 Class 2 Device Recall BD Cathena" Safety IV Catheter BD Multiguard" Technologysee related information
Date Initiated by FirmSeptember 02, 2026
Date PostedNovember 23, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0530-2027
Recall Event ID 99848
510(K)NumberK251155 
Product Classification Catheter, intravascular, therapeutic, short-term less than 30 days - Product Code FOZ
ProductBD Cathena" Safety IV Catheter BD Multiguard" Technology 20 GA x 1.00 IN REF 386862 UDI-DI code: (01)00382903868629 BD Cathena Safety IV Catheter is intended to be inserted into a patient s peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids.
Code Information Model Number: 386862 UDI-DI code: (01)00382903868629 Lot Number: 5335702
FEI Number 1710034
Recalling Firm/
Manufacturer
Becton Dickinson Infusion Therapy Systems, Inc.
9450 S State St
Sandy UT 84070-3213
For Additional Information ContactFallon McLoughlin
801-565-2300
Manufacturer Reason
for Recall
Needle Safety shield for Intravenous catheter may not properly engage and allow the needles to pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
FDA Determined
Cause 2
Process control
ActionOn 09/02/2026, an "URGENT: Medical Device Recall" letter was emailed to customers informing them that, BD has confirmed the safety shield may not properly engage on the affected product listed above. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed. Customer are instructed to: 1. Immediately review inventory for Catalog Number 386862, Lot number 5335702. Destroy all unused product subject to the recall following institution s process for destruction. 2. Review all areas where the affected product may be in use. Immediately discontinue use of the affected product and destroy any remaining unused inventory. 3. Provide a copy of this notification to any customers and/or facilities you may have distributed devices to. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. For Questions or further assistance: North American Regional Complaint Center - Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Quantity in Commerce200000 units
DistributionU.S. Nationwide distribution in the states of AZ, CA, FL, GA, IA, IN, LA, MA, MD, MS, NC, NY, PA, TX, and UT.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.