 Class 2 Device Recall Hamiltonsee related information
Date Initiated by FirmMay 24, 2006
Date PostedJuly 22, 2006
Recall Status1 Terminated 3 on December 21, 2006
Recall NumberZ-1249-06
Recall Event ID 35592
510(K)NumberK022679 K052863 
Product Classification Ventilator - Product Code CBK
ProductHamilton brand RAPHAEL Ventilator (Software Versio 2.2x), Product is distributed by Hamilton Medical, Inc., 4990 Energy Way,Reno, NV 89502-Serial Numbers 2975 to 5360,
Code Information Serial Numbers 2975 to 5360
Recalling Firm/
Manufacturer
Hamilton Medical Inc
4990 Energy Way
P O Box 30008
Reno NV 89520-3008
For Additional Information ContactArthur Gutzeit
775-858-3200 Ext. 211
Manufacturer Reason
for Recall
ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
FDA Determined
Cause 2
Other
ActionOn May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.
Quantity in Commerce2,385 for all models
DistributionWorldwide
Total Product Life CycleTPLC Device Report