS24360003MA, SQS24200002MA, SQS24400003MA, SQS23320001MA, SQS23320002MA, SQS23320003MA, SQS23320004MA, SQS23320005MA, SQS23320006MA, SQS23320007MA, SQS23320008MA, SQS24170071MA, SQS24170072MA, SQS24170073MA, SQS24170074MA, SQS24330007MA, SQS24330008MA, SQS24330009MA, SQS24330010MA, SQS24360006MA, SQS24360007MA, SQS24360010MA, SQS24410001MA, SQS24490008MA, SQS24490010MA, SQS24490015MA, SQS24490018MA, SQS24490021MA, SQS24490022MA, SQS24490023MA, SQS24490024MA, SQS24490025MA, SQS24490028MA, SQS24490029MA, SQS24490030MA, SQS24490031MA, SQS24490005MA, SQS24490006MA, SQS24490011MA, SQS24490012MA, SQS24490014MA, SQS24490016MA, SQS24490017MA, SQS24490019MA, SQS24490020MA, SQS24490026MA, SQS24490027MA, SQS24340021MA, SQS24340022MA, SQS24340023MA, SQS24310003MA, SQS24490001MA, SQS24490002MA, SQS24490003MA, SQS24490004MA, SQS24490007MA, SQS24490009MA, SQS24490013MA, SQS24360004MA, SQS24360005MA, SQS24360008MA, SQS24360009MA, SQS23430011MA, SQS24070001MA, SQS24160042MA, SQS24160043MA, SQS24340014MA, SQS24340015MA, SQS24340016MA, SQS24340017MA, SQS24340018MA, SQS24340019MA, SQS24340020MA, SQS24490032MA, SQS24140005MA, SQS24140006MA, SQS24050036MA, SQS24050037MA, SQS24050038MA, SQS24050039MA, SQS24050040MA, SQS24050041MA, SQS24050042MA, SQS24050043MA, SQS24050044MA, SQS24050045MA, SQS24040014MA, SQS24500001MA, SQS24410002MA, SQS24410003MA, SQS24410004MA, SQS24410005MA, SQS24410006MA, SQS24410007MA, SQS24410008MA, SQS24410009MA, SQS24410010MA, SQS24410011MA, SQS24410012MA, SQS24090005MA, SQS24110005MA, SQS23430014MA, SQS23430015MA, SQS23430016MA, SQS23430017MA, SQS23430018MA, SQS23430019MA, SQS24180099MA, SQS24180100MA, SQS24180101MA, SQS24180102MA, SQS24180103MA, SQS24180104MA, SQS24470004MA, SQS24470005MA, SQS23480009MA, SQS23480010MA, SQS24510020MA, SQS24510029MA, SQS24510030MA, SQS24150017MA, SQS24290022MA, SQS24510031MA, SQS24070012MA, SQS24070013MA, SQS24070014MA, SQS24070015MA, SQS24070016MA, SQS24070017MA, SQS24070018MA, SQS24070019MA, SQS24070020MA, SQS24070021MA, SQS24380005MA, SQS24380006MA, SQS24380007MA, SQS24380008MA, SQS24380009MA, SQS24380010MA, SQS24380011MA, SQS24380012MA, SQS24380013MA, SQS24380014MA, SQS24060010MA, SQS24060011MA, SQS24060012MA, SQS24060013MA, SQS24060014MA, SQS24060015MA, SQS24060016MA, SQS24060017MA, SQS24060018MA, SQS24060019MA, SQS24060020MA, SQS23260016MA, SQS23260017MA, SQS23260018MA, SQS23260019MA, SQS23260020MA, SQS23260021MA, SQS23260022MA, SQS23260023MA, SQS23260024MA, SQS23260025MA, SQS23260026MA, SQS23260027MA, SQS23260028MA, SQS23260029MA, SQS24040001MA, SQS24040002MA, SQS24220001MA, SQS24220002MA, SQS24220003MA, SQS24220004MA, SQS24500031MA, SQS24500032MA, SQS23500002MA, SQS23500003MA, SQS24060025MA, SQS24060026MA, SQS24060027MA, SQS24060028MA, SQS24060029MA, SQS24060030MA, SQS24500018MA, SQS24500019MA, SQS24500020MA, SQS24500021MA, SQS24500022MA, SQS24500023MA, SQS24500024MA, SQS24450004MA, SQS24300023MA, SQS24300024MA, SQS24040017MA, SQS23480002MA, SQS23480003MA, SQS23480004MA, SQS23480005MA, SQS23480006MA, SQS23480007MA, SQS23480008MA, SQS24020006MA, SQS25020001MA, SQS24170069MA, SQS24170070MA, SQS23500005MA, SQS24430004MA, SQS24430005MA, SQS24430006MA, SQS24500033MA, SQS23280039MA, SQS23280040MA, SQS23280041MA, SQS23280042MA, SQS23280043MA, SQS23280044MA, SQS23380001MA, SQS24200005MA, SQS24200006MA, SQS24310001MA, SQS23260006MA, SQS23260007MA, SQS23260008MA, SQS23260009MA, SQS23260010MA, SQS24160046MA, SQS24160047MA, SQS24160048MA, SQS24160049MA, SQS24160050MA, SQS24160051MA, SQS24160052MA, SQS24040018MA, SQS24040019MA, SQS24470002MA, SQS24470003MA, SQS24190113MA, SQS24190114MA, SQS24260003MA, SQS24260004MA, SQS24260005MA, SQS24260006MA, SQS23490001MA, SQS24150007MA, SQS24150008MA, SQS24150009MA, SQS24150010MA, SQS24470001MA, SQS24120013MA, SQS24120011MA, SQS24120012MA, SQS24120014MA, SQS24180098MA, SQS24060021MA, SQS24060022MA, SQS24060023MA, SQS24450005MA, SQS24450006MA, SQS24450007MA, SQS23320011MA, SQS23320012MA, SQS23320014MA, SQS23320015MA, SQS23320016MA, SQS23440005MA, SQS23440006MA, SQS23440007MA, SQS24230007MA, SQS24230008MA, SQS24230009MA, SQS24230010MA, SQS24230011MA, SQS24230012MA, SQS24230013MA, SQS24230014MA, SQS24230015MA, SQS24260002MA, SQS23320013MA, SQS24280027MA, SQS24280028MA, SQS24480007MA, SQS24480008MA, SQS24480009MA, SQS24480010MA, SQS24480011MA, SQS24480012MA, SQS24480013MA, SQS24480014MA, SQS24480015MA, SQS24480016MA, SQS24480017MA, SQS24210009MA, SQS24210010MA, SQS24210011MA, SQS24210012MA, SQS24210013MA, SQS24210014MA, SQS24210015MA, SQS24210016MA, SQS24210017MA, SQS24340114MA, SQS24340115MA, SQS24390004MA, SQS24010028MA, SQS24010029MA, SQS24010030MA, SQS24010031MA, SQS24010032MA, SQS24010033MA, SQS24010034MA, SQS24010035MA, SQS24010036MA, SQS24010037MA, SQS24010038MA, SQS24010039MA, SQS24040016MA, SQS24280025MA, SQS24170058MA, SQS23440001MA, SQS23440002MA, SQS24030005MA, SQS24030006MA, SQS24030007MA, SQS24030008MA, SQS24170075MA, SQS24170076MA, SQS24170077MA, SQS24170078MA, SQS24200003MA, SQS24200004MA, SQS24310004MA, SQS24310005MA, SQS24420027MA, SQS24420028MA, SQS24420029MA, SQS24420030MA, SQS24140004MA, SQS24500004MA, SQS24500005MA, SQS24500006MA, SQS24500007MA, SQS24500008MA, SQS24500009MA, SQS24180096MA, SQS24180097MA, SQS24200001MA, SQS24390002MA, SQS23380008MA, SQS23380009MA, SQS23380010MA, SQS23380011MA, SQS23380012MA, SQS23380013MA, SQS23380014MA, SQS23380015MA, SQS23380016MA, SQS23310006MA, SQS23310007MA, SQS23310008MA, SQS23310009MA, SQS24510017MA, SQS24510018MA, SQS24510019MA, SQS25040037MA, SQS25040038MA, SQS24130004MA, SQS24130005MA, SQS24130006MA, SQS24130007MA, SQS24130008MA, SQS24130009MA, SQS24130010MA, SQS23440025MA, SQS23440026MA, SQS24040011MA, SQS24410013MA, SQS24410014MA, SQS24440014MA, SQS24440015MA, SQS24390003MA, SQS24380015MA, SQS23380017MA, SQS23380018MA, SQS23380019MA, SQS23380020MA, SQS23380021MA, SQS23380022MA, SQS23370001MA, SQS23370002MA, SQS24490040MA, SQS23080031MA, SQS23080034MA, SQS23420001MA, SQS23420002MA, SQS23420003MA, SQS23420004MA, SQS23420005MA, SQS23420006MA, SQS24210004MA, SQS23340003MA, SQS23470001MA, SQS23470002MA, SQS23470003MA, SQS23470004MA, SQS23470005MA, SQS23470006MA, SQS23470007MA, SQS23470008MA, SQS23470009MA, SQS23470010MA, SQS23470011MA, SQS23470012MA, SQS23470013MA, SQS23470014MA, SQS23470015MA, SQS23470016MA, SQS23470017MA, SQS23470018MA, SQS23470019MA, SQS24370007MA, SQS24370008MA, SQS24300021MA, SQS24300022MA, SQS23480001MA, SQS24260008MA, SQS24280024MA, SQS24160053MA, SQS23480011MA, SQS23480012MA, SQS23480013MA, SQS23480014MA, SQS23480015MA, SQS23480016MA, SQS23480017MA, SQS23480018MA, SQS24240001MA, SQS24240002MA, SQS24240003MA, SQS24240004MA, SQS24240005MA, SQS24240006MA, SQS24120022MA, SQS24120023MA, SQS24120024MA, SQS24120025MA, SQS23480019MA, SQS23480020MA, SQS23480021MA, SQS23480022MA, SQS23480023MA, SQS23480024MA, SQS23480025MA, SQS23480026MA, SQS23480027MA, SQS24110003MA, SQS24110004MA, SQS23310001MA, SQS23310002MA, SQS23310003MA, SQS23310004MA, SQS24010040MA, SQS24250001MA, SQS24410017MA, SQS24320004MA, SQS24320005MA, SQS24260001MA, SQS24510028MA, SQS23430001MA, SQS23430002MA, SQS23430003MA, SQS23430004MA, SQS23430005MA, SQS23430006MA, SQS23430007MA, SQS23430008MA, SQS24180106MA, SQS23010002MA, SQS23010003MA, SQS23010004MA, SQS23010005MA, SQS23080030MA, SQS23080032MA, SQS23080035MA, SQS23080036MA, SQS23080037MA, SQS23080038MA, SQS23080039MA, SQS23230001MA, SQS23230002MA, SQS23380002MA, SQS23440003MA, SQS23440004MA, SQS24340024MA, SQS24340025MA, SQS24340026MA, SQS24380016MA, SQS24380017MA, SQS24380018MA, SQS24390010MA, SQS24390011MA, SQS24390012MA, SQS23320009MA, SQS23320010MA, SQS24160044MA, SQS24160045MA, SQS23430009MA, SQS24510021MA, SQS24040013MA, SQS24190112MA, SQS24150019MA, SQS24150020MA, SQS24150021MA, SQS24460006MA, SQS24460007MA, SQS24310002MA, SQS24510032MA, SQS24510033MA, SQS23190042WA, SQS23190043WA, SQS23190044WA, SQS23310033WA, SQS23290045WA, SQS23290046WA, SQS24080009WA, SQS24080010WA, SQS24080011WA, SQS24080012WA, SQS24080013WA, SQS24080014WA, SQS24080015WA, SQS24080016WA, SQS24080017WA, SQS24080018WA, SQS23210007WA, SQS24490035WA, SQS24510037WA, SQS24440074WA, SQS24050064WA, SQS23480023WA, SQS25030083WA, SQS23410009WA, SQS23410010WA, SQS23410011WA, SQS23160012WA, SQS23280006WA, SQS24010002WA, SQS24010003WA, SQS24010004WA, SQS23400003WA, SQS24470039WA, SQS24470040WA, SQS24320033WA, SQS24510045WA, SQS23480024WA, SQS23020025WA, SQS23020026WA, SQS23020027WA, SQS23020028WA, SQS23020029WA, SQS23050036WA, SQS23050037WA, SQS23050038WA, SQS23050039WA, SQS23050040WA, SQS23050041WA, SQS23050042WA, SQS23050043WA, SQS23050044WA, SQS23050045WA, SQS23060059WA, SQS23060060WA, SQS23060061WA, SQS23060062WA, SQS23060063WA, SQS23070065WA, SQS23070066WA, SQS23070067WA, SQS23070068WA, SQS23070069WA, SQS23070070WA, SQS23080025WA, SQS23080026WA, SQS23080027WA, SQS23080028WA, SQS23080029WA, SQS24190273WA, SQS24370001WA, SQS24370002WA, SQS24490036WA, SQS24490037WA, SQS24210001MA, SQS24210002MA, SQS24210003MA, SQS24370001MA, SQS23070001WA, SQS23070002WA, SQS23150008WA, SQS23150009WA, SQS23150010WA, SQS24050062WA, SQS24050063WA, SQS24100016WA, SQS24390008WA, SQS24390009WA, SQS24390010WA, SQS24390011WA, SQS24390012WA, SQS24470043WA, SQS24290001WA, SQS23400014WA, SQS23110001WA, SQS24050002WA, SQS24050003WA, SQS24050004WA, SQS24170133WA, SQS24170134WA, SQS24170135WA, SQS24440001WA, SQS24440002WA, SQS24440003WA, SQS24440004WA, SQS23420077WA, SQS23420080WA, SQS23470003WA, SQS23310024WA, SQS23310025WA, SQS23310026WA, SQS24320036WA, SQS24320037WA, SQS24320038WA, SQS24320039WA, SQS24320040WA, SQS24150092WA, SQS24150094WA, SQS24150095WA, SQS24150096WA, SQS24150097WA, SQS24150098WA, SQS24150099WA, SQS24150101WA, SQS23420078WA, SQS24320001WA, SQS24150093WA, SQS24150100WA, SQS23420079WA, SQS24280001WA, SQS24280002WA, SQS24280003WA, SQS24280004WA, SQS24070025WA, SQS24280016WA, SQS24280017WA, SQS23200001MA, SQS24190110MA, SQS24290021MA, SQS23450002WA, SQS23450003WA, SQS23450004WA, SQS23450005WA, SQS23450006WA, SQS23450007WA, SQS23450008WA, SQS23450009WA, SQS24200007MA, SQS24200008MA, SQS24200009MA, SQS23450010WA, SQS23450011WA, SQS23450012WA, SQS25040124WA, SQS25040125WA, SQS25040126WA, SQS25040127WA, SQS25040128WA, SQS23090003WA, SQS23090004WA, SQS23090007WA, SQS23090010WA, SQS23090011WA, SQS23090012WA, SQS24490035MA, SQS24490036MA, SQS24490037MA, SQS24490038MA, SQS24490039MA, SQS23090005WA, SQS23090006WA, SQS23090008WA, SQS23090009WA, SQS25030076WA, SQS25030077WA, SQS23090001WA, SQS23090002WA, SQS23160001WA, SQS23160002WA, SQS23160003WA, SQS23290007WA, SQS23290008WA, SQS23290009WA, SQS23290010WA, SQS23140008WA, SQS23140009WA, SQS24050016WA, SQS24050046WA, SQS24280019WA, SQS24280028WA, SQS24280029WA, SQS24050009WA, SQS24050010WA, SQS24050011WA, SQS24050012WA, SQS24050013WA, SQS24050014WA, SQS24080044WA, SQS24090016WA, SQS24090017WA, SQS24090018WA, SQS24090019WA, SQS24090020WA, SQS24090021WA, SQS24090022WA, SQS24090023WA, SQS24090024WA, SQS24090025WA, SQS24090026WA, SQS24090027WA, SQS24090028WA, SQS24090029WA, SQS24090030WA, SQS24090031WA, SQS24090032WA, SQS24090033WA, SQS24090034WA, SQS24090035WA, SQS24090036WA, SQS24090037WA, SQS24090038WA, SQS24090039WA, SQS24090040WA, SQS24090041WA, SQS24090042WA, SQS24090043WA, SQS24090044WA, SQS24090045WA, SQS24090049WA, SQS24090050WA, SQS24090051WA, SQS24090052WA, SQS24090053WA, SQS24090054WA, SQS24090055WA, SQS24090056WA, SQS24090057WA, SQS24090058WA, SQS24090059WA, SQS23020015WA, SQS23020016WA, SQS23020017WA, SQS23020018WA, SQS23020019WA, SQS23020020WA, SQS23020021WA, SQS23020022WA, SQS23020023WA, SQS23020024WA, SQS23440057WA, SQS23440058WA, SQS23440059WA, SQS23440060WA, SQS23440061WA, SQS23440062WA, SQS23440063WA, SQS23440064WA, SQS23440065WA, SQS23440066WA, SQS23440067WA, SQS23440068WA, SQS23440069WA, SQS23440070WA, SQS23440071WA, SQS23440072WA, SQS23440073WA, SQS23440074WA, SQS23440075WA, SQS23440076WA, SQS23440077WA, SQS23440078WA, SQS23440079WA, SQS23440080WA, SQS24190234WA, SQS24200030WA, SQS24280014WA, SQS24470017WA, SQS24470018WA, SQS24470019WA, SQS24470020WA, SQS24470021WA, SQS24470022WA, SQS24470023WA, SQS24470024WA, SQS24470025WA, SQS24470026WA, SQS23010010WA, SQS23010011WA, SQS23070020WA, SQS23060010WA, SQS24150102WA, SQS23400010WA, SQS23400011WA, SQS23400012WA, SQS23400013WA, SQS23270004WA, SQS23440008MA, SQS23440009MA, SQS23440010MA, SQS23440011MA, SQS23440012MA, SQS23440013MA, SQS23440014MA, SQS23440015MA, SQS23440016MA, SQS23440017MA, SQS23440018MA, SQS23440019MA, SQS24020007MA, SQS24020008MA, SQS24020009MA, SQS24020010MA, SQS24020011MA, SQS24020012MA, SQS24020013MA, SQS24020014MA, SQS24020015MA, SQS24020016MA, SQS24020017MA, SQS24020018MA, SQS24020019MA, SQS24020020MA, SQS24020021MA, SQS24020022MA, SQS24020023MA, SQS24020024MA, SQS24020025MA, SQS24020026MA, SQS24020027MA, SQS24020028MA, SQS24020029MA, SQS24020030MA, SQS24020031MA, SQS24040020MA, SQS24040021MA, SQS24040022MA, SQS24040023MA, SQS24040024MA, SQS24040025MA, SQS24040026MA, SQS24040027MA, SQS24040028MA, SQS24040029MA, SQS24040030MA, SQS24040031MA, SQS24040032MA, SQS24040033MA, SQS24040034MA, SQS24040035MA, SQS24040036MA, SQS24040037MA, SQS24040038MA, SQS24040039MA, SQS24040040MA, SQS24040041MA, SQS24040042MA, SQS24040043MA, SQS24040044MA, SQS24110034MA, SQS24110035MA, SQS24110036MA, SQS24110037MA, SQS24110038MA, SQS24110039MA, SQS24110040MA, SQS24110041MA, SQS24110042MA, SQS24110043MA, SQS24110044MA, SQS24110045MA, SQS24110046MA, SQS24110047MA, SQS24110048MA, SQS24110049MA, SQS24110050MA, SQS24110051MA, SQS24110052MA, SQS24110053MA, SQS24110054MA, SQS24110055MA, SQS24110056MA, SQS24110057MA, SQS24110058MA, SQS24280021MA, SQS24280023MA, SQS24290024MA, SQS24290025MA, SQS24290026MA, SQS24290027MA, SQS24290028MA, SQS24290029MA, SQS24290030MA, SQS24290031MA, SQS24290032MA, SQS24290033MA, SQS25020005MA, SQS25020006MA, SQS25020007MA, SQS25020008MA, SQS25020009MA, SQS25020010MA, SQS25020011MA, SQS25020012MA, SQS25020013MA, SQS25020014MA, SQS24290001MA, SQS24290002MA, SQS24290003MA, SQS24290004MA, SQS24290005MA, SQS24290006MA, SQS24290007MA, SQS24290008MA, SQS24290009MA, SQS24290010MA, SQS23050001WA, SQS23050003WA, SQS23140003WA, SQS23190041WA, SQS23050008WA, SQS23050009WA, SQS24280008WA, SQS24410004WA, SQS24440032WA, SQS24440033WA, SQS24440034WA, SQS24440035WA, SQS24440036WA, SQS24440037WA, SQS24440038WA, SQS24440039WA, SQS24440040WA, SQS24430017WA, SQS23140023WA, SQS23490017WA, SQS24270036WA, SQS24270037WA, SQS24270038WA, SQS23060011WA, SQS24460001WA, SQS23490001WA, SQS23280025WA, SQS24370003WA, SQS24070024WA, SQS24030023WA, SQS23470029WA, SQS24190272WA, SQS23140021WA, SQS23140022WA, SQS23170007WA, SQS24160119WA, SQS23140020WA, SQS23350002WA, SQS24200031WA, SQS24250004WA, SQS24330008WA, SQS24330009WA, SQS24230008WA, SQS23330021WA, SQS23330022WA, SQS23020011WA, SQS23280022WA, SQS24020033WA, SQS24020034WA, SQS24020035WA, SQS24390019WA, SQS24390020WA, SQS24390021WA, SQS24390022WA, SQS24390023WA, SQS24390024WA, SQS24390025WA, SQS24390026WA, SQS24450018WA, SQS23060006WA, SQS23060007WA, SQS23060008WA, SQS23060009WA, SQS24240026WA, SQS24280031WA, SQS24050057WA, SQS24050091WA, SQS24340017WA, SQS24050115WA, SQS24340019WA, SQS24490034WA, SQS24340016WA, SQS24170125WA, SQS23020006WA, SQS23020007WA, SQS23020008WA, SQS23020009WA, SQS23020030WA, SQS23020031WA, SQS23350009WA, SQS24050067WA, SQS24050068WA, SQS23420004WA, SQS23420005WA, SQS24510040WA, SQS24240028WA, SQS24240029WA, SQS24240030WA, SQS24320027WA, SQS24030016WA, SQS24330019WA, SQS24330020WA, SQS23160013WA, SQS24270039WA, SQS24440042WA, SQS24340041WA, SQS24280018WA, SQS23410036WA, SQS24450017WA, SQS23070021WA, SQS23070022WA, SQS24010018WA, SQS24010019WA, SQS24330021WA, SQS24330022WA, SQS24330023WA, SQS24390037WA, SQS24390038WA, SQS24390039WA, SQS24390040WA, SQS24280027WA, SQS24280030WA, SQS23480003WA, SQS23280001WA, SQS23280011WA, SQS24280012WA, SQS24390004WA, SQS24390005WA, SQS24390006WA, SQS24390007WA, SQS23020001WA, SQS23020002WA, SQS23020003WA, SQS23020004WA, SQS24030021WA, SQS24470027WA, SQS24330010WA, SQS24330011WA, SQS23210009WA, SQS23210010WA, SQS23220001WA, SQS24150103WA, SQS24150104WA, SQS23350034WA, SQS23350035WA, SQS23350036WA, SQS24030020WA, SQS24510046WA, SQS24510047WA, SQS23450017WA, SQS23450018WA, SQS23450019WA, SQS23450020WA, SQS23450021WA, SQS23450022WA, SQS24190232WA, SQS24160120WA, SQS23180003WA, SQS23180004WA, SQS23350008WA, SQS23010007WA, SQS23010008WA, SQS23010009WA, SQS23070075WA, SQS23070076WA, SQS24040051WA, SQS24040052WA, SQS23270030WA, SQS23270031WA, SQS23270032WA, SQS23270033WA, SQS23270034WA, SQS23270035WA, SQS24360014WA, SQS24360015WA, SQS24360016WA, SQS24360017WA, SQS24360018WA, SQS24360019WA, SQS24360020WA, SQS23010006WA, SQS24050055WA, SQS23450013WA, SQS23450014WA, SQS24320011WA, SQS24320012WA, SQS24320013WA, SQS24430026WA, SQS24430027WA, SQS24430028WA, SQS24430029WA, SQS24430030WA, SQS24430031WA, SQS24430032WA, SQS24430033WA, SQS24430034WA, SQS24430035WA, SQS24430036WA, SQS24430037WA, SQS24430038WA, SQS24430039WA, SQS24430040WA, SQS24430041WA, SQS24430042WA, SQS24430043WA, SQS24430044WA, SQS24430045WA, SQS24440064WA, SQS24440065WA, SQS24490007WA, SQS24490008WA, SQS24490009WA, SQS24490010WA, SQS24490011WA, SQS24490012WA, SQS24490013WA, SQS24490014WA, SQS24490015WA, SQS24490016WA, SQS24490017WA, SQS24490018WA, SQS24490019WA, SQS24490020WA, SQS24490021WA, SQS24490022WA, SQS24490023WA, SQS24490024WA, SQS24490025WA, SQS24490026WA, SQS24490027WA, SQS24490028WA, SQS24490029WA, SQS24490030WA, SQS24490031WA, SQS24490032WA, SQS24490038WA, SQS24510009WA, SQS24510010WA, SQS24510011WA, SQS24510012WA, SQS24510013WA, SQS24510014WA, SQS24510015WA, SQS24510016WA, SQS24510017WA, SQS24510018WA, SQS24510019WA, SQS24510020WA, SQS24510021WA, SQS24510022WA, SQS24510023WA, SQS24510024WA, SQS24510025WA, SQS24510026WA, SQS24510032WA, SQS24510033WA, SQS24510035WA, SQS24510038WA, SQS24510039WA, SQS24510048WA, SQS25010001WA, SQS25020001WA, SQS25020002WA, SQS25020008WA, SQS25020009WA, SQS25020010WA, SQS25020017WA, SQS25020018WA, SQS25030050WA, SQS25030051WA, SQS25030052WA, SQS25030053WA, SQS25030054WA, SQS25030055WA, SQS25030056WA, SQS25030057WA, SQS25030058WA, SQS25030059WA, SQS25030060WA, SQS25030061WA, SQS25030062WA, SQS25030081WA, SQS25030082WA, SQS25030112WA, SQS25040129WA, SQS25060134WA, SQS25070140WA, SQS23300009WA, SQS23300010WA, SQS23300011WA, SQS23300012WA, SQS23300013WA, SQS23300014WA, SQS23300015WA, SQS23300016WA, SQS23300017WA, SQS23300018WA, SQS23300019WA, SQS23300020WA, SQS23300021WA, SQS23300022WA, SQS23300023WA, SQS23300024WA, SQS23300025WA, SQS23300026WA, SQS23050024WA, SQS23050025WA, SQS23050026WA, SQS23440011WA, SQS23440012WA, SQS23440013WA, SQS23440014WA, SQS23440015WA, SQS23440016WA, SQS23440044WA, SQS23440045WA, SQS24120064WA, SQS24120065WA, SQS24120066WA, SQS24120067WA, SQS24120068WA, SQS24120069WA, SQS24120070WA, SQS24120071WA, SQS24120072WA, SQS24120073WA, SQS24120074WA, SQS24120075WA, SQS24120076WA, SQS24120077WA, SQS24120078WA, SQS24280023WA, SQS24280024WA, SQS24280025WA, SQS24280026WA, SQS24420025MA, SQS24420026MA, SQS25040034MA, SQS23290014MA, SQS25040031MA, SQS25040036MA, SQS23340004MA, SQS25040032MA, SQS25040033MA, SQS25040035MA, SQS24410051WA, SQS24410052WA, SQS24450024WA, SQS25030109WA, SQS25030110WA, SQS25030111WA, SQS23120002WA, SQS24280038WA, SQS24150081WA, SQS23360081WA, SQS23460003WA, SQS23460004WA, SQS25020013WA, SQS24510001MA, SQS23280033WA, SQS23280034WA, SQS24330006WA, SQS24330007WA, SQS24360013WA, SQS23460001WA, SQS23460002WA, SQS23130003WA, SQS23130004WA, SQS23130005WA, SQS23130006WA, SQS23260001MA, SQS23260003MA, SQS23260004MA, SQS23260005MA, SQS23260011MA, SQS23260012MA, SQS23260013MA, SQS23260014MA, SQS23260015MA, SQS24280034WA, SQS24280035WA, SQS23260002MA, SQS25020039WA, SQS23050020WA, SQS23440017WA, SQS23370004MA, SQS23370005MA, SQS24280001MA, SQS24280002MA, SQS24280003MA, SQS24280004MA, SQS24280005MA, SQS24280006MA, SQS24280007MA, SQS24280008MA, SQS24280009MA, SQS24280010MA, SQS24280011MA, SQS24280012MA, SQS24280013MA, SQS24280014MA, SQS24280015MA, SQS24280016MA, SQS24280017MA, SQS24280018MA, SQS24280019MA, SQS24280020MA, SQS23490005WA, SQS23490006WA, SQS23490007WA, SQS23490008WA, SQS23280031WA, SQS24150080WA, SQS23120007WA, SQS23120008WA, SQS23120009WA, SQS23120010WA, SQS23120011WA, SQS23120012WA, SQS23050017WA, SQS23050021WA, SQS23170035WA, SQS23490024WA, SQS23490025WA, SQS23170032WA, SQS24280043WA, SQS24320002WA, SQS24320003WA, SQS23050019WA, SQS23500024WA, SQS23500025WA, SQS24150077WA, SQS24150078WA, SQS24170139WA, SQS24080036WA, SQS23360074WA, SQS23360078WA, SQS23360079WA, SQS23360080WA, SQS23490023WA, SQS24120006WA, SQS24120007WA, SQS24120008WA, SQS23520051WA, SQS24280044WA, SQS24280045WA, SQS24320044WA, SQS23500020WA, SQS23360076WA, SQS24280040WA, SQS24280041WA, SQS24450010WA, SQS24450011WA, SQS24320047WA, SQS24410005WA, SQS23060017WA, SQS23060015WA, SQS24510002MA, SQS24510003MA, SQS24510004MA, SQS24510005MA, SQS24510006MA, SQS24510007MA, SQS24510008MA, SQS24510009MA, SQS24510010MA, SQS24510011MA, SQS24510012MA, SQS24510013MA, SQS24510014MA, SQS24510015MA, SQS24410006WA, SQS24410007WA, SQS24410008WA, SQS24410009WA, SQS24510027WA, SQS24510028WA, SQS24510029WA, SQS24510031WA, SQS25010004WA, SQS25010005WA, SQS25010006WA, SQS25010007WA, SQS25010008WA, SQS25010009WA, SQS25010010WA, SQS25010011WA, SQS25010012WA, SQS25010013WA, SQS25030099WA, SQS25030100WA, SQS25030101WA, SQS25030102WA, SQS25030103WA, SQS25030104WA, SQS25030105WA, SQS25030106WA, SQS25030107WA, SQS25030108WA, SQS25040115WA, SQS25040116WA, SQS25040117WA, SQS25040118WA, SQS25040119WA, SQS24320008WA, SQS24320006WA, SQS23170036WA, SQS23280032WA, SQS24150090WA, SQS24170136WA, SQS24170137WA, SQS23050022WA, SQS24320007WA, SQS24320045WA, SQS23370003MA, SQS23120001WA, SQS23490014WA, SQS24280036WA, SQS23130001WA, SQS23280028WA, SQS23500021WA, SQS24320046WA, SQS24320004WA, SQS23500026WA, SQS23500027WA, SQS23500028WA, SQS23500029WA, SQS23520052WA, SQS23520053WA, SQS23520054WA, SQS23520055WA, SQS24030001WA, SQS24030002WA, SQS24030003WA, SQS24030004WA, SQS24120003WA, SQS24120004WA, SQS24120005WA, SQS24120061WA, SQS24120062WA, SQS24120063WA, SQS23280029WA, SQS23280030WA, SQS23120004WA, SQS23120005WA, SQS24150083WA, SQS24150085WA, SQS24150086WA, SQS24150088WA, SQS24150089WA, SQS24150091WA, SQS23280035WA, SQS23280036WA, SQS23500022WA, SQS23500023WA, SQS24280039WA, SQS24280042WA, SQS24280046WA, SQS24280047WA, SQS23050018WA, SQS23060018WA, SQS23280027WA, SQS24150079WA, SQS24320005WA, SQS24060001MA, SQS23120003WA, SQS23120006WA, SQS23060020WA, SQS24280037WA, SQS23360073WA, SQS23060016WA, SQS23490009WA, SQS23490010WA, SQS23490011WA, SQS23490012WA, SQS23490013WA, SQS23170002WA, SQS23170033WA, SQS23170034WA, SQS23360077WA, SQS23170001WA, SQS23170003WA, SQS23170004WA, SQS23170005WA, SQS23170006WA, SQS23170031WA, SQS24170140WA, SQS23130002WA, SQS23360072WA, SQS23360075WA, SQS24150082WA, SQS24150084WA, SQS24150087WA, SQS24170138WA, SQS24280048WA, SQS23060019WA, SQS24080032WA, SQS24080033WA, SQS24080034WA, SQS24080035WA, SQS24080037WA, SQS24120002WA, SQS24240007MA, SQS23200004WA, SQS24250004MA, SQS24250005MA, SQS24250006MA, SQS24250007MA, SQS24250008MA, SQS23030001WA, SQS23030002WA, SQS23030003WA, SQS24010011MA, SQS24010012MA, SQS24010013MA, SQS24010014MA, SQS24010015MA, SQS24280026MA, SQS24120015MA, SQS24120016MA, SQS24120017MA, SQS24120018MA, SQS24120019MA, SQS24170067MA, SQS24170068MA, SQS24340116MA, SQS24380001MA, SQS23200002WA, SQS23460001MA, SQS23460002MA, SQS23210008MA, SQS24240008MA, SQS24240009MA, SQS24290034MA, SQS24290035MA, SQS24290036MA, SQS24290037MA, SQS24290038MA, SQS24290039MA, SQS24290040MA, SQS24290041MA, SQS24290042MA, SQS24290043MA, SQS24290044MA, SQS24290045MA, SQS24290046MA, SQS24290047MA, SQS24230001MA, SQS24230002MA, SQS24230003MA, SQS24230004MA, SQS24230005MA, SQS24230006MA, SQS24050015WA, SQS23290013WA, SQS24190247WA, SQS24190248WA, SQS24190249WA, SQS24190250WA, SQS24500002WA, SQS24500003WA, SQS24500004WA, SQS24500005WA, SQS24500006WA, SQS24500007WA, SQS24500008WA, SQS24500009WA, SQS24500010WA, SQS24500011WA, SQS25020029WA, SQS25020030WA, SQS25020031WA, SQS25020032WA, SQS25020033WA, SQS24010005WA, SQS24010006WA, SQS24010007WA, SQS23070025WA, SQS23070026WA, SQS24150109WA, SQS23150024WA, SQS23150025WA, SQS23170017WA, SQS24170129WA, SQS24170130WA, SQS24170131WA, SQS24170132WA, SQS23170018WA, SQS24030017WA, SQS24030018WA, SQS24030019WA, SQS23170016WA, SQS23470009WA, SQS23290012WA, SQS23360059WA, SQS25020003WA, SQS25020004WA, SQS25020020WA, SQS25020021WA, SQS25020022WA, SQS25020023WA, SQS25020024WA, SQS25020025WA, SQS25020026WA, SQS25020027WA, SQS25020028WA, SQS25040130WA, SQS25040131WA, SQS25040132WA, SQS25040133WA, SQS23070074WA, SQS23350012WA, SQS23350015WA, SQS24390013WA, SQS24390014WA, SQS24390015WA, SQS24390016WA, SQS23350013WA, SQS23350011WA, SQS23350014WA, SQS23020013WA, SQS23070072WA, SQS23070071WA, SQS23440042WA, SQS24140010WA, SQS24140011WA, SQS24140012WA, SQS24140013WA, SQS24140014WA, SQS23470007WA, SQS23470008WA, SQS23440043WA, SQS23160014WA, SQS24150111WA, SQS24090048WA, SQS24450022WA, SQS23140024WA, SQS24460002WA, SQS24460003WA, SQS23060039WA, SQS23060040WA, SQS24440069WA, SQS24440070WA, SQS24440071WA, SQS23150006WA, SQS23280008WA, SQS24100009WA, SQS25010028WA, SQS25070139WA, SQS23290011WA, SQS23360060WA, SQS23360061WA, SQS23360062WA, SQS23360063WA, SQS23360064WA, SQS23360065WA, SQS23360066WA, SQS23360067WA, SQS23360068WA, SQS23360069WA, SQS24060138WA, SQS23180008WA, SQS23180009WA, SQS24100011WA, SQS24520001WA, SQS24520002WA, SQS24440031WA, SQS25010002WA, SQS24480012WA, SQS23330023WA, SQS23480002WA, SQS24070022WA, SQS24270047WA, SQS24450025WA, SQS23070073WA, SQS24080024WA, SQS24080025WA, SQS24080026WA, SQS24080027WA, SQS24430021WA, SQS24430022WA, SQS24430023WA, SQS24480007WA, SQS24480008WA, SQS24480009WA, SQS24480010WA, SQS24480011WA, SQS23430007WA, SQS23430008WA, SQS23430009WA, SQS23430010WA, SQS23140025WA, SQS23070027WA, SQS23070028WA, SQS24050066WA, SQS23310029WA, SQS24320042WA, SQS24320043WA, SQS23020010WA, SQS23020032WA, SQS23330003WA, SQS23330004WA, SQS23330001WA, SQS23330002WA, SQS24270058WA, SQS24270059WA, SQS23350033WA, SQS23350028WA, SQS23350029WA, SQS23070017WA, SQS23070029WA, SQS23070077WA, SQS23480061WA, SQS23480062WA, SQS23480063WA, SQS24050056WA, SQS24030013WA, SQS24030014WA, SQS24030015WA, SQS24440028WA, SQS24440029WA, SQS24440030WA, SQS24050092WA, SQS24100012WA, SQS23350045WA, SQS24410001WA, SQS24040076WA, SQS24470045WA, SQS23030021WA, SQS23480008WA, SQS24430024WA, SQS24440009WA, SQS23250001WA, SQS23250002WA, SQS23250003WA, SQS24160122WA, SQS24160123WA, SQS24160124WA, SQS24200034WA, SQS23090040WA, SQS24380008WA, SQS23430002WA, SQS24410002WA, SQS24410003WA, SQS23350046WA, SQS23070003WA, SQS23310020WA, SQS23360071WA, SQS24050088WA, SQS23070014WA, SQS23150001WA, SQS23150002WA, SQS23150003WA, SQS23150004WA, SQS23150005WA, SQS23150014WA, SQS23150015WA, SQS23300033WA, SQS23070079WA, SQS23400015WA, SQS24080029WA, SQS24080030WA, SQS25020019WA, SQS23050023WA, SQS24440008WA, SQS24470034WA, SQS24270063WA, SQS24270064WA, SQS24270065WA, SQS24340012WA, SQS23480064WA, SQS24420109WA, SQS24420110WA, SQS23150016WA, SQS23150017WA, SQS23150018WA, SQS23490004WA, SQS23480020WA, SQS23490015WA, SQS23480065WA, SQS24020001WA, SQS24320020WA, SQS24320021WA, SQS24320022WA, SQS24320023WA, SQS24320024WA, SQS24320025WA, SQS24320026WA, SQS24140067WA, SQS23460017WA, SQS23460018WA, SQS23330008WA, SQS23310016WA, SQS24290014WA, SQS24040079WA, SQS24010017WA, SQS24440010WA, SQS24050001WA, SQS23060001WA, SQS23060002WA, SQS23060003WA, SQS23330009WA, SQS24190241WA, SQS24190242WA, SQS24190243WA, SQS24340005WA, SQS23470006WA, SQS23050004WA, SQS23070015WA, SQS24040075WA, SQS23480021WA, SQS23480022WA, SQS24030022WA, SQS23190001WA, SQS23480049WA, SQS24050114WA, SQS23170019WA, SQS24050089WA, SQS24320048WA, SQS23080012WA, SQS24280050WA, SQS24280051WA, SQS24150110WA, SQS24440045WA, SQS24440046WA, SQS24090003WA, SQS23360095WA, SQS23500003WA, SQS24340013WA, SQS23480019WA, SQS23500004WA, SQS24150112WA, SQS24150113WA, SQS23350043WA, SQS24460012WA, SQS24160121WA, SQS24440073WA, SQS23330005WA, SQS23330006WA, SQS23330007WA, SQS24440005WA, SQS24440006WA, SQS24440007WA, SQS24290015WA, SQS24290016WA, SQS24370004WA, SQS23360097WA, SQS23470001WA, SQS23470002WA, SQS23060064WA, SQS23360096WA, SQS24090012WA, SQS24340020WA, SQS23070012WA, SQS23070013WA, SQS24440066WA, SQS23050007WA, SQS23070016WA, SQS23070078WA, SQS23280002WA, SQS23280037WA, SQS23280038WA, SQS23300030WA, SQS23300032WA, SQS23300035WA, SQS23300031WA, SQS24150114WA, SQS24440014WA, SQS24440015WA, SQS24440016WA, SQS24290017WA, SQS23150019WA, SQS23280007WA, SQS23330024WA, SQS23050033WA, SQS23050034WA, SQS23050035WA, SQS23500006WA, SQS23500007WA, SQS23500008WA, SQS23430001WA, SQS23470004WA, SQS23470005WA, SQS24440022WA, SQS24440023WA, SQS24440024WA, SQS24440025WA, SQS24440026WA, SQS24440027WA, SQS23050032WA, SQS24330029WA, SQS24330030WA, SQS23450001WA, SQS24120020MA, SQS24120021MA, SQS24430001MA, SQS23060004WA, SQS23060005WA, SQS23450015WA, SQS23450016WA, SQS24200020WA, SQS24200021WA, SQS24200022WA, SQS24090008WA, SQS24090009WA, SQS24090010WA, SQS23310010WA, SQS24470028WA, SQS24470033WA, SQS24080043WA, SQS23310011WA, SQS24450003WA, SQS24450004WA, SQS24450005WA, SQS24450006WA, SQS24450007WA, SQS24450020WA, SQS24200032WA, SQS24290007WA, SQS24290008WA, SQS23150021WA, SQS23150022WA, SQS23150023WA, SQS23400001WA, SQS23400002WA, SQS24090011WA, SQS24090046WA, SQS24090047WA, SQS24110017WA, SQS24270062WA, SQS23180007WA, SQS24190244WA, SQS24190245WA, SQS24320041WA, SQS24430001WA, SQS23030019WA, SQS23030020WA, SQS23030022WA, SQS24320034WA, SQS24320035WA, SQS24450019WA, SQS23330018WA, SQS23480004WA, SQS23480005WA, SQS24200007WA, SQS23330019WA, SQS23330020WA, SQS24040053WA, SQS24040054WA, SQS24040055WA, SQS23480050WA, SQS24200033WA, SQS24450021WA, SQS23180005WA, SQS23180006WA, SQS23480007WA, SQS24040077WA, SQS24040078WA, SQS24470001WA, SQS24470002WA, SQS24470038WA, SQS24500001WA, SQS24210012WA, SQS24230007WA, SQS24390017WA, SQS24390018WA, SQS24420111WA, SQS23060042WA, SQS23060043WA, SQS23060044WA, SQS23060045WA, SQS23060046WA, SQS23170027WA, SQS23170028WA, SQS23170029WA, SQS23170030WA, SQS23310018WA, SQS23350044WA, SQS23350047WA, SQS23350048WA, SQS23350049WA, SQS23350050WA, SQS23350051WA, SQS23360058WA, SQS23480040WA, SQS23500001WA, SQS23500002WA, SQS24010016WA, SQS24080008WA, SQS24080021WA, SQS24080022WA, SQS24150072WA, SQS24460013WA, SQS24230009WA, SQS24230010WA, SQS23150007WA, SQS23160007WA, SQS23450023WA, SQS23480018WA, SQS24020032WA, SQS24080023WA, SQS24120027WA, SQS24190233WA, SQS24220005WA, SQS24330018WA, SQS24340014WA, SQS24350001WA, SQS24350002WA, SQS23370002WA, SQS23370003WA, SQS23370004WA, SQS23370005WA, SQS23370006WA, SQS23370007WA, SQS23370008WA, SQS23370009WA, SQS23370010WA, SQS23370011WA, SQS23370012WA, SQS23370013WA, SQS23370014WA, SQS24060139WA, SQS24110008MA, SQS24110009MA, SQS24110015MA, SQS24110016MA, SQS24110017MA, SQS24430002MA, SQS24430003MA, SQS24460002MA, SQS24460003MA, SQS24460004MA, SQS24460005MA, SQS23050005WA, SQS23080013WA, SQS23080014WA, SQS23300002WA, SQS24020049WA, SQS24020050WA, SQS24020051WA, SQS24340001WA, SQS24340002WA, SQS24340003WA, SQS23360070WA, SQS24290013WA, SQS24150073WA, SQS24150075WA, SQS23160006WA, SQS23160010WA, SQS23170037WA, SQS23170038WA, SQS23170039WA, SQS23170040WA, SQS23170041WA, SQS23150028WA, SQS23160011WA, SQS23440018WA, SQS24050051WA, SQS23440022WA, SQS24150074WA, SQS23300006WA, SQS23300007WA, SQS23070030WA, SQS23070031WA, SQS24270008WA, SQS24270009WA, SQS24270010WA, SQS24270011WA, SQS24270012WA, SQS24270013WA, SQS24270014WA, SQS24270015WA, SQS24270016WA, SQS24270017WA, SQS24270018WA, SQS24270019WA, SQS24270020WA, SQS24270021WA, SQS24270022WA, SQS24270023WA, SQS24270024WA, SQS24270025WA, SQS24270026WA, SQS24270027WA, SQS24290002WA, SQS24290003WA, SQS24290004WA, SQS24290005WA, SQS24290006WA, SQS23070033WA, SQS23150029WA, SQS24200023WA, SQS24200024WA, SQS24200025WA, SQS24200026WA, SQS24200027WA, SQS24270001WA, SQS24270002WA, SQS24270003WA, SQS24270004WA, SQS24270005WA, SQS24380003WA, SQS24380004WA, SQS24380005WA, SQS24380006WA, SQS24380007WA, SQS24390027WA, SQS24390028WA, SQS24390029WA, SQS24390030WA, SQS24390031WA, SQS23300027WA, SQS23300004WA, SQS23300005WA, SQS23070032WA, SQS23070034WA, SQS23150030WA, SQS23150031WA, SQS24290009WA, SQS24230002WA, SQS24230003WA, SQS24230004WA, SQS24230005WA, SQS24230006WA, SQS24290011WA, SQS24290012WA, SQS23150032WA, SQS23160004WA, SQS23160005WA, SQS23210011WA, SQS23210012WA, SQS23210014WA, SQS23210015WA, SQS23300008WA, SQS23300028WA, SQS23440019WA, SQS23440020WA, SQS23440021WA, SQS24050050WA, SQS24290010WA, SQS24050047WA, SQS24050048WA, SQS24050049WA, SQS23210013WA, SQS24390032WA, SQS24390033WA, SQS24390034WA, SQS24390035WA, SQS24390036WA, SQS24430046WA, SQS24430047WA, SQS24430048WA, SQS24430049WA, SQS25010003WA, SQS24330004WA, SQS24270035WA, SQS24340070MA, SQS24340071MA, SQS24340030MA, SQS24340031MA, SQS24270040WA, SQS24270041WA, SQS24340081MA, SQS24270042WA, SQS24270046WA, SQS24340064MA, SQS24340066MA, SQS24260004WA, SQS24340035MA, SQS24340063MA, SQS24500035MA, SQS24270030WA, SQS24270031WA, SQS24240001WA, SQS24240002WA, SQS24240003WA, SQS24260024WA, SQS24240004WA, SQS24240039WA, SQS24240017WA, SQS24240018WA, SQS24240019WA, SQS24240020WA, SQS24240021WA, SQS24240022WA, SQS24240023WA, SQS24340067MA, SQS24340069MA, SQS24340072MA, SQS24340074MA, SQS24340076MA, SQS24240042WA, SQS24240043WA, SQS24240044WA, SQS24240045WA, SQS24240047WA, SQS24240048WA, SQS24240049WA, SQS24240050WA, SQS24240051WA, SQS24240052WA, SQS24240053WA, SQS24240054WA, SQS24240046WA, SQS24270032WA, SQS24270033WA, SQS24340038MA, SQS24340039MA, SQS24340040MA, SQS24240011WA, SQS24240012WA, SQS24240013WA, SQS24260001WA, SQS24260002WA, SQS24260017WA, SQS24260018WA, SQS24270034WA, SQS24340032MA, SQS24340034MA, SQS24340041MA, SQS24340042MA, SQS24340043MA, SQS24340057MA, SQS24340073MA, SQS24340077MA, SQS24240059WA, SQS24240060WA, SQS24240061WA, SQS24240062WA, SQS24240067WA, SQS24240068WA, SQS24240069WA, SQS24270006WA, SQS24270007WA, SQS24330001WA, SQS24330002WA, SQS24330003WA, SQS24360011MA, SQS24270028WA, SQS24270029WA, SQS24340055MA, SQS24340058MA, SQS24340059MA, SQS24340060MA, SQS24340061MA, SQS24340062MA, SQS24160115WA, SQS24160116WA, SQS24160117WA, SQS24240038WA, SQS24260025WA, SQS24240014WA, SQS24240015WA, SQS24240016WA, SQS24260019WA, SQS24260020WA, SQS24270043WA, SQS24340036MA, SQS24240073WA, SQS24240074WA, SQS24260005WA, SQS24260006WA, SQS24260007WA, SQS24260008WA, SQS24340078MA, SQS24240063WA, SQS24240064WA, SQS24240065WA, SQS24240066WA, SQS24240036WA, SQS24240040WA, SQS24240031WA, SQS24240032WA, SQS24240033WA, SQS24240034WA, SQS24240035WA, SQS24200011MA, SQS24200012MA, SQS24200014MA, SQS24200017MA, SQS24200019MA, SQS24240005WA, SQS24460004WA, SQS24460005WA, SQS24460006WA, SQS24460007WA, SQS24460008WA, SQS24460009WA, SQS24460010WA, SQS24460011WA, SQS25020004MA, SQS24330005WA, SQS24320001MA, SQS24320002MA, SQS24320003MA, SQS24340027MA, SQS24340028MA, SQS24340037MA, SQS24340050MA, SQS24340051MA, SQS24340053MA, SQS24340054MA, SQS24340056MA, SQS24340093MA, SQS24340033MA, SQS24340052MA, SQS24340094MA, SQS24340097MA, SQS24340029MA, SQS24340106MA, SQS24340107MA, SQS24340108MA, SQS24340110MA, SQS24240006WA, SQS24240007WA, SQS24240008WA, SQS24240009WA, SQS24240010WA, SQS24210024WA, SQS24210025WA, SQS24210026WA, SQS24210027WA, SQS24210028WA, SQS24210029WA, SQS24210030WA, SQS24210031WA, SQS24210032WA, SQS24210033WA, SQS24210034WA, SQS24200010MA, SQS24200013MA, SQS24200015MA, SQS24200016MA, SQS24200020MA, SQS24260012WA, SQS24260013WA, SQS24260014WA, SQS24340065MA, SQS24340075MA, SQS24340083MA, SQS24340084MA, SQS24340085MA, SQS24340090MA, SQS24340092MA, SQS24240070WA, SQS24240071WA, SQS24240072WA, SQS24260010WA, SQS24260009WA, SQS24260011WA, SQS24340109MA, SQS24260003WA, SQS24340082MA, SQS24340086MA, SQS24340101MA, SQS24340105MA, SQS24240055WA, SQS24240056WA, SQS24240057WA, SQS24240058WA, SQS24340046MA, SQS24340047MA, SQS24340048MA, SQS24340049MA, SQS24340079MA, SQS24340087MA, SQS24340088MA, SQS24340089MA, SQS24340091MA, SQS24340098MA, SQS24340099MA, SQS24340102MA, SQS24340103MA, SQS24340104MA, SQS24270044WA, SQS24270045WA, SQS24340080MA, SQS24240037WA, SQS24260021WA, SQS24260022WA, SQS24260023WA, SQS24340095MA, SQS24340096MA, SQS24340100MA, SQS24500034MA, SQS24340111MA, SQS24340068MA, SQS24250001WA, SQS24250002WA, SQS24250003WA, SQS23110002WA, SQS23110003WA, SQS24050058WA, SQS24050059WA, SQS24050060WA, SQS24050061WA, SQS23350052WA, SQS23350053WA, SQS23350056WA, SQS23350057WA, SQS23160008WA, SQS23160009WA, SQS24440020WA, SQS23180001WA, SQS24090001WA, SQS24090002WA, SQS23480048WA, SQS24390002WA, SQS23480044WA, SQS23480045WA, SQS23480046WA, SQS23480047WA, SQS24430015WA, SQS23060041WA, SQS23150020WA, SQS23020005WA, SQS23310027WA, SQS23310028WA, SQS24190246WA, SQS24210011WA, SQS25030073WA, SQS24010001WA, SQS24140001WA, SQS24010009WA, SQS24040046WA, SQS24040047WA, SQS24070023WA, SQS24040048WA, SQS23170021WA, SQS23180010WA, SQS23440020MA, SQS23440021MA, SQS23440022MA, SQS24050031MA, SQS24050032MA, SQS24050033MA, SQS24050034MA, SQS24050035MA, SQS24140002MA, SQS24140003MA, SQS23280024WA, SQS24110006MA, SQS24010048MA, SQS24040015MA, SQS24270061WA, SQS23500006MA, SQS23080021WA, SQS23080023WA, SQS23080024WA, SQS23170013WA, SQS24450008WA, SQS24450009WA, SQS24450016WA, SQS23080022WA, SQS23170010WA, SQS23170011WA, SQS23170012WA, SQS23170014WA, SQS23170015WA, SQS23420001WA, SQS23420002WA, SQS23420003WA, SQS24340018WA, SQS24150107WA, SQS24150108WA, SQS24500012WA, SQS24500013WA, SQS24500014WA, SQS24500015WA, SQS24500016WA, SQS24500017WA, SQS24500018WA, SQS24500019WA, SQS24500020WA, SQS24500021WA, SQS23100001WA, SQS23100002WA, SQS23100003WA, SQS23100004WA, SQS23180036WA, SQS23180037WA, SQS23190038WA, SQS23190039WA, SQS23190040WA, SQS24080028WA, SQS24080031WA, SQS24150106WA, SQS24440067WA, SQS25060135WA, SQS23360086WA, SQS23150054WA, SQS23160055WA, SQS23410040WA, SQS24220006WA, SQS24430019WA, SQS24430020WA, SQS23180013WA, SQS24240041WA, SQS24200004WA, SQS24200005WA, SQS24200006WA, SQS23080005WA, SQS23410013WA, SQS24360011WA, SQS24360012WA, SQS24032005WA, SQS24032006WA, SQS24032007WA, SQS23350030WA, SQS23350031WA, SQS23350032WA, SQS23310031WA, SQS23310032WA, SQS23030033WA, SQS24042000WA, SQS24042001WA, SQS24042002WA, SQS24480001WA, SQS24480002WA, SQS24360024WA, SQS23300003WA, SQS23480037WA, SQS24030009WA, SQS24030010WA, SQS24030011WA, SQS24030012WA, SQS23330010WA, SQS23330011WA, SQS24032009WA, SQS24032010WA, SQS24032011WA, SQS23080010WA, SQS24360021WA, SQS24360022WA, SQS24360023WA, SQS24340007WA, SQS23470018WA, SQS23470019WA, SQS23470020WA, SQS23470021WA, SQS23080001WA, SQS23310030WA, SQS23480027WA, SQS23480028WA, SQS24340008WA, SQS24340009WA, SQS24032012WA, SQS24150076WA, SQS24220001WA, SQS24032003WA, SQS23460015WA, SQS23080004WA, SQS23500014WA, SQS23500015WA, SQS23500016WA, SQS23500017WA, SQS23500018WA, SQS24042003WA, SQS23500013WA, SQS23500019WA, SQS24470047WA, SQS24150069WA, SQS24150070WA, SQS24280010WA, SQS24280011WA, SQS23080002WA, SQS23210018WA, SQS23210019WA, SQS23470030WA, SQS24450001WA, SQS23180012WA, SQS23180014WA, SQS23350001WA, SQS23360088WA, SQS23360089WA, SQS23360090WA, SQS23360091WA, SQS24280005WA, SQS24280006WA, SQS24280007WA, SQS23360092WA, SQS23360093WA, SQS23210005WA, SQS23210006WA, SQS24230001WA, SQS23080009WA, SQS24430018WA, SQS23370016WA, SQS24022009WA, SQS23410041WA, SQS24042034WA, SQS24042035WA, SQS24042036WA, SQS24042037WA, SQS24042038WA, SQS24042039WA, SQS24042040WA, SQS24042041WA, SQS24042042WA, SQS23410001WA, SQS23410002WA, SQS23410003WA, SQS23410004WA, SQS23410005WA, SQS23020014WA, SQS23470031WA, SQS24380001WA, SQS23500011WA, SQS23500012WA, SQS23080006WA, SQS24470003WA, SQS24470044WA, SQS24470046WA, SQS24480016WA, SQS24490001WA, SQS24490002WA, SQS24490004WA, SQS24490005WA, SQS24022000WA, SQS24022001WA, SQS24022002WA, SQS24022003WA, SQS24022004WA, SQS24022005WA, SQS24022006WA, SQS24022007WA, SQS24022008WA, SQS24220004WA, SQS23240015WA, SQS23240016WA, SQS23240017WA, SQS23240018WA, SQS24160118WA, SQS24032013WA, SQS24450002WA, SQS24150071WA, SQS24280009WA, SQS24430025WA, SQS24440068WA, SQS23080003WA, SQS23290015WA, SQS23410037WA, SQS23410038WA, SQS23480041WA, SQS24140068WA, SQS23010012WA, SQS23470027WA, SQS24490003WA, SQS23470022WA, SQS23470023WA, SQS23470024WA, SQS23470025WA, SQS23470026WA, SQS23480033WA, SQS23480034WA, SQS23480035WA, SQS23480036WA, SQS23460005WA, SQS23460006WA, SQS23460008WA, SQS23460009WA, SQS23460010WA, SQS23470010WA, SQS23470011WA, SQS23470012WA, SQS23470013WA, SQS23470014WA, SQS23470015WA, SQS24360008WA, SQS24360009WA, SQS24360010WA, SQS23370015WA, SQS23370017WA, SQS23480042WA, SQS23480043WA, SQS23500009WA, SQS23500010WA, SQS24490006WA, SQS23480038WA, SQS24360002WA, SQS24360003WA, SQS24360004WA, SQS24360005WA, SQS24360006WA, SQS24360007WA, SQS24340021WA, SQS24340022WA, SQS24340023WA, SQS24340024WA, SQS24340025WA, SQS24340026WA, SQS24340027WA, SQS24340028WA, SQS24340029WA, SQS24340030WA, SQS24340031WA, SQS24340032WA, SQS24340033WA, SQS24340034WA, SQS24340035WA, SQS24340036WA, SQS24340037WA, SQS24340038WA, SQS24340039WA, SQS24340040WA, SQS23460007WA, SQS23480006WA, SQS23470032WA, SQS23470033WA, SQS23480025WA, SQS23480026WA, SQS24010010WA, SQS23080008WA, SQS24480013WA, SQS24480014WA, SQS24480015WA, SQS24480017WA, SQS24480018WA, SQS24032008WA, SQS23080007WA, SQS24370005WA, SQS24370006WA, SQS24370007WA, SQS24430016WA, SQS23360087WA, SQS24340006WA, SQS24510001WA, SQS24510002WA, SQS24510003WA, SQS24510004WA, SQS24510005WA, SQS24510006WA, SQS24510007WA, SQS24510008WA, SQS23480009WA, SQS23480010WA, SQS23480011WA, SQS23480012WA, SQS23480013WA, SQS23480014WA, SQS23480015WA, SQS23480016WA, SQS23480017WA, SQS24440041WA, SQS23290016WA, SQS23290017WA, SQS24480003WA, SQS24480004WA, SQS24220003WA, SQS23480001WA, SQS24210007WA, SQS23410039WA, SQS23410042WA, SQS23080018WA, SQS23370001WA, SQS24200001WA, SQS24200002WA, SQS24200003WA, SQS23410017WA, SQS23470016WA, SQS23470017WA, SQS23500030WA, SQS24220002WA, SQS23140002WA, SQS24150041MA, SQS24010020MA, SQS24080019WA, SQS24080020WA, SQS24410050WA, SQS23080016WA, SQS23080017WA, SQS23210016WA, SQS23210017WA, SQS23280021WA, SQS23350054WA, SQS23350055WA, SQS23460011WA, SQS23460012WA, SQS23460013WA, SQS23460014WA, SQS23460016WA, SQS24120001WA, SQS24140018WA, SQS24210001WA, SQS24210002WA, SQS24320009WA, SQS24320010WA, SQS24340010WA, SQS24340011WA, SQS24450023WA, SQS24450012WA, SQS24450013WA, SQS24450014WA, SQS24450015WA, SQS24470004WA, SQS24470005WA, SQS24470006WA, SQS24470007WA, SQS24420041WA, SQS24420042WA, SQS24420043WA, SQS24420044WA, SQS24420045WA, SQS24420046WA, SQS24420047WA, SQS24420048WA, SQS24420049WA, SQS24420050WA, SQS24420051WA, SQS24410029WA, SQS24410030WA, SQS24410031WA, SQS24410032WA, SQS24410033WA, SQS24410034WA, SQS24410035WA, SQS24410017WA, SQS24410018WA, SQS24410019WA, SQS24410020WA, SQS24410021WA, SQS24410022WA, SQS24410023WA, SQS24410024WA, SQS24410025WA, SQS24410026WA, SQS24410027WA, SQS24410028WA, SQS24410036WA, SQS24410037WA, SQS24410038WA, SQS24410039WA, SQS24410040WA, SQS24410041WA, SQS24410042WA, SQS24410043WA, SQS24410044WA, SQS24410045WA, SQS24410046WA, SQS24410053WA, SQS24410054WA, SQS24410055WA, SQS24410056WA, SQS24410057WA, SQS24410058WA, SQS24410059WA, SQS24410060WA, SQS24410061WA, SQS24410062WA, SQS24410063WA, SQS24410064WA, SQS24410065WA, SQS24410066WA, SQS24410067WA, SQS24410070WA, SQS24420017WA, SQS24420018WA, SQS24420019WA, SQS24420020WA, SQS24420021WA, SQS24420022WA, SQS24420023WA, SQS24420024WA, SQS24420025WA, SQS24420026WA, SQS24420027WA, SQS24420028WA, SQS24420029WA, SQS24420030WA, SQS24420035WA, SQS24420037WA, SQS24420039WA, SQS24420040WA, SQS24420061WA, SQS24420062WA, SQS24420063WA, SQS24420064WA, SQS24420065WA, SQS24420066WA, SQS24420067WA, SQS24420068WA, SQS24420069WA, SQS24420070WA, SQS24420071WA, SQS24420072WA, SQS24420073WA, SQS24420090WA, SQS24420091WA, SQS24420092WA, SQS24420093WA, SQS24420094WA, SQS24420095WA, SQS24420096WA, SQS24420097WA, SQS24420098WA, SQS24420099WA, SQS24420100WA, SQS24420101WA, SQS24420102WA, SQS24420052WA, SQS24420053WA, SQS24420054WA, SQS24420055WA, SQS24420056WA, SQS24420057WA, SQS24420058WA, SQS24420059WA, SQS24420060WA, SQS24410094WA, SQS24410010WA, SQS24410011WA, SQS24410012WA, SQS24410013WA, SQS24410014WA, SQS24410015WA, SQS24410090WA, SQS24410091WA, SQS24410092WA, SQS24410093WA, SQS24410095WA, SQS24410096WA, SQS24410097WA, SQS24420001WA, SQS24420002WA, SQS24420003WA, SQS24420004WA, SQS24420005WA, SQS24420006WA, SQS24420007WA, SQS24420008WA, SQS24420009WA, SQS24420010WA, SQS24420011WA, SQS24420012WA, SQS24420013WA, SQS24420014WA, SQS24420015WA, SQS24420016WA, SQS24410068WA, SQS24410071WA, SQS24410074WA, SQS24410075WA, SQS24410076WA, SQS24410077WA, SQS24410078WA, SQS24410079WA, SQS24410080WA, SQS24410081WA, SQS24410082WA, SQS24410083WA, SQS24410084WA, SQS24410085WA, SQS24410086WA, SQS24410087WA, SQS24410088WA, SQS24410089WA, SQS24050078WA, SQS24420031WA, SQS24420032WA, SQS24420033WA, SQS24420034WA, SQS24420036WA, SQS24420038WA, SQS24050079WA, SQS24410073WA, SQS24410069WA, SQS24410072WA, SQS24190238WA, SQS24190239WA, SQS24200019WA, SQS24500003MA, SQS24010046MA, SQS24170089MA, SQS24170090MA, SQS24170091MA, SQS24170092MA, SQS24170093MA, SQS24400004MA, SQS24400005MA, SQS24400006MA, SQS24260007MA, SQS24450008MA, SQS24450009MA, SQS24450010MA, SQS24450011MA, SQS24450012MA, SQS24450013MA, SQS24450014MA, SQS24450015MA, SQS24450016MA, SQS24450017MA, SQS24450018MA, SQS24450019MA, SQS24450020MA, SQS24010047MA, SQS24340001MA, SQS24340002MA, SQS24340003MA, SQS24340004MA, SQS24340005MA, SQS24340006MA, SQS24340007MA, SQS24340008MA, SQS24340009MA, SQS24340010MA, SQS24340011MA, SQS24340012MA, SQS24340013MA, SQS23270001WA, SQS23270002WA, SQS23270003WA, SQS23300029WA, SQS23440056WA, SQS24060137WA, SQS24200013WA, SQS24200014WA, SQS24200015WA, SQS24200016WA, SQS24200017WA, SQS24200018WA, SQS24440043WA, SQS24440044WA, SQS24470032WA, SQS24280013WA, SQS24280020WA, SQS24280021WA, SQS23400004WA, SQS23400005WA, SQS23400006WA, SQS23400007WA, SQS23400008WA, SQS23400009WA, SQS23180011WA, SQS23180015WA, SQS23180016WA, SQS23180018WA, SQS23490018WA, SQS23490019WA, SQS23490020WA, SQS23490021WA, SQS23490022WA, SQS24050053WA, SQS24050054WA, SQS23180017WA, SQS23180021WA, SQS23260004WA, SQS23260005WA, SQS23260006WA, SQS23260007WA, SQS24050052WA, SQS23480066WA, SQS23480067WA, SQS23480068WA, SQS24390001WA, SQS23180019WA, SQS23180020WA, SQS23260008WA, SQS24050113WA, SQS24340004WA, SQS23220003WA, SQS24470008WA, SQS24470009WA, SQS23060012WA, SQS23060013WA, SQS23060014WA, SQS23070024WA, SQS23140001WA, SQS23150033WA, SQS23150034WA, SQS23150035WA, SQS23150036WA, SQS23150037WA, SQS23150038WA, SQS23150039WA, SQS23150040WA, SQS23150041WA, SQS23150042WA, SQS24320030WA, SQS24320031WA, SQS24320032WA, SQS23310019WA, SQS23310021WA, SQS23310022WA, SQS23310023WA, SQS24420103WA, SQS24420104WA, SQS24420105WA, SQS24420106WA, SQS24420107WA, SQS24420108WA, SQS24420112WA, SQS24420113WA, SQS23100015WA, SQS24150105WA, SQS24280022WA, SQS24340015WA, SQS24140015WA, SQS24140016WA, SQS24140017WA, SQS23420010WA, SQS23420012WA, SQS23420014WA, SQS23420018WA, SQS23420021WA, SQS23420026WA, SQS23420027WA, SQS23420034WA, SQS23420043WA, SQS23420044WA, SQS23420051WA, SQS23430003WA, SQS23430006WA, SQS24010008WA, SQS23490016WA, SQS23480039WA, SQS24330025WA, SQS24330026WA, SQS24330027WA, SQS23100013WA, SQS23100014WA, SQS23420033WA, SQS23420040WA, SQS23420042WA, SQS23420049WA, SQS23420050WA, SQS23420056WA, SQS23420069WA, SQS23430004WA, SQS24210008WA, SQS24420135WA, SQS24420136WA, SQS23190002WA, SQS23480029WA, SQS23480030WA, SQS23480031WA, SQS23480032WA, SQS23522002WA, SQS24032014WA, SQS24032015WA, SQS24032016WA, SQS24032017WA, SQS24032018WA, SQS24032019WA, SQS24032020WA, SQS24032021WA, SQS24032022WA, SQS24032023WA, SQS24042026WA, SQS23170008WA, SQS23170009WA, SQS24280015WA, SQS23420035WA, SQS23420045WA, SQS23420058WA, SQS23420063WA, SQS23420068WA, SQS23420070WA, SQS23430005WA, SQS23522000WA, SQS23522003WA, SQS23522005WA, SQS23522017WA, SQS23522027WA, SQS23522028WA, SQS24042005WA, SQS24042012WA, SQS24042015WA, SQS24042017WA, SQS24042018WA, SQS24042020WA, SQS24042022WA, SQS24042023WA, SQS24042025WA, SQS24042028WA, SQS23240019WA, SQS23240020WA, SQS23420037WA, SQS23420038WA, SQS23420053WA, SQS23420055WA, SQS23420061WA, SQS23420072WA, SQS23420074WA, SQS23190003WA, SQS24130001WA, SQS23280026WA, SQS24440063WA, SQS23420041WA, SQS23420048WA, SQS23420064WA, SQS24100010WA, SQS24050065WA, SQS24110059WA, SQS24110060WA, SQS24470029WA, SQS24470030WA, SQS24470031WA, SQS23420006WA, SQS23350010WA, SQS24070026WA, SQS23420008WA, SQS23420009WA, SQS23420011WA, SQS23420015WA, SQS23420016WA, SQS23420017WA, SQS23420019WA, SQS23420020WA, SQS23420022WA, SQS23420024WA, SQS23420025WA, SQS23420029WA, SQS23420039WA, SQS23420057WA, SQS23420060WA, SQS23522008WA, SQS23522034WA, SQS24050098WA, SQS24050100WA, SQS24050106WA, SQS24510041WA, SQS24050071WA, SQS24050072WA, SQS24050107WA, SQS23420071WA, SQS23420023WA, SQS23420067WA, SQS23420075WA, SQS24210003WA, SQS24210004WA, SQS24210005WA, SQS24210006WA, SQS23420028WA, SQS23522001WA, SQS23522004WA, SQS23522007WA, SQS23522009WA, SQS23522011WA, SQS23522014WA, SQS23522020WA, SQS23522021WA, SQS23522023WA, SQS23522025WA, SQS23522029WA, SQS23522035WA, SQS23522036WA, SQS24042004WA, SQS24042006WA, SQS24042007WA, SQS24042008WA, SQS24042009WA, SQS24042011WA, SQS24042013WA, SQS24042016WA, SQS24042021WA, SQS24042024WA, SQS24042027WA, SQS24042029WA, SQS24042030WA, SQS24042031WA, SQS24042033WA, SQS24050070WA, SQS24050096WA, SQS24050097WA, SQS24050099WA, SQS24050101WA, SQS24050102WA, SQS24050103WA, SQS24050104WA, SQS24050105WA, SQS24050108WA, SQS24050109WA, SQS24050110WA, SQS24050111WA, SQS24190235WA, SQS24190230WA, SQS24190231WA, SQS23060038WA, SQS23420036WA, SQS23420054WA, SQS24330024WA, SQS23420013WA, SQS23420052WA, SQS23420059WA, SQS23420062WA, SQS23420065WA, SQS23420066WA, SQS23060047WA, SQS23060048WA, SQS23522015WA, SQS23522019WA, SQS23522022WA, SQS24050069WA, SQS24050073WA, SQS24050074WA, SQS24050075WA, SQS24050076WA, SQS24050077WA, SQS23080019WA, SQS23080020WA, SQS23220002WA, SQS24240024WA, SQS24240025WA, SQS24270060WA, SQS24190271WA, SQS24280049WA, SQS24280052WA, SQS24430002WA, SQS24430003WA, SQS24430004WA, SQS24430005WA, SQS24430006WA, SQS24430007WA, SQS24430008WA, SQS24430009WA, SQS24430010WA, SQS24430011WA, SQS24430012WA, SQS24430013WA, SQS24430014WA, SQS23200003WA, SQS24210009WA, SQS24210010WA, SQS23290018WA, SQS23420046WA, SQS23420047WA, SQS23420073WA, SQS23420076WA, SQS23500031WA, SQS23500032WA, SQS23500033WA, SQS23500034WA, SQS23500035WA, SQS23500036WA, SQS23500037WA, SQS23500038WA, SQS23500039WA, SQS23500040WA, SQS23500041WA, SQS23500042WA, SQS23500043WA, SQS23500044WA, SQS23500045WA, SQS23500046WA, SQS23500047WA, SQS23500048WA, SQS23500049WA, SQS23500050WA, SQS23522006WA, SQS23522010WA, SQS23522012WA, SQS23522013WA, SQS23522016WA, SQS23522018WA, SQS23522024WA, SQS23522026WA, SQS23522030WA, SQS23522031WA, SQS23522032WA, SQS23522033WA, SQS23522037WA, SQS23522038WA, SQS23522039WA, SQS24042010WA, SQS24042014WA, SQS24042019WA, SQS24042032WA, SQS24140019WA, SQS24140020WA, SQS23050027WA, SQS23050028WA, SQS23050029WA, SQS23050030WA, SQS23050031WA, SQS24030036WA, SQS24040045WA, SQS24510034WA, SQS24440021WA, SQS23230004WA, SQS23230005WA, SQS23230006WA, SQS23230007WA, SQS23230013WA, SQS23240014WA, SQS24330012WA, SQS24330013WA, SQS24330014WA, SQS24330015WA, SQS24330016WA, SQS24330017WA, SQS24270057WA, SQS23410012WA, SQS23160015WA, SQS24170126WA, SQS24170127WA, SQS24170128WA, SQS24040049WA, SQS23020012WA, SQS24280032WA, SQS24280033WA, SQS24050025WA, SQS24050116WA, SQS24050117WA, SQS24120033WA, SQS24120034WA, SQS24120035WA, SQS24120036WA, SQS24120037WA, SQS24120038WA, SQS24120039WA, SQS24120040WA, SQS24120041WA, SQS24180164WA, SQS24180165WA, SQS24180166WA, SQS24180197WA, SQS24180198WA, SQS24180199WA, SQS24180200WA, SQS24180201WA, SQS24180202WA, SQS24180203WA, SQS24180204WA, SQS24180205WA, SQS24180206WA, SQS24180207WA, SQS24180208WA, SQS24180209WA, SQS24180210WA, SQS24180211WA, SQS24180212WA, SQS24180213WA, SQS24180214WA, SQS24180215WA, SQS24180216WA, SQS24180217WA, SQS24180218WA, SQS24180219WA, SQS24190220WA, SQS24190221WA, SQS24190222WA, SQS24190223WA, SQS24190224WA, SQS24190225WA, SQS24190226WA, SQS24190227WA, SQS24190228WA, SQS24190229WA, SQS23060049WA, SQS23060050WA, SQS23060051WA, SQS23070018WA, SQS23070019WA, SQS23180022WA, SQS23180023WA, SQS23180024WA, SQS23180025WA, SQS23180026WA, SQS23180027WA, SQS23180028WA, SQS23180029WA, SQS23180030WA, SQS23180031WA, SQS23280003WA, SQS23280004WA, SQS23280012WA, SQS23280016WA, SQS23280017WA, SQS23280018WA, SQS23280019WA, SQS23280020WA, SQS23410030WA, SQS23410031WA, SQS23410032WA, SQS23410033WA, SQS23410034WA, SQS23410035WA, SQS24020002WA, SQS24020003WA, SQS24020004WA, SQS24020005WA, SQS24040056WA, SQS24040057WA, SQS24040058WA, SQS24040059WA, SQS24040060WA, SQS24040061WA, SQS24040062WA, SQS24040063WA, SQS24040064WA, SQS24040065WA, SQS24040066WA, SQS24040067WA, SQS24040068WA, SQS24040069WA, SQS24040070WA, SQS24040071WA, SQS24040072WA, SQS24040073WA, SQS24040074WA, SQS24050005WA, SQS24050006WA, SQS24050007WA, SQS24050008WA, SQS24060118WA, SQS24060119WA, SQS24060120WA, SQS24060121WA, SQS24060122WA, SQS24060123WA, SQS24060124WA, SQS24060125WA, SQS24060126WA, SQS24060127WA, SQS24060128WA, SQS24060129WA, SQS24060130WA, SQS24060131WA, SQS24060132WA, SQS24060133WA, SQS24060134WA, SQS24060135WA, SQS24060136WA, SQS24200008WA, SQS24200009WA, SQS24200010WA, SQS24200011WA, SQS24200012WA, SQS24010041WA, SQS24010042WA, SQS24010043WA, SQS24010044WA, SQS24010045WA, SQS24050017WA, SQS24050018WA, SQS24050019WA, SQS24050020WA, SQS24050021WA, SQS24050022WA, SQS24050023WA, SQS24050024WA, SQS24050026WA, SQS24050027WA, SQS24050028WA, SQS24050029WA, SQS24050030WA, SQS24130002MA, SQS24180107MA, SQS24180108MA, SQS24390001MA, SQS24180109MA, SQS24190111MA, SQS24290081MA, SQS24290082MA, SQS24280029MA, SQS23310005MA, SQS25020020MA, SQS25040039MA, SQS25040040MA, SQS25040041MA, SQS25040042MA, SQS24410015MA, SQS24410016MA, SQS24130003MA, SQS24080001WA, SQS24190251WA, SQS24190252WA, SQS24190253WA, SQS24190254WA, SQS24190255WA, SQS24190256WA, SQS24190257WA, SQS24190258WA, SQS24190259WA, SQS24190260WA, SQS24190261WA, SQS24190262WA, SQS24190263WA, SQS24190264WA, SQS24190265WA, SQS24190266WA, SQS24190267WA, SQS24190268WA, SQS24190269WA, SQS24190270WA, SQS24420074WA, SQS24420075WA, SQS24420076WA, SQS24420077WA, SQS24420078WA, SQS24420079WA, SQS24420080WA, SQS24420081WA, SQS24420082WA, SQS24420083WA, SQS24420084WA, SQS24420085WA, SQS24420086WA, SQS24420087WA, SQS24420088WA, SQS24420089WA, SQS24440047WA, SQS24440048WA, SQS24440049WA, SQS24440050WA, SQS24440051WA, SQS24440052WA, SQS24440053WA, SQS24440054WA, SQS24440055WA, SQS24440056WA, SQS24440057WA, SQS24440058WA, SQS24440059WA, SQS24440060WA, SQS24440061WA, SQS25010014WA, SQS25010015WA, SQS25010016WA, SQS25010017WA, SQS25010018WA, SQS25010019WA, SQS25010020WA, SQS25010021WA, SQS25010022WA, SQS25010023WA, SQS25010024WA, SQS25010025WA, SQS25010026WA, SQS25010027WA, SQS25030089WA, SQS25030090WA, SQS25030091WA, SQS23140010WA, SQS23140011WA, SQS23140015WA, SQS23170024WA, SQS23440004WA, SQS23440005WA, SQS23440006WA, SQS23440008WA, SQS23440009WA, SQS23440010WA, SQS24170152WA, SQS24170153WA, SQS24170154WA, SQS24170155WA, SQS24170156WA, SQS24170157WA, SQS24170158WA, SQS24170159WA, SQS24170160WA, SQS24170161WA, SQS24420114WA, SQS24420115WA, SQS24420116WA, SQS24420117WA, SQS24420118WA, SQS24420119WA, SQS24420120WA, SQS24420121WA, SQS24420122WA, SQS24420123WA, SQS24420124WA, SQS24420125WA, SQS24420126WA, SQS24420127WA, SQS24420128WA, SQS24420129WA, SQS24420130WA, SQS24420131WA, SQS24420132WA, SQS24420133WA, SQS24420134WA, SQS23480069WA, SQS23480070WA, SQS23490002WA, SQS23490003WA, SQS24080002WA, SQS24080003WA, SQS24080004WA, SQS24080005WA, SQS24080006WA, SQS24080007WA, SQS23140018WA, SQS23290005WA, SQS23440002WA, SQS23440003WA, SQS23140016WA, SQS23290004WA, SQS23140017WA, SQS25030092WA, SQS25030093WA, SQS25030094WA, SQS25030095WA, SQS25030096WA, SQS25030097WA, SQS25030098WA, SQS23290006WA, SQS23140014WA, SQS23140013WA, SQS23290003WA, SQS23140019WA, SQS23170022WA, SQS23170025WA, SQS23170023WA, SQS23170026WA, SQS23290002WA, SQS23140012WA, SQS23440001WA, SQS23440007WA, SQS23140004WA, SQS23140005WA, SQS23140006WA, SQS23140007WA, SQS23150026WA, SQS23150027WA, SQS23150043WA, SQS23150044WA, SQS23150045WA, SQS23150046WA, SQS23150047WA, SQS23150048WA, SQS23150049WA, SQS23150050WA, SQS23150051WA, SQS23150052WA, SQS23150053WA, SQS24050112WA, SQS24140066WA, SQS24270049WA, SQS24270050WA, SQS24270051WA, SQS24270052WA, SQS24270053WA, SQS24270054WA, SQS24270055WA, SQS24270056WA, SQS24330028WA, SQS23070005WA, SQS23070006WA, SQS23070007WA, SQS23070008WA, SQS23070009WA, SQS23070010WA, SQS23070011WA, SQS23070023WA, SQS23280013WA, SQS23280014WA, SQS23280015WA, SQS24050093WA, SQS24050094WA, SQS24050095WA, SQS24120026WA, SQS24470010WA, SQS24470011WA, SQS24470012WA, SQS24470013WA, SQS23310012WA, SQS23310013WA, SQS23310014WA, SQS23310015WA, SQS24110007MA, SQS24110011MA, SQS24110012MA, SQS24140007WA, SQS24140008WA, SQS24140009WA, SQS24110010MA, SQS24430050WA, SQS24500025MA, SQS24500026MA, SQS24500027MA, SQS24500028MA, SQS24500029MA, SQS24500030MA, SQS24510022MA, SQS24510023MA, SQS24510024MA, SQS24510025MA, SQS24510026MA, SQS24510027MA, SQS24510034MA, SQS24510035MA, SQS25020002MA, SQS23470028WA, SQS23350003WA, SQS23350004WA, SQS23350005WA, SQS23350006WA, SQS23350007WA, SQS24170162WA, SQS24170163WA, SQS24470041WA, SQS24470042WA, SQS25020005WA, SQS25020006WA, SQS25020007WA, SQS25040121WA, SQS25040122WA, SQS25060136WA, SQS25060137WA, SQS25060138WA, SQS23350016WA, SQS23350017WA, SQS23350018WA, SQS23350019WA, SQS23350020WA, SQS23350021WA, SQS23350022WA, SQS23350037WA, SQS23350038WA, SQS23350039WA, SQS23080011WA, SQS23080015WA, SQS23240002WA, SQS23240003WA, SQS23240004WA, SQS23240005WA, SQS23240006WA, SQS23240007WA, SQS23240008WA, SQS23240009WA, SQS24090013WA, SQS24090014WA, SQS24090015WA, SQS23440046WA, SQS23440047WA, SQS23440048WA, SQS23440049WA, SQS23440050WA, SQS23440051WA, SQS23440052WA, SQS23440053WA, SQS23440054WA, SQS23440055WA, SQS24320014WA, SQS24320015WA, SQS24320016WA, SQS24320017WA, SQS24320018WA, SQS24320019WA, SQS23180002WA, SQS25010029WA, SQS25010030WA, SQS23410014WA, SQS23410015WA, SQS23410016WA, SQS24080038WA, SQS24080039WA, SQS24080040WA, SQS24080041WA, SQS24080042WA, SQS23030011WA, SQS23030012WA, SQS23030013WA, SQS23030014WA, SQS23030015WA, SQS23030016WA, SQS23030017WA, SQS23030018WA, SQS24190236WA, SQS24190237WA, SQS23410026WA, SQS23410027WA, SQS23410028WA, SQS23410029WA, SQS23070004WA, SQS23050014WA, SQS23360084WA, SQS24100066WA, SQS24100067WA, SQS24100068WA, SQS24100069WA, SQS24100070WA, SQS24100071WA, SQS24100072WA, SQS24100073WA, SQS24100074WA, SQS24100075WA, SQS24100076WA, SQS24100077WA, SQS24100078WA, SQS24100079WA, SQS24100080WA, SQS24100081WA, SQS24100082WA, SQS23360082WA, SQS23360083WA, SQS23360085WA, SQS23280009WA, SQS23350023WA, SQS23350024WA, SQS23350025WA, SQS23350026WA, SQS23350027WA, SQS23350040WA, SQS23350041WA, SQS23350042WA, SQS23410006WA, SQS23410007WA, SQS23410008WA, SQS23280010WA, SQS24040050WA, SQS24080045WA, SQS23170020WA, SQS24040040WA, SQS24040041WA, SQS23060021WA, SQS23060022WA, SQS23060023WA, SQS23060024WA, SQS23060025WA, SQS23060026WA, SQS23060027WA, SQS23060028WA, SQS23060029WA, SQS23060030WA, SQS23060031WA, SQS23060032WA, SQS23060033WA, SQS23060034WA, SQS23060035WA, SQS23060036WA, SQS23060037WA, SQS23060052WA, SQS23060053WA, SQS23060054WA, SQS23060055WA, SQS23060056WA, SQS23060057WA, SQS23060058WA, SQS24100013WA, SQS24100014WA, SQS24100015WA, SQS24110019WA, SQS24110020WA, SQS24110021WA, SQS24110022WA, SQS24110023WA, SQS24110024WA, SQS24110025WA, SQS24110026WA, SQS24110027WA, SQS24110028WA, SQS23410018WA, SQS23410019WA, SQS23410020WA, SQS23410021WA, SQS23410022WA, SQS23410023WA, SQS23410024WA, SQS23410025WA, SQS23450044WA, SQS23450045WA, SQS23450046WA, SQS24110018WA, SQS24100007WA, SQS24100001WA, SQS24100002WA, SQS24100003WA, SQS24100004WA, SQS24100005WA, SQS24100006WA, SQS23050013WA, SQS23050015WA, SQS23050016WA, SQS23330017WA, SQS23180032WA, SQS23180033WA, SQS23180034WA, SQS23180035WA, SQS24110029WA, SQS24110030WA, SQS23440023WA, SQS23440024WA, SQS23440025WA, SQS23440026WA, SQS23440027WA, SQS23440028WA, SQS23440029WA, SQS23440030WA, SQS23440031WA, SQS23440032WA, SQS23440033WA, SQS23440034WA, SQS23440035WA, SQS23440036WA, SQS23440037WA, SQS23440038WA, SQS23440039WA, SQS23440040WA, SQS23440041WA, SQS24110031WA, SQS24110032WA, SQS24110033WA, SQS24120028WA, SQS24120029WA, SQS24120030WA, SQS24120031WA, SQS24120032WA, SQS23030004WA, SQS23030005WA, SQS23030006WA, SQS23030007WA, SQS23030008WA, SQS23030009WA, SQS23030010WA, SQS23450024WA, SQS23450025WA, SQS23450026WA, SQS23450027WA, SQS23450028WA, SQS23450029WA, SQS23450030WA, SQS23450031WA, SQS23450032WA, SQS23450033WA, SQS23450034WA, SQS23450035WA, SQS23450036WA, SQS23450037WA, SQS23450038WA, SQS23450039WA, SQS23450040WA, SQS23450041WA, SQS23450042WA, SQS23450043WA, SQS24100008WA, SQS23420007WA, SQS23420030WA, SQS23420031WA, SQS23420032WA, SQS23430012MA, SQS23430013MA, SQS24440001MA, SQS24250002MA.
Recalling Firm/
Manufacturer
GE Medical Systems China Co., Ltd.
Dev. Zone
National Hi-Tech; No. 19 Changjiang Road; Xin
Wuxi China
For Additional Information ContactGE HealthCare Service
800-437-1171
Manufacturer Reason
for Recall
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
FDA Determined
Cause 2
Device Design
ActionGE Healthcare issued an URGENT MEDICAL DEVICE notice to its consignees on 03/21/2025 distributed via traceable means. The notice explained the safety issue with the device, potential risk, and requested the following actions be taken: The anesthesia system may continue to be used following the instructions below: Perform the Ventilation Screening Test (specified in Appendix 1) for each affected Carestation system. If the Carestation system passes the Ventilation Screening Test, you can continue to use the device in accordance with the instructions in the User Reference Manual (URM). If the Carestation system fails the Ventilation Screening Test and must be used prior to the system being corrected by GE HealthCare, follow the instructions below: 1. Use the device in accordance with the instructions in the User Reference Manual (URM) with these changes: i. Use only Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) mode to mechanically ventilate a patient. ii. Do not use Volume Control Ventilation (VCV) mode to mechanically ventilate a patient. NOTE: Manual mode of the anesthesia system can be used to provide manual ventilation or allow spontaneous ventilation of the patient. 2. Ensure users are made aware not to use Volume Control Ventilation (VCV) mode on the device until the device has been corrected by GE HealthCare. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions, and retain this document for your record. GE HealthCare will correct all devices that fail the Ventilation Screening Test at no cost. A GE HealthCare representative will contact you to arrange for the correction. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. Notice sent on 04/21/25 with updated ventilation screening test instructions.
Quantity in Commerce4339 units
DistributionWorldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = BSZ
-
-