• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Stratus CS STAT Fluorometric Analyzer MYO DilPak (CMYOD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Stratus CS STAT Fluorometric Analyzer MYO DilPak (CMYOD)see related information
Date Initiated by FirmApril 03, 2006
Date PostedMay 19, 2006
Recall Status1 Terminated 3 on August 07, 2006
Recall NumberZ-0884-06
Recall Event ID 35112
510(K)NumberK984065 
Product Classification Myoglobin, Antigen, Antiserum, Control - Product Code DDR
ProductStratus CS STAT Fluorometric Analyzer MYO DilPak (CMYO-D) in vitro diagnostic.
Code Information Lot # 816023002
FEI Number 2517506
Recalling Firm/
Manufacturer
Dade Behring , Inc.
500 GBC Dr., Mailstop 514
PO BOX 6101
Newark DE 19714-6101
For Additional Information ContactFran Dillon
302-631-6951
Manufacturer Reason
for Recall
Foil sealing of the DilPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.
FDA Determined
Cause 2
Other
ActionCustomers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.
Quantity in Commerce194 cartons of 5 DilPaks
DistributionNationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DDR
-
-