| | Class 3 Device Recall HeartSine SAM 350P, HeartSine SAM 360P and HeartSine SAM 450P |  |
| Date Initiated by Firm | July 27, 2026 |
| Date Posted | September 25, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3270-2026 |
| Recall Event ID |
98483 |
| PMA Number | P160008 |
| Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
| Product | HeartSine samaritan PAD (Public Access Defibrillator);
HeartSine SAM 350P
350P US Device, US English - 2 Min CPR
UDI-DI codes: 0506016712067
350P US Device, US English (American Airlines) - 2 Min CPR
UDI-DI code: 0506016712207
350P US Device, US English with Gateway - 2 Min CPR
UDI-DI code: 0506016712804
HeartSine SAM 360P
360P US Device, US English - 2 Min CPR
UDI-DI code: 0506016712068
HeartSine SAM 450P
SAM 450P US Device, US English (2 min CPR)
UDI-DI code: 0506016712069 |
| Code Information |
Product - 350P US Device, US English - 2 Min CPR
Item code: 350-STR-US-10
UDI-DI code: 0506016712067
Serial Numbers:
24D98056051D24
24D98057467D24
24D98057468D24
24D98057469D24
24D98068091E24
24D98068097E24
24D98070022F24
24D98070023F24
24D98070024F24
24D98070025F24
24D98070026F24
24D98070027F24
24D98070028F24
24D98070029F24
24D98070030F24
24D98070031F24
24D98070032F24
24D98070033F24
24D98070034F24
24D98070035F24
24D98070036F24
24D98070037F24
24D98070038F24
24D98070040F24
24D98070041F24
24D98070042F24
24D98070043F24
24D98070046F24
24D98070047F24
24D98070048F24
24D98070049F24
24D98070050F24
24D98070051F24
24D98070052F24
24D98070053F24
24D98070054F24
24D98070055F24
24D98070056F24
24D98070057F24
24D98070058F24
24D98070060F24
24D98070061F24
24D98070062F24
24D98070063F24
24D98070064F24
24D98070065F24
24D98070066F24
24D98070068F24
24D98070069F24
24D98070071F24
24D98070072F24
24D98070074F24
24D98070075F24
24D98070076F24
24D98070077F24
24D98070078F24
24D98070079F24
24D98070080F24
24D98070081F24
24D98070082F24
24D98070083F24
24D98070084F24
24D98070085F24
24D98070087F24
24D98070088F24
24D98070089F24
24D98070091F24
24D98070092F24
24D98070093F24
24D98070094F24
24D98070095F24
24D98070096F24
24D98070097F24
24D98070100F24
24D98070102F24
24D98070103F24
24D98070104F24
24D98070107F24
24D98070108F24
24D98070109F24
24D98070110F24
24D98070111F24
24D98070114F24
24D98070115F24
24D98070117F24
24D98070118F24
24D98070119F24
24D98070120F24
24D98070121F24
24D98070122F24
24D98070123F24
24D98070126F24
24D98070127F24
24D98070128F24
24D98070129F24
24D98070130F24
24D98070132F24
24D98070133F24
24D98070134F24
24D98070135F24
24D98070136F24
24D98070138F24
24D98070139F24
24D98070140F24
24D98070141F24
24D98070143F24
24D98070144F24
24D98070145F24
24D98070148F24
24D98070149F24
24D98070150F24
24D98070151F24
24D98070157F24
24D98070160F24
24D98070161F24
24D98070162F24
24D98070163F24
24D98070166F24
24D98056785D24
24D98057279D24
24D98057648D24
24D98053900E24
24D98054669E24
24D98054543E24
24D98054034E24
24D98050678E24
24D98050536E24
24D98055040E24
24D98055328E24
24D98050945E24
24D98055733E24
24D98070147F24
24D98070039F24
24D98070101F24
24D98070105F24
24D98070146F24
24D98070045F24
24D98070067F24
24D98070116F24
24D98070099F24
24D98070044F24
24D98070098F24
24D98070059F24
24D98070073F24
24D98070090F24
24D98070142F24
24D98070124F24
24D98070113F24
24D98070131F24
24D98070112F24
24D98070125F24
24D98070412F24
24D98026634F24
24D98026619F24
24D98048208F24
24D98026934F24
24D98026936F24
24D98048576F24
24D98048559F24
24D98048502F24
24D98048594F24
24D98048571F24
24D98048600F24
24D98048566F24
24D98048609F24
24D98048588F24
24D98048578F24
24D98048554F24
24D98048499F24
24D98048503F24
24D98048504F24
24D98048574F24
24D98048582F24
24D98048491F24
24D98048556F24
24D98048597F24
24D98048598F24
24D98048570F24
24D98048712F24
24D98048560F24
24D98048500F24
24D98048595F24
24D98048575F24
24D98048709F24
24D98048603F24
24D98048495F24
24D98048617F24
24D98048492F24
24D98048497F24
24D98048558F24
24D98048605F24
24D98048710F24
24D98048565F24
24D98048586F24
24D98048604F24
24D98048561F24
24D98048606F24
24D98048589F24
24D98048555F24
24D98048564F24
24D98048614F24
24D98048599F24
24D98048590F24
24D98048591F24
24D98048583F24
24D98048601F24
24D98048580F24
24D98048573F24
24D98048618F24
24D98048577F24
24D98048567F24
24D98048501F24
24D98048610F24
24D98048607F24
24D98048711F24
24D98048557F24
24D98048616F24
24D98048585F24
24D98048572F24
24D98048608F24
24D98048592F24
24D98048593F24
24D98048569F24
24D98048619F24
24D98048579F24
24D98048602F24
24D98048498F24
24D98048496F24
24D98048581F24
24D98048587F24
24D98048562F24
24D98048596F24
24D98048611F24
24D98048568F24
24D98048615F24
24D98048584F24
24D98048563F24
24D98026922F24
24D98026921F24
24D98048625F24
24D98026935F24
24D98026838F24
24D98026836F24
24D98026834F24
24D98026928F24
24D98026868F24
24D98026869F24
24D98026933F24
24D98048388F24
24D98028430G24
24D98062685G24
24D98071182G24
24D98071082G24
24D98038091G24
24D98071057G24
24D98038146G24
24D98038133G24
24D98074601G24
24D98074572G24
24D98074570G24
24D98074606G24
24D98074541G24
24D98024371H24
24D98024818H24
24D98024808H24
24D98002300H24
24D98002281H24
24D98024650H24
24D98019809H24
24D98019376H24
24D98040257G24
24D98040243G24
24D98040260G24
24D98040262G24
24D98040258G24
24D98040256G24
24D98040263G24
24D98040261G24
24D98001851H24
24D98038364G24
24D98071566G24
24D98071559G24
24D98071562G24
24D98038340G24
24D98038333G24
24D98038334G24
24D98038326G24
24D98071497G24
24D98071498G24
24D98071502G24
24D98019747H24
24D98038331G24
24D98038335G24
24D98038330G24
24D98071567G24
24D98071568G24
24D98038319G24
24D98038318G24
24D98038315G24
24D98038311G24
24D98038304G24
24D98038338G24
24D98038322G24
24D98038306G24
24D98038302G24
24D98038309G24
24D98038303G24
24D98038308G24
24D98071563G24
24D98038352G24
24D98038339G24
24D98038321G24
24D98038314G24
24D98038313G24
24D98038320G24
24D98038323G24
24D98071501G24
24D98038387G24
24D98071499G24
24D98071573G24
24D98071571G24
24D98071581G24
24D98071558G24
24D98071560G24
24D98038295G24
24D98071561G24
24D98071565G24
24D98071564G24
24D98071500G24
24D98071495G24
24D98071496G24
24D98071570G24
24D98038139G24
24D98038307G24
24D98038324G24
24D98038317G24
24D98038316G24
24D98038305G24
24D98038312G24
24D98038293G24
24D98038300G24
24D98038301G24
24D98038297G24
24D98038144G24
24D98038145G24
24D98038337G24
24D98038344G24
24D98038343G24
24D98038345G24
24D98038327G24
24D98038342G24
24D98038296G24
24D98038289G24
24D98038294G24
24D98038290G24
24D98038299G24
24D98064983E24
24D98064976E24
24D98064990E24
24D98050963E24
24D98067351E24
24D98067347E24
24D98067346E24
24D98055977E24
24D98055973E24
24D98055974E24
24D98050964E24
24D98067337E24
24D98055964E24
24D98055976E24
24D98060043E24
24D98055967E24
24D98055972E24
24D98055969E24
24D98055978E24
24D98055966E24
24D98055981E24
24D98055968E24
24D98055971E24
24D98055979E24
24D98055982E24
24D98055970E24
24D98050987E24
24D98050978E24
24D98055940E24
24D98050966E24
24D98050967E24
24D98050956E24
24D98050981E24
24D98067336E24
24D98067335E24
24D98050982E24
24D98050977E24
24D98050988E24
24D98064993E24
24D98055950E24
24D98055949E24
24D98055951E24
24D98055942E24
24D98055943E24
24D98055952E24
24D98064996E24
24D98067354E24
24D98067349E24
24D98064998E24
24D98065000E24
24D98055965E24
24D98055941E24
24D98055947E24
24D98055948E24
24D98064999E24
24D98067353E24
24D98064997E24
24D98064995E24
24D98060037E24
24D98055975E24
24D98055963E24
24D98050972E24
24D98038087G24
24D98038102G24
24D98071042G24
24D98038090G24
24D98038100G24
24D98038053G24
24D98038098G24
24D98038095G24
24D98038085G24
24D98038104G24
24D98055781E24
24D98055765E24
24D98055766E24
24D98055754E24
24D98055764E24
24D98055767E24
24D98055763E24
24D98055753E24
24D98055757E24
24D98055762E24
24D98055755E24
24D98055756E24
24D98055882E24
24D98055884E24
24D98055885E24
24D98055990E24
24D98055989E24
24D98055984E24
24D98055985E24
24D98055983E24
24D98060040E24
24D98060038E24
24D98055997E24
24D98055996E24
24D98055998E24
24D98060039E24
24D98055956E24
24D98055958E24
24D98055991E24
24D98050892E24
24D98050973E24
24D98055992E24
24D98040242G24
24D98055955E24
24D98060041E24
24D98055959E24
24D98055953E24
24D98055954E24
24D98055993E24
24D98055995E24
24D98055961E24
24D98055994E24
24D98055960E24
24D98055957E24
24D98055962E24
24D98055986E24
24D98055988E24
24D98055987E24
24D98055891E24
24D98055874E24
24D98055890E24
24D98055877E24
24D98055881E24
24D98055880E24
24D98055875E24
24D98055900E24
24D98055887E24
24D98055889E24
24D98055886E24
24D98050889E24
24D98055888E24
24D98050888E24
24D98055883E24
24D98050877E24
24D98055876E24
24D98067338E24
24D98055878E24
24D98055879E24
24D98071065G24
24D98051941G24
24D98051959G24
24D98051965G24
24D98051963G24
24D98051972G24
24D98051954G24
24D98051976G24
24D98051945G24
24D98051958G24
24D98051973G24
24D98051962G24
24D98051957G24
24D98051964G24
24D98051944G24
24D98051943G24
24D98051956G24
24D98051953G24
24D98051975G24
24D98051960G24
24D98051961G24
24D98051952G24
24D98051955G24
24D98051942G24
24D98051974G24
24D98071039G24
24D98071060G24
24D98061798G24
24D98061800G24
24D98071063G24
24D98071062G24
24D98071052G24
24D98071054G24
24D98061799G24
24D98071050G24
24D98071055G24
24D98071058G24
24D98061801G24
24D98040236G24
24D98040241G24
24D98040235G24
24D98040238G24
24D98040240G24
24D98040239G24
24D98040234G24
24D98040265G24
24D98040226G24
24D98040212G24
24D98040213G24
24D98040228G24
24D98040224G24
24D98040264G24
24D98040225G24
24D98040209G24
24D98040254G24
24D98040249G24
24D98040255G24
24D98040253G24
24D98040251G24
24D98040227G24
24D98040218G24
24D98040215G24
24D98040231G24
24D98040252G24
24D98040223G24
24D98040216G24
24D98040233G24
24D98040217G24
24D98040882G24
24D98040222G24
24D98040219G24
24D98040232G24
24D98040250G24
24D98040214G24
24D98040230G24
24D98040220G24
24D98040221G24
24D98002445H24
24D98040813H24
24D98040809H24
24D98040816H24
24D98040879H24
24D98040878H24
24D98040810H24
24D98093349H24
24D98093145G24
24D98093165G24
24D98093367H24
24D98040266G24
24D98093175G24
24D98093176G24
24D98093178G24
24D98093203H24
24D98093169G24
24D98093206H24
24D98093157G24
24D98093194H24
24D98040267G24
24D98093167G24
24D98093369H24
24D98093204H24
24D98093333H24
24D98093155G24
24D98093357H24
24D98093347H24
24D98093168G24
24D98093351H24
24D98093166G24
24D98093348H24
24D98093170G24
24D98093171G24
24D98093164G24
24D98093336H24
24D98093208H24
24D98093159G24
24D98093163G24
24D98093173G24
24D98093172G24
24D98093334H24
24D98093177G24
24D98093350H24
24D98093331H24
24D98093366H24
24D98093332H24
24D98093205H24
24D98093162G24
24D98093160G24
24D98093188G24
24D98093200H24
24D98093192H24
24D98040821H24
24D98093143G24
24D98093198H24
24D98093193H24
24D98093343H24
24D98093196H24
24D98093210H24
24D98093207H24
24D98093202H24
24D98093346H24
24D98093197H24
24D98093158G24
24D98093146G24
24D98093144G24
24D98093199H24
24D98093345H24
24D98093201H24
24D98093190H24
24D98093195H24
24D98093191H24
24D98093187G24
24D98093209H24
24D98093161G24
24D98093174G24
24D98040823H24
24D98093156G24
24D98093189G24
24D98039964B24
24D98026855F24
24D98039958B24
24D98052243G24
24D98020465H24
24D98093992H24
24D98026895F24
24D98026879F24
24D98029982A24
24D98048955G24
24D98026897F24
24D98068161E24
24D98053692E24
24D98053766E24
24D98050976E24
Product - 350P US Device, US English (American Airlines) - 2 Min CPR
Item code: 350-STR-AA-AV
UDI-DI code: 0506016712207
Serial Numbers:
24D98067253D24
24D98067252D24
24D98067251D24
24D98067249D24
24D98056938D24
24D98067250D24
24D98067248D24
24D98067153D24
24D98067160D24
24D98067155D24
24D98067163D24
24D98067180D24
24D98067182D24
24D98067219D24
24D98067186D24
24D98067220D24
24D98067221D24
24D98067215D24
24D98067156D24
24D98067200D24
24D98067195D24
24D98067242D24
24D98067231D24
24D98067234D24
24D98067244D24
24D98067197D24
24D98067232D24
24D98067194D24
24D98067166D24
24D98067165D24
24D98067161D24
24D98067169D24
24D98067241D24
24D98067162D24
24D98067192D24
24D98067159D24
24D98067198D24
24D98067157D24
24D98067203D24
24D98067191D24
24D98067158D24
24D98067246D24
24D98067225D24
24D98067247D24
24D98067206D24
24D98067173D24
24D98067204D24
24D98067213D24
24D98067209D24
24D98067187D24
24D98067207D24
24D98067178D24
24D98067217D24
24D98067214D24
24D98067222D24
24D98067189D24
24D98067212D24
24D98067171D24
24D98067170D24
24D98067177D24
24D98067223D24
24D98067237D24
24D98067172D24
24D98067245D24
24D98067210D24
24D98067211D24
24D98067205D24
24D98067233D24
24D98067208D24
24D98067174D24
24D98067226D24
24D98067190D24
24D98067164D24
24D98067168D24
24D98067167D24
24D98067175D24
24D98067176D24
24D98067181D24
24D98067185D24
24D98067184D24
24D98067216D24
24D98067188D24
24D98067183D24
24D98067218D24
24D98067179D24
24D98067228D24
24D98067229D24
24D98067202D24
24D98067236D24
24D98067235D24
24D98067201D24
24D98067239D24
24D98067230D24
24D98067243D24
24D98067196D24
24D98067193D24
24D98067240D24
24D98067227D24
24D98067224D24
24D98067144D24
24D98067238D24
24D98067121D24
24D98067131D24
Product - 350P US Device, US English with Gateway - 2 Min CPR
Item code: 350-STR-US-GW
UDI-DI code: 0506016712804
Serial Numbers:
24D98073174E24
24D98033695E24
24D98033694E24
24D98033696E24
24D98033711E24
24D98033688E24
24D98033686E24
Product - 360P US Device, US English - 2 Min CPR
Item code: 360-STR-US-10
UDI-DI code: 0506016712068
Serial Numbers:
24E90017943A24
24E90016508A24
24E90017446A24
24E90017448A24
24E90017447A24
24E90016896A24
24E90017445A24
24E90017444A24
24E90017664A24
24E90017257A24
24E90017482A24
24E90017492A24
24E90017662A24
24E90017479A24
24E90017488A24
24E90017661A24
24E90017480A24
24E90017307A24
24E90017659A24
24E90017490A24
24E90017481A24
24E90017486A24
24E90017665A24
24E90017663A24
24E90017258A24
24E90017308A24
24E90017484A24
24E90017489A24
24E90017656A24
24E90017309A24
24E90017224A24
24E90017226A24
24E90016745A24
24E90016512A24
24E90017143A24
24E90017134A24
24E90017136A24
24E90018007A24
24E90017956A24
24E90017954A24
24E90017955A24
24E90017950A24
24E90017940A24
24E90017948A24
24E90017947A24
24E90017937A24
24E90017941A24
24E90017938A24
24E90017942A24
24E90017939A24
24E90017946A24
24E90017995A24
24E90017990A24
24E90018001A24
24E90017993A24
24E90017978A24
24E90018341A24
24E90017985A24
24E90018327A24
24E90018345A24
24E90018340A24
24E90018342A24
24E90017987A24
24E90017980A24
24E90017986A24
24E90017992A24
24E90017981A24
24E90017977A24
24E90017976A24
24E90018339A24
24E90018326A24
24E90018000A24
24E90018328A24
24E90017314A24
24E90017991A24
24E90017996A24
24E90017994A24
24E90017984A24
24E90017312A24
24E90017322A24
24E90018344A24
24E90018343A24
24E90017989A24
24E90017319A24
24E90017982A24
24E90017998A24
24E90017999A24
24E90017979A24
24E90017273A24
24E90017264A24
24E90017271A24
24E90017263A24
24E90017262A24
24E90017267A24
24E90017274A24
24E90017259A24
24E90017272A24
24E90017266A24
24E90017265A24
24E90017634A24
24E90003313B24
24E90016891A24
24E90017310A24
24E90018357A24
24E90003897B24
24E90003900B24
24E90003929B24
24E90003958B24
24E90003931B24
24E90003936B24
24E90003917B24
24E90003921B24
24E90003068B24
24E90003067B24
24E90003493B24
24E90003494B24
24E90003306B24
24E90003312B24
24E90003309B24
24E90013330B24
24E90004172B24
24E90005978B24
24E90018657B24
24E90001548B24
24E90014938B24
24E90015411B24
24E90002912B24
24E90001072B24
24E90015420B24
24E90001638B24
24E90000739B24
24E90000958B24
24E90006176B24
24E90006182B24
24E90006179B24
24E90006181B24
24E90004468B24
24E90004449B24
24E90004476B24
24E90004622B24
24E90004471B24
24E90004457B24
24E90004398B24
24E90004629B24
24E90004389B24
24E90004455B24
24E90004450B24
24E90004386B24
24E90004470B24
24E90004474B24
24E90004391B24
24E90004396B24
24E90004448B24
24E90004390B24
24E90004388B24
24E90004621B24
24E90004453B24
24E90004451B24
24E90004397B24
24E90004392B24
24E90004478B24
24E90004473B24
24E90004393B24
24E90004623B24
24E90004452B24
24E90004624B24
24E90004387B24
24E90004477B24
24E90004456B24
24E90004626B24
24E90004472B24
24E90004454B24
24E90004625B24
24E90004469B24
24E90005977B24
24E90004724B24
24E90004755B24
24E90004752B24
24E90004730B24
24E90004749B24
24E90004726B24
24E90004721B24
24E90004750B24
24E90004751B24
24E90004728B24
24E90004722B24
24E90004723B24
24E90004727B24
24E90006184B24
24E90014632B24
24E90002115B24
24E90004710B24
24E90004695B24
24E90004707B24
24E90004716B24
24E90004642B24
24E90004711B24
24E90004640B24
24E90004700B24
24E90004717B24
24E90004733B24
24E90004704B24
24E90004697B24
24E90004699B24
24E90004731B24
24E90004641B24
24E90004698B24
24E90004634B24
24E90004693B24
24E90004696B24
24E90004719B24
24E90004706B24
24E90004631B24
24E90004638B24
24E90004705B24
24E90004636B24
24E90004635B24
24E90004637B24
24E90004735B24
24E90004702B24
24E90004734B24
24E90004639B24
24E90004715B24
24E90004718B24
24E90004694B24
24E90004633B24
24E90004732B24
24E90004701B24
24E90004708B24
24E90004632B24
24E90002370B24
24E90004612B24
24E90004613B24
24E90004616B24
24E90004618B24
24E90004615B24
24E90004617B24
24E90004619B24
24E90006185B24
24E90006177B24
24E90006183B24
24E90006180B24
24E90006178B24
24E90004464B24
24E90004466B24
24E90004462B24
24E90004458B24
24E90005791B24
24E90004399B24
24E90004609B24
24E90003808B24
24E90004607B24
24E90004467B24
24E90004400B24
24E90004461B24
24E90004460B24
24E90003835B24
24E90005794B24
24E90004459B24
24E90004531B24
24E90005784B24
24E90005792B24
24E90003807B24
24E90004605B24
24E90004530B24
24E90004599B24
24E90004532B24
24E90005783B24
24E90004463B24
24E90004602B24
24E90004598B24
24E90003809B24
24E90005788B24
24E90005793B24
24E90004600B24
24E90004601B24
24E90004594B24
24E90004593B24
24E90005790B24
24E90003818B24
24E90005781B24
24E90005785B24
24E90004465B24
24E90004610B24
24E90006156B24
24E90005890B24
24E90005906B24
24E90005897B24
24E90006153B24
24E90005892B24
24E90005896B24
24E90005898B24
24E90005902B24
24E90005913B24
24E90006154B24
24E90005894B24
24E90006155B24
24E90005899B24
24E90005910B24
24E90005915B24
24E90005901B24
24E90005893B24
24E90005900B24
24E90005891B24
24E90005904B24
24E90005905B24
24E90005895B24
24E90005903B24
24E90005907B24
24E90005795B24
24E90005797B24
24E90006150B24
24E90005883B24
24E90005880B24
24E90006145B24
24E90006151B24
24E90005881B24
24E90006143B24
24E90005889B24
24E90006198B24
24E90006197B24
24E90006193B24
24E90006200B24
24E90006148B24
24E90006152B24
24E90005885B24
24E90006149B24
24E90006188B24
24E90005886B24
24E90005888B24
24E90005882B24
24E90005884B24
24E90006146B24
24E90006190B24
24E90006144B24
24E90014657B24
24E90006158B24
24E90006168B24
24E90006157B24
24E90006173B24
24E90006174B24
24E90006160B24
24E90006159B24
24E90006175B24
24E90006169B24
24E90006165B24
24E90006186B24
24E90006172B24
24E90006162B24
24E90006161B24
24E90006164B24
24E90006192B24
24E90006167B24
24E90006163B24
24E90015664B24
23E90139374B24
24E90012588A24
24E90005976B24
24E90017923B24
24E90015530B24
24E90001785B24
24E90015665B24
24E90004170B24
24E90001501B24
24E90001288B24
24E90005937B24
24E90005925B24
24E90015201A24
24E90017790A24
24E90017788A24
24E90017789A24
24E90017787A24
24E90017795A24
24E90018540A24
24E90017791A24
24E90018538A24
24E90018542A24
24E90017793A24
24E90018536A24
24E90018543A24
24E90017867A24
24E90018535A24
24E90017798A24
24E90017792A24
24E90017794A24
24E90018541A24
24E90018534A24
24E90018533A24
24E90018530A24
24E90015382A24
24E90018529A24
24E90017888A24
24E90015383A24
24E90017785A24
24E90017889A24
24E90000689A24
24E90017892A24
24E90017782A24
24E90018532A24
24E90018527A24
24E90018526A24
24E90017361A24
24E90017362A24
24E90017360A24
24E90017352A24
24E90017366A24
24E90017359A24
24E90017363A24
24E90017354A24
24E90017353A24
24E90017367A24
24E90017365A24
24E90017343A24
24E90017364A24
24E90017355A24
24E90017234A24
24E90017085A24
24E90017123A24
24E90017127A24
24E90016474A24
24E90017461A24
24E90017374A24
24E90017080A24
24E90017478A24
24E90017121A24
24E90016475A24
24E90017078A24
24E90017462A24
24E90017456A24
24E90017476A24
24E90017473A24
24E90016698A24
24E90017378A24
24E90016903A24
24E90017372A24
24E90017472A24
24E90017467A24
24E90017231A24
24E90017126A24
24E90017457A24
24E90017235A24
24E90017079A24
24E90017368A24
24E90017081A24
24E90017124A24
24E90016457A24
24E90016467A24
24E90017465A24
24E90017460A24
24E90017120A24
24E90017466A24
24E90017455A24
24E90017458A24
24E90017371A24
24E90017230A24
24E90016894A24
24E90017237A24
24E90017370A24
24E90017377A24
24E90017083A24
24E90017236A24
24E90017125A24
24E90016460A24
24E90016893A24
24E90017376A24
24E90017087A24
24E90017477A24
24E90017122A24
24E90017471A24
24E90017128A24
24E90017463A24
24E90017375A24
24E90017119A24
24E90017475A24
24E90016476A24
24E90016471A24
24E90017459A24
24E90017238A24
24E90017369A24
24E90017373A24
24E90017233A24
24E90017232A24
24E90016477A24
24E90017452A24
24E90017464A24
24E90017082A24
24E90017086A24
24E90018004A24
24E90018336A24
24E90018286A24
24E90018287A24
24E90014848A24
24E90018334A24
24E90018338A24
24E90018302A24
24E90018277A24
24E90017299A24
24E90017507A24
24E90017301A24
24E90017494A24
24E90017505A24
24E90017324A24
24E90017512A24
24E90017325A24
24E90017252A24
24E90017524A24
24E90017255A24
24E90017501A24
24E90017328A24
24E90017517A24
24E90017323A24
24E90017485A24
24E90017250A24
24E90017518A24
24E90017563A24
24E90017326A24
24E90017503A24
24E90017297A24
24E90017508A24
24E90017493A24
24E90017253A24
24E90017294A24
24E90017541A24
24E90017510A24
24E90017515A24
24E90017498A24
24E90017513A24
24E90017509A24
24E90017521A24
24E90017526A24
24E90017511A24
24E90017254A24
24E90017500A24
24E90017251A24
24E90017499A24
24E90017293A24
24E90017495A24
24E90017523A24
24E90017300A24
24E90017292A24
24E90017519A24
24E90017516A24
24E90017504A24
24E90017295A24
24E90017522A24
24E90017256A24
24E90017525A24
24E90017520A24
24E90017298A24
24E90017514A24
24E90017502A24
24E90017497A24
24E90017296A24
24E90017506A24
24E90016454A24
24E90016439A24
24E90017110A24
24E90017222A24
24E90017116A24
24E90017225A24
24E90017109A24
24E90017227A24
24E90017115A24
24E90017117A24
24E90017228A24
24E90017229A24
24E90016443A24
24E90017220A24
24E90017113A24
24E90017221A24
24E90017118A24
24E90017223A24
24E90017114A24
24E90016747A24
24E90017111A24
24E90016989A24
24E90016990A24
24E90016991A24
24E90012680A24
24E90016978A24
24E90016981A24
24E90016979A24
24E90016988A24
24E90016980A24
24E90016837A24
24E90016840A24
24E90016842A24
24E90016805A24
24E90016810A24
24E90016721A24
24E90016571A24
24E90016703A24
24E90017075A24
24E90016570A24
24E90018424A24
24E90017896A24
24E90017893A24
24E90018429A24
24E90018427A24
24E90018442A24
24E90017815A24
24E90017408A24
24E90018425A24
24E90018431A24
24E90018428A24
24E90017894A24
24E90018443A24
24E90017405A24
24E90017812A24
24E90017811A24
24E90017131A24
24E90017130A24
24E90017621A24
24E90017138A24
24E90018300A24
24E90017141A24
24E90017132A24
24E90017142A24
24E90017624A24
24E90018281A24
24E90017242A24
24E90017133A24
24E90018284A24
24E90017241A24
24E90017248A24
24E90018282A24
24E90017619A24
24E90018285A24
24E90017618A24
24E90018337A24
24E90017244A24
24E90017239A24
24E90017623A24
24E90017140A24
24E90018283A24
24E90017243A24
24E90017625A24
24E90017139A24
24E90017622A24
24E90017620A24
24E90018279A24
24E90017246A24
24E90017245A24
24E90017617A24
24E90017129A24
24E90017247A24
24E90017240A24
24E90018426A24
24E90018487A24
24E90018488A24
24E90018523A24
24E90018420A24
24E90013066A24
24E90013060A24
24E90013067A24
24E90017402A24
24E90017406A24
24E90013058A24
24E90017403A24
24E90013064A24
24E90013063A24
24E90013062A24
24E90017404A24
24E90017707A24
24E90013059A24
24E90018565A24
24E90018559A24
24E90018560A24
24E90018571A24
24E90018013A24
24E90018547A24
24E90017096A24
24E90018568A24
24E90018566A24
24E90018554A24
24E90018567A24
24E90018005A24
24E90018561A24
24E90016862A24
24E90018569A24
24E90018564A24
24E90018006A24
24E90018570A24
24E90018563A24
24E90018555A24
24E90018552A24
24E90018572A24
24E90018574A24
24E90018573A24
24E90018576A24
24E90018580A24
24E90018575A24
24E90018577A24
24E90018581A24
24E90018587A24
24E90018584A24
24E90018583A24
24E90018551A24
24E90018413A24
24E90018591A24
24E90018590A24
24E90018546A24
24E90018335A24
24E90018588A24
24E90018589A24
24E90018298A24
24E90018410A24
24E90016585A24
24E90018585A24
24E90018548A24
24E90018586A24
24E90018433A24
24E90018441A24
24E90018432A24
24E90018440A24
24E90018435A24
24E90018434A24
24E90018436A24
24E90018439A24
24E90018322A24
24E90018017A24
24E90017318A24
24E90018307A24
24E90018317A24
24E90018309A24
24E90018305A24
24E90018306A24
24E90018008A24
24E90018274A24
24E90018009A24
24E90018280A24
24E90018010A24
24E90018295A24
24E90018296A24
24E90017288A24
24E90018331A24
24E90018315A24
24E90018022A24
24E90017283A24
24E90017289A24
24E90018288A24
24E90018012A24
24E90018316A24
24E90018320A24
24E90018351A24
24E90018224A24
24E90017282A24
24E90018016A24
24E90017286A24
24E90018329A24
24E90017302A24
24E90018311A24
24E90018550A24
24E90018323A24
24E90017304A24
24E90018304A24
24E90018310A24
24E90018332A24
24E90018330A24
24E90018321A24
24E90018318A24
24E90018289A24
24E90018319A24
24E90018314A24
24E90018325A24
24E90018308A24
24E90018018A24
24E90018011A24
24E90018312A24
24E90018290A24
24E90017305A24
24E90017281A24
24E90018294A24
24E90017285A24
24E90017303A24
24E90017290A24
24E90018292A24
24E90017291A24
24E90018275A24
24E90018333A24
24E90018324A24
24E90017279A24
24E90018023A24
24E90018293A24
24E90018291A24
24E90018014A24
24E90018015A24
24E90017280A24
24E90018303A24
24E90018297A24
24E90000677A24
24E90000682A24
24E90000664A24
24E90000690A24
24E90000692A24
24E90000683A24
24E90000678A24
24E90015394A24
24E90015011A24
24E90003106B24
24E90003103B24
24E90003129B24
24E90015386A24
24E90015391A24
24E90003105B24
24E90003127B24
24E90015389A24
24E90003565B24
24E90014964A24
24E90015236A24
24E90015229A24
24E90015231A24
24E90014963A24
24E90015219A24
24E90014962A24
24E90015215A24
24E90014961A24
24E90014622A24
24E90015233A24
24E90015217A24
24E90015235A24
24E90015225A24
24E90015232A24
24E90015222A24
24E90015230A24
24E90015221A24
24E90015226A24
24E90015224A24
24E90014965A24
24E90015480A24
24E90015287A24
24E90015288A24
24E90015507A24
24E90000393A24
24E90015289A24
24E90015290A24
24E90015284A24
24E90015285A24
24E90015286A24
24E90015282A24
24E90015283A24
24E90000394A24
24E90000391A24
24E90015468A24
24E90000593A24
24E90000395A24
24E90015281A24
24E90000392A24
24E90003123B24
24E90000302A24
24E90003125B24
24E90003122B24
24E90000301A24
24E90003126B24
24E90003131B24
24E90003130B24
24E90015373A24
24E90000598A24
24E90000621A24
24E90003661B24
24E90000451A24
24E90015555A24
24E90015492A24
24E90000616A24
24E90003102B24
24E90000614A24
24E90000390A24
24E90003111B24
24E90003658B24
24E90000458A24
24E90003660B24
24E90000557A24
24E90015493A24
24E90000457A24
24E90015496A24
24E90015471A24
24E90003110B24
24E90015498A24
24E90015474A24
24E90000606A24
24E90003663B24
24E90000456A24
24E90000459A24
24E90000605A24
24E90015497A24
24E90015475A24
24E90015500A24
24E90000615A24
24E90000601A24
24E90015491A24
24E90015556A24
24E90000622A24
24E90015501A24
24E90000453A24
24E90000620A24
24E90015535A24
24E90003132B24
24E90000452A24
24E90003657B24
24E90000602A24
24E90000617A24
24E90003659B24
24E90000603A24
24E90000604A24
24E90015490A24
24E90015536A24
24E90015499A24
24E90015532A24
24E90015502A24
24E90000600A24
24E90015296A24
24E90015310A24
24E90015291A24
24E90015302A24
24E90014780A24
24E90014765A24
24E90014768A24
24E90014791A24
24E90014784A24
24E90014783A24
24E90014767A24
24E90014769A24
24E90014764A24
24E90014770A24
24E90014785A24
24E90014789A24
24E90003119B24
24E90003116B24
24E90003117B24
24E90003120B24
24E90003112B24
24E90003115B24
24E90002151B24
24E90003121B24
24E90003118B24
24E90002150B24
24E90014800A24
24E90014829A24
24E90014804A24
24E90014797A24
24E90014816A24
24E90014803A24
24E90014777A24
24E90014793A24
24E90014814A24
24E90014801A24
24E90014794A24
24E90014802A24
24E90014792A24
24E90014796A24
24E90014831A24
24E90014833A24
24E90014836A24
24E90014832A24
24E90000627A24
24E90000629A24
24E90014835A24
24E90014837A24
24E90000628A24
24E90000632A24
24E90000633A24
24E90014820A24
24E90000634A24
24E90014819A24
24E90014834A24
24E90000630A24
24E90000631A24
24E90014818A24
24E90014823A24
24E90015561A24
24E90015571A24
24E90015559A24
24E90000277A24
24E90000513A24
24E90014773A24
24E90000279A24
24E90015545A24
24E90000282A24
24E90015562A24
24E90000283A24
24E90000378A24
24E90014771A24
24E90015546A24
24E90000280A24
24E90000381A24
24E90015560A24
24E90000368A24
24E90000284A24
24E90000281A24
24E90015558A24
24E90014772A24
24E90000296A24
24E90015573A24
24E90015563A24
24E90015570A24
24E90015572A24
24E90015544A24
24E90015557A24
24E90000508A24
24E90000509A24
24E90000377A24
24E90000374A24
24E90003092B24
24E90003098B24
24E90003096B24
24E90003099B24
24E90003100B24
24E90003093B24
24E90002328B24
24E90002149B24
24E90002333B24
24E90001962B24
24E90002143B24
24E90003077B24
24E90002332B24
24E90001963B24
24E90002330B24
24E90002324B24
24E90002329B24
24E90003555B24
24E90001964B24
24E90002148B24
24E90001965B24
24E90002142B24
24E90002326B24
24E90002141B24
24E90001961B24
24E90002331B24
24E90003078B24
24E90002327B24
24E90003681B24
24E90003684B24
24E90003665B24
24E90003682B24
24E90003680B24
24E90003672B24
24E90003671B24
24E90004131B24
24E90001997B24
24E90003769B24
24E90003791B24
24E90004108B24
24E90003792B24
24E90003762B24
24E90004127B24
24E90003763B24
24E90004130B24
24E90004105B24
24E90003785B24
24E90004110B24
24E90003996B24
24E90003786B24
24E90003771B24
24E90004107B24
24E90003788B24
24E90003794B24
24E90003787B24
24E90004091B24
24E90003999B24
24E90004160B24
24E90003998B24
24E90003765B24
24E90004093B24
24E90003774B24
24E90004109B24
24E90004120B24
24E90003994B24
24E90004129B24
24E90004092B24
24E90003773B24
24E90003761B24
24E90003764B24
24E90004106B24
24E90003760B24
24E90003766B24
24E90003768B24
24E90003789B24
24E90003759B24
24E90003995B24
24E90004104B24
24E90003770B24
24E90004089B24
24E90003758B24
24E90003772B24
24E90004159B24
24E90003767B24
24E90001974B24
24E90001975B24
24E90001968B24
24E90001969B24
24E90001967B24
24E90001970B24
24E90001977B24
24E90001972B24
24E90004098B24
24E90004101B24
24E90004097B24
24E90004102B24
24E90004099B24
24E90004094B24
24E90004095B24
24E90004100B24
24E90004017B24
24E90003090B24
24E90003745B24
24E90002337B24
24E90002336B24
24E90002338B24
24E90003740B24
24E90003085B24
24E90003091B24
24E90003084B24
24E90003086B24
24E90003088B24
24E90002341B24
24E90002335B24
24E90003087B24
24E90002334B24
24E90002340B24
24E90003083B24
24E90002339B24
24E90003739B24
24E90002169B24
24E90002167B24
24E90002164B24
24E90002166B24
24E90001987B24
24E90001978B24
24E90001985B24
24E90002168B24
24E90001986B24
24E90004168B24
24E90002170B24
24E90002173B24
24E90003988B24
24E90004164B24
24E90004169B24
24E90003989B24
24E90004165B24
24E90002172B24
24E90003987B24
24E90002315B24
24E90003747B24
24E90003775B24
24E90004003B24
24E90002346B24
24E90003781B24
24E90003778B24
24E90003776B24
24E90003784B24
24E90004006B24
24E90004007B24
24E90002307B24
24E90002305B24
24E90002342B24
24E90002309B24
24E90003779B24
24E90001992B24
24E90002311B24
24E90004009B24
24E90002310B24
24E90002301B24
24E90002321B24
24E90002323B24
24E90003777B24
24E90002312B24
24E90004005B24
24E90004000B24
24E90002348B24
24E90004002B24
24E90001988B24
24E90004010B24
24E90004015B24
24E90002304B24
24E90004001B24
24E90002322B24
24E90003780B24
24E90004012B24
24E90002303B24
24E90001989B24
24E90004013B24
24E90002345B24
24E90001996B24
24E90003782B24
24E90003783B24
24E90002344B24
24E90002316B24
24E90004016B24
24E90002302B24
24E90004011B24
24E90003757B24
24E90003738B24
24E90002343B24
24E90001993B24
24E90003743B24
24E90001995B24
24E90003744B24
24E90002306B24
24E90002320B24
24E90002318B24
24E90004008B24
24E90004004B24
24E90002308B24
24E90003742B24
24E90003223B24
24E90003217B24
24E90003216B24
24E90003226B24
24E90003215B24
24E90003317B24
24E90003231B24
24E90003232B24
24E90003230B24
24E90003314B24
24E90003229B24
24E90003227B24
24E90003506B24
24E90003225B24
24E90003228B24
24E90015902A24
24E90012610A24
23E90144949A24
24E90013313A24
24E90013327A24
24E90013309A24
24E90017874A24
24E90017870A24
24E90017884A24
24E90017885A24
24E90017887A24
24E90017883A24
24E90017879A24
24E90017886A24
24E90017872A24
24E90017871A24
24E90017869A24
24E90018481A24
24E90018482A24
24E90018483A24
24E90018484A24
24E90017066A24
24E90017057A24
24E90016241A24
24E90017112A24
24E90017211A24
24E90017160A24
24E90015986A24
24E90017320A24
24E90017533A24
24E90017334A24
24E90018091A24
24E90017277A24
24E90017534A24
24E90017333A24
24E90017329A24
24E90017335A24
24E90017331A24
24E90017330A24
24E90017332A24
24E90017261A24
24E90017275A24
24E90017268A24
24E90017260A24
24E90017269A24
24E90017199A24
24E90017198A24
24E90017097A24
24E90017091A24
24E90017206A24
24E90017196A24
24E90017092A24
24E90017207A24
24E90017181A24
24E90017214A24
24E90017182A24
24E90017183A24
24E90017215A24
24E90017210A24
24E90017099A24
24E90017213A24
24E90017098A24
24E90017216A24
24E90017218A24
24E90017217A24
24E90017219A24
24E90017106A24
24E90017100A24
24E90017186A24
24E90017187A24
24E90017101A24
24E90017105A24
24E90017629A24
24E90017535A24
24E90017635A24
24E90017537A24
24E90017536A24
24E90017532A24
24E90017529A24
24E90017528A24
24E90017628A24
24E90017627A24
24E90017633A24
24E90017179A24
24E90017539A24
24E90017626A24
24E90017527A24
24E90017632A24
24E90017205A24
24E90017090A24
24E90017192A24
24E90017191A24
24E90017209A24
24E90017095A24
24E90017193A24
24E90017208A24
24E90017094A24
24E90017197A24
24E90017201A24
24E90017204A24
24E90017631A24
24E90017538A24
24E90017630A24
24E90017089A24
24E90017195A24
24E90017088A24
24E90017194A24
24E90017202A24
24E90017203A24
24E90018658A24
24E90017107A24
24E90017212A24
24E90017184A24
24E90017185A24
24E90017180A24
24E90017188A24
24E90017103A24
24E90017189A24
24E90017200A24
24E90015279A24
24E90015275A24
24E90015277A24
24E90001278A24
24E90001281A24
24E90001283A24
24E90001284A24
24E90001256A24
24E90000644A24
24E90000645A24
24E90015280A24
24E90001280A24
24E90000648A24
24E90000646A24
24E90001253A24
24E90000639A24
24E90000640A24
24E90000636A24
24E90000635A24
24E90001287A24
24E90000642A24
24E90001285A24
24E90000643A24
24E90001286A24
24E90001381A24
24E90000641A24
24E90001251A24
24E90000638A24
24E90000637A24
24E90001254A24
24E90001289A24
24E90001269A24
24E90001267A24
23E90116066K23
24E90012533A24
24E90012531A24
24E90012532A24
24E90015379A24
24E90015378A24
24E90015385A24
24E90015384A24
24E90015376A24
24E90015380A24
24E90015381A24
24E90012523A24
24E90012525A24
24E90012526A24
24E90012528A24
24E90012527A24
24E90016374A24
24E90016772A24
24E90016367A24
24E90016372A24
24E90016386A24
24E90016767A24
24E90016769A24
24E90016373A24
24E90016385A24
24E90016370A24
24E90016768A24
24E90016770A24
24E90016368A24
24E90016773A24
24E90016766A24
24E90016366A24
24E90016371A24
24E90016384A24
24E90016383A24
24E90016379A24
24E90016380A24
24E90016387A24
24E90016381A24
24E90016376A24
24E90016382A24
24E90012673A24
24E90012669A24
24E90012671A24
24E90012670A24
24E90012674A24
24E90012668A24
24E90012672A24
24E90012675A24
24E90016362A24
24E90016364A24
24E90012682A24
24E90012683A24
24E90016774A24
24E90012684A24
24E90012679A24
24E90016365A24
24E90012681A24
24E90012678A24
24E90016351A24
24E90012677A24
24E90016348A24
24E90016349A24
24E90012676A24
24E90016347A24
24E90016350A24
24E90016342A24
24E90015371A24
24E90015427A24
24E90015431A24
24E90015400A24
24E90015404A24
24E90015396A24
24E90015365A24
24E90015398A24
24E90001064A24
24E90015399A24
24E90015402A24
24E90015403A24
24E90015397A24
24E90015395A24
24E90000547A24
24E90015318A24
24E90015388A24
24E90000545A24
24E90000544A24
24E90000543A24
24E90000541A24
24E90015435A24
24E90015433A24
24E90015387A24
24E90015428A24
24E90015392A24
24E90015393A24
24E90015390A24
24E90015430A24
24E90015429A24
24E90015434A24
24E90015426A24
24E90000574A24
24E90000572A24
24E90000576A24
24E90000571A24
24E90000568A24
24E90000569A24
24E90000573A24
24E90015211A24
24E90015407A24
24E90000575A24
24E90015406A24
24E90000570A24
24E90015409A24
24E90015413A24
24E90015408A24
24E90015415A24
24E90015375A24
24E90015369A24
24E90015367A24
24E90015370A24
24E90015372A24
24E90015368A24
24E90015366A24
24E90000556A24
24E90015551A24
24E90015548A24
24E90015537A24
24E90000578A24
24E90015549A24
24E90015533A24
24E90000554A24
24E90000563A24
24E90000580A24
24E90000553A24
24E90015554A24
24E90000555A24
24E90015547A24
24E90000564A24
24E90015550A24
24E90015531A24
24E90015542A24
24E90000577A24
24E90000581A24
24E90000579A24
24E90000551A24
24E90015541A24
24E90015553A24
24E90015552A24
24E90015538A24
24E90000552A24
24E90015543A24
24E90015540A24
24E90000539A24
24E90015423A24
24E90015419A24
24E90015417A24
24E90000538A24
24E90015416A24
24E90015418A24
24E90000533A24
24E90015451A24
24E90015463A24
24E90015459A24
24E90015465A24
24E90015338A24
24E90015340A24
24E90015334A24
24E90015339A24
24E90015337A24
24E90015342A24
24E90015341A24
24E90003902B24
24E90003263B24
24E90003260B24
24E90015438A24
24E90015439A24
24E90015437A24
24E90015444A24
24E90015440A24
24E90015445A24
24E90015436A24
24E90015441A24
24E90000285A24
24E90003072B24
24E90003954B24
24E90003914B24
24E90003950B24
24E90003947B24
24E90003556B24
24E90003915B24
24E90003935B24
24E90003080B24
24E90003927B24
24E90003930B24
24E90003082B24
24E90003945B24
24E90003905B24
24E90003946B24
24E90003944B24
24E90003942B24
24E90003896B24
24E90003894B24
23E90117536A24
23E90117549A24
23E90117546A24
24E90015886A24
24E90015957A24
24E90015896A24
24E90015838A24
24E90015846A24
24E90015877A24
24E90015880A24
24E90015878A24
24E90015980A24
24E90015884A24
24E90015962A24
24E90015897A24
24E90015898A24
24E90015981A24
24E90015988A24
24E90015841A24
24E90015968A24
24E90015875A24
24E90015983A24
24E90015879A24
24E90015979A24
24E90015978A24
24E90015900A24
24E90015876A24
24E90015994A24
24E90015958A24
24E90015895A24
24E90015984A24
24E90015889A24
24E90015989A24
24E90015967A24
24E90015871A24
24E90015888A24
24E90015882A24
24E90015899A24
24E90015891A24
24E90015892A24
24E90015956A24
24E90015987A24
24E90015985A24
24E90015951A24
24E90015955A24
24E90015982A24
24E90015901A24
24E90015974A24
24E90015972A24
24E90015973A24
24E90015971A24
24E90015977A24
24E90015970A24
24E90015975A24
24E90016360A24
24E90016359A24
24E90016355A24
24E90016358A24
24E90016361A24
24E90016353A24
24E90016352A24
24E90016354A24
24E90017997A24
23E90130321K23
23E90139221K23
24E90003282B24
24E90003288B24
24E90003285B24
24E90003235B24
24E90003299B24
24E90003272B24
24E90003300B24
24E90003290B24
24E90003271B24
24E90003233B24
24E90003294B24
24E90003266B24
24E90003277B24
24E90012592A24
24E90012589A24
24E90012586A24
24E90012561A24
24E90012608A24
24E90012556A24
24E90012606A24
24E90012613A24
24E90012555A24
24E90012577A24
24E90015923A24
24E90003254B24
24E90003256B24
24E90003888B24
24E90003984B24
24E90003964B24
24E90003887B24
24E90003940B24
24E90003941B24
24E90003283B24
24E90003279B24
24E90003203B24
24E90003296B24
24E90003284B24
24E90003278B24
24E90012591A24
24E90012594A24
24E90012590A24
24E90015921A24
24E90012599A24
24E90017658A24
24E90017949A24
24E90017988A24
24E90003073B24
24E90003071B24
24E90018226A24
24E90018225A24
24E90018389A24
24E90018393A24
24E90018391A24
24E90018227A24
24E90017713A24
24E90016944A24
24E90016952A24
24E90016945A24
24E90016951A24
24E90016942A24
24E90016955A24
24E90016948A24
24E90016953A24
24E90017442A24
24E90017725A24
24E90017474A24
24E90017449A24
24E90017439A24
24E90017454A24
24E90016895A24
24E90017440A24
24E90017443A24
24E90016900A24
24E90016897A24
24E90016901A24
24E90016902A24
24E90016904A24
24E90017441A24
24E90017714A24
24E90018169A24
24E90018374A24
24E90018378A24
24E90018167A24
24E90018170A24
24E90018168A24
24E90016700A24
24E90018384A24
24E90018387A24
24E90018383A24
24E90018631A24
24E90018633A24
24E90018544A24
24E90018645A24
24E90018634A24
24E90018623A24
24E90018189A24
24E90018625A24
24E90018134A24
24E90018642A24
24E90018171A24
24E90018172A24
24E90018632A24
24E90018644A24
24E90018130A24
24E90018185A24
24E90018164A24
24E90018186A24
24E90018627A24
24E90018622A24
24E90018630A24
24E90018626A24
24E90018133A24
24E90018156A24
24E90018647A24
24E90018135A24
24E90018646A24
24E90018144A24
24E90018163A24
24E90018639A24
24E90018636A24
24E90018188A24
24E90018641A24
24E90018624A24
24E90018643A24
24E90018628A24
24E90018161A24
24E90018637A24
24E90018162A24
24E90018132A24
24E90018128A24
24E90018640A24
24E90018165A24
24E90018159A24
24E90018194A24
24E90018187A24
24E90018160A24
24E90018129A24
24E90018201A24
24E90018126A24
24E90018629A24
24E90018545A24
24E90018638A24
24E90018192A24
24E90018635A24
24E90017410A24
24E90017418A24
24E90017450A24
24E90017428A24
24E90017421A24
24E90017451A24
24E90017433A24
24E90017760A24
24E90017409A24
24E90017432A24
24E90017424A24
24E90017430A24
24E90017427A24
24E90017429A24
24E90017417A24
24E90017423A24
24E90017431A24
24E90017426A24
24E90017422A24
24E90018556A24
24E90018553A24
24E90018370A24
24E90018562A24
24E90018155A24
24E90018558A24
24E90018557A24
24E90018054A24
24E90018208A24
24E90018049A24
24E90018218A24
24E90018594A24
24E90018609A24
24E90018604A24
24E90017926A24
24E90018607A24
24E90018617A24
24E90017930A24
24E90018062A24
24E90018615A24
24E90018610A24
24E90017925A24
24E90018620A24
24E90018614A24
24E90017927A24
24E90017761A24
24E90018595A24
24E90017931A24
24E90017936A24
24E90017933A24
24E90018600A24
24E90018621A24
24E90017935A24
24E90017762A24
24E90018597A24
24E90018613A24
24E90018602A24
24E90017934A24
24E90018605A24
24E90018606A24
24E90018599A24
24E90018616A24
24E90018619A24
24E90017928A24
24E90018603A24
24E90018596A24
24E90018223A24
24E90018608A24
24E90018618A24
24E90017932A24
24E90018598A24
24E90018612A24
24E90018408A24
24E90018611A24
24E90017764A24
24E90017766A24
24E90018047A24
24E90018352A24
24E90018355A24
24E90018045A24
24E90018041A24
24E90018350A24
24E90018367A24
24E90018354A24
24E90018347A24
24E90018348A24
24E90018358A24
24E90018048A24
24E90018356A24
24E90018042A24
24E90018346A24
24E90018046A24
24E90018368A24
24E90018360A24
24E90018043A24
24E90018039A24
24E90018349A24
24E90018044A24
24E90018549A24
24E90018040A24
24E90018359A24
24E90015823A24
24E90015940A24
24E90015933A24
24E90015936A24
24E90015934A24
24E90015935A24
24E90015937A24
24E90018250A24
24E90018178A24
24E90018183A24
24E90018230A24
24E90018030A24
24E90018031A24
24E90018237A24
24E90018236A24
24E90018033A24
24E90018229A24
24E90018228A24
24E90018037A24
24E90018038A24
24E90018028A24
24E90018034A24
24E90018248A24
24E90018179A24
24E90018184A24
24E90018035A24
24E90018036A24
24E90018251A24
24E90018029A24
24E90018032A24
24E90018233A24
24E90018232A24
24E90017747A24
24E90017739A24
24E90017735A24
24E90017716A24
24E90017715A24
24E90018243A24
24E90018245A24
24E90018244A24
24E90018239A24
24E90018238A24
24E90018241A24
24E90017742A24
24E90017743A24
24E90017741A24
24E90017740A24
24E90017738A24
24E90018242A24
24E90018247A24
24E90018246A24
24E90018240A24
24E90017736A24
24E90017737A24
24E90003251B24
24E90003258B24
24E90003261B24
24E90003259B24
24E90003257B24
24E90003262B24
24E90003255B24
24E90003912B24
24E90003948B24
24E90003076B24
24E90003951B24
24E90003075B24
24E90003074B24
24E90003952B24
24E90003949B24
24E90003923B24
24E90003081B24
24E90003922B24
24E90003069B24
24E90003064B24
24E90003544B24
24E90004153B24
24E90004163B24
24E90004144B24
24E90004090B24
24E90003953B24
24E90003253B24
24E90004166B24
24E90003957B24
24E90003889B24
24E90003890B24
24E90003943B24
24E90003267B24
24E90003286B24
24E90003298B24
24E90003270B24
24E90003291B24
24E90003234B24
24E90003293B24
24E90003194B24
24E90003195B24
24E90003268B24
24E90003269B24
24E90003295B24
24E90003273B24
24E90003264B24
24E90003265B24
24E90003289B24
24E90003275B24
24E90003276B24
24E90003274B24
24E90017306A24
24E90017487A24
24E90017315A24
24E90003911B24
24E90016892A24
24E90017491A24
24E90018366A24
24E90017983A24
24E90017321A24
24E90017313A24
24E90003919B24
24E90008782D24
24E90008892D24
24E90008785D24
24E90008904D24
24E90008873D24
24E90008783D24
24E90008786D24
24E90008885D24
24E90008878D24
24E90008874D24
24E90008784D24
24E90008893D24
24E90008891D24
24E90008884D24
24E90008890D24
24E90008882D24
24E90008905D24
24E90008902D24
24E90008887D24
24E90008875D24
24E90008777D24
24E90008876D24
24E90011868D24
24E90011867D24
24E90011880D24
24E90011863D24
24E90105958H24
24E90105950H24
24E90105716H24
24E90105726H24
24E90105725H24
24E90105728H24
24E90105639H24
24E90105722H24
24E90105723H24
24E90105724H24
24E90008973D24
24E90008965D24
24E90008970D24
24E90008974D24
24E90008972D24
24E90008968D24
24E90008971D24
24E90008958D24
24E90008964D24
24E90008982D24
24E90008961D24
24E90008787D24
24E90009007D24
24E90008975D24
24E90008980D24
24E90008984D24
24E90008960D24
24E90008962D24
24E90008963D24
24E90008959D24
24E90008907D24
24E90008914D24
24E90009009D24
24E90008981D24
24E90008979D24
24E90009010D24
24E90008977D24
24E90008986D24
24E90008978D24
24E90024769D24
24E90024761D24
24E90024357D24
24E90024758D24
24E90023975D24
24E90018737D24
24E90018741D24
24E90018742D24
24E90018747D24
24E90024356D24
24E90024358D24
24E90024363D24
24E90024361D24
24E90021240D24
24E90024360D24
24E90021242D24
24E90021239D24
24E90024757D24
24E90023951D24
24E90021349D24
24E90021241D24
24E90023415C24
24E90023420C24
24E90023417C24
24E90023411C24
24E90006363D24
24E90020999D24
24E90020993D24
24E90020997D24
24E90024767D24
24E90024762D24
24E90024760D24
24E90018739D24
24E90024759D24
24E90024766D24
24E90023986D24
24E90023990D24
24E90023988D24
24E90023995D24
24E90023987D24
24E90023994D24
24E90019739D24
24E90018744D24
24E90018738D24
24E90003928B24
24E90018353A24
24E90018363A24
24E90017316A24
24E90003310B24
24E90018361A24
24E90003983B24
24E90003492B24
24E90003909B24
24E90017945A24
24E90003893B24
24E90018362A24
24E90017657A24
24E90017483A24
24E90018364A24
24E90018369A24
24E90017311A24
Product - SAM 450P US Device, US English (2 min CPR)
Item code: 450-STR-US-10
UDI-DI code: 0506016712069
Serial Numbers:
24G90001624D24
24G90001620D24
24G90001610D24
23G90004172G23
23G90108237L23
23G90108235L23
23G90108212L23
23G90108209L23
23G90108236L23
23G90108213L23
23G90108208L23
23G90108215L23
23G90108210L23
23G90108211L23
23G90108495L23
23G90108226L23
23G90108322L23
23G90108326L23
23G90108222L23
23G90108499L23
23G90108225L23
23G90108497L23
23G90108498L23
23G90108223L23
23G90108492L23
23G90108501L23
23G90108240L23
23G90108227L23
23G90108242L23
23G90108239L23
23G90108684L23
23G90108221L23
23G90108247L23
23G90108479L23
23G90108241L23
23G90108244L23
23G90108246L23
23G90108218L23
23G90108313L23
23G90108219L23
23G90108477L23
23G90108238L23
23G90108473L23
23G90108341L23
23G90108368L23
23G90108345L23
23G90108502L23
23G90108339L23
23G90108365L23
23G90108363L23
23G90108347L23
23G90108344L23
23G90108342L23
23G90108527L23
23G90108524L23
23G90108528L23
23G90108525L23
23G90108513L23
23G90108506L23
23G90108537L23
23G90108531L23
23G90108536L23
23G90108532L23
23G90108538L23
23G90108526L23
23G90108507L23
23G90108523L23
23G90108517L23
23G90108515L23
23G90108533L23
23G90108205L23
23G90108402L23
23G90108203L23
23G90108398L23
23G90108408L23
23G90108199L23
23G90108417L23
23G90108415L23
23G90108409L23
23G90108441L23
23G90108692L23
23G90108437L23
23G90108207L23
23G90108411L23
23G90108399L23
23G90108412L23
23G90108401L23
23G90108403L23
23G90108416L23
23G90108699L23
23G90108440L23
23G90108200L23
23G90108198L23
23G90108404L23
23G90108201L23
23G90108204L23
23G90108655L23
23G90108693L23
23G90108202L23
23G90108428L23
23G90108419L23
23G90108442L23
23G90108410L23
23G90108418L23
23G90108405L23
23G90108706L23
23G90108707L23
23G90108505L23
23G90108709L23
23G90108704L23
23G90108708L23
23G90108705L23
23G90008119L23
23G90108259L23
23G90108256L23
23G90108457L23
23G90107062J23
23G90108274L23
23G90108271L23
23G90009047L23
23G90107662J23
24G90013199H24
24G90013402H24
24G90013387H24
24G90013395H24
24G90013388H24
24G90010488H24
24G90013392H24
24G90013397H24
24G90013386H24
24G90013390H24
24G90013403H24
24G90013391H24
24G90013401H24
24G90013389H24
24G90013195H24
24G90013396H24
24G90013399H24
24G90013404H24
24G90013400H24
24G90013652H24
24G90013201H24
24G90013635H24
24G90013622H24
24G90013617H24
24G90013650H24
24G90013628H24
24G90013620H24
24G90013192H24
24G90013193H24
24G90013627H24
24G90013630H24
24G90013224H24
24G90013223H24
24G90013632H24
24G90013625H24
24G90013636H24
24G90013221H24
24G90013197H24
24G90013629H24
24G90013631H24
24G90013637H24
24G90013633H24
24G90013634H24
24G90013623H24
24G90013626H24
24G90013385H24
24G90013383H24
24G90013382H24
24G90013381H24
24G90013378H24
24G90013384H24
24G90013612H24
24G90013661H24
24G90013648H24
24G90013655H24
24G90013656H24
24G90013665H24
24G90007696D24
24G90007698D24
24G90007706D24
24G90007704D24
24G90007701D24
24G90008495D24
24G90008493D24
24G90008494D24
24G90008496D24
24G90008491D24
24G90017726C24
24G90008499D24
24G90007690C24
24G90012744D24
24G90009451D24
24G90012735D24
24G90009476D24
24G90009446D24
24G90009455D24
24G90009462D24
24G90009449D24
24G90009463D24
24G90009480D24
24G90009461D24
24G90009544D24
24G90009583D24
24G90009441D24
24G90009456D24
24G90009443D24
24G90009580D24
24G90009447D24
24G90009383D24
24G90009457D24
24G90009458D24
24G90009445D24
24G90009577D24
24G90009584D24
24G90009431D24
24G90009468D24
24G90009430D24
24G90009433D24
24G90009434D24
24G90009432D24
24G90009438D24
24G90009437D24
24G90009473D24
24G90009435D24
24G90009465D24
24G90009467D24
24G90009484D24
24G90009439D24
24G90009440D24
24G90009448D24
24G90009481D24
24G90009474D24
24G90009459D24
24G90009472D24
24G90009573D24
24G90009426D24
24G90009471D24
24G90009453D24
24G90009454D24
24G90009477D24
24G90009450D24
24G90009478D24
24G90009387D24
24G90009452D24
24G90008310D24
24G90008303D24
24G90008307D24
24G90008306D24
24G90008313D24
24G90008314D24
24G90008304D24
24G90008301D24
24G90008302D24
24G90008276D24
24G90008316D24
24G90008318D24
24G90008251D24
24G90008337D24
24G90008311D24
24G90012963H24
24G90001550D24
24G90001553D24
24G90001558D24
24G90001605D24
24G90001614D24
24G90001611D24
24G90001617D24
24G90013288H24
24G90013250H24
24G90013294H24
24G90013287H24
24G90012927H24
24G90013218H24
24G90012916H24
24G90012899H24
24G90012936H24
24G90012935H24
24G90012929H24
24G90012890H24
24G90012934H24
24G90012872H24
24G90012925H24
24G90013209H24
24G90012895H24
24G90012915H24
24G90012930H24
24G90012914H24
24G90013370H24
24G90013325H24
24G90013366H24
24G90013216H24
24G90013369H24
24G90013364H24
24G90013365H24
24G90013329H24
24G90012922H24
24G90013311H24
24G90013324H24
24G90012886H24
24G90013320H24
24G90013371H24
24G90013215H24
24G90013234H24
24G90013292H24
24G90013290H24
24G90013296H24
24G90013289H24
24G90013291H24
24G90013256H24
24G90013253H24
24G90013270H24
24G90013259H24
23G90102088J23
23G90102097J23
23G90103309J23
23G90109899J23
23G90102063J23
23G90103956J23
23G90109879J23
23G90103954J23
23G90109885J23
23G90109896J23
23G90103844J23
23G90109886J23
23G90109887J23
23G90103576J23
23G90103564J23
23G90103574J23
23G90103214J23
23G90103562J23
23G90103581J23
23G90103592J23
24G90016039A24
24G90016029A24
24G90016024A24
23G90103066J23
24G90016042A24
24G90016019A24
24G90016041A24
24G90016022A24
24G90016040A24
24G90016028A24
24G90016020A24
24G90016044A24
24G90016030A24
24G90016021A24
23G90103100J23
23G90103542J23
23G90103095J23
24G90016046A24
24G90016037A24
24G90016018A24
24G90016038A24
24G90016031A24
24G90016025A24
24G90016026A24
24G90016017A24
24G90016032A24
24G90016027A24
24G90016034A24
24G90016033A24
24G90016036A24
24G90010725A24
24G90010733A24
24G90016721A24
24G90016720A24
24G90016722A24
24G90010723A24
24G90016724A24
24G90016726A24
24G90010734A24
24G90016728A24
24G90010834A24
24G90010828A24
24G90010836A24
24G90010837A24
24G90010815A24
24G90010816A24
24G90015905A24
24G90010817A24
24G90015765A24
24G90015908A24
24G90010808A24
24G90010814A24
24G90010761A24
24G90015904A24
24G90010810A24
24G90010811A24
24G90010819A24
24G90016080A24
24G90010771A24
24G90010824A24
24G90016086A24
24G90010820A24
24G90010791A24
24G90010772A24
24G90016084A24
24G90010793A24
24G90010821A24
24G90010800A24
24G90010804A24
24G90016079A24
24G90016088A24
24G90010805A24
24G90010794A24
24G90010781A24
24G90010749A24
24G90016716A24
24G90016709A24
24G90016585A24
24G90016584A24
24G90016583A24
24G90016710A24
24G90016712A24
24G90016580A24
24G90016718A24
24G90016579A24
24G90016703A24
24G90016699A24
24G90016696A24
24G90016689A24
24G90016694A24
24G90016704A24
24G90016695A24
24G90016692A24
24G90016702A24
24G90000135B24
24G90000134B24
24G90000141B24
24G90000142B24
24G90000133B24
24G90000138B24
24G90000139B24
24G90000132B24
24G90011286B24
24G90011062B24
24G90011056B24
24G90011063B24
24G90011069B24
24G90011068B24
24G90000109B24
24G90000108B24
24G90000107B24
24G90000113B24
24G90017604B24
24G90017605B24
24G90017616B24
24G90017618B24
24G90017606B24
24G90017059B24
24G90017058B24
24G90017060B24
24G90011055B24
24G90017081B24
24G90017079B24
24G90000117B24
24G90017080B24
24G90000119B24
24G90012576B24
24G90000146B24
24G90000150B24
24G90012573B24
24G90000152B24
24G90012572B24
24G90000147B24
24G90000144B24
24G90012507B24
24G90017233B24
24G90017236B24
24G90017237B24
24G90012504B24
24G90012511B24
24G90012564B24
24G90012570B24
24G90012508B24
24G90012509B24
24G90012505B24
24G90012510B24
24G90012551B24
24G90012568B24
24G90012506B24
24G90012512B24
24G90012571B24
24G90012567B24
24G90012569B24
24G90012562B24
24G90012563B24
24G90000032B24
24G90000014B24
24G90017252B24
24G90000049B24
24G90000035B24
24G90000034B24
24G90017082B24
24G90017083B24
24G90012557B24
24G90012559B24
24G90012556B24
24G90012555B24
24G90017013B24
24G90012552B24
24G90017011B24
24G90012554B24
24G90012558B24
24G90012560B24
24G90012561B24
24G90017012B24
24G90017015B24
24G90017014B24
24G90012553B24
24G90017017B24
24G90017641B24
24G90017018B24
24G90017102B24
24G90017103B24
24G90017640B24
24G90012501B24
24G90017650B24
24G90017639B24
24G90017632B24
24G90017645B24
24G90017095B24
24G90017633B24
24G90017649B24
24G90012503B24
24G90017646B24
24G90017643B24
24G90017097B24
24G90017100B24
24G90017265B24
24G90017651B24
24G90017099B24
24G90017638B24
24G90017648B24
24G90017105B24
24G90017644B24
24G90017637B24
24G90017109B24
24G90017110B24
24G90017101B24
24G90017107B24
24G90017263B24
24G90017636B24
24G90017108B24
24G90017635B24
24G90017104B24
24G90017634B24
24G90017264B24
24G90017647B24
24G90017098B24
24G90012502B24
24G90017642B24
24G90017278B24
24G90017448B24
24G90017449B24
24G90017280B24
24G90017271B24
24G90017446B24
24G90017268B24
24G90017270B24
24G90017279B24
24G90017523B24
24G90017503B24
24G90017507B24
24G90017504B24
24G90017506B24
24G90017512B24
24G90017511B24
24G90017509B24
24G90017510B24
24G90017505B24
24G90017508B24
24G90017091B24
24G90017608B24
24G90017087B24
24G90017090B24
24G90017056B24
24G90017053B24
24G90017057B24
24G90017055B24
24G90017610B24
24G90017609B24
24G90017540B24
24G90017549B24
24G90017534B24
24G90017537B24
24G90017518B24
24G90017535B24
24G90017094B24
24G90017514B24
24G90017539B24
24G90017516B24
24G90017547B24
24G90017521B24
24G90017541B24
24G90017552B24
24G90017550B24
24G90017548B24
24G90017522B24
24G90017536B24
24G90017513B24
24G90017520B24
24G90017551B24
24G90017542B24
24G90017543B24
24G90017538B24
24G90017545B24
24G90017546B24
24G90017544B24
24G90017515B24
24G90017530B24
24G90017085B24
24G90017531B24
24G90017527B24
24G90017529B24
24G90017524B24
24G90017084B24
24G90017526B24
24G90017528B24
24G90017525B24
24G90017532B24
24G90013283H24
24G90013284H24
24G90013285H24
24G90013281H24
24G90013213H24
24G90015930A24
24G90015928A24
24G90015912A24
24G90015922A24
24G90015926A24
24G90015915A24
24G90015796A24
24G90010661A24
24G90015948A24
24G90010659A24
24G90015944A24
24G90010718A24
24G90010717A24
24G90010701A24
24G90010700A24
24G90010720A24
24G90010707A24
24G90010716A24
24G90010706A24
24G90016687A24
24G90016685A24
24G90016681A24
24G90016684A24
24G90016679A24
24G90016683A24
24G90016686A24
23G90103802J23
23G90103607J23
23G90103588J23
23G90103550J23
23G90103553J23
23G90103549J23
23G90103656J23
23G90103707J23
23G90103554J23
23G90103560J23
23G90103557J23
23G90103556J23
23G90103558J23
23G90103697J23
23G90103696J23
24G90012099B24
24G90012091B24
24G90012093B24
24G90011536B24
24G90011801B24
24G90011913B24
24G90011912B24
24G90011793B24
24G90011804B24
24G90012235B24
24G90011797B24
24G90011795B24
24G90011925B24
24G90011926B24
24G90012071B24
24G90012066B24
24G90012063B24
24G90012069B24
24G90012067B24
24G90012065B24
24G90012229B24
24G90005706B24
24G90017736C24
24G90008497D24
24G90012749D24
24G90012747D24
24G90009475D24
24G90009483D24
24G90009571D24
24G90009550D24
24G90009460D24
24G90009576D24
24G90009582D24
24G90009578D24
24G90009581D24
24G90009575D24
24G90009574D24
24G90009579D24
24G90009466D24
24G90009444D24
24G90009436D24
24G90009442D24
24G90009479D24
24G90009464D24
24G90009482D24
24G90009572D24
24G90008256D24
24G90008250D24
24G90008246D24
24G90008245D24
24G90011960B24
24G90011970B24
24G90011967B24
24G90011709B24
24G90011703B24
24G90011707B24
24G90011708B24
24G90011705B24
24G90011710B24
24G90005646B24
24G90005672B24
24G90005673B24
24G90005670B24
24G90011792B24
24G90011975B24
24G90011979B24
24G90011973B24
24G90011977B24
24G90011790B24
24G90011811B24
24G90011535B24
24G90011789B24
24G90011799B24
24G90011922B24
24G90011920B24
24G90011814B24
24G90011921B24
24G90011816B24
24G90011820B24
24G90011766B24
24G90011767B24
24G90011817B24
24G90011903B24
24G90011821B24
24G90011818B24
24G90011531B24
24G90011779B24
24G90011777B24
24G90011534B24
24G90011771B24
24G90011773B24
24G90011947B24
24G90011938B24
24G90011939B24
24G90011945B24
24G90011940B24
24G90011952B24
24G90011802B24
24G90011808B24
24G90011971B24
24G90011950B24
24G90011955B24
24G90011810B24
24G90011969B24
24G90011803B24
24G90011809B24
24G90011949B24
24G90011948B24
24G90011953B24
24G90011956B24
24G90005308B24
24G90005307B24
24G90005311B24
24G90005663B24
24G90005304B24
24G90005305B24
24G90005298B24
24G90005668B24
24G90005666B24
24G90005661B24
24G90005669B24
24G90011785B24
24G90011786B24
24G90011781B24
24G90011780B24
24G90005701B24
24G90005708B24
24G90005684B24
24G90005687B24
24G90005691B24
24G90005694B24
24G90005696B24
24G90005688B24
24G90005266B24
24G90005262B24
24G90005693B24
24G90005282B24
24G90005273B24
24G90005704B24
24G90011676B24
24G90005275B24
24G90005278B24
24G90011931B24
24G90011933B24
24G90011932B24
24G90011934B24
24G90011539B24
24G90011696B24
24G90011698B24
24G90011700B24
24G90010623A24
24G90010624A24
24G90010622A24
24G90010613A24
24G90010612A24
24G90015855A24
24G90015861A24
24G90015865A24
24G90015870A24
24G90011194B24
24G90011005B24
24G90011004B24
24G90010992B24
24G90011006B24
24G90011187B24
24G90010988B24
24G90011029B24
24G90011030B24
24G90011035B24
24G90011226B24
24G90011034B24
24G90011076B24
24G90010980B24
24G90011044B24
24G90011051B24
24G90011038B24
24G90011050B24
24G90011039B24
24G90011037B24
24G90011047B24
24G90011045B24
24G90011212B24
24G90011214B24
24G90011024B24
24G90011344B24
24G90011346B24
24G90011221B24
24G90011206B24
24G90011205B24
24G90010649A24
24G90010651A24
24G90010641A24
24G90010647A24
24G90010643A24
24G90010645A24
24G90010646A24
24G90015801A24
24G90010650A24
24G90010640A24
24G90010626A24
24G90013178H24
24G90012910H24
24G90012961H24
24G90012964H24
24G90001606D24
24G90001603D24
24G90001629D24
24G90001604D24
24G90001609D24
24G90001613D24
24G90012921H24
24G90001666D24
24G90001670D24
24G90001549D24
24G90001563D24
24G90001564D24
24G90001567D24
24G90001552D24
24G90001466D24
24G90001568D24
24G90001545D24
24G90001546D24
24G90001586D24
24G90001587D24
24G90001656D24
24G90001639D24
24G90001641D24
24G90001642D24
24G90001635D24
24G90001644D24
24G90001657D24
24G90001584D24
24G90001593D24
24G90001647D24
24G90001640D24
24G90001636D24
24G90001659D24
24G90001637D24
24G90001638D24
24G90001664D24
24G90001658D24
24G90001588D24
24G90001591D24
24G90001599D24
24G90001598D24
24G90001626D24
24G90001654D24
24G90001652D24
24G90001601D24
24G90001597D24
24G90001590D24
24G90001594D24
24G90001589D24
24G90010736A24
24G90010726A24
24G90015860A24
24G90016004A24
24G90016000A24
24G90010731A24
24G90010737A24
24G90016003A24
24G90010727A24
24G90016002A24
24G90010728A24
24G90010729A24
24G90015859A24
24G90016006A24
24G90015858A24
24G90015857A24
24G90010735A24
24G90016005A24
24G90015856A24
24G90010724A24
24G90016001A24
24G90010732A24
24G90015869A24
24G90015864A24
24G90010835A24
24G90015866A24
24G90015867A24
24G90015851A24
24G90015862A24
24G90015863A24
24G90015853A24
24G90010830A24
24G90015854A24
24G90015852A24
24G90010833A24
24G90016069A24
24G90015881A24
24G90010843A24
24G90010841A24
24G90010842A24
24G90010839A24
24G90016072A24
24G90016075A24
24G90010845A24
24G90016068A24
24G90010813A24
24G90016078A24
24G90010809A24
24G90016070A24
24G90010844A24
24G90010758A24
24G90010818A24
24G90010780A24
24G90010822A24
24G90010807A24
24G90010796A24
24G90010785A24
24G90010827A24
24G90010792A24
24G90010788A24
24G90010803A24
24G90010826A24
24G90010823A24
24G90010991B24
24G90010755A24
24G90010750A24
24G90010752A24
24G90016521A24
24G90016515A24
24G90011188B24
24G90015880A24
24G90015877A24
24G90015872A24
24G90015879A24
24G90015871A24
24G90015875A24
24G90015876A24
24G90011192B24
24G90011189B24
24G90010990B24
24G90017074B24
24G90011196B24
24G90010989B24
24G90010997B24
24G90010999B24
24G90010987B24
24G90011336B24
24G90017624B24
24G90017621B24
24G90000129B24
24G90011183B24
24G90011179B24
24G90011186B24
24G90017495B24
24G90011330B24
24G90000136B24
24G90000137B24
24G90011013B24
24G90011031B24
24G90011033B24
24G90011032B24
24G90011293B24
24G90000128B24
24G90011074B24
24G90010984B24
24G90000089B24
24G90000087B24
24G90000091B24
24G90010978B24
24G90011061B24
24G90011011B24
24G90011058B24
24G90011059B24
24G90011012B24
24G90011070B24
24G90000084B24
24G90000085B24
24G90000123B24
24G90011025B24
24G90000106B24
24G90000111B24
24G90011054B24
24G90011046B24
24G90011049B24
24G90012582B24
24G90011213B24
24G90000121B24
24G90000118B24
24G90000122B24
24G90017075B24
24G90017078B24
24G90011340B24
24G90000145B24
24G90012575B24
24G90012589B24
24G90012577B24
24G90000148B24
24G90012596B24
24G90012593B24
24G90012597B24
24G90012600B24
24G90012594B24
24G90012599B24
24G90012598B24
24G90012595B24
24G90012592B24
24G90012513B24
24G90011359B24
24G90011018B24
24G90011341B24
24G90011017B24
24G90011352B24
24G90011016B24
24G90000101B24
24G90000050B24
24G90017496B24
24G90017486B24
24G90017488B24
24G90017484B24
24G90017493B24
24G90017595B24
24G90017073B24
24G90017076B24
24G90017077B24
24G90011701B24
24G90011702B24
24G90011211B24
24G90011208B24
24G90011706B24
24G90011704B24
24G90005650B24
24G90011225B24
24G90005676B24
24G90005649B24
24G90011216B24
24G90005647B24
24G90005648B24
24G90005659B24
24G90005658B24
24G90005674B24
24G90011207B24
24G90011198B24
24G90005671B24
24G90005678B24
24G90011204B24
24G90005656B24
24G90011202B24
24G90005655B24
24G90011220B24
24G90005679B24
24G90005653B24
24G90005675B24
24G90005677B24
24G90005657B24
24G90017586B24
24G90017585B24
24G90017587B24
24G90017594B24
24G90017592B24
24G90011755B24
24G90017588B24
24G90011794B24
24G90011761B24
24G90011923B24
24G90011822B24
24G90017591B24
24G90011533B24
24G90011776B24
24G90011778B24
24G90011521B24
24G90017574B24
24G90017578B24
24G90011942B24
24G90017573B24
24G90017566B24
24G90011944B24
24G90011936B24
24G90017583B24
24G90017580B24
24G90017565B24
24G90017567B24
24G90011943B24
24G90017581B24
24G90017569B24
24G90017568B24
24G90017579B24
24G90017576B24
24G90017584B24
24G90017575B24
24G90017572B24
24G90017582B24
24G90011951B24
24G90017564B24
24G90011954B24
24G90017562B24
24G90011744B24
24G90017559B24
24G90011746B24
24G90012516B24
24G90017557B24
24G90017560B24
24G90017561B24
24G90017563B24
24G90012522B24
24G90011747B24
24G90005494B24
24G90012528B24
24G90005493B24
24G90012530B24
24G90012525B24
24G90005492B24
24G90012537B24
24G90012523B24
24G90012527B24
24G90012533B24
24G90012531B24
24G90012526B24
24G90012524B24
24G90011957B24
24G90005491B24
24G90005664B24
24G90005260B24
24G90005299B24
24G90005667B24
24G90005261B24
24G90005665B24
24G90005660B24
24G90005662B24
24G90005301B24
24G90005300B24
24G90005303B24
24G90011788B24
24G90011783B24
24G90017481B24
24G90012521B24
24G90017602B24
24G90017596B24
24G90012519B24
24G90012515B24
24G90012517B24
24G90017600B24
24G90017483B24
24G90017599B24
24G90012520B24
24G90017601B24
24G90012514B24
24G90010644A24
24G90010642A24
24G90010631A24
24G90010658A24
24G90015943A24
24G90010715A24
24G90010692A24
24G90010722A24
24G90010694A24
24G90010691A24
24G90010687A24
24G90010683A24
24G90010721A24
24G90010719A24
24G90010710A24
24G90010693A24
24G90010699A24
24G90010684A24
24G90010698A24
24G90010711A24
24G90010703A24
24G90010712A24
24G90010708A24
24G90010704A24
24G90010713A24
24G90010685A24
24G90010686A24
24G90010695A24
24G90010689A24
24G90010705A24
24G90017554B24
24G90017629B24
24G90017631B24
24G90017630B24
24G90017555B24
24G90005692B24
24G90005690B24
24G90005699B24
24G90005695B24
24G90005685B24
24G90005707B24
24G90005697B24
24G90005689B24
24G90005686B24
24G90005682B24
24G90005683B24
24G90005680B24
24G90005681B24
24G90005703B24
24G90005277B24
24G90005269B24
24G90005700B24
24G90005705B24
24G90005702B24
24G90005281B24
24G90005279B24
24G90005267B24
24G90011550B24
24G90011930B24
24G90011935B24
24G90011695B24
24G90011547B24
24G90011699B24
24G90012913H24
24G90017999H24
24G90012874H24
24G90012926H24
24G90012868H24
24G90012900H24
24G90012945H24
24G90012875H24
24G90012893H24
24G90012955H24
24G90012907H24
24G90012908H24
24G90012909H24
24G90012870H24
24G90012871H24
24G90017996H24
24G90018002H24
24G90012902H24
24G90018000H24
24G90012928H24
24G90012892H24
24G90012896H24
24G90012924H24
24G90012887H24
24G90012957H24
24G90012923H24
24G90018030H24
24G90018021H24
24G90018025H24
24G90018034H24
24G90011427H24
24G90011429H24
24G90018033H24
24G90011430H24
24G90013672H24
24G90018017H24
24G90011428H24
24G90018031H24
24G90011426H24
24G90011431H24
24G90018024H24
24G90018028H24
24G90018018H24
24G90018032H24
24G90018009H24
24G90013678H24
24G90006847D24
24G90006848D24
24G90006839D24
24G90006837D24
24G90006854D24
24G90006836D24
24G90006838D24
24G90001671D24
24G90001682D24
24G90001684D24
24G90001673D24
24G90001676D24
24G90001675D24
24G90001677D24
24G90001674D24
24G90001683D24
24G90001678D24
24G90001672D24
24G90001681D24
24G90001679D24
24G90001562D24
24G90001542D24
24G90001560D24
24G90001538D24
24G90001551D24
24G90001557D24
24G90001565D24
24G90001559D24
24G90001536D24
24G90001547D24
24G90001555D24
24G90001548D24
24G90001556D24
24G90001561D24
24G90001537D24
24G90010350D24
24G90010345D24
24G90010347D24
24G90010346D24
24G90010344D24
24G90010348D24
24G90010351D24
24G90010349D24
24G90001883D24
24G90001884D24
24G90001686D24
24G90001891D24
24G90001655D24
24G90001543D24
24G90001886D24
24G90001544D24
24G90001888D24
24G90001885D24
24G90001887D24
24G90001687D24
24G90012952H24
24G90012939H24
24G90012894H24
23G90100150B24
24G90001669D24
24G90001668D24
24G90011980B24
24G90011964B24
24G90011966B24
24G90011965B24
24G90011968B24
24G90011976B24
24G90011974B24
24G90011978B24
24G90011774B24
24G90011775B24
24G90011946B24
24G90011937B24
24G90011806B24
24G90011958B24
24G90011784B24
24G90011782B24
24G90011787B24
24G90005268B24
24G90005276B24
24G90001651D24
24G90013295H24
24G90013220H24
24G90013257H24
24G90013183H24
24G90013271H24
24G90013286H24
24G90013175H24
24G90013293H24
24G90013179H24
24G90013174H24
24G90013217H24
24G90007313H24
24G90012917H24
24G90012912H24
24G90012873H24
24G90012876H24
24G90012932H24
24G90013265H24
24G90012891H24
24G90012897H24
24G90013258H24
24G90012869H24
24G90012931H24
24G90012918H24
24G90013269H24
24G90012898H24
24G90013375H24
24G90012919H24
24G90012953H24
24G90012906H24
24G90012888H24
24G90013210H24
24G90012904H24
24G90012940H24
24G90012905H24
24G90012903H24
24G90012901H24
24G90013322H24
24G90013323H24
24G90013357H24
24G90013368H24
24G90012960H24
24G90013327H24
24G90013326H24
24G90013317H24
24G90013374H24
24G90012937H24
24G90013373H24
24G90012966H24
24G90013367H24
24G90013321H24
24G90013372H24
24G90013328H24
24G90012879H24
24G90012885H24
24G90013319H24
24G90013316H24
24G90013318H24
24G90013229H24
24G90013225H24
24G90013227H24
24G90013211H24
24G90013222H24
24G90013226H24
24G90013297H24
24G90013331H24
24G90013254H24
24G90013260H24
24G90013255H24
24G90013248H24
24G90013249H24
24G90013230H24
23G90109799J23
23G90103856J23
23G90102098J23
23G90103852J23
23G90103850J23
23G90102095J23
23G90102087J23
23G90102086J23
23G90103848J23
23G90102096J23
23G90102458K23
23G90102460K23
23G90102462K23
23G90102461K23
23G90102455K23
23G90102464K23
23G90102295K23
23G90102838K23
23G90102463K23
23G90102836K23
23G90109816J23
24G90016443A24
24G90016350A24
24G90016352A24
23G90103310J23
23G90103308J23
23G90103307J23
23G90103316J23
23G90109861J23
23G90102843K23
23G90103339J23
23G90102848K23
23G90102860K23
23G90109897J23
23G90102846K23
23G90102858K23
23G90102280K23
23G90102276K23
23G90102844K23
23G90102853K23
23G90102212K23
23G90102277K23
23G90102856K23
23G90102851K23
23G90102281K23
23G90102279K23
23G90103341J23
23G90103511J23
23G90103991J23
23G90103652J23
23G90103998J23
23G90103958J23
23G90103957J23
23G90109893J23
23G90103858J23
23G90103959J23
23G90109884J23
23G90102103J23
23G90102061J23
23G90102062J23
23G90103649J23
23G90103668J23
23G90109880J23
23G90103953J23
23G90103999J23
23G90103997J23
23G90103955J23
23G90104000J23
23G90109878J23
23G90103996J23
23G90103364J23
23G90103983J23
23G90102368K23
23G90102365K23
23G90102864K23
23G90102473K23
23G90102477K23
23G90102467K23
23G90102862K23
23G90102469K23
23G90102255K23
23G90103518J23
23G90103594J23
23G90103570J23
23G90103369J23
23G90103215J23
23G90103216J23
23G90103572J23
23G90103543J23
23G90103084J23
23G90103222J23
23G90103537J23
23G90103544J23
23G90103539J23
23G90103540J23
23G90103087J23
23G90103541J23
23G90103234J23
23G90103536J23
23G90103538J23
24G90016023A24
23G90102228K23
23G90102239K23
23G90102223K23
23G90102182K23
23G90102224K23
23G90102230K23
23G90102232K23
23G90102229K23
23G90102222K23
23G90102241K23
23G90102237K23
23G90102213K23
23G90102184K23
23G90102233K23
24G90016627A24
23G90102236K23
23G90102226K23
23G90102225K23
23G90102235K23
23G90102221K23
23G90102231K23
23G90102234K23
23G90102274K23
23G90102227K23
23G90102238K23
24G90016723A24
24G90010730A24
24G90016727A24
24G90016077A24
24G90016076A24
24G90016082A24
24G90016087A24
24G90016760A24
24G90016763A24
24G90016765A24
24G90016758A24
24G90016767A24
24G90016757A24
24G90016768A24
24G90016764A24
24G90016762A24
24G90016761A24
24G90016766A24
24G90016715A24
24G90016713A24
24G90016586A24
24G90016582A24
24G90016717A24
24G90016505A24
24G90016517A24
24G90016511A24
24G90016504A24
24G90016520A24
24G90016440A24
24G90016496A24
24G90016503A24
24G90016392A24
24G90016502A24
24G90016394A24
24G90016439A24
24G90016380A24
24G90016494A24
24G90016438A24
24G90016512A24
24G90016385A24
24G90016497A24
24G90016509A24
24G90016390A24
24G90016381A24
24G90016389A24
24G90016388A24
24G90016488A24
24G90016490A24
24G90016384A24
24G90016386A24
24G90016441A24
24G90016437A24
24G90016500A24
24G90016510A24
24G90016495A24
24G90016491A24
24G90016492A24
24G90016379A24
24G90016393A24
24G90016516A24
24G90016387A24
24G90016501A24
24G90016487A24
24G90016391A24
24G90016506A24
24G90016382A24
24G90016508A24
24G90016493A24
24G90016507A24
24G90016486A24
24G90016499A24
24G90016513A24
24G90016383A24
24G90016519A24
24G90016489A24
24G90016498A24
24G90016518A24
24G90016334A24
24G90016474A24
24G90016475A24
24G90016376A24
24G90016332A24
24G90016325A24
24G90016473A24
24G90016338A24
24G90016339A24
24G90016335A24
24G90016373A24
24G90016479A24
24G90016331A24
24G90016341A24
24G90016481A24
24G90016330A24
24G90016327A24
24G90016480A24
24G90016482A24
24G90016485A24
24G90016378A24
24G90016377A24
24G90016337A24
24G90016484A24
24G90016324A24
24G90016326A24
24G90016375A24
24G90016374A24
24G90016476A24
24G90016329A24
24G90016478A24
24G90016483A24
24G90016781A24
24G90016782A24
24G90016769A24
24G90016771A24
24G90016770A24
24G90016778A24
24G90016773A24
24G90016776A24
24G90016777A24
24G90016774A24
24G90016775A24
24G90016780A24
24G90001667D24
24G90017613B24
24G90017612B24
24G90017620B24
24G90011588B24
24G90011600B24
24G90011591B24
24G90011589B24
24G90011590B24
24G90017611B24
24G90017623B24
24G90017619B24
24G90017502B24
24G90000130B24
24G90011290B24
24G90011292B24
24G90011075B24
24G90000090B24
24G90000086B24
24G90010979B24
24G90011067B24
24G90011060B24
24G90000099B24
24G90000112B24
24G90000098B24
24G90000104B24
24G90017052B24
24G90017050B24
24G90017043B24
24G90017048B24
24G90017044B24
24G90017047B24
24G90017045B24
24G90017051B24
24G90017603B24
24G90017046B24
24G90000116B24
24G90000120B24
24G90000115B24
24G90000114B24
24G90000143B24
24G90000149B24
24G90000151B24
24G90017498B24
24G90017500B24
24G90000102B24
24G90000103B24
24G90000051B24
24G90017061B24
24G90017501B24
24G90017062B24
24G90017499B24
24G90017497B24
24G90017489B24
24G90017490B24
24G90017491B24
24G90017492B24
24G90017487B24
24G90017069B24
24G90017066B24
24G90017071B24
24G90017072B24
24G90017065B24
24G90017068B24
24G90017028B24
24G90017041B24
24G90017037B24
24G90017036B24
24G90017039B24
24G90017022B24
24G90017025B24
24G90017038B24
24G90017019B24
24G90017026B24
24G90017020B24
24G90017027B24
24G90017024B24
24G90017040B24
24G90017021B24
24G90017029B24
24G90017031B24
24G90017030B24
24G90017450B24
24G90011529B24
24G90011596B24
24G90011712B24
24G90011752B24
24G90011753B24
24G90011598B24
24G90011764B24
24G90011733B24
24G90011593B24
24G90011594B24
24G90011592B24
24G90011751B24
24G90017590B24
24G90011763B24
24G90017593B24
24G90011759B24
24G90011765B24
24G90011599B24
24G90011512B24
24G90011511B24
24G90011760B24
24G90017589B24
24G90011757B24
24G90011762B24
24G90011750B24
24G90011758B24
24G90011768B24
24G90011769B24
24G90011739B24
24G90011736B24
24G90011528B24
24G90011595B24
24G90011754B24
24G90011737B24
24G90011524B24
24G90011526B24
24G90011530B24
24G90011714B24
24G90011711B24
24G90011525B24
24G90011522B24
24G90011713B24
24G90011507B24
24G90011516B24
24G90011520B24
24G90011506B24
24G90017494B24
24G90011504B24
24G90011514B24
24G90017571B24
24G90011505B24
24G90011502B24
24G90017577B24
24G90011519B24
24G90011509B24
24G90011513B24
24G90011518B24
24G90011515B24
24G90011501B24
24G90011508B24
24G90011510B24
24G90011503B24
24G90011517B24
24G90011749B24
24G90017042B24
24G90011748B24
24G90017558B24
24G90017556B24
24G90011741B24
24G90011740B24
24G90011743B24
24G90011745B24
24G90017607B24
24G90011742B24
24G90017064B24
24G90017067B24
24G90017482B24
24G90017598B24
24G90017597B24
24G90008404D24
24G90008387D24
24G90008358D24
24G90008383D24
24G90008380D24
24G90008363D24
24G90008360D24
24G90008364D24
24G90008361D24
24G90008359D24
24G90008378D24
24G90013282H24
23G90107047J23
23G90107058J23
23G90107057J23
23G90107712J23
23G90102166K23
23G90107046J23
23G90107054J23
23G90107060J23
23G90107061J23
23G90107053J23
23G90107056J23
23G90105394K23
23G90107713J23
23G90105644K23
24G90010702A24
24G90010697A24
24G90010696A24
23G90102248K23
23G90102250K23
23G90102466K23
23G90102449K23
23G90103012J23
23G90103804J23
23G90103801J23
23G90103803J23
23G90103015J23
23G90102445K23
23G90102465K23
23G90102447K23
23G90103800J23
23G90103799J23
23G90103712J23
23G90103605J23
23G90103606J23
23G90103609J23
23G90103597J23
23G90103659J23
23G90103684J23
23G90103660J23
23G90103608J23
23G90102354K23
23G90102448K23
23G90107563K23
23G90102246K23
23G90102377K23
23G90102243K23
23G90102247K23
23G90102244K23
23G90107561K23
23G90103545J23
23G90103552J23
23G90102440K23
23G90102438K23
23G90102351K23
23G90102825J23
23G90103555J23
23G90103694J23
23G90103705J23
23G90107448J23
23G90107447J23
23G90102633J23
23G90102822J23
23G90102375K23
23G90103686J23
23G90103688J23
23G90103687J23
23G90103681J23
23G90103680J23
23G90103658J23
23G90103678J23
23G90103677J23
23G90103691J23
23G90103689J23
23G90103683J23
23G90102398K23
23G90102395K23
23G90103692J23
23G90103690J23
23G90103703J23
23G90103699J23
23G90103702J23
23G90102832K23
23G90102385K23
23G90102376K23
23G90102383K23
23G90102363K23
23G90102249K23
23G90102380K23
23G90102381K23
23G90102378K23
23G90107710J23
23G90102831J23
24G90017035B24
24G90017034B24
24G90017033B24
24G90017628B24
24G90017032B24
24G90017553B24
24G90011677B24
24G90011678B24
24G90011667B24
24G90011664B24
24G90011663B24
24G90011675B24
24G90011673B24
24G90011680B24
24G90011671B24
24G90011672B24
24G90011679B24
24G90011662B24
24G90011661B24
24G90011674B24
24G90011545B24
24G90011543B24
24G90011542B24
24G90011541B24
24G90011684B24
24G90011685B24
24G90011655B24
24G90011691B24
24G90011665B24
24G90011666B24
24G90011669B24
24G90011670B24
24G90011668B24
24G90011656B24
24G90011657B24
24G90011694B24
24G90011693B24
24G90011692B24
24G90011654B24
24G90011546B24
24G90011548B24
24G90011653B24
24G90011651B24
24G90011690B24
24G90011683B24
24G90011658B24
24G90011652B24
24G90011686B24
24G90011660B24
24G90011659B24
24G90011689B24
24G90011688B24
24G90011687B24
24G90011681B24
24G90001661D24
24G90001663D24
24G90001653D24
24G90001649D24
24G90001619D24
24G90001632D24
24G90001650D24
24G90001600D24
24G90012956H24
24G90012944H24
24G90013069H24
24G90013071H24
24G90013068H24
24G90012949H24
24G90012942H24
24G90018108H24
24G90018109H24
24G90018106H24
24G90018089H24
24G90018091H24
24G90018086H24
24G90001616D24
24G90011191B24
24G90011197B24
24G90010994B24
24G90010995B24
24G90010993B24
24G90010985B24
24G90011066B24
24G90011077B24
24G90011010B24
24G90011009B24
24G90011048B24
24G90011036B24
24G90011360B24
24G90011357B24
24G90011351B24
24G90011354B24
24G90011021B24
24G90011358B24
24G90011356B24
24G90012095B24
24G90011209B24
24G90012097B24
24G90012105B24
24G90012098B24
24G90011200B24
24G90011199B24
24G90011203B24
24G90011201B24
24G90011910B24
24G90011904B24
24G90005274B24
24G90005265B24
24G90001595D24
24G90001628D24
24G90001585D24
24G90001631D24
24G90001633D24
24G90001643D24
24G90001646D24
24G90001630D24
24G90001634D24
24G90001648D24
24G90001618D24
24G90001592D24
24G90004246F24
24G90004248F24
24G90007430F24
24G90007429F24
24G90007422F24
24G90007425F24
24G90007428F24
24G90007426F24
24G90004244F24
24G90007496F24
24G90007424F24
24G90004249F24
24G90001831F24
24G90007431F24
24G90007494F24
24G90001825F24
24G90001834F24
24G90007432F24
24G90007499F24
24G90007500F24
24G90007490F24
24G90007493F24
24G90001830F24
24G90007497F24
24G90007423F24
24G90007378F24
24G90007427F24
24G90004243F24
24G90009500F24
24G90009499F24
24G90004242F24
24G90004245F24
24G90007419F24
24G90004247F24
24G90001827F24
24G90001828F24
24G90001823F24
24G90001824F24
24G90001826F24
24G90001822F24
24G90010355F24
24G90001829F24
24G90001833F24
24G90007495F24
24G90001608D24
23G90000314H23
23G90000312H23
23G90000316H23
23G90000313H23
23G90102256K23
23G90009051H23
23G90009407H23
23G90009403H23
23G90009410H23
23G90009413H23
23G90009449H23
23G90009626H23
23G90009425H23
23G90009624H23
23G90009629H23
23G90009769H23
23G90009771H23
23G90009630H23
23G90009622H23
23G90009053H23
23G90009073H23
23G90009071H23
23G90009429H23
23G90009550H23
23G90009473H23
23G90009472H23
23G90009614H23
23G90000371H23
23G90009474H23
23G90009619H23
23G90009668H23
23G90000373H23
23G90000375H23
23G90009554H23
23G90009476H23
23G90009618H23
23G90009610H23
23G90009611H23
23G90009556H23
23G90000372H23
23G90009549H23
23G90009606H23
23G90009609H23
23G90009612H23
23G90009552H23
23G90009553H23
23G90009669H23
23G90000374H23
23G90009030H23
23G90000306H23
23G90009524H23
23G90009632H23
23G90000308H23
23G90000369H23
23G90000304H23
23G90000370H23
23G90009593H23
23G90000303H23
23G90000302H23
23G90009280H23
23G90009275H23
23G90009279H23
23G90009586H23
23G90009281H23
23G90009282H23
23G90009589H23
23G90009585H23
23G90009597H23
23G90009594H23
23G90009276H23
23G90009277H23
23G90009599H23
23G90009753H23
23G90009528H23
23G90009526H23
23G90009640H23
23G90009651H23
23G90009652H23
23G90009650H23
23G90009271H23
23G90009038H23
23G90009285H23
23G90009443H23
23G90009445H23
23G90009603H23
23G90009601H23
23G90009607H23
23G90009613H23
23G90009608H23
23G90009604H23
23G90009522H23
23G90009748H23
23G90009034H23
23G90009283H23
23G90009438H23
23G90009436H23
23G90009602H23
23G90009284H23
23G90009649H23
23G90009646H23
23G90009536H23
23G90009653H23
23G90009535H23
23G90009537H23
23G90009525H23
23G90009529H23
23G90009760H23
23G90009268H23
23G90009272H23
23G90009273H23
23G90009639H23
23G90009523H23
23G90009664H23
23G90009663H23
23G90009662H23
23G90009655H23
23G90009661H23
23G90009581H23
23G90009579H23
23G90009660H23
23G90009656H23
23G90009659H23
23G90009657H23
23G90009658H23
23G90009583H23
23G90009584H23
23G90009582H23
24G90010323D24
24G90010321D24
24G90010310D24
24G90009619D24
24G90010035D24
24G90010029D24
24G90010027D24
24G90010449D24
24G90003355D24
24G90010032D24
24G90010033D24
24G90018912D24
24G90007433F24
24G90012089B24
24G90012094B24
24G90012120B24
24G90012118B24
24G90011227B24
24G90015878A24
24G90015874A24
24G90015868A24
24G90008247D24
24G90001625D24
24G90011961B24
24G90011770B24
24G90001627D24
24G90001615D24
24G90011800B24
24G90011772B24
24G90001596D24
24G90012061B24
24G90001607D24
24G90012965H24
24G90001554D24
24G90001645D24
24G90012121B24
24G90009715D24
24G90011798B24
24G90011807B24
24G90011537B24
24G90011532B24
24G90011193B24
24G90011190B24
24G90011007B24
24G90011000B24
24G90010996B24
24G90011001B24
24G90011002B24
24G90011332B24
24G90010986B24
24G90011310B24
24G90011283B24
24G90011311B24
24G90011312B24
24G90011313B24
24G90017211B24
24G90011307B24
24G90011309B24
24G90011185B24
24G90011182B24
24G90011180B24
24G90011181B24
24G90011178B24
24G90011329B24
24G90011334B24
24G90011284B24
24G90011288B24
24G90011285B24
24G90011316B24
24G90011287B24
24G90011071B24
24G90011299B24
24G90011297B24
24G90011015B24
24G90011301B24
24G90011014B24
24G90011008B24
24G90017232B24
24G90011296B24
24G90011303B24
24G90011300B24
24G90011304B24
24G90011027B24
24G90011298B24
24G90011210B24
24G90011337B24
24G90011215B24
24G90000005B24
24G90000002B24
24G90000003B24
24G90017234B24
24G90017235B24
24G90000006B24
24G90011020B24
24G90011328B24
24G90011019B24
24G90011022B24
24G90011023B24
24G90000052B24
24G90000027B24
24G90017257B24
24G90000062B24
24G90000019B24
24G90000037B24
24G90000068B24
24G90000028B24
24G90000077B24
24G90000074B24
24G90017241B24
24G90000057B24
24G90000073B24
24G90000072B24
24G90000008B24
24G90000054B24
24G90000044B24
24G90000061B24
24G90000022B24
24G90000056B24
24G90000041B24
24G90000063B24
24G90000023B24
24G90000012B24
24G90000055B24
24G90000070B24
24G90000020B24
24G90000060B24
24G90000043B24
24G90017250B24
24G90000075B24
24G90000013B24
24G90017249B24
24G90000040B24
24G90000047B24
24G90017244B24
24G90017256B24
24G90000053B24
24G90000059B24
24G90000046B24
24G90000007B24
24G90017240B24
24G90000058B24
24G90017245B24
24G90017243B24
24G90017251B24
24G90000045B24
24G90017255B24
24G90000029B24
24G90000033B24
24G90000031B24
24G90000009B24
24G90000071B24
24G90000069B24
24G90000036B24
24G90017261B24
24G90017262B24
24G90017258B24
24G90017260B24
24G90017259B24
24G90017267B24
24G90017276B24
24G90017274B24
24G90017286B24
24G90011224B24
24G90011218B24
24G90017287B24
24G90011223B24
24G90017275B24
24G90017273B24
24G90011222B24
24G90017225B24
24G90017231B24
24G90017219B24
24G90017212B24
24G90017230B24
24G90017220B24
24G90017228B24
24G90017229B24
24G90017224B24
24G90017218B24
24G90017215B24
24G90017222B24
24G90017213B24
24G90017221B24
24G90017227B24
24G90017223B24
24G90017216B24
24G90017217B24
24G90000010B24
24G90011295B24
24G90000039B24
24G90011219B24
24G90012297B24
24G90012119B24
24G90012920H24
24G90012962H24
24G90001623D24
24G90012968H24
24G90001566D24
24G90012889H24
24G90001602D24
|
| FEI Number |
3004123209
|
Recalling Firm/ Manufacturer |
HeartSine Technologies Ltd
207 Airport Road West
Belfast
United Kingdom
|
| For Additional Information Contact | Joyce Staggs 425-867-5497 |
Manufacturer Reason for Recall | Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check. |
FDA Determined Cause 2 | Process control |
| Action | On 07/28/2026, Stryker sent several versions of an Urgent: Medical device recall letter that outlines two issues:
1) An issue with the membrane tail, an internal component, could cause an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs.
2) A limited number of HeartSine samaritan PAD devices were distributed without the required maintenance insert. This insert contains periodic maintenance instructions implemented as an additional risk control measure. While most devices contain the maintenance insert, this voluntary correction is being initiated to ensure all end users have access to the supplemental maintenance information.
For the missing required maintenance insert (issue 2), Customers are instructed to:
1. Identify impacted devices by verifying if your device serial numbers are included in the list at the link https://heartsinerecall.com/HeartSineSamaritanPAD/LookUp. For this field action notice, select AED + Pad-Pak.
2. Download a copy of the missing maintenance insert when acknowledging the recall and store with your devices user manual until a replacement is received.
For questions or concerns, contact Stryker Customer Service at +1 8007879537, option 2, from 8 AM to 7 PM Eastern Time, Monday to Friday, your Authorized Distributor, or HeartSine Technologies Technical Support at heartsinesupport@stryker.com. |
| Quantity in Commerce | 31,834 |
| Distribution | U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = MKJ
|
|
|
|