| | Class 2 Device Recall Salem |  |
| Date Initiated by Firm | September 18, 2025 |
| Date Posted | October 17, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0207-2026 |
| Recall Event ID |
97539 |
| 510(K)Number | K810156 |
| Product Classification |
Tubes, gastrointestinal (and accessories) - Product Code KNT
|
| Product | Salem Sump PVC Tubes:
Product Code/Product Name:
264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm)
264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm)
264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm)
264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm)
5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm)
5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm)
5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm)
5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm)
8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm)
8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm)
8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm) |
| Code Information |
Product Code/Product Name/UDI-DI:
264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 10884521067196
264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 10884521067202
264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 10884521067219
264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 10884521067226
5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 50192253008171
5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 50192253008195
5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 50192253008201
5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm) 50192253008218
8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm) 10192253012453
8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm) 10192253012972
8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm) 10192253012996
All Lots |
Recalling Firm/ Manufacturer |
Cardinal Health 200, LLC 3651 Birchwood Dr Waukegan IL 60085-8337
|
| For Additional Information Contact | Customer Service 800-292-9332 |
Manufacturer Reason for Recall | Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214. |
FDA Determined Cause 2 | Use error |
| Action | On September 18, 2025, the firm notified customers via Urgent Medical Device Product Correction Update letters. This is an update to the notice sent in July 2024 for RES 95214. The present recall includes all lots of additional product codes listed.
The letter provides customers with additional guidance on how to properly use the Anti-Reflux Valve (ARV):
Important Instructions:
After placement of the Salem Sump tube, always inject 10ml-20ml of air before seating the Anti Reflux Valve (ARV) into the blue sump vent lumen.
Instructions:
1. Seat the BLUE END of the ARV into the blue sump vent lumen of the Salem Sump tube. Insert the ARV gently into the Salem Sump tube to avoid difficult removal.
2. To cap tube, seat the WHITE END of the ARV into the suction lumen. Insert the ARV gently into the Salem Sump tube to avoid difficult removal.
3. To remove, pull valve in the same direction of assembly. Do not pull at an angle to avoid breaking the valve.
Please visit https://www.mycardinalmsds.com for updated electronic IFUs. |
| Quantity in Commerce | all lots |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|