Class 2 Device Recall LIFEPAK 12 |
|
Date Initiated by Firm |
March 22, 2006 |
Date Posted |
April 04, 2006 |
Recall Status1 |
Terminated 3 on September 06, 2011 |
Recall Number |
Z-0710-06 |
Recall Event ID |
34947 |
510(K)Number |
K010918
|
Product Classification |
Automated External Defibrillators (Non-Wearable) - Product Code MKJ
|
Product |
LIFEPAK 12 defibrillator/monitor |
Code Information |
Domestic Serial Numbers: 34106383, 33996555, 33992033, 33992030, 33992029, 33992031, 33992027, 33990244, 33990241, 33990245, 33990242, 33990240, 34096406, 34086957, 34039438, 34100891, 33917921, 34100498, 34019560, 34011939, 34011940, 33985062, 33985063, 33985064, 33985065, 33985066, 33985067, 33985068, 33985070, 33985069, 33985074, 33985061, 33985071, 33985073, 33915256, 34016156, 34016153, 34016154, 34016155, 34016157, 34016158, 33989218, 33989220, 33989216, 33989219, 33989217, 34039448, 34048193, 34100900, 34020194, 34020192, 34060159, 33988136, 33988137, 34020193, 33995357, 33949669, 34074066, 33913884, 34014205, 11201950, 34091725, 34064561, 33913893, 33913894, 33913895, 34004873, 33995317, 33995318, 33995316, 33995315, 33995319, 34064439, 34064440, 34064441, 34020190, 34020507, 34020187, 34020189, 34020188, 34019563, 34019572, 34019573, 33904159, 33905092, 34095289, 34095297, 34095301, 33991999, 33992000, 33992003, 33992023, 33992028, 33992016, 33991998, 34087619, 34087620, 34087605, 34087610, 34087612, 34087035, 34087892, 34087602, 34087615, 34087616, 34087617, 34087618, 34086700, 34086709, 34086715, 34087613, 34086714, 34087609, 34087607, 34087614, 34087608, 34086718, 34086706, 34086719, 34086711, 34086713, 34086705, 34086717, 34086702, 34087606, 34086703, 34087033, 34086708, 34086712, 34087603, 34086707, 34087604, 34086716, 34087601, 34087034, 34091720, 33949930, 34091120, 34039430, 34039429, 34015786, 34011974, 33982989, 33899585, 33995356, 33995351, 33995350, 33995346, 33995360, 33995355, 33995369, 33995349, 33995348, 33995352, 33995347, 33995359, 33995353, 33995345, 33995358, 33995354, 12737672, 33913878, 34094631, 34049603, 34049601, 34049604, 34049611, 34049606, 34049607, 34049609, 34049612, 34049608, 34049610, 34049605, 34049602, 30859396, 33978921, 33978922, 33978923, 33978924, 34039468, 34039433, 34039470, 34083705, 34096379, 33915357, 33988150, 33992032, 33931367, 33931365, 33931364, 33931363, 33931362, 33931366, 34052702, 34052703, 33795379, 34095303, 34095298, 34095299, 34095304, 34095300, 34096375, 34095314, 34095313, 34095312, 34095307, 34095306, 34095305, 33862271, 34019587, 34020177, 34019586, 34019585, 34090338, 34090337, 34049130, 34043663, 34086675, 34086676, 34086671, 34086672, 34086674, 34086673, 33832318, 33888386, 33888372, 33888377, 33888376, 33888378, 33888379, 33888383, 33888385, 33888384, 33888373, 34044798, 34064654, 34019555, 34019554, 33802295, 33990042, 33982418, 33982417, 34074060, 34074061, 34095296, 34039463, 33950592, 33983052, 34099021, 34095323, 34082638, 34078647, 34078648, 34078646, 34082590, 34082591, 34048199, 34048200, 34049140, 34049141, 34048196, 34048198, 34045124, 34048194, 34010325, 33899587, 34096386, 34074089, 33974887, 33974888, 34078076, 34100491, 34100492, 34099179, 34099180, 34099178, 34099177, 34099176, 34090571, 34067899, 33995913, 33995914, 33974374, 33974373, 34013030, 34083205, 34019584, 10732394, 10732404, 34079488, 34084291, 34084290, 34004900, 34019558, 34019561, 33913867, 33913879, 33913888, 33995281, 33995280, 34090554, 34090556, 34090555, 34052280, 33991462, 34065273, 34048197, 34048195, 34034132, 34004904, 34082921, 34060126, 34060127, 34060128, 34086704, 34086701, 34086710, 33992506, 34074081, 34074073, 33855207, 33995302, 33995301, 34054626, 34054625, 33904163, 12776983, 33997196, 33987972, 33987973, 33928602, 34010320, 33974366, 34095302, 33938189, 34040207, 34039481, 34060174, 34060158, 34060161, 34060160, 34064667, 34003616, 34039469, 34060746, 34034134, 34106173, 33995035, 34074053, 34010323, 34090366, 34087611, 34090303, 34090302, 34091128, 34091129, 34095293, 33897593, 34078783, 34078784, 34078785, 34074072, 33992007, 33992020, 33993956, 33992022, 33992002, 33992021, 33992024, 33992025, 33992026, 33992004, 33992009, 33992015, 33992001, 33990248, 33992006, 33992005, 33992008, 33992014, 33992018, 33992017, 33990243, 33990246, 33990247, 33992011, 33992012, 33992013, 33992019, 33992010, 34019904, 33899702, 33960797, 33960796, 33960799, 33960798, 33960795, 33995049, 34061036, 34090589, 33931359, 33931358, 33931355, 33931356, 33931357, 34079490, 34079489, 34079833, 34079492, 34079491, 34079487, 34079486, 34079485, 33936114, 34068040, 34068041, 34078675, 33925842, 33925844, 33925846, 33925841, 33925843, 33925845, 33935283, 33935282, 33935281, 33935280, 33928109, 34090546, 34039464, 34082588, 33911101, 33911100, 33965863, 34074054, 34070493, 33997671, 33997672, 34078070, 34064659, 33996557, 33996553, 33996551, 33996550, 33996556, 33996549, 33996849, 33996558, 33996554, 33996552, 33993444, 34095321, 34095327, 34095331, 34095330, 34095326, 34095329, 34095320, 34095324, 34095328, 34095322, 33983339, 33976090, 33983346, 33983345, 33983338, 33983337, 33983336, 33983335, 32525390, 33995263, 34020191, 34019826, 34019557, 34019827, 34019556, 13583628, 34005592, 34005591, 34005587, 34005588, 34005589, 34005590, 33969580, 33969582, 33969581, 33969583, 33969584, 33995222, 34045762, 34099186, 34099185, 33991446, 33903858, 33997197, 34084471, 34075613, 34075614. International Serial Numbers: 34098600, 34098601, 34090290, 34049509, 34049514, 34049515, 34083701, 34083702, 34044802, 33982442, 33982438, 33982439, 33934527, 33903205, 33903206, 33903208, 33903204, 33903198, 33903203, 33903207, 33903201, 33903210, 33903202, 33903199, 33903200, 33934528, 33934531, 33934529, 33934526, 33934530, 33928168, 33903217, 33928174, 33928165, 33982454, 33982437, 33982443, 33982440, 33903216, 33903215, 33982453, 33982452, 33982451, 33982450, 33982441, 33982448, 33982447, 33982446, 33982445, 33982444, 33982449, 33903209, 33928169, 33928160, 33928161, 33928162, 33928164, 33928170, 33928171, 33928172, 33928173, 33928175, 33928176, 33928177, 33928167, 33903211, 33903212, 33903213, 33903214, 33928159, 33950039, 33950038, 33950048, 33952773, 33952774, 33952775, 33952771, 34105152, 33950032, 33876945, 33876930, 33821315, 33876931, 34078216, 33876941, 33876943, 33876942, 33876938, 33876944, 33876952, 33876928, 33876946, 33876951, 33876950, 33876948, 33876940, 33876933, 33876935, 33876936, 33876937, 33950037, 33929937, 33950034, 33950043, 33950042, 33950041, 33950040, 33950053, 33950052, 33950051, 33950049, 33950045, 33950046, 33950047, 33950035, 33950050, 33950054, 33876949, 33876947, 33876939, 33876934, 33876932, 34079098, 33876929, 33950044, 33950033, 33950031, 33950030, 33950036, 34078082, 34078078, 34078081, 34078080, 34079097, 34083393, 34078079, 33985301, 33985300, 33985299, 33985307, 33986568, 33952772, 34079096, 34078215, 34078217, 34078225, 33982471, 33985298, 33985303, 33985304, 33986537, 33985302, 34009595, 34009594, 34009593, 34083729, 34083732, 34077752, 34106204, 34091376, 13669402, 33928605, 33985022, 33983183, 33979582, 33979583, 33974289, 33974293, 33952829, 33952830, 33798619, 33913915, 33798616, 33913909, 33918168, 33918735, 33918736, 33872713, 33918165, 33918166, 33918167, 33918181, 33795390, 33798614, 33896029, 33887441, 33798612, 33795399, 33795397, 33932561, 34094348, 34094349, 34091746, 33986283, 34039425, 33986294, 33983009, 33795389, 33983184, 33983178, 33983010, 33961531, 34074827, 34087714, 33974286, 34049521, 34049526, 34049527, 34049528, 34049529, 34049530, 34053704, 34049523, 33913897, 33913904, 34008743, 34008737, 34003664, 34003634, 34020680, 34020681, 34020164, 34020510, 34034118, 34034725, 33983556, 33974288, 33974287, 33979581, 33974290, 33974291, 34049522, 33983555, 33983554, 33979580, 33974229, 33974292, 33986320, 33913913, 34049520, 33986261, 33974295, 34053733, 34049525, 34053699, 34049524, 33913899, 33913900, 33913901, 33913902, 33985306, 33913905, 34069183, 33985305, 33913898, 33913903, 33913906, 33986313, 33986285, 34003817, 34049519, 34003650, 34003796, 34004911, 34008686, 34008744, 34008742, 34008740, 34008739, 34008738, 34008687, 34008745, 34008741, 34003658, 34020516, 34084458, 34084468, 34074427, 34084092, 34087711, 34087496, 33937399, 33937400, 33937398, 34090742, 33937401, 33937404, 34074835, 33937402, 34083715, 33937403, 33937397, 34090743, 33937405, 33910410, 33910411, 34090741, 34083714, 34044811, 34044812, 34039416, 34074102, 33979590, 34074107, 33983550, 34039419, 34073371, 34039413, 33958404, 33958774, 34074106, 33855054, 33921591, 33921592, 33880013, 33880012, 33880011, 33895654, 33900603, 33899646, 33895655, 33961098, 33961097, 33978886, 34074101, 34106223, 33973952, 33978885, 33979004, 33978888, 33978887, 33983551, 34039415, 34039409, 34039408, 33979564, 33979567, 34079036, 33895653, 33895652, 33895651, 33895650, 33895649, 33893664, 33892562, 33945926, 34074108, 34064246, 34039418, 33895656, 34064247, 34074105, 33968953, 34074099, 34074595, 34074596, 34074104, 34074098, 33949217, 33968972, 33969016, 34074103, 34008392, 34039421, 34012451, 34039422, 34040777, 34040525, 33969015, 34039406, 33968973, 34039405, 33968970, 33968962, 33968966, 33968964, 33969014, 34039407, 33968968, 34034122, 34034124, 34034127, 34034129, 34073370, 33983559, 34073365, 34073366, 34073376, 34065724, 34073364, 33905086, 33905087, 33905088, 33905089, 34003929, 34073369, 34073372, 34073373, 34073374, 34073375, 33968961, 33968963, 33983561, 34079035, 34043644, 34079040, 34079039, 34079037, 34073368, 34073367, 34073363, 34079038, 33928097, 34039410, 34043602, 33983560, 34039411, 34039412, 34043601, 34039414, 33928392, 33968957, 33968956, 33968955, 33968954, 34004876, 34043599, 34044807, 34004928, 34004927, 33968971, 34039417, 34003928, 34004922, 34004924, 34039423, 33968959, 34106425, 34106224, 34106225, 34086724, 34091392, 34086725. |
Recalling Firm/ Manufacturer |
Medtronic Emergency Response Systems, Inc. 11811 Willows Rd Ne Redmond WA 98052-2003
|
For Additional Information Contact |
Todd Bandy 425-867-4514
|
Manufacturer Reason for Recall |
Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
On 3/22/06, the firm began sending out letters dated March 2006, and flagged URGENT MEDICAL DEVICE CORRECTION. The letters advised customers of the issue and that a service representative would visit and inspect the therapy connector sockets, and replace the therapy connector and the mating cable if short drill sockets are found. |
Quantity in Commerce |
942 devices |
Distribution |
Nationwide, and Canada, Netherlands, Germany, Australia, Guam, Hong Kong, New Zealand, New Caledonia, Argentina, Brazil, Chile, Costa Rica, Mexico, Uruguay, Venezuela. Domestic distribution includes hospitals, clinics, |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MKJ and Original Applicant = MEDTRONIC PHYSIO-CONTROL CORP.
|
|
|
|