• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Abbott

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Abbottsee related information
Date Initiated by FirmJune 09, 2003
Date PostedJune 27, 2003
Recall Status1 Terminated 3 on August 27, 2004
Recall NumberZ-1134-03
Recall Event ID 26450
Product Classification Set, Administration, Intravascular - Product Code FPA
Product96-8366 Drip Chambers, Convertible with Pierce Pin .138 Assembly; Abbott, Ashland, Ohio 44805; 650 drip chambers per case; Abbott Commodity 968366, Ashland Product #85-55960
Code Information Abbott Commodity 96-8366, Ashland Product #85-55960, lot numbers 85C51LBMP, 89A195IMP, 89B335IMP, 89B345IMP, 90A105IMP, 90A125IMP, 91A145IMP, 91A155IMP, 91A165IMP, 92A125IMP
FEI Number 1415939
Recalling Firm/
Manufacturer
Abbott Laboratories HPD/ADD
100/200 Abbott Park Road
Abbott Park IL 60064
For Additional Information ContactAbbott Customer Service
800-222-6883
Manufacturer Reason
for Recall
The drip chambers and/or adapters in the sets were manufactured with non-radiation grade plastic resin, resulting in discoloration of the resin.
FDA Determined
Cause 2
Other
ActionThe IV administration sets were recalled by letters dated 6/9/03, advising the accounts of the discoloration of the sight chamber and adapters due to the use of non-radiation grade plastic resin in the manufacture of these components of the sets and requesting the return of the sets for credit. These letters were accompanied by a customer ship history/business reply form, which listed the specific set list numbers and lot numbers shipped to the direct account. The direct accounts were requested to send copies of the enclosed recall letter with the listing of all affected sets to their subaccounts. Abbott sent recall letters dated 6/13/03 to those accounts who received I.V. set list 018810458, lot 910824W packed in cases with Liposyn II 20% Kits, list 97920403, lot 92552DF and list 97930403, lot 94553DF. These accounts were informed of the problem with the administration sets and were requested to remove the sets from the Liposyn II cases and return the sets for credit. Abbott sent recall letters dated 7/7/03 to distributors and veterinary clinics receiving the 12 additional lots of primary I.V. sets, list numbers 018810448 and 089620448, informing them of the problem with the drip chambers and requesting them to return the affected lots for credit. The distributors were requested to notify their customers. Abbott sent recall letters dated 7/7/03 to the third party manufacturers receiving the affected drip chamber components, informing them of the problem with the drip chambers and requesting the return of the affected lots. Abbott sent recall letters dated 8/19/03 to hospitals and distributors receiving the 9 additional lots of I.V. sets, list numbers 119440402, 119480402, 120300412, 122590402 and 123120401, informing them of the problem with the drip chambers and requesting them to return the affected lots for credit. The distributors were requested to notify their customers. Abbott sent recall letters dated 5/13/04 to hospitals and distributor
Quantity in Commerce911,750 drip chambers
DistributionNationwide and Internationally to Abbott affiliates in Argentina, Aruba, Australia, the Bahamas, Barbados, Bermuda, Bolivia, Brazil, Canada, Cayman Islands, Chile, Columbia, Costa Rica, Croatia, Curacao, Dominican Republic, Ecuador, Egypt, El Salvador, France, Greece, Guam, Guatemala, Haiti, Honduras, Hong Kong, Hungary, Italy, Jordan, Kuwait, Mexico, New Zealand, Nicaragua, Pakistan, Panama, Paraguay, Peru, Suadia Arabia, Singapore, Taiwan, Trinidad, Turkey, United Arab Emirates and Uruguay
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-