| | Class 2 Device Recall Terumo APS 1 |  |
| Date Initiated by Firm | October 12, 2004 |
| Date Posted | December 30, 2006 |
| Recall Status1 |
Terminated 3 on February 01, 2007 |
| Recall Number | Z-0331-2007 |
| Recall Event ID |
36396 |
| 510(K)Number | K022947 |
| Product Classification |
Perfusion System - Product Code DTQ
|
| Product | Terumo Advanced Perfusion System 1 Pressure Monitoring; Catalog number 802112. |
| Code Information |
Serial numbers 00017 through 00032, 00036, 00043, 00045 through 00057, 00059 through 00062, 00065 through 00071, 00073 through 00079, 00081 through 00099, 00102 through 00122, 00125 through 000251, 00253 through 00292, 00296 through 00309, 00311 through 00349, 00352 through 00479, 00484 and 00489. |
| FEI Number |
1828100
|
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corp 6200 Jackson Rd Ann Arbor MI 48103-9586
|
| For Additional Information Contact | 800-521-2818 |
Manufacturer Reason for Recall | Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen. |
FDA Determined Cause 2 | Other |
| Action | U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed. |
| Quantity in Commerce | 4,740 for all affected products |
| Distribution | Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DTQ
|
|
|
|