Class 2 Device Recall Allegro 6800 Imaging table. |
|
Date Initiated by Firm |
February 28, 2007 |
Date Posted |
December 13, 2007 |
Recall Status1 |
Terminated 3 on December 18, 2008 |
Recall Number |
Z-0241-2008 |
Recall Event ID |
38307 |
Product Classification |
Imaging Table (Operating), - Product Code LHN
|
Product |
OSI, Allegro 6800 Imaging Table (Operating), single pedestal AC or battery operated table, Orthopedic Systems, Inc., Union City, CA |
Code Information |
Serial Numbers: 12-19, 21-61, 64-67, 70-71, 73-83, 85-93, 95, 97 and 99-107. |
Recalling Firm/ Manufacturer |
Mizuho Orthopedic Systems Inc 30031 Ahern Ave Union City CA 94587-1234
|
For Additional Information Contact |
510-429-1500
|
Manufacturer Reason for Recall |
Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly.
|
FDA Determined Cause 2 |
Software design |
Action |
The firm issued a Product Correction Notification letter by mail on 02/28/2007 and follow up with phone calls. The letter notified users that a required no-cost upgrade of the removal and replacement of a microprocessor that controls the logic for the floor locking mechanism will be conducted by the recalling firm. |
Quantity in Commerce |
117 units |
Distribution |
Worldwide: USA product was distributed to 16 wholesale dealers, 212 medical facilities, 14 distributors throughout the US in the following states: AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, MA, MD, ME, MI, MO, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, WI, WV and WY and to the following countries Taiwan, Australia, German, Japan, Canada, Saudi Arabia, and The UK. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.